Handout METHOD VERIFICATION AND VALIDATIONS
Handout METHOD VERIFICATION AND VALIDATIONS
Handout METHOD VERIFICATION AND VALIDATIONS
AND VALIDATIONS
BY
PUAN SALMA ARIFIN
MS ISO/IEC 17025:2017
requirements
Verification and validation of methods
are stated in the clause 7: Process
Requirements
Specifically mentioned in clause 7.2
Selection, Verification and Validation of
methods
Method Verification
requirements:
The laboratory shall verify that it can properly
perform methods before introducing them by
ensuring that it can achieve the required
performance.
Method verification:
validity of test method, valid to use.
METHOD VERIFICATION
vi) Precision
Repeatability and reproducibility
Intra-laboratory
> Duplicate sample with 5 replicates per lot of sample
and per microbial level
> microbial level of spike (control, low, medium, high)
> calculate the standard deviation by logarithm the
results.
METHOD VALIDATION
Precision continue…
Inter-laboratory
> minimum 2 laboratories participate
> calculate the standard deviation, Sr values
> compare the values obtained between lab
False positive and false negative
sample that gives interfering substances( affect this
results) has to be removed.
METHOD VALIDATION
Sensitivity
Same as recovery and accuracy
Ruggedness Testing
>small changes in operating conditions and
environment, example : incubation temperature.
METHOD VALIDATION
Sample Matrixes
> ideally, 5 matrixes for each type of food category.
Level
➢ microbial level of spike : unspiked, low, middle, and high
spike
➢ analyzed at least 5 times per matrixes
Interfering Microorganism
➢ Related to natural contaminants : natural microflora.
➢ If any interfere happen, sample (matrixes) shall be
removed from list
➢ Use matrixes with high and low levels of incurred
microflora.
METHOD VALIDATION
Acceptance Criteria
➢ No specific acceptance criteria
➢ For Quantitative method, calculate the results from
validated method and compare with standard
method results or
➢ For Qualitative method, compared the results
obtained from validated method against to inoculum
of microbial suspension results.
METHOD VALIDATION
Quantitative
Procedure
➢ Prepare sample matrixes as per required for each type
of matrixes
➢ Prepare inoculum for spiking level (unspike (blank),
low(0.1ml), middle (0.5ml), high(1 ml))
➢ Obtained the initial count for each level of spike
➢ Spike the sample matrixes
➢ Each level = 5 times (5 replicates)
METHOD VALIDATION
Quantitative
Procedure cont…
> Proceed with the test by following the procedure of
the method to be validated.
➢ Count the results obtained
➢ Calculate the results in terms of comparing the results
with initial count in percentage term.
QUALITATIVE METHOD
VALIDATION
Accuracy – comparing results obtained from
validated method with standard method and
microbial detected is correctly identified as per test
method required.
Recovery – use middle or high spike for qualitative
method. Comparing by proportion of positive results.
Homogenity of the sample is crucial to obtained
results.
QUALITATIVE METHOD
VALIDATION
Limits of detection – theoretically, one microbe is to be
detected for particular method to be validated. Limit
sample is 25g.
Limits of quantitation and determination – one
analytical particle per chosen analytical sample size
QUALITATIVE METHOD
VALIDATION
Precision
Repeatability and reproducibility
Intra-laboratory
> Duplicate sample with 10 replicates per lot of sample
and per microbial level
> microbial level of spike (control, low, medium, high)
QUALITATIVE METHOD
VALIDATION
Inter laboratory reproducibility
> minimum 2 laboratories participate
> 10 replicates of sample matrixes with spike level of low,
middle and high.
Sensitivity
>proportion of test sample that contain microbial and
response positively to the test
Specificity
> Proportion of test sample that do not contain microbial
and response negatively to the test.
QUALITATIVE METHOD
VALIDATION
Sample matrixes
> one or more different samples
Microbial analyte
> Ideally a variety of strains of the target microorganisms
would be tested.
> If possible, use isolated microbial (natural contaminant)
for validation.
QUALITATIVE METHOD
VALIDATION
State of stress (concentration) of microbial analyte
(microbe)
➢ microbes may be subject in a particular matrix; dry
inocula are used for dry foods and wet inocula are
used for wet foods.
Spiking levels
> Each strain is to be used at a minimum of 3 levels per
analytical sample (25g/ml); 1-10cfu; 10-50cfu; blank; no
addition. Obtained the initial count of spike level.
QUALITATIVE METHOD
VALIDATION
Acceptance criteria
> depend on matrixes type
> inoculum level must be stated from < 10 to 100%
QUALITATIVE METHOD
VALIDATION
Procedure
➢ Prepare sample (10 replicates of different matrixes)
➢ Prepare inoculum of microbial and obtained initial
count
➢ Spike with microbial levels (low, middle, high)
➢ Proceed with the test to be validated
➢ Obtained the results
➢ Compare with the results by standard method
➢ Results as presence or abscence
METHOD VALIDATION
Report of method validation:
An official report need to be established
containing:
>Title
>Date of conducted and completed
>Name and signed of person
conducted the test
> Name and signed of person verified
the test
>procedure of verification
➢ Results (calculation)
➢ Discussion and conclusion
EXAMPLE OF METHOD VERIFICATION
Thank You