SERVICE MANUAL

Affinity® Three
Birthing Bed
and
Affinity® Four
Birthing Bed
From Hill-Rom

Product No. P3700


For Parts or Technical Assistance MAN272 REV 3
USA (800) 445-3720 Canada (800) 267-2337
International: Contact your distributor.
Affinity® Three Birthing Bed
and
Affinity® Four Birthing Bed
Service Manual
Revisions

Revision Pages Affected Date


Original Issue October, 2000
A All October, 2002
3 All November 2006

MAN272 REV 3

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page i
Service Manual (MAN272 REV 3)
Revisions

© 2006 by Hill-Rom Services, Inc. ALL RIGHTS RESERVED.

No part of this text shall be reproduced or transmitted in any form or by any


means, electronic or mechanical, including photocopying, recording, or by any
information or retrieval system without written permission from Hill-Rom
Services, Inc. (Hill-Rom).

Third Edition

First Printing 2000

Printed in the USA

Affinity® is a registered trademark of Hill-Rom Services, Inc.

Allen™ is a trademark of Industrial Fasteners, Inc.

CSA® is a registered trademark of Canadian Standards Association.

Freedom Hill™ is a trademark of Hill-Rom Services, Inc.

Hill-Rom® is a registered trademark of Hill-Rom Services, Inc.

Liberty Hill™ is a trademark of Hill-Rom Services, Inc.

Loctite® is a registered trademark of Loctite Corporation.

Mechlok® is a registered trademark of P.L. Porter Company.

Oilite® is a registered trademark of Beemer Precision, Incorporated.

Plexiglas® is a registered trademark of Rohm and Haas Company.

SideCom® is a registered trademark of Hill-Rom Services, Inc.

Slo-Blo® is a registered trademark of Littelfuse, Inc.

Stow and Go™ is a trademark of Hill-Rom Services, Inc.

Torx® is a registered trademark of Textron, Inc.

Underwriters Laboratories Inc.® is a registered trademark of Underwriters


Laboratories Inc.

Page ii Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Revisions

The information contained in this manual is subject to change without notice.


Hill-Rom makes no commitment to update or keep current, the information
contained in this manual.

The only product warranty intended by Hill-Rom is the express, written


warranty accompanying the bill of sale to the original purchaser. Hill-Rom
makes no other warranty, express or implied, and in particular, makes no
warranty of merchantability or fitness for a particular purpose.

Additional copies of this manual can be obtained from Hill-Rom.

To order additional copies of this manual (MAN272), refer to the back cover
for contact information. For countries not listed on the back cover, contact
your distributor.

NOTE:
The back cover is a comprehensive list of Technical Support contact
information for Hill-Rom. The product discussed in this manual may not be
available in all of the countries listed.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page iii
Service Manual (MAN272 REV 3)
Revisions

NOTES:

Page iv Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Table of Contents

Chapter 1: Introduction
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Document Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Operating Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Head Section Inclination. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Foot Section Inclination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
High Low Sleeping Surface (Hilow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Battery Back-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Auxiliary Outlet (no longer available) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Automatic Tilt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Headboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Docking/Wall Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Lockout Control Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Night Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Foot Support Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Labor Grips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Siderails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Trend-Like Position Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
CPR Release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
The Integrated Air Support System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Mobility/Braking and Steer System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10
Siderails, Restraints, and Patient Monitoring . . . . . . . . . . . . . . . . . . . . . . 1 - 11
Dust Covers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 12

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page v
Service Manual (MAN272 REV 3)
Table of Contents

SideCom® Communication System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 12


Air Supply System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 13
Bed Positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 13
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 16
Physical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 16
Electrical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 17
Automatic Night Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 18
Siderail Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 19
P.C. Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 19
Regulations, Standards, and Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 20
Model Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 20
Safety Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 21
Bed Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 21
Siderails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 21
Brake and Steer Pedals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 21
Fluids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 22
Water Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 22
Lockout Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 22
CPR Release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 22
Product Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 29
Warning and Caution Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 32
Chapter 2: Troubleshooting Procedures
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Initial Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Quick Reference Problem/Solution Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Function Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Hilow System Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Trend-Like Position Functional Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Head System Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 7
Foot System Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 8

Page vi Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Table of Contents

Air System Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 9


Other Bed Functional Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 9
Final Actions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Hilow Drive Operation Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Hilow Up Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Hilow Down Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 14
Head Drive Operation Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 16
Head Up Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 17
Head Down Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 19
Foot Drive Operation Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 21
Foot Up Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 22
Foot Down Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 24
Hilow Up Malfunction (When The Foot Section Is Lowered) . . . . . . . . . . . . . . . 2 - 26
Brake Caster Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 27
Bed Air Surface Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 28
Night Light Does Not Illuminate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 34
Siderail Mechanism Does Not Hold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 35
Trend-Like Position Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 36
Battery Back-up Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 37
CPR Release Malfunction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 39
Chapter 3: Theory of Operation
Electrical System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
Chapter 4: Removal, Replacement, and Adjustment Procedures
Tool and Supply Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Top Motor Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
Bottom Motor Cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 6

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page vii
Service Manual (MAN272 REV 3)
Table of Contents

Hilow Drive Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 7


Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 7
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 10
Hilow Limit Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 11
Checking the Hilow Limit Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 11
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 12
Head Drive Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 17
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 17
Foot Drive Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 18
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 18
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 20
Foot Limit Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 21
Checking the Foot Limit Switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 21
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 22
CPR Release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 25
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 25
Trend-Like Position Sensing Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 27
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 27
Trend-Like Position Gas Spring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 29
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 29
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 30
Head Section Gas Spring and Damper (CPR Assist) . . . . . . . . . . . . . . . . . . . . . . 4 - 31
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 31
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 32
Foot Support Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 34
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 34
Night Light and Cable Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 36

Page viii Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Table of Contents

Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 36
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 36
Mattress Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 37
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 37
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 37
Labor Grips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 38
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 38
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 39
Foot Section (Lift Off) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 40
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 40
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 40
Foot Section (Slide Off) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 42
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 42
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 43
Line Voltage Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 44
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 44
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 45
Battery Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 46
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 46
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 46
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 47
Battery Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 47
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 47
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 48
Battery Charge Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 48
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 48
Bed Surface Air System Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 50
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 50
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 52
Air System Manifold Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 53
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 53

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page ix
Service Manual (MAN272 REV 3)
Table of Contents

Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 53
Air System Control P.C. Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 54
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 54
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 55
Logic Control P.C. Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 56
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 56
Modifying the Lockout Switch Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 58
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 60
Transformer Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 61
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 61
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 62
Auxiliary Transformer Assembly (Affinity® Three Bed) . . . . . . . . . . . . . . . . . . 4 - 63
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 63
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 64
Lockout Switch Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 65
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 65
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 66
Siderail Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 67
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 67
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 68
Patient Siderail (Inboard) P.C. Switch Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 69
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 69
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 70
Caregiver Siderail (Outboard) P.C. Switch Board . . . . . . . . . . . . . . . . . . . . . . . . 4 - 71
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 71
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 72
Central Braking System—P3700A Model Only . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 73
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 73
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 76
Central Braking System—P3700B Model Only. . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 77
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 77

Page x Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Table of Contents

Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 79
Caster Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 80
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 80
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 81
Brake and Steer Caster Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 81
Brake Caster Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 82
Stow and Go™ Foot Section Wire Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 83
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 83
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 85
Chapter 5: Parts List
Service Parts Ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Exchange Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
In-Warranty Exchanges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
Out-of-Warranty Exchanges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 5
Recommended Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 7
Caster Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 9
Base Module—P/N63951S (Sheet 1 of 2) (Affinity® Three Bed). . . . . . . . . . . . 5 - 10
Base Module—P/N63951S (Sheet 2 of 2) (Affinity® Three Bed). . . . . . . . . . . . 5 - 14
Main Frame Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 16
Siderail Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 20
Electronics Module (Sheet 1 of 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 24
Electronics Module (Sheet 2 of 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 27
Electronics Module (Sheet 3 of 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 28
Electrical Cable Routing (Sheet 1 of 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 30
Electrical Cable Routing (Sheet 2 of 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 32
Sleep Deck Module (Sheet 1 of 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 34
Sleep Deck Module (Sheet 2 of 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 38
Sleep Deck Module (Sheet 3 of 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 40
Label Module (Sheet 1 of 4)—P3700A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 44
Label Module (Sheet 2 of 4)—P3700A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 46

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page xi
Service Manual (MAN272 REV 3)
Table of Contents

Label Module (Sheet 3 of 4)—P3700A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 48


Label Module (Sheet 4 of 4)—P3700A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 49
Foot Support Module—P3700A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 50
Calf Support Module—P3700A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 54
Air Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 56
Headboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 58
Caster Module—P3700B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 60
Base Module—P3700B (Sheet 1 of 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 62
Base Module—P3700B (Sheet 2 of 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 66
Sleep Deck—P3700B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 68
Foot Support Module—P3700B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 70
Calf Support Module—P3700B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 73
Label (Sheet 1 of 4)—P3700B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 76
Label (Sheet 2 of 4)—P3700B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 78
Label (Sheet 3 of 4)—P3700B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 80
Label (Sheet 4 of 4)—P3700B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 82
Chapter 6: General Procedures
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Steam Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Cleaning Hard to Clean Spots. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Disinfecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Draping the Mattress. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Caring for the Wood Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Servicing the Bed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 4
Component Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 5
P.C. Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 5
Lubrication Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 6
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 7
Inspecting the Pivot Point Fasteners . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 7
Inspecting the Foot Supports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 8
Preventive Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 10

Page xii Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Table of Contents

Preventive Maintenance Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 13


Using Loctite® Adhesive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 15
Purple Loctite® Adhesive #222 (P/N SA4842) . . . . . . . . . . . . . . . . . . . . . . . 6 - 16
Blue Loctite® Adhesive #242 (P/N SA3618) . . . . . . . . . . . . . . . . . . . . . . . . 6 - 16
Red Loctite® Adhesive #262 (P/N SA4841) . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 16
Red Loctite® Adhesive #271 (P/N SA4840) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 16
Chapter 7: Accessories
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Labor Bar (P3613) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 4
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 4
Oxygen Tank Holder (P276). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
Telescoping Calf Supports (P35745A) or
Telescoping Full Leg Supports (P7634C) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 6
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 6
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 7
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 7

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page xiii
Service Manual (MAN272 REV 3)
Table of Contents

NOTES:

Page xiv Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
1
Chapter 1
Introduction

Purpose
This manual provides requirements for the Affinity® Three Birthing Bed and
Affinity® Four Birthing Bed normal operation and maintenance. It also
includes parts lists (in chapter 5) for ordering replacement components.

Audience
This manual is intended for use by facility-authorized personnel only. Failure
to observe this restriction can result in severe injury to people and serious
damage to equipment.

Reference Documents
For more information (such as operating instructions, features, and product symbols),
refer to the “Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User
Manual (USR025).”

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 1
Service Manual (MAN272 REV 3)
Document Symbols
Chapter 1: Introduction

Document Symbols
This manual contains different typefaces and symbols to make the content
easier to read and understand:
• Standard text—used for regular data.
• Boldface text—emphasizes a word or phrase.
• NOTE:—sets apart special data or important instruction clarification.
• WARNING or CAUTION

– A WARNING identifies situations or actions that may have an


effect on patient or user safety. To ignore a warning could cause
patient or user injury.
– A CAUTION identifies special procedures or precautions that
persons must obey to help prevent equipment damage.
• CAUGHT HAZARD WARNING

• CHEMICAL HAZARD WARNING

• ELECTRICAL SHOCK HAZARD WARNING

Page 1 - 2 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction
1
Introduction
The Affinity® Three Birthing Bed and Affinity® Four Birthing Bed are
intended to be used as a birthing bed within the acute care Labor and Delivery
Unit. The Affinity® Three Birthing Bed and Affinity® Four Birthing Bed are a
three-motor birthing type bed with multiple articulation features (see “Bed
Position” on page 1-21). A variety of options are available including nurse call,
air system, night light, auxiliary outlets, and siderail controls. See “Features”
on page 1-3 for additional information concerning some of the Affinity® Three
Birthing Bed and Affinity® Four Birthing Bed features.

NOTE:
Any reference in this manual to a side of the bed is from the point of view of a
patient lying in the bed on her back. This information appears in the manual
wherever it is pertinent.

Operating Precautions
Before operating the bed, be sure that you have read and understand in detail
the contents of this manual. It is important that you read and strictly adhere to
the safety precautions (see “Safety Tips” on page 1-21).

For additional operating precautions for the bed and its accessories, refer to the
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual
(USR025).

Features
The bed incorporates three independent motors to raise and lower the
following bed features:
• Bed height
• Foot section
• Head section (The seat section automatically tilts with head section
operation.)

These functions are governed by P.C. board logic and controlled by the use of
the siderail bed controls or an optional hand-held, six-function pendant.

Head Section Inclination

The bed is mechanized so that the patient or attendant may elevate the head
section by pressing a switch. The head up/head down switches are momentary

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 3
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

contact type switches and are pictorially labeled to indicate their function. The
switches are fixed in relation to the head section and within easy access of both
the occupant and attendant, regardless of the degree of inclination.

Foot Section Inclination

The bed is mechanized so that the patient or attendant may elevate the foot
section by pressing a switch. The foot up/foot down switches are momentary
contact type switches and are pictorially labeled to indicate their function. The
switches are fixed in relation to the head section and within easy access of both
the occupant and attendant, regardless of the degree of inclination.

High Low Sleeping Surface (Hilow)

The bed is mechanized so that an attendant can raise or lower the sleeping
surface to facilitate examination or bed ingress or egress. The hilow up/down
switches are momentary contact type switches and are pictorially labeled to
indicate their function. The switches are fixed in relation to the head section
and face toward the attendant position only. The hilow up/down switches are
not accessible to the occupant.

Battery Back-up

The bed is designed with battery back-up as a standard feature. The battery
allows the hilow, foot, and head motors to be activated from the siderail
controls without power being supplied to the bed. In addition, the battery
powers the nurse call function, but it does not power any other bed functions.

Battery Back-up LED Indicators


• ON = Battery status is operational or when bed is plugged into the
appropriate power source.
• FLASHING = Battery needs charging.
• OFF = Battery is discharged below the level required to run the motors.

CAUTION:
Remove the battery if the bed will not be in service for extended periods
of time. Failure to do so could result in damage to the life of the battery
or damage to the bed. Contact the appropriate maintenance personnel.

If the battery has been completely discharged, it may take up to 36 hours to


recharge to operational status.

Page 1 - 4 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

To ensure the battery is always charged, plug the bed into the appropriate
1
power source whenever possible. If the battery is discharged or unplugged, the
bed is still functional when plugged into the appropriate AC receptacle.

Disp osal

Power comes from a lead acid battery, which needs to be


disposed of properly according to your local regulations.

For assistance in disposing of the battery, contact your Pb


maintenance technician.

Auxiliary Outlet (no longer available)

The two auxiliary outlets provide electrical power for accessories.

Automatic Tilt

As the head section is raised, the seat gradually tilts up from 0 to 15 degrees.
As the head section is lowered, the seat gradually returns to a flat position.

Headboard

The headboard is mounted to a metal frame. The headboard has built-in


handles, which aid in steering control and increase mobility. The headboard is
removable.

Docking/Wall Protection

Standard roller bumpers, located at the head end of the bed, protect walls and
headwall systems from damage.

CAUTION:
Do not position the head of the bed against a wall or beneath any
fixtures. This is to prevent damage due to the arcing motion of the bed
while the bed is being raised and lowered.

Lockout Control Switch

The lockout control switch is located at the head end of the bed on the frame
(see figure 1-1 on page 1-6). This control is used to deactivate the patient
(inboard) and caregiver (outboard) siderail controls.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 5
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

Figure 1-1. Lockout Switches

Night Light

The night light is located under the bed. A photocell control automatically
turns the light on when the room darkens and turns it off when it gets brighter.

Foot Support Assemblies

The bed is equipped with built-in foot support assemblies designed to remain
on the unit during normal usage. The assemblies can be simultaneously
positioned up or down, through use of the motor-powered, foot section yoke.
This foot section yoke controls the overall height up and down. Each foot
support can be independently positioned by using the mechanical release latch
at the end of the foot supports.

The following positional adjustments can be made:


• To position the patient’s legs, squeeze the release lever and rotate the foot
support. The foot support rotates upward and outward from 0° to 85°.
• To return to the storage position, squeeze the release latch again, and return
the supports to the desired position.

Page 1 - 6 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

Each foot support is attached to the bed and can only be removed by the
1
appropriate personnel.

Labor Grips

To raise the labor grip, grasp the grip, and rotate it up from under the bed until
it clicks. This signals that it has locked into position. Ensure that the labor grip
is properly locked by giving it a tug.

To lower the labor grip, pull the release handle, and lower the grip under the
bed.

Siderails

To raise the siderails (A), grasp the top center of the siderail, and pull out and
up from beneath the bed (see figure 1-2 on page 1-7). Rotate the siderail
upward until it clicks. The clicking sound signals that the siderail has locked
into position. Ensure that the siderail is up and properly locked into position by
giving it a gentle tug.

Figure 1-2. Siderails

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 7
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

To lower/store the siderail, pull the release handle (B), and allow the siderail to
lower into the down or storage position.

The siderail stores in a position that does not extend outward from the width of
the bed. This feature aids in preventing transfer gaps when two beds are placed
side by side.

Trend-Like Position Operation

Trend-Like positioning to 8° is achieved mechanically from any bed height


position by depressing one of two Trend-Like position handles. Trend-Like
position handles are located near the head end of the bed under the head
section. Depressing either one of the handles releases gas-assisted springs
(cylinders) and allows for easy positioning.

To place the sleep surface in Trend-Like position, push down on the handle,
and guide the bed to the desired angle. If the desired Trend-Like position
cannot be achieved because the bed is too low, the bed’s hilow function will
automatically raise the bed until the proper position is reached.

To return the bed to the level position, pull up on the handle. Mechanical limits
stop the travel when the sleep surface is level. Quickly release it when the bed
is in the desired position.

Depending on the position of the head section and the weight of the patient, the
amount of activation force required to place the bed in or out of the Trend-Like
position varies. For example, with a patient occupying the bed and the head
section raised to 45 , the bed will more easily go out of Trend-Like position,
O

versus going into the Trend-Like position. With a patient occupying the bed
and the head section in the flat position, the bed will more easily go into the
Trend-Like position, versus going out of the Trend-Like position.

NOTE:
The bed must be in the flat (level) position and out of the Trend-Like position,
for the hilow to work properly.

CPR Release

The head section has a releasing mechanism for rapid emergency lowering.
The release mechanism requires intentional action to release it. The bed
mechanism recognizes the de-coupling and automatically lowers the head and
seat section; then it recouples the head and seat mechanism. This ensures
proper head section lifting after the emergency.

Page 1 - 8 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

A CPR release handle is located on either side of the bed, beneath the head
1
section. To operate, pull the CPR handle and hold. With a patient in the bed,
the head section will lower to a flat position within approximately 7 seconds.

The Integrated Air Support System

The integrated air support system allows the patient or attendant to select the
desired level of firmness in the lumbar and seat sections of the mattress. Both
the patient and attendant can control these functions from the siderail. The
lumbar bladder, when fully inflated, appears as a large round bump in the
mattress. The seat bladder, when fully inflated, does not have a specific
appearance. Proper operation of the seat bladder is best determined by lying on
the bed and activating the seat inflate function.

Seat Section Mattress

The seat section mattress has auto-inflate and auto-deflate functions. The
appropriate button need be pressed only once to achieve full inflation or
deflation of the mattress.

Seat Section Mattress Auto-Inflate or Deflate Control

To automatically inflate or deflate the seat section, press the caregiver seat
inflate button (plus sign) or caregiver seat deflate button (minus sign) once.
The seat section will inflate or deflate for approximately 20 to 25 seconds and
then automatically stop.

If a complete inflation or deflation cycle is not necessary, press the


caregiver seat inflate or caregiver seat deflate button a second time to stop the
inflation or deflation.

Back Section Lumbar Mattress

The back section lumbar mattress does not have auto-inflate or auto-deflate
functions. The appropriate button must be pressed and held until the desired
inflation is attained.

Back Section Lumbar Mattress Control

To inflate the back section, press and hold the back inflate button (plus sign)
until the desired firmness is attained.

To deflate the back section, press and hold the back deflate button (minus sign)
until the desired softness is attained.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 9
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

Mobility/Braking and Steer System

The bed is mounted on four precision-bearing swivel casters. Each caster has a
molded polyurethane wheel with a diameter of 6" (15 cm) or 8" (20 cm)
(Affinity® Three Birthing Bed only). Wheels have precision-bearing axles and
swivels for high mobility. They are easily removed for cleaning or replacing.

The caster braking system is activated by foot pedals located on either side of
the bed to enable one foot to lock two casters (wheels and swivels). Activation
of the steer system locks the swivel on one caster (steer lock on the foot end) to
enable easy steering of the bed, on the Affinity® Four Birthing Bed, all four
casters lock. The steer system is operated by foot pedals located on either side
of the bed. The foot pedals are located in convenient, accessible locations on
both sides of bed and are clearly identified. The brake pedal is designated by an
orange dot, and the steer pedal is designated by a green dot.

Brake and Steer Pedals

Placing the brake/steer pedal in steer locks the left side, foot end caster into a
position that is parallel to the bed. When transporting the bed, the caster should
trail the direction of travel (see figure 1-3 on page 1-10). This trailing position
is when the center of the caster’s wheel axle is behind the caster stem during
transport.

Figure 1-3. Steer Caster Position

Page 1 - 10 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction
1
WARNING:
A bed that is transported with the steer caster locked in an incorrect
position can drift from side to side during transport. Personal injury or
equipment damage could occur.

If the caster is leading the direction of travel, reposition it. Do this by placing
the brake/steer pedal in neutral, swiveling the caster 180 degrees, and placing
the brake/steer pedal back in the steer position.

Siderails, Restraints, and Patient Monitoring

The bed is equipped with head section siderails as standard equipment. They
can be placed in the up or down position. In the down position, the rails store
under the frame of the bed providing a zero transfer gap for safe surface-to­
surface transfer and better patient access. The bed is configured so that the
following functions may be included as part of the siderails:
• Patient and attendant head and foot actuators
• Hilow actuators (attendant only)
• Nurse call
• Entertainment center actuators (lighting, TV, UTV, and radio)
• Mattress controls

The siderails should always be in a full upright position and latched when a
patient is unattended. When raising the siderails, an audible click indicates that
the siderail is completely raised and locked in place.

We recognize that certain healthcare situations may indicate the need for
specialized siderail configurations. In response to this need we offer, upon
request, several siderail accessories.

Siderails are intended to be a reminder, not a patient restraining device.


Appropriate medical personnel should determine the level of restraint
necessary to ensure that a patient remains safely in bed. Consult the restraint
manufacturer’s instruction for use to verify the correct application of each
restraining device.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 11
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

Whenever medicated or confused patients are involved, Hill-Rom recommends


the following minimum actions:
• Develop guidelines for all such patients that indicate:
– Which patients should be restrained and the appropriate restraint to
utilize.
– The proper method to monitor a patient, whether restrained or not,
including time interval, visual check of restraint, etc.
• Develop training programs for all caregivers concerning the proper use and
application of restraints.
• Maintain the bed at its lowest position whenever a caregiver is not in the
room.

Dust Covers

All bed mechanisms are enclosed for protection from damage, aid in infection
control, and patient room housekeeping efficiencies.

SideCom® Communication System

Every bed with siderail controls is pre-wired to accommodate the SideCom®


Communication System. The SideCom® Communication System enables
integration of the backlit nurse call, entertainment (radio and TV), and lighting
in the siderails. The optional Universal TV is also available.

Nurse Call (Optional)

On beds equipped with the optional nurse call feature, use the nurse call button
to place a call to the nurse call system. Above the nurse call button, the call
indicator light illuminates on the patient’s siderail controls, indicating a call
has been made.

Siderail Communication Watchdog System

Beds equipped with the SideCom® Communication System include a feature


called the siderail communication watchdog system. It sends a nurse call if for
any reason the siderail controls do not communicate properly with the bed.

Page 1 - 12 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Introduction

Air Supply System


Chapter 1: Introduction
1
Major components in the air supply system include:
• Air compressor
• Manifold solenoid block
• Mattress drive circuit board
• Air mattress

Bed Positions
The bed has three sections: head, seat, and foot.

The bed positions are shown as follows:


• Trend-Like Position (see figure 1-4 on page 1-13)
• Head Position (see figure 1-5 on page 1-14)
• Foot Position (see figure 1-6 on page 1-14)
• Hilow Positions (see figure 1-7 on page 1-15)

Figure 1-4. Trend-Like Position

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 13
Service Manual (MAN272 REV 3)
Introduction
Chapter 1: Introduction

Figure 1-5. Head Position

Figure 1-6. Foot Position

Page 1 - 14 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Introduction

Figure 1-7. Hilow Positions


Chapter 1: Introduction
1

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 15
Service Manual (MAN272 REV 3)
Specifications
Chapter 1: Introduction

Specifications

Physical Description
For bed specifications, see table 1-1 on page 1-16.

Table 1-1. Specifications

Feature Dimension
Total Length 90" (229 cm)
Length from roller bumpers to break in seat 61 7/8" (157 cm)
section
Length from break in the seat section to 28 1/8" (71 cm)
end of bed
Maximum Width (siderails stored) 36" (91 cm)
Maximum Width (siderails up) 39" (99 cm)
Maximum Headboard Height 52" (132 cm)
Maximum Siderail Height (without mat­ 14¾" (37.47 cm)
tress)
Minimum Under-Bed Clearance 5" (13 cm), 3" 7.6 cm) for Stow
and Go™ Beds
Wheel Base 50" x 29" (127 cm x 74 cm)
Mattress Width 35" (89 cm)
Mattress Length 78" (198 cm)
Mattress Thickness (head/seat) 4" (10 cm)
Mattress Thickness (foot) 3" (8 cm)
Detachable Power Cord, IEC 320/interface US, 84" (213 cm)/international,
98" (249 cm)
Caster Size 6" (15 cm) or 8" (20 cm) (Affin­
ity® Three Birthing Bed only)
Total Weight (maximum) 480 lb (218 kg)
Head Section Inclination (maximum) 65°
Seat Section Inclination (maximum) 15°
Bed Height Range 18" (46 cm) to 34" (86 cm)
Bed Height Range (with mattress) 22" (56 cm) to 38" (97 cm)
Trend-Like position (maximum) 8°
(Continued on next page.)

Page 1 - 16 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Specifications

Table 1-1. Specifications (cont’d.)


Chapter 1: Introduction
1
Feature Dimension
Bed Lift capacity (maximum safe working 500 lb (227 kg)
load)
Foot Section Weight capacity (maximum) 400 lb (181 kg)
Head Section Lift capacity (maximum) 200 lb (91 kg)
Maximum Height of Seat Section (in 40" (102 cm)
Trend-Like position)

Table 1-2. Environmental Conditions for Transport and Storage

Condition Range
Temperature -40°F (-40°C) to 158°F (70°C)
Relative humidity 95% non-condensing
Pressure 15 "Hg (50 kPa) to
31 "Hg (106 kPa)

Table 1-3. Environmental Conditions for Use

Condition Range
Temperature 59°F (15°C) to 104°F (40°C)
ambient temperature
Relative humidity range 10% to 85% non-condensing

Electrical Description

Table 1-4. Mains Power Requirements

Condition Range
Rated voltage 120V AC/230V AC
Power/input 4 A/2 A
Frequency 50/60 Hz

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 17
Service Manual (MAN272 REV 3)
Specifications
Chapter 1: Introduction

Table 1-5. Fuse Specifications


Condition Range
Air system fuse (air system optional) 2 A, 250 V~, 5 x 20 mm, UL
198G Fast Acting
Battery fuse 10 A, 32 V~, ATO
Mains fuses (120 V bed model) 4 A, 125 V~, 5 x 20 mm, UL
198G Slo-Blo® or equivalent
a

Mains fuses (230 V bed model) 2 A, 250 V~, 5 x 20 mm,


IEC127 Sheet III, Time Delay
a. Slo-Blo® is a registered trademark of Littelfuse, Inc.

Table 1-6. Electrical Leakage

Condition Range
Ground resistance < 0.20Ω
Leakage current < 100μA for 120V models
< 150μA for 230V models

The electrical power system is insulated from the metal parts of the bed. No
additional electrical components, such as isolation transformers, are required
to make this bed meet applicable electrical codes. All patient support surfaces
are fully grounded through a three-wire power cord having a hospital-grade,
3-prong, grounded plug.

Each bed is factory-tested for complete operation with and without a load.
Each bed is tested for insulation integrity and micro-current leakage.

All electrical components of this bed have been approved by Underwriters


Laboratories Inc.® (UL) and Canadian Standards Association (CSA® ) for
1 2

this application.

Automatic Night Light

The bed is equipped with a night light that automatically activates as the
ambient light decreases. The light illuminates the foot fall area under the bed.

1. Underwriters Laboratories Inc.® is a registered trademark of Underwriters Laboratories Inc.


2. CSA® is a registered trademark of Canadian Standards Association

Page 1 - 18 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Specifications

Siderail Controls
Chapter 1: Introduction
1
Patients and attendants operate the various functions of this bed by means of
finger-touch switches located in a convenient fixed position on the siderails at
both sides of the bed. These controls move with the head section for easy
access and visual contact regardless of the head section elevation. The control
switches are electronically interlocked to prevent electrical damage to the
motor caused by contradictory direction signaling.

Head

Head section operating controls, identified by graphic symbols, are visible and
accessible to both the patient and the attendant. The up/down travel of the head
section is controlled by separate momentary type switches.

Foot

Foot section operating controls, identified by graphic symbols, are visible and
accessible to both the patient and the attendant. The up/down travel of the foot
section is controlled by separate momentary type switches.

On the original version of the bed, the foot section travel pauses when it
reaches the flat or bed level position, indicating the bed is in the level position.
Travel then proceeds to the desired position if the foot section switch remains
actuated. The new version of the bed does not have this pause feature.

Hilow

The hilow controls, identified by graphic symbols, are visible and accessible
only to the nurse/attendant. The up/down travel of the sleep deck is controlled
by separate momentary type switches.

Lockout

This bed has the capability of locking out the inboard and outboard hilow,
head, and foot functions if so prescribed by the attending physician. The
lockout switch is located at the head end of the bed’s main frame. The lockout
function interrupts the patient/attendant siderail controls until the lockout
switch is turned off.

P.C. Boards

All P.C. board functions are tested at the completion of circuit board assembly
and at final bed assembly. A line transient filter is included to protect the
boards from excessive line surges.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 19
Service Manual (MAN272 REV 3)
Model Identification
Chapter 1: Introduction

Regulations, Standards, and Codes


The bed is designed and manufactured according to equipment classifications
and the standards in table 1-7 on page 1-20.

Table 1-7. Regulations, Standards, and Codes


Technical and Quality Assurance Stan­ EN 60601-1 and amendments
dards UL 2601-1
CSA C22.2 No. 601.1
IEC 60601-2-38
IEC 60601-1-2
EN ISO 9001 and EN 46001
Equipment Classifications per EN 60601-1 Class I equipment, internally powered
equipment
Degree of Protection Against Electric Type B
Shock of the Applied Part
Degree of Protection Against Ingress of IPX2
Water
Degree of Protection Against the Presence Ordinary equipment, not for use in a
of Flammable Anaesthetic Mixtures flammable atmosphere
Mode of Operation Continuous operation with intermittent
loading, 3 minutes ON/30 minutes OFF
Classification according to Directive Class I
93/42/EEC

Model Identification
For bed model identification, see table 1-8 on page 1-20.

Table 1-8. Model Identification

Model Number Description


P3700A Affinity® Three Birthing Bed
P3700B Affinity® Four Birthing Bed

Page 1 - 20 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Safety Tips
Chapter 1: Introduction
1
Safety Tips
Train and educate your staff on the hazards associated with electric beds. Do
not allow personnel to have their entire body below the sleep surface and
within the confines of the bed. Unplug the bed from its power source and
engage the lockout control prior to cleaning or servicing it. If service personnel
need to get under the bed, they must block up the hilow portion as an added
precaution. Ensure that the foot section is properly mounted to the yoke.

We urge you to incorporate these safety tips into your procedures for the safety
of both patients and staff.

Bed Position
To reduce the severity of falls by patients, always leave the bed in the low
position when the patient is unattended.

Siderails
Leave the siderails fully up and locked when the patient is left unattended.
When raising the siderails, be sure that you hear the click that signals the up
and locked condition. Give the siderails a tug to check that they are firmly in
position.

Siderails are intended as a reminder, not a restraint device. Appropriate


medical personnel should determine the level of restraint necessary to ensure a
patient remains safely in bed.

Brake and Steer Pedals


Always keep the casters in the brake position when the bed is occupied.
Patients often use the bed for support when getting in or out of bed, and serious
injuries can result if the bed moves. After brakes are set, rock the bed gently to
ensure that they are locked.

Put the casters in the steer mode when moving the bed. This will make the bed
easier to position or transport.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 21
Service Manual (MAN272 REV 3)
Safety Tips
Chapter 1: Introduction

Fluids
When massive spills occur in the area of the P.C. boards, motors, or
transformers, immediately:

1. Unplug the bed from its power source.

2. Take care of the patient.

3. Clean the fluid from the bed.

4. Have maintenance check out the bed completely. Fluids can short out
controls, making the bed inoperable or cause the bed to operate erratically.
Component failure caused by fluids can even cause the bed to operate
without warning, causing injury.

5. Do not place the bed back into service until it is unquestionably dry and
tested safe to operate.

Water Mattress
The excessive weight associated with water mattresses puts an undue stress on
the motor drives. In most cases, the patient’s weight plus the water mattress
weight exceeds the recommended bed capacity. Even more important is the
fact that water mattresses are subject to rupture, which would allow large
amounts of water to come into contact with the electrical components of the
bed. We recommend not using water mattresses.

Lockout Switch
Whenever a patient or visitor should be restricted from operating the siderail
controls, activate the lockout switch located at the head end of the bed (on the
main frame). The lockout switch is for the convenience of the staff and the
safety of the patient. Use the lockout switch when appropriate.

CPR Release
Only healthcare professionals should use the emergency CPR release. The two
release handles are located under the head section of the bed near the seat
section.

To activate the CPR release, pull the handle away from the bed. Continue to
pull out on the handle until the head section is flat. When this is complete,
attend to the patient. The bed will automatically flatten the head and seat
sections and reset itself to be ready to use after the emergency.

Page 1 - 22 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Safety Tips
Chapter 1: Introduction
1
WARNING:
A bed that is transported with the steer caster locked in an incorrect
position can drift from side to side during transport. Personal injury or
equipment damage could occur.

WARNING:
Only facility-authorized maintenance personnel should troubleshoot the
bed. Troubleshooting by unauthorized personnel could result in
personal injury or equipment damage.

WARNING:
Unplug the bed from its power source and disconnect the battery
back-up before checking ohms/resistance measurements. Failure to
disconnect line voltage to the bed can damage the VOM and cause
equipment damage or personal injury.

WARNING:
Refer to your VOM owner’s manual for complete and detailed
information regarding the operation of your VOM. Failure to do so could
result in personal injury or equipment damage.

WARNING:
You must use the 2" x 4" pieces of lumber to support the bed. The bed
will fall during this procedure if not supported. Failure to do so could
result in personal injury or equipment damage.

WARNING:
Prop up the head section before removing the cotter pins and clevis
pins. Failure to do so could result in personal injury or equipment
damage.

WARNING:
You must use the 2" x 4" pieces of lumber to support the foot yoke. The
foot yoke will fall during this procedure if not supported. Failure to do so
could result in personal injury or equipment damage.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 23
Service Manual (MAN272 REV 3)
Safety Tips
Chapter 1: Introduction

WARNING:
The head end of the main frame is supported by the two Trend-Like
position gas springs. Any servicing will require that a support device be
placed just beneath the head end of the main frame. Failure to do so
could result in personal injury or equipment damage.

WARNING:
Do not attempt to remove the gas spring with the head section fully
raised and the spring compressed. Lower the head section until the
cylinder of the gas spring is free in the slide bracket of the main frame,
and support the head section securely before removing the cotter pins
and clevis pins. Failure to do so could result in personal injury or
equipment damage.

WARNING:
Failure to properly mount the foot section to the yoke latches could
result in equipment damage or personal injury.

WARNING:
Failure to properly mount the foot section to the yoke slide brackets
could result in equipment damage or personal injury.

WARNING:
Adhere to all electrical safety precautions when servicing the bed's
electrical system. Failure to do so could result in personal injury or
equipment damage.

WARNING:
Ensure that the bed is stable before removing the caster. Failure to do
so could result in personal injury or equipment damage.

WARNING:
Follow the product manufacturer’s instructions. Failure to do so could
result in personal injury or equipment damage.

WARNING:
Adhere to the Infection Control Policies and Procedures from Hill-Rom.

Page 1 - 24 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Safety Tips
Chapter 1: Introduction

Failure to do so could result in the spread of infection.


1
WARNING:
Powered bed mechanisms can cause serious injury. Operate the bed
only with persons clear of mechanisms. Failure to do so could result in
personal injury or equipment damage.

WARNING:
Unplug the bed from its power source and engage the lockout control
during routine maintenance or cleaning. Refer to the Affinity® Three
Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025)
and specific sections in this service manual for additional precautions.
Failure to do so could result in personal injury or equipment damage.

WARNING:
Ensure that all electrical/mechanical loads are removed prior to
maintenance/repair of the bed’s drive system or other mechanical
assemblies. Failure to do so could result in personal injury or equipment
damage.

WARNING:
Visually inspect the bushings annually. If wear is apparent, replace
them. Failure to do so could result in personal injury or equipment
damage.

WARNING:
Only facility-authorized personnel should perform preventive
maintenance on the bed. Preventive maintenance performed by
unauthorized personnel could result in personal injury or equipment
damage.

WARNING:
Inspect the pivot point fasteners semi-annually. Failure to do so could
result in personal injury or equipment damage.

WARNING:
Use primers with adequate ventilation. Avoid skin contact and

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 25
Service Manual (MAN272 REV 3)
Safety Tips
Chapter 1: Introduction

prolonged or repeated breathing of vapors. Do not allow primers to be


trapped under rings, watch bands, etc. Observe all directions on the
primer can. Failure to do so could result in personal injury.

WARNING:
Avoid excessive or repeated skin contact with the liquid. Repeated
contact with the liquid could result in personal injury.

WARNING:
The labor bar is intended to be used in the prescribed manner only.
Failure to use this product as outlined may result in personal injury or
equipment damage.

WARNING:
Insufficient tightening will allow the cradles to slip and lose their original
position. Personal injury could occur.

WARNING:
The supports are intended to be used in the prescribed manner only.
Failure to use this product as outlined may result in personal injury or
equipment damage.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

SHOCK HAZARD:
The voltage in the electrical system presents an electrical shock
hazard. Perform standard electrical service procedures before
attempting service within the P.C. board enclosure. Adhere to all
electrical safety precautions when servicing the bed’s electrical system.
Failure to do so could result in personal injury or equipment damage.

SHOCK HAZARD:
Do not expose the unit to excessive moisture which would allow liquid
to pool. Personal injury or equipment damage could occur.

Page 1 - 26 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Safety Tips
Chapter 1: Introduction
1
CAUTION:
Remove the battery if the bed will not be in service for extended periods
of time. Failure to do so could result in damage to the life of the battery,
or damage to the bed. Contact the appropriate maintenance personnel.

CAUTION:
Do not position the head of the bed against a wall or beneath any
fixtures. This is to prevent damage due to the arcing motion of the bed
while the bed is being raised and lowered.

CAUTION:
Ensure that the night light is not damaged when the bed is being
lowered. Failure to do so could result in equipment damage.

CAUTION:
Do not pull on the mattress material when unfastening the mattress
retaining snaps. Unfasten the mattress retaining snaps at the snap
location. Failure to do so could result in equipment damage.

CAUTION:
Ensure that the fuse housing is properly oriented for the voltage rating
(110-120V or 220-240V) on the bed that you are servicing. Failure to do
so could result in equipment damage.

CAUTION:
Do not cut or remove the cable ties that secure the battery leads to the
electronics pan. This ensures the proper connection of the batteries
during the replacement procedure. Possible equipment damage could
occur if the cable ties are removed.

CAUTION:
To prevent component damage, ensure that your hands are clean, and
only handle a P.C. board by its edges. Failure to do so could result in
equipment damage.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 27
Service Manual (MAN272 REV 3)
Safety Tips
Chapter 1: Introduction

CAUTION:
For shipping and storage, place the removed P.C. board in an antistatic
protective bag. Failure to do so could result in equipment damage.

CAUTION:
Support the siderail during the removal procedure. Failure to do so
could result in equipment damage.

CAUTION:
Ensure that the siderail does not drop when the pins are removed.
Failure to do so could result in damage to the wiring going to the
siderail.

CAUTION:
Do not use harsh cleansers or detergents such as scouring pads and
heavy-duty grease removers, or solvents such as toluene, xylene, and
acetone. Equipment damage could occur.

CAUTION:
Ensure that the metal platform is dry before placing the mattress back
onto the bed. Failure to do so could result in equipment damage.

CAUTION:
Mattress damage caused by improper draping and/or cleaning
procedures is not covered by warranty.

CAUTION:
Standard OB packs and paper drapes will not keep the sheets dry.

CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could result in component damage.

CAUTION:
Do not use silicone-based lubricants. Equipment damage could occur.

Page 1 - 28 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Product Symbol Definitions
Chapter 1: Introduction
1
Product Symbol Definitions
For Affinity® Three Birthing Bed symbol definitions, see table 1-9 on page 1­
29.

Table 1-9. Product Symbol Definitions

Symbol Description
Type B applied part according to EN 60601-1.

IPX2 According to IEC 529


CAUTION: Consult accompanying documents.

Conforms to the European Medical Device Directive


93/42/EEC.

Certified by Underwriters Laboratories Inc.® in accordance


a

with UL2601-1, CAN/CSA C22.2 NO.601.1, EN 60601-1,


IEC 601-2-38 and IEC 601-1-2.

Approval mark from VDE Prüf-und-Zertifizierungsinstitut in


accordance with EN 60601-1, IEC 601-2-38, and IEC 601-1-2.

CPR function—Identifies the release lever that can be used to


manually drop the inclined head section, so that
cardiopulmonary resuscitation can be performed without delay.

Trend-Like function

a. Underwriters Laboratories Inc.® is a registered trademark of Underwriters Laboratories Inc.

(Continued on next page.)

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 29
Service Manual (MAN272 REV 3)
Product Symbol Definitions
Chapter 1: Introduction

Table 1-9. Product Symbol Definitions (cont'd)

Symbol Description
Lockout control label—The switch in the up position indicates
the lockout control is on. The switch in the down position
indicates the lockout control is off.

Lockout control status—when the lockout control status light is


on, the lockout function is activated.

Battery charge status

Nurse call

Identifying mains fuse

Alternating current

Equipotentiality

(Continued on next page.)

Page 1 - 30 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Product Symbol Definitions

Table 1-9. Product Symbol Definitions (cont'd)


Chapter 1: Introduction
1
Symbol Description
Safe working load

= kg

Electric shock hazard

Keep feet clear of this area. Potential for injury.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 31
Service Manual (MAN272 REV 3)
Warning and Caution Labels
Chapter 1: Introduction

Warning and Caution Labels

Figure 1-8. Warning and Caution Labels (Sheet 1 of 3)

Page 1 - 32 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Warning and Caution Labels
Chapter 1: Introduction

Figure 1-9. Warning and Caution Labels (Sheet 2 of 3)


1

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 33
Service Manual (MAN272 REV 3)
Warning and Caution Labels
Chapter 1: Introduction

Figure 1-10. Warning and Caution Labels (Sheet 3 of 3)

Page 1 - 34 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Warning and Caution Labels

NOTES:
Chapter 1: Introduction
1

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 35
Service Manual (MAN272 REV 3)
Warning and Caution Labels
Chapter 1: Introduction

Page 1 - 36 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Chapter 2
Troubleshooting Procedures
2
Getting Started

WARNING:
Only facility-authorized personnel should troubleshoot the bed.
Troubleshooting by unauthorized personnel could result in personal
injury or equipment damage.

Begin each procedure in this chapter with step 1. Follow the sequence outlined
(each step assumes the previous step has been completed). In each step, the
normal operation of the product can be confirmed by answering Yes or No to
the statement. Your response will lead to another step in the procedure, a repair
analysis procedure (RAP), or a component replacement. If more than one
component is listed, replace them in the given order.

To begin gathering information about the problem, start with Initial Actions.

To isolate or identify a problem and to verify the repair after completing each
corrective action (replacing or adjusting a part, seating a connector, etc.),
perform the Function Checks.

To verify the repair, perform the Final Actions after the Function Checks.

If troubleshooting procedures do not isolate the problem, call Hill-Rom


Technical Support at (800) 445-3720 for assistance.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 1
Service Manual (MAN272 REV 3)
Test Equipment
Chapter 2: Troubleshooting Procedures

Test Equipment
You will need a digital or analog multimeter (VOM) with fine tip probes to
troubleshoot the bed.

WARNING:
Refer to your VOM owner’s manual for complete and detailed
information regarding the operation of your VOM. Failure to do so could
result in personal injury or equipment damage.

Figure 2-1 on page 2-2 represents a common digital VOM. The three basic
electrical functions that you will test are alternating current (AC), direct
current (DC), and ohms/resistance.

Figure 2-1. VOM

Figure 2-1 on page 2-2 displays the correct connection for the fine tip probes.
The red probe plugs into the port marked “V Ω.” The black probe plugs into
the port marked “COM.” The troubleshooting repair analysis procedure (RAP)
indicates where on the bed to connect the red probe and black probe.

Page 2 - 2 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Initial Actions
Chapter 2: Troubleshooting Procedures

WARNING:
Unplug the bed from its power source and disconnect the battery
back-up before checking ohms/resistance measurements. Failure to
disconnect line voltage to the bed can damage the VOM and cause
equipment damage or personal injury.

Initial Actions
2
To gather information from operators concerning problems with the bed, use
Initial Actions. Note symptoms or other information concerning the problem
that the operator describes. This information helps identify the probable cause.

1. Someone who can explain the problem is available.


Yes No
↓ → Go to “Function Checks” on page 2-4.

2. Ask that person to demonstrate or explain the problem. The problem can be
duplicated.
Yes No
↓ → Go to “Function Checks” on page 2-4.

3. The problem is a result of improper operator action.


Yes No
↓ → Go to “Function Checks” on page 2-4.

4. Instruct the operator to refer to the procedures in the Affinity® Three


Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025). Do
the “Function Checks” on page 2-4.

Quick Reference Problem/Solution Matrix


If a problem with the bed system is readily identified, use table 2-1 on page 2-4
to quickly go to the applicable troubleshooting procedure. If the problem is not
readily identified, go to “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 3
Service Manual (MAN272 REV 3)
Function Checks
Chapter 2: Troubleshooting Procedures

Table 2-1. Quick Reference Problem/Solution Matrix

Problem Solution
Hilow Drive Operation Failures RAP 2.1
Hilow Up Malfunction RAP 2.2
Hilow Down Malfunction RAP 2.3
Head Drive Operation Failures RAP 2.4
Head Up Malfunction RAP 2.5
Head Down Malfunction RAP 2.6
Foot Drive Operation Failures RAP 2.7
Foot Up Malfunction RAP 2.8
Foot Down Malfunction RAP 2.9
Hilow Up Malfunction (When The Foot Section Is Lowered) RAP 2.10
Brake Caster Malfunction RAP 2.11
Bed Air Surface Malfunction RAP 2.12
Night Light Does Not Illuminate RAP 2.13
Siderail Mechanism Does Not Hold RAP 2.14
Trend-Like Position Malfunction RAP 2.15
Battery Backup Malfunction RAP 2.16
CPR Release Malfunction RAP 2.17

Function Checks
Function checks determine whether the bed is operating properly. All caregiver
control panel functions are available from both the right and left siderails.
When checking redundant function controls, activate each of the siderails to
determine if the fault is contained in one or both of the siderails.

1. Initial Actions have been performed.


Yes No
↓ → Go to “Initial Actions” on page 2-3.

2. The bed is plugged into an appropriate power source.

Page 2 - 4 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Function Checks
Chapter 2: Troubleshooting Procedures

Yes No
↓ → Plug the bed into an appropriate power source.

3. The lockout switch is in the “ON” (unlocked) position before proceeding


with the function checks.

2
Yes No
↓ → Place the lockout switch in the “ON” (unlocked) position. Go to
step 4.

4. Inspect the siderail cable connections to the logic control P.C. board cable.
The cables are connected properly.
Yes No
↓ → Connect the siderail cable to the cable from the logic control P.C.
board. Go to step 5.

5. Visually inspect the bed for loose connections on cable assemblies and
electrical wires. All cable connections and wiring are properly secured.
Yes No
↓ → Secure the cable connections and wiring. Go to step 6.

6. Inspect the bed for signs of obvious damage. The bed appears to be all
right.
Yes No
↓ → Repair the damage, and proceed to the section, “Hilow System
Functional Check” on page 2-5.

Hilow System Functional Check


1. Use the bed function controls to place the bed in the following positions:
• Position the bed in the mid-height position with the hilow function.
• Raise the foot section to the high position with the foot function.
• Lower the head to the low position with the head function.
• Ensure that the bed is out of the Trend-Like position (the sleep surface is
level).

2. Momentarily activate the hilow up switch. The bed rises.


Yes No
↓ → Go to RAP 2.2.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 5
Service Manual (MAN272 REV 3)
Function Checks
Chapter 2: Troubleshooting Procedures

3. Momentarily activate the hilow down switch. The bed lowers.


Yes No
↓ → Go to RAP 2.3.

4. Activate the hilow up and down switch. The hilow motor is free of
excessive noise or any mechanical binding/grinding.
Yes No
↓ → Go to RAP 2.1.

5. Raise, and then lower the sleep surface using the hilow function. The bed
sleep surface stops and does not drift downward or coast.
Yes No
↓ → Go to RAP 2.1.

6. Raise the bed to the high position using the hilow function. Determine the
hilow up limit setting by measuring at the side of the bed from the floor to
the top of the mainframe (see figure 4-5 on page 4-12). The hilow limit
(LBU) measurement is 31" ± ½" (79 cm ± 12.7 mm).
Yes No
↓ → Adjust the hilow limit switch (LBU) (refer to procedure 4.4). If
this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

7. Lower the bed to the low position using the hilow function. The bed stops
at the (LBD) limit stop. The bed height is 16" ± ½" (41 cm ± 12.7 mm).
Yes No
↓ → Adjust the hilow limit switch (LBD) (refer to procedure 4.4). If
this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

Trend-Like Position Functional Check


1. Position the bed into a mid height position with the hilow function.
Activate the Trend-Like position release handle at the head end of the bed,
and place the bed in and out of the Trend-Like position. The bed goes in
and out of Trend-Like position smoothly, and the sleep surface is secure
when the handle is released:
Yes No
↓ → Adjust the Trend-Like position sensing switch (refer to
procedure 4.9). If this solves the problem, go to “Final Actions”
on page 2-10. Otherwise, call Hill-Rom Technical Support at
(800) 445-3720.

Page 2 - 6 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Function Checks
Chapter 2: Troubleshooting Procedures

Head System Functional Check


1. Ensure that the sleep surface is level, and place the bed in the following
positions:
• Place the bed in a mid-height position with the hilow function.
• Place the head section in a mid-height position with the head function.

2. Momentarily activate the head up switch. The head section rises.


2
Yes No
↓ → Go to RAP 2.5.

3. Momentarily activate the head down switch. The head section lowers.
Yes No
↓ → Go to RAP 2.6.

4. Activate the head up or down switch. The head motor is free of excessive
noise or any mechanical binding/grinding.
Yes No
↓ → Go to RAP 2.4.

5. Raise the head section to the high position with the head up switch. Note
the position of the sleep surface. The head section stops and does not drift
downward or coast.
Yes No
↓ → Go to RAP 2.4.

6. Raise the head section to the high position with the head up switch. Evenly
distribute a maximum of 50 lb (23 kg) of weight on the head section of the
sleep surface simulating a 50 lb (23 kg) patient. Activate the CPR function
by pulling the CPR release lever located at the side of the bed. The head
section lowers to a flat position within approximately 7 seconds.
Yes No
↓ → Adjust the CPR release cables (refer to procedure 4.8). If this
solves the problem, go to “Final Actions” on page 2-10.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 7
Service Manual (MAN272 REV 3)
Function Checks
Chapter 2: Troubleshooting Procedures

Foot System Functional Check


1. Place the bed in the following positions:
• Position the bed in the mid-height position with the hilow function.
• Use the foot function to raise the foot section until the mattress is level with
the seat section mattress.
• Lower the head section to the low position with the head function.
• Ensure that the bed is out of the Trend-Like position (the sleep surface is
level).

2. Press the foot up switch, and activate the foot section. The foot section
rises when the foot up switch is activated.
Yes No
↓ → Go to RAP 2.8.

3. Press the foot down switch. The foot section lowers when the foot down
switch is activated.
Yes No
↓ → Go to RAP 2.9.

4. Raise the foot section to the high position using the foot up switch. The top
of the foot section stops approximately 2-1/2" (6.35 cm) above the seat
section.
Yes No
↓ → Adjust the foot limit switch (refer to procedure 4.7). If this solves
the problem, go to “Final Actions” on page 2-10. Otherwise, call
Hill-Rom Technical Support at (800) 445-3720.

5. Lower the foot section to the low position with the foot down switch. The
foot drive stops when the foot lift arm is 1/4" (6.4 mm) above the foot end
hilow lift arm (see figure 4-9 on page 4-21).
Yes No
↓ → Adjust the foot limit switch (refer to procedure 4.7). If this solves
the problem, go to “Final Actions” on page 2-10. Otherwise, call
Hill-Rom Technical Support at (800) 445-3720.

6. Lower the bed to the low position with the hilow down switch. Lower the
foot section with the foot down switch. The bed automatically rises from
the low position.
Yes No
↓ → Go to RAP 2.10.

Page 2 - 8 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Function Checks
Chapter 2: Troubleshooting Procedures

7. Activate the foot up or down switch. The foot motor is free of excessive
noise or any mechanical binding/grinding.
Yes No
↓ → Go to RAP 2.7.

2
Air System Functional Check
1. Press the seat inflate button, and activate the bed air surface. The seat
section mattress inflates.
Yes No
↓ → Go to RAP 2.12.

2. Press and hold the seat deflate button, and activate the bed air surface. The
seat section mattress deflates.
Yes No
↓ → Go to RAP 2.12.

3. Press the back inflate button, and activate the bed air surface. The lumbar
section inflates.
Yes No
↓ → Go to RAP 2.12.

4. Press the back deflate button to activate the bed air surface. The lumbar
section deflates.
Yes No
↓ → Go to RAP 2.12.

Other Bed Functional Checks


1. The lockout switch locks out the designated inboard/outboard controls
when placed in the “OFF” position (locked), and enables the functions
when placed in the “ON” position (unlocked). The lockout functions work
properly.
Yes No
↓ → Replace the faulty rocker switch (refer to procedure 4.26). If this
solves the problem, go to “Final Actions” on page 2-10.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 9
Service Manual (MAN272 REV 3)
Final Actions
Chapter 2: Troubleshooting Procedures

2. Operate all radio, TV, nurse call, and lighting functions in both siderails.
All functions operate properly.
Yes No
↓ → Replace the P.C. switch board containing the faulty function(s)
for caregiver controls (refer to procedure 4.29), or for patient
controls (refer to procedure 4.28). If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, call Hill-Rom
Technical Support at (800) 445-3720.

3. Perform the brake/steer performance check. Place the brake steer pedal into
the brake position and then the steer position. The brake and steer functions
operate properly.
Yes No
↓ → Go to RAP 2.11, and adjust the brake and steer functions (refer
to procedure 4.30). If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, call Hill-Rom Technical
Support at (800) 445-3720.

4. Operate the labor grips on both sides of the bed. The labor grips latch when
in the full up position and then release when the release handle is activated.
Yes No
↓ → Replace the faulty labor grips (refer to procedure 4.15).

5. Operate the siderail latching mechanism by raising and then lowering the
siderail. The siderail latches when in the full up position and lowers slowly
when the release handle is activated and allowed to free fall.
Yes No
↓ → Replace the faulty siderail assembly (refer to procedure 4.27).

6. Unplug the bed from its power source, and verify that the battery back-up
is operating properly by activating bed functions. The battery back-up
provides sufficient power to operate the bed functions.
Yes No
↓ → Go to RAP 2.16.

7. Go to “Final Actions” on page 2-10.

Final Actions
1. Complete the required preventive maintenance procedures. See
“Preventive Maintenance” on page 6-7.

2. Complete all required administrative tasks.

Page 2 - 10 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Final Actions
Chapter 2: Troubleshooting Procedures

2.1 Hilow Drive Operation Failures


The hilow motor runs, but other problems with the motor are suspected
(i.e., it is noisy, it drifts, etc.).

1. Raise and lower the bed with the hilow function. Look for evidence of
physical damage to the drive system (i.e., check for a bent motor shaft,
loose metal, etc.). The drive system passes inspection. 2
Yes No
↓ → Replace the hilow motor (refer to procedure 4.3). If this solves
the problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 2.

2. Inspect the mounting clevis pins and cotter pins on the hilow motor drive.
The clevis pins are installed properly.
Yes No
↓ → Install the hilow motor clevis pins properly to secure the hilow
motor to the bed frame. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 3.

3. Raise and lower the bed under a load condition. The hilow drive runs
quietly.
Yes No
↓ → Inspect for damage to the hilow motor. Replace if necessary
(refer to procedure 4.3). If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 4.

4. Using the hilow function, raise and lower the bed. The hilow motor retains
its position, and does not drift or coast.
Yes No
↓ → Replace the hilow motor (refer to procedure 4.3). If this solves
the problem, go to “Final Actions” on page 2-10. Otherwise, call
Hill-Rom Technical Support at (800) 445-3720.

5. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 11
Service Manual (MAN272 REV 3)
Final Actions
Chapter 2: Troubleshooting Procedures

2.2 Hilow Up Malfunction


The hilow motor does not raise the bed when the hilow up switch is activated.

1. The bed is plugged into the appropriate power source, and the lockout
switch is off.
Yes No
↓ → Plug the bed into an appropriate power source and turn the
lockout switch off. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 2.

2. Activate the hilow up switch on the opposite siderail. The hilow motor
runs.
Yes No
↓ → Go to step 4.

3. Activate another bed function switch on the siderail with the non-
functioning hilow up switch. The other bed functions work properly.
Yes No
↓ → Replace the P.C. switch board containing the faulty function(s)
for caregiver controls (refer to procedure 4.30) or for patient
controls (refer to procedure 4.29). If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 4.

4. Activate the hilow up switch, and listen for an audible relay click. An
audible clicking sound can be heard.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 5.

5. Allow the hilow motor to cool. Activate the hilow motor up switch. The
hilow motor runs.
Yes No
↓ → Replace the hilow motor (refer to procedure 4.3). If this solves
the problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 7.

6. Go to “Final Actions” on page 2-10.

Page 2 - 12 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.2 Hilow Up Malfunction
Chapter 2: Troubleshooting Procedures

7. Set your VOM to measure V DC. At the logic control P.C. board connector
P15, place your black probe into pin 1 (black wire) and your red probe into
pin 2 (blue wire). Press the hilow up switch. The voltage is between
25V DC and 30V DC.
Yes No

2
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 8.

8. Replace the hilow motor (refer to procedure 4.3).


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

9. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 13
Service Manual (MAN272 REV 3)
2.3 Hilow Down Malfunction
Chapter 2: Troubleshooting Procedures

2.3 Hilow Down Malfunction


The hilow motor does not lower the bed when the hilow down switch is
activated.

1. The bed is plugged into the appropriate power source, and the lockout
switch is off.
Yes No
↓ → Plug the bed into an appropriate power source and turn the
lockout switch off. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 2.

2. Activate the hilow down switch on the opposite siderail. The hilow motor
runs.
Yes No
↓ → Go to step 4.

3. Activate another bed function switch on the siderail with the non-
functioning hilow down switch. The other bed functions work properly.
Yes No
↓ → Replace the P.C. switch board containing the faulty function(s)
for caregiver controls (refer to procedure 4.30) or for patient
controls (refer to procedure 4.29). If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 4.

4. Activate the hilow down switch, and listen for an audible relay click. An
audible clicking sound can be heard.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 5.

5. Allow the hilow motor to cool. Activate the hilow motor down switch. The
hilow motor runs.
Yes No
↓ → Replace the hilow motor (refer to procedure 4.3). If this solves
the problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 7.

6. Go to “Final Actions” on page 2-10.

Page 2 - 14 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.3 Hilow Down Malfunction
Chapter 2: Troubleshooting Procedures

7. Set your VOM to measure V DC. At the logic control P.C. board connector
P15, place your black probe into pin 1 (black wire) and your red probe into
pin 2 (blue wire). Press the hilow down switch. The voltage is between
25V DC and 30V DC.
Yes No

2
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 8.

8. Replace the hilow motor (refer to procedure 4.3).


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

9. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 15
Service Manual (MAN272 REV 3)
2.4 Head Drive Operation Failures
Chapter 2: Troubleshooting Procedures

2.4 Head Drive Operation Failures


The hilow motor runs, but other problems with the motor are suspected
(i.e., it is noisy, it drifts, etc.).

1. Raise and lower the head section with the head function. Look for evidence
of physical damage to the drive system (i.e., check for a bent motor shaft,
loose metal, etc.). The drive system passes inspection.
Yes No
↓ → Replace the head motor (refer to procedure 4.5). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 2.

2. Inspect the mounting clevis pins and cotter pins on the head motor drive.
The clevis pins are installed properly.
Yes No
↓ → Install the head motor clevis pins properly to secure the head
motor to the bed frame. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 3.

3. Raise and lower the head section under a load condition. The head drive
runs quietly.
Yes No
↓ → Inspect for damage to the head motor. Replace if necessary (refer
to procedure 4.5). If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 4.

4. Using the head function, raise and lower the head section. The head motor
retains its position, and does not drift or coast.
Yes No
↓ → Replace the head motor (refer to procedure 4.5). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise, call
Hill-Rom Technical Support at (800) 445-3720.

5. Go to “Final Actions” on page 2-10.

Page 2 - 16 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.5 Head Up Malfunction
Chapter 2: Troubleshooting Procedures

2.5 Head Up Malfunction


The head motor does not raise the head section when the head up switch is
activated.

1. The bed is plugged into the appropriate power source, and the lockout
switch is off.
Yes No
2
↓ → Plug the bed into an appropriate power source and turn the
lockout switch off. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 2.

2. Check the CPR release handles on either side of the bed. The CPR handles
are in the proper position, and no sheets or blankets are caught between the
head section and the CPR release handle causing the CPR release to
actuate.
Yes No
↓ → Remove the sheet or blanket that is causing the CPR release to
actuate. If this solves the problem, go to “Final Actions” on page
2-10. Otherwise, continue to step 3.

3. Activate the head up switch on the opposite siderail. The head motor runs.
Yes No
↓ → Go to step 5.

4. Activate another bed switch on the siderail with the non-functioning head
up switch. The other bed functions work properly.
Yes No
↓ → Replace the P.C. switch board containing the faulty function(s)
for caregiver controls (refer to procedure 4.30) or for patient
controls (refer to procedure 4.29). If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 5.

5. Activate the head up switch, and listen for an audible relay click. An
audible clicking sound can be heard.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 6.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 17
Service Manual (MAN272 REV 3)
2.5 Head Up Malfunction
Chapter 2: Troubleshooting Procedures

6. Allow the head motor to cool. Activate the head motor up switch. The head
motor runs.
Yes No
↓ → Replace the head motor (refer to procedure 4.5). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 8.

7. Go to “Final Actions” on page 2-10.

8. Set your VOM to measure V DC. At the logic control P.C. board connector
P4, place your black probe into pin 1 (brown wire) and your red probe into
pin 2 (blue wire). Press the head up switch. The voltage is between
25V DC and 30V DC.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 9.

9. Replace the head motor (refer to procedure 4.5).


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

10. Go to “Final Actions” on page 2-10.

Page 2 - 18 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.6 Head Down Malfunction
Chapter 2: Troubleshooting Procedures

2.6 Head Down Malfunction


The head motor does not lower the head section when the head down switch is
activated.

1. The bed is plugged into the appropriate power source, and the lockout
switch is off.
Yes No
2
↓ → Plug the bed into an appropriate power source and turn the
lockout switch off. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 2.

2. Activate the head down switch on the opposite siderail. The head motor
runs.
Yes No
↓ → Go to step 4.

3. Activate another bed switch on the siderail with the non-functioning head
down switch. The other bed functions work properly.
Yes No
↓ → Replace the P.C. switch board containing the faulty function(s)
for caregiver controls (refer to procedure 4.30) or for patient
controls (refer to procedure 4.29). If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 4.

4. Activate the head down switch, and listen for an audible relay click. An
audible clicking sound can be heard.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 5.

5. Allow the head motor to cool. Activate the head down switch. The head
motor runs.
Yes No
↓ → Replace the head motor (refer to procedure 4.5). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 7.

6. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 19
Service Manual (MAN272 REV 3)
2.6 Head Down Malfunction
Chapter 2: Troubleshooting Procedures

7. Set your VOM to measure V DC. At the logic control P.C. board connector
P4, place your black probe into pin 1 (brown wire) and your red probe into
pin 2 (blue wire). Press the head down switch. The voltage is between 25V
DC and 30V DC.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 8.

8. Replace the head motor (refer to procedure 4.5).


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

9. Go to “Final Actions” on page 2-10.

Page 2 - 20 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.7 Foot Drive Operation Failures
Chapter 2: Troubleshooting Procedures

2.7 Foot Drive Operation Failures


The foot motor runs, but other problems with the motor are suspected (i.e., it is
noisy, it drifts, etc.).

1. Raise and lower the foot section with the foot function. Look for evidence
of physical damage to the drive system (i.e., check for a bent motor shaft,
loose metal, etc.). The drive system passes inspection. 2
Yes No
↓ → Replace the foot motor (refer to procedure 4.6). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 2.

2. Inspect the mounting clevis pins and cotter pins on the foot motor drive.
The clevis pins are installed properly.
Yes No
↓ → Install the foot motor clevis pins properly to secure the foot
motor to the bed frame. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 3.

3. Raise and lower the foot section under a load condition. The foot drive runs
quietly.
Yes No
↓ → Inspect for damage to the foot motor. Replace if necessary (refer
to procedure 4.6). If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 4.

4. Using the foot function, raise the foot section up and down. The foot motor
retains its position, and does not drift or coast.
Yes No
↓ → Replace the foot motor (refer to procedure 4.6). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise, call
Hill-Rom Technical Support at (800) 445-3720.

5. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 21
Service Manual (MAN272 REV 3)
2.8 Foot Up Malfunction
Chapter 2: Troubleshooting Procedures

2.8 Foot Up Malfunction


The foot motor does not raise the foot section when the foot up switch is
activated.

1. The bed is plugged into the appropriate power source, and the lockout
switch is off.
Yes No
↓ → Plug the bed into an appropriate power source and turn the
lockout switch off. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 2.

2. Activate the foot up switch on the opposite siderail. The foot motor runs.
Yes No
↓ → Go to step 4.

3. Activate another bed switch on the siderail with the non-functioning foot
up switch. The other bed functions work properly.
Yes No
↓ → Replace the P.C. switch board containing the faulty function(s)
for caregiver controls (refer to procedure 4.30) or for patient
controls (refer to procedure 4.29). If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 4.

4. Activate the foot up switch, and listen for an audible relay click. An
audible clicking sound can be heard.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 5.

5. Allow the foot motor to cool. Activate the foot up switch. The foot motor
runs.
Yes No
↓ → Replace the foot motor (refer to procedure 4.6). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 7.

6. Go to “Final Actions” on page 2-10.

Page 2 - 22 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.8 Foot Up Malfunction
Chapter 2: Troubleshooting Procedures

7. Set your VOM to measure V DC. At the logic control P.C. board connector
P6, place your black probe into pin 1 (black wire) and your red probe into
pin 2 (blue wire). Press the foot up switch. The voltage is between 25V DC
and 30V DC.
Yes No

2
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 8.

8. Replace the foot motor (refer to procedure 4.6).


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

9. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 23
Service Manual (MAN272 REV 3)
2.9 Foot Down Malfunction
Chapter 2: Troubleshooting Procedures

2.9 Foot Down Malfunction


The foot motor does not lower the foot section when the foot down switch is
activated.

1. The bed is plugged into the appropriate power source, and the lockout
switch is off.
Yes No
↓ → Plug the bed into an appropriate power source and turn the
lockout switch off. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 2.

2. Activate the foot down switch on the opposite siderail. The foot motor
runs.
Yes No
↓ → Go to step 4.

3. Activate another bed switch on the siderail with the non-functioning foot
down switch. The other bed functions work properly.
Yes No
↓ → Replace the P.C. switch board containing the faulty function(s)
for caregiver controls (refer to procedure 4.30) or for patient
controls (refer to procedure 4.29). If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 4.

4. Activate the foot down switch, and listen for an audible relay click. An
audible clicking sound can be heard.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 5.

5. Allow the foot motor to cool. Activate the foot down switch. The foot
motor runs.
Yes No
↓ → Replace the foot motor (refer to procedure 4.6). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 7.

6. Go to “Final Actions” on page 2-10.

Page 2 - 24 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.9 Foot Down Malfunction
Chapter 2: Troubleshooting Procedures

7. Set your VOM to measure V DC. At the logic control P.C. board connector
P6, place your black probe into pin 1 (black wire) and your red probe into
pin 2 (blue wire). Press the foot down switch. The voltage is between
25V DC and 30V DC.
Yes No

2
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 8.

8. Replace the foot motor (refer to procedure 4.6).


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

9. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 25
Service Manual (MAN272 REV 3)
2.10 Hilow Up Malfunction (When The Foot Section Is Lowered)
Chapter 2: Troubleshooting Procedures

2.10 Hilow Up Malfunction (When The Foot Section Is Lowered)


The hilow function (or motor) fails to automatically raise the bed from the low
position when the foot section is lowered to its lowest point.

1. Activate the hilow function from the siderails and pendant control. The
hilow function works properly.
Yes No
↓ → Go to “Hilow Up Malfunction” on page 2-12 or “Hilow Down
Malfunction” on page 2-14. If this solves the problem, go to
“Final Actions” on page 2-10. Otherwise, continue to step 2.

2. Put the bed into the low position. Lower the foot section to the low
position. The bed automatically rises.
Yes No
↓ → Check the hilow and foot limit switches. Refer to “Hilow Limit
Switch” on page 4-11 or “Foot Limit Switch” on page 4-21. If
this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

3. Go to “Final Actions” on page 2-10.

Page 2 - 26 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.11 Brake Caster Malfunction
Chapter 2: Troubleshooting Procedures

2.11 Brake Caster Malfunction


1. The brake caster pedal locks in the brake position when you push on it with
your foot.
Yes No

2
↓ → Inspect the caster for excessive wear. If necessary, replace the
damaged caster (refer to procedure 4.32). If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise, go to
step 2.

2. The casters lock in the brake position and do not roll or rotate when you
push in the brake pedal.
Yes No
↓ → Adjust the caster assembly (refer to procedure 4.32). If this
solves the problem, go to “Final Actions” on page 2-10.
Otherwise, go to step 3.

3. Another part of the bed interferes with the brake mechanism.


Yes No
↓ → If the brakes still do not work, call Hill-Rom Technical Support
at (800) 445-3720.

4. Take appropriate actions to eliminate the interference.


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

5. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 27
Service Manual (MAN272 REV 3)
2.12 Bed Air Surface Malfunction
Chapter 2: Troubleshooting Procedures

2.12 Bed Air Surface Malfunction


The bed air surface does not function properly.

1. Activate all bed air surface switches in both siderails. Only one switch does
not work.
Yes No
↓ → Go to step 4.

2. Replace the P.C. switch board containing the faulty function for caregiver
controls (refer to procedure 4.29) or for patient controls (refer to procedure
4.28). This solves the problem.
Yes No
↓ → Go to step 4.

3. Go to “Final Actions” on page 2-10.

4. Activate all bed air surface switches in both siderails. All of the switches in
one siderail work while several of the switches in the opposite siderail fail.
Yes No
↓ → Go to step 7.

5. Replace the P.C. switch board(s) containing the faulty functions for
caregiver controls (refer to procedure 4.29) or for patient controls (refer to
procedure 4.28). This solves the problem.
Yes No
↓ → Go to step 7.

6. Go to “Final Actions” on page 2-10.

7. Activate one of the bed air function switches. Audible clicks of the relay
from the air system control board can be heard.
Yes No
↓ → Go to step 9.

8. Go to step 12.

9. Set your VOM to measure V DC. At the logic control P.C. board connector
P8, probe pins 7, 8, 9, 11, and 12 while pressing each siderail air function
button (see table 2-2 on page 2-29). The pin voltages are correct.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

Page 2 - 28 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.12 Bed Air Surface Malfunction
Chapter 2: Troubleshooting Procedures

Table 2-2. Three Station Manifold I/O Signal Reference


Logic Control P.C. Board Connector P8
Input Command: Mattress Function
Output Wire BF/ BS/ BF/
Pin No. SF SS BF BS AI BS/SF

Pump
Signal
11
Color
GRN X X
SF SS SS
X X X 2
Lum 7 BLU X X X X
Seat 8 VIO X X X X X
In/out 9 RED X X X
Super­ 12 ORG X X X X X X X
visory
X= ON= 5V DC (approximately)
Legend: SF= Seat firm SS= Seat soft BF= Back firm BS= Back soft AI= Auto inflate

10. Replace the cable or connectors from the logic control P.C. board to the air
system control P.C. board. This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

11. Go to “Final Actions” on page 2-10.

12. Disconnect the air connectors at the seat section. Activate the back inflate
button, and test for air pressure on the patients left-hand side. Activate the
back deflate button, and test for air release on the patient’s left-hand side.
Activate the seat inflate button, and test for air pressure on the patients
right hand side. The air pressure and air release is correct at both of the
connectors.
Yes No
↓ → Go to step 15.

13. Ensure that all hoses and bladders are free of kinks, pinching, and leaks.
This solves the problem.
Yes No
↓ → Go to step 15.

14. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 29
Service Manual (MAN272 REV 3)
2.12 Bed Air Surface Malfunction
Chapter 2: Troubleshooting Procedures

15. Pressure and suction are present at the connectors for all functions.
Yes No
↓ → Go to step 17.

16. Go to step 27.

17. Press any bed inflate button. The pump runs.


Yes No
↓ → Go to step 20.

18. Disconnect the air line (A) from the air compressor (B) to the top of the
manifold (C) (see figure 2-2 on page 2-30). Press any bed inflate button to
activate the pump. The air compressor creates suction in the air line.
Yes No
↓ → Replace the air compressor (refer to procedure 4.20). If this
solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 27.

19. Go to step 27.

Figure 2-2. Air System

Page 2 - 30 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.12 Bed Air Surface Malfunction
Chapter 2: Troubleshooting Procedures

20. Set your VOM to measure V AC. At the air system control P.C. board
connector P19, probe pin 1 (black) and pin 3 (white). The voltage is
approximately 115V AC.

NOTE:
For high voltage checks, do not unplug the connectors.

NOTE:
Mains power is 120V AC if domestic, 230V AC if European.
2
Yes No
↓ → Go to step 24.

21. Set your VOM to measure V AC. At the air system control P.C. board
connector P20, probe pin 1 (black) and pin 3 (white). Press any bed inflate
button. The voltage is approximately 115V AC.

NOTE:
For high voltage checks, do not unplug the connectors.

NOTE:
Mains power is 120V AC if domestic, 230V AC if European.
Yes No
↓ → Replace the air system control P.C. board (refer to procedure
4.20). If this solves the problem, go to “Final Actions” on page
2-10. Otherwise, continue to step 27.

22. Replace the air compressor (refer to procedure 4.20). This solves the
problem.
Yes No
↓ → Go to step 27.

23. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 31
Service Manual (MAN272 REV 3)
2.12 Bed Air Surface Malfunction
Chapter 2: Troubleshooting Procedures

24. Set your VOM to measure V AC. At the logic control P.C. board connector
P16, probe pin 1 (black) and pin 3 (white). The voltage is approximately
115V AC if domestic, or 230V AC if European.

NOTE:
For high voltage checks, do not unplug the connectors.

NOTE:
Mains power is 120V AC if domestic, 230V AC if European.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 27.

25. Replace the cable or connectors between the logic control P.C. board and
the air system control P.C. board. This solves the problem.
Yes No
↓ → Go to step 27.

26. Go to “Final Actions” on page 2-10.

27. Set your VOM to measure V AC. At the air system control P.C. board
connector P21, probe pin 1 (white) and at the air system control P.C. board
connector P22, probe the pins listed in table 2-3 on page 2-33 while
pressing the appropriate siderail air function buttons (see table 2-3 on page
2-33). The voltage is approximately 115V AC.

NOTE:
For high voltage checks, do not unplug the connectors.

NOTE:
Mains power is 120V AC if domestic, 230V AC if European.

NOTE:
See “3-Station Manifold I/O Signal Reference” table 2-2 on page 2-29.
Yes No
↓ → Replace the air system control P.C. board (refer to procedure
4.22). If this solves the problem, go to “Final Actions” on page
2-10. Otherwise, call Hill-Rom Technical Support at (800) 445­
3720.

Page 2 - 32 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.12 Bed Air Surface Malfunction
Chapter 2: Troubleshooting Procedures

Table 2-3. Solenoid Test—Air System Control P.C. Board Connector P22

Pin# Wire Color Press Function


1 Yellow or black Back firm
2 Brown Seat firm
5
6
Red
Violet
Seat soft
Seat soft
2
8 Pink Seat soft
9 Tan Seat firm

28. Set your VOM to measure V DC. At the logic control P.C. board connector
P8, probe pins 7, 8, 9, 11, and 12 while pressing each siderail air function
button (see table 2-2 on page 2-29). The pin voltages are correct.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

29. Replace the cable or connectors from the logic control P.C. board to the air
system control P.C. board.
This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

30. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 33
Service Manual (MAN272 REV 3)
2.13 Night Light Does Not Illuminate
Chapter 2: Troubleshooting Procedures

2.13 Night Light Does Not Illuminate


1. The night light comes on when the ambient light dims.
Yes No
↓ → Replace the lamp bulb or fixture if damaged (refer to procedure
4.13). If this solves the problem, go to “Final Actions” on page
2-10. Otherwise, continue to step 2.

2. Verify that the photocell functions. The night light comes on when the
ambient light dims.
Yes No
↓ → Replace the photocell. If this solves the problem, go to “Final
Actions” on page 2-10. Otherwise, continue to step 3.

3. The system functions normally.


Yes No
↓ → Verify that the night light lamp cable is plugged into the logic
control P.C. board connector, P1. If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 4.

4. The system functions normally.


Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

5. Go to “Final Actions” on page 2-10.

Page 2 - 34 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.14 Siderail Mechanism Does Not Hold
Chapter 2: Troubleshooting Procedures

2.14 Siderail Mechanism Does Not Hold


1. From the low (stow) position, raise the siderail by pulling the siderail up
until it clicks in the locked position. It rises and clicks into place.
Yes No

2
↓ → Inspect for obstructions, missing hardware, or loose fasteners. If
found, replace the siderail. See “Siderail Assembly” on page 4­
67. If this solves the problem, go to “Final Actions” on page
2-10. Otherwise, continue to step 2.

2. Grasp the top of the siderail, and push and pull firmly on the siderail. The
siderail remains latched in the locked position.
Yes No
↓ → Inspect for obstructions or loose fasteners. If found, replace the
siderail. See “Siderail Assembly” on page 4-67. If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 3.

3. Stow the siderail as follows: grasp and pull the release handle out, and
allow the siderail to swing down in a controlled, dampened, manner to the
fully stowed position. The siderail goes down into the stored position.
Yes No
↓ → Inspect for obstructions in the center arm, loose fasteners,
missing spring, or binding. If found, replace the siderail. See
“Siderail Assembly” on page 4-67. If this solves the problem, go
to “Final Actions” on page 2-10. Otherwise, continue to step 4.

4. Another part of the bed interferes with the siderail.


Yes No
↓ → If the siderail still does not work properly, call Hill-Rom
Technical Support at (800) 445-3720.

5. Take appropriate actions to eliminate the interference.


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

6. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 35
Service Manual (MAN272 REV 3)
2.15 Trend-Like Position Malfunction
Chapter 2: Troubleshooting Procedures

2.15 Trend-Like Position Malfunction


1. If unable to place the bed in the Trend-Like position, press the hilow up
switch to raise the bed to a higher position. Push down on the Trend-Like
position release handle. The head end is 8° below the foot end.
Yes No
↓ → Verify that the Trend-Like position release handles are properly
adjusted and functional. See “Trend-Like Position Sensing
Switch” on page 4-27 for access and safety precautions in
servicing the Trendelenburg features. If this solves the problem,
go to “Final Actions” on page 2-10. Otherwise, call Hill-Rom
Technical Support at (800) 445-3720.

2. Go to “Final Actions” on page 2-10.

Page 2 - 36 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.16 Battery Back-up Malfunction
Chapter 2: Troubleshooting Procedures

2.16 Battery Back-up Malfunction


1. The battery has been properly charged and is providing power to the bed.
Yes No
↓ → Charge the battery. Refer to “Battery Back-up” on page 1-4. If

2
this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 2.

2. Using the VOM, check the battery fuse for electrical continuity. The
battery fuse is in good condition.
Yes No
↓ → Replace the battery fuse (refer to procedure 4.19). If this solves
the problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 3.

3. Visually inspect for loose connections on cable assemblies and electrical


wires. All cable connections and wiring are properly secured.
Yes No
↓ → Secure the cable connections and wiring. If this solves the
problem, go to “Final Actions” on page 2-10. Otherwise,
continue to step 4.

4. Set your VOM to measure V DC. At the logic control P.C board, connect
the ground lead of the VOM to the negative side of C1 and the positive lead
to the positive side of D15 (the side without a silver band). The voltage is
between 27.6 and 28.0 V DC.
Yes No
↓ → Adjust the battery charge circuit (refer to procedure 4.19). If this
solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 5.

5. Set your VOM to measure V AC. At the logic control P.C board connector
P2, probe pin 1 (yellow) and pin 2 (red). The voltage is approximately 12V
AC or 24V AC.
Yes No
↓ → Replace the transformer assembly (refer to procedure 4.24). If
this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 6.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 37
Service Manual (MAN272 REV 3)
2.16 Battery Back-up Malfunction
Chapter 2: Troubleshooting Procedures

6. Set your VOM to measure V DC. Probe connector P5 at the logic control
P.C. board. The voltage is between 25 V DC to 30V DC.
Yes No
↓ → Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 7.

7. Replace the two batteries in the battery back-up assembly (refer to


procedure 4.19). This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

Go to “Final Actions” on page 2-10.

Page 2 - 38 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.17 CPR Release Malfunction
Chapter 2: Troubleshooting Procedures

2.17 CPR Release Malfunction


1. Pull and hold the CPR release handle. The head section goes into the full
flat position. (When unoccupied, the head section may require slight
downward hand pressure to rotate completely into the full flat position.)

2
Yes No
↓ → Check the adjustment of the CPR function (refer to procedure
4.8). If this solves the problem, go to “Final Actions” on page
2-10. Otherwise, go to step 2.

2. Visually inspect the head drive motor assembly for damage or missing
parts. The head drive motor assembly is in good working condition.
Yes No
↓ → Replace the head drive motor assembly (refer to procedure 4.5).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, go to step 3.

3. Another part of the bed interferes with the CPR release mechanism.
Yes No
↓ → If the head section still does not release, call Hill-Rom Technical
Support at (800) 445-3720.

4. Take appropriate actions to eliminate the interference.


This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.

5. Go to “Final Actions” on page 2-10.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 39
Service Manual (MAN272 REV 3)
2.17 CPR Release Malfunction
Chapter 2: Troubleshooting Procedures

NOTES:

Page 2 - 40 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
2.17 CPR Release Malfunction
Chapter 2: Troubleshooting Procedures

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 41
Service Manual (MAN272 REV 3)
2.17 CPR Release Malfunction
Chapter 2: Troubleshooting Procedures

Page 2 - 42 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Chapter 3
Theory of Operation

Electrical System
The following wiring diagrams detail the electrical circuits and the P.C. boards
on the bed. Use these diagrams as troubleshooting aids.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 3 - 1
Service Manual (MAN272 REV 3)
Electrical System
Chapter 3: Theory of Operation

Figure 3-1. Bed Wiring Diagram—P/N 63993

Refer to fold-out FO 3-1 and FO 3-2 at the rear of this manual.

Page 3 - 2 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Electrical System
Chapter 3: Theory of Operation

Figure 3-2. Logic Control P.C. Board—P/N 68121 Top

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 3 - 3
Service Manual (MAN272 REV 3)
Electrical System
Chapter 3: Theory of Operation

Figure 3-3. Logic Control P.C. Board—P/N 68121 Bottom

Page 3 - 4 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Electrical System
Chapter 3: Theory of Operation

Figure 3-4. Mattress Driver P.C. Board P/N 41188

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 3 - 5
Service Manual (MAN272 REV 3)
Theory of Operation
Chapter 3: Theory of Operation

Theory of Operation
A theory of operation is not available for the bed.

Page 3 - 6 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Theory of Operation
Chapter 3: Theory of Operation

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 3 - 7
Service Manual (MAN272 REV 3)
Theory of Operation
Chapter 3: Theory of Operation

Page 3 - 8 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Chapter 4
Removal, Replacement, and
Adjustment Procedures

Tool and Supply Requirements


To service the bed, the following tools and supplies are required:
• Digital or analog multimeter (VOM)
• Ratchet
• T10 Torx® head bit
1

• T25 Torx® head bit


• 7/16" socket
• ¾" socket
• 5/16" wrench
4
• 7/16" wrench
• ½" wrench
• 11/16" wrench
• Adjustable wrench
• 3/32" Allen™ wrench
2

• 5/32" Allen™ wrench


• Small screwdriver
• Short phillips head screwdriver
• Needle nose pliers
• External snap-ring tool
• E-ring tool
• Split-ring tool

1. Torx® is a registered trademark of Textron, Inc.


2. Allen™ is a trademark of Industrial Fasteners, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 1
Service Manual (MAN272 REV 3)
Tool and Supply Requirements
Chapter 4: Removal, Replacement, and Adjustment Procedures

• Small wire cutters


• Hammer
• 1/8" drift punch
• Tape measure
• Marker pen
• Bed jack
• Bed stands, or (4) pieces of 2" x 4" x 32" lumber
• Anti-seize lubricant
• Blue Loctite® #2421

• Red Loctite® #262


• Isopropyl alcohol
• Turpentine

1. Loctite® is a registered trademark of Loctite Corporation.

Page 4 - 2 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.1 Top Motor Cover
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.1 Top Motor Cover


Tools required: Ratchet T25 Torx® head bit
1

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the hilow function, raise the bed to the high position.

4. Using the head function, raise the head section to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Using a ratchet and T25 Torx® head bit, remove the seven screws (A)
securing the top motor cover (B) (see figure 4-1 on page 4-4).

7. Remove the two headboard bushings (C) from the top motor cover (B). 4
8. Remove the top motor cover (B) from the bed.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 3
Service Manual (MAN272 REV 3)
4.1 Top Motor Cover
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-1. Top Motor Cover Assembly

Replacement
To install the top motor cover (B), reverse the removal procedure.

Page 4 - 4 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.2 Bottom Motor Cover
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.2 Bottom Motor Cover


Tools required: Ratchet T25 Torx® head bit
1

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the hilow function, raise the bed to the high position.

4. Using the head function, raise the head section to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Remove the top motor cover (refer to procedure 4.1).

7. Disconnect the night light and cable assembly (A) from logic control
P.C. board connector P1 (see figure 4-2 on page 4-6). 4
8. Disconnect the pendant control cable (B) from logic control P.C. board
connector P14 (see figure 4-2 on page 4-6).

9. Using a ratchet and T25 Torx® head bit, remove the six screws (C)
securing the bottom motor cover (D).

10. Remove the bottom motor cover (D) from the bed.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 5
Service Manual (MAN272 REV 3)
4.2 Bottom Motor Cover
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-2. Bottom Motor Cover Assembly

Replacement
1. To install the bottom motor cover (D), reverse the removal procedure.

2. Do the “Function Checks” on page 2-4.

Page 4 - 6 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.3 Hilow Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.3 Hilow Drive Motor Assembly


Tools required: T25 Torx® head bit Ratchet
1

Needle nose pliers


Bed stands, or (2) pieces of 2" x 4" x 32" lumber

Removal
1. Set the brake/steer pedal to the brake position.

2. Using the hilow function, raise the bed to the high position.

3. Using the head function, raise the head section to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

4. Unplug the bed from its power source, and engage the lockout control.

5. Remove the top motor cover (refer to procedure 4.1).

6. Remove the bottom motor cover (refer to procedure 4.2).


4
WARNING:
You must use the 2" x 4" pieces of lumber to support the bed. The bed
will fall during this procedure if not supported. Failure to do so could
result in personal injury or equipment damage.

7. Position the two 2" x 4" x 32" pieces of lumber (A) across the base frame
(B) beneath the lift arms (C) (see figure 4-3 on page 4-8).

8. Ensure that the 4" flat side of the lumber (A) lies flat on the base frame (B).

9. Before lowering the bed, visually inspect underneath the bed for
obstructions that may damage the night light when the bed is lowered.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 7
Service Manual (MAN272 REV 3)
4.3 Hilow Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-3. Bed Support Location

CAUTION:
Ensure that the night light is not damaged when the bed is being
lowered. Failure to do so could result in equipment damage.

10. Deactivate the lockout function, and lower the bed until both lift arms (C)
are resting on the 2" x 4"s.

11. Using a ratchet and T25 Torx® head bit, remove the four screws (D)
1

securing the electronics plate weldment (E) to the bed frame (see figure 4-4
on page 4-9).

12. To improve access to the hilow drive motor assembly (F) on the right side,
carefully lift and move the electronics plate weldment (E) to the patient’s
left-hand side of the bed.

13. Unplug the hilow drive motor assembly (F) power cable (G) from logic
control P.C. board (H) connector P15.

1. Torx® is a registered trademark of Textron, Inc.

Page 4 - 8 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.3 Hilow Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-4. Hilow Drive Motor Assembly

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 9
Service Manual (MAN272 REV 3)
4.3 Hilow Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

14. Using the needle nose pliers, remove two cotter pins (I) from the two clevis
pins (J) that secure the hilow drive motor assembly (F) to the bed frame.

15. Remove the two clevis pins (J) from the hilow drive motor assembly (F),
and from the bed frame.

16. Remove the hilow drive motor assembly (F).

Replacement
1. To install the replacement hilow drive motor assembly (F), reverse the
removal procedure.

2. Do the “Function Checks” on page 2-4.

Page 4 - 10 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.4 Hilow Limit Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.4 Hilow Limit Switch


Tools required: Small screwdriver Tape measure

The hilow limit switch enables the correct maximum raised position of the bed.
Its adjustment should be checked whenever the bed’s hilow system is serviced,
such as when replacing the motor and during preventive maintenance. This
switch controls the bed’s travel limits and the Trend-Like function limits.

There are five limits controlled by this switch. Adjusting the bed travel to the
correct maximum height provides the correct adjustment of the other four
limits.

Checking the Hilow Limit Switch


1. Ensure that the bed is operational with the hilow system correctly installed
and connected.

2. Set the brake/steer pedal to the brake position.

3. Ensure that the bed is out of the Trend-Like position.

4
4. Using the hilow function, raise the bed to the high position.

5. Using the head function, raise the head section to the high position.

6. Using the tape measure, measure from the top center point of the main
frame (at the head end of the bed) to the floor. The distance (A) should be
31" ± ½" (787.4 ± 12.7 mm) (see figure 4-5 on page 4-12).

7. If the distance (A) is correct, go to step 8. If the distance (A) is not correct,
adjust the hilow limit switch by following the procedures in the section,
“Adjustment” on page 4-12.

8. Do the “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 11
Service Manual (MAN272 REV 3)
4.4 Hilow Limit Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-5. Hilow Limit Switch

Adjustment
1. If the distance (A) is not 31" ± ½" (787.4 ± 12.7 mm), adjust the hilow limit
switch by performing the following steps:
a. Using the hilow function, raise the bed until the distance (A) is 31"
(787.4 mm).
b. Using the small screwdriver, turn the screw (C) on the hilow limit
switch (LBU), located on the logic control P.C. board, until the green
light comes on (see figure 4-6 on page 4-13).
NOTE:
The green light indicates the maximum limit of the hilow function's travel. If
the green light is out, the hilow function can travel farther until the light comes
on or the bed reaches its mechanical limits.
c. Continue to adjust the screw clockwise and counterclockwise in small
increments, until an adjustment is attained between the green light
coming on and going off.

Page 4 - 12 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.4 Hilow Limit Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-6. Logic Control P.C. Board—P/N 68121

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 13
Service Manual (MAN272 REV 3)
4.4 Hilow Limit Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

d. Using the hilow function, lower the bed until dimension B is 16"
(406.4 mm) (see figure 4-5 on page 4-12).
e. Using the small screwdriver, turn the screw (D) on the hilow limit
switch (LBD), located on the logic control P.C. board, until the green
light comes on (see figure 4-6 on page 4-13).
f. Continue to adjust the screw clockwise and counterclockwise in small
increments, until an adjustment is attained between the green light
coming on and going off.
g. Using the hilow function, lower and then raise the bed.
h. Check the bed height at the high position. Verify that the distance (A) is
31" ± ½" (787.4 ± 12.7 mm) (see figure 4-5 on page 4-12).
i. Check the bed height at the low position. Verify that dimension B is
16" ± 1/2" (406.4 ± 12.7 mm).
j. Repeat step a through step i until the high position of the bed is correct.
NOTE:
An adjustment to the hilow limit switch requires verification of the foot limit
switch adjustment.

2. Check the foot limit switch adjustment as instructed in the section, “Foot
Limit Switch” on page 4-21.

3. Do the “Function Checks” on page 2-4.

Page 4 - 14 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.5 Head Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.5 Head Drive Motor Assembly


Tools required: T25 Torx® head bit Ratchet
1

Needle nose pliers Small wire cutters


Bed stands, or (2) pieces of 2" x 4" x 32" lumber

Removal
1. Set the brake/steer pedal to the brake position.

2. Using the hilow function, raise the bed to the high position.

3. Using the head function, raise the head section to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

4. Unplug the bed from its power source, and engage the lockout control.

5. Remove the top motor cover (refer to procedure 4.1).

6. Remove the bottom motor cover (refer to procedure 4.2).

7. Perform step 7 through step 10 in the section titled “Hilow Drive Motor
4
Assembly” on page 4-7.

CAUTION:
Ensure that the night light is not damaged when the bed is being
lowered. Failure to do so could result in equipment damage.

8. Using a ratchet and T25 Torx® head bit, remove the four screws (A)
securing the electronics plate weldment (B) to the bed frame (see figure 4­
7 on page 4-16).

9. To improve access to the head drive motor assembly (C), carefully lift and
move the electronics plate weldment (B) to the patient’s left-hand side of
the bed.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 15
Service Manual (MAN272 REV 3)
4.5 Head Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-7. Head Drive Motor Assembly

Page 4 - 16 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.5 Head Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

10. Unplug the head drive motor assembly (C) power cable (D) from logic
control P.C. board connector P4.

11. Using the small wire cutters, cut the two cable ties (E) that secure the CPR
cable assembly (F) to the head drive motor shaft (G).

12. Using the needle nose pliers, remove the cotter pin (H), that secures the
CPR cable assembly (F) to the head drive motor release pin (I).

13. Note the orientation of the CPR cables to ensure that the CPR cables are
oriented correctly during the replacement procedure.

14. Remove the CPR cable assembly (F) from the head drive motor.

WARNING:
Prop up the head section before removing the cotter pins and clevis
pins. Failure to do so could result in personal injury or equipment
damage.

15. Lift the head section (J) frame upward, and prop it up (support it).

16. Using the needle nose pliers, remove the cotter pins (K) from the two clevis
pins (L) that secure the head drive motor assembly (C) to the bed frame. 4
17. Remove the head drive motor assembly (C) from the bed frame.

Replacement
1. To install the replacement head drive motor assembly (C), reverse the
removal procedure with attention to the following:
a. Ensure that the CPR cable routing is correct and free from binding.
b. Ensure that all cable ties, clamps, and connectors are properly installed.

2. Do the “Function Checks” on page 2-4.

Adjustment
To adjust the CPR function, refer to “CPR Release” on page 4-25.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 17
Service Manual (MAN272 REV 3)
4.6 Foot Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.6 Foot Drive Motor Assembly


Tools required: T25 Torx® head bit Ratchet
1

Needle nose pliers


Bed stands, or (4) pieces of 2" x 4" x 32" lumber

Removal
1. Set the brake/steer pedal to the brake position.

2. Using the hilow function, raise the bed to the high position.

3. Using the head function, raise the head section to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

4. Unplug the bed from its power source, and engage the lockout control.

5. Remove the top motor cover (refer to procedure 4.1).

6. Remove the bottom motor cover (refer to procedure 4.2).

7. Perform step 7 through step 10 in “Hilow Drive Motor Assembly” on


page 4-7.

WARNING:
You must use the 2" x 4" pieces of lumber to support the bed. The bed
will fall during this procedure if not supported. Failure to support the bed
could result in personal injury or equipment damage.

CAUTION:
Ensure that the night light is not damaged when the bed is being
lowered. Failure to do so could result in equipment damage.

1. Torx® is a registered trademark of Textron, Inc.

Page 4 - 18 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.6 Foot Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

WARNING:
You must use the 2" x 4" pieces of lumber to support the foot yoke. The
foot yoke will fall during this procedure if not supported. Failure to do so
could result in personal injury or equipment damage.

8. Place the remaining 2" x 4" pieces of lumber below the foot yoke.

9. Deactivate the lockout function, and lower the foot section until the foot
yoke is resting on the 2" x 4" pieces of lumber.

10. Using a ratchet and T25 Torx® head bit, remove the four screws (A)
1

securing the electronics plate weldment (B) to the bed frame (see figure 4­
8 on page 4-19).

Figure 4-8. Foot Drive Motor Assembly

11. To improve access to the foot drive motor assembly (C) on the left side,
carefully lift and move the electronics plate weldment (B) to the patient’s
right-hand side of the bed.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 19
Service Manual (MAN272 REV 3)
4.6 Foot Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

12. Unplug the foot drive motor assembly (C) power cable (D) from logic
control P.C. board connector P6.

13. Using the needle nose pliers, remove two cotter pins (E) from the two
clevis pins (F) that secure the foot drive motor assembly (C) to the bed
frame.

14. Remove the two clevis pins (F) from the foot drive motor assembly (C),
and from the bed frame.

15. Remove the foot drive motor assembly (C).

Replacement
1. To install the replacement foot drive motor assembly (C), reverse the
removal procedure.

2. Do the “Function Checks” on page 2-4.

Page 4 - 20 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.7 Foot Limit Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.7 Foot Limit Switch


Tools required: Small screwdriver Tape measure

The foot limit switch adjustment should be performed whenever the bed’s foot
system is serviced, such as when the foot motor is replaced, when the hilow
switch is adjusted, and during preventive maintenance.

Checking the Foot Limit Switch


1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the hilow function, raise the bed to the high position.

4. Using the head function, raise the head section to the high position.

5. Using the foot function, lower the foot section to the low position.

6. Using the tape measure, measure the gap (A) between the foot section left-
hand lift arm (B) and the hilow left-hand lift arm (C). The distance should

4
be exactly ¼" (6.4 mm). (see figure 4-9 on page 4-21).

Figure 4-9. Checking the Foot Limit Switch

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 21
Service Manual (MAN272 REV 3)
4.7 Foot Limit Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

7. If the distance (A) is correct, go to step 8. If the distance (A) is not correct,
adjust the foot limit switch by following the procedures in the section,
“Adjustment” on page 4-22.

8. Do the “Function Checks” on page 2-4.

Adjustment
1. If the distance (A) is not exactly ¼" (6.4 mm), adjust the foot limit switch
by performing the following steps:
a. Using the foot function, raise or lower the foot section until the gap (A)
measures ¼" (6.4 mm) between the foot section left-hand lift arm (B)
and the hilow left-hand lift arm (C).
b. Remove the top motor cover (refer to procedure 4.1).
c. Using the small screwdriver, turn the screw (E) on the foot limit switch
(LFD), located on the logic control P.C. board, until the green light
comes on (see figure 4-10 on page 4-23).
NOTE:
The green light indicates the maximum limit of the foot section's travel. If
the green light is out, the foot section can travel farther until the light
comes on or the foot section reaches its mechanical limits.
d. Continue to adjust the screw clockwise and counterclockwise in small
increments, until an adjustment is attained between the green light
coming on and going off.
e. Using the measuring tape, measure the height of the foot section in the
low position.
f. Using the foot function, raise the foot section up 8¼" (209.6 mm) (D)
from the measurement in the low position (see figure 4-9 on
page 4-21).
g. Using the small screwdriver, turn the screw (F) on the foot limit switch
(LFU), located on the logic control P.C. board, until the green light
comes on (see figure 4-10 on page 4-23).
h. Continue to adjust the screw clockwise and counterclockwise in small
increments, until an adjustment is attained between the green light
coming on and going off.
i. Using the foot function, raise and then lower the foot section.

Page 4 - 22 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.7 Foot Limit Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-10. Logic Control P.C. Board—P/N 68121

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 23
Service Manual (MAN272 REV 3)
4.7 Foot Limit Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

j. Check the gap (A) between the foot section left-hand lift arm (B) and
the hilow left-hand lift arm (C) at the low position. Verify that the
distance (A) is exactly ¼" (6.4 mm).
k. Repeat step a through step j until the correct foot level position is
obtained.

2. Do the “Function Checks” on page 2-4.

Page 4 - 24 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.8 CPR Release
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.8 CPR Release


Tools required: Adjustable wrench

Adjustment
1. Set the brake/steer pedal to the brake position.
2. Ensure that the bed is out of the Trend-Like position.
3. Using the hilow function, place the bed in a mid-height position.
4. Using the head function, raise the head section to its highest position.

5. Ensure that the siderails are in the up position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
6. Unplug the bed from its power source, and engage the lockout control.
7. Remove the top motor cover (refer to procedure 4.1).
8. Ensure that the CPR cables (A) are routed inside the head section
weldment (B) (see figure 4-11 on page 4-25).
Figure 4-11. CPR Release
4

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 25
Service Manual (MAN272 REV 3)
4.8 CPR Release
Chapter 4: Removal, Replacement, and Adjustment Procedures

9. Ensure that the CPR cables (A) are secured loosely with the cable tie (C)
located on the head drive motor assembly.

NOTE:
The cable tie should be snug enough to secure the cable, but not tight enough to
pinch the cable sheath.

10. Using a wrench, adjust the CPR cable tensioning nuts (D) until the CPR
release works properly on both sides of the bed.

11. Do the “Function Checks” on page 2-4.

Page 4 - 26 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.9 Trend-Like Position Sensing Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.9 Trend-Like Position Sensing Switch


Tools required: 5/16" wrench

There are two contact-type microswitches installed in the Trend-Like position


mechanism. Only one switch requires adjustment. Proper adjustment of the
Trend-Like position sensing switch (TS) keeps the bed from going too low
while in the Trend-Like position. The Trend-Like position handle switch (TH)
installed in the release arm mechanism is activated whenever the Trend-Like
position handle is pulled and does not require adjustment.

Adjustment
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the hilow function, place the bed in a mid-height position.

4. Using the head function, raise the head section to its highest position.

5. Ensure that the siderails are in the up position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
4
personal injury or equipment damage.

6. Unplug the bed from its power source, and engage the lockout control.

7. Remove the top motor cover (refer to procedure 4.1).

8. If the bed is equipped with the optional air system, the 5/16" Trend-Like
position switch adjusting screws are accessible. However, easier access
may be obtained by removing the air package (refer to procedure 4.20).

9. Using a 5/16" wrench, loosen the two screws (A) that secure the switch
bracket (B) to the Trend-Like position release mechanism (E) (gas head
and release lever) (see figure 4-12 on page 4-28).

10. Using the Trend-Like position handle, place the bed in the Trend-Like
position while observing the actuator arm (D) of the TS switch (C).

NOTE:
When the bed is in the Trend-Like position, the TS switch should be activated
by the Trend-Like position gas spring cylinder.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 27
Service Manual (MAN272 REV 3)
4.9 Trend-Like Position Sensing Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-12. Trend-Like Position Assembly

11. Ensure that the actuator arm (D) is fully activated. (The arm is fully
activated when the switch makes an audible click.)

12. Using a 5/16" wrench, tighten the two screws (A) that secure the switch
bracket (B) to the Trend-Like position release mechanism (E).

13. After the adjustment is complete, ensure that the Trend-Like position
system is operating correctly. See “Function Checks” on page 2-4.

14. Reverse the removal procedures to reinstall the optional air system (if so
equipped) and the top motor cover.

15. Ensure that all Trend-Like functions work properly. See “Function
Checks” on page 2-4.

Page 4 - 28 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.10 Trend-Like Position Gas Spring
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.10 Trend-Like Position Gas Spring


Tools required: 11/16" wrench Split-ring removal tool

WARNING:
The head end of the main frame is supported by the two Trend-Like
position gas springs. Any servicing will require that a support device be
placed just beneath the head end of the main frame. Failure to do so
could result in personal injury or equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in

4
personal injury or equipment damage.

4. Unplug the bed from its power source.

5. Remove the top motor cover (refer to procedure 4.1).

6. Using the split-ring removal tool, remove the split-ring (A) from the post
(B) at the cylinder end (C) of the Trend-Like position gas spring (E)
(see figure 4-13 on page 4-30).

7. Slide the cylinder end (C) of the Trend-Like position gas spring (E) off the
post (B).

NOTE:
It may be necessary to apply upward force on the head end of the main frame
to free the cylinder from the post.

8. Using the 11/16" wrench, loosen the jam nut (D) on the rod end (F) of the
gas spring (E).

9. Turn the rod end (F) counterclockwise to remove it from the release
assembly (G).

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 29
Service Manual (MAN272 REV 3)
4.10 Trend-Like Position Gas Spring
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-13. Trend-Like Position Gas Spring Assembly

Replacement
1. Spin the jam nut (D) all the way down on the new gas spring's rod end (F).

2. Thread the rod end (F) of the new gas spring (E) into the release assembly
(G) until the plunger in the end of the rod just touches, or is no more than
1/16" away from the release arm (H).

3. Using the 11/16" wrench, tighten the jam nut (D) to lock the rod end (F) in
this position.

4. Slide the cylinder end (C) of the Trend-Like position gas spring (E) onto
the post (B), and install the split-ring (A) to secure it.

5. Check for proper Trend-Like position operation by activating the Trend-


Like position handle.

6. Ensure that the gas springs release when the handle is in the middle of its
travel.

7. Install the top cover over the drive system.

8. Do the “Function Checks” on page 2-4.

Page 4 - 30 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.11 Head Section Gas Spring and Damper (CPR Assist)
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.11 Head Section Gas Spring and Damper (CPR Assist)


Tools required: T25 Torx® head bit Ratchet
1

Needle nose pliers Screwdriver

Removal
1. Perform step 1 through step 5, step 8, and step 9 of “Head Drive Motor
Assembly” on page 4-15.

WARNING:
Do not attempt to remove the gas spring with the head section fully
raised and the spring compressed. Lower the head section until the
cylinder of the gas spring is free in the slide bracket of the main frame,
and support the head section securely before removing the cotter pins
and clevis pins. Failure to do so could result in personal injury or
equipment damage.

2. Lower the head section frame (A) until the cylinder end of the gas spring
(D) is free in the main frame slide bracket (F). Support the head section
frame securely (see figure 4-14 on page 4-32).

3. Using the needle nose pliers, remove the cotter pins (B) from the two clevis
pins (C) that secure the head section gas spring (D) and damper (E)
4
cylinder ends to the bed frame weldment (F).

4. Using the screwdriver, remove the two e-rings (G) that secure the head
section gas spring (D) and damper (E) rod ends to the head section frame
weldment (H).

5. Remove the head section gas spring (D) and damper (E) from the bed.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 31
Service Manual (MAN272 REV 3)
4.11 Head Section Gas Spring and Damper (CPR Assist)
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-14. Head Section Gas Spring and Damper (CPR Assist)

Replacement
1. To install the replacement head section gas spring (D) and damper (E),
reverse the removal procedure.

2. Do the “Function Checks” on page 2-4.

Page 4 - 32 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.12 Foot Support Module
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.12 Foot Support Module


Tools required: 5/32" Allen™ wrench 7/16" wrench
1

T25 Torx® head bit Ratchet


2

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Remove the foot section from the bed. Refer to “Foot Section (Lift Off)”
on page 4-40 or “Foot Section (Slide Off)” on page 4-42.

4. Using the Allen™ wrench, remove the set screw (A) that secures the foot
support module (B) to the yoke assembly (C) (see figure 4-15 on
page 4-33)

5. Remove the foot support module (B) from the yoke assembly (C).

Figure 4-15. Foot Support Module

1. Allen™ is a trademark of Industrial Fasteners, Inc.


2. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 33
Service Manual (MAN272 REV 3)
4.12 Foot Support Module
Chapter 4: Removal, Replacement, and Adjustment Procedures

Replacement
1. To install the replacement foot support module (B), reverse the removal
procedure.

2. Do the “Function Checks” on page 2-4.

Adjustment
1. Do one of the following:
a. Grasp the loops at the bottom of the bellows wireform (A), pull it open
and lift it off the foot support bellows (B) and the foot support heel cup
(C) (see figure 4-16 on page 4-35).
b. Remove the screw (I) that attaches the strap (A) to the foot support
bellows (B) and the foot support heel cup (C). Remove the strap (A).

2. Using the ratchet and the T25 Torx® head bit, remove the four screws (D)
1

attaching the foot support heel cup (C) to the foot rest weldment (E).

3. Remove the foot support heel cup (C) from the foot rest weldment (E) and
the foot support bellows (B).

4. Using the 7/16" wrench, adjust the cable tensioning nuts (F) so that
the cable ends ride freely in the mounting holes of the release handle
weldment (G) without binding.

5. Reassemble the foot support heel cup (C), Torx® head screws (D), and
bellows wireform (A) to the foot rest weldment (E) and the foot support
bellows (B).

6. Ensure that the foot support release handle (H) functions properly.

1. Torx® is a registered trademark of Textron, Inc.

Page 4 - 34 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.12 Foot Support Module
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-16. Disassembled Foot Support

Affinity® Three Birthing Bed

Affinity® Four Birthing Bed

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 35
Service Manual (MAN272 REV 3)
4.13 Night Light and Cable Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.13 Night Light and Cable Assembly


Tools required: T25 Torx® head bit Ratchet
1

Removal
1. Remove the top motor cover (refer to procedure 4.1).

2. Disconnect the night light and cable assembly (A) from logic control
P.C. board connector P1 (see figure 4-17 on page 4-36).

Figure 4-17. Night Light and Cable Assembly

3. Using a ratchet and T25 Torx® head bit, remove the two screws (B) that
secure the night light and cable assembly (A) to the bottom cover (C).

4. Remove the night light and cable assembly (A) from the bed.

Replacement
1. To install the replacement night light and cable assembly (A), reverse the
removal procedure.

2. Do the “Function Checks” on page 2-4.

1. Torx® is a registered trademark of Textron, Inc.

Page 4 - 36 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.14 Mattress Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.14 Mattress Assembly


Tools required: None

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Carefully lift the mattress from the head end of the bed.

CAUTION:
Do not pull on the mattress material when unfastening the mattress
retaining snaps. Unfasten the mattress retaining snaps at the snap
location. Failure to do so could result in equipment damage.

4. Locate and unfasten the two mattress retaining snaps beneath the seat
section.

5. If the mattress has air bladders, disconnect the two hoses located beneath
the seat section.

6. Lift the side panels on the foot section and unfasten the four snaps on the
foot mattress.
4
Replacement
1. To install the replacement mattress, reverse the removal procedure.

2. Ensure that the two air hoses are properly installed below the seat section.

3. Do the “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 37
Service Manual (MAN272 REV 3)
4.15 Labor Grips
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.15 Labor Grips


Tools required: T25 Torx® head bit Ratchet
1

Lubri-film spray (P/N 36958)


External snap-ring removal tool

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Remove the mattress to gain access to the seat panel. Refer to section the
“Mattress Assembly” on page 4-37.

4. Using a ratchet and T25 Torx® head bit, remove the screws (A) that secure
the seat panel (B) to the seat pan (C) (see figure 4-18 on page 4-38).

Figure 4-18. Labor Grips

1. Torx® is a registered trademark of Textron, Inc.

Page 4 - 38 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.15 Labor Grips
Chapter 4: Removal, Replacement, and Adjustment Procedures

5. Using the external snap-ring removal tool, remove the external snap ring
(D) from the pull handle.

6. Pull and hold the release (E) for the labor grip handle (F).

7. Slide the labor grip handle (F) from its socket.

Replacement
1. To install the new labor grip handle (F), reverse the removal procedure.

2. Apply lubri-film spray (P/N 36958) to the arm of the labor grip that slides
into the socket.

3. Do the “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 39
Service Manual (MAN272 REV 3)
4.16 Foot Section (Lift Off)
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.16 Foot Section (Lift Off)


Tools required: None

Removal
1. Pivot the foot mattress side panels (A) up (see figure 4-19 on page 4-41).

2. Grasp the mattress handles (B) or (G) located on both sides of the mattress.

3. Pull straight up and then toward yourself.

4. Stand the foot section up on the pull handle (C) base or on the perineal
stand.

Replacement

WARNING:
Failure to properly mount the foot section to the yoke latches could
result in equipment damage or personal injury.

1. Grasp the mattress handles (B) or (G) located on both sides of the mattress.

2. Align the foot section slides (D) with their receptacles (F) on the bed.

3. Lower the foot section (E) onto the bed.

4. Tug upward on the pull handle (C) to verify that the foot section is securely
mounted to the bed.

Page 4 - 40 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.16 Foot Section (Lift Off)
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-19. Foot Section (Lift Off)

Affinity® Three Birthing Bed

Affinity® Four Birthing Bed

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 41
Service Manual (MAN272 REV 3)
4.17 Foot Section (Slide Off)
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.17 Foot Section (Slide Off)


Tools required: None

Removal
1. Pivot the foot mattress side panels (A) up (see figure 4-19 on page 4-41).

2. Pull the foot section release handles (D) to release the latch mechanisms
(E) from the channel slides (F) (see figure 4-20 on page 4-42).

Figure 4-20. Foot Section (Slide Off)

3. Grasp the mattress handles (B) located on both sides of the mattress
(see figure 4-19 on page 4-41).

4. Slide the foot section straight out toward yourself.

NOTE:
The foot section pull handle doubles as a base to allow the foot section to stand
alone.

5. Stand the foot section up on the pull handle (C) base.

Page 4 - 42 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.17 Foot Section (Slide Off)
Chapter 4: Removal, Replacement, and Adjustment Procedures

Replacement

WARNING:
Failure to properly mount the foot section to the yoke slide brackets
could result in equipment damage or personal injury.

1. Grasp the mattress handles (B) located on both sides of the mattress
(see figure 4-19 on page 4-41).

2. Align the foot section channel slides (F) with the mounting brackets (G) on
the bed(see figure 4-20 on page 4-42).

3. Slide the foot section onto the bed.

4. Ensure that the foot section latch mechanism (E) locks into the channel
slide (F).

5. Tug out on the pull handle (C) to verify that the foot section is securely
mounted to the bed.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 43
Service Manual (MAN272 REV 3)
4.18 Line Voltage Fuses
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.18 Line Voltage Fuses


Tools required: Small screwdriver

Three mains fuses protect the high voltage circuit. Two fuses (line voltage
fuses) are located in the fuse holder secured to the head end of the main frame.
The third fuse is located on the air system control P.C. board.

WARNING:
Adhere to all electrical safety precautions when servicing the bed's
electrical system. Failure to do so could result in personal injury or
equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position.

4. Using the hilow function, raise the bed to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Remove the top motor cover (refer to procedure 4.1).

7. Using the screwdriver, remove the fuse housing (A) from the power entry
module (B) (see figure 4-21 on page 4-45).

8. Remove the appropriate line voltage fuse (C) from the fuse housing (A).

Page 4 - 44 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.18 Line Voltage Fuses
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-21. Fuse Location

4
Replacement

CAUTION:
Ensure that the fuse housing is properly oriented for the voltage rating
(110-120V or 220-240V) on the bed that you are servicing. Failure to do
so could result in equipment damage.

1. Orient the fuse housing properly for the voltage rating (110-220V or
220-240V) on the bed you are servicing.

2. Insert a serviceable line voltage fuse (C) into the fuse housing (A).

3. To install the replacement line voltage fuse (C), reverse the removal
procedures.

4. Do the “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 45
Service Manual (MAN272 REV 3)
4.19 Battery Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.19 Battery Assembly


Tools required: T25 Torx® head bit Ratchet
1

VOM Small screwdriver

WARNING:
Adhere to all electrical safety precautions when servicing the bed's
electrical system. Failure to do so could result in personal injury or
equipment damage.

Batteries

Removal

1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position.

4. Using the hilow function, raise the bed to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Remove the top motor cover (refer to procedure 4.1).

CAUTION:
Do not cut or remove the cable ties that secure the battery leads to the
electronics pan. This ensures the proper connection of the batteries
during the replacement procedure. Possible equipment damage could
occur if the cable ties are removed.

7. Unplug the battery cable assembly (A) from logic control P.C. board
connector P5 (see figure 4-22 on page 4-47).

8. Unplug the battery cable assemblies (A and B) from the four connection
points (C) on the two batteries (D).

1. Torx® is a registered trademark of Textron, Inc.

Page 4 - 46 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.19 Battery Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-22. Battery Assembly

9. Using a ratchet and T25 Torx® head bit, remove the screw (E) that secures
1 4
the battery holder bracket (F) to the electronics plate weldment (G).

10. Remove the two batteries (D) from the electronics plate weldment (G).

Replacement

1. To install the replacement batteries (D), reverse the removal procedures.

2. Do the “Function Checks” on page 2-4.

Battery Fuse

Removal

1. Carefully pull the battery fuse housing (H) from the electronics plate
weldment (G).

2. Remove the battery fuse (I) from the battery fuse housing (H).

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 47
Service Manual (MAN272 REV 3)
4.19 Battery Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Replacement

1. To install the replacement battery fuse (I), reverse the removal procedures.

2. Do the “Function Checks” on page 2-4.

Battery Charge Circuit


Beds equipped with a 65345 logic control P.C. board and 67208 interface P.C.
board may require adjustment of the battery charge circuit for optimal battery
function. Beds equipped with the 68121 logic control P.C. board will not
require adjustment of the battery charge circuit.

WARNING:
Adhere to all electrical safety precautions when servicing the bed's
electrical system. Failure to do so could result in personal injury or
equipment damage.

Adjustment

1. Unplug the bed from its power source.

2. Unplug the battery cable assembly (A) from logic control P.C. board
connector P5 (see figure 4-22 on page 4-47).

3. Set your VOM to measure V DC.

4. Connect the ground lead of the VOM to the negative side of C1 (J)
(see figure 4-23 on page 4-49).

5. Connect the positive lead of the VOM to the positive side of D15 (K).

NOTE:
The positive side of D15 is the side without a silver band.

6. Reconnect the bed to an appropriate power source.

7. Using the small screwdriver, turn the screw on potentiometer R2 (L) until
the VOM reads between 27.6 and 28.0 V DC.

8. Unplug the bed from its power source.

9. Reconnect the battery cable assembly to P5.

10. Do the “Function Checks” on page 2-4.

Page 4 - 48 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.19 Battery Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-23. Logic Control P.C. Board—P/N 65345

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 49
Service Manual (MAN272 REV 3)
4.20 Bed Surface Air System Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.20 Bed Surface Air System Assembly


Tools required: Small wire cutters (2) ½" wrenches

NOTE:
When servicing the air system, it is important to ensure that the lumbar and
seat inflate/deflate hoses are connected to the correct fittings on the air
package. The hose from the left side of the bed is the lumbar hose; the hose
from the right side is the seat hose.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the hilow function, raise the bed to the high position.

4. Using the head function, raise the head section to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Remove the top motor cover (refer to procedure 4.1).

7. Remove the bottom motor cover (refer to procedure 4.2).

8. Disconnect the hose (A) leading from the right side of the mattress to the
air manifold fitting (B) (see figure 4-24 on page 4-51).

9. Disconnect the hose (C) leading from the left side of the mattress to the air
manifold fitting (D).

Page 4 - 50 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.20 Bed Surface Air System Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-24. Bed Surface Air System Assembly

10. Disconnect the transformer assembly power cable from air system control
4
P.C. board (F) connector P19 (E) (see figure 4-25 on page 4-52).

11. Disconnect the cable (G) from logic control P.C. board connector P8 to air
system control P.C. board (F) connector P18 (H).

12. Using the small wire cutters, cut and remove the cable tie (I) that secures
the linear pump power cable (J), air system power cable, and air system
control P.C. board cable (G) to the mounting bracket (K).

13. Using the two ½" wrenches, remove the bolts (L) and nuts (M) that secure
the air system to the bed frame (see figure 4-24 on page 4-51).

14. Remove the air system from the bed.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 51
Service Manual (MAN272 REV 3)
4.20 Bed Surface Air System Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-25. Bed Surface Air System Cable Connections

Replacement
1. To install the replacement air system, reverse the removal procedure.

2. Ensure that all air system functions work properly.

3. Do the “Function Checks” on page 2-4.

Page 4 - 52 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.21 Air System Manifold Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.21 Air System Manifold Assembly


Tools required: Phillips head screwdriver

Removal
1. Remove the air system assembly (refer to procedure 4.20).

2. Disconnect the air hoses from the manifold assembly (A) (see figure 4-26
on page 4-53).

Figure 4-26. Air System Manifold Assembly

3. Identify and disconnect the wiring connectors (B) from the manifold
assembly (A).

4. Using the phillips head screwdriver, remove the two screws (C) that secure
the manifold assembly (A) to the mounting plate (D).

Replacement
1. To install the replacement manifold assembly (A), reverse the removal
procedure.

2. Do the “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 53
Service Manual (MAN272 REV 3)
4.22 Air System Control P.C. Board Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.22 Air System Control P.C. Board Assembly


Tools required: Phillips head screwdriver Needle nose pliers

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the hilow function, raise the bed to the high position.

4. Using the head function, raise the head section to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source.

6. Remove the bottom motor cover (refer to procedure 4.2).

7. Identify and disconnect all wiring connectors (A) from the air system
control P.C. board (B) (see figure 4-27 on page 4-55).

8. Using the phillips head screwdriver, remove the screw (C) that secures the
air system control board (B) to the air system bracket (D).

9. Using the needle nose pliers, compress the air system control P.C. board
(B) mounting standoffs (E).

CAUTION:
To prevent component damage, ensure that your hands are clean, and
only handle a P.C. board by its edges. Failure to do so could result in
equipment damage.

CAUTION:
For shipping and storage, place the removed P.C. board in an antistatic
protective bag. Failure to do so could result in equipment damage.

10. Handling it by its edges with clean hands, maneuver and remove the air
system control P.C. board (B) from the bed. Place it in an antistatic
protective bag.

Page 4 - 54 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.22 Air System Control P.C. Board Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-27. Air System Control P.C. Board

Replacement
4
1. To install the replacement air system control P.C. board (B), reverse the
removal procedures.

2. Ensure that the wiring connectors are firmly connected to the air system
control P.C. board receptacles.

3. Do the “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 55
Service Manual (MAN272 REV 3)
4.23 Logic Control P.C. Board Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.23 Logic Control P.C. Board Assembly


Tools required: Needle nose pliers Isopropyl alcohol

SHOCK HAZARD:
The voltage in the electrical system presents an electrical shock
hazard. Perform standard electrical service procedures before
attempting service within the P.C. board enclosure. Adhere to all
electrical safety precautions when servicing the bed’s electrical system.
Failure to do so could result in personal injury or equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the hilow function, raise the bed to the high position.

NOTE:
If the hilow function is inoperable, skip step 3.

4. Using the head function, raise the head section to the high position. If the
head function is inoperable, perform the following step:
a. Pull the CPR handle, lift the head section far enough to gain access to
the motor cover, and then release the CPR handle.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Remove the top motor cover (refer to procedure 4.1).

7. Identify and disconnect the wiring connectors from the logic control P.C.
board (A), beginning with the battery cable connection P5 (see figure 4-28
on page 4-57).

Page 4 - 56 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.23 Logic Control P.C. Board Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

CAUTION:
To prevent component damage, ensure that your hands are clean, and
only handle a P.C. board by its edges. Failure to do so could result in
equipment damage.

8. Using the needle nose pliers, compress the locking tabs on the five logic
control P.C. board mounting standoffs (B), pulling the logic control P.C.
board (A) slightly up and over the locking tab on each mounting standoff.

Figure 4-28. Logic Control P.C. Board Assembly

CAUTION:
For shipping and storage, place the removed P.C. board in an antistatic
protective bag. Failure to do so could result in equipment damage.

9. Handling the logic control P.C. board (A) by its edges with clean hands,
maneuver and remove it from the electronics plate weldment (C).
a. If replacing a 65345 logic control P.C. board and 67208 interface P.C.
board, discard them.
b. If replacing a 68121 logic control P.C. board, place it in an antistatic
protective bag.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 57
Service Manual (MAN272 REV 3)
4.23 Logic Control P.C. Board Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Modifying the Lockout Switch Cable


When replacing a 65345 logic control P.C. board and 67208 interface P.C.
board with a 68121 logic control P.C. board, the lockout switch cable must be
modified as follows:

1. Locate the P3 connector of the lockout switch cable. When used with a
65345 logic control P.C. board and 67208 interface P.C. board, the order of
the four wires in the connector housing is red, red adjacent, and black,
black adjacent (see figure 4-29 on page 4-58). Pin 1 may be either red or
black.

NOTE:
If the order of the four wires in the connector housing is red, black, red, black
alternating, the lockout switch cable has already been modified for use with a
68121 logic control P.C. board. Skip to “Replacement” on page 60.

2. Using the needle nose pliers, compress the locking tabs (D) holding the
connector shell (E) and the backshell (F) of the plastic connector housing
together, and pull the two shells about ¼" apart. This procedure unlocks the
pins in the housing.
Figure 4-29. Lockout Switch Cable P3 Connector

Page 4 - 58 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.23 Logic Control P.C. Board Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

3. Pull the red middle wire (G) with its attached pin, and the black middle
wire (H) with its attached pin, out of the backshell (F).

4. Reverse the order of the two middle wires and pins, and reinstall them into
the backshell, pushing on each wire until you hear a click and the pin is
seated tightly. The order of the four wires in the connector housing should
now be red, black, red, black alternating, for use with a 68121 logic control
P.C. board. Pin 1 may be either red or black.

5. Push the connector shell (E) and the backshell (F) together until you hear a
click and the locking tabs (D) are locked, holding the two parts of the
connector housing firmly together.

6. Clean the underside of the top motor cover with a cloth soaked in isopropyl
alcohol, remove the protective backing from the 6399301 wiring diagram
(I), and install the new wiring diagram in place of the existing wiring
diagram located on the underside of the top motor cover (J) (see figure 4­
30 on page 4-59).

NOTE:
It is not necessary to remove the old wiring diagram, as long as the new one
covers it completely.

Figure 4-30. Logic Control P.C. Board Wiring Diagram 4

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 59
Service Manual (MAN272 REV 3)
4.23 Logic Control P.C. Board Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Replacement
1. To install the replacement logic control P.C. board (A), reverse the removal
procedures, finishing with the battery cable connection P5.

2. Ensure that the wiring connectors are firmly connected to the logic control
P.C. board (A) receptacles.

3. Do the “Function Checks” on page 2-4.

Page 4 - 60 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.24 Transformer Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.24 Transformer Assembly


Tools required: T25 Torx® head bit Ratchet
1

SHOCK HAZARD:
The voltage in the electrical system presents an electrical shock
hazard. Perform standard electrical service procedures before
attempting service within the P.C board enclosure. Adhere to all
electrical safety precautions when servicing the bed’s electrical system.
Failure to do so could result in personal injury or equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position.

4. Using the hilow function, raise the bed to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in 4
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Remove the top motor cover (refer to procedure 4.1).

7. Identify and disconnect the wiring connectors from the transformer


assembly (A) (see figure 4-31 on page 4-62).

8. Using the ratchet and T25 Torx® head bit, remove the four Torx® screws
(B) that secure the transformer assembly (A) to the bed frame (C).

9. Maneuver and remove the transformer assembly (A) from the bed.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 61
Service Manual (MAN272 REV 3)
4.24 Transformer Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-31. Transformer Assembly

Replacement
1. To install the replacement transformer assembly (A), reverse the removal
procedures.

2. Ensure that the wiring connectors are firmly connected to the transformer
assembly (A).

3. Do the “Function Checks” on page 2-4.

Page 4 - 62 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.25 Auxiliary Transformer Assembly (Affinity® Three Bed)
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.25 Auxiliary Transformer Assembly (Affinity® Three Bed)


Tools required: T25 Torx® head bit Ratchet
1

SHOCK HAZARD:
The voltage in the electrical system presents an electrical shock
hazard. Perform standard electrical service procedures before
attempting service within the P.C. board enclosure. Adhere to all
electrical safety precautions when servicing the bed’s electrical system.
Failure to do so could result in personal injury or equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position.

4. Using the hilow function, raise the bed to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in 4
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Remove the top motor cover (refer to procedure 4.1).

7. Identify and disconnect the wiring connectors (A) from the auxiliary
transformer assembly (B) (see figure 4-32 on page 4-64).

8. Using the ratchet and T25 Torx® head bit, remove the two Torx® screws
(C) that secure the auxiliary transformer assembly (B) to the bed frame (D).

9. Maneuver and remove the auxiliary transformer assembly (B) from the
bed.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 63
Service Manual (MAN272 REV 3)
4.25 Auxiliary Transformer Assembly (Affinity® Three Bed)
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-32. Auxiliary Transformer Assembly

Replacement
1. To install the replacement auxiliary transformer assembly (B), reverse the
removal procedures.

2. Ensure that the wiring connectors (A) are firmly connected to the auxiliary
transformer assembly (B).

3. Do the “Function Checks” on page 2-4.

Page 4 - 64 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.26 Lockout Switch Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.26 Lockout Switch Assembly


Tools required: None

SHOCK HAZARD:
The voltage in the electrical system presents an electrical shock
hazard. Perform standard electrical service procedures before
attempting service within the P.C. board enclosure. Adhere to all
electrical safety precautions when servicing the bed’s electrical system.
Failure to do so could result in personal injury or equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position.

4. Using the hilow function, raise the bed to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in 4
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Remove the top motor cover (refer to procedure 4.1).

7. Identify and disconnect the wiring connector (A) from the lockout switch
assembly (B) (see figure 4-33 on page 4-66).

8. Compress the tabs (C) on the lockout switch assembly (B).

9. Maneuver and remove the lockout switch assembly (B) from the bed.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 65
Service Manual (MAN272 REV 3)
4.26 Lockout Switch Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-33. Lockout Switch Assembly

Replacement
1. To install the replacement lockout switch assembly (B), reverse the
removal procedures.

2. Ensure that the wiring connector (A) is firmly connected to the lockout
switch assembly (B).

3. Do the “Function Checks” on page 2-4.

Page 4 - 66 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.27 Siderail Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.27 Siderail Assembly


Tools required: Small screwdriver Ratchet
Needle nose pliers 7/16" socket

WARNING:
Adhere to all electrical safety precautions when servicing the bed’s
electrical system. Failure to do so could result in personal injury or
equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in

4
personal injury or equipment damage.

4. Unplug the bed from its power source.

5. Disconnect the siderail cable (A) from the cable (B) to the logic control
P.C. board (see figure 4-34 on page 4-68).

6. Disconnect the ground strap (H).

CAUTION:
Support the siderail during the removal procedure. Failure to do so
could result in equipment damage.

7. Using the needle nose pliers or a small screwdriver, remove the four
e-rings (C) that secure the siderail assembly (D) to the head section.

8. Using a ratchet and 7/16" socket, remove the two nuts (E) and bolts (G)
that secure the siderail mounting bracket (F) to the head section.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 67
Service Manual (MAN272 REV 3)
4.27 Siderail Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-34. Siderail Assembly

CAUTION:
Ensure that the siderail does not drop when the pins are removed.
Failure to do so could result in damage to the wiring going to the
siderail.

9. Remove the pins (I) from the siderail assembly (D).

10. Set the siderail in a place where it will not be damaged.

Replacement
1. To install the replacement siderail assembly (D), reverse the removal
procedure.

2. Do the “Function Checks” on page 2-4.

Page 4 - 68 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.28 Patient Siderail (Inboard) P.C. Switch Board
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.28 Patient Siderail (Inboard) P.C. Switch Board


Tools required: T25 Torx® head bit Ratchet
1

T10 Torx® head bit

WARNING:
Adhere to all electrical safety precautions when servicing the bed’s
electrical system. Failure to do so could result in personal injury or
equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position.

4. Using the hilow function, raise the bed to the high position.

4
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source, and engage the lockout control.

6. Using the ratchet and T25 Torx® head bit, remove the eight plastic screws
(A) that secure the patient siderail cover (B) to the siderail assembly (C)
(see figure 4-35 on page 4-70).

7. Carefully remove the patient siderail cover (B) from the siderail
assembly (C).

8. Note the orientation of the wires and cable assemblies to ensure proper
installation during the replacement procedure.

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 69
Service Manual (MAN272 REV 3)
4.28 Patient Siderail (Inboard) P.C. Switch Board
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-35. Patient Siderail (Inboard) P.C. Switch Board

9. Using the ratchet and T10 Torx® head bit, remove the four screws (D) that
1

secure the P.C. switch board (E) to the patient siderail cover (B).

10. Slide the P.C. switch board (E) off the patient siderail cover (B) tabs, and
remove it from the patient siderail cover (B).

11. Disconnect the two SideCom® Communication System cable assemblies


(F) from the P.C. switch board (E).

Replacement
1. To install the replacement P.C. switch board (E), reverse the removal
procedures.

2. Do the “Function Checks” on page 2-4.

1. Torx® is a registered trademark of Textron, Inc.

Page 4 - 70 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.29 Caregiver Siderail (Outboard) P.C. Switch Board
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.29 Caregiver Siderail (Outboard) P.C. Switch Board


Tools required: T25 Torx® head bit Ratchet
1

T10 Torx® head bit

WARNING:
Adhere to all electrical safety precautions when servicing the bed’s
electrical system. Failure to do so could result in personal injury or
equipment damage.

Removal
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the head function, raise the head section to the high position.

4. Using the hilow function, raise the bed to the high position.

4
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

5. Unplug the bed from its power source.

6. Using the ratchet and T25 Torx® head bit, remove the eight plastic screws
(A) that secure the patient siderail cover (B) to the siderail assembly (C)
(see figure 4-36 on page 4-72).

7. Carefully separate the patient siderail cover (B) from the siderail
assembly (C).

8. Note the orientation of the wires and cable assemblies to ensure proper
installation during the replacement procedure.

9. Using the ratchet and T25 Torx® head bit, remove the three screws (G)
that secure the caregiver siderail cover (H) to the siderail assembly (C).

10. Carefully remove the caregiver siderail cover from the siderail
assembly (C).

1. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 71
Service Manual (MAN272 REV 3)
4.29 Caregiver Siderail (Outboard) P.C. Switch Board
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-36. Caregiver Siderail (Outboard) P.C. Switch Board

11. Note the orientation of the wires and cable assemblies to ensure proper
installation during the replacement procedure.

12. Using the ratchet and T10 Torx® head bit, remove the four screws (D) that
1

secure the P.C. switch board (E) to the caregiver siderail cover (H).

13. Slide the P.C. switch board (E) off the caregiver siderail cover (H) tabs,
and remove it from the caregiver siderail cover (H).

14. Disconnect the two SideCom® Communication System cable assemblies


(F) from the P.C. switch board (E).

Replacement
1. To install the replacement P.C. switch board (E), reverse the removal
procedures.

2. Do the “Function Checks” on page 2-4.

1. Torx® is a registered trademark of Textron, Inc.

Page 4 - 72 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.30 Central Braking System—P3700A Model Only
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.30 Central Braking System—P3700A Model Only


Tools required: Phillips head screwdriver 1/8" drift punch
7/16" wrench Blue Loctite® #242
1

Hammer ½" wrench


E-ring removal tool Anti-seize compound
Needle nose pliers Marker pen
(2) pieces of 2" x 4" x 36" lumber
Small screwdriver
Red Loctite® #262

The brake/steer caster is located at the left-hand foot end of the base frame.
The brake caster is located at the right-hand head end of the base frame. To
service the central braking system mechanism, lay the bed on its side. This
requires the siderail to be secured and the brake/steer pedal to be removed.

Removal
1. Determine on which side the bed is to be placed.

2. Secure the siderail by lowering it beneath the bed and tying it in this
position, or removing it as an assembly from the bed (refer to procedure
4.27).

3. Place the brake/steer (foot) pedal (A) into the steer position
4
(see figure 4-37 on page 4-74).

4. Using a phillips head screwdriver, remove the two screws (B) securing the
leg covers (C) over the brake/steer and brake casters.

5. Remove the leg covers (C).

6. Remove the rocker arm (D) located above either the brake/steer caster or
the brake caster by performing the following steps:

1. Loctite® is a registered trademark of Loctite Corporation.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 73
Service Manual (MAN272 REV 3)
4.30 Central Braking System—P3700A Model Only
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-37. Central Braking System

Page 4 - 74 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.30 Central Braking System—P3700A Model Only
Chapter 4: Removal, Replacement, and Adjustment Procedures

a. Remove the e-ring (E) from one side of the groove pin (F) securing the
rocker arm (D) above the caster.
b. Remove the groove pin (F) to free the rocker arm (D).
c. Remove the spacers (X) on each side of the rocker arm (D).
d. Remove the hairpin (G) and connector pin (H) that secures the metal
brake strip (I) to the rocker arm (D).
e. Lift out and remove the rocker arm (D).

7. Repeat step 6 for the remaining rocker arm.

8. Place the brake/steer pedal (A) into the neutral position.

9. Remove the headboard and the foot section from the bed.

10. Place the pads on the floor to protect the bed from damage.

11. Carefully turn the bed onto the side that has the siderail assembly secured.

12. Using a phillips head screwdriver, remove the two screws (J) and the end
caps (K) covering the cross tube (behind the brake/steer pedal) at both
sides of the base frame.

13. Using a phillips head screwdriver, remove the two screws (L) from the
4
brake rod support bracket (M) on each side of the base frame.

14. Using a phillips head screwdriver, remove the four screws (N) securing the
detent housing plate (O) at the center of the cross channel.

15. Remove the two hairpins (P) and clevis pins (Q) that connect the shaft of
both the brake/steer pedals to the detent mechanism assembly (R).

16. Rotate either brake/steer pedal (A) to maneuver the cam portion (Z) of its
shaft (U) from the cam pivot (W).

17. Pull the brake/steer pedal shaft (U) free of the cross channel and detent
mechanism assembly (R).

18. Repeat step 16 and step 17 for the opposite side.

19. Remove the detent mechanism assembly (R) from the cross channel.

20. Slide the brake rod support bracket (M) and bushing (T) from the end of
the shaft (U).

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 75
Service Manual (MAN272 REV 3)
4.30 Central Braking System—P3700A Model Only
Chapter 4: Removal, Replacement, and Adjustment Procedures

21. Remove the brake strips (I) from the base frame.

22. Using a phillips head screwdriver, remove the two screws (Y) and
locknuts (V) that fasten each brake strip (I) to its cam pivot (W).

Replacement
1. To install the replacement central braking system mechanism, reverse the
removal procedure with attention to the following:
a. Ensure that the brake/steer strip (straight metal strap) and the brake
strip (bent metal strap) are correctly assembled on their respective
sides.
b. Ensure that each brake strip (I) is correctly attached to its cam
pivot (W).
c. Complete sections “Brake and Steer Caster Adjustment” on page 4-81
and “Brake Caster Adjustment” on page 4-82.

2. Ensure that all brake and steer functions work properly.

3. Do the “Function Checks” on page 2-4.

Page 4 - 76 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.31 Central Braking System—P3700B Model Only
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.31 Central Braking System—P3700B Model Only


Tools required: Phillips head screwdriver 1/8" drift punch
7/16" wrench Blue Loctite® #242
1

Hammer ½" wrench


E-ring removal tool Anti-seize compound
Needle nose pliers Marker pen
(2) pieces of 2" x 4" x 36" lumber
Small screwdriver T25 Torx® screwdriver
2

Red Loctite® #262

The brake/steer caster is located at the left-hand foot end of the base frame.
The other three casters are brake casters. To service the central braking system
mechanism, lay the bed on its side. This requires the siderail to be secured and
the brake/steer pedal to be removed.

Removal
1. Determine on which side the bed is to be placed.

2. Secure the siderail by lowering it beneath the bed and tying it in this
position, or removing it as an assembly from the bed (refer to procedure
4.27).

3. Place the brake/steer pedal (A) into the steer position (see figure 4-38 on
4
page 4-78).

4. Remove the two screws (B) securing the leg covers (C) to the bed (D).

5. Remove the leg covers (C).

6. Remove the foot-end rocker arm (E) as follows:


a. Remove the e-ring (F) from one side of the groove pin (G) securing the
rocker arm (E) above the caster.
b. Remove the groove pin (G) to free the rocker arm (E).
c. Remove the spacers (H) on each side of the rocker arm (E).
d. Remove the hairpin (I) and connector pin (J) that secures the metal
brake strip (K) to the rocker arm (E).
e. Lift out and remove the rocker arm (E).

1. Loctite® is a registered trademark of Loctite Corporation.


2. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 77
Service Manual (MAN272 REV 3)
4.31 Central Braking System—P3700B Model Only
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-38. Central Braking System

7. Repeat step 6 for the head-end rocker arm.

8. Place the brake/steer pedal (A) into the neutral position.

9. Remove the headboard and the foot section from the bed.

10. Place the pads on the floor to protect the bed from damage.

11. Carefully turn the bed onto the side that has the siderail assembly secured.

12. Remove the two screws (L) and the end caps (M) covering the cross tube
(behind the brake/steer pedal) at both sides of the base frame.

13. Remove the two screws (N) from the brake rod support bracket (O) on each
side of the base frame.

14. Remove the four screws (P) securing the detent housing plate (Q) at the
center of the cross channel.

15. Remove the two hairpins (R) and clevis pins (S) that connect the shaft of
both the brake/steer pedals to the detent mechanism assembly (T).

Page 4 - 78 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.31 Central Braking System—P3700B Model Only
Chapter 4: Removal, Replacement, and Adjustment Procedures

16. Pull the brake/steer pedal (A) free of the cross channel and detent
mechanism assembly (T).

17. Repeat step 3 through step 17 for the opposite side.

18. Remove the detent mechanism assembly (T) from the cross channel.

19. Slide the brake rod support bracket (O) and bushing (U) from the end of the
brake/steer pedal (A).

20. Remove the brake strips (K and V) from the base frame.

Replacement
1. To install the replacement central braking system mechanism, reverse the
removal procedure with attention to the following:
a. Ensure that the brake/steer strip (straight metal strap) and the brake
strip (bent metal strap) are correctly assembled on their respective
sides.
b. Ensure that each brake stripS (K and V) are correctly attached to its
cam pivot.
c. Complete sections “Brake and Steer Caster Adjustment” on page 4-81
and “Brake Caster Adjustment” on page 4-82.
4
2. Ensure that all brake and steer functions work properly.

3. Do the “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 79
Service Manual (MAN272 REV 3)
4.32 Caster Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.32 Caster Assembly


Tools required: T25 Torx® head bit 5/32" Allen™ wrench
1 2

Ratchet 3/16" Allen™ wrench


Bed jack Turpentine

P3700A—The bed has two casters that have a braking feature. One of these
casters also has a steer feature. The brake/steer caster is located at the left-hand
foot end of the bed (as determined by lying in the bed on one's back). The
brake (only) caster is located at the right-hand head end of the bed. These two
casters may require adjustment after service, or because of normal wear. The
other two casters are basic swivel casters which require no adjustments.

P3700B—The foot-left caster is a brake/steer caster and the remaining casters


are brake casters.

Placing the brake/steer pedal in steer locks the left-hand foot end caster into a
position that is parallel to the bed (as described in chapter 1).

Removal
1. Set the brake/steer pedal to the brake position.

2. Using the ratchet and T25 Torx® head bit, remove the two screws (A)
securing the leg cover (B) over the caster (C) (see figure 4-39 on
page 4-81).

3. Remove the leg cover (B).

4. Using the 3/16" Allen ™ wrench, remove the caster mounting screw (E).

5. Place a jack under the base frame, and raise the base frame high enough to
remove the caster (C).

WARNING:
Ensure that the bed is stable before removing the caster. Failure to do
so could result in personal injury or equipment damage.

6. Remove the caster (C) from the bed.

1. Torx® is a registered trademark of Textron, Inc.


2. Allen™ is a trademark of Industrial Fasteners, Inc.

Page 4 - 80 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.32 Caster Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-39. Caster Assembly

Replacement
4
1. To install the replacement caster (C), reverse the removal procedure.

2. Ensure that the caster (C) rolls and swivels freely, and that all brake and
steer functions work properly.

Brake and Steer Caster Adjustment


1. Set the brake/steer pedal to place the system in neutral.

2. Perform step 1 through step 3 in “Removal” on page 4-80.

3. Using the 5/32" Allen™ wrench, turn the setscrew (D) clockwise while
swiveling the caster (C) about 2" from side to side, until you hear a light
ratcheting sound.

4. Turn the setscrew (D) counterclockwise until the ratcheting stops.

5. Set the brake/steer pedal to place the system in the brake position.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 81
Service Manual (MAN272 REV 3)
4.32 Caster Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures

6. Ensure that the caster (C) tire does not roll.

7. If the caster (C) tire rolls, tighten the setscrew (D) slightly.

8. If, after performing step 7, the brake still seems to slip, check the caster (C)
tire for wax buildup. Clean the tire with turpentine if necessary.

9. Set the brake/steer pedal to place the system in steer.

10. Swivel the brake/steer caster (C).

11. The caster (C) should lock into a position parallel with the length of the
bed.

12. Set the brake/steer pedal in the neutral position.

13. Rotate the brake/steer caster (C) 180°. The caster should again lock into a
position parallel with the length of the bed.

14. Set the brake/steer pedal to place the system in neutral.

15. Swivel the brake/steer caster (C) from side to side, while checking for signs
of ratcheting.

16. If necessary, loosen the setscrew (D) to allow the caster (C) to swivel
without ratcheting, but so that it can still be locked in the steer position.

Brake Caster Adjustment


1. Perform step 1 through step 8 in “Brake and Steer Caster Adjustment” on
page 4-81.

Page 4 - 82 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.33 Stow and Go™ Foot Section Wire Form
Chapter 4: Removal, Replacement, and Adjustment Procedures

4.33 Stow and Go™ Foot Section Wire Form


Tools required:

Removal
1. Set the brakes.

2. Raise the bed to the full up position.

3. Raise the foot section to the full up position.

WARNING:
Failure to unplug the bed can cause personal injury or equipment
damage.

4. Unplug the bed from its power source.

5. Remove the placenta basin (A) from the foot section (B) (see figure 4-40
on page 4-84).

6. Remove the foot section (B).

7. Remove the two screws (C) that secure the wire form (D) to the foot
section weldment (E).
4
8. Remove the two nuts (F) from the bolts (G).

9. Support the weight of the wire form with one hand.

10. Remove one bolt (G).

11. Remove the other bolt (G).

12. Remove the wire form (D) from the bed.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 83
Service Manual (MAN272 REV 3)
4.33 Stow and Go™ Foot Section Wire Form
Chapter 4: Removal, Replacement, and Adjustment Procedures

Figure 4-40. Wire Form Removal

Page 4 - 84 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
4.33 Stow and Go™ Foot Section Wire Form
Chapter 4: Removal, Replacement, and Adjustment Procedures

Replacement
1. Do the “Removal” procedure in reverse order.

2. Do the “Function Checks” on page 2-4.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 4 - 85
Service Manual (MAN272 REV 3)
4.33 Stow and Go™ Foot Section Wire Form
Chapter 4: Removal, Replacement, and Adjustment Procedures

NOTES:

Page 4 - 86 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Chapter 5
Parts List

Service Parts Ordering


Using the parts lists in this manual, identify the part number(s) you require.
Find the product number and serial number on the product identification label
(A) (see figure 5-1 on page 5-1).
Figure 5-1. Product Identification Label Location

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 1
Service Manual (MAN272 REV 3)
Service Parts Ordering
Chapter 5: Parts List

Call Hill-Rom Technical Support at (800) 445-3720 with the following


information:
• Six-digit customer account number
• Purchase order number
• Product number
• Serial number
• Part number(s)

To promptly order parts, request part prices and availability, or follow up on a


service order, use the following Hill-Rom fax number:

(812) 934-8472

Terms:
• Net 30 days
• F.O.B. Batesville, IN
• Prepaid shipping charges added to invoice
• All orders shipped UPS ground unless specified

Address all inquiries to:


ATTN TECHNICAL SUPPORT—PARTS
HILL-ROM COMPANY, INC.
1069 STATE ROUTE 46 E
BATESVILLE IN 47006-9167

Address all return goods to:


ATTN SERVICE STORES
DISTRIBUTION CENTER DOOR D23
HILL-ROM COMPANY, INC.
COUNTY ROAD 300E
BATESVILLE IN 47006-9167

NOTE:
To eliminate possible delays or incorrect billings, do not return any items
without a Return Material Authorization (RMA) number. When a return is
requested, an RMA packet is included with each order. This packet includes an
RMA number, instructions, and a shipping label. If an RMA number is not
available, obtain one by phoning Hill-Rom Technical Support at
(800) 445-3720.

Page 5 - 2 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Exchange Policy
Chapter 5: Parts List

Exchange Policy
The following are policies for in-warranty and out-of-warranty exchanges from
Hill-Rom.

In-Warranty Exchanges
In some cases, Hill-Rom will request that parts/products be returned for
inspection. When this occurs, you are expected to return parts/products within
30 days of receipt of the exchange part. If you fail to return the inoperative
parts/products within the 30 day period, Hill-Rom will invoice your facility for
the full selling price of the parts/products.

NOTE:
The preceding billing procedure pertains only to parts/products that Hill-Rom
requests to be returned.

In some cases, the invoice accompanying the parts will show the full selling
price (only for internal use at Hill-Rom). Do not confuse this price with your
price.

Do not return any parts without an RMA number. When parts/products have
been requested to be returned, Hill-Rom will include an RMA packet with the
parts/products shipment. If an RMA number is not available, obtain one by
phoning Hill-Rom Technical Support at (800) 445-3720.

Out-of-Warranty Exchanges

5
You are expected to return the inoperative parts/products within 30 days of
receipt of the exchange part. Hill-Rom will include an RMA packet with the
parts/products shipment. If an RMA number is not available, obtain one by
phoning Hill-Rom Technical Support at (800) 445-3720. Hill-Rom will
invoice your facility for the full selling price of the parts/products. Upon return
of the inoperative parts/products, Hill-Rom will issue a credit to your facility
for the difference between the exchange price and the full selling price of
the parts/products.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 3
Service Manual (MAN272 REV 3)
Exchange Policy
Chapter 5: Parts List

NOTES:

Page 5 - 4 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Warranty
Chapter 5: Parts List

Warranty
HILL-ROM, INC.
LIMITED WARRANTY
Hill-Rom, Inc. (Hill-Rom) has a long tradition of providing superior products and service to our customer. Our goal is “Total
Customer Satisfaction”. In that spirit, Hill-Rom is proud to offer the following warranty.
GENERAL WARRANTY (APPLICABLE UNLESS A SPECIFIC WARRANTY IS LISTED)
Hill-Rom warrants to the original purchaser that its products and replacement parts shall be free from defects in material and
workmanship for a period of one (1) year from date of delivery. Hill-Rom’s obligation under this warranty is expressly
limited to supplying replacement parts and/or service for, or replacing, at its option, any product which is, in the sole
discretion of Hill-Rom, found to be defective. In addition to the foregoing one year warranty, Hill-Rom warrants to the
original purchaser that the frame and welds on its products will be free from structural defects for the life of the product. Any
product upgrade or modification initiated by Hill-Rom does not affect the original product warranty.
SPECIFIC WARRANTIES
MATTRESS WARRANTIES
Hill-Rom warrants to the original purchaser that its mattress product shall be free from defects in material and workmanship
for a period of two (2) years from date of delivery. However, electro mechanical mattress components (compressors, valves,
printed circuit boards, hoses, and couplers) are covered by the general one (1) year warranty.
EXPENDABLES WARRANTIES
A sixty (60) day limited warranty from date of delivery applies to expendable parts such as cushions, coverlets, software
diskettes, locator badge batteries, dome light incandescent bulbs, overhead fluorescent tubes, heating elements, temperature
probes, filter sheets, and microspheres. This warranty is limited to replacement of the parts covered.
TO OBTAIN PARTS AND SERVICE
In the United States, call Hill-Rom Technical Support Department at (800) 445-3720, Monday through Friday. In Canada,
call Hill-Rom Technical Support Department at (800) 267-2337, Monday through Friday. Outside the United States and
Canada, call your authorized Hill-Rom Distributor. In order to expedite service, we request you furnish the following
information: customer identification number, product model number, serial number, and description of problem. A qualified
specialist will provide, via telephone (United States and Canada), or FAX (Outside the United States and Canada),
troubleshooting assistance for facility personnel and provide necessary parts to make repairs. If troubleshooting determines
the need for on-site technical service, a qualified service representative will be dispatched. Replacement of non-technical
items will be the responsibility of the customer. If requested by Hill-Rom, products or parts for which a warranty claim is
made shall be returned prepaid to Hill-Rom’s factory.
OUT OF WARRANTY EXCHANGE POLICY
After the expiration of the original warranty, upon request, Hill-Rom will ship as a replacement, components such as selected:

5
motors and printed circuit boards, for like units returned to Hill-Rom by the original purchaser at a substantial savings. Please
call Hill-Rom Technical Support Department for current pricing.
PARTS AVAILABILITY POLICY
Hill-Rom will offer parts for new and remanufactured products for ten (10) years from date of sale; for communications
products for five (5) years from date of sale.
Note: Some original component parts and assemblies may not be available; functional equivalents may be substituted.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES
AND IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS OF PURPOSE. HILL-ROM’S OBLIGATION UNDER THESE
WARRANTIES SHALL NOT INCLUDE ANY LIABILITY FOR LOSS OF PROFITS, DIRECT, INDIRECT OR
CONSEQUENTIAL DAMAGES OR DELAYS. Some states, provinces, or countries do not allow the exclusion or
limitation of incidental or consequential damages, so the above exclusion or limitation may not apply. Any improper or
negligent use, any alterations or repairs not in accordance with Hill-Rom’s manuals or performed by others in such manner as
in Hill-Rom’s sole judgment affects the product materially and adversely, shall void these warranties. These warranties do not
cover failures due to misuse, abuse, neglect, or lack of routine maintenance. No employee or representative of Hill-Rom is
authorized to change these warranties in any way or grant any other warranty unless in writing and signed by a Hill-Rom
officer. These warranties provide specific legal rights; but, there may be other available rights, which vary from state to state,
province to province, or country to country.
Revised October 20, 1998
Hill-Rom Company, Inc., 1069 State Route 46 E, Batesville, IN 47006-9167

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 5
Service Manual (MAN272 REV 3)
Warranty
Chapter 5: Parts List

NOTES:

Page 5 - 6 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Recommended Spare Parts
Chapter 5: Parts List

Recommended Spare Parts


For a recommended spare parts list to service five or more units, see
table 5-1 on page 5-7.

Table 5-1. Recommended Spare Parts

Part Number Quantity Description


450846048 (3700) 2 Caster, 6" swivel
450848048 (3700) 2 Caster, 8" swivel (Affinity® Three
Bed)
45084-6648 (3700) 1 Caster, 6" brake
450848648 (3700) 1 Caster, 8" brake
450846448 (3700) 1 Caster, 6" brake/steer
450848448 (3700) 1 Caster, 8" brake/steer
63920 (3700) 2 Hilow/Foot drive
63927 (3700) 1 Head drive
4314317 (3700) 5 Fuse, timelag, 4A, 125V AC,
5 mm x 20 mm
4360519 (3700) 5 Fuse, timelag, 2A, 250V AC,
5 mm x 20 mm
65345 (3700) 1 PCB assembly, control
67208 (3700) 1 Interface board assembly
63945 (3700) 1 Cable assembly, battery
6394201 (3700)
6642910S (3700)
1
5
Cable assembly, battery-to-battery
Fuse, 10A, 32V DC, automotive style 5

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 7
Service Manual (MAN272 REV 3)
Recommended Spare Parts
Chapter 5: Parts List

NOTES:

Page 5 - 8 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Caster Module
Chapter 5: Parts List

Caster Module
Figure 5-2. Caster Module

Table 5-2. Caster Module

Item Number Part Number Quantity Description


1
2
450846048 (3700)
450848048 (3700)
2
2
Caster, 6" swivel (Affinity® Three Bed)
Caster, 8" swivel (Affinity® Three Bed)
5
3 45084-6648 (3700) 1 Caster, 6" brake
4 450848648 (3700) 1 Caster, 8" brake (Affinity® Three Bed)
5 450846448 (3700) 1 Caster, 6" brake/steer
6 450848448 (3700) 1 Caster, 8" brake/steer (Affinity® Three
Bed)

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 9
Service Manual (MAN272 REV 3)
Base Module—P/N63951S (Sheet 1 of 2) (Affinity® Three Bed)
Chapter 5: Parts List

Base Module—P/N63951S (Sheet 1 of 2) (Affinity® Three Bed)


Figure 5-3. Base Module—P/N 63951S (Sheet 1 of 2) (Affinity® Three Bed)

Page 5 - 10 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Base Module—P/N63951S (Sheet 1 of 2) (Affinity® Three Bed)
Chapter 5: Parts List

Table 5-3. Base Module—P/N 63951S (Sheet 1 of 2) (Affinity® Three Bed)

Item Number Part Number Quantity Description


1 43878 (3700) 16 Screw, button hd Torx® a

2 41002 (3700) 4 Spacer, rocker arm,


3 41269 (3700) 4 Leg cover
4 32425 (3700) 2 Set screw, hex socket hd
5 43030 (3700) 4 Spring washer, caster
6 32572 (3700) 4 Screw
7 18890 (3700) 4 Truarc ring
8 41001PL (3700) 2 Groove pin, rocker arm
9 4258358 (3700) 2 Brake/steer disc—green
10 41237 (3700) 2 Pad, right
11 4258360 (3700) 2 Brake/steer disc—orange
12 41238 (3700) 2 Pad, left
13 43940 (3700) 1 Brake pedal assembly, lh
14 9034002 (3700) 2 Bearing, DU flanged
15 44017PL (3700) 2 Bracket, brake rod support
16 439460148 (3700) 1 End cap, cross tube, lh
17 46260 (3700) 4 Screw, button hd Torx®
18 32909 (3700) 8 Locknut
19 90055-10 (3700) 4 Screw, pan hd truss
20 43953 (3700) 2 Cam pivot
21
22
755 (3700)
43955PL (3700)
4
1
Locknut
Brake strip
5
23 757 (3700) 2 Pin, connector
24 44352 (3700) 4 Hairpin
25 34715 (3700) 2 Rocker arm assembly
26 43957 (3700) 1 Plate, detent housing
27 44350 (3700) 2 Clevis pin
28 44023 (3700) 1 Detent mechanism assembly
29 6722548 (3700) 2 Base frame weldment
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 11
Service Manual (MAN272 REV 3)
Base Module—P/N63951S (Sheet 1 of 2) (Affinity® Three Bed)
Chapter 5: Parts List

Item Number Part Number Quantity Description


30 4098301 (3700) 4 Washer, celcon
31 9026312 (3700) 2 Bearing, DU
32 470850148 (3700) 4 Base boss
33 9047714 (3700) 4 Screw, shoulder, hex socket hd
34 4130848 (3700) 4 End cap
35 44433 (3700) 1 Brake pedal assembly, rh
36 4394602 (3700) 1 End cap, cross tube, rh
37 43956PL (3700) 1 Brake/steer strip
38 6988301S (3700) 1 Base module, complete assembly,
domestic
39 6988302S (3700) 1 Base module, complete assembly, inter­
national
NOTE:
Items 38 and 39 not shown in figure 5-3 on page 5-10.

Page 5 - 12 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Base Module—P/N63951S (Sheet 1 of 2) (Affinity® Three Bed)
Chapter 5: Parts List

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 13
Service Manual (MAN272 REV 3)
Base Module—P/N63951S (Sheet 2 of 2) (Affinity® Three Bed)
Chapter 5: Parts List

Base Module—P/N63951S (Sheet 2 of 2) (Affinity® Three Bed)


Figure 5-4. Base Module—P/N 63951S (Sheet 2 of 2) (Affinity® Three Bed)

Page 5 - 14 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Base Module—P/N63951S (Sheet 2 of 2) (Affinity® Three Bed)
Chapter 5: Parts List

Table 5-4. Base Module—P/N 63951S (Sheet 2 of 2) (Affinity® Three Bed)

Item Number Part Number Quantity Description


1 60133 (3700) 12 Hex rivnut
2 61653PL (3700) 1 Clevis pin
3 6314308PL (3700) 1 Clevis pin
4 63920 (3700) 1 Hilow/Foot drive
5 61615 (3700) 4 Rue ring cotter
6 9026304 (3700) 4 Bearing, DU
7 6392648 (3700) 1 Intermediate frame weldment
8 66039 (3700) 4 Nut, jam
9 41022 (3700) 8 Pivot block
10 43772 (3700) As required Lithium base grease
11 6722348 (3700) 1 Lift arm crank weldment
12 62228 (3700) 8 Screw, hex hd
13 6722448 (3700) 1 Lift arm follower weldment
14 41024 (3700) 4 Pivot block plate
15 27509 (3700) 4 Roll pin
16 6314304PL (3700) 2 Clevis pin
17 6396248 (3700) 1 Link, hilow lift arms
18 65987 (3700) 2 Guard, lift arm
19 393 (3700) 4 Screw, pan hd
20 68189 4 Retaining strap
21 9001808 8 Screw with lock patch
5

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 15
Service Manual (MAN272 REV 3)
Main Frame Module
Chapter 5: Parts List

Main Frame Module


Figure 5-5. Main Frame Module

Page 5 - 16 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Main Frame Module
Chapter 5: Parts List

Table 5-5. Main Frame Module

Item Number Part Number Quantity Description


1 6391648 (3700) 1 Cover, top motor
2 41250 (3700) 1 Headboard bushing, oval
3 43878 (3700) 17 Screw, button hd Torx®a

4 41078 (3700) 1 Headboard bushing, round


5 41243S48 (3700) 1 Roller bumper and bracket assembly
6 25208 (3700) 2 Axle
7 90016-06 (3700) 4 Cap screw, hex hd
8 3455748 (3700) 2 Roller bumper
9 25210 (3700) 4 Pushnut
10 41243 (3700) 2 Bracket, roller bumper
11 3517 (3700) 2 Roll pin
12 38135 (3700) 2 Knob assembly
13 6391948 (3700) 1 Mainframe weldment
14 61615 (3700) 4 Rue ring cotter, 3/8"
15 61653PL (3700) 3 Clevis pin
16 6314306PL (3700) 1 Clevis pin
17 63927 (3700) 1 Head drive
18 63920 (3700) 1 Hilow/Foot drive
19 6591548 (3700) 1 Foot drive extension tube weldment
20 6161501 (3700) 7 Rue ring cotter
21
22
6398848 (3700)
41273 (3700)
1
1
Bottom cover weldment
Seat side shroud, lh
5
23 393H (3700) 4 Screw, pan hd
24 6393348 (3700) 1 Foot section pivot weldment
25 41023 (3700) 4 Pivot block, trend
26 48868 (3700) 2 Clevis pin
27 18921 (3700) 1 Screw, pan hd truss
28 37879 (3700) 1 Ground strap assembly
29 13206 (3700) 1 Screw, pan hd truss
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 17
Service Manual (MAN272 REV 3)
Main Frame Module
Chapter 5: Parts List

Item Number Part Number Quantity Description


30 4886801 (3700) 1 Clevis pin (used starting Jan. 2001)
or or
90260-48 (3700) Shoulder screw, hex socket hd,
(used prior to Jan. 2001)
31 41263 (3700) 1 Foot section link bar weldment
32 41413 (3700) 1 Shroud, front
33 46772-48 (3700) 1 Placenta basin
34 41111S (3700) 1 Yoke casting assembly
35 48868 (3700) 4 Clevis pin (used starting Jan. 2001)
or or
90260-28 (3700) Shoulder screw, hex socket hd,
(used prior to Jan. 2001)
36 9026301 (3700) 20 Bushing, DU
37 4887448 (3700) 3 Foot section lift arm
38 41272 (3700) 1 Seat side shroud, rh
39 19678 (3700) 4 Truarc ring
40 36826 (3700) 2 Ring, retaining
41 9026307 (3700) 4 Bearing, DU self-lube
42 43039 (3700) 2 Link, seat pivot
43 41089PL (3700) 4 Pin, foot section
44 27509 (3700) 4 Roll pin
45 41036 (3700) 4 Bushing, foot section pivot
46 41037PL (3700) 4 Plate, foot section bushing
47 9016128 (3700) 8 Cap screw, hex socket hd
48 20605 (3700) 4 Locknut
49 41253 (3700) 1 Bracket, Trendelenburg limit switch
50 34401 (3700) 2 Spacer
51 19512 (3700) 2 Switch, snap, spdt
52 31642 (3700) 4 Screw
53 41450 (3700) 2 Trendelenburg gas spring
54 9685 (3700) 2 Roll pin
55 41252 (3700) 2 Trendelenburg gas head

Page 5 - 18 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Main Frame Module
Chapter 5: Parts List

Item Number Part Number Quantity Description


56 41091 (3700) 2 Trendelenburg release lever
57 43056 (3700) 2 Spring, extension
58 43049 (3700) 2 Screw, shoulder, pan hd cross
59 40497 (3700) 2 Nut, keps
60 831 (3700) 2 Locknut
61 9006116 (3700) 2 Screw
62 41073PL (3700) 4 Trendelenburg activator
63 41087PL (3700) 2 Trendelenburg follower
64 41135PL (3700) 2 Trendelenburg cam
65 41079PL (3700) 2 Trendelenburg stop
66 20802 (3700) 2 Nut, keps
67 4540 (3700) 2 Washer
68 9025910 (3700) 2 Shoulder screw, hex socket hd,
5/16-18 x 5/8"
69 SA4840 (3700) As required Red Loctite® #271 a

70 9025916 (3700) 2 Screw, shoulder, hex socket hd


71 43041 (3700) 2 Trendelenburg limit spacer
72 15463 (3700) 2 Bolt
73 40915PL (3700) 1 Trendelenburg handle hex bar
74 4630 (3700) 2 Oilite® bushing
b

75 4089648 (3700) 2 Trendelenburg handle assembly


76
77
4115363 (3700)
42539 (3700)
2
2
Cover, Trendelenburg handle
Spiral pin, 3/16" x 1¼" 5
78 4257363 (3700) 2 Trendelenburg handle foam
79 4116363 (3700) 2 Trendelenburg handle plug
80 65914 2 Spacer
a. Loctite® is a registered trademark of Loctite Corporation.
b. Oilite® is a registered trademark of Beemer Precision, Incorporated.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 19
Service Manual (MAN272 REV 3)
Siderail Module
Chapter 5: Parts List

Siderail Module
Figure 5-6. Siderail Module

Page 5 - 20 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Siderail Module
Chapter 5: Parts List

Table 5-6. Siderail Module)

Item Number Part Number Quantity Description


1 62625 (3700) 4 Screw, pan hd, self-locking
2 4214101 (3700) 24 Screw, pan hd Torx® hi-lo
a

3 6860002 (3700) 1 Siderail cover, patient, rh


4 472730448 (3700) 16 Screw, nylon
5 6503802 (3700) 1 P.C. board assembly, caregiver siderail
control, rh
6 6401301 (3700) 2 Cable assembly, siderail jumper, 11
7 6401302 (3700) 2 Cable assembly, siderail jumper, 12
8 6528602 (3700) 1 P.C. board assembly, patient siderail
control, rh
9 4727110PL (3700) 4 D-pin
10 35325 (3700) 18 E-ring
11 42142 (3700) 6 Screw, pan hd Torx® hi-lo
12 49192PL (3700) 4 Pin, latch pivot
13 20802 (3700) 4 Nut, keps
14 47323 (3700) 2 Dampener
15 44408 (3700) 4 Screw, hex washer hd
16 63250 (3700) 2 Cover, center arm
17 4727102PL (3700) 4 D-pin
18 6887248 (3700) 2 Siderail mounting bracket weldment
(Replaces 4725948 siderail mounting
bracket weldment. When replacing a
4725948 weldment with a 6887248 5
weldment, also replace 49192PL latch
pivot pin with 68861PL latch pivot
D-pin, and two 35325 E-rings with two
18890 E-rings)
19 47257PL (3700) 2 Latch, siderail
20 49111 (3700) 2 Spring, latch bias
21 19124 (3700) 2 Cable tie
22 62753 (3700) 2 Center arm, siderail
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 21
Service Manual (MAN272 REV 3)
Siderail Module
Chapter 5: Parts List

Item Number Part Number Quantity Description


23 47144PL (3700) 2 Shaft, pivot, siderail release lever
24 3005804 (3700) 2 Cable assembly, audio taper
25 6397501 (3700) 2 Siderail arm, lh
26 63977 (3700) 2 Upper rail
27 47256 (3700) 2 Lever, siderail release
28 6397648 (3700) 2 Siderail center weldment
29 67060 (3700) 2 Seal, siderail
30 4096301 (3700) 2 Knob, volume
31 6528601 (3700) 1 P.C. board assembly, patient siderail
control, lh
32 6397901 (3700) 1 Siderail cover, patient, lh
33 6397502 (3700) 2 Siderail arm, rh
34 38873 (3700) 2 Speaker
35 28562 (3700) 8 Palnut
36 6860001 (3700) 2 Siderail cover, caregiver
37 6503801 (3700) 1 P.C. board assembly, caregiver siderail
control, lh
38 63969 (3700) 4 Bushing
39 66861 (3700) 4 Hex gasket, top cane
40 60619 (3700) 8 Nut, nylock
41 9001656 (3700) 8 Cap screw, hex hd
42 64011 (3700) 2 Cable assembly, siderail
43 49193 (3700) 2 E-ring
44 66860 (3700) 2 Gasket, top cane
45 4565820 (3700) 2 Ground strap, 20" (51 cm)
46 43878 (3700) 2 Screw, button hd Torx® a

47 68537 (3700) 2 Plug, volume control


a. Torx® is a registered trademark of Textron, Inc.

Page 5 - 22 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Siderail Module
Chapter 5: Parts List

Item Number Part Number Quantity Description


48 18890 (3700) 4 E-ring (Used only with 6887248 siderail
mounting bracket weldment and
68861PL latch pivot D-pin.)
or or
35325 (3700) E-ring (Used only with 4725948 siderail
mounting bracket weldment and
49192PL latch pivot pin.)
49 68861PL (3700) 2 D-pin, latch pivot (Used only with
6887248 siderail mounting bracket
weldment and 18890 E-rings.)
or or
49192PL (3700) Pin, latch pivot (Used only with 4725948
siderail mounting bracket weldment and
35325 E-rings.)
50 6665801 2 Blank label
51 64011 2 Cable assembly, siderail
52 64010 2 Cable assembly, siderail to P.C. board

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 23
Service Manual (MAN272 REV 3)
Electronics Module (Sheet 1 of 3)
Chapter 5: Parts List

Electronics Module (Sheet 1 of 3)


Figure 5-7. Electronics Module (Sheet 1 of 3)

Page 5 - 24 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Electronics Module (Sheet 1 of 3)
Chapter 5: Parts List

Table 5-7. Electronics Module (Sheet 1 of 3)

Item Number Part Number Quantity Description


1 6342201 (3700) 1 Power cord, detachable 110V
or or
6342202 (3700) Power cord, detachable 220V,
Australia/New Zealand
or or
6342203 (3700) Power cord, Continental Europe
or or
6342204 (3700) Power cord, UK/Ireland
or or
6342205 (3700) Power cord, Switzerland
or or
6342206 (3700) Power cord, Denmark
2 46015 (3700) 2 Screw, bottom rail end cap
3 64002 (3700) 1 Power entry module
4 4314317 (3700) 2 Fuse, timelag, 4A, 125V AC
or or
4360519 (3700) Fuse, timelag, 2A, 250V AC
5 66230 (3700) 1 Cable assembly, ground
6 43878 (3700) 18 Screw, button hd Torx® a

7 4142501S (3700) 1 Cable assembly, pendant control


8 6398848 (3700) 1 Bottom cover weldment
9 63166 (3700) 2 Screw, button hd Torx®
10 66129 (3700) 1 Night light and cable assembly
6679903
4734501
1
1
Assembly, lens, bulb and holder
Bulb
5
11 4142301S (3700) 1 Cable assembly, Trendelenburg
mechanical limits
12 19124 (3700) 2 Cable tie
13 6395948 (3700) 1 Electronic plate weldment
14 35663 (3700) 5 Standoff, P.C. board
15 4840501 (3700) 2 Battery, lead acid, sealed, 12V, 7.2A
16 6583048 (3700) 1 Bracket, battery hold-down
17 63945 (3700) 1 Cable assembly, battery
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 25
Service Manual (MAN272 REV 3)
Electronics Module (Sheet 1 of 3)
Chapter 5: Parts List

Item Number Part Number Quantity Description


18 68121 (3700) 1 P.C. board assembly, logic control
(replaces 65345 logic control P.C. board
assembly and 67208 interface P.C. board
assembly)
19 63960 (3700) 1 Transformer assembly
20 63944 (3700) 1 Cable assembly, lockout
21 63981 (3700) 1 Switch, lockout
22 66278 (3700) 1 Transformer assembly, OB light option
23 29891 (3700) 1 Cable clamp
24 4759 (3700) 1 Screw

Page 5 - 26 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Electronics Module (Sheet 2 of 3)
Chapter 5: Parts List

Electronics Module (Sheet 2 of 3)


Figure 5-8. Electronics Module (Sheet 2 of 3)

Table 5-8. Electronics Module (Sheet 2 of 3)

5
Item Number Part Number Quantity Description
1 40421 (3700) 2 Cable tie mount
2 66053 (3700) 2 Bracket, DC box mount
3 4214101 (3700) 2 Screw, pan hd Torx® hi-lo
a

4 Reference only 2 Box, auxiliary outlet


5 43878 (3700) 6 Screw, button hd Torx®
6 19124 (3700) 4 Cable tie
7 64009 (3700) 1 Cable assembly, OB light
8 66278 (3700) 1 Transformer assembly, OB light option
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 27
Service Manual (MAN272 REV 3)
Electronics Module (Sheet 3 of 3)
Chapter 5: Parts List

Electronics Module (Sheet 3 of 3)


Figure 5-9. Electronics Module (Sheet 3 of 3)

Page 5 - 28 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Electronics Module (Sheet 3 of 3)
Chapter 5: Parts List

Table 5-9. Electronics Module (Sheet 3 of 3)

Item Number Part Number Quantity Description


1 43878 (3700) 2 Screw, button hd Torx® a

2 4759 (3700) 5 Screw


3 41226PL (3700) 1 Junction box top
4 6454701 (3700) 1 PCB assembly, relay junction
or or
6454702 (3700) PCB assembly, relay junction with UTV
5 66253 (3700) 1 Cable assembly, 36-conductor
6 41455 (3700) 1 Insulator, junction box
7 41306PL (3700) 1 Junction box bottom
8 3976301 (3700) 4 Standoff
9 41298 (3700) 1 Washer, nylon
10 34512 (3700) 1 Dummy plug
11 4128201 (3700) 1 Dome plug
12 6702148 (3700) 1 Protector cover
13 43031 (3700) 1 Bracket (non-SideCom® beds only)
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 29
Service Manual (MAN272 REV 3)
Electrical Cable Routing (Sheet 1 of 2)
Chapter 5: Parts List

Electrical Cable Routing (Sheet 1 of 2)


Figure 5-10. Electrical Cable Routing (Sheet 1 of 2)

Page 5 - 30 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Electrical Cable Routing (Sheet 1 of 2)
Chapter 5: Parts List

Table 5-10. Electrical Cable Routing (Sheet 1 of 2)

Item Number Part Number Quantity Description


1 64002 (3700) 1 Power entry module
2 66129 (3700) 1 Night light and cable assembly
3 4142301S (3700) 1 Trendelenburg mechanical limits cable
assembly
4 43878 (3700) 2 Screw, button hd Torx® a

5 19124 (3700) 5 Cable tie


6 4142501S (3700) 1 Cable assembly, pendant control
7 63981 (3700) 1 Switch, lockout
8 44279 (3700) 1 Night light sensor and cable assembly
9 63960 (3700) 1 Transformer assembly
10 66278 (3700) 1 Transformer assembly, OB light option
11 17292 (3700) 2 Cable clamp
12 63944 (3700) 1 Cable assembly, lockout
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 31
Service Manual (MAN272 REV 3)
Electrical Cable Routing (Sheet 2 of 2)
Chapter 5: Parts List

Electrical Cable Routing (Sheet 2 of 2)


Figure 5-11. Electrical Cable Routing (Sheet 2 of 2)

Page 5 - 32 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Electrical Cable Routing (Sheet 2 of 2)
Chapter 5: Parts List

Table 5-11. Electrical Cable Routing (Sheet 2 of 2)

Item Number Part Number Quantity Description


1 6394201 (3700) 1 Cable assembly, battery-to-battery
2 44125 (3700) 1 Plug (used on 230V beds only)
3 44128 (3700) 1 Lockwasher, serrated (used on 230V
beds only)
4 44126 (3700) 1 Nut (used on 230V beds only)
5 44127 (3700) 1 Washer—color code green/yellow (used
on 230V beds only)
6 44464 (3700) 1 Label, potential equalization (used on
230V beds only)
7 63945 (3700) 1 Cable assembly, battery
8 19124 (3700) 2 Cable tie
9 43878 (3700) 7 Screw, button hd Torx® a

10 17292 (3700) 6 Cable clamp


11 6642910S (3700) 1 Fuse, 10A, 32V DC, automotive style
12 27873 (3700) 1 Cable clamp (used prior to
panel-mounted pendant control cable
connection only)
13 4142501S (3700) 1 Cable assembly, pendant control
14 10595 (3700) 2 Screw, pan hd
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 33
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 1 of 3)
Chapter 5: Parts List

Sleep Deck Module (Sheet 1 of 3)


Figure 5-12. Sleep Deck Module (Sheet 1 of 3)

Page 5 - 34 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 1 of 3)
Chapter 5: Parts List

Table 5-12. Sleep Deck Module (Sheet 1 of 3)

Item Number Part Number Quantity Description


1 41059 (3700) 1 Panel, V seat
2 40958 (3700) 2 Snap, mattress retaining
35308 2 Finish washer
3 41060 (3700) 1 Panel, U seat
4 777 (3700) 2 Hex locknut
5 41047 (3700) 4 Washer, nylon
6 43552 (3700) 2 Ring, retaining
7 40817 (3700) 2 Ring, retaining
8 6392148 (3700) 1 Seat section weldment
9 659660148 (3700) 1 Labor grip weldment, lh
10 430220263 (3700) 2 Foam tube, ¾" ID x 25"
11 4116363 (3700) 2 Trendelenburg handle plug
12 6394801PL (3700) 1 Labor grip release weldment, lh
13 42617 (3700) 2 Washer, spring
14 90338-10 (3700) 2 Screw, shoulder
15 62919 (3700) 2 Cap, release handle, vinyl dipped
16 6314306PL (3700) 1 Clevis pin
17 61615 (3700) 2 Rue ring cotter
18 61653PL (3700) 1 Clevis pin
19 9026307 (3700) 2 Bearing, DU self-lube
20
21
9026006 (3700)
4127048 (3700)
2
1
Screw, shoulder
Front shroud, U seat
5
22 4127148 (3700) 1 Front shroud, V seat
23 43878 (3700) 4 Screw, button hd Torx®
a

24 6394802PL (3700) 1 Labor grip release weldment, rh


25 40812 (3700) 2 Spring
26 40811 (3700) 2 Pin, locking
27 36958 (3700) As required Lubricant
28 65967 (3700) 2 Clevis pin
29 49257 (3700) 2 Hairpin
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 35
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 1 of 3)
Chapter 5: Parts List

Item Number Part Number Quantity Description


30 659660248 (3700) 1 Labor grip weldment, rh
31 64006PL (3700) 1 Pin, head/seat drive
32 65986 5 Screw

Page 5 - 36 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 1 of 3)
Chapter 5: Parts List

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 37
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 2 of 3)
Chapter 5: Parts List

Sleep Deck Module (Sheet 2 of 3)


Figure 5-13. Sleep Deck Module (Sheet 2 of 3)

Page 5 - 38 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 2 of 3)
Chapter 5: Parts List

Table 5-13. Sleep Deck Module (Sheet 2 of 3)

Item Number Part Number Quantity Description


1 65986 (3700) 13 Screw
2 41061 (3700) 1 Head deck
3 41344 (3700) 2 Bumper plug
4 67088 (3700) 2 Slide
5 6393901PL (3700) 1 Handle weldment, CPR, lh
6 6393801 (3700) 1 CPR handle, lh
7 35326 (3700) 2 E-ring
8 9033907 (3700) 4 Screw, locking
9 64007 (3700) 2 Release cable assembly
10 65912 (3700) 1 Damper, CPR assist
11 64012 (3700) 1 Gas spring, CPR assist
12 61440 (3700) 1 Washer
13 44352 (3700) 1 Hairpin
14 40827 (3700) As required Dual-sided adhesive tape
15 19124 (3700) 3 Cable tie
16 6394348 (3700) 1 Bracket, CPR cable attachment
17 6395348 (3700) 1 Bracket, CPR cable housing
18 6393802 (3700) 1 CPR handle, rh
19 6393902PL (3700) 1 Handle weldment, CPR, rh
20 6395548 (3700) 1 Head section weldment
21 4106848 (3700) 1 Splash channel
5

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 39
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 3 of 3)
Chapter 5: Parts List

Sleep Deck Module (Sheet 3 of 3)


Figure 5-14. Sleep Deck Module (Sheet 3 of 3)

Page 5 - 40 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 3 of 3)
Chapter 5: Parts List

Table 5-14. Sleep Deck Module (Sheet 3 of 3)

Item Number Part Number Quantity Description


1 42575 (3700) 1 Wireform, V seat
2 42576 (3700) 1 Wireform, U seat
3 1012 (3700) 2 Washer
4 9001828 (3700) 2 Cap screw, hex hd
5 6724448 (3700) 1 Foot section weldment, lift-off (used
with 67415 yoke spacer and 4257401 cap
screw only)
or or
448170148 (3700) Foot section weldment, lift-off (used
with 42574 cap screw only)
6 9001828 (3700) 4 Cap screw, hex hd
7 1012 (3700) 8 Washer
8 41204S (3700) 1 Latch weldment, lh
9 4435 (3700) 4 Locknut
10 4257401 (3700) 6 Cap screw, hex socket flat hd (used with
6724448 foot section weldment and
67415 yoke spacer only)
or or
42574 (3700) Cap screw, hex socket flat hd (used with
448170148 foot section weldment only)
11 61510 (3700) 1 Latch, foot, lh
12 67415 (3700) 2 Spacer, yoke (used with 6724448 foot

5
section weldment and 4257401 cap
screw only)
13 41104 (3700) 2 Bumper, foot section
14 90058-06 (3700) 4 Screw, truss hd cross, patchlock
15 44768 (3700) 4 Screw, flat hd machine
16 41052 (3700) 4 Stud snap
17 15250 (3700) 4 Locknut, washer-base
18 46765 (3700) 2 Hole plug
19 61509 (3700) 1 Latch, foot, rh
20 41203 (3700) 1 Latch weldment, rh
21 615140148 (3700) 1 Foot section weldment, slide-off

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 41
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 3 of 3)
Chapter 5: Parts List

Item Number Part Number Quantity Description


22 9023428 (3700) 4 Locknut
23 6356601PL (3700) 1 Bracket weldment, lh
24 63611 (3700) 4 Screw, hex socket flat hd
25 18890 (3700) 2 Truarc ring
26 61532PL (3700) 2 Pivot stud
27 61534 (3700) 1 Spring, non-corrosive torsion
28 63565 (3700) 1 Pivot shaft
29 61525 (3700) 2 Bearing
30 35326 (3700) 4 E-ring
31 6153401 (3700) 1 Spring, torsion, lh
32 61537 (3700) 1 Bracket, slide, lh
33 42574 (3700) 6 Screw
34 6341601PL (3700) 1 Latch weldment, lh
35 61529PL (3700) 2 Connecting rod
36 6153002PL (3700) 1 Release handle, lh
37 6153001PL (3700) 1 Release handle, rh
38 61533 (3700) 2 Vinyl cap
39 61531PL (3700) 4 Link rivet
40 6341602PL (3700) 1 Latch weldment, rh
41 61536 (3700) 1 Bracket, slide, rh
42 46765 (3700) 4 Hole plug
43 6356602PL (3700) 1 Bracket weldment, rh
44 (not 6781702S 1 Foot section, Lift-off, complete—
shown) straight cut
6781701S 1 Foot section, Lift-off, complete—U cut
45 (not 6781704S Foot section, slide-off, complete—
shown) straight cut
6781703S Foot section, slide-off, complete—U cut

Page 5 - 42 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Sleep Deck Module (Sheet 3 of 3)
Chapter 5: Parts List

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 43
Service Manual (MAN272 REV 3)
Label Module (Sheet 1 of 4)—P3700A
Chapter 5: Parts List

Label Module (Sheet 1 of 4)—P3700A


Figure 5-15. Label Module (Sheet 1 of 4)—P3700A

Page 5 - 44 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Label Module (Sheet 1 of 4)—P3700A
Chapter 5: Parts List

Table 5-15. Label Module (Sheet 1 of 4)—P3700A

Item Number Part Number Quantity Description


1 6399301 (3700) 1 Wiring diagram
2 67020 (3700) 2 Label, caution, electrical shock
3 6702000 (3700) 2 Label, warning, international
4 6583102 (3700) 1 Label, power inlet
5 6399602 (3700) 1 Label, Trendelenburg, rh
6 66870 (3700) 1 Label, battery warning
7 6399601 (3700) 1 Label, Trendelenburg, lh
8 65989 (3700) 1 Label, battery caution
9 64004 (3700) 1 Label, base frame
10 4583600 (3700) 2 Label, foot clear, international
11 45836 (3700) 2 Label, foot clear, Euro
12 64005 (3700) 1 Label, base, Hill-Rom
13 45834 (3700) 2 Label, foot support warning
14 4583400 (3700) 2 Label, foot support warning,
international
15 63995 (3700) 2 Label, CPR
16 65382 (3700) 2 Label, auxiliary light
17 65833 (3700) 2 Label, siderail warning
18 6583300 (3700) 2 Label, siderail warning, international
19 66654 (3700) 2 Overlay, caregiver, blank
20
21
66655 (3700)
6665801 (3700)
2
2
Overlay, patient, blank
Label, speaker control, blank 5
22 6665802 (3700) 2 Label, volume control, blank

NOTE:
Items 15 through 22 are not shown in figure 5-15 on page 5-44.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 45
Service Manual (MAN272 REV 3)
Label Module (Sheet 2 of 4)—P3700A
Chapter 5: Parts List

Label Module (Sheet 2 of 4)—P3700A


Figure 5-16. Label Module—Siderail (Sheet 2 of 4)—P3700A

Page 5 - 46 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Label Module (Sheet 2 of 4)—P3700A
Chapter 5: Parts List

Table 5-16. Label Module—Siderail (Sheet 2 of 4)—P3700A

Item Number Part Number Quantity Description


1 6396507 (3700) 1 Overlay, caregiver, rh
2 6396503 (3700) 1 Overlay, caregiver, rh
3 6396500 (3700) 1 Overlay, caregiver, rh
4 6396501 (3700) 1 Overlay, caregiver, rh
5 6396509 (3700) 1 Overlay, caregiver, rh
6 6396697 (3700) 1 Overlay, patient, rh
7 6396607 (3700) 1 Overlay, patient, rh
8 6396693 (3700) 1 Overlay, patient, rh
9 6396603 (3700) 1 Overlay, patient, rh
10 6396600 (3700) 1 Overlay, patient, rh
11 6396609 (3700) 1 Overlay, patient, rh
12 6396601 (3700) 1 Overlay, patient, rh
13 6396797 (3700) 1 Overlay, patient, lh
14 6396707 (3700) 1 Overlay, patient, lh
15 6396793 (3700) 1 Overlay, patient, lh
16 6396703 (3700) 1 Overlay, patient, lh
17 6396700 (3700) 1 Overlay, patient, lh
18 6396709 (3700) 1 Overlay, patient, lh
19 6396701 (3700) 1 Overlay, patient, lh
20 6396807 (3700) 1 Overlay, caregiver, lh
21
22
6396803 (3700)
6396800 (3700)
1
1
Overlay, caregiver, lh
Overlay, caregiver, lh
5
23 6396801 (3700) 1 Overlay, caregiver, lh
24 6396809 (3700) 1 Overlay, caregiver, lh

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 47
Service Manual (MAN272 REV 3)
Label Module (Sheet 3 of 4)—P3700A
Chapter 5: Parts List

Label Module (Sheet 3 of 4)—P3700A


Figure 5-17. Label Module (Sheet 3 of 4)—P3700A

Table 5-17. Label Module (Sheet 3 of 4)—P3700A

Item Number Part Number Quantity Description


1 6153501 (3700) 1 Label, foot release, rh
2 63609 (3700) 2 Label, slide off foot section caution
3 6360900 (3700) 2 Label, slide off foot section caution,
international
4 45832 (3700) 1 Label, foot latch warning
5 4583200 (3700) 1 Label, foot latch warning, international
6 6153502 (3700) 1 Label, foot release, lh
7 6709011 (3700) 2 Label, weight limit, foot support
8 6709003 (3700) 2 Label, weight limit, calf support

Page 5 - 48 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Label Module (Sheet 4 of 4)—P3700A
Chapter 5: Parts List

Label Module (Sheet 4 of 4)—P3700A


Figure 5-18. Label Module (Sheet 4 of 4)—P3700A

Table 5-18. Label Module (Sheet 4 of 4)—P3700A

Item Number Part Number Quantity Description


1
2
67091 (3700)
46770 (3700)
1
1
Grounding reliability
Label CE
5
3 66859 (3700) 1 Label, lockout
4 42446 (3700) 1 Label, VDE (if applicable)
5 6274006 (3700) 1 Label, serial number blank
6 44464 (3700) 1 Label, potential equalization
7 45830 (3700) 1 Label, caution
8 4583000 (3700) 1 Label, caution, international
9 65831 (3700) 1 Label, power inlet
10 6583101 (3700) 1 Label, power inlet, international

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 49
Service Manual (MAN272 REV 3)
Foot Support Module—P3700A
Chapter 5: Parts List

Foot Support Module—P3700A


Figure 5-19. Foot Support Module—P3700A

Page 5 - 50 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Foot Support Module—P3700A
Chapter 5: Parts List

Table 5-19. Foot Support Module—P3700A

Item Number Part Number Quantity Description


1 6563703 (3700) 1 Foot support module with calf support,
lh
2 6563701S (3700) 1 Foot support module, lh
3 6697001S (3700) 1 Calf support module, lh
4 65700 (3700) 4 Oilite® bushing
a

5 6572401PL (3700) 1 Cam post weldment, lh


6 66966 (3700) 2 Set screw, cup point
7 19124 (3700) 2 Cable tie
8 6580101 (3700) 1 Mechlok® , lh b

9 35325 (3700) 4 E-ring


10 65799 (3700) 2 Cable, foot abduction
11 6604202 (3700) 4 Clevis pin
12 61615 (3700) 4 Rue ring cotter
13 65829 (3700) 2 Toe boot
14 62919 (3700) 2 Cap, release handle, vinyl dipped
15 39172 (3700) As required Adhesive
16 65699PL (3700) 2 Release handle weldment
17 67686 (3700) 8 Plug
18 64466 (3700) 8 Screw, Torx® c

19 62959 (3700) 2 Handle/bumper, foot support


20
21
6604101 (3700)
67794 (3700)
1
2
Heel cup, foot support, lh
Wireform, bellows 5
22 9025806 (3700) 1 Screw, shoulder
23 65800 (3700) 2 Cable, foot tilt
24 66152 (3700) 2 Gas spring, foot support
25 6696501 (3700) 1 Mechlok®, foot tilt, lh
26 656320148 (3700) 1 Foot rest weldment, lh
27 6582801 (3700) 1 Bellows, foot support, lh
a. Oilite® is a registered trademark of Beemer Precision, Incorporated.
b. Mechlok® is a registered trademark of P.L. Porter Company.
c. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 51
Service Manual (MAN272 REV 3)
Foot Support Module—P3700A
Chapter 5: Parts List

Item Number Part Number Quantity Description


28 66430 (3700) 2 Washer, plastic
29 65964 (3700) 2 Snap ring
30 6580102 (3700) 1 Mechlok® , rh
a

31 6572402PL (3700) 1 Cam post weldment, rh


32 6604102 (3700) 1 Heel cup, foot support, rh
33 6696502 (3700) 1 Mechlok®, foot tilt, rh
34 656320248 (3700) 1 Foot rest weldment, rh
35 6582802 (3700) 1 Bellows, foot support, rh
36 6563704 (3700) 1 Foot support module with calf support,
rh
37 6563702S (3700) 1 Foot support module, rh
38 6697002S (3700) 1 Calf support module, rh
39 68112 (3700) 2 Foot support post plug
a. Mechlok® is a registered trademark of P.L. Porter Company.

Page 5 - 52 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Foot Support Module—P3700A
Chapter 5: Parts List

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 53
Service Manual (MAN272 REV 3)
Calf Support Module—P3700A
Chapter 5: Parts List

Calf Support Module—P3700A


Figure 5-20. Calf Support Module—P3700A

Page 5 - 54 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Calf Support Module—P3700A
Chapter 5: Parts List

Table 5-20. Calf Support Module—P3700A

Item Number Part Number Quantity Description


1 68117 (3700) 2 Knob assembly, calf support
2 37145 (3700) 2 Lock stud
3 37146 (3700) 4 Knob stud
4 8651 (3700) As required Loctite® #262
a

5 3714463 (3700) 4 Knob


6 66975 (3700) 2 Threaded insert
7 38677 (3700) 2 Roll pin
8 67258 (3700) 2 Ring, retaining, internal
9 6446601 (3700) 2 Screw, Torx® b

10 66979 (3700) 2 Pin, calf support


11 6696801 (3700) 1 Mount, calf support, lh
12 66966 (3700) 2 Set screw, cup point
13 67260 (3700) 2 Set screw cup point
14 66971 (3700) 2 Key, calf support
SA1736 2 Key kit
15 67259 (3700) 2 Spring
16 6697301 (3700) 1 Pull handle, lh
17 6696701 (3700) 1 Arm, calf support, lh
18 47598 (3700) 2 Ball and rod
19 23261 (3700) 2 Roll pin
20
21
70742 (3700)
19918 (3700)
2
4
Bracket, calf support
Washer
5
22 19224 (3700) 4 Bolt
23 4126048 (3700) 2 Former, calf support
24 67247 (3700) 2 Cover
25 6697302 (3700) 1 Pull handle, rh
26 6696802 (3700) 1 Mount, calf support, rh
27 6696702 (3700) 1 Arm, calf support, rh
a. Loctite® is a registered trademark of Loctite Corporation
b. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 55
Service Manual (MAN272 REV 3)
Air Module
Chapter 5: Parts List

Air Module
Figure 5-21. Air Module

Page 5 - 56 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Air Module
Chapter 5: Parts List

Table 5-21. Air Module

Item Number Part Number Quantity Description


1 4698031 (3700) 1 Air hose
2 46980-35 (3700) 1 Air hose, lumbar seat, rh
3 19124 (3700) 6 Cable tie
4 831 (3700) 4 Locknut
5 90016-40 (3700) 4 Cap screw, hex hd
6 66135 (3700) 1 Transformer and cable assembly, 220V
7 64008 (3700) 1 Bracket, transformer mount—220V
8 41466 (3700) 2 Quick coupler, male
9 3815606 (3700) 2 Hole plug, ¾" light neutral
10 64095 (3700) 1 Linear air pump
11 6686648 (3700) 1 Bracket, compressor mount
12 47070 (3700) 1 Rivnut
13 947 (3700) 4 Lockwasher, #8 internal tooth
14 987 (3700) 4 Screw
15 35663 (3700) 4 Standoff, P.C. board
16 4118802 (3700) 1 PCB assembly, mattress driver
17 66216 (3700) 1 Power cable, solenoid
18 9005811 (3700) 1 Screw
19 6621501 (3700) 1 Cable, digital mattress board
20 47103 (3700) 1 Muffler assembly
21
22
4760602 (3700)
4698008 (3700)
1
1
Tubing, vacuum
Air hose
5
23 64096 (3700) 1 Fitting, elbow—¼" hose
24 46981 (3700) 3 Standoff
25 4710501 (3700) 1 Manifold, 3 station
26 15380 (3700) 2 Screw
27 (not SA4462 1 Air coupling, female
shown)

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 57
Service Manual (MAN272 REV 3)
Headboard
Chapter 5: Parts List

Headboard
Figure 5-22. Headboard

Table 5-22. Headboard

Item Number Part Number Quantity Description


1 43878 (3700) 2 Screw, button hd Torx® a

2 37664 (3700) 2 Clip, IV


3 31773 (3700) 10 Screw
4 40995 (3700) 1 Handle weldment, headboard
5 42600 (3700) 2 Plug
6 42588 (3700) 2 Grip, handle, foam
7 41114 (3700) 1 Headboard, wood (oak or cherry)
or or
112769 (3700) Headboard, wood—Liberty Hill™
or (cherry)
112770 (3700) or
Headboard, wood—Freedom Hill™
(oak)

Page 5 - 58 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Headboard
Chapter 5: Parts List

Item Number Part Number Quantity Description


8 (not shown) 46395 2 Handle cover
9 (not shown) P3629 1 Headboard, blow molded
10 (not 64835105 1 Insert, laminate
shown)
a. Torx® is a registered trademark of Textron, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 59
Service Manual (MAN272 REV 3)
Caster Module—P3700B
Chapter 5: Parts List

Caster Module—P3700B
Figure 5-23. Caster Module—P3700B

Table 5-23. Caster Module—P377B

Item Number Part Number Quantity Description


1 45084-6648 3 Caster, 6" brake
2 450846448 1 Caster, 6" brake/steer

Page 5 - 60 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Caster Module—P3700B
Chapter 5: Parts List

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 61
Service Manual (MAN272 REV 3)
Base Module—P3700B (Sheet 1 of 2)
Chapter 5: Parts List

Base Module—P3700B (Sheet 1 of 2)


Figure 5-24. Base Module—P3700B (Sheet 1 of 2)

Page 5 - 62 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Base Module—P3700B (Sheet 1 of 2)
Chapter 5: Parts List

Table 5-24. Base Module—P3700B (Sheet 1 of 2)

Item Number Part Number Quantity Description


1 4258360 2 Disk, orange
2 4258358 2 Disk, green
3 41237 2 Pedal, rh
4 41238 2 Pedal, lh
5 32909 8 Locknut
6 142626 2 Pedal weldment
7 142628 2 Brake cam
8 43951 4 Pin
9 142639 2 Flanged bushing
10 143966 2 Brake pedal support
11 434602 1 End cap cross tube, rh
12 142635 2 Brake link, head
13 41308 4 End cap
14 90477 4 Shoulder bolt
15 4708501 4 Base boss cap
16 9026312 4 DU bushing
17 44352 6 Cotter pin
18 757 4 Clevis pin
19 43878 16 Screw
20 41269 4 Leg cover
21
22
158890
41002
8
8
Retaining ring
Spacer, rocker arm
5
23 32425 4 Bolt
24 34715 4 Rover arm
25 41001 4 Groove pin, rocker arm
26 32572 4 Screw
27 43030 4 Washer
28 40980301 4 Washer
29 142633 1 Brake link, foot, lh
30 4394601 1 End cap cross tube, lh
31 43957 1 Detent housing place

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 63
Service Manual (MAN272 REV 3)
Base Module—P3700B (Sheet 1 of 2)
Chapter 5: Parts List

Item Number Part Number Quantity Description


32 142638 1 Spring
33 40827 2 Double sided tape
34 43949 1 Detent plunger
35 43952 1 Detent mechanism housing
36 43951 4 Pin, grove
37 44350 2 Clevis pin
38 142634 1 Brake link, foot, rh

Page 5 - 64 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Base Module—P3700B (Sheet 1 of 2)
Chapter 5: Parts List

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 65
Service Manual (MAN272 REV 3)
Base Module—P3700B (Sheet 2 of 2)
Chapter 5: Parts List

Base Module—P3700B (Sheet 2 of 2)


Figure 5-25. Base Module—P3700B (Sheet 2 of 2)

Page 5 - 66 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Base Module—P3700B (Sheet 2 of 2)
Chapter 5: Parts List

Table 5-25. Base Module—P3700B (Sheet 2 of 2)

Item Number Part Number Quantity Description


1 6314308 1 Clevis pin
2 63920 1 Hilow foot drive motor
3 44352 2 Cotter pin
4 61653 1 Clevis pin
5 142933 1 Intermediate frame weldment
6 68188 4 Bushing
7 40827 4 Double sided tape
8 67223 1 Lift arm weldment
9 6314304 2 Clevis pin
10 9026304 4 DU bushing
11 63962 2 Link, hilow
12 66039 4 Nut
13 61615 4 Rue ring
14 143032 1 Lift arm follower
15 9001808 8 Screw, cap
16 68199 4 Retaining strap
17 27509 4 Roll pin

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 67
Service Manual (MAN272 REV 3)
Sleep Deck—P3700B
Chapter 5: Parts List

Sleep Deck—P3700B
Figure 5-26. Sleep Deck—P3700B

Page 5 - 68 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Sleep Deck—P3700B
Chapter 5: Parts List

Table 5-26. Sleep Deck—P3700B

Item Number Part Number Quantity Description


1 42575 1 V-cut wire form
2 42576 1 U-cut wire form
3 9001828 4 Screw
4 1012 10 Washer
5 142644 1 Latch weldment, lh
6 4435 4 Locknut
7 4257401 6 Screw
8 142642 1 Foot section receiver, lh
9 143885 1 Perineal stand, U-cut
10 9001836 2 Screw
11 144069 1 Perineal stand, V-cut
12 67244 1 Foot section weldment
13 46765 2 Plug
14 142645 1 Latch weldment, rh
15 142772 1 Placenta basin
16 143183 1 Lock block, rh
17 9016114 2 Screw (short)
18 144070 1 Foot section
19 142720 1 Lock, lh
20 145105 2 Slide guide assembly
21
22
145106
142641
1
1
Underframe assembly
Foot latch, rh
5
23 144184 4 Screw (long)

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 69
Service Manual (MAN272 REV 3)
Foot Support Module—P3700B
Chapter 5: Parts List

Foot Support Module—P3700B


Figure 5-27. Foot Support Module—P3700B

Page 5 - 70 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Foot Support Module—P3700B
Chapter 5: Parts List

Table 5-27. Foot Support Module—P3700B

Item Number Part Number Quantity Description


1 62959 2 Foot support handle
2 69391 6 Screw
3 43394 2 Clevis pin
4 60086 2 Screw
5 65829 2 Toe boot
6 142961 2 Handle, release
7 66152 2 Gas spring
8 138735 2 Screw
9 6604202 2 Clevis pin
10 61615 2 Rue ring
11 35325 6 E-ring
12 66966 2 Set screw
13 65700 4 Bushing
14 65800 2 Cable, foot tilt
15 65799 2 Cable, foot abduction
16 19124 2 Wire tie
17 66430 2 Washer, plastic
18 65964 2 Snap ring
19 142752 2 Heel cup
20 142825 2 Clamp, bellows
21
22
39172
68112
As required
2
Adhesive
Hole plug
5
23 142959 1 Foot rest weldment, lh
24 6572401 1 Cam post weldment, lh
25 6580101 1 Mechlok® , foot left, lh
a

26 142958 1 Bellows, foot support, lh


27 142377 1 Release handle weldment, lh
28 6696801 1 Mechlok®, foot tilt, lh
29 142960 1 Foot rest weldment, rh
30 6572402 1 Cam post weldment, rh
31 6580102 1 Mechlok®, foot lift, rh

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 71
Service Manual (MAN272 REV 3)
Foot Support Module—P3700B
Chapter 5: Parts List

Item Number Part Number Quantity Description


32 143010 1 Bellows, foot support, rh
33 143225 1 Release handle weldment, rh
34 6696502 1 Mechlok®, foot tilt, rh
35 (not 145103 1 Bellows assembly (includes 4, 20 and
shown) 32), rh
36 (not 145104 1 Bellows assembly (includes 2, 20 and
shown) 26), lh
a. Mechlok® is a registered trademark of P.L. Porter Company.

Page 5 - 72 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Calf Support Module—P3700B
Chapter 5: Parts List

Calf Support Module—P3700B


Figure 5-28. Calf Support Module—P3700B

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 73
Service Manual (MAN272 REV 3)
Calf Support Module—P3700B
Chapter 5: Parts List

Table 5-28. Calf Support Module—P3700B

Item Number Part Number Quantity Description


1 142388 2 Calf support pad
2 142621 1 Calf support arm, rh
3 69391 2 Screw (short)
4 138735 2 Screw (long)
5 3517 2 Roll pin
6 142620 1 Calf support arm, lh

Page 5 - 74 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Calf Support Module—P3700B
Chapter 5: Parts List

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 75
Service Manual (MAN272 REV 3)
Label (Sheet 1 of 4)—P3700B
Chapter 5: Parts List

Label (Sheet 1 of 4)—P3700B


Figure 5-29. Label (Sheet 1 of 4)—P3700B

Page 5 - 76 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Label (Sheet 1 of 4)—P3700B
Chapter 5: Parts List

Table 5-29. Label (Sheet 1 of 4)—P3700B

Item Number Part Number Quantity Description


1 144122 1 Overlay, caregiver, rh, NC, BC, Std, Ent,
air
2 144120 1 Overlay, caregiver, rh, BC, air
3 144133 1 Overlay, caregiver, rh, NC, BC, Std, Ent
4 144121 1 Overlay, caregiver, rh, NC, BC
5 144123 1 Overlay, caregiver, rh, BC
6 144134 1 Overlay, patient, rh, NC, BC, Ent, UTV,
air
7 144128 1 Overlay, patient, rh, NC, BC, Ent, air
8 144127 1 Overlay, patient, rh,
9 144124 1 Overlay, patient, rh, BC, air
10 144130 1 Overlay, patient, rh, NC, BC, Ent, UTV
11 144126 1 Overlay, patient, rh, NC, BC, Ent,
12 144125 1 Overlay, patient, rh, NC, BC
13 144129 1 Overlay, patient, rh, BC
14 144142 1 Overlay, patient, lh, NC, BC, Ent, UTV,
air
15 144139 1 Overlay, patient, lh, NC, BC, Ent, air
16 144138 1 Overlay, patient, lh,
17 144135 1 Overlay, patient, lh, BC, air
18 144141 1 Overlay, patient, lh, NC, BC, Ent, UTV
19
20
144137
144136
1
1
Overlay, patient, lh, NC, BC, Ent,
Overlay, patient, lh, NC, BC
5
21 144140 1 Overlay, patient, lh, BC
22 144149 1 Overlay, caregiver, lh, NC, BC, Std, Ent,
air
23 144143 1 Overlay, caregiver, lh, BC, air
24 144145 1 Overlay, caregiver, lh, NC, BC, Std, Ent
25 144144 1 Overlay, caregiver, lh, NC, BC
26 144150 1 Overlay, caregiver, lh, BC

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 77
Service Manual (MAN272 REV 3)
Label (Sheet 2 of 4)—P3700B
Chapter 5: Parts List

Label (Sheet 2 of 4)—P3700B


Figure 5-30. Label (Sheet 2 of 4)—P3700B

Page 5 - 78 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Label (Sheet 2 of 4)—P3700B
Chapter 5: Parts List

Table 5-30. Label (Sheet 2 of 4)—P3700B

Item Number Part Number Quantity Description


1 6399301 1 Wiring diagram
2 128992 1 Patent label
3 6702000 1 Electrical warning
4 67020 1 Shock (international beds)
5 66870 1 Battery warning
6 6399601 1 Trend, lh
7 45836 4 Foot clear warning
8 4583600 Foot clear warning (international beds)
9 144152 2 Logo
10 144153 1 Brake/steer, lh
11 144160 1 Company
12 144154 1 Brake/steer, rh
13 45834 2 Foot support warning
14 4583400 2 Foot support warning (international beds)
15 65989 1 Battery caution
16 6583102 1 Auto fuse
17 6399602 1 Trend, rh

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 79
Service Manual (MAN272 REV 3)
Label (Sheet 3 of 4)—P3700B
Chapter 5: Parts List

Label (Sheet 3 of 4)—P3700B


Figure 5-31. Label (Sheet 3 of 4)—P3700B

Page 5 - 80 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Label (Sheet 3 of 4)—P3700B
Chapter 5: Parts List

Table 5-31. Label (Sheet 3 of 4)—P3700B

Item Number Part Number Quantity Description


1 68537 2 Volume, blank
2 6665801 2 Speaker, blank
3 65833 2 Siderail warning
4 6583300 2 Siderail warning (international beds)
5 63995 2 CPR
6 6709011 2 Foot support safe working load
7 144155 1 Calf support release, lh
8 6709003 2 Calf support safe working load
9 144158 1 Calf support, rh
10 144156 1 Calf support, lh
11 144157 1 Calf support release, rh
12 4583200 2 Warning

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 81
Service Manual (MAN272 REV 3)
Label (Sheet 4 of 4)—P3700B
Chapter 5: Parts List

Label (Sheet 4 of 4)—P3700B


Figure 5-32. Label (Sheet 4 of 4)—P3700B

Page 5 - 82 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Label (Sheet 4 of 4)—P3700B
Chapter 5: Parts List

Table 5-32. Label (Sheet 4 of 4)—P3700B

Item Number Part Number Quantity Description


1 66859 1 Lockout
2 46770 1 CE
3 6274006 1 Serial number
4 144692 1 Hold and push to stow
5 45832 2 Latch warning
6 4583200 2 Latch warning (international beds)
7 144151 1 Lift-off
8 144693 2 No step
9 45830 Caution
10 4583000 1 Caution (international beds)
11 144822 1 Transport label

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 5 - 83
Service Manual (MAN272 REV 3)
Label (Sheet 4 of 4)—P3700B
Chapter 5: Parts List

NOTES:

Page 5 - 84 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Chapter 6
General Procedures

Cleaning

WARNING:
Follow the product manufacturer’s instructions. Failure to do so could
result in personal injury or equipment damage.

SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.

SHOCK HAZARD:
Do not expose the unit to excessive moisture which would allow liquid
to pool. Personal injury or equipment damage could occur.

CAUTION:
Do not use harsh cleansers or detergents such as scouring pads and
heavy-duty grease removers, or solvents such as toluene, xylene, and
acetone. Equipment damage could occur.

CAUTION:
Ensure that the metal platform is dry before placing the mattress back
onto the bed. Failure to do so could result in equipment damage.

If there is no visible soilage with possible body fluids, we recommend that you
clean the unit with a mild detergent and warm water. If disinfection is desired,
you may use a combination cleanser/disinfectant as explained in
“Disinfecting” on page 6-2. In either case, ensure that the metal platform is dry
6
before placing the mattress back onto the bed.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 1
Service Manual (MAN272 REV 3
Cleaning
Chapter 6: General Procedures

Steam Cleaning
Do not use any steam cleaning device on the unit. Excessive moisture can
damage mechanisms in this unit.

Cleaning Hard to Clean Spots


To remove difficult spots or stains, we recommend that you use standard
household cleansers and a soft bristle brush. To loosen heavy, dried-on soil,
you may first need to saturate the spot.

Disinfecting

WARNING:
Adhere to the Infection Control Policies and Procedures from Hill-Rom.
Failure to do so could result in the spread of infection.
When there is visible soilage and also between patient use, we recommend that
you disinfect the unit using an EPA registered (US only), tuberculocidal,
disinfectant.
Dilute the disinfectant as specified on the manufacturer's label.

Draping the Mattress

CAUTION:
Mattress damage caused by improper draping and/or cleaning
procedures is not covered by warranty.

Correct draping techniques are essential in helping to preserve the life of the
mattress. Drapes must be fluid repellent. The full-size, B7831, labor and
delivery drape effectively covers the lower three-quarters of the bedding
throughout labor. Additional pads or towels placed under the patient will keep
fluid from reaching the edges of the drape. This keeps the sheets clean and dry
while preventing fluid exposure to the mattress.

CAUTION:
Standard OB packs and paper drapes will not keep the sheets dry.

Repeated soaking and scrubbing of mattress materials will accelerate wear and
eventually destroy mattress seals, causing fluids to leak into the cushions.

Page 6 - 2 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Cleaning
Chapter 6: General Procedures

Caring for the Wood Components


Wood is selected for use on beds because of its beauty and warmth. All
Hill-Rom wood products are treated with a resin-based sealer and finish that
provide resistance to abrasion, staining, and fluids. Many
disinfectant/cleansers have a softening effect on any painted or finished
surface if used in high concentrations. Diluted ammonia or detergent, may be
used.

The bed should be cleaned by wiping a soft dampened cloth over the surface,
followed by wiping with a dry cloth. At no time should a wet cloth be allowed
to lay on the surface. Any liquid spilled on the surface should be wiped up
immediately. Any liquid allowed to lie on the surface unattended may damage
the finish.

For protection of the finish, we recommend using a liquid furniture polish.


Polish about once a month, and wipe off any excess with a soft dry cloth. Have
any nicks or scrapes repaired to prevent water damage.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 3
Service Manual (MAN272 REV 3)
Servicing the Bed
Chapter 6: General Procedures

Servicing the Bed

WARNING:
Powered bed mechanisms can cause serious injury. Operate the bed
only with persons clear of mechanisms. Failure to do so could result in
personal injury or equipment damage.

WARNING:
Unplug the bed from its power source and engage the lockout control
during routine maintenance or cleaning. Refer to the Affinity® Three
Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025)
and specific sections in this service manual for additional precautions.
Failure to do so could result in personal injury or equipment damage.

WARNING:
Ensure all electrical/mechanical loads are removed prior to
maintenance/repair of the bed’s drive system or other mechanical
assemblies. Failure to do so could result in personal injury or equipment
damage.

When you work with the bed in the high position, set the brakes and place
2 x 4’s between the base frame and lift arms. This will help prevent injury in
case someone accidentally actuates the bed down switch.

If service on the bed requires that it be placed on its side, be sure to store and
pad the siderails to prevent damage. Also, remove the brake/steer pedal to
prevent damage.

Page 6 - 4 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Component Handling
Chapter 6: General Procedures

Component Handling

CAUTION:
To prevent component damage, ensure that your hands are clean, and
only handle a P.C. board by its edges. Failure to do so could result in
equipment damage.

CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could result in component damage.

CAUTION:
For shipping and storage, place the removed P.C. board in an antistatic
protective bag. Equipment damage can occur.

P.C. Boards
When servicing P.C. boards, follow good handling practices. Mishandling a
P.C. board can cause the following:
• P.C. board damage
• Shortened P.C. board life
• Unit malfunctions

Observe the following P.C. board handling rules:


• Ensure that hands are clean and free of moisture, oily liquids, etc.
• Only handle a P.C. board by its outer edges.
• Do not touch the P.C. board components. Finger contact with the board
surface and/or with its components can leave a deposit that will result in
board (and component) deterioration.
• When working with electronics, wear an appropriate antistatic strap, and
ensure that it is properly grounded.
• Service the removed P.C. board at a static-free workstation that is properly
grounded.
6
• For shipping and storage, place the removed P.C. board in an antistatic
protective bag.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 5
Service Manual (MAN272 REV 3)
Lubrication Requirements
Chapter 6: General Procedures

Lubrication Requirements

WARNING:
Follow the product manufacturer’s instructions. Failure to do so could
result in personal injury or equipment damage.

CAUTION:
Do not use silicone-based lubricants. Equipment damage could occur.

Hill-Rom uses maintenance-free DU bushings at major load and pivot points


throughout the bed. These bushings are designed to run dry without
lubrication. If the bushing squeaks during operation, apply anti-seize
compound to its bearing surface.

WARNING:
Visually inspect the bushings annually. If wear is apparent, replace
them. Failure to do so could result in personal injury or equipment
damage.

The following lubricants can be safely used on the DU bushings:


• P/N SA4269 anti-seize lube (aerosol can)
• P/N 37929 anti-seize lube (brush top lid)

Oilite® bearings and bushings are utilized in several places on the system. By
1

retaining oil, the pores give a self-lubricating quality to the bearings and
bushings. If any silicone-based lubricant is applied to the bearings and
bushings or anywhere else on the system, this self-lubricating quality is
neutralized.

It is safe to apply the following lubricants to the system (see table 6-1 on page
6-6):

Table 6-1. Lubricants

Part Number Description


8252 (100) 2 oz m-1 oil (apply to Oilite® bearings and bushings)
SA3351 (100) 4 ounce lithium grease

1. Oilite® is a registered trademark of Beemer Precision, Incorporated.

Page 6 - 6 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Preventive Maintenance
Chapter 6: General Procedures

Preventive Maintenance

WARNING:
Only facility-authorized personnel should perform preventive
maintenance on the bed. Preventive maintenance performed by
unauthorized personnel could result in personal injury or equipment
damage.

The Affinity® Three Birthing Bed and Affinity® Four Birthing Beds require
an effective maintenance program. We recommend that you perform semi
annual preventive maintenance (PM) along with a quarterly battery check and
testing for Joint Commission on Accreditation of Healthcare Organizations
(JCAHO). PM and testing not only meet JCAHO requirements but will help
ensure a long, operative life for the bed. PM will minimize downtime due to
excessive wear.
The following PM schedule (see table 6-2 on page 6-10) guides you through a
normal PM procedure on the bed. During this PM process, check each item on
the schedule, and make the necessary adjustments.
Follow the PM schedule with the corresponding PM checklist (see table 6-3 on
page 6-13). This checklist is designed to keep a running maintenance history
and subsequent repair costs for one bed. However, your facility can modify
this checklist or design another to fit your needs. Two effective ways to reduce
downtime and ensure the patient remains comfortable are keeping close
records and maintaining the bed.

Inspecting the Pivot Point Fasteners

WARNING:
Inspect the pivot point fasteners semi-annually. Failure to do so could
result in personal injury or equipment damage.

1. Inspect the pivot point fasteners (see figure 6-1 on page 6-8).

2. If the pivot point fasteners (A) are finger-loose, or a gap is visible between
the bolt head (B) and the pivot block plate (C), perform the following steps:
a. Remove the bolt.
6
b. Apply Loctite® adhesive to the bolt threads.
1

c. Reinstall and tighten the bolt.

1. Loctite® is a registered trademark of Loctite Company, Inc.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 7
Service Manual (MAN272 REV 3)
Preventive Maintenance
Chapter 6: General Procedures

Figure 6-1. Pivot Point Location (Intermediate Frame)

Inspecting the Foot Supports


Tools required: Tape measure

The foot support assembly requires periodic inspection and lubrication.


Perform the following to inspect the foot support assembly:

1. Rotate the foot support (B) upward (see figure 6-2 on page 6-9).

2. Ensure that the foot supports (B) latch in each ratchet notch.

3. To ensure proper latching, deliver a solid, downward impact at two to three


places on the foot support (B).

4. Pull the release latch (A) to release the foot support (B).

5. Ensure that the release latch (A) operates smoothly, and that it re-engages
the foot support (B) when released at anytime during the downward
movement.

Page 6 - 8 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Preventive Maintenance
Chapter 6: General Procedures

Figure 6-2. Foot Support Assembly

6. Pull the release latch (A) to swing the foot support (B) outward.

7. To ensure proper engagement of the release latch (A), release the latch two
to three times during the outward motion.

8. Return the foot support (B) to its original position.

9. Remove the foot support (B) from the foot yoke (refer to procedure 4.12).
The removal should be smooth and easy.

10. Check the foot support (B) for any missing, loose, misplaced and wrong
parts.

11. Check the finish on the surfaces of the foot support (B).

12. Ensure that the plastic foot pad sits flush and centered with the foot rest
casting, and that no gaps are evident. 6
13. Ensure that all foot support (B) functions work properly after the
completion of the inspection.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 9
Service Manual (MAN272 REV 3)
Preventive Maintenance
Chapter 6: General Procedures

Preventive Maintenance Schedule

Table 6-2. Preventive Maintenance Schedule

Function Procedure
Siderail controls Check the switches in the siderails to ensure they are functioning
correctly. Also check for intermittent operation. For replacement
instructions, see “Patient Siderail (Inboard) P.C. Switch Board” on
page 4-69.
Siderail frame Check the siderail frame to ensure proper latching and down storage.
Repair as necessary (see “Logic Control P.C. Board Assembly” on
page 4-56).
Caster tires and Check the tires for cuts, wear, tread life, etc.
central brake and
Apply the brake, and check to ensure that the bed will not move. If the
steer
bed moves, inspect it for wear, and adjust if required. See “Caster
Assembly” on page 4-80.
Apply the steering pedal and check the steering to ensure proper
locking action when activated. See “Caster Assembly” on page 4-80.
Communications Test all SideCom® Communication System features (radio, TV, light,
entertainment, and nurse call functions) for proper operation.
Inspect the communication cable for cuts, nicks, or breaks.
Inspect the male pins and female receptacle in the connecting plug.
CPR release Test the CPR release for proper operation and reset of the head drive
system. When the CPR release is pulled, the bed should lower from
any position into a flat position within 7 seconds with at least 50 lb
(23 kg) of weight on the head section.
Electrical test Test the bed for electrical leakage. Ground resistance must be less
than 0.20Ω. Leakage current must be less than 100μA for 120V
models, and less than 150μΑ for 230V models.
Battery test Test the battery voltage at connector P5 at the control P.C. board.
Voltage should be 25V DC to 30V DC.

Page 6 - 10 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Preventive Maintenance
Chapter 6: General Procedures

Table 6-2. Preventive Maintenance Schedule

Function Procedure
Trend-Like function Make sure that the Trend-Like function handles are accessible
limits on both sides of the bed, and check for proper handle functions as
follows:
• Push down on the Trend-Like function handle. The head end
lowers from the level position to 8°, relative to the foot end.
• Pull up on the Trend-Like function handle. The head rises from the
Trend-Like position to the level position.
• Repeat the steps using the Trend-Like function handle on the other
side of the bed.
Hilow limits Operate the hilow section to the full upper and lower limits. See
“Specifications” on page 1-16. Check that the position sensors
function properly.
Head limits Operate the head section to the full upper and lower limits. See
“Specifications” on page 1-16. Check that the position sensors
function properly.
Foot limits Operate the foot section to the full upper and lower limits. See
“Specifications” on page 1-16. Check that the position sensors
function properly.
LED indicators Check all LED indicators on the caregiver control panel for proper
operation.
Night light Check the night light for proper functioning. The sensor for this light
is very sensitive to ambient light and does not turn on the light unless
placed in complete darkness. To test, cover the sensor completely
with a folded dark cloth.
Pendant control Check the pendant control for proper operation. Visually inspect the
cord to ensure it has no cuts, nicks, or breaks. Replace if necessary.
Pivot points Inspect the shoulder bolts through the pivot blocks on the frame (see
figure 6-1 on page 6-8). Apply Loctite® adhesive, and tighten if
a

necessary.
Inspect pivot points for wear, and lubricate if noisy.
a. Loctite® is a registered trademark of Loctite Corporation, Inc.
6

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 11
Service Manual (MAN272 REV 3)
Preventive Maintenance
Chapter 6: General Procedures

Table 6-2. Preventive Maintenance Schedule

Power cord and plug Inspect the power cord and plug for cuts, nicks, or breaks.
Lockout switches Test each switch separately for proper function.
Foot section Inspect for proper slide engagement of the foot section to the foot
yoke casting assembly.
Inspect the Stow and Go™ Foot Section (to include the wire forms,
casters, rollers, and mattress deck) for damage, wear, missing parts,
proper function. Repair as needed.
Air system Test the air system functions (air bladders) for proper inflation and
deflation. Also inspect the mattress for punctures, cuts or tears.
Check the air compressor assembly, hose connections, and "O" rings
for signs of leakage.
Electrical compart­ Inspect the covers for signs of fluid leakage into the electrical com­
ment covers partment.
Foot supports Check the latch/release mechanism for proper operation. Verify
positioning to ensure there is no binding. Ensure the foot support
rotates up to a 85° position. Inspect the foot supports (see “Inspecting
the Foot Supports” on page 6-8). Repair or replace as required.
Calf supports Check the latch/release mechanism for proper operation. Verify free
rotation to ensure there is no binding. When the release handle is
rotated to the right (clockwise), the ball joint should tighten; when
rotated to the left (counterclockwise), the ball joint should loosen.
Repair or replace as required.
Labor grips Check that the labor grips latch properly and are functional.
Headboard Check the headboard appearance and functionality.
Mattress inspection Inspect the ticking for punctures or any other type of compromise.
Inspect the interior of the mattress for any contaminates.
Inspect air bladders (if present) for leaks, and ensure proper air supply
connection.
Overall appearance Touch up the paint where necessary.
Inspect the labels, and replace if necessary.
Check the general aesthetics of the bed.
Observe the symmetry of the bed by viewing it from each corner
while looking toward the opposite corner (i.e., view from left side,
foot end looking toward right side head end). Ensure that the foot and
head are not bent out of alignment.

Page 6 - 12 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Preventive Maintenance
Chapter 6: General Procedures

Preventive Maintenance Checklist

Table 6-3. Preventive Maintenance Checklist

Date
Function

Siderail controls
Manufacturer

Siderail frame
Hill-Rom

Casters
Central brake and steer
Communications
CPR release
Electrical test
Battery test
Trend-Like function
Model Number

limits
Hilow limits
Head limits
Foot limits
LED indicators
Night light
Pendant control
Pivot points
Serial Number

Power cord and plug


Lockout switches
Foot section
Air system
Electrical compartment
covers
Labor Time:
this Page
Total Cost for

Repair Cost:

Inspected by:
6
Remarks:
Applicable
N=Not
O=Okay
Replace
R=Repair or
A=Adjust
C=Clean
L=Lube
Legend

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 13
Service Manual (MAN272 REV 3)
Preventive Maintenance
Chapter 6: General Procedures

Table 6-3. Preventive Maintenance Checklist

Date
Function
Hill-Rom

Foot supports
Manufacturer

Calf supports
Labor grips
Headboard
Mattress inspection
Overall appearance
Model Number
Serial Number

Labor Time:
this Page
Total Cost for

Repair Cost:

Inspected by:
Remarks:
Applicable
N=Not
O=Okay
Replace
R=Repair or
A=Adjust
C=Clean
L=Lube
Legend

Page 6 - 14 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Using Loctite® Adhesive
Chapter 6: General Procedures

Using Loctite® Adhesive


1

The proper use of Loctite® retaining compounds during service is essential to


enable certain fasteners to be correctly retained.

The Loctite® adhesive/sealant retaining compounds, are anaerobic, single-


component resins supplied as a liquid. These resins remain in a liquid state in
the presence of air, but harden into insoluble plastic film solids when confined
between clean, close-fitting metal surfaces. They can be used in conjunction
with natural rubber, glass, ceramics, and thermosetting plastic surfaces, such as
phenolic and polyester. Also, they do not usually harm nylon, polyethylene, or
saran. However, in their liquid state, they sometimes soften or craze vinyl,
cellulosic, styrene, methacrylate plastic, such as Plexiglas® , and varnished
2

finishes. Cured resins do not affect any of these materials. It is important to


note that an opened bottle of these products has a definite shelf-life of about 6
months, which affects its serviceability.

The resins referred to within this manual usually cure to their hardened state on
metal parts without need of primers. However, the use of a recommended
primer will help degrease the part and leave a light deposit of a catalyst to
speed curing. The temperature of the parts also affects their cure time. At 40° F
(4° C) it may take weeks to cure, but at 75° F (24° C), parts cure within the
time length indicated on the bottle.

WARNING:
Use primers with adequate ventilation. Avoid skin contact and
prolonged or repeated breathing of vapors. Do not allow primers to be
trapped under rings, watch bands, etc. Observe all directions on the
primer can. Failure to do so could result in personal injury.

When using the product recommended during the service procedure:

1. Disassemble the parts, remove the old Loctite®, and clean the parts.
Reapply the new Loctite® during reassembly.

2. Apply to clean surfaces only. If necessary, degrease the parts.

3. Use liquid sparingly (usually one or two drops on threaded fasteners).


Be sure to apply where thread engagement begins. 6
1. Loctite® is a registered trademark of Loctite Corporation.
2. Plexiglas® is a registered trademark of Rohm and Haas Company.

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 15
Service Manual (MAN272 REV 3)
Using Loctite® Adhesive
Chapter 6: General Procedures

4. Assemble the parts and allow them to cure. Observe the time stated on
bottle label before placing back into service.

WARNING:
Avoid excessive or repeated skin contact with the liquid. Repeated
contact with the liquid could result in personal injury.

If difficulty is experienced in removal or disassembly of a part with Loctite®


#262 or #271 applied, heating the screw, bolt, nut, etc. aids in its removal.

Purple Loctite® Adhesive #222 (P/N SA4842)


1

Typical usage is as a low strength adhesive for screws. It is also suitable for
sealing threads, fittings, and cylindrical parts. It normally cures in 2 to 6 hours
at 75° F (24° C) and permits disassembly under condition of maximum thread
engagement.

Blue Loctite® Adhesive #242 (P/N SA3618)


Typical usage is as a medium strength adhesive for nuts, screws, keys, and
splines. It is not normally used on fasteners smaller than ¼" (6.4 mm) with
more than ½" (12.7 mm) engagement if removal is desired. It is also suitable
for hydraulic, pneumatic, and pipe fittings. It normally cures in 2 to 6 hours at
75° F (24° C) and permits disassembly under condition of maximum thread
engagement.

Red Loctite® Adhesive #262 (P/N SA4841)


Typical usage is as a medium to high strength adhesive for nuts, screws, etc.
Normally it is not used on fasteners smaller than ¼" (6.4 mm) with more than
½" (12.7 mm) engagement if removal is desired. It normally cures in 0.5 to 2
hours at 75° F (24° C) and permits disassembly under condition of maximum
thread engagement.

Red Loctite® Adhesive #271 (P/N SA4840)


Typical usage is as a high strength adhesive for studs and screws. It is not
normally used on fasteners smaller than 3/8" (9.5 mm) with more than ½"
(12.7 mm) engagement if removal is desired. It is also suitable for hydraulic,
pneumatic, and pipe fittings. It normally cures in 0.5 to 2 hours at 75° F (24°
C) and does not permit easy disassembly.

1. Loctite® is a registered trademark of Loctite Corporation.

Page 6 - 16 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Using Loctite® Adhesive
Chapter 6: General Procedures

NOTES:

Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 6 - 17
Service Manual (MAN272 REV 3)
Using Loctite® Adhesive
Chapter 6: General Procedures

Page 6 - 18 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Chapter 7
Accessories

Accessories
For bed accessories, see table 7-1 on page 7-1.

Table 7-1. Accessories List

Product Number Description


P9625 Anesthesia screen (maximum safe working load of
5 lb (2.3 kg))
P9620 Armboard (maximum safe working load of 30 lb
(14 kg))
P3704 Calf supports, attached (maximum safe working load
of 60 lb (27 kg))
P35745A Calf supports, telescoping (maximum safe working
load of 60 lb (27 kg))
P3612EA Comfort pad
B7831 Disposable drape
P3623 Foley hook kit (maximum safe working load of 20 lb
(9 kg))
P451A Foot rest bar (maximum safe working load of 150 lb
(68 kg)) (Not available with Stow and Go™ Foot Sec­
tions)
P7625C Full leg supports, attached (maximum safe working
load of 60 lb (27 kg))
P7634C Full leg supports, telescoping (maximum safe working
load of 60 lb (27 kg))
P278, P27802 Instrument tray (lift/slide) (maximum safe working
load of 25 lb (11 kg)) (Not available with Stow and
Go™ Foot Sections)
P156 ISS mobile stand kit (maximum safe working load of
40 lb (18 kg))
(Continued on next page.)

7
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 7 - 1
Service Manual (MAN272 REV 3)
Accessories
Chapter 7: Accessories

Table 7-1. Accessories List (cont'd)

Product Number Description


P159 ISS offset bar (maximum safe working load of 40 lb
(18 kg))
P158 ISS transfer pole (maximum safe working load of 40
lb (18 kg))
P3613 Labor bar (maximum safe working load of 100 lb
(45 kg))
P276 O2 tank holder (maximum safe working load of 30 lb
(14 kg))
P5362 Pendant control
P222101 Permanent IV pole (maximum safe working load of
25 lb (11 kg))
P495 Phone
P496 Phone adaptor
SA1528 Sheets, V-cut (complete)
SA1529 Sheets, straight (complete)
SA4585 Sheets, V-cut (two-piece)
SA4586 Sheets, straight (two-piece)
P3705 Siderail pads
P3628B Slipcover, cut head section, air
P3630EA Slipcover, foot section
P3626B Slipcover, straight edge head section, without air
P3627B Slipcover, straight edge head section, air
P3625B Slipcover, V-cut head section, without air
P3619 Wedge assembly
P3710 Lumbar wedge
P3629B Headboard, plastic, blow-molded
P3617B† Headboard, wood—standard (oak or cherry)
P923617B Headboard, wood—Liberty Hill™ (cherry)
P933617B Headboard, wood—Freedom Hill™ (oak)

† Specify wood finish.

Page 7 - 2 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Accessories
Chapter 7: Accessories

7.1 Labor Bar (P3613)


Tools required: None

Installation

WARNING:
The labor bar is intended to be used in the prescribed manner only.
Failure to use this product as outlined may result in personal injury or
equipment damage.

1. Determine the orientation of the labor bar that is required for the position
of the patient.

2. Insert the labor bar (A) into the leg support sockets (B) between the seat
and foot cushions (see figure 7-1 on page 7-3).

Figure 7-1. Labor Bar

7
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 7 - 3
Service Manual (MAN272 REV 3)
Accessories
Chapter 7: Accessories

Adjustment
Using the foot section controls, raise or lower the labor bar (A) to the desired
height.

NOTE:
The labor bar can be used in either direction for different positions and
purposes.

Removal
To remove the labor bar (A), reverse the installation procedure.

Page 7 - 4 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
7.2 Oxygen Tank Holder (P276)
Chapter 7: Accessories

7.2 Oxygen Tank Holder (P276)


Tools required: None

The oxygen tank holder attaches to the head end of the frame in a vertical
position. The oxygen tank holder accommodates one E-size oxygen tank with a
regulator. The mounting points are located to allow the attached oxygen tank
holders to pivot.

Installation
1. Install the mounting bar vertically into a mounting socket at the head end
of the frame.

2. Place one E-size oxygen tank in the holder.

3. Tighten the holder thumbscrew.

NOTE:
The thumbscrew keeps the oxygen tank from rotating in the holder.

Removal
1. Loosen the thumbscrew that holds the tank securely in the holder.

2. Lift the tank out of the holder.

3. Lift up on the tank holder, and remove it from the mounting sockets.

7
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 7 - 5
Service Manual (MAN272 REV 3)
7.3 Telescoping Calf Supports (P35745A) or Telescoping Full Leg Supports (P7634C)
Chapter 7: Accessories

7.3 Telescoping Calf Supports (P35745A) or Telescoping Full Leg


Supports (P7634C)
Tools required: None

Installation
1. Set the brake/steer pedal to the brake position.

2. Ensure that the bed is out of the Trend-Like position.

3. Using the foot function, position the foot section in a mid-height position.

4. Position and insert the telescoping supports (A) into the mounting post (B)
(see figure 7-2 on page 7-6).

Figure 7-2. Telescoping Calf or Full Leg Supports

5. Ensure that the telescoping supports (A) are fully seated and indexed in the
appropriate position.

NOTE:
The ball joints can be indexed inward or outward for different sized patients.

Page 7 - 6 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
7.3 Telescoping Calf Supports (P35745A) or Telescoping Full Leg Supports (P7634C)
Chapter 7: Accessories

Adjustment
1. Loosen the swivel clamp screws (C) to enable the cradles (D) to be
swiveled to the desired direction.

2. Tighten the swivel clamp screws (C) to secure the cradles (D) in the
desired position.

WARNING:
Insufficient tightening will allow the cradles to slip and lose their original
position. Personal injury could occur.

WARNING:
The supports are intended to be used in the prescribed manner only.
Failure to use this product as outlined may result in personal injury or
equipment damage.

Removal
To remove the telescoping supports, reverse the installation procedure.

7
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 7 - 7
Service Manual (MAN272 REV 3)
7.3 Telescoping Calf Supports (P35745A) or Telescoping Full Leg Supports (P7634C)
Chapter 7: Accessories

NOTES:

Page 7 - 8 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 3)
Wiring Diagram/Schematic—P3700A
Wiring Diagram/Schematic—P3700B
Global Headquarters US
Hill-Rom Company, Inc.
1069 State Route 46 E
Batesville, IN 47006-9167
Tel: 800-445-3720
www.hill-rom.com

US Rental Therapy St. Paul, MN International


Hill-Rom Company, Inc. Hill-Rom Company, Inc. Hill-Rom Company, Inc.
Tel: 800-638-2546 Tel: 651-490-1468 or 800-426-4224 Tel: +1 (0)812 934 8173
www.thevest.com Fax: +1 (0)812 934 7191
www.hill-rom.com
international@hill-rom.com

Australia Belgique/België Canada


Hill-Rom Australia Pty. Ltd. Hill-Rom Medical Services BV Hill-Rom Canada
Tel: +61 (0)2 8814 3000 Tel: +31 (0)347 / 32 35 32 Tel: 800-267-2337
Fax: +61 (0)2 8814 3030 Fax: +31 (0)347 / 32 35 00

Deutschland España
Hill-Rom Shanghai Hill-Rom GmbH Hill-Rom Iberia S.L.
Tel: +86 (0)21 5396 6933 Tel: +49 (0)211 16450 0 Tel: +34 (0)93 685 6009
Fax: +86 (0)21 5383 3136 Fax: +49 (0)211 16450 182 Fax: +34 (0)93 666 5570

France Hong Kong Ireland


Hill-Rom SAS Hill-Rom Asia Ltd. Hill-Rom Ltd.
Tel: +33 (0)2 97 50 92 12 Tel: +852 (0)2297-2395 Tel: +353 (0)1 413 6005
Service: +33 (0)820 01 23 45 Fax: +852 (0)2297-0090 Fax: +353 (0)1 413 6030
Fax: +33 (0)2 97 50 92 00 dublin.sales@hill-rom.com

Italia
Hill-Rom S.p.A. Hill-Rom Japan c/o Hill-Rom Japan
Tel: +39 (0)02 / 950541 Tel: +81 (0)3 5715 3420 Tel: +81 (0)3 5715 3420
Fax: +39 (0)02 / 95328578 Fax: +81 (0)3 5715 3425 Fax: +81 (0)3 5715 3425

Nederland New Zealand Nordic Region:


Hill-Rom Medical Services BV Hill-Rom Australia Pty. Ltd. Sverige, Denmark, Norge
Tel: +31 (0)347 / 32 35 32 Tel: +61 (0)2 8814 3000 Hill-Rom AB
Fax: +31 (0)347 / 32 35 00 Fax: +61 (0)2 8814 3030 Tel: +46 (0)8 564 353 60
Fax: +46 (0)8 564 353 61
se.marketing@hill-rom.com

Österreich Portugal South East Asia


Hill-Rom Austria GmbH Hill-Rom Iberia S.L. Hill-Rom Singapore
Tel: +43 (0)2243 / 28550 Tel: +34 (0)93 685 6009 Tel: +65 (0)6391 1322
Fax: +43 (0)2243 / 28550-19 Fax: +34 (0)93 666 5570 Fax: +65 (0)6391 1324
austria@hill-rom.com