Service Manual-2017.07.19 PKL - PPC - 125
Service Manual-2017.07.19 PKL - PPC - 125
Service Manual-2017.07.19 PKL - PPC - 125
1.2
Take out the accessory,we can check it according to the packing list.
1.3
Open the front panel of wooden box.take out the remained accessory.it is also in the packing list.
1.4
There are 2 screws which are use to fix the analyzer from the bottom.
So in the next step,here is the operation.
A:Lean the wooden box, loose the 2 screws at the button ,please check the picture below.
1.8 After remove the proctecting materal, instrument was placed like this.
2. Instrument introduction
2.1 Top side view
2.2 Left side view
3.Software Installation
3.0 Paste the software into D/E /F Disk.
3.1 Register the software
a:go to the document folder. Click setupfile as the picture.
c:double click the Analyzer_200.exe and run the software as the picture below.
3.1-Software Registration
3.1.1Run the software, for the first time enter. just input Password [ADmin], do not need check
Login ID.
3.1.7 The password number comes from the computer time. for example my computer time
is 2016-7-12(year2016,July,12th).the principle is plus year(2016)+month(7)+date(12) together.so
the password is [2035]
After we type in the password 2035,we go to the inerface as below.
3.1.8 Input ‘Computer ID’ and select the [Start date],set [Valid dates] (how many days can
use), press Code, and [SAVE], then press [Cancel] to return the main screen.
Click the icon of CO again.input the password :8888 . then we can choose the system to open
or close. Remember to save after you choose open.
3.1.9 Back to [MAINTENANCE]—[Parameter setting], you will found the code and days.
Adjustment
1.basic operation
1.make sure the signal cable and power cable are well connected then turn on the analyzer.
5.then go back to icon of initializen to initialize the analyzer.the machnical parts will go to the
related correct position.
3.Wash Arm adjustment
A:check and adjust wash arm position.
Click the icon to CRU water dispense then you could see the water station goes down to the
surface of cuvette. And please check the position to the wash needle ,they must be in the
middle position of the cuvette.
1. Then we click the icon CRU Down and start to make the wash station go to the bottom.
2. we loose all the fixing screws of the washing arm the seven needles will touch the cuvette
bottom .
3.fixing the screws for the third to seventh pair of the washing arm.
4.decrease the parameter for CRU Down . for example on step 1 the parameter is 611. Now we
change it to 610. Then save the parameters , and re-click CRU Down, and start .
4.Stirl Adjustment
2.check and adjust stir needle position.
Click cuvette icon and start then stiring needle will go to the top of cuvette 1 as the picture
showed.
Click the icon of stiring deep .the stir needle will go to the bottom side of the cuvette.
Mostly 2 step to the bottom side.
5.Sample Needle Adjustment
When Analyzer was delivered to customer .there is no need to install the halogen lamp for the
lamp was well fixed inside.
Secondly light source of the halogen bulb must face the cuvette.
Otherwise the light signal will be weakened. (please tighten the screw after you insert the
halogen bulb).
Open front cover, then check the points of Amplifier board. Use DC mv of multimeter, the voltage
should be 2mv~5mv.Factory adjust to 3mv.
If it is out of range. We can adjust it according to the potentiometer.
run ABS check . all result should be 35000~45000. (this value with new lamp and new cuvettes,
after used, the accept value is 25000~55000,and the best value is 30000~50000)
8.Parameters Setting
9.Inner structure
1. loose all the screws of the analyzer ,and take out the cover .
2.secondly. loose the air plug button once we loose this button, the pressure of the tubes can be
decreased. Then we are able to remove the tubes.
3.take out the tray
6.unplug the tubes which connect between the 2 peltier cooling holder.
7.unplug the tube which connect to the copper tube cooling fan.
8.take out 2 power cable
17.after changing the peltier, we put back the sample reagent holder and tighten the
screws ,tubes ,the power cables.
18. then we can adjust the sample and reagent mechanical position precisely.
Appendix 1 Level sensor board sensitivity