Unit 2
Unit 2
Unit 2
Calorimetry is used to measure the heat released into or absorbed from the environment
during a chemical reaction. There are three main classes of calorimeters: isothermal, which
uses a heat sink to keep the reaction environment at a constant temperature as the reaction
occurs; adiabatic, which allows reactions to change the environment and the change in
temperature is measured as time goes on; and temperature scanning, in which the
difference in the amount of heat required to increase the temperature of the sample being
tested and a reference sample is determined.
Applications:
Thermal analysis and calorimetry are used in electronics industries to help determine and
maintain stability, safety, and quality control of their products. Other applications include:
Materials science
Pharmaceutical
Food analysis
Energy
Analytical Chemistry
and more
Differential Scanning Calorimetry (DSC)
DSC is the generic term for the following two measurement methods.
Heat Flux DSCs
A technique in which the temperature of the sample unit, formed by a sample and
reference material, is varied in a specified program, and the temperature difference
between the sample and the reference material is measured as a function of
temperature.
Power Compensation DSC
A technique in which difference of thermal energy that is applied to the sample and
the reference material per unit of time is measured as a function of the temperature
to equalize their temperature, while temperature of the sample unit, formed by the
sample and reference material, is varied in a specified program.
Description of DSC
• The instrument used for TGA analysis is a programmed precision balance for a rise
in temperature (called as Thermobalance, see Figure 23.01). Thermobalance
consists of an electronic microbalance (important component), a furnace, a
temperature programmer and a recorder.
Applications of TGA:
Purpose
The ARC is used to characterize the reactive nature of a chemicals. The substance is place in
the small container (35 ml) shown below and then installed in the apparatus (also shown
below). The ARC has two heating modes: (1) heat and search, and (2) heat.
In the heat and search mode, the substance is heated by fixed temperature increments, say 5
deg. C., and then the ARC waits at the specified temperature until an exotherm is detected. If
an exotherm is detected, the ARC goes into adiabatic mode and follows the exotherm. If an
exotherm is not detected, then the ARC moves to the next higher temperature.
In the heat mode, the temperature is increased continuously until an exotherm is detected, at
which point the ARC goes into adiabatic mode and follows the exotherm.
Data
The data from the ARC test includes the temperature and pressure as a function of time.
These data can be used to calculate the following parameters used to characterize reactive
chemicals:
maximum self-heat rate
maximum pressure rate
onset temperature
reaction order and Arrhenius parameters
others as required
Sensitivity
The ARC is capable of detecting self-heat rates as low as 0.01 deg. C per minute and as high
as 10 deg. C per minute. The major problem with the ARC is the high thermal inertia of the
reactor vessel, which may produce misleading results under some conditions.
Reaction calorimeter:
The first specialized reaction calorimeter was designed around the principle of
heat flow, and it is still considered an industry standard to this day. The
measurement vessel is essentially a jacketed reactor, and signals from temperature
probes are used to calculate heat flux through the reactor wall, using the following
equation (under steady-state conditions):
Where: Q is rate of heat transfer [W]; U is overall heat transfer coefficient [W
(m2 K)-1]; A is heat transfer area [m2]; Tr is process temperature; Tj is reactor jacket
temperature.
This technique has one major disadvantage. Since the U and A can change
during the experiment, it is necessary to perform calibration to determine the UA
before and after the experiment. When UA is known, the data is processed using the
values obtained to correct it. The UA is determined by using a calibration heater of
known power.
This means this method has limited use when UA changes considerably
(change in viscosity, density, reactor fill volume). Those changes are often non-
linear, and they are very challenging to correct (torque measurements etc.). Such
corrections are normally sufficient for safety studies, however, heat flow calorimeters
struggle when big changes of UA are observed. Because of the calibration pulses,
the method can’t provide real-time data (as they are corrected afterward using the
UA obtained from calibration).
The typical workflow of the Isothermal heat flow calorimeter is the following:
Where: Q is the rate of heat transfer [W]; I is the current supplied to heater
[A]; I0 is the current supplied to the heater at steady conditions before experiment
[A], U is the voltage supplied to the heater [V].
The system uses a cooling jacket with constant flow and temperature of the
coolant, which is typically set 5 – 20 °C below desired process temperature. The
process temperature is then maintained by the electrical heater with a large surface
area. As the heat load of the reaction changes, the power of the heater is controlled
to maintain isothermal conditions – so the inverse power supplied to the heater
equals the heat generated / absorbed by the reacting system in the vessel.
1. Heating: Reach the desired temperature with the heater (constant high power)
and circulator on
2. Settling: Once the desired temperature is reached, the heater remains on
constant power and the circulator controls the temperature to reach steady
state conditions
3. Zeroing: Circulator temperature is fixed, and heater power is varied to keep
the reactor at the desired temperature
4. Reaction takes place with desired process conditions (additions etc.), while
the exotherm is controlled by varying the power of the heater. The inverse
value of the heater power shows actual reaction power
5. Data processing
This method isn’t very often used in reaction calorimetry, but is very valuable
in accessories, as this principle is used with reflux condensers used in reaction
calorimetry, to calculate the cooling power of the condenser. However, it usually
requires higher exotherms (to create the temperature gradient between jacket
inlet/outlet) and is very sensitive to heat loss to the environment, which can be
challenging to determine. However, compared to PCC and HFC it is not generally
affected by changes in UA.
The main difference in the reactor is that the temperature control is achieved
by use of Peltier element, which controls the temperature in the reactor via its metal
base. The metal base contains a heat flow transducer, and since it is the only and
well-defined area where heat transfer between reactor contents and the environment
is achieved, it is not very susceptible to changes in UA.
Where Q is the rate of heat transfer [W]; λ is the heat flow transducer heat
conductivity [W (m K)-1]; A is heat flow transducer area [m 2] and dT/dX is the
measured temperature gradient over the transducer disk [K m-1].
From the equation, it is visible that none of the parameters are dependent on
reactor contents, and don’t change during the experiment; this eliminates many
shortcomings of the methods above and introduces advantages over them. For an
additional level of precision, the whole reactor is placed inside in liquid bath with
precise temperature control, which is kept at the same temperature as the reactor.
A major advantage of true heat flow calorimetry is that the reactor doesn’t
require routine calibration; the reactor is calibrated during the manufacturing
process.
A major benefit of true heat flow calorimetry for the user is that that the
baseline doesn’t change over the course of the experiment (as with other methods
due to changes in UA), so it’s not affected by changes in viscosity, fill volume, etc.,
and it delivers a higher level of accuracy and sensitivity (resolution down to 0.0001W
for the Chemisens HighSens Calorimeter).
The RSST (Figures 1 and 2) consists of a spherical glass reaction vessel and
immersion heater (optional), its surrounding jacket heater and insulation,
thermocouples and a pressure transducer, a stainless steel containment vessel that
serves as both a pressure simulator and safety vessel, and, not shown, a magnetic
stirrer base, a control box containing the heater power supply, temperature
amplifiers, and a data acquisition and control panel. The sample cell volume is 10 ml
and the containment volume is 350 ml. A key feature of the apparatus is its low
effective heat capacity relative to that of the sample whose value, expressed as the
capacity ratio, is approximately 1.04 (i.e., quite adiabatic). This feature allows the
measured data to be directly applied to process scale.
The RSST quickly and safely determines the potential for runaway reactions
and measures the rates of temperature and, in case of gassy reactions, pressure
increases to allow reliable determinations of the energy and gas release rates. A
recent additional feature of the RSST allows flow regime characterization, i.e.,
"foamy" versus "non-foamy" behavior, to be determined under actual runaway
conditions (Fauske, 1998). This information can be combined with analytical tools
(summarized later) to assess reactor vent size requirements.
EXPLOSIVE TESTING:
Deflagration Test:
While the action of deflagration is to push the air in front it, objects do
not explode because the rate of combustion is relatively slow. Because the
action of detonation is so rapid, however, detonations result in shattering or
pulverizing objects in their path.
Deflagration:
internal combustion engine (used in any vehicle that uses fossil fuels such
as gasoline, oil, or diesel fuel)
gas stove (fueled with natural gas)
fireworks and other pyrotechnics
gunpowder in a firearm
Deflagration burns outward radially and requires fuel to spread. Thus, for
example, a wildfire starts with a single spark and then expands in a circular
pattern if there is fuel available. If there is no fuel, the fire simply burns out.
The speed at which deflagration moves depends upon the quality of the
available fuel.
Detonation test:
TNT (trinitrotoluene)
nitroglycerine
dynamite
picric acid
C4
The minimum ignition energy (MIE) of a dust cloud is the lowest energy
value of a high-voltage capacitor discharge required to ignite the most readily
ignitable dust/air mixture at atmospheric pressure and room temperature. The
dust concentration and the ignition delay are systematically varied until a
minimum value of the ignition energy is found. The tested energy levels are
1000, 300, 100, 30, 10, 3 en 1 mJ.
Importance
It is well known that explosible dust clouds can be ignited by electric
sparks and arcs that occur in switches and motors, and in short-circuiting
caused by damaged cables. In addition, some categories of electrostatic
discharges may initiate dust explosions in industry.
For an assessment of the hazard situation in dust-processing installations,
knowledge of the minimum ignition energy is indispensable. This value can
possibly determine the extent and hence the cost of protective measures.
Friction Sensitivity
Twenty trials should be conducted using the steel shoe or fiber shoe.
The test material using the steel shoe should not react in any of the twenty
trials. Other friction sensitivity tests that are most common include ABL
Friction test (recommended by UN) and BAM friction test (recommended by
both US DOD and UN).
For some explosives, a wetting agent is added for safe handling and
transportation. This type of sample is tested with the minimum content of
wetting agent. About 10 mg of the test sample is placed on the plate. The
stationary porcelain cylinder is lowered to the top the sample by using the
loading arm. A selected weight [mass] is attached to the arm to apply the
force, which may vary from 4.9 to 353.2 N. The movement of the plate, or the
stroke length, is 10 mm forward and backward from the initial position. The
occurrence of the initiation due to friction is determined from crackling sound,
appearance of smoke, or by the characteristic burning smell. Although both
methods can be used in evaluating safety in transport, in other hazard
assessments only the BAM test is specified (Fig. 1.28).
Gap Tests-Shock Wave:
The gap test is conducted to obtain the minimum shock wave pressure that
can cause complete detonation of the explosive. The design of the experimental
system depends on the size of the booster and the sample and is called either Small
Scale Gap Test or Large Scale Gap Test system. The choice of the small or large
scale gap test facility depends on the pressure and duration of positive phase of
pressure of the generated shock wave. The determine the sensitivity to the shock
wave, the complete detonation of the explosive charge must occur, which is
generally indicated by the formation a clean hole in the steel witness plate that is
9.53 mm thick. A small scale gap test system is shown in Fig. 1.30.