Service Manual For XL-100 With ISE
Service Manual For XL-100 With ISE
Service Manual For XL-100 With ISE
Service Manual
Model: XL-100
Foreword
This manual is organized in a progressive sequence for easy study and reference. It is an
instructional aide to provide a reference for easy operation and general maintenance of
this analyzer.
The manual contains detailed description of the XL-100 analyzer features and
specifications, including software on the operational PC. The analyzer is used with
operational PC and printer, and can interact with the host computer. All the samples and
reagents for measurements including samples obtained from patients are controlled by
bar codes enabling the analyzer to perform the entire process of the analysis
automatically.
Use of the analyzer with proper knowledge will ensure quality test results and trouble free
analyzer operation and performance.
Assumptions are made that before making an attempt to operate the analyzer, the
operator is familiar with the operation of analyzer and has:
In case of problems beyond the scope of the service engineer, the Transasia Bio-
medicals Limited is available for assistance and at the following:
Copyright
© 2010 “Transasia Bio-medicals Limited "All rights reserved. This document is subject to
change. This document and the information herein are provided for the sole use of the
intended recipient(s) and for information purposes only.
This document contains contents which are the confidential and proprietary information of
Transasia Bio-medicals Limited. This document may not be modified, reproduced, or
transmitted in any form or by any means, electronic or mechanical, for any purpose,
without prior written permission from Transasia Bio-medicals Limited.
Contact Us
We welcome your comments about the document. You can contact us through our e-mail
at the following.
E-mail : transasia@transasia.co.in
Web site : www.transasia.co.in
Contents
1. Safety ............................................................................................................................ 9
1.1. Warning Labels ........................................................................................................9
1.2. Warning for Safe Use ............................................................................................ 12
2. Introduction ................................................................................................................ 13
2.1. Purpose and Scope ............................................................................................... 13
2.2. System Description ................................................................................................ 13
2.2.1. Overview ..................................................................................................................... 13
2.2.2. Main Features ............................................................................................................. 14
2.2.3. Operating Principle ...................................................................................................... 15
2.2.4. Equipment Layout and Dimensions............................................................................. 16
2.2.5. Sequence of Operation ............................................................................................... 17
3. Technical Specifications............................................................................................ 20
3.1. General Specification ............................................................................................. 20
3.2. Installation Conditions ............................................................................................ 21
3.3. Sampling Unit ........................................................................................................ 22
3.4. Reagent Unit .......................................................................................................... 22
3.5. Reaction Unit ......................................................................................................... 23
3.6. Optical Absorption Measurement Unit ................................................................... 24
3.7. Data Processing .................................................................................................... 24
3.8. Control Unit ............................................................................................................25
3.9. Accessories and Consumables.............................................................................. 25
4. Pre-installation Requirements................................................................................... 27
4.1. Installation Conditions ............................................................................................ 27
4.2. Site Requirements ................................................................................................. 30
4.2.1. Proper Room Temperature and Humidity ................................................................... 30
4.2.2. External Interferences ................................................................................................. 30
4.2.3. Proper Room Lighting ................................................................................................. 31
4.2.4. Proper Laboratory Cleanliness (Practice) ................................................................... 31
4.2.5. Flow of Liquids in ISE Tubing’s ................................................................................... 31
4.2.6. Electrical Requirements .............................................................................................. 31
4.3. Floor Requirements ............................................................................................... 32
5. Installation .................................................................................................................. 33
5.1. Inspections and Accessories Checklist .................................................................. 33
5.2. Basic Operational Information................................................................................ 35
5.2.1. Receiving Instructions ................................................................................................. 35
5.2.2. Warranty Information ................................................................................................... 35
5.3. Unpacking the Analyzer ......................................................................................... 35
5.4. Installing the Components of Analyzer .................................................................. 36
5.4.1. Connecting Communication Cable between Computer and Analyzer ........................ 36
5.4.2. Installing the Sample and Reagent Tray (SRGT) ........................................................ 36
5.4.3. Installing the Rotor Tray .............................................................................................. 37
5.4.4. Installing DI Water and Waste Can ............................................................................. 38
5.5. Equipment Startup Procedure................................................................................ 39
5.6. Software Installation Procedure ............................................................................. 40
5.6.1. Pre-requisite ................................................................................................................ 40
5.6.2. Installing MultiXL Software .......................................................................................... 49
1. Safety
This chapter provides the user with necessary information on the warning labels.
The user is requested to read before installation of the analyzer.
Bio-hazard
Electric Shock
High Temperature
Injury
Action to be taken as
directed by the
operator manual
The following warning labels are affixed on the places that are the potentially
hazardous.
2. Introduction
2.1. Purpose and Scope
This is the service manual for the equipment Fully Automatic Clinical Analyzer XL-
100. It provides the brief knowledge on the technical components and the detailed
information on the software functionality of the analyzer. The operator of this
machine uses this document for reference.
This manual does not intend to teach troubleshooting skills. It is designed to
provide a competent operator trained on the XL-100 analyzer, with the necessary
information.
Incident or white light contains the entire spectrum, objects that appear colored,
absorb light at a particular wavelength and reflect others, thus giving different
colors. That color is a function of its wavelength
Light having a wavelength of less than 400 nm is termed Ultraviolet, where as light
having a wavelength greater than 800 nm is described as Infrared, both ultraviolet
and infrared lights are invisible to the human eyes. Light corresponding to
wavelengths between 400 nm and 800 nm is visible to the human eye and is
termed as Visible light.
3. Technical Specifications
3.1. General Specification
Item Description
Throughput 100 tests/hour for a cycle time of 36 seconds
With ISE (250 tests)
System type Discrete, open, automated, random access, patient prioritized, 1/2
Reagent system
Sample Serum
Urine
Others (CSF, Plasma, etc.)
Measurement Turbidimetric Immunoassay
principle Colorimetric (Rate/End Point)
Ion Selective Electrodes (optional)
Applicable Photometric assays
analytes Enzyme, lipid, protein, sugar, nitrides, inorganic substances,
complements and others
Turbidimetric assays
IgG, IgA, IgM, C3, C4, RF, CRP, ASO, Transferrin and others
On board tests 42 test items maximum
25 positions fixed
15 flexible
2 additional positions at room temp.
Programmable Unlimited
parameters
Assay modes 1-Point
2-Point
Rate-A
Rate-B
Sample volume 2-70 µl (adjustable in 0.1 µl step)
Reaction volume 180 µl
Reaction 37 °C
temperature Temperature stability: ± 0.5 °C
Reaction time Depends on the designated cycle time and number of reagents used
For 1 step assay (using R1)
612 seconds (10.2 min)
For 2 step assay (using R1 and R2)
1st reaction 306 seconds (5.1 minutes) + 2nd reaction 306
seconds (5.1 minutes)
Test selection Setting of tests one by one or with profile key for each sample
Group order entry is possible
Setting from host computer via interface having ASTM protocol
compliance (optional)
Setting according to bar-code data on sample container (Optional fixed
bar code reader)
Mixers Mixing by probe
3 levels -Low, Medium & High
Maintenance Programmable maintenance actions: Probe Wash& maintenance
Barcode Sample tube barcode ID (NW7, code 39, code 128, 2 of 5 interleaved, 2
identification of 5 standard, ISBT-code 128) – 6 to 18 digits
reagent barcode ID,
Water supply unit Water consumption: Less Than 1.8 liters/hour
Manufactures and supplies: Type 2 quality (by NCCLS standards) ion
exchange water (optional)
Quality control Unlimited
Walk Away Time 1 hour
System warm-up 10 minutes warm-up at instrument power on
time 5 minutes warm up for rotor at start of the run if new rotor is used.
Safety mechanism Vertical obstruction detection
Capacitance based liquid level sensing
Froth error detection
Noise level Less than 65 dB
Display PC Monitor
Keyboard External Standard PC keyboard
System interface Analyzer – PC: USB
A4 size DeskJet/Laser/Dot Matrix printer
STAT samples STAT samples are measured preferentially (To be placed in any of
the 25 sample positions on the outer rim)
Calibrators may be placed at any of these 25 positions
Interrupt permitted even during analysis
Sampling Micro-pipette with level sensor
Pipetting system with plunger, driven by stepper motor
Sample volume: 2-70 µl (adjustable in 0.1 µl step)
Preheated probe common for Sample and Reagents
Pipetting mode Preheated probe with Sapphire Pump, driven by stepping motor
Single preheated Micro-pipette with level sensor
probe for dispensing Washing solution
Samples and
Reagents Outside: Deionized water
Inside: DI water passed through preheated probe.
Equipped with vertical obstruction detection facility to prevent probe
crash
Sample dilution Dilution ratio: 2 to 150 times
A cuvette is used as dilution vessel
Set amount of diluent is dispensed into a cuvette and set amount of
sample is dispensed into it by sample/reagent probe
Dilution possible for repeat run
Direct reduced/increased volume runs are also possible
Repeat run Reduced/increased volume repeat run also possible
Sample identification Position ID for all samples
Sample bar-code ID
Dead Volume ≤ 350 µl for 5ml and 7ml Sample Tube
≤ 125 µl for 2ml Sample Cup
temperature
Reagent bottles 25 fixed positions for R1 or R2 of 50/20ml inner rim.
15 flexible positions for R1 or R2 of 20 ml, outer rim
2 external positions at ambient temp for diluents, wash solutions,
reagents.
5ml Tube Adapters on 20ml bottles
Reagent dispensing Pipetting system with Sapphire Pump, driven by stepping motor
Single preheated Micro-pipette with level sensor ( Same probe as for sampling)
probe Washing solution
for dispensing Outside: De-ionized water
Samples and
Reagents Inside: De-ionized water passed through preheated probe.
Equipped with vertical obstruction detection facility to prevent probe
crash
Reagent steps 1 step or 2 step
Reagent volume Reagent 1: 50 – 300 µl (adjustable in 1 µl step)
Reagent 2: 0, or 10 – 200 µl (adjustable in 1 µl step)
Dead volume 2.5 ml for 50 ml bottles
1.5 ml for 20 ml bottles
Mixers Mixing by probe with 3 levels
Low, Medium & High
Reagent Position ID
identification Barcode ID
Residual volume Calculated by countdown system as measured by capacitance type
information level sensor and displayed on screen
Reagent protection Reagent tray cover protection from evaporation, dust, and direct light
Carry over actions Reagent, Detergent or System Wash given for Carry Over Pairs
Quality control Within day as well as day-to-day X-Mean and R control diagram (L-J
Graphs)
Mean, SD, %CV, R are calculated for each chemistry
Item Description
Installation Parts Standard sample/Reagent tray.
Waste reservoir.
Water supply reservoir.
Packaging Instrument fixed on seaworthy wooden pallet with wooden covers &
soft material inserts.
Accessories in cardboard box.
Consumables Probe wash solution.
Rotors.
Sample cup.
Halogen lamp.
ISE electrodes (Optional).
ISE calibrant and cleaning solutions (Optional).
ISE pump tubes (Optional).
4. Pre-installation Requirements
Before installing the analyzer, you must follow the pre-installation requirements to
ensure the accuracy and precision of the analyzer.
The proper location is an important consideration. Poor location can lead to
malfunction of the analyzer.
e. Wipe the nozzle tips of SRPT several times with cloth or alike with
rubbing alcohol before the analyzer is used. At this time, do not forget to
put medical rubber gloves or alike on. Pay attention to prevent bare
skins of hands or arms from being touched by or pricked with the nozzle
tip.
4. The caution should always be exercised when replacing the halogen lamp.
a. Replace the halogen lamp by a new one after a lapse of 30 minutes after
the analyzer is turned off, to avoid the danger of burns. Keep hands
away from the glass part of the new halogen lamp. Make sure that there
is no crack or breakage in the glass part.
5. The following cautions should be exercised during the operation of analyzer:
a. Pay attention not to exceed the time and volume necessary for
diagnosis.
b. Keep monitoring the behavior of the whole system in order to detect any
malfunction.
c. Take immediate corrective measures including shutdown of operation
when any malfunction is detected in the analyzer.
d. Avoid possibilities of any direct access by the patients.
6. The following cautions should be exercised after the use of the analyzer:
a. Turn off the power after every operation so that control is restored to its
previous state as directed.
b. Do not remove the line cord plugs from receptacles by pulling the cords
so that no undue stress is developed in the cords.
c. If the auto maintenance at shutdown is not performed, wipe the nozzle
tips of SRPT several times with cloth or alike with rubbing alcohol before
the analyzer is used. At this time, do not forget to put medical rubber
glove. Pay attention to prevent bare skins of hands or arms from being
touched by or pricked with the nozzle tip.
d. Pay attention to the storage area.
Keep the analyzer out of the rain and any other water splash.
Avoid areas that are adversely affected by atmospheric pressure,
temperature, humidity, ventilation, sunlight, dust and air containing
salt, sulfur, etc.
Pay attention to inclination, vibration, shock (including shock during
transportation), etc.
Avoid areas adjacent to the storage room of chemicals or areas that
are likely to generate gasses. Avoid areas that are likely to be subject
to inclination, vibration and shock.
Organize and store parts and cords associated with the analyzer
after they have been cleaned.
Keep the analyzer clean not to cause any inconvenience to the next
use.
7. In the event of trouble, do not fiddle with the analyzer. You must contact the
authorized service personnel for troubleshooting.
8. Maintenance and checks.
a. It is important for the analyzer and its associated parts to be periodically
checked.
b. Ensure that the analyzer operates normally and correctly, when it is
reused after being unused for some time.
9. The following cautions should be taken when using and handling the
reagents:
a. After unpacking the reagents, be sure not to allow dust, dirt or bacteria
to come in touch with the reagent.
b. Do not use reagents that are expired.
c. Handle a reagent gently to avoid formation of bubbles.
d. Take care not to spill the reagent. If it spills, wipe it off immediately using
a wet cloth.
10. Follow other instructions described in the package insert on each reagent.
a. If a reagent happens to enter your eye, wash it off immediately using
plenty of water, and take medical treatment at once.
b. If you swallow it inadvertently, call for a doctor immediately and drink
plenty of water.
Warning: Some reagents are strong acids or alkalis. Exercise great
care so that your hands and clothing do not come into contact with
reagents. If your hands or clothing come into contact with either
reagent, immediately wash them off with soap and water. If a reagent
comes into contact with your eye(s), immediately rinse with water for
at least 15 minutes.
11. Prohibit any alteration and/or modification to the analyzer without permission
by manufacturer.
12. The following precautions should be taken for preventing infection due to
sample handling:
4.2.6.2. Grounding
It is absolutely necessary that the perfect grounding must be provided to the power
source with all applicable local requirement (only a certified grounded, 3 pin power
plug should be used).
5. Installation
The analyzer is fully inspected before leaving the factory and carefully packed to
withstand shocks in transit.
On receiving the analyzer, check the package externally; make sure that there is
no external visible damage to the shipping container if there is damage please
make a note of it, if possible photograph it and inform the Transasia immediately.
Verify with the invoice, if all the boxes have arrived.
NOTE: In case of damage, the necessary procedures for insurance
claim are to be initiated at customer site.
1. Remove the front panel of the wooden box by loosening the bolts. The front
panel on the wooden box is marked.
2. Remove the top and side panels of the wooden box as a whole section by
loosening the bolts from the back panel side.
3. Remove the four “Z” brackets, which are holding the analyzer on the pallet.
4. Lift the instrument with the help of handle provided on both sides of the
analyzer. Handles needs to be pulled out underneath the panels.
5. Place the analyzer at the desired place keeping the floor requirement given
above in view.
1. Unpack the USB interface cable from the accessories box, and connect its
one end to the USB port of the computer and other end to the instrument’s
USB port as shown in figure below.
2. Connect the one end of the power cord to the analyzer’s power outlet at the
back side and other end to the main switch as shown in figure below:
ON/OFF
Switch
USB
interface Power
cable cord
2. Gently place the sample and reagent tray into the container, make sure that
the index pin on the tray, slides into the index hole provided on the sample
and reagent tray.
3. Unpack the Sample and Reagent tray lid, and place over the Sample and
Reagent Tray.
Make sure that the index pin on the Reagent lid slides into the index hole
provided on the Sample and Reagent tray lid.
3. Now tighten the rotor tray with the knob provided, as shown in figure below.
4. Unpack the rotor tray lid, and place over the rotor tray.
6. Take another tube of same size and connect the one end of it to the
analyzer outlet marked DI WATER – LAMP. Connect another end of the
tube on canister of DI Water CAN connector marked as DI WATER LAMP.
7. Take another tube of size 4 x 6. Connect one end of this tube to the
remaining yellow color connector on analyzer with the other end on
connector provided on Waste CAN.
Float sensor
cable with
To DI water blue sleeve to
CAN DI water CAN
To DI water
CAN for
Lamp Float sensor
Cooling cable with
yellow sleeve
to Waste CAN
To Waste
CAN
NOTE:
Take care that all the above-mentioned tubes to reach
their respective containers without any sharp bends or
obstructions.
The tube lengths needs to be cut to appropriate length as
required on site to avoid sharp bends or twisting of tubes
at later stage during refill or waste removal.
3. Turn on the switch that is provided on the back side of the instrument.
4. Start the MultiXL software.
5. Turn on the instrument switch provided on the right side of the analyzer.
After turning on the switch, all the assemblies initialize to its home position.
6. Now the instrument is powered-ON, and ready for further RUN.
NOTE: The main power switch should be ON at all times
except in an emergency. The main power supplies
electricity to ISE unit for Automatic priming & SRGT unit
cooling.
5.6.1. Pre-requisite
5.6.1.1. System Configuration
Processor Pentium Core2 Duo
Operating system TBM image of Windows Embedded 7.
Hard disk 120 GB or above.
RAM 2 GB or above.
Monitor 15 inch TFT color.
Supporting 1024 X 768 resolution
Printer Laser/ DeskJet.
5.6.1.2. Important PC Settings for Windows 7
NOTE: PC Settings will be applied automatically while installing the
customized image of Windows 7 Embedded as Operating System.
* A customized image of Windows 7 Embedded is provided as
Operating System for Analyzer PC.
Remove all memory resident software including anti-virus software from the
analyzer PC, if installed.
Remove firewall and other security software and / or settings from the
analyzer PC.
Do not run any other application on the analyzer PC during batch run on
analyzer.
Disable Screen-savers and Power Management on Analyzer PC before
starting the application software.
Ensure that the default printer (Laser Jet / DeskJet) is configured and
connected to the analyzer PC.
Delete the “Microsoft XPS Document Writer” from the system. To do this,
click Start button, and click Control Panel. Now, under the Category view,
go to Hardware and Sound > Devices and Printers.
Right click on the Microsoft XPS Document Writer, and choose Remove
Device to delete.
Disable Screen-savers and Power Management on Analyzer PC before
starting the Application Software.
Use the following procedure for disabling the screen saver and power
management:
a. Open Screen Saver Settings by clicking the Start button, clicking
Control Panel, clicking Appearance and Personalization under
Category view, clicking Personalization, and then clicking Screen
Saver.
b. To turn off all screen savers, under Screen Saver, select (None) from
the drop-down list, and then click OK.
c. Now click on the link Change power settings (Applicable only if option
is available). The following screen will be displayed.
Similarly apply same settings to Power saver option. Repeat the step c
to e.
User account control settings
Follow these steps to change the user account settings:
f. Open User Account Control Settings by clicking the Start button, and
then clicking Control Panel. Under Small or Large icons view, click on
User Accounts and then, again click User Accounts. The following
window will be displayed.
Use the following procedure for setting the regional and language options:
Ensure that the following regional and language settings are appropriate.
Note that these are the critical settings for the communication with the
analyzer.
Click on Start and open Control Panel. Under Category view, click on
Clock, Language and Region. The following Region and Language
screen will be displayed.
i. Following should be the settings for Formats tab:
5. Click Accept.
On clicking, the following screens will be displayed in a sequence.
Wait until the setup wizard is visible on the screen as shown below.
6. Click Next.
The following screen will be displayed. The installer will install the software
in the default location C:\Program Files\MultiXL. To install to a different
location enter the appropriate location or click Browse. The installation in a
location, different from the default one, is not recommendable.
8. On clicking Next, the installation is started and the status will be displayed,
as shown below.
Once the installation completes, the software icon will be created on the
desktop and the following screen will be displayed.
2. Double-click DatabaseUtility.exe.
3. Click on CHECK DATABASE.
This command will check for the existing database on your computer. If no
database is available, a message is displayed on the following screen.
5. Now click OK and select the blank database file MultiXL.BKP from the
software CD, and then click Open.
After some time, the database will be created successfully and the
following screen will be displayed.
7. The above screen indicates that the database is created successfully. Click
Close to exit from the utility screen.
8. Now you can access the MultiXL application. (Refer section 5.6.5 Accessing
MultiXL Software for more details).
4. Now click OK and select the blank database file MultiXL.BKP from the
software CD, and then click Open.
After some time, the database will be upgraded successfully and the
following screen will be displayed.
5. Select the path to store the backup file, and then click Save.
On clicking, the following screen will be displayed to confirm the successful
completion of backup database.
On clicking No, the Live and Archive data will be deleted without saving the
database.
8. Click Save to save the database.
On clicking, a copy of database will be saved in the desired location.
9. The following screen will be displayed after successful deletion of the database.
6. Click OK and select the blank database file MultiXL.BKP from the
application CD.
7. Click Open.
The following screen will be displayed after successful creation of
database.
4. After archiving the database successfully, the following screen will be displayed.
Once the OK is pressed, the following user login screen will be displayed. This
screen will be displayed every time when you start the application
Enter the following details, and then click OK. You may use this Login details for the
first time and create the other Login IDs with appropriate Access Rights.
Login ID : Guest or guest.
Password : Guest
QC/Calibrations – This screen is used for viewing the QC Data and Twin Plot
graph. It is also used for viewing the calibration curve for a particular test and for
scheduling blanks / standards / calibrators / controls.
Consumables – This screen is used for definition of Reagents, Standards,
Blanks, Controls, Diluents and Wash Solution.
Status Monitor – This screen is used for performing calibration/patient run,
viewing Reagent Level of different Tests and performing Sample & Reagent
Barcode scans. It is also used to view online reaction curve.
Search – This screen is used for searching Patient Results, Calibration / Control
Results, Patients, Consumables or Test Details
Reports – This screen is used for viewing Patient Reports, calculating Statistical
data using Test Statistics screen, View the results on Result Reprint screen, View
the calibration of a test over a period of time on Calibration Trace, View the
current calibration on Calibration Monitor, Log of all the errors on Error Log
screen, reaction curve of the results obtained on Reaction Curve screen and
Reagent consumption on Other screen.
Master – This screen is used for defining miscellaneous parameters like Area,
Laboratory, Doctor’s Details, Analyst Details etc.
Utility – This screen is used for defining Reagent Positions, Backup Data, to
view Offline Results and recalculate the results obtained.
Service Check – This screen is used only by Service Personnel.
Maintenance – This screen is used for performing Maintenance operations at the
start and end of the day.
Settings – This screen is used for defining system parameters, carryover pairs,
test sequence, rerun flags and to assign user rights.
Archive - This option is used to the archive the MultiXL data. On clicking this
option, the data will be moved to the MultiXL archive database. This option will be
visible only when the data exceeds specified threshold levels.
Shut Down – This screen is used to turn off the application software.
The menus can be accessed using the shortcut keys shown in brackets.
The currently selected option is highlighted with Yellow color.
Following screenshot shows list of Common buttons used:
For example: In the screenshot below, button is placed near a box with caption
‘Test’. If this button is clicked, small window opens up for selecting a particular test.
Following is an indication bar available on most of the screens. This provides help /
warning messages to the user.
NOTE: The sign * near any of the fields on the software screens
indicate that the field is mandatory. The sign ** near any of the
column in the master option indicate that row is a default row and
can’t be deleted by the user.
2. On clicking, a text file is created which contains password details of the user
and sends to the location C:\MultiXLLOG. The following screen will be
displayed.
3. Now go to location C:\MultiXLLOG, and send the generated text file to the
service engineer.
4. Click OK to close.
5. Click OK.
6. Click SAVE to save the settings.
The captions on the screen will appear in the selected language.
Put the required amount of cleaning solution and two empty cups at tray
position 4 and 6. After the completion of maintenance activity, the
application will be closed and analyzer is switched off.
2. In case the malfunction occurred is fatal and the instrument is not able to
come to its stable state then the user needs to switch OFF the Mains Switch
provided on the right side of the instrument. This will turn OFF the
instrument and will not allow further hazard. The user can turn ON the
instrument again so as to restart the functionality.
6. Relocation
This section provides instructions for decontaminating, draining out the analyzer,
disconnecting peripheral devices and packing the analyzer securely for
transportation, and is intended for use only by Transasia trained Field Service
Engineer.
7. Insert the wooden box from the top of the analyzer packing to complete the
packing and fasten the nut bolt with the pallet.
8. Put the stickers like fragile indication; do not tumble; do not stack on top this
box, Protect from sunlight and rain, Precession instrument.
9. Copy of the certificate of decontamination.
CITY: COUNTRY:
CUSTOMERS NAME:
CUSTOMERS ADDRESS:
CITY: COUNTRY:
DATE:
7. Functional Description
This chapter provides the user with necessary background on the analyzer for its
use. The user is requested to read before starting operation.
This section contains the description of each unit constituting the system.
SRPT Arm
ISE port
Reaction Tray
Place for
ISE unit Sample and
(optional) Reagent Tray
(SRGT)
The sample and reagent probe is connected to the syringe pump for sample
aspiration via PTFE tube. This is also equipped with heating element which helps
reagent or samples to be preheated close to 37ºC prior to dispense in cuvettes.
The sample or reagent on the SRGT tray is aspirated by the pipette and then
dispensed into the cuvettes (reaction cells) in the RCT unit. After dispensing of
Sample or Reagent 2, the probe mixes the reaction mixture with the help of
aspiration and dispensing the mixture as programmed by the user (High, medium
or low). Probe has smooth finish surface from outside and is polished from inside
to minimize any sample or reagent carry over.
When an optional ISE unit is fitted and the ISE measurement is performed, the
SRPT aspirates sample and dispenses it into the sample port of the ISE unit for
ISE measurement.
20 ml 50 ml
Reagent Reagent
bottle bottle
The resolution of the barcode label should be 0.25 mm. Length of the barcode
should be 42 mm and width should be 10 mm.
Number of digits should be 3 to 18 but the combination of digit and type should be
within the specified length of the barcode label.
Electrodes Reagent
Pack
Waste
CAL B
CAL A
Calibrant-B
Calibrant-B is used at the time of two-point calibration.
The two-point calibration should be carried out at the beginning of the day and at
least once every 8 hours or after completion of 50 samples.
Cleaning Solution
The cleaning solution needs to be dispensed into the unit to avoid deposition of
protein on the electrodes.
As necessary, 500µl of the wash solution is dispensed into a sample cup and it is
placed on position of the ASP tray.
This function should be carried out twice a day, once in the beginning of the day
before the Calibration and at the end of day. When more than 50 samples of
measurement are carried out, washing must be carried out.
Diluent
The diluent is used to dilute urine to one-tenth in concentration. It is placed in a
reagent bottle that is placed in the RGT unit at the user defined position. The
necessary volume for diluting one sample is 200µl. The dilution is carried out using
a cuvette in the RCT unit and therefore one cycle of chemistry analysis is allocated
to this processing.
Sampling Volume at each Measurement
In the case of analytic Sample Volume for Serum 70 µl
measurement Sample Volume for Urine 140 µl
In the case of full calibration Calibrant-A: 180µl, Calibrant-B: 180µl
In the case of 1-point Calibrant-A: 180µl
calibration (Serum cycle)
In the case of 1-point Calibrant-B: 230µl
calibration (Urine cycle) Calibrant-A: 100µl
Reduced power rail count means fewer power regulators and lower cost
regulators due to more tolerant Vcc specifications.
True 3.3V I/Os result in high noise tolerance with no external input buffers.
LVCMOS and LVTTL drive strength eliminate the need for signal output
buffers and line drivers.
Due to digital implementation of DCM, fewer external components are
required.
Block Diagram of Spartan – 3E FPGA as follows:
A flexible micro blaze system architecture enables you to effectively achieve four
product goals:
Reduce overall product cost
Achieve performance targets
Eliminate risk of obsolescence
Use industry-standard software development frameworks for productivity
Spartan – 3E Architecture
The FPGA is an integrated circuit that contains many (64 to over 10,000) identical
logic cells that can be viewed as standard components. Each logic cell can
independently take on any one of a limited set of personalities. The individual cells
are interconnected by a matrix of wires and programmable switches. A user's
design is implemented by specifying the simple logic function for each cell and
selectively closing the switches in the interconnect matrix. The array of logic cells
and interconnects form a fabric of basic building blocks for logic circuits. Complex
designs are created by combining these basic blocks to create the desired circuit.
The logic cell architecture varies between different device families. Generally
speaking, each logic cell combines a few binary inputs (typically between 3 and
10) to one or two outputs according to a Boolean logic function specified in the
user program. In most families, the user also has the option of registering the
combinatorial output of the cell, so that clocked logic can be easily implemented.
The cell's combinatorial logic may be physically implemented as a small look-up
table memory (LUT) or as a set of multiplexers and gates. LUT devices tend to be
a bit more flexible and provide more inputs per cell than multiplexer cells at the
expense of propagation delay.
Field Programmable means that the FPGA's function is defined by a user's
program rather than by the manufacturer of the device. A typical integrated circuit
performs a particular function defined at the time of manufacture. In contrast, the
FPGA's function is defined by a program written by someone other than the device
manufacturer. Depending on the particular device, the program is either ‘burned’
in permanently or semi-permanently as part of a board assembly process, or is
loaded from an external memory each time the device is powered up. This user
programmability gives the user access to complex integrated designs without the
high engineering costs associated with application specific integrated circuits.
Individually defining the many switch connections and cell logic functions would be
a daunting task. Fortunately, this task is handled by special software. The
software translates a user's schematic diagrams or textual hardware description
language code then places and routes the translated design. Most of the software
packages have hooks to allow the user to influence implementation, placement
and routing to obtain better performance and utilization of the device. Libraries of
more complex function macros (e.g. adders) further simplify the design process by
providing common circuits that are already optimized for speed or area.
It is 256K X 16 bits (4MB bit RAM) High Speed Asynchronous SRAM working on
3V3 power supply.
It is being used to download FPGA code in to RAM at Power ON to interface as
per instruction from Micro blaze to process task.
C) Flash Memory: M25P16:
It is 16Mbit Serial flash Memory with 50MHz SPI interface.
Micro blaze Program code is being downloaded in to Flash Memory during Power
ON. SPI interface is established between Micro blaze & Application software
through Flash Memory with SPI interface.
D: Data Input pulses to be monitored at TP11.
CLK: CLK to Flash Ram to be monitored at TP9.
S: Chip select to be monitored at TP10.
Q: Output of flash Ram is to be monitored at TP12.
D) EEPROM: AT24C1024
It is 1 M bits Two wire Serial EEPROM. SCL is a CLK input & SDA is Data.
Calibration data of interfaces are being stored in to EEPROM. During next
operation interfaces will behave as per updated data.
E) Power Supply Section:
CPU PCB assembly’s interfaces are working on +3.3V Dc, +2.5V DC, +1.2V DC.
1. +3.3 V Dc: +5V VCC as an input supply at TP1 to IC U1 – LD1085 – Pin3:
Low drop positive voltage regulator. Output voltage 3V3 DC at Pin2 of U1
(+/-2%) to be monitored at TP2 w.r.t. GND TP4, indicated by LED D1. 3V3
as VCCO connected to Xilinx FPGA.
2. +2.5 V Dc: +5V VCC as an input supply at TP1 to IC U2 – LD1117 25 –
Pin3: Low drop positive voltage regulator. Output voltage 2V5 DC at Pin2 of
U2 (+/-2%) to be monitored at TP5 w.r.t. GND TP4, indicated by LED D2.
2V5 as VCCAUX connected to Xilinx FPGA.
3. +1.2 V Dc: +5V VCC as an input supply at TP6 to IC U3 – LD1085 – Pin3:
Low drop positive voltage regulator. Output voltage 1V2 DC at Pin2 of U3
(+/- 2%) to be monitored at TP7 w.r.t. GND TP8. 1V2 as VCCINT connected
to Xilinx FPGA.
Following Functions have been carried out with the help of CPU PCB
ASSEMBLY:
1. Generate Clock /Direction signals to drive the stepper motors
2. Generate signals to drive the Valves/Pumps,
3. Read liquid level isolated signals & generate error signals accordingly.
4. Read opto isolator signals generated from interfaces.
5. Preheater Probe temperature control.
6. Serial communication with RGT temperature Controller to monitor RGT
Temperature, RCT controller to control RCT temperature at 37 Deg. +/- 0.5
PWM L298 is being used to control heater. U24: Heater1, U62: Heater 2,
Sensor Input is connected at Connector J23. Connector J10 is being used
to Program Microcontroller U22 through JTAG. LED 15, LED 16 will be
showing status of Sensor. If sensor gets open then respective LED will start
flashing & Heater will be OFF. RCT Heater ON/OFF status will be indicated
with the help of LED’s 14 – Heater H1,LED 17 : Heater H2 ,
5. LLC: Liquid level sensing section the signal sense through LLC junction
board at J9 Sensor Input buffered through U12 & connected to CPU board
& Respective Output from CPU for error indication through Darlington
transistor array U11 to LLC Junction board to drive respective LED’s.
Connector J9 for LLC Signals from / To CPU PCB - To/From LLC junction
board.
6. Power Supply: +5V,-5V DC from power supply cards to the LLS card
through Interface PCB assembly.
The flow of Power supply for the LLS is from Interface card through CPU
card to LLS PCB Assembly.
Connector J26: LLS Power Supply to LLS PCB Assembly Located at ARM
routed through Interface card.
From +24V DC – SMPS IS +5V is being generated with the help of U55 – IC
7805-D2Pack , +3V3 generated from U21 – LD1085CDT to Power Up 3V3
interfaces/IC’s.
7. RS 232 - Serial Communication Interface :
a. RS232 Communication :
i) RCT heater temperature: RCT heater temperature signals from U22
Pin29 & 30 are send to CPU PCB Assembly at RCT HTR_RX,RCT
HTR_TX
ii) RGT Temperature: RGT Temperature Controller serial interface has
been connected to J38. Serial Interface Baud rate 19200, with “Request
to Send “Command Protocol from PC to read RGT temperature
Max 3232 U36 is being used for RGT temperature serial communication
between RGT controller & CPU PCB Assembly as RGT-Cooler_TX,
RGT-cooler_RX.
b. Photometer Communication: MAX3232 U34 is being used to transmit
& receive Photometer data from / to CPU PCB Assembly Connector J35
for Photometer communication.
Opto 6N137: High Speed opto are being used for Transmit U33 &
Receive U35.
C PC Communication : USB
MAX3232 U29 is being used to transmit & receive PC data from / To
CPU PCB Assembly Connector J28 for PC communication.
Opto 6N137: High Speed opto are being used for Transmit U30 &
Receive U32.
Connector J28 is connected to J4 on USB connector PCB assembly
ELLC2003R4
d. Barcode Scanner : SRGT
MAX3232 U31 is used to transmit & receive barcode data From / To
CPU pcb for Barcode scanner serial communication.
Connector J30: For SRGT Barcode scanner.
e. ISE Communication :
MAX3232 U28 is used for ISE serial communication.
Connector J27: ISE serial communication from ISE module to transmit &
receive ISE data from / to CPU PCB.
8. LLS Interface: LLS signals are opto isolated with the help of U26. as Auto
Zero – U26-A, LLS Out U26 -B, VOD through U26 – C.
Auto Zero signal is being generated from CPU PCB Assy during Probe is
positioned at top level of bottle, cup etc. Probe touches to liquid in cup,
bottle etc LLS signal is generated from LLS PCB Assy & send to CPU PCB
assembly for next action. During probe travelling any obstruction occurres
VOD signal is being generated from LLS PCB assembly & sent to CPU PCB
assembly for next action.
Location: Interface PCB Assy – ELS62009R7 is located on Base Plate at Front
Side with the help of Bracket.
Ref.: Component Layout Interface PCB Assy – ELS62009R7
measurements. The Analog data is being converted in to 16bit digital format with
the help of built in ADC of microcontroller C8051F067.
The processed data will be sending to the CPU board through PHM serial
communication via - interface board. It is located inside a black casting box below
the photometer diode PCB assembly beside the RCT Assy.
Location: Photometer signal Processor PCB assembly in Photometer Box.
Positive fixed voltage Regulator) TO220 Package to get +5V at TP6- Pin1 of
connector J9 w.r.t. GND Pin2 of connector J9
LED D4 indication for +5V
Rectified voltage regulated through LDO U4 – LM2990T-5.0/T (-5V Negative fixed
voltage Regulator) TO220 package to get -5V at TP7 w.r.t. GND pin2 of connector
J9.
LED D5 indication for -5V
D) ISOLATED SUPPLY: +5V_ISO
Transformer secondary output voltage 0-9 VAC at connector J4 on Power Supply
PCB is being rectified – BR4 & regulated through LDO U6 – TL75M05 (+5V
Positive fixed voltage Regulator) TO220 Package to get +5V _ISO at Pin J5 Pin1
w.r.t. ISO5V_GND Pin2 of connector J5.
LED D6 indication for +5V_ISO.
E) Ventilation Fan Supply: +12V
Transformer secondary output voltage 0-9 VAC at connector J11 on Power Supply
PCB is being rectified – BR5 & Filtered by C28. Unregulated +12V power supply at
connector J12 Pin1 +12V w.r.t. GND Pin2 of J12 is being used for Ventilation fan
of Blood Analyzer.
LED D7 indication for +12V ventilation fan Supply.
Location: Power Supply PCB is located on the bracket mounted below of
Photometer assembly mounting plate at RHS of Analyzer.
Refer: Silkscreen layout of Top - SST & bottom side – SSB of Power Supply PCB
ELS62011R3.
Action Observations
Go to the SERVICE CHECK >, click on Arm probe will be in up position at trough.
the ARM button. Click on ARM UP &
EXECUTE respectively.
Keep 20 ml empty bottle at position 1. Arm should rotate to position 1 of reagent tray.
In service check, execute command for
rotate arm to reagent inner.
Switch ON all the ARM TROUGH The empty 20 ml bottle should fill till the neck in
VALVE & ARM PROBE CLEAN approx. 30 seconds.
VALVE together. Observe a small fountain of water in trough.
During ARM PROBE CLEAN VALVE is ON,
observe the flow of water through probe as a
smooth laminar flow (Uninterrupted straight
jet).Verify that the water flow is smooth.
NOTE: Verify that all the valves turned off before exit from service
check menu to avoid any spillage of DI water.
Two temperature sensors are used for monitoring & control the temperature of the
rotor.
The terminals of the photometer lamp assy are mounted onto the base of RCT
Heater assy.
RCT Base RCT base consisted of Heater coils – 02 nos, Digital Temp. sensor
– 02 nos, Thermal cut off switch – 01 no. Through wire with Molex
connector Heater Supply & Temp. Sensor signals have been send
to Interface PCB assy for monitoring & control RCT temp. +37 Deg.
+/- 0.5 Deg.
Digital Temp. Sensor assy To sense RCT base temp. & sending to Interface pcb assy to
control temp. 37 Deg. +/- 0.5 Deg.
Heater Assy. Heater coil – 02 nos. fabricated in RCT base. Termination wires
taken out for Termination.
Lamp Assy Terminal. Lamp Assy terminals are mounted on Shim plate. Shim plate
mounted on RCT base. Lamp assy is connected to Lamp assy
terminals mounted on shim plate.
13. Maintenance
This section provides the necessary and minimal maintenance procedures in order
to ensure that the analyzer operates correctly and provides the accurate
measurement results.
5. Immerse the rotor in 3% nitric acid or Hydrochloric Acid solution and keep it
for 5 minutes.
9. Place the rotor in clean space by resting it on handles provided for lifting &
allow drying at room temperature.
NOTE:
1. Deproteinization is required only when tests for ions
such as magnesium, calcium etc are required.
2. Organic solvents (alcohol, benzene) or alkaline
solutions must not be used. They must be left to dry
completely before being reused.
3. High temperatures must not be used during drying.
4. The rotors must be rejected if they are noticeably
deteriorated. The optical status of a rotor can be
verified by means of the Reactions rotor verification
utility on the maintenance tab of MultiXL software. The
process is defined in section 10. Maintenance Menu.
5. The useful lifetime of each rotor depends drastically
on its use and care.
Regular maintenance of the analyzer will ensure trouble free operation and
consistent quality test results throughout its working.
wavelengths.
8 Perform a precision check and note down the %CV for an end point and
kinetic test.
9 Clean the barcode reader window in the piece of RGT base unit
10 Archive MultiXL database
11 Backup MultiXL database – main Database, Archive database
1. Check that the main switch located on the left hand side of the analyzer is at
"ON" position.
2. Check that the main fuses are not burnt.
3. When the main fuses are checked, turn the main switch off without fail and
then pull out the plug of power supply cable from its receptacle on the
analyzer. Open up the fuse cover and pull the fuses out.
4. Check that the circuit breaker of the power supply system to which the
analyzer is connected is not cut off.
Fuse cover
obstructing
movement
movement
3. Go to Service
Check: Arm
Menu. Click on
<Initialize>
button. Push the
probe gently to
cut the VOD opto
so that VOD
Error will be
generated and
Arm initializes.
4. If the
initialization or
VOD generation
fails, call the
Service Engineer
Arm 1D R1* Reagent 1 1. Reagent is not 1. Place the
absent - Pos. kept in the reagent reagent at the
bottle or reagent required reagent
bottle not kept at position
the defined position
2. Reagent is below 2. Check the
the Dead volume level of Reagent
and ensure that it
is above the
Dead volume
3. Arm position in 3. Call Service
reagent tray Engineer
4. LLS circuit and
its connector
problem
Arm 1D D* Diluent absent - 1. Reagent is not 1. Place the
Pos. kept in the reagent reagent at the
bottle or reagent required reagent
bottle not kept at position
the defined position
2. Reagent is below 2. Check the
the Dead volume level of Reagent
and ensure that it
is above the
Dead volume
3. Arm position in 3. Call Service
reagent tray Engineer
4. LLS circuit and
its connector
problem
Arm 1E @R1 Arm Initialize 1. Sample position 1. Switch OFF
Rotational error Opto signal the analyzer;
Rotate arm by
hand and make
sure that nothing
is obstructing
Arm rotation
4. Call Service
Engineer
Arm 2F @R2 Arm Initialize 1. Arm VOD opto, 1. Switch OFF
Up/Down error arm position optos, the analyzer;
up/down optos and Move SPT arm
rotation optos up and down by
hand and make
sure that nothing
is obstructing
SPT movement
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute
commands
3. Up/Down and 3. If the
rotation stepper initialization or
motor and its rotation fails, call
connection Service Engineer
Arm 312 @S Arm Rotational 1. position Opto 1. Switch OFF
Error - Trough to signal the analyzer;
Dilution Cuvette Rotate arm by
hand and make
sure that nothing
is obstructing
Arm rotation
2. Interface board 2. Then switch
and its connectors ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute
commands
3. Up/Down and 3. If the
rotation stepper initialization or
motor and its rotation fails, call
connection Service Engineer
Arm 313 @S Arm Rotational 1. position Opto 1. Switch OFF
Error - Trough to signal the analyzer;
Dilution Cuvette Rotate arm by
hand and make
sure that nothing
is obstructing
Arm rotation
2. Interface board 2. Then switch
and its connectors ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette >
Execute
commands
Arm rotation
Execute < R2
Cuvette to R1
Cuvette >
Execute
commands
3. Up/Down and 3. If the
rotation stepper initialization or
motor and its rotation fails, call
connection Service Engineer
Arm 324 @S Arm Rotational 1. Position Opto 1. Switch OFF
Error - Sample signal the analyzer;
to Cuvette Rotate arm by
hand and make
sure that nothing
is obstructing
Arm rotation
2. Interface board 2. Then switch
and its connectors ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to sample
outer> Execute
<Sample Outer to
R2 Cuvette>
Execute
commands
3. Up/Down and 3. If the
rotation stepper initialization or
motor and its rotation fails, call
connection Service Engineer
Arm 325 @S Arm Rotational 1. Position Opto 1. Switch OFF
Error - Sample signal the analyzer;
to Cuvette Rotate arm by
hand and make
sure that nothing
is obstructing
Arm rotation
2. Interface board 2. Then switch
and its connectors ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to sample
outer> Execute
<Sample Outer to
R2 Cuvette>
Execute
commands
3. Up/Down and 3. If the
Arm 352 @S Arm Down Error 1. Arm VOD opto & 1. Switch OFF
- Dilution arm optos and then switch
Cuvette ON the
instrument.
Check if the error
comes again
2. Interface card 2. Remove the
and its connector cover of the
sample arm and
check and clean
obstacle opto
3. Probe goes 3. Check the Arm
down but doesn’t assembly
find LLS signal due
to problem in LLS
card or its
connector
4. Probe assembly 4. Go to [Service
Check: Arm]
menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <Arm
Down> Execute
commands
5. If still it is
giving error, call
Service Engineer
Arm 353 @S Arm Down Error 1. Arm VOD opto & 1. Switch OFF
- Dilution arm optos and then switch
Cuvette ON the
instrument.
Check if the error
comes again
2. Interface card 2. Remove the
and its connector cover of the
sample arm and
check and clean
obstacle opto
3. Probe goes 3. Check the Arm
down but doesn’t assembly
find LLS signal due
to problem in LLS
card or its
connector
4. Probe assembly 4. Go to [Service
Check: Arm]
menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <Arm
Down> Execute
commands
5. If still it is
giving error, call
Service Engineer
Arm 356 @S Arm Down Error 1. Arm VOD opto & 1. Switch OFF
- Dilution arm optos and then switch
Cuvette ON the
instrument.
Check if the error
comes again
2. Interface card 2. Remove the
and its connector cover of the
sample arm and
check and clean
obstacle opto
3. Probe goes 3. Check the Arm
down but doesn’t assembly
find LLS signal due
to problem in LLS
card or its
connector
4. Probe assembly 4. Go to [Service
Check: Arm]
menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <Arm
Down> Execute
commands
5. If still it is
giving error, call
Service Engineer
Arm 357 @S Arm Down Error 1. Arm VOD opto 1. Switch OFF
- Dilution and arm optos and then switch
Cuvette ON the
instrument.
Check if the error
comes again
2. Interface card 2. Remove the
and its connector cover of the
sample arm and
check and clean
obstacle opto
3. Probe goes 3. Check the Arm
down but doesn’t assembly
find LLS signal due
to problem in LLS
card or its
connector
4. Probe assembly 4. Go to [Service
Check: Arm]
menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <Arm
Down> Execute
commands
5. If still it is
giving error, call
Service Engineer
Arm 36 @S Arm Up error - 1. Arm VOD opto, 1. Switch OFF
Sample position arm position optos, the analyzer;
up/down optos and Move SPT arm
rotation optos up and down by
hand and make
sure that nothing
is obstructing
SPT movement
2. Probe assembly 2. Then switch
ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to Sample
outer> Execute
<Arm Down>
Execute <Arm
Up> Execute
commands
3. Interface card 3. If the
and its connector initialization or
4. Up/down and up/down
rotation stepper movement fails,
motor and its call Service
connections Engineer
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <Arm
Down> Execute
<Arm Up>
Execute
commands
3. Interface card 3. If the
and its connector initialization or
4. Up/down and up/down
rotation stepper movement fails,
motor and its call Service
connections Engineer
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <Arm
Down> Execute
<Arm Up>
Execute
commands
3. Interface card 3. If the
and its connector initialization or
4. Up/down and up/down
rotation stepper movement fails,
motor and its call Service
connections Engineer
Arm 37 @S Arm Down error 1. Arm VOD opto & 1. Switch OFF
- Sample arm optos and then switch
Cuvette ON the
instrument.
Check if the error
comes again
4. If still it is
giving error, call
Service Engineer
Arm 39 @S Arm Down error 1. Arm VOD opto 1. Switch OFF
- Trough during and arm optos and then switch
sample ON the
operation instrument.
Check if the error
comes again
2. Interface card 2. Remove the
and its connector cover of the
sample arm and
check and clean
obstacle opto
4. Probe assembly 3. Check the Arm
assembly
4. If still it is
giving error, call
Service Engineer
Arm 3A !S Arm VOD error - 1. VOD Opto 1. Check whether
Sample Pos. Sensor the selected
option is tube and
instead of tube a
cup is placed
2. The connectors 2. Check the Arm
alignment in
[Service Check].
If it is hitting at
the edge, then
align the probe
using the
calibrate facility.
3. Sample arm 3. Remove the
position in Sample cover of the
tray sample arm and
check and clean
VOD opto
4. Go to Service
Check: Arm
Menu. Click on
<Initialize>
button. Push the
Probe gently to
cut the VOD opto
so that VOD
Error will be
generated and
Arm initializes.
5. If the
initialization or
VOD generation
fails, call the
Service Engineer
Arm 3A2 !DILN Arm VOD Error - 1. VOD Opto 1. Check whether
Dilution Cuvette Sensor the selected
option is tube and
instead of tube a
cup is placed
2. The connectors 2. Check the Arm
alignment in
[Service Check].
If it is hitting at
the edge, then
align the probe
using the
calibrate facility.
3. Sample arm 3. Remove the
position in Sample cover of the
tray sample arm and
check and clean
VOD opto
4. Go to Service
Check: Arm
Menu. Click on
<Initialize>
button. Push the
Probe gently to
cut the VOD opto
so that VOD
Error will be
generated and
Arm initializes.
5. If the
initialization or
VOD generation
fails, call the
Service Engineer
Arm 3A3 !DILN Arm VOD Error - 1. VOD Opto 1. Check whether
Dilution Cuvette Sensor the selected
option is tube and
instead of tube a
cup is placed
2. The connectors 2. Check the Arm
alignment in
[Service Check].
If it is hitting at
the edge, then
Error will be
generated and
Arm initializes.
4. If the
initialization or
VOD generation
fails, call the
Service Engineer
Arm 3B DILN! Arm VOD error - 1. VOD Opto 1. Check the arm
During Diluted Sensor alignment in
Sample [Service Check].
dispense If it is hitting at
the edge , then
align the probe
using the
calibrate facility.
2. The connectors 2. Remove the
cover of the arm
and check and
clean VOD opto
3. Sample arm 3. Go to Service
position in Check: Arm
Dispense cuvette Menu. Click on
during dilution <Initialize>
button. Push the
probe gently to
cut the VOD opto
so that VOD
Error will be
generated and
Arm initializes.
4. If the
initialization or
VOD generation
fails, call the
Service Engineer
Arm 3B0 S! Arm VOD Error - 1. VOD Opto 1. Check the arm
Cuvette Sensor alignment in
[Service Check].
If it is hitting at
the edge , then
align the probe
using the
calibrate facility.
2. The connectors 2. Remove the
cover of the arm
and check and
clean VOD opto
3. Sample arm 3. Go to Service
position in Check: Arm
Dispense cuvette Menu. Click on
<Initialize>
4. Call Service
Engineer
Arm 3F @S Arm Initialize 1. Arm VOD opto, 1. Switch OFF
Up/Down error arm position optos, the analyzer;
up/down optos and Move SPT arm
rotation optos up and down by
hand and make
sure that nothing
is obstructing
SPT movement
Reagent Tray];
Give <Initialize>
command.
3. Interface card 3. If initialization
and its connector fails, call Service
Engineer
SRGT tray 77 @SRGT SRGT Tray ARM assembly UP 1. Make sure that
rotational error- opto & Direction arm assembly is
Due to arm opto. not down in RGT
down in Tray tray during
during R1 service check for
aspiration reagent i.e.
service command
1) RGT initialize
2) RGT tray
rotate to “X”
position. If arm
assembly is not
down then check
up opto &
direction opto
signal.
2. Check opto
connectors are
connected
properly or not.
SRGT tray 78 @SRGT SRGT Tray ARM assembly UP 1. Make sure that
rotational error- opto & Direction arm assembly is
Due to arm opto. not down in RGT
down in Tray tray during
during R2 service check for
aspiration reagent i.e.
service command
1) RGT initialize
2) RGT tray
rotate to “X”
position. If arm
assembly is not
down then check
up opto &
direction opto
signal.
2. Check opto
connectors are
connected
properly or not.
SRGT tray 7A @SRGT SRGT Tray ARM assembly UP 1. Make sure that
rotational error- opto & Direction arm assembly is
Due to arm opto. not down in RGT
down in Tray tray during
during Sample service check for
aspiration reagent i.e.
service command
1) RGT initialize
2) RGT tray
rotate to “X”
position. If arm
assembly is not
down then check
up opto &
direction opto
signal.
2. Check opto
connectors are
connected
properly or not.
SRGT tray 7F @SRGT SRGT Tray ARM assembly UP 1. Make sure that
Interlock Error opto & Direction arm assembly is
opto. not down in RGT
tray during
service check for
reagent i.e.
service command
1) RGT initialize
2) RGT tray
rotate to “X”
position. If arm
assembly is not
down then check
up opto &
direction opto
signal.
2. Check opto
connectors are
connected
properly or not.
Arm 81 R1& Froth Sensed at 1. Froth in reagent 1. Remove the
Reagent Pos bottle froth in reagent
bottle.
2. Reagent is near 2. Check the
the Dead volume level of Reagent
and ensure that it
is above the
Dead volume.
3. LLS circuit and 3. Ensure that
its connector ARM is aligned at
problem the center of the
bottle.
4. Call Service
Engineer
SRGT tray 82 @SRGT SRGT tray ARM assembly UP 1. Make sure that
rotational error opto & Direction arm assembly is
opto. not down in RGT
tray during
service check for
reagent i.e.
service command
1) RGT initialize
2) RGT tray
rotate to “X”
position. If arm
assembly is not
down then check
up opto &
direction opto
signal.
2. Check opto
connectors are
connected
properly or not.
SRGT tray 8F @SRGT SRGT tray ARM assembly UP 1. Make sure that
interlock error opto & Direction arm assembly is
opto. not down in RGT
tray during
service check for
reagent i.e.
service command
1) RGT initialize
2) RGT tray
rotate to “X”
position. If arm
assembly is not
down then check
up opto &
direction opto
signal.
2. Check opto
connectors are
connected
properly or not.
RCT Tray 91 RCT1 RCT tray 1. Position opto 1. Switch OFF
Initialize assembly of RCT the analyzer;
Rotational error tray Rotate RCT by
hand and make
sure that nothing
is obstructing the
rotation
2. Stepper motor 2. Then switch
and its connections ON the
instrument; Go to
[Service Check:
RCT Tray]; Give
<Initialize>
command
3. Interface card 3. If initialization
and its connector fails, otherwise
4. Arms are down call Service
in RCT Engineer
RCT Tray 97 RCT4 RCT rotational ARM assembly UP 1. Make sure that
error -Due to opto & Direction arm assembly is
R1 dispense opto. not down in RCT
interlock tray during
service check for
reagent i.e.
service command
1) RCT initialize
2) RCT tray
rotate to “X”
position. If arm
assembly is not
down then check
up opto &
direction opto
signal.
2. Check opto
connectors are
connected
properly or not.
RCT Tray 98 RCT6 RCT rotational . ARM assembly 1. Make sure that
error - Due to UP opto & Direction arm assembly is
R2 dispense opto. not down in RCT
interlock tray during
service check for
reagent i.e.
service command
1) RCT initialize
2) RCT tray
rotate to “X”
position. If arm
assembly is not
down then check
up opto &
direction opto
signal.
2. Check opto
connectors are
connected
properly or not.
Hardware Error 99 HErr Hardware Error Check the Take the
assembly which corrective action
gave error prior to for the error
Hardware Error. which occurred
prior to Hardware
Error
RCT Tray 9F RCT3 RCT Interlock ARM assembly UP 1. Make sure that
error opto & Direction arm assembly is
opto & ARM home not down in RCT
opto tray during
service check for
RCT tray i.e.
service command
1) RCT initialize
2) RCT tray
rotate to “X”
position
2. Check opto
connectors are
connected
properly or not.
Arm A1 @STR2 Arm Rotational 1. position Opto 1. Switch OFF
Error for R2 signal the analyzer;
stirring - Trough Rotate arm by
to Cuvette hand and make
sure that nothing
is obstructing
Arm rotation
2. Interface board 2. Then switch
and its connectors ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <R2
Cuvette to
Trough> Execute
commands
3. Up/Down and 3. If the
rotation stepper initialization or
motor and its rotation fails, call
connection Service Engineer
Arm A2 @R1 Arm Down error 1. Arm VOD opto & 1. Switch OFF
- Cuvette during arm optos and then switch
R1 operation ON the
instrument.
Check if the error
comes again
4. If still it is
giving error, call
Service Engineer
Arm A3 @STR2 Arm Rotational 1. position Opto 1. Switch OFF
Error for R2 signal the analyzer;
stirring - Cuvette Rotate arm by
To Trough hand and make
sure that nothing
is obstructing
Arm rotation
2. Interface board 2. Then switch
and its connectors ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <R2
Cuvette to
Trough> Execute
commands
3. Up/Down and 3. If the
rotation stepper initialization or
motor and its rotation fails, call
connection Service Engineer
Arm A4 @STR2 Arm Up error - 1. Arm VOD opto, 1. Switch OFF
Trough during arm position optos, the analyzer;
R2 Stirring up/down optos and Move arm up and
rotation optos down by hand
and make sure
that nothing is
obstructing Arm
movement
2. Probe assembly 2. Then switch
ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
Down in Trough>
Execute <Arm
Up> Execute
commands
3. Interface card 3. If the
and its connector initialization or
4. Up/down and up/down
rotation stepper movement fails,
motor and its call Service
connections Engineer
connected
properly or not.
RCT Tray B4 RCT Interlock ARM assembly UP 1) RCT initialize
Error - Due to opto & Direction 2) RCT tray
Arm up Error at opto & ARM home rotate to “X”
Sample Cuvette opto position
during Stirring
3.)Move arm to
RCT and down in
RCT
If initialization or
rotation fails call
service engineer
2. Check opto
connectors are
connected
properly or not.
RCT Tray B5 RCT Interlock ARM assembly UP 1) RCT initialize
Error - Due to opto & Direction 2) RCT tray
Arm up Error at opto & ARM home rotate to “X”
R2 Cuvette opto position
during Stirring
3.)Move arm to
RCT and down in
RCT
If initialization or
rotation fails call
service engineer
2. Check opto
connectors are
connected
properly or not.
Arm C1 @STR1 Arm Rotational 1. position Opto 1. Switch OFF
Error for Sample signal the analyzer;
stirring - Trough Rotate arm by
to Cuvette hand and make
sure that nothing
is obstructing
Arm rotation
2. Interface board 2. Then switch
and its connectors ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <R2
Cuvette to
Trough> Execute
commands
3. Up/Down and 3. If the
rotation stepper initialization or
motor and its rotation fails, call
Arm C2 @STR2 Syringe Down Syringe home opto 1.Give init
error for signal. command. 2.Give
Reagent 2 syringe up /down
command.
3. If the
initialization or
up/ down
operation fails,
call Service
Engineer
Arm C3 @STR1 Arm Rotational 1. position Opto 1. Switch OFF
Error for Sample signal the analyzer;
stirring - Cuvette Rotate arm by
To Trough hand and make
sure that nothing
is obstructing
Arm rotation
2. Interface board 2. Then switch
and its connectors ON the
instrument; Go to
[Service Check:
Arm] menu; Give
<Initialize> and
<Arm Up>
Execute <Arm
trough to R2
Cuvette>
Execute <R2
Cuvette to
Trough> Execute
commands
3. Up/Down and 3. If the
rotation stepper initialization or
motor and its rotation fails, call
connection Service Engineer
Arm C4 @STR1 Arm Up error - 1. Arm VOD opto & 1. Switch OFF
Trough during arm optos and then switch
Sample Stirring ON the
instrument.
Check if the error
comes again
2. Interface card 2. Remove the
and its connector cover of the
sample arm and
check and clean
obstacle opto
3. Probe assembly 3. Check the Arm
assembly
4. If still it is
giving error, call
Service Engineer
Arm C7 @STR1 Arm Down error 1. Arm VOD opto & 1. Switch OFF
- Cuvette during arm optos and then switch
screen by clicking on Reports > Error Log button. The following screen will be
displayed:
1. The user can select the date range by changing the From and To Date.
2. The user can select operation (Service, Maintenance, Run or All operations)
during which the errors occurred.
3. Once the above selection is done, user needs to click on SHOW button to
view all the errors.
4. In the grid following are the different fields present:
a. Date - Date and Time of the occurrence of the error
b. Error Code: Displays the error code for the errors occurred.
c. Description - Description of the error occurred
d. Action - Action taken on the error occurrence is displayed in this column
5. To print the details of the error log, user can click on PRINT button
19 RANG H
1. This flag is issued with patient and control serum results to
indicate that the absorbance of the sample is higher than
the absorbance of the highest concentration calibrator in
the calibration table for increasing direction test.
2. This flag is also issued with patient and control serum
results if the absorbance of the sample is lower than the
absorbance of the highest concentration calibrator in the
calibration table for decreasing direction test.
20 RANG L 3. This flag is issued with patient and control serum results to
indicate that the absorbance of the sample is lower than
the absorbance of the blank (or lowest concentration
calibrator) in the calibration table for increasing direction
test.
4. This flag is also issued with patient and control serum
results if the absorbance of the sample is higher than the
absorbance of blank (or lowest concentration calibrator) for
a decreasing direction test.
21 H Measured value is larger than upper limit set for normal value
range for the corresponding age, sample type and category.
22 L Measured value is smaller than lower limit set for normal value
range for the corresponding age, sample type and category.
23 CALC! Calculation Item calculation does not take place for any of the
following reasons
5. Denominator is 0 (zero) in the process of calculation for
compensation.
6. The test to be used for Calculation Item has not been
measured yet.
7. Any test to be used for Calculation Item has data/calibration
alarms (such as Chk Calib)
8. Any test to be used for Calculation Item errors (S*, R1* etc)
24 ABSLIM 9. The absorbance value between M2S and M2E exceeded
the Reaction Absorbance limit.
10. This flag should be issued only for End-Point Chemistries
(For Rate-Chemistries, this flag will not be issued due to
extension logic program)
25 PANL Low Panic value error. This flag is issued with a sample result
to indicate that the patient result is lower than the programmed
Panic Limit Min. ISE tests too will be sent for a rerun
26 PANH High Panic value error. This flag is issued with a sample result
to indicate that the patient result is higher than the
programmed Panic Limit Max. ISE tests too will be sent for a
rerun
27 LINxx Linearity abnormal (checked only for Rate A and Rate B
assays). When the reaction during measurement points M2S
and M2E is non-linear beyond the set limit for linearity of
reaction this flag is given and the percent linearity of reaction
is indicated by a two digit number xx after “LIN”.
28 Lim0 This flag is applicable for Rate Chemistries, only during the
extension logic and when Reaction Absorbance Limit is
present. If there are no points available for calculation, then
this flag is issued
29 Lim1 This flag is applicable for Rate Chemistries, only during the
extension logic and when Reaction Absorbance Limit is
present. If there is only one point available for calculation, then
this flag is issued
30 Lim2 This flag is applicable for Rate Chemistries, only during the
extension logic and when Reaction Absorbance Limit is
present. If there are 2 points available for calculation, then this
flag is issued
31 ??? This flag is issued when the denominator becomes zero
during calculation or an overflow error occurs in logarithmic or
exponential calculation
32 Rgt Abs This flag indicates that the reagent 1 absorbance is lower than
Min the programmed Reagent Absorbance Min
33 Rgt Abs This flag indicates that the reagent 1 absorbance is greater
Max than the programmed Reagent Absorbance Min
34 CD This flag indicates that a clot has been detected during
sampling for that test. The result “NA” is associated with the
flag.
35 @TMP This flag is issued when the RCT temperature was out of
range while the measurement was in process.
36 TO This flag indicates Time Out while receiving result from the
machine.
37 Cal* This flag is issued with the patient and control results to
indicate that the results are being calibrated with previous
calibration data.
38 EVH Extrapolation % is set in system parameters
This flag is issued when sample result is higher than the conc.
of the highest calibrator then this flag is attached to the patient
result.
39 EVL Extrapolation % is set in system parameters
This flag is issued when sample result is lower than the conc.
of the Lowest calibrator then this flag is attached to the patient
result.
40 S$ This flag is issued when the froth is detected on sample
position. If this error occurred three consecutive times, then
the tests for that particular sample will be stopped and goes
into pending list.
43 W* This flag is issued with the result when the detergent wash
solution is found absent during carry over pair process.
44 COMM* This flag is issued with the result if any reading is missed out
between the measurement point (M1 Start and M2 Start).
This menu enables the user to view the cuvette blank absorbance values (obtained
with DI water in the cuvette or in air) at any particular wavelength.
The screen displays the cuvette blank for the requested wavelength. Wavelength
can be selected by the pull-down option provided on the left side of the screen.
The <Next> and <Previous> buttons can also be used to view the cuvette blanks
for the next and previous wavelength. There is also a Graph option available for
display. The cuvette blank table consists of three sections.
3. Click OK.
On clicking, the probe wash procedure will be started.
4. After the action is completed the analyzer gets initialized.
13.5.3.2.3. Prime Wash
This option is used at the beginning of the day, if user wants to have priming again.
The SYRINGE valve of the Probe is kept ON and the Syringe pump is moved to
extreme down position to remove the air trapped inside the tubing’s and syringe
pump.
The following operation occurs after the button is clicked:
1. Machine Initializes.
2. The syringe valves probe clean opens.
3. Syringe pump is taken to its extreme down condition.
4. Syringe valve is made ON/OFF several times to create turbulence for probe
cleaning & trapped air removal.
5. After approximately 2 minutes, Syringe pump is initialized for its home
position & valve made OFF.
The following steps should be done to carry out the Reagent bottle calibration:
1. User should select the Reagent bottle type from the SELECT CONTAINER
TYPE list.
2. Pipette the exact amount displayed for the Dead Volume in the Reagent
Bottle.
3. Place the Reagent bottle according to the bottle type on the position
specified in the list.
4. Click on the Calibrate button.
5. If the reagent volume present in the specified reagent bottle is not
approximately equal to the volume with which it is to be calibrated, then
message “Calibrated Value Out of Range of Specified Value” is displayed.
6. If the calibration is within range then it is automatically stored in the
software.
The following steps should be done to carry out the Sample tube calibration:
1. User should select the Sample Container type from the SELECT
CONTAINER TYPE list.
2. Place the containers according to the container type on the position
specified in the list.
3. Pipette the exact amount of volume in the Sample container as per the
Dead Volume required.
4. Click on the Calibrate button.
5. Once the calibration process is completed then it is automatically stored in
the software.
13.5.3.3.3. For Default Calibration
The following steps should be done to reset the Dead Volume Calibration to
Default:
User should click on Default button.
1. Select the CONTAINER TYPES from the list for which the Dead volume
calibration needs to be reset.
13.5.3.4. ISE
This option is available only when Ion Selective Electrode unit is installed on the
analyzer to perform routing maintenance, purging, cleaning and calibration on the
ISE unit.
This cycle is used to calibrate the peristaltic pumps of the ISE Module.
BUBBLE CALIBRATION
The bubble calibration command is used to allow the module to reestablish
a baseline for detecting air-liquid interfaces. It can also be used as a
diagnostic tool to see if the bubble detector is functioning properly.
MAINTENANCE
This is used to clear fluid from the flow path of the ISE Module, and to
pause the Sip Cycle.
The Sip Cycle is used to refresh the Calibrant A in front of the electrodes.
Every 30 minutes after the last sample is run, the ISE Module will
automatically run a Sip Cycle. No command is required from the host
analyzer to initiate a Sip Cycle. The ISE Module automatically clears the
flow path, next the ISE Module dispenses 36 μL of Calibrant B into the
Sample Entry Port, and it pulls it past the electrodes using the waste pump.
The ISE Module then dispenses 95 μL of Calibrant A into the sample entry
port, and positions it in front of the electrodes.
PRIME A
This command is used to prime Calibrant A solution from the reagent pack.
PRIME B
This command is used to prime Calibrant B solution from the reagent pack.
SHOW BUBBLE CALIBRATION
This command is used to display the bubble calibration values.
SHOW LAST SLOPE CALCULATED
This command is used to display last calibration values for the Na, K, Cl,
and Li.
SHOW ISE PUMP CALIBATION
This command is used to display the ISE pump calibration values.
2. Click START.
On clicking, the following window will be displayed.
2. Click START.
Put the required amount of cleaning solution on the mentioned reagent and
sample position, and then click OK.
3. On clicking, the probe will check the level of wash solution. If not enough
wash solution found, it will abandon the operation.
4. If enough wash solution is found, probe will pick up 250µl of wash solution
and dispense in empty cup at position 4. This process will be repeated twice.
5. The probe will get at position 2 and aspirate approx. 300 µl of wash solution
and soak in the same position for approx. 3 minutes.
6. The probe will start gargling with wash solution by aspirating and dispensing
the wash solution 3 times.
7. Arm will initialize and get itself cleaned at trough position.
8. The probe will now get in to the empty cup at position 6 & dispense approx.
500µl of DI water.
9. Probe will gargle with dispensed DI water 3 times at this position.
10. Arm will initialize and get itself washed at trough.
11. User needs to remove the sample cups with the remains of wash solution
and DI water and dispose it off.
13.5.4.1. Arm
Initialize
This will initialize the arm assembly as per the available options selected from the
list.
Dummy Run
On clicking this button, user can define the Sample, Diluent and reagent position.
After defining the respective posintions, on clicking START button, Dummy run is
initiated. On clicking STOP button, Dummy run can be stopped when required.
Execute
This will move the respective assembly in the selected path. This is used to check
the motion of an arm assembly in a particular path. Also for checking the opto
conditions.
13.5.4.4. Syringe
Initialize:
This is used to initialize the selected syringe assembly.
Aspirate:
You can check the ASPIRATE steps by entering the desired volume in the field
provided. The maximum volume that can be entered for ASPIRATION movement
is 375ul.
Dispense:
You can check the DISPENSE steps by entering the desired volume in the field
provided. The maximum volume that can be entered for DISPENSE movement is
375ul.
13.5.4.6. Calibration
This section is used to Calibrate ARM centering at various applicable positions.
Pre-requisite for this operation is to bring the ARM to required position using the
ARM options.
Clock Wise: Set direction of rotation for ARM as Clockwise.
Anti Clock Wise: Set direction of rotation for ARM as Anticlockwise.
+++: This button allows Course step movement of ARM.
+: This button allows FINE step movement of ARM.
Accept: Saves the Course / FINE adjusted value of ARM in to local memory.
Values are not saved in Analyzer until “Save Calib.” Button is clicked.
Cancel: Cancels the locally Accepted Value(s) of Course / FINE adjustment(s) of
ARM.
Default: Cancels and reloads the default position of ARM at respective position of
calibration.
Save Calib: This option is mandatory after calibrating the ARM. The locally
Accepted values are Saved in the Analyzer Non-volatile memory and are used by
Analyzer henceforth.
13.5.4.8. Communication
This option is used for checking the communication between PC and the analyzer.
The service personnel select the COM port and then click on the Check button.
All waste liquid are discharged into the external tank for high concentration waste
liquid.
The Module is completely self-contained. All the sample and calibrant positioning
within the module is controlled by an integral microprocessor, which assures
reliable electrode operation and maximum lifetime. The electrolyte measurement
system’s microprocessor applies proprietary mathematical algorithms to electrode
output voltages, converting them to clinical units of mmol/L.
Urine Diluent: This is required for urine samples. Urine samples must be diluted
by a factor of 10 (urine sample to urine diluent ratio of 1:9) to perform urine
measurement. The operator must keep the urine diluent on the Reagent Tray.
14.12. Troubleshooting
Symptom Cause Corrective actions
Drift Error Flag 1. May occur when new electrode Purge Purge the Cal A and recalibrate the
Single the Calibrant A is installed. If the module.
Electrode electrode is new it may initially drift as it
rehydrates over the course of 15 minutes
Drift Error Flag 1. May occur when new electrode or Purge the Calibrant A and recalibrate
Multiple reagent pack is installed on system. the ISE Module
Electrode
2. Reference electrode Replace reference electrode and
recalibrate
Air in Sample 1. Insufficient sample pipetted into the Host instrument must deliver 70µl.
ISE Module sample entry port. Increase dispensed sample volume.
Air in Sample 1. Cal B and Cal A are segmented with a) Electrodes are not properly seated
and Cal A air or compressed. Check compression
plate, spring and seal. Remove and
reassemble electrodes.
2. Fibrin or salt is plugging the electrode a) Use Cleaning procedure for module.
flow path. b) Remove electrode and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
Air in Cal B and 1. Cal B and Cal A are segmented with a) Electrodes are properly seated.
Air in Cal A air Check compression plate, spring and
seal.
b) Ensure that all electrodes and o-
rings are properly installed.
c) Ensure tubing between reagent pack
and sample entry port is connected
properly.
d) Replace tubing between reagent
pack and sample entry port.
e) Reagent low or out.
f) Use Cleaning procedure for module.
2. Fibrin or salt is plugging the electrode a) Use Cleaning procedure for module.
flow path. b) Remove electrodes and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
14.13.1. Pre-requisite
MultiXL software having ISE Reagent Pack Inventory feature with the New
ISE Reagent Pack. For existing partially utilized pack, the inventory will not
be correct.
NOTE: Software is using a counter to calculate ISE
inventory. Any manipulation, like releasing and attaching the
reagent pack connector, will falsify the inventory.
ISE Reagent Pack from one analyzer should not be interchanged with Pack
on other analyzer. In such event, the inventory will not be correct.
ISE Module should be always ON (for periodic sip cycles).
14.13.2. Procedure
Follow this procedure for installing the new reagent pack:
1. Connect the new Reagent Pack on the analyzer.
2. Click on ISE ON button in the Maintenance screen. The software will detect
the new pack and perform priming (Prime A and Prime B) operation
automatically till ISE module sends OK status (air removed), maximum up to
10 times.
3. Once the auto priming operation performed correctly, the lot number and expiry
date of the new reagent pack will be displayed on the Lot No. and Expiry Date
fields along with the following information:
ISE Pack Installation Date
Lot Number
Expiry Date
Distributor code (This information is stored in the reagent pack’s Dallas
chip and not visible to user.)
2. Tubing
leakage
ISE ISEC* Air in Cleaner 1. Required 1. Change ISE Pack
volume 2. Ensure proper tubing
insufficient
2. Tubing
leakage
ISE ISEM* Air in Segment 1. Required 1. Change ISE Pack
volume 2. Ensure proper tubing
insufficient
2. Tubing
leakage
ISE ISEP* Pump Cal 1. Required 1. Change ISE Pack
volume 2. Ensure proper tubing
insufficient
3. Check motor wiring
2. Tubing
leakage 4. Change motor
3. Motor
problem
ISE ISEF* No Flow 1. Motor 1. Check motor wiring
problem 2. Change motor
2. Tube 3. Replace Tubing
choked/bent
ISE ISED* Bubble Detector 1. Bubble Change ISE Module
Detector
module
problem
3. Connect the USB end of platform cable to one of the USB ports of PC and
other end to the programmer box. Connect the flying wires adapter cable to
the 2mm connecter slot present on the programmer box. See figure below.
16. Select the M25P16 from the part name list box, and then click OK. See
figure below.
17. On clicking, the FPGA graphical chip will appear with the selected mcs file
name. Place the mouse pointer on the graphical chip, right click and choose
Program. See figure below.
18. The Programming Property dialog box will appears. Keep it as default
settings, and click OK. See figure below.
If any ERRORS are reported during the process, reprogram the CPU using
same method.
20. After successful programming process, turn OFF the power of CPU board
and place the jumper JP2 between 2 and 3. The CPU board is ready to
operate for the analyzer logic. Plug on CPU PCB assembly on interface
PCB assembly with appropriate orientation.
21. Follow the same steps for programming the CPU board connected on
interface board, without connecting the power connecter as CPU board gets
power up from interface board itself.
5. Once the appropriate parameters are selected, click Connect. This will make
the software connection with the target.
6. Select Download Hex File/Go/Stop tab, select the appropriate file from the
desired location, and then click Download.
2. Now, press the switch and keeping it pressed, expose it to barcode “ENTER”
shown in figure below.
Switch
Green light
ENTER
A melody sound will be heard and the beam will go OFF. Release the
switch. Press it again and keeping it pressed, expose it to barcode “ISP”
shown in figure below.
ISP
A melody sound will be heard and green light on the PCB will start
blinking, release the switch.
3. The following screen will be displayed for the first time.
5. Select option “Install from a list or specific location (Advanced)” and click
“Next”.
The following screen will be displayed.
11. Select folder “D200\D200 (E)\Tools\D200 (E)” from CD and double click on
the icon shown in figure below.
13. Select the “Virtual COM Port ” known from the Device manager shown in
figure above and Click “OK”
Fig 17
14. Go to the toolbar, select File > Open.
A progress bar shown in fig 21 below, will appear indicating that the .dat
file is being downloaded
17. Disconnect USB mini connector from CANMAX scanner and reconnect it
again. Press switch, and keeping it pressed expose it to barcode “ENTER”
shown in figure below.
Switch
ENTER
A melody sound will be heard. Release the switch. Press switch, and
keeping it pressed expose it to barcode “ENABLE” shown figure below.
ENABLE
A melody sound will be heard. Release the switch. Press switch, and
keeping it pressed expose it to barcode “Setting End” shown in fig 25,
below, a melody sound will be heard. Release the switch.
END
This ends the configuration process of Canmax Barcode scanner.
3. Using your right hand, loosen the rotor knob in anticlockwise direction.
5. Now remove the four screws available on the retaining plate using screwdriver.
9. Now remove the lamp from the lamp holder as shown in figure below.
Lamp
10. Take new lamp clean it properly to avoid any finger impression. Place
halogen lamp gently and tighten the grub screw. Fix lamp connection and
tighten Lamp terminals.
5. Remove the 2 pin connector from the LLS (Liquid Level Sensing) PCB.
6. Remove the probe tube gently. Ensure that water does not spill on the LLS
(Liquid Level Sensing) PCB.
7. With the help of screwdriver, loosen the screw that is located on the front side of
the probe.
9. Clean the probe or take new probe, and insert the probe gently and properly
into its place. Now, tighten the front screw to fix the probe.
10. Insert probe tube back into its place and connect all connectors.
11. After replacing the probe, you must check the physical alignment of the probe.
Ensure probe height (Refer section 16.3 Probe Height Alignment Procedure
for more details).
Switch on the analyzer, Initialize all the assemblies. Check the internal
probe cleaning by switching on the respective Arm Probe Clean Valve,
ensure no leakage.
Confirm arm alignment before start the bio-chemistry run.
Reference 12mm
height gauge
2. Now initialize the arm assemblies using Service Check screen under Arm
section.
4. Check the gap or interference between probe tip and gauge center. Probe tip
should touch to 12mm gauge to get the require probe height.
5. Once it is done, fix the probe by tightening the probe screw.
7. Place bar-coded bottle at position 11 in Tray. Mount the Barcode reader such
that the beam falls at the center of position 11 of the Tray. Bar-coded bottle
position 11 will be facing the barcode scanner. The beam will go ON and
remain ON if the barcode scanner is not able to read the barcode. If alignment
is done proper that is if the barcode scanner is able to read the bar-coded tube
properly, the beam should go ON and OFF with beeping sound.
8. If the barcode scanner is not able to read the barcode tube at position 11, move
the scanner slightly towards the right from the center maintaining the distance of
70 +5mm, such that the scanner Beam goes OFF with beep and then continues
ON – OFF followed by BEEP
9. Place the inner bar-coded bottles, run the barcode scan in Service Check. To
run the barcode scan, click on “Barcode Scan”.
10. Now remove the inner bottles and place the outer bottles and again run the
barcode scan.
11. Finally, place all the bar-coded bottles and conform that both inner and outer
barcodes are read consistently for 10 times.
Probe calibration.
Time delay (Relay restart after cutoff)
Keypad lock.
Relay status on probe failure.
Restore factory defaults.
End Programming and keypad functions.
Use the following procedure to set the required parameters:
Serial
Action Observation
Number
1 In Multi-XL software go to <SERVICE LED on SN1: OFF, SN2,
CHECK > Select <ARM >< INITIALIZE>. SN3 opto PCB should be
ON.
16.6.7. PC Communication
This command will check the communication between PC & Analyzer. Select the
appropriate Serial port for communication - port 1, 2, 3, 4, virtual Com port & click
on <check>
If communication is being established between Analyzer & PC, then MultiXL
screen will display “Communication OK.” else will display “Communication Not
OK”
ADAPTOR OUTER
TUBES TEST PLASTIC (12MM X
120161
5 75MM) 10
REAGENT BOTTLE (20ml) ASSY.
182581
6 WITH LIGHT BROWN COLOUR CAP 25
REAGENT BOTTLE (50ml) ASSY.
182584
7 WITH DARK BROWN COLOUR CAP 25
8 182220 PHOTOMETER LAMP ASSY FOR XL 2
9 106846 ROTOR 120 CELL 5mm LIGHT PATH 10
10 100660 PVC BOAT SUPPORT 25
107180 25 MICRON FILTER FOR DI WATER
11 4LIT BOTTLE 2
KIT PREVENTIVE MAINTENANCE -
107540
12 2R XL100 1
13 101677 CLEANER FOR PROBE 1
FLOAT BASED LEVEL SENSOR -
WASTE (MFS VT 02 NO SH) with
107528
2Mtr. Wire, CONN. 2 PIN MOLEX
14 5557-02 & YELLOW SLEEVE _D13 1
FLOAT BASED LEVEL SENSOR - DI
WATER (MFS VT 02 NO SH) WITH
107527
2MTR. WIRE, CONN. 2 PIN MOLEX
15 5557-02 & BLUE SLEEVE _D13 1
18. Miscellaneous
This chapter deals with additional information as follows.
Control ON / OFF
Switch “ON”
4 Terminal Block (L & 220 ± 5VAC 110 ± 2.5VAC
N)
19. Appendix
19.1. TBM Image Installation
This section explains you the procedure of installing the TBM image (Windows
Embedded 7 Operating System) into the analyzer PC.
NOTE: The analyzer PC comes with preinstalled TBM image.
For any reason, if the TBM image is not installed or you want to upgrade the
existing operating system (OS) into TBM image, please read the following
information carefully for installing the TBM image to analyzer PC.
19.1.1. Scenario
There could be two scenarios under installing the TBM image:
New (Clean Installation)
Upgrade/ Reinstall
19.1.1.2. Upgrade/Re-installation
It is kind of an upgrade installation in which the analyzer PC already has some
operating system (for example: Windows XP, Windows 7 or previous TBM image)
installed. In this case, the TBM image setup will only format the C: drive and install
the TBM image on the same. This type of installation must be used if you want
to install TBM image on C: drive but want to keep the data on other drives
untouched.
1. Restart the PC and continuously press F12 (or key that is mentioned on the
screen to go to boot menu) to go to boot menu. Select system setup in the
boot menu to find all the options for the system setup.
2. Find the boot sequence option in the system setup and make sure that the
sequence of boot for CD-ROM is higher than hard drive.
3. If any changes are made to the existing sequence, click Apply to save the
changes and restart the PC.
Use the following steps to install the TBM image available on DVD to the analyzer
PC:
1. Insert the DVD containing TBM image setup into the DVD drive and restart
the PC.
2. It will show a message “Press any key to boot from CD or DVD”. As
suggested, press any key so as to boot from the DVD we entered.
3. Now, the PC will boot from the DVD containing TBM image. It will show the
Loading Windows... progress bar message on the screen.
4. After loading windows, wait till it shows command prompt window i.e.
X:\Windows\System32>.
5. Now, you need to get to the DVD and existing OS drive letter so that we can
start the TBM setup.
(Refer section 19.1.3 Identifying Various Drives for details on finding the
DVD and OS drive letter on the analyzer PC. Note down DVD drive letter
and Drive on which you want to install TBM image of OS).
6. Type the appropriate DVD drive letter followed by colon ‘:’ to enter into the
DVD drive.
For example, if E is the DVD drive letter then type the following command
and press Enter key.
X:\Windows\System32> E:
7. Now, type the following command in the command prompt window and press
Enter key.
TBMSetupWithClean
Caution: This will delete all the data from the PC.
Type y for yes to continue with the installation or type n to cancel the
installation.
On clicking n, the installation will be canceled and the command prompt will
be displayed.
8. After confirmation, it will clean the entire hard disk and then create two
partitions i.e. C and D drive with C drive size as 35 GB approximately, and
the remaining as D drive. It will then start applying the TBM image on the
PC. This process may take few minutes depending on the configuration of
PC.
9. Once the installation of TBM image is completed, a confirmation message
for the same will be provided on the screen and then the target computer will
restart automatically as shown below.
10. After restart, it might again ask to boot from DVD. Do not press any key and
let the PC continue to load the windows automatically. It is ok to take the
boot DVD out.
11. It will apply the required settings and may boot again.
12. Finally, it will ask to enter the product key. Enter the product key and click
Next button.
13. Follow and accept the remaining instructions to complete the TBM image
setup on the target computer.
NOTE: TBM setup image contains driver for specific hardware
configuration, especially for Network Interface and Display. If
the analyzer PC has different hardware components, you may
need to install the corresponding drivers from the PC’s
motherboard utility CD.
TBMSetup
A message will be asked to type the OS drive letter followed by colon.
For example, if C is the OS drive letter then type C: in the command
prompt and press Enter key. If you do not want to continue with the
installation then type exit and press Enter key.
After drive letter is entered in the command prompt, it will ask to confirm
the TBM image installation on the specified drive, type y for yes in the
command prompt to continue with the installation. If you do not want to
continue with the installation type n to exit and return to command prompt.
The same has been shown in the figure.
8. This will format the drive that was provided in step above and then start
installing the TBM image on the PC. This process may take few minutes
depending on the configuration of analyzer PC.
9. Once the installation of TBM image is completed, a confirmation message
for the same will be provided on the screen and then the analyzer PC will
restart automatically as shown below.
10. It might again ask to boot from DVD. Do not press any key and let the PC
continue to load the windows automatically. It is ok to take the boot DVD
out.
11. It will apply the required settings and may boot again.
12. Finally, it will ask to enter the product key. Enter the product key and click
Next button.
13. Follow and accept the remaining instructions to complete the TBM image
setup on the target computer.
NOTE:
1. In the upgrade installation scenario, if the previous
installation is Windows 7, it is possible that after
installing TBM image, it will show one extra volume
which was the hidden partition in previously
installed Windows 7 OS. The typical size of this
partition or volume is normally 100 to 200 MB.
2. TBM setup image contains driver for specific
hardware configuration, especially for Network
Interface and Display. If the analyzer PC has
different hardware components, you may need to
install the corresponding drivers from the PC’s
motherboard utility CD.
6. Cancel and Close the Notepad window after confirming the DVD drive
letter.
Cancel and Close the Notepad window after confirming the DVD drive letter