Castings, Carbon, Low-Alloy, and Martensitic Stainless Steel, Ultrasonic Examination Thereof
Castings, Carbon, Low-Alloy, and Martensitic Stainless Steel, Ultrasonic Examination Thereof
Castings, Carbon, Low-Alloy, and Martensitic Stainless Steel, Ultrasonic Examination Thereof
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1. Scope 1.6 This standard does not purport to address all of the
2
1.1 This practice covers the standards and procedures for safety concerns, if any, associated with its use. It is the
the pulse-echo ultrasonic examination of heat-treated carbon, responsibility of the user of this standard to establish appro-
low-alloy, and martensitic stainless steel castings. priate safety, health, and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
1.2 This practice is to be used whenever the inquiry, 1.7 This international standard was developed in accor-
contract, order, or specification states that castings are to be dance with internationally recognized principles on standard-
subjected to ultrasonic examination in accordance with Prac- ization established in the Decision on Principles for the
tice A609/A609M. Development of International Standards, Guides and Recom-
1.3 This practice contains two procedures. Procedure A is mendations issued by the World Trade Organization Technical
the original A609/A609M practice and requires calibration Barriers to Trade (TBT) Committee.
using a series of test blocks containing flat-bottomed holes. It
also provides supplementary requirements for angle beam 2. Referenced Documents
testing. Procedure B requires calibration using a back wall 2.1 ASTM Standards:3
reflection from a series of solid calibration blocks. A217/A217M Specification for Steel Castings, Martensitic
NOTE 1—Ultrasonic examination and radiography are not directly Stainless and Alloy, for Pressure-Containing Parts, Suit-
comparable. This examination technique is intended to complement Guide able for High-Temperature Service
E94/E94M in the detection of discontinuities. E94/E94M Guide for Radiographic Examination Using In-
1.4 Supplementary requirements of an optional nature are dustrial Radiographic Film
provided for use at the option of the purchaser. The supple- E317 Practice for Evaluating Performance Characteristics of
mentary requirements shall apply only when specified indi- Ultrasonic Pulse-Echo Testing Instruments and Systems
vidually by the purchaser in the purchase order or contract. without the Use of Electronic Measurement Instruments
1.5 The values stated in either SI units or inch-pound units 2.2 Other Document:4
are to be regarded separately as standard. The values stated in SNT-TC-1A Recommended Practice for Nondestructive
each system may not be exact equivalents; therefore, each Testing Personnel Qualification and Certification
system shall be used independently of the other. Combining
values from the two systems may result in nonconformance 3. Ordering Information
with the standard. 3.1 The inquiry and order should specify which procedure is
1.5.1 Within the text, the SI units are shown in brackets. to be used. If a procedure is not specified, Procedure A shall be
1.5.2 This practice is expressed in both inch-pound units used.
and SI units; however, unless the purchase order or contract
specifies the applicable M-specification designation (SI units), 3.2 The purchaser shall furnish the following information:
the inch-pound units shall apply. 3.2.1 Quality levels for the entire casting or portions
thereof,
1
This practice is under the jurisdiction of ASTM Committee A01 on Steel,
Stainless Steel and Related Alloys and is the direct responsibility of Subcommittee
3
A01.18 on Castings. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Nov. 1, 2023. Published November 2023. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1970. Last previous edition approved in 2018 as A609/A609M – 12 Standards volume information, refer to the standard’s Document Summary page on
(2018). DOI: 10.1520/A0609_A0609M-12R23. the ASTM website.
2 4
For ASME Boiler and Pressure Vessel Code applications, see related Specifi- Available from American Society for Nondestructive Testing (ASNT), P.O. Box
cation SA-609 of Section II of that Code. 28518, 1711 Arlingate Ln., Columbus, OH 43228-0518, http://www.asnt.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
4. Apparatus
4.1 Electronic Apparatus:
4.1.1 An ultrasonic, pulsed, reflection type of instrument
that is capable of generating, receiving, and amplifying fre-
quencies of at least 0.5 to 5 MHz.
4.1.2 The ultrasonic instrument shall provide linear presen-
tation (within 65 %) for at least 75 % of the screen height
(sweep line to top of screen). Linearity shall be determined in NOTE 1—Opposite ends of reference block shall be flat and parallel
accordance with Practice E317 or equivalent electronic means. within 0.001 in. [0.025 mm].
4.1.3 The electronic apparatus shall contain a signal attenu- NOTE 2—Bottom of flat-bottom hole shall be flat within 0.002 in.
[0.051 mm] and the finished diameter shall be 1⁄4 + 0.002 in. [6.4 + 0.050].
ator or calibrated gain control that shall be accurate over its
NOTE 3—Hole shall be straight and perpendicular to entry surface
useful range to 610 % of the nominal attenuation or gain ratio within 0°, 30 min and located within 1⁄32 in. [0.80 mm] of longitudinal
to allow measurement of signals beyond the linear range of the axis.
instrument. NOTE 4—Counter bore shall be 1⁄2 in. [15.0 mm] diameter by 1⁄8 in.
[5 mm] deep.
4.2 Search Units:
4.2.1 Longitudinal Wave, internally grounded, having a 1⁄2 FIG. 1 Ultrasonic Standard Reference Block
to 1 in. [13 to 25 mm] diameter or 1 in. [25 mm] square
piezo-electric elements. Based on the signals-to-noise ratio of TABLE 1 Dimensions and Identification of Reference Blocks in
the response pattern of the casting, a frequency in the range the Basic Set (See Fig. 1)
from 0.5 to 5 MHz shall be used. The background noise shall Metal Overall Width or Block
Hole Diameter
not exceed 25 % of the distance amplitude correction curve in 1⁄64 ths, in.
Distance Length Diameter Identifi-
(DAC). Transducers shall be utilized at their rated frequencies. [mm] (B), in.A (C), in. (D), min, cation
[mm] [mm] in. [mm] Number
4.2.2 Dual Element, 5 MHz, 1⁄2 by 1 in. [13 by 25 mm], 12° 16 [6.4] 1 [25] 13⁄4 [45] 2 [50] 16-0100
included angle search units are recommended for sections 1 in. 16 [6.4] 2 [50] 23⁄4 [70] 2 [50] 16-0200
[25 mm] and under. 16 [6.4] 3 [75] 33⁄4 [95] 2 [50] 16-0300
16 [6.4] 6 [150] 63⁄4 [170] 3 [75] 16-0600
4.2.3 Other frequencies and sizes of search units may be 16 [6.4] 10 [255] 103⁄4 [275] 4 [100] 16-1000
used for evaluating and pinpointing indications. 16 [6.4] B B + 3⁄4 [B + 20] 5 [125] 16-B00B
A
Tolerance ±1⁄8 in. [3 mm].
4.3 Reference Blocks: B
Additional supplemental blocks for testing thickness greater than 10 in.
4.3.1 Reference blocks containing flat-bottom holes shall be [250 mm], see 4.3.3.
used to establish test sensitivity in accordance with 8.2.
4.3.2 Reference blocks shall be made from cast steels that
give an acoustic response similar to the castings being exam-
ined. 4.3.5 Each reference block shall be permanently identified
along the side of the block indicating the material and the block
4.3.3 The design of reference blocks shall be in accordance
identification.
with Fig. 1, and the basic set shall consist of those blocks listed
in Table 1. When section thicknesses over 15 in. [380 mm] are 4.4 Couplant—A suitable couplant having good wetting
to be inspected, an additional block of the maximum test characteristics shall be used between the search unit and
thickness shall be made to supplement the basic set. examination surface. The same couplant shall be used for
4.3.4 Machined blocks with 3⁄32 in. [2.4 mm] diameter calibrations and examinations.
flat-bottom holes at depths from the entry surface of 1⁄8 in.
[3 mm], 1⁄2 in. [13 mm], or 1⁄2 t and 3⁄4 in. [19 mm], or 3⁄4 t 5. Personnel Requirements
(where t = thickness of the block) shall be used to establish the 5.1 Personnel performing ultrasonic examination in accor-
DAC for the dual element search units (see Fig. 2). dance with this practice shall be qualified and certified in
accordance with a written procedure conforming to Recom- markers, note the flat-bottom hole indication height for each of
mended Practice No. SNT-TC-1A or another national standard the applicable blocks on the instrument screen. Draw a curve
acceptable to both the purchaser and the supplier. through these marks on the screen or on suitable graph paper.
The maximum signal amplitude for the test blocks used shall
6. Casting Conditions peak at approximately three fourths of the screen height above
6.1 Castings shall receive at least an austenitizing heat the sweep by use of the attenuator. This curve shall be referred
treatment before being ultrasonically examined. to as the 100 % distance amplitude correction (DAC) curve. If
6.2 Test surfaces of castings shall be free of material that the attenuation of ultrasound in the casting thickness being
will interfere with the ultrasonic examination. They may be as examined is such that the system’s dynamic range is exceeded,
cast, blasted, ground, or machined. segmented DAC curves are permitted.
6.3 The ultrasonic examination shall be conducted prior to 8.3 The casting examination surface will normally be
machining that prevents an effective examination of the cast- rougher than that of the test blocks; consequently, employ a
ing. transfer mechanism to provide approximate compensation. In
order to accomplish this, first select a region of the casting that
7. Test Conditions has parallel walls and a surface condition representative of the
7.1 To ensure complete coverage of the specified casting rest of the casting as a transfer point. Next, select the test block
section, each pass of the search unit shall overlap by at least whose overall length, C (Fig. 1), most closely matches the
10 % of the width of the transducer. reflection amplitude through the block length. Place the search
unit on the casting at the transfer point and adjust the
7.2 The rate of scanning shall not exceed 6 in./s [150 mm/s]. instrument gain until the back reflection amplitude through the
7.3 The ultrasonic beam shall be introduced perpendicular casting matches that through the test block. Using this transfer
to the examination surface. technique, the examination sensitivity in the casting may be
expected to be within 630 % or less of that given by the test
8. Procedure blocks.
8.1 Adjust the instrument controls to position the first back 8.4 Do not change those instrument controls and the test
reflection for the thickness to be tested at least one half of the frequency set during calibration, except the attenuator or
distance across the instrument screen. calibrated gain control, during acceptance examination of a
8.2 Using the set of reference blocks spanning the thickness given thickness of the casting. Make a periodic calibration
of the casting being inspected and overlays or electronic during the inspection by checking the amplitude of response
10.1 This practice is intended for application to castings PROCEDURE B – BACK WALL REFLECTION
with a wide variety of sizes, shapes, compositions, melting CALIBRATION PROCEDURE
processes, foundry practices, and applications. Therefore, it is
impractical to specify an ultrasonic quality level that would be 11. Apparatus
universally applicable to such a diversity of products. Ultra- 11.1 Apparatus shall be kept on a regular six-month main-
sonic acceptance or rejection criteria for individual castings tenance cycle during which, as a minimum requirement, the
should be based on a realistic appraisal of service requirements vertical and horizontal linearities, sensitivity, and resolution
and the quality that can normally be obtained in production of shall be established in accordance with the requirements of
the particular type of casting. Practice E317.
10.2 Acceptance quality levels shall be established between 11.2 Search Units—Ceramic element transducers not ex-
the purchaser and the manufacturer on the basis of one or more ceeding 1.25 in. [32 mm] diameter or 1 in.2 [645 mm2] shall be
of the following criteria: used.
10.2.1 No indication equal to or greater than the DAC over 11.3 Search Units Facing—A soft urethane membrane or
an area specified for the applicable quality level of Table 2. neoprene sheet, approximately 0.025 in. [0.64 mm] thick, may
10.2.2 No reduction of back reflection of 75 % or greater be used to improve coupling and minimize transducer wear
that has been determined to be caused by a discontinuity over caused by casting surface roughness.
an area specified for the applicable quality level of Table 2.
10.2.3 Indications producing a continuous response equal to 11.4 Calibration/Testing—The same system, including the
or greater than the DAC with a dimension exceeding the urethane membrane, used for calibration shall be used to
maximum length shown for the applicable quality level shall be inspect the casting.
unacceptable. 11.5 Other Inspections—Other frequencies and type search
10.2.4 Other criteria agreed upon between the purchaser and units may be used for obtaining additional information and
the manufacturer. pinpointing of individual indications.
10.3 Other means may be used to establish the validity of a 11.6 Couplant—A suitable liquid couplant, such as clean
rejection based on ultrasonic inspection. SAE 30 motor oil or similar commercial ultrasonic couplant,
The following supplementary requirement shall be applied only when agreed upon between the
purchaser and the supplier to achieve an effective examination of a critical casting area that cannot be
effectively examined using a longitudinal beam as a result of casting design or possible discontinuity
orientation.
S1. Angle Beam Examination of Steel Castings S1.2.1 Construct the distance amplitude correction curve by
S1.1 Equipment: utilizing the responses from the side-drilled holes in the basic
S1.1.1 Search Units—Angle beam search units shall pro- calibration block for angle beam examination as shown in Fig.
duce an angle beam in steel in the range from 30 to 75° S1.1 and Table S1.1.
inclusive, measured to the perpendicular of the entry surface of S1.2.1.1 Resolve and mark the amplitudes of the 1⁄4 t and
1⁄2 t side-drilled holes from the same surface. The side-drilled
the casting being examined. Search units shall have a fre-
quency of 0.5 to 5 MHz. hole used for the 1⁄4 t amplitude may be used to establish the
3⁄4 t amplitude from the opposite surface or a separate hole may
S1.1.2 Calibration Blocks—A set of blocks, as shown in
Fig. S1.1, with as-cast surface equivalent to SCRATA Com- be used.
parator A35 and of a thickness comparable to the sections being S1.2.1.2 Connect the 1⁄4 t, 1⁄2 t, and 3⁄4 t amplitudes to
examined with side-drilled holes at 1⁄4 t, 1⁄2 t, and 3⁄4 t (where establish the applicable DAC.
t = thickness of the block) shall be used to establish an S1.2.2 The basic calibration blocks shall be made of mate-
amplitude reference line (ARL). rial that is acoustically similar to the casting being examined.
S1.2 Calibration of Equipment: S1.2.3 Do not use basic calibration blocks with as-cast
surface equivalent to SCRATA Comparator A3 to examine
5
Available from Steel Founders’ Society of America (SFSA), 780 McArdle Dr., castings with surface rougher than SCRATA Comparator A3.
Unit G, Crystal Lake, IL 60014, http://www.sfsa.org. Use a machined calibration block for machined surfaces.
L = length of block determined by the angle of search unit and the vee-path used.
T = thickness of basic calibration block (see Table S1.1).
D = depth of side-drilled hole (see Table S1.1).
d = diameter of side-drilled hole (see Table S1.1).
t = nominal production material thickness.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/