DW077
DW077
DW077
LASER KEYPAD
B. SOFTWARE CHECK:
Note: The laser beam must be rotating for the On-Trak sensors to work.
1. On the computer screen, click on the green “arrow” button to start measuring the beam location.
2. Close the “Station Calibration Screen” window. This procedure only uses the “Cone Error
Adjustment Screen” window.
1
Step 1: Click on the
Green Arrow to start Step 2: Click here to
measuring close the “Station
Calibration Screen”
Note: To fully calibrate the DW077 laser, the X, Y, and Z axis must be calibrated separately, one at a
time.
2. Turn on the laser unit, allow it to self-level, and put it in rotation mode.
3. On the computer screen, look at the value in the “Delta Position” box. As soon as it stabilizes,
write down the value, including the +/- sign and label it “+ X”.
2
Step 3: Write down value in “Delta
Position Box for +X direction
4. Rotate the laser unit on the tripod head 180 degrees, so that the “X” marking is lined up
STRAIGHT AWAY from the On-Trak targets.
5. Once the reading stabilizes, write down the new “Delta Position” value, including the +/- sign, and
label it “-X”.
6. Subtract the “-X” value from the “+X” value, and write this down as “X Error”.
7. If the “X Error” is within the limits shown in the table below, then the unit is already calibrated.
If the “X Error” is outside of the limits, then the unit must be recalibrated.
3
D. RECALIBRATING THE X-AXIS:
1. To calculate the desired target value for “Delta Position”, take the average of the two values
previously written down for +X and –X. Write down this target value.
2. The laser unit should still be aligned with the “X” marking facing away from the On-Trak.
3. Turn the laser unit off.
4. Put the laser unit into “X” calibration mode by pushing and holding the “UP” arrow button on the
laser keypad. While still pushing the “UP” arrow, push and release the “ON” button on the laser
keypad. Release the “UP” arrow button.
5. The laser unit will automatically move through its range of X movement, then come to rest.
6. Push the rotation button (on the keypad or on the remote control) to spin the laser head.
7. Use the REMOTE CONTROL “UP” and “DOWN” arrow buttons to move the laser beam until the
“Delta Position” reading matches the target value calculated in step 1 above.
IMPORTANT NOTE: IF THE BEAM GETS TO THE END OF ITS ADJUSTMENT RANGE
BEFORE CALIBRATION REQUIREMENTS ARE MET, THEN THE UNIT CAN NOT BE
RECALIBRATED. THE COMPENSATOR WILL NEED TO BE REPLACED.
8. IMPORTANT NOTE: IF THE BEAM GETS TO THE END OF ITS ADJUSTMENT RANGE
BEFORE CALIBRATION REQUIREMENTS ARE MET, THEN THE UNIT CAN NOT BE
4
RECALIBRATED. THE COMPENSATOR WILL NEED TO BE REPLACED.
2. Position a tripod at least 40 feet (12 m) from a wall where a stationary laser dot can be projected.
3. Tape a piece of graph paper on the wall where the laser dot will be. The graph paper should have
1mm X 1mm grid lines.
4. Position the Optical Auto Level halfway between the tripod and the wall, almost in line with the
laser beam, but not where it will block the beam.
Laser beam
5
A piece of paper on
the laser head allows
the beam to be sighted
with the optical level
5. Adjust the elevation of the Optical Auto Level so the horizontal cross-hair is lined up exactly on
the center of the red dot at the laser unit.
DANGER: NEVER LOOK THROUGH AN OPTICAL LEVEL DIRECTLY AT A LASER
SOURCE. SERIOUS EYE INJURY COULD RESULT.
6. Remove the paper from the laser unit’s rotary head so the beam can project a red dot on the far
wall.
7. Rotate the Optical Level around to view the dot on the graph paper mounted on the far wall.
8. Measure how far the center of the red dot is above or below the cross-hair. This is the amount of
calibration error of the laser unit’s Z axis.
9. Table of acceptable calibration error
Distance to Target 40 ft. (12m) 50 ft. (15m) 70 ft. (21m) 100 ft. (30m)
Allowable Error +/- 1mm +/- 1.5mm +/- 2mm +/- 3mm
If the error is greater than these acceptance limits, then the laser unit must be recalibrated
IMPORTANT NOTE: IF THE BEAM GETS TO THE END OF ITS ADJUSTMENT RANGE
BEFORE CALIBRATION REQUIREMENTS ARE MET, THEN THE UNIT CAN NOT BE
RECALIBRATED. THE COMPENSATOR WILL NEED TO BE REPLACED.