Test Equipment Solutions Datasheet

Test Equipment Solutions Ltd specialise in the second user sale, rental and distribution of
quality test & measurement (T&M) equipment. We stock all major equipment types such as
spectrum analyzers, signal generators, oscilloscopes, power meters, logic analysers etc from
all the major suppliers such as Agilent, Tektronix, Anritsu and Rohde & Schwarz.

We are focused at the professional end of the marketplace, primarily working with customers
for whom high performance, quality and service are key, whilst realising the cost savings that
second user equipment offers. As such, we fully test & refurbish equipment in our in-house,
traceable Lab. Items are supplied with manuals, accessories and typically a full no-quibble 1
year warranty. Our staff have extensive backgrounds in T&M, totalling over 150 years of
combined experience, which enables us to deliver industry-leading service and support. We
endeavour to be customer focused in every way right down to the detail, such as offering free
delivery on sales, presenting flexible technical + commercial solutions and supplying a loan
unit during warranty repair, if available.

As well as the headline benefit of cost saving, second user offers shorter lead times, higher
reliability and multivendor solutions. Rental, of course, is ideal for shorter term needs and
offers fast delivery, flexibility, try-before-you-buy, zero capital expenditure, lower risk and off
balance sheet accounting. Both second user and rental improve the key business measure of
Return On Capital Employed.

We are based at Aldermaston in the UK from where we supply test equipment worldwide.
Our facility incorporates Sales, Support, Admin, Logistics and our own in-house Lab.

All products supplied by Test Equipment Solutions include:

- No-quibble parts & labour warranty (we provide transport for UK mainland addresses).
- Free loan equipment during warranty repair, if available.
- Full electrical, mechanical and safety refurbishment in our 40GHz in-house Lab.
- Certificate of Conformance (calibration available on request).
- Manuals and accessories required for normal operation.
- Free insured delivery to your UK mainland address (sales).
- Support from our team of seasoned Test & Measurement engineers.
- ISO9001 quality assurance.

Test Equipment Solutions Ltd


Unit 3 Zodiac House
Calleva Park
Aldermaston
Berkshire
RG7 8HN

T: 01183 800 800


F: 01183 800 804

Email: info@TestEquipmentHQ.com
Web: www.TestEquipmentHQ.com
MULTIFUNCTION MicroCal
Accurate and Reliable
CALIBRATORS
365 Days a Year
MicroCal 200/200+
Table IN - OUT Range MicroCal 2000+
High Accuracy
Sensor or Total range Accuracy Range Resolution
MicroCal 200
Accuracy
MicroCal 200+
Accuracy
MicroCal 2000+
Resolution Accuracy
Multifunctional Process Calibrators
parameter (% of rdg) (% of rdg) (% of rdg)
Tc J -210 ... 1200°C -190 ... 1200°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-350 ... 2200°F -310 ... 2192°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Tc K -270 ... 1370°C -160 ... 1260°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
454 ... 2500°F -256 ... 2300°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Tc T -270 ... 400°C -130 ... 400°C 0.01°C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-454 ... 760°F -238 ... 752°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Tc R -50 ... 1760°C 150 ... 1760°C 0.1 °C ±(0.02% + 0.2 °C) ± (0.01% + 0.2 °C) 0.01°C ±(0.005% + 0.2 °C)
-60 ... 3200°F 302 ... 3200°F 0.1 °F ±(0.02% + 0.36°F) ± (0.01% + 0.36 °F) 0.1 °F ±(0.005% + 0.36°F)
Tc S -50 ... 1760°C 170 ... 1760°C 0.1 °C ±(0.02% + 0.2 °C) ± (0.01% + 0.2 °C) 0.01°C ±(0.005% + 0.2 °C)
-60 ... 3200°F 338 ... 3200°F 0.1 °F ±(0.02% + 0.36°F) ± (0.01% + 0.36 °F) 0.1 °F ±(0.005% + 0.36°F)
MULTIFUNCTION
Tc B 50 ... 1820°C 920 ... 1820°C 0.1 °C ±(0.02% + 0.3 °C) ± (0.01% + 0.3 °C) 0.01°C ±(0.01% + 0.3 °C) CALIBRATORS
140 ... 3310°F 1688 ... 3308°F 0.1 °F ±(0.02% + 0.54°F) ± (0.01% + 0.54 °F) 0.1 °F ±(0.01% + 0.54°F)
Tc C 0 ... 2300°C 0 ... 2000°C 0.1 °C ±(0.02% + 0.2 °C) ± (0.01% + 0.2 °C) 0.1 °C ±(0.01% + 0.2 °C)
32 ... 4172°F 32 ... 3632°F 0.1 °F ±(0.02% + 0.36°F) ± (0.01% + 0.36 °F) 0.1 °F ±(0.01% + 0.36 °F)
Tc G 0 ... 2300°C 190 ... 2300°C 0.1 °C ±(0.02% + 0.3 °C) ± (0.01% + 0.3 °C) 0.1 °C ±(0.01% + 0.3 °C)
32 ... 4172°F 374 ... 4172°F 0.1 °F ±(0.02% + 0.54°F) ± (0.01% + 0.54 °F) 0.1 °F ±(0.01% + 0.54 °F)
Portable, Rugged and
Tc D 0 ... 2300°C 0 ... 2130°C 0.1 °C ±(0.02% + 0.3 °C) ± (0.01% + 0.3 °C) 0.1 °C ±(0.01% + 0.3 °C)
32 ... 4172°F 32 ... 3866°F 0.1 °F ±(0.02% + 0.54°F) ± (0.01% + 0.54 °F) 0.1 °F ±(0.01% + 0.54 °F) Easy to Use
Tc U -200 ... 400°C -160 ... 400°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-330 ... 760°F -256 ... 752°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F) Measures and Sources 14 Type
Tc L -200 ... 760°C -200 ... 760°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
of TCs and 10 RTDs
-330 ... 1400°F -328 ... 1400°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Tc N -270 ... 1300°C 0 ... 1300°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-450 ... 2380°F 32 ... 2372°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F) 24 Vdc Loop Power Supply
Tc E -270 ... 1000°C -200 ... 1000°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-454 ... 1840°F -328 ... 1832°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
In-Line Digit Setting
Tc F 0 ... 1400°C 0 ... 1400°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
32 ... 2560°F 32 ... 2552°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Pannel Mounting Kit
Pt100 IEC -200 ... 850°C -200 ... 850°C 0.01°C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.01°C ±(0.005% + 0.05°C)
OIML, a 3926 -330 ... 1570°F -328 ... 1562°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.005% + 0.09°F)
MicroCal PM200
Pt100 -200 ... 650°C -200 ... 650°C 0.01°C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.01°C ±(0.005% + 0.05°C)
a 3902 -330 ... 1210°F -328 ... 1210°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.005% + 0.09°F) Optional Module
Pt100 JIS -200 ... 600°C -200 ... 600°C 0.01°C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.01°C ±(0.005% + 0.05°C) to Generate/Measure Pressure
SAMA -330 ... 1120°F -328 ... 1112°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.005% + 0.09°F)
Pt 200 -200 ... 850°C -200 ... 850°C 0.1 °C ±(0.02% + 0.15°C) ± (0.01% + 0.15 °C) 0.01°C ±(0.005% + 0.15°C)
CalpMan 2007
-330 ... 1570°F -328 ... 1562°F 0.1 °F ±(0.02% + 0.27°F) ± (0.01% + 0.27 °F) 0.1 °F ±(0.005% + 0.27°F)
Pt 500 -200 ... 850°C -200 ... 530°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C) Software for Automatic
-330 ... 1570°F -328 ... 986°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F) Calibration and Documenting in
Pt1000 IEC -200 ... 850°C -200 ... 850°C 0.01°C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C) Compliance with ISO9001
OIML -330 ... 1570°F -328 ... 1562°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)

CU10 -70 ... 150°C -70 ... 150°C 0.1 °C ±(0.02% + 0.4 °C) ± (0.01% + 0.4 °C) 0.1 °C ±(0.01% + 0.4 °C)
-100 ... 310°F -94 ... 302°F 0.1 °F ±(0.02% + 0.72°F) ± (0.01% + 0.72 °F) 0.1 °F ±(0.01% + 0.72°F)
CU100 -180 ... 150°C -180 ... 150°C 0.1 °C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.1 °C ±(0.01% + 0.05°C)
-300 ... 310°F -292 ... 302°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.01% + 0.09°F)
Ni100 -60 ... 180°C -60 ... 180°C 0.1 °C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.1 °C ±(0.01% + 0.05°C)
-80 ... 360°F -76 ... 356°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.01% + 0.09°F)
Ni120 0 ... 150°C 0 ... 150°C 0.1 °C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.1 °C ±(0.01% + 0.05°C)
32 ... 310°F 32 ... 302°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.01% + 0.09°F)

mV (L) -20 ... +200mV 1 µ V ±(0.02% + 2 µ V) ± (0.01% + 2 µ V) 0.1 µV ±(0.0035% + 1 µ V)


mV (H) -0.2 ... +2 V 10 µ V ±(0.02% + 10 µ V) ± (0.01% +10 µ V) 1 µ V ±(0.005% + 10 µ V)
V 10 µ bt !
No dou Accurate
-2 ... +20 V 0.1 mV ±(0.02% + 0.08mV) ± (0.01% + 0.08mV) V ±(0.005% + 0.08mV)
mA (In) -5 ... +50 mA 0.1 µA ±(0.02% + 0.4 µ A) ± (0.01% + 0.4 µ A) 0.1 µA ±(0.005% + 0.4 µ A)
st
mA (Out) 0 ... +50 mA 0.1 µA ±(0.02% + 0.4 µ A) ± (0.01% + 0.4 µ A) 0.1 µA ±(0.005% + 0.4 µ A)
e of the Mo rdg)
Bulletin 08-44.3 E MicroCal 200_2000 - ed. 04-2007

Ω IN 0 ... 500 Ω 1 mΩ ±(0.02% + 12 mΩ ) ± (0.01% +12 mΩ ) 1 mΩ ±(0.005% + 12 mΩ ) O n 5%


(0.003 or
Specifications may change without notice

0 ... 5000 Ω 0.01 Ω ±(0.02% +120 mΩ ) ± (0.01% +120 mΩ ) 0.01 Ω ±(0.005% +120 mΩ )
b le C alibrat

OUT 0 ... 500 Ω 1 mΩ ±(0.02% + 20 mΩ ) ± (0.01% + 20 mΩ ) 1 mΩ ±(0.005% + 12 mΩ )
Por t a y!
5000 Ω 0.01 Ω 0.01 Ω le Toda
Availab
0 ... ±(0.02% +200 mΩ ) ± (0.01% +200 mΩ ) ±(0.005% +120 mΩ )
Frequency 1 ... 200 Hz 0.001 Hz ±(0.005% + 0.001 Hz) ±(0.005% + 0.001 Hz) 0.001 Hz ±(0.005% + 0.001 Hz)
1 ... 2000 Hz 0.01 ±(0.005% + 0.001 Hz) ±(0.005% + 0.001 Hz) 0.01 Hz ±(0.005% + 0.001 Hz)
1 ... 20000 Hz 0.1 Hz ±(0.005% + 0.001 Hz) ±(0.005% + 0.001 Hz) 0.1 Hz ±(0.005% + 0.001 Hz)
Pulse counter
6
0 ... 10 counts 1 count infinite infinite 1 count infinite
Instrum
Pulse (Out) 0 ... 6000 pulse/min 1 pulse/min 1 pulse / min 1 pulse / min 1 pulse / min 1 pulse / min ron e
ot

Q
nt

0 ... 36000 pulse/h 1 pulse/h 1 pulse / min 1 pulse / min 1 pulse / h 1 pulse / min
r
Eu

Note: • The MicroCal 2000+ relative accuracies shown above are stated for 360 • Typical 2 years relative accuracy can be estimated by multiplying the “% of
Ce

m
ste

• The MicroCal 200/200+ relative accuracies shown above are stated for 360 days and the operative conditions are from +70°F to +77°F reading" specifications by 1.4. fie
rti

y
days and the operative conditions are from +64°F to +82°F • Typical 90 days relative accuracy can be estimated by dividing the "% of • All input ranges: additional error ±1 digit. d Quality S
reading" specifications by 1.6. • Eurotron traceability chart and uncertainty can be supplied on request.

www.eurotron.com
12 All descriptions are related to a fully optioned instrument. See last page for the different configurations
MULTIFUNCTION
MicroCal MicroCal MULTIFUNCTION
CALIBRATORS To Keep your Test and Process Equipment High Technology for Daily Use CALIBRATORS
Accurately Calibrated for ISO 9000 Compliance in Field and Laboratory

General SPECIFICATIONS

The MicroCal 200/200+ and MicroCal 2000+ are multifunctional calibrators IN/OUT parameters: Signal type: mV, V, mA, Ω , KΩ , frequency, pulses
with insulated channels. The instruments are portable and developed to meet thermocouples type: J, K, T, R, S, B, N, C, E, U, L, F, G, D
the needs of instrumentation engineers and Quality managers, both in resistance thermometers: Pt100 IEC, OIML, USLAB, US, SAMA, JIS, Pt200, 500,
1000, 1000 OIML, Ni100, Ni120, Cu10, Cu100
laboratory and in field work. Reference junction compensation: internal automatic: from -10 °C to +55 °C (14°F to 130°F)
This units are accurate, rugged, compact and easy to use. They are the ideal external adjustable: from -50 °C to +100 °C (-58°F to 210°F)
solution to simulate and measure simultaneously: Voltage, current, resistance, remote with external Pt100: from -10°C to +100 °C (14°F to 210°F)
thermocouple, resistance thermometers, frequency and pulse. Rj compensation drift: ± 0.015°C/°C (from -10 °C to +55 °C)
Rj compensation error: internal : ±0.15°C (±0.3°F)
Advanced flexibility and high performance has been achieved using a 32 bit remote : ±0.3°C (±0.6°F)
microprocessor and a fast A/D conversion technology. The calibrator’s internal Common mode rejection: >140 dB at ac operation
memory stores all data for normalized IEC, DIN and JIS thermoelectric sensors Normal mode rejection: >70 dB at 50 or 60 Hz
Temperature stability: MicroCal 200/200+: for temperature exceeding the band +18°C to +28°C (from +64 °C to +84 °F)
for both IPTS68 and ITS90 International Temperature Scale
full scale: ± 8 ppm/°C
zero: ± 0.2 mV /°C
MicroCal 2000+: for temperature exceeding the band +21°C to +25°C (from +70 °C to +77 °F)
The microprocessor performs automatic Table top
=
full scale: ± 3 ppm/°C
linearization and cold junction The case is equipped with 2 pivot feet to zero: ± 0.2 mV /°C
compensation to assure high accuracy. It is change the vertical viewing angle when Output impedance (emf output): < 0.5 Ω with 0.5 mA maximum current
possible to set the calibrator to execute using the instrument on the top of a table.
Input impedance (mV, V and Tc ranges): >10 MΩ
menu-driven calibration procedures for Input impedance (mA ranges): <140 Ω @ 1 mA
your instruments in field work. Quality System Source resistance effects: ±1 mV error for 1000 Ω source resistance
Research, development, production, Rtd and Ω simulation excitation current: MicroCal 200/200+: from 0.01 to 5 mA
Both Channel 1 (Out) and Channel 2 (In) inspection and certification activities are MicroCal 2000+: from 0.01 to 2 mA
have the following operative mode defined by methods and procedures of the Rtd and Ω
measurement excitation current: ~ 0.4 mA @ 400Ω
capabilities: Eurotron Quality System inspected for ~ 0.04 mA @ 4000Ω
= millivolts compliance and certified ISO9001 by Rtd connection: 2, 3, and 4 wires
= volts GASTEC, a Dutch notified body. Rtd cable compensation: up to 100 Ω (each wire)
= milliamperes (active and passive loop) Rtd cable compensation error (Pt100): ±0.005°C/ Ω of total wire
= Ohms EMC Conformity Maximum load resistance: 1000 Ω @ 20 mA
= temperature with thermocouples Display: graphic LCD 240 x 64 dots display with LED backlight device
The instrument case, made in shock-
= temperature with resistance Measurement sampling time: 250 ms
resistant injection molded ABS + Output noise (at 300 Hz): <2 mVpp for ranges up to 200 mV f.s.,
thermometers polycarbonate has an internal metal <10 mVpp for ranges up to 2 V f.s.
= frequency
coating to fulfill the prevision of the directive <80 mVpp for ranges up to 20 V f.s.
= pulse
89/336/CEE Electromagnetic Digital interface: full bidirectional TTL (a RS232 adapter normal or insulated, is available as an option)
Compatibility. Channel 1_Channel 2 insulation: 50 Vdc
Remote auxiliary inputs are available for: Calculation functions: hold, max, min, offset, average
Relative humidity
= and temperature Selection °C/°F/K: through the configuration procedure
module In/Out data memory: 20 data with manual or automatic recall
= Two internal sensors and built-in hand Logging mode: 1500 input data items (optional memory card for memory extension)
pump pressure module Convert function: displays the electrical equivalent of the engineering unit
Scale factor: 5 different settings with zero and span programmable within -399999 and +999999
The ergonomic case design allows the user Square root: in combination with scale factor
Calibration: self learning technique with automatic procedure
the calibrator in three different ways:
Power supply: external charger and rechargeable Ni-Cd battery
= Portable
Self contained operation: 6h on Tc and mV input/output (backlight Off)
Two different leather cases, with a cover 3.5h with 20 mA simulation (backlight Off)
and shoulder strap, are available upon Recharging time: 5h at 90% and 6h at 99% with instrument switched off.
request for instrument alone or The battery charging is active only with the instrument switched off.
instrument, printer and accessories. Line operation: 100 - 120 - 230 Vac through the external battery charger
These are extremely useful for practical Line transformer insulation: 2500 Vac
use since they leave one hand free for Firmware release identification: release code on the display
instrument tuning. Operating environment temperature range: from -10 °C to +55 °C (from 14 °C to 130 °F)
= Panel mounting Storage temperature range: from -30 °C to +60 °C (from -22 °C to 140 °F)
It requires a panel cutout of 9.7” x 3.5”. Case: Injection moulded ABS with internal metal coating
The instrument bezel flange butts against Dimensions: 264 x 96 x 172 mm (10.4”x3.8”x6.8”) DIN size
the front of the mounting plate; two Weights: nett: 8.8 lb. gross: 12.1 lb.
lateral mounting brackets fit over the
instrument rear panel

6 11
MicroCal 200/200+ MULTIFUNCTION
CALIBRATORS
MicroCal 2000+
"Excellence in measurements"

Instrument Highlight

Display & Keyboard


Comparative Table MicroCal MicroCal MicroCal The large graphic display with back light
200 200+ 2000+ allows a simultaneous indication of the
Accuracy ±0.02% rdg ± 0.01%rdg ± 0.0035% rdg measured and the simulated value (large
digit), together with all data related to the
32-bit microprocessor and flash memory for
firmware upgrading through serial interface signal as: engineering unit, type of sensor
All normalized IEC, DIN, JIS thermocouples or signal, temperature scale, cold
Pt, Ni, Cu resistance thermometers
junction selection, battery level, etc.
A swap key allow to change the position
mA, mV, V, W, frequency, pulse, counter
of the IN and OUT parameters on the
IPTS 68 and ITS 90 linearization
display.
Current output mode directly
The operative set-up mode is simplified
on active or passive loops
by a sequence of menu pages that only
Bidirectional serial interface Option
require <Select> and <Enter>
Communication bus for extension
to pressure and other optional modules
instructions.
A full set of instruction pages are memory
Optional dedicated external impact
type printerOptionOptionOption stored for direct operator assistance.
Pannel mounting brackets Option Option Option The relevant instruction may be recalled
through the <Help> key.
Traceable Report of Calibration
RAM extension with PCMCIA memory card
Non-volatile memory with real-time clock Option
Logging and direct real-time graph
with movable cursor to read the required Option Option
actual value
mV and V I/O display 6 digits 6 digits 6 ½ digits (0.1 µ
V)

Real Time Graphic Logging

MicroCal 200/200+
6 digit display

MicroCal 2000+
6 ½ digit display
7
MULTIFUNCTION MicroCal 200/200+
CALIBRATORS MicroCal 2000+

Firmware
The firmware is stored inside a flash
memory and allows a fast and easy
upgrade of the instrument using a
standard PC. Firmware includes the
following capabilities:
• multiple measurement and
generation mode
• signal processing: filter, average,
min, max
• PC downloadable test procedure
• DataLogger and graphic recorder
• switch test routine
• ramping and stepping for dynamic
testing
• user definable linearization
• user entry of probe specific
calibration coefficients.

Scale Factor - Square Root


All non-temperature ranges are fully
programmable to read both input and
output values in terms of engineering
compensation temperature (from -58 or to calibrate totalizers and counters.
units. Four programmable
to +210°F) by keyboard. The instrument can be configured to
alphanumeric characters are
measure frequency and count pulse
available on the display to show the
Transmitter Simulation (totalizer mode). Technical units in Hz,
symbol of the parameter being
The calibrator can be connected to pulse/h and pulse/m. The threshold is
measured (i.e. mbar, % RH, % CO, etc.)
your system inputs to simulate 4-20 mA adjustable from 0 to 20 V with 0.01 V
The unit allows the mA reading and
transmitters. It has an adequate power resolution.
output to be related to flow when using
to drive 20 mA into a load of 1000 Ω in
a DP transmitter across a calibrated
source mode. For fast loop checks, Calculated Readings
flange.
preprogrammed steps can be The units allow measurements of
Rj Compensation recalled to output 4, 8, 12, 16 and 20 unstable input signals through a
Accurate and fast response mA or % equivalent. Since the programmable averaging or min, max
compensation, through a special low instrument has two separate channels. identification. A hold feature is also
thermal capacity design of binding It is possible to read the error directly in included from a keyboard or external
posts, incorporates a thin film, high % (In/Out). The operator can set and contact instructions.
accuracy Pt100 as cold junction change temperature values while
reference. The internal reference obtaining the equivalent output mA. Programmable Signal Converter
allows the maximum accuracy for the The mA mode may be connected The instrument can be used as a
14 to 130°F directly either on passive or on active t e m p o r a r y s i g n a l c o n v e r t e r
current loops. replacement. Any input signal
temperature
(including remote auxiliary inputs) can
range. An external
Frequency - Counts be converted into any of the available
Pt100 sensor can
be connected for The "Out" mode is designed to output signals. The galvanic insulation
s p e c i a l generate zero based square pulse, between the input and output
application (from with an adjustable amplitude, at a channels allow to use this feature in
14 to +210°C). It is frequency up to 20 KHz. A preset field work applications as well.
possible to number of pulses may be
manually set the programmed and transmitted to test

8
MicroCal MULTIFUNCTION
To Document the Inspection CALIBRATORS
"Excellence in measurements" and Calibration Activities

ORDERING CODE

direct numeric entry;


= MicroCal 200
direct keyboard access to n. 20
= 3916 - A - B - C - D
programmable
= memory stored
values. MicroCal 200+
3918 - A - B - C - D

MicroCal 2000+
End
step 3928 - A - B - C - D
Each instrument is supplied with USB cable, PC
Software Package (Linman, Logman), leather
carrying case with shoulder strap, Report of Calibration
and instruction manual.

Start Table A Options:


0 None
Rise Soack Fall Time 3 TTL/RS232 insulated converter

Table B Line setting - Line cord plug


Report of Calibration 1 120 V 50/60Hz - USA plug
2 230 V 50/60Hz - Schuko plug
Each MicroCal unit is factor y 3 230 V 50/60Hz - UK plug
calibrated and certified against the 4 230 V 50/60Hz - European plug
highest Eurotron Standards, which are 5 100 V 50/60Hz - USA/Japan plug
9 special
periodically certified by an
Internationally recognized laboratory Table C Accessories
0 None
to ensure performance. The units are
6 Brackets for panel mounting
also shipped with a Report of
4 Wire Resistance Thermometer Table D Report of Calibration
Calibration stating the nominal and
Although resistance and temperature 1 Eurotron certificate
actual values and the deviation errors.
with resistance thermometer may be
measured with 2 or 3 wire standard
MicroCal PM200 Pressure Module
connections. The instrument is also
An optional pressure module is
designed for 4-wire measurements for
available with one or two internal
higher accuracy and the ability to use
sensors. The MicroCal PM200 is an
as a reference for your calibrations.
intelligent module and it is used to
extend the MicroCal capability to
Remote Temperature Probe
pressure in/out. MicroCal PM200 can
A high accuracy probe is available on
incorporate pressure/vacuum pump,
request for general purpose
volume adjuster and ventilation valve.
temperature measurement and/or
External sensors can be used to
remote cold junction compensation.
measure pressure up to 700bar. See
MicroCal PM200 bulletin for more
Simulation Capability
specifications.
Menu-driven set up to generate: MicroCal PM200
= a continuous or step ramp output Multifunction electrical signals
+ pressure calibrator
where the total time, the starting
point, the final point and the size of
the steps are requested by the set-up
procedure to run the program;
= a continuous programmable cycle
rises, soaks, falls; (see figure)
= the signal value setting uses unique
in-line, single-digit setting mode or a

9
MULTIFUNCTION
MicroCal
CALIBRATORS Easy to use
Easy to Order

CalPMan 2007 Software Automatic procedure run


¡ OFF-LINE Procedure
Built-in communication with
¡
For over 20 years, Eurotron Instruments has documenting calibrators The CalPMan 2007 can also manage the
combined its own ISO 9001, electric, Traceable and editable calibration
¡ off-line procedures with all the Eurotron
thermoelectric and pressure documents documenting calibrators. The test data
measurements, and development of Database backup/Restore ability
¡ can be transferred directly into the
software application experiences. instrument and the test can run by the
CalpMan 2007 has been designed to be The CalpMan 2007 Advanced has the calibrator itself without any connection to
used industries, where there are both following advanced features: a computer. The final results will be stored
laboratory and maintenance on field ¾ Up to 3 instruments of simultaneous into the instrument, and they can be
needs. The CalPMan 2007 is a WindowsTM management; transferred from it directly into the
2000/XP software designed to plan, ¾ Advanced control of the test point; CalPMan 2007 database. As transferred,
manage and document all the ¾ Multi-parameter procedures run; is it possible to print out the certificates
calibrations and the certifications of the ¾ Simultaneous certification of more than and manage the instrument database.
process instrumentation. The software can one temperature sensors;
manage the automatic procedures of ¾ Password manager and advanced
the Eurotron calibrators. data security ability.

Two software versions are available: the


CalpMan 2007 and the CalpMan 2007
Advanced.
The main features are:
¡Easy to use;
¡Compliant with ISO9001 requirements;
¡Standards and instruments expired date
management;

FIVE STEPS FOR IN-FIELD DOCUMENTING CALIBRATION


2 Transfer the procedure
into the calibrator The MicroCal 200/200+ and 2000+ Series are
documenting calibrators able to load test
strategies using the CalPMan 2007 software. It
1
Prepare the procedure 3 can run them and store the results
using the CalPMan Select the instrument automatically, calculate the errors, highlight the
Software (TAG) and run the procedure out of range points, and download data on a
PC to store them and to generate different
reports. The calibrator internal memory can store
Print Transfer the results a whole week procedures.
the report 5 4 on a PC

Calibration IN
Procedure 5.01 bar

InP
ç
ç 5.00 bar

Run Dis Esc Tag Selection


Tag: TRX0GH-00
Pressure transmitter for PLC
PFK3JM-04 located on sector Q28
inside panel PNL8WC-23
Before: 25/07/97 17:07:00 (FAIL)
After : 25/07/97 17:27:00 (OK )

10
MULTIFUNCTION
MicroCal MicroCal MULTIFUNCTION
CALIBRATORS To Keep your Test and Process Equipment High Technology for Daily Use CALIBRATORS
Accurately Calibrated for ISO 9000 Compliance in Field and Laboratory

General SPECIFICATIONS

The MicroCal 200/200+ and MicroCal 2000+ are multifunctional calibrators IN/OUT parameters: Signal type: mV, V, mA, Ω , KΩ , frequency, pulses
with insulated channels. The instruments are portable and developed to meet thermocouples type: J, K, T, R, S, B, N, C, E, U, L, F, G, D
the needs of instrumentation engineers and Quality managers, both in resistance thermometers: Pt100 IEC, OIML, USLAB, US, SAMA, JIS, Pt200, 500,
1000, 1000 OIML, Ni100, Ni120, Cu10, Cu100
laboratory and in field work. Reference junction compensation: internal automatic: from -10 °C to +55 °C (14°F to 130°F)
This units are accurate, rugged, compact and easy to use. They are the ideal external adjustable: from -50 °C to +100 °C (-58°F to 210°F)
solution to simulate and measure simultaneously: Voltage, current, resistance, remote with external Pt100: from -10°C to +100 °C (14°F to 210°F)
thermocouple, resistance thermometers, frequency and pulse. Rj compensation drift: ± 0.015°C/°C (from -10 °C to +55 °C)
Rj compensation error: internal : ±0.15°C (±0.3°F)
Advanced flexibility and high performance has been achieved using a 32 bit remote : ±0.3°C (±0.6°F)
microprocessor and a fast A/D conversion technology. The calibrator’s internal Common mode rejection: >140 dB at ac operation
memory stores all data for normalized IEC, DIN and JIS thermoelectric sensors Normal mode rejection: >70 dB at 50 or 60 Hz
Temperature stability: MicroCal 200/200+: for temperature exceeding the band +18°C to +28°C (from +64 °C to +84 °F)
for both IPTS68 and ITS90 International Temperature Scale
full scale: ± 8 ppm/°C
zero: ± 0.2 mV /°C
MicroCal 2000+: for temperature exceeding the band +21°C to +25°C (from +70 °C to +77 °F)
The microprocessor performs automatic Table top
=
full scale: ± 3 ppm/°C
linearization and cold junction The case is equipped with 2 pivot feet to zero: ± 0.2 mV /°C
compensation to assure high accuracy. It is change the vertical viewing angle when Output impedance (emf output): < 0.5 Ω with 0.5 mA maximum current
possible to set the calibrator to execute using the instrument on the top of a table.
Input impedance (mV, V and Tc ranges): >10 MΩ
menu-driven calibration procedures for Input impedance (mA ranges): <140 Ω @ 1 mA
your instruments in field work. Quality System Source resistance effects: ±1 mV error for 1000 Ω source resistance
Research, development, production, Rtd and Ω simulation excitation current: MicroCal 200/200+: from 0.01 to 5 mA
Both Channel 1 (Out) and Channel 2 (In) inspection and certification activities are MicroCal 2000+: from 0.01 to 2 mA
have the following operative mode defined by methods and procedures of the Rtd and Ω
measurement excitation current: ~ 0.4 mA @ 400Ω
capabilities: Eurotron Quality System inspected for ~ 0.04 mA @ 4000Ω
= millivolts compliance and certified ISO9001 by Rtd connection: 2, 3, and 4 wires
= volts GASTEC, a Dutch notified body. Rtd cable compensation: up to 100 Ω (each wire)
= milliamperes (active and passive loop) Rtd cable compensation error (Pt100): ±0.005°C/ Ω of total wire
= Ohms EMC Conformity Maximum load resistance: 1000 Ω @ 20 mA
= temperature with thermocouples Display: graphic LCD 240 x 64 dots display with LED backlight device
The instrument case, made in shock-
= temperature with resistance Measurement sampling time: 250 ms
resistant injection molded ABS + Output noise (at 300 Hz): <2 mVpp for ranges up to 200 mV f.s.,
thermometers polycarbonate has an internal metal <10 mVpp for ranges up to 2 V f.s.
= frequency
coating to fulfill the prevision of the directive <80 mVpp for ranges up to 20 V f.s.
= pulse
89/336/CEE Electromagnetic Digital interface: full bidirectional TTL (a RS232 adapter normal or insulated, is available as an option)
Compatibility. Channel 1_Channel 2 insulation: 50 Vdc
Remote auxiliary inputs are available for: Calculation functions: hold, max, min, offset, average
Relative humidity
= and temperature Selection °C/°F/K: through the configuration procedure
module In/Out data memory: 20 data with manual or automatic recall
= Two internal sensors and built-in hand Logging mode: 1500 input data items (optional memory card for memory extension)
pump pressure module Convert function: displays the electrical equivalent of the engineering unit
Scale factor: 5 different settings with zero and span programmable within -399999 and +999999
The ergonomic case design allows the user Square root: in combination with scale factor
Calibration: self learning technique with automatic procedure
the calibrator in three different ways:
Power supply: external charger and rechargeable Ni-Cd battery
= Portable
Self contained operation: 6h on Tc and mV input/output (backlight Off)
Two different leather cases, with a cover 3.5h with 20 mA simulation (backlight Off)
and shoulder strap, are available upon Recharging time: 5h at 90% and 6h at 99% with instrument switched off.
request for instrument alone or The battery charging is active only with the instrument switched off.
instrument, printer and accessories. Line operation: 100 - 120 - 230 Vac through the external battery charger
These are extremely useful for practical Line transformer insulation: 2500 Vac
use since they leave one hand free for Firmware release identification: release code on the display
instrument tuning. Operating environment temperature range: from -10 °C to +55 °C (from 14 °C to 130 °F)
= Panel mounting Storage temperature range: from -30 °C to +60 °C (from -22 °C to 140 °F)
It requires a panel cutout of 9.7” x 3.5”. Case: Injection moulded ABS with internal metal coating
The instrument bezel flange butts against Dimensions: 264 x 96 x 172 mm (10.4”x3.8”x6.8”) DIN size
the front of the mounting plate; two Weights: nett: 8.8 lb. gross: 12.1 lb.
lateral mounting brackets fit over the
instrument rear panel

6 11
MULTIFUNCTION MicroCal
Accurate and Reliable
CALIBRATORS
365 Days a Year
MicroCal 200/200+
Table IN - OUT Range MicroCal 2000+
High Accuracy
Sensor or Total range Accuracy Range Resolution
MicroCal 200
Accuracy
MicroCal 200+
Accuracy
MicroCal 2000+
Resolution Accuracy
Multifunctional Process Calibrators
parameter (% of rdg) (% of rdg) (% of rdg)
Tc J -210 ... 1200°C -190 ... 1200°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-350 ... 2200°F -310 ... 2192°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Tc K -270 ... 1370°C -160 ... 1260°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
454 ... 2500°F -256 ... 2300°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Tc T -270 ... 400°C -130 ... 400°C 0.01°C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-454 ... 760°F -238 ... 752°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Tc R -50 ... 1760°C 150 ... 1760°C 0.1 °C ±(0.02% + 0.2 °C) ± (0.01% + 0.2 °C) 0.01°C ±(0.005% + 0.2 °C)
-60 ... 3200°F 302 ... 3200°F 0.1 °F ±(0.02% + 0.36°F) ± (0.01% + 0.36 °F) 0.1 °F ±(0.005% + 0.36°F)
Tc S -50 ... 1760°C 170 ... 1760°C 0.1 °C ±(0.02% + 0.2 °C) ± (0.01% + 0.2 °C) 0.01°C ±(0.005% + 0.2 °C)
-60 ... 3200°F 338 ... 3200°F 0.1 °F ±(0.02% + 0.36°F) ± (0.01% + 0.36 °F) 0.1 °F ±(0.005% + 0.36°F)
MULTIFUNCTION
Tc B 50 ... 1820°C 920 ... 1820°C 0.1 °C ±(0.02% + 0.3 °C) ± (0.01% + 0.3 °C) 0.01°C ±(0.01% + 0.3 °C) CALIBRATORS
140 ... 3310°F 1688 ... 3308°F 0.1 °F ±(0.02% + 0.54°F) ± (0.01% + 0.54 °F) 0.1 °F ±(0.01% + 0.54°F)
Tc C 0 ... 2300°C 0 ... 2000°C 0.1 °C ±(0.02% + 0.2 °C) ± (0.01% + 0.2 °C) 0.1 °C ±(0.01% + 0.2 °C)
32 ... 4172°F 32 ... 3632°F 0.1 °F ±(0.02% + 0.36°F) ± (0.01% + 0.36 °F) 0.1 °F ±(0.01% + 0.36 °F)
Tc G 0 ... 2300°C 190 ... 2300°C 0.1 °C ±(0.02% + 0.3 °C) ± (0.01% + 0.3 °C) 0.1 °C ±(0.01% + 0.3 °C)
32 ... 4172°F 374 ... 4172°F 0.1 °F ±(0.02% + 0.54°F) ± (0.01% + 0.54 °F) 0.1 °F ±(0.01% + 0.54 °F)
Portable, Rugged and
Tc D 0 ... 2300°C 0 ... 2130°C 0.1 °C ±(0.02% + 0.3 °C) ± (0.01% + 0.3 °C) 0.1 °C ±(0.01% + 0.3 °C)
32 ... 4172°F 32 ... 3866°F 0.1 °F ±(0.02% + 0.54°F) ± (0.01% + 0.54 °F) 0.1 °F ±(0.01% + 0.54 °F) Easy to Use
Tc U -200 ... 400°C -160 ... 400°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-330 ... 760°F -256 ... 752°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F) Measures and Sources 14 Type
Tc L -200 ... 760°C -200 ... 760°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
of TCs and 10 RTDs
-330 ... 1400°F -328 ... 1400°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Tc N -270 ... 1300°C 0 ... 1300°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-450 ... 2380°F 32 ... 2372°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F) 24 Vdc Loop Power Supply
Tc E -270 ... 1000°C -200 ... 1000°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
-454 ... 1840°F -328 ... 1832°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
In-Line Digit Setting
Tc F 0 ... 1400°C 0 ... 1400°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C)
32 ... 2560°F 32 ... 2552°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)
Pannel Mounting Kit
Pt100 IEC -200 ... 850°C -200 ... 850°C 0.01°C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.01°C ±(0.005% + 0.05°C)
OIML, a 3926 -330 ... 1570°F -328 ... 1562°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.005% + 0.09°F)
MicroCal PM200
Pt100 -200 ... 650°C -200 ... 650°C 0.01°C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.01°C ±(0.005% + 0.05°C)
a 3902 -330 ... 1210°F -328 ... 1210°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.005% + 0.09°F) Optional Module
Pt100 JIS -200 ... 600°C -200 ... 600°C 0.01°C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.01°C ±(0.005% + 0.05°C) to Generate/Measure Pressure
SAMA -330 ... 1120°F -328 ... 1112°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.005% + 0.09°F)
Pt 200 -200 ... 850°C -200 ... 850°C 0.1 °C ±(0.02% + 0.15°C) ± (0.01% + 0.15 °C) 0.01°C ±(0.005% + 0.15°C)
CalpMan 2007
-330 ... 1570°F -328 ... 1562°F 0.1 °F ±(0.02% + 0.27°F) ± (0.01% + 0.27 °F) 0.1 °F ±(0.005% + 0.27°F)
Pt 500 -200 ... 850°C -200 ... 530°C 0.1 °C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C) Software for Automatic
-330 ... 1570°F -328 ... 986°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F) Calibration and Documenting in
Pt1000 IEC -200 ... 850°C -200 ... 850°C 0.01°C ±(0.02% + 0.1 °C) ± (0.01% + 0.1 °C) 0.01°C ±(0.005% + 0.1 °C) Compliance with ISO9001
OIML -330 ... 1570°F -328 ... 1562°F 0.1 °F ±(0.02% + 0.18°F) ± (0.01% + 0.18 °F) 0.1 °F ±(0.005% + 0.18°F)

CU10 -70 ... 150°C -70 ... 150°C 0.1 °C ±(0.02% + 0.4 °C) ± (0.01% + 0.4 °C) 0.1 °C ±(0.01% + 0.4 °C)
-100 ... 310°F -94 ... 302°F 0.1 °F ±(0.02% + 0.72°F) ± (0.01% + 0.72 °F) 0.1 °F ±(0.01% + 0.72°F)
CU100 -180 ... 150°C -180 ... 150°C 0.1 °C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.1 °C ±(0.01% + 0.05°C)
-300 ... 310°F -292 ... 302°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.01% + 0.09°F)
Ni100 -60 ... 180°C -60 ... 180°C 0.1 °C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.1 °C ±(0.01% + 0.05°C)
-80 ... 360°F -76 ... 356°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.01% + 0.09°F)
Ni120 0 ... 150°C 0 ... 150°C 0.1 °C ±(0.02% + 0.05°C) ± (0.01% + 0.05 °C) 0.1 °C ±(0.01% + 0.05°C)
32 ... 310°F 32 ... 302°F 0.1 °F ±(0.02% + 0.09°F) ± (0.01% + 0.09 °F) 0.1 °F ±(0.01% + 0.09°F)

mV (L) -20 ... +200mV 1 µ V ±(0.02% + 2 µ V) ± (0.01% + 2 µ V) 0.1 µV ±(0.0035% + 1 µ V)


mV (H) -0.2 ... +2 V 10 µ V ±(0.02% + 10 µ V) ± (0.01% +10 µ V) 1 µ V ±(0.005% + 10 µ V)
V 10 µ bt !
No dou Accurate
-2 ... +20 V 0.1 mV ±(0.02% + 0.08mV) ± (0.01% + 0.08mV) V ±(0.005% + 0.08mV)
mA (In) -5 ... +50 mA 0.1 µA ±(0.02% + 0.4 µ A) ± (0.01% + 0.4 µ A) 0.1 µA ±(0.005% + 0.4 µ A)
st
mA (Out) 0 ... +50 mA 0.1 µA ±(0.02% + 0.4 µ A) ± (0.01% + 0.4 µ A) 0.1 µA ±(0.005% + 0.4 µ A)
e of the Mo rdg)
Bulletin 08-44.3 E MicroCal 200_2000 - ed. 04-2007

Ω IN 0 ... 500 Ω 1 mΩ ±(0.02% + 12 mΩ ) ± (0.01% +12 mΩ ) 1 mΩ ±(0.005% + 12 mΩ ) O n 5%


(0.003 or
Specifications may change without notice

0 ... 5000 Ω 0.01 Ω ±(0.02% +120 mΩ ) ± (0.01% +120 mΩ ) 0.01 Ω ±(0.005% +120 mΩ )
b le C alibrat

OUT 0 ... 500 Ω 1 mΩ ±(0.02% + 20 mΩ ) ± (0.01% + 20 mΩ ) 1 mΩ ±(0.005% + 12 mΩ )
Por t a y!
5000 Ω 0.01 Ω 0.01 Ω le Toda
Availab
0 ... ±(0.02% +200 mΩ ) ± (0.01% +200 mΩ ) ±(0.005% +120 mΩ )
Frequency 1 ... 200 Hz 0.001 Hz ±(0.005% + 0.001 Hz) ±(0.005% + 0.001 Hz) 0.001 Hz ±(0.005% + 0.001 Hz)
1 ... 2000 Hz 0.01 ±(0.005% + 0.001 Hz) ±(0.005% + 0.001 Hz) 0.01 Hz ±(0.005% + 0.001 Hz)
1 ... 20000 Hz 0.1 Hz ±(0.005% + 0.001 Hz) ±(0.005% + 0.001 Hz) 0.1 Hz ±(0.005% + 0.001 Hz)
Pulse counter
6
0 ... 10 counts 1 count infinite infinite 1 count infinite
Instrum
Pulse (Out) 0 ... 6000 pulse/min 1 pulse/min 1 pulse / min 1 pulse / min 1 pulse / min 1 pulse / min ron e
ot

Q
nt

0 ... 36000 pulse/h 1 pulse/h 1 pulse / min 1 pulse / min 1 pulse / h 1 pulse / min
r
Eu

Note: • The MicroCal 2000+ relative accuracies shown above are stated for 360 • Typical 2 years relative accuracy can be estimated by multiplying the “% of
Ce

m
ste

• The MicroCal 200/200+ relative accuracies shown above are stated for 360 days and the operative conditions are from +70°F to +77°F reading" specifications by 1.4. fie
rti

y
days and the operative conditions are from +64°F to +82°F • Typical 90 days relative accuracy can be estimated by dividing the "% of • All input ranges: additional error ±1 digit. d Quality S
reading" specifications by 1.6. • Eurotron traceability chart and uncertainty can be supplied on request.

www.eurotron.com
12 All descriptions are related to a fully optioned instrument. See last page for the different configurations