DF52 Manual Servicio Tecnico B
DF52 Manual Servicio Tecnico B
DF52 Manual Servicio Tecnico B
syringe with the silicone tube to Connector 2. Pull out Tube P6 from the L-type connector at the
top end of Connector 1. Drain the water from the sheath flow device with a syringe and pull out
Tube P21 and P4c from the Connector 3 and 4 at the bottom of the flow chamber.
Figure 2-51 Pulling out the tubing
1 - Tube P6 2 - Tube P1
60
2 Instrument Configuration
Installation
Follow the corresponding disassembly steps in reverse order.
l Put on antistatic gloves and an electrostatic-shielding ring while operating to avoid damaging the
LED.
l All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
l During transportation and installation, the flow chamber exposed at the bottom of the optical
system needs to be protected from being crushed or pressed.When pulling out the tubes, hold
the flow chamber by hand to prevent it from being dislocated by external force.
Purpose
The optical preamp panel can be disassembled and replaced by following the procedures specified
in this section; however, in each case, Preliminary Steps need to be taken before you begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l New optical preamplifier panel
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right-side door and disassemble the top cover panel.
4. Take down the cover panel of the optical system box.
Disassembly
1. Pull out the three wires from the optical preamp panel.
2. Use a #2 (Ph2) Phillips screwdriver to disassemble the four M3×8 cross-recessed pan-head
combination screws which affix the optical preamp panel, and carefully detach the cover panel of
the optical system box.
61
2 Instrument Configuration
Figure 2-53 Taking down the cover panel of the optical system box
Installation
Follow the corresponding disassembly steps in reverse order.
l Put on antistatic gloves and an electrostatic-shielding ring while operating to avoid damaging the
LED.
l All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
Purpose
The laser driver panel can be disassembled and replaced by following the procedures specified in
this section; however, in each case, Preliminary Steps need to be taken before you begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l 2.5mm hex-socket screwdriver
l New laser driver panel
Preliminary Steps
1. Turn off the power switch on the left side of the device.
62
2 Instrument Configuration
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right-side door and disassemble the top cover panel.
Disassembly
1. Use a 2.5mm hex-socket screwdriver to disassemble the four M3×6 stainless-steel inner hex
screws that affix the baffle plate of the laser driver panel.
2. Pull out the two wires from the laser driver panel.
3. Pull out the two wires from the laser driver panel. Use a #2 (Ph2) Phillips screwdriver to
disassemble the four M3×8 cross-recessed pan-head combination screws, and carefully detach
the laser driver panel.
Figure 2-54 Replacing the Laser Driver Panel
1 - M3×6 stainless-steel inner hex 2 - Baffle plate of the laser driver panel
screw (×4)
Installation
Follow the corresponding disassembly steps in reverse order.
63
2 Instrument Configuration
l Put on antistatic gloves and an electrostatic-shielding ring while operating to avoid damaging the
LED.
l All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
Purpose
The fine tuning of the flow chamber can be performed by following the procedures specified in this
section; however, in each case, Preliminary Steps need to be taken before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Preliminary Steps
l Open the right-side door and disassemble the top cover panel.
l Take down the cover panel of the optical system box.
l Turn on the device and launch the software application; then enter the Optical Tuning screen.
Tuning Procedures
1. Use a #2 (Ph2) Phillips screwdriver to loosen the two M3×8 cross-recessed pan-head
combination screws that affix the locking plates of the flow chamber.
Figure 22-55 Fine tuning of the flow chamber
64
2 Instrument Configuration
testing.
Ø If the resultant scattergram shows a square or diamond with the upper and lower sides
parallel to each other, the device does not need fine-tuning.
Ø If the scattergram shows that the upper and lower sides are not parallel, then take the next
step.
3. If the slanting directional arrow shown in the scattergram is ↗, then rotate the adjustment knob
of the flow chamber counter-clockwise; if the slanting direction shown in the scattergram is ↘,
then rotate the adjustment knob of the flow chamber clockwise (the adjustment knob should be
lightly rotated).
4. Use the 7um nominal particles as a sample for DIFF testing.
Ø If the resultant scattergram shows a square or diamond with the upper and lower sides
parallel to each other, the device does not need fine-tuning.
Ø If the scattergram shows that the upper and lower sides are not parallel, then repeat Step 3.
5. After tuning, tighten the two locking screws of the flow chamber, and reinstall the cover panel of
the optical system box as well as the four affixing screws; then lock them in place.
Purpose
The high-angle receiving plate can be disassembled and replaced by following the procedures
specified in this section; however, in each case, Preliminary Steps need to be taken before you
begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l 2.5mm hex-socket screwdriver
l New high-angle receiving plate
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right-side door and disassemble the top cover panel.
Disassembly
1. Disassemble 2 M3 ×8 stainless steel internal hexagon screws used for fixing the fixed seat of the
side panel card by 2.5mm internal hexagonal wrench, and then remove the fixed seat from the
side basal plate.
2. Pull out 2 connecting wires on the high-angle receiving plate.
3. Disassemble 2 M2.5 ×6 cross recess head screws used for fixing the high-angle receiving plate
by #2 (Ph2) Philips screwdriver, and remove the laser driver plate carefully.
65
2 Instrument Configuration
1 - M3×6 inner hex screw (×2) 2 - Fixed seat of side panel card
66
2 Instrument Configuration
There are only sporadic The flow chamber Check if the collecting corner plate has two
background scattergram is dislocated. bilaterally symmetric light spots in the shape of red
signals when testing the The hydraulics vertical stripes.
DIFF signal of venous have no sheath l If there is only one red vertical stripe or the
blood. flow. two stripes are significantly different in
brightness, make small adjustment to the
location of the flow chamber and return the
nominal-particle scattergram to normal.
l If there are two red vertical stripes which are
not significantly different in brightness, check
if the pinch valve is working; if it is, check if
Valve 10, 11, 12, 13, and 18 are also working.
The neutral-particle signal The flow chamber Make a small adjustment to the location of the flow
(the long horizontal stripe is dislocated. chamber and return the nominal-particle
at the top right corner) in scattergram to normal.
the scattergram is
considerably slanted and
the three signals are
mostly overlapping with
inaccurate classification
when testing the DIFF
signal of fresh venous
blood.
There is a lot of noise The reagent is l If the background noise has a certain shape
when testing the contaminated. and is concentrated at the lower half of the
background signal. Tiny air bubbles scattergram, keep the DIFF1 and DIFF2
get into the fluids at room temperature and perform three
hydraulics. rounds of the reagent replacement sequence
each for the diluent, DIFF1 and DIFF2.
l If the background signal has no shape and
randomly spreads all over the scattergram,
replace the sheath flow syringe.
If the problem persists, replace the optical
assembly.
67
3 Repairs
3 Repairs
3.1 Overview
The repair engineer can repair the analyzer using standard tools. See the following section for
repairing procedures (including the tools needed). If any repairing step requires a validating step, the
repair engineer should strictly follow the procedure and take the validating step.
When following the repairing procedure as specified in this section, the operator should put on
rubber gloves and clean both hands with disinfectant after performing any repair work.
Purpose
While performing regular machine maintenance and inspection, the left-side door needs to be
opened to disassemble the sample syringe assembly, reagent testing panel, or electromagnetic
valve assembly, which are all in the left section of the machine.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Opening
1. As shown in Figure 3-1, use #2 (Ph2) cross screw driver to take out four M3×8 cross recessed
pan-head combination screws on the back panel, and take off the back panel.
68
3 Repairs
2. As shown in Figure 3-2, use #2 (Ph2) cross screw driver to take out three M3×8 cross recessed
pan-head combination screws on the left side panel, and take off the left side panel.
Figure 3-2 Disassembling the left side panel
69
3 Repairs
Purpose
While performing machine maintenance and inspection, the right side door needs to be opened to
disassemble the preheating bath assembly, pinch valve, sampling assembly, liquid pump assembly,
impedance counting reaction bath assembly (WBC & RBC), negative-pressure chamber, or
electromagnetic valve & pump assembly, which are all in the right section of the machine.
Tools/Spare Parts
Slot-type screwdriver
Opening
As shown in Figure 3-3, insert the slot-type screwdriver into the slot in the right side door lock and
rotate 90° counterclockwise, then manually open the right side door.
Figure 3-3 Opening the right side door
Purpose
While performing regular machine maintenance and inspection, the panel cover needs to be taken
down for disassembling the sample probe, cleaning swab, sampling assembly, or indicator light
panel.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
70
3 Repairs
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. The left side panel has been disassembled. See 3.2.1 Disassembling the Left Side Panel.
3. The right-side door is opened. See 3.2.2 Open the Right Side Door.
Disassembly
1. Disassembly the back panel: use a #2 (Ph2) Phillips screwdriver to disassemble the three M3×6
cross-recessed pan-head combination screw on the back panel, and take off the back panel, as
shown in Figure 3-4.
Figure 3-4 Disassembling the back panel
2. Diassembling top cover: Open the right side door and disassemble 4 M3×8 cross-recessed
countersunk-head screws on both sides of top cover by #2 (Ph2) Philips screwdriver, the top
cover can be removed, as shown in Figure 3-5.
71
3 Repairs
When removing the panel cover, make sure to pull out the wires connected to the display circuit
panel on the panel cover.
72
3 Repairs
Purpose
The display screen can be disassembled and replaced by following the procedures specified in this
section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you
begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l Display screen
Disassembly
1. Pull out both interfaces at the shield panel of the LCD screen interface panel, use #2 (Ph2) cross
screw driver to remove four M3×8 cross recessed pan-head combination screws on the shield
panel, take off the shield panel, then detach the black buckles on both ends, gently pull out the
golden ribbon cable, as shown in Figure 3-7.
Figure 3-7 Replacement of display screen (1)
73
3 Repairs
2. Use #2 (Ph2) cross screw driver to remove eight M3×8 cross recessed pan-head combination
screws on the screen support, take off the screen support, display screen and touch screen
respectively, as shown in Figure 3-8.
Figure 3-8 Replacement of display screen (2)
Purpose
The sample probe can be disassembled and replaced by following the procedures specified in this
section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you
begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l Sample probe
Disassembly
1. Pull out the swab's snap ring to detach it from the lower bracket panel. See Figure 3-9.
74
3 Repairs
75
3 Repairs
Purpose
The optocoupler can be disassembled and replaced by following the procedures specified in this
section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you
begin.
Tools/Spare Parts
l 2.5mm hex-socket screwdriver
l Optocoupler
Disassembly
Use the 2.5mm hex wrench to remove the M3 hex screw affixing the optocoupler, then pull out the
plug attached to the optocoupler wire to disassemble the optocoupler to be replaced. See Figure
3-11.
Figure 3-11 Replacement of optocoupler
Purpose
The sampling assembly can be disassembled and replaced by following the procedures specified in
this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before
you begin.
76
3 Repairs
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l diagonal cutting pliers
Disassembly
1. Use the diagonal cutting pliers to cut off all the nylon binding tapes affixing the fluid tubes
(aspiration tubes of the sample probe and swab tubes), then pull out the fluid tubes of the sample
probe and the swab to detach them from the sampling assembly.
2. Pull out all the motor wires and optocoupler plug from the horizontal moving module of the
sample assembly, and use a #2 (Ph2) Phillips screwdriver to take out the three M3×6 cross
recessed countersunk head screws affixing the bracket for drag chains to detach the
sampling-specific drag chain from the sampling assembly. See Figure 3-12.
Figure 3-12 Replacement of the sampling assembly in X- or Y-direction (1)
77
3 Repairs
Tools/Spare Parts
l Phillips screwdriver
l Power assembly with the same specifications
Disassembly
1. Use Phillips screwdriver to take out four M3×8 cross recessed pan-head combination screws on
the back of the device power supply assembly, then remove one M3×8 cross recessed pan-head
combination screw on the front, as shown in Figure 3-14.
78
3 Repairs
Purpose
The valve assembly can be disassembled and replaced by following the procedures specified in this
section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you
begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l The replacement valve with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. If it is only the electromagnetic valve on the left or right clapboard that needs to be replaced,
79
3 Repairs
open the corresponding left- or right-side door; if the one on the front panel needs to be replaced,
remove the front panel cover by following the instructions in 3.2.3 Disassembling the Panel
Cover.
Disassembly
1. Disassemble the peripheral fluid tubes connected to the valve assembly.
2. Use a #2 (Ph2) Phillips screwdriver to disassemble the two M3×8 cross-recessed pan-head
combination screws which affix the valve assembly, and carefully take out the valve assembly
while making sure to unplug the attached wires. See Figure 3-15.
Figure 3-15 Replacement of valve assembly
Installation
l Make sure to the use correct model of valve and create a reliable connection
l All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
l This machine involves a great number of valves. Only one valve is discussed as an example
here, and all the other valves follow the same maintenance procedures.
Purpose
The liquid pump assembly can be disassembled and replaced by following the procedures specified
80
3 Repairs
in this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken
before you begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l 2.5mm hex-socket screwdriver
l The replacement liquid pump with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right side door.
Disassembly
1. Pull out the peripheral fluid tubes and wire connectors attached to the liquid pump.
2. Use a #2 (Ph2) Phillips screwdriver to loosen the two M4×10 cross-recessed pan-head
combination screws which affix the liquid pump, and move the liquid pump assembly to the left till
the big hole on the affixing panel of the liquid pump is aligned to the affixing screw, then carefully
hold up the liquid pump assembly to take it out. See Figure 3-16.
Figure 3-16 Replacement of liquid pump assembly (1)
3. The further dismantling of the liquid pump assembly is shown in Figure 3-17.
81
3 Repairs
Installation
l All the wirings need to be set up according to their original positions to prevent them from being
crushed or otherwise damaged, and to prevent the working liquid pump from being shaken,
which affects the fluid tubes.
l Pay special attention to the tubing joints and ensure that the connections are sound.
Purpose
The negative-pressure chamber assembly can be disassembled and replaced by following the
procedures specified in this section; however, in each case, 3.2 Preparatory Work Before Repairs
need to be taken before you begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l 2.5mm hex-socket screwdriver
l The replacement parts or assembly of the negative-pressure chamber with the corresponding
specifications
82
3 Repairs
Preliminary Steps
Open the right side door.
Disassembly
1. Disassemble the peripheral fluid tubes connected to the negative-pressure chamber.
2. Use a #2 (Ph2) Phillips screwdriver to disassemble the two M4×10 cross-recessed pan-head
combination screws which affix the negative-pressure chamber, and carefully take out the
negative-pressure chamber assembly. See Figure 3-18.
Figure 3-18 Replacement of negative-pressure chamber assembly
Installation
l Make sure to place the seal ring into the seal tank to keep the pressure chamber airtight.
l All the tubing needs to be set up according to the relevant standards or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
83
3 Repairs
Purpose
The syringe can be disassembled and replaced by following the procedures specified in this section;
however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you begin.
Tools/Spare Parts
l 2.5mm hex-socket screwdriver
l The replacement syringe assembly with the same specifications
Preliminary Steps
l Open the left side door.
Disassembly
1. Pull out the peripheral fluid tubes connected to the syringe assembly to be replaced.
2. Use the 2.5mm hex-socket screwdriver to disassemble the two M3×8 stainless-steel inner hex
screws affixing the syringe assembly. Then take out the syringe assembly. See Figure 3-19.
Figure 3-19 Replacement of syringe
2
1
Installation
Follow the corresponding disassembly steps in reverse order.
84
3 Repairs
Tools/Spare Parts
l 2.5mm hex-socket screwdriver
l #2 (Ph2) Phillips screwdriver
l The replacement syringe assembly with the same specifications
Preliminary Steps
Open the left side door.
Disassembly
1. Pull out the peripheral fluid tubes connected to the syringe module to be replaced.
2. Use #2 (Ph2) cross screw driver to remove four M3×8 cross recessed pan-head combination
screws, take off the syringe assembly. See Figure 3-20.
Figure 3-20 Replacement of motor (1)
1
2
3. Use the 2.5mm hex-socket screwdriver to disassemble the two M3×8 stainless-steel inner hex
screws affixing the syringe module. Then take out the syringe. See Figure 3-19.
4. Take out the two M3×8 inner hex screws on the shield cover, then remove the white bolt from the
motor. Next, remove the four M3×8 inner hex screws from the back of the motor for the
replacement. See Figure 3-21.
85
3 Repairs
Installation
Follow the corresponding disassembly steps in reverse order.
Purpose
The WBC counting bath assembly can be disassembled and replaced by following the procedures
specified in this section; however, in each case, Preliminary Steps needs to be taken before you
begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l 2.5mm hex-socket screwdriver
l The replacement WBC counting bath assembly with the same specifications
Preliminary Steps
1. Run the draining sequence of the software to drain any residual liquid inside the machine.
2. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
3. Open the right side door.
86
3 Repairs
Disassembly
1. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw in the lower section of the shield cover to remove the cover; to take out the
screw, move slightly upwards for 3mm or so and apply force in a direction perpendicular to the
right clapboard. See Figure 3-22.
Figure 3-22 Dismantling and replacing the WBC bath assembly (1)
87
3 Repairs
4. Use 2.5mm hex wrench to remove the two M3×8 stainless-steel hex-socket screws affixing the
upper section of WBC counting bath assembly. See Figure 3-24.
Figure 3-24 Dismantling and replacing the WBC bath assembly (3)
1 - WBC counting chamber assembly 2 - M3×8 stainless-steel inner hex screw (×2)
5. Carefully and slowly take out the WBC counting bath assembly.Pay attention to the wiring in the
process of moving to avoid breaking the wires.
6. Remove the protective ring from the shield case and open the wire clip to take out the inner wire
along the stretching direction of the wire.Then take out the WBC counting bath assembly as a
whole.
7. The further dismantling of the WBC counting chamber assembly. See Figure 3-25.
Figure 3-25 Dismantling and replacing the WBC bath assembly (4)
88
3 Repairs
Installation
l All the wirings need to be set up according to the working procedures or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
l Pay special attention to the tubing joints and ensure that the connections are sound.
l The aperture, seal gasket, and seal ring need to be correctly positioned and oriented in the right
direction.
Purpose
The RBC counting bath assembly can be disassembled and replaced by following the procedures
specified in this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be
taken before you begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l 2.5mm hex-socket screwdriver
l The replacement WBC counting bath assembly with the same specifications
Preliminary Steps
1. Run the draining sequence of the software to drain any residual liquid inside the machine.
2. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
3. Open the right side door.
Disassembly
1. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw in the lower section of the shield cover to remove the cover; to take out the
screw, move slightly upwards for 3mm or so and apply force in a direction perpendicular to the
right clapboard.
Refer to the corresponding section in 3.8.1 Dismantling and replacing the WBC bath
assembly.
2. Pull out the peripheral fluid tubes attached to the RBC counting bath assembly.
89
3 Repairs
Pay attention to handling the residual fluid; try to prevent it from flowing onto the other
components in the machine, and wipe clean if necessary.
3. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw for grounding on the middle clapboard of the shield case cover and pull out
the grounding wire. See Figure 3-26.
Figure 3-26 Dismantling and replacing the RBC bath assembly (1)
1 - RBC counting chamber assembly 2- M3×8 stainless-steel inner hex screw (×2)
90
3 Repairs
5. Carefully and slowly take out the RBC counting bath assembly. Pay attention to the wiring during
the process of moving to avoid breaking the wires.
6. Remove the protective ring from the shield case and open the wire clip to take out the inner wire
along the stretching direction of the wire. Then take out the RBC counting bath assembly as a
whole.
7. The further dismantling of the RBC counting chamber assembly. See Figure 3-28.
Figure 3-28 Dismantling and replacing the RBC bath assembly (4)
Installation
l All the wirings need to be set up according to the working procedures or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
l Pay special attention to the tubing joints and ensure that the connections are sound.
l The aperture, seal gasket, and seal ring need to be correctly positioned and oriented in the right
direction.
91
3 Repairs
Tools/Spare Parts
l Phillips screwdriver
l Corresponding specifications needed to replace the main control panel
Preliminary Steps
1. Pull out the power cord plug from the rear panel of the analyzer, and turn off the power switch.
2. Disassemble the screws on the back panel of housing and remove the back panel of the
housing.
Disassembly
1. Pull out the peripheral wires and connectors attached to the main control panel.
2. As shown in Figure 3-29, disassemble four M4×8 cross recessed pan-head combination screws
fixing the main control panel, and remove the main control panel.
Figure 3-29 Replacement of the main control panel
1
2
92
3 Repairs
Installation
Follow the corresponding disassembly steps in reverse order.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l The replacement reagent testing panel with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. Open the left side door.
Disassembly
1. Remove the four M3×8 cross recessed pan-head combination screws affixing the protective
cover of the reagent testing panel and remove the protective cover. See Figure 3-30.
Figure 3-30 Replacement of reagent testing panel (1)
1 2
93
3 Repairs
Installation
l All the wirings need to be set up according to the working procedures or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
l Pay attention to the position of wires when installing the sensor panel so as to prevent the wires
from breaking.
94
3 Repairs
Purpose
The temperature sensor can be disassembled and replaced by following the procedures specified in
this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before
you begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l 2 hex-socket screwdriver
l The replacement temperature sensor with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. Open the left side door and the front panel cover.
Disassembly
Loosen the screws in the bracket for temperature sensor to take the sensor out. Then replace it with
a new sensor and tighten the screws. See Figure 3-32.
Figure 3-32 Disassembling the temperature sensor
95
3 Repairs
Installation
Follow the corresponding disassembly steps in reverse order.
Purpose
The temperature sensor can be disassembled and replaced by following the procedures specified in
this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before
you begin.
Tools/Spare Parts
l #2 (Ph2) Phillips screwdriver
l 2 hex-socket screwdriver
l The replacement temperature sensor with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. Open the left side door and the front panel cover.
Disassembly
Unplug the wire stock and tubing connected to the pressure sensor, and take the pressure sensor
out. Then replace it with a new sensor. See Figure 3-33.
96
3 Repairs
3 - Sheath-flow pressure
sensor connecting wire
Installation
Follow the corresponding disassembly steps in reverse order.
97
4 Software Upgrade
4 Software Upgrade
4.1 Preparation
1. Get officially provided upgrade package, and prepare a USB flash disk for storing the upgrade
package.
2. Copy the upgrade package to the root directory of the USB flash disk (Do not paste it into any
newly-created folder).
3. Insert the USB flash disk with upgrade package into the USB interface of the host.
4. Use service authority to log in to the host.
98
4 Software Upgrade
The host will detect the USB flash disk. 2~5 seconds later, if the USB flash disk cannot be
recognized, there will be corresponding prompt in the interface, please insert the USB flash disk
again; if the USB flash disk can be recognized, there will be dialog box in the interface as follows.
99
4 Software Upgrade
There will be prompt with the upgrading progress on the screen. If the upgrade is successful, the
progress bar will display 100%. See the picture below.
4. Turn off the power supply of the analyzer, and restart the analyzer.
5. Use a hard object, such as a toothpick or pencil to click the focal points successively appearing
on the screen (5 points in total).
After clicking 5 focal points, the touch screen calibration is complete.
6. Turn off the power supply of the analyzer, and restart the analyzer.
100
5 Comprehensive Device Tuning
Since the replacement and maintenance of some components can lead to changes in relevant tuning
parameters, this section introduces the tuning procedures for the parameters that may be affected.
3. Next, click Check to run the initialization of the sample probe. Check if the bottom of the sample
probe is still aligned with the bottom of the swab (see Figure 5-1). If it is, click OK to complete the
tuning; if not, repeat Steps 1~2.
101
5 Comprehensive Device Tuning
102
5 Comprehensive Device Tuning
2. Fill in the cell corresponding to MCV with the MCV reference value for quality control.
3. Perform the QC test for three times in a row for the first run.
The results for each time will be automatically displayed.
Ø If the CV falls within reasonable parameters, the screen will show the CV and Gain values for
the first run. Go to step 4.
Ø If it does not fall within reasonable parameters, you’ll be prompted to redo the calibration.
Please click OK to close the message box. Then click Clear to delete the data, and repeat
the step 3.
4. Perform the QC test for three times in a row for the second run.
The results for each time will be automatically displayed.
Ø If the CV falls within reasonable parameters, the screen will show the CV and Gain values for
the second run, and show the final results. Please click "OK" to complete gain calibration.
Ø If it does not fall within reasonable parameters, you’ll be prompted to redo the calibration.
Please click OK to close the message box. Then click Clear to delete the data, and repeat
the step 4.
103
5 Comprehensive Device Tuning
3. Input the calibrator lot No. JZQX-01 in the text box for Lot No., and click the Exp. Date control to
set the expiry date of the calibrator.
4. Input the target value of the current calibrator parameter.
The calibration mode is Whole Blood by default, so there is no need to set the mode.
5. Set the well-shaken calibrator under the sample probe, then press the aspiration key on the
analyzer to start the calibration counting.
6. Repeat step 5 for a total of 12 times to get 12 results of calibration counting.
Ø After the counting is complete, a dialog box will pop up indicating that the test is complete.
Click the Save button to save the calibration result.
Ø If significant differences are found across results, a dialog box for data abnormality will pop
up.Please redo the calibration.
104
5 Comprehensive Device Tuning
105
5 Comprehensive Device Tuning
4. Set the IP address and other network information of the analyzer according to the actual
situation.
Ø If the network is accessed through a router on the site, please select Obtain an IP address
automatically and Obtain DNS server address automatically.
Ø If the network is accessed through a network switch, or the analyzer is directly connected to
the LIS on the site, please select Use the following address, so as to manually set the IP
address and subnet mask of the analyzer. The IP addresses of the analyzer and LIS must be
in the same network segment. Furthermore, their subnet masks shall be the same, while
other parameters can maintain null.
5. Click OK to save the settings and close the dialog box.
3. Input the IP address and port of LIS workstation in Network Settings area.
Find the IP address and port of LIS in the network setup interface in the LIS workstation; if IP
address can’t be found, try the method below:
106
5 Comprehensive Device Tuning
The IPv4 address in the red box is the IP address of LIS workstation.
The IP address 192.168.8.44 of the LIS workstation shown as above is used as an example, real IP
should be in the same network segment with LIS server.
The LIS icon in the upper right side on the analyzer screen turns from gray to black ,
which indicates auto hematology analyzer software is connected to LIS successfully.
If the icon stays gray, the connection fails. Please check if the IP address and port of LIS is
correct and reconnect as the steps above; if the problem still exists, please contact the hospital
network administrator to handle it.
107
6 Alarms and Solutions
The syringe fails to This syringe problem can occur while it is being
leave its initial moved.Please refer to the following solution:
position. 1. Follow the instructions in 3.7 Sheath Flow Syringe
Syringe fails to return Assembly Replacement to disassemble the
to its initial position. syringe, then remove the dust cover and
optocoupler. Plug the optocoupler into the connector
The syringe takes too of the syringe optocoupler. The user enters the
many steps to return Status > Sensor screen and covers the center of
Syringe
1 to its initial position. the optocoupler with a piece of paper. Check if the
problem
optocoupler status shown in the screen is blocked; if
The sample syringe yes, then the optocoupler is working properly.
is busy.
2. Follow the instructions in 3.7.2 Replacing the
Motor to disassemble the syringe motor and replace
it with a new motor. Then go to syringe self-test
Sample syringe under Service > Self-test > Syringe; if the syringe
timeout is working, then the maintenance has been
successful.
The horizontal motor 1. The user enters the Status > Sensor screen and
fails to leave its initial covers the center of the optocoupler with a piece of
position. paper. Check if the optocoupler status shown in the
screen is blocked. If it is, then the optocoupler is
The horizontal motor working properly; if not, then refer to 3.4.2
fails to return to its Replacing the Optocoupler on how to replace the
Horizontal initial position. optocoupler.
2 motor
problem The motor fails to 2. Refer to 3.4.3 Replacing the Sampling Assembly
move to the WBC in X- or Y-direction to disassemble the sampling
position. assembly and make sure the motor wiring is secure.
If the wiring is OK, then disassemble the motor and
The motor fails to replace it with a new one. After the installation, make
move to the RBC proper adjustments to the belt's tension, then go to
position. the syringe self-test under Service > Self-test >
108
6 Alarms and Solutions
Horizontal motor
timeout
The optocoupler of
the horizontal motor
is not working
properly.
The vertical motor 1. The user enters Status > Sensor screen and covers
fails to leave its initial the center of the vertical optocoupler with a piece of
position. paper. Check if the corresponding optocoupler
status shown in the screen is blocked. If it is, then
The vertical motor the optocoupler is working properly; if not, then refer
fails to return to its to 3.4.2 Replacing the Optocoupler on how to
initial position. replace the optocoupler.
The motor fails to 2. Refer to 3.4.3 Replacing the Sampling Assembly
move into position to in X- or Y-direction to disassemble the motor from
isolate the air the sampling assembly and make sure motor wiring
bubbles. is secure. If the wiring is OK, then disassemble the
motor and replace it with a new one. After the
Vertical The motor fails to installation, make the proper adjustments to the belt
3 motor move to the DIFF tension, then go to the syringe self-test under
problem bath position. Service > Self-test > Syringe. If the syringe is
working, then it has been successfully fixed.
The motor fails to
move to the counting
bath position.
Vertical motor
timeout
109
6 Alarms and Solutions
The 12V +12V power is not Refer to 3.9 Main Control Panel Replacement to
power is working properly. replace the main control panel.
6
not working -12V power is not
properly working properly.
110
6 Alarms and Solutions
Abnormal pressure of
positive-pressure
chamber (lower than
normal)
Abnormal pressure of
positive-pressure
chamber (higher than
Pressure normal)
9 chamber
problem The Refer to 2.2.12.10 Problems in Creating the
negative-pressure Negative Pressure for troubleshooting.
chamber fails to
create pressure
Abnormal pressure of
negative-pressure
chamber (lower than
normal)
Abnormal pressure of
negative-pressure
chamber (higher than
normal)
111
6 Alarms and Solutions
LYC-2 expired.
Insufficient LYC-1.
DIL-C running out. Check for the availability of the relevant reagent. If the
reagent is available, then check if the optocoupler of
15 No reagent LYC-1 running out or the reagent testing panel is on. If not, adjust the screw
air bubbles in inlet of the potentiometer and rotate counterclockwise to
tubing.
112
6 Alarms and Solutions
113
7 Maintenance Inventory
7 Maintenance Inventory
114
7 Maintenance Inventory
33 32010006A Power
115
7 Maintenance Inventory
116
P/N: 65.02.0120A (5.0)