MODEL: SI-100: Manual
MODEL: SI-100: Manual
MODEL: SI-100: Manual
Manual
SI-100 MANUAL
WARNING
Read this manual carefully before using the instrument. The instrument
will perform as designed only if it is used and maintained in accordance
with the manufacturer's instruction. Otherwise, it could fail to perform as
designed and persons who rely on this instrument for their safety could
sustain serious personal injury or death.
Warning
➢ Remove any debris on the surfaces of the sensor part before use.
➢ Test regularly whether the alarms go off properly.
➢ Use the device under the conditions instructed, including the temperature, humidity and
pressure range. The use out of the instructed conditions may cause malfunction or failure.
➢ The sensors inside the device may indicate the gas concentration differently according to the
environment such as temperature, pressure and humidity. Make sure to calibrate the detector
under the same or similar environment to the specification.
➢ Severe shock to the device may cause failures of the device or sensors.
➢ Alarm values are set according to international standards. Therefore, the values must be changed
by an authorized expert.
Caution
Warranty
We, SENKO CO., LTD warrant replacement or repair for the products of SI series for 24 months from the
shipment date of the product(s). However, the parts, whose life can be shortened by use, such as sensors,
batteries and lamps are not under the warranty. Also, Free repair and replacement is not available in case of
purchases of our products through unauthorized channels, mechanical damage and deformation by user’s
misuse, and calibration and replacements of parts without following the instruction in the manual. If any defect
or quality problem occurs to the products during the warranty period, the user should notify it to the
manufacturer. In this case, all the expenses except freight cost are paid by SENKO. Repair, replacement and
freight cost for the products, whose warranty is already over, are paid by the user. SENKO CO., LTD does not
have any responsibility for indirect, or accidental loss which occurs while using our products, and the warranty
is limited to the exchange of parts and products. The warranty is subject to the users who have bought
products from the authorized agency and office appointed by SENKO CO., LTD, and warranty repairs must be
made through the designated A/S center of SENKO CO., LTD with a skilled technician.
2
SI-100 MANUAL
Table of Contents
1. Introduction ................................................................................................................................................................. 4
2. Instrument Overview............................................................................................................................................... 6
5. Certificate ................................................................................................................................................................... 15
3
SI-100 MANUAL
1. Introduction
1.1. Specification
✓ Various sensors can be applied
4
SI-100 MANUAL
5
SI-100 MANUAL
2. Instrument Overview
1
2
6
SI-100 MANUAL
3. How to install
Installing a detector at a site, opening the cover of a detector, or operating it may cause fire or
explosion depending on the environment. Therefore, you should proceed with your work after turning off
the power and examining whether explosive residual gas is around you or not.
You can easily install a detector on a wall or ceiling like the above drawing by using M8 Bolts or other bolts.
7
SI-100 MANUAL
DC 24V
4~20 mA
GND
Normal : Dry
LOW : DRY LOW : 24V
1-2 pin connect : Dry HIGH : DRY HIGH : DRY
4. Measuring Mode
Power supply
8
SI-100 MANUAL
Showing
Version
Counting for
30 seconds
Failing to
Entering
initiate
Measuring
measuring
mode
mode
When this gas detector repeatedly fails to initiate measuring mode, first you should check the conditions
of the connection between gas sensors and the detector. When it is impossible for the detector to initiate
measuring mode even when the connection to the sensors is proper, the gas detector should require checks
and repairs. Therefore, please contact SENKO CO., LTD or your supplier.
9
SI-100 MANUAL
Model Concentration
Gas Type
Unit
The displays are sequentially shown whenever you press each mode key ( ) by using a magnetic bar.
10
SI-100 MANUAL
➢ Press MODE key ( ) for 2 seconds by using the magnetic bar in Set Display to enter Advanced
setting.
➢ Press MODE key ( ) by using the magnetic bar in Advanced setting screen, in order to go to
another menu.
➢ Press DOWN key ( ) by using the magnetic bar to go back to the main menu (SET 1, 2 or 3).
➢ Without magnetic operation for 60 seconds, the display returns to the main menu (SET 1, 2 or 3).
11
SI-100 MANUAL
“2sec”
Blinking Blinking
“2sec”
“1sec”
Blinking Blinking
“2sec”
“1sec”
➢ Press MODE key ( ) for 2 seconds by using the magnetic bar in Set Display, in order to enter
Advanced setting.
➢ When you press UP key ( ) by using the magnetic bar in Advanced setting display, setting values
flicker, which means they can be changed.
➢ Press UP key ( ) when setting values flicker, in order to change the values.
➢ Press MODE key ( ) for 2 seconds to save the changes of setting values. These changes are saved
with the indication, “SAVED”.
✓ When GAS TYPE is changed, related settings (GAS UNIT, RANGE, ALAMR, etc.) are changed as well.
In order to operate properly, sensor replacement for the gas type and calibration is required.
✓ When you press DOWN key ( ) in each mode, the mode goes back to its previous step.
12
SI-100 MANUAL
WARNING
Do not conduct Zero Calibration unless you are sure you are in fresh, uncontaminated air. Otherwise,
inaccurate reading can occur, then show that a hazardous atmosphere is safe. If you have any doubts as to
the quality of the surrounding air, Calibration with N2 gas should be done. Failure to follow this warning can
result in serious personal injury or death.
4.5.1 How to connect Calibration cap
➢ Correctly attach the calibration cap to the sensor part of the gas detector (SI-100) as the picture
above (using flow meter, etc., flow rate: about 300 SCCM) and inject the calibration gas.
Display in
ZERO/SPAN
Calibration
➢ In Calibration mode, the ADC values of ZERO and SPAN are displayed. When you conduct
ZERO/SPAN Calibration, Cal. Count starts counting. In the meantime, ADC Value (Present) is
displayed on the above part and ADC Value (Before Cal.) on the below part.
13
SI-100 MANUAL
➢ Press MODE key ( ) 4 times by using the magnetic bar in Measuring mode, in order to enter
Calibration mode.
➢ Press MODE key ( ) for 2 seconds by using the magnetic bar in Calibration mode, in order to
enter ZERO calibration when “ZERO” flickers. Then, press MODE key ( ) for 2 seconds once again
to initiate ZERO calibration.
➢ When ZERO calibration is completed, the display automatically shifts to SPAN calibration. The
process is same as ZERO calibration.
✓ If ZERO calibration ADC value (EX: 12667) goes beyond the SPAN calibration ADC value (EX: 14928),
you may suspect failure of calibration or malfunction of the gas sensor. In this case, you should
carry out recalibration or receive formal A/S to take appropriate measures for the case.