Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Laser Cooling and Trapping

Download as pdf or txt
Download as pdf or txt
You are on page 1of 16

Advanced Optics Laboratory

Laser Cooling and Trapping


1 INTRODUCTION
Laser cooling and trapping of neutral atoms is a rapidly expanding area of physics research that has
seen dramatic new developments over the last decade. These include the ability to cool atoms down to
unprecedented kinetic temperatures (as low as one micro Kelvin) and to bold samples of a gas isolated in
the middle of a vacuum system for many seconds. This unique new level of control of atomic motion is
allowing researchers to probe the behavior of atoms in a whole new regime of matter where deBroglie
wavelengths are much larger than the Bohr radius. Undoubtedly one of the distinct appeals of this
research is the leisurely and highly visible motion of the laser cooled and trapped atoms.
In this experiment you will operate a laser trap that is equal or superior in performance to what is
used in many current research programs. This experiment uses the lasers and saturated absorption
spectrometers used in the laser spectroscopy experiment' and thus you should have done that experiment
before doing this one. A small fraction (- 10%) of the beams of each of the two lasers goes to their
respective saturated absorption spectrometers. This allows for precise detection and control of the laser
frequencies, which is essential for cooling and trapping. The remainder of the laser light goes into the
trapping cell.
Section 2 of this write-up provides a brief introduction to the relevant physics of the atom trap,
section 3 discusses the laser stabilization, section 4 explains the optical layout for sending the laser beams
into the cell to create the trap, section 5 explains the trapping cell construction and section 6 discusses the
operation of the trap, measurement of the number of trapped atoms, and measurement of the time the
atoms remain in the trap.

2 THEORY AND OVERVIEW


We will present a brief description of the relevant physics of the vapor cell magneto-optical trap. For
more information, a relatively non-technical discussion is given in Ref. 2, while more detailed
discussions of the magneto-optical trap and the vapor cell trap can be found in Ref. 3 and Ref. 4,
respectively.

2.1 L ASER C O O L I N G
The primary force used in laser cooling and trapping is the recoil when momentum is transferred
from photons scattering off an atom. This radiation-pressure force is analogous to that applied to a
bowling ball when it is bombarded by a stream of ping pony balls. The momentum kick that the atom
receives from each scattered photon is quite small; a typical velocity change is about 1 cm/s. However,
by exciting a strong atomic transition, it is possible to scatter more than 107 photons per second and
produce large accelerations (104 g). The
radiation-pressure force is controlled in such
n laser + Dn Doppler a way that it brings the atoms in a sample to a
Scattering Rate (force)

velocity near zero ("cooling"), and holds


them at a particular point in space
("trapping").
The cooling is achieved by making the
n laser - Dn Doppler photon scattering rate velocity-dependent
using the Doppler effect.[5] The basic
principle is illustrated in Figure 1. If an atom
is moving in a laser beam, it will see the laser
frequency n laser shifted by an amount
n laser n atom Frequency
-(v / c)n laser , where v is the velocity of the
Figure 1. Atomic scattering rate versus laser frequency. atom along the direction of the laser beam. If

Laser Cooling Page 1/16


Advanced Optics Laboratory

Atom the laser frequency is below the atomic


resonance frequency, the atom, as a result of
this Doppler shift, will scatter photons at a
higher rate if it is moving toward the laser
beam (v negative), than if it is moving away.
Laser beam If laser beams impinge on the atom from all
six directions, the only remaining force on the
Velocity atom is the velocity-dependent part, which
Force opposes the motion of the atoms. This
provides strong damping of any atomic
motion and cools the atomic vapor This
arrangement of laser fields is often known as
"optical molasses”.[6]
Will scatter photons at a higher rate than
those moving in the same direction as the
Velocity beam. This leads to a larger force on the
counter propagating atoms.
Force
Figure 2. With the laser tuned to below the peak of atomic
resonance. Due to the Doppler shift, atoms moving in the
2.2 M AGNETO - OPTICAL T R A P
direction opposite the laser beam will scatter photons at a higher Although optical molasses will cool
rate than those moving in the same direction as the beam. This atoms, the atoms will still diffuse out of the
leads to a larger force on the counter-propagating atoms. region if there is no position dependence to
the optical force. Position dependence can be
introduced in a variety of ways. Here we will
B only discuss how it is done in the
"magneto-optical trap" (MOT), also known as
the "Zeeman shift optical trap," or "ZOT." The
s+ s - position-dependent force is created by using
0 appropriately polarized laser beams and by
applying an inhomogeneous magnetic field to
the trapping region. Through Zeeman shifts of
m=-1 m=0 m=+1
the atomic energy levels, the magnetic field
J=1 regulates the rate at which an atom in a
particular position scatters photons from the
s- s+ s- s+ s- s+ various beams and thereby causes the atoms to
be pushed to a particular point in space. In
J=0 addition to holding the atoms in place, this
B<0 B=0 B>0
force force=0 force greatly increases the atomic density since
many atoms are pushed to the same position.
Figure 3. One dimensional explanation of the MOT. Laser beams Details of how the trapping works are rather
with opposite helicity polarizations impinge on an atom from complex for a real atom in three dimensions,
opposite directions. The lasers excite the J= 0 to J = 1 transition. so we will illustrate the basic principle using
The laser beam from the right only excites the m = -1 excited state, the simplified case shown in Figure 2.
and the laser from the left only excites the m = +1 state. As an
atom moves to the right or left, these levels are shifted by the In this simplified case we consider an
magnetic field thereby affecting the respective photon scattering atom with a J = 0 ground state and a J = 1
rates. The net result is a position-dependent force that pushes the excited state, illuminated by circularly
atoms into the center. polarized beams of light coming from the left
and the right. Because of its polarization, the
beam from the left can only excite transitions to the m = +1 state, while the beam from the right can only
excite transitions to the m = -1 state. The magnetic field is zero in the center, increases linearly in the
positive x direction, and decreases linearly in the negative x direction. This field perturbs the energy
levels so that the Dm = +1transition shifts to lower frequency if the atom moves to the left of the origin,

Laser Cooling Page 2/16


Advanced Optics Laboratory

s+ while the Dm = -1transition shifts to higher frequency.


If the laser frequency is below all the atomic transition
frequencies and the atom is to the left of the origin, many
Magnetic photons are scattered from the s + laser beam, because it
coils is close to resonance. The s - laser beam from the right,
however, is far from its resonance and scatters few
s+ photons. Thus the force from the scattered photons
pushes the atom back to the zero of the magnetic field. If
s- s + the atom moves to the right of the origin, exactly the
opposite happens, and again the atom is pushed toward
s- the center where the magnetic field is zero. Although it is
somewhat more complicated to extend the analysis to
three dimensions, experimentally it is simple, as shown
in Figure 4. As in optical molasses, laser beams
illuminate the atom from all six directions. Two
Polarized symmetric magnetic field coils with oppositely directed
light currents create a magnetic field that is zero in the center
s- and changes linearly along the x, y, and z axes. If the
circular polarizations of the lasers are set correctly, a
Figure 4. Schematic of the MOT. Lasers beams are linear restoring force is produced in each direction.
incident from all six directions and have helicities Damping in the trap is provided by the cooling forces
(circular polarizations) as shown. Two coils with discussed in section 2.1. It is best to characterize the trap
opposite currents produce a magnetic field that is zero "depth" in terms of the maximum velocity that an atom
in the middle and changes linearly along all three axes. can have and still be contained in the trap. This
maximum velocity vmax is typically a few times Gl ( Gl
is the velocity at which the Doppler shift equals the natural linewidth Gof the trapping transition, where
l is the wavelength of the laser light).
A much more complicated three-dimensional calculation using the appropriate angular momentum
states for a real atom will give results which are qualitatively very similar to those provided by the above
F¢ analysis if: (1) the atom is excited on a transition where the upper
state total angular momentum is larger than that of the lower state
3
(F Æ F ¢ = F + 1) and (2) v ≥ (hG / 2 m)1/ 2 where m is the mass of the
2 5 2P atom. This velocity is often known as the "Doppler limit" velocity.' If
3/2
1 the atoms are moving more slowly than this, "sub-Doppler" cooling
0 and trapping processes become important, and the simple analysis
can no longer be used.[7] We will not discuss these processes here,
780 nm but their primary effect is to increase the cooling and trapping forces
hyperfine 780 nm
trapping for very slow atoms in the case of F Æ F + 1 transitions.
pumping
diode laser diode laser We will now consider the specific case of rubidium (Figure 5).
Essentially all the trapping and cooling is done by one laser which is
F tuned slightly (1-3 natural linewidths) to the low frequency side of
the 5S1/ 2 F = 2 Æ 2P3/ 2 F ¢ = 3 transition of 87Rb. (For simplicity we
2 2 will only discuss trapping of this isotope. The other stable isotope,
5 S1/2 85
1 Rb, can be trapped equally well using its F = 3 Æ F ¢ = 4
transition.) Unfortunately, about one excitation out of 1000 will
Figure 5. 87Rb energy level diagram cause the atom to decay to the F=1 state instead of the F=2 state. This
showing the trapping and hyperfine takes the atom out of resonance with the trapping laser. Another laser
pumping transitions. The atoms are (called the "hyperfine pumping laser") is used to excite the atom from
observed by detecting the 780 nm the 5S F = 1 to the 5P F ¢ = 1 or 2 state, from which it can decay back
fluorescence as they decay back to the
ground state.

Laser Cooling Page 3/16


Advanced Optics Laboratory

to the 5S F = 2 state where it will again be excited by the trapping laser.


In a vapor cell trap, the MOT is established in a low pressure cell containing a small amount of
rubidium vapor.[4] The rubidium atoms in the low energy tail ( v < vmax ª 20 m / s) of the
Maxwell-Boltzmann distribution are captured in the laser trap. If the trap is fumed on at t=0, the number
N of atoms in the trap will increase with the same functional form as that of a capacitor charging,

N (t ) = N0 (1 - e - t / t ) , (1)

where t is the time constant for the trap to fill to its steady state value N 0 and is also the average time an
atom will remain in the trap before it is knocked out by a collision. This time is just the inverse of the loss
rate from the trap due to collisions. Under certain conditions, collisions between the trapped atoms can be
important, but for conditions that are usually encountered, the loss rate will be dominated by collisions
with the room temperature background gas. These "hot" background atoms and molecules (rubidium and
contaminants) have more than enough energy to knock atoms out of the trap. The time constant t can be
expressed in terms of the cross sections s , densities n, and velocities of rubidium and non-rubidium
components as:

1
= nRbs Rb vRb + nnons non vnon . (2)
t
The steady-state number of trapped atoms is that value for which the capture and loss rates of the
trap are equal. The capture rate is simply given by the number of atoms which enter the trap volume (as
defined by the overlap of the laser beams) with speeds less than vmax . It is straightforward to show that
4
this is proportional to the rubidium density, vmax , and the surface area A of the trap. When the
background vapor is predominantly rubidium, the loss and capture rates are both proportional to
rubidium pressure. In this case N0 is simply:
4
A Ê vmax ˆ
N0 = 0.1 , (3)
s Rb ÁË vavg ˜¯

where vavg = (2 kT / m) , the average velocity of the rubidium atoms in the vapor. If the loss rate due to
1/ 2

collisions with non-rubidium background gas is significant, Eq. (3) must be multiplied by the factor
( )
nRbs Rb vavg / nRbs Rb vavg + nnons non vnon . The densities are proportional to the respective partial pressures.
Finally, if the loss rate is dominated by collisions with non-rubidium background gas, the number of
atoms in the trap will be proportional to the rubidium pressure divided by the non-rubidium pressure, but
t will be independent of the rubidium pressure.
As a final note on the theory of trapping and cooling, we emphasize certain qualitative features that
are not initially obvious. This trap is a highly over-damped system; hence damping effects are more
important for determining trap performance than is the trapping force. If this is kept in mind it is much
easier to gain an intuitive understanding of the trap behavior. Because it is highly over-damped, the
critical quantity vmax is determined almost entirely by the Doppler slowing which provides the damping.
Also, the cross sections for collisional loss are only very weakly dependent on the depth of the trap, and
therefore the trap lifetime is usually quite insensitive to everything except background pressure. As a
result of these two features, the number of atoms in the trap is very sensitive to laser beam diameter,
power, and frequency, all of which affect the Doppler cooling and hence vmax However, the number of
trapped atoms is insensitive to factors which primarily affect the trapping force but not the damping, such
as the magnetic field (stray or applied) and the alignment and polarizations of the laser beams. For

Laser Cooling Page 4/16


Advanced Optics Laboratory

example, changing the alignment of the laser beams will dramatically affect the shape of the cloud of
trapped atoms since it changes the shape of the trapping potential. However, these very differently
shaped clouds will still have similar numbers of atoms until the alignment is changed enough to affect the
volume of the laser beam overlap. When this happens, the damping in three dimensions is changed and
the number of trapped atoms will change dramatically. Of course, if the trapping potential is changed
enough that there is no potential minimum (for example, the zero of the magnetic field is no longer within
the region of overlap of the laser beams), there will be no trapped atoms. However, as long as the
damping force remains the same, almost any potential minimum will have about the same number of
atoms and trap lifetime.

2.3 O VERVIEW OF THE TRAPPING APPARATUS.


Figure 6 shows a general schematic of the trapping apparatus. It consists of two Vortex™ diode
lasers, two saturated absorption spectrometers, a trapping cell, and a variety of optics. The optical
elements are lenses for expanding the laser beams, mirrors and beamsplitters for splitting and steering the
Output
Photo- signal
diodes

Hyperfine
Variable
pumping laser attenuator

Pump
beam

Laser driver

Rb
cell
Probe
beams
Thick M
beamsplitter

Vortex™
diode laser

l/4
BS BS
l/4

Thick
M M
beamsplitter

l/4
Unused BS l/4 Vacuum
Rb
Pump
beam cell
cell

Probe Photo-
beams Variable diodes
attenuator

Output
signal

Amplifier

Ramp M
generator

Vortex™
diode laser M
M
Servo
electronics
Laser driver

Trapping
laser
Figure 6. Overall optical layout for laser trap experiment including both saturated absorption
spectrometers.

Laser Cooling Page 5/16


Advanced Optics Laboratory

beams, and waveplates for controlling their


polarizations. To monitor the laser frequency, a small
fraction of the output of each laser is split off and sent
to a saturated absorption spectrometer. An electronic
error signal from the trapping laser's saturated
absorption spectrometer is fed back to the laser to
actively stabilize its frequency. The trapping cell is a
small vacuum chamber with an ion vacuum pump, a
rubidium source, and windows for transmitting the
laser light. In the following sections we will discuss
the various components of the apparatus and the
operation of the trap.
Trapping
Tr
cell The key to the cooling are the three mutually
perpendicular laser beams that cross within the
trapping cell, from which another three are derived by
retro-reflecting them (see the detail of Figure 7).
Figure 6 shows the three laser beams derived using
three beamsplitters. The arrangement may be
Figure 7. Detail of laser beams that are sent through the somewhat different in your setup.
trapping cell. To simplify the figure, the l/4 waveplates
are not shown. Two beam paths are in the - horizontal
plane and a third is brought down and is then reflected up 3 LASER STABILIZATION
vertically through the bottom of the cell. The retro- As mentioned above, two lasers are needed for
reflected beams are tilted slightly to avoid feedback to the the trap. A few milliwatts of laser power are plenty for
diode laser. hyperfine pumping ( F = 1 Æ F ¢ = 1,2 ), but the
number of trapped atoms is nearly proportional to the
amount of power in the trapping laser ( F = 2 Æ F ¢ = 3 ). Setting up and using the trap is much easier
with at least 5 mW of trapping laser power, although it can work with less. The trapping laser must have
an absolute frequency stability of a few megahertz. This requires you to actively eliminate fluctuations in
the laser frequency, which are usually due to changes in the length of the laser cavity caused by
mechanical vibrations or temperature drifts. This is accomplished by using the saturated absorption signal
to detect the laser frequency and a feedback loop to hold the length of the laser cavity constant.
The laser servo control system is far less complicated as first appears. As intimidating as the servo
box inputs, outputs, switches and dials seem, the principles are relatively straightforward. The Vortex™
laser frequency is controlled by a piezo voltage. As its frequency is tuned and scanned near a hyperfine
level of interest the output from saturated spectroscopy setup (that is, from the photodiode preamplifier)
looks more or less like Figure 8 (or perhaps an
Final sweep range
prior to locking upside-down and/or backwards version of it). The
Error signal voltage (V)

purpose of the laser servo controller is to maintain


the laser frequency to the “red” side of atomic
resonance, about halfway down the curve shown
Vref in the figure. At that point the photodiode
preamplifier produces an output that has been
labeled as Vref.The servo controller box stabilizes
the laser frequency by adjusting the piezo voltage
such that the photodiode amplifier output remains
desired n atom Piezo voltage at Vref, or nearly so. Briefly it does so by
n laser (Frequency) decreasing the piezo voltage when the photodiode
voltage falls above Vref and increases the piezo
Figure 8. Photodiode preamplifier output voltage (error
voltage when the photodiode rises above Vref.
signal voltage) as a function of Vortex™ piezo (PZT)
voltage near the hyperfine line of interest. The servo system The servo needs some help from you to get
is meant to stabilize the laser to the desired frequency. started, however. The insides of the servo

Laser Cooling Page 6/16


Advanced Optics Laboratory

PZT (piezo)
servo gain
adjust
Error in
(from photodiode Polarity +/-
preamplifier) V err Vset - Verr Monitor out
(to oscilloscope)
BNC
-1 BNC
Servo
amplifier Servo loop
open/closed
Vset

Ramp gain adjust


coarse fine

Ramp in
(from ramp
generator) Ramp on/off
PZT/piezo out
(to Vortex™ controller)
BNC
BNC
Ramp
amplifier

Modulation in PZT bias


(normally unused) (coarse and fine)

BNC
Figure 9. Laser frequency stabilization servo control box. Shaded and dashed boxes indicate the knobs and switches on the
controller unit.
controller box are schematically illustrated in Figure 9. The shaded, dashed-outline boxes label the
various switches and knobs that are on the controller.

3.1 S WEEP AND SETPOINT ADJUSTMENTS


The first order of business is to produce a triangular wave input using the function generator. Start
with a low amplitude and turn all biases and offsets to zero. A few tens of Hertz, up to about 100 Hz is a
good frequency range to work with. You will connect the function generator output to the BNC on the
servo controller box (labeled “Ramp In”) but before you do: 1. Switch the servo loop to “open” (switch
down) and the “Ramp” switch to “on” (switch up). 2. Turn the coarse ramp gain adjust knob all the way
down (counter-clockwise). 3. Connect the piezo output to an oscilloscope and not to the Vortex™
controller. Now connect the function generator to to the ramp intput connector on the servo controller
unit. Turn the ramp coarse gain all the way up and then adjust the output of the function generator so that
the maximum piezo voltage allowed by the Vortex™ controller is not exceeded (±4.5 V). From now on
you should not have to adjust anything on the function generator except perhaps its frequency. Set the
piezo voltage on the Vortex™ controller to midway (50 V). Turn down the ramp gain on the servo
controller box and connect the piezo output to the Vortex™ piezo input.

Laser Cooling Page 7/16


Advanced Optics Laboratory

The next item of business is to zoom in on the rubidium hyperfine atomic transition of interest (e.g.,
the 5S F = 2 Æ 5P F ¢ = 3 transition of 87Rb.) To do so you will use the ramp gain adjust and PZT bias
controls. Connect the photodiode preamplifier output to the “Error In” BNC of the servo controller. Also
observe the photodiode preamplifier output with the scope. Set the ramp gain to a relatively high value
until you see the various rubidium lines as described in the Doppler Free Saturated Absorption
Spectroscopy experiment. Center the line of interest on the oscilloscope. You can do so by adjusting the
PZT bias knob (coarse, at first) on the servo controller —you can also do it by setting the piezo voltage
on the Vortex™ controller directly. Actually, the latter is a bit safer, as the input specifications will not
inadvertently be exceeded. To zero-in on the line requires an iterative process. The idea is to reduce the
sweep voltage of the piezo by decreasing the ramp gain, then re-centering the scan range to the desired
frequency with the piezo offset or the Vortex™ controller piezo setting, then reducing the ramp gain
again, and so onUsing the second channel of the oscilloscope used to monitor the photodiode
preamplifier output, connect the “Monitor output” of the servo controller box. Be sure to note the
position of the scope trace zero for this channel. At the same time, notice the value Vref of the photodiode
preamplifier output when it is halfway up the transition. The next step is to adjust the “Setpoint” value of
the servo controller. It determines value of the voltageVset shown in Figure 9. Notice what the circuit
does —the input to the servo amplifier is the difference, Verr - Vset , where Verr is the photodiode
preamplifier output voltage. This difference is amplified and then (will later be) fed back to the piezo
voltage input of the Vortex™ controller. You should adjust the Setpoint knob so that the monitor output
passes through zero exactly at the point where the laser is halfway up the transition curve. You can
overlap the monitor and photodiode preamplifier traces on the oscilloscope, then adjust the setpoint knob
so former traces crosses zero at the appropriate place.

3.2 A CQUIRING LASER LOCK


You are almost ready to “close the loop” and let the frequency servo do its job. First turn the “Servo
gain” knob to midway between its minimum and maximum. Second you must further zero-in on
frequency region of interest by adjusting the piezo sweep using the same iterative process as you did
above. Continue reducing the ramp gain in steps until the laser frequency sweep covers just a small
region around the laser frequency of interest. Figure 8’s illustration is meant to be suggestive of this
process –the increasing thickness of lines is meant to show the decreasing frequency scans of the piezo
voltage.
Now you can close the servo loop by putting the loop switch in the up position. If the laser is
properly locked the photodiode preamplifier output should be moving up and down a bit but stay near the
value Vref. However, it is not entirely likely that the servo was able to “catch” lock on the first try. Both
the servo gain and the polarity may need to be adjusted. The polarity switch adjust the sign of the
feedback –you can see from Figure 9 that it merely enters or removes a –1 in the feedback loop. In
principle one can track through the electronics an know which polarity, + or -, is appropriate for the
feedback signal. In practice, however, it is usually easier just to flip the polarity switch to discover
which position is correct. If you choose the wrong sign, then the laser becomes stabilized to the wrong
side of the transition and laser cooling will not work!
If the servo is not working, it is also possible that the servo gain is either too low or too high. Too
low and there is simply not enough piezo voltage to retun the laser to the desired frequency when it drifts
away. If the servo gain is too high the servo can go “crazy” by oscillating, or the servo can suddenly
jump and lock to some other (undesired) transition. It is usually effective to turn the servo gain low and
then turn it up until the servo goes, or nearly goes, into oscillation. Then set the gain just a bit lower and
redo the locking procedure.
Detailed discussions of the saturated absorption spectrometer, the feedback loop circuit, and the
procedure for locking the laser frequency are given in Ref. 1, along with examples of spectra obtained.
You should already be familiar with all but the locking of the laser frequency. After a little practice one
can lock the laser frequency to the proper value within a few seconds.

Laser Cooling Page 8/16


Advanced Optics Laboratory

Here we just mention a few potential problems and solutions Check that the baseline on the saturated
absorption spectrum well off resonance is not fluctuating by more than several percent of the
F = 2 Æ F ¢ = 3 peak height. If there are larger fluctuations, they are likely caused by light feeding back
into the laser or by the probe beams vibrating across the surface of the photodiodes. Changing the
alignment and making sure that all the optical components are rigidly mounted will normally solve these
problems. Under quiet conditions and reasonably constant room temperature, the laser should stay
locked for many minutes and sometimes hours at a time. Bumping the table or the laser or making a loud
noise will likely knock the laser out of lock

3.3 S TABILITY REQUIREMENT


[Optional note: Even under the best conditions, the system we have described is likely to have
residual frequency fluctuations of around 1 MHz. The trap will work fine with this level of stability but
these fluctuations will cause some noise in the fluorescence from the trapped atoms. For most
experiments this is not serious, but it can limit some measurements. If one has a little knowledge of servo
systems, it is quite straightforward to construct a second feedback loop that adjusts the laser current. This
current feedback loop should have a roll-on filter and high gain. (Because the laser current loop can be
much faster than the PZT loop, it can have a much higher gain.) This results in the current loop feedback
dominating for frequencies above a few hertz. However, the PZT loop gain is largest for very low
frequencies, and thus handles the DC drifts. The combination of the two servo loops will make the laser
frequency much more stable and will also make the lock extremely robust. We have had lasers with
combined PZT and current servo loops remain frequency-locked for days at a time and resist all but the
most violent jarring of the laser. A circuit diagram for the combined PZT and current servo loop can be
obtained from C. Wieman. This extra servo loop is a complication, however, which is not necessary for
your experiments.]
The requirements for the frequency stability of the hyperfine pumping laser are much less stringent
than those for the trapping laser. For many situations it is adequate to simply set the frequency near the
peak of the 5S F = 1 Æ 5P F ¢ = 2 transition by hand Over time it will drift off, but if the room
temperature does not vary too much, it will only be necessary to bring the laser back on to the peak by
slightly adjusting the PZT offset control on the ramp box every 5 to 10 minutes. If better control is
desired, the frequency of this laser can be locked to the peak of the F = 1 Æ F ¢ = 1,2, and 3 Doppler-
broadened (and hence unresolved) absorption line by modulating the laser frequency and using
phase-sensitive detection with a lock-in amplifier. The output of the lock-in amplifier is then fed back to
the PZT to keep the laser frequency on the peak. There is a lock-in amplifier in the lab that can be used if
desired. This provides a relatively crude frequency lock that is quite adequate for the hyperfine pumping
laser.

4 OPTICAL SYSTEM
The beams from the lasers must now be sent into the cell to form the trap. The basic requirement is
to send light beams from the trapping laser into the cell in such a way that the radiation-pressure force has
a component along all six directions. To motivate the discussion of the optical design for this lab, we first
mention the design used in most of the traps in research programs. In these research traps, the light from
the trapping laser first passes through an optical isolator and then through beam shaping optics that make
the elliptical diode laser beam circular and expand it to between 1 and 1.5 cm in diameter. The beam is
then split into three equal intensity beams using dielectric beam splitters. These three beams are circularly
polarized with quarter-wave plates before they pass through the trapping cell where they intersect at right
angles in the center of the cell. After leaving the trapping cell each beam goes through a second
quarter-wave plate and is then reflected back on itself with a mirror. This accomplishes the goal of having
three orthogonal pairs of nearly counter-propagating beams, with the reflected beams having circular
polarization opposite to the original beams.

Laser Cooling Page 9/16


Advanced Optics Laboratory

4.1 G EOMETRY
For this lab you use a modified version of this setup, as shown in Fig. 5. A number of the
expensive optical components (optical isolator, dielectric beam splitters, and large aperture high quality
quarter-wave plates) have been eliminated from the research design. The light from the trapping diode
laser is sent into a simple two-lens telescope that expands it. This is then split into three beams. The
operation of the trap is insensitive to the relative amounts of power in each of the beams. [You should
check how much the number of trapped atoms change with a factor-of-two difference in relative powers.]
However, the beam size as set by the telescope is of some importance. The number of trapped atoms
increases quite rapidly as the beam size increases and, as discussed below, the larger the beam, the less
critical the alignment. However, if the beams are too large they will not fit on the mirrors and it becomes
harder to see the beams due to their reduced intensity. Beams of about 1.5 cm diameter work well, but
avoid using beams much smaller than 1.5 cm in diameter because of the decrease in the number of
trapped atoms and the increased alignment sensitivity. The lenses in the lab will make a simple telescope
that expands the beam from the diode laser to this size.
Two of the three beams remain in a horizontal plane and are sent into the cell as shown in Figure 7.
The third is angled down and reflects up from the bottom of the cell. To simplify the adjustment of the
polarizations, the light should be kept linearly polarized until it reaches the quarter-wave plates. This will
be the case as long as all the beams have their axes of polarization either parallel (p) or perpendicular (s)
to the plane of incidence of each mirror. This is easy to achieve for the beams in the horizontal plane, but
difficult for the beam that comes up through the bottom of the cell. However, with minimal effort to be
close to this condition, the polarization will remain sufficiently well linearly polarized. If you think this
might be a problem you can check the ellipticity of the polarization by using a photodiode and a rotatable
linear polarizer. An eccentricity of 10 or greater on the polarization ellipse is adequate.
After they pass through the cell, the beams are reflected approximately (but not exactly!) back on
themselves. The reason for having an optical isolator in the research design is that even a small amount of
laser light reflected back into the laser will dramatically shift the laser frequency and cause it to jump out
of lock. In the absence of an optical isolator, this will always happen if the laser beams are reflected
nearly back on themselves. Feedback can be avoided by insuring that the reflected beams are steered
away from the incident beams so that they are spatially offset by many (5-10) beam diameters when they
arrive back at the position of the laser. Fortunately, for operation of the trap the return beam need only
overlap most of the incident beam in the cell, but its exact direction is unimportant. Thus by making the
beams large and placing the retro mirrors close to the trap (within 10 cm for example) it is possible to
have the forward and backward going beams almost entirely overlap even when the angle between them
is substantially different from 180°. This design eliminates the need for the very expensive ($2500)
optical isolator and, as an added benefit, makes the operation of the trap very insensitive to the alignment
of the return beams.
It is easy to tell if feedback from the return beams is perturbing the laser by watching the signal from
the saturated absorption spectrometer on the oscilloscope. If the amplitude of the fluctuations is affected
by the alignment of the reflected beam or is reduced when the beam to the trapping cell is blocked,
unwanted optical feedback is occurring.

4.2 P OLARIZATION
The next task is to set the polarizations of the three incident beams. The orientations of the respective
circular polarizations are determined by the orientation of the magnetic field gradient coils. The two
transverse beams that propagate through the cell perpendicular to the coil axis should have the same
circular polarization, while the beam that propagates along the axis of the coils should have the opposite
circular polarization. Although in principle it is possible to initially determine and set all three
polarizations correctly with respect to the magnetic field gradient, in practice it is much simpler to set the
three polarizations relative to each other and then try both directions of current through the magnetic field
coils to determine which sign of magnetic field gradient makes the trap work. To set the relative

Laser Cooling Page 10/16


Advanced Optics Laboratory

polarization of the three beams, first identify the same (fast or slow) axis of the three quarter-wave (l/4)
plates. For the two beams that are to have the same polarization, this axis is set at an angle of 45°
clockwise with respect to the linear polarization axis when looking along the laser beam. For the axial
beam, the axis is oriented at 45° counterclockwise with respect to the linear polarization. This orientation
needs only to be set to within about ±10°. The orientation of the l/4 plates through which the beams pass
after they have gone through the trapping cell ("retro l/4 plates") is arbitrary. No matter what the plates'
orientations are, after the beams have passed though them twice, the light's circular polarization will be
reversed.
An optional experiment is to see what happens when you replace one or more of the three retro l/4
plates with retro-reflecting right angle mirrors. [1] Although this combination of mirrors does not provide
ideal l/2 retardance, it is fairly close. An added benefit of this approach is that two reflections off a
mirror usually result in much less light loss than one reflection and two passes through a l/4 plate.

4.3 H YPERFINE PUMPING LASER OPTICS


Minimal optics is needed for the hyperfine pumping laser. You need only send it through the two
lens beam expanding telescope in the lab to make a large roughly collimated beam (typically with a 2-3
cm diameter) and send it into the cell from a direction which will minimize the scattered light into the
detectors that observe the trapped atoms. The trap is insensitive to nearly everything about the hyperfine
pumping light, including its polarization. Why should you expect this to be the case based on how the
trap works?

5 TRAPPING CELL CONSTRUCTION


The primary concern in the construction of the trapping cell is that ultrahigh vacuum (UHV) is
required. Although trapped atoms can be observed at pressures of 10-5 Pa (~10-7 Torr), trap lifetimes long
enough for most experiments of interest require pressures in the 10^6 to 10-7 Pa range. There are three
main elements in the trapping cell: (1) a pump to remove unwanted background gas - mostly water,
hydrogen, and helium (helium can diffuse through glass), (2) a controllable source of rubidium atoms,
and (3) windows to transmit the laser light and allow observation of the trapped atoms. The major
components are depicted in Figure 10.

5.1 V ACUUM PUMP


The cell is attached to an ion pump. In this pump the atoms are ionized in a high voltage discharge
and then embedded in the electrodes, thereby removing them from the system. The 5000 V needed to
make the pump operate is provided by a power supply that is attached to the back of the pump by a wire.
Although this wire has a grounded outer shield, and therefore in principle quite safe, it is wise to always
avoid touching it or the pump near the high voltage connector. The ion pump has the minor drawback
that it requires a large magnetic field, which is provided by a permanent magnet. The fringing fields from
this magnet can extend into the trap region and will affect the trap to some extent. Although the trap will
usually work without it, we put a layer of 0.75 mm (0.03 in) magnetically permeable steel sheet around
the pump to shield the trap from this field. For this same reason it is advisable to avoid having magnetic
bases very near the trap.

5.2 RUBIDIUM SOURCES


We will now discuss how to produce the correct pressure of rubidium vapor in the cell. The vapor
pressure of a room temperature sample of rubidium is about 5 x 10-5 Pa. This is much higher than the
10-6 to 10-7 Pa (~10-8 to 10-9 Torr) of rubidium vapor pressure that is optimum for trapping. At higher
pressures the trap will still work, but the atoms remain in the trap a very short time, and it is often

Laser Cooling Page 11/16


Advanced Optics Laboratory

Epoxied
glass cell

zero length
Pumpout adapter
valve

Ion Pump

Rubidium
dispenser
bottom
window
Figure 10. Drawing of trapping cell. The tubes on the cross have been elongated in the drawing
for ease of display. The Rb dispenser is inside a stainless steel vacuum tube; we show a
cut-away view in this drawing.

difficult to see them because of the bright fluorescence from the un-trapped background atoms. Also, the
absorption of the trapping beams when passing through the cell will be significant. Thus you should
maintain the rubidium at well below its room temperature vapor pressure. Because it is necessary to
continuously pump on the system to avoid the buildup of hydrogen and helium vapor, it is necessary to
have a constant source of rubidium to maintain the correct pressure. Through chemical reactions and
physisorption, the walls of the cell usually remove far more rubidium than the ion pump does and the rate
of pumping by the walls depends on how well they are coated with rubidium.
In this experiment the rubidium vapor is produced by a commercial “rubidium dispenser” (locally
sometimes also referred to as a “rubidium getter”. The dispenser is several milligrams of a rubidium
compound that is contained in a small (1.0 x 0.2 x 0.2 cm) stainless steel oven. Two of these ovens are
spot-welded onto two pins of a vacuum feed-through, one of which is illustrated in Figure 10. When
current (3-5 A) is sent through the oven, rubidium vapor is produced. The higher the current, the higher
the rubidium pressure in the chamber. With this system it is unnecessary to coat the entire surface with
rubidium, and is in fact undesirable since the dispenser is likely to be exhausted before the surface is
entirely saturated. You should be able to produce enough rubidium pressure to easily see the background
fluorescence with as little as 3.4 A through the dispenser oven. The current through the dispenser is
controlled by the power supply under the table. Be careful not to turn the current higher than 6.5 amps
under any circumstances! If you do it will burn out the dispenser and ruin the experiment! You can see
how much rubidium is in the cell by sending the laser beams into the cell, setting the laser frequency to
the trapping transition (the saturated absorption cell should glow brightly) and look for fluorescence from
the atoms in the cell. If you can see any glow from inside the cell that goes away when the laser
frequency is tuned off the transition, there is more than enough rubidium for trapping.
After the current through the dispenser has been turned on, the rubidium vapor comes to an
equilibrium pressure with a time constant of about 5 min. When the current through the dispenser is
turned off, the rubidium pressure drops with a time constant of about 4 s if the dispenser has been on for
only a short while. With prolonged use of the dispenser, this time constant can increase up to a few
minutes as rubidium builds up on the walls, but it decreases to the original value if the dispenser is left
off for several days.

Laser Cooling Page 12/16


Advanced Optics Laboratory

The superior rubidium pressure control provided by the dispenser makes it possible to use high
pressures to easily observe fluorescence from the background vapor. This allows one both to check
whether the laser and cell are operating properly, and to have rapid response in the number of atoms
while optimizing trapping parameters. Once these tasks are complete, simply reducing the dispenser
current provides low pressures almost immediately. Low pressures are desirable for many experiments
because they yield relatively long trap lifetimes (seconds) and little background light from the
fluorescence of un-trapped atoms.

5.3 M AGNETIC FIELD GRADIENT


With the cell installed and the laser beams aligned, the final ingredients for trapping are the magnetic
field coils. A gradient of up to about 0.20 T/m (20 G/cm; normal trap operation is at 10-15 G/cm) is
needed. This is provided by two freestanding coils 1.3 cm in diameter with 25 turns each of 24 gauge
magnet wire and a separation of 3.3 cm. The coils are mounted on either side of the cell such that the
current travels through the loops in opposite directions and the coil axes are collinear with one of the laser
beam axes.

6 OPERATION OF THE TRAP AND MEASUREMENTS ON IT

6.1 O BSERVATION SYSTEM


An inexpensive CCD TV camera and monitor are used to observe the trapping cell. This will show
the cloud of trapped atoms as a very bright white glow in the center of the cell. Because of the poor
response of the eye at 780 nm, the trapped atoms can be seen by eye only if the room is quite dark. For
aligning the trapping laser beams you can use the IR phosphorescent card or a piece of white paper if the
room is darkened. A 1 cm2 photodiode with a simple current-to-voltage amplifier is used for making
quantitative measurements on the trapped atoms. It is placed at any convenient position that is close to the
trap, has an unobstructed view, and receives relatively little scattered light from the windows. The
photodiode is used to detect the 780 nm fluorescence from the atoms as they spontaneously decay to the
ground state from the 5P3/2 level. This measurement can be quantified and used to determine the number
of trapped atoms. The same or a similar photodiode can also be used to look at the absorption by the
trapped atoms and to monitor the rubidium pressure in the cell by measuring the absorption of a probe
laser beam.
Although the trap fluorescence is large enough to easily detect with the photodiode, it can be
obscured by fluorescence from the background rubidium vapor or by scattered light from the cell
windows. Over a large range of pressures, the fluorescence from the background gas will be smaller than
that from the trapped atoms. However, the scatter from the windows is likely to be significant under all
conditions. The scattered light background will simply be a constant offset on the photodiode signal. For
studying small numbers of atoms in the trap, however, the noise on this background can become a
problem. In this case, you can use a lens to image the trap fluorescence onto a mask which blocks out the
unwanted scattered light but allows the light from the trapped atoms to reach the photodiode.

6.2 T RAP OPERATION


When the cell and all the optics are in place, the first step is to turn on the dispenser to put rubidium
into the cell. Initially, monitor the absorption of a weak probe beam through the cell to determine the
rubidium pressure. Although the trap will operate over a wide range of pressures, a good starting point is
to have about 1%/cm absorption on the F=2 to upper states transition in the region of the trap. At this
pressure it is possible to see dim lines of fluorescence where the trapping beams pass through the cell
when the trapping laser is tuned to one of the rubidium transitions. It is often easier to identify this
fluorescence by slowly scanning the laser frequency and looking for a change in the amount of light in
the cell. While absorption measurements are valuable for the initial setup and for quantitative

Laser Cooling Page 13/16


Advanced Optics Laboratory

measurements, in the standard operation of the trap, use the setting on the dispenser current and/or the
observation of the fluorescence to check that the pressure is reasonable.
After an adequate rubidium pressure has been detected, the magnetic field gradient is turned on and
the lasers are set to the appropriate transitions. If the apparatus is being used for the first time, it will be
necessary to try both directions of current through the field coils to determine the correct sign for
trapping. The trapped atoms should appear as a small bright cloud, much brighter than the background
fluorescence. Pieces of dust on the windows may appear nearly as bright, but they will be more localized
and can be easily distinguished by the fact they do not change with the laser frequencies or magnetic
field. If the trap does not work (and the direction of magnetic field and the laser polarizations are set
correctly) the lasers are probably not set on the correct transitions. The trapped atom cloud may vary in
size; it can be anywhere from less than 1 mm in diameter to several millimeters. Blocking any of the
beams is also a simple method for distinguishing the trapped atoms from the background light.

6.3 M EASUREMENTS
Although many other more complicated measurements could be made with the trap, you should start
by making the two most straightforward measurements: the number of trapped atoms and the time that
atoms remain in the trap. If you have time after completing these, you can consider other experiments.
These measurements are both made by observing the fluorescence from the trapped atoms with a
photodiode. The number of atoms is determined by measuring the amount of light coming from the
trapped atoms and dividing by the amount of light scattered per atom, which is calculated from the
excited state lifetime. The time the atoms remain in the trap is found by observing the trap filling time and
using Eq.(1). To make a reliable measurement of the number of trapped atoms, it is crucial to accurately
separate the fluorescence of the trapped atoms from the scattered light and the fluorescence of the
background vapor. To do this one must compare the signal difference between having the trap off and
on. Therefore, the trap must be disabled in a way that has a negligibly small effect on the background
light. We have found that turning off or, even better, reversing the magnetic field is usually the best way
to do this. The magnetic field may alter the background fluorescence, but this change is generally smaller
than the signal of a typical cloud of trapped atoms. Check how big an effect this is. Once you have
determined the photocurrent due to just the trapped atoms, the total amount of light emitted can be found
using the photodiode calibration of 0.3 mA/mW and calculating the detection solid angle. The rate R at
which an individual atom scatters photons is given by

pG( I / Is )
R= (4)
1 + ( I / Is ) + 4( D / G )
2

where I is the sum of the intensities of the six trapping beams, G is the 6 MHz natural linewidth of the
transition, D is the detuning of the laser frequency from resonance, and I s, is the 4.1 mW/cm2 saturation
intensity. The simplest way to find D is to ramp over the saturated absorption spectrum and, when
looking at the locking error signal, find the position of the lock point (zero crossing point) relative to the
peak of the line. The frequency scale for the ramp can be determined using the known spacing between
two hyperfine peaks. A typical number for R is 6 x 106 photons / (s ¥ atoms) . One can optimize the
number of atoms in the trap by adjusting the position of the magnetic field coils, the size of the gradient,
the frequencies of both trapping and hyperfine pumping lasers, the beam alignments, and the polarization
of the beams. More than 107 trapped atoms have been obtained when the rubidium pressure is large
enough to dominate the lifetime.
The filling of the trap can also be observed using the same photodiode signal. This is best done by
suddenly turning on the current to the field coils to produce a trap. The fluorescence signal from the
photodiode will then follow the dependence given by Eq. (1), as shown in Figure 11. This can be most
easily observed by sending the photodiode signal into a storage scope. The value of the 1/e trap lifetime t

Laser Cooling Page 14/16


Advanced Optics Laboratory

(the characteristic time an atom remains trapped) can then


Florescence (arb. units)
be determined from this curve. Eq. (1) tells us that it is the
same as the time for the trap to fill to 1/e of its final value.
1 Lifetimes between a fraction of a second to a few seconds
are reasonable.
By changing the current through the dispenser, you
should vary the rubidium pressure and see how it affects
the number and lifetime of the trapped atoms. This can be
0 compared with the predictions of Eqs. (1) and (2) and can
be used to determine the collision cross sections if you also
-2 0 2 4 6 measure the rubidium density in the cell. There are many
Time (sec) other experiments that can be done with the trapped atoms.
The choices are only limited by your imagination and/or
Figure 11. Typical curve of number of trapped available equipment. You might look at how the number of
atoms versus time after the trap is turned on at O atoms and the lifetime depends on the various parameters
see, as shown on oscilloscope. such as magnetic field, laser intensity, etc. You could also
think about how to measure the spring constant and
damping constant for the trap. You can also turn off the magnetic field and watch the atoms spread out.
In this situation, known as optical molasses and discussed in references 6 and 7, the laser light provides
damping of the velocity but no trapping. In these conditions the atoms reach the lowest possible
temperatures, but it is necessary to use additional field coils to cancel the magnetic fields from the earth
and the ion pump. If you do, you will be able to watch the atoms spread out slowly (in a fraction of a
second or longer) if the laser is detuned well to the red of the resonance line.
We conclude with both a challenge and a warning about all such possible experiments: you are likely
to observe phenomena that have not yet been studied or are just being studied in research labs around the
world. Often you will not find explanations (or will find incorrect explanations) in the current research
literature, your instructor will not be able to provide much help, and the subject is too new for you to find
textbooks providing answers. On the other hand, this means that if you do observe new phenomena and
can explain them you will be able to publish your results in a scientific journal.

7 REFERENCES
1. K. B. MacAdam, A. Steinbach, and C. Wieman, "A narrow-band tunable diode laser
system with grating feedback and a saturated absorption spectrometer for Cs and Rb," Am.
J. Phys., 60, 1098-1111 (1992) and the Senior Lab manual.
2. S. L. Gilbert and C. E. Wieman, "Laser Cooling and Trapping for the Masses," Optics and
Photonics News, 4, 8-14 (1993).
3. The initial demonstration and discussion is E. L. Raab, M. Prentiss, A. Cable, S. Chu, and D.
Pritchard, "Trapping of neutral sodium atoms with radiation pressure," Phys. Rev. Lett. 59,
2631-2634 (1987). More detailed analysis of the trap is given in A. M. Steane, M. Chowdhury, and
C. J. Foot, "Radiation force in the magneto-optical trap," J. Opt. Soc. Am B 9, 2142 (1992), while
a discussion of many of the novel aspects of the behavior of atoms in the trap is given in D. Sesko,
T. Walker and C. Wieman, "Behavior of neutral atoms in a spontaneous force trap," J. Opt. Soc.
Am. B 8, 946-958 (1991).
4. C. Monroe, W. Swann, H. Robinson, and C. Wieman, "Very cold atoms in a vapor cell," Phys.
Rev. Lett. 65, 1990, 1571-1574; K. Lindquist, M. Stephens, and C. Wieman, "Experimental and
theoretical study of the vapor-cell Zeeman optical trap," Phys. Rev. A 46, 4082-4090 (1992).
5. T.W. Hansch and A. L. Schawlow, "Cooling of gases by laser radiation," Opt. Commun. 13,
68-69 (1975).

Laser Cooling Page 15/16


Advanced Optics Laboratory

6. S. Chu, L. Hollberg, J. Bjorkholm, A. Cable, and A. Ashkin, "Three-dimensional viscous


confinement and cooling of atoms by resonance radiation pressure," Phys. Rev. Lett. 55, 48-51
(1985).
7. J. Dalibard and C. Cohen-Tannoudii, "Laser cooling below the Doppler limit by polarization
gradients: simple theoretical models," J. Opt. Soc. Am. B 6, 2023-2045 (1989); P. Ungar, D.
Weiss, E. Riis, S. Chu, "Optical molasses and multilevel atoms: theory," J. Opt. Soc. Am. B 6,
2058-2072 (1989).
8. L. Orozco, SUNY at Stony Brook, private communication.

8 Prelab Exercises
1. (Section 2.1)
Calculate the force on, and the acceleration of, the atoms when they are scattering photons at the
maximum rate from a laser beam. Include this result in your write-up. An atom can absorb and reemit a
photon every 2 natural lifetimes when it is in very intense laser light that is at the resonance frequency of
the atom.

2. (Section 2.2)
a) What is the purpose of the re-pumping laser? What would happen in the MOT without it? b)
Your MOT uses the 5S1/ 2 F = 2 Æ 5P3/ 2 F ¢ = 3 transition for the trapping and cooling transition and
requires a re-pumping laser for the 5S1/ 2 F = 1 Æ 5P3/ 2 F ¢ = 2 transition. What other transition scheme
might work for trapping and cooling 87Rb with two lasers? Explain.
3. (Section 3)
Calculate how large a change in the length of a 3 cm laser cavity will shift the frequency by 1 MHz.
Hint: an integer number m of half wavelengths must fit in the cavity so 3 cm = ml / 2 , and c = ln .

4. (Section 5.3)
Recall that the magnetic field is provided by two freestanding coils 1.3 cm in diameter with 25 turns
each of 24 gauge magnet wire and a separation of 3.3 cm. How much current do you need to get a
gradient of 15 G/cm? How much power? (Hint, the coils should get warm, but not burn up!).

Laser Cooling Page 16/16

You might also like