Man 4820 4103 UltiMate 3000 RSLCnano Man48204103 EN
Man 4820 4103 UltiMate 3000 RSLCnano Man48204103 EN
Man 4820 4103 UltiMate 3000 RSLCnano Man48204103 EN
UltiMate™3000
RSLCnano
Standard Applications Guide
Revision: 3.1
All other trademarks are property of Thermo Fisher Scientific and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to
use in the product operation. The document is copyright protected; any reproduction of the
whole or any part of this document is strictly prohibited, except with the written authorization
of Thermo Fisher Scientific Inc.
This manual is provided "as is." The contents of this manual are subject to changes, without
notice, in future revisions.
Thermo Fisher Scientific Inc. makes no representations that this document is complete,
accurate, or error-free. Thermo Fisher Scientific Inc. assumes no responsibility and will not be
liable for any errors, omissions, damage, or loss that might result from any use of this document,
even if the information in the document is followed correctly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a
purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale.
The Terms and Conditions of Sale shall govern all conflicting information between the two
documents.
Manufacturer's address
Dionex Softron GmbH, Part of Thermo Fisher Scientific
Dornierstrasse 4, D-82110 Germering, Germany
Contacting Us
Contacting Us
There are several ways to contact us:
Ordering Information
For ordering information or sales support for HPLC products, contact
your local Thermo Fisher Scientific sales organization. For contact
information, go to Contact Us on http://www.thermofisher.com.
Technical Assistance
For technical support for HPLC products, contact your local Thermo
Fisher Scientific support organization. For contact information, go to
Contact Us on http://www.thermofisher.com.
Contents
1 Using this Manual ............................................................................ 9
1.1 About this Manual ................................................................................................ 10
1.2 Conventions .......................................................................................................... 11
1.2.1 Special Notices and Informational Notes................................................. 11
1.2.2 Typographical Conventions...................................................................... 12
1.3 Reference Documentation.................................................................................... 13
3 FAQs ...............................................................................................83
3.1 NC_Pump Solvent Recalibration – Best Practice .................................................. 84
3.1.1 ProFlow Flow Meter................................................................................. 84
3.1.2 Classic Flow Meter ................................................................................... 84
3.2 Interpreting a Chromatogram............................................................................... 85
3.3 Troubleshooting Nano LC Peptide Applications ................................................... 86
3.4 The Use of TFA and FA .......................................................................................... 88
3.5 Minimizing Baseline Effects .................................................................................. 90
3.5.1 Drift .......................................................................................................... 90
3.5.2 Unstable Baseline..................................................................................... 91
3.6 Typical WPS-3000TPL RS Autosampler Settings for Standard Injection
Routines. ......................................................................................................................... 92
4 Appendix.........................................................................................93
4.1 Customized Sample Injection Routines ................................................................ 94
4.1.1 Introduction to User Defined Program (UDP) Injection Routines ........... 94
4.1.2 Important Considerations when Writing a UDP ...................................... 94
4.1.3 The UDP Commands ................................................................................ 94
4.1.4 Example µL pickup UDP for Maximum Sample Pickup. ........................... 95
4.1.5 Variable Injection Volumes ...................................................................... 96
4.2 Common Application Related Consumables ........................................................ 97
4.2.1 Columns ................................................................................................... 97
4.2.2 nanoViper Capillaries, Sample Loops and Connectors............................. 99
4.3 UltiMate 3000 RSLCnano convenience bundles, list of contents ....................... 101
4.4 Hardware Accessories......................................................................................... 102
This manual contains important information about the correct care and
use of the UltiMate 3000 RSLCnano. Please read this manual carefully
before installing or running any of the applications described. Keep this
manual close to the UltiMate 3000 RSLCnano for future reference and
pass it on to any subsequent user.
1.2 Conventions
This section describes the conventions used throughout this manual
Viewpoint
If not otherwise stated, the expressions left and right in this manual
always refer to the viewpoint of a person that is facing the instrument
from the front.
The electronic version (PDF) of the manual contains numerous links that
you can click to go to other locations within the manual. These include:
Index entries
2 Application Setup
This chapter provides details on each of the application kits available for
the UltiMate 3000 RSLCnano system.
Use a new nano connector sleeve (P/N 6720.0391) each time the
connection is made.
NOTICE: Do NOT use a PTFE sleeve (P/N 160486; supplied with the
columns). The size does not match the nano connector union (Figure 3)
and the pressure resistance is much lower.
1. Slide the black nut and transparent union onto one of the ends of
the fused silica, and the other black nut onto the other fused silica
end Figure 4a.
2. Slide the nano connector sleeve onto one end of the fused silica
until it reaches the middle of the sleeve. Slide the other end of the
fused silica into the connector sleeve. Make sure that the
connection is dead volume free (i.e. that the ends meet in the
middle Figure 4b).
3. Tighten both sides of the black nut equally to ensure that the nano
connector sleeve is in the center of the transparent union Figure
4c.
Figure 4: Nano connector assembly. (a) The two black nuts and
transparent union are mounted on the two outlets to be connected.
(b) A dead volume free connection with the nano connector. (c) The
complete fitting with the black nuts and union housing the nano
connector.
2.1.4 Interfacing the UltiMate 3000 RSLCnano with the Nanospray Flex™ Ion
Source
LC−MS based applications using linear columns commonly use the
Nanospray Flex Ion source (see Figure 41) to interface with Thermo
Scientific Mass Spectrometers. As of January 2019, 1.5 m of fused silica
(20 µm and 50 µm) and a tile for cutting fused silica capillaries have
been included in the pump (NCS-3500RS and NCP-3200RS) accessory
kits. This capillary should be used to connect the outlet of the column or
UV flow cell, if included in the setup, with the emitter installed in the ion
source. The connection between the capillary outlet and the emitter is
realized using a 1/32” micro tight® union assembly included with the
Nanospray Flex Ion Source. For capillaries and columns with 280 µm O.D.
a black sleeve (P/N SC903) should be used with the microtight fitting.
For 360 µm O.D. capillaries and columns (e.g. the Acclaim PepMap RSLC
C18 75 cm PepMap RSLC, P/N 164939) a beige sleeve (P/N SC603)
should be used (both types are included with the ion source). For more
details on connecting the UltiMate 3000 RSLCnano with a Thermo
Scientific mass spectrometer, please refer to “The Complete and Easy
Guide to Configuring Your Thermo Scientific Nano LC for Mass
Spectrometric Analysis”.
NOTICE The sample dilution protocol described here differs from the
product sheet and is designed to offer the user a starting point. The
sample concentration required to run a particular application may
deviate from the sample concentrations given below. Sample dilutions
may also need to be prepared in a different buffer to that given below.
Please check the required sample concentration and dilution conditions
for the application and prepare the sample accordingly!
For nano flow applications, dilute the stock solution to 500 fmol /
µL using mobile phase A (direct injection) or loading buffer (pre-
concentration) as follows:
TIP: To limit the risk of peptide or protein adsorption on the walls of the
vials, Thermo Fisher Scientific recommends using vials containing glass
inserts (Polypropylene vials for WPS with glass insert, 250 µL, set of 100,
P/N 6820.0027).
µ-precolumns are small trap cartridges that are inserted into a cartridge
holder, connected to the switching valve by two 30 µm ID x 100mm
nanoViper capillaries. The stationary phase is retained by a frit at both
ends of the cartridge allowing the mobile phase to flow through it in
both directions without disrupting the column packing. Therefore,
µ-precolumns can be used in both forward- and back-flush operation
(see section 2.2.2 for details). The bed volume is large, but short, giving
it higher absolute loadability compared to nano traps, but the short bed
could result in earlier sample breakthrough for hydrophilic components.
Backpressure is lower compared to nano trap columns and therefore µ-
precolumns can accommodate higher loading flows and are often
preferred when large sample volumes need to be injected.
TIP: Note that for Acclaim PepMap RSLCnano columns, the difference
between the pressures on the nano trap column and analytical column is
smaller than with the combination of the µ-precolumn and the analytical
column.
For nano trap columns, the packing material is only retained by the frit
at one end of the trap column. In order not to damage nano traps, only
forward-flush can be used.
The choice between forward- and back-flush for the µ-precolumn design
is made on the following criteria.
SRD-3400, (Optional):
SRD-3200 with degassing
or SR-3000 without degassing.
WPS-3000TPL RS
- Temperature controlled autosampler
equipped with a
860 bar switching valve
- Optional:
8-port valve (350 bar) for micro-
fractionation applications
The ProFlow flow meter controls pump flow using thermal flow
sensors built into the flow meter. It is a unit dedicated to nano and
low capillary flow rates (50 nL / min – 1500 nL / min) and allows a
pump pressure rating of 900 bar at the full flow rate range for all
common solvents used for reversed phased LC applications.
Column compartment
NOTICE The maximum pump pressure available for the column is 900
bar if a ProFlow flow meter is installed.
Column compartment
NOTICE The maximum pump pressure available for the column is 800
bar with a classic flow meter installed.
2.3.2.5 NCP-3200RS
NCP-3200RS module featuring
- NC pump
Figure 12: NCP 3200RS pump with ProFlow flow meter installed
TIP The NCS-3500RS and NCP-3200RS are both compatible with the
ProFlow and classic flow meters.
TIP An upgrade kit from classic nano to ProFlow is available for both
NCS-3500RS and NCP-3200RS. Please order P/N 6041.7850 (ProFlow
flow meter) and P/N 6041.3003 (Upgrade Kit for ProFlow flow meter).
Please see the ProFlow quick installation guide for more details.
NOTICE The Chromeleon and SII software versions show in Figure 14 are
the minimum requirements. Later versions are fully compatible.
The real seal wash solvent should be installed and primed prior to
powering up the modules (see section 2.3.4.5 for details on how to
prepare the rear seal wash solvent).
Use the PEEK solvent inlet filter frits for both the NC pump and the
loading pump solvent lines. Do not use metal filters.
The online degasser for the loading pump must be employed when:
Notice: 60 minutes are required after pump power up for the flow
meter to reach a stable operating temperature. It is recommended to
execute pump and flow meter purges during this time. Flow meter zero
offset adjustments / calibration routines can only be started after the
module has been powered on for 60 minutes.
Once the software is installed, connect the USB cable(s) to the PC.
Verify that the correct flow meter type is displayed under the flow
meter tab within the instrument configuration panel and that the
valve(s) are correctly configured (oven/valves tab of the NC module).
NOTICE The sample loop volume will vary according to the application.
Ensure that the sample loop volume in the instrument configuration
matches the hardware.
For best results, use premixed Fisher Chemical Optima LC-MS grade
solvents:
NC Pump solvent A - Water with 0.1% Formic Acid (FA) P/N LS118-500
NC Pump solvent B - 80/20 (v/v) Acetonitrile / Water with 0.1% FA P/N
LS122-500
Alternative for NC Pump solvent B (ProFlow only) - Acetonitrile with
0.1% FA P/N LS120-500
The pump block purge time (Table 2) depends on the flow meter type
and whether the solvents are being refreshed or changed:
The flow meter purge time (Table 3) depends on the flow meter type
and the application scale (nano, capillary or micro):
2.3.4.7 Performing the Adjust Zero Balance Test / Pressure Transducer Test
The type of test required will depend on the flow meter installed.
The test “Adjust Zero Balance” is located under the “Wellness” button
on the NC pump tab on the ePanel. A wizard will guide you through the
procedure.
Classic flow meter -> pressure transducer test
TIP The descriptions above refer to the test locations in SII / CM 7.2. For
information on where to find the test in CM 6.8 / DCMSLink please refer
to the NCx-3x00 operating instructions.
Custom solvent calibrations are only valid for the channel on which they
are carried out. To perform a custom solvent calibration, click on the
“Wellness” button on the NC_Pump panel and then”Calibrate Solvent”.
A wizard guides the user through the calibration procedure.
TIP Custom solvent calibrations are also stored locally on the flow meter
hardware and are valid for the life of the flow meter.
The solvent type for the A and B channels should match the solvents
used in the application. To select the solvent type, go to the
PumpModule tab of the ePanel. Under NC Pump -> More Options select
the desired solvent type for each channel from the drop down menu
under ‘solvents’.
Solvent Viscosity %
Water 100
Acetonitrile 50
MeOH 75
IPA 220
Table 4: Viscosity values relative to water for common solvents
TIP The viscosity measurement test takes about 15 minutes. The user
should remain present throughout the test, as prompt user intervention
is required part way through. Once the test is finished, the measured
viscosity values can be stored by selecting the ‘apply’ button for the
respective channel. The results should be examined for plausibility (see
Table 4 for reference values).
SRD-3400 5035.9245
NCS-3500RS 5041.0010A
VWD-3400RS (optional) 5074.0010
3 nL flow cell 6074.0270
WPS-3000TPL RS 5826.0020
Application kit: 6720.0300
# Item P/N
a 75 µm I.D. x 15 cm, packed with Acclaim PepMap RSLC C18, 2 µm, 100Å, nanoViper 164534
nanoViper sample loop 1 µL, FS/PEEK sheathed I.D. x L 100 µm x 127 mm 6826.2401
Polypropylene vials for WPS with glass insert, 250 µL, 25 pcs 6820.0027
Property Setting
Oven temperature 35 °C
WPS temperature 5 °C
The advantage of such a setup is the ease of use and a minimum loss of
peptides, especially hydrophilic ones. The prerequisites of this setup are
i) desalted samples, since all the sample that is injected will enter the
MS, and ii) an investment of extra analysis time to accommodate the
complete loading of sample at low flow rates.
SRD-3400 5035.9245
NCS-3500RS 5041.0020
VWD-3400RS (optional) 5074.0010
45 nL flow cell 6074.0280
WPS-3000TPL RS 5826.0020
Application kit: 6720.0305
# Item P/N
a 300 µm I.D. x 15 cm, packed with Acclaim PepMap RSLC C18, 2 µm, 164537
100Å, nanoViper
nanoViper sample loop 5 µL, FS/PEEK sheathed I.D. x L 200 µm x 159 mm 6826.2405
Polypropylene vials for WPS with glass insert, 250 µL, 25 pcs 6820.0027
Property Setting
Oven temperature 40 °C
WPS temperature 5 °C
Flow Rate 5 µL / min during gradient, 10 µL / min during wash and equilibration
Table 8: Test conditions for direct injection of 1 pmol Cytochrome C onto a capillary column
SRD-3400 5035.9245
NCS-3500RS 5041.0010A
1x 10-port sw. valve 6041.0001A
VWD-3400RS (optional) 5074.0010
3 nL flow cell 6074.0270
WPS-3000TPL RS 5826.0020
Application kit: 6720.0310
TIP The schematic shows a 10-port switching valve. This application can
also be performed using a 6-port valve.
# Item P/N
a 75 µm I.D. x 15 cm, packed with Acclaim PepMap RSLC C18, 2 µm, 100Å, nanoViper 164534
b 300 µm I.D. x 5 mm, packed with Acclaim PepMap100 C18, 5 µm, 100Å (set of 5
160454
cartridges)
µ-Precolumn holder, 5 mm, with 30 µm I.D. connecting tubing, nanoViper fittings 164649
# Item P/N
4 PTFE tubing, 500 µm I.D. 100 cm, used in waste tubing 6720.0077
1/16" Universal Fingertight Fitting, one-piece design, extra long thread (4 pieces) 6720.0015
Polypropylene vials for WPS with glass insert, 250 µL, 25 pcs 6820.0027
Property Setting
Loading time 0.5 minutes (may vary according to required injection volume / routine)
Oven temperature 35 °C
WPS temperature 5 °C
NC Flow Rate 300 nL / min (ProFlow or classic flow meter with nano flow selector)
Table 10 Test conditions for pre-concentration injection of 1 pmol Cytochrome C onto a
nano column
SRD-3400 5035.9245
NCS-3500RS 5041.0020
1x.10 port sw valve 6041.0001A
VWD-3400RS (optional) 5074.0010
3 nL flow cell 6074.0270
WPS-3000TPL RS 5826.0020
Application kit: 6720.0320
TIP The schematic shows a 10-port switching valve. This application can
also be performed using a 6-port valve.
# Item P/N
a PepSwift Monolithic Capillary Column 200 µm I.D. x 5 cm, (PS-DVB), nanoViper 164557
PTFE tubing, 500 µm I.D. x 100 cm, used as waste tubing 6720.0077
1/16" Universal Fingertight Fitting, one-piece design, extra long thread, 4 pieces 6720.0015
Polypropylene vials for WPS with glass insert, 250 µL, 25 pcs 6820.0027
Property Setting
Property Setting
Sample Cytochrome C digest, 1 pmol / µL, in 98% mobile phase A, 2% mobile phase B.
Note: The sample must be diluted in the loading solvent
Loading time 3 minutes (may vary depending on the required injection volume / routine)
Oven temperature 60 °C
WPS temperature 5 °C
TIP When the trap column is switched in line with the analytical column,
a large absorption (injection) peak is detected at 214 nm. This is due to
the different UV absorbance of the ion-pairing agents (e.g. HFBA vs. TFA)
in the loading and analytical solvent.
SRD-3400 5035.9245
NCS-3500RS 5041.0020
1x.10 port sw. valve 6041.0001A
VWD-3400RS 5074.0010
45 nL flow cell 6074.0280
WPS-3000TPL RS 5826.0020
Application kit: 6720.0315
TIP The schematic shows a 10-port switching valve. This application can
also be performed using a 6-port valve.
# Item P/N
a 300 µm I.D. x 15 cm, packed with Acclaim PepMap RSLC C18, 2 µm, 100Å, nanoViper 164537
b 300 µm I.D. x 5 mm, packed with Acclaim PepMap100 C18, 5 µm, 100Å (set of 5
160454
cartridges)
µ-Precolumn holder, 5 mm, with 30 µm I.D. connecting tubing, nanoViper fittings 164649
1 nanoViper capillary FS/PEEK sheathed 1/32" I.D. x L 50 µm x 350 mm 6041.5540
# Item P/N
4 PTFE tubing, 500 µm I.D. 100 cm, used in waste tubing 6720.0077
1/16" Universal Fingertight Fitting, one-piece design, extra long thread (4 pieces) 6720.0015
Polypropylene vials for WPS with glass insert, 250 µL, 25 pcs 6820.0027
Property Setting
Loading time 3 min (may vary according to required injection volume / routine)
Gradient 5% for 1 minute, 5 to 35% B in 10 minutes, 35% B to 90% B in 1 minute, for further
details see TN-72277
Oven temperature 35 °C
WPS temperature 5 °C
Separation
column and
temperature 7 μm internal
control diameter
glass emitter
nanoViper
fitting
The EASY-Spray column simply slots into the EASY-Spray source and is
connected to the LC outlet using a viper union. In-built column
temperature control ensures optimal retention stability and consistent
chromatographic performance. Figure 32 shows the EASY-Spray source
with an EASY-Spray column installed.
Source body
EASY-Spray column
nanoViper connection
Proximity alignment
NOTICE The EASY-Spray Ion Source interface is available in two formats, the EASY-
Spray Ion Source (P/N ES081) and EASY-Spray Ion Source NG (P/N ES082). The source
required will depend on the mass spectrometer type (see Table 15 below).
80
Relative Abundance
0
60 12.00 12.10 12.20
Time (min)
100 B PWHH
40 3.0 sec
20
0 0
14.20 14.30 14.40
5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Time (min)
Time (min)
SR-3000 5035.9200
NCP-3200RS 5041.0030A
WPS-3000TPL RS 5826.0020
Application kit: 6720.0395
SRD-3400 5035.9245
NCS-3500RS 5041.0010A
1x 10-port sw.valve 6041.0001A
WPS-3000TPL RS 5826.0020
Application kit: 6720.0395
TIP The gradient delay resulting from the connecting capillaries varies
according to the length and inner diameter of the tubing. As a rule of
thumb, every 10 cm of 20 µm I.D. tubing contributes 30 nL or 6 seconds
delay at 300 nL/ minute.
Item P/N>
300 µm I.D. x 5 mm, packed with Acclaim PepMap100 C18, 5 µm, 100Å 160454
(set of 5 cartridges)
PTFE tubing, 500 µm I.D. x 100 cm, used as waste tubing 6720.0077
Item P/N>
1/16" Universal Fingertight Fitting, one-piece design, with extra long 6720.0015
thread, 4 pieces
Polypropylene vials for WPS with glass insert, 250 µL, 25 pieces 6820.0027
EASY-Spray transfer lines are available in both nano fIow (P/N ES791A)
and capillary /micro flow (P/N ES792A) compatible formats.
SRD-3400 5035.9245
NCS-3500RS 5041.0010A
2x 10-port sw.valve 6041.0001A
VWD-3400RS 5074.0010
3 nL flow cell 6074.0270
WPS-3000TPL RS 5826.0020
Application kit: 6720.0325
TIP The schematic shows 10-port switching valves. This application can
also be performed on 6-port valves. Ensure that the relative positions on
the connections are correct and update the valve switching in the
instrument setup and method as necessary.
# Item P/N
a 300 µm I.D. x 10 cm, packed with Poros 10 S with connections, 130 µm I.D.
164565
FS sheathed inlet and outlet, nanoViper
b 75 µm I.D. x 15 cm, packed with Acclaim PepMap RSLC C18, 2 µm, 100Å, nanoViper 164534
c 300 µm I.D. x 5 mm, packed with Acclaim PepMap100 C18, 5 µm, 100Å (set of 5
160454
cartridges)
µ-Precolumn holder, 5 mm, with 30 µm I.D. connecting tubing, nanoViper fittings 164649
# Item P/N
5 PTFE tubing, 500 µm I.D. 100 cm, used in waste tubing 6720.0077
1/16” Universal Fingertight Fitting, one piece design, long thread, 4 pieces 6720.0015
Polypropylene vials for WPS with glass insert, 250 µL, 25 pcs 6820.0027
The loading time and desalting time are highly dependent on the
sample quantity and purity. They can be adjusted to meet customer
needs. However, the desalting step must be kept long enough to
avoid the formation of adducts between salt and sample.
The salt plugs listed here have been chosen for the separation of the
protein mix digest. The best sequence of plugs will highly depend on
the affinity of the peptides present in the sample with the IEX
column.
Property Setting
Sample Protein mix digest 100 pmol, lyophilized, prepared according to instruction sheet
Loading time 5 min (may vary according to required injection volume / routine)
Desalting time 7 min (started after the loading time has passed)
Oven Temperature 35 °C
WPS temperature 5 °C
2. 100 mM NaCl (e.g. prepare the 100 mM solution of the first table
two times)
5 mM 50 µL 950 µL 1000 µL
2 mM 20 µL 980 µL 1000 µL
1 mM 10 µL 990 µL 1000 µL
Table 20: Guide for the preparation of the Salt Plugs (dilutions from 100mM NaCl)
Figure 39: Setup for a Tandem nano LC experiment including the optional UV detector
NOTICE Control of the post column nano valve can either be performed
using the WPS-3000FC autosampler or by using an external USB
controlled universal electric actuator (P/N EUHB) and corresponding
mounting hardware (P/N CMH12H) from VICI® Valco Instruments, to
control the valve. If the WPS-3000FC sampler is used, the nano injection
kit (P/N 6824.0030) should be installed and the lower valve on the WPS-
3000FC replaced with the 1/32” nano switching valve (P/N 6820.6232).
All necessary parts are included in the kit. A detailed description of how
to set-up, configure and control both hardware variants is given in
TN72899.
# Item P/N
a 75 µm I.D. x 15 cm, packed with Acclaim PepMap RSLC C18, 2 µm, 100Å, nanoViper 164534
b 300 µm I.D. x 5 mm, packed with Acclaim PepMap100 C18, 5 µm, 100Å (set of 5
160454
cartridges)
µ-Precolumn holder, 5 mm, with 30 µm I.D. connecting tubing, nanoViper fittings 164649
1 nanoViper capillary FS/PEEK sheathed 1/32" I.D. x L 20 µm x 350 mm 6041.5240
5 PTFE tubing, 500 µm I.D. 100 cm, used in waste tubing 6720.0077
1/16” Universal Fingertight Fitting, one piece design, long thread, 4 pieces 6720.0015
Fused silica tubing I.D. 20µm O.D. 280µm, 5 meters for nano LC connections 160475
Polypropylene vials for WPS with glass insert, 250 µL, 25 pcs 6820.0027
Property> Setting
Oven Temperature 35 °C
WPS temperature 5 °C
Complete LC method parameters for this set-up are also available for
download from the AppsLab library.
If UV detection is required, the cap flow cell (45 nL; P/N 6074.0280) is
recommended for all micro LC applications.
Item P/N
Fused silica tubing I.D. 20 µm ±3 µm/O.D. 280 µm ±10 µm, 5 meters 160475
Fused silica tubing I.D. 50 µm ±3 µm/O.D. 280 µm ±10 µm, 5 meters 160477
PTFE tubing, 250 µm I.D., low pressure connection of 280 µm O.D. 6720.0030
fused silica capillaries, 5 pieces
1/16" Valco Ferrule and Nut, stainless steel, 10 pc. (for 10-port valve) 161103
PEEK sleeves, precision cut and polished for connections with fused 6720.0064
silica tubing (280 µm O.D.), 5 pieces
PEEK sleeves, precision cut and polished for connections with fused 6720.0078
silica tubing (360 µm O.D.), 5 pieces
PEEK sleeves, precision cut and polished for connections with fused 6720.0075
silica tubing (280 µm O.D.), 10 pieces
PEEK sleeves, precision cut and polished for connections with 6720.0076
Microtight Union (360 µm O.D.), 10 pieces
TIP The connection between the LC separation and the MS should have
the lowest volume possible to minimize dispersion. Keep in mind that
internal diameter has a much bigger effect on dispersion then the
length of the connecting tubing. Ensure that tubing with the correct
diameter is used at all times.
MS compatibility
LTQ™ and VelosTM Series
Orbitrap™ Series
Exactive™ series
Legacy TSQ™ series (Quantum Access
Max, Vantage, Ultra etc.)
MS compatibility
2.15.1.2 Heated Electrospray Ionization (HESI-II) Probe and H-ESI Spray Insert
MS Compatibility
LTQ™ and Velos™ Series
Orbitrap™ Series
Exactive™ series
Legacy TSQ™ series (Quantum Access
Max, Vantage, Ultra etc.)
H-ESI Spray Insert for Ion Max NG Source
P/N 80000-60321
Figure 42: Ionization probes for the Ion Max and Ion Max NG sources.
Figure 43: Ionization probes for the Ion Max and Ion Max NG sources
3 FAQs
NOTICE The pump blocks and flow meter should be purged regularly
with fresh solvents. Purging is especially important, before any
calibration routines are carried out. Please see section 2.3.4.6 for details.
The finite volume of an HPLC system results in time offset between the
formation of a gradient, its delivery onto the column and the detection
of the gradient change by the (UV and / or MS) detector. This so-called
gradient delay can be visualized by comparing the programmed gradient
to the UV signal. Figure 44 shows the gradient delay between pump and
UV detector.
The dwell volume represents the volume between the autosampler and
the nano columns. Since there are usually one (direct injection) or two
(pre-concentration) valve switches involved in the application,
introducing additional capillaries and connections, the dwell volume and
gradient delay are not the same volume in direct injection and pre-
concentration setups.
Most commonly, trifluoro acetic acid (TFA) and formic acid (FA) are
used. In this manual and in most LC-MS applications, FA is preferred as
its use minimizes ion-suppression effects. TFA is a stronger ion-pairing
agent and results in better chromatography, but can result in ionization
suppression. The use of TFA is generally restricted to the loading buffer
or when increased retention (compared to FA) is necessary. When
performing the applications described in this manual with TFA instead of
FA, use the amounts given in Table 24. Figure 46 demonstrates the
effect of ion pairing agent choice on the chromatographic separation of
Cytochrome C.
3.5.1 Drift
Ensure that the UV lamp has been switched on for sufficient time in
order that the lamp temperature can stabilize. Chromeleon can detect
this and will give a warning during the ‘Ready Check‘ whenever the UV
lamp temperature is not stable. In such cases, the UV detector can be
used but may not perform optimally.
Solvent A Solvent B
0.1% 0.08%
0.05% 0.04%
The age of both lamp and flow cell can have a significant influence on
baseline drift. New lamps and flow cells may show some drift during the
so-called ‘burn in’ period. Allowing sufficient equilibration time for the
lamp is necessary for obtaining a stable baseline.
short period with a strong acidic solution; see the Variable Wavelength
Detector’s Operating Instructions for more details.
DrawSpeed Sets the speed of the syringe used for drawing 0.2 µL / sec
the sample.
DrawDelay Sets the time that the needle remains in the 5 sec
vial after drawing the sample
WashSpeed Sets the speed of the syringe for the wash 4 µL / sec
cycle.
Loop Overfill Loop overfill factor used in full loop only 2.0
4 Appendix
UDPs provide the user with full control over every aspect of the injection
routine and enable every type of injection, from simply filling the loop
and injecting the sample to complete liquid handling, such as in-well
digestion.
Step ‘1’ is often omitted, as samples are usually ready to inject when
placed in the autosampler.
Step 4 – Injecting the sample and starting the run (commands 10 – 11)
Step 5 – Washing the syringe and needle and preparing for the next
sample (commands 12-14)
The following steps describe how to adapt the method given in section
4.1.4 to permit injection volumes according to those defined in the
sample table.
Step 2 - Navigate to the line where the sample vial volume is defined
4.2.1 Columns
The most common analytical columns used with the UltiMate 3000
RSLCnano system are listed below. For a comprehensive guide of all
available column formats please refer to the Thermo Scientific
Chromatography columns and consumables catalogue.
Item P/N
Acclaim PepMap RSLC C18, 2 μm, 100 Å, 50 μm I.D. x 15 cm, nanoViper 164562
Acclaim PepMap RSLC C18, 2 μm, 100 Å, 50 μm I.D. x 25 cm, nanoViper 164709
Acclaim PepMap RSLC C18, 2 μm, 100 Å, 50 μm I.D. x 50 cm, nanoViper 164710
Acclaim PepMap RSLC C18, 2 μm, 100 Å, 75 μm I.D. x 15 cm, nanoViper 164534
Acclaim PepMap RSLC C18, 2 μm, 100 Å, 75 μm I.D. x 25 cm, nanoViper 164536
Acclaim PepMap RSLC C18, 2 μm, 100 Å, 75 μm I.D. x 50 cm, nanoViper 164540
Acclaim PepMap RSLC C18, 2 μm, 100 Å, 75 μm I.D. x 75 cm, nanoViper 164939
Item P/N
300 µm i.d. x 5 mm, packed with Acclaim PepMap100 C18, 5 µm, 100Å (set
160454
of 5 cartridges)
Item P/N
EASY-Spray column, 75 cm x 150 µm I.D., PepMap RSLC C18, 2 µm, 100Å ES806A
EASY-Spray emitter, nano-flow (20µm I.D. transfer line, 7µm I.D. emitter) ES791A
EASY-Spray emitter, micro-flow (75µm I.D. transfer line, 20µm I.D. emitter) ES792A
Table 28: EASY-Spray columns
/N
Item P/N
Item P/N
10 μm 20 μm 50 μm 75 μm 100 μm 150 μm
[Green] [Orange] [Brown] [Black] [Red] [Purple]
850
Description P/N
Item P/N
Solvent Rack 5035.9245 SRD-3400 with degasser 5035.9245 SRD-3400 with degasser 5035.9245 SRD-3400 with degasser 5035.9245 SRD-3400 with degasser
NCS-3500RS nano pump with 5041.0010A NCS-3500RS nano pump with 5041.0020 NCS-3500RS Cap pump with 5041.0010A NCS-3500RS nano pump with
Pump 5041.0010A
integrated column compt. integrated column compt. integrated column compt. integrated column compt.
Autosampler 5826.0020 WPS-3000RPL RS nano/cap 5826.0020 WPS-3000RPL RS nano/cap 5826.0020 WPS-3000RPL RS nano/cap 5826.0020 WPS-3000RPL RS nano/cap
VWD-3400RS 4 channels
Detector 5074.0010 - - - - - -
without flow cell
Accessory 2 6041.0001A 2P-10P valve for NCS-3500 6041.0001A 2P-10P valve for NCS-3500 6041.0001A 2P-10P valve for NCS-3500 6041.0001A 2P-10P valve for NCS-3500
Accessory 3 6720.0310 Preconcentration nano kit 6720.0310 Preconcentration nano kit 6720.0315 Preconcentration cap kit 6720.0395 Easy-Spray connection kit
Accessory 4 6000.1004 Signal cable, 6pol.5m 6000.1004 Signal cable, 6pol.5m 6000.1004 Signal cable, 6pol.5m 6000.1004 Signal cable, 6pol.5m
Column to MS tubing ID x OD 6041.5295 Column to MS tubing ID x OD x 6826.2405 nanoViper sampler loop for
Accessory 8 6041.5295 - -
x L, 50µm x 360µm x 1m L, 50µm x 360µm x 1m WPS-TPL RS, 5 µL
Accessory 9 161089 Cytochrome C digest sample 161089 Cytochrome C digest sample 161089 Cytochrome C digest sample - -
Item P/N
Item P/N