FT Desfibrilador-Monitor Comen S8

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Specification: S 8

SHENZHEN COMEN MEDICAL INSTRUMENTS CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming New District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax:+86-755-26431232
Email: info@szcomen.com
Web: www.comen.com

2019 V 1.1
Defibrillator/ Monitor
S8

Standard Configuration: Battery indicator


12-lead ECG, RESP, Thermal Recorder Maintain indicator
Optional: Error indicator
NIBP, TEMP, PR, EtCO2, IBP, SPO2 QRS beep and alarm sound
Safety Standards Operating key sound
ISO 13485:2003 approved, CE marking according to Interfacing
MDD93/42/EEC USB interface
Physical Characteristics RJ45 interface
Size: 323mm×277mm×338mm AC power input
Weight 7.2 kg VGA interface
Screen Size: 8.4” TFT screen Multi-functional connector
Resolution 800 × 600 Date storage
Waveforms: 2 waveforms Alarm Event: 200 groups
Operation Environment Patient profiles: 1000 groups
Temperature: 0~45℃ Wave Review: 48 hours
Humidity: 10%~95%, non-condensation NIBP Review: 2000 groups
Water Resistance: IP44 (without external power) Trend Graph: 160 hours
Solids Resistance: IP4X Trend Table: 160 hours
Power requirement: 100-240V~, 50/60Hz±1Hz ECG report: 500 cases of 12-lead ECG
Battery type: Rechargeable Lithium-ion battery diagnosis report (Up to 5 case
Battery capacity: 7500mAh, d.c.14.8V reports per patient)
Battery number: Max 2 Voice recording: Max 240 min in total;
Battery recharging (Up to 60 min for each patient)
Time: Less than 2 hours to 80% and less Marked events Available
than 3 hours to 100% with Power-off storage:
equipment power off Yes
Battery backup: Monitoring Mode: 12 hours; Alarm: User-adjustable High and Low
(Two new, fully Defib Mode: 420 times (360J 3-level Limits;
charged battery) charge at intervals of 1minute Prioritized audible and visual
without recording); alarm
Pacing Mode: 9 hours (50 Ω load Network: Connected to Central Monitoring
impedance, Pacing rate: 80bpm, System by hardwire/wireless
Pacing output: 60mA, without Recorder
recording) Type: Built-in; Thermal array
Brightness: Manual from 1 to 10 Channel: 4 channel waveforms
Indicator: Real-time recording: 3s, 5s, 8s, 16s, 32s, auto
Two alarm indicators Speed: 25mm/s, 50mm/s
Power indicator Record width: 80mm
Resolution: 8dot/mm (Horizontal and Energy transfer begins within
vertical) 60ms of the QRS peak;
Background grid: Configurable Energy transfer begins within
External printer: Yes 25ms of the External Sync signal
AED
Defibrillator Output Energy: User configurable
Operating mode: Manual Mode, AED Mode, AED Shock Series: Configurable
Synchronous defibrillation Noninvasive Pacing
Waveform: Biphasic truncated exponential Waveform: Monophasic square wave pulse
waveform, with impedance Pulse Width: 20 ms
compensation Accuracy: ±5%
Defibrillation pathway: External defibrillation & Internal Pacing Mode: Demand or fixed
defibrillation Pacing rate: 40 ppm to 170 ppm
Electrode type: External defibrillation electrode Accuracy: ±1ppm or ±1.5% (whichever is
plate, multifunctional electrode greater)
pads and internal defibrillation Pacing output: 0 mA to 200 mA
electrode plate Accuracy: ±5% or ±5mA, whichever is
External defibrillation greater
electrode plate: Supports charging, discharging, 4:1 pacing: Pacing pulse frequency reduced
energy selection and other by factor of 4 when activated
operational functions; Charging Monitoring ECG
completion indicator Lead Type: 3 leads ECG, 5 leads ECG, 12 leads
Charge Time: Less than 5 seconds to 200 Joules ECG, AUTO
(Battery power) with a new, fully charged battery Lead selection: 12-Lead I; II; III; aVR; aVL;aVF;
Less than 8 seconds to 360 Joules V1~V6
with a new, fully charged battery 5-lead: I; II; III; aVR; aVL; aVF; V
Charge Time: Less than 7 seconds to 200 3-lead: I; II; III
(AC power) Joules; Multi-lead
Less than 11 seconds to 360 synchronization
Joules analysis: Available
Energy accuracy: ±1.5J or ±10% of setting, ECG size: Auto, 1.25 mm/mV (×0.125),
whichever is greater, into 50Ω 2.5 mm/mV (×0.25), 5 mm/mV
±2J or 15% of setting, whichever (×0.5), 10 mm/mV (×1),
is greater, into 25Ω, 75Ω, 100 20 mm/mV (×2), 40 mm/mV (×4),
Ω, 125Ω, 150Ω, 175Ω, Accuracy: Less than ±5%
Patient Impedance Sweep speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s,
Range: 20~250Ω(External defibrillation); 50 mm/s
15~250Ω(Internal defibrillation) Accuracy: Less than ±10%
Defibrillation proof: Type CF: ECG, RESP, SpO2, NIBP, Heart Rate
IBP, TEMP, PR; Measurement & alarm
Type BF: CO2 range: Adult: 15~300bpm
Manual Mode Pediatric/Neonate:15~350bpm
External defibrillators: 1J~360J Resolution: 1 bpm
Internal defibrillation: 1J~50J Accuracy: ±1% or ±1bpm (whichever is
greater)
Synchronous Bandwidth: MON: 0.5~40Hz
Cardioversion: DIA: 0.05~150Hz
Treat: 1~20Hz Accuracy: 7~150rpm: ±2rpm or ±2%
ST: 0.05~40Hz (whichever is greater)
CMRR: MON: >105dB 0~6rpm: unspecified
DIA: >90dB Apnea Alarm: Adult: 10s~60s Ped/Neo: 10s~40s
Treat: >105dB Accuracy: ±5s
ST: >105dB Alarm: Audible and visual alarm; alarm
Input Impedance: ≥5MΩ events reviewable
Input signal range: ±8mV NIBP
HR trigger value 200μV Method Automatic oscillometric
Lead off detection Work mode: Manual / Automatic/Continuous
current: Measuring electrode: <0.1μV Measurement Time: Adjustable (1~720min)
Driving electrode: <1μV Maximum
Pacemaker pulse measurement time Adu/Ped: 120s; Neo: 85s
suppression switch: Manual selection when the Measurement Unit: mmHg / kPa selectable
pacemaker is turned on Measurement types: Systolic, Diastolic, Mean
Analog output: Magnification: 1:1000; Range of systolic
Accuracy: ±5% pressure: Adult Mode:40~270mmHg
Bandwidth: 0.5Hz~40Hz Pediatric Mode:40~200mmHg
Delay: ≤35ms Neonate Mode 40~135mmHg
ST Detection: -2.0mV~+2.0mV Range of diastolic
Resolution: 0.01mV pressure: Adult Mode:10~215mmHg
Accuracy: -0.8mV ~ +0.8mV: ±0.02mV or Pediatric Mode:10~150mmHg
±10%; Others: Unspecified Neonate Mode 10~100mmHg
System noise: Less than 25μV Range of mean
Calibration voltage 1 mV; Accuracy: ±5% pressure: Adult Mode:20~235mmHg
Arrhythmia Analysis: 26 Types Pediatric Mode:20~165mmHg
Pacemaker detection: detectable Neonate Mode 20~110mmHg
Defibrillation ECG Over pressure
Lead Type: Single lead ECG protection: Both Hardware and software
Heart Rate over pressure protection
measurement & alarm Accuracy: ±3mmHg
range: Adult: 15~300bpm Resolution: 1 bpm
Pediatric/Neonate:15~350bpm Alarm: Systolic, Diastolic, Mean
Resolution: 1 bpm PR from NIBP: 40bpm~240bpm
Accuracy: ±1% or ±1bpm (whichever is Resolution: 1bpm
greater) Accuracy: ±3% or ±3bpm, whichever is
Bandwidth: Defib: 1~20Hz greater
CMRR: Defib: >105dB Nellcor SpO2
Input Impedance: ≥5MΩ Measurement range: 0~100%
Input signal range: ±8mV Resolution: 1%
HR trigger value 200μV Accuracy: ±2% (70~100%, Adu/Ped, non-
Arrhythmia Analysis: 5 Types motion)
Respiration ±3% (70~100%, Neo, non-motion)
Method: RA-LL Impedance Method 1~69% unspecified
RR measurement Alarm range: 20~100%
range: Adult: 0~120rpm
Pediatric/Neonate: 0 ~150bpm PR Measurement
Range: 20~300bpm Temperature (Dual Channel)
Measurement & alarm
Resolution: 1bpm
range: 0~50°C
Accuracy: ±3bpm (20~250bpm)
TEMP sensor: Standard configuration-skin
Unspecified (251~300bpm)
TEMP sensor
Alarm range: 20~350bpm
Resolution: 0.1°C
Masimo SpO2
Accuracy: ±0.1°C (exclusive of error of
Measurement & alarm
sensor)
range 1~100%
Channel type: T1, T2, TD (Temperature
Resolution: 1%
Difference)
Accuracy: ±2% (70~100%, Ped/Adu,
MASIMO EtCO2 (Sidestream)
non-motion
Measurement range: 0~190mmHg, 0~25% (at
±3% (70~100%, Neo,
760mmHg)
non-motion);
Accuracy: ± (2.25mmHg +4% of reading)
1~69% unspecified
Resolution: 1mmHg
Alarm range 1~100%
awRR range: 0~150rpm
PR Measurement
awRR accuracy: ±1rpm
Range 25~240bpm
Response time: <240msec (10% to 90%)
Resolution: 1bpm
Delay time: <2s
Accuracy: ±3%(non-motion)
Respironics EtCO2 (Sidestream)
±5% (motion);
Measurement range: 0~150mmHg, 0to 25% (at
Alarm range: 20~350bpm
760mmHg)
PI value: Resolution: 0.02~20% Accuracy: ± 2 mmHg (0 – 40 mmHg)
0.01% (0.02%~9.99%) ± 5% of reading (41 – 70 mmHg)
0.1% (10.0%~20.0%) ± 8% of reading (71 –100 mmHg)
Accuracy: unspecified ±10% of reading (101~150
SIQ: Available mmHg)
Comen SpO2 Resolution: 1mmHg
Measurement & alarm awRR range 0~150rpm
range: 0~100% awRR accuracy: ±1rpm
Resolution: 1% Response time: <240msec (10% to 90%)
Accuracy: ±2% (70~100%, Ped/Adu, Delay time: <2s
non-motion)
IBP
±3% (70~100%, Neo,
Channel: 2 Channels
non-motion);
Measured Pressure: ART, PA, CVP, RAP, LAP, ICP, LV,
0~69% unspecified
AO, UAP, BAP, FAP, UVP, IAP, P1,
PR Measurement
P2, P3, P4
Range: 20~254bpm
Measurement Unit: mmHg/ kPa/ cmH2O selectable
Resolution: 1bpm
Measurement range: ART: 0~300mmHg
Accuracy: ±2bpm
PA: -6~120 mmHg
Alarm range: 20~350bpm
CVP: -10~40mmHg
PI value: 0.05~20%
RAP: -10~40mmHg
Resolution: 0.01% (0.05%~9.99%)
LAP: -10~40mmHg
0.1% (10.0%~20.0%)
ICP: -10~40mmHg
Accuracy: unspecified
LV: 0~300mmHg
SIQ: Available
AO: 0~300mmHg
UAP: 0~300mmHg
BAP: 0~300mmHg Resolution: 1bpm
FAP: 0~300mmHg Accuracy: ±1% or ±1bpm, whichever is
UVP: -10~ 40mmHg greater
IAP: -10~40mmHg PPV/SPV
P1, P2: -50~300mmHg measurement: Available
Accuracy: ±2% or ±1mmHg (whichever is PAWP measurement: Available
greater)
Resolution: 0.1kPa or 1mmHg
(-50mmHg~+300mmHg)
Alarm Range: -50mmHg~+300mmHg
PR from IBP: 20bpm~350bpm
*Notice: Specifications subject to changes without prior
notice. All rights reserved by Comen