SPD 20 A
SPD 20 A
SPD 20 A
UV-VIS DETECTOR
FOR
SHIMADZU HIGH PERFORMANCE
LIQUID CHROMATOGRAPH
SPD-20A/20AV
SPD-20A UFLC
SERVICE MANUAL
CAUTION !
This service manual has been edited for those that have received service related
training.
Those that have not been trained, should not use this manual.
SHIMADZU
Analytical & Measuring Instruments Division
228-90095F
CONTENTS
Chapter 1 Introduction........................................................................................................................6
1.1 Introduction...........................................................................................................................................................................6
Chapter 2 Specifications......................................................................................................................7
Chapter 8 Troubleshooting..............................................................................................................46
8.1 Using symptoms to identify the problem................................................................................................................46
8.2 Using the error display to identify the problem....................................................................................................46
8.3 Problems with electrical circuits and points to check.........................................................................................47
8.3.1 SPD20-MAIN circuit board....................................................................................................................................47
8.3.2 SPD20-PWR(POWER) circuit board.................................................................................................................49
【Amendment history】
D “5.3 List of factory functions SIGNAL MODE” was corrected.
“4.7.2 List of primary internal parts for optical unit List of primary internal
parts for optical unit” was corrected.
“9.2 Optical repair parts list”was corrected.
F “6.3.10 Fine adjustment of mirror M1” was added to optical system adjustment
process, “6.3.3 Optical adjustment jigs (targets)” was added, the paper target copy was
discontinued, and “7.1.3 Changing the gain of operation amplifier” was added as
additional explanations. 7.2.13 Add Notes of Clause of Linearity Calibration.
6.3.12 Add the Method of Lighting of Hg Lamp by Clause of Adjustment of Hg Lamp.
228-90095F
Chapter 1 Introduction
1.1 Introduction
This service manual applies to SPD-20A, SPD-20AV and SPD-20A UFLC
detectors. The difference between the construction of 20A and 20AV models is
the positioning of the tungsten (W) lamp, the W lamp power supply, and the
mercury (Hg) lamp used for wavelength calibration. The difference between the
-20A and -20A UFLC is the specification of the flow cell (-20A: standard cell,
-20A UFLC: semi-micro cell). The figures in this manual illustrate model 20AV,
but the explanatory text applies to both and having the instrument available to
look at should prevent any confusion. However, please note that parts with the
same name for all models may be differentiated using different part numbers.
CAUTION:
• This service manual includes only the table of contents entries and no
corresponding text for subjects addressed in the instruction manual. Please
use this manual in combination with the instruction manual.
6
228-90095F
Chapter 2 Specifications
Please read a see instruction manual.
When the angle of mirror M1 is in the position shown, the light from the D2 lamp
is reflected and focused through the quartz window and enters the slit. When the
angle position of mirror M1 is switched, the light from the W lamp passes
through a stray light cut-filter (short wavelengths) and heat-absorbing filter, then
in the same manner is reflected and focused by mirror M1 and passes through the
quartz window and enters the slit. A filter assembly (high order cut-filter or
blocking filter) is inserted in front of the slit, based on the wavelength setting.
The blocking filter is inserted to detect zero order light. The high order cut-filter
is inserted in front of the slit to filter higher order spectra in the specified
wavelength region corresponding to the wavelength setting. The filter is inserted
in front of the slit at the 371-700 nm range for model 20A and 701 nm -- 900 nm
for model 20AV. After the light beam passes through the slit, it is reflected and
focused by mirror M2 and hits a holographic grating of 1600 grooves per
millimeter, where it is separated into its spectrum. The light separated by the
grating emerges as parallel light beams that are reflected by mirror M3 and are
wide enough to straddle and enter both of the sample and reference cells. The
spectrum is generated in an orientation perpendicular to the line bisecting both
cells, so both cells receive identical light wavelengths. After passing through each
7
228-90095F
cell, the respective light enters a silicon photodiodes (D1 and D2) where it is
converted to an electrical signal that is proportional to the light intensity.
The light entering the cells, or in other words, the wavelength being measured
during analysis, is selected by changing the angle of the grating. A stepping
motor is used to move the grating via a pulley and belt. When the instrument
power switch is turned on, mirror M1 moves to the angle position for the W lamp,
then again to the D2 lamp position. The lamp switching photomicrosensor detects
the position and the D2 lamp will turn on. When the mirror M1 position is
switched, the home position for the filter assembly is simultaneously detected.
Once the lamp goes on, the grating stepping motor first searches for the home
position of the wavelength setting, then verifies the wavelength using the D2
lamp and Hg lamp emission lines. This instrument measures the amount of light
passing through the cell as the wavelength is varied and automatically identifies
the wavelengths corresponding to the emission line spectrum peaks. If these
wavelengths are within one nanometer of the emission line wavelengths, then
“CHECK GOOD” will appear on the display. Then the grating rotates according
to the wavelength setting and the lamp position switches also. The sample cell
has a flow-through type structure. The reference cell is simply a hole that lets
light pass through. The Mercury lamp is used to check or calibrate wavelengths.
The characteristic emission line of Hg lamps at 253.7 nm is used as a reference
for checking accuracy in the ultraviolet region.
8
228-90095F
9
228-90095F
10
228-90095F
(Measurementside)
11
228-90095F
Highvoltage
triggercircuit
Transformerpowersupply
D2lamp
Fixed current
D2 circuit
Fixed voltage
D2 heater circuit
12
228-90095F
4.3 Inside the front cover, and its top and left sides
Please read a see instruction manual.
13
228-90095F
14
228-90095F
Reference: Procedure for tightening lens fixing screws (part 16) (factory method)
1) Verify that the lens surface fits completely inside the counterbore in the
housing. Do not use gaskets that are too warped. Doing so can raise the
lens and cause it to break during tightening.
2) Use a torque screwdriver to temporarily tighten screws to 0.4 Nm (4
kgf·cm).
3) Use a torque screwdriver to finish tightening screws to 0.8 Nm (8
kgf·cm).
4) Verify there is no lens breakage.
5) Retighten screws to 0.8 Nm (8 kgf·cm) eight hours later.
15
228-90095F
16
228-90095F
17
228-90095F
18
228-90095F
19
228-90095F
20
228-90095F
21
228-90095F
Screws
(3) A quartz window is located behind the spring. Pull the spring out slowly
and remove the window.
(4) Wipe any dirt from the window using a Kimwipe, etc., and reassemble the
parts.
22
228-90095F
23
228-90095F
6.3.2 Preparations
Before adjusting any component, please prepare the following.
24
228-90095F
25
228-90095F
6.3.4 Adjusting M1
> In order to make it factory mode
(1) Turn the power on while pressing the “func” key.
> push func →Factory→and Set λ=0nm
(2) Turn the D2 lamp ON.
(3) Attach the M1 and M2 jigs.
(4) (For model SPD-20AV, use a clip or other means to secure the lever pin
attached to the M1 assembly and the stopper pin.
(5) With the slit assembly removed, adjust the M1 assembly so the light
image hits the center of the target mark on the M2 jig.
Lateral adjustments to image on M2 jig (Paper Target):
Loosen screws “A” on the M1 assembly and rotate the assembly.
Vertical adjustments to image on M2 jig:
Adjust the angle of elevation using the screws “C”. Do not loosen set
screws “B” and the nuts that secure them. After making adjustments,
tighten screws “A”.
26
228-90095F
27
228-90095F
28
228-90095F
6.3.7 Adjusting M2
(1) Remove the M2 jig and attach the M2 assembly.
Attach the M3 jig to assembly M3. Press the function key and set “Z
WAVE”, in the factory group, to 3.
(Wavelength still = 0nm)
(2) Roughly adjust the rotation of M2 and GR so the 0-order light from M2
passes via GR and is laterally centered on M3 (within Φ10mm circle).
Next, attach the GR jig and adjust the M2 assembly rotation so the light
image is laterally centered on the GR target mark. Roughly adjust the
vertical angle of M2 at this point. (Rotation: Adjust by loosening screws
“A” and rotating. Elevation: Adjust by loosening nut and turning set screw
“B”.)
(3) After making adjustments, tighten screws “A” and remove the GR jig, but
don’t tighten the nut at this point.
29
228-90095F
(3) Adjust the grating by rotating it in direction “A” shown above until the
0-order light image reaches a lateral position on the M3 jig that is 1 mm
(±0.3 mm) to the left of center.
(4) Remove the M3 jig.
30
228-90095F
6.3.9 Adjusting M3
(1) Attach the cell jig.
(2) Adjust the rotation and elevation angle of the M3 assembly so the 0-order
light image hits the center of the target on the cell jig.
(3) Remove the M1 jig, attach a dummy cell, and double check that the long
narrow 0-order light image is over the cell hole.
Caution:Be careful not to touch the mirror or grating surfaces with your hands.
31
228-90095F
mirror M1.
(4) Set the pin of the lever (rotating table)
where the mirror M1 ASSY is mounted and
the stopper pin on the lamp housing using a
clip or something similar.
M1 ASSY
(5) Adjust the direction of the mirror M1 ASSY
so that two optical images on the slit (figure
below) go into the slit. (Unfasten screw A on
the mirror M1 ASSY and rotate the mirror
M1 ASSY.)
(6) After adjustment is completed, tighten screw A to fix it and detach the jig
for fine adjustment of M1 and clip.
D2 lamp
Slit
スリット
Slit
Rotation
Pin of the lever
Optical
光の像 Stopper pin
images
M1 ASSY Screw A
Clip
32
228-90095F
(3) Set ”FILTER POSI” to “4” and check that the second-order light cut-filter
covers the entire front surface of the slit.
(4) Set ”FILTER POSI” to “7” and check that the deep green filter covers the
entire front surface of the slit.
33
228-90095F
Brightest Image
34
228-90095F
(Model SPD-20AV)
The Hg lamp will project multiple images onto the slit from multiple
reflections in the VIS filter, as shown below. Adjust the Hg lamp by moving
it in the direction of the arrow shown below, until the brightest Hg lamp
image hits the center of the monochromator.
Brightest Image
35
228-90095F
36
228-90095F
* 1: The light intensity standard (upper limit) differs according to the type of the
MAIN board (PCB, SPD20-MAIN ASSY); new or former. The new and former
boards can be distinguished by the BOARD identification no.*3 serigraphed on
the MAIN board.
* 2: The gain of the pre-amplifier on the RE side is maintained at a low level on
instruments where a micro cell is attached. Thus, the light intensity does not
saturate even after the optical system is adjusted using the added jig (jig for fine
adjustment of M1), leaving enough allowance. Therefore, modification is
unnecessary.
* 3: How to distinguish new and former MAIN boards
They can be distinguished by the BOARD identification no. serigraphed on the
MAIN board.
37
228-90095F
Upper limit of the light intensity when the identification no. is “228-42031B”
Cell SE RE
When using a standard cell 2200 mV 2200 mV If the value is over the upper
When using a semi-micro 1400 mV 2200 mV limit, the operational
cell amplifier needs to be
modified.
When using a micro cell Modification is unnecessary.
Upper limit of the light intensity when the identification no. is “228-42031C”
or“228-53021”
Cell SE RE
When using a standard cell 3100 mV 3100 mV If the value is over the upper
When using a semi-micro 2000 mV 3100 mV limit, the operational
cell amplifier needs to be
modified.
When using a micro cell Modification is unnecessary.
38
228-90095F
* The lamp and optical elements (e.g. mirror, grating) of instruments where the
pre-amplifier has been modified may have to be replaced in several months to
several years depending on use frequency. In this case, the light intensity may
exceed the standard value due to a variation of the lamp or optical element
attached. However, no additional modification is necessary for instruments for
which the pre-amplifier has been modified once because there is enough
allowance in light intensity saturation voltage.
* If wax contained in soldering material remains on the surface of the board after
modification, it may cause poor insulation and eventually noise. If the board is
soiled, wipe it off with Kimwipe with a small amount of ethanol or IPA.
* Take out the PB-1 ASSY from the optical unit and modify the soldered face of the
PB-4 with an alignment of gain resistors mounted on the PB-1 ASSY using a
jumper wire* and solder as shown in the figure below.
PB-1ASSY
Optical unit
39
228-90095F
40
228-90095F
time settings have not been entered, dates for log entries will use total
system usage time values, expressed in the form T000001 (total usage
time of one hour)
7.2.5 Password
Instrument calibration and hardware related settings are performed using
the calibration support group functions. Password security is provided to
prevent operators from changing these settings. Passwords are set to
“00000” when shipped from the factory, but customers (instrument
administrators) can change it freely. In the event a customer forgets a
changed password, perform the following.
① Startup the instrument in special mode
② Enter “00000” for the password setting (the special mode allows either
the correct password or 00000)
③ Set a new password
41
228-90095F
Instrument model
SPD-10A SPD-10Avp SPD-20A SPD-M10Avp Remarks
SPD-10A Operating life
Yes Yes* Yes* No
(670-14603) = 1000 hrs
SPD-10Avp,-20A
Yes Yes Yes No
(22834016-02)
SPD-M10Avp
Yes Yes Yes Yes
(228-34016)
* Note the guideline for initial replacement is 2000 hours.
42
228-90095F
(Procedure)
① Connect the instrument to be updated to the computer via the
OPT-232C adapter and start the computer.
② Start the SPD system in the version update mode (turn on the power
while pressing the VP key). The instrument display will show
“Monitor-20A Receive Ready“.
③ Open the version update application file “Fupdate.exe”.
④ Choose a COM port, select the ROM firmware (spdv xxx .rom: where
“xxx” indicates the version number), and begin updating by clicking
“Start”. The instrument display will show “Monitor-20A wait
Completion“.
⑤ Once the update is complete, the computer will display “ROM UPDATE
IS COMPLETED”.
⑥ The instrument display will not change even after the update has been
completed. Briefly turn off the power and restart the instrument.
43
228-90095F
(2) Turn the rotary power supply voltage switch to the desired supply
voltage.
(Alternatively, the rotary switch may be removed, set to the desired
power supply setting and reattached.)
44
228-90095F
(1) [4] inputs [LINEAR CALIB] while on display, and make linearity
compensation into initial value.
(2) Perform wavelength proofreading.
(3) Perform linearity compensation again.
(4) Please once end re-start a power supply.
45
228-90095F
Chapter 8 Troubleshooting
8.1 Using symptoms to identify the problem
Please read a see instruction manual.
46
228-90095F
47
228-90095F
48
228-90095F
49
228-90095F
50
228-90095F
51
228-90095F
9.3.2
52
228-90095F
53
228-90095F
54
228-90095F
55
228-90095F
56
228-90095F
57
228-90095F
58
228-90095F
59
228-90095F
60
228-90095F
61
228-90095F
62
228-90095F
63
228-90095F
64
228-90095F
65
228-90095F
66
228-90095F
67
228-90095F
No. Part Number Part Name No. Part Number Part Name
1 228-32982 Cell housing 14 228-33376-30 Tubing, PEEK 1.6 x 0.25
2 228-25440 Plate ”A” 15 026-02034 Spring pin, SST 2 x 8
Screw, SST sems P2 M3 x
3 228-32983 Angle 16 020-46140
20
4 228-32984 Plate 17 020-46535 Screw, SST sems P3 M3 x 8
5 228-14520 Cell fixing screw 18 020-46634 Screw, SST sems P4 M3 x 6
6 228-33347 Nut 21 228-23424 Cover, cell
7 228-14571 Gasket, PTFE 22 228-24181 Retaining bush
8 228-14572 Lens (R4, RD6) 23 228-32985 Cover
Screw, SST pan head
9 228-14569 Gasket 24 020-02906
M2 x 4
10 228-14568 Lens fixing screw 25 228-33255 Sponge
11 228-32998-02 Gasket, PTFE 6 x 2.5 x 0.2 26 228-31638 Ferrule 1.6F PEEK
12 228-25277-04 Label, inert 27 016-37722-06 FEP tube, NFL016
13 228-33376 Tubing, PEEK 1.6 x 0.25 28 016-37722-06 FEP tube, NFL016
68
228-90095F
69
228-90095F
70