Kunkin - KP184 - Exe v1 - 0 - 17 - 0 User Manual
Kunkin - KP184 - Exe v1 - 0 - 17 - 0 User Manual
Kunkin - KP184 - Exe v1 - 0 - 17 - 0 User Manual
Installation Instructions
• Download the setup executable installer to any directory with
permissions.
• Double click it to execute the setup installer.
• Follow all the prompts that appear.
• It is suggested that you follow all the defaults, but you can select an
alternate directory for the install.
• You are good to go.
Main Screen
Page 1
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
The standard Kunkin KP184 operating parameters are omitted from this
manual, please refer to the equipment manual for reference.
Page 2
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
The screen values are always refreshed 3 times per second. The Data
capture sampling period refers to the .csv and plot graph data creation.
The captured data includes elapsed time in seconds, voltage, current,
capacity (Ah) and energy (Wh). I probably should add Power in Watts.
Check the CRC LSB First option if you experience difficulties establishing the
serial communication with KP184 later models (Firmware 2020 and up – can
be check at the Kunkin screen at startup)
File settings group: Default file names and folders for captured data, graph
images and battery capacity discharge test reports.
A verification is made to check if the folder and path are valid. In error case it
will default to “My Documents\KP184 Modbus\”.
All the data file names (.csv; .png; .pdf) are used as base file names,
numbers are automatically added in subsequent files to avoid overwriting
files.
Please use a decent USB serial adapter in the absence of a legacy COM
port. A FTDI or Silabs processor chips are far better than a low the cost
Page 3
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
This unregistered product works in trial mode for 15 minutes, after the expired
period the data capture features are disabled. If you find out that this software
product is useful, you can register sending your registration ID from this
configuration dialog to interflexo@sapo.pt along with $30 USD by PayPal.
Interflexo will then email you back an unlocking registration KEY after
payment clearance.
All controls have tool tips (hints) when you carefully hover the cursor over the
controls.
Page 4
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
Remote/Local Toggles the voltage sense mode. Local uses the power
connectors and is affected by the voltage drop (proportional to the electronic
load current) on the cables, connectors, etc. Remote uses the independent
front BNC connector for voltage sense and is not affected by the voltage drop
because it does not carry any meaningful current. You may connect the
alligators directly to the battery (being tested) terminals, please respect the
polarity. The assembly bellow costs about 7 Eur.
Page 5
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
CV/CC/CW/CR Signals and Cycles through the four load modes. Constant
Voltage, Constant Current, Constant Power and Constant Resistance.
Toggles the Power up load state. OFF (dimmed) the electronic load
is always OFF at power up. ON the electronic load remembers the last state
when was power down and uses it at power up.
0.000V Threshold voltage for load OFF programmed value. Please notice
the tool tip (hint) to check for hardware or software handling type. When using
the software type the indicator flashes red when stopping the load.
0.0 0.0 A/ms Current slew down rate and current slew up rate when
switching ON/OFF.
Page 6
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
General Tab:
Set the:
and use the corresponding send button. The new setting is sent to
Kunkin KP184. This button can appear enabled or disabled according to the
setting entered string syntax validation and COM port communication status.
Page 7
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
key_sound
power_up_state
voltage_sense
load_mode
cv_setting
cc_setting
cr_setting
cw_setting
threshold_volt
slew_up_rate
slew_dn_rate
dynamic_mode_1
level_1_cur
Page 8
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
level_1_tim
level_2_cur
level_2_tim
dynamic_mode_2
battery_mode
bat_end_volt
bat_go_half_cur
bat_cap_units
bat_test_over_signal
Saves the plot graph on a previously configured PDF file name and
folder. The file name is used as a base file name, numbers are automatically
added in subsequent files to avoid overwriting files. It can be used during data
Page 9
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
acquisition. A matching “.png” image file is also added for easier document
insertion. After file creation the PDF file is automatically opened on Adobe
Acrobat Reader or configured web browser.
This PDF report is built based on a supplied html template file named
“Report.html” that must exist on the program directory. You can make minor
and careful changes to this template file to fit your particular needs. No
support will be given to accommodate these changes. This html template file
is used by the xhtml2pdf Python module. Please refer to the module
documentation here:
https://xhtml2pdf.readthedocs.io/en/latest/
Page 10
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
The battery capacity discharge test may have ended automatically when the
end test voltage was reached, time limit was reached or test was user
terminated or temporary suspended by a load OFF command.
The report is saved on a previously configured PDF file name and folder. The
file name is used as a base file name, numbers are automatically added in
subsequent files to avoid overwriting files. After file creation the PDF file is
automatically opened on Adobe Acrobat Reader or configured web browser.
On the next page you can find an example of a dummy report test page.
Page 11
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
Page 12
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
Calendar date picker for timer to turn Load ON/OFF. Adjust this
setting prior to COM port connection. The settings below can be adjusted
during serial communications.
Timer disabled
These last three buttons cycle through each other when the user clicks the
button. The tool tips (hints) are always updated.
The input text box accepts a time in the format hh:mm:ss. Adjust this setting
preferably with Timer disabled.
The input text box content is always checked for errors and in case of invalid
value the current time is used and the timer is disabled.
Page 13
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
Battery Tab:
Toggles battery
capacity test mode ON/OFF.
Mode ON unlocks Battery
tab.
Fill the Load Set Value and End Test Voltage (LVC) and
The software supervises LVC value and finishes the test even if Kunkin
KP184 feels like going further.
The Max Duration input text box allows the user to specify a maximum test
duration in seconds. Zero disables this feature.
Page 14
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
This occurrence happens on several other features. Once again the Kunkin
software behaves in the exact same manner. There are no errors reported
through the Modbus protocol.
“Stop test at once” option ends the test when the programmed end voltage
condition is reached.
Page 15
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
“Go half current until the next under voltage” option lowers the discharge
current to half the initially configured value at least 10mV prior to the test end
voltage reached mark. The next time the test programmed end voltage
condition is reached the test is ended. A specific log is added for this event.
In case of high voltage instability due to bad connections or other reasons the
10mV step may not be enough to avoid tripping prematurely the test end
voltage condition. Please check your electric connections.
The radio buttons show the parameter value and allow the user to change it.
The options are:
• Buzz once
• Buzz until a key is pressed on the Kunkin KP184
• The PASS output electric signal is enabled after the test is finished
• The plot graph displays a line for the voltage and another for the
capacity.
• On the LCD area the large Power indicator shows the battery capacity
(Ah). Clicking on it resets the capacity value to 0.0Ah.
• On the LCD area the Ah/Wh indicator allows the user to switch
between Ah/Wh when clicked. The alternate units and Power are
always displayed the battery tab.
Page 16
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
In Resist Tab:
Toggles Internal
resistance test mode
ON/OFF. Mode ON unlocks
In Resist tab.
The current steps have about 1 second duration to allow for voltage settling.
The result in ohms (Ω) is displayed on the bottom under “Internal resistance”.
The user can start data acquisition previous to the test start on the General
tab and log the process.
Page 17
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
In CV C Lim Tab:
Kunkin is hardware
configured to CC and a PID
configurable loop holds the
configured CV setting by
permanently adjusting the CC
load.
GEN/CVL/... Toggle the test mode indicator to access CVL test. It unlocks
In CVL tab.
Page 18
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
Cycle Tab:
The dynamic Cycle mode allows the user to program a custom variable CC
load profile of unlimited steps, ramps and repetitions.
Page 19
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
Toggle the test mode indicator to access the special test modes GEN / CVL /
CYC. CYC unlocks the Cycle Tab
The format is simple and can be edited on the multiline text box.
Any empty lines are ignored. All lines beginning with ‘#’ are treated as a
comment.
The programmed profile can be repeated any number of times and stops
automatically or indefinetly and must be stopped with Load Switch button.
Use the checkbox and edit box to enter the cycle repetition information.
The load commands are sent in real time through the serial interface. Use
115200 baud to reduce the lag as much as possible. The minimum
programmable time step is 0.1s
Page 20
User Manual KP184.exe v1.0.17.0 Interflexo, Lda – Mar 2021
Log Tab:
Page 21