Manual Book Hunterlab
Manual Book Hunterlab
Manual Book Hunterlab
1
User’s Manual for Aeros and EasyMatch Essentials v 1.1
This documentation contains proprietary information of Hunter Associates Laboratory, Inc. Its
reproduction, in whole or in part, without express written consent of Hunter Associates Laboratory, Inc.
is prohibited.
Aeros and EasyMatch are registered trademarks of Hunter Associates Laboratory, Inc.
Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
Caution: If the equipment is used in a manner not specified by the HunterLab, the overall safety may be
impaired. - The instrument is for indoor use only and not suitable for a wet location.
Caution: There is a potential of a UV Light hazard in using this instrument. Please avoid looking directly
at the light. The frequency of this flashing light is in the range of sensitivity for those prone to epileptic
seizures.
Safety Notes
For your safety when using the Aeros, you should pay attention to the following types of statements in
this User’s Manual:
▪ General safety instruction that should be observed at all times while operating the
instrument.
▪ Specific safety instruction critical to the type of instrument operation being explained in the
manual where the caution appears.
▪ Use of this equipment in a manner not specified by the manufacturer may impair the
protection afforded by the equipment.
▪ Danger of electric shock if liquids are spilled and fire if volatile or flammable liquids are
spilled. Use care when measuring liquid samples.
▪ This instrument sensor moves up and down during standardization and creating a
measurement profile. Please keep fingers and other items out of the way of the sensor.
▪ The turntable will rotate if turned on in Read Options > Measurement configuration. Please
take care to remove fingers, jewelry and clothing to prevent damage.
▪ The Aeros is for indoor use only at an altitude of up to 2000m and pollution degree 2.
2
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Contents
SETTING UP THE AEROS ...................................................................................... 7
Standard Accessories ................................................................................................................................ 7
Power Jack ................................................................................................................................................ 7
Power Switch ............................................................................................................................................ 8
Keyboard and Mouse ................................................................................................................................ 8
Front and Rear USB Connectors ............................................................................................................... 8
Ethernet Port ............................................................................................................................................ 8
Moving the Unit ........................................................................................................................................ 9
3
User’s Manual for Aeros and EasyMatch Essentials v 1.1
SPECIFICATIONS ...................................................................................... 75
4
User’s Manual for Aeros and EasyMatch Essentials v 1.1
INDEX ...................................................................................... 85
TABLE OF FIGURES ...................................................................................... 87
5
User’s Manual for Aeros and EasyMatch Essentials v 1.1
6
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Standard Accessories
• Calibration Box with calibrated white tile, black glass and green diagnostics tile
• Certificate of Traceability
• Power Supply
• Aeros Quick Start Guide
• 12-in and 6-in Sample Dishes
• USB Flash Drive
Power Jack
• The instrument is supplied with a 24 VDC (3.75A) power supply. The power supply is
plugged into the back of the instrument as shown along with the Ethernet port and the USB
port.
Power
Power
Supply
Connection
7
User’s Manual for Aeros and EasyMatch Essentials v 1.1
CAUTION
Note: Use only the power cord included with this instrument or a replacement obtained
from HunterLab. Be certain that the power cord is in good condition before connecting it.
Power Switch
• To turn the instrument on, press the rocker switch on the back of the instrument.
Ethernet Port
Ethernet Port
• This port is used to connect the Aeros to:
▪ Computer or to a network with the purpose of sending data (ASCII) to a server.
▪ Connect with EasyMatch QC and EasyMatch ER Validation and Compliance
▪ Remote Support
▪ Network printer.
8
User’s Manual for Aeros and EasyMatch Essentials v 1.1
9
User’s Manual for Aeros and EasyMatch Essentials v 1.1
10
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• The instrument is automatically connected and this connection is reported on the status bar.
Next, the unit must be Standardized.
• There are two ways to initiate Standardization.
▪ Press the Workspace icon and select Standardization. User will be prompted
to remove all samples from under the sensor.
▪ Alternately, press the Standardization status area on the status bar to initiate
Standardization.
• Move sensor to Top: Press OK to move the sensor to the highest position.
11
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ Read Black Glass: When the sensor stays at the top, attach the standardization box
to the sensor. Then attach the black glass to standardization box and press Read.
▪ Read White Tile: Remove the black glass and attach the calibrated white tile to
standardization box. Press Read to continue.
12
User’s Manual for Aeros and EasyMatch Essentials v 1.1
13
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Note: The active workspace name is displayed in the lower right corner of the screen.
▪ Configure the New Workspace: With this new workspace open, you can start to
change the settings in Color Scales, Read Options, Standard and Tolerances, Views
and View Options. All these changes will be automatically saved in this new
workspace.
▪ Default Settings: Following are the default settings for a new workspace:
Table 1. Workspace Parameters
Parameter Selection
ColorScales CIE L*a*b*
Illuminant D65/10
Indices None
Differences None
Read Options > Options Prompt for Sample Name, Auto Save
Read Options > Measurement Turntable ON, Auto Height selected,
Configuration Measurement time = 5 sec
Standard and Tolerances None
Views Color Data Table only
View Options for Color Data Table Latest Data First selected, Precision = 2
▪ When Auto Height is selected as default, the sensor head will adjust its position
automatically for each sample measurement. Users can also choose to read
samples with the sensor head at a fixed distance. Create Sample Profile is covered
in Workspace > Read Options > Measurement Configuration portion of this
manual.
14
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ Now your instrument is ready to read your product under this new workspace. If
you’d like to start a new job for this product, you can press New Job and load this
configured workspace to continue.
• Main Measurement Screen: The Color Data Table view shows the configured Color Scale
results for the standard and sample measurements in the job. The configured tolerances
can be applied to the Job and Pass/Fail results will also be displayed. To add the color
differences, indices and tolerances to the Color Data Screen, see Workspace > Color Scales
and Workspace > Standard and Tolerances.
• To add a product standard and tolerances, see Standards and Tolerances. To change the
color scale, etc., see Workspace: Color Scales. The setups are saved automatically in
Workspace.
• A long press on the Sample name will show a menu with the following options:
▪ Set as Standard – to set the sample as Standard
▪ Rename – to rename the sample
▪ Delete – to delete the sample.
15
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• A long press on the Standard name will show a menu with the following options:
▪ Edit – to edit the standard. If Edit is selected, the Workspace > Standard and
Tolerances dialog box is presented to allow for editing the name, assigning
tolerances or changing the type of standard.
▪ Delete – to delete the standard. The deleted Standard is reverted into the samples
list with its original name.
16
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Workspace
Moves & System
Sensor Settings
to Top
View
Jobs
Options
View
Flippers
Current
Read Workspace &
Shortcut to
New Job
Sensor Information
Tools: Read
• Switching between Views can be accomplished by using the semi-transparent Next and Prev
buttons placed at the side edges of the screen or by swiping left or right with two fingers on
the screen.
17
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Tools: Information
i
• The sensor type and serial number is shown at the bottom left side of the System Bar when
the i is pressed. When application security is enabled and the user logs into Essentials, the
User Account will also be shown in the Information box.
• This menu shows the configuration options for the active view. A total of six views are
available. Each view shows a different option. Views can be added or removed in
Workspace > Views.
• The Workspace menu sets up the data screen with measurement scales, read options,
standards, tolerances and views.
18
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Systems Settings initiates Standardization, Diagnostics, Preferences, and the User Manager
for System Security.
Tools: Jobs
• A job is a collection of all the sample measurements and a workspace used for a task,
product, or customer. Jobs are the ‘readings’ of EasyMatch Essentials. Jobs can be created
for many different reasons, such as to hold data for a certain customer or a specific product
line. Each operator may maintain their own job with preferences or create separate jobs for
different operations. A workspace is a collection of the measurement parameters for a job
along with tolerances and the standard, i.e. analogous to word processing documents
containing text and formatting. Each job has only one workspace.
19
User’s Manual for Aeros and EasyMatch Essentials v 1.1
20
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Jobs
Jobs vs. Workspace: A job consists of samples measured according to a specific workspace. A
workspace is a template with measurement conditions such as standard & tolerances, Color Scale,
Index, illuminant, etc. There can be only one workspace associated with a job. The main tool bar
provides the options to create a new job, open an existing job and save a job.
Jobs: New
• When click New Job (Shortcut: press Workspace name in the lower right status bar), the
Load Workspace dialog will pop out. The default selected workspace is the current
workspace. User can change the workspace and click Load then the selected workspace is
opened in the new job. Once the Workspace has been loaded to the New Job, the
Workspace name associated with this job cannot be edited.
Jobs: Open
• Open a saved Job: A list of available jobs under the current path are displayed for selection.
If the job that is needed exists in another folder, then it is an option to change the folder.
When the job to be opened is displayed select the file and press Open. We also provide a
short cut for Jobs: Open in main tool bar.
• Shortcut: press job name in the status bar on the lower right.
21
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Jobs: Print
• Print an open Job using the parameters set up under Workspace & Systems Settings >
Preferences.
Figure 20. Workspace & System Settings > Preferences > Print
22
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Drivers included in the Aeros are shown below. Additional printer drivers can be added under
Workspace > Diagnostics > Advanced.
Table 2: Printer Drivers Included
Printer Driver
Canon Canon Print Service 4.4+
HP HP Print Service Plugin 4.1+
Epson Epson Print Enabler 4.4+
Konica Minolta Konica Minolta Print Service Plugin 4.4+
Kyocera Kyocera Print Service Plugin 4.4+
Lexmark Lexmark Print Service Plugin 4.4+
Sharp Sharp Print Service Plugin 4.4+
Xerox Xerox Print Service Plugin 4.4+
• Printing can be downloaded to a pdf file by selecting, Save As PDF. Once this is selected, the
parameters for the output are presented. Please save the file to the download folder.
23
User’s Manual for Aeros and EasyMatch Essentials v 1.1
24
User’s Manual for Aeros and EasyMatch Essentials v 1.1
25
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ Import Job
This option allows the user to browse and select a Job file(s) (.ezm) from the USB
flash drive and import into the instrument. If a file name already exists, then the
name will be incremented numerically.
▪ Import Standard
This option allows the user to browse and select a Standard(s)(extension .std) from
the USB flash drive and import into the database. If required, the Standard Name
can be changed.
▪ Import Workspace
This option allows the user to browse and select a Workspace(s) (extension .wsp)
from the USB flash drive and import into the database. If the workspace already
exists, then the user is prompted to specify a different name.
To use the above functions, a USB flash drive must be present in the port.
▪ Import Diagnostics: This option allows the user to browse and select a Diagnostics
file from the USB flash drive for import into the instrument database.
▪ Import Others: This function is available to import other items such as a logo for the
printed report. Once the logo is imported, go to Workspace & System Settings >
Preferences > Print to add the logo to a report.
26
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Jobs: Data Management: Export. This feature allows the user to export the below data from
instrument into a USB flash drive. Data can be one file or multiple files. All selected files
should be in the same file path location. The following data can be exported:
▪ Job
▪ Standard
▪ Workspace
▪ Diagnostics
▪ Others (e.g all files in Hunterlab folder, and all pdf reports in Download folder)
▪ Export Job: This option allows the user to browse and select an existing Job(s)
(.ezm) or the current active Job data and copy into a USB flash drive either in CSV or
EZM file format. While exporting into EZM format, the current active Workspace
settings can be applied. The color data shown in the Color Data View and the
Spectral Data is saved in a CSV file.
▪ Export Standard
This option allows the user to browse and select an existing Standard(s) in the
database and copy into the USB flash drive as a file (.std).
27
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ Export Workspace
This option allows the user to browse and select an existing Workspace(s) in the
database and copy into the USB flash drive as a file (.wsp). To use the above functions,
a USB flash drive must be present in the port.
▪ Export PDF
This allows the user to select a PDF file from the Downloads folder to export. Switch
to the Download folder in the dropdown list and then select the pdf files to export.
• Jobs: Data Management: Email. Saved Jobs can be emailed if there is an active internet
connection. When the Email option is clicked, the following screen is shown prompting the
user to browse and select a user and enter the recipient mail address. Data can be one file
or multiple files. All selected files for one email should be in the same file path location. You
can email any file in HunterLab folder as well as in the downloads folder.
▪ Mail Settings
Click Mail Settings button to configure the SMTP mail server configuration (Port,
Server) as shown below. The mail settings configuration is mandatory to enable the
mail job feature in the application. When done, press Send.
28
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Jobs: Data Management: Delete. The Delete function will allow deletion of Jobs,
Standards, Workspace, Diagnostics and others. Data can be one file or multiple files. All
selected files must be in the same file path location. In addition, one can delete PDF files
from the Downloads folder.
• Jobs: Data Management: Backup/Restore. The Backup function will copy the entire
Hunter Lab folder to a thumb drive. Restore enables the user to copy the backup folder of a
thumb drive and upload to the Aeros.
Jobs: Help
To access the onboard manual, use Jobs: Help. Novice Help can also be enabled under Preferences>
General.
Jobs: About
The About menu provides information about HunterLab and the current software version.
To update the software version from a USB flash drive, insert the USB flash drive into the port on the
front of the instrument. Open the Jobs > About menu and press Update to continue. After update,
open Essentials and it will prompt to enter or create an Administrator Account. If needed, you can edit
this account in User Manager later.
29
User’s Manual for Aeros and EasyMatch Essentials v 1.1
For detailed information on firmware and more, please press the Info button on the screen.
30
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Under the Workspace & Systems Settings, the following tasks can be accomplished:
• The Scales Tab showsthe five scales available for measurement. Select the absolute scale or
color difference scales (if a standard is selected). Press Apply and begin to read your
samples.
31
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• The Illuminant/Observer tab displays combination selections for these parameters. To see
all of the choices, you can scroll through the selections by viewing the screen.
• To select indices, go to the Indices Tab and check the corresponding box on the right side.
Multiple selections are avaible. To see more choices, the screen can be scrolled. Press
Apply to continue.
• To select differences, go to the Differences Tab and check the corresponding box on the
right side. Press Apply to continue.
32
User’s Manual for Aeros and EasyMatch Essentials v 1.1
2/10 Zoom
Average2
Std. Deviation2
Not Available on Color Plot, 2Trend Plot Only, 3Color Data Table Only
1
33
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ Average
Select the number of readings to average to produce the final measurement. The
total number of readings to be averaged can be no less than two. Press Apply to
Once the Read button is pressed, the instrument will display a unique dialog box to
Read and Average the readings. The second reading is taken using the dialog box
button, Read. Once all the readings are taken, press Average to obtain the results.
Average and Continuous Read are mutually exclusive.
▪ Continuous Read Interval
This feature performs measurements continuously. In Continuous Read mode,
measurements are initiated and stopped using the Read Button. The minimum
value of the Read interval is 3 seconds and it will read as fast as it can update.
When in Continuous Read mode, the Read Button is enhanced with a checkmark.
When taking measurements, the Read button is greyed out. When waiting to take
the next measurement, the Read button turns green.
34
User’s Manual for Aeros and EasyMatch Essentials v 1.1
To stop the Continuous Read, press the Read button when the button is green..
▪ Auto Save Job
This selection will automatically save a job. Once this feature is selected, a dialog
box will be displayed to name the job. If there is no name for a job yet, the file
name will be default with the date, time, instrument and workspace.
1. One standard data point: In this case, the single data point is compared to
the expected value. The Gain remains at 1.0 and the Bias is corrected:
Bias= Expected Value- Measured Value
2. Two data points: In this case, the two readings are compared to the
expected values.
Bias Correction=Expected Value 1-(Measured Value 1*Gain)
35
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Select Index
Bias
Configuration
The indices with bias correction include: 457nm Brightness, BCU, HCCI,
SCAA, Tint E313, WI E313, Y Brightness, YI D195, YI E313, Z%.
36
User’s Manual for Aeros and EasyMatch Essentials v 1.1
37
User’s Manual for Aeros and EasyMatch Essentials v 1.1
38
User’s Manual for Aeros and EasyMatch Essentials v 1.1
When profile is complete, press OK to continue. The sample height will be shown
next to the Create Sample Profile button.
The Measurement Time is the amount of time used to average the readings
together. The longer the time, the more the averaging.
• Standard is saved with standard name, standard color value and standard tolerances to the
database by pressing the button at the bottom of the screen. When there is a standard
applied in a job, you must delete it first if you want to change the standard type (Recall.
Physical/Adhoc and Numeric). You can click the Calc Auto Tolerances here to calculate the
tolerances of standards.
• Tolerances can be entered manually for a selected scale, index and difference.
39
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Tolerances will be displayed on the measurement screen if enabled under the View Options
for the Color Data and the Color Plot Screens.
• Pass/Fail based on these tolerances can be used on the EZ View as well as Color Data View.
• Hitch Standardization
❖ Hitch Standardization is a process by which two or more instruments of similar
design can be made to read the same color values on a group of specimens. This
process is very useful in expanding the communications of color around the world or
between vendor and customer.
❖ The process of Hitch Standardization (also known as transfer standardization)
involves assigning one instrument to be the reference, or master, unit and
mathematically adjusting the secondary, or slave, unit(s) to read the “same” values.
In this way, two or more instruments can be hitched together. Hitching a secondary
unit to a reference instrument requires that a specimen be read on both units and
the values compared and adjusted accordingly. This specimen, known as the hitch
standard, is first read on the reference instrument and its values recorded as
spectral data or colorimetric (tristimulus) data. The hitch standard is then physically
moved to the secondary instrument where it is reread and the values from the
reference unit are input into the secondary instrument’s processor.
❖ Steps for Hitch Standardization:
1. Read a standard.
2. Go to Workspace > Standard and Tolerances and select Hitch.
40
User’s Manual for Aeros and EasyMatch Essentials v 1.1
AutoTolerances Setting
▪ Using Tab 4, AutoTolerances are calculated for a Color Scale using CMC. The default
values of l:C – 2:1 with auto correction factor = 0.75 and commercial factor = 1.
However, these ratios can be modified as needed.
41
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Note: If AutoTolerances are selected, the user cannot manually enter tolerances.
Tolerances
▪ Tolerances can be entered for Scales, Indices and Differences.
42
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Workspace: Views
• Views: This option can be used to select the views to be presented in the application.
Simply check on the box of the screen needed. Press Apply to save one or all of the screens.
The default screen is the Color Data Table. To navigate between screens once the
selections have been applied, use the View Flippers on the left and right of the screen.
43
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ Read Black Glass: Attach the standardization box to the sensor. Then attach the black glass
and press READ to continue.
▪ Read White Tile: Remove the black glass and attach the calibrated white tile to the
standardization box. Press READ to continue.
Figure 65. Change to White Tile
44
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ Remove the calibration box when standardization is complete. The instrument is ready to
Read Samples.
45
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Leave the white tile at the port and press OK to initiate the repeatability
readings. The white tile is read 30 times and the individual results reported. A
table of the difference between the current reading and Standard is shown after
every measurement. By comparing each reading to the tolerance, a Pass/Fail
assessment is shown.
46
User’s Manual for Aeros and EasyMatch Essentials v 1.1
When all 30 readings have been made, the final test result is shown and saved
automatically. To print the results, press the print button or click to Open the file
and then Print.
• Once the target values have been entered, press Next. Standardize the
instrument and attach the Green Tile. Press OK to continue.
47
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Ten readings are taken and compared to the tolerance as an average. This test
is then automatically saved and can be printed by pressing Print.
Advanced Tests
Advanced Tests are primarily for use by HunterLab’s Service Department. However, this menu
supports the upload of Printer Drivers, Parking of the sensor for Shipping, and remote support
setting.
▪ Sensor Head Position: To test the movement of the sensor head up or down. To
prepare for shipping, use the Park for Shipping command under System.
▪ Read Signal, Dark, Zero: This function will enable the Service Department to
determine proper performance of the instrument.
48
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ The Turntable can be tested next by pressing Start. Press again to stop the
turntable.
▪ Enable Log: Once check enabled, this feature records the instrument actions for
tracking purposes. When complete the user returns to this screen and exports the
data to a thumb drive. Once the data export is completed the data size becomes ‘0’
again.
▪ Standardize: Initiates standardization from the Diagnostics screen.
▪ Measure: Initiates the measurement of a sample from the Diagnostics screen. The
reflectance spectra data will be showed in this dialog.
▪ Park for Shipping: Move sensor down to secures the optics for transportation.
▪ Printer Driver: To upload a new print driver or apk file, download the apk file
needed from the internet onto a flash drive. Place the flash drive into the
instrument (front port) so that it can access the list of apk files. Select the driver to
upload and press OK.
49
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• The Aeros will install the new printer driver and it will be then available to use.
50
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Restart Comm can be used to reset the ethernet communications for EasyMatch
QC.
• Support Region
Note: Your instrument must be connected to the internet.
a. Select Support Region. A dialog is displayed with three options for region
selection. Select USA, Europe or Asia Pacific and the URL string is modified
accordingly. Press OK to continue.
b. Select Restart Support to view the Netops Host Screen. From the top right
side of the Netops Host screen, select the 3 dots. From the list menu, select
Restart.
51
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Exit Arrow
c. To ensure that your application is successfully restarted, make sure that you
see the message Webconnect: ‘Hunterlabs’ Online. If this message does not
appear, please contact our support team. To exit press the arrow at the
bottom of the screen.
52
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ To Load the Last Workspace at Startup check this box and press Apply.
▪ To Load the Last Job at Startup, check this box and press Apply.
▪ The Standardization Time Interval is a useful reminder to restandardize. Press
Apply to set the new interval. When the time has lapsed, a prompt to restandardize
will be displayed before measurements can be taken.
▪ Set the screen Brightness using a sliding scale and press Apply.
▪ Set the Date and Time, time zone, and format use the Adjust Clock feature.
▪ Enable Novice Tooltips by checking on the box. Once enabled, on screen tips are
displayed for 3 seconds. To display again, roll over the lightbulb icon on the lower
right part of the screen.
53
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ Enable Application Security. This selection is available after the User Manager
has been set up. Please refer to the System Settings: User Manager for more
information.
▪ When this is selected, the application will require valid login credentials at
startup. On successful login, the user name will be shown in the status bar. If
‘Use Last Login Credentials’ is checked, the user will be automatically logged in
on subsequent startups.
▪ To Configure and Enable the Network Data Export and Network Settings,
please connect the Aeros to a computer as described in Special Functions
(Chapter 7). In Network settings, one can use an Ethernet cable or setup a Wifi
Connection.
Preferences: Print
The Print page allows the user to configure:
▪ The Readings to print
▪ The option to Preview before print.
▪ Print report Title and Logo. To apply a logo, import the logo first to the
HunterLab folder.
▪ Orientation of the report (Portrait or Landscape) orientation.
▪ To save changes, press Apply.
54
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Once the groups have been established, then individual users with User Names and
Passwords can be setup for both Administrator and User Groups.
• Users in Administrative Groups have all features enabled. For Users in User Groups,
Privileges can be setup as shown below. Press Update Profile to complete.
55
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• After enabling security, each user must enter a name and password when logging into the
Aeros. For convenience, the user can check the box under Workspace > Preferences>
General to use the ‘last login credentials’.
56
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• If needed, the administrative user can delete groups / users and reset passwords of all
Groups & Users.
57
User’s Manual for Aeros and EasyMatch Essentials v 1.1
58
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Views are selected using a dialog box under Workspace. Simply check on the box of the screen needed.
Press Apply to save one or all of the screens. The default screen is the Color Data Table. To navigate
between screens once the selections have been applied, use the View Flippers on the left and right of
the screen.
Once the views have been selected, then View Options are available to add additional information to
the screen. Each screen has a unique set of options associated with it.
Views: EZ View
• This view provides a simple display of Standard vs. Sample and Pass/Fail results.
❖ View Options includes the selection of No Color Scale, Pass/Fail, Precision and
Font Size.
59
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Options such as Tolerances, Data Order, Interval, Date, Distance, Turntable Movement,
Height, Time and Pass/Fail can be selected for viewing using the View Options.
60
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• A long press on the Sample can enable the user to make the sample into a Standard, change
the name or delete the reading.
❖ To delete a Sample (or Standard), select Delete and then confirm the action.
• A long press on the Standard will enable the user to edit or delete the Standard. Edit opens
the Standard and Tolerances dialog box. Delete will delete the standard from the current
workspace.
61
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Selections include Absolute or Difference, Reflectance or K/S. Enter the start and stop
wavelength, the interval and the precision and press OK to continue.
62
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Press Clear All to remove all the samples to display. Press Select All to enable display of all
samples. To select an individual sample, click on the respective Sample in the list located on
the right edge of the screen.
• The Samples List is paginated. Click the left and right arrow buttons below the samples list
to navigate between pages.
• Press and hold on the left/right page number arrows under the sample list to show a small
dialog box. This dialog allows you to select the number of records per page to display and
the page number to display.
• Spectral Plot Options: There are three choices for spectral plot options:
❖ K/S – mathematical calculation based on reflectance and determined at each
wavelength for the standard and sample.
❖ Reflectance - Displays the reflectance value at each wavelength.
❖ % Strength – Percentage of the ratio of the K/S of the sample to the K/S of the
standard.
• Uncheck the Options: Show Background, to display the plot with white background color.
63
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Check Options: Auto Range to automatically scale the contents to fit. If Auto Range is not
selected, then enter the Y- and X-axis range to display.
▪ View Options for the Trend Plot include the type of display, the statistics and the
number of readings per display.
64
User’s Manual for Aeros and EasyMatch Essentials v 1.1
▪ View Options:Traces set the ranges for the traces or allow selection of Auto Range.
Trace 1 to 3 uses the current Color Measurement Scale and Trace 4 will allow for
measurement of differences or an index. The user can select which Traces to view
and set control limits as a percent.
65
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• The tolerance plot is available in rectangular and elliptical color space. The Pass/Fail sample
points are shown in green and red when in difference mode, respectively. In Absolute
Mode, they are shown in green.
66
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Special Functions
67
User’s Manual for Aeros and EasyMatch Essentials v 1.1
68
User’s Manual for Aeros and EasyMatch Essentials v 1.1
69
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• Select Configure WiFi Settings and the WiFi configuration dialog will be prompted.
• Please search and connect to the available WiFi and write down the IP address
showing in this dialog. After the WiFi configuration, please click the floating Back
Button to go back to Essentials app.
• Go back to Workspaces > Preferences > Auto Export Measurement and check
‘Config’. Choose Act as Server and Port number as 11111. You can also choose a
delimiter to mark your data.
70
User’s Manual for Aeros and EasyMatch Essentials v 1.1
71
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Ethernet Port
❖ Plug the other end into the Computer or into the Ethernet adapter
• Open Command Prompt in the PC.
Type in ipconfig, find the right ethernet (in this case, it is Ethernet adapter Ethernet) and
write down autoconfiguration IPv4 Address as well as the Subnet Mask.
72
User’s Manual for Aeros and EasyMatch Essentials v 1.1
❖ Press Apply on the Ethernet Configuration and then Apply on the Preferences Page
to complete.
❖ Turn the instrument off and then back on.
❖ Go to Preferences and select Auto Network Data Export.
❖ For a direct connection between Aeros and data collection computer, set up the
Aeros as a Client.
❖ Enter the computer IP address here, in this case 169.254.113.144 and the Port as
10001. Press Apply on the screen to continue.
73
User’s Manual for Aeros and EasyMatch Essentials v 1.1
• If the user answers ‘YES, all measurements are recovered into a new job or appended to a
saved job.
74
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Specifications
The specifications and characteristics of the instrument are given in this chapter. For best performance,
the instrument should be placed where there is ample work space with medium or subdued illumination
and no drafts. The operating conditions (temperature and humidity ranges) are given in the Operating
Conditions section below.
Note: Do not leave Aeros in an area where temperature or humidity extremes are
possible.
Operating Conditions
Physical Characteristics
Weight 23.0 kg (50.0 lbs)
Dimensions 56 cm x 38 cm x 51 cm
(Height x width x depth) (22.0 in x 15 in x 20 in)
Maximum Sample Height 65 – 140 mm (2.5 – 5.5 in)
Communications I/O:
USB Front & Rear Panel Connectivity to printer,
keyboard, mouse and other
peripherals;
Ethernet RJ45 Print directly to standalone or
network printers; email directly
from the instrument; stream data
to LIMS and SPC systems;
Remote Access Support Enabled
System Power 100 – 240 VAC, 47 – 63 Hz to
universal power supply @ 24
VDC/3.75A
75
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Instrument Performance
Range: 400-700 nm
Spectral Data
Reporting Interval (nm): 10 nm
Spectral Resolution <3 nm
Reporting Interval 10 nm
Continuous at 7 measurements
Sampling Rate
per second
Photometric Range 0-150%
Measurement Duration 5 seconds (1 rotation)
Measurements per Rotation 35
Inter-instrument Agreement on ∆E*L*a*b* < 0.30 (Avg)
BCRA II Tile Set ∆E* (L*a*b*) < 0.50 (Max) on
Colorimetric Repeatability ∆E*< 0.025 Max on White Tile
Measurement
Color Data, Spectral Data, Spectral
Data Views Plot, EZ View, Tristimulus Color Plot,
Trend Plot
A, C, D50, D55, D65, D75, F02, F07,
Illuminants
F11, TL84, ULT30, ULT35
Observers 2° and 10°
CIE L*a*b*, Hunter Lab, CIE L*C*h,
Color Scales
CIE Yxy, CIE XYZ and differences
Color Difference Indices ∆E*, ∆E, ∆E CMC, ∆E 2000
E313 Yellowness, E313 Whiteness, YI
D1925, Y Brightness, Z%, 457nm
Indices and Metrics
Brightness, Baking Contrast Units,
Tint, HCCI, SCCA
Data Storage 500,000 Records max
76
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Regulatory Notice
77
User’s Manual for Aeros and EasyMatch Essentials v 1.1
78
User’s Manual for Aeros and EasyMatch Essentials v 1.1
79
User’s Manual for Aeros and EasyMatch Essentials v 1.1
80
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Warranty
HunterLab warrants that all instruments it manufactures are free from defects in material and
workmanship under normal use. This warranty is limited to repairing or replacing any defective
hardware or software that may cause the instrument to perform outside of its specified tolerances.
This warranty is one year from date of shipment of new instruments and two months from the date of
shipment of repaired instruments.
Note: Printers and computers are covered under the original manufacturer’s warranty.
The warranty is void if the user has made unauthorized repairs, improperly installation, operated, or
subjected the instrument to conditions outside of the operating conditions specified in the product
documentation.
The HunterLab warranty does not cover consumable items such as lamps, fuses, batteries, etc. An
instrument registration card is available online at https://www.hunterlab.com/ras-registration.html.
It is important that the instrument owner fills this out on receipt of equipment.
Questions concerning operation, maintenance, or repair of your equipment can be directed to the
Service Department at Service@hunterlab.com. Additional information can be obtained at
http://support.hunterlab.com.
Shipping Claims
All materials are sold F.O.B. from Reston, Virginia (unless otherwise specified) and HunterLab
responsibility ends upon delivery to the first carrier. All claims for loss or damage must be rendered by
the consignee against the carrier within fifteen days of receipt of goods. A copy of this notice must also
be forwarded to HunterLab within five days of its receipt.
• Breakage or Damage
Per the contract terms and conditions of the carrier, the responsibility of the shipper ends at
the time and place of shipment. The carrier then assumes full responsibility. Perform the
following procedures in the case that your instrument arrives broken or damaged.
• Freight or Express
❖ Notify your local carrier.
❖ Hold the damaged goods with their container and packaging for inspection
by the examining agent. Do not return any goods to HunterLab prior to
inspection by and authorization of the carrier.
81
User’s Manual for Aeros and EasyMatch Essentials v 1.1
❖ File a claim against the carrier. Substantiate this claim with the examining
agent’s report. A certified copy of our invoice is available upon request.
The original B/L is attached to our original invoice. If the shipment is
prepaid, write for a receipted transportation bill.
❖ Advise HunterLab regarding replacement.
Shortage
• Perform the following procedure if your order appears to be missing items.
▪ Check the packing list notations. The apparent shortage may be a backordered item
and may be marked as an intentional short-ship.
▪ Re-inspect the container and packing material, particularly to locate smaller items.
▪ Ascertain that the item was not removed by unauthorized personnel prior to
complete unpacking and checking.
▪ Notify HunterLab immediately of the shortage in writing.
Incorrect Shipment
• Perform the following procedure if material received does not correspond with your order.
▪ Notify HunterLab immediately, referencing your order number and item.
▪ Hold incorrect items until return shipping instructions are received.
82
User’s Manual for Aeros and EasyMatch Essentials v 1.1
HunterLab offers complete repair services for all instruments it manufactures. Call HunterLab for the
service facility nearest your location. If your equipment is not functioning properly, contact HunterLab
Service for maintenance or repair instructions. Many times, this on-the-spot diagnosis is all that is
required.
If repair is required, the instrument may be returned to a HunterLab service facility. For schedule and
terms for repairs, call HunterLab Service. Please read the next section, “When You Need Assistance,”
prior to contacting HunterLab.
Customers are responsible for incoming and outgoing freight charges for instrument returned to
HunterLab for all repairs, including warranty repairs.
83
User’s Manual for Aeros and EasyMatch Essentials v 1.1
84
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Index
85
User’s Manual for Aeros and EasyMatch Essentials v 1.1
86
User’s Manual for Aeros and EasyMatch Essentials v 1.1
Table of Figures
Figure 1. Calibration Box 7
Figure 2. Rear View with Connectors 7
Figure 3. USB Port on Rear of Instrument 8
Figure 4. Measurement Screen 11
Figure 5. Sensor Moves to Top 12
Figure 6. Read the Black Glass for Bottom-of-Scale 12
Figure 7. Read the White Tile for Top-of-Scale 12
Figure 8. Completed Standardization 13
Figure 9. Standardization Status Bar 13
Figure 10. New Workspace 13
Figure 11. Name New Workspace 14
Figure 12. Read Options: Measurement Configuration 15
Figure 13. Changing, Renaming or Deleting a Sample 15
Figure 14. Edit/Delete a Standard 16
Figure 15. User Interface Screen for Aeros & Essentials 17
Figure 16. Sensor Serial Number 18
Figure 17. Job Menu 21
Figure 18. Open A Job 22
Figure 19. Save A Job 22
Figure 20. Workspace & System Settings > Preferences > Print 22
Figure 21. Save as PDF 23
Figure 22. Save PDF to Downloads File 23
Figure 23. Download File Contents 24
Figure 24. Data Management Menu 24
Figure 25. Recall Measurements 25
Figure 26. Import Job 26
Figure 27. Import Workspace 26
Figure 28. Export Current Job 27
Figure 29. Export Standard 27
Figure 30. Select Download Folder for PDF File Export 28
Figure 31. Enter an Address to Email a Job 28
Figure 32. Enter SMTP Mail Server Information 29
Figure 33. Job > About the Software 29
Figure 34. Instrument Info 30
Figure 35. Workspace Parameters 31
Figure 36. Color Measurement Scales 31
Figure 37. Illuminant/Observer Configuration 32
Figure 38. Index Configuration 32
Figure 39. Color Measurement Differences 32
Figure 40. Read Options 34
Figure 41. Reading and Averaging 34
Figure 42. Continuous Read 35
Figure 43. Auto Save Job 35
Figure 44. Slope & Bias Correction 36
Figure 45. Input Gain & Bias 36
Figure 46. Prompt for Sample (Standard) Name 37
Figure 47. Input Sample Name 37
Figure 48. Measurement Configuration 38
Figure 49. Creating a Sample Profile 38
87
User’s Manual for Aeros and EasyMatch Essentials v 1.1
88
User’s Manual for Aeros and EasyMatch Essentials v 1.1
89