Gemini VII 2390 Operator Manual V3.03
Gemini VII 2390 Operator Manual V3.03
Gemini VII 2390 Operator Manual V3.03
Gemini VII
Operator’s Manual
V3.03
239-42828-01
Feb 2014
Adobe Acrobat Reader is a registered trademark of Adobe Systems, Inc.
Alconox is a registered trademark of the Alconox Company.
Micromeritics is a registered trademark of Micromeritics Instrument Corporation.
Microsoft, Windows, and Windows 7, and Windows Vista are registered trademarks of Microsoft Corporation.
Gemini is a registered trademark of Micromeritics Instrument Corporation.
© Micromeritics Instrument Corporation 2009-2014. All rights reserved. Printed in the U.S.A.
The software described in this manual is furnished under a license agreement and may be used or copied only in
accordance with the terms of the agreement.
WARRANTY
MICROMERITICS INSTRUMENT CORPORATION warrants for one year from the date of shipment each
instrument it manufactures to be free from defects in material and workmanship impairing its usefulness under
normal use and service conditions except as noted herein.
Our liability under this warranty is limited to repair, servicing and adjustment, free of charge at our plant, of any
instrument or defective parts when returned prepaid to us and which our examination discloses to have been
defective. The purchaser is responsible for all transportation charges involving the shipment of materials for
warranty repairs. Failure of any instrument or product due to operator error, improper installation, unauthorized
repair or alteration, failure of utilities, or environmental contamination will not constitute a warranty claim. The
materials of construction used in MICROMERITICS instruments and other products were chosen after extensive
testing and experience for their reliability and durability. However, these materials cannot be totally guaranteed
against wear and/or decomposition by chemical action (corrosion) as a result of normal use.
Repair parts are warranted to be free from defects in material and workmanship for 90 days from the date of
shipment.
No instrument or product shall be returned to MICROMERITICS prior to notification of alleged defect and
authorization to return the instrument or product. All repairs or replacements are made subject to factory
inspection of returned parts.
MICROMERITICS shall be released from all obligations under its warranty in the event repairs or modifications
are made by persons other than its own authorized service personnel unless such work is authorized in writing by
MICROMERITICS.
The obligations of this warranty will be limited under the following conditions:
1. Certain products sold by MICROMERITICS are the products of reputable manufacturers, sold
under their respective brand names or trade names. We, therefore, make no express or implied
warranty as to such products. We shall use our best efforts to obtain from the manufacturer, in
accordance with his customary practice, the repair or replacement of such of his products that may
prove defective in workmanship or materials. Service charges made by such manufacturer are the
responsibility of the ultimate purchaser. This states our entire liability in respect to such products,
except as an authorized person of MICROMERITICS may otherwise agree to in writing.
2. If an instrument or product is found defective during the warranty period, replacement parts may,
at the discretion of MICROMERITICS, be sent to be installed by the purchaser, e.g., printed
circuit boards, check valves, seals, etc.
3. Expendable items, e.g., sample tubes, detector source lamps, indicator lamps, fuses, valve plugs
(rotor) and stems, seals and O-rings, ferrules, etc., are excluded from this warranty except for
manufacturing defects. Such items which perform satisfactorily during the first 45 days after the
date of shipment are assumed to be free of manufacturing defects.
Purchaser agrees to hold MICROMERITICS harmless from any patent infringement action brought against
MICROMERITICS if, at the request of the purchaser, MICROMERITICS modifies a standard product or
manufactures a special product to the purchaser’s specifications.
MICROMERITICS shall not be liable for consequential or other type damages resulting from the use of any of
its products other than the liability stated above. This warranty is in lieu of all other warranties, express or
implied, including, but not limited to, the implied warranties of merchantability or fitness for use.
TABLE OF CONTENTS
1. GENERAL INFORMATION
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Organization of the Operator’s Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Equipment Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Cryogen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Degasser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Vacuum Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
User-defined Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Report Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
2. USER INTERFACE
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Instrument Components and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Sample Compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Turning the Analyzer On and Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Turning the Analyzer On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Turning the Analyzer Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Using the Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Shortcut Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Shortcut Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
File Status and Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
File Name Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Menu Structure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Main Menu Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Window Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Help Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
3. OPERATING PROCEDURES
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Specifying Sample Defaults. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Advanced Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Basic Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Restricted Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
5. FILE MENU
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Common Fields and Buttons - File Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Open . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Save As. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Save As (t-Curve or Alpha-S Curve) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Save All . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Close. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Close All. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Export. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Convert. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Sample Information Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Degas Conditions Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Analysis Conditions Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Adsorptive Properties Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Report Options Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
Summary Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
Isotherm Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
BET/Langmuir Surface Area Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
Freundlich Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27
Temkin Isotherm Report Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28
t-Plot Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-30
Alpha-S Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-33
f-Ratio Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34
BJH Adsorption/Desorption Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-36
Tabular Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-38
Plot Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-40
Dollimore-Heal Adsorption/Desorption Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-41
Tabular Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-45
Plot Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-45
DFT Pore Size Report Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-46
DFT Surface Energy Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-49
Dubinin Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-49
Tabular Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-50
Transformed Isotherm Plot Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-51
Pore Volume Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-52
MP-Method Report Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-53
Tabular Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-54
Plot Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-54
Options Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-55
Sample Log Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-55
Validation Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-56
Entered or Collected Data File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-57
6. UNIT MENU
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Common Fields and Buttons - Unit Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Start Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
QuickStart Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Start Po Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Blank Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Enable Manual Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Instrument Schematic Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
System Valves. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Instrument Schematic Shortcut Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Show Instrument Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Show Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Show Instrument Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
Unit Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
Zero Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
Match Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
Reset Pressure Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
Sample Servo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
Balance Servo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
Save to File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
Load from File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
Service Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
7. REPORTS MENU
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Common Fields and Buttons - Reports Menu Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Start Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Close Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Open Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
SPC Report Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Regression Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Control Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Heat of Adsorption Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Report Features and Shortcuts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Report Header . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
Report Tool Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
Report Shortcut Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17
Report Header Shortcuts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17
Tabular Reports Shortcuts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
Graph Shortcuts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
Other On-Screen Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
Zoom Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
Axis Cross-Hair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
Report Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
Isotherm Linear Plot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-25
8. OPTIONS MENU
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Option Presentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Sample Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Restricted Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Graph Grid Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Service Test Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
A. FORMS
Sample Data Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
B. ERROR MESSAGES
2400 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
2500 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7
4000 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-10
6000 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-22
C. CALCULATIONS
Saturation Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
Relative Pressure Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
Quantity Adsorbed Calculations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Free Space. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Measured. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Calculated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Equilibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
Thermal Transpiration Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
BET Surface Area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-6
Langmuir Surface Area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-7
t-Plot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-8
BJH Pore Volume and Area Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-10
Explanation of Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-10
Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-11
Compendium of Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-18
Horvath-Kawazoe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-19
Slit Pore Geometry (original HK). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-19
Cylinder Pore Geometry (Saito/Foley) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-20
Sphere Pore Geometry (Cheng/Yang) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-21
Cheng/Yang Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-22
Interaction Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-23
Additional Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-24
Interaction Parameter Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-25
Dubinin-Radushkevich . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-27
Dubinin-Astakhov. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-29
MP-Method. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-33
Freundlich Isotherm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-35
Temkin Isotherm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-35
DFT (Density Functional Theory) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-36
The Integral Equation of Adsorption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-36
Application to Surface Energy Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-37
Application to Pore Size Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-37
Performing the Deconvolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-38
Regularization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-38
SPC Report Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-39
Regression Chart Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-39
Control Chart Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-40
Summary Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-40
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-42
D. ETHERNET CONFIGURATION
Configuring the Ethernet Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Windows 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Windows Vista . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5
F. DFT MODELS
Models Based on Statistical Thermodynamics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
Theoretical Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
Molecular Simulation Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-2
Molecular Dynamics Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-2
Monte Carlo Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-2
Density Functional Formulation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-3
Models Included. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-7
Non-Local Density Functional Theory with Density-Independent Weights . . . . . . . . F-7
Non-Local Density Functional Theory with Density-Dependent Weights . . . . . . . . . F-7
Models Based on Classical Theories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-13
Surface Energy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-13
Pore Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-13
Models Included. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-14
Kelvin Equation with Halsey Thickness Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-14
Kelvin Equation with Harkins and Jura Thickness Curve . . . . . . . . . . . . . . . . . . . . . . F-15
Kelvin Equation with Broekhoff-de Boer Thickness Curve . . . . . . . . . . . . . . . . . . . . F-16
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-17
1. INDEX
1. GENERAL INFORMATION
Introduction
This manual provides details on software installation, program menu options and operating instruc-
tions. This chapter contains information on the Gemini VII hardware and software components.
Chapter Description
Appendix A FORMS
Provides a sample information worksheet used to assist in obtaining a sample
mass.
Appendix C CALCULATIONS
Provides calculations used in reports.
Index INDEX
Provides quick access to a subject matter.
Conventions
This manual uses the icons shown below to identify notes of importance, warnings and cautions.
Warnings contain information to help prevent actions that may cause personal injury.
Cautions contain information to help prevent actions that may damage the analyzer or
components.
Equipment Description
The Gemini VII Series analyzers are easy to use, fully automatic, and provide single and multipoint
surface area and pore size measurements. Three models are available: the Gemini 2390a, 2390p, and
2390t.
The Gemini 2390a and 2390p are enclosed in the same size cabinet. The only physical difference is the
Po (saturation pressure) tube which is installed on the Gemini 2390p, allowing continuous measure-
ment of the saturation pressure.
The Gemini 2390t is in a slightly larger cabinet allowing the use of a larger Dewar and longer sample
tubes for extended analyses. This model also is equipped with a Po tube.
Most features are available on all three models, with the exception of a continuous Po measurement
and some reports. Refer to the following table for a brief summary of some of the model reports and
features.
The basic function of the Gemini analyzers is to determine the surface area of solid materials. Analyses
are performed using a static volumetric technique in which the adsorptive flows into a tube containing
the sample and into a balance tube simultaneously. The internal volume and the temperature surround-
ing both tubes are maintained at identical conditions. The only difference is the presence of the sample
in the sample tube.
Analysis log
Dubinin reports
MP-Method reports
DFT reports
Operation
The sample and balance tubes are immersed in a single liquid nitrogen bath which maintains isother-
mal conditions for both tubes. The analysis gas is then delivered to the sample tube by a servo valve
mechanism. The delivery rate of analysis gas flow into the balance tube is controlled by another servo
valve connected to a differential pressure transducer. This differential pressure transducer measures the
pressure imbalance between the sample and balance tubes, which is caused by the adsorption of the
analysis gas onto the sample. As the sample adsorbs analysis gas, the pressure drops in the sample
tube. The servo valve continuously restores the pressure balance between the two tubes by admitting
more gas into the sample tube. The end result is that the Gemini maintains a constant pressure of anal-
ysis gas over the sample while varying the rate of analysis gas delivery to match exactly the rate at
which the sample can adsorb the gas.
The Gemini provides rapid and accurate sample analysis. As many as 50 data points can be collected
and reported in the BET, Langmuir, or t-Method range. The 2390p model of the Gemini VII Series can
report discrete point adsorption isotherms consisting of up to 1,000 adsorption points and, with the
Gemini 2390t, up to 1,000 adsorption and desorption isotherm points.
The Gemini requires only pure nitrogen as the analysis gas, even for most low surface area samples.
The analyzer requires no flow regulator, gas mixer, or mixed gases to operate. Surface areas as low as
0.01 m2/g are easily determined with excellent precision using nitrogen gas as the adsorbate. Option-
ally, helium may be used to determine any slight differences in volume between the balance tube and
the sample tube (differential free space). The Gemini uses this information to automatically compen-
sate even for these small differences in calculating analysis results.
The design of the Gemini avoids many errors which may occur with other surface area analysis
equipment:
• Any free space associated with the sample tube is offset by an identical amount in the balance tube.
• Any thermal gradient in the coolant bath is the same for both the sample and balance tubes and
produces a net effect of zero.
• Non-ideal gas behavior occurs in both the sample and balance tubes, cancelling negative effects on
sample analysis.
• Gas-mixture separation, which may occur in flowing-gas systems due to thermal diffusion, cannot
occur because mixed gases are not used.
Cryogen
Liquid nitrogen is used as the cryogen to cool the sample during analysis. An easy-to-use liquid nitro-
gen transfer system, the Model 021, eliminates the need to pressurize storage Dewars.
Degasser
Micromeritics has the following degassing units available for preparing samples:
• SmartPrep 065
The SmartPrep uses flowing-gas to remove adsorbed contaminants from the sample surface and
pores. It contains six sample ports, each one independently controlled for greater flexibility. It
contains two serial ports, one for connecting to the computer and a second for connecting an
additional SmartPrep. Up to four SmartPreps can be connected in sequence allowing the capability
of up to 24 preparation ports at one time.
The SmartPrep also can be controlled through the analyzer application and is the recommended
degassing unit for the analyzer.
• VacPrep 061
The VacPrep degasses up to six samples at up to 400 ºC with either flowing gas or evacuation.
• FlowPrep 060
Vacuum Pump
An external vacuum pump equipped with an anti-suckback valve is required for sample analysis with
the Gemini analyzer. Any vacuum source achieving vacuums better than 20 x 10-3 mmHg at the instru-
ment inlet may be used.
A device to reduce oil vapor backstreaming is recommended.
The vacuum pump used with the Gemini must have an anti-suckback valve to prevent
oil from being admitted to the instrument should the power fail while the system is
under vacuum. Pumps available from Micromeritics are equipped with an anti-
suckback valve.
An oil-based or oil-free vacuum pump can be used with the Gemini analyzer. Appropriate vacuum
pumps are available from Micromeritics. Refer to ORDERING INFORMATION, page 10-1.
Software
The Gemini VII program allows other computer applications to run while an automatic operation is in
progress. The Help menu provides access to the online manual and instructional videos on using the
software.
The Gemini VII program offers a Windows interface with an easy way to collect, organize, archive and
reduce raw data and store sample information files for later use. The reports can be generated to
screen, paper, or exported for use in other applications. Cut and paste graphics, scalable and editable
graphs, and customized reports are easily generated.
User-defined Reports
Report options can be specified when creating the sample information file. When an analysis is run,
data gathered during the analysis process are compiled into the predefined reports. Reports can also be
defined and generated after an analysis has been run. Each selected report is displayed on its own tab
and reflects data collected during the analysis.
Report Features
• After analysis, reports can be viewed, printed, and/or copied and pasted into other documents.
• The report zoom feature provides the viewing of fine graph details and the ability to shift the axes.
• Reports can be customized with a choice of fonts and a logo added to the report header.
Gases
Compressed gases are required for sample analysis with the Gemini analyzer. Gas bottles and a rack
for storage of these bottles should be located near the analyzer. As an alternative, outlets from a central
source may be provided nearby. Gas sources must be able to supply flow rates of 4 liters per minute at
15 psig (103.4 kPa).
Appropriate two-stage regulators which have been leak-checked and specially cleaned are required.
Pressure relief valves should be set to no more than 30 psig. Gas regulators are available from
Micromeritics. Refer to ORDERING INFORMATION, page 10-1.
When helium is used for differential free-space measurement and nitrogen is used as the adsorbate gas,
they should be of the following purity or better:
• Helium - purity of 99.9% (For analysis of materials with very low surface areas, Micromeritics
recommends use of helium with purity of 99.995%.)
• Nitrogen - purity of 99.9% (For analysis of materials with very low surface areas, Micromeritics
recommends use of nitrogen with purity of 99.995%.)
Specifications
The analyzer has been designed and tested to meet the following specifications:
Characteristic Specification
Applicability
Pressure Measurement
Vacuum System
External: Vacuum source achieving 20 × 10-3 mmHg (or better) at the instrument
inlet for oil-based or oil-free pumps.
For oil-based pumps, an anti-suckback valve is required to prevent oil
from being admitted into the Gemini should there be a power failure. A
device to reduce oil vapor backstreaming is also recommended.
Environment
Electrical
Frequency: 50 to 60 Hz
Gases
Adsorbate: Optimized for nitrogen in a liquid nitrogen sample bath. Gemini may
be used with non-corrosive adsorbate gases having vapor pressures at
both room and bath temperatures that are acceptably high relative to
the resolution of the 1,000 mmHg pressure transducer. Typically,
oxygen, argon, carbon dioxide, butane, methane, and other light
hydrocarbons will produce useful data above absolute pressures of a
few mmHg.
Physical
Computer
2. USER INTERFACE
Introduction
This chapter contains information on the analyzer hardware and software components of the
Gemini VII.
• Shortcuts
• File name conventions
• Menu Structure
Component Description
Power indicator Blinks when power is applied to the analyzer; illuminates when analysis
program is initiated and ready for operation.
Front Panel
Power
Indicator
Sample compartment
door
Component Description
Power indicator Blinks when power is applied to the analyzer; illuminates when analysis
program is initiated and ready for operation.
Sample Compartment
Po port and tube (2390p and 2390t models)
Elevator
Component Description
Po port and tube The Po (saturation pressure) tube comes installed on the Gemini II 2390p
and 2390t.
Diagram Diagram of port numbers and locations as well as the positions of the
balance and sample tubes. The empty tube represents the balance tube
installed on the left port. The filled tube represents the sample tube
installed on the right port.
Elevator Allows placement of the Dewar around the sample and Po tubes. The
elevator raises automatically when the analysis is started and lowers
automatically upon completion. During analysis, the elevator lowers after
the free-space measurement to allow evacuation, then raises and continues
the analysis.
Rear Panel
Keypad connector
Ethernet port
USB ports
RS232 connector
On/Off Switch
Power connector
Component Description
Adsorptive inlet port For attaching the adsorptive (analysis) gas supply.
Keypad connector Not used with the for Gemini VII version.
Ethernet port For connecting an Ethernet cable for communication between the analyzer
and the computer.
2. Place the analyzer ON/OFF switch in the ON position. The pump will turn on automatically if it is
plugged into the electrical outlet.
Do not turn off the analyzer while initialization is in progress. Doing so may damage the
instrument.
3. Turn on the pump. The pump must warm for approximately two hours before performing
analyses. For analyses that require very precise results, allow the analyzer to warm up while the
sample is being prepared.
It is recommended that the analyzer be powered on at all times. If it does become necessary to turn it
off, perform the following steps:
1. Select File > Exit from the analyzer program (or use the Alt+F4 keyboard shortcut).
Yes No
Select Yes and the analysis program closes. The analysis continues and data continue to be
collected. Queued reports under the print manager will print. If a power failure occurs and an
uninterruptible power supply (UPS) is not attached to the analyzer, the data collected after exiting
the analysis program are lost.
2. Place the computer, monitor and printer ON/OFF switches in the OFF position.
Shortcut Menus
Shortcut menus are available for:
• onscreen graphs and tabular reports. Refer to Report Shortcut Menus, page 7-17.
Shortcut Keys
Shortcut keys can be used to activate some menu commands. Shortcut keys or key combinations (when
applicable) are listed to the right of the menu item.
Another shortcut method of accessing a menu or function is to use the Alt key plus the underlined let-
ter in the menu command. For example, to access the File menu, press Alt + F. Then press the
underlined letter on the submenu. For example, Alt + F opens the File menu; then press O to access the
Sample Information file option.
If the underscored letters do not display menus and dialogs, press the Alt key on the
keyboard.
F6 Cascade windows
F7 Tile windows
F8 Start a report
Files
Analyzing Sample information files that are currently being used for analysis.
Complete Sample information files used in an analysis that has been completed.
No Analysis Sample information files which have not been used to perform an analysis.
Prepared Sample information files used in an automatic degas operation but not for an
analysis.
Preparing Sample information files currently being used in an automatic degas operation.
The following table lists the default file extensions assigned to files in the Gemini VII
program:
The following file types are available for reports saved from the Print window:
Report .REP
Spreadsheet .XLS
ASCII .TXT
Menu Structure
All program functions are located on menus accessed from the menu bar. Each menu contains com-
mands and, in some cases, a submenu. A submenu is indicated when the command is followed by an
arrow.
A SmartPrep menu option
displays on the menu bar if a
SmartPrep unit is used
with the analyzer.
Arrow indicates
a submenu is
available
Unit [n] Use to perform analyses, calibrations and other instrument operations.
Refer to UNIT MENU, page 6-1.
Reports Use to run reports and view the results. Refer to REPORTS MENU, page
7-1.
Options Use to edit sample defaults, specify system configuration and data
presentation formats. Refer to OPTIONS MENU, page 8-1.
Window Use to arrange the windows and icons on the computer screen and display
a list of open windows. Refer to Window Menu, page 2-11.
Help Use to access online operator’s manual, Micromeritics web page, and
videos. Refer to Help Menu, page 2-12.
Window Menu
The Window menu lists open files and provides the ability to rearrange the way open windows display.
A check mark appears to the left of the active window.
Window Menu
Description
Option
Cascade Stacks open windows in a fanned format so that each window title bar is
visible. F6 can also be used as a keyboard shortcut.
Tile Resizes open windows and arranges them horizontally so that multiple
windows can be viewed at once. F7 can also be used as a keyboard
shortcut.
Help Menu
The Help menu provides access to the online operator’s manual, the Micromeritics web page, online
instructional videos, and information about the analyzer.
Gemini VII on the Provides an Internet link to additional Gemini VII information:
Web Gemini VII on the Web
Evaluating Data
3. OPERATING PROCEDURES
Introduction
This chapter contains information on:
Format Description
Advanced Displays all parts of sample information and parameter files. Navigate to
parameter screens by clicking the tabs across the top of the window.
Basic Displays the entire sample information file in a single window with no tabs
across the top. This option is used once the parameter files have been created.
The previously entered or default parameter files are then accessible using
dropdown lists.
Restricted Displays the sample information file in a single window similar to the Basic
format with certain functions disabled. The Restricted format requires a
password when entering or exiting the Restricted format. It is typically used in
laboratories where analysis conditions must remain constant, for example, in the
pharmaceutical industry. The Advanced format is not available from the
Restricted format.
Advanced Format
The values specified in the parameter portions of the default sample file (Degas Conditions, Analysis
Conditions, Adsorptive Properties, and Report Options) are saved as the defaults for new sample files.
To navigate from one set of parameters to another, click the parameter tab at the top of the window.
Perform the following steps to establish defaults for a sample information file in the Advanced format.
2. In the Sequence Number alpha/numeric text box, specify a default file name string. The
Sequence Number field must contain a minimum of three (3) numbers. This number is
incrementally sequenced and displays in the File name text box in the File > Open > Sample
Information window.
3. The text box to the right of the Sample field, is used to enter a format for the sample
identification. Include the $ symbol to automatically include the Sequence Number field text as
part of the sample identification.
4. Enter Operator, Submitter and Bar Code identification information in the respective text boxes.
To omit the field and information in the sample files, select the Omit checkbox to the right of the
field.
The field labels for the Sample, Operator, Submitter and Bar Code fields can be modified
by overwriting the default field labels.
5. In the Mass group box, enter default values for the sample’s mass and density. Both of these
values may be edited at the time of analysis. Refer to Sample Information Files, page 5-8.
6. In the Type of Data group box, select if data are to be automatically collected or manually
entered.
7. The optional user-defined fields in the User Parameters group box may be used to enter and
track information from another instrument or source, along with other statistical process control
data. Refer to Sample Information Files, page 5-8 .
8. After completing the Sample Description window, click the parameter tabs to edit other sample
information file parameters. The saved parameter defaults become the defaults for new parameter
files.
Basic Format
When specifying defaults in the Basic format, select from existing default parameters. The defaults
established in Basic format also serve as defaults for files created using the Restricted format.
To define parameter files before entering sample defaults in Basic format, refer to Defining Sample
Information Files, page 3-6.
Perform the following steps to establish defaults for a sample information file in the Basic format.
3. In the Sequence Number alpha/numeric text box, specify a default file name string. The
Sequence Number field must contain a minimum of three (3) numbers. This number is
incrementally sequenced and displays in the File name text box in the File > Open > Sample
Information window.
4. The text box to the right of the Sample field, is used to enter a format for the sample
identification. Include the $ symbol to automatically include the Sequence Number field text as
part of the sample identification.
The field label for the Sample field can be modified by overwriting the default field label.
5. In the Mass group box, enter default values for the sample’s mass and density. Both of these
values may be edited at the time of analysis. Refer to Sample Information Files, page 5-8.
6. Select the dropdown arrow to the right of the Degas conditions, Analysis conditions, Adsorptive
properties, and Report options fields to select default parameter files that were saved while in
Advanced format and saved to the directory specified in Options > Parameter Files Directory.
Restricted Format
Restricted format presents the sample information file in the same manner as the Basic format except
that the operator cannot switch to Advanced format. This format is password protected and is typically
used in laboratories where sample parameters must remain constant, for example, in pharmaceutical
laboratories.
Advanced Format
1. Go to File > Open > Sample Information or use the F2 keyboard shortcut.
2. Accept the next sequenced file number or enter a new name in the File name text box. Click
Open.
5. Enter Operator, Submitter and Bar Code identification information in the respective text boxes.
6. In the Mass group box, enter default values for the sample’s mass and density. Both of these
values may be edited at the time of analysis. Refer to Sample Information Files, page 5-8.
7. In the Type of Data group box, select if data are to be automatically collected or manually
entered.
9. After completing the Sample Description window, click the parameter tabs to edit other sample
information file parameters.
• If using the Basic format, switch to the Advanced format to edit parameter file values.
• If using the Restricted format, parameter files cannot be edited.
1. Go to File > Open > Sample Information or use the F2 keyboard shortcut.
2. In the File name field, accept the next sequenced file number or enter a new file name. Click
Open.
3. Click OK to create a new file. If Advanced displays in the dropdown list at the bottom of the
window and there are tabs across the top of the window, click the Advanced dropdown arrow and
select Basic to change to Basic format. If Advanced does not show in the dropdown list, then you
are in Restricted format.
The Restricted
format does not
show an
Advanced option.
5. In the Mass group box, enter default values for the sample’s mass and density. Both of these
values may be edited at the time of analysis. Refer to Sample Information Files, page 5-8.
6. Click the dropdown arrows to select default parameter files for Degas conditions, Analysis
conditions, Adsorptive properties, and Report options.
If there is an existing sample file containing the values to be used in the current sample
file, click Replace All to choose the file and copy the values into the current file.
2. To sort the file list by status, click the Mic Status column header. Select a file with a status of
Complete, Analyzing, or Entered and click Open (or double click the file name). The reporting
window displays.
3. Save the changes in the sample information file by clicking the Save button.
2. To sort the file list by status, click the Mic Status column header.
3. Select the file to open in the Name column and click Open (or double click the file name).
4. The Sample Information window displays in the specified format (Advanced, Basic, or
Restricted). The following is shown in Advanced format.
• Degas conditions
• Analysis conditions
• Adsorptive properties
• Report options
Default parameter files can be used for multiple analyses without having to re-enter the values each
time an analysis is performed.
Predefined parameter files are included with the program and can be edited as needed or new parame-
ter files can be created. The Replace button can be used to replace values from an existing file to
overwrite values in the current file. Changes can be made as needed to the new file while the original
file remains unchanged.
New parameter files must be saved in the directory specified in Options > Parameter
Files Directory for the files to appear in dropdown lists in Basic and Restricted format.
Degas Conditions
Degassing is a required step in preparation for the analysis; however, this section is
only applicable if using the SmartPrep Degasser. This section contains degassing
instructions that will be sent to the SmartPrep Degasser equipment.
5. Use the Insert button to enter up to five stages of degassing (soak temperature, ramp rate, and
soak time).
6. Click Save, then click Close. The file saves with a .DEG file extension.
Analysis Conditions
Analysis conditions specify the data used to guide an analysis.
1. Go to File > Open > Analysis Conditions or use the F3 keyboard shortcut.
5. Indicate the points for the pressure table. Use the Insert, Delete, Clear, and Append buttons to
assist in entering data.
7. Click Save, then click Close. The file saves with a .ANC file extension.
Adsorptive Properties
Adsorptive properties provide the adsorptive (analysis gas) characteristics for the analysis.
1. Go to File > Open > Adsorptive Properties or use the F4 keyboard shortcut.
4. Enter a description in the Adsorptive text box (for example, the gas and the temperature).
5. Enter the mnemonic for the Adsorptive gas (for example, CO2) in the Mnemonic text box.
• Non-ideality factor
• Density conversion factor
• Molecular cross-sectional area
7. Click Save, then click Close. The file saves with a .ADP file extension.
Report Options
Report Options files specify the type of reports that will be generated from an analysis or from manu-
ally entered data. They also contain report details, for example, axis scale, axis range and column
headings. Report options files may contain tabular reports, plots, or both, as well as user-defined report
tables.
Report options files may be defined to include overlay options. This system allows the overlay of up to
eight plots of different samples onto a ninth plot of the same type or overlay one plot type onto a differ-
ent plot type from the same analysis. Refer to Generating Graph Overlays, page 3-35.
1. Go to File > Open > Report Options or use the F5 keyboard shortcut.
5. Select Show report title and enter a title to appear on the report header. Deselect this option to
omit a report title.
6. Select the Show graphic checkbox to insert a graphic in the report header. Use the Browse button
to locate a .BMP or a .EMF file. Specify the graphic size in the Height and Width text fields.
Deselect this option to omit a report graphic.
7. The Selected Reports list box displays the reports that may be generated.
a.) Select the checkbox to the left of the report to include in this file. A selected box displays a
checkmark in the box.
b.) To specify report options, highlight the report in the Selected Reports list box and click
Edit. Makes changes as necessary. Click OK.
8. Click Save, then click Close. The file saves with a .RPO file extension.
Clean the sample tube Cleaning and Labeling Sample Tubes, page 3-17
Create the sample file Defining Sample Information Files, page 3-6
Load sample on sample port Installing the Sample Tube, page 3-21
Fill Dewar and check LN2 level Installing the Analysis Dewar, page 3-22
3. Use 5 grams of Alconox (or other suitable detergent) per 500 mL of warm water and fill the
ultrasonic unit with enough water to cover the sample tubes and filler rods (if used). Ensure the
detergent is dissolved before placing the sample tubes and filler rods into the water. If too much
detergent is used, it may be difficult to rinse from the sample tubes.
4. Fill the sample tubes with warm water and place them in the ultrasonic cleaning unit. Then place
the filler rods in the unit. Turn on the ultrasonic cleaning unit for approximately fifteen minutes.
5. Use rubber gloves to ensure no oils or residue are transferred to the clean tubes and filler rods and
remove the sample tubes and filler rods from the unit.
6. Clean the interior of the sample tubes with the brush supplied with the system.
7. Rinse the sample tubes and filler rods thoroughly with hot water. Then rinse again with isopropyl
alcohol or acetone. If isopropyl alcohol or acetone is not available, deionized water may be used.
8. Stand the sample tubes on the sample tube rack and place the filler rods in a basket or in the rack.
Bake in a vacuum oven for two hours at 110 °C.
9. Remove the sample tubes and filler rods from the oven and allow to cool.
Do not insert the filler rods at this time. Filler rods are inserted just before the sample
tube is installed on the analysis port.
• weigh the empty Sample Tube Set (sample tube and stopper or seal frit) before degas
• weigh the Sample Tube Set with sample before degas and subtract the weight of the empty Sample
Tube Set
• weigh the Sample Tube Set with sample after degas and subtract the weight of the empty Sample
Tube Set
• weigh the Sample Tube Set with sample after analysis and subtract the weight of the empty Sample
Tube Set
A FORMS for recording the weights and calculating the mass is included in Appendix A. Make copies
as needed.
2. Place the sample weighing support on the balance. Tare the balance and allow it to stabilize at
zero (0).
3. Place the sample tube set on the sample weighing support and place it on the balance.
4. Record the stabilized weight on the Sample Data Worksheet as [A] Mass for empty sample tube
set. Remove the sample weighing support and sample tube set from the balance.
Do not touch the sample with bare hands while performing the following steps. Doing
so could affect the accuracy of results.
5. Place a sample container on the balance and slowly pour the sample into the container.
6. Remove the rubber stopper (or seal frit) from the sample tube.
7. Use the sample tube funnel (provided in the accessories kit) and pour the sample from the
weighing container into the sample tube.
Funnel
9. Weigh the sample tube set containing the sample and record the value on the Sample Data
Worksheet as [B] Sample tube set plus sample mass (Before Degas).
10. Subtract the [A] Mass for empty sample tube set from the [B] Mass of sample tube set plus sample
and record this value as the [C] Sample mass (Before Degas).
If using the SmartPrep degasser, a menu displays on the program menu bar. Degas the sample using
menu commands and information entered in the Degas Conditions window. Refer to the SmartPrep
operator’s manual for operating instructions.
1. Weigh the sample tube set containing the sample and record the weight on the Sample Data
Worksheet as [B] Sample tube set plus sample mass (After Degas).
2. Subtract the [A] Mass for empty sample tube set (Before Degas) from the [B] Sample tube set plus
sample mass (After Degas) to obtain the sample’s mass. Record this value as [C] Sample mass
(After Degas).
2. Place the connector nut, ferrule, and O-ring onto the sample tube stem.
O-ring
Ferrule
Connector nut
Sample tube
The sample tube ferrule is tapered slightly on one end. The ferrule may be installed onto the
sample tube with the tapered end in the up or down position.
If analyzing samples with a surface area of 1.0 m2 or less, use filler rods in the sample
and balance tubes. Refer to ANALYZING SAMPLES WITH TOTAL SURFACE
AREA OF 1.0M2 OR LESS, page E-1.
3. Attach the sample tube to the analysis port. Ensure it is fully in the port. Secure it in place by
screwing the connector nut onto the analysis port. Hand-tighten the connector nut.
4. If the balance tube is not installed, attach it to the balance port following the above procedures. If
a filler rod was used in the sample tube, one must be used in the balance tube.
It is not necessary to remove and replace the balance tube between analyses unless it
has been contaminated or if using a different size sample tube.
5. Place the Dewar cover under the sample and balance tubes and slide upward until it is three inches
from sample nuts.
Always handle glass Dewars with care. Any product incorporating a vacuum is a
potential safety hazard and should be treated with caution. Always observe the
precautions listed below.
• Use a plastic stirring rod when stirring substances in a Dewar containing liquefied gases (or other
materials of extremely low temperature). Do not use a glass or metal stirring rod unless it has a
protective coating.
• Do not remove the mesh covering from the Dewar flask. This covering helps to maintain a secure
grip on the Dewar.
• Do not handle heavy objects above the Dewar. If unavoidable, place a protective cover over the
Dewar’s opening. If an object of sufficient weight is accidentally dropped into the Dewar, shattering
may occur.
• Use a plastic stirring rod when stirring substances in a Dewar containing liquefied gases (or other
materials of extremely low temperature).
• Do not handle heavy objects above the Dewar. If unavoidable, place a protective cover over the
Dewar’s opening.
2. Fill the Dewar with the analysis bath liquid (liquid nitrogen):
Incorrect fluid levels can lead to measurement errors. Check the level of the bath liquid
before each analysis.
3. Insert the dipstick into the Dewar and check the level of the analysis bath liquid. Condensation
should not exceed the Level Indicator mark.
Wetness or frozen
condensation indicates
bath liquid level
4. For best results, if the Dewar has not been in use for a while, allow approximately 30 minutes for
the temperature of the Dewar to stabilize with the bath liquid. Then recheck the level of the bath
liquid. Add additional liquid if necessary.
Dewar support
(not used with
Gemini VII 2390t)
A Dewar support is not necessary for the larger Dewar used with the Gemini VII 2390t. Place the
Dewar directly on the elevator.
Performing an Analysis
Begin analysis after the sample has been degassed and transferred to the analysis port. The system pro-
vides these types of analyses:
Start Analysis
Ensure the analyzer compartment door is closed before beginning an analysis. If the
analyzer is operated at an excessive pressure, the sample or balance tube could
become dislodged from its port, possibly causing personal injury or damage to the
equipment.
2. Click Browse to select a sample information file or accept the next sequenced number.
3. Verify the information populated into the sample identification, density and mass fields. This
information is pulled from the selected file. The density value is applicable only if using the
Calculate method for the free space determination.
4. Click Report after analysis to automatically generate reports when the analysis is complete. On
the Report Settings window, select the report destination. Click OK.
5. Click Start to start the analysis. A window displays data as they are collected. A short delay is
encountered before the port status changes from the Idle state at the bottom of the screen.
6. When the analysis is complete, remove the sample tube and store (or dispose) of the sample
material as applicable.
QuickStart Analysis
When using the QuickStart option, it is not necessary to prepare sample files in advance. Sample files
are assigned automatically and given default conditions. File contents can be replaced if necessary
using the Browse button.
2. File names are assigned automatically and given default conditions. To replace the parameters
with those in another sample information file, click Browse and select the sample file.
3. Verify the information populated into the sample identification, density, and mass fields. This
information is pulled from the selected sample file. The density value is applicable only if using
the Calculate method for the free space determination.
4. Click Report after analysis to automatically generate reports when the analysis is complete. On
the Report Settings window, select the report destination. Click OK.
5. Click Start to start the analysis. A window displays data as they are collected. A short delay is
encountered before the port status changes from the Idle state at the bottom of the screen.
6. When the analysis is complete, remove the sample tube and store (or dispose) of the sample
material as applicable.
Use caution when removing the sample tube if using a hanging filler rod. The sample
tube O-ring or Dewar cover may snag the filler rod retaining ring. Loosen the snag
gently; excessive force may break the tip of the filler rod.
Blank Analysis
Use this method to perform a blank analysis with no sample data. The blank analysis is used primarily
for diagnostic purposes to look for instrument problems. The blank analysis should result in zero
adsorptive. An amount other than zero could indicate a possible leak.
1. Open a new Sample Information file using Unit [n] > Blank Analysis.
3. Click Report after analysis to automatically generate reports when the analysis is complete. On
the Report Settings window, select the report destination. Click OK.
4. Click Start to start the analysis. A window displays data as they are collected. A short delay is
encountered before the port status changes from the Idle state at the bottom of the screen
When analysis is complete, remove the sample tube and store (or dispose) of the sample material as
applicable.
Use caution when removing the sample tube if using a hanging filler rod. The sample
tube O-ring or Dewar cover may snag the filler rod retaining ring. Loosen the snag
gently; excessive force may break the tip of the filler rod.
Generating Reports
Reports > Start Report > [file] (or use the F8 keyboard shortcut)
1. Select the report in the Name column. The selected report name appears in the File name text
box. To select more than one file, hold down the Ctrl key on the keyboard while selecting the
files. Click Open.
2. Select the report destination in the Report Settings window and click OK. Refer to Printing
Files, page 3-30.
3. If only one file was selected in step 1, the Select Reports window displays. Verify the reports to
generate and click OK. If multiple files were selected, this window does not display.
4. Click a tab across the top of the window to review each report.
Printing Files
Provides the option to print the contents of one or more sample files to either the screen, a printer, or to
a file.
2. From the Name column, select a file to print. To select more than one file, hold down the Ctrl key
on the keyboard while selecting the files. To sort a column, click the column header.
3. Click Open.
• Copies - select the number of copies to print. This field is only enabled when Print is selected.
• File - select the destination directory and enter a file name in the File name text box. Select to
save the file as a .REP, .XLS, or .TXT file format.
5. Click Print. The following example shows a sample information file printed to the screen.
Listing Files
Provides the option to create a list of sample file information, for example, file name, date / time the
file was created or last edited, file identification and file status.
2. From the Name column, select a file to list. To select more than one file, hold down the Ctrl key
on the keyboard while selecting the files. To sort a column, click the column header.
3. Click List.
• Copies - select the number of copies to print. This field is only enabled when Print is selected.
• File - select the destination directory and enter a file name in the File name text box. Select to
save the file as a .REP, .XLS, or .TXT file format.
5. Click OK. The following example shows a sample information file printed to the screen.
Description
Operator
Submitter
Sample Mass
* Type of Data
0 = automatically collected
1 = entered (relative pressures)
2 = entered (absolute pressures)
2. From the Name column, select a file to export. To select more than one file, hold down the Ctrl
key on the keyboard while selecting the files. To sort a column, click the column header.
3. Click Open. The file is automatically converted to a .SMP file and the following window
displays.
• Copies - select the number of copies to print. This field is only enabled when Print is selected.
• File - select the destination directory and enter a file name in the File name text box. Select to
save the file as a .REP, .XLS, or .TXT file format.
Converting Files
Converts StarDriver files (.MGD extension) to a sample information file with a .SMP file extension.
Only those files with a .MGD file extension display in the Name column. Multiple files can be
selected by holding down the Ctrl key on the keyboard while selecting the files.
2. From the Name column, select a file to convert. To select more than one file, hold down the Ctrl
key on the keyboard while selecting the files. To sort a column, click the column header.
3. Click Open. The file is automatically converted to a .SMP file and the following window
displays.
• Multiple Sample Overlays - overlay up to eight plots of the same type with that of the current plot.
• Multiple Graph Overlays - overlay two different types of plots from one sample. This type of
overlay is available only for:
– BJH Adsorption/Desorption
– DFT Pore Size/Surface Energy
– Dollimore-Heal Adsorption/Desorption
– Horvath-Kawazoe
– M-P Method
Only the Advanced format can be used to generate overlays. Select Options > Options Presentation >
Advanced to access the Advanced format or select Advanced from the dropdown list at the bottom of
the screen.
If a file with a status other than Preparing, Prepared or No Analysis is selected, the Isotherm
plot displays. Select Advanced from the dropdown list at the bottom of the screen.
2. Click the Report Options tab at the top of the screen. Refer to the following table for instructions
for the selected report.
• BET Surface Area a.) In the Selected Reports list box, highlight one of the report
• Langmuir Surface Area options shown on the left and click Edit.
• Freundlich
• Temkin b.) On the Report Options window, select the Overlay samples
• t-Plot checkbox for the Transform plot and/or the Isotherm plot.
• Alpha-S Verify other fields. Click OK to return to the Report Options
• f-Ratio tab.
a.) In the Selected Reports list box, highlight one of the report
options shown on the left and click Edit.
• BJH Adsorption
• BJH Desorption
b.) Select the report variable from the Select Reports group box
• Dollimore-Heal
and click Edit.
Adsorption
• Dollimore-Heal c.) Click the dropdown arrow on the Overlay field and select the
Desorption Samples option. Verify other fields. Click OK to return to the
• MP-Method Report Options window.
4. Click Browse to the right of the Sample [n] field. Select the file containing the graph to overlay
and click Open. After selecting up to eight files, click OK to return to the Report Options tab.
Be advised when working with an existing file that a copy of the file be used rather than
the original.
6. Select Reports > Start Report. The Start Report window displays with the name of the edited
and saved file in the File name text box. To select more than one file, hold down the Ctrl key on
the keyboard while selecting the files.
7. Click Open.
8. Select the report destination in the Report Settings window and click OK. Refer to
Printing Files, page 3-30.
9. If only one file was selected, the Selected Reports window displays. Verify the reports to
generate and click OK. If multiple files were selected, the following window will not display.
10. The report screen displays with tabs across the top of the screen. Click each tab to view the
reports.
• BJH Adsorption
• BJH Desorption
• DFT Pore Size
• DFT Surface Energy
• Dollimore-Heal Adsorption
• Dollimore-Heal Desorption
• Horvath-Kawazoe
• MP-Method
The following instructions use BJH Adsorption as an example. Screen appearance will vary depending
on the selected report.
2. Select a sample file to overlay graphs of other samples. To select more than one file, hold down
the Ctrl key on the keyboard while selecting the files. To sort a column, click the column header.
Click Open.
If a file with a status other than Preparing, Prepared or No Analysis is selected, the Isotherm
plot displays. Select Advanced from the dropdown list at the bottom of the screen.
3. Click the Report Options tab. In the Selected Reports list box, highlight the type of report to
overlay with a graph and click Edit.
Highlight the
report, then click
Edit.
4. From the Report Options window, in the Selected Reports list box, select a sub-report and click
and click Edit.
Click Edit
5. Click the dropdown arrow on the Overlay field and select a second variable to overlay. Click OK
to return to the Report Options window.
Be advised when working with an existing file that a copy of the file be used rather
than the original.
8. Select Reports > Start Report. Select the file created in the previous steps and click Open. To
select more than one file, hold down the Ctrl key on the keyboard while selecting the files.
9. Click Open and select the report destination in the Report Settings window and click OK.
10. If only one file was selected in step 8, the Selected Reports window displays. Verify the reports
to generate and click OK. If multiple files were selected, the following window will not display.
11. The report screen displays with tabs across the top of the screen. Click each tab to view the
reports.
Introduction
This chapter provides instructions for:
The analysis program can also be installed for offline data manipulation on a computer other than the
one controlling the analyzer. This allows:
Review the Micromeritics PROGRAM License Agreement for restrictions on the use of another copy
of the analysis program.
When finished working with the setup CD, store it in a safe location.
• Configure an Ethernet port on the computer to communicate with the analyzer. Refer to
ETHERNET CONFIGURATION, page D-1.
• Attach the analyzer to the computer’s configured Ethernet port and turn on the power to the
analyzer.
The application should not be installed on a network drive with shared access.
Multiple users cannot operate the application at the same time.
CDs containing calibration files end with a suffix of 99. Update CDs end with a suffix of 00.
1. Insert the installation CD into the CD-ROM drive. The program automatically starts the
installation. If the installation does not immediately start, navigate to the CD-Rom drive, locate
and double-click the setup.exe file.
System Installation videos for hardware components and verification of operation are available by
clicking the Click here to view the system installation videos link on the Ethernet Introduction
screen.
3. On the Ethernet Introduction screen, verify that the prerequisites have all been met. If not, click
Cancel and complete the steps, then restart the installation program. Do not proceed with
installation until these tasks have been completed. If the prerequisites have been met, click Next.
4. On the Choose Users screen, select one of the following, then click Next:
• Install for anyone using this computer - anyone who logs onto this computer will have
access to the analyzer program.
• Install just for me - only the person logged onto the computer at the time of installation will
have access to the analyzer program.
5. On the Choose Install Location screen, verify that enough disk space is available to continue the
installation. If not, click Cancel and free up the minimum required disk space. If the disk space is
OK, click Next to accept the default installation directory or click Browse to change the
installation location, then click Next.
6. On the Choose Start Menu Folder screen, select a folder for the program shortcut. Select one of
the following:
7. When installation completes, the Gemini VII Installed screen displays. Click Next.
8. The Analyzer Configuration screen is used to configure the system for the each analyzer and
serial number to be installed.
If... Then..
moving an analyzer from select 0 when using the Moving an Analyzer from One PC to
another computer to this Another PC, page 4-16 option.
one
a.) In the Step 2 group box, enter the serial number for each analyzer to be attached to this
computer. The analyzer’s serial number label is located near the power switch.
9. On the Ethernet Ports screen, select one option from the following table:
If... Then...
the configured port is listed Select the port then click Next to proceed to the next step.
10. The Analyzer S/N [n] Installed screen displays. Click Next. A screen displays for each unit
installed.
11. If installing a SmartPrep at this time, on the Configure SmartPrep screen, select the number of
SmartPrep instruments to be installed and the Com Port connected to the first SmartPrep. Click
Next. If not installing the SmartPrep, select None for step 1 and click Next.
12. The Product Release Information screen displays containing information about updates and
changes included in the release. Click Next.
13. The Product Registration screen displays. Click the appropriate link to register the product or to
log into your portal or click Next to register later.
14. The Completing the Gemini VII Setup screen displays. Select one of the following:
• Reboot now
• I want to manually reboot later
Select the Create a desktop icon checkbox to have the system install a program shortcut onto the
desktop, then click Finish to close the Setup program.
1. Ensure that the analysis program is not operating and the analyzer is idle.
2. Insert the setup CD into the CD-ROM drive. The setup program starts automatically.
If the program does not start automatically, navigate to the CD drive and double-click the
setup.exe file.
4. On the Select Task screen, select the task to perform, then click Next. Use the following sections
to complete the selected task.
The setup program automatically detects which type of installation applies and customizes the selec-
tion accordingly.
1. Insert the setup CD into the CD-ROM drive. The setup program starts automatically.
If the program does not start automatically, navigate to the CD drive and double-click the
setup.exe file.
3. On the Select Task screen, select one of the following software options (only the applicable
option displays) and click Next:
If the option Downgrade software to version (previous number) displays, contact your
Micromeritics representative before proceeding. Reverting to a previous version can
damage existing sample files.
5. The Gemini VII Reinstalled screen displays. Select one of the following:
• Reboot now
• I want to manually reboot later
Select the Create a desktop icon checkbox to have the system install a program shortcut onto the
desktop. Click Finish.
Adding an Analyzer
An Ethernet switch is required when connecting multiple analyzers. After connecting the Ethernet
switch, continue with the Software, page 4-14 section.
Ethernet Switch
1. Connect the power cord of the Ethernet switch to an appropriate power outlet.
2. For each analyzer to be added, connect an Ethernet cable from the Ethernet labeled port on the
rear of the analyzer to a numbered port on the Ethernet switch taking care not to use the uplink
port on the switch.
4. Ensure the power is turned on to the computer, the Ethernet switch, and each analyzer.
5. Ensure the power is turned on to the computer, the Ethernet switch, and each analyzer.
Software
1. Start the setup program. The Welcome screen displays. Click Next.
3. On the Add Analyzer screen, enter the serial number of the analyzer being added. Click Next.
4. The Analyzer S/N [n] Installed screen indicates a successful install. Click Next.
1. Ensure the analyzer program is installed on the destination computer. If this is a new install, refer
to Installing the Software, page 4-2 and select ’0’ as the number of analyzers to be installed.
2. Start the setup program on the Source computer. On the Welcome screen, click Next.
3. On the Select Task screen, select Move an analyzer from one PC to another PC. Click Next.
4. On the Move Analyzer screen, use the dropdown list to select the source analyzer to be moved.
Click Next.
5. On the new Move Analyzer screen, select the new destination for the analyzer to be moved. This
location is where the Analyzer SN [n] Installer.exe file will be located. To accept the default
location, click Next or click Browse to select a new destination, then click Next.
6. The analyzer serial number [n] files are moved to the install program and the Analyzer Installer
Created screen displays. Click Finish.
a.) Remove the CD from the CD-ROM drive and store in a safe location.
b.) Disconnect the Ethernet cable from the Source computer and reconnect it to the configured
Ethernet port on the Destination computer. If moving the analyzer to a computer that already
has an analyzer attached, an Ethernet switch is required.
c.) If the move destination was a local drive, navigate to the drive and copy the Analyzer SN [n]
Installer.exe file to a media device such as a USB stick. Install the file on the destination
computer, then double-click the Analyzer SN [n] Installer.exe file.
If the move destination was a network drive, navigate to the location then double-click the
Analyzer SN [n] Installer.exe file.
Sample and parameter files are not copied and moved with the analyzer. Use a file
management program such as Explorer or a backup/restore utility to move these
files.
Removing an Analyzer
1. Start the setup program. On the Welcome screen, click Next.
3. On the Remove Analyzer screen, from the dropdown list, select the unit and serial number of the
analyzer to be removed. Click Next.
This operation removes the selected instrument from the list of attached instruments.
It does not remove calibration and status files associated with the analyzer nor does it
remove sample and parameter data files.
2. On the Select Task screen, select Change analyzer setup. Click Next.
3. On the Ethernet Ports screen, select one option from the following table:
If... Then...
the configured port is listed Select the port then click Next to proceed to the next step.
2. On the Select Task screen, select Reinstall calibration files for an analyzer. Click Next.
3. On the Reinstall Calibration Files screen, select the unit and serial number from the dropdown
list. Click Next.
2. On the Select Task screen, select Import an analyzer from a previous installation on this PC.
Click Next.
3. On the Import Analyzer screen, click the Browse button to locate the installation directory of the
previously installed instrument, then select the serial number of the instrument files to be
imported. Click Next.
4. When the previously installed files have been imported, the Analyzer SN [n] Installed screen
displays. Click Next.
2. On the Select Task screen, select Add or Modify SmartPrep. Click Next.
3. On the Configure SmartPrep screen, select the number of SmartPrep units to be installed and the
Com Port for the first SmartPrep unit. Click Next.
4. On the Completing the Gemini VII Setup screen, to place a shortcut icon on the desktop, select
the Create a desktop icon checkbox. Click Finish to complete the process.
1. Double-click the uninstall.exe file located in the directory where the Gemini VII files are
installed.
4. The Completing the Gemini VII Uninstall screen displays. Click Finish to close the setup
program.
5. FILE MENU
Introduction
This chapter contains information specific to the File menu options used to manage sample and param-
eter files. This chapter provides details of File menu options, commonly used functions and buttons,
and field descriptions.
Menu bar
Arrow indicates
submenu
Common field and button descriptions are listed in a Common table at the beginning
of their respective chapters. Field and button descriptions not listed in the Common
table are listed in their appropriate heading.
Refer to the Appendix section of this manual for further details on report calculations
(CALCULATIONS, page C-1) and DFT models (DFT MODELS, page F-1).
Autoscale checkbox When working with report parameters screens, when enabled, allows the X-
and Y-axes to be scaled automatically.
Axis Range When working with report parameters screens, the From / To fields are
enabled when Autoscale options are not selected. Enter the starting and
ending values for the X- and/or Y-axes.
Browse button Click to search for a file. Select from a file in the Name column and click
Open or double click the file name in the Name column to open (or import)
the file.
Close button Closes the active window. If changes were made to the file and not yet
saved, a prompt displays providing the option to save the file.
Delete button When working with report parameters screens, click Delete to remove the
selected report. Deleted reports will have to be regenerated if deleted in
error.
Destination • Preview - sends the file to the screen. Click Print on the report screen to
group box send the file to the printer.
• Copies - select the number of copies to print. This field is only enabled
when Print is selected.
• File - saves the report as a file. Click the Browse icon to the right of the
text field to select the directory where the new file will be stored. Enter
the new file name in the File name text box.
• File Type - select to save the new file with a .TXT, .XLS or .REP file
extension. This field is only enabled when File is selected.
.REP (Report system) - saves the report in a format that can be opened
within the program.
.TXT (ASCII text) - saves the report as a common machine language
file.
.XLS (Spreadsheet file) - saves the report in a format that can be opened
within a spreadsheet program.
Edit button When working with report parameters screens, highlight the item in the
Selected Reports list box and click Edit to modify report details.
File name text box Select the file to print (or open) from the Name column. The file name will
appear in the File name text box. Click Open or double-click the file name
in the Name column to open the file. Multiple files can be selected by
holding down the Ctrl key on the keyboard while selecting multiple files.
From / To text boxes When working with report parameters screens, enter the From and To range
for autoscaling the X- and/or Y- axes.
Open button Click to open the selected file (or double-click the file name in the Name
column to open the file).
Preview button Click to preview predefined reports. Click the tabs across the top of the
window to preview each selected report. Refer to Report Tool Bar, page 7-
14. When an analysis has not been run on a sample, this button is disabled.
Print button Sends the report to the selected destination (screen, printer or file).
Replace button Click to select another file where the values will replace the current file’s
values. The original file will remain unchanged.
Save As menu option Saves the active file under a new file name.
Table buttons • Insert - inserts one row above the selected row.
• Delete - deletes the selected row.
• Clear - clears all table entries and displays only one default value.
• Append - inserts one row at the end of the table.
Open
Opens a sample file, parameter file, or report option.
• Sample Information files - the File name text box contains the next sequential file name generated
by the program. The sample information file extension is .SMP.
• Parameter files - the File name text box contains an asterisk (*) and a default file extension
depending on the type of parameter file selected. Default file extensions are:
Save
Saves the active window under the current file name.
Save As
Saves the active window as another file type or file name.
• Saves a subset (parameter) of the sample file in the active window as a standalone parameter file, for
example, select Analysis Conditions from the Save As menu to create a standalone parameter file of
the analysis conditions portion of the active sample file.
Alpha-S - saves the relative pressures and resulting quantities adsorbed as an ASCII text file with a
.ALS file extension. These data are derived by dividing the isotherm by the quantity adsorbed at 0.4
relative pressure.
Save All
Saves all active windows under the current file names.
Close
Closes the active window. If changes were made and not yet saved, a prompt displays providing the
option to save the file.
Close All
Closes all active windows. If changes were made and not yet saved, a prompt displays providing the
option to save the file.
Print
Prints the contents of one or more sample or reports options files to either the screen, a printer, or to a
file.
List
Provides the option to create a list of sample or report options file information, for example, file name,
date / time the file was created or last edited, file identification and file status.
Export
Exports isotherm data in a sample information file as a .TXT or .XLS file. When saved to a file, the
data can be imported into other applications. The output file contains elapsed time, absolute pressure,
relative pressure, and specific quantity adsorbed in column format.
Convert
Converts the following sample file types to be compatible with the Gemini VII application:
StarDriver carbon.MGD
Exit
Exits the analysis program.
• If changes have been made to a file and not yet saved, a prompt displays providing the option to save
the file.
• If an analysis is in progress, a prompt displays providing the option to exit the program.
Each analysis must be linked with a sample information file before the analysis can proceed. A sample
information file can consist of multiple parameter files; however, parameter files can also stand alone.
Parameter files allow repeated use of the file, for example, if the same analysis conditions exist for
multiple analyses, create an Analysis Conditions file containing the recurring conditions. When the
sample file is created, select the Analysis Conditions file for the analysis conditions. Once it becomes
part of the new sample file, edit the new file as needed without affecting the original Analysis Condi-
tions file. Sample Information files can be created or opened in Advanced, Basic, or Restricted
formats.
Specify or change the default format by selecting Options > Option Presentation or
select Basic / Advanced from the dropdown list at the bottom of the window. Refer to
Specifying Sample Defaults, page 3-1 for a description of the Advanced, Basic, and
Restricted formats.
Operator / Submitter Enter the name of the operator or submitter performing the analysis.
text boxes
Bar Code text box Enter bar code reader information if a bar code reader is connected to the
computer’s USB port. If a bar code reader is not used, this alpha/numeric
field can be used to enter additional information about the sample; such as, a
sample lot number, sample ID, chain of custody, etc.
Mass group box • Enter - when selected, enables the Sample Mass field. Enter a value for
the sample mass.
• Calculate - enables the Empty tube and Sample + tube fields. Enter the
values necessary to calculate the sample mass. Equation used to calculate
sample mass:
• Density - value is used for the Calculated free space method only. Use
0.000 for a blank analysis.
Type of Data group • Automatically collected - select if the type of data will be automatically
box collected while an analysis is running.
User Parameters These fields are primarily used for the SPC (Statistical Process Control)
group box reporting to specify sample characteristics or its manufacturing process but
may be used for other data by entering specific analysis conditions or
sample criteria.
These parameters display on the Sample Description window, in Reports >
SPC Report Options, and the Summary Report.
Modify the default field labels (alpha/numeric) in Options > Sample
Defaults.
Comments text box Enter comments about the sample or its analysis conditions. Comments will
appear in the report header.
Add Log Entry Click to enter information to appear in the sample log report. Click the
button button again to enter multiple log entries.
Replace All button Click to select another file where the values will replace all values for the
Sample Information file. The original file will remain unchanged.
Close button
Preview button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Save button
The Degas Conditions tab provides details for setting up the Degas Conditions parameter file. This
information will be automatically applied during the degassing procedure if using the SmartPrep
Degasser. Use this option when the SmartPrep Degasser is installed.
Conditions table Enter up to five stages of degas conditions. Available columns are:
• Temperature - soaking temperature with flowing gas.
Close button
Preview button
Replace button Refer to the field and button table in Common Fields and Buttons - File
Menu Options, page 5-1.
Save button
Table buttons
The Analysis Conditions tab provides details for setting up the sample analysis conditions file.
Use this window to manually enter the pressures or use the Insert Predefined button to insert default
system values.
A pressure table consists of relative pressure points (and possibly calculation assignments) at which
data (isotherms) are to be collected. The relative pressures may span the entire range of 0.00000001 to
0.995 p/po. There must be one adsorption branch followed optionally by one desorption branch.
Several analysis conditions files containing complete pressure tables are included with the analysis
program. To use these files, it is recommended to use the File > Save As option to save the file as a dif-
ferent name and edit the new file.
The pressure table for automatically collected data includes the pressure points for data collection and
(when the Use calculation assignments checkbox is selected) identifies calculations through which
the data are processed.
Replace button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Absolute pressure Select to specify pressure targets in mmHg, mbar, or kPa rather than
dosing checkbox relative pressure. This option is typically selected when using adsorptives
at analysis conditions above the critical point of the gas; for example, H2
adsorption on carbon at liquid nitrogen temperature.
Use calculation Select to use assigned collected data points. If this option is not selected,
assignments checkbox each report (with the exception of Langmuir and BET) will use a range of
pressures as selected in the report options. The Langmuir and BET reports
interpolate to entered relative pressures on the report options.
The outlier points can be selected so they will not be reported.
Insert Range Click to enter starting pressure, ending pressure, the number of points to
button insert within the specified range and whether to have linear or geometric
progression.
0.01
0.02
0.04
0.08
0.16
0.99
0.98
0.96
0.92
0.84
Insert Predefined Insert Predefined button - this button is disabled if Absolute pressure
button dosing is selected on the previous window. The predefined pressures are
logical only if nitrogen is used as the adsorptive. These options can be used
to select the system’s default pressures which can then be edited on the
Entered Pressures window.
Click the dropdown arrow to select predefined pressure points for surface
area, t-Plot micropore, and/or BJH adsorption/desorption.
Free Space button Click to enter the type of free space measurement.
po and T button Click to select options for obtaining the saturation pressure (Po) and
analysis bath temperature.
po and T button • Entered - select to manually enter the Po in the text box.
(continued)
• Most recent measured - select to use the most recently measured
saturation pressure.
• Measure Po for each point (models 2390p and 2390t only) - select if
Po is to be measured for each point.
Cancel button
Close button
OK button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Save button
Table buttons
The Adsorptive Properties window is used to specify the characteristics of the gases used during an
analysis.
Adsorptive text box Enter a description of the adsorptive gas or select the adsorptive using the
Replace button.
Non-ideality factor Enter a factor to compensate for the forces of attraction between molecules
text box in a real gas. This value is used for a calculated free space.
Density conversion Enter the density conversion factor which is determined by obtaining the
factor text box ratio of the gas density (STP) to the liquid volume.
Close button
Preview button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Replace button
Save button
Use to specify report options for collected (from an analysis) or manually entered data. They also help
in customizing report details such as axis scale, axis range, column headings, and components of thick-
ness curve equations. Refer to REPORTS MENU, page 7-1.
Customized report options files can be created then loaded into the original converted sample file
allowing the quick and easy generation of reports.
To include the parameter files in the dropdown list on the Basic and Restricted Sample
Information windows, save the file to the directory specified in Options > Parameter Files
Directory. Refer to UNIT MENU, page 6-1.
Show report title Select and enter a report title to appear on the report header.
checkbox and text box
Show graphic Select to show a graphic on the report header. Click the Browse button to
checkbox and text box locate the graphic in either .BMP or .EMF format.
• Height / Width - enter the height and width of the selected graphic.
These values determine the graphic appearance on the generated report.
Apply thermal Select to correct the temperature-induced pressure difference between the
transpiration manifold and the chilled sample tube. This option is most significant for
correction checkbox pressures less than approximately 1.0 mmHg. Do not use filler rods in the
sample tube when applying correction for thermal transpiration.
Always use thermal transpiration when performing micropore analyses.
Refer to CALCULATIONS, page C-1 for additional information on
thermal transpiration.
Inside diameter of Enabled when Apply thermal transpiration correction is selected. Enter
sample tube text box the inside diameter of the sample tube used in the analysis.
Selected Reports Select the checkbox to the left of the report names to include in the report.
list box
Calculated pore width (w), pore radius (R) or pore diameter (D) for BJH
reports. Select Options > Units to specify default calculations.
Browse button
Close button
Edit button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Preview button
Replace button
Save button
Summary Report
The Summary Report provides a condensed listing of selected data results. In the Selected Reports
list box, highlight Summary, then click Edit. Select the data types to include in the Summary report.
In the Pore Volume column, if Adsorption total or Desorption total is selected, the P/Po field is
enabled. Enter the relative pressure used to calculate the total pore volume.
If Use calculation assignments is not selected on the Collected Data window, the isotherm is interpo-
lated to the value in the P/Po fields and the point is used for the total pore volume calculation.
Otherwise, the point selected with calculation assignment is used. Pass/Fail criteria can be specified
for up to four parameters.
Select All / • Pass/Fail [n] button - select the S A: Single-point BET checkbox to
Deselect All buttons enable this button. Click the Pass/Fail [n] button to display the Pass/
(continued) Fail Options window for selection of pass/fail criteria.
Item [n] group boxes • Upper/Lower options and text boxes - specify upper and lower limits
for the selected parameter. A range can be left open by not selecting the
limit. In the text box to the right of Upper / Lower, enter a message
line of operator instructions to be displayed if a failure is encountered.
Cancel button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Select Reports Select the checkbox to the left of each applicable report to include the
group box option on the final report.
Options buttons Click to display related linear plot options. All plot windows contain
identical fields.
Options buttons • Autoscale X-axis - linear X-axes begin at zero. Logarithmic X-axes
(continued) begin at an appropriate value. The X-axis field shows the relative or
absolute pressure.
• Autoscale Y-axis - Y-axes begin at zero. The Y-axis field shows the
quantity of gas adsorbed.
Weight group box Enabled when the Pressure Composition plot checkbox is selected on the
Isotherm Report Options window.
• Adsorbate Molecular Weight - this plot is useful for plotting pressure
as a function of Weight % adsorbed (i.e. H2 adsorbed on carbon).
Cancel button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
• The BET calculation obtains the sample surface area value by determining the monolayer volume
of adsorbed gas from the isotherm data.
• The Langmuir calculation determines the surface area of a sample by relating the surface area to
the volume of gas adsorbed as a monolayer.
BET (or Langmuir) fit Enter values in the From / To fields to indicate the fit pressure range.
pressure range text
boxes
Pressures button Click to enter a range of pressure points or to modify table values for
pressure points.
BET (or Langmuir) Select to generate a traditional BET (Langmuir) surface area plot used to
Transform plot determine monolayer volume and BET C constant.
• X-axis - linear X-axes begin at zero. The X-axis field shows the
relative pressure for BET and show absolute pressure for Langmuir.
• Y-axis - Y-axes begin at zero. The Y-axis field shows BET (Langmuir)
transformation.
BET (or Langmuir) Select to use the BET (Langmuir) monolayer volume and constant to
Isotherm plot produce an isotherm.
Uses the BET (Langmuir) monolayer volume and constant to produce an
isotherm.
• X-axis - linear X-axes begin at zero. The X-axis field shows the
relative pressure for BET and show absolute pressure for Langmuir.
• Y-axis - Y-axes begin at zero. The Y-axis field shows the quantity of
gas adsorbed.
Cancel button
Close button
From / To text boxes
OK button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Preview button
Save button
Table buttons
Specify monolayer Select and enter the monolayer capacity of the sample.
capacity checkbox
Absolute pressure Enter a pressure range when calculation assignments are not used.
range text boxes
Tabular Report
checkbox
Refer to BET/Langmuir Surface Area Report Options, page 5-24.
Overlay Samples
checkbox
Freundlich Isotherm Plots the absolute pressure vs quantity adsorbed. Shows best fit line.
plot checkbox
• X-axis - linear X-axes begin at zero. The X-axis field shows the
absolute pressure.
• Y-axis - Y-axes begin at zero. The Y-axis field shows the quantity of
gas adsorbed.
Cancel button
From / To text boxes Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Specify differential Select and enter the differential heat of adsorption at zero surface
heat of adsorption coverage. This allows inclusion of all Temkin constants.
checkbox
Absolute pressure Enter a pressure range when calculation assignments are not used.
range text boxes
Temkin Isotherm plot Overlays the Temkin isotherm with the analysis data.
checkbox
• X-axis - linear X-axes begin at zero. The X-axis field shows the
absolute pressure.
• Y-axis - Y-axes begin at zero. The Y-axis field shows the quantity of
gas adsorbed.
Tabular Report
checkbox
Refer to BET/Langmuir Surface Area Report Options, page 5-24.
Overlay Samples
checkbox
Cancel button
From / To text boxes Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Thickness Curve Select the thickness curve. Click Edit to modify the values in the equation
group box for the selected curve. The Frenkel-Halsey-Hill thickness curve can be
applied using the Halsey option.
Reference option - select Reference and click the Edit button to define a
t-curve by entering the relative pressure and thickness values. One
predefined curve is shipped with the analysis program and is found in the
Reference directory.
Thickness Curve
group box (continued)
To import values from an existing thickness curve (.THK file), click Open
and select the file containing the values. The table to be imported must be
saved as ASCII text with a .THK extension. It must have a two-column
format with the relative pressures in the first column and the thickness
values in the second column. Columns must be separated by a space or a
tab.
Kruk-Jaroniec-Sayari / Halsey / Harkins and Jura / Broekhoff-de
Boer / Carbon Black STSA - select the thickness curve option and click
the Edit button. Modify the equation for the selected curve as needed.
Surface Area Select the surface area value used for thickness calculations. BET is the
group box most commonly used option.
Fitted thickness range Enter the minimum and maximum thicknesses to include in the thickness
text boxes curve. The values entered for the thicknesses can be expressed in
angstroms or nanometers. Select Options > Units to specify default units.
Refer to UNIT MENU, page 6-1.
Pressure Range button Click to enter minimum and maximum relative pressures.
If Use calculation assignments is not selected on the Analysis
Conditions window, all of the non-outlier points of the collected data
within the specified range are used for calculating the data for this report.
If Use calculation assignments is selected, collected data points assigned
to this report type are used.
Surface area Enter the value to correct for surface areas that are not smooth. This brings
correction factor the values for BET surface area and micropore surface area into
text box accordance. For most samples, the default value of 1.000 is adequate.
Tabular Report
checkbox
Refer to BET/Langmuir Surface Area Report Options, page 5-24.
Overlay Samples
checkbox
Cancel button
From / To text boxes Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Table buttons
Alpha-S Method
The Alpha-S plot converts the standard adsorption isotherm into a dimensionless isotherm using the
quantity adsorbed at a relative pressure of 0.4. In the Selected Reports list box, highlight Alpha-S
Method, then click Edit.
One predefined curve is shipped with the analysis program and is located in the Reference directory.
Use the table to enter relative pressure and the alpha-s values.
Open button To import values from an existing thickness curve (.ALS ASCII file), click
Open, and select the file containing the values.
The table to be imported must be saved as ASCII text with a .ALS
extension. It must have a two-column format with the relative pressures in
the first column and the alpha-s values in the second column. Columns
must be separated by a space or a tab.
Relative pressure Enter the pressure range when calculation assignments are not used.
range text boxes
Fitted Alpha-S range Enter minimum and maximum relative pressures to determine the fit.
text boxes
Ref. surface area Enter the surface area from the reference curve. This value is used to
text box calculate the sample surface area.
Tabular Report
checkbox
Refer to BET/Langmuir Surface Area Report Options, page 5-24.
Overlay Samples
checkbox
Browse button
Cancel button
From / To buttons Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Save As button
Table buttons
f-Ratio Method
The f-Ratio report uses the measured isotherm and normalizes it using a reference isotherm. In the
Selected Reports list box, highlight f-Ratio Method, then click Edit.
Reference isotherm Click Browse to select a sample file to use as a reference for the isotherm.
Select a file containing an isotherm measured from a non-porous sample of
the same material as the current sample. When the referenced file is
selected, the file name appears to the left of the Browse button.
Relative pressure Enter a pressure range when calculations assignments are not used.
range text boxes
Tabular Report
checkbox
Refer to BET/Langmuir Surface Area Report Options, page 5-24.
Overlay Samples
checkbox
Browse button
Cancel button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
From / To text boxes
OK button
The fields for both report options are identical unless otherwise specified. In the Selected Reports list
box, highlight BJH Adsorption (or BJH Desorption), then click Edit.
BJH Correction Select the type of correction to apply to calculations. The selected type
group box displays in the report header.
• Standard - uses original BJH models.
• Kruk-Jaroniec-Sayari - good for reference thickness curves.
• Faas - good for statistical thickness curves.
Minimum / Maximum Enter the minimum and maximum diameter (radius or width) of pores to
BJH width text boxes include in the BJH reports.
Fraction of pores During adsorption calculations, the analysis program assumes that all
open at both ends text pores are closed at one end. Occasionally, a percentage of pores may be
box open at both ends causing disagreement in the adsorption and desorption
data or in the values for total volume and total BJH pore volume. Enter the
fraction of pores open at both ends to compensate for this error.
This field is not available for the BJH Desorption Report Options
window.
Adsorptive button Displays the BJH Adsorptive Options window. The recommended
adsorptives and their values are shown. Up to eight adsorptive / adsorbate
property factor combinations may be specified.
Smooth differentials Select the checkbox to smooth all differential calculations thus eliminating
checkbox variations in the differential computation caused by noise in the input data.
Cumulative Reports • Larger - select to report the total volume found in pores larger than the
options current pore size.
• Smaller - select to report the total volume found in pores smaller than
the current pore size.
Pressure Range button Click to enter minimum and maximum relative pressures.
If Use calculation assignments is not selected on the Analysis
Conditions window, all of the non-outlier points of the collected data
within the specified range are used for calculating the data for this report.
If Use calculation assignments is selected, collected data points assigned
to this report type are used.
Cancel button
Edit button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Fixed pore size table Select to specify exact pore sizes for volume or area data.
Click the Table button to modify the fixed pore size table. Reference
Table and Columns below for information on the use of these buttons.
Collected points Select to include all relative pressure points collected. Reference the
Columns button referenced below.
Columns button Click the dropdown arrow and select the data types to include in the report.
These selections indicate the column order and data contents for the report.
Table button The fixed pore size table must contain a minimum of two points. The
points must be strictly decreasing. Enabled only when Fixed pore size
table is selected.
Cancel button
OK button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Table buttons
Plot Options
The fields for all plot options are identical for specifying plotting methods and customizing plots.
Highlight any plot option in the Selected Reports list box and click Edit.
Autoscale checkbox
Cancel button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Dollimore-Heal Adsorption/Desorption fields and buttons are identical to the BJH Adsorption/
Desorption Report Options, page 5-36.
Or Dollimore-Heal
Desorption Report
Options
The Horvath-Kawazoe method plots individual peaks for different pore sizes even if the difference
between one pore size and the next is only one Angstrom (0.10 nm) or less. In the Selected Reports
list box, highlight Horvath-Kawazoe, then click Edit.
Pore Geometry Select the option that best represents the physical geometry of the
group box micropores in the sample material. When Sphere is selected, options in the
Interaction Parameter group box are disabled.
Interaction Parameter Select to determine which interaction parameter will be used in the report.
group box These options are disabled if Sphere is selected in the Pore Geometry
group box.
• Computed - select to calculate using the parameters on the Horvath-
Kawazoe Physical Properties window (click the Properties button to
display the Physical Properties window). The interaction parameter is
recalculated each time a parameter on the Physical Properties window
is edited.
• Entered - select to calculate using the value entered in the text box.
Properties button Click to view or edit the constants describing the physical properties of the
adsorbent and adsorptive.
• Diameter - enter the diameter of the gas phase (or adsorbent) atom.
Properties button • Diameter at zero energy - enter the diameter of an atom at zero
(continued) interaction energy: (2/5)1/6 × diameter.
• Density - enter the density per unit area of the adsorbent (or
adsorptive).
Apply Cheng/Yang Select to apply the Cheng/Yang correction to the pore size analysis. This
correction checkbox correction substitutes the Langmuir equation of state for Henry’s Law in
the Horvath-Kawazoe derivation.
Smooth differentials
checkbox
Refer to BJH Adsorption/Desorption Report Options, page 5-36.
Pressure Range button
Cancel button
Edit button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Plot Options
Highlight a plot option in the Selected Reports list box and click Edit to customize the plotting
method.
X-Axis / Y-Axis • X-Axis - the X-axis field shows pore radius or diameter in angstroms
or nanometers.
• Variable dropdown list - click to select a Y-axis variable for the report.
Autoscale checkbox
Cancel button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Edit button
OK button
Models list box Lists the models that meet the criteria specified and which match
the adsorbate and temperature of the sample data. If no models
appear, no models meet the selected criteria. One model must be
selected.
Reports to Generate Click the checkbox to select the reports to generate. To edit
list box graph details, highlight the graph option and click Edit. The Log
Goodness of Fit and Goodness of Fit graphs cannot be edited.
• Plot Type group box - select the method for data display.
Cancel button Refer to Common Fields and Buttons - File Menu Options,
page 5-1.
From / To text boxes
OK button
DFT Surface Energy Report Options fields and buttons are identical to the DFT Pore Size Report
Options, page 5-46.
Report Type Select the checkbox to specify report types. If Astakhov is selected, either
group box select the Optimize exponent checkbox or enter an appropriate exponent
value in the text box.
Fitted Radushkevich Enter the minimum and maximum limits on relative pressures included in
(Astakhov) relative the line fit.
pressure range
text boxes
Pressure Range button Refer to BJH Adsorption/Desorption Report Options, page 5-36.
Adsorptive button
Cancel button
Edit button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Log (po/p )^n - the value for [n] is the optimized exponent if Optimized Astakhov exponent is
selected on the Dubinin Report Options window. If not, then the value for [n] is the entered exponent
value.
The transformed Dubinin isotherm is the logarithm of quantity adsorbed as a function of the log of rel-
ative pressure raised to a power. Isotherms for which the Dubinin method is applicable produce
straight lines when transformed in this way.
Overlay Samples Refer to BET/Langmuir Surface Area Report Options, page 5-24.
checkbox
Autoscale x-axis / Select an option to have the X- and/or Y-axes scaled automatically. Both
Autoscale y-axis axes begin at 0; the system uses the highest values collected during
checkboxes analysis as the ending points for axis ranges.
To enter beginning and ending values manually, deselect these
checkboxes.
X-axis - shows the quantity of gas adsorbed at standard temperature and
pressure.
Y-axis - shows the log of relative pressure.
Cancel button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Autoscale x-axis / Select an option to have the X- and/or Y-axes scaled automatically. Both
Autoscale y-axis axes begin at 0; the system uses the highest values collected during
checkboxes analysis as the ending points for axis ranges.
To enter beginning and ending values manually, deselect these
checkboxes.
Cancel button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Pore size can be expressed in angstroms or nanometers. Select Options > Units to
specify the unit. Refer to UNIT MENU, page 6-1.
Pressure Range button Refer to BJH Adsorption/Desorption Report Options, page 5-36.
Cancel button
Edit button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
The MP-Method reports hydraulic radius only. If Pore size in diameter is selected from
the Unit Selection window, pore size in radius will be reported.
Plot Options
Highlight a plot option in the Selected Reports list box and click Edit to customize the plotting
method.
Autoscale checkboxes
Cancel button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
OK button
Options Report
The Options report provides a subset of the contents report. The Options report cannot be edited. It
provides pertinent information for:
• Degas conditions
• Adsorptive properties
• Analysis conditions
• Free space
• Saturation pressure (Po) and temperature
• Analysis method
• Isotherm collection
Validation Report
Use this report to have the data examined by the software to determine if the results are within typical
ranges.
If the data selected for validation are determined to be out of range, a warning displays and suggestions
are given for corrective action. This information is also detailed in the report and plotted on the graph
as a unique plot symbol.
If Automatically Collected is selected, a Collected Data tab displays after the analysis is completed.
Pressures group box For manually entered data only, select if the pressures are Relative or
Absolute.
Close button
Save button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Preview button
Table buttons
6. UNIT MENU
Introduction
This chapter contains information specific to the Unit menu options. These options are used to run
analyses on one or more units attached to a controlling computer. This chapter provides details of Unit
menu options, and descriptions of commonly used functions and buttons.
The title bar of the main window displays as Unit [n] - S/N [n] for each attached unit. Each unit has a
status window displayed in different colors.
The Unit menu does not appear on the menu bar if the analysis program is being used
for offline data operations on a computer other than the one controlling the analyzer.
Common field and button descriptions are listed in a Common table at the beginning
of their respective chapters. Field and button descriptions not listed in the Common
table are listed in their appropriate heading.
Browse button Click to search for a file. Select from a file in the Name column and click
Open or double click the file name in the Name column to open (or
import) the file.
Cancel button Discards any changes made to the screen or cancels the current process.
Start Analysis
Unit [n] > Start Analysis
Use to begin an analysis. Select an existing analysis file or create a new one. Then specify analysis
defaults.
Port Activity
window
Status
window
Mass text field Use to modify the sample mass weight. If the field is disabled, click Edit
then click the Enter option in the Mass group box on the Sample
Description tab.
Sample + Tube text Use to modify the weight of the sample plus the sample tube. If the field is
field disabled, click Edit then click the Enter option in the Mass group box on
the Sample Description tab.
Empty Tube text field Use to modify the weight of the empty sample tube. If the field is disabled,
click Edit then click the Enter option in the Mass group box on the
Sample Description tab.
Start / Next button Starts the analysis and displays a graph during analysis. When the first
analysis is complete, the Start button changes to Next.
Browse button Refer to Common Fields and Buttons - Unit Menu Options, page 6-2.
Close button
QuickStart Analysis
Unit [n] > QuickStart Analysis
Use to analyze samples of the same type using the same analysis conditions, adsorptive properties, and
report options. The QuickStart Analysis option automatically creates a sample information file for
and assigns the default parameters. Edit the parameters of a file using the Edit button or select another
file using the Browse button.
Additional analyses can be scheduled by clicking Next after the completion of the first series of analy-
ses. The Next button appears after the first set of analyses is complete. The sample cannot be removed
or added until the analysis has completed.
QuickStart Analysis fields and buttons are identical to the Start Analysis, page 6-2.
Start Po Measurement
Use to measure the saturation pressure.
Model 2390a Measure the Po only in the Sample tube. Ensure a sample tube with no
filler rod is attached to the sample port then click Start to begin the
measurement.
Models 2390p and A Po tube is installed on the Model 2390p and 2390t. Measure the Po in
2390t the Po tube or in the Sample tube. Ensure a sample tube with no filler rod
is attached to the sample port, then click Start to begin the measurement.
The window closes at the end of the measurement. The value obtained is stored and recorded on the
Unit Configuration window as the most recently measured Po and is available for subsequent sample
analyses. The Po option is selected from the Analysis Conditions tab by clicking the Po and T button.
Blank Analysis
Unit [n] > Blank Analysis
Use to obtain sample tube parameters when using a calculated free space. A blank analysis can also be
used for diagnostic purposes.
Blank Analysis fields are identical to the Start Analysis, page 6-2 window with the exception of Den-
sity, Sample + Tube, Mass, and Tube fields.
Use to enable the manual control of certain system valves and elevator components. When this option
is enabled, a checkmark appears to the left of the menu item. Enable this option by selecting Enable
Manual Control on the menu.
If the following instrument schematic is not immediately visible, select Unit [n] > Show Instrument
Schematic.
Open Sample File Click to open the Open Sample Information File window.
button
Component Description
Servo Valve
Elevator
Component Description
Reservoir for the balance tube (left side) and sample tube.
Represents the balance tube (left side) and the sample tube.
Saturation pressure (Po) tube. Not displayed for Gemini VII 2390a.
System Valves
Select Unit [n] > Show Instrument Schematic to display the instrument schematic.
Valves Description
1 Vacuum valve
2 Adsorptive valve
3 Helium valve
When the cursor changes to this icon, right click to display options for the
selected component:
• For Valves:
Open - opens the selected valve. The valve symbol changes to green. An
alternate method is to either double click the valve or select the valve
and press the keyboard spacebar to turn it off/on.
Open Both - opens valves 4b and 4s or valves 5b and 5s. The valve
symbol changes to green. An alternate method is to either double click
the valve or select the valve and press the keyboard spacebar to turn it
off/on.
Close - closes the selected valve. The valve symbol changes to yellow.
An alternate method is to either double click the valve or select the valve
and press the keyboard spacebar to turn it off/on.
Close Both - closes valves 4b and 4s or valves 5b and 5s. The valve
symbol changes to yellow. An alternate method is to either double click
the valve or select the valve and press the keyboard spacebar to turn it
off/on.
Pulse - select to quickly turn the valve on and off allowing the operation
to proceed in small increments.
• For Balance Servo Valve:
Enable - enables the balance servo valve. Operations for the Sample
servo will also be performed by the Balance servo.
Disable - disables the balance servo valve. Operations for the Sample
servo will also be ignored by the Balance servo.
When the cursor changes to this icon, right click to display options for the
elevator:
• Raise - raises the elevator. Select Raise and press the keyboard space
bar to raise/lower the elevator.
• Lower - lowers the elevator. Select Lower and press the keyboard space
bar to raise/lower the elevator.
Use to display an analyzer schematic. To operate the valves and elevator from this window, Manual
Controls must be enabled (Unit [n] > Enable Manual Control).
Refer to Enable Manual Control, page 6-8 for additional details on this schematic.
Show Status
Unit [n] > Show Status
If multiple instruments are attached to the computer, the status window for each
instrument is displayed in a different color.
The buttons across the bottom of the window are described in Enable Manual Control, page 6-8.
The status window shows the current status for each port. Use this option if:
• there is an automatic operation in progress and the sample file needs to be edited.
• there are multiple units attached to the computer. Select Show Status on each unit menu and have
the status for all units displayed at one time.
Analysis status bar • Preliminary - evacuation and free space measurement. Displays green
during progression.
Step details Provides details of the current step in the analysis process.
Use to display a log of recent analyses, calibrations, errors or messages. This information is logged for
a 7-day period for analyses and a 30-day period for messages and calibrations.
Add Log Entry button Click to manually enter information into the instrument log window, for
example, Changed the port filter. To enable the Add Log Entry button,
enter information into the text box, then click Add Log Entry. The entry
then appears in the information log.
Report button Click to display the Log Report Settings window to specify report output
options.
Unit Configuration
Unit [n] > Unit Configuration
Use to display hardware/software configurations, calibrations and gas selections of the connected
analyzer.
Configuration Displays the IP address used by the analysis program and the serial number
group box of the selected analyzer.
Software Versions Displays the software versions of the MIC BIOS, controller and analysis
group box program.
Last Measured Po Displays the value for the last measured saturation pressure.
Volume Displays the volume correction of the attached analyzer. If a different value
Correction text box is entered in this field, it is stored with sample data at the start of an
analysis. The value in this field is used in a calculated free space.
Change IP button Click to display the Unit IP Setup window. Shows the IP address and
Subnet mask assigned during installation. Do not edit these fields unless
instructed by a Micromeritics service representative.
Restore Factory Deletes all stored analysis data and resets instrument parameters to factory
Defaults button defaults. It does not affect sample and parameter files.
Cancel button Refer to Common Fields and Buttons - Unit Menu Options, page 6-2.
OK button
Diagnostics
Unit [n] > Diagnostics
Use to clean and verify gas lines when connecting or changing a gas. Refer to Cleaning and Verifying
the Gas Line, page 9-22. Also used to perform diagnostic tests requested by a service representative.
This option is typically used and/or requested by a service representative. These tests create and save
files to the Gemini VII\Service\USERDIAG directory. The service representative may request an
emailed or faxed copy of the file for diagnostic purposes.
Repeat button Click the repeat the previous test. Enabled when the previous test is
complete.
Start / Next button Starts the analysis and displays a graph during analysis. When the first
analysis is complete, the Start button changes to Next.
Skip button Click to skip an analysis. This button is only visible when an analysis is in
progress. Select the ports containing the analysis to skip.
Suspend / Resume Click to suspend or resume an analysis in progress. Select the ports
buttons containing the analysis to suspend or resume.
Report button Click to view a list of reports to be generated during the test.
Item [n] dropdown list Click to select the data available for the current window.
Item [n] dropdown list Click to select the data available for the current window.
Cancel button Refer to Common Fields and Buttons - Unit Menu Options, page 6-2.
Refer to Report Tool Bar, page 7-14 for additional information on viewing, customizing, and printing
reports.
Calibration
Unit [n] > Calibration
Use to perform system calibrations. Disabled calibration options can be accessed only with the assis-
tance of a service representative and are not covered in this manual.
Zero Pressure
Unit [n] > Calibration > Zero Pressure
Use to evacuate the manifold and zero out the selected transducers. Refer to Zero Pressure, page 9-30.
Match Transducers
Unit [n] > Calibration > Match Transducers
Use to zero and match the selected transducers to the main manifold transducer. Refer to Match
Transducers, page 9-31.
Sample Servo
Unit [n] > Calibration > Sample Servo
Use to calibrate the servo valve to the sample pressure transducer. The sample pressure transducer
should be calibrated before starting this calibration procedure. The servo valve should always be reca-
librated after a pressure calibration has been performed. Refer to Sample Servo, page 9-31.
Balance Servo
Unit [n] > Calibration > Balance Servo
Use to calibrate the balance servo valve. Ensure that empty sample tubes are installed on the sample
and balance ports.
Save to File
Unit [n] > Calibration > Save to File
Use to save the current calibration settings to a backup file which can later be reloaded using Unit [n]
> Calibration > Load from File menu option.
File name dropdown list The default file naming convention for calibration files can be used
or the filename can be changed. The default file name is interpreted
as:
0217 - 2011-06-10.CAL
It is recommended that the current calibration settings be saved using Unit [n] > Calibration > Save to
File prior to loading another calibration file. When loading a previously saved calibration file, a
backup of the current file is created and saved as [SN]last.cal. The backup file is overwritten each time
a new one is created.
Changing the calibration may affect the instrument’s performance.
Service Test
Unit [n] > Service Test
Use for service tests performed only with the assistance of a trained Micromeritics service representa-
tive. These tests provide the service representative with troubleshooting tools and readouts.
This option is enabled only when the analysis program is operating in Service Test mode. Refer to
Service Test Mode, page 8-4.
7. REPORTS MENU
Introduction
This chapter contains information specific to the Reports menu options used to customize and run
reports. This chapter provides details of Reports menu options, commonly used functions and buttons,
field-by-field descriptions, and sample reports.
Cancel button Closes the window and cancels any unsaved changes.
Destination Refer to Common Fields and Buttons - File Menu Options, page 5-1.
group box
File name text box Select the file to print (or open) from the Name column. The file name will
appear in the File name text box. Click Open or double-click the file name
in the Name column to open the file. Multiple files can be selected by
holding down the Ctrl key on the keyboard while selecting multiple files.
From / To text boxes Enter the From and To range for autoscaling the X- and/or Y-axes.
Open button Click to open the selected file (or double-click the file name in the Name
column to open the file).
Show graphic Select to show a graphic on the report header. Click the Browse button to
checkbox and text box locate the graphic in either .BMP or .EMF format.
• Height / Width - enter the height and width of the selected graphic.
These values determine the graphic appearance on the generated report.
Start Report
Reports > Start Report > [file] (or use the F8 keyboard shortcut)
Close Reports
Reports > Close Reports (or use the F9 keyboard shortcut)
Use to close all open report windows. This option is unavailable if reports are being generated.
Open Report
Reports > Open Report > [file]
Use to generate reports with various SPC (Statistical Process Control) options. All selected variables
must be computed for each sample file used in an SPC report; therefore, it is more efficient to select
only the necessary variables.
The selected items appear as options on the Reports > Regression Report screen as selections in the
dropdown boxes and are used in graph selection in Reports > Control Chart.
The selected items also appear as options on the Reports > Control Chart screen. (Click the Graph [n]
button, then click the Statistic dropdown arrow.)
Regression Report
Reports > Regression Report
Use to generate an SPC (Statistical Process Control) Regression report to determine the interdepen-
dency between two variables. Up to three dependent variables (Y-axis) may be plotted against a single
independent variable (X-axis). The degree of correlation between the variables is also reported.
Show report title checkbox Select and enter a report title to appear on the report header.
and text box
Show graphic Select to show a graphic on the report header. Click the Browse button
checkbox and text box to locate the graphic in either .BMP or .EMF format.
• Height / Width - enter the height and width of the selected graphic.
These values determine the graphic appearance on the generated
report.
X- and Y-Axis Use to designate the X- and Y-axes variables. The variables in the
Variable dropdown lists are those selected in the Reports > SPC Report Options
dropdown lists window. Use these options to plot the regression of up to three Y-axis
variables against the X-axis variable.
Axis Range Enter the beginning and ending values for the X- and Y-axis ranges.
text boxes These fields are disabled if Autoscale is selected.
Autoscale checkboxes When enabled, allows the X- and Y-axes to be scaled automatically.
Tabular report checkbox Select to generate a tabular report of the included samples. A tabular
report contains the numeric values contributed by each sample.
Recalculate archived Select to have archived SPC values recalculated ensuring any changes
SPC results checkbox made to the SPC Report Options are included in the new report.
Selecting this option lengthens the time required to generate the report.
Label data checkbox Select to label the points on the plot to correspond with the values in
the sample files.
Samples button Click to select sample files for report generation. Multiple files can be
selected by holding down the Ctrl key on the keyboard while selecting
the files.
• Selected Files - files added from the Available Files list box.
• Use all files in this directory - click to select all files in the
Available Files list box. The Add and Remove buttons become
disabled.
• Add / Remove buttons - select a file in the Available Files list box
and click Add to move the file to the Selected Files list box. Or
select a file in the Selected Files list box and click Remove to move
the file back to the Available Files list box. Or double-click the file
name to move the file from one list box to the other.
Report button Click to view the report for the items selected.
Save as Default button Click to save selected report options as default report settings.
Browse button
Cancel button Refer to Common Fields and Buttons - Reports Menu Options,
page 7-2.
Destination group box
From / To text boxes
OK button
Control Chart
Reports > Control Chart
Use to generate an SPC (Statistical Process Control) control chart report which plots the changes in a
statistic.
X Axis Order by Select the order in which X-axis statistics are placed.
group box
• Time - sorts by the time the files were analyzed.
• Minutes - sorts by the minutes elapsed from the first file placed on the
list, which is the earliest-analyzed file.
• Days - sorts by the number of days elapsed from the first file placed on
the list, which is the earliest-analyzed file.
Graph [n] buttons • Center Line group box - displays placement options for the center line
(continued) in the graph. Choose Entered to specify placement of the line.
• Limit Lines group box - displays limiting lines options. Lines can be
placed at some multiple of the standard deviation or at specified
positions (Entered). When Entered is selected, enter the High limit
and Low limit fields with appropriate values.
Browse button
Cancel button Refer to Common Fields and Buttons - Reports Menu Options, page 7-
2.
Report button
Use to select sample files, define quantities and generate a Heat of Adsorption report. The isosteric
heat of adsorption is an important parameter for characterizing the surface heterogeneity and for pro-
viding information about the adsorbent and the adsorption capacity. Multiple adsorption isotherms are
obtained on the same sample using the same adsorptive but at different temperatures to obtain the heat
of adsorption.
Table Contains files added by using the Add Samples button. Also provides the
quantity adsorbed.
2. Double-click the file in the Name column or select the file name and
click Open.
Remove Sample Click to remove the selected sample from the list.
button
Edit Quantities Use to specify the range of surface coverage to include in the report.
button
• Insert Range button - click to specify the starting and ending quantities
adsorbed and number of points to insert.
Isostere plot Select to generate a graph showing quantities of gas adsorbed versus the
checkbox temperature.
Heat of adsorption Select to generate the Heat of Adsorption data in a graphical format.
plot checkbox
Browse button
Cancel button Refer to Common Fields and Buttons - Reports Menu Options, page 7-
2.
OK button
Report button
Save button
Reports
List Box
Generated
Report
Tabs
Header
Data display
(either graphical
or text)
Tool Bar
Report Header
All reports contain a header displaying file statistics.
If configured, the report header can also contain a graphic and a title.
• Tabular and graphical reports contain sample and instrument statistics such as analysis date / time,
analysis conditions, etc.
• The headers contain notes of sample file changes occurring after analysis.
• Summary report headers contain the same information as tabular and graphical reports with the
exception of notes.
Each report has a tool bar and selectable tabs across the top of the report header. To view a specific
report, select its tab or select the report in the Reports list box and click Show.
Reports can be customized and manipulated using the tool bar, shortcut menus, the zoom feature, or
axis cross-hairs.
Reports list box Contains a list of all generated reports. The same reports display as tabs
across the top of the report header unless the report has been hidden
using the Hide button.
Show button Jumps to the selected report in the Reports list box (or select the report
tab to show the report). If the report tab has been hidden using the Hide
button, click Show to display the report and tab.
Delete button Deletes the selected report. Deleted reports will have to be regenerated if
deleted in error.
Hide button Hides (or temporarily removes) the selected report from the tabbed view.
The report name remains in the Reports list box. To redisplay the tab,
select the report in the Reports list box and click Show.
Name dropdown list and Properties button - select the printer and
click the Properties button to change printout orientation, etc.
Save button Saves all reports of the active file using the sample file name with a
.REP file extension.
Save As button Saves all selected reports to the indicated file format:
• .REP (Report system) - saves the report in a format that can be
opened within the program.
Default Style button Click to specify default report parameters for fonts and curve properties.
Font Type list box - allows font type and attributes to be edited for
the selected item. Select an item in the list, click Edit, and select
from various font options. Click OK when done.
• Save button - click to save the changes as the new default settings.
• Close button - click to close the window and save the changes for the
current report.
Option Description
Edit Select to edit the report title and/or graphic to display in the report header.
Copy header as text Select to copy the report header as text. Text is copied to the clipboard and
can then be pasted into other documents.
Display tabular report shortcuts by right-clicking in the body of the tabular report. Column shortcuts
require right-clicking on the column to be modified.
Option Description
Resize column Right-click on the column to be resized. Select Resize Column on the
shortcut menu and enter the new column width in inches.
Rename column Right-click on the column to be renamed. Select Rename Column on the
shortcut menu and enter the new column name.
Move column Right-click on the column to be moved. Select Move Column on the
shortcut menu and select Left or Right for the move.
Align column Right-click on the column to be aligned. Select Align Column on the
shortcut menu and select Left, Right or Center.
Show column Displays a list of all hidden columns. Select a column to show.
Table data font Right-click in the report data. Select Table data font on the shortcut menu.
Deselect the Use default font to enable font options. Select new font
attributes for the report data. To return to the default fonts, select the Use
default font checkbox.
Table header font Right-click in the report data. Select Table header font on the shortcut
menu. Deselect the Use default font to enable font options. Select new
font attributes for the header. To return to the default fonts, select the Use
default font checkbox.
Edit title Select to edit the report title and/or title font attributes.
Copy table as text Select to copy the report contents to the clipboard as tab delimited text. It
can then be pasted into another document.
Graph Shortcuts
Display graph report shortcuts by right-clicking in the body of the graph report.
Option Description
Autoscale all axes Returns the report to full view after using the zoom feature.
Reset axis limits to Removes the cross-hair and returns the graph back to the initial setting.
initial setting
• Title text box - use to change the title of the selected curve.
• Style dropdown list - use to select another style for the collected data
curve.
• Label dropdown list - select where the graph point labels will display
(left, right, center, etc.) on the SPC report.
Graph Label
Title text box - use to modify the label of the selected axis.
Title font button - use to modify the font for the selected axis label.
Deselect the Use default font to enable font options. Select the font
attributes and click OK.
• Scale group box - use to change the graph display.
Edit legend Select to change the legend location and font. Click Font to modify legend
fonts. Deselect the Use default font to enable font options.
Edit title Select to change the graph title and font. Deselect the Use default font to
enable font options.
Copy Graph Copies the graph to the clipboard. It can then be pasted into other
applications.
Copy Data Copies the report data to the clipboard. It can then be pasted into other
applications as tab-delimited columns of text.
Zoom Feature
Use the zoom feature to closer examine graph details. To use this feature:
2. Hold down the left mouse button and drag the cursor across the graphical area to be enlarged. A
box will display in the area to be enlarged.
3. Release the mouse button. The enlarged area fills the graph area. To return to normal view, right-
click in the graph area and select Autoscale all axes or Reset all axes to initial setting on the
shortcut menu.
Axis Cross-Hair
X-axis
position
1. Click the left mouse button on the graph to view the cross-hair coordinates.
2. To remove the cross-hair, right-click in the graph area and select Autoscale all axes or Reset all
axes to initial setting from the shortcut menu.
Report Examples
This section of the manual contains samples of some of the available reports. Most of the
reports can be customized.
t-Plot
8. OPTIONS MENU
Introduction
This chapter contains information specific to the Options menu selections used to configure the system
by setting defaults for sample and parameter files.
Option Presentation
Options > Option Presentation
Use to change the default editing format - Restricted, Basic or Advanced. Each format type displays
sample information and options differently. For descriptions of the presentation types, refer to Sample
Defaults, page 8-2. When using Restricted format, a password is required to change to Advanced
format.
Sample Defaults
Options > Sample Defaults
Use to specify default parameters for sample information files and report option files in either
Advanced, Basic, or Restricted format. The Restricted format requires a password when entering or
exiting the Restricted format. It is s typically used in laboratories where analysis conditions must
remain constant, for example, in the pharmaceutical industry. The Advanced format is not available
from the Restricted format.
Restricted Format
Restricted format presents the sample information file in the same manner as the Basic format except
that the operator cannot switch to Advanced format. This format is password protected and is typically
used in laboratories where sample parameters must remain constant, for example, in pharmaceutical
laboratories.
Units
Options > Units
Use to specify how data should appear on the application windows and reports. This menu option is
not available if using Restricted format.
Use to select how grid lines appear on reports. This menu option is not available if using Restricted
format.
X-Axis / Y-Axis Select major and/or minor lines to display in reports for the logarithmic and
options linear scales. To remove the gridlines, deselect these options.
Grid Line Styles Select if the major and/or minor grid lines should appear as solid or dotted
options lines.
Service Test mode is a password protected option used to perform certain service tests with the assis-
tance of a trained Micromeritics service representative. This password is supplied by your service
representative. After Service Test Mode has been enabled, the service tests are accessible from the Unit
menu (Unit > Service Test).
Introduction
The analyzer has been designed to provide efficient and continuous service; however, certain mainte-
nance procedures should be followed to obtain the best results over the longest period of time. This
chapter includes maintenance and calibration procedures.
Troubleshooting
Most operational problems are caused by:
• Leaks (commonly around the sample tube O-ring at the analysis port)
• Sample weighing errors
• Use of too much analysis bath fluid in the Dewar at the start of an analysis
• Entry of incorrect system volume for analysis
• Impure gas supply
When unexpected analysis results occur, check the above first. Some common operational problems
that are not indicated on the screen and their respective causes and solutions are provided in the fol-
lowing table:6
Specified pressure not Sample or balance tube not Reattach sample and balance
reached. properly attached. tubes. Ensure both are securely
attached to ports. Replace O-
rings if defective. Refer to
Replacing Port Filter (Frit)
and O-ring, page 9-6.
Unable to reach a Vacuum pump oil level is low or Inspect the oil to see if it is low
satisfactory vacuum. needs to be changed. or needs changing. Refer to
Inspecting and Changing
Vacuum Pump Oil, page 9-8.
Centering ring has become too Check the centering ring at the
flat and unable to hold a vacuum. pump intake port. Replace if
necessary. Refer to Replacing
the Alumina in the Oil Vapor
Trap, page 9-11.
Nitrogen or helium drained Leak(s) in the gas line Perform the Adsorptive (or
from tank, or depleted in a connection. Helium) Line test to determine
short period of time. the location of the leak. Refer to
Diagnostics, page 6-17 for
details on these tests.
Unit does not work when Power cord not fully inserted at Insert power plug firmly into
power switch is turned on. one end. power source and analyzer power
connector.
Plug prongs bent so that contact Gently move power plug at outlet
not made at outlet. while watching power indicator
light. If indicator light comes on,
have electrician replace outlet or
plug.
Expected results are not Sample improperly degassed. Verify that the degassing
within range. temperature and degas time were
set properly.
Preventive Maintenance
The following table lists the preventive maintenance procedures to keep the analyzer operating at peak
performance. Instructions for each procedure follow the table. Micromeritics also recommends that
preventive maintenance procedures and calibration be performed by a Micromeritics Service Repre-
sentative every 12 months.
** For oil-free pumps only: After 12 to 18 months of operation, the diaphragm in the pump will wear
out and the pump becomes completely inoperable. To prevent instrument downtime due to an
inoperable pump, it is recommend that the diaphragm be replaced by a Micromeritics Service Rep-
resentative every 12 months.
Do not allow liquid to penetrate the casing of the analyzer. Doing so could result in
damage to the unit.
Ice and suspended frost particles may accumulate in the bottom of an analysis port Dewar. Particles or
deposits exceeding 1/4-in. (0.64-cm) in depth may jam between the bottom of the sample tubes and the
bottom of the Dewar, causing the Dewar not to raise fully. Accumulations of fine particles impede liq-
uid nitrogen circulation around the bottom of the sample tubes. This causes the sample temperature to
be slightly higher which, in turn, can cause pore volume measurement errors in those samples exhibit-
ing high isotherm slope above 0.97 relative pressure.
Accumulated ice is likely to melt and form a pool of water in the Dewar if all liquid nitrogen evapo-
rates. The water must be removed, otherwise it will solidify when liquid nitrogen is added and could
press on the bottom of the sample tube causing breakage.
To ensure problems do not develop due to ice accumulation, check the Dewar after each use. Clean the
Dewar on a weekly basis.
2. Go to Unit [n] > Show Instrument Schematic to display the instrument schematic, then go to
Unit [n] > Enable Manual Control.
3. Right-click on the elevator icon and select Lower to lower the elevator to its lowest position.
4. Remove the Dewar and pour the liquid nitrogen from the Dewar into an appropriate cryogenic
container.
Do not pour liquid nitrogen directly into a sink. Doing so may cause drain pipes to
burst.
5. Rinse the Dewar with warm water to melt any ice accumulation which may remain in the Dewar,
then dry thoroughly.
1. Select Unit [n] > Show Instrument Schematic, then Unit [n] > Enable Manual Control.
2. Right-click on the valve of the appropriate port and select Close to close the valve.
3. Use a wrench to remove the sample tube fitting from the port. Pry out the filter and O-ring.
O-ring
Filter (frit)
To avoid degassing problems, the filter and O-ring should be clean and should not be
touched with bare hands.
4. Replace the filter and O-ring. Carefully reassemble the sample tube fitting and reinstall on the
sample port. Tighten by hand, then with a wrench to prevent leaks.
Before removing (or installing) a sample tube, ensure that the port valve is closed.
Observe the instrument schematic to verify valve status.
1. Holding the sample tube firmly with one hand, loosen the sample tube connector nut by turning
counterclockwise.
Be careful not to let the sample tube connector nut drop onto the bottom of the tube as
it may break the tube.
Be sure to use the proper size frit for the sample port. The instrument will not operate
properly if the incorrect frit size is used.
2. Carefully pull the sample tube down until it is free from the port. It may be necessary to grasp the
sample tube with both hands.
O-ring
Ferrule
Connector nut
3. Remove the O-ring from the top of the sample tube and replace it with a new one.
If the O-ring remains inside the sample port, use a pair of tweezers or needle-nose
pliers to remove it.
4. After the new O-ring is in place, insert the sample tube back into the sample port until it is fully
seated.
5. Slide the sample tube connector nut up the tube (the ferrule and O-ring will move along with the
connector nut). Then, turning clockwise, hand-tighten the connector nut to the sample connector.
The oil in the vacuum pump should be changed every three months when the efficiency of the vacuum
pump declines (requiring increased time to reach vacuum levels) or if it becomes discolored. The oil is
easily inspected to determine if a change is necessary.
Oil-level window
Use oil supplied by Micromeritics or refer to the vacuum pump manual for other acceptable oils.
2. Loosen the wing nut on the clamp at the top of the oil vapor trap. Swing the clamp open and
remove the trap from the hose.
3. Grasp the handle on top of the vacuum pump and place it on a work table.
b.) Remove the plug from the drain spout and allow the oil to drain into the waste container.
Oil-fill port
6. Slowly add oil to the port until the level is midway between the indicator lines in the oil-level
window when the pump is running.
Midway level
Do not allow oil to rise above the midway position when the pump is running. Doing so
may cause oil to splash into the oil filter and contaminate it.
7. Check the washer or O-ring used at the oil-filling port and replace if necessary.
9. Check the alumina in the oil vapor trap. If most of the pellets are no longer white, replace the
alumina before reattaching the vacuum pump.
12. Allow the pump to run a few hours (overnight if possible) to eliminate air and moisture from the
fresh fluid and to produce efficient vacuum operations.
Do not perform the following procedure on used alumina. The resultant oil vapors may
cause a fire or an explosion.
1. Disconnect the vacuum pump from the analyzer and place it on a work table.
2. Loosen the wing nut on the clamp at the top of the oil vapor trap. Swing the clamp open and
remove the trap from the hose.
3. Loosen the wing nut on the clamp at the bottom of the oil vapor trap. Swing the clamp open and
remove the trap.
4. Remove one end fitting from the trap body and dispose of the used alumina in an appropriate
manner.
5. Wash the trap body with a detergent-based soap. Rinse with water, then rinse with isopropyl or
ethyl alcohol. Set the trap aside and allow to dry thoroughly.
Exposure of the trap body to oil vapor may cause small cracks on the inside surface of
the trap body. Under normal circumstances, these cracks will not cause problems or
leaks.
Do not perform the following procedure on used alumina. The resultant oil vapors may
cause a fire or an explosion.
b.) Pour approximately 180 grams of fresh alumina into a glass or metal container
(approximately 250 mL if a graduated beaker is used). Place the container in the oven.
d.) Remove the baked alumina from the oven and allow it to cool until luke warm. A desiccator
may be used to speed the cooling process.
7. Use a small spatula to gently pry the O-ring from the end fittings of each end of the trap body.
10. Ensure the trap body is dry and the alumina is lukewarm. Pour the alumina pellets into the trap
until level with the top of the trap body.
11. Screw the other end fitting back onto the trap and tighten securely by hand.
12. Lightly tap both ends of the trap body on the work surface to remove remaining dust from the
pellets.
13. Inspect the centering ring before placing it back on the intake port. If it appears to be flattened,
replace it. A flattened centering ring can cause vacuum leaks.
There are two types of centering rings. Use the one with the smaller opening at the intake port.
Centering ring
16. Open the clamp and place it around the flange of the intake port and the flange of the trap. Swing
the clamp fastening screw toward the intake port until it fits into the slot in the other half of the
clamp. Tighten the wing nut securely by hand.
17. Reconnect the hose from the analyzer to the oil vapor trap.
a.) Place the clamp around the vacuum pump hose flange and vapor trap.
b.) Swing the c-clamp fastening screw around until it fits into the slot on the other half of the
clamp. Tighten the wing nut securely by hand.
18. Plug the pump power cord into the power source.
19. Allow the pump to run a few hours (overnight if possible) to eliminate air and moisture from the
fresh oil and to produce efficient vacuum operations.
Exhaust filters are used to minimize the release of oil vapors. The gases are diluted
substantially upon release from the vacuum pump; however, it may be necessary in
some locations to provide a fume hood for protection from hazardous gases and
vapors released into the work area.
1. Turn the power to the unit off by either removing the plug from the power source or by turning the
power switch to the OFF position.
2. Loosen the wing nut of the clamp at the vacuum pump exhaust port. Swing the clamp away from
the exhaust port and remove it.
3. Remove and discard the exhaust filter. Do not remove the centering ring.
6. Open the clamp and place it around the exhaust port flange and the exhaust filter flange. Swing
the clamp fastening screw toward the exhaust port until it fits into the slot in the other half of the
clamp. Tighten the wing nut securely by hand.
7. Turn the power to the unit on by replacing the plug back into the power source or by turning the
power switch to the ON position.
2. Click the down arrow in the Test field and select System Leak Test Rev. [rev].
3. Select Report after test and Preview as the destination. Click OK.
4. Click Next. Data will be inserted into the window as they are collected.
5. A popup window indicates the test is complete. Click OK, then click Close.
7. Verify that all tests have a Passed status by clicking the tabs and looking for the Passed status for
each test run.
8. Click Save As and accept the default name displayed in the File name field.
9. Click Save. Make a note of the directory and filename, then click Close.
10. E-mail the file to the service representative requesting the report.
Connecting Gases
This procedure is also shown in one of the How-To videos located on the Help menu.
• Place gas bottles close to the analyzer. Using gas line extenders on gas bottles located in remote
areas may degrade gas quality and reduce pressure.
• Use a retaining strap (or other appropriate tether) to secure the gas bottle.
• Carefully route the gas lines from the bottle to the analyzer avoiding overlapping or entangling gas
lines.
• Label the gas line at the instrument inlet for proper identification and maintenance.
The following instructions describe a typical installation. Some configurations require additional com-
ponents, for example, regulator expansion kits, when one gas source will be used for several operations
or when the gas bottle cannot be located close to the analyzer.
Gas bottle
shut-off valve
Regulator
shut-off
valve
2. Both gauges should read at or near zero. If not, open the regulator gas shut-off valve to release
gas. It is not necessary to disconnect the gas line from the regulator or the instrument.
3. Use an appropriate wrench to loosen the nut at the regulator/gas bottle connection then remove
the regulator from the bottle.
4. Replace the protective cap on the depleted bottle. Disconnect the retaining strap and remove the
bottle from its current location.
When connecting hazardous gases, ensure there is proper ventilation and always
follow the safety procedures established for your lab.
A power failure or loss of cryogen can result in dangerous pressures in the sample
tube. The analyzer uses pressure relief valves to vent this pressure into the instrument
cabinet and return the instrument to a safe condition. When using toxic or flammable
gases, additional venting of the cabinet may be required.
1. Use an appropriate cylinder wrench to remove the protective cap from the replacement bottle.
2. Attach the gas regulator to the gas bottle connector. Hand-tighten the nut then use an appropriate
wrench to tighten an additional 3/4 turn.
3. Check for leaks at the high-pressure side of the regulator and in the connector.
HIgh-pressure Gauge
Gas bottle
shut-off valve
Regulator
Regulator connector nut
control knob
b.) Slowly open the gas bottle shut-off valve then close it.
Gas bottle
shut-off valve
Regulator
shut-off valve
a.) Turn the regulator clockwise until the low pressure gauge shows a few pounds of pressure.
Regulator
Gas bottle
shut-off valve
shut-off valve
Regulator
control knob
a.) Turn the regulator control knob clockwise until the low-pressure gauge reads 15 psig
(103 kPag).
c.) Open the gas bottle shut-off valve and flow gas for 10 to 30 seconds.
6. If the gas line to the instrument inlet was previously disconnected, reconnect it now.
Before beginning, confirm that the state for valves and the low-pressure gauge are as follows:
Low-pressure gauge
14 - 15 psig
2. Click the down arrow to the right of the Test field and select Clean and Verify Analysis Gas
Line Rev [n]. This procedure uses the analysis gas as an example. Select Clean and Verify
Helium Gas Line Rev [n] if the helium bottle is being verified.
3. Click Report after analysis and select Preview as the destination. Click OK.
4. Click Start. From the View dropdown list, select either Operation, Instrument Log, or
Instrument Schematic. Refer to Start Analysis, page 6-2.
5. A series of prompts display on the screen requiring operator response. The length of time a test
will run is also indicated on the screen.
6. A popup window indicates the test is complete. Click OK. The report displays on the screen.
On the report, the Leak Rate field on the Instrument to Regulator Shut-off Valve -1 report
should read Passed. A Passed reading indicates the gas line from the instrument to the regulator is
clean and leak-free.
If the Leak Rate field indicates Failed, a leak is indicated between this connection. Check the
connections from the instrument to the gas regulator valve. Tighten as necessary, then run the test
again.
7. Click the Instrument to Gas Bottle Shut-off Valve - 2 tab to display its report.
On the report, the Leak Rate field should read Passed. A Passed reading indicates the gas line
from the instrument to the regulator is clean and leak-free.
If the Leak Rate field indicates Failed, a leak is indicated between this connection. Check the
connections from the regulator shut-off valve to the gas bottle shut-off valve. Tighten as necessary,
then run the test again.
On the report, the Minimum field should read Passed. A Passed reading indicates all valves are in
the proper state for operation.
If the Minimum field indicates Failed, one or more valves is not in the proper position. Set the
valves as shown below and ensure the appropriate pressure is displayed on the low-pressure gauge.
Low-pressure gauge
14 - 15 psig
If running the test again, close the gas bottle valve before starting the test.
9. Click Close to close the test report. Click Close again to close the test.
Helium is required for performing a blank analysis since a measured free space is required. If helium is
unavailable, contact a Micromeritics Service Representative for an alternative method of performing
this procedure.
1. Install empty, straight-wall sample tubes (free of cracks or chips) of the same size in the sample
and balance ports. Ensure the O-rings are present and in good condition.
2. Place the Dewar support on the elevator, then place a Dewar of water at ambient temperature on
the Dewar support.
A Dewar support is not necessary for the larger Dewar used with the Gemini VII 2390t. Place the
Dewar directly on the elevator.
4. Click Browse to select an existing file or to create a new one. Click Open.
5. Click Edit to make and save modifications or replace all parameters in the open file using the
Replace All button to select another file. Click Save and Close when done.
6. Click Report after analysis to automatically generate reports to the specified destination when
the analysis is complete. On the Report Settings window, select the report destination. Click OK.
7. Click Start to start the analysis. A window displays data as they are collected. A short delay is
encountered before the port status changes from the Idle state at the bottom of the screen. Review
the analysis results to see if they are within the tolerances shown in the following table. If these
results are not achieved, call a Micromeritics Service Representative.
0.1 ±0.008
0.2 ±0.010
0.3 ±0.012
0.4 ±0.014
0.5 ±0.016
0.6 ±0.018
0.7 ±0.020
0.8 ±0.022
0.9 ±0.024
1. Select File > Open > Sample Information and create the new sample file, for example,
CARBON1.SMP.
2. Click the Replace All button and select a reference analysis file, for example: Carbon.smp.
3. Click Open. Information from the reference analysis file populates fields in the new file.
2. On the Analysis screen, click Browse to select the test file created in the previous steps. Click
Open.
3. Click Edit to make and save modifications or replace all parameters in the open file using the
Replace All button to select another file. Click Save and Close when done.
4. Click Report after analysis to automatically generate reports to the specified destination when
the analysis is complete. On the Report Settings window, select the report destination. Click OK.
5. Click the Start button. Graphs are displayed as the analyses are performed.
When the analyses are complete, compare the BET Surface Area shown on the Summary Report
with the BET Surface Area shown on the reference material bottle. The values should match
within the tolerance level shown on the bottle.
If the BET Surface Area values do not match, refer to the following causes and actions.
Cause Action
The sample was not degassed properly. Degas the sample again.
The gas lines are not clean. Perform the procedure for cleaning and verifying
gas lines, then try again.
The measured free space is too high. This Use helium that is 99.999% pure, then try again.
indicates the helium is not pure enough.
To review calibration details of the analyzer, go to Unit [n] > Unit Configuration.
The following calibrations can be performed without the assistance of a service representative:
• Zero pressure
• Match transducers
• Servo valve
Zero Pressure
Use this option to evacuate the system and zero the transducers.
3. Ensure that all transducers are selected, then click Start. The window closes when the operation is
complete.
Match Transducers
Use this option to zero and match the selected transducers to the main manifold transducer.
A sample or Po tube must be installed on each selected port before clicking Start.
Sample Servo
Ensure the pressure transducer has been calibrated before performing this procedure.
Go to Unit [n] > Unit Configuration and view the calibration information. Contact your
service representative if calibration dates are not listed.
Balance Servo
Use this option to calibrate the balance servo valve.
Ensure that empty tubes are installed on the sample and balance ports before
performing this procedure.
Ensure that empty tubes are installed on the sample and balance ports.
The Gemini components and accessories can be ordered using one of the following methods:
060-00023-00 FlowPrep 060, degasses up to six samples at up to 400 ºC with flowing gas
(requires gas regulator).
061-00023-00 VacPrep 061, degasses up to six samples at up to 400 ºC; uses flowing gas or
evacuation by vacuum (evacuation requires a vacuum pump and a regulator)
065-00000-00 SmartPrep 065, Windows interface allowing programmable ramp and soak
rates; degasses up to six samples with flowing gas
236-33005-00 PrepSeal Accessory Kit; for degassing and transferring air-sensitive samples
to an analysis port without contamination
239-33601-00 Dewar, 2.2 liters, stainless steel; includes cover (Gemini 2390a and 2390p)
239-33602-00 Dewar, 4.0 liters, stainless steel; includes cover (for Gemini 2390t)
004-33602-00 Pressure Relief Kit (recommended for both gas supplies); prevents excessive
gas pressure in the event of regulator failure (not to be used with toxic gases)
Reference Material
236-61702-00 Sample Tube, bulb; 3/8 in. (0.95 cm) OD × 6.1 in. (15.5 cm) long, 3/4 in.
(1.91 cm) OD bulb
236-61703-00 Sample Tube, straight-wall; 3/8 in. (0.95 cm) OD × 6.1 in. (15.5 cm) long
238-61703-00 Sample Tube, straight-wall; 3/8 in (0.95 cm) OD × 8.1 (20.5 cm) in. long (for
Gemini 2390t)
062-00000-11 Vacuum Pump with built-in anti-suckback valve, 100/120 VAC, includes hose
kit
062-00000-23 Vacuum Pump with built-in anti-suckback valve, 220/240 VAC, includes hose
kit
062-00003-00 Vacuum Pump, hybrid (oil free); includes vacuum hose, fittings, and manual.
Cables
Miscellaneous
A. FORMS
Use this worksheet to record the values necessary to calculate the sample mass.
Degas Information
_________________________________________________________________________________
_________________________________________________________________________________
_________________________________________________________________________________
_________________________________________________________________________________
_________________________________________________________________________________
Gemini VII Appendix B
B. ERROR MESSAGES
Error messages for the Gemini program are listed numerically. If the Action response instructs you to
contact a service representative, record the error message and make backup copies of any files
involved in the operation.
2400 Series
2430- Error accessing file [file name], error code = [n].
Cause A: Media may be damaged.
Action A: Clean the media drive. If this does not eliminate the problem, attempt
operation using a backup copy of the file.
Action: Copy files not used regularly from the hard disk to an external media,
delete them from the hard disk and try the operation again
2433- New entries have been found in this directory. Refresh the directory
information?
Cause: Several analyzer files (sample information, analysis conditions,
adsorptive properties or report options) have been added to this
directory by some function other than the analyzer program.
Action: Select Yes to update the directory information with data from each
new file. This operation may take a minute.
Select No to locate the file manually. This option may be feasible if a
large number of files have been copied into the directory and the
name of the file is known.
2434- File [file name] — Subset # [n] wrote wrong amount of data.
Cause: An internal processing and/or hardware error has occurred.
Action: Type a valid path name (including the proper extension) and press
Enter.
Action: Enter an existing file specification or select a file name from the list
box.
2448- File directory [path name] is invalid. Resetting to the installation directory.
Cause: A working directory specified in the .INI file is invalid or has been
moved or deleted.
Action: The installation directory will be substituted. The next time a file is
opened, use the Directories list to move to the correct directory.
Action: Refer to an operating system manual and set the limit for open files to
50 or greater.
Action: Wait a short time and attempt to stop the application again.
2459- An instrument is busy. A delay in restarting this application could result in loss
of new data. Continue with program Exit? [Yes, No]
Cause: An attempt was made to exit the application while an analysis was in
progress. While this is possible, the data collected when the
application is inactive will not be permanently recorded until the
application is re-started. A power failure to the instrument could
cause some data to be lost.
Action: If not concerned with the potential for loss of data should a power
failure occur, click Yes to continue; otherwise click No.
Action: Correct the problem as described above then restart the application.
Action: Finish editing the file, save and close it then start the analysis.
Action: Locate the application using the file (in the Micromeritics application,
use the Windows menu item to get a list of all windows, one of which
may contain this file).
2483- An analysis cannot be performed on [file name]. It is open for editing and con-
tains errors.
Cause: An attempt was made to use a sample file containing errors that is
currently open.
Action: Go to the window containing the file, correct the errors and save it.
2484- The edit session for [file name] must be saved before the analysis. Save
changes and continue with the analysis? [Yes, No]
Cause: An attempt was made to start an analysis using a file that contains
unsaved changes and is open for editing.
Action: Select Yes to save the changes and proceed with the analysis. Select
No to cancel the analysis and continue editing the Sample
Information file.
2486- Could not construct [name] report type. Program will terminate.
2487- Could not start report generator. Error code [n]. Program will
terminate.
Cause A: Full rights to the application’s folders and files is required.
2488- File [file name] cannot be opened for editing. It is already in use.
Action: Check the Windows list to locate the other edit session.
2489- File [file name] cannot be opened for writing. It is already in use.
Cause: The specified file in a Save As operation is already open for edit.
Action: Run the analyzer setup program, select Change analyzer setup and
create the control file used by the analyzer.
2491- Highlighted fields contain errors. Please correct the errors before closing dia-
log box.
Cause: The highlighted fields contain invalid entries. The dialog box cannot
be closed until all errors are corrected.
Action: Check the entries, correct the errors and close the window.
2500 Series
2500- All sample file names that can be created using the sequence number pattern
already exist. You may want to modify the next sequence number.
Cause: No more sample information files can be created using the currently
entered file name sequence number.
Action: Select Options > Sample defaults and enter another sequence
number.
2505- Error logger cannot be initialized. Error code [n]. Program will exit.
Cause: An internal processing error has occurred.
2506- Sample file [file name] has a 'No Analysis' status and cannot be used for this
operation.
Cause: The selected sample file does not have collected data and cannot be
used for operations, for example, reporting.
2508- [sample file] Overlay file [name] was not found. It will not be included in the
reports.
Cause: The specified overlay file could not be found.
2509- [sample file] Error opening file [name] [error]. Reports cannot be produced.
Cause: An error occurred while the program was opening a file necessary to
the report operation.
Action: Use the name given in the error message to investigate. Contact a
Micromeritics service representative if this error message continues.
2510- [sample file] Error parsing reports from file [name]. Reports cannot be
produced.
Cause A: One or more data entry fields in the sample file may contain an
invalid character (such as a single quote or double quotes).
Action A: Review the data entry fields (for example, the Sample field) and
remove the invalid character.
Cause B: The system was unable to create the usual temporary files during the
report, possibly due to insufficient disk space.
2511- Print job [name] has been cancelled due to insufficient disk space. Delete
unnecessary files and restart the report.
Cause: The disk drive does not have required space for the temporary file.
Action: Delete unnecessary files from the disk. At least five megabytes of free
space is required for normal operation.
Action: Ensure the media containing the calibration file has no problems.
2515- Warning: Changing the calibration information will affect the performance of
the instrument. Only qualified service personnel should do this. Do you wish to
proceed?
Cause: The process of performing a calibration operation was started.
2517- Canceling this dialog will reset the calibration state to what it was when this
dialog was first opened. Are you sure you want to cancel?
Cause: The calibration has not been accepted.
2532- The instrument contains a different software version. Do you want to reset it?
Cause: The application has discovered a different version of software
operating in the analyzer.
Action: If there are no analyzers other than the Gemini connected to the
computer, select Yes and allow the updated software to load.
2599- The selected file has an extension that is not supported by this operation.
Cause: The selected file does not have a supported file extension.
Action: Open the adsorptive properties file. Open the FPI file selector and
select another file with a supported file extension.
4000 Series
4003- Error Converting Pressures.
4004- Error Computing Volume Adsorbed.
Cause: An internal processing and/or hardware error occurred during report
generation.
4005- Pressures were not smoothed. Not enough pressures below 0.10 P/Po.
Cause: The Smooth pressures below 0.10 P/Po option was selected on the
Report Options window. There must be at least 10 pressures within
this range for smoothing to occur.
Action: Select another file which contains collective data and restart the
report.
Action: Select only files that have been created by the proper program.
4023- The file [file name] cannot be prepared for analysis. It is open for editing and
contains errors.
Cause: An attempt was made to start an analysis using a file that contains
errors and is open for editing.
Action: Finish editing this file, save and close it, then start the analysis.
4027- Fewer than two sample files have data suitable for heat of adsorption reports.
Cause: Less than two of the selected sample files for heat of adsorption
reports contain appropriate data.
Action: Edit the Quantity Adsorbed table or select other sample files.
Action: Access the Dubinin Report Adsorptive options in the sample file and
enter an appropriate value for the analysis gas.
4029- At least two fitted data points are needed for Alpha-S calculations.
Cause: Fewer than two data points fall within the selected Alpha-S range.
Action: Edit the calculation assignments or the fitted Alpha-S range or use a
different reference curve.
Action: This message was preceded by a different error message. Refer to the
cause/action of the preceding message.
4031- Not enough points with a relative pressure in the range [n,n]
Cause: Fewer than two data points selected for the Dubinin report fall within
the selected relative pressure range.
Action: Edit the calculation assignments or the fitted relative pressure range.
4035- Cannot calculate optimized Astakhov exponent: Not enough points with a rel-
ative pressure in the range [(pressure), (pressure)].
Cause: There are fewer than two valid data points in the relative pressure
range specified. Astakhov reports will not be produced.
Action: From the Collected Data window, select the data points to be
analyzed.
4037- Computations failed while processing the primary data set. No reports will be
produced.
Cause: The preparation of data for reporting could not be successfully
completed. No Horvath-Kawazoe reports will be produced. This
message will always be preceded with another one containing
additional information.
Action: Refer to the number of the error message which preceded this one for
an explanation.
4038- Fewer than 2 points available for the Langmuir Qm computation. Cheng/Yang
correction will not be applied.
Cause: The Cheng/Yang correction to the Horvath-Kawazoe equation
requires the value of the monolayer volume as calculated from the
Langmuir equation. The isotherm must include at least two points
above 0.02 relative pressure for the Langmuir equation to be applied.
4039- The isotherm does not meet the constraints of the Cheng/Yang assumption.
Cheng/Yang correction will not be applied.
Cause: The Cheng/Yang correction to the Horvath-Kawazoe equation
requires the value of the monolayer volume as calculated from the
Langmuir equation. The isotherm does not correlate to the Langmuir
assumption with a coefficient of 0.98 or more. The correction is not
applicable to this isotherm or to the range of the data points selected.
Action B: Generate the Langmuir report for the same data points selected for the
Horvath-Kawazoe report. If the Langmuir correlation coefficient can
be brought above 0.98 by removing some points at high relative
pressure, remove them and reproduce the Horvath-Kawazoe reports.
4040- The value of Qm computed from the Langmuir equation is too low. The pore
size will not be computed for all data points.
Cause: The Cheng/Yang correction to the Horvath-Kawazoe equation
requires the value of the monolayer volume as calculated from the
Langmuir equation. The computed value is less than the volume
adsorbed at the largest relative pressure included in the analysis. The
correction is not applicable to this isotherm or to the range of the data
points selected.
Action: The analysis will be performed and the Cheng/Yang correction will
be applied to all points with a volume adsorbed less than the value of
Vm. The pore size will not be calculated for data points with an
invalid volume adsorbed. Deselect Apply Cheng/Yang correction on
the Horvath-Kawazoe Report Options window to clear this message.
Action: Change the data points selected for the Horvath-Kawazoe report to
include only relative pressures at or below the knee of the isotherm or
change the Horvath-Kawazoe report options so that the Cheng/Yang
correction is not applied.
4042- 0.0 cannot be a starting or ending pressure for a geometric progression from
low pressure.
Cause: An attempt was made to generate a pressure table from a
geometrically progressing range.
4043- 1.0 cannot be a starting or ending pressure for a geometric progression toward
saturation.
Cause: An attempt was made to generate a pressure table from a
geometrically progressing range.
4044- Points in the Langmuir report pressure table lie outside the collected data.
Cause: Calculation assignments are not being used and more than one of the
report pressure table points is above the range of the collected data
and more than one is below.
Action: Change the report pressure table to be more consistent with the
collected data.
4045- Points in the report pressure table lie outside the collected data.
Cause: Calculation assignments are not being used and more than one of the
report pressure table points is above the range of the collected data
and more than one is below.
Action: Change the report pressure table to be more consistent with the
collected data.
Action: If the problem persists, restart the computer and optionally perform a
media integrity check (using ScanDisk).
Action: If the problem persists, restart the computer and optionally perform a
media integrity check (using ScanDisk).
4048- Warning: An error occurred while restoring the heat of adsorption report editor.
Cause: The state of the heat of adsorption report editor could not be restored.
Default settings will be used.
Action: No action.
4049- The sample [file name] does not have enough data. A minimum of two adsorp-
tion points is required.
Cause: A sample file has been included in the Heat of Adsorption report that
does not have enough data.
4050- None of the requested quantities adsorbed is within the range of the collected
data of more than one sample file.
Cause: The Heat of Adsorption report failed because the specified quantities
are not within the range of the collected data.
Action: Edit the quantities adsorbed so that they are within the range of the
collected data or select other sample files.
4051- The sample [file name] does not have any data in the range of the requested
quantities adsorbed.
Cause: The sample’s data cannot be interpolated to any of the quantities
adsorbed.
Action: Edit the quantities adsorbed so that they are within the range of the
collected data.
4052- Fewer than two points are selected for this report.
Cause: At least two points are required for the BET calculations.
4053- At least two data points must be selected for t-Plot calculations.
Cause: At least two points are required for the t-Plot calculations.
4054- Fewer than two data points are inside the fitted thickness range.
Cause: At least two points must be within the fitted thickness range for the t-
Plot calculations.
Action B: Edit the fitted thickness range in the t-Plot report editor.
4056- A positive Langmuir surface area was not calculated. Please check your Lang-
muir report.
Cause: Fewer than two points were assigned to the requested surface area
calculation in the collected data table.
4057- At least two data points are needed for Freundlich calculations.
Cause: Less than two data points have been selected for the Freundlich
report; at least two are required.
4057- At least two data points are needed for Freundlich calculations.
Cause: Less than two data points have been selected for the Freundlich
report; at least two are required.
4063- The data requested on this report are not available. No subreports selected.
Cause: There is no information in the sample log to report.
Action: A sample file was selected which no instrument operations has been
used. Select a sample file with a status of Prepared, Preparing,
Analyzing or Complete to obtain a valid sample log report.
4067- No data points are within the range of pressures in the reference isotherm.
Cause: There are no collected data points within the range of pressures in the
reference isotherm.
Action: Select data points that are in the range of the reference isotherm or
select a more appropriate reference isotherm.
Action: Edit the collected data window and select points for the f-Ratio
report.
Action: Exit the program and reinstall the software, then try again.
4071- The selected pressures points do not form a valid set for deconvolution.
Cause: The data points selected for analysis do not contain enough
information to allow a DFT data reduction.
Action A: Edit data points in the table of the Collected Data window or select
another sample file. At least two points with strictly increasing
pressures and volumes adsorbed are required for a DFT Plus data
reduction.
Action B: Edit the pressure range on the DFT report options window.
4072- The range of pressures selected is too small to deconvolute using this model.
Cause: A null result was found using the selected model.
Action: Edit data points in the table of the Collected Data window or select
another sample file. At least two points with strictly increasing
pressures and volumes adsorbed are required for a DFT data
reduction.
4073- The analysis gas [name] does not match the model gas [name].
Cause: The model assumes a specific gas and the sample file uses a different
one.
4074- The analysis temperature [nn] does not match the model temperature [nn].
Cause: The temperature for the selected model did not match the analysis
temperature.
4075- The models cannot be located in the models folder. Reinstall the software.
Cause: The models could not be located. They may have been inadvertently
deleted or moved.
Cause: The Isotherm report for the named overlay file has Per gram selected
for the Volume Adsorbed and the Isotherm report for the primary file
has a surface area option selected for the Volume Adsorbed.
Action A: Edit the Isotherm report for the named overlay file and select a
surface area option for Volume Adsorbed.
Action B: Click Overlays on the Report options window of the primary file and
remove the named overlay file from the list.
4078- Slope and Y-Intercept cannot be determined from the selected points.
Cause: The Langmuir report cannot be generated from the selected points.
4079- Points found with negative pressure values will not appear on the graph.
Cause: Collected data contains negative pressure values.
Cause: The sample file in the Heat of Adsorption report list does not exist.
Action: Go to Report > Heat of Adsorption. Click Add Samples, then select
the sample file.
Action: Go to Report > Heat of Adsorption. Select the correct sample file,
then click Remove Sample. Rerun the Heat of Adsorption report.
4400- The computer does not have the communications port specified for the Smart-
Prep. Cannot initialize.
Cause: The communications port associated with this unit was not valid.
Action: Run the setup program and set up the unit on a valid port.
4401- The communications port specified for the SmartPrep is already in use. Cannot
initialize.
Cause: The communications port associated with this instrument is in use by
some other program in the system.
Action: Close the other program to release the port. Restart the analysis
application.
4402- The communications port specified for the SmartPrep cannot be accessed.
Cannot initialize.
4403- Cannot communicate with SmartPrep Unit [n] - S/N: [nn].
Cause: The communications port associated with this unit was not valid.
Action: Run the setup program and set up the unit on a valid port.
4404- The application version of the SmartPrep Unit [n] - S/N: [nn] is invalid.
Cause: The controller software running on the designated instrument is
invalid.
Action: Use the SmartPrep setup program to download the proper controller
software to the instrument, or if unavailable, contact a Micromeritics
service representative.
4405- Fatal communications error with SmartPrep Unit [n] - S/N: [nn].
Cause: There was a fatal error in the serial communications between the
application and the SmartPrep Instrument Controller. All displays for
that SmartPrep will be closed.
Action: Ensure that the SmartPrep is properly chained to the computer on the
communications port configured in the Setup program. Stop and re-
start the application. Contact a Micromeritics service representative if
this error message continues.
6000 Series
6000- An error occurred while loading the application control information. Data entry
cannot be performed. [Code number]
Cause: An error occurred accessing the control information disk file required
by this application.
Action: The disk drive may have failed or be corrupt. Run diagnostics on the
disk drive.
6008- The sample has an invalid status and cannot be used for this analysis.
Cause: A sample file with a status other than No Analysis or Prepared was
selected.
6023- Desorption points in this sample file cannot be processed by this instrument.
Continue with adsorption points?
Cause: The selected sample file in this analysis has desorption points
selected in the pressure table. The analyzer (Gemini 2390a or 2390p)
cannot report desorption data.
Action A: Click Yes to continue with the analysis and report adsorption data
only.
Click No to choose a different file, or to edit the current file.
Action B: If using a Gemini 2390t, select Start Analysis for that unit.
Cause: The requested report cannot be generated with the Gemini 2390a.
Cause: There was a fatal error in communication between the analyzer and
the software. All displays for that instrument will be closed.
Action: Ensure the Ethernet cable is securely connected. Exit the application
and then restart it. Contact a Micromeritics service representative if
this error message continues.
6026- Measure Po for each point is not supported by this model. Choose another
sample file for analysis or change the Po option in the analysis conditions.
Cause: Measure Po for each point was selected as the saturation method.
This option is only available for the Gemini 2390p or 2390t.
Action: Edit the Po options in the analysis conditions and select Entered or
Most recent measured as the saturation pressure method.
Cause: The Dewar does not contain enough cryogen to take an accurate
reading in the Po tube.
Action: Refill the Dewar with cryogen and restart the analysis.
Action A: Cancel the analysis, open the sample file, and choose Entered or
Most recent measured for the Po value. Restart the analysis.
Action C: If a Gemini 2390p or 2390t is attached, select Start Analysis for that
unit.
Action: Select Unit > Calibration > Load from File and select a file
containing calibration data. Contact a Micromeritics service
representative if this error message continues.
Cause: Upon initialization, the analyzer was unable to locate the calibration
file.
Action: Select Unit > Calibration > Load from File and select a file
containing calibration data. Contact a Micromeritics service
representative if this error message continues.
Action: Select Unit > Calibration > Load from File and select a file
containing calibration data. Contact a Micromeritics service
representative if this error message continues.
Cause: A test of the sample pressure transducer showed a low reading. The
current operation will continue.
Cause A: Adsorbate gas pressure is too high. The current operation will
continue.
Action A: Verify that the gas inlet pressure regulator setting does not exceed 15
to 18 psi (103.4 to 124.1 kPa). Contact a Micromeritics service
representative if this error message continues.
Cause A: The gas pressure in the Po tube is too high. The current operation will
continue.
Action A: Verify that the gas inlet pressure regulator setting does not exceed 15
to 18 psi (103.4 to 124.1 kPa). Contact a Micromeritics service
representative if this error message continues.
Cause A: The rate of adsorption by the sample exceeded the response capacity
of the analyzer, preventing normal refilling of the gas reservoirs.
Cause: The analyzer failed to reach a suitable vacuum level within the
specified evacuation time.
Action A: Remove and reattach the sample and balance tubes, ensuring that both
are securely in the port.
Action C: If actions A and B don’t correct the problem, degas the sample again
and repeat the analysis.
Cause A: The adsorbate gas supply was depleted by the analyzer during
analysis or termination.
Action A: Verify that the inlet valve on the gas bottle is open and that the gas
pressure regulator setting is between 15 and 18 psig (103.4 and 124.1
kPa). Check the adsorbate gas bottle to see if it is empty; replace if
necessary.
Action B: If additional data points are required, repeat the analysis with a
smaller amount of sample (to avoid exceeding 100 cm3 STP total
volume adsorbed).
Cause: The elevator was unable to reach the raised or lowered position.
Cause: The adsorbate gas supply was depleted during and analysis or a Po
measurement.
Action: Verify that the inlet valve on the gas bottle is open and that the gas
pressure regulator setting is between 15 and 18 psig (103.4 and 124.1
kPa). Check the adsorbate gas bottle to see if it is empty; replace if
necessary.
Action A: Verify the preparatory steps for the calibration were performed
properly. For example, the sample tube may not have been empty.
If all preparatory steps were followed and no other message was
displayed, contact a Micromeritics service representative.
Cause B: One of the steps in the calibration failed. An additional error message
is displayed.
Action B: Refer to the Cause/Action for the error message that displayed with
this one.
C. CALCULATIONS
Saturation Pressure
Saturation pressure (Po) is selected on the Po and Temperature Options dialog. It may be entered or
measured in the Po tube or on port 3.
Po 2 – Po 1 t I – to 1
Po I = Po 1 + ----------------------------------------------------------
to 2 – to 1
• For points taken after the previous Po, use the previous Po:
PoI = Po1
PoI = PoE
P1
P rel = --------
1 Po I
where
Free Space
Measured
Measured free-space volumes are calculated using the following equations:
V MAN P1
V FW = -------------- ------ – 1 T STD
T MAN P2
P2
V FC = V FW ------
P 3
V FC – V FW
V BATH = --------------------------
T BATH
1 – ---------------
T AMB
where:
Calculated
The calculated free space is determined by subtracting the gas capacity of the volume occupied by the
sample from the measured free space of the empty tube.
T STD
V FW = V WB + V – VS -------------
T AMB
T STD
V FC = V CB + V – V S ----------------
T BATH
where
Equilibration
Equilibration is reached when the pressure change per equilibration time interval (first derivative) is
less than 0.01% of the average pressure during the interval. Both the first derivative and average pres-
sure are calculated using the Savitzky-Golay1 convolution method for polynomial functions. The
equations below are those used to compute weighted average and first derivative, respectively, for the
6th point of an 11-point window.
– 36 P 11 + P 1 + 9 P 10 + P 2 + 44 P 9 + P 3 + 69 P 8 + P 4 + 84 P 7 + P 5 + 89 P 6
P AVG = -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
429
5 P 11 – P 1 + 4 P 10 – P 2 + 3 P 9 – P 3 + 2 P 8 – P 4 + P 7 – P 5
P CHG = ----------------------------------------------------------------------------------------------------------------------------------------------------------------
110
P CHG
P PCPI = 100% --------------
P AVG
where the numerical constants are from the Savitzky-Golay convolution arrays, and
If a non-zero value that is too small is entered for the maximum equilibration time, the
points are collected before equilibration is reached.
If PAVG is greater than 0.995 times the current Po, equilibration will not take place until
the Minimum equilibration delay for P/Po 0.995 has expired, in addition to the standard
equilibration criteria.
2
P SD MD 3
Y = ------------------------------------ 10
2.33 T
1+GY
= -----------------------
1+HY
1
F = ------------------------------------------------------
2
Y + Y +
T ABT 0.5
PC = 1 – F – F ------------ P
T RM
where:
P rel
B 1 = ------------------------------------------------
1
-
1.0 – P rel N ads
1 1
A least-squares fit is performed on the (PrelI ,BI) designated pairs where PrelI is the independent vari-
able and BI is the dependent variable. The following are calculated:
Using the results of the above calculations, the following can be calculated:
23
CSA 6.023 10
SABET = ------------------------------------------------------------------------------------------------------------------------
3 18 2 2
22414 cm STP 10 nm /m S + Y INT
where
CSA = analysis gas molecular cross-sectional area (nm2), user-entered on the Adsorp-
tive Properties dialog
BET C value:
S + Y INT
C = --------------------
Y INT
1 1
V M = --------------------- = --------------------
C Y INT S + Y INT
P rel
L I = ------------I
N ads
I
A least-squares fit is performed on the (PrelI, LI) designated pairs where PrelI is the independent vari-
able and LI is the dependent variable. The following are calculated:
Using the results of the above calculations, the following can be calculated:
23
CSA 6.023 10
SA LAN = -------------------------------------------------------------------------------------------------
3 18 2 2
-
22414 cm STP 10 nm /m S
where
CSA = analysis gas molecular cross-sectional area (nm2), user-entered on the Adsorp-
tive Properties dialog
1
V M = ---
S
Langmuir b Value:
b = Y INT V M
SA LAN S ERR
LAN ERR = -----------------------------------
S
t-Plot
For each point designated for t-Plot4 calculations, the following calculations are done:
HP2 HP3
t I = HP1 --------------------- (Halsey5)
ln P rel
I
or
HJP1 HJP3
t I = -------------------------------------------- (Harkins and Jura6)
HJP2 – log P rel
I
or
2
t I = STSA1 P rel + STSA2 P rel + STSA3
1 1
where:
A least-squares analysis fit is performed on the (tI, NadsI) data pairs where tI is the independent variable
and NadsI is the dependent variable. Only the values of tI between tMIN and tMAX, the minimum and
maximum thickness, are used. The following are calculated:
Using the results of the above calculations, the following can be calculated:
3 10 3 3
S cm /g – A STP 10 A/m D cm liquid/cm STP -
SA EXT = ---------------------------------------------------------------------------------------------------------------------------------------------------------
6 3 3
F 10 cm /m
where
F = surface area correction factor, user-entered on the t-Plot Report Options dialog
D = density conversion factor, user-entered on the Adsorptive Properties dialog
SA P = SA TOT + SA EXT
where SATOT is the BET surface area if the user enabled the BET report exclusively, or Langmuir sur-
face area if the user enabled the Langmuir report exclusively. If neither report has been selected, SATOT
is the BET surface area value calculated using a set of default parameters.
3 3 3
V P = Y INT cm /g STP D cm liquid/cm STP
The data used in these calculations must be in order of strictly decreasing numerical value. Points
which do not meet this criterion are omitted. The remaining data set is composed of relative pressure
(Pr), quantity adsorbed (Va) pairs from (Pr1, Va1) to (PrN, VaN) where (PrN = 0, VaN = 0) is assumed as
a final point. Each data pair represents an interval boundary (or de-sorption step boundary) for inter-
vals i=1 to i=N-1 where N = total number of (Pr, Va) pairs.
Generally, the desorption branch of an isotherm is used to relate the amount of adsorbate lost in a
desorption step to the average size of pores emptied in the step. A pore loses its condensed liquid
adsorbate, known as the core of the pore, at a particular relative pressure related to the core radius by
the Kelvin7 equation. After the core has evaporated, a layer of adsorbate remains on the wall of the
pore. The thickness of this layer is calculated for a particular relative pressure from the thickness equa-
tion. This layer becomes thinner with successive decreases in pressure, so that the measured quantity
of gas desorbed in a step is composed of a quantity equivalent to the liquid cores evaporated in that
step plus the quantity desorbed from the pore walls of pores whose cores have been evaporated in that
and previous steps. Barrett, Joyner, and Halenda8 developed the method (known as the BJH method)
which incorporates these ideas. The algorithm used is an implementation of the BJH method.
Explanation of Terms
A pore filled with condensed liquid nitrogen has three zones:
a.) The core - evaporates all at once when the critical pressure for that radius is reached; the
relationship between the core radius and the critical pressure is defined by the Kelvin
equation.
b.) The adsorbed layer - composed of adsorbed gas that is stripped off a bit at a time with each
pressure step; the relationship between the thickness of the layer and the relative pressure is
defined by the thickness equation.
c.) The walls of the cylindrical pore - the diameter of the empty pore is required to determine
the pore volume and pore area. End area is neglected.
Pore
Wall
Insignificant
compared to
core length
Calculations
The volumes adsorbed (Va) are converted to the liquid equivalent volumes (V1, cm3/g):
VII = Va I D
where D is the Density Conversion Factor from the Adsorptive Properties dialog.
The relative pressure (Pr1) is assumed to be close to unity so that substantially all the pores in the sam-
ple are filled.
The corresponding Kelvin core radius is calculated. Only pores smaller than this size will be included:
–A
RcI = -----------------------------------------
1 + F ln Pr I
where
This radius will be adjusted for the thickness of the adsorbed layer during subsequent calculation steps.
The thickness of the remaining adsorbed layer at this relative pressure is calculated:
HP2 HP3
Tw1 = HP1 -------------------
ln Pr 1
where
HP1, HP2, and HP3 are Halsey Parameters 1, 2, and 3 (respectively) from the Halsey Thickness Equa-
tion dialog.
These calculations illustrate the use of the Halsey thickness equation. If the Harkins/Jura equation was
selected, substitute the following wherever the thickness equation appears:
HJ1 HJ3
Tw 1 = --------------------------------------
HJ2 – log Pr 1
where
HJ1, HJ2, and HJ3 are Harkins-Jura Parameters 1, 2, and 3 (respectively) from the Harkins-Jura Thick-
ness Equation dialog.
The following calculations (a-c) are made for each relative pressure interval based on the increment of
volume desorbed during that interval. The variable I refers to the interval number, that is I=1 for the
first interval from Pr1 to Pr2, and so on. J refers to each previous interval during which new pores were
found. K refers to the total number of intervals in which new pores have been found. K is also the num-
ber of lines reported on the BJH table for collected data.
a.) The thickness of the adsorbed layer at the end of the interval is calculated as follows:
HP2 HP3
Tw I + 1 = HP1 -------------------------
ln Pr I + 1
(For the last pressure interval from the lowest PrI to zero relative pressure, TwI+1 = 0.)
For the first pressure interval, there are no previously opened pores so the volume desorbed
from walls of previously opened pores is zero (Vd1 = 0), and the remainder of Step (a) is
skipped.
The change in thickness of the wall layer due to desorption from previously opened pores is
calculated:
Tw = Tw 1 – Tw I + 1
The annular cross-sectional area of the wall layer desorbed is calculated for all previously
opened pores:
2 2 – 16 2 2
CSA J = Rc j + Tw – Rc j 10 cm /A
The total volume of gas desorbed from walls of previously opened pores is calculated:
b.) The physical processes occurring for this pressure interval are determined as follows:
(1.) If VdI is greater than the current increment of volume desorbed (VlI - VlI+1),
desorption from walls only is occurring. Total surface of walls exposed thus far (cm2/
g) is calculated as follows:
–8
SA W = LPJ Davg 10 J
cm/A for all previously opened pores
where
A new layer thickness (Tw) that will not overcompensate for the actual volume
desorbed in this interval is calculated:
8
V1 I – V1 I + 1 10 A/cm
Tw = -----------------------------------------------------------------
SAw I
Since no cores are evaporated in this pressure interval, no new pores are revealed.
Thus no ending Kelvin radius and average pore diameter are calculated for this inter-
val. Note that this means the report may have fewer tabulated intervals on the
collected data report than experimental pressure intervals.
(2.) If VdI is less than the volume increment desorbed during this interval Vl 1 – Vl l + 1 , the
remaining volume is due to new pores with core evaporation taking place in this inter-
val. K, the number of intervals with new pores exposed, is increased by 1. (For the
interval from the lowest Pr1 to zero relative pressure, no new pore volume is calculated
and the rest of Step b is skipped.)
The volume desorbed from newly opened pores in this interval is calculated as
follows:
Vc I = VI I – VI I + 1 – Vd I
The Kelvin radius for the end of the interval is calculated as follows:
–A
RcK + 1 = -------------------------------------------------
1 + F ln Pr I + 1
All new pores opened in this interval are represented by one pore having a length-
weighted average pore diameter and a corresponding length sufficient to account for
the required volume of adsorbate. The weighted average pore diameter is calculated as
follows:
2 Rc K + Rc K + 1 Rc K Rc K + 1
D avg = -----------------------------------------------------------------------------------
2 2
-
K
Rc K + Rc K + 1
DavgK is the diameter of a pore which would have a surface area that is the average of
the areas for pores radius RcK and RcK+1, if its length was the mean of the lengths at
those radii.
–1 – 2A
------------------------------------
-
P avg = ln
K 1 + F D avg
K
HP2 HP3
Tw avg = HP1 ------------------------
K ln P avg
K
The decrease in thickness of the wall layer by desorption from the walls of new pores
during the lower portion of the pressure interval is calculated as follows:
Td = Tw avg – TW I + 1
K
Davg K 2
– 16 2 2
CSAc K = ----------------- + Td 10 cm /A
2
Vc I
LP K = -----------------
CSAc K
c.) Pore diameters and radii are adjusted for the change in thickness of the adsorbed wall layer
during this interval. If new pores were opened during this interval, the average diameter is
adjusted by the change in layer thickness during the second portion of the desorption interval
as follows:
The layer thickness change during the whole interval is added to diameters of previously
opened pores as follows:
D avg = D avg + 2 Tw
Jnew Jold
The layer thickness change desorbed during this interval also is added to the radii corre-
sponding to the ends of the pressure intervals as follows:
After the above calculations have been performed, the diameters corresponding to the ends of
the intervals are calculated as follows:
Dp J = 2 Rc J
The remaining calculations are based on DP(I), Davg(I), and LP(I). These calculations are only
done for Davg(I) values that fall between the Minimum BJH diameter and the Maximum BJH
diameter specified by the operator on the BJH Adsorption Report Options dialog or the BJH
Desorption Report Options dialog.
D avg 2 16 2 2
Vp I = Lp I -------------I 10 cm /A
2
VP CUM
1
= VpJ for J 1
–2 – 10
SA I = LP I 10 m/cm D avg 10 m/A
I
SA CUM
10
= SAJ for J 1
VP I
dV/dD I = --------------------------------
Dp I – Dp I + 1
Dp I
dDv/d log D I = VP I / log ----------------
Dp I + 1
SA I
dA/dD I = --------------------------------
Dp I – Dp I + 1
Dp I
dA/d log D I = SA I / log ----------------
Dp I + 1
For fixed pore size tables (if selected), the following calculations are performed:
Dp F + Dp F
DF avg = -----------------------------------
J J+1
-
J 2
For the intervals with between the Minimum BJH diameter and the Maximum BJH
diameter.
where INTERP(x) is the value interpolated from the function X = DpJ+1 and Y =
VPCUMJ, using an AKIMA semi-spline interpolation.
where VpFCUM0 = 0.
where INTERP(x) is the value interpolated from the function X = DpJ+1 and Y =
SACUMJ.
where SAFCUM0 = 0.
where INTERP(x) is the value interpolated from the function X = DavgJ and Y = dV/
dDJ.
where INTERP(x) is the value interpolated from the function X = DavgJ and Y = dV/
dlog(D)J.
where INTERP(x) is the value interpolated from the function X = DavgJ and Y = dA/
dDJ.
where INTERP(x) is the value interpolated from the function X = DavgJ and Y = dA/
dlog(D)J.
Compendium of Variables
Va = volume adsorbed expressed as a volume (cm3/g STP)
Vl = liquid equivalent volume of volume adsorbed (cm3/g)
D = density conversion factor from the Adsorptive Properties dialog cm3/cm3 STP)
Pr = relative pressure
Dp = pore (or core) diameter (A)
Rc = Kelvin radius (A) of core
A = adsorbate property factor; from the BJH Adsorptive Options dialog
F = fraction of pores open at both ends; from the BJH Adsorption Report Options
dialog or the BJH Desorption Report Options dialog
Tw = thickness of adsorbed layer desorbed during interval (A)
Tw = thickness of remaining adsorbed wall (A)
HP1, HP2, and HP3 are Halsey Parameters from the Halsey Thickness Equation dialog.
HJ1, HJ2, and HJ3 are Harkins and Jura Parameters from the Harkins/Jura Thickness Equation dialog.
STSA1, STSA2, and STSA3 are STSA parameters from the STSA Thickness Equation dialog.
Horvath-Kawazoe
A relative pressure lower limit is determined such that L-d0 never equals zero. All pressure points less
than this limit are discarded. For each collected relative pressure point, values of L are chosen in an
iterative manner, and the relative pressure (P/Po) determined by solving one of the following
equations:
32 4 4 4 10 4 10
K- IP 10 JA /J cm
P- = ------
ln ------ ---------------------------------------------------
4
------------------------------3- – ------------------------------9- – ---------------
3
- + ----------------
9
Po RT L – 2 d0 3 L – d0 9 L – d0 3 d0 9 d0
where
where:
ZS = sample equilibrium diameter at zero interaction energy (Å) from the Horvath-
Kawazoe Physical Properties dialog
ZA = zero interaction energy diameter from the Horvath-Kawazoe Physical
Properties dialog
DA + DS A
d0 = -------------------------------
2
where:
where
– 4.5 – k- 2
k = ------------------- k – 1 , 0 = 1.0
k
2
k = –-------------------
1.5 – k-
, = 1.0
k k–1 0
* * 3 32 4 4
6 N 1 12 + N 2 22 L 10 J A /J cm
ln ------- = ---------------------------------------------------------------------------------------------------------
P -
Po 3
RT L – d 0
d0 6 1 1 d 0 12 1 1
– ----- ------ T 1 + --- T 2 + ----- ------ T 3 + ------ T 4
L 12 8 L 90 80
where
where:
1 - -------------------
1
T1 = ------------------
3
– 3
1 – S 1 + S
1 1
T2 = -------------------
2
– ------------------2-
1 + S 1 – S
1 - -------------------
1
T3 = ------------------
9
– 9
1 – S 1 + S
1 1
T4 = -------------------
8
– ------------------8-
1 + S 1 – S
L – d0
where S = --------------
L
Cheng/Yang Correction
This factor corrects for the nonlinearity of the isotherm. It adds an additional term to the equations for
the different geometrics:
1 1
ln ------- = G L – 1 – --- ln ------------
P
Po 1–
where
Interaction Parameter
The interaction parameter (IP) results from the following calculations:
2
3 MC A A
A A = -----------------------------------------------
2
where
IP = N A A A + N S A S
where:
Refer to Interaction Parameter Components later in this Appendix for recommended values.
Additional Calculations
Based on the previous calculations, the following can be calculated:
AL I = L I – DS
V CUM = V I D
I
where
V CUM
V HALF = ----------------N-
2
log D G – log D L
D MED = 10 log D L + log V HALF – log V L -----------------------------------------------
log V G – log V L
where
VCUMN = total cumulative pore volume (VCUMI) for points designated for Horvath-
Kawazoe calculations
VHALF = 50% of total cumulative pore volume
VL = cumulative pore volume (VCUMI) for first point less than VHALF
VG = cumulative pore volume (VCUMI) for first point greater than VHALF
DL = pore width (LI) that corresponds to VL
DG = pore width (LI) that corresponds to VG
Interaction
Gas Bath Sample Type Parameter
Temperature Calculated
(K) Value*
* The interaction parameter is entered in the Horvath-Kawazoe Report Options dialog in the
following field:
Interaction parameter: (calculated value) x 10-43 ergs-cm4
The following values were used to calculate the values in Table C-1.
Carbon-Graphite Zeolite
DS= 3.40 DS= 3.04
Nitrogen Argon
DA= 3.00 DA= 2.95
Carbon Dioxide
DA= 3.23
Aluminophosphate Aluminosilicate
DS= 2.60 DS= 2.76
Dubinin-Radushkevich
The Dubinin-Radushkevich13 equation is as follows:
2 2
BT
log V = log Vo – --------------- log Po
-------
P
where:
For each point designated for Dubinin-Radushkevich calculations, the following calculations are done:
LV = log (V)
2
LP = log -------
Po
P
The intercept, log(Vo) can be found by performing a least-squares fit on the (LP,LV) designated pairs
where LP is the independent variable and LV is the dependent variable. Assuming the adsorption of
gas is restricted to a monolayer, Vo is the monolayer capacity. Based on this assumption, the following
are calculated:
Using the results of the above calculations, the following can be calculated:
Vo = 10YI
23
Vo 6.023 10
SDP = ----------------------------------------------------------------------
3 18 2 2
22414 cm 10 nm /m
where
= molecular cross sectional area of gas (nm2) from the Adsorptive Properties
dialog
Dubinin-Astakhov
The Dubinin-Astakhov equation is a follows:
RT N N
log V = log Vo – --------- log Po
-------
E 0 P
where
For each point designated for Dubinin-Astakhov calculations, the following calculations are done:
LV = log(V)
N
LP = log Po
-------
P
A least-squares fit is performed on the (LP,LV) designated pairs where LP is the independent variable
and LV is the dependent variable. If the user selected Yes for the Optimize Astakhov Exponent
prompt, a systematic search for the optimum value of N is conducted by recalculating the linear regres-
sion and selecting the value of N that gives the smallest standard error of the y-intercept. The exponent
N is optimized to within 10-4. If the optimum value for N is not found in this range, an exponent of 2 is
used. The following are calculated:
Using the results of the above calculations, the following can be calculated:
V0 = 10YI
WI = (VI x D)
where
D = density conversion factor (cm3 liquid/cm3 STP) from the Adsorptive proper-
ties dialog
W0 = (V0 x D)
where
D = density conversion factor (cm3 liquid/cm3 STP) from the Adsorptive Proper-
ties dialog
2.303 R T
E = --------------------------------------
1/N
-
2.303 S
Modal Equivalent Pore Diameter (Å):
3 3 3 1/3
3N 1/N 10 nm /A
D MODE = 2 ---------------- -----------------------------
3N + 1 Eo
where
= affinity coefficient of the analysis gas relative to the Po gas from the Dubinin
Adsorptive Options dialog
1/3
3N + 1 1/3N E0 3N + 1
exp – ----------------
dV -
------------------ Max = 0.5 3N + 1 W o ---------------- ----------------------------
-
dD mode 3N 3 3 3 3N
10 nm /A
E0 1/3
3N + 1
SDA = 1000 2.0 W 0 ------ ----------------
k 3N
(x+1) = 1+b1x + b2x2 + b3x3 + b4x4 + b5x5 + b6x6 + b7x7 + b8x8 + (x) | (x) | 3 x 10-7
where
b1 = -0.57719 1652
b2 = 0.98820 5891
b3 = -0.89705 6937
b4 = 0.91820 6857
b5 = -0.75670 4078
b6 = 0.48219 9394
b7 = -0.19352 7818
b8 = 0.03586 8343
and where
3N + 1
x + 1 = ----------------
3N
3 3 3 N D i – 3N + 1 10 3 nm 3 /A 3
N
D i –3N
10 nm /A
----- exp – ----------------------------- -----
dV-
-------- = 0.5 W 0 3N -----------------------------
dD i E0 2 E0 2
MP-Method
For each point designated for MP-method14 calculations, the following calculations are done:
HP2 HP3
t I = HP1 --------------------- (Halsey5)
ln P rel
I
or
HJP1 HJP3
t I = -------------------------------------------- (Harkins and Jura6)
HJP2 – log P rel
I
where:
With the (tI,VI) data pairs, the Akima semi-spline interpolation method is used to interpolate quantity
adsorbed values based on thickness values that are evenly spaced 0.2 Angstrom apart starting at the
first outlier point. Outliers are defined as those points have the maximum instantaneous slope within an
iteratively shrinking subset of all points. The remaining pore surface area calculation result is the slope
of the line defined by two consecutive interpolated points. The slopes of each pair of consecutive
points from the origin to the last point must be monotonically decreasing and non-negative. With the
interpolated points set the following can be calculated:
tI + tI – 1
R I = --------------------
2
where
Incremental pore surface area occluded for the Ith point (m2/g):
S INC = S I – 1 – S I
I
Cumulative pore surface area occluded for the Ith point (m2/g):
dA S INC
-------- = -------------------
I
dR I tI – tI – 1
4 2 2 –8
V INC = S INC 10 cm /m R I 10 cm/Å
I I
dV V INC
-------- = -------------------
I
dR I tI – tI – 1
Freundlich Isotherm
The Freundlich isotherm has the form:
Q 1/m
------- = CP
Qm
where
1/m
Q = Q m CP
1
log Q = log Q m C + ---- log P
m
Temkin Isotherm
The Temkin isotherm has the form,
Q RT
------- = --------- ln A 0 P
Qm q0
where
RTQ S
Q = -------------- ln A 0 + ln -------
P
q0 Po
Thus, the plot of the natural log of absolute pressure vs. quantity adsorbed yields a straight line with
slope RTQm/q0 and intercept (ln A) RTQm/q0.
This energetic heterogeneity greatly affects the shape of the adsorption isotherm with the result that
simple theories such as the Langmuir and BET formulas can, at best, give only approximate estimates
of surface area. Porous solids virtually are never characterized by a single pore dimension, but instead
exhibit a more or less wide distribution of sizes. The observed adsorption isotherm for a typical mate-
rial is therefore the convolution of an adsorption process with the distribution of one or more
properties which affect that process. This was first stated mathematically by Ross and Olivier12 for the
case of surface energy distribution and has become known as the integral equation of adsorption.
where
Equation (1), a Fredholm integral of the first kind, is a member of a class of problems known as ill-
posed, in that there are an infinite number of functional combinations inside the integral that will pro-
vide solutions. Even when the kernel function is known, experimental error in the data can make
solving for even a single distribution function a difficult task. Solving for multiple distribution func-
tions requires more data than provided by a single adsorption isotherm.
Qp = d q p f (2)
where
The exact form of the energy-dependent term depends on the form of the model isotherms expressed in
the kernel function and is provided in the model description.
Qp = dH q p H f H (3)
where
Numerical values for the kernel functions in the form of model isotherms can be derived from modern
statistical mechanics such as density functional theory or molecular simulations, or can be calculated
from one of various classical theories based on the Kelvin equation. Several types are found in the
models library.
Qp = q p Z i f Z i (4)
i
Given a set of model isotherms, q(p,Z), from a model chosen from the models library and an experi-
mental isotherm, Q(p), contained in a sample information file, the software determines the set of
positive values f(Z) that most nearly, in a least squares sense, solves equation (4). The distributed prop-
erty, surface energy or pore size, is then displayed on the Report Options window as a selection of
tables or graphs.
Regularization
DFT allows a selectable regularization (also referred to as smoothing) constraint to be applied during
the deconvolution process to avoid over-fitting in the case of noisy data or ill-fitting models. The
method used is based on co-minimization of the second derivative of the distribution. The relative
weight given to this term is determined by the value of the regularization parameter, which is set on the
DFT Pore Size or Surface Energy dialog and also is shown in the header of reports. The value of the
regularization parameter varies from zero (for no second derivative constraint) to ten (indicating a
weight equal to minimizing the residuals), or even larger. When the distribution and residuals obtained
change little with the value of the regularization parameter, it indicates that the chosen model provides
a good representation of the data. Conversely, a large sensitivity to the regularization parameter might
indicate inadequate data or a poor choice of model to represent the data.
xi
XAve = ---------
N
yi
YAve = ---------
N
x i – XAve y i – YAve
Slope = --------------------------------------------------------------
2
-
x i – XAve
The coefficient of Correlation for this line is also calculated in the usual way. (Refer to Mathematical
Handbook for Scientists and Engineers, by Granino A. Korn and Theresa M. Korn, Copyright 1961,
1968 McGraw Hill, Sec. 18.4.)
2
x i – XAve
x = -------------------------------------
N
2
y i – YAve
y = ------------------------------------
N
x i – XAve y i – YAve
Cov x y = ---------------------------------------------------------------
N
Cov x y
Correlation Coefficient = -----------------------
x y
yi
Mean = ---------
N
2
y – Mean -
StdDev = -----------------------------------
N–1
StdDev
CoefVar = -------------------
Mean
Summary Report
The following calculations and the results of previous calculations (as noted) are used to generate the
summary report:
23
Va 1 – Pr CSA 6.023 10
S 1PT = -------------------------------------------------------------------------------------------------
3 18 2 2
22414 cm STP 10 nm /m
where
Pr = pressure closest to 0.3 of the relative pressure points designated for surface
area calculations.
Va = volume corresponding to Pr
MPSTOT = SCUMI, (see MP-method Calculations) for the last collected data point used in the
MP-method Calculations, and the range of hydraulic pore radii over which the cumulative
surface area was computed.
MPVTOT = VCUMI, (see MP-method calculations) for the last collected data point used in the
MP-method calculations, and the range of hydraulic pore radii over which the cumulative
pore volume was computed.
MPV TOT –6 3 10
MPDAVER = ---------------------- 10 m /cm 10 Å/m
MPS TOT
References
1. Savitzky, A. and Golay, M.J.E., Anal. Chem. 36, 1627 (1964).
2. Brunauer, S.; Emmett, P.H.; and Teller, E., J. Am. Chem. Soc. 60, 309 (1938).
3. Langmuir, I., J. Am. Chem. Soc. 38, 2267 (1916); J. Am. Chem. Soc. 40, 1361 (1918); Phys. Rev 8,
149 (1916).
4. deBoer, J. H., et al, J. Catalysis 3, 32, 38, 44, 268 (1964); J. Catalysis 4, 319, 643, 649 (1965);
Cranston, R. and Inkley, F., Adv. Catalysis 9, 143 (1957).
6. Harkins, W.D. and Jura, G., J. Chem. Phys. 11, 431 (1943).
7. Kelvin, J. (published under the name of Sir William Thomson), Phil. Mag. 42, 448-452 (1871).
8. Barrett, E.P.; Joyner, L.S.; and Halenda, P.P., J. Am. Chem. Soc. 73, 373-380 (1951).
9. Horvath, G. and Kawazoe, K., J. Chem. Eng. Japan 16(6), 470 (1983).
10. Saito, A. and Foley, H. C., AlChE Journal 37(3), 429 (1991).
11. Cheng, Linda S. and Yang, Ralph T., Chemical Engineering Science 49(16), 2599-2609 (1994).
12. Ross and Olivier, J.P., “On Physical Adsorption,” J. Wiley and Sons, New York (1964).
13. Dubinin, M., Carbon 21, 359 (1983); Dubinin, M., Progress in Surface and Membrane Science 9,
1, Academic Press, New York (1975); Dubinin, M. and Astakhov, V., Adv. Chem. Ser. 102, 69
(1971); Lamond, T. and Marsh, H., Carbon 1, 281, 293 (1964); Medek, J., Fuel 56, 131 (1977);
Polanyi, M., Trans. Faraday Soc. 28, 316 (1932); Radushkevich, L., Zh. fiz. Kemi. 33, 2202
(1949); Stoeckli, H., et al, Carbon 27, 125 (1989).
14. Mikhail, R., Brunauer, S. and Bodor, E., J. Colloid and Interface Sci. 24, 45 (1968).
D. ETHERNET CONFIGURATION
This section contains instructions for configuring an Ethernet port in either Microsoft® Windows 7 or
Windows Vista. The instructions do not apply to any other operating system.
The windows in this section are Microsoft operating system windows, which may
be subject to change. The windows are included for reference purposes only.
Methods to access to these windows may vary. Refer to the Windows help system
for assistance, if necessary.
Windows 7
1. Select the Windows icon from the Status bar at the bottom of the screen, then select Control
Panel.
5. Click Properties.
7. Click Use the following IP address then enter the following in the IP address field:
192 168 77 100.
8. The following numbers should display in the Subnet mask field: 255.255.255.0. If not, enter
them. Leave all other fields blank.
Windows Vista
Windows 7 Professional is recommended for the best user experience.
1. Select the Windows icon from the Status bar at the bottom of the screen, then select Control
Panel.
Click here
3. Click Manage network connection from the Tasks list on the left side of the window then click
Local Area Connection for your analyzer.
4. Click Properties.
6. Click Use the following IP address and enter the following in the IP address field:
192 168 77 100.
7. The following numbers should display in the Subnet mask field: 255.255.255.0. If not, enter
them. Leave all other fields blank.
Introduction
The unique balanced measurement method used in the Gemini permits small amounts of surface area
to be measured with nitrogen gas that otherwise would be measurable only with krypton. Low surface
area samples often displace many times more nitrogen than they adsorb, especially if composed of
low-density materials of large particle size. The standard, built-in, helium free-space difference mea-
surement and mathematical compensation routine typically removes the effects of more than 99% of
this sample displacement, but the small amount remaining uncompensated can still be significant.
A technique for adding a compensating volume with negligible surface area into the balance tube has
been developed. It can reduce the size of the initial imbalance to low levels and allow precise measure-
ment of adsorbed gas.
Some users find that a combination of the two methods — using glass beads and filler rods — pro-
duces even better results. This, however, will depend on the type of sample you are analyzing.
Filler rods and glass beads are included in your accessories kit. Either straight-wall or bulb sample
tubes may be used.
2. Load a second sample tube of the same size with glass beads that have a total volume
approximately the same as the sample volume.
w
v = ----
where
b.) Determine the number (n) of glass beads needed to equal the sample volume:
v
n = ------------------------3-
0.014 cm
where
3. Outgas the sample in the sample tube at an appropriate temperature for an appropriate amount of
time.
4. Install the sample tube (containing the outgassed sample) onto the analysis port and the sample
tube (containing the glass beads) onto the balance port.
5. Specify a one-point measurement (P/Po = 0.05 to 0.1) so that the initial free-space measurement
can quickly be determined; then perform the measurement.
6. Using the measured free space absolute value and the following relationship, determine the mass
of glass beads to remove from (or add to) the balance tube to reduce the free-space imbalance:
3 3
free space cm 2.515 g/cm -
---------------------------------------------------------------------------- = mass of glass beads g
3.53
where
Note that the volume of one glass bead is approximately 0.014 cm3. Therefore, if the
measured free space is less than 0.02 cm3, it is unnecessary to correct the free space.
7. Use a beaker of warm water to bring the balance tube to room temperature before removing it
from the balance port of the Gemini to remove (or add) glass beads. This prevents condensation
of moisture from the laboratory atmosphere onto the cold glass beads.
• If the measured free space is negative (–), add the calculated mass of glass beads into the
balance tube.
• If the measured free space is positive (+), remove the calculated mass of glass beads from the
balance tube.
9. Reinstall the balance tube onto the balance port of the Gemini analyzer.
For subsequent samples of the same material, you may simply use the same mass of
sample that was used for the initial sample so that the original bead quantity may be
left undisturbed on the balance port.
11. Close the sample compartment door and start your analysis.
1. Clean your sample tube, balance tube, and filler rods; label your sample and balance tubes.
2. Hold the balance tube horizontally, and carefully slide the longer filler rod into the tube.
Filler Rod Sample Balance Tube
Do not hold the rod vertically and drop the rod into the tube; this may break the tube.
3. Prepare your sample and place it into the sample tube; insert the shorter filler rod into the sample
tube.
If you are using bulbous sample tubes, use the same size filler rod as used in the balance tube.
4. Ensure that the filler rods are equidistant from the top of each tube. If they are not, add or remove
sample until they are the same distance from the top of the tube.
Packing of some powders may restrict gas access to the powder and cause slower
equilibration and/or lower results.
5. Attach the sample tube to the analysis port and the balance tube to the balance port.
1. After determining and loading the amount of glass beads to use, insert a filler rod into the balance
tube.
2. Insert a filler rod into the sample tube. Ensure that the filler rods are equidistant from the top of
each tube. If they are not, add or remove sample until they are the same distance from the top of
the tube.
If you are using bulbous sample tubes, use the same size filler rod as used in the balance tube.
3. Attach the sample tube to the analysis port and the balance tube to the balance port.
F. DFT MODELS
Theories are developed by scientists in an attempt to explain a class of observed behavior. In the exper-
imental physical sciences, theories are often expressed in terms of a model that can be visualized and
described mathematically. Early models of physical adsorption were quite simple, both conceptually
and mathematically, for very practical reasons — hand computations were required. Today we can
explore complex models that describe adsorption systems on the atomic scale of size and sub-picosec-
ond time frame. This is not because scientists are smarter, but because of available tools. The DFT
models are created by classical approaches to adsorption as well as models based on modern statistical
thermodynamics.
Theoretical Background
Traditional adsorption theories attempt to describe experimental adsorption isotherms with an isotherm
equation containing a small number of parameters. At a minimum, these parameters include the extent
of the surface, such as the monolayer capacity (Qm), and the molar intensity of the gas-surface interac-
tion, such as the Langmuir “K” constant or the BET “C” constant. In some equations, additional
parameters take into account the lateral interaction of adsorbed molecules with each other. Other theo-
ries, such as the Dubinin-Astakhov approach, also include parameters for the effect of adsorbent
porosity.
Instead of this classical kinetic or phenomenological approach, we can use a molecular-based statisti-
cal thermodynamic theory that allows us to relate the adsorption isotherm to the microscopic
properties of the system: the fluid-fluid and fluid-solid interaction energy parameters, the pore size, the
pore geometry, and the temperature.
The following example is given so that you may understand how such a theory is constructed.
A clean sample of a solid material containing slit-shaped pores of a single width is placed in an evacu-
ated space. It is kept at a fixed temperature as a known quantity of pure argon gas is admitted into the
space surrounding the sample. The pressure within the space is recorded over time. In this situation,
the pressure falls rapidly from its initial value and gradually approaches a steady reading, called the
equilibrium pressure. The amount adsorbed corresponds to the quantity of gas effectively removed
from the gas phase by the solid surface. A graph that plots amount adsorbed versus equilibrium pres-
sure is called an adsorption isotherm.
Under such conditions, the argon atoms that randomly enter the pore space feel the presence of the
solid surface as the action of an external attractive force (the dispersion forces or Van der Waal’s
forces) and spend more time near the surface. As a result, the space near the surface acquires a greater
average density of argon atoms than regions farther removed.
If the equilibrium distribution of the gas atoms near the surface could be described as a function of
pressure and the molecular properties of the components of the system, then a model could be con-
structed for the adsorption isotherm for the system. Modern physical chemistry provides several ways
to calculate this distribution. All these methods are based on the fundamental thermodynamic law that
such a system adopts a configuration of minimum free energy at equilibrium. Also needed is a descrip-
tion of the pairwise interaction energy between atoms, U(s), commonly given by a Lennard-Jones
potential:
12 6
U s = 4 --- – ---
s s
where
This method can be thought of as a way to determine the chronological record of the movement of
each particle in the system using time steps of 10-14 seconds. Although the mathematics are simple,
the number of calculations required for a system of even a few hundred particles is astronomical and
challenges even the fastest computers.
If the move results in a configuration of lower total energy, then the move is completed and another
particle is randomly selected to be moved.
If the move results in a configuration of higher energy, a probability for that event is calculated, and a
random number between zero and one is generated. If the generated number is smaller than the proba-
bility of the event, then the move is accepted; otherwise, another particle is selected and the process is
repeated. This process continues until the average total energy of the system no longer decreases; at
this point, average configuration data are accumulated to yield the mean density distribution of parti-
cles in the system.
Monte Carlo simulations require considerably less computation time than molecular dynamic simula-
tions and can yield the same results; however, neither method provides a really practical way to
calculate complete isotherms.
The system being modeled consists of a single pore represented by two parallel walls separated by a
distance H. The pore is open and immersed in a single component fluid (adsorptive) at a fixed temper-
ature and pressure. Under such conditions, the fluid responds to the walls and reaches an equilibrium
distribution. In this condition (by the definition of equilibrium), the chemical potential at every point
equals the chemical potential of the bulk fluid. The bulk fluid is a homogenous system of constant den-
sity; its chemical potential1 is determined by the pressure of the system using well-known equations.
The fluid near the walls is not of constant density; its chemical potential is composed of several posi-
tion-dependent contributions that must total at every point to the same value as the chemical potential
of the bulk fluid.
As noted previously, at equilibrium, the whole system has a minimum (Helmholtz) free energy, known
thermodynamically as the grand potential energy (GPE). Density functional theory describes the ther-
modynamic grand potential as a functional of the single-particle density distribution; therefore,
calculating the density profile that minimizes the GPE yields the equilibrium density profile. The cal-
culation method requires the solution of a system of complex integral equations that are implicit
functions of the density vector. Since analytic solutions are not possible, the problem must be solved
using iterative numerical methods. Although calculation using these methods still requires supercom-
puting speed, the calculation of many isotherm pressure points for a wide range of pore sizes is a
feasible task. The complete details of the theory and the mathematics can be found in the papers listed
under References at the end of this appendix.
The following graphs and accompanying text illustrate the results of using density functional theory to
predict the behavior of a model system.
Figure D-1 shows the density profile for argon at a carbon surface as calculated by density functional
theory for a temperature of 87.3 K and a relative pressure of about 0.5.
1.Chemical potential may be thought of as the energy change felt by a probe particle when it is inserted into the
system from a reference point outside the system. It can also be defined as the partial derivative of the grand
potential energy with respect to density (or concentration).
Figure D-1. Density Profile for Argon on Carbon at 87.3 K and a Relative Pressure of 0.5
This figure represents a cross-section of the region near the surface. Note the layerwise distribution of
adsorbate; the first monolayer is sharply defined and a third layer can be distinguished. The area under
the profile curve represents the amount adsorbed per unit area at this pressure. The positions of the
maxima are separated by a distance determined by the size of the adsorptive atom.
Given the density profile, the amount adsorbed at the stated pressure can be easily calculated as the
integral over the profile. Repeating this calculation over a range of pressures yields the adsorption iso-
therm for the model. If the value of H is very large, the isotherm obtained corresponds to that of an
external, or free, surface. If H is smaller, a range of pressures is reached where two minima exist for
the grand potential, showing the presence of two metastable phases having different density distribu-
tions but the same chemical potential. The phase with the lower GPE is the stable one. As the pressure
is increased, a point is reached where the other phase becomes the stable one. This phase transition
reflects condensation of adsorbate in the pore; the pressure at which it occurs is called the critical pore-
filling pressure. This pressure is analogous to the condensation pressure predicted by the Kelvin equa-
tion in the classical model of pore filling.
Figure D-2 shows how the profiles change with pressure for a model pore with H = 40 Å. The insets
show the density profiles for the corresponding points of the isotherm.
Figure D-2. Model Isotherm for Argon at 87.3 K in a 40 Å Slit in a Carbon Substrate
The profiles show the density distribution from one wall to the center of the slit; the other half of the
distribution is a mirror image of the profile shown.
As the pressure is first increased from zero, almost all the adsorbed atoms occupy a position close to
the surface.
• Inset a shows the profile corresponding to point a on the isotherm where the surface is about half
covered.
• At point b, the first layer is so full that it is more favorable for atoms to start a new layer.
• At point c, a third layer is forming. Point c, for this size slit, is the critical pore-filling pressure. In
inset c, the profile shows the density decreasing to near zero (actually the bulk gas density) at 4 or 5
molecular diameters from the surface.
• Inset d shows the profile converging on a density similar to that of bulk liquid argon in the center of
the pore, indicating a phase transition.
Note that the adsorption isotherms for pores larger than the one shown in Figure D-2 is identical up to
point c. The lower branch of the isotherm simply continues to a higher pressure for larger pores. This
trend is illustrated in Figure D-3, where isotherms for some larger size pores are shown. It is clear that
pore size is uniquely characterized by a corresponding critical pore-filling pressure. At large pore
sizes, density functional theory produces results for the critical filling pressures that are in good agree-
ment with those produced by the Kelvin equation.
Figure D-3. Model Isotherms for Some Larger Pore Widths Argon on Carbon at 87.3 K
Figure D-4 shows model isotherms for pores in the micropore size range. Note the logarithmic scale
for pressure.
Figure D-4. Model Isotherms in the Micropore Size Range of Pore Width
Argon on Carbon at 87.3 K
Pores of 4 Å width, barely larger than the argon atom (3.38 Å), fill at pressures below 1 millitorr. Pores
below 15 Å fill before a monolayer is completed on the surface of the larger pores. In the micropore
size range, the pore volume fills more gradually with pressure and the total shape of the isotherm is
important in characterizing the pore size.
Models Included
Geometry: Slit
Substrate: Carbon (graphite)
Category: Porosity
Method: Nitrogen at 77 K; Argon at 87 K
Using the methods of non-local density functional theory, two sets of isotherms have been calculated
to serve as kernel functions for the characterization of porous solids from adsorption data. The model
isotherms are stored in binary format files. These models assume a slit-like pore geometry. The pore
size range from 4.0 to 4000 Å is covered in 91 classes in a geometric progression. The class intervals
are rounded to the nearest 0.02 molecular diameters. A model for the free or external surface is
included to account for unfilled pores. Each of the 92 model isotherms has been calculated at 181 pres-
sure points from near 1x10-6 to near 1.00 relative pressure.
These models are identical to those supplied with the original DOS version of DFT software. Some
slight difference from the DOS results may be noted when they are applied to the same data due to
improvements in the deconvolution algorithm and better regularization of the current software.
Using the modified Tarazona prescription described by Olivier (refer to References, numbers 3 and 4),
model isotherms were calculated for a wide range of adsorptive energies to a relative pressure of 0.6.
The model makes no provision for pore filling in the micropore region. If the sample solid contains
small mesopores, the isotherm data should be truncated (using the Select Data Points window) to a
suitably low relative pressure to avoid trying to fit this region; mesopore filling reports as a large area
of low energy in the calculated distribution of adsorptive potential.
The surface energy is reported in terms of the effective Lennard-Jones interaction parameter, ie., for
the adsorptive/adsorbent pair divided by Boltzmann’s constant. The units are therefore Kelvin.
Geometry Cylinder
Substrate Oxide
Category: Porosity
Method: Nitrogen at 77 K; Argon at 87 K
Model isotherms were calculated using a combination of statistical mechanical calculations and exper-
imental observations for macroporous silicas and MCM-41 mesoporous silicas as well as zeolites. The
pore-filling pressures were determined as a function of the pore size from adsorption isotherms on
MCM-41 materials characterized by X-ray and other techniques. The variation of the pore fluid den-
sity with pressure and pore size has been accounted for by density functional theory calculations. The
N2 model reports pore sizes ranging from 3.8 to 387 Å and the AR model from 3.8 to over 500 Å.
Reference: M. Jaroniec, M. Kruk, J.P. Olivier, and S. Koch, “A New Method for the
Accurate Pore Size Analysis of MCM-41 and Other Silica-Based
Mesoporous Materials,” Proceedings of COPS-V, Heidelberg, Germany
(1999).
Geometry Cylinder
Substrate Crystalline Silicate
Category: Porosity
Method: Nitrogen at 77 K
Model isotherms were calculated using a combination of statistical thermodynamic Non-Local Density
Functional Theory (NLDFT) calculations and experimental isotherms for reference samples of mont-
morillionite. The construction method for the hybrid models was analogous to that described in the
first reference below (Jaroniec et al,1999). The additional references add additional theoretical details
as well as examples of the application of the model to pillared clay catalysts. This model reports pore
widths from 3.8 to 387 Å.
References: Mietec Jaroniec, Michal Kruk, James P. Olivier and Stefan Koch, “A
New Method for the Characterization of Mesoporous Silicas,”
Proceedings of COPS-V, 1999, Studies in Surface Science, Vol 128,
Characterization of porous Solids V , Unger, et al, Eds, Elsevier,
Amsterdam, 2000.
James P. Olivier and Mario L. Occelli, “Surface Area and Microporosity
of a Pillared Interlayered Clay (PILC) from a Hybrid Density Functional
Theory (DFT) Method,” The Journal of Physical Chemistry B; 2001,
105(3), 623-629.
M. L. Occelli, J. P. Olivier, J. A. Perdigon-Melon, and A. Auroux,
“Surface Area, Pore Volume Distribution, and Acidity in Mesoporous
Expanded Clay Catalysts from Hybrid Density Functional Theory (DFT)
and Adsorption Microcalorimetry Methods,” Langmuir 2002, 18, 9816-
9823.9b.
James P. Olivier, “The Importance of Surface Heterogeneity in
Developing Characterization Methods.” 6th International Symposium on
the Characterization of Porous Solids, Studies in Surface Science and
Catalysis 144, Elsevier, 2002.
James P. Olivier and Mario L. Occelli, “Surface Area and Microporosity
of Pillared Rectorite Catalysts from a Hybrid Density Functional Theory
Method,” Microporous and Mesoporous Materials 2003, 57, 291-296.
Geometry Slit
Substrate Carbon
Category: Porosity
Method: Carbon dioxide at 273 K
Model isotherms were calculated using the non-local prescription of Tarazona, employing molecular
parameters derived from the known bulk properties of carbon dioxide.
This model was produced in the same manner as the N2 Modified Density Functional model listed ear-
lier, except applicable to argon adsorbed at 87.3 K.
Geometry Cylinder
Substrate Oxide
Category: Porosity
Method: Nitrogen at 77 K
Model isotherms were calculated using the prescriptions of Tarazona for density dependent weighting
functions and a cylindrical pore geometry. The wall potential used is k = 30 K, typical for a silica or
alumina surface.
This model file is particularly useful for sizing zeolites or zeolite containing materials that have sub-
stantial micropore volume. The reported pore size range is 3.8 to 387 Å.
Geometry Slit
Substrate Carbon
Category: Porosity
Method: Nitrogen at 77 K; Argon at 87 K
Model isotherms were calculated using the prescriptions of Tarazona for density dependent weighting
functions and a slit-like pore geometry. These models are slightly different from N2-DFT and Ar-DFT
models that were calculated using NLDFT with density independent weighting functions.
Model isotherms were calculated using the prescriptions of Tarazona for density dependent weighting
functions assuming 2D model of finite slit pores having a diameter-to-width aspect ratio of 6.
This model is particularly useful for microporous carbon materials. The reported pore size range is
from 3.5 to 250 Å.
Model isotherms were calculated using the same methods and assumptions that were used in the model
above except in this model, the aspect ratio is equal to 12.
These two finite pore models may be used as a research tool in conjunction with independent analyti-
cal techniques such as high-resolution transmission electron microscopy (HRTEM) and/or X-ray
diffraction (XRD) to obtain comprehensive information about the structure of studied carbon material.
Geometry Cylinder
Substrate Carbon
Category: Porosity
Method: Nitrogen at 77 K; Argon at 87 K
Model isotherms were calculated using the prescriptions of Tarazona for density dependent weighting
functions and cylindrical pore geometry. The pore wall potential is described by the Lennard-Jones
potential of interaction between a gas molecule and the graphitic surface of an infinitely long cylinder.
This model is particularly useful for characterizing carbon single-wall nanotubes. The reported pore
size range is from 3.5 to 1000 Å.
Geometry Cylinder
Substrate Carbon
Category: Porosity
Method: Nitrogen at 77 K; Argon at 87 K
Model isotherms were calculated using the prescriptions of Tarazona for density dependent weighting
functions and cylindrical pore geometry. The pore wall potential is described by the Lennard-Jones
potential of interaction between a gas molecule and multiple concentric graphitic surfaces of infinitely
long cylinders.
This model is particularly useful for characterizing carbon multi-wall nanotubes. The reported pore
size range is from 3.5 to 1000 Å.
Geometry Cylinder
Substrate Zeolite
Category: Porosity
Method: Argon at 77 K
Model isotherms were calculated using the prescriptions of Tarazona for density dependent weighting
functions and cylindrical pore geometry. The pore wall potential is described by the Lennard-Jones
potential of interaction between a gas molecule and the oxide surface of an infinitely long cylinder.
This model is particularly useful for characterizing oxides and H+ and (NH4)+ exchanged zeolites. The
reported pore size range is from 3.5 to 300 Å.
Geometry Cylinder
Substrate Zeolite
Category: Porosity
Method: Argon at 77 K
Model isotherms were calculated using the prescriptions of Tarazona for density dependent weighting
functions and cylindrical pore geometry. The pore wall potential is described by the Lennard-Jones
potential of interaction between a gas molecule and the oxide surface of an infinitely long cylinder.
This model is similar to the model above, but it more appropriate is for characterizing alkali metal
exchanged zeolites. The reported pore size range is from 3.5 to 300 Å.
Surface Energy
The use of classical theories to extract adsorptive potential distribution is mostly of historical interest.
At a minimum, the equation must contain a parameter dependent on adsorption energy and another
dependent on monolayer capacity or surface area. This is sufficient to permit the calculation of the set
of model isotherms that is used to create a library model. The Langmuir equation has been used in the
past, as have the Hill-de Boer equation and the Fowler-Guggenheim equation. All of these suffer from
the fact that they only describe monolayer adsorption, whereas the data may include contributions
from multilayer formation.
Pore Size
It is well established that the pore space of a mesoporous solid fills with condensed adsorbate at pres-
sures somewhat below the prevailing saturated vapor pressure of the adsorptive. When combined with
a correlating function that relates pore size with a critical condensation pressure, this knowledge can be
used to characterize the mesopore size distribution of the adsorbent. The correlating function most
commonly used is the Kelvin equation. Refinements make allowance for the reduction of the physical
pore size by the thickness of the adsorbed film existing at the critical condensation pressure. Still fur-
ther refinements adjust the film thickness for the curvature of the pore wall.
The commonly used practical methods of extracting mesopore distribution from isotherm data using
Kelvin-based theories, such as the BJH method, were for the most part developed decades ago and
were designed for hand computation using relatively few experimental points. In general, these meth-
ods visualize the incremental decomposition of an experimental isotherm, starting at the highest
relative pressure or pore size. At each step, the quantity of adsorptive involved is divided between pore
emptying and film thinning processes and exactly is accounted for. This computational algorithm fre-
quently leads to inconsistencies when carried to small mesopore sizes. If the thickness curve used is
too steep, it finally will predict a larger increment of adsorptive for a given pressure increment than is
actually observed; since a negative pore volume is non-physical, the algorithm must stop. Conversely,
if the thickness curve used underestimates film thinning, accumulated error results in the calculation of
an overly large volume of (possibly nonexistent) small pores.
The use of equation (1) represents an improvement over the traditional algorithm. Kernel functions
corresponding to various classical Kelvin-based methods have been calculated for differing geometries
and included in the list of models.
Models Included
Geometry Slit
Substrate Average
Category: Porosity
Method: Nitrogen at 77 K
The kernel function is calculated using the Halsey equation with standard parameters:
– 5.00 1/3
t = 3.54 ----------------------
ln P/Po
Geometry Cylinder
Substrate Average
Category: Porosity
Method: Nitrogen at 77 K
The calculation is the same as above except that cylindrical geometry is assumed.
Geometry Slit
Substrate Average
Category: Porosity
Method: Nitrogen at 77 K
The kernel function is calculated using the Harkins and Jura equation with standard parameters:
13.99 1/2
t = ---------------------------------------------
0.034 – log P/Po
Geometry Cylinder
Substrate Average
Category: Porosity
Method: Nitrogen at 77 K
The calculation is the same as above except that cylindrical geometry is assumed.
Geometry Slit
Substrate Average
Category: Porosity
Method: Nitrogen at 77 K
The kernel function is calculated using the Broekhoff-de Boer equation with standard parameters:
0 – 16.11 – 0.1137
t
log p/p = ---------------
2
- + 0.1682
t
Geometry Cylinder
Substrate Average
Category: Porosity
Method: Nitrogen at 77 K
The calculation is similar to the above except that cylindrical geometry is assumed, and the film thick-
ness depends on pore size (see reference).
Reference: Specifically, equations 20 and 21 in: J.C.P. Broekhoff and J.H. de Boer,
“The Surface Area in Intermediate Pores,” Proceedings of the
International Symposium on Surface Area Determination, D.H. Everett,
R.H. Ottwill, eds., U.K. (1969).
References
The papers listed below provide additional information on DFT models:
3. “Modeling Physical Adsorption on Porous and Nonporous solids Using Density Functional
Theory,” J.P. Olivier, Journal of Porous Materials, 3, 9-17 (1995).
INDEX
A Ar Carbon Finite Pores, As=12, 2D-NLDFT model, F-11
Ar Carbon Finite Pores, As=6, 2D-NLDFT model, F-10
Absolute pressure dosing, 5-12
Ar Carbon Slit Pores by NLDFT model, F-10
Accessories, ordering, 10-1
AR Cylindrical Pores - Oxide Surface model, F-8
Adding an analyzer, 4-13
AR DFT model, F-7
Adsorption isotherm, F-1
AR Modified Density Functional Model, F-9
Adsorption points, 1-5
Argon Cylinder, multi-wall nanotube by NLDFT model,
Adsorptive properties
F-12
command, 5-17
Argon Cylinder, single-wall nanotube by NLDFT model,
create file, 3-14
F-11
Advanced format
Avogadro's number, C-19
sample information file, 3-6
Axis, view coordinates, 7-23
Advanced format, Sample defaults, 3-2
All file status, 2-8
Alpha-S Method, 5-33 B
Alumina
Inspecting and Changing Alumina, video, 2-13 Balance tube, on schematic, 6-9
replacing in oil vapor traps, 9-11 Basic
Analysis format, 3-1
bath, 1-6 sample information file, 3-7
blank, 3-27, 9-26 BET surface area
Dewar, 1-9, 3-22 calculations, C-6
Dewar, checking, 9-5 report, 5-24
gas, 1-5, 1-7 BJH
performing, 3-25 Adsorption/Desorption report, 5-36
preparing for, 3-11 calculations, C-10
QuickStart method, 3-26 plot report options, 5-40
small surface area sample, E-1 pore volume and area distribution, C-10
Analysis conditions tabular report options, 5-38
calculation assignments, 5-12 Blank analysis
command, 5-11 command, 6-7
creating the file, 5-11 performing, 3-27, 9-26
defined, 3-13 Boltzmann’s constant, F-7
Analysis program Broekhoff-de Boer thickness curve, 5-31
installing, 4-2
uninstalling, 4-31
C
using, 2-6
version number, 2-13 Calculations, C-1
Analyzer Calibration
adding, 4-13 command, 6-19
changing Ethernet connection, 4-22 information, 6-16
cleaning, 9-5 load file, 6-23
configurations, 6-16 match transducers, 9-31
description, 1-4 reinstalling calibration file, 4-24
maintaining, 9-4 save to file, 6-22
moving, 4-16 servo valve, 6-21, 9-31, 9-32
physical dimensions, 1-9 zero pressure, 6-20, 9-30
removing, 4-20 Carbon Black STSA thickness curve, 5-31
turning on/off, 2-5 Carbon Cylinder
Analyzing file status, 2-8 multi-wall nanotube by NLDFT model, F-12
W
Wall potential, F-10
Warnings, defined, 1-3
Weighing sample, 3-18
Window menu
arrange icons, 2-11
tile and cascade, 2-11
OPERATOR MANUAL
Smart VacPrep Operator Manual
Nov 2013
Micromeritics is a registered trademark of Micromeritics Instrument Corporation.
Microsoft, Windows, and Windows 7 are registered trademarks of Microsoft Corporation.
Smart VacPrep is a registered trademark of Micromeritics Instrument Corporation.
_____________________________________________________________________________________
© Micromeritics Instrument Corporation 2013. All rights reserved. Printed in the U.S.A.
The software described in this manual is furnished under a license agreement and may be used or
copied only in accordance with the terms of the agreement.
Warranty
WARRANTY
MICROMERITICS INSTRUMENT CORPORATION warrants for one year from the date of shipment
each instrument it manufactures to be free from defects in material and workmanship impairing its
usefulness under normal use and service conditions except as noted herein.
Our liability under this warranty is limited to repair, servicing and adjustment, free of charge at our
plant, of any instrument or defective parts when returned prepaid to us and which our examination
discloses to have been defective. The purchaser is responsible for all transportation charges involving the
shipment of materials for warranty repairs. Failure of any instrument or product due to operator error,
improper installation, unauthorized repair or alteration, failure of utilities, or environmental
contamination will not constitute a warranty claim. The materials of construction used in
MICROMERITICS instruments and other products were chosen after extensive testing and experience
for their reliability and durability. However, these materials cannot be totally guaranteed against wear
and / or decomposition by chemical action (corrosion) as a result of normal use.
Repair parts are warranted to be free from defects in material and workmanship for 90 days from the
date of shipment.
MICROMERITICS shall be released from all obligations under its warranty in the event repairs or
modifications are made by persons other than its own authorized service personnel unless such work is
authorized in writing by MICROMERITICS.
The obligations of this warranty will be limited under the following conditions:
1. Certain products sold by MICROMERITICS are the products of reputable manufacturers, sold
under their respective brand names or trade names. We, therefore, make no express or implied
warranty as to such products. We shall use our best efforts to obtain from the manufacturer, in
accordance with his customary practice, the repair or replacement of such of his products that
may prove defective in workmanship or materials. Service charges made by such manufacturer
are the responsibility of the ultimate purchaser. This states our entire liability in respect to such
products, except as an authorized person of MICROMERITICS may otherwise agree to in writ-
ing.
2. If an instrument or product is found defective during the warranty period, replacement parts may,
at the discretion of MICROMERITICS, be sent to be installed by the purchaser, e.g., printed cir-
cuit boards, check valves, seals, etc.
3. Expendable items, e.g., sample tubes, detector source lamps, indicator lamps, fuses, valve plugs
(rotor) and stems, seals and O-rings, ferrules, etc., are excluded from this warranty except for
manufacturing defects. Such items which perform satisfactorily during the first 45 days after the
date of shipment are assumed to be free of manufacturing defects.
Purchaser agrees to hold MICROMERITICS harmless from any patent infringement action brought
against MICROMERITICS if, at the request of the purchaser, MICROMERITICS modifies a standard
product or manufactures a special product to the purchaser’s specifications.
MICROMERITICS shall not be liable for consequential or other type damages resulting from the use of
any of its products other than the liability stated above. This warranty is in lieu of all other warranties,
express or implied, including but not limited to, the implied warranties of merchantability or fitness for
use.
For Chemisorption - Indicates the marked item is applicable to chemisorption analyzers only.
For Physisorption - Indicates the marked item is applicable to physisorption analyzers only.
Related Topics
Related Topics provides links to additional information about the subject matter. Related topic links are
located at the end of each topic.
CONTACT US
Micromeritics Instrument Corporation
4356 Communications Drive
www.Micromeritics.com
---------------
U.S. Inquiries
U.S. Sales: (770) 662-3633
ussales@micromeritics.com
---------------
orders@micromeritics.com
---------------
Warranty i
Contact Us v
Index Index - 1
The Smart VacPrep is an advanced six-port degassing system that utilizes vacuum to degas samples by
heating and evacuation. Each of the ports may be operated independently of one another. Samples may
be added or removed from degas ports without disturbing the treatment of other samples undergoing
preparation. When starting a new sample, ports already degassing samples are automatically suspended
unless they are in the slow evacuation stage. When unrestricted evacuation is attained by the new
samples, the previously suspended samples will automatically resume their evacuation. Degassing will
automatically terminate when each sample has completed all programmed steps.
The Smart VacPrep remembers the previously selected set of degas conditions and allows them to be
reused. However, the option of programming each port with a different set of parameters is available. A
record of the sample preparation time and temperature can be recorded as part of the sample data file.
The degas program may be started or terminated by using the analyzer software on the computer
attached to the analyzer or with push buttons on the front panel of the Smart VacPrep. The push button
control on each port provides the ability to attach the sample tube with heating mantle, and immediately
start degassing a sample.
There are two methods of starting the degas process from the analyzer software:
l Start Degas menu option. Accessed from the Smart VacPrep > Unit [n] > Start Degas menu selec-
tion. Provides the option to use degas conditions from selected sample files or a degas conditions files
for each Smart VacPrep port. When using the Start Degas menu option, the degas process is started
using the Start button on the Start Degas window. This option is more convenient for samples that
have specific degas requirements that should be associated with the sample file when it is created.
l QuickStart menu option and Smart VacPrep panel buttons. Accessed from the Smart VacPrep >
Unit [n] > QuickStart Degas Conditions menu selection. Provides the option to enter commonly used
degas conditions for each Smart VacPrep port without the need for a sample information file.
This option provides three ways to load information into QuickStart Degas Conditions:
Using the Start Degas option does not change the degas conditions set up in QuickStart. QuickStart
degas conditions are remembered for the next push button operation.
Sample tubes are attached to the sample ports. Heating mantles are then installed and secured to the
sample tube with a metal clip. If using straight wall sample tubes, heating mantles with elastic cords
may be purchased separately.The mantle is plugged into the power and thermocouple connectors on the
front panel. The degassing process can then be started either by pressing the Select and Start buttons on
the Smart VacPrep unit or by going to Smart VacPrep > Unit [n] > Degas > Start Degas.
A Check Seal or TransSeal can be used to transfer air-sensitive samples from the Smart VacPrep to the
analyzer's port without atmospheric contamination. To order these parts, reference the Parts and
Accessories section for the analyzer. Instructions for using the Check Seal or TranSeal are included
with the ordered part.
Green. Indicates the station is idle and not hot. The setpoint temperature
is below 30 ºC and the actual temperature is below 50 ºC.
Yellow. Indicates the station is in use, but not currently being heated. The
heating mantle may still be above 50 ºC.
Related Topics
See "Set the Voltage for the Smart VacPrep" on the next page
Computer
Windows 7 Professional or higher operating system is recommended for the best user
experience. If the computer is to be connected to a network, a second Ethernet port on the
computer must be used for that purpose.
1. Use a flat head screwdriver to gently pry open the cover of the compartment.
2. Firmly press down on the voltage indicator and roll the indicator to display the appropriate voltage
for your environment. Selections are: 100 Vac, 120 Vac, 230 Vac, and 240 Vac.
3. Firmly press the compartment cover shut. Ensure it closes securely. The selected voltage will
appear in the window on the compartment cover.
If a software update is required for the analyzer to support the Smart VacPrep, the software can be
downloaded from the internet at this location:
www.micromeritics.com/Smart-VacPrep-Software.aspx
The degasser software is installed as a part of the analyzer installation. During installation, an
IP address will be assigned to the unit allowing the unit to connect automatically during startup. Up to
three additional degasser units can be installed.
Menu options for the Smart VacPrep are provided on the Smart VacPrep menu.
Related Topics
A Smart VacPrep unit must be removed from one application before it can be added to another
application if they are running on the same subnet.
1. When prompted, enter the serial number of the unit being added. Duplicate serial numbers are not
allowed.
2. Click Add. Initialization begins.
The QuickStart Degas Conditions option provides the ability to enter commonly used degas conditions
for each Smart VacPrep port without the need for a sample information file. Degassing can then be
easily started using the front panel buttons on the Smart VacPrep unit.
This option provides three ways to load information into QuickStart Degas Conditions:
When using the QuickStart menu option, the degas process is started using the Select button and Start
buttons for each port on the Smart VacPrep unit.
QuickStart degas conditions are remembered for the next push button operation.
of the Degas Conditions field. If the file is not shown, select Browse , locate the file, then
click Load.
l To copy QuickStart Degas Conditions from another Smart VacPrep unit, select the unit and
port number to the right of the Copy QuickStart Degas Conditions from Smart VacPrep Unit
field..
l To enter degas conditions manually, complete the fields on the window. Reference the Degas
Conditions section of the Operator Manual
4. Click Save. Repeat these steps for each port to be configured. If multiple ports will have the
same degas conditions parameters, click Edit on the port and use the Copy QuickStart Degas
Conditions from Smart VacPrep Unit [n] Port [n field to load the settings from the first port.
5. Repeat these steps for each applicable unit and port.
6. If using a Check Seal, verify that the Check Seal opener is installed in the Smart VacPrep port.
Follow the instructions included with the Check Seal for instructions on inserting the opener.
7. Load the sample into the sample tube. If using a Check Seal or TranSeal, insert it into the sample
tube.
8. Load the sample on the Smart VacPrep sample port.
9. Attach the heating mantle using the metal clip. If a straight wall tube is used, heating mantles with
elastic cords are available. Reference the Parts and Accessories section of the analyzer Operator
Manual.
This option allows the degassing of up to six samples simultaneously. A sample can be added to any idle
port and a degas operation started. When adding a sample, the degas operation in progress will be
suspended until the sample reaches the target pressure, then degassing on all ports will resume. If other
samples are still in the initial evacuation phase, the new samples will wait until the fast evacuation is
started. Samples can be removed from any of the six ports without disturbing the degas operations
already in process. Starting several samples as a batch allows them to perform the initial evacuation at
the same time.
Degas conditions remain the same from degas to degas unless the Degas Conditions file is changed or
replaced by the operator. A manual evacuation can be performed on any idle degas port.
The six ports are represented by row numbers. Ports that are busy are grayed out and disabled.
1. Click Browse to search for a sample file or a degas conditions file. Select the file, then click
Open. Do this for each Smart VacPrep port to be used.
2. If using a Check Seal, verify that the Check Seal opener is installed in the Smart VacPrep port.
Follow the instructions included with the Check Seal for instructions on inserting the opener.
3. Load the sample into the sample tube. If using a Check Seal or TranSeal, insert it into the sample
tube.
4. Load the sample on the Smart VacPrep sample port.
5. Attach the heating mantle using the metal clip. If a straight wall tube is used, heating mantles
with elastic cords are available. Refer to the Parts and Accessories section of the analyzer oper-
ator manual.
6. Install the safety shield.
7. Click Start on the Start Degas window to begin the degas process for the selected Smart
VacPrep ports.
8. Click Start to begin the degas process for the selected Smart VacPrep ports.
9. Allow the samples to cool before transferring them to the analysis ports to start the analysis.
Smart VacPrep Start Degas Fields and Buttons Table
Field or Button Description
Browse Searches for a file. Select a file from the Name column or from the
library, then click Open. Alternatively, double click the file name to
open (or import) the file.
Cancel Discards any changes or cancels the current process.
Clear Clears the field.
Sample Displays the path and file name of the selected .SMP file.
Start Starts the degas operation for the selected ports.
Degas ports can also be canceled using buttons on the Smart VacPrep
unit. Press Select on the unit to select the port, then press the Cancel but-
ton.
Use to display an analyzer schematic. To operate the valves and elevator from this window, manual
control must be enabled (Smart VacPrep > Unit [n] > Enable Manual Control).
The Smart VacPrep Status window allows the monitoring of degas operations for a manual degas
operation.
If the outgassing rate indicated that the sample has been freed of
contaminants (minimal pressure increases), click Skip to advance to the
next state of the degassing operation. For example, if degassing is
checked after the setpoint is attained, Skip advances you to the ramping
stage.
Various calibration procedures and service tests are included in the operating program. These procedures
and tests are designed to provide a service representative with instrument readouts, as well as to assist in
troubleshooting potential problems.
Related Topics
Use to perform system calibration. This option evacuates the system and zeros the pressure transducer.
Disabled calibration options can be accessed only with the assistance of an authorized Micromeritics
service representative.
Use to calibrate the servo valve to the manifold pressure transducer. The servo valve should always be
recalibrated after a pressure calibration has been performed. The pressure transducer should be
calibrated before starting this calibration procedure.
1. Go to Smart VacPrep > Unit [n] > Calibration > Servo Valve.
2. Click Start. The window closes when the calibration is complete. Click Cancel to stop the cal-
ibration process.
Servo Valve Fields and Buttons Table
Field or Button Description
Cancel Discards any changes or cancels the current process.
Start Starts the operation.
Related Topics
Displays previously run diagnostic tests. Separate directories store tests run once, daily, weekly, and
monthly.
Allows the specification of one-time or periodic running of a sequence of diagnostic tests. A separate list
of tests is saved for each of the possible test frequencies. Tests are categorized and flagged as requiring
intervention or not. If tests requiring intervention are scheduled, the operator has the option of skipping
these tests if the operator does not respond within a specified time after an initial prompt is displayed,
before the test is started. Events are logged in the instrument log for all starting, ending, and skipped
tests.
Use to enable the manual control of certain system valves and elevator components. When this option is
enabled, a checkmark appears to the left of Unit [n] > Enable Manual Control.
If the schematic is not immediately visible, go to Unit [n] > Show Instrument Schematic.
1. Use a flat head screwdriver to gently pry open the cover of the compartment.
2. Use the flat head screwdriver to gently pry the fuses out of the holder.
3. Insert a new fuse into each holder, then push the holders back into the slot until snug.
4. Press the compartment cover shut. Ensure it closes securely.
2. If using a rubber stopper, remove it from the sample tube.If using a Check Seal or TransSeal, do
not remove them from the sample tube. Place the port connector nut, ferrule, and O-ring onto the
sample tube.
3. If using a Check Seal, verify that the Check Seal opener is installed in the sample port.
4. To attach the sample tube set to the degas port, push the sample tube in to a full stop. Secure the
sample tube in place by sliding the connector nut, ferrule, and O-ring up onto the degas port and
turning the connector nut clockwise. Tighten the nut securely by hand.
5. Place a heating mantle over the sample tube bulb and secure the mantle in place with a mantle
clip.
6. Insert the heating mantle thermocouple plug into the thermocouple connector.
7. Insert the heating mantle power plug into the power connector and twist the power plug to lock
securely.
8. After installing the sample tubes, install the safety shield around the sample tubes and heating
mantles to minimize the risk of receiving a burn by touching hot components.
The equipment shown in this section may differ slightly from yours.
However, the process is the same unless otherwise noted.
This procedure is also shown in the How-To Tutorials located on the Help
menu.
The oil in the vacuum pump should be changed every three months when the efficiency of the vacuum
pump declines (requiring increased time to reach vacuum levels) or if it becomes discolored.
Drain the oil while the pump is warm and disconnected from the power
source.
Use oil supplied by Micromeritics or refer to the vacuum pump manual for other acceptable oils.
3. Grasp the handle on top of the vacuum pump and place the pump on a work surface.
4. Place a waste container under the drain spout.
5. Remove the drain plug and drain the oil into the waste container.
6. Replace the drain plug.
7. Remove the drain plug from the oil-fill port on top of the pump.
8. Slowly add oil to the port until the level is midway between the indicator lines in the oil-level win-
dow when the pump is running.
Do not allow oil to rise above the midway position when the pump is run-
ning. Doing so may cause oil to splash into the oil filter causing con-
tamination.
7. Check the washer or O-ring used at the oil-filling port and replace if necessary.
8. Insert the oil-fill plug and turn clockwise to tighten.
9. Check the alumina in the oil vapor trap. If most of the pellets are no longer white,
replace the alumina before reattaching the vacuum pump. See "Replace the Alu-
mina in the Oil Vapor Trap" on the facing page.
10. Reconnect the vacuum pump hose.
11. Reconnect the power cord to the power source.
12. Turn the vacuum pump on and recheck the oil level.
13. Allow the pump to run a few hours (overnight if possible) to eliminate air and mois-
ture from the fresh fluid and to produce efficient vacuum operations.
Related Topics
See "Replace the Alumina in the Oil Vapor Trap" on the facing page
The equipment shown in this section may differ slightly from yours.
However, the process is the same unless otherwise noted.
The activated alumina in the oil vapor trap becomes saturated during use. The alumina should be
inspected periodically and replaced when most of the alumina pellets are no longer white.
1. Disconnect the vacuum pump from the analyzer and place it on a work table.
2. Loosen the wing nut on the clamp at the top of the oil vapor trap. Swing the clamp open and
remove the trap from the hose.
3. Loosen the wing nut on the clamp at the bottom of the oil vapor trap. Open the clamp and remove
the trap.
4. Remove one end fitting from the trap body and dispose of the used alumina in an appropriate
manner.
5. Wash the trap body with a detergent-based soap. Rinse with water, then rinse with isopropyl or
ethyl alcohol. Set the trap aside and allow to dry thoroughly.
Exposure of the trap body to oil vapor may cause small cracks on the
inside surface of the trap body. Under normal circumstances, these
cracks will not cause problems or leaks.
6. Inspect the O-rings. If dusty, clean with a lint-free tissue. If damaged, replace with a new O-ring.
7. Screw one of the end fittings onto the trap body.
8. Ensure the trap body is dry and the alumina is lukewarm. Pour the alumina pellets into the trap
until level with the top of the trap body.
9. Screw the other end fitting back onto the trap and tighten securely by hand.
10. Lightly tap both ends of the trap body on the work surface to remove remaining dust from the pel-
lets.
11. Inspect the centering ring before replacing it onto the intake port. If it appears to be flattened,
replace it. A flattened centering ring can cause vacuum leaks. There are two types of centering
rings. Use the one with the smaller opening at the intake port.
12. Place the centering ring on the intake port.
13. Place the trap on the centering ring.
14. Open the clamp and place it around the flange of the intake port and the flange of the trap. Swing
the clamp fastening screw toward the intake port until it fits into the slot in the other half of the
clamp. Tighten the wing nut securely by hand.
15. Reconnect the hose from the analyzer to the oil vapor trap.
a. Place the clamp around the vacuum pump hose flange and vapor trap.
16. Plug the pump power cord into the power source. Allow the pump to run a few hours (overnight if
possible) to eliminate air and moisture from the fresh oil and to produce efficient vacuum
operations.
Related Topics
The equipment shown in this section may differ slightly from yours.
However, the process is the same unless otherwise noted.
The gases used by the analyzer are exhausted by the oil-filled vacuum pump. The filter minimizes the
release of oil vapor and should be replaced when it becomes so saturated with oil that it is ineffective.
Exhaust filters are used to minimize the release of oil vapors. The gases
are diluted substantially upon release from the vacuum pump; however,
it may be necessary in some locations to provide a fume hood for pro-
tection from hazardous gases and vapors released into the work area.
1. Turn off the power to the pump and disconnect the pump from the power source.
2. Loosen the clamp wing nut at the vacuum pump exhaust port. Swing the clamp away from the
exhaust port and remove it.
3. Remove and discard the exhaust filter. Ensure the centering ring remains in place on the exhaust
port.
4. Place the new filter on the centering ring.
5. Open the clamp and place it around the exhaust port flange and the exhaust filter flange. Swing
the clamp fastening screw toward the exhaust port until it fits into the slot in the other half of the
clamp. Tighten the wing nut securely by hand.
6. Turn on the power to the pump.
Related Topics
8 ERROR MESSAGES
Program error messages are listed numerically. If the Action response indicates to contact a
Micromeritics service representative, record the error message, then make backup copies of any files
involved in the operation.
4407 Error searching for installed Smart VacPreps. The registry could not be read..
Cause: The application was not installed properly.
Action: Reinstall the application. If the problem persists, contact your Micromeritics service rep-
resentative.
4408 The %s in %s already has Smart VacPrep S/N %d. The Smart VacPrep
must be removed from the %s before it can be added.
Cause: The Smart VacPrep was already installed for another application / unit.
Action: Remove the Smart VacPrep from the installed unit before adding it to the preferred unit.
4410 The .INI file could not be updated with configuration for Smart
VacPrep S/N %s."
Cause: The application .INI file is opened by another application and could not be updated.
Action: Close all open applications and add the Smart VacPrep again using the Smart VacPrep
menu.
4424 The file %s on port %d could not be opened. Check if the sample file is
already in use for editing or analysis.
Cause: The selected sample file is already open by this or another application.
Cause: The selected sample file is damaged.
Action: Select another sample file.
4425 The sample %s on port %d has an invalid status and cannot be used
for degassing.
Cause: The status of the file is not consistent with the current operation.
Action: Select a sample file that has not been used for an analysis. Only sample files with a
status of No Analysis or Prepared may be selected.
INDEX F
Fuses
B replace on the Smart VacPrep 6 - 2
Calibration
Smart VacPrep 4 - 1 M
Smart VacPrep, pressure offset 4 - 1
Smart VacPrep, servo valve 4 - 2 Mainteane
Configuration fuse replacement on the Smart VacPrep 6 -
Smart VacPrep 2 - 3 2
Contact Us v Maintenance
install sample tube, Ssmart VacPrep 6 - 2
Manual control
D Smart VacPrep 6 - 1
Dashboard
Smart VacPrep 3 - 7 P
Smart VacPrep, manifold outgas rate 3 - 7
Smart VacPrep, roughing pump 3 - 7 Parts and accessories
Degas Smart VacPrep 7 - 1
using the Smart VacPrep 3 - 5 Pressure offset
Degas conditions Smart VacPrep 4 - 1
heating phase 3 - 4 Pump
soak time 3 - 4 Edwards 6 - 3
temperature ramp rate 3 - 4 inspect and change oil 6 - 3
Diagnostics
Smart VacPrep 5 - 1
Smart VacPrep, schedule tests 5 - 1 S
Smart VacPrep, test report 5 - 1
Sample tube
install, Smart VacPrep 6 - 2
E Schematic
shortcut menus 3 - 10
Edwards vacuum pump SmartVacPrep 3 - 9
inspect and change oil 6 - 3 Servo valve
oil vapor trap 6 - 7 SmartVacPrep, calibrate 4 - 2
replace alumina 6 - 7 Shortcuts
replace exhaust filter 6 - 11 schematic 3 - 10
Elements used in this manual iii Smart VacPrep
Error messages 8 - 1 about 1 - 1
Evacuate ports, Smart VacPrep 3 - 1
backfill ports 3 - 1
calibrate servo valve 4 - 2
calibrate, pressure offset 4 - 1
calibration 4 - 1
configuration 2 - 3
diagnostic test report 5 - 1
Diagnostics 5 - 1
evacuate ports 3 - 1
install additional unit 2 - 1
install sample tube 6 - 2
Parts and accessories 7 - 1
QuickStart degas conditions 3 - 2
remove a unit 2 - 1
schedule diagnostic tests 5 - 1
schematic 3 - 9
show dashboard 3 - 7
show status 3 - 11
software
about 2 - 1
specifications 1 - 3
start degas 3 - 5
troubleshooting and maintenance 6 - 1
use 3 - 1
Soak time, degas conditions 3 - 4
Specifications
Smart VacPrep 1 - 3
System valves 3 - 10, 6 - 1
T
Temperature ramp rate 3 - 4
Troubleshooting and maintenance
Smart VacPrep 6 - 1
Smart VacPrep, enable manual control 6 - 1
V
Voltage
set for the Smart VacPrep 1 - 4
W
WARRANTY i