Stopwatch
Stopwatch
1 2
Reading Readings of m
Parameter L.C Hr Min Sec m sec Hr Min Sec
in Sec std sec
CERTIFICATE No.:
CAL DATE:
DUE DATE:
3 4 5 Average 1
m
Hr Min Sec Hr Min Sec m Sec Hr Min Sec m Sec Hr Min Sec m Sec In Sec
Sec
0 0 50 74 0 0 50 84 0 0 50 98 00 00 50 82 51
Calibration Result
Reading On UUC Expanded
Reading On Standard UUC Reading
Error in Sec Uncertainty
in Sec Hrs Min Sec 1/100 Sec in Sec
in ±,Sec
50.000 00 00 50 82 50.82 0.82 0.117
100.000 00 00 100 79 100.79 0.79 0.399
300.000 00 00 300 83 300.83 0.83 1.001
900.000 00 00 900 50 900.50 0.50 1.122
1800.000 00 00 1800 88 1800.88 0.88 2.164
3600.000 00 00 3600 81 3600.81 0.81 4.802
1. The reported expanded uncertainty is stated as a standard uncertainty multiplied by coverage factor ‘k=2’ at 95.45% confidence level.
2. The standards used for calibration were calibrated by using reference standard traceable to National/International standard.
3. The calibration result reported in this certificate are valid at the time of & under the stated conditions of measurement
4. Calibration status Sticker affixed on the instruments.
5. This Certificate refers only to the instrument calibrated and shall not be reproduced except ,in full without the written approval from
BECL
Page 13 of 16
IN Value
Type-A Contribution 0.000490 Sec
Type-B Contribution
UB1 Uncertanity of Standard used for calibration -
B1:= 0.001 sec sec 0.00050 Sec
= 5.71E+01
= 6.21E+05