Duo2 Service Manual English
Duo2 Service Manual English
Duo2 Service Manual English
W12003
Duo 2™
Multi Mode Mattress System
Service Manual
Revisions
W12003
First Edition
Printed in Europe
The only product warranty intended by Hill-Rom is the express, written warranty
accompanying the bill of sale to the original purchaser. Hill-Rom makes no other
warranty, express or implied, and in particular, makes no warranty of
merchantability or fitness for a particular purpose.
1 Power Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 9
Blower Does Not Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Blower Does Not Stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Deteq™ Module Control Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 17
Alarm Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 20
2 Air Leak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 22
Alternating Therapy Mode Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 30
CPR Function Not Operating Correctly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 36
Transport Mode Electrical Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 40
Vacuum Mode Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 41
Fan Not Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 42
3 Chapter 3: Theory of Operation
Voltage Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Schematics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Logic P.C. Board Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Logic P.C. Board Layout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5
7
Page iv Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Table of Contents
2 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 25
Deteq™ Module Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 26
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 26
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 27
Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 28
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 28
3 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 28
Main Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 29
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 29
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 30
Heel Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 31
4 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 31
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 32
In-Line Check Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page vii
Table of Contents
1 NOTES:
7
Page viii Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Chapter 1 1
Introduction
Chapter Contents
2
Purpose. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Organisation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Chapter 2: Troubleshooting Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
3
Chapter 3: Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Chapter 4: Removal, Replacement, and Adjustment Procedures . . . . . . . . . . . 1 - 3
Chapter 5: Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Chapter 6: General Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Chapter 7: Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4 4
Typographical Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Enclosed Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
5
Product Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 11
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 12
Electrical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 13
6
Symbol Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 14
Safety tips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 15
Warning and Caution Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 16
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 1
Chapter 1: Introduction
1 NOTES:
7
Page 1 - 2 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Purpose
Chapter 1: Introduction
Purpose
1
This manual provides requirements for the Duo 2™ Multi Mode Mattress System
normal operation and maintenance. It also includes a parts list (in chapter 5) for
ordering replacement components.
Audience
This manual is intended to be used by facility authorised maintenance personnel
2
only. Failure to observe this restriction can result in severe injury to people and
serious damage to equipment.
Organisation
This service manual contains seven chapters. 3
Chapter 1: Introduction
In addition to a brief description of this service manual, chapter 1 also provides a
product overview.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 3
Organisation
Chapter 1: Introduction
1 Chapter 7: Accessories
A list of additional products, that can be used in conjunction with Duo 2™ Multi
Mode Mattress System is available in chapter 7. Installation procedures for these
accessories are also included.
7
Page 1 - 4 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Typographical Conventions
Chapter 1: Introduction
Typographical Conventions
1
This manual contains different typographical conventions designed to enhance
readability and understanding of its content.
• Standard text - used for standard text throughout the manual.
• Boldface text - emphasizes a word or phrase.
• NOTE:—sets apart special information or important instruction clarification. 2
• The symbol below highlights a WARNING or CAUTION:
5
• The symbol below highlights an ELECTRICAL SHOCK HAZARD
WARNING:
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 5
Introduction
Chapter 1: Introduction
1 Introduction
4
Since the development of Clinitron® Air Fluidised Therapy in the 1960’s,
Hill-Rom has been developing and manufacturing a range of products for the
prevention and treatment of pressure ulcers and other complications associated
with immobility.
Duo 2™ Multi Mode Mattress System is the latest innovation to be introduced.
5 For the first time two clinically accepted therapies are combined in a clinically
proven system.
At the touch of a button, you can select either alternating low pressure therapy or
continuous low pressure therapy to suit the changing clinical needs of your patient.
Whichever mode you select, you can be confident that the therapeutic mattress is
automatically regulated at the lowest possible support pressure. This is due to the
6 patented Deteq™ intelligent pressure control system, which constantly senses and
adapts cushion pressures to each individual patient. So whatever the size, weight
or bed position of your patients, you are giving them the most comfortable and
effective therapy available, automatically.
Our prevalence survey experience with over 50,000 patients across Europe shows
that the patient’s heels are particularly at risk of skin breakdown (31% of patients).
Duo 2™ Multi Mode Mattress System incorporates RealHeel™, an ultra low
7 pressure heel zone in both continuous and alternating therapy modes, designed to
treat as well as prevent skin damage in this extremely vulnerable area.
All of these advanced functions are controlled by the most compact, lightweight
and easy to use control pendant. Also incorporated are maximum inflation and
1
transport modes, alarm silence, and CPR rapid deflation, all without the need for
obtrusive air hose attachments.
Duo 2™ Multi Mode Mattress System offers a double layer multi mode support
surface which incorporates a special low pressure zone (the last 8 cushions) for the
patient’s heels. The double layer support surface is 25 cm deep and has a fully
enclosed cover which is moisture vapour permeable. The material possesses a two
way stretch ability helping towards the reduction of "hammocking" and can be
2
safely disinfected between patients.
Deteq™ technology also provides an inbuilt alarm that indicates when the patient
is close to touching the sub-mattress. Should this happen, the patient may be too
heavy for example (greater than 150 kg), the backrest should be lowered until the
alarm stops.
Beneath the mattress is an under mattress composed of 90 mm (4") layer of foam. 3
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 7
Features
Chapter 1: Introduction
1 Features
The standard Duo 2™ Multi Mode Mattress System is delivered complete with:
• A therapeutic mattress with 21 air cushions.
• A Deteq™ control module.
Enclosed Cover
The enclosed cover is made of Protex™ fabric which has low friction, stretches
two ways, and is breathable.
3 The cover contains an additive that is bacteriostatic, fungistatic and antimicrobial.
It can be wiped down and washed.
7
Page 1 - 8 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Specifications
Chapter 1: Introduction
Specifications
1
Intended use
Duo 2™ Multi Mode Mattress System is a therapeutic support specifically
designed to help prevent and treat wounds, especially pressure sores from stage 1
to 4 including surgical grafts in high risk patients.
Product Classification
2
Class IIa medical device in accordance with the Council Directive 93/42/EEC.
Protection type against electric shock:
Class I equipment in accordance with EN 60601-1, 1995 edition.
Protection degree against electric shock:
Defibrillation-proof type B applied parts in accordance with Amendment 2 of
EN 60601-1, 1995 edition.
3
Protection of mattress against harmful ingress of water:
Equipment protected against powerful water jets, IPX6 in accordance with EN
60529, 1991 edition.
Protection of control pendant against harmful ingress of water:
Ordinary equipment (enclosed equipment without protection against ingress of
water), IPX0 in accordance with IEC 529. 4
Radio disturbance characteristics:
Class B equipment in accordance with EN 55011, 1998 edition.
Flammability rating for high risk locations:
Class D mattress in accordance with recommendation Groupes Permanents
d'Etude des Marchés/Chimiques et Para-chimiques (GPEM/CP) N° D1-90, 1991
edition. 5
Conformity
The conformity of Duo 2™ Multi Mode Mattress System with the essential
requirements of the Council Directive 93/42/EEC is demonstrated by compliance
with the following product safety standards:
EN 60601-1, 1995 edition:
Medical electrical equipment - Part 1: General requirements for safety.
6
EN 60601-1-2, 1995 edition:
Medical electrical equipment - Part 1: General requirements for safety - Part 2:
Collateral standard: Electromagnetic compatibility - requirements and tests.
EN 61000-3-2, 1995 edition:
Electromagnetic compatibility (EMC) - Part 3: Limits - Section 2: Limits for
harmonic current emissions (equipment input current lower or equal to 16 A per
phase).
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 9
Specifications
Chapter 1: Introduction
7
Page 1 - 10 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Specifications
Chapter 1: Introduction
Physical Specifications 1
WARNING:
Please ensure that the furniture (bed frame or seat) used can safely
accommodate the patient and the accessories. Failure to do so could
result in personal injury or equipment damage
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 11
Specifications
Chapter 1: Introduction
1 Environmental Specifications
7
Page 1 - 12 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Specifications
Chapter 1: Introduction
WARNING:
1
Do not operate the equipment in the presence of flammable anaesthetic
mixture with air, oxygen or nitrous oxide. Personal injury or equipment
damage could occur.
Electrical Specifications
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 13
Symbol Quick Reference
Chapter 1: Introduction
3 - Maximum inflation
mode symbol/button
- CPR/STOP button
- Deteq™ status
Diagnostics connection
(for maintenance technician only)
5 Deflate position
Normal position
- Dangerous voltage
- Defibrillation-proof type B applied parts
W12003_003
7
Page 1 - 14 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Safety tips
Chapter 1: Introduction
Safety tips
1
WARNING:
Please ensure that the furniture (bed frame or seat) used can safely
accommodate the patient and the accessories. Failure to do so could result
in personal injury or equipment damage.
2
WARNING:
Do not operate the equipment in the presence of flammable anaesthetic
mixture with air, oxygen or nitrous oxide. Personal injury or equipment
damage could occur.
WARNING:
Only facility-authorised maintenance personnel should troubleshoot the
3
Duo 2™ Multi Mode Mattress System. Troubleshooting by unauthorised
personnel could result in personal injury or equipment damage.
WARNING:
Adhere to the Infection Control Policies and Procedures from Hill-Rom.
Failure to do so could result in the spread of infection. 4
WARNING:
Ensure the frame brake is on to avoid unplanned frame movements.
Failure to do so could result in personal injury.
WARNING:
Follow the product manufacturer’s instructions. Failure to do so could result
5
in personal injury or equipment damage.
WARNING:
Handle the sodium hypochlorite (bleach) with care. Dispose of the sodium
hypochlorite in accordance with the proper disposal procedure as specified
by the local regulating authority. Failure to do so could result in personal
injury.
6
SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
personal injury or equipment damage. 7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 15
Safety tips
Chapter 1: Introduction
1 SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
2 SHOCK HAZARD:
Do not expose the unit to excessive moisture that would allow for liquid
pooling. Personal injury or equipment damage could occur.
CAUTION:
Following prolonged storage of the mattress, it is advisable to select
maximum inflation mode to allow the mattress to expand its cushions fully.
3 Allow the mattress to run in maximum inflation mode for two minutes then
select the therapy mode you wish. Failure to do so could result in
incomplete inflation of the mattress.
CAUTION:
The air filter door will not close if the lever is not returned to the horizontal
position. Damage to the door or its clasps may occur if it is forced.
4 CAUTION:
Do not use harsh cleansers/detergents such as scouring pads and heavy
duty grease removers or solvents such as acetone. Equipment damage
could occur.
5 CAUTION:
The control pendant is not protected against ingress of water. Disconnect it
before cleaning. Failure to do so could result in equipment damage.
CAUTION:
Avoid the use of cleaning solutions containing Dimethyl formamide,
Tetrahydrofuran, Phenol, Cresol or concentrated inorganic acids, as these
CAUTION:
Avoid highly alkaline and strong organic based detergents and
disinfectants, as these chemicals may attack the cushion fabric dye,
causing transfer of colour.
7
Page 1 - 16 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Safety tips
Chapter 1: Introduction
CAUTION:
1
Do not use hypochlorite (bleach) above a concentration of 500 ppm as this
will discolour and weaken the fabric. Failure to do so could result in
equipment damage.
CAUTION:
The mattress is protected against a jet of water from a standard nozzle of
12.5 mm diameter at a distance 2.5 m to 3.0 m, with a delivery rate 100
2
l/min, spraying the mattress from all practical directions. Check the
capability of the cleaning equipment prior to using it, otherwise damage to
the mattress may occur.
CAUTION:
Do not use phenolics on the Protex™ cover. This could result in damage to
equipment and loss of waterproof qualities.
3
CAUTION:
Do not use hypochlorite (bleach) above a concentration of 500 ppm as this
will discolour and weaken the fabric.
CAUTION:
4
Mattress connections under the therapeutic mattress must be sealed (see
“Tool and Supply Requirements” on page 6-15) prior to laundry. Failure to
do so could result in equipment damage.
CAUTION:
Ensure the mattress is completely dry before storage. Failure to do so
could result in equipment damage.
5
CAUTION:
Do not mangle the Protex™ cover. Equipment damage could result.
CAUTION:
Do not use heat above 67°C (153°F) as this may damage the urethane
6
lining and heat seals of the cushion.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 1 - 17
Warning and Caution Labels
Chapter 1: Introduction
6
W12003_004
7
Page 1 - 18 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Chapter 2 1
Troubleshooting Procedures
Chapter Contents
2
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Initial Actions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Function Checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Final Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 7
Power Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 9
3
Blower Does Not Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Blower Does Not Stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Deteq™ Module Control Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 17
Alarm Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 20
Air Leak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 22 4
Alternating Therapy Mode Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 30
CPR Function Not Operating Correctly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 36
Transport Mode Electrical Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 40
Vacuum Mode Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 41
Fan Not Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 42 5
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 1
Chapter 2: Troubleshooting Procedures
1 NOTES:
7
Page 2 - 2 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Getting Started
Chapter 2: Troubleshooting Procedures
Getting Started
1
WARNING:
Only facility-authorised maintenance personnel should troubleshoot the
Duo 2™ Multi Mode Mattress System. Troubleshooting by unauthorised
personnel could result in personal injury or equipment damage.
Begin each procedure in this chapter with step 1. Follow the sequence outlined
2
(each step assumes the previous step has been completed). In each step, the normal
operation of the product can be confirmed by answering Yes or No to the
statement. Your response will lead to another step in the procedure, a repair
analysis procedure (RAP), or a component replacement. If more than one
component is listed, replace them in the given order. If the problem still exists,
replace the next component listed, and try to operate the system again.
Start with Initial Actions to begin gathering information about the problem.
3
Perform the Function Checks to isolate or identify a problem and to verify the
repair after completing each corrective action (replacing or adjusting a part,
seating a connector, etc.).
When reading the voltage, the first test point is for the RED (+) digital multi-meter
(DMM) lead, and the second entry is for the BLACK (-) DMM lead. If no second
entry exists, the BLACK DMM lead must be connected to any good power supply
4
ground or zero volt test point, TP 3, on the logic P.C. board (see figure 3-3 on page
3-5).
Perform the Final Actions after the Function Checks to verify the repair. 5
If troubleshooting procedures do not isolate the problem, call your national
Hill-Rom representative (see back cover).
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 3
Initial Actions
Chapter 2: Troubleshooting Procedures
1 Initial Actions
Use Initial Actions to gather information from operators concerning problems with
the Duo 2™ Multi Mode Mattress System. Note the symptoms or other
information concerning the problem that the operator describes. This information
helps identify the probable cause.
2. Ask that person to demonstrate or explain the problem. The problem can be
duplicated.
Yes No
4. Instruct the operator to refer to the procedures in the Duo 2™ Multi Mode
4 Mattress System User Manual. Perform the “Function Checks” on page 2-5 to
ensure proper operation of the Duo 2™ Multi Mode Mattress System.
Problem Solution
Power Malfunction RAP 2.1
7
Page 2 - 4 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Function Checks
Chapter 2: Troubleshooting Procedures
Function Checks
1
1. Initial Actions have been performed.
Yes No
↓ → Go to “Initial Actions” on page 2-4.
5. The mattress continues to inflate. The Deteq™ module status LED changes
from red to green. 5
Yes No
↓ → Go to RAP 2.4.
7. Connect a manometer to the main manifold test port. Unplug the system from
6
its power source and note the value. The transport mode LED illuminates.
Yes No
↓ → Go to RAP 2.9.
8. Wait 5 min, and read the manometer again. The value remains within
1" H2O (1.9 mm Hg) of the original reading.
Yes
↓
No
→ Go to RAP 2.6.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 5
Function Checks
Chapter 2: Troubleshooting Procedures
1 9. Plug the system into an appropriate power source with the system still in
maximum inflation mode. Disconnect the manometer, and seal the test port.
Connect the manometer to the heel zone test port. Unplug the system from its
power source, and note the value. Wait 5 min and read the manometer again.
The value remains within 1" H2O (1.9 mm Hg) of the original reading.
Yes No
↓ → Go to RAP 2.6.
2 10. Plug the system into an appropriate power source and select alternating therapy
mode. Place a weight on the mattress or lie on the mattress (place a hospital
sheet on mattress to protect the mattress cover). The first cushion at the head
end remain inflated; the next 12 cushions operate in a one-in-four alternating
cycle; the final eight cushions operate in a one-in-two alternating cycle.
Yes No
↓ → Go to RAP 2.7.
3 11. Press the CPR/STOP button.
The mattress deflates within 20 s.
Yes No
↓ → Go to RAP 2.8.
6 NOTE:
If your problem is not listed, call Hill-Rom Technical Support (see the rear cover
for your national Hill-Rom representative).
7
Page 2 - 6 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Final Actions
Chapter 2: Troubleshooting Procedures
Final Actions
1
1. Complete the required interval maintenance (routine, pre-installation)
procedures. See “Pre-Installation Maintenance” on page 6-3 and “Routine
Maintenance” on page 6-4.
3. At the customer site, notify the user that work has been completed on the
2
system.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 7
Final Actions
Chapter 2: Troubleshooting Procedures
Alternating solenoids:
EV2; EV3; EV4; EV5
2 Compressor on
Heel Manifold - fill
3 Bottoming
≥ 22" H2O (41 mm Hg)
Deteq™ module sensor
fault or hardware
>15 minute time-out malfunction
Battery test
CPR active
+12VDC = Yellow = Green
Colour key:
Blower stop @ ≈ 6" H2O
4 Not used = Red = Off
(11 mm Hg)
W12003_005
Notes:
• The Leak LED indicates that the compressor has run continuously for more than 15 min
which indicates that the system has a leak or a hardware fault causing the compressor to
run continuously.
• The Bottoming LED indicates that the patient is not supported correctly, and is bottom-
ing for more than 1 minute.
5 • The Deteq™ module sensor fault or hardware malfunction LED indicates that the servo
control on the logic P.C. board has detected a discontinuity in the coil circuit or a logic
P.C. board hardware fault.
• The max inflate/22" H2O (41 mm Hg) LED indicates when the main manifold is at or
above 22" H2O (41 mm Hg).
7
Page 2 - 8 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.1 Power Malfunction
Chapter 2: Troubleshooting Procedures
3. Inspect the power cable and connectors for loose connections, cuts, or damage
to the cable. The power cable and connectors are in good condition.
Yes No
3
↓ → Repair or replace the power cable.
SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
5
care when working close to connections. Failure to do so could result in
personal injury or equipment damage.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 9
2.1 Power Malfunction
Chapter 2: Troubleshooting Procedures
3 B
A
W12003_006
4 Yes
↓
No
→ Go to step 7.
SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
6 C
C
7
Page 2 - 10 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.1 Power Malfunction
Chapter 2: Troubleshooting Procedures
W12003_007
7. Set the DMM to the DC volts scale. Measure the voltage on J7 of the power
supply P.C. board at pins 1 and 9 for 12VDC (see figure 3-6 on page 3-8) and
5
(see figure 3-7 on page 3-8). The voltage is 12VDC.
Yes No
↓ → Check the fuse F4 for continuity, replace the damaged fuse (see
figure 2-4 on page 2-11), and test.
8. Set the DMM to the DC volts scale. Measure the voltage on power supply P.C.
board at TP1 and TP2 for 13.8VDC (see figure 3-6 on page 3-8) and (see
figure 3-7 on page 3-8).
6
The voltage is 13.8VDC.
Yes No
↓ → Check the fuse F3 for continuity, replace the damaged fuse (see
figure 2-4 on page 2-11), and test. If this solves the problem, go to
“Final Actions” on page 2-7. Otherwise, go to step 9.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 11
2.1 Power Malfunction
Chapter 2: Troubleshooting Procedures
1 9. Set the DMM to the DC volts scale. Measure the voltage between between TP2
and TP3 on the logic P.C. board for 12VDC and TP3 and TP1 for 13.8VDC
(see figure 3-3 on page 3-5).
Both voltages are present.
Yes No
↓ → Replace the interconnection cable assembly (J7 to J12) (C) (see
figure 2-5 on page 2-12). If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, go to step 10.
5
W12003_006
6 Yes
↓
No
→ Replace the logic P.C. board (refer to procedure 4.8). If this solves
the problem, go to “Final Actions” on page 2-7. Otherwise, call
Hill-Rom Technical Support (see the rear cover for your national
Hill-Rom Representative).
7
Page 2 - 12 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.2 Blower Does Not Start
Chapter 2: Troubleshooting Procedures
2
SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
personal injury or equipment damage.
c. Plug the system into an appropriate power source.
d. The 6" H2O (11 mm Hg) yellow LED does not come on.
The blower does not start.
3
Yes No
↓ → Check fuse F1 for continuity, replace the damaged fuse (see figure
2-4 on page 2-11) and test.
3. Replace the blower and test. It starts. This solves the problem.
Yes No
↓ → Go to step 5.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 13
2.2 Blower Does Not Start
Chapter 2: Troubleshooting Procedures
8. Replace the power supply P.C. board (refer to procedure 4.7). If this solves the
problem, go to “Final Actions” on page 2-7. Otherwise, go to step 9.
10. Disconnect the control pendant and check for 12VDC on the power supply P.C.
4 board, between TP2 and the anode of diode D12. Voltage is 12VDC.
Yes No
↓ → Replace the interconnection cable assembly (see figure 2-5 on page
2-12), and test. This solves the problem.
11. Replace the power supply P.C. board (refer to procedure 4.7) and test. This
solves the problem.
5 Yes
↓
No
→ Call Hill-Rom Technical Support (see the rear cover for your
national Hill-Rom Representative).
7
Page 2 - 14 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.3 Blower Does Not Stop
Chapter 2: Troubleshooting Procedures
2. Go to step 5.
5. Replace the vacuum valve microswitch (refer to procedure 4.12). This solves
6
the problem.
Yes No
↓ → Go to step 7.
8. Replace the power supply PC board and test. This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support (see the rear cover for your
7
Page 2 - 16 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.4 Deteq™ Module Control Fault
Chapter 2: Troubleshooting Procedures
2
SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
personal injury or equipment damage.
c. Plug the system into an appropriate power source.
On the diagnostics tester, the Deteq™ module sensor LED illuminates red.
Yes No
3
↓ → Go to RAP 2.6.
W12003_009 7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 17
2.4 Deteq™ Module Control Fault
Chapter 2: Troubleshooting Procedures
5 W12003_010
6 Yes
↓
No
→ Replace the Deteq™ module coil socket cable harness (D) J9 to J14.
If this solves the problem, go to“Final Actions” on page 2-7.
Otherwise, go to step 4.
5. Call Hill-Rom Technical Support (see the rear cover for your national Hill-
Rom representative).
1
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 19
2.5 Alarm Malfunction
Chapter 2: Troubleshooting Procedures
5 W12003_011
SHOCK HAZARD:
6 Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
personal injury or equipment damage.
e. Plug the system into an appropriate power source.
f. Select maximum inflation mode.
The audible alarm sounds.
7 Yes
↓
No
→ Go to step 3.
5
W12003_012
4. Replace the pendant P.C. board and test (refer to procedure 4.10). This solves
the problem.
Yes
↓
No
→ Call Hill-Rom Technical Support (see the rear cover for your
6
national Hill-Rom representative).
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 21
2.6 Air Leak
Chapter 2: Troubleshooting Procedures
2. Replace the damaged mattress component, and test. If this solves the problem,
go to “Final Actions” on page 2-7. Otherwise, go to step 3.
3 b. Remove the cover of the main test port (A), and connect a manometer (see
Figure 2-10 on page 2-22).
5
W12003_013
6 e. Wait 5 min.
The value drops by more than 1" H2O (1.9 mm Hg).
Yes No
↓ → The system is within specification. If this solves the problem, go to
“Final Actions” on page 2-7. Otherwise, go to step 4.
C
c. Open the ASE (refer to procedure 4.1).
W12003_014 4
SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
personal injury or equipment damage. 5
d. Plug the system into an appropriate power source, and select maximum
inflation mode.
NOTE:
The compressor, after an initial operation of 1 s or 2 s, should not operate for at
least 1 min.
e. Verify either by listening for the click of the relay on the power supply P.C.
board, or by carefully placing your hand on the body of the compressor and
6
checking for vibration. The compressor or the relay operates within 1 min.
Yes No
↓ → The air leak must be on the mattress. Go to step 7.
1 b. Clamp the supply tube from the main manifold to the heel manifold with an
arterial clamp (D) (see Figure 2-12 on page 2-24).
7
Page 2 - 24 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.6 Air Leak
Chapter 2: Troubleshooting Procedures
From E
blower
3
To manual vacuum valve
W12003_016
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 25
2.6 Air Leak
Chapter 2: Troubleshooting Procedures
3 F
W12003_017
d. Remove the cover of the main test port (A) and connect a manometer (see
f. After 1 min the drop in the reading is < 0.5" H2O (0.9 mm Hg).
Yes No
1
↓ → Repair or replace the head zone of the therapeutic mattress, and test.
If this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 9.
4
H
G
W12003_018
b. Plug the system into an appropriate power source and select maximum
5
inflation mode.
c. Allow the cushions to fully inflate.
d. Unplug the system from its power source.
e. Check for a drop in the manometer reading.
f. After 1 min the drop in the reading is < 0.5" H2O (0.9 mm Hg).
Yes No
6
↓ → Repair or replace Zones 1 and 3 of the therapeutic mattress, and test.
If this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 10.
3 J
I
W12003_019
b. Plug the system into an appropriate power source and select maximum
4 inflation mode.
c. Allow the cushions to fully inflate.
d. Unplug the system from its power source.
e. Check for a drop in the manometer reading.
f. After 1 min the drop in the reading is < 0.5" H2O (0.9 mm Hg).
Yes No
7
Page 2 - 28 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.6 Air Leak
Chapter 2: Troubleshooting Procedures
K
K
3
W12003_020
c. Plug the system into an appropriate power source and select maximum
inflation mode.
d. Allow the cushions to fully inflate.
4
e. Unplug the system from its power source.
f. Check for a drop in the manometer reading.
g. After 1 min the drop in the reading is < 0.5" H2O (0.9 mm Hg).
Yes No
↓ → The air leak must be in heel zone of the therapeutic mattress. Repair
or replace, and test. If this solves the problem, go to “Final Actions” 5
on page 2-7. Otherwise, call Hill-Rom Technical Support (see the
rear cover for your national Hill-Rom representative).
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 29
2.7 Alternating Therapy Mode Fault
Chapter 2: Troubleshooting Procedures
3
Head end
4
W12003_021
b. The 5 kg (11 lb) weight (B) centred on top of the second and third cushions
5 of the heel zone, i.e. over the “O” in the “Duo 2™” logo.
7
Page 2 - 30 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.7 Alternating Therapy Mode Fault
Chapter 2: Troubleshooting Procedures
2
W12003_022
• When the mode changes to alternating, the LED for zone 1 lights up (see
figure 2-20 on page 2-31).
4
W12003_023
• After 1 min 30 s, the LEDs for all four zones are off (see figure 2-21 on
page 2-31).
W12003_022
6
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 31
2.7 Alternating Therapy Mode Fault
Chapter 2: Troubleshooting Procedures
1 • After 2 min 30 s, the LED for zone 4 lights up (see figure 2-22 on page
2-32).
W12003_026
3 • After 4 min, the LEDs for all four zones are off (see figure 2-23 on page
2-32).
4
W12003_022
• After 5 min, the LED for zone 3 lights up (see figure 2-24 on page 2-32).
5 Figure 2-24. Alternating Therapy Mode T+5 min
6
W12003_025
7
Page 2 - 32 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.7 Alternating Therapy Mode Fault
Chapter 2: Troubleshooting Procedures
• After 6 min 30 s, the LEDs for all four zones are off (see figure 2-25 on
page 2-33).
1
Figure 2-25. Alternating Therapy Mode T+6 min 30 s
W12003_022
• After 7 min 30 s, the LED for zone 2 lights up (see figure 2-26 on page
2-33). 3
Figure 2-26. Alternating Therapy Mode T+7 min 30 s
4
W12003_024
• After 9 min, the LEDs for all four zones are off (see figure 2-27 on page
2-33).
6
W12003_022
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 33
2.7 Alternating Therapy Mode Fault
Chapter 2: Troubleshooting Procedures
1 • After 10 min, the LED for zone 1 lights up (see figure 2-28 on page 2-34).
W12003_023
3 Yes
↓
No
→ Replace the logic P.C. board and test (refer to procedure 4.8). If this
solves the problem, go to “Final Actions” on page 2-7. Otherwise,
go to step 3.
6. Locate connector J10 (see fold out FO 3-1 at the rear of the manual), and
disconnect.
7. Use a DMM set to resistance measurement and connect directly to the solenoid
coil you wish to test. The resistance is approximately 36 Ω.
5 Yes
↓
No
→ Replace the solenoid coil and test (refer to procedure 4.9). If this
solves the problem, go to “Final Actions” on page 2-7. Otherwise,
go to step 8.
6 SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
personal injury or equipment damage.
a. Plug the system into an appropriate power source.
b. Select the alternating therapy mode.
7 c. Wait until the LED corresponding to the faulty zone turns on.
d. Using a DMM set to VDC, connect directly to the solenoid coil you wish to
test (see fold out FO 3-1 at the rear of the manual).
1
The voltage is 6.5VDC (±10%).
Yes No
↓ → Go to step 10
9. Replace the solenoid coil and test (refer to procedure 4.9). If this solves the
problem, go to “Final Actions” on page 2-7. Otherwise, call Hill-Rom
Technical Support (see the rear cover for your national Hill-Rom
2
Representative).
10. Using a DMM set to VDC, measure the voltage on the appropriate contacts of
J10 on the logic P.C. board (see fold out FO 3-1 at the rear of the manual).
The voltage is 6.5VDC (±10%).
Yes
↓
No
→ Replace the logic P.C. board and test (refer to procedure 4.8). If this
solves the problem, go to “Final Actions” on page 2-7.
3
11. Replace the main manifold ribbon cable (see figure 5-11 on page 5-22) and
test. This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support (see the rear cover for your
national Hill-Rom Representative).
4
12. Go to “Final Actions” on page 2-7.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 35
2.8 CPR Function Not Operating Correctly
Chapter 2: Troubleshooting Procedures
B A
3
Head end
4
W12003_021
b. The 5 kg (11 lb) weight (B) centred on top of the second and third cushions
5 of the heel zone, i.e. over the “O” in the “Duo 2™” logo.
7
Page 2 - 36 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.8 CPR Function Not Operating Correctly
Chapter 2: Troubleshooting Procedures
• During the first 45 s, the LEDs for the compressor, the main manifold, and heel
manifold are yellow, the LEDs for the battery +12V is green, and the LED for
1
CPR is red (see figure 2-30 on page 2-37).
= Yellow
= Red
= Green
= Off W12003_027
4
• Forty five seconds after activating CPR, the LEDs for the main manifold and
heel manifold are off (see figure 2-31 on page 2-38).
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 37
2.8 CPR Function Not Operating Correctly
Chapter 2: Troubleshooting Procedures
3
= Yellow = Green
= Red = Off W12003_028
• After more than 45 s, the LED for bottoming has changed to red
4 (see figure 2-32 on page 2-38).
= Yellow = Green
= Red = Off
7 W12003_029
• After more than 15 min, the LED for leak has changed to red (see figure 2-33
on page 2-39).
1
Figure 2-33. CPR: T > 15 min
= Yellow
= Red
= Green
= Off W12003_030 4
3. The sequence of LEDs is followed within ± 10 % of the noted amount of time.
Yes No
↓ → Replace the logic P.C. board and test (refer to procedure 4.8). If this
solves the problem, go to “Final Actions” on page 2-7. Otherwise,
go to step 4.
4. Call Hill-Rom Technical Support (see the rear cover for your national Hill- 5
Rom Representative).
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 39
2.9 Transport Mode Electrical Fault
Chapter 2: Troubleshooting Procedures
3 ↓ → Replace the battery in the ASE and test (refer to procedure 4.11). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 3.
7
Page 2 - 40 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
2.10 Vacuum Mode Fault
Chapter 2: Troubleshooting Procedures
SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
personal injury or equipment damage.
5
a. Plug the system into an appropriate power source.
b. Using a DMM set to DC volts, connect the - lead to TP3 on the logic P.C.
board and the + lead to pin 2 of J8 on the power supply P.C. board.
The voltage is 12VDC.
Yes
↓
No
→ Replace the power supply P.C. board and test (refer to procedure
6
4.7). If this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 5.
5. Call Hill-Rom Technical Support (see the rear cover for your national Hill-
Rom Representative).
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 2 - 41
2.11 Fan Not Operating
Chapter 2: Troubleshooting Procedures
SHOCK HAZARD:
Voltage is present on exposed electrical connectors when power is on. Use
care when working close to connections. Failure to do so could result in
personal injury or equipment damage.
b. Plug the system into an appropriate power source.
c. Disconnect J9 from the logic P.C. board.
d. Using a DMM set to DC volts, connect the - lead to TP3 on the logic P.C.
3 board and the + lead to pin 5 of J9 on the logic P.C. board.
The voltage is 12VDC.
Yes No
↓ → Replace the logic P.C. board and test (refer to procedure 4.8). If this
solves the problem, go to “Final Actions” on page 2-7. Otherwise,
go to step 2.
4 2. Replace the cooling fan (refer to procedure 4.13). This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support (see the rear cover for your
national Hill-Rom Representative).
7
Page 2 - 42 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Chapter 3 1
Theory of Operation
Chapter Contents
2
Voltage Distribution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Logic P.C. Board Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Logic P.C. Board Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5
Control Pendant P.C. Board Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
3
Control Pendant P.C. Board Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7
Power Supply P.C. Board Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8
Power Supply P.C. Board Schematic. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8
Diagnostics Tester Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 9
Pneumatic Flow Charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 10 4
Subsystem Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 17
Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 18
Vacuum Mode Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 18
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 18
Power Supply P.C. Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 20
Logic P.C. Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 20
5
Control Pendant P.C. Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 20
Alarms (Audible and Visual) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 21
Pneumatic. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 23
Double Layer Mattress. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 23
6
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 1
Chapter 3: Theory of Operation
1 NOTES:
7
Page 3 - 2 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Voltage Distribution
Chapter 3: Theory of Operation
Voltage Distribution
1
Figure 3-1. AC and DC Voltage Distribution and Plug Jack Layout
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 3
Schematics
Chapter 3: Theory of Operation
1 Schematics
7
Page 3 - 4 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Schematics
Chapter 3: Theory of Operation
6
W12003_031
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 5
Schematics
Chapter 3: Theory of Operation
4 W12003_032
7
Page 3 - 6 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Schematics
Chapter 3: Theory of Operation
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 7
Schematics
Chapter 3: Theory of Operation
4 W12003_033
CPR on
5 Pendant on
Low Voltage
6 Blower
Blower on
Stop blower
High Voltage
Pump
Stop CPR
Stop P.Max
Cde pump
7
Vacuum
Vacuum on
W12003_034
W12003_035
4
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 9
Pneumatic Flow Charts
Chapter 3: Theory of Operation
2 Manual
vacuum
Pressure
Sensors
valve
Blower
Safety release
valve 30" H2O
3 Heel manifold
Compressor
Leak Heel Heel Fill
valve zone zone valve
4 Main manifold
5
Zone 1
Zone 2
Zone 3
Zone 4
Deteq™
Deteq™
mattress
mattress
Head end
6
Therapeutic
mattress
Deteq™
mattress W12003_036
7
Page 3 - 10 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Pneumatic Flow Charts
Chapter 3: Theory of Operation
Heel manifold
3
Compressor
Leak Heel Heel Fill
valve zone zone valve
Main manifold
4
Zone 1
Zone 2
Zone 3
Zone 4
Deteq™
Deteq™
mattress
mattress
5
Head end
Therapeutic
mattress
Deteq™
6
W12003_037
mattress
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 11
Pneumatic Flow Charts
Chapter 3: Theory of Operation
2 valve
Blower
Safety release
valve 30" H2O
3 Heel manifold
Main manifold
4
Zone 1
Zone 2
Zone 3
Zone 4
Deteq™
Deteq™
mattress
mattress
5
Head end
6 Therapeutic
mattress
Deteq™
mattress W12003_038
7
Page 3 - 12 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Pneumatic Flow Charts
Chapter 3: Theory of Operation
Start of
Figure 3-12. Alternating Therapy Mode Cycle Logic 1
alternating cycle
Heel zone Head zone
Zone 4
2
3
Zone 3
4
Zone 2
5
Zone 1
End of
alternating cycle = 10 min
Foot Head
6
W12003_039
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 13
Pneumatic Flow Charts
Chapter 3: Theory of Operation
2 valve
Blower
Safety release
valve 30" H2O
3 Heel manifold
Compressor
Leak Heel Heel Fill
valve zone zone valve
Main manifold
4
Zone 1
Zone 2
Zone 3
Zone 4
Deteq™
Deteq™
mattress
mattress
5
Head end
6 Therapeutic
mattress
Deteq™
mattress W12003_040
7
Page 3 - 14 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Pneumatic Flow Charts
Chapter 3: Theory of Operation
Blower
2
Safety release
valve 30" H2O
Heel manifold
3
Leak Heel Heel Fill Compressor
valve zone zone valve
Main manifold
4
Zone 1
Zone 2
Zone 3
Zone 4
Deteq™
Deteq™
mattress
mattress
5
Head end
Therapeutic
mattress
Deteq™
6
W12003_041
mattress
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 15
Pneumatic Flow Charts
Chapter 3: Theory of Operation
2 valve
Blower
Safety release
valve 30" H2O
3 Heel manifold
Compressor
Leak Heel Heel Fill
valve zone zone valve
Main manifold
4
Zone 1
Zone 2
Zone 3
Zone 4
Deteq™
Deteq™
mattress
mattress
5
Head end
6 Therapeutic
mattress
Deteq™
mattress W12003_042
7
Page 3 - 16 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Subsystem Theory
Chapter 3: Theory of Operation
Subsystem Theory
1
The operation of the Duo 2™ Multi Mode Mattress System can be divided into
three categories or subsystems:
• Mechanical
• Electrical
• Pneumatic 2
The combination of functions in these three subsystems provides the operational
capabilities of the system. To perform all operational functions, these three
subsystems interact with each other.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 17
Subsystem Theory
Chapter 3: Theory of Operation
1 Mechanical
The mechanical subsystem is composed of the vacuum mode valve.
The handle of the vacuum mode valve is located behind the air filter door. It is
2 used to quickly deflate the cushions to store the system. The handle opens the
mechanical valve inside the air supply enclosure to let the air pass from the
manifold directly to the blower by the venturi effect. When the mattress is empty,
the handle must be returned to the normal operating position because the air filter
door is designed to remain open if the handle is not in the normal operating
position.
Electrical
3 Controls are kept to a minimum (see figure 3-16 on page 3-18).
4 regulation
Power
status LED Alternating mode
Green = On button and LED
Red = Off
5 Maximum inflate
mode button and
Continuous mode LED
button and LED
Alarm silence CPR/Stop mode
button and LED button and LED
6 W12003_043
7
Page 3 - 18 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Subsystem Theory
Chapter 3: Theory of Operation
Battery
Compressor 9 coil
sensor
3
Driver
DC
Power power
Logic CPR
supply Control User
P.C. board control
P.C. board pendant
4
Driver
(optional)
Hour Axial
Vacuum meter fan
switch
Heel manifold
5
solenoid valves
Blower
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 19
Subsystem Theory
Chapter 3: Theory of Operation
A rechargeable battery remembers the last status of the system and provides power
for CPR mode in case of power disconnection.
3 The heart of the system is the unique control logic circuitry which provides
optimum patient support therapy. The Deteq™ module sensor uses induction
between aluminium and an energised coil. As the aluminium foil is moved closer
or further away from the sensor coil, the impedance of the coil varies. This is used
as the feedback circuit to control the blower/pump output.
The control pendant operates as the main control of Duo 2™ Multi Mode Mattress
System while showing feedback to the operator of system status and problems.
5 The control pendant allows the operator to select the following modes:
• Continuous therapy mode
• Alternating therapy mode
• Maximum inflation mode
The CPR/STOP button is fitted and, when activated, electrically opens all solenoid
valves for approximately 45 s to assist in deflating the mattress.
6 The control pendant is designed to allow the user to operate it in two different
ways when power is disconnected: to differentiate between transport mode and
power off.
Transport/Stand By Mode
Transport mode is engaged by unplugging the system from its power source. The
7 mattress remains inflated for approximately two hours, and the 12VDC battery
powers the logic circuit and the control pendant until the system is plugged into an
appropriate power source.
1
Power Off
When the alarm on the control pendant activates, there are a number of different
causes which are shown below.
If the mattress is not able to achieve correct inflation (some causes are: cushion
leaks, hose disconnections, manifold test port caps not secured, compressor not
operating correctly), the servo control continuously drives the compressor to try to
compensate for these problems. After 15 minutes of continuous compressor
operation, the audible/visual alarm will be activated. In all cases, the alarm may be
silenced for 15 minutes.
6
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 3 - 21
Subsystem Theory
Chapter 3: Theory of Operation
Foil
aluminium
2
Captor P+D=constant
coils
Distance
measurement Integrator
3 P1 Comparator
adjustment
AIR
Pressure
measurement Window
comparator
4 P5 Driver
adjustment
Compressor
Voltage TP7
A Zone Driver
5 B Zone
Window
Solenoid
valve
comparator coil
Air leak
Patient Patient
150kg (330lbs) 40kg (88lbs)
6
Small distance (d) + high pressure (P) = constant Large distance (D) + low pressure (p) = constant
W12003_045
7
Page 3 - 22 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Subsystem Theory
Chapter 3: Theory of Operation
Pneumatic 1
Air enters the air supply enclosure (ASE) through the air filter. The air filter
removes all airborne particles greater than 30 microns. The air travels directly into
the blower and, to the compressor inlet (see figure 3-9 on page 3-10).
On exiting the blower, the air travels directly into the main manifold and from the
compressor, through a non return valve, into the main manifold.
At the air output on each manifold, there are pressure test ports for calibration
2
purposes.
A small axial fan pushes air into the therapeutic mattress encasement to minimize
any buildup in humidity and ensure that air is continuously changed in the
therapeutic mattress.
Therapeutic
mattress
Deteq™
mattress
Heel spacing mattress
Air supply
4
Head foam Sensor coils foam Foot foam enclosure
Deteq™
mattress
Foil Foam 5
Air supply enclosure
1 The therapeutic cushions are composed of a series of 13 air cushions, none with air
vents. The top cushion at the head end of the therapeutic mattress is not used in
alternating therapy mode. They are at the same air pressure as the Deteq™
module. Twelve cushions are used for the alternating therapy mode. The heel zone
comprises eight cushions none with air vents. The heel zone air is taken through
the main manifold to the heel manifold, the air in the heel zone is regulated by the
heel manifold solenoid valves.
2 The Deteq™ aluminium foil is located between the therapeutic mattress and the
Deteq™ module, and is held in place by two lateral fabric strips fitted with snap
fasteners.
7
Page 3 - 24 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Chapter 4 1
Removal, Replacement, and
Adjustment Procedures
2
Chapter Contents
1 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 17
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 18
Control Pendant P.C. Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 20
2 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 21
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 21
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 21
Vacuum Valve Microswitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 22
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 22
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
3 Cooling Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 24
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 24
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 25
Deteq™ Module Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 26
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 26
4 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 27
Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 28
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 28
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 28
Main Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 29
5 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 29
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 30
Heel Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 31
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 31
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 32
In-Line Check Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
6 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
Maximum Inflation Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 34
Continuous Mode Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 35
Heel Zone Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 37
1. Working on the patient left of the mattress, unclip the therapeutic mattress flap
from the Deteq™ module.
2. Expose the two quick connect couplings at the foot left side of the heel
module.
2
3. Press the releases on the quick connect couplings and release the two hose
connections (A) (see figure 4-1 on page 4-3).
5
W12003_053
4. Unzip the heel module coverlet on the patient left to expose the main six
mattress connections and the two air vent connections.
5. Press the releases on the quick connect couplings, and release the six hose
connections.
6
6. Fold back the therapeutic mattress and the heel module cover to fully expose
the ASE.
1 1
7. Separate the two Velcro® straps that secure the ASE to the base of the heel
module.
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2
8. Using a 3 mm Allen™ hex key, loosen the eight screws (B) securing the lid of
the ASE (see figure 4-2 on page 4-4).
5 W12003_048
9. Lift away the lid of the ASE, and store safely with the eight screws.
2. Loosen the two thumbscrews (A) and gently pull out the control pendant
connector (see figure 4-3 on page 4-5). 2
Figure 4-3. Control Pendant Connector Removal
Replacement
W12003_049
5
1. Perform the removal procedure in reverse order.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 4 - 5
4.3 Air Filter Door
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
1. At the patient foot right, unclip the air filter door cover.
2. Open the air filter door by unfastening the two clasps (A) (see figure 4-4 on
page 4-6).
3 B
4
A
B
5 3. Using the needle nose pliers at each end of the outside hinge point of the air
filter door, pull off the “push-on” fixings (B).
Replacement
6 1. Perform the removal procedure in reverse order.
7
Page 4 - 6 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.4 Vacuum Valve
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
2
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2. Locate connector J8 (F) on the power supply P.C. board (G) and disconnect
(see figure 4-9 on page 4-13). 3
3. Open the air filter door.
4. Using a No. 2 Pozidriv® screwdriver, loosen and remove the valve lever screw
(A) and pull away the lever (B) (see figure 4-5 on page 4-7).
B C
6
A
C W12003_051
1 5. Using a No. 2 Pozidriv® screwdriver, loosen and remove the three valve
assembly screws (C) and lift away the valve assembly (D) (see figure 4-6 on
page 4-8).
4 W12003_052
Replacement
1. Perform the removal procedure in reverse order.
7
Page 4 - 8 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.5 Blower
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.5 Blower
Tools required: Small screwdriver
1
1
No. 2 Pozidriv® screwdriver
2
3 mm Allen™ hex key
Removal
SHOCK HAZARD:
2
Voltage is present on exposed electrical connectors and P.C. board when
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2. Locate connector J4 (A) on the power supply P.C. board (B) and disconnect
(see figure 4-7 on page 4-9).
3
Figure 4-7. Blower Assembly Removal
W12003_053
6
3. Using a No. 2 Pozidriv® screwdriver, loosen and remove the 3 blower inlet
screws (C).
4. Gently pull away the the rubber hose connection between the manifold (D) and
the blower outlet (E), and lift away the blower assembly (F).
1 Replacement
1. Perform the removal procedure in reverse order.
7
Page 4 - 10 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.6 Compressor
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.6 Compressor
Tools required: 1
3 mm Allen™ hex key
1
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
power is on. Use care when working close to connections. Failure to do so
2
could result in personal injury or equipment damage.
2. Locate connector J5 (A) on the power supply P.C. board (B) and disconnect
(see figure 4-8 on page 4-11).
D A
B
C
4
W12003_054
2
3. Tear apart the Velcro® strap (C) that secures the compressor to the base of the
ASE.
6
4. Locate the two hoses and pull them off the inlet and outlet (D) of the
compressor (E).
1 Replacement
1. Perform the removal procedure in reverse order.
7
Page 4 - 12 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.7 Power Supply P.C. Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
2
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2. Using a No. 1 Pozidriv® screwdriver, loosen and remove the two cover
mounting screws (H) and the cover (K) (see figure 4-10 on page 4-14). 3
3. Disconnect connectors J3 (A), J4 (B), J5 (C), J6 (D), J7 (E), and J8 (F) on the
power supply P.C. board (G) (see figure 4-9 on page 4-13).
A
B C 4
G
5
E
F D
6
W12003_055
4. Using a No. 1 Pozidriv® screwdriver, loosen and remove the two P.C. board
mounting screws (I).
I
I
J
3
W12003_056
7
Page 4 - 14 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.8 Logic P.C. Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
2
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2. Disconnect connectors J9 (A), J10 (B), J11 (C), J12 (D), and J13 (E) on the
logic P.C. board (see figure 4-11 on page 4-15). 3
Figure 4-11. Logic P.C. Board Connector Removal
F
F A 4
C
D
5
E
W12003_057
3. Locate the hoses at the two transducers (F) on the logic P.C. board, and very
6
gently pull the hoses off while supporting the transducer.
4. Using a No. 1 Pozidriv® screwdriver, loosen and remove the P.C. board
mounting screw (G) (see figure 4-12 on page 4-16).
3
W12003_058
5. Lift away the logic P.C. board (H) by pulling towards the compressor and then
4 upwards.
Replacement
1. Perform the removal procedure in reverse order.
7
Page 4 - 16 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.9 Solenoid Coil
Chapter 4: Removal, Replacement, and Adjustment Procedures
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
power is on. Use care when working close to connections. Failure to do so
2
could result in personal injury or equipment damage.
2. Locate the solenoid coil (A) to be removed, and disconnect the push-on
connector board (B) (see figure 4-13 on page 4-17).
C
5
A
W12003_059
NOTE:
This process is the same for all solenoid coils.
6
3. Pull clip (C) from the top of the solenoid coil (A).
1 Replacement
1. Perform the removal procedure in reverse order.
7
Page 4 - 18 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.10 Control Pendant P.C. Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
1. Place the control pendant face down on a surface.
NOTE:
Protect the control pendant face using bubble-wrap or a similar soft material.
2
2. Using a No. 2 Pozidriv® screwdriver, loosen and remove the four rear housing
screws (A) (see figure 4-14 on page 4-19).
5
W12003_060
3. Locate connector J2 (A) on the control pendant P.C. board (B) and disconnect
(see figure 4-15 on page 4-20).
4. Locate connector J1 (C) on the control pendant P.C. board (B) and disconnect. 6
3 C
A
W12003_061
1
5. Using a No. 1 Pozidriv® screwdriver, loosen and remove the two P.C. board
4 mounting screws (D).
Replacement
1. Perform the removal procedure in reverse order.
4.11 Battery
Tools required: 1
3 mm Allen™ hex key
1
Removal
1. Open the ASE (refer to procedure 4.1).
2. Locate the battery (A) close to the vacuum mode lever (B) (see figure 4-16 on
page 4-21). 2
Figure 4-16. Battery removal
A C
C
4. Push the two clips (C) aside and lift away the battery (see figure 4-16 on page
4-21).
Replacement
1. Perform the removal procedure in reverse order.
6
2. To ensure proper operation of the Duo 2™ Multi Mode Mattress System,
perform “Function Checks” on page 2-5.
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2. Locate connector J8 (F) on the power supply P.C. board (G) and disconnect
(see figure 4-9 on page 4-13).
4 4. Using a No. 1 Pozidriv® screwdriver, loosen and remove the valve lever screw
(A) and pull away lever (B) (see figure 4-17 on page 4-22).
6
B C
C
7
W12003_051
6. Use a small screwdriver as a lever to release the rear bracket (D), and lift away
the valve assembly (E) and rear bracket (see figure 4-18 on page 4-23).
G
3
G
4
H
F
W12003_063
7. Using a small screwdriver, locate the countersunk screws (F), and using a
4 mm open end spanner, loosen and remove the two microswitch nuts, (G) and
lift away the microswitch (H) and cable (I). 5
Replacement
1. Perform the removal procedure in reverse order.
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
3 2. Locate connector J9 (A) on the logic P.C. board, and disconnect (see figure 4-
19 on page 4-24).
NOTE:
The Deteq™ sensor is also wired to connector J9, use a small screwdriver to
release the 2 pins of the cooling fan cable from the connector.
3. Pull hoses (B) from the upper part of cooling fan assembly (C).
4 4. Pull the cooling fan assembly (C) upwards from the four plastic studs and lift it
away.
5 B
6
B
7 W12003_064
Replacement 1
1. Perform the removal procedure in reverse order.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 4 - 25
4.14 Deteq™ Module Sensor
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
1. Working on the patient left of the mattress, unclip the mattress flap from the
heel module.
2. Expose the two quick connect couplings (A) at the foot left side of the heel
module; press the releases on the quick connect couplings, and release the two
hose connections (see figure 4-20 on page 4-26).
5 W12003_047
3. Unzip the heel module cover on the patient left to expose the main six mattress
connections.
4. Press the releases on the quick connect couplings, and release the six hose
connections.
6 5. Fold back the therapeutic mattress and the cover to fully expose the ASE.
6. Locate the Deteq™ module coil connector (B) at the patient left panel of the
ASE (see figure 4-21 on page 4-27).
7. Turn the collar (C) of the connector counterclockwise to unlock the connector
and pull it out.
7
Page 4 - 26 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.14 Deteq™ Module Sensor
Chapter 4: Removal, Replacement, and Adjustment Procedures
2
E
D
B
3
C
W12003_065
8. Gently lift the cable (D) out of the submattress sections until you reach the
Deteq™ module coil subassembly (E).
1
4
9. Separate the two Velcro® straps that secure the Deteq™ module coil
subassembly to the base of the mattress cover.
Replacement
1. Perform the removal procedure in reverse order.
5
2. To ensure proper operation of the Duo 2™ Multi Mode Mattress System,
perform “Function Checks” on page 2-5.
Removal
1. At the patient foot right, unclip the air filter door cover and open the air filter
door by unfastening the two clasps (A) (see figure 4-22 on page 4-28).
3 B
4 A
7
Page 4 - 28 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.16 Main Manifold
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
2
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2. Locate connector J10 (B) on the logic P.C. board, and disconnect (see figure 4-
11 on page 4-15). 3
3. Using a No. 1 Pozidriv® screwdriver, loosen and remove the three manifold
assembly screws (A) (see figure 4-23 on page 4-29).
C
4
B
B
A
6
W12003_067
4. Gently pull away the five moulded rubber hoses (B) from the rear of the
manifold (C).
1 5. Gently pull away the moulded manifold hose from the base of the manifold.
Replacement
1. Perform the removal procedure in reverse order.
7
Page 4 - 30 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.17 Heel Manifold
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
2
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2. Locate connector J11 (C) on the logic P.C. board, and disconnect
(see figure 4-11 on page 4-15). 3
3. Locate the three moulded rubber hoses (A) at the rear of the heel manifold, and
carefully pull off the hoses (see figure 4-24 on page 4-31).
A C 4
B 6
W12003_068
4. Using a No. 1 Pozidriv® screwdriver, loosen and remove the two manifold
assembly screws (B).
Replacement
1. Perform the removal procedure in reverse order.
7
Page 4 - 32 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.18 In-Line Check Valve
Chapter 4: Removal, Replacement, and Adjustment Procedures
Removal
SHOCK HAZARD:
Voltage is present on exposed electrical connectors and P.C. board when
2
power is on. Use care when working close to connections. Failure to do so
could result in personal injury or equipment damage.
2. Locate the in-line check valve under the blower outlet moulded hose.
Replacement
1. Perform the removal procedure in reverse order.
NOTE:
This procedure is for use prior to using procedure 4.21.
NOTE:
The manometer must have a minimum range of 0" H2O to 25" H2O (0 mm Hg to
47 mm Hg) and an accuracy of ± 0.1" H2O (± 0.2 mm Hg).
1. Connect the manometer to the main zone test port (A) (see figure 4-25 on page
4-34).
W12003_069
5 2. Select maximum inflation mode and allow the mattress to become firm prior to
taking measurements.
3. Verify that the pressure is 25.0" H2O ± 2.0" H2O (46.7 mm Hg ± 3.7 mm Hg).
7
Page 4 - 34 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.20 Continuous Mode Adjustment
Chapter 4: Removal, Replacement, and Adjustment Procedures
1. Select continuous therapy mode and allow the mattress to stabilize prior to
taking measurements. 2
2. Connect the manometer to the main zone test port (A) (see figure 4-25 on page
4-34).
3. Loosen the two lock screws on the test connector cover (A) and connect the
diagnostics tester (B) (see figure 2-2 on page 2-10).
4. Place: 3
a. The 25 kg (55 lb) test load (A) centred, counting from the head end, above
the ninth bladder and over four cushions at the minimum (see figure 4-27
on page 4-37).
5
Head end
6
W12003_021
b. The 5 kg (11 lb) weight (B) centred on top of the heel zone, i.e. over the
“O” in the “Duo 2™” logo.
5. Push a small screwdriver into the seat cushion connector located under the heel
zone bus to lower the pressure inside the main zone of the mattress to below
7" H2O.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 4 - 35
4.20 Continuous Mode Adjustment
Chapter 4: Removal, Replacement, and Adjustment Procedures
1 NOTE:
The pressure must drop below 7" H2O to allow the mattress to inflate up to the
preset level.
6. When the compressor stops, take a measurement. The value should read
7.0" H2O ± 0.4" H2O (13.1 mm Hg ± 0.8 mm Hg).
7. If outside acceptable range, adjust P1 (D) located to the right of the diagnostics
tester connection using the trimmer tool to read 7.0" H2O ± 0.2" H2O
(4.7 mm Hg ± 0.4 mm Hg) (see figure 4-25 on page 4-34).
NOTE:
To increase the pressure, turn P1 counterclockwise, and to decrease the pressure,
turn P2 clockwise.
7
Page 4 - 36 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.21 Heel Zone Adjustment
Chapter 4: Removal, Replacement, and Adjustment Procedures
1. Select continuous therapy mode and allow the mattress to stabilize prior to
taking measurements. 2
2. Connect the manometer to the heel zone test port (B) (see figure 4-25 on page
4-34).
3. Loosen the two lock screws on the test connector cover (A) and connect the
diagnostics tester (B) (see figure 2-2 on page 2-10).
4. Place: 3
a. The 25 kg (55 lb) test load (A) centred, counting from the head end, above
the ninth bladder and over four cushions at the minimum (see figure 4-27
on page 4-37).
5
Head end
6
W12003_021
b. The 5 kg (11 lb) weight (B) centred on top of the heel zone, i.e. over the
“O” in the “Duo 2™” logo.
5. Take a measurement. The value should read 2.5" H2O ± 0.5" H2O
(4.7 mm Hg ± 0.9 mm Hg). 7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 4 - 37
4.21 Heel Zone Adjustment
Chapter 4: Removal, Replacement, and Adjustment Procedures
1 NOTE:
Adjustments are set only when the “Heel Manifold - fill” LED and “Heel Manifold
- leak” LED on the diagnostics tester are both off (see figure 2-1 on page 2-8).
6. If outside acceptable range, adjust P2 (C) located to the left of the diagnostics
tester connection using the trimmer tool to read 2.5" H2O ± 0.2" H2O
(4.7 mm Hg ± 0.4 mm Hg) (see figure 4-25 on page 4-34).
NOTE:
To increase the pressure turn P2 clockwise, and to decrease the pressure turn P2
counterclockwise.
7
Page 4 - 38 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
4.22 Alternating Mode Verification
Chapter 4: Removal, Replacement, and Adjustment Procedures
2. Connect the manometer to the heel zone test port (B) (see figure 4-25 on page
4-34).
2
3. Loosen the two lock screws on the test connector cover (A) and connect the
diagnostics tester (B) (see figure 2-2 on page 2-10).
4. Place:
a. The 25 kg (55 lb) test load (A) centred, counting from the head end, above
the ninth bladder and over four cushions at the minimum (see figure 4-27
3
on page 4-37).
b. The 5 kg (11 lb) weight (B) centred on top of the heel zone, i.e. over the
“O” in the “Duo 2™” logo.
5. Ensure that one of the zone LEDs on the diagnostics tester is illuminated (see
figure 2-1 on page 2-8).
6. Take a measurement. The value should read 2.5" H2O ± 0.5" H2O
4
(4.7 mm Hg ± 0.9 mm Hg).
NOTE:
Values are set only when the “Heel Manifold - fill” LED and “Heel Manifold -
leak” LED on the diagnostics tester are both off (see figure 2-1 on page 2-8).
7. If outside acceptable range, perform the heel zone adjustment procedure (See
“Heel Zone Adjustment” on page 4-37).
5
8. Ensure that none of the zone LEDs on the diagnostics tester is illuminated (see
figure 2-1 on page 2-8).
9. Take a measurement. The value should read 7.0" H2O ± 0.4" H2O
(13.1 mm Hg ± 0.8 mm Hg).
11. Repeat steps 5 to 10 for the remaining 3 parts of the alternating cycle.
12. When complete, remove the manometer and seal the test port and remove the
diagnostics connection and fit the test connector cover.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 4 - 39
4.22 Alternating Mode Verification
Chapter 4: Removal, Replacement, and Adjustment Procedures
1 NOTES:
7
Page 4 - 40 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Chapter 5 1
Parts List
Chapter Contents
2
Service Parts Ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
Recommended Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
Overall Mattress Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Therapeutic Mattress Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8
Control Pendant Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 10
3
Air Supply Enclosure (Sheet 1 of 5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 12
Air Supply Enclosure (Sheet 2 of 5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 14
Air Supply Enclosure (Sheet 3 of 5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 16
Air Supply Enclosure (Sheet 4 of 5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
Air Supply Enclosure (Sheet 5 of 5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 20 4
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 21
Main Manifold Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 22
Heel Manifold Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 23
Vacuum Valve Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 24
Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 26
Diagnostics Tester . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 28
5
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 1
Chapter 5: Parts List
1 NOTES:
7
Page 5 - 2 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Service Parts Ordering
Chapter 5: Parts List
4
W12003_070
Call the After Sales Service of your national Hill-Rom representative (refer to the
last page of this manual).
To help expedite the processing of your parts order, please have the following
5
information available:
• Customer account number
• Purchase order number
• Serial number
• Part number(s)
6
Delivery:
Please address all enquiries to your National Hill-Rom representative (refer to the
back cover of this manual).
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 3
Recommended Spare Parts
Chapter 5: Parts List
2 M05088
Part Number
1
Quantity Description
Compressor 220-240VAC
M05091 1 Blower (230VAC)
E17160 1 Cushion, therapeutic mattress
E06184 1 Bus connector
E17161 1 Cushion, Heel zone
3 M01233
F20007
1
1
Pendant assembly, complete
Heel zone manifold
F20006 1 Main manifold
M04169 1 Diagnostics cover plate assembly
A36061 1 Logic P.C. board
A52002 1 Deteq™ sensor assembly
4 F10015
F10016
1
1
Air filter
Dust filter
a
E17101 1 Velcro® strap for compressor
A39005 1 Battery, 12V/0.8Ah
A47117 1 P.C. boards link cable
A36060 1 Power supply P.C. board, 220/240VAC
5 A36059
A47116
1
1
Control pendant P.C. board
Power inlet harness
A16037 1 400mAF fuse
A16046 1 125mAF fuse
A16052 1 2AT fuse
A16036 1 630mAT fuse
6 A16039
M04192
2
1
1AF Fuse
Vacuum valve assembly
A32006 6 Solenoid coil, 12VDC
C02067 2 Push-on cap for hinge pin of air filter
door
a. Velcro® is a registered trademark of Velcro Industries, BV (a Dutch corporation)
7
Page 5 - 4 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Recommended Spare Parts
Chapter 5: Parts List
NOTES: 1
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 5
Overall Mattress Assembly
Chapter 5: Parts List
6
W12003_071
7
Page 5 - 6 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Overall Mattress Assembly
Chapter 5: Parts List
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 7
Therapeutic Mattress Assembly
Chapter 5: Parts List
4
Note: View flipped over for clarity
5
Detail A
6
Detail B Detail C W12003_072
7
Page 5 - 8 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Therapeutic Mattress Assembly
Chapter 5: Parts List
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 9
Control Pendant Assembly
Chapter 5: Parts List
4 W12003_073
5 3
4
E06165
A36059
1
1
Front housing - control pendant
Control pendant P.C. board
5 E06168 1 Control pendant housing gasket
6 E06166 1 Rear housing - control pendant
7 Z01338 1 Label, IPX0
8 E06167 1 Hanging bracket
6 9
10
A47114
Z01337
1
1
Cable assembly, control pendant
Label, CE mark
11 C01270 6 Screw, housing
12 C04001 2 Plain washer, D4
13 C06023 1 Nylon hose clip, SNP 2
b
14 C01248 2 Screw, No.4 x 8, self-tapping, Pozidriv®
NOTES: 1
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 11
Air Supply Enclosure (Sheet 1 of 5)
Chapter 5: Parts List
6
W12003_074
7
Page 5 - 12 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Air Supply Enclosure (Sheet 1 of 5)
Chapter 5: Parts List
4 E06180 1
pressure sensor
Moulded flexible hose - main manifold to
2
heel manifold
5 M05098 1 Fan assembly
6 E06173 1 Moulded flexible hose - compressor to
non-return valve
7
8
E04224
E06176
1
1
Calibrated outlet for compressor
Moulded flexible hose - non-return valve
to main manifold
3
9 E06177 1 Moulded flexible hose - main manifold to
check valve
10 E06171 1 Moulded flexible hose - main manifold to
vacuum valve
11
12
F02085
C04001
2
3
Female quick connect coupling
Plain washer, D4
4
13 C01272 3 Manifold screw, 35 x 20
14 F23005 1 Non-return valve, 3" H2O
15 F06011 1 Check valve, 1½ p.s.i.
16 Z01342 1 Label, 2 x 1A fuses
17
18
A99025
F11002
1
1
IEC cable connector retention kit
Air Filter inlet kit
5
19 C07038 4 Rivet
20 A47116 1 IEC inlet assembly with filter and fuses
Not shown A16039 2 1AF Fuse
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 13
Air Supply Enclosure (Sheet 2 of 5)
Chapter 5: Parts List
6
W12003_075
7
Page 5 - 14 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Air Supply Enclosure (Sheet 2 of 5)
Chapter 5: Parts List
11 C01262 4
board
Screw, M3 x 6, Pozidriv®
3
12 A36060 1 Power supply P.C. board, 220/240 VAC
13 C01251 4 Screw, 4.2 x 9.5, Pozidriv®
14 E05331 2 Battery Clip
15 E06172 1 Moulded flexible hose - vacuum valve to
16 Z01343 1
blower
Battery label
4
17 E08131 1 Rapid deflate exit flange
a. Pozidriv® is a registered trademark of Phillips Screw Company.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 15
Air Supply Enclosure (Sheet 3 of 5)
Chapter 5: Parts List
6
W12003_076
7
Page 5 - 16 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Air Supply Enclosure (Sheet 3 of 5)
Chapter 5: Parts List
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 17
Air Supply Enclosure (Sheet 4 of 5)
Chapter 5: Parts List
6
W12003_077
7
Page 5 - 18 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Air Supply Enclosure (Sheet 4 of 5)
Chapter 5: Parts List
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 19
Air Supply Enclosure (Sheet 5 of 5)
Chapter 5: Parts List
4 W12003_078
5 3
4
A47117
E08157
1
1
P.C. board link cable
Compressor, anti-vibration mount
b
5 E17101 1 Velcro® strap for compressor
6 M05088 1 Compressor 220-240VAC
a. Pozidriv® is a registered trademark of Phillips Screw Company.
b. Velcro® is a registered trademark of Velcro Industries, BV (a Dutch corporation)
7
Page 5 - 20 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Fuses
Chapter 5: Parts List
Fuses
1
Figure 5-10. Fuses
W12003_079
4
Table 5-10. Fuses
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 21
Main Manifold Assembly
Chapter 5: Parts List
4 W12003_080
5 3
4
F20006
F02087
1
1
Main manifold
Hose connector, 90° elbow, 3/8 NPT
5 A32006 6 Solenoid coil, 12VDC
6 A99030 6 Fixing clip, manifold coil
7
Page 5 - 22 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Heel Manifold Assembly
Chapter 5: Parts List
W12003_081
4
Table 5-12. Heel Manifold Assembly
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 23
Vacuum Valve Assembly
Chapter 5: Parts List
4 W12003_082
6 8 C03033 2
harness
Nut, M2
9 C04024 2 Lock Washer, D2
10 C01252 2 Screw, M2 x 12, countersunk head
11 M04192 1 Vacuum valve assembly
a. Pozidriv® is a registered trademark of Phillips Screw Company.
7
Page 5 - 24 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Vacuum Valve Assembly
Chapter 5: Parts List
NOTES: 1
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 25
Labels
Chapter 5: Parts List
1 Labels
6
Note: Label is inside ASE
W12003_083
7
Page 5 - 26 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Labels
Chapter 5: Parts List
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 5 - 27
Diagnostics Tester
Chapter 5: Parts List
1 Diagnostics Tester
4 W12003_084
5 3
4
A36049
C03004
1
8
Diagnostics tester P.C. board
Nut, M4
5 C04059 4 Lock washer, D4
6 C01207 4 Screw, M4 x 12 countersunk head
7 Z01306 1 Display label, diagnostics tester
8 G05001 1 Diagnostics tester
7
Page 5 - 28 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Chapter 6 1
General Procedures
Chapter Contents
2
Pre-Installation Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 4
Infection Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 5
Transportation, Setup, and Pickup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 6
Transportation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 6
3
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 6
Pickup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 9
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 11
General Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 12
Steam Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 12 4
Cleaning Hard to Clean Spots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 12
Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 12
Laundering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 13
Therapeutic Mattress and Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 13
Tool and Supply Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 15 5
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 6 - 1
Chapter 6: General Procedures
1 NOTES:
7
Page 6 - 2 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
6.1 Pre-Installation Maintenance
Chapter 6: General Procedures
NOTE:
Dispose of the battery in accordance with local regulations. 2
Perform the Pre-Installation Maintenance (PIM) procedure each time the Duo 2™
Multi Mode Mattress System is removed from a facility. The purpose of the PIM is
to ensure that the system is in optimum condition for the next patient.
1. Follow the steps on the PIM form, filling it out as the PIM is performed.
4. Once the form is completed, signed, and recorded, file it according to current
service location policy.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 6 - 3
6.2 Routine Maintenance
Chapter 6: General Procedures
WARNING:
Adhere to the Infection Control Policies and Procedures from Hill-Rom.
Failure to do so could result in the spread of infection.
1. Contact the customer to advise of your approximate arrival time. Ask that the
customer arrange to remove the patient from the unit during the service time
(approximately 15 minutes).
3. Wipe down the air cushion, control pendant, and cable with an approved or
hospital-approved disinfectant.
5. After you complete all the above steps, follow the service location policy to
7
Page 6 - 4 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
6.3 Infection Control
Chapter 6: General Procedures
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 6 - 5
6.4 Transportation, Setup, and Pickup
Chapter 6: General Procedures
Transportation
Place the folded Duo 2™ Multi Mode Mattress System in the transport case.
Setup
WARNING:
Adhere to the Infection Control Policies and Procedures from Hill-Rom.
Failure to do so could result in the spread of infection.
3. Place the Duo 2™ Multi Mode Mattress System at the head end of the frame
(A) (see figure 6-1 on page 6-6).
Foot end
4 A
5 B
6 W12003_086
4. Unlock the buckle (B), and with one person on each side holding the handle
(C) for foot end, lift, move, and place at the foot end of the frame
(see figure 6-2 on page 6-7).
7 5. Secure with the two nylon straps (D), ensuring that the strap placement does
not interfere with the operation of the bed functions (head-up or hilow).
2
D
D
3
W12003_087
6. Secure the three side snaps (E) of the therapeutic mattress (F) to their matching
Deteq™ module (G) snaps and close the zippers (H) on each side of the
Deteq™ module (see figure 6-3 on page 6-7).
4
Figure 6-3. Securing the Mattress
F
5
H
G
E
W12003_088
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 6 - 7
6.4 Transportation, Setup, and Pickup
Chapter 6: General Procedures
1 7. Remove the control pendant from its protective location and attach it to the
foot board of the frame using the universal clamp (I) (see figure 6-4 on page
6-8).
4 W12003_089
CAUTION:
6 not become trapped or trail on the floor. The mattress will start to inflate
immediately once power is applied.
10. Initially a small blower will start and, even with the patient lying on the
mattress, rapidly inflates the surface. The patient will be well supported after
only one minute, then the blower will shut down and a small silent air pump
will take over.
11. In all cases, the patient's sacrum should be centred on the mattress adjacent to
the Deteq™ logo (it is printed on each side of the mattress cover).
1
12. The control pendant will display the status of the system continuously.
Pickup
WARNING:
Adhere to the Infection Control Policies and Procedures from Hill-Rom.
2
Failure to do so could result in the spread of infection.
1. Place the system into continuous therapy mode on the control pendant.
WARNING:
Ensure the frame brake is on to avoid unplanned frame movements.
Failure to do so could result in personal injury.
3
2. Follow the infection control protocol for product removal.
4
B
A 6
W12003_090
3. Open the filter door (A) at the right side of the ASE. Turn the vacuum valve
lever (B) one quarter turn counterclockwise (see figure 6-5 on page 6-9).
4. Press firmly on the heel zone to open the heel manifold solenoid valves.
5. The air escapes from the mattress which will be empty in approximately seven
to ten minutes.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 6 - 9
6.4 Transportation, Setup, and Pickup
Chapter 6: General Procedures
1 6. When the mattress is empty, turn the vacuum valve knob to the horizontal
position.
CAUTION:
The air filter door will not close if the lever is not returned to the horizontal
position. Damage to the door or its clasps may occur if it is forced.
7. Close the air filter door and fasten it into place with the two clasps.
9. Place the power cord on top of the mattress at the head end.
10. Take the control pendant, and place it between the two layers of the mattress.
3 NOTE:
Whilst the packing procedure can be done by one person, we recommend using
two people.
4 b. Lift the foot section up and fold the mattress towards the head in a “Z”
shape.
c. Wrap the packing strap around the mattress, lock the buckles, and tighten
securely together.
7
Page 6 - 10 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
6.5 Cleaning
Chapter 6: General Procedures
6.5 Cleaning 1
WARNING:
Follow the product manufacturer’s instructions. Failure to do so could result
in personal injury or equipment damage.
SHOCK HAZARD:
2
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture that would allow for liquid
pooling. Personal injury or equipment damage could occur. 3
CAUTION:
Do not use harsh cleansers/detergents such as scouring pads and heavy
duty grease removers or solvents such as acetone. Equipment damage
could occur.
CAUTION:
4
The control pendant is not protected against ingress of water. Disconnect it
before cleaning. Failure to do so could result in equipment damage.
CAUTION:
Avoid the use of cleaning solutions containing Dimethyl formamide,
Tetrahydrofuran, Phenol, Cresol or concentrated inorganic acids, as these 5
chemicals will damage the cushion fabric.
NOTE:
These chemicals are not normally found in standard hospital cleaning solutions.
CAUTION:
Avoid highly alkaline and strong organic based detergents and 6
disinfectants, as these chemicals may attack the cushion fabric dye,
causing transfer of colour.
Clean the unit with a lightly dampened cloth and ordinary disinfectants.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 6 - 11
6.5 Cleaning
Chapter 6: General Procedures
1 General Cleaning
We recommend that you clean the unit with detergent and warm water. Do not use
excessive liquid or harsh cleansers.
Steam Cleaning
Do not use any steam cleaning device on the unit. Excessive moisture can damage
mechanisms in this unit.
3 Disinfecting
CAUTION:
Do not use hypochlorite (bleach) above a concentration of 500 ppm as this
will discolour and weaken the fabric. Failure to do so could result in
equipment damage.
4 1. When there is visible soilage and also between patient use, we recommend that
you disinfect the unit using a tuberculocidal disinfectant.
CAUTION:
The control pendant is not protected against ingress of water. Disconnect it
before cleaning. Failure to do so could result in equipment damage.
5 CAUTION:
The mattress is protected against a jet of water from a standard nozzle of
12.5 mm diameter at a distance 2.5 m to 3.0 m, with a delivery rate 100
l/min, spraying the mattress from all practical directions. Check the
capability of the cleaning equipment prior to using it, otherwise damage to
the mattress may occur.
6 2. Dilute the disinfectant as specified on the manufacturer's label and follow the
manufacturer’s instructions for “wetting” times for hard surface disinfection.
CAUTION:
Do not use phenolics on the Protex™ cover. This could result in damage to
equipment and loss of waterproof qualities.
Laundering
CAUTION:
Mattress connections under the therapeutic mattress must be sealed (see
4
“Tool and Supply Requirements” on page 6-15) prior to laundry. Failure to
do so could result in equipment damage.
NOTE:
8 air outlet plugs (F02071) must be used to seal the mattress.
CAUTION:
Ensure the mattress is completely dry before storage. Failure to do so
could result in equipment damage.
6
3. Ensure adequate time for the Protex™ cover to completely dry.
CAUTION:
Do not mangle the Protex™ cover. Equipment damage could result.
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 6 - 13
6.5 Cleaning
Chapter 6: General Procedures
1 CAUTION:
Do not use heat above 67°C (153°F) as this may damage the urethane
lining and heat seals of the cushion.
7
Page 6 - 14 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
6.6 Tool and Supply Requirements
Chapter 6: General Procedures
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 6 - 15
6.6 Tool and Supply Requirements
Chapter 6: General Procedures
1 NOTES:
7
Page 6 - 16 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Chapter 7 1
Accessories
Chapter Contents
2
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
Seat Cushion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 6
3
7
Duo 2™ Multi Mode Mattress System Service Manual (W12003) Page 7 - 1
Chapter 7: Accessories
1 NOTES:
7
Page 7 - 2 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Accessories
Chapter 7: Accessories
Accessories
1
7.1 Seat Cushion
The seat cushion is an option available with the Duo 2™ Multi Mode Mattress
System.
Installation 2
1. Perform the following:
a. Place the seat cushion on the patient's chair, and secure the seat cushion by
using the strap provided (A) (see figure 7-1 on page 7-3).
1
b. Open the Velcro® flaps (B) on the seat cushion, disconnect and extend the
hose (C).
W12003_091
c. Lift the cover (D) of the unit at the foot left end to reveal the heel zone
access zipper (E).
6
d. Open the zipper (E).
e. Locate the seat cushion outlet (F) (see figure 7-2 on page 7-4).
3
W12003_092
f. Insert the seat cushion connector (G); an audible click will confirm
4 connection.
7
Page 7 - 4 Duo 2™ Multi Mode Mattress System Service Manual (W12003)
Accessories
Chapter 7: Accessories
1
e. Fold the Velcro® flaps (K) over the hose and fasten.
W12003_093
3
4. Select continuous low pressure mode on the control pendant.
5. When the patient is returned to the mattress, the Deteq™ system will
automatically take control and manage the mattress pressure to the optimum
value.
4
1 Removal
1
1. To deflate the seat cushion, pull apart the covers at the Velcro® joint on the
side adjacent to the hose cover.
2. Locate the blue deflate valve (L), and twist the cap (K) a ½ turn counter-
clockwise to open (see figure 7-4 on page 7-6).
4 W12003_094
3. Fold the cushion from the opposite side to expel the air; then return the cap to
the sealed position, fasten the Velcro® joint, and store.
J3 - AC supply
1 + Coil
2 - Coil 1 NC - CPR switch
3 Not used 2 Common - CPR switch
4 Not used
5
6
Fan
0VDC 1 + Coil
J2 - CPR switch
7 0VDC 2 - Coil
8 Hour meter (optional) Control Pendant
P.C. board 1 0VDC
J9 - Coil/fan/hour meter J14 - Deteq™ module coil 2 0VDC
3 Pendant ON
4 CPR
1 Solenoid - Heels fill 1
5 Alternating
2 Solenoid - Heel 2/4 2 Solenoid - Zone 4
6 Maximum inflate
3 Solenoid - Heel 1/3 3 Solenoid - Zone 3
7 Deteq™ fault
4 Solenoid - Heels dump 4 Solenoid - Zone 2
8
5 + Battery 5 Solenoid - Zone 1
9 +12VDC
6 + Battery 6 + Deteq™ dump
10 +12VDC
7 + Battery
8 + Battery
J1 - Logic P.C. board cable
J11 - Heel solenoids J10 - Main mattress solenoids
W12003_095
FO 3-1
Logic P.C. Board Schematic
W12003_096
FO 3-2
Control Pendant P.C. Board Schematic
W12003_097
FO 3-3
F France S Sverige GR Ellás
Hill-Rom Hill-Rom Medibed AB Meditime
Z.I. du Talhouët PO BOX 3035 Chatziioannidis Nikolaos S.A.
BP 14 Djursholmsvägen 29 Egnatia 142 Str.
56330 Pluvigner 18303 Täby 54622 Thessaloniki
: +33 (0)2 97 50 92 12 : +46 (0)8 756 07 07 : +30 (0)310 263 300
Service Clients : +33 (0)802 01 23 45 Fax : +46 (0)8 732 69 38 Fax : +30 (0)310 263 329
Fax : +33 (0)2 97 50 92 00 e-mail: se.marketing@hill-rom.com