Rev A - Thermo Scientific Digital Dry Bath With Advanced Microprocessor Control - Operation Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 21

Thermo Scientific Digital Drybath

Operating Manual
0000848 • Revision A • October 2021
IMPORTANT! Before using this product, read this entire operation manual carefully. Users should follow all of
the operational guidelines contained in this manual and take all necessary safety precautions while using this
product. Failure to follow these guidelines could result in potentially irreparable bodily harm and/or property
damage.

Caution all internal adjustments and maintenance must be performed by qualified service personnel.

Material in this manual is for information purposes only. Thermo Fisher Scientific is committed to a continuing
program of product development and improvement, and reserves the right to change information, such as
specifications, appearance, and dimensions, described in this document without notice. Thermo Fisher
Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo Fisher
Scientific be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.

No part of this manual may be reproduced or transmitted in any form or by any means, including
photocopying, recording, or using information storage and retrieval systems, for any purpose other than the
purchaser's own use, without the express written permission of the manufacturer.

Any other product names and services identified in this manual are trademarks or registered trademarks of
their respective owners. No such use, or the use of any trade name, is intended to convey endorsement or
other affiliation with Thermo Fisher Scientific.

©2021 Thermo Fisher Scientific. All rights reserved.

All rights reserved. Without prior written consent from Thermo Fisher Scientific, any part of the manual should
not be duplicated, reproduced or translated into any other languages.

Caution: All rights reserved. Thermo Fisher Scientific reserves the right to modify this manual at any time
without notice.

Before the initial use of this instrument, please carefully read this manual.
Contents
Models......................................................................... 1

Safety Information ........................................................ 2

Introduction ................................................................. 4

Features....................................................................... 5
Normal Working Conditions...................................... 5
Transportation and Storage Conditions .................... 5
Basic Parameters ..................................................... 5
Regulatory Compliance ............................................ 6

Preparation Work......................................................... 8
Structure Diagram .................................................... 8
Display Panel Diagram ............................................. 8
Instructions for Buttons ............................................ 8

Operation Guidance..................................................... 9
Examination Before Startup...................................... 9
Startup..................................................................... 9
Setting Temperature and Heating Time .................... 9
Start and Stop Instrument ........................................ 10
Temperature Measurement and Calibration.............. 10

Maintenance ................................................................ 11

Failure Analysis and Troubleshooting ........................... 12

Modular Block Accessories.......................................... 13

Warranty ...................................................................... 15

Frequently Asked Questions ........................................ 16


Models
This manual covers the models shown below.

Table 1. Models

NA Model EU Model Voltage Description

88870001 N/A 100-120 Vac TS Drybath Stdrd 1 blck 100-120 V

88870002 N/A 100-120 Vac TS Drybath Stdrd 2 blck 100-120 V

88870003 N/A 100-120 Vac TS Drybath Stdrd 4 blck 100-120 V

N/A 88870004 200-240 Vac TS Drybath Stdrd 1 blck 200-240 V

N/A 88870005 200-240 Vac TS Drybath Stdrd 2 blck 200-240 V

N/A 88870006 200-240 Vac TS Drybath Stdrd 4 blck 200-240 V

Digital Drybath Models | 1


Safety Information
Alert Signals • The power cord

CAUTION: Important information is contained in The instrument should use the power line cord
any item and should be carefully read. Failure to attached to it. If the power line is destroyed, it
observe the instruction would result in damage or must be replaced but not be repaired. The
abnormal function of the instrument. replacement should be carried out with the power
line of the same type and same specification
(reach out to technical service team for further
WARNING: The warning message requires information). No items should be put on the power
extremely careful operation of a certain step. cord when the instrument is in operation.
Failure to observe the instruction may result in
• Connecting and disconnecting to/from
serious personal injury.
power source
The user should hold the plug to remove from
During operation, maintenance and repair of this instrument,
power source. When connecting the plug, user
the following basic safety notes should be observed. In case
should make sure it has been fully plugged in;
of failure to follow these instructions, the warnings or notes
when removing the plug, do not pull the power line
indicated herein, the basic protection provided by the
forcefully. Do not position the equipment so that it
instrument, its safety criteria of design and manufacture, and
is difficult to operate the disconnecting device
its predicted use range would be impaired. If the equipment is
(plug).
used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired. • Placement of the instrument
This instrument should be fixed in a low
CAUTION: This instrument is for indoor use. RH (≤ 80%) and low dust place away from
• The ground connection water source (e.g. sink or water pipeline) and
the room should be well ventilated, and free of
To avoid the electric shock, the input power corrosive gas or interference of strong
line must be reliably grounded. The instrument magnetic field. The instrument should not be
is equipped with the three-pin plug that has placed in a wet or dusty location.
the third pin (the pin connects the ground),
therefore, the plug should be used with the The openings on this instrument are for
grounded power socket only. This is a kind of ventilation circulation and in order to avoid
safety device. If the plug cannot be plugged over-heat of the instrument body they should
into the power socket, please ask the not be blocked or covered. When a single set
electrician to install a correct power socket, so of instrument is used, the interval between
as to make the grounded plug to work for ventilation opening before and after the
safety. instrument and its nearest object should not
be less than 25 cm. Also, don’t use the
Thermo Fisher Scientific is not responsible for instrument on loose or soft surface, or the air
any injury as a result of the user’s failure to inlet of instrument bottom might be blocked.
observe the following requirements.
• Observe voltage/power requirements
Before the instrument is connected to the AC
power source, the voltage of the power
source should be the same with the required
voltage of the instrument (a deviation of ±10%
is allowed). The rated load of the power
socket should not be lower than the
requirements of the instrument.

2 | Safety Information Digital Drybath


Excessive temperature will impair
performance or result in failure of the
instrument. This instrument should not be
used in locations subjected to direct sun light.
The instrument should be kept away from hot
gas, oven and all other heat sources. If the
instrument is to be stored for a long time, the
power plug should be withdrawn and the
instrument covered with soft cloth or plastic
film to avoid entrance of dust. The product is
powered by connecting the mains plug to a
standard socket-outlet. Always place the
product in such a way that it is easy for the
operator to disconnect the product from the
mains supply.
• Explanation of symbols
Attention, read user manual before use.

WARNING: There is a sign of


“CAUTION: HOT SURFACE!
ATTENTION: SURFACE HOT” on the
instrument. The metal part (module)
near the sign should not be touched
with any part of the body when the
instrument is operating in a high
temperature state or just finished
operation to avoid burns!

Alternating Current

Protective Conductor Terminal

CAUTION: In any of the following cases,


immediately turn off the power supply, withdraw
the power plug from the power socket, contact the
supplier for service:
• Liquid drops into the inside of the instrument.
• The instrument is rained on or water is spilled
on it.
• The instrument works abnormally, especially
if generating an abnormal sound or odor.
• The instrument is dropped or its casing is
damaged.
• The function of instrument obviously
changes.

Digital Drybath Safety Information | 3


Introduction
The product is a dry bath with advanced microcomputer
control, which can be widely applied to sample reservation,
enzyme reservation and reaction, DNA amplification,
electrophoresis degeneration and serum coagulation, etc.
The characteristics of the product are as follows:
• The digital display for control of the temperature.
• Heat blocks help avoid possible contamination while
containing the sample inside a tube.
• The heat blocks are easy to replace, clean and disinfect
and are suitable for various tubes sizes.
• Built-in over-temperature protection device warrants
sample and user safety.
• Temperature deviation adjustment.

4 | Introduction Digital Drybath


Features
This chapter introduces the usage, transportation, storage CAUTION: Before using the instrument, please
conditions of the instrument, as well as its basic parameters, make sure the working condition meets the above
performance and functions. requirements. Especially that the power line is
reliably grounded.

Normal Working
Conditions Transportation and
Storage Conditions
Ambient temperature 10°C ~ 30°C Ambient temperature: -20°C~+55°C
Relative humidity ≤ 80% Relative humidity: ≤ 80%
Power supply 100-120 V ~ 50/60 Hz or
200-240 V ~ 50/60 Hz
Altitude Up to 2000 m
Pollution Degree 2
Overvoltage category II Indoor use
Mains supply voltage ±10%
fluctuations

Basic Parameters
Table 2. 100-120 V

NA Catalog Number 88870001 88870002 8887000303

Digital 1 block Digital 2 block Digital 4 block


Model Name
100-120 V 100-120 V 100-120 V

Temperature Range Ambient Temperature +5°C ~ 130°C (Ambient Temperature 25°C)

Temperature Uniformity ±1.0°C@37°C

Temperature Accuracy ±0.5°C@37°C

Temperature Fluctuation ±0.5°C

≤ 25 min (rise from 30°C to


Temperature Rise Time ≤ 20 min (rise from 30°C to 130°C)
130°C)

288 × 200 × 100 mm 318 × 200 × 100 mm 450 × 200 × 100 mm


Size (L × W × H)
11.3 × 7.87 × 3.94 in 12.5 × 7.87 × 3.94 in 17.7 × 7.87 × 3.94 in

Weight 2.9 kg (6.39 lbs) 3.3 kg (7.28 lbs) 4.7 kg (10.36 lbs)

Fuse Protector (ø5×20) F 250 V 2.5 A F 250 V 5 A F 250 V 8 A

Digital Drybath Features | 5


Table 3. 200-240 V

EU Catalog
88870004 88870005 88870006
Number

Digital 1 block Digital 2 block Digital 4 block


Model Name
200-240 V 200-240 V 200-240 V

Temperature Range Ambient Temperature +5°C ~ 130°C (Ambient Temperature 25°C)

Temperature
±1.0°C
Uniformity

Temperature Accuracy ±0.5°C@37°C

Temperature
±0.5°C
Fluctuation

Temperature Rise ≤ 25 min (rise from 30°C to


≤ 20 min (rise from 30°C to 130°C)
Time 130°C)

288 × 200 × 100 mm 318 × 200 × 100 mm 450 × 200 × 100 mm


Size (L × W × H)
11.3 × 7.87 × 3.94 in 12.5 × 7.87 × 3.94 in 17.7 × 7.87 × 3.94 in

Weight 2.9 kg (6.39 lbs) 3.3 kg (7.28 lbs) 4.7 kg (10.36 lbs)

Fuse Protector (ø5×20) F 250 V 2.5 A F 250 V 2.5 A F 250 V 3.15 A

CAUTION: The instrument is only used for heating


operation: from low temperature to high
temperature.

Regulatory Compliance CE Safety EN 61010-1 CE


Safety EN 61010-2-010

European Union UL 61010-1/CSAC22.2 NO.61010-1


ROHS2.0 2011/65/EU
The European voltage models of this product
meet all the applicable requirements of the
European Directives and therefore display the CE
Marking. These Directives are captured in the EU
Korean Registration
Declaration of Conformity which may be obtained
from the manufacturer. 사용자안내문
이기기는엄무용환경에서사용할
목적으로적함성평가를받은기기
로서가정용환경에서사용하는경우전파

Product Safety 간섬의우려가있습니다

This product family has been tested to applicable product WARNING: EMC Registration is done on this
safety standards by a Nationally Recognized Test Laboratory equipment for business use only. It may cause
(NRTL) and may bear the NRTL’s mark of safety compliance interference when the product would be used
to those applicable standards. in home. This warning statement applies a
product for business use.
CEEMC EN 61326-1

6 | Features Digital Drybath


Electromagnetic Compatibility • Have a certified technician remove the Refractory
Ceramic insulation from the unit then dispose properly.
FCC Statement (USA)

This device complies with Part 15 Subpart B of


the FCC Rules. Operation is subject to the
following two conditions:
(1) this device may not cause harmful
interference, and
(2) this device must accept any interference
received, including interference that may cause
undesired operation.
Any changes or modifications not expressly
approved by the manufacturer could void the
user’s authority to operate the equipment.

Note: This equipment has been tested and found to comply


with the limits for a Class A digital device, pursuant to part 15
of the FCC Rules. These limits are designed to provide rea-
sonable protection against harmful interference when the
equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own
expense.
Canadian ISED IC Notice
• This ISM digital apparatus complies with Canadian ICES-
001, Class A
• Cet appareil ISM est conforme á la norme NMB-001 du
Canada, Classe A
End of Life Care
Some considerations and suggestions are listed below for
proper disposal of this product. While addressing these
actions for safe recycling and disposal, please follow all
guidelines, Safety Data Sheets (SDS), or regulations
applicable to your country and region.
• This product has materials and components that may be
recycled or reused according to local guidelines and
regulations.
• Remove any batteries present before disposal. Batteries,
battery packs, and accumulators should not be disposed
of as unsorted household waste. Please use the public
collection system to return, recycle, or treat them in
compliance with the local regulations.
• Remove all samples and items before defrosting a unit to
room ambient temperatures.
• Clean up any chemical or biological safety hazards using
appropriate methods.

Digital Drybath Features | 7


Preparation Work
This chapter introduces the structure of the instrument, user
interface and functions of all buttons and preparation before Instructions for Buttons
startup. Read the content in the chapter carefully before using
the instrument for the first time. Setting or adjusting button: Press the button
to display present setting temperature and
press the button again to display present time.
Structure Diagram After the unit has reached set point and
temperature is stable, calibration can be set
Housing by pressing the button.

Under the status of setting, press “ ” or “ ”


Front Cover to modify the flashing figure. The instrument
will enter temperature calibration mode if user
Display press the two buttons simultaneously under
Window the status of stopping.

When setting or adjusting, pressing the button


can move the flashing figure. During the
Figure 1. Structure Diagram operation, the display can be switched
between the temperature and the time.

Product Label For starting up or stopping program.

Power Socket,
Fuse Housing,
Power Switch
Block Lifter - Use
to lift blocks
Figure 2. Back Unit

Display Panel Diagram


Heating Indicator Light
Warning Indicator
Light

Digital Display

Figure 3. Display Panel

8 | Preparation Work Digital Drybath


Operation Guidance
This chapter introduces the displaying window and the
operation of buttons of the dry bath with constant Setting Temperature and
temperature.
Heating Time
Note: The illumination pattern of the ‘Heat Light’ indicates the
Operation status. Set Temperature and Hold-Time Setpoint:

• Heat Light OFF (Unit Stopped) = “Set/Cal” button used 1. Connect power cord to appropriate electrical supply.
for Setpoint adjustment. 2. Make Power switch, located near power cord on rear, to
• Heat Light blinks/flashes as Block is heating or cooling the ‘ON’ position.
toward setpoint. 3. Confirm the Heat-indicator is ‘off’. If 'ON' or 'blinking’,
• Heat Light ON (Unit Started and stable at setpoint) = press “Stop”.
“Set/Cal” used for Calibration. 4. Set-up Temperature:
a. Press the “SET/CAL” button. The instrument is in the
Temperature Setpoint mode with the display window
Examination Before showing the previous temperature set value, and the left

Startup digit flashing.


b. Press “ ” to increase or “ ” to decrease the flashing
Please confirm the following content before plugging the plug digit of the Setpoint.
into the power socket.
c. Press “SHIFT” to move the flashing right, to the next digit.
1. The power source is in accordance with the voltage
requirement of the instrument (please refer to the d. Repeat Steps 4b and 4c to complete Temperature
chapter II for the requirement of the voltage). setpoint selection.

2. Make sure the plug has been fully plugged into the power 5. Setting Hold-Time Setpoint:
socket. a. After the temperature is selected, press “SET/CAL”
3. The power line is reliably grounded. button. The instrument is in the Hold-time Setpoint mode
with the display window showing the previous Hold-time
set value, and the left digit flashing.
CAUTION: If the display of the instrument is
abnormal after startup, please turn off the power b. Press “ ” to increase or “ ” to decrease the flashing
source immediately and contact with the supplier. digit of the Hold-Time. Choose a time from 00:01 to
99:59, hours and minutes.

c. Press “SHIFT” to move the flashing right, to the next digit.

Startup d. Repeat Steps 5b and 5c to complete Hold-Time setpoint


selection. For Continuous operation, Choose 99:59 then
increase to display “- -:- -”.
Press the power switch and the instrument will power on. All
indicator lights and display serial ports will be turned on. e. After the Time setting is complete, Wait for 5 seconds
About 3 ~ 5 seconds later, temperature display window will and the instrument will automatically exit the Setpoint
show the real-time temperature. All indicator lights will be mode. The display will now show the current measured
turned off. temperature.

6. Press “Start” button to begin heating the Block. The


green “Heat Light” will blink as Block is heating toward
set point.
7. Allow 30-minutes for temperature to stabilize. The heat
indicator stops blinking and remains constantly “ON” to
indicate the temperature is stable, and instrument is
ready to use.

Digital Drybath Operation Guidance | 9


Instruction: There is limitation for temperature and time 7. The Heat-indicator may start 'blinking' as the
setting. The setting temperature range is: 15 ~ 135°C; the temperature is stabilized to the calibrated temperature.
time range is: 0 ~ 99 h 59 min and continuous mode "- -:- -".
8. When the Heat-indicator is 'ON' steady, the temperature
is stable.

Start and Stop Instrument 9. After the instrument stable operated for additional
10 minutes, compare the display temperature and
measurement again.
START: Press “START/STOP” to start operation, the indicator If the temperature gap is > 0.2°C, repeat all steps of 6.
light “HEAT” will be flashing. When the instrument achieves
the setting temperature, the “HEAT” light will stop flashing.
This indicates that the unit will enter hold temperature status. CAUTION:
The digital display window will display the real-time • To ensure the accuracy of temperature, the
temperature and the timing function by pressing “SHIFT” instrument shall be calibrated after
button. When the holding time equals to the setting time, the temperature stabilizes for 30 minutes.
instrument will alarm 5 beeps.
• Calibrate the instrument with qualified Grade
STOP: Under the working status, press “START/STOP” to II standard filled thermometer.
stop working. At that time, corresponding working indicator
light “HEAT” will be turned off. The digital display window • Calibration point: Fill the middle hole of the
displays the real-time temperature. heat block with paraffin oil and soak the
thermometer bulb.

Temperature Measurement
and Calibration
1. The instrument has been operational for minimum of
30-minutes at desired setpoint.
2. An independent temperature measurement device is in
the designed 5 mm hole in Block, along with paraffin oil,
or similar heat-transfer oil.
3. Confirm the Heat-indicator is ‘ON’ steady, not ‘blinking’.
4. Compare the Display and independent measurements.
And check if the temperature gap is > 0.2°C or not.
5. If less than 0.2°C, no calibration should be made.
6. If more than 0.2°C, calibrate the Temperature by these
steps:
a. Press and hold the “SET/CAL” button. The instrument is
in the Temperature Calibration mode with the display
window showing the measured temperature value, and
the left digit flashing.
b. Press “ ” to increase, or “ ” to decrease, the flashing
digit to match the same digit of the independent
measurement.

c. Press “SHIFT” to move the flashing right, to the next digit.

d. Repeat Steps 6b and 6c to complete Temperature


caibration.

e. After the temperature calibration is completed, press


“SET/CAL” button to exit the calibration mode and
continue the Start mode.

10 | Operation Guidance Digital Drybath


Maintenance
The holes of the blocks should be regularly cleaned with the
damp cloth to ensure the test tube be well contacted with the
wall of the holes so as to have good heat conduction. If the
surface of the instrument is polluted, it can be cleaned with a
slightly damp soft cloth. Generally unit falls under protective
system of IP20 pursuant to IEC 60529.

WARNING: When cleaning the instrument, the


power supply should be shut off and unplugged.
The instrument surface should be cleaned with a
non-corrosive cleaning agent.

Digital Drybath Maintenance | 11


Failure Analysis and Troubleshooting
This chaper introduces possible failures of this instrument,
explanation and troubleshooting.

CAUTION: The user is not allowed to open the


housing of the instrument for inspection during the
warranty period. If any failures that require opening
the housing for inspection occur, the supplier or
manufacturer should be contacted timely.

Table 4. Failure Analysis and Troubleshooting

No. Error Cause Solution

Power supply fails to be Check power supply and


connected connect it.

Display window is not turned on Fuse burned out Replace fuse


1
after switching on power supply Damage of the switch Replace the switch.

Contact Thermo Fisher


Others
Scientific.

Temperature display window Check if the connecting line


2 shows "001" and the buzzer Short circuit of sensor of sensor is damaged, and
alarm in a di...di...sound replace it.

Temperature display window Check if the connecting line


3 shows "002" and the buzzer Open circuit of sensor of sensor is damaged, and
alarm in a di...di...sound replace it.

Temperature display does not Sensor is damaged or in poor Contact Thermo Fisher
4
conform to actual temperature contact Scientific.

12 | Failure Analysis and Troubleshooting Digital Drybath


Modular Block Accessories
Table 5. Modular Block Accessories

Block Dimensions
Catalog No. Heating Blocks No. of Holes
W x D x H (mm/in)

124 x 76 x 39 /
88870101 For 6 mm test tubes 46
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870102 For 10 mm test tubes 28
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870103 For 1.5 ml test tubes 28
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870104 For 2.0 ml test tubes 28
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870105 For 12 mm & 13 mm test tubes 24
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870106 For 15 mm & 16 mm test tubes 15
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870107 For 17 mm & 18 mm test tubes 12
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870108 For 20 mm test tubes 8
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870109 For 25 mm test tubes 6
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870110 For 0.5 ml test tubes 40
4.9 x 3.0 x 1.5

124 x 76 x 39 /
88870111 For 96 well Elisa plate 1
4.9 x 3.0 x 1.5

For mixed size test tubes 32 holes for 6 mm 124 x 76 x 39 /


88870112 32 + 21
test tubes 21 holes for 10 mm test tubes 4.9 x 3.0 x 1.5

For mixed size test tubes 18 holes for 1.5 ml 124 x 76 x 39 /


88870113 18 +10
test tubes 10 holes for 2.0 ml test tubes 4.9 x 3.0 x 1.5

For mixed size test tubes 3 holes for 25 mm


124 x 76 x 39 /
88870114 test tubes 12 holes for 13 mm test tubes 3+12+6
4.9 x 3.0 x 1.5
6 holes for 6 mm test tubes

Digital Drybath Modular Block Accessories | 13


Table 5. Modular Block Accessories

Block Dimensions
Catalog No. Heating Blocks No. of Holes
W x D x H (mm/in)

For mixed size test tubes 30 holes for 0.5 ml 124 x 76 x 39 /


88870115 30+20
test tubes 20 holes for 0.2 ml test tubes 4.9 x 3.0 x 1.5

124 x 76 x 113 /
88870116 For 15 ml flat bottom test tubes 15
4.9 x 3.0 x 4.55

124 x 76 x 113 /
88870117 For 50 ml flat bottom test tubes 4
4.9 x 3.0 x 4.55

124 x 76 x 113 /
88870118 For 15 ml conical bottom test tubes 15
4.9 x 3.0 x 4.55

124 x 76 x 113 /
88870119 For 50 ml conical bottom test tubes 4
4.9 x 3.0 x 4.55

124 x 76 x 39 /
88870120 For 96 well Non-skirted PCR plate 1
4.9 x 3.0 x 1.5

124 x 76 x 55 /
88870121 For 96 well half/full skirted PCR plate 1
4.9 x 3.0 x 2.17

88870122 Temperature probe, PT1000, stainless steel NA NA

* Dry bath blocks can be autoclaved.

14 | Modular Block Accessories Digital Drybath


Warranty
When used in laboratory conditions and according to these
operation instructions and maintenance, this product is
warranted for 24 months against defective materials or
workmanship. The 24 month warranty period begins from the
delivery date of this product.
For product quality or performance issues, contact Thermo
Fisher Scientific Customer Service.

Digital Drybath Warranty | 15


Frequently Asked Questions
Scenario 1
Does unit turn off once the program time is complete:
Example- If unit is set for one program to run at 45°C for
2 hrs.
Q-Would the unit turn off upon completion of this 2 hrs?
A-The instrument doesn't turn off by itself, it just stops
running.
Q-If not- would the unit hold the set temperature beyond this
2 hrs?
A-After this 2 hrs, the unit doesn't hold the set temperature. It
will cool to ambient temp.
Q-Would unit return to ambient temp?
A-The unit will return to ambient temp.

Scenario 2
Suppose the unit is running a set program and there is power
outage. Once the power resumes-
Q-Would unit restart automatically once power is restored
A-The instrument will automatically turn on.
Q-Would unit resume the program from where it was left?
A-It can't.
Q-Would customer need to start the program manually?
A-Yes.
Q-Would the unit keep the saved program in memory?
A-The unit has no memory function.
Q-Would customer need to create new program?
A-Need to reset the program.

16 | Frequently Asked Questions Digital Drybath


WEEE Compliance
WEEE Compliance. This product is required to comply with the European Union’s Waste Electrical &
Great Britain
Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked with the following symbol. Thermo Fisher
Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and
this product should be disposed of or recycled through them. Further information on our compliance with
these Directives, the recyclers in your country, and information on Thermo Fisher Scientific products which
may assist the detection of substances subject to the RoHS Directive are available at thermofisher.com/
WEEERoHS.rrrrrrrrrt

WEEE Konformittät. Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE)
Deutschland
Richtlinie 2012/19/EU erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet. Thermo Fisher
Scientific hat Vereinbarungen getroffen mit Verwertungs-/Entsorgungsanlagen in allen EU-
Mitgliederstaaten und dieses Produkt muss durch diese Firmen widerverwetet oder entsorgt werden.
Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo Fisher Scientific, dieVerwerter
und Hinweise die Ihnen nützlich sein können, dsss3Axdie Thermo Fisher Scientific Produkte zu
identizfizieren, die unter diese RoHS. Anweisungfallen, finden Sie unter thermofisher.com/WEEERoHS.

Conformità WEEE. Questo prodotto deve rispondere alla direttiva dell’ Unione Europea 2012/19/EU in
Italia
merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici (WEEE). È marcato col seguente simbolo.Thermo
Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio/smaltimento in ognuno degli
Stati Membri Europei. Questo prodotto verrà smaltito o riciclato tramite queste medesime. Ulteriori
informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive, l’elenco delle ditte di
riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Fisher Scientific che possono essere utili
alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito thermofisher.com/
WEEERoHS.

Conformité WEEE. Ce produit doit être conforme à la directive euro-péenne (2012/19/EU) des Déchets
France
d’Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant. Thermo Fisher
Scientific s’est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de
l’union européenne et ce produit devraitêtre collecté ou recyclé par celles-ci. Davantage d’informations sur
laconformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les
informations sur les produits Thermo Fisher Scientific qui peuvent aider le détection des substances
sujettes à la directive RoHS sont disponibles sur thermofisher.com/WEEERoHS.
North America’s Technical Support Team
Phone: 1-866-984-3766, Option#2
Service Email: Servicesupport.led.asheville@thermofisher.com
Presales Email: LED.Presales@thermofisher.com
Webpage: www.unitylabservices.com

Europe Middle East Africa Technical Support Team


German national toll free: 0800 1 112 110, Fax: 0800 1 112 114
German international: +49 6103 408 0, Fax: +49 6103 408 12 13
Email: unity.de@thermofisher.com and let.support@thermofisher.com

Singapore Technical Support Team


Service Hotmail is Service.LPG.ROA1@thermofisher.com

Thermo Fisher Scientific Inc.


22 Alpha Rd,
Chelmsford, MA
01824, USA

Find out more at thermofisher.com/drybaths


Australia: (613) 9757-4300 In Australia: (1300) 735-295
China: (86) 21-6865-4588 Germany: (49) 6184-90-6321
India: (91) 22-4157-8800 Japan: (81) 045-453-9175
North America: 1-978-232-6000 Toll Free: 1-800-225-1480
Singapore: (65) 6778-6876

© 2021 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its
subsidiaries unless otherwise specified. 1021