Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

PeakTech 3360

Download as pdf or txt
Download as pdf or txt
You are on page 1of 29

PeakTech® 3355 / 3360

Bedienungsanleitung /
operation manual /
Mode d’emploi /
Istruzioni per l'uso /
Manual de instrucciones

Digital Multimeter /
Multimètre digital /
Multimetro digitale /
Multímetro digital
1. Safety Precautions
This product complies with the requirements of the following
European Community Directives: 2004/108/EC (Electromagnetic
Compatibility) and 2006/95/EC (Low Voltage) as amended by
2004/22/EC (CE-Marking). Overvoltage category III 1000V; over-
voltage category IV 600V; pollution degree 2.

CAT I: For signal level, telecommunication, electronic with small


transient over voltage

CAT II: For local level, appliances, main wall outlets, portable
equipment

CAT III: Distribution level, fixed installation, with smaller transient


over-voltages than CAT IV.

CAT IV: Units and installations, which are supplied overhead lines,
which are stand in a risk of persuade of a lightning, i.e.
main-switches on current input, over-voltage-diverter,
current use counter.

To ensure safe operation of the equipment and eliminate the


danger of serious injury due to short-circuits (arcing), the following
safety precautions must be observed.

Damages resulting from failure to observe these safety precautions


are exempt from any legal claims whatever.

* Do not use this instrument for high-energy industrial installation


measurement. This instrument is intended for use in installation
over-voltage category III.
* Do not exceed the maximum permissible input ratings (danger
of serious injury and/or destruction of the equipment).

-34-
* The meter is designed to withstand the stated max voltages. If it
is not possible to exclude without that impulses, transients,
disturbance or for other reasons, these voltages are exceeded
a suitable prescale (10:1) must be used.
* Replace a defective fuse only with a fuse of the original rating.
Never short-circuit fuse or fuse holding.
* Disconnect test leads or probe from the measuring circuit
before switching modes or functions.
* Do not conduct voltage measurements with the test leads
connected to the mA/A- and COM-terminal of the equipment.
* The 10A-range is protected. To avoid damage or injury, use the
meter only in circuits limited by fuse or circuit breaker to 10A or
2000VA.
* To avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any resistance
measurements.
* Do not conduct current measurements with the leads connected
to the V/Ω-terminals of the equipment.
* Check test leads and probes for faulty insulation or bare wires
before connection to the equipment.
* To avoid electric shock, do not operate this product in wet or
damp conditions. Conduct measuring works only in dry clothing
and rubber shoes, i. e. on isolating mats.
* Never touch the tips of the test leads or probe.
* Comply with the warning labels and other info on the
equipment.
* Always start with the highest measuring range when measuring
unknown values.
* Do not subject the equipment to direct sunlight or extreme
temperatures, humidity or dampness.
* Do not subject the equipment to shocks or strong vibrations.
* Do not operate the equipment near strong magnetic fields
(motors, transformers etc.).
* Keep hot soldering irons or guns away from the equipment.
* Allow the equipment to stabilize at room temperature before
taking up measurement (important for exact measurements).
* Do not input values over the maximum range of each
measurement to avoid damages of the meter.
-35-
* Do not turn the rotary function switch during voltage or current
measurement, otherwise the meter could be damaged.
* Use caution when working with voltages above 35 V DC or
25 V AC. These Voltages pose shock hazard.
* Replace the battery as soon as the battery indicator “BAT”
appears. With a low battery, the meter might produce false
reading that can lead to electric shock and personal injury.
* Fetch out the battery when the meter will not be used for long
period.
* Periodically wipe the cabinet with a damp cloth and mid
detergent. Do not use abrasives or solvents.
* The meter is suitable for indoor use only
* Do not operate the meter before the cabinet has been closed
and screwed safely as terminal can carry voltage.
* Do not store the meter in a place of explosive, inflammable
substances.
* Do not modify the equipment in any way
* Do not place the equipment face-down on any table or work
bench to prevent damaging the controls at the front.
* Opening the equipment and service – and repair work must only
be performed by qualified service personnel
* Measuring instruments don’t belong to children hands.

1.1. Input limits

Input limits 1000 V DC/ACrms less than


10 sec.
mA range AC/DC 400 mA AC/DC 1000 V / 0,5 A
(fused)
A range 10 A AC/DC 1000 V / 10 A
(fused)
Frequency, Resistance, 600 V AC/DCrms (P 3355)
Capacitance, Duty Cycle, Diode 1000 V AC/DCrms (P 3360)
test, Continuity and
Temperature
-36-
1.2. Safety Symbols

This symbol adjacent to another symbol, terminal or


operating device indicates that the operator must refer to
an explanation in the operating instructions to avoid
personal injury or damage to the meter.

This symbol advices the user that the terminals so


marked must not be connected to a circuit point at which
the voltage, with respect to earth ground, exceeds (in this
case) 1000 V AC or VDC

This symbol adjacent to one or more terminals identifies


them as being associated with ranges that may, in normal
use, be subjected to particulary hazardous voltages. For
maximum safety, thermometer and its test leads should
not be handled when these terminals are energized.

Indicates protection class II, double insulation

IP67 Dust- and waterproof

µA The maximum current that can measure with this

mA terminal is 400 mA AC/DC. This terminal is protected by


500 mA/1000 V fuse

A The maximum current that can measure with this terminal


is 10 A AC/DC. This terminal is protected by 10 A/1000 V
fuse.

-37-
CAUTION!
Note on using the supplied safety test leads according the
IEC / EN 61010-031:2008:

Measurements in the field of overvoltage category CAT I or CAT II


can be performed with test leads without sleeves with a maximum
of up to 18mm long, touchable metallic probe, whereas for
measurements in the field of overvoltage category CAT III or CAT
IV test leads with put on sleeves, printed with CAT III and CAT IV
must be used, and therefore the touchable and conductive part of
the probes have only max. 4mm of length.

CAUTION !
This CAUTION symbol indicates a potentially hazardous situation,
which if not avoided, may result in minor or moderate injury, or
damage to the product or other property.

2. Technical Data
2.1. Specifications

Display P 3355
26 mm , 3 ¾ digit, 4000 counts with
automatic polarity indication and
backlight

P 3360
19 mm, 4 ¾ digit, 40000 county with
automatic polarity indication, backlight
and 42 segment bargraph

Overrange indicaton: "OL"

Low battery indication: Battery symbol indicates low battery


condition (P 3360)

Measuring rate: 2 times / sec.

-38-
Auto / manual Ranging

Auto power off: about 15 min.

P 3360 : It is possible to turn off the APO (find on


page 53)

Overload protection: on all ranges

measuring functions: True RMS measurement (only P 3360)


Auto- or manual range selection
Data Hold
Relative zero
Min/Max and Peak-Hold (only P 3360)

Overload protection: on all ranges

Operating Temperature: 41°F to 104°F (5°C to 40°C)

Storage Temperature: -4°F to 140°F (-20...+60° C)

Accuracy Temperature: 64°F to 82°F (18°C to 28°C) to


maintain guaranteed accuracy

Relative Humidity: < 80% operating

Dimensions (W x H x D): 85 x 185 x 55 mm

Weight: approx. 400 g

Power source: 9 V battery (Neda 1604)

Accessories supplied: Operation manual, test leads, socket


seals, Battery, K-type thermocouple,
Adaptor for thermocouple, carrying
case

-39-
2.2. Electrical Specifications

DC Volts

Model Range Resolution Accuracy


P 3355 400 mV 100 µV ± 0,5% rdg. + 2 dgt
4 V 1 mV
± 1,2% r dg. + 2 dgt
40 V 10 mV
400 V 100 mV
1000 V 1 V ± 1,5% rdg. + 2 dgt
P 3360 400 mV 10 µV
4 V 100 µV ± 0,06% rdg. + 2 dgt
40 V 1 mV
400 V 10 mV
1000 V 100 mV ± 0,1% rdg. + 5 dgt
Overload protection 1000V AC/DCrms
Input Impedance: (P 3355) 7,8 MΩ / (P 3360) 10 MΩ

AC Volts

Model Range Resolution Accuracy


P 3355 400 mV 100 µV ± 1,0% rdg. + 5 dgt
4 V 1 mV
40 V 10 mV ± 1,0% rdg. + 3 dgt
400 V 100 mV
1000 V 1 V
P 3360 400 mV 100 µV ± 1,0% rdg. + 5 dgt
4 V 1 mV
40 V 10 mV ± 1,0% v.M. + 3 dgt
400 V 100 mV
1000 V 1 V
Overload protection 1000V AC/DCrms
Input Impedance: (P 3355) 7,8 MΩ / (P 3360) 3 MΩ
AC Response: (P 3355) 50 to 400Hz / (P 3360) 50 to
1000Hz

-40-
DC Current

Model Range Resolution Accuracy


P 3355 400 µA 0,1 µA ± 1,0 % rdg. + 3 dgt.
4000 µA 1 µA .
40 mA 10 µA ± 1,5% rdg. + 3 dgt.
400 mA 100 µA
10 A 10 mA ± 2,5% rdg. + 5 dgt.
P 3360 400 µA 0,01 µA
4000 µA 0,1 µA .
40 mA 1 µA ± 1,0% rdg. + 3 dgt.
400 mA 10 µA
10 A 1 mA
Overload protection: 0,5A / 1000V and 10A / 1000V
Max. Input: 400mA DC on mA – ranges and
10A on A – range

AC current

Model Range Resolution Accuracy


P 3355 400 µA 0,1 µA ± 1,5 % rdg. + 5 dgt.
4000 µA 1 µA .
40 mA 10 µA ± 1,8% rdg. + 5 dgt.
400 mA 100 µA
10 A 10 mA ± 3,0% rdg. + 7 dgt.
P 3360 400 µA 0,1 µA
4000 µA 1 µA .
40 mA 10 µA ± 1,5% rdg. + 3 dgt.
400 mA 100 µA
10 A 10 mA
Overload protection: 0,5A / 1000V and 10A / 1000V
Max. input: 400mA DC on mA – Range and
10A on A – Range
AC Response : (P 3355) 50 to 400Hz /
(P 3360) 50 to 1000Hz

-41-
Resistance

Model Range Resolution Accuracy


P 3355 400 Ω 100 mΩ ± 1,2% rdg. + 4 dgt.
4 kΩ 1 Ω ± 1,0% rdg. + 2 dgt.
40 kΩ 10 Ω .
400 kΩ 100 Ω ± 1,2% rdg. + 2 dgt
4 MΩ 1 kΩ .
40 MΩ 10 kΩ ± 2,0% rdg. + 3 dgt.
P 3360 400 Ω 10 mΩ ± 0,3% rdg. + 9 dgt.
4 kΩ 100 mΩ
40 kΩ 1 Ω ± 0,3% rdg. + 4 dgt.
400 kΩ 10 Ω
4 MΩ 100 Ω
40 MΩ 1 kΩ ± 2,0% rdg. + 10 dgt.
Overload protection: 600 V DC/ACrms

Capacitance

Model Range Resolution Accuracy


P 3355 4 nF 1 pF ± 5,0% rdg. + 20 dgt
40 nF 10 pF ± 5,0% rdg. + 7 dgt
400 nF 100 pF
4 µF 1 nF ± 3,0% rdg. + 5 dgt
40 µF 10 nF
100 µF 100 nF ± 5,0% rdg. + 5 dgt
P 3360 40 nF 1 pF ± 3,5% rdg. + 40 dgt
400 nF 10 pF
4 µF 100 pF
40 µF 1 nF ± 3,5% rdg. + 10 dgt
400 µF 10 nF
4000 µF 100 nF ± 5,0% rdg. + 10 dgt
40 mF 1 µF
Overload protection: 600 V DC/ACrms

-42-
Frequency

Model Range Resolution Accuracy


P 3355 10 Hz 1 mHz ± 1,5% rdg. +5 dgt
100 Hz 10 mHz
1000 Hz 100 mHz
10 kHz 1 Hz ± 1,2% rdg.+3 dgt
100 kHz 10 Hz
1000 kHz 100 Hz
10 MHz 1 kHz ±1,5% rdg.+4dgt
P 3360 40 Hz 1 mHz
400 Hz 10 mHz
4 kHz 100 mHz
40 kHz 1 Hz ± 0,1% rdg. +1 dgt
400 kHz 10 Hz
4 MHz 100 Hz
40 MHz 1 kHz
100 MHz 10 kHz not specified
Overload protection: 600 V DC/ACrms
Sensitivity: <0,5V RMS at ≤1MHz /
>3V RMS at >1MHz (P 3355)
Sensitivity: <0,8V RMS at ≤100kHz /
>5V RMS at >100kHz (P 3360)

-43-
Duty cycle

Model Range Resolution Accuracy


P 3355 0,1...99,9 % 0,1% ± 1,2% rdg. +2 dgt.
Pulse width: >100 µs,< 100 ms
Frequency: 5Hz-150kHz
Sensitivity: <0,5Vrms
P 3360 0,1...99,9 % 0,01% +/- 1,2% rdg. +2 dgt.
4-20mA% 0,01% +/-50 dgt.
-25...125% 0 mA = -25%; 4 mA = 0 %;
20 mA = 100 %, 24 mA = 125 %
Pulse width: >100 µs,< 100 ms
Frequency: 5Hz-150kHz
Sensitivity: <0,5Vrms
Overload protection: 600 V DC/ACrms

Temperature

Model Range Resolution Accuracy


P 3355 -20 ... +760°C 1°C ± 3,0% rdg. + 3,0°C
-4...+1400°F 1°F ± 3,0% rdg. + 3,0°F.
P 3360 -50...+1000°C 1°C ± 1,0% rdg. + 2,5°C
-58...+1832°F 1°F ± 1,0% rdg. + 4,5°F
Overload protection: 600 V DC/ACrms
Temperature Accuracy : Probe accuracy not included.

-44-
Continuity

Model Audible Threshold Test current


P 3355 < 150 Ω <0,3 mA
P 3360 < 35 Ω < 0,35 mA
Overload protection: 600 V DC/ACrms

Diode Test

Model Test current Open circuit


voltage
P 3355 0,3 mA 1,5 V
P 3360 0,9 mA 2,8 V
Overload protection: 600 V DC/ACrms

-45-
P 3360

1.

2. 11.

3. 10.

9.
4.

5.

8.

6. 7.

1. LCD Display
2. Relative pushbutton
3. Range pushbutton
4. Mode pushbutton
5. Function switch
6. µA/mA/10 A (positive) input jack
7. COM (negative) input jack
8. Positive input jack for voltage, Hz, Duty cycle, Ohms, Diode,
Continuity, Capacitance, and temperature measurements
9. Data Hold and Backlight pushbutton
10. PEAK button
11. MIN / MAX button
-47-
3.1. Symbols in the LCD Display

•))) Continuity
->|- Diode test
BAT Battery status
n nano (10-9)
µ micro (10-6)
m milli (10-3)
A current
k kilo (103)
F capacitance (Farads)
M mega (106)
Ω Resistance
Hz Frequency
% Duty cycle
AC Alternating current
DC Direct current
°F Temperature in Fahrenheit
°C Temperature in Centigrade
MAX Maximum
MIN Minimum
PEAK Peak hold
V voltage
REL Relative
AUTO Autoranging
HOLD Display hold

4. Operating instructions
Warning!
Risk of electrocution. High-voltage circuits, both AC and DC are
very dangerous and should be measured with great care.

1. Always push the power pushbutton to the OFF position when


the meter is not in use. This meter has Auto OFF that
automatically shuts the meter OFF if 15 minutes elapse
between uses.

-48-
2. If "OL" appears in the display during a measurement, that value
you are measuring exceeds the range you have selected.
Change to higher range.

Note:
On some low AC and DC voltage ranges, with the test leads not
connected to a device, the display may show a random, changing
reading. This is normal and is caused by the high-input sensitivity.
The reading will stabilize and give a proper measurement when
connected to a circuit.

3. Press the "POWER" pushbutton to turn the meter on and off.

4.1. Autoranging / manual range selection

When the meter is first turned on, it automatically goes into Auto-
Ranging. This automatically selects the best range for the
measurements being made and is generally the best mode for most
measurements. For measurement situations requiring that the
range be manually held, perform the following:

1. Press the "RANGE" button. The "AUTO" indicator will


extinguish and the currently selected range will be held.

2. Press the "RANGE" button to step through the available ranges


until you select the range you want.

3. Press and hold the "RANGE" button for 2 seconds to exit the
manual ranging mode and return to "AUTO" mode.

4.2. Backlight

The backlight function is used to illuminate the display when the


meter is used at night or in dimly lighted area.

1. Press the "HOLD" button (P 3355) or the backlight button


(P 3360) for more than 2 seconds, the "HOLD" indicator will
appear and backlight will be activated.
-49-
Note (P 3360):
The backlight automatically turns off after 30 seconds.

2. Press the "HOLD" button momentarily to remove the "HOLD"


function. (P 3360).

3. Press the "HOLD" button (P 3355) or the backlight button (P


3355) for more than 2 seconds to remove the backlight function.

4. Press the "HOLD" button momentarily to remove the "HOLD"


function. (P 3360)

4.3 Auto Power Off

The auto power off function will turn the meter off after 15 minutes
operation time. To disable this function (only P 3360) press and
hold the MODE button and then power on the meter. The display
shows “ APO d “. To activate the APO function the meter must be
power off. Then power on the meter.

5. Measuring
5.1. Relative

The relative measurement feature allows you to make


measurements relative to a stored reference value. A reference
voltage, current, etc. can be stored and measurements made in
comparison to that value. The displayed value is the difference
between the reference and the measured value.

1. Perform any measurement as described in the operating


instructions.

2. Press the "REL" button to store the reading in the display and
the "REL" indicator will appear on the display.

3. The display will now indicate the difference between the stored
value and the measured value.
-50-
5.2. Data Hold

The data hold function allows the meter to "freeze" a measurement


for later reference.

1. Press the "HOLD" button to freeze the display, the "HOLD"


indicator will appear in the display.

2. Press the "HOLD" button to return to normal operation.

5.3. DC Voltage measurements

Caution:
Do not measure DC voltages if a motor on the circuit is being
switched ON or OFF. Large voltage surges may occur during the
ON or OFF operations that can damage the meter.

1. Set the function switch to the "V" position.

2. If "AC" appears in the display, press the "MODE"


pushbutton until "DC" appears in the display.

3. Insert the black test lead banana plug into the negative COM
jack and the red test lead banana plug into the positive V jack.

4. Touch the test probe tips to the circuit under test. Be sure to
observe the correct polarity (red lead to positive, black lead to
negative).

5. Read the voltage in the display. The display will indicate the
proper decimal point and value. If the polarity is reserved, the
display will show (-) minus before the value.

-51-
5.4. AC voltage measurements

Warning:
Risk of Electrocution. The probe tips may not be long enough to
contact the live parts inside some 230 V outlets for appliances
because the contacts are recessed deep in the outlets. As a result,
the reading may show 0 volts when the outlet actually has voltage
on it. Make sure the probe tips are contacting the metal contacts
inside the outlet before assuming that no voltage is present.

Caution:
Do not measure AC voltages if a motor on the circuit is being
switched ON or OFF. Large voltage surges may occur during the
ON or OFF operations that can damage the meter.

1. Set the function switch to the "V" position.

2. Press the "MODE" button until "AC" appears in the display.

3. Insert the black test lead banana plug into the negative COM
jack and the red test lead banana plug into the positive V jack.

4. Touch the test probe tips to the circuit under test.

5. Read the voltage in the display. The display will indicate the
proper decimal point, value and symbol.

5.5. DC current measurement

1. Insert the black test lead banana plug into the negative COM
jack.

2. For current measurements up to 4000 µA DC, set the function


switch to the "µA" position and insert the red test lead banana
plug into the mA-µA jack.

-52-
3. For current measurements up to 400 mA DC, set the function
switch to the "mA" position and insert the red test lead banana
plug into the mA-µA jack.

4. For current measurements up to 10 A DC, set the function


switch to the 10 A position and insert the red test lead banana
plug into the 10 A jack.

5. Press the “MODE” button until "DC" appears in the display.

6. Remove power from the circuit under test and open the circuit
at the point where you wish to measure current.

7. Touch the black test probe tip to the negative side of the circuit
and touch the red test probe tip to the positive side of the
circuit.

8. Apply power to the circuit.

9. Read the current in the display. The display will indicate the
proper decimal point, value and symbol.

5.6. AC current measurements

Warning:
To avoid electric shock do not measure AC current on any circuit
whose voltage exceeds 250 V AC.

1. Insert the black test lead banana plug into the negative COM
jack.

2. For current measurements up to 4000 µA AC, set the function


switch to the "µA" position and insert the red test lead banana
plug into the mA-µA jack.

3. For current measurements up to 400 mA DC, set the function


switch to the "mA" position and insert the red test lead banana
plug into the mA-µA jack.
-53-
4. For current measurements up to 10 A DC, set the function
switch to the 10 A position and insert the red test lead banana
plug into the 10 A jack.

5. Press the “MODE” button until "AC" appears in the display.

6. Remove power from the circuit under test and open the circuit
at the point where you wish to measure current.

7. Touch the black test probe tip to the negative side of the circuit
and touch the red test probe tip to the positive side of the
circuit.

8. Apply power to the circuit.

9. Read the current in the display. The display will indicate the
proper decimal point, value and symbol.

5.7. Resistance measurements

Warning:
To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any resistance
measurements. Remove the batteries and unplug the line cords.

1. Set the function switch to the "OHM" position.

2. Insert the black test lead banana jack into the negative COM
jack and the red test lead banana plug into the positive Ω jack.

3. Touch the test probe tips across the circuit or part under test. It
is best to disconnect one side of the part under test so the rest
of the circuit will not interfere with the resistance reading.

4. Read the resistance in the display. The display will indicate the
proper decimal point, value and symbol.

-54-
When you short the test leads in the 400 Ω range, your meter
display a small value (no more than 0.3 Ω). This value is due to
your meter´s and test leads internal resistance. Make a note of this
value and subtract it from small resistance measurements for better
accuracy.

5.8. Continuity check

Warning:
To avoid electric shock, never measure continuity on circuits or
wires that have voltage on them.

1. Set the function switch to the "OHM" position.

2. Insert the black test lead banana jack into the negative COM
jack and the red test lead banana-plug into the positive Ω jack.

3. Press the "MODE" push button until the " )))" symbol appears
in the display.

4. Touch the test probe tips to the circuit or wire you wish to
check.

5. If the resistance is less than 150 ohms (P 3355) or 35 ohms


(P 3360), the audible signal will sound. The display will also
show the actual resistance.

5.9. Diode test

Warning:
To avoid electric shock, do not test any diode that has voltage on it.

1. Set the function switch to " " position.

2. Press the "MODE" button until the " " symbol appears in
the display.
-55-
3. Insert the black test lead banana plug into the negative COM
jack and the red test lead banana plug into the positive Ω jack.

4. Touch the test probe tips to the diode or semiconductor junction


you wish to test. Note the meter reading.

5. Reverse the probe polarity by switching probe position. Note


this reading.

6. The diode or junction can be evaluated as follows:

A: If one reading shows a value and the other reading show


OL, the diode is good.
B: If both readings are OL, the device is open.
C: If both readings are very small or 0, the device is shorted.

Note: The value indicated in the display during the diode check is
the forward voltage.

5.10. Frequency / Duty cycle measurement

1. Set the function switch to the "Hz" position.

2. Insert the black test lead banana jack into the negative COM
jack and the red test lead banana plug into the positive Hz jack.

3. Touch the test probe tips to the circuit under test.

4. Read the frequency in the display. The digital reading will


indicate the proper decimal point, symbols (Hz, kHz, MHz) and
value.

5. Press the "MODE" button to display duty cycle in %.

-56-
5.11. Capacitance measurements

Warning:
To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any capacitance
measurements. Remove the batteries and unplug the line cords.

1. Set the function switch to the "CAP" ┤├ position.

2. Insert the black test lead banana jack into the negative COM
jack and the red test lead banana plug into the positive ┤├
jack.

3. Touch the test leads to the capacitor to be tested. The display


will indicate the proper decimal point, value and symbol.

5.12. Temperature measurements

Warning:
To avoid electric shock, disconnect both test probes from any
source of voltage before making a temperature measurement.

Insert the adapter in the input socket for temperature


measurements.

Insert the K-type thermocouple into the temperature socket, making


sure to observe the correct polarity. Touch the temperature probe
head to the part whose temperature you wish to measure. Keep the
probe touching the part under test until the reading stabilize (about
30 seconds). Read the temperature in the display. The digital
reading will indicate the proper decimal point and value.

Warning:
To avoid electric shock, be sure the thermocouple has been
removed before changing to any other measurement function.

-57-
6. Replacing the fuses
WARNING!
To avoid electric shock, disconnect all the test probes before
removing the fuse. Replace only with the same type of fuse. Not
note remove the top cover. Service should be performed only by
qualified personnel.

CAUTION!
For continued protection against fire or other hazard, replace only
with fuse of the specified voltage and current ratings.

Follow these steps to replace the fuse:

1. Press ON/OFF button to turn the meter off and disconnect the
test probes.

2. Remove the Battery cover (screws B) and the battery.

3. Remover the back cover by unscrewing the 6 screws (A)

4. Remove the blown fuse. Use only a fuse with the proper size
and value 500mA/1000V (6,3x32mm) or 10A/1000V (10x38mm)

5. Replace the back cover, the Battery and the battery cover.

-58-
WARNING!
Do not operate your meter until the back cover is in place and fully
closed.

7. Maintenance
Warning:
To avoid electric shock, disconnect the test leads from any source
of voltage before removing the back cover or the battery/fuse door.

Warning:
To avoid electric shock, do not operate your meter until the battery/
fuse door are in place and fastened securely. This multimeter is
designed to provide years of dependable service, if the following
care instructions are performed.

1. Keep the meter dry. If it gets wet, wipe it off.

2. Use and store the meter in normal temperatures. Temperature


extremes can shorten the life of the electronic parts and distort
or melt plastic parts.

3. Handle the meter gently and carefully. Dropping it can damage


the electronic parts or the case.

4. Keep the meter clean. Wipe the case occasionally with a


damp cloth. Do not use chemicals, cleaning solvents or
detergents.

5. Use only fresh batteries or the recommended size and type.


Remove old or weak batteries so they do not leak and damage
the unit.

6. If the meter is to be stored for a long period of time, the


batteries should be removed to prevent damage to the unit.

-59-
8. Troubleshooting
There may be times when your meter does not operate properly.
Here are some common problems that you may have and some
easy solutions to them.

Meter does not operate:

1. Always read all the instructions in this manual before use.

2. Check to be sure the battery is properly installed.

3. Check to be sure the battery is good.

4. If the battery is good and the meter still doesn’t operate,


check to be sure that both ends of the fuse are properly
installed.

9. Replacing the battery


Warning:
To avoid electric shock, disconnect the test leads from any source
of voltage before removing the back cover or the battery/fuse door.

1. Disconnect the test leads from the meter.

2. Open the battery/fuse door by loosening one screw on the


battery/fuse door using a screw-driver.

3. Clip the new battery into battery holder, observing the correct
polarity.

4. Place the battery into the battery/fuse compartment.

5. Put the battery/fuse door back in place. Secure with the screw.

-60-
Warning:
To avoid electric shock, do not operate your meter until the back
cover and the battery / fuse door is in place and fastened securely.

Note:
If your meter does not work properly, check the fuses and batteries
to make sure that they are still good and that they are properly
inserted.

9.2. Statutory Notification about the Battery Regulations

The delivery of many devices includes batteries, which for example


serve to operate the remote control. There also could be batteries
or accumulators built into the device itself. In connection with the
sale of these batteries or accumulators, we are obliged under the
Battery Regulations to notify our customers of the following:

Please dispose of old batteries at a council collection point or return


them to a local shop at no cost. The disposal in domestic refuse is
strictly forbidden according to the Battery Regulations. You can
return used batteries obtained from us at no charge at the address
on the last side in this manual or by posting with sufficient stamps.

Batteries, which contain harmful substances, are


marked with the symbol of a crossed-out waste bin,
similar to the illustration shown left. Under the
waste bin symbol is the chemical symbol for the
harmful substance, e.g. „Cd“ for cadmium, „Pb“
stands for lead and „Hg“ for mercury.

You can obtain further information about the Battery Regulations


from the Bundesministerium für Umwelt, Naturschutz und
Reaktorsicherheit (Federal Ministry of Environment, Nature
Conservation and Reactor Safety).

-61-
All rights, also for translation, reprinting and copy of this manual or
parts are reserved.

Reproduction of all kinds (photocopy, microfilm or other) only by


written permission of the publisher.

This manual considers the latest technical knowing. Technical


changings which are in the interest of progress reserved.

We herewith confirm, that the units are calibrated by the factory


according to the specifications as per the technical specifications.
We recommend to calibrate the unit again, after 1 year.

© PeakTech® 01/2013/Ho/Pt.

-62-

You might also like