Extech 380926 Manual
Extech 380926 Manual
Extech 380926 Manual
SAFETY
Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to
the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present
Double insulation
WARNING: This indicates that a potentially hazardous condition which, if not avoided, could result
in death or serious injury.
CAUTION: This indicates that a potentially hazardous condition which, if not avoided, could result
in injury or damage to the meter Reply to request.
Safety Precautions
1. WARNING: Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user’s manual before operating the meter.
2. Make sure any covers or battery doors are properly closed and secured.
3. Always remove the test leads before replacing the battery or fuses.
4. Inspect the condition of the test leads and the meter itself for any damage before operating the
meter. Repair or replace any damage before use.
5. Do not exceed the maximum rated input limits.
6. Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.
7. Always discharge capacitors and remove power from the device under test before performing
Capacitance, Diode, Resistance or Continuity tests.
8. Remove the battery from the meter if the meter is to be stored for long periods.
2 380926 -EN-V7.1-3/12
METER DESCRIPTION
Front panel
1. Current Sense Jaws
2. Function switch
3. Keypad
4. LCD display
5. Input jacks
Symbols
AC current or Voltage
DC current or Voltage
Continuity
H Display hold
REL Relative
AUTO Auto range
Diode
Display Backlight
Units
mV, V millivolt, volt (voltage)
Ω, kΩ, MΩ ohm, kilohm, megohm (resistance)
% percent (duty cycle)
μA, mA, A microamps, milliamps, amps (current)
Hz, kHz hertz, kilohertz (frequency)
nF, μF nanofarad, microfarad (capacitance)
3 380926 -EN-V7.1-3/12
OPERATING INSTRUCTIONS
Current Measurements using the Clamp, DC/AC
WARNING: Make sure that the test leads are disconnected from the meter’s terminals before
making current measurements with the clamp jaw.
WARNING: For safety, use the rubber input terminal cover when measuring current using the
Clamp.
1. Set the Function switch to 2000A position.
2. Press the FUNC key to select AC or DC current
3. For DC current measurements, push & hold
the DCA ZERO button until the reading
indicates zero.
4. Press the trigger to open jaw. Fully enclose
the conductor to be measured (see
diagram).
5. The clamp meter will automatically select
the proper range (Auto Range). To select
the range manually, press the RANGE
button before pressing DCA ZERO button.
6. Read the measured value from the LCD
display. INCORRECT
More than one CORRECT
conductor in jaws One conductor in jaws
NOTE: If the DCA zero button is used to zero the
meter, the clamp meter remains in the auto-range mode. If the REL button is used, the
clamp meter will be placed in the manual range mode.
µA and mA
Current
4 380926 -EN-V7.1-3/12
Voltage Measurements, DC/AC
1. Set the Function Switch to the V position.
2. Press the FUNC key to select AC or DC.
3. Insert the black test lead to the COM input jack
and the red test lead into the V input jack.
4. Connect the test leads in PARALLEL with the
circuit to be measured (see diagrams).
5. Read the measured value from the LCD display.
Resistance Measurements
CAUTION: Before taking any in-circuit resistance measurements remove
power from the circuit under test and discharge all capacitors.
1. Set the Function switch to the Ω position.
2. Press the FUNC key until the ohms symbol appears on LCD.
3. Insert the black test lead to the COM input jack and the red test lead to
the Ω input jack.
4. Connect test leads to the device under test (see diagram).
5. Read the measured value from the LCD display.
Continuity Test
1. Set the Function switch to the position.
2. Press the FUNC key until the "Ω" and " " symbols appear on the display.
3. Insert the black test lead to the COM input jack and the red test lead to the Ω input jack.
4. Connect the test lead tips to the device to be measured (refer to diagram for resistance
measurements above).
5. Read the measured value from the LCD display.
6. If the resistance is < 10Ω approx. an audible signal will be heard.
5 380926 -EN-V7.1-3/12
Diode Test
1. Set the Function switch to the diode position.
2. Press the FUNC key until the symbol appears on the LCD.
3. Insert the black test lead to the COM input jack and the red test lead to the input jack.
4. Connect the test lead tips to the diode. A typical diode forward voltage drop will indicate 0.3
to 0.7V. The reverse drop will indicate "OL", indicating high impedance. Proper diode check
should include both forward and reverse tests.
RED BLACK BLACK RED
Frequency Measurement
1. Set the Function switch to the Hz position.
2. Press the FUNC key until the "Hz" symbol appears on the LCD.
3. Insert the black test lead to the COM input jack and the red test lead to the
Hz input jack.
4. Connect the tips of the test leads to the device to be measured (see
diagram).
5. Read the frequency value in Hz on the LCD.
Capacitance Measurement
1. Set the Function switch to the position.
2. Press the FUNC key until the "nF" symbol appears on the LCD.
3. Insert the black test lead to the COM input jack and the red test lead to the
red input terminal.
4. Connect the tips of the test leads to the device to be measured (see
diagram).
5. Read the capacitance value on the display.
NOTE: When making very low capacitance measurements, use the REL function
to zero any stray capacitance.
6 380926 -EN-V7.1-3/12
Automatic / Manual Range Selections
The meter defaults to the autoranging mode when turned on. "AUTO" will appear in the display.
To select manual ranging, press the RANGE key. Momentary presses of the RANGE key will
step through the ranges. To return to the Auto Range mode, press and hold the RANGE key for 2
seconds.
Relative Reading Measurements
Press the REL key to enter the relative mode. "REL" will appear in the display. In the relative
mode, the meter stores the reading that was on the display at the time of the REL key is pressed
and displays the difference between the measured value and the stored value
Press the REL key to return to normal mode.
Data Hold / Backlight key
Press the HOLD key momentarily to freeze the present reading on the LCD."H" will appear in the
display. Press HOLD again to return to normal operation.
Backlight
Press and hold the Hold key for 2 seconds to activate the backlighting. The backlight will
automatically turn off to conserve energy after approx 12 seconds.
MAINTENANCE
Battery Replacement
When the low battery symbol appears on the LCD, replace the meter’s 9V battery.
1. Remove power to the meter and remove test leads from meter
2. Remove the Phillips head screw (back of meter) and open the battery compartment.
3. Remove and replace the 9V battery.
4. Replace the compartment cover and rear screw.
You, as the end user, are legally bound (EU Battery ordinance) to return all used
batteries, disposal in the household garbage is prohibited! You can hand over your
used batteries / accumulators at collection points in your community or wherever
batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at
the end of its lifecycle
Fuse Replacement
NOTE: Fuse Rating: 500mA (5mm x 20mm diameter)
1. The meter is provided with one overload protect 500mA fuse for current measurements
(direct input NOT clamp measurements).
2. To replace the fuse, open the meter case by removing the battery cover and battery and then
the four screws holding the rear case.
3. The fuse is located on the Main PCB.
4. Replace fuse and secure meter case.
7 380926 -EN-V7.1-3/12
SPECIFICATIONS
Function Range Resolution Accuracy Remarks
AC/DC Current 400.0A 0.1A ±(2.0% + 5d) 45Hz to 1KHz
(Clamp on) 2000A 1A ±(2.0% + 8d)
AC Voltage 4.000V 1mV ±(1.2% + 5d) 45Hz to 1KHz
True RMS 40.00V 10mV Input Impedance:
400.0V 0.1V 10Mohms
1000V 1V
DC Voltage 400.0mV 0.1mV ±(0.5% + 2d) Input Impedance:
4.000V 1mV ±(1.0% + 2d) 10Mohms
40.00V 10mV
400.0V 0.1V
1000V 1V
Resistance 400.0 Ω 0.1Ω ±(1.0% + 5d)
4.000 kΩ 1Ω
40.00 kΩ 10Ω
400.0 kΩ 100Ω
4.000 MΩ 1kΩ ±(2.0% + 2d)
40.00 MΩ 10kΩ ±(3.5% + 5d)
AC/DC Current 400.0µA 0.1µA ±(1.2% + 5d) 45Hz to 1KHz
(Direct Input) 4000µA 1µA
40.00mA 0.01mA
400.0mA 0.1mA
Frequency 5Hz 0.001Hz ±(1% + 5d)
50Hz 0.01Hz
500Hz 0.1Hz
5kHz 1Hz
50kHz 10Hz
100kHz 100Hz
Capacitance 50nF 10pF ±(3% + 5d) Accuracy stated
500nF 100pF after a relative "REL"
5µF 0.001µF zero performed
50µF 0.01µF
Duty Cycle 1 to 99% 0.1% ±(1% + 5d)
Continuity Audible tone; <10 ohms approximately
Open circuit voltage; 0.5V approximately
8 380926 -EN-V7.1-3/12
Conductor Size 2.0”(50mm) maximum
Measurement parameters ACA, DCA, ACV, DCV, Resistance, Diode, Frequency,
Capacitance, Duty Cycle, Continuity
Current Sensor Hall Effect
Zero adjust Automatic except for DCA (Push-button)
Diode Test Test current of 0.6mA typical;
Open circuit voltage <1.6V DC typical.
Battery type 9V NEDA 1604
Range Selection Auto or manual
Display 0.6” (15mm) 5000 Count Backlit LCD
Overload Indication "OL"
Power Consumption 5mA approx.
Low Battery Indication Battery icon appears on LCD
Sampling rate 1 reading every 0.35 seconds approx.
Standards CE, IEC 1010-1 CAT III 1000V and CAT IV 600V
o o o
Operating Temperature/Humidity 32 to 122 F (0 C to 50 C) / <80%
Dimensions 10 x 2.9 x 1.5” (255 x 73 x 38mm)
Weight 0.85 lbs. (380g)
9 380926 -EN-V7.1-3/12
Warranty
EXTECH INSTRUMENTS CORPORATION (a FLIR company) warrants this instrument to be free of defects in parts and
workmanship for one year from date of shipment (a six month limited warranty applies to sensors and cables). If it should
become necessary to return the instrument for service during or beyond the warranty period, contact the Customer
Service Department for authorization. Visit our website www.extech.com for contact information. A Return Authorization
(RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges,
freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting
from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or
repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a
specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is
limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether
written or oral, is expressed or implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also provides NIST certification for most
products. Call the Customer Service Department for information on calibration services available for this product. Extech
recommends that annual calibrations be performed to verify meter performance and accuracy.
10 380926 -EN-V7.1-3/12