Fully Automated Clinical Chemistry Analyzer

Parameters at a Glance Technical Specifications


System Function: Automatic, discrete, random access with STAT sample priority
Throughput: Up to 240 tests/hour
Measuring principles: Spectrophotometry
Item Description Assay Method No. of test Pack Size Item Description Assay Method No. of test Pack Size
Methodology: End-point, fixed-time, kinetic, Single/dual reagent chemistries,
Glucose 1 Point End 969 5x65mL Magnesium 1 Point End 169 2x30mL
monochromatic/bichromatic linear/non-linear multipoint calibration
Glucose Hexokinase 2 Point End 814 3x63/3x20mL CK-NAC Rate A 171 2x20/1x13mL
Photometric System: HCFG rear spectrophotometry
Cholesterol 1 Point End 969 5x65mL ASO 2 Point End 156 2x20/2x8mL
Reagent/Sample Handling
Triglycerides 1 Point End 969 5x65mL CRP 2 Point End 204 1x50/1x20mL
Reagent/Sample tray: Multi-functional reagent and sample carousel with flexibile positions;
HDL-C Direct 2 Point End 517 3x53/3x20mL RF 2 Point End 165 2x20/2x9mL
24 hour non-stop refrigerated compartment (2~12˚C)
LDL-C Direct 2 Point End 273 3x30/3x11mL HbA1c Direct 2 Point End 235 1x50/1x20mL
Reagent bottle volume: 20ml & 70ml
Urea U.V 2 Point Rate 713 3x65/3x20mL Microalbumin 2 Point End 167 2x20/2x6mL
Reagent volume: R1: 10~350µl, R2: 10-200µl, step by 1µl
Enzymatic creatinine 2 Point End 633 3x52/3x20mL CRP Ultra 2 Point End 167 2x20/2x11mL
Sample volume: 3~35µl, step by 0.1µl
Uric Acid 1 Point End 844 4x60mL IgA 2 Point End 156 2x20/2x8mL
Reagent/Sample probe: 1, with Liquid level detection, collision detection and inventory checking
SGOT Rate A 814 3x63/3x20mL IgE 2 Point Rate 141 2x20/2x6mL Probe cleaning: Automatic washing both interior and exterior; Carry-over < 0.1%
SGPT Rate A 814 3x63/3x20mL IgG 2 Point End 165 2x15/2x15mL Sample dilution: Pre-dilution and post-dilution
Albumin 1 Point End 775 4x65mL IgM 2 Point End 190 2x20/2x9mL Dilution vessel: UV plastic semi permanent cuvettes
Total Protein 1 Point End 348 2x50mL Lp (a) 2 Point Rate 167 2x20/2x6mL External Bar Code Reader (optional) For sample and reagent programming; capable to communicate
Bilirubin Total TAB 2 Point End 667 3x63/3x8mL C3 2 Point End 160 2x19/2x5mL with LIS in a bi-directional mode
Bilirubin Driect 2 Point End 667 3x63/3x8mL C4 2 Point End 160 2x19/2x5mL Reaction System
Alkaline phosphatase Rate A 814 3x63/3x20mL Apo A1 2 Point End 133 2x20/2x5mL Reaction rotor: Rotating tray, containing 120 cuvettes (1 set 20 pieces)
α-Amylase Rate A 347 2x55mL Apo B 2 Point End 133 2x20/2x5mL Cuvette: Optical diameter of 6mm
Gamma GT Rate A 271 1x65/1x20mL Ceruloplasmin 2 Point End 114 2x20/2x5mL Reaction volume: 150~550µl
Calcium (Arsenazo ) 1 Point End 231 2x40mL Cystatin C 2 Point End 167 2x20/2x6mL Reaction temperature: 37±0.1˚C
Inorganic Phosphorus 1 Point End 459 2x65mL Ferritin Rate A 156 2x20/2x8mL Reaction disk constant temperature: Circulating water
LDH – P Rate A 273 3x30/3x11mL Prealbumin 2 Point End 119 2x20/2x4mL Mixing system: Teflon coated stirrer with triple speed mechanism
Laundry system: Adopting 8 stops, 12 steps by warm water rinsing
Optical System

Company reserves the right to change any design and technical features of the product at any time, If needed.
Light Source: 12V/20W, Quartz halogen lamp with hydro cooling system
Wavelength: 340nm, 380nm, 405nm, 450nm, 480nm, 505nm, 546nm, 570nm, 600nm, 660nm,
700nm, 750 or 800nm
Monochromator: Grating Photometry
Linear range: 0~3.3Abs
Detector: Silica photo-diode array
Calibration and QC
Calibration method Linear (one-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline,
Exponential, Polynomial
Parabola Control rules: Westgard multi-rule, L-J Chart
Operation Unit
PC Operation system: Windows® XP, Windows® 7, Windows® 10
PC configuration: CPU > 2.8 Ghz (dual core processor), Memory >1GB; Harddisk >160GB
Interface: TCP/IP Network connection, standard RS-232C and USB interface
Working Conditions
ISO 9001:2015
Power Supply: 100~240 VAC | 50/60Hz | Power 650VA
EN ISO 13485:2016
Ambient Temperature: 15~25˚C
Relative Humidity: 40% - 85%
Water consumption: 6L/hour
ACCURACY GUARANTEED SYSTEM REAGENTS Dimension: 998x752x517mm (LxWxH)
Agappe provides the genuine system pack reagents for genuine results Weight: Approx. 120 Kg Fully Automated Clinical Chemistry Analyzer

AGAPPE DIAGNOSTICS SWITZERLAND GmbH

V0-09/18
Knonauerstrasse 54 - 6330 Cham, Switzerland.
Tel: +41 41 780 60 10 | Fax: +41 41 780 60 11 |info@agappeswiss.com | www.agappeswiss.com
Fully Automated Clinical Chemistry Analyzer Fully Automated Clinical Chemistry Analyzer Fully Automated Clinical Chemistry Analyzer

Mispa CCXL is a fully automated clinical chemistry analyzer with throughput of 240 T/H. Mispa CCXL is a
Sample/Reagent Pipetting Mechanism logical choice for customers in various segments, ensuring convenience, flexibility, affordability & accuracy. Mixing System
Ÿ Probe with liquid level detection and collision protection Ÿ Triple speed stirrer mechanism for optimal mixing
Ÿ The syringes are made of long life high precision ceramic piston with Ÿ Teflon coated stirrer ensures no carry over effect
low maintenance Photometric throughput of 240 Tests/hour Multifunctional sample & reagent carousel Ÿ Flat paddle stirrer offers homogeneous mixing
Ÿ Analyzer has special degassing device to remove air dissolved in tube Ÿ Enhanced durability due to minimal maintenance
system for accurate pipetting
Ÿ Probe with internal and external washing
Ÿ 60nm polished probe with nano coating technology which effectively HCFG photometry system Waterfall washing system
reduces cross-contamination

Efficient laundry system adopting 8 stops 12 steps Constant temperature reaction cuvettes
Photometry system Semi-Permanent Reaction Cuvette
Ÿ HCFG (Holographic Concave Flat field Grating) rear Ÿ 1 Set 20 Pieces
spectrophotometry reducing ambient light interference Hydro cooling system for enhanced lamp durability Teflon coated stirrer with triple speed mechanism Ÿ 6 Sets 120 Pieces
Ÿ Photospot technology to reach super trace analysis Ÿ UV Plastic semi-permanent cuvettes ensures cost saving, anti-
Ÿ Water cooling method for long lamp life corrosion and enhanced durability of 9 to 12 months
Ÿ Specially designed lamp placement to reduce signal
attenuation and interference
Ÿ 12 wavelengths ranging from 340nm to 800nm

Multi-function sample & reagent carousel Calibration and QC Program


Ÿ Total 67 positions including reagent and sample, user defined Ÿ Linear and non linear calibration with 9 types of calibration curve
proportion of reagent and sample positions Ÿ 6 different levels of calibrator for each item can be programmed
Ÿ 24 hour continuous cooling condition ensures the quality of Ÿ Calibration tracing possibility depicting calibration K value
reagent control and calibrator variation trends, help reduce system errors
Ÿ Can accommodate 20ml and 70ml reagent bottles Ÿ QC with Westgard multi rules
Ÿ Primary sample tube can be used Ÿ QC plot with L-J and cumulative statistics
Ÿ Barcode reading facility Ÿ Automatic error reporting compliant with lab QC management
Ÿ Single/double reagent testing

User Friendly Software


Constant Temperature Reaction Cuvettes Ÿ Simple and user friendly software
Ÿ Real time online help system
Ÿ Adopts a recycling water constant temperature device
Ÿ Multiple self monitoring to ensure high data efficiency
Ÿ Automatically changing water and adding de-foamer
Ÿ Multiple report formats with automatic print function
Ÿ PID thermostat technology to ensure temperature 37 ± 0.10C
Ÿ Serum indices available
Ÿ Laundry adopting 8 Stops, 12 steps, two time recycling detergent
Ÿ Carryover prevention program available
and warm water rinsing
Ÿ Sample auto dilution with user define conditions (3 - 115 times)

AUTOMATION AT ITS FINEST PROBES WITH NANO COATING TECHNOLOGY