E70
1 2-Channel Electrocadiog raph

("-L

I inch colorTFT display, high resolution.

Data management of 300 groups paiient data

Support saving data on SD card or USB flash disk.

Alphanumeric keyboard, individual capital and numeral key.


TECH N IGAL SPECI FICATION : E70
Size and Weight: Display of ECG Collection
Size: 400mmx330mmx 146mm (WxHxD) Signal input: 12-Lead, defibrillator-proof, Pacemaker pulse
Weight: S 6Kgs rejection
Degree of protection against electric shock: 4000V, Type CF
Display:
applied part
Type, 8.0" Color TFT LCD
Polarization voltage resistance: >+500mV
Resolution: 800x600 pixels
Response to frequency: 0.05H2 -150H2 (-3dB)
Power Supply: Baseline filter: 0.05H2- 0.15H2. 0.25H2. 0.5H2
Input voltage: AC: 100V-240Y, 50{zl60Hz ECG filter :25H2, 35H2. 45Hz
Input power: 60VA Lowpass filter: 75H2. 100H2. 150H2
Standard: Comply with IEC 60601-1 & IEC 60601-1-2 Gain selection:2.5, 5, 10. 20, 10/5, AGC (mm/mV)
Gain accuracy: t5o/o
Battery
Time base selection: Smm/s, 6.25mm/s, 1Omm/s,
Type: 11.lV 2000mAh rechargeable lithium ion battery
12.5mm/s, 25mm/s, 50mm/s
Operating time: Used continuously for 1.5 hours (print 200
Input impedance: >50MO (10H2)
shares of cardiogram)
Current of input loop: <50 n A
Charge time: 4 hours
Patient leakage current: <10 tr A
Indication of battery capability: With
CMRR: >109 dB (with AC filter)
Input/Output Specification Time for response to wave displaying: <5s
Keyboard: Standard PC104 (position) Time for baseline recovered after switching leads: S1s
Touch screen: Option Defibrillator-proof: 5000V 360J
Network connector: 1 X RI-45 socket Recovering time for defibrillator-proof <5 s
USB connector:2X USB Host (support hot-swap) Recovering time for defibrillator electrode
SD card; 1 X Standard SD card socket polarization 310 s
Energy reduced <10o/o
Data Storage
Voltage transfer <1V
Fixed Memory:100 groups of ECG data
Standard complied with: IEC 60601-2-51
Optional Memory; SD card or U disk
Storage mode: Background storage automatically Measure and Diagnosis of ECG Waveform
Storage format: xml HR range: 30bpm -350bpm
HR precision: +1bpm (10s average)
Recorder
Coefficient error: <5olo, 0.333
Recording way: Thermal dot array
Measure info of ECG waveform : P time limit, PR interphase,
Horizontal resolution: 40 dots/mm
QRS time limit, QT interphase, QTC interphase,
Vertical resolution: 8 dots/mm
Recording speed:5 mm/s.6.25 mm/s, 10 mm/s. 12.5 mm/s.
RR interphase, RV5swing, SVlswing,
RV6swing, SV2swing, RV5+SVlswing, P axis,
25 mm/s. 50 mm/s, error: +3olo.
Recording linearity: <0.5mm/30M
QRSaxis,Taxis.
Diagnosis analyze: 2140 kinds
Paper length: 295mm/280mm
Paper width: 210mm/216mm Environ mental Specifications
Paper type: Folded thermo sensitive printing paper Operating
Reel thermo sensitive printing paper Temperature : (s-40)c
Recording precision: *5olo (X axes) , Relative humidity : 25o/o -93o/o (non condensing)
+50/o (y axes) Atmospheric Pressure (700-1060) hPa
'
Recording type: more than 10 kind of printing methods.
(3x4, 3x4+1R. 3x4+3R.6x2. 6x2+1R. Transportation and Storage
12x1 , Manual 3 path, Manual 6path, Temperature ' (-20-+55) C
manual 12path, single-lead rhythm, Relative humidity : 25o/o-95olo (non condensing)
three-leads rhythm and so on) . Atmospheric Pressure , (500-1060) hPa

BLT reserves the right to change specifications without notice.