User Manual BTS 310 Plus
User Manual BTS 310 Plus
User Manual BTS 310 Plus
English
TEUS-00004-03-ING
May-2004
Esteemed customer,
Thank you for acquiring our photometer. Its features, we are sure, will make it a valued instrument
for your laboratory. Although it can be operated in a logical and simple way through its menu
options, we recommend you to read this manual carefully. It will help you for performing the
installation and the maintenance correctly, and will allow you to get the maximum benefit from its
many possibilities.
TABLE OF CONTENTS
1. INTRODUCTION .................................................................................... 7
2. DESCRIPTION OF THE INSTRUMENT ............................................... 8
2.1. DESCRIPTION OF THE KEYBOARD ............................................................. 8
2.1.1. Numeric Keys ....................................................................................................... 8
2.1.2. Function keys ....................................................................................................... 8
2.1.3. Cursor keys .......................................................................................................... 9
2.2. DISPLAY DESCRIPTION ................................................................................ 9
2.3. DESCRIPTION OF THE COMMUNICATIONS ................................................ 9
2.3.1. Communication connector .................................................................................. 9
2.3.2. Configuration of the communication ................................................................. 9
2.3.3. Information transmitted ....................................................................................... 9
2.4. VERSIONS PROVIDING PRINTER ................................................................ 9
2.5. VERSION PROVIDING SIPPER ..................................................................... 9
2.5.1. Sipping cycle ........................................................................................................ 9
2.5.2. Elements of the SIPPER.................................................................................... 10
2.5.3. Programmable parameters ................................................................................ 10
2.6. MEASUREMENTS WITH COMMON CUVETTES OR TUBES .................... 10
2.7. MODE OF OPERATION BY MENUS ............................................................. 10
2.7.1. Viewing a menu on the display and selecting an option ................................ 11
2.7.2. Selecting a job. Selection path. ....................................................................... 11
2.7.3. Exiting from a job ............................................................................................... 12
2.8. LOCATIONS FOR PROGRAMMING TESTS ................................................ 12
2.9. UNITS ............................................................................................................ 12
2.10. FILTERS ...................................................................................................... 12
2.11. INPUTTING DATA ....................................................................................... 12
2.11.1. Alphanumeric input .......................................................................................... 12
2.11.2. Numeric input ................................................................................................... 12
2.11.3. Selection by cursors ........................................................................................ 13
2.12. INTERNAL QUALITY CONTROL................................................................ 13
3. GENERAL METHOD OF OPERATION............................................... 14
3.1. STARTING OPERATION. THE MAIN MENU. ............................................... 14
3.2. ABSORBANCES ........................................................................................... 15
3.3. CONCENTRATIONS ..................................................................................... 16
3.3.1. Temporary modification of parameters ............................................................ 16
3.3.2. Heading listings (only versions with printer) .................................................. 16
3.3.3. Measurement process with base line .............................................................. 16
3.3.3.1. Process with reagent blank ........................................................................... 16
3.3.3.2. Process without reagent blank ..................................................................... 17
3.3.4. Measurement process with reagent blank...................................................... 17
3.3.5. Measurement procedures ................................................................................ 17
3.3.5.1. End point, increasing ..................................................................................... 17
3.3.5.2. End point, decreasing .................................................................................... 18
3.3.5.3. Delta kinetics ................................................................................................... 18
3.3.5.4. Differential mode ............................................................................................ 18
3.3.5.5. Fixed time ........................................................................................................ 18
3.3.5.6. Multistandard................................................................................................... 19
3.3.5.7. Multiple-test procedures ............................................................................... 19
3.3.6. Resolution in measurements, display, printer and calculations ................... 21
3.3.8. Error messages .................................................................................................. 21
3.3.7. Ending a measurement cycle ............................................................................ 21
3.4. PROGRAMMING ........................................................................................... 22
3.4.1. Programming tests ............................................................................................. 22
3.4.1.1. Creating a Test ................................................................................................ 22
3.4.1.1.1. General procedure ...................................................................................... 22
3.4.1.1.2. How to program parameters ....................................................................... 22
3.4.1.1.3. Exit menu ...................................................................................................... 26
3.4.1.2. Reviewing a Test ............................................................................................ 26
3.4.1.3. Delete a test .................................................................................................... 26
3.4.1.4. Copy a test ....................................................................................................... 26
3.4.1.5. Move a test ...................................................................................................... 27
3.4.2. Personalization of the photometer .................................................................. 27
3.4.2.1. Language ......................................................................................................... 27
3.4.2.2. Laboratory name (only in versions with printer) ......................................... 27
3.4.2.3. Screen contrast ............................................................................................... 28
3.4.2.4. Selection of the wash mode (only in versions with sipper) ....................... 28
3.4.2.5. List headings (only in versions with printer) ............................................... 28
3.4.2.6. Adjusting the printer (only in versions with printer) ................................... 28
3.4.2.7. Patient code .................................................................................................... 28
3.4.3. Programming the filters table ........................................................................... 28
3.4.4. Programming units ............................................................................................. 29
3.4.5. Programming communications ......................................................................... 29
3.4.5.1. Configuration of RS-232-C ............................................................................. 29
3.4.5.2. Default parameters ......................................................................................... 29
3.5. LISTING ......................................................................................................... 29
3.5.1. Listing the display .............................................................................................. 30
3.5.1.1. Listing tests ..................................................................................................... 30
3.5.1.2. Listing units ..................................................................................................... 30
3.5.2. Listing in the printer (only in versions with printer) ...................................... 30
3.5.2.1. Listing tests ..................................................................................................... 30
3.5.2.1.1. Listing of the tests names ........................................................................... 30
3.5.2.1.2. Listings of tests parameters ....................................................................... 30
3.5.2.2. Listing the personalization parameters ....................................................... 30
3.5.2.3. Listing the filters table ................................................................................... 30
3.5.2.4. Listing units ..................................................................................................... 31
3.5.2.5. Listing the communication parameters ........................................................ 31
3.5.2.6. Listing the pump parameters ......................................................................... 31
3.6. OTHER OPTIONS ......................................................................................... 31
3.6.1. Service ................................................................................................................ 31
3.6.2. Check lamp ......................................................................................................... 31
3.6.3. Adjust of the peristaltic pump (only in versions with sipper)........................ 31
3.6.3.1. Checking adjust of the sample volume ........................................................ 32
3.6.3.2. Manual adjust .................................................................................................. 32
3.6.3.2.1. Sample volume ............................................................................................. 32
3.6.3.2.2. Sample positioning ...................................................................................... 32
3.6.3.2.3. Pump delay time ........................................................................................... 32
3.6.3.3. Automatic adjust .............................................................................................. 33
3.6.3.3.1. Sample volume ............................................................................................. 33
3.6.3.3.2. Sample positioning ...................................................................................... 33
3.6.4. Load a new program .......................................................................................... 33
3.6.5. Delete the concentration values ..................................................................... 33
APPENDIX I: MENUS TREE. VERSION WITH PRINTER AND SIPPER. ................................................. 34
APPENDIX I: MENUS TREE. VERSION WITH PRINTER AND WITHOUT SIPPER. ................................ 35
APPENDIX I: MENUS TREE. VERSION WITHOUT PRINTER AND WITHOUT SIPPER. ......................... 36
1. INTRODUCTION
This photometer has been designed for use in "in vitro" The functionality of this photometer is completed by a series
diagnosis, and should therefore only be used by of automatic calculation modes that cover most of the
professionals. Its function is based on measuring the possibilities of photometric analysis: end point, differential
analyte concentration present in a sample by means of mode, fixed time and kinetics. The calculations also make
absorbency measurement, after application of one or several it possible to use standard or factor and bichromatics.
reagents.
The instrument is manufactured in three versions providing
Despite the large number and the complexity of tasks that different accesories:
the photometer is capable of performing, its operation is
extraordinarily simple through the menu system and the - Version with printer and sipper
underlying logic of its interface, so that an easy-to-assimilate - Version with printer and without sipper
dialogue is established between the user and the instrument. - Version without printer and without sipper
7
User Manual
The photometer sucks the sample up through a teflon tube. 2.1.2. Function keys
To accomplish this a peristaltic pump is employed that is
controlled by the microprocessor. When the sample has The function keys are those which allow the general operation
been positioned in the flow tray, the program will maintain of the program jobs or activate some general functions when
the programmed temperature thermostatically, if the using the photometer.
technique requires this. A bundle of monochromatic light is
passed through the sample to be measured and enters a ENTER/YES Is used to confirm the data input through the
photodiode which converts the light to a current. numeric keyboard, or to answer YES to the questions the
program may pose. Other specific uses are explained
Absorbency is a relative measurement, i.e. two throughout the manual.
measurements are made, one of the sample and the other,
normally of distilled water, is a reference measurement. ESC/NO Is used to quit one job in order to go to the
preceeding one in the program, such as to pass from one
Absorbency is calculated using the following relation: menu to the previous one, or to answer NO to the questions
the program may pose. Other specific uses are explained
Reference current throughout the manual.
Absorbency = log
Sample current
END This key lets you quit most of the jobs in the apparatus,
going to the main menu. It is useful to avoid going through
The concentration of the analyte of the sample will be the intermediate menus and to abandon long processes that
proportional to its measured absorbency. you do not desire but that you have started by error.
2.1. DESCRIPTION OF THE KEYBOARD WASH (only in versions with sipper) This key lets you
operate the pump independently of the program, allowing
2.1.1. Numeric Keys you to perform washes of the tubing and of the cuvette at any
time. This key may operate continuously or on a fixed
The keys 0 to 9, together with the decimal point and the C volume mode.
key, constitute the numeric keyboard. Keys 0 to 9 and the
decimal point serve for inputting numerical data when the PUMP (only in versions with sipper) This key is used to start
program requires it. The C key is used to correct the the SIPPING CYCLE, aspirating the sample to be measured.
The key has an indicator which lights up every time the pump
is ready to start a new cycle (it is not necessary to wait for
1 3
8
the display nor the printer in order to sip the next sample). In order to have a right communication between PC and the
photometer, all these parameters must be the same in both
2.1.3. Cursor keys apparatus
The Default values are:
There are four cursors on the keyboard, two called VERTI-
CAL CURSORS (CURSOR UP and CURSOR DOWN) and Bauds: 9600
two called HORIZONTAL CURSORS (CURSOR RIGHT and Timeout: 0
CURSOR LEFT). These cursors have diverse functions in Terminal: 0
2
the program according to the job the photometer is performing,
so they will be explained throughout the manual when 2.3.3. Information transmitted
necessary.
With this PC program two different types of information can
2.2. DISPLAY DESCRIPTION be transmitted.
The display is made of liquid crystal (LCD) and consists of 1- The memorised values of concentration measured. These
240 x 64 pixels. The distribution of text and data on it will results are applied to perform the final report.
depend on the job being performed.
2- The photometer program can be changed from the PC.
2.3. DESCRIPTION OF THE COMMUNICATIONS
2.4. VERSIONS PROVIDING PRINTER
The photometer results can be transmitted to a PC(Personal
Computer) through a communications cable. A PC program Some versions of the photometer include a thermo-printer,
is supplied with the photometer, allowing the management allowing text to be printed rapidly and silently. The printer
of all communications and data received between the PC works bi-directionally, optimizing the path of the head and
and the photometer. consequently the print time. Paper advance is performed
rapidly by means of an independent motor.
2.3.1. Communication connector
Text is printed in forty columns, producing an uncrowded
In the rear part of the apparatus, you can see the connector appearance. The print format will depend on each job.
(37) for the computer connection, in a box marked COM1.
The connector is a RS-232 with the following line connections:
2.5. VERSION PROVIDING SIPPER
All control lines and electrical levels of the serial connector
are compatible with the standard E.I.A. RS-232. The SIPPER has the duty of handling the samples to be
measured. It is one of the most delicate parts of the
2.3.2. Configuration of the communication instrument and must be treated with the utmost care. This
section explains the functioning of this circuit, its constituent
In the serial port the following parameters can be configured: elements and the parameters that control it.
-Baud rate, is the transmission velocity
-Timeout, is the resent time
-Terminal number
1 NC
5 1 2 RXD
3 TXD
4 NC
9 6 5 GND
9
User Manual
This cycle involves aspirating the sample to be measured Three parameters control the sipper functioning and they must be
and transporting it to the flow cuvette. It consists of three programmed to obtain the expected performance characteristics.
consecutive steps: a) SAMPLE VOLUME. It is a number that allows to adjust
the volume of sample that will be sipped.
SIPPING: The sample is aspirated in by the sip tubing up to
the programmed volume. b) PUMP DELAY. It is the number of seconds the pump will
wait from the moment the sipping finishes up to the
PUMP DELAY: Waiting period from the moment the sipping moment the pump activates again to position the sample.
finishes. The tube or vessel with the sample must be
pulled out the sip tubing during this time. c) POSITIONING. This is also a number that determines the
transportation of the sample to the cuvette, ensuring that
POSITIONING: The pump activates again transporting the it is suitably positioned to be read.
sample to the cuvette and positioning it suitably for
its measurement. These parameters depend on the length, the diameter and
the aging of the tubing, and must be controlled and adjusted
The measured samples are transported to the waste bottle periodically.
when new samples are sipped or a wash is performed.
1 CREATE NEW TEST At the end of selecting several options, a state is reached
2 REVIEW where the apparatus executes a job. For example, if we
3 DELETE select 1 (CREATE NEW TEST) from the previous menu, the
< 4 COPY following will be displayed:
In order to see the options below the last line, press the
CURSOR DOWN, and the list will move down one line, CREATE TEST
leaving: LOCATION ........ = _
1st available .... 25
< 2 REVIEW
3 DELETE
4 COPY
5 MOVE indicating that we are in the task of CREATE A NEW TEST.
Note that the same marker appears now in the first row, To start this task, it has been necessary to select three
indicating that there is more text above it. Press the consecutive options. The expression:
CURSOR UP to move up one line.
PROGRAMMING/TESTS/CREATE NEW TEST
In the case of a long list, when there is text above and below
the first and fourth rows of the display, respectively, these formed by the selected options, is called a PATH, and from
markers appear in both rows. this point on and throughout this manual, it will be the way
of indicating the set of consecutive options in the linked
Using the CURSORS UP and DOWN, all the options in any menus that have to be selected in order to reach the task
list can be viewed. being explained.
A menu option is selected by pressing the numeric key Sometimes, suspension points will appear in the middle of
corresponding to the number of the desired option. It is not a path (OPTION/.../OPTION), meaning that it is necessary
necessary to press ENTER. to input a parameter between both options.
11
User Manual
12
C and the complete number will be deleted. 2.12. INTERNAL QUALITY CONTROL
To confirm the number (and quit the process) press ENTER.
The number will be permanently stored. All clinical analysis laboratories should have internal quality
control procedures so that they may verify the correct
Often, the following titles appear on the display: functioning of each measurement procedure in use, and for
the validation or rejection of measurements made.
Actual value: Which is the current value of the parameter
(previous data input or default value). Since the photometer is a manual instrument, there is no
New value: The one which is input and that will be stored on special option for effecting such control. Thus, it becomes
pressing ENTER. the responsibility of laboratory personnel to carry out an
internal quality control programme, in which intercalated
control samples would be measured daily along with other
2.11.3. Selection by cursors samples for later comparison with their respective expected
values. In this way it would be possible to detect the
In some cases, the user must choose an item among a pre- presence of occasional anomalies in measurement
defined set of them. This can be achieved by means of the procedures. It is recommended that one employs control
HORIZONTAL or VERTICAL CURSORS according to the materials titrated by Biosystems.
case. By pressing the CURSORS, the different options will
appear on the display. The manner of confirming them is
usually by pressing ENTER, although it may differ from case
to case, and will be explained throughout the manual.
13
User manual
SELFTEST When errors are not detected in the selftest process, the
following display appears in about 5 seconds:
in process
Information checked Caution message <model> is the reference to the apparatus model
Ver n.m is the version of the program installed.
Serial number: is the serial number of the apparatus
Program
ROM ERROR Press ENTER and, in versions providing printer, the display
will show:
LOCATION CHECKSUM
ERROR
In case you do not desire to print the date press ESC/NO
Press ENTER to continue the instrument operation.
14
Seleccionar una de estas opciones, dependiendo del tra- c) Sample volume (only in versions with sipper). Input the
bajo que quiera realizarse (ver Apéndice I): volume (in mL) of sample to be sipped (section
3.4.1.1.2.m)
Select one of these options depending on the task you want d) Stabilization time (only in versions with sipper). Input the
to perform (see appendix I): time (in seconds) you want to elapse between the end of
the SIPPING CYCLE and the measurement (section
1. ABSORBANCES: To measure absorbances. 3.4.1.1.2.i).
e) Temperature. Input the temperature at which it is desired
2.CONCENTRATIONS: To measure concentrations to thermostat the cuvette or “0” for no thermostating
according to a predefined test which has been programmed (section 3.4.1.1.2.p).
in a location.
Once the temperature have been selected, the display will
3.PROGRAMMING: To programme tests and other show:
parameters which will control different functions of the
instrument.
15
User manual
- Samples consecutively numbered and their respective 3.3.2. Heading listings (only versions with printer)
absorbance values.
PATH: CONCENTRATIONS/.../LIST
3.3. CONCENTRATIONS
With this option, you can print again the heading of the Test
PATH: CONCENTRATIONS for the retrieved location. This is specially useful when
parameters have been modified in MODIFY PARAMETERS
This option allows the calculation of concentrations by
using any of the TESTS previously programmed and stored
in the 75 available LOCATIONS. On selecting this option, 3.3.3. Measurement process with base line
the following display appears:
PATH: CONCENTRATIONS/.../BASE LINE
Input the number of the location desired and press ENTER. 3.3.3.1. Process with reagent blank
In versions providing printer, the instrument will print the In the following group of procedures:
name of the test, the location number and the calculation
mode, as well as the values of the different parameters only - End point or multistandard, increasing slope
if this option has been selected in LIST HEADING (section - Differential mode
3.4.2.5.).
The absorbance is generally adjusted to ZERO against the
Once the location number has been input, two different REAGENT BLANK (section 3.3.4), but sometimes it is
menus may appear on the display, depending on the convenient to know the absorbance of the reagent blank.
calculation mode and on the reaction slope of the selected This value may be obtained by selecting BASE LINE in the
Test. menu. Upon selection, the display will show:
a) In measurement procedures where the reagent blank is
low and it is usual to adjust to ZERO with it, such as END
POINT and MULTISTANDARD (increasing) or
DIFFERENTIAL MODE, the following menu will appear: THERMOSTATING
Wait X minutes
1 MODIFY PARAMETERS
2 BASE LINE
3 REAGENT BLANK
4 LIST This display will not appear when the instrument has been
previously operating at the same temperature and the
b) In measurement procedures where the reagent blank is cuvette holder is already thermostated, nor when
not used (FIXED TIME, KINETICS) or has a high value thermostatation is not required.
(DECREASING reactions) and it is usual to adjust to ZERO
with water, the following menu will appear: Nextly, the display requires:
1 MODIFY PARAMETERS
2 BASE LINE Insert
3 LIST
BLANK, and the concentration is calculated by multiplying display requests: LIST DELTA?, offering the possibility of
the absorbance by a factor or with reference to the standard. reviewing the delta values (diagram B, appendix III). Answer
ENTER/YES to print them or ESC/NO and the display will
Factor: Cs = As x F request: DELETE DELTA?. You can eliminate one or
several deltas by pressing ENTER/YES and following the
Cst procedure:
Standard: Cs = As x
Ast a) Select the delta you wish to delete by using the VERTI-
CAL CURSORS.
The display shows the measured absorbances and the b) Press ENTER/YES to delete the delta on the display, or
calculated concentration of each sample. C to retrieve a previously deleted delta.
c) Once all the desired deltas have been deleted (or retrieved
In versions providing printer, the samples are consecutively if previously deleted), press ESC. The concentration will
numbered and their respective absorbance and concentration be re-calculated using the selected deltas and printed
are printed. again.
d) You may then list and select again deltas (pressing
3.3.5.2. End point, decreasing ENTER/YES) or go to the next sample by pressing ESC/
NO.
The absorbance of the reagent blank and of each sample is
read against the BASE LINE, and the concentration is 3.3.5.4. Differential mode
calculated by multiplying the difference between the
absorbances of the REAGENT BLANK and the sample by Two tubes are required per sample, one for the reaction of
a factor or with reference to the standard. the sample blank, and another for the overall reaction. After
adjusting to zero with the REAGENT BLANK, the instrument
Factor: Cs = (Arb - As) x F sequentially requires the sample blank and the corresponding
sample, in order to perform the measurements (diagram C,
Cst appendix III). The concentration of the sample is calculated
Standard: Cs = (Arb - As) x by substracting the absorbance of the sample blank from
(Arb - Ast) the absorbance of the sample, and multiplying by a factor
or with reference to the standard.
The display and printer provide the same data than in end
point, increasing. Factor: Cs = (As - Asb) x F
LIN LIN
POLYGONAL SPLINE
LIN LIN
LOG LOG
REGRESSION REGRESSION
LINE PARABOLA
LOG LIN
The instrument then waits for the interval time to perform the
Once zeroed with the base line, the display requests the second reading of the first and the following samples,
number of samples to be processed. The number must be displaying a waiting message. The display will show again
within the range displayed, a funtion of the delay and the the sample number and a beep warns that it must be sipped.
interval times programmed. Input the number and press A second reading of all the samples is thus obtained.
ENTER.
In the case of multi-kinetics, the process will continue until
The multiple-test process begins after pressing ENTER the programmed number of intervals is completed.
again. The display shows the number of the sample to be
inserted and beeps at the precise time it must be sipped; At the end of the measurement process, the instrument will
once the programmed delay time is completed, the instrument display and print (versions providing printer) the results
requests the second sample and sequentially the rest until obtained for each sample as described for the single-test
the frist reading of all the samples is accomplished. In test procedures (see Sections 3.3.5.3. and 3.3.5.5.). The results
using a standard, it is requested and processed at the of the diverse samples and the single delta values are
beginning of the test, like in the single-test procedures. available using the VERTICAL CURSORS.
20
The same standard curve is shown in different formats 3.3.7. Ending a measurement cycle
depending on the calculation mode and the axis scale.
Any measurement procedure may be ended at any moment
3.3.6. Resolution in measurements, display, by pressing END. The process will be interrupted and the
printer and calculations program will return to the MAIN MENU.
The absorbances are measured and printed (in versions If you are measuring a standard or samples and you press
providing printer) with a resolution of 0.0001 A, although they ESCAPE, the measurement will be interrupted and the
are shown on the display rounded to the third decimal. program will return to “insert REAGENT BLANK” or to “insert
Calculations are made taking the measured absorbances BASE LINE”.
with a resolution of 0.0001 A, and then round to a maximum
of 4 digits to express concentrations.
Sample value beyond range Absorbency is beyond the * Repeat sample reading
Absorbency > 2.200 range of the instrument
* Possible bubble in flow tray * Replace reagent
* Deteriorated reagent
21
User manual
This option of the MAIN MENU allows you to program any On creating a Test, the parameters appear with default
of the programmable parameters of the photometer. The values.
following menu appears on the display:
You can go from one display to the previous or to the next
1 TESTS one by means of the VERTICAL CURSORS, or forward by
2 PERSONALIZE means of the ENTER key, in which case the numeric
3 FILTERS TABLE parameters that had been input are confirmed and stored.
4 UNITS Programming finishes when you reach the last parameter
5 COMMUNICATIONS and you press ENTER (or by pressing ESCAPE at any
moment), the exit menu will appear:
WARNING 1 SAVE
The pre-programmed techniques are compati- 2 QUIT (NO SAVE)
ble with Biosystems reagents. Reagents whose 3 REVIEW
measurement procedures, as described in the
instructions for use, are compatible with the
specifications of the instrument may also be EPF = END POINT WITH FACTOR.
employed. EPS = END POINT WITH STANDARD.
DKF = DELTA KINETICS WITH FACTOR.
DKS = DELTA KINETICS WITH STANDARD.
DMF = DIFFERENTIAL MODE WITH FACTOR.
DMS = DIFFERENTIAL MODE WITH STANDARD.
3.4.1. Programming tests FTF = FIXED TIME WITH FACTOR.
FTS = FIXED TIME WITH STANDARD.
PATH: PROGRAMMING/TESTS MS = MULTISTANDARD
MKS = MULTINETIC WITH STANDARD
This option lets you program the parameters of a Test, in MKF = MULTINETIC WITH FACTOR
each of the 75 locations available in the photometer. On MFS = MULTIFIXED TIME WITH STANDARD
selecting TESTS, the following menu appears: MFF = MULTIFIXED TIME WITH FACTOR
1 CREATE NEW TEST
2 REVIEW
3 DELETE
3.4.1.1.2. How to program parameters
4 COPY
5 MOVE PATHS: PROGRAMMING/TESTS/CREATE NEW TEST
PROGRAMMING/TESTS/REVIEW
CONCENTRATIONS/.../MODIFY PARAMETERS
3.4.1.1. Creating a Test
The main function of the photometer is to calculate concentra-
PATH: PROGRAMMING/TESTS/CREATE NEW TEST tions, using the analysis Tests programmed within it and
stored at a numbered location. The input or modification of
the parameters that constitute these Tests is explained in
3.4.1.1.1. General procedure this section. A diagram in appendix IV illustrates the whole
procedure and also shows the different options and the value
This option allows you to program a new location with the limits for programming.
different parameters which constitute a Test. The program
indicates which is the first free location. Press ENTER to a) LOCATION
accept it or key in a different number. If the selected location
is already programmed, an information message appears Input, using the NUMERIC KEYBOARD, the number of
on the display. the location in which it is desired to create, review or
modify a Test. The display indicates which is the first
Once the location number has been selected, a series of available location in the case of CREATE. Press ENTER
displays are presented to program each parameter. The to confirm the selected location.
way of programming each one depends on the parameter,
and will be explained in the following section. The program To CREATE A NEW TEST, a non-programmed location
22
PARAMETER CALCULATIONMODE
TESTNAME * * * * * * * * * * * * *
UNITS * * * * * * * * * * * * *
CALCULATIONMODE * * * * * * * * * * * * *
READINGMODE * * - - - - - - * - - - -
REFERENCEFILTER * * - - - - - - * - - - -
READINGFILTER * * * * * * * * * * * * *
FACTOR * - * - * - * - - - * - *
STABILIZ. TIME # * * - - * * - - * - - - -
INCUBATIONTIME - - * * - - * * - * * * *
INTERVALTIME - - * * - - * * - * * * *
INTERVALSNUMBER - - * * - - - - - * * - -
SAMPLE VOLUME # * * * * * * * * * * * * *
REACTIONSLOPE * * * * - - * * * * * * *
STANDARD - * - * - * - * - * - * -
STANDARDNUMBER - - - - - - - - * - - - -
STANDARD REPLICATES - - - - - - - - * - - - -
SAMPLEREPLICATES - - - - - - - - * - - - -
AXIS* - - - - - - - - * - - - -
STANDARD CONCENTRATIONS - - - - - - - - * - - - -
DILUCTIONFACTOR - - - - - - - - * - - - -
TEMPERATURE * * * * * * * * * * * * *
DELAY - - - - - - - - - * * * *
Table 3.1 Requested parameters according to the calculation mode
23
User manual
Select using the HORIZONTAL CURSORS the calculation Only in the END POINT and MULTISTANDARD modes.
mode to be employed among the following available: Select by means of the HORIZONTAL CURSORS one of
the options.
END POINT WITH FACTOR
END POINT WITH STANDARD MONOCHROMATIC: A single measurement is made of
DELTA KINETICS WITH FACTOR each sample at the selected wavelength.
DELTA KINETICS WITH STANDARD
DIFFERENTIAL MODE WITH FACTOR BICHROMATIC: Two measurements of each sample are
DIFFERENTIAL MODE WITH STANDARD made (figure 3.2), one at the main wavelength (usually at
FIXED TIME WITH FACTOR the maximum absorbance of the liquid being measured),
FIXED TIME WITH STANDARD and another at the secondary or reference wavelength
MULTISTANDARD (generally far from the maximum absorbance of the liquid
MULTIKINETIC WITH FACTOR being measured). The absorbance value used for
MULTIKINETIC WITH STANDARD calculations is in this case the difference between both
MULTIFIXED TIME WITH FACTOR absorbances (A = Amain - Areference).
MULTIFIXED TIME WITH STANDARD
POLYGONAL 1 340
SPLINE 2 405
REGRESSION LINE 3 505
REGRESSION PARABOLA 4 546
5 600
Select by means of the HORIZONTAL CURSORS one of 6 635
the options and press ENTER. 7 670
8 ---
Abs A= Am - Ar
Am
Ar
l
main l reference l
Only in BICHROMATIC reading mode. Input by means of Only in the mode KINETICS. Number of intervals (deltas)
the NUMERIC KEYBOARD, or select using the HORI- desired. Equivalent to the total number of measurements
ZONTAL CURSORS, the secondary or reference minus 1. Input the number using the NUMERIC
wavelength in nanometers. If the filter with the selected KEYBOARD.
wavelength has not been programmed in the FILTERS
TABLE, an error message will appear in the display. Table m) SAMPLE VOLUME (Only in version with sipper)
3.3 shows the filters the photometer is usually equiped
with, and their respective positions. Volume (in mL) of sample to be sipped for the
measurements. Minimum programmable value is 100
g) READING FILTER mL. Take into account that the less the volume sipped,
the more will be the carryover contamination between
Input by means of the NUMERIC KEYBOARD, or select consecutive samples. Table 3.4 shows an evaluation of
using the HORIZONTAL CURSORS, the wavelength in carryover for different sample volumes. Take into account
nanometers at which it is desired to make the that these values may considerably vary depending on
measurements. In BICHROMATIC mode, it refers to the the nature of the liquid being measured. Input the volume
main wavelength (figure 3.2). If the filter with the selected using the NUMERIC KEYBOARD.
wavelength has not been programmed in the FILTERS
TABLE, an error message will appear in the display. Table
3.3 shows the filters the photometer is usually equipped
with, and their respective positions. SAMPLE VOLUME (%)
o) STANDARD
j) INCUBATION TIME
Appears for the calculation modes using a single stan-
Time elapsed from pressing ENTER (or PUMP in version dard. Input, by means of the NUMERIC KEYBOARD, the
with sipper) and the first measurement in the modes concentration value of the standard, in the same units
KINETICS and FIXED TIME. Input the time using the programmed to express the results.
NUMERIC KEYBOARD.
p) STANDARD NUMBER
k) INTERVAL TIME
Appears for the calculation mode MULTISTANDARD.
Time elapsed between two consecutive measurements of Input the total number of standards to be used by means
absorbance in the mode KINETICS, and time elapsed of the NUMERIC KEYBOARD.
between the first and the second measurement in the
mode FIXED TIME. Input the time using the NUMERIC q) STANDARD REPLICATES
KEYBOARD.
Appears only in the mode MULTISTANDARD. Input the
number of replicates to be required for each standard and
for the reagent blank using the NUMERIC KEYBOARD.
25
User manual
r) SAMPLE REPLICATES parameters that may have been input, select the option
QUIT (NO SAVE).
Appears only in the mode MULTISTANDARD. Input the
number of replicates to be required for each sample using It is also possible to review the programmed parameters by
the NUMERIC KEYBOARD. selecting REVIEW. You can modify any value during this
process and validate it at the end of the reviewing by
s) AXIS selecting SAVE.
26
specially useful to program a test that is very similar to an * Version with printer and sipper
existing one. The test can be copied in a new location and 1 LANGUAGE
afterwards it can be reviewed to modify a few parameters. 2 SCREEN CONTRAST
To make the copy, the following display appears: 3 LABORATORY NAME
4 WASH MODE
5 HEADING LIST
COPY TEST 6 PRINTER CONTROL
Source location = _ 7 PATIENT CODE
Target location =
* Version with printer and without sipper
1 LANGUAGE
Enter into “Source location” the number of the location you 2 SCREEN CONTRAST
want to copy, and into “Target location” the new location 3 LABORATORY NAME
where you want it to be copied. On pressing ENTER a 4 HEADING LIST
second time, the program copies the contents of the source 5 PRINTER CONTROL
location into the target location. 6 PATIENT CODE
If the source location is not programmed, as well as if the * Version without printer and without sipper
target location is already programmed, caution messages
are issued and the program does not confirm the location 1 LANGUAGE
numbers until they are correct. 2 SCREEN CONTRAST
3 PATIENT CODE
3.4.1.5. Move a test
If the source location is not programmed, as well as if the 3.4.2.2. Laboratory name (only in versions
target location is already programmed, caution messages with printer)
are issued and the program does not confirm the location
numbers until they are correct. PATH: PROGRAMMING/PERSONALIZE/LABORATORY
NAME
3.4.2. Personalization of the photometer
This option enables you to program a title of maximum 10
PATH: PROGRAMMING/PERSONALIZE alphanumeric characters, using the ALPHANUMERIC
INPUT procedure. Once the title has been input, press
This option enables the user to adapt the photometer to his ENTER.
personal preferences in some aspects. A diagram in
appendix VI illustrates the programming process of these The programmed text will be printed each time the photometer
personalized parameters. With this option, one of the is switched on.
following menus appears:
27
User manual
3.4.2.3. Screen contrast This option enables you to adjust the intensity of black
print on the thermal paper of the printer in such a way that
PATH: PROGRAMMING/PERSONALIZE/SCREEN slight variations in the paper quality can be compensated
CONTRAST for. It also allows you to de-activate the printer.
This option allows adjusting the screen contrast. The Once selected this option, the following appears on the
contrast change from 1 to 10, from less to more intensity. display:
By means of VERTICAL CURSORS it is possible to change
the value, this is shown on the display. PRINTER CONTROL
CONTRAST.........= _
3.4.2.4. Selection of the wash mode (only in THERMAL PRINTER
1 CHECK
versions with sipper)
PATH: PROGRAMMING/PERSONALIZE/WASH MODE The print intensity (contrast) is defined by a number from 1
to 10 which may be varied by means of the HORIZONTAL
The WASH key is used to circulate a wash solution or water CURSORS. By pressing key 1, a line of asterisks is printed
through the sipping circuit for the purpose of cleaning it at allowing you to check the selected intensity.
the end of a series of measurements or between samples.
By means of this option, it is possible to select using the By using the VERTICAL CURSORS you can deactivate
HORIZONTAL CURSORS, two washing modes: the printer (or activate it, if previously de-activated).
CONTINUOUS: The pump continuosly functions while the Quit with ENTER.
WASH key is held down.
Once your option is selected on the display, quit by This option allows you to activate or deactivate the patient
pressing ENTER. code. When the patient code is active, you have to enter a
code before entering each sample. This code identifies the
sample results in the printer and in the data information
3.4.2.5. List headings (only in versions with transmitted to the PC through the serial port.
printer)
On selecting this option, the menu appears: This allows you to list the information in the display
29
User manual
PATH: LIST/DISPLAY (In versions with printer) If you select this option, the following menu appears:
LIST (In versions without printer)
1 TEST NAMES
2 TEST CONTENTS
On selecting this option, the following menu appears:
If there is no programmed location, an information message
1 TESTS is displayed instead of the menu.
2 UNITS
3.5.2.1.1. Listing of the tests names
PATH: LIST/DISPLAY/TESTS (In versions with printer) The numbers of the programmed locations will be listed,
LIST/TESTS (In versions without printer) along with the name of the Tests that each one contains.
LIST TEST
3.5.1.2. Listing units From ............ =_
To................. =
PATH: LIST/DISPLAY/UNITS (In versions with printer)
LIST/UNITS (In versions without printer)
Key in the numbers of the first and last location you want
If you select this option, the number and title of the first to list and press ENTER. The complete data contained in
programmed unit appears on the display. You can review each location will be immediately printed.
these data of all the programmed units by using the
VERTICAL CURSORS. Pressing the RIGHT CURSOR at If no test is programmed in the selected range, a caution
any moment returns you to the first programmed unit, and message appears.
pressing the LEFT CURSOR brings you to the last
programmed unit. Press ESC to quit. 3.5.2.2. Listing the personalization
parameters
3.5.2. Listing in the printer (only in versions PATH: LIST/PRINTER/PERSONALIZE
with printer)
On selecting this option, a list is printed showing the titles
PATH: LIST/PRINTER or options that have been programmed in personalization
(section 3.4.2).
On selecting this option, the following menu appears:
3.5.2.3. Listing the filters table
1 TESTS
2 PERSONALIZE PATH: LIST/PRINTER/FILTERS TABLE
3 FILTERS TABLE
4 UNITS This option enables you to list a table with the number and
5 COMMUNICATIONS the wavelength of each programmed filter.
6 PUMP PARAMETERS (only in versions with sipper) (Section 3.4.3).
30
3.5.2.4. Listing units a) Fill the cuvette with distilled water and insert it into the
cuvette holder. Use the smallest cuvette you normally
PATH: LIST/PRINTER/UNITS work with. In version providing sipper fill the flow cuvette
with distilled water by pressing WASH.
This option lists a table with the number and title of each
programmed unit. b) Select the menu OTHER OPTIONS and press CHECK
(Section 3.4.4). LAMP.
PATH: LIST/PRINTER/COMMUNICATIONS
10.254 nA PASSED
This option enables you to list the parameters of the channel
that are programmed at the time (section 3.4.5.). d) In case the lamp has not been properly installed, the
display will show:
3.5.2.6. Listing the pump parameters
CHECK LAMP
PATH: LIST/PRINTER/PUMP PARAMETERS
On selecting this option, a list is printed showing the values 0.254 nA NO PASSED
that have been programmed in pump parameters (section
3.7.3).
The value shows the photodiode current with the 340nm
3.6. OTHER OPTIONS filter, it is expressed with nanoamperes (nA).
PATH: OTHER OPTIONS In this case, repeat the procedure for changing the lamp
described in section 5.3. but turning the lamp-holder 180
This option of the main menu includes some less frecuently degrees, and repeat the checking. If the lamp does not pass
used jobs. the test again, it is defective and must be replaced with
another one.
The following menu appears on the display:
3.6.3. Adjust of the peristaltic pump (only in
versions with sipper)
1 SERVICE
2 CHECK LAMP PATH: OTHER OPTIONS/ADJUST PUMP
3 ADJUST PUMP
4 LOAD A NEW PROGRAM Through usage and with the passage of time, the silicone
5 DELETE CONCENTRATIONS tubing of the peristaltic pump suffers slight deformation
which leads to the SAMPLE VOLUME not being exactly the
3.6.1. Service same as that which had been programmed. Because of this,
it is advisable to perform a adjust of the peristaltic pump
PATH: OTHER OPTIONS/SERVICE periodically and also after changing the tubing of the pump.
This option permits the instrument to be adjusted in those Besides adjusting the SAMPLE VOLUME, the
areas reserved for the technical assistance service. When POSITIONING of the sample in the cuvette, and the DELAY
this option is selected, the apparatus requests an ACCESS TIME, can also be adjusted.
PASSWORD only known by authorized staff.
Upon selecting this option, the following sub-menu appears
3.6.2. Check lamp on the display:
31
User manual
3.6.3.1. Checking adjust of the sample Key in the estimated number and press ENTER. The new
volume value will be stored and used in the next sipping cycles to
calculate the sample volume. You may check it by going to
PATH: OTHER OPTIONS/ADJUST PUMP/CHECK the previous menu (press ESC) and selecting CHECK.
This option allows you to enter a SAMPLE VOLUME 3.6.3.2.2. Sample positioning
between 100 and 5000 mL, and to perform pump cycles in
order to check the programmed parameters. On entering PATH: OTHER OPTIONS/ADJUST PUMP/MANUAL/
into the option, key in the value of the SAMPLE VOLUME, POSITIONING
and press ENTER. Then the instrument indicates:
The value for this parameter is pre-programmed in factory,
and is the number of steps the pump must run, after sipping
CHECK PUMP and the pump delay time, to position the sample in the
cuvette. It is said that the sample is positioned when
practically all the sipped sample has passed through the
Insert TUBE cuvette, cleaning the previous sample, and only about 5 mm
of the tubing immediatelly before the cuvette is filled with
sample. If it is seen that a longer part of the sipped sample
From this moment on, you can perform as many sipping remains without entering into the cuvette, this number must
cycles as you wish by pressing PUMP. Take into account be increased, in the opposite case it must be decreased. In
that here no photometric measurement is performed. the manual procedure the number must be a rough estimate.
During the cycles check that the sipped volume is the The following display appears:
required and that the sample is correctly transported to the
cuvette.
POSITIONING
3.6.3.2. Manual adjust Actual value = 178
New value ...=_
PATH: OTHER OPTIONS/ADJUST PUMP/MANUAL
32
3.6.3.3. Automatic adjust appears the received bytes and the saved bytes.
The following screen appears:
PATH: OTHER OPTIONS/ADJUST PUMP/AUTOMATIC
CHANGE PROGRAM
With this option, the following menu appears: 1 CONFIGURATION
2 EXIT
1 SAMPLE VOLUME RECEIVED=2039 SAVING
2 POSITIONING SAVED=156789
3.6.3.3.1. Sample volume At the end of the process the following message appears:
33
User manual
1 ABSORBANCES 1 MONOCHROMATIC
2 BICHROMATIC
1 LANGUAGE
2 PERSONALIZE 2 SCREEN CONTRAST
3 FILTERS TABLE 3 LABORATORY NAME
4 UNITS 4 WASH MODE
5 LIST HEADINGS
6 PRINTER CONTROL
7 PATIENT CODE
5 COMMUNICATIONS
6 USER OPTIONS 1 CONFIG RS-232-C
2 DEFAULT
PARAMETER
(1) This option may not appear, depending on the CALCULATION MODE and TYPE OF REACTION
34
APPENDIX I: MENUS TREE. VERSION WITH PRINTER AND
WITHOUT SIPPER.
1 ABSORBANCES 1 MONOCHROMATIC
2 BICHROMATIC
1 LANGUAGE
2 PERSONALIZE 2 SCREEN CONTRAST
3 FILTERS TABLE
3 LABORATORY NAME
4 UNITS
4 LIST HEADINGS
5 PRINTER CONTROL
6 PATIENT CODE
5 COMMUNICATIONS
1 CONFIG RS-232-C
6 USER OPTIONS 2 DEFAULT
PARAMETER
(1) This option may not appear, depending on the CALCULATION MODE and TYPE OF REACTION
35
User manual
1 ABSORBANCES 1 MONOCHROMATIC
2 BICHROMATIC
1 LANGUAGE
2 PERSONALIZE 2 SCREEN CONTRAST
3 FILTERS TABLE 3 PATIENT CODE
4 UNITS
4 LIST 1 TESTS
2 UNITS
(1) This option may not appear, depending on the CALCULATION MODE and TYPE OF REACTION
36