400amp True RMS AC/DC Clamp Meter: Model EX613
400amp True RMS AC/DC Clamp Meter: Model EX613
400amp True RMS AC/DC Clamp Meter: Model EX613
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user
must refer to the manual for further information.
Double insulation
2 EX613-EU-EN-V2.4-08/10
SAFETY NOTES
• Do not exceed the maximum allowable input range of any function.
• Do not apply voltage to meter when resistance function is selected.
• Set the function switch OFF when the meter is not in use.
• Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS
• Set function switch to the appropriate position before measuring.
• When measuring volts do not switch to current/resistance modes.
• Do not measure current on a circuit whose voltage exceeds 600V.
• When changing ranges always disconnect the test leads from the circuit under test.
CAUTIONS
• Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user manual before operating the meter.
• Always remove the test leads before replacing the battery or fuses.
• Inspect the condition of the test leads and the meter itself for any damage before
operating the meter. Repair or replace any damage before use.
• Use great care when making measurements if the voltages are greater than 25VAC
rms or 35VDC. These voltages are considered a shock hazard.
• Always discharge capacitors and remove power from the device under test before
performing Diode, Resistance or Continuity tests.
• Voltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of connection to the recessed electrical contacts. Other means should be
used to ensure that the terminals are not "live".
• If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
3 EX613-EU-EN-V2.4-08/10
Description
Meter Description
1. Current clamp
2. Clamp opening trigger
3. HOLD/BACKLIGHT button
4. MODE / ZERO button
5. Backlit LCD Display
6. Type K input jacks
7. Multimeter input jacks
8. Function switch
9. PEAK button
10. MAX/MIN button
11. Range button
12. Thermocouple display button
13. Battery compartment (rear)
14. Non-Contact Voltage Detector
15. NCV LED indicator
16. Input shutter
4 EX613-EU-EN-V2.4-08/10
Operation
NOTES: Read and understand all Warning and Caution statements in this operation
manual prior to using this meter. Set the function select switch to the OFF position when
the meter is not in use.
Input Shutter
The Input Shutter inhibits simultaneous connection to the thermocouple jacks and the
DMM input jacks. This is a safety feature which prevents a potentially hazardous condition
from existing during high voltage measurements. Slide the shutter up for test lead
measurements or slide it down for thermocouple temperature measurements.
5 EX613-EU-EN-V2.4-08/10
AC/DC Current Measurements
6 EX613-EU-EN-V2.4-08/10
Resistance Measurements
Note: Remove power from the device under test before making resistance measurements
1. Slide the input shutter to the up position.
Continuity Test
1. Connect as described for resistance measurements.
3. Touch the test probe tips across the circuit or component under test.
Diode Test
1. Connect as described for resistance measurements
2. Press the MODE button to select diode test .
3. Touch the test probe tips to the diode or semiconductor junction under test. Note the
meter reading.
4. Reverse the test lead polarity by reversing the red and black leads. Note this reading.
5. The diode or junction can be evaluated as follows:
If one reading displays a value (typically 0.400V to 01.800V) and the other
reading displays OL, the diode is good.
If both readings display OL the device is open.
If both readings are very small or ‘0’, the device is shorted.
7 EX613-EU-EN-V2.4-08/10
Capacitance Measurements
WARNING: To avoid electric shock, discharge the capacitor before measuring.
3. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive jack.
4. Touch the black test probe tip to one side of the device.
Touch the red test probe tip to the other side of the device.
5. Read the capacitance value in the display.
3. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive Hz jack.
4. Touch the black test probe tip to one side of the device.
Touch the red test probe tip to the other side of the device.
5. Read the Frequency value on the upper large display.
Read the Duty Ratio on the lower small display.
6. Press the MODE button to display the Duty Ratio on the large
display.
8 EX613-EU-EN-V2.4-08/10
μA DC/AC Current Measurements
9 EX613-EU-EN-V2.4-08/10
Data Hold
To freeze the LCD reading, press the HOLD button. While data hold is active, the HOLD
icon appears on the LCD. Press the HOLD button again to return to normal operation.
MAX/MIN
1. Press the MAX/MIN button to activate the MAX/MIN recording mode. The display icon
"MAX" will appear. The meter will begins recording and displaying the maximum
value measured.
2. Press the MAX/MIN button and “MIN” will appear. The meter will display the minimum
value measured during the recording session.
3. Press the MAX/MIN button and “MAX MIN” will appear. The meter will display the
present reading, but will continue to update and store the max and min readings.
4. To exit MAX/MIN mode press and hold the MAX/MIN button for 2 seconds.
Peak Hold
When ACA or ACV is selected, pressing the PEAK button enables the peak capture circuit.
The meter will now capture and display the maximum and minimum peaks of the
waveform.
RANGE
In the Voltage, Resistance, Capacitance, Frequency or uA function the meter automatically
selects the best range for the measurements being made. For measurement situations
requiring that a range be manually selected, perform the following:
1. Press the RANGE button. The “AUTO” display icon will turn off.
2. Press the RANGE key to step through the available ranges. Observe the decimal
point and units displayed until the preferred range is located.
3. To exit the Manual Ranging mode and return to Autoranging, press and hold the
RANGE key for 2 seconds.
LCD Backlight
The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas.
Press and hold the HOLD/ button for 2 seconds to turn the backlight on. The backlight
will automatically turn off after 30 seconds.
10 EX613-EU-EN-V2.4-08/10
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the
test leads from the input terminals, and turn OFF the meter before opening the case. Do
not operate the meter with an open case.
Battery Replacement
1. Remove the Phillips head screw that secures the rear battery door
2. Open the battery compartment
3. Replace the 9V battery
4. Secure the battery compartment door
5.
You, as the end user, are legally bound (EU Battery ordinance) to return all used
batteries, disposal in the household garbage is prohibited! You can hand over
your used batteries / accumulators at collection points in your community or
wherever batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the
device at the end of its lifecycle
Fuse Replacement
1. Remove the battery
2. Remove the Phillips head screws (2) that secures the rear cover.
3. Replace the fuse with one of equal rating. (500mA, 660V fast blow [SIBA 70-180-40])
4. Replace the rear cover and battery
11 EX613-EU-EN-V2.4-08/10
Specifications
Accuracy
Function Range Resolution
(% of reading + digits)
AC Current 400.0 AAC 0.1A
50/60 Hz ±(1.5% + 5 digits)
True RMS 40.00 AAC 0.01A
DC Current 400.0 ADC 0.1A
±(1.5% + 5 digits)
40.00 ADC 0.01A
AC/DC μA 400.00μA 0.01μA DC: ±(1.0% + 2 digits)
Current
4000.0μA 0.1μA AC: ±(1.5% + 2 digits)
AC Voltage 400.0 mVAC 0.1mV ±(1.0% + 20 digits)
50/60 Hz
True RMS 4.000 VAC 0.001V
40.00 VAC 0.01V
±(2.0% + 5 digits)
400.0 VAC 0.1V
600 VAC 1V
DC Voltage 400.00 mVDC 0.01mV
4.0000 VDC 0.0001V
±(0.1% + 2 digits)
40.000 VDC 0.001V
400.00 VDC 0.01V
600.0 VDC 0.1V ±(1.0% + 2 digits)
Resistance 400.00Ω 0.01Ω ±(0.8% + 20 digits)
4.0000kΩ 0.0001kΩ
40.000kΩ 0.001kΩ
±(0.8% + 4 digits)
400.00kΩ 0.01kΩ
4.0000MΩ 0.0001MΩ
40.000MΩ 0.001MΩ ±(2.5% + 10 digits)
Capacitance 400.00nF 0.01nF ±(5.0% + 40 digits)
4000.0nF 0.1nF ±(3.0% + 10 digits)
400.00µF 0.01µF ±(3.5% + 10 digits)
4000.0µF 0.1µF
±(5.0% + 10 digits)
40.000mF 0.001mF
Frequency 400.00Hz 0.01Hz ±(1.0% + 3 digits)
(clamp)
Sensitivity: 5Arms minimum
12 EX613-EU-EN-V2.4-08/10
Accuracy
Function Range Resolution
(% of reading + digits)
Frequency 40.000Hz 0.001Hz ±(0.3% + 3 digits)
(test leads)
400.00Hz 0.01Hz
4000.0Hz 0.1Hz
40.000kHz 0.001kHz
±(0.3% + 2 digits)
400.00kHz 0.01kHz
4000.0kHz 0.1kHz
40.000MHz 0.001MHz
100.00MHz 0.01MHz Not specified
Sensitivity: 5 to 5kHz; 0.8Vrms min., 5kHz to 150kHz; 5Vrms min
Duty Cycle 0.5% to 99.0% 0.1% ±(1.2% + 2 digits)
Pulse width: 100μs to 100ms, Frequency: 5Hz to 150kHz
Temperature -58 to -4°F
±13°F/7°C
Type K -50 to -19°C
-4 to 31°F
±(1.0% + 2°F/1°C)
-20 to -1°C
32°F
±2°F/1°C
0°C 0.1° <1000°
33 to 211°F 1° >1000°
±(1.0% + 2°F/1°C)
1 to 100°C
212 to 718°F
±(1.5% + 3°F/2°C)
101 to 399°C
719 to 1832°F
±(2.5% + 7°F/4°C)
400 to 1000°C
Specification does not include probe accuracy
13 EX613-EU-EN-V2.4-08/10
General Specifications
Clamp jaw opening 32mm (1.25") approx.
Display Dual 40,000/4,000 count backlit LCD
Non-Contact Voltage 100 to 600VAC
Continuity check Threshold 50Ω; Test current < 0.5mA
Diode test Test current of 0.3mA typical;
Open circuit voltage [ 2.8VDC typical
Low Battery indication Battery symbol is displayed
Over-range indication ‘OL’ display
Measurement rate 2 readings per second, nominal
Peak detector >1ms
Thermocouple sensor Type K thermocouple required
Fuse 500mA, ceramic fast blow
Input Impedance 10MΩ (VDC and VAC)
AC bandwidth 50 to 400Hz (AAC and VAC)
AC response True rms (AAC and VAC)
Crest Factor 3.0 in 40A and 400A ranges, 1.4 in 1000A range (50/60Hz
and 5% to 100% of range)
Operating Temperature 5°C to 40°C (41°F to 104°F)
Storage Temperature -20°C to 60°C (-4°F to 140°F)
Operating Humidity Max 80% up to 31°C (87°F) decreasing linearly to 50% at
40°C (104°F)
Storage Humidity <80%
Operating Altitude 2000 meters (7000ft) maximum.
Battery One (1) 9V Battery (NEDA 1604)
Auto power OFF After approx. 30 minutes, with disable
Dimensions & Weight 241x96x44.5mm (9.5x3.8x1.75”); 386g (13.6 oz)
Safety For indoor use and in accordance with the requirements for
double insulation to IEC1010-1 (2001): EN61010-1 (2001)
Overvoltage Category III 600V and Category II 1000V,
Pollution Degree 2.
Approvals CE,
14 EX613-EU-EN-V2.4-08/10