How To Solve The Problem of HGB Results Unstable?: Global Customer Service Center
How To Solve The Problem of HGB Results Unstable?: Global Customer Service Center
How To Solve The Problem of HGB Results Unstable?: Global Customer Service Center
3 Q&A
1.1 Principle of HGB measurement process
LED Lamp
(λ=525nm) Solution Filter Photodiode
◼ Colorimetric principle
DH26/DH36/DH56/DH76 DF50/DF50VET/DF50CRP
◼ Signal collection
1. Blank Photocurrent (Colorless Diluent in chamber)
2. Sample Photocurrent (Red color sample solution in chamber)
1.2 Key components for HGB counting process
3 Q&A
2. Abnormal HGB results
× means abnormal
If the WBC & RBC √ means normal
result is OK
Blood aspiration or 2.1 Counting 2.3 Diluent 2.4 Waste Main board issue or
2.2 Lyse issue
dispensing issue chamber issue issue discharge issue HGB unit issue
2.1 WBC and RBC results are both abnormal
If WBC & RBC result repeatability are both unstable, then please check the dispensing valve, syringe and sample probe.
2. Lyse issue
2. Suggest to use Dymind original lyse to ensure the performance of the analyzer
If both the WBC & RBC result repeatability is stable, but HGB results are still abnormal, then please check the followings.
1. Replace the HGB unit itself
2. Replace the main board
https://www.wjx.cn/jq/87790763.aspx https://www.wjx.cn/jq/87983790.aspx
After training, please scan the QR code or enter in the link to fill in the survey to get more scores
THANK YOU
www.dymind.com/en.