Rev A - Thermo Scientific Digital Dry Bath With Advanced Microprocessor Control - Operation Manual
Rev A - Thermo Scientific Digital Dry Bath With Advanced Microprocessor Control - Operation Manual
Rev A - Thermo Scientific Digital Dry Bath With Advanced Microprocessor Control - Operation Manual
Operating Manual
0000848 • Revision A • October 2021
IMPORTANT! Before using this product, read this entire operation manual carefully. Users should follow all of
the operational guidelines contained in this manual and take all necessary safety precautions while using this
product. Failure to follow these guidelines could result in potentially irreparable bodily harm and/or property
damage.
Caution all internal adjustments and maintenance must be performed by qualified service personnel.
Material in this manual is for information purposes only. Thermo Fisher Scientific is committed to a continuing
program of product development and improvement, and reserves the right to change information, such as
specifications, appearance, and dimensions, described in this document without notice. Thermo Fisher
Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo Fisher
Scientific be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.
No part of this manual may be reproduced or transmitted in any form or by any means, including
photocopying, recording, or using information storage and retrieval systems, for any purpose other than the
purchaser's own use, without the express written permission of the manufacturer.
Any other product names and services identified in this manual are trademarks or registered trademarks of
their respective owners. No such use, or the use of any trade name, is intended to convey endorsement or
other affiliation with Thermo Fisher Scientific.
All rights reserved. Without prior written consent from Thermo Fisher Scientific, any part of the manual should
not be duplicated, reproduced or translated into any other languages.
Caution: All rights reserved. Thermo Fisher Scientific reserves the right to modify this manual at any time
without notice.
Before the initial use of this instrument, please carefully read this manual.
Contents
Models......................................................................... 1
Introduction ................................................................. 4
Features....................................................................... 5
Normal Working Conditions...................................... 5
Transportation and Storage Conditions .................... 5
Basic Parameters ..................................................... 5
Regulatory Compliance ............................................ 6
Preparation Work......................................................... 8
Structure Diagram .................................................... 8
Display Panel Diagram ............................................. 8
Instructions for Buttons ............................................ 8
Operation Guidance..................................................... 9
Examination Before Startup...................................... 9
Startup..................................................................... 9
Setting Temperature and Heating Time .................... 9
Start and Stop Instrument ........................................ 10
Temperature Measurement and Calibration.............. 10
Maintenance ................................................................ 11
Warranty ...................................................................... 15
Table 1. Models
CAUTION: Important information is contained in The instrument should use the power line cord
any item and should be carefully read. Failure to attached to it. If the power line is destroyed, it
observe the instruction would result in damage or must be replaced but not be repaired. The
abnormal function of the instrument. replacement should be carried out with the power
line of the same type and same specification
(reach out to technical service team for further
WARNING: The warning message requires information). No items should be put on the power
extremely careful operation of a certain step. cord when the instrument is in operation.
Failure to observe the instruction may result in
• Connecting and disconnecting to/from
serious personal injury.
power source
The user should hold the plug to remove from
During operation, maintenance and repair of this instrument,
power source. When connecting the plug, user
the following basic safety notes should be observed. In case
should make sure it has been fully plugged in;
of failure to follow these instructions, the warnings or notes
when removing the plug, do not pull the power line
indicated herein, the basic protection provided by the
forcefully. Do not position the equipment so that it
instrument, its safety criteria of design and manufacture, and
is difficult to operate the disconnecting device
its predicted use range would be impaired. If the equipment is
(plug).
used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired. • Placement of the instrument
This instrument should be fixed in a low
CAUTION: This instrument is for indoor use. RH (≤ 80%) and low dust place away from
• The ground connection water source (e.g. sink or water pipeline) and
the room should be well ventilated, and free of
To avoid the electric shock, the input power corrosive gas or interference of strong
line must be reliably grounded. The instrument magnetic field. The instrument should not be
is equipped with the three-pin plug that has placed in a wet or dusty location.
the third pin (the pin connects the ground),
therefore, the plug should be used with the The openings on this instrument are for
grounded power socket only. This is a kind of ventilation circulation and in order to avoid
safety device. If the plug cannot be plugged over-heat of the instrument body they should
into the power socket, please ask the not be blocked or covered. When a single set
electrician to install a correct power socket, so of instrument is used, the interval between
as to make the grounded plug to work for ventilation opening before and after the
safety. instrument and its nearest object should not
be less than 25 cm. Also, don’t use the
Thermo Fisher Scientific is not responsible for instrument on loose or soft surface, or the air
any injury as a result of the user’s failure to inlet of instrument bottom might be blocked.
observe the following requirements.
• Observe voltage/power requirements
Before the instrument is connected to the AC
power source, the voltage of the power
source should be the same with the required
voltage of the instrument (a deviation of ±10%
is allowed). The rated load of the power
socket should not be lower than the
requirements of the instrument.
Alternating Current
Normal Working
Conditions Transportation and
Storage Conditions
Ambient temperature 10°C ~ 30°C Ambient temperature: -20°C~+55°C
Relative humidity ≤ 80% Relative humidity: ≤ 80%
Power supply 100-120 V ~ 50/60 Hz or
200-240 V ~ 50/60 Hz
Altitude Up to 2000 m
Pollution Degree 2
Overvoltage category II Indoor use
Mains supply voltage ±10%
fluctuations
Basic Parameters
Table 2. 100-120 V
Weight 2.9 kg (6.39 lbs) 3.3 kg (7.28 lbs) 4.7 kg (10.36 lbs)
EU Catalog
88870004 88870005 88870006
Number
Temperature
±1.0°C
Uniformity
Temperature
±0.5°C
Fluctuation
Weight 2.9 kg (6.39 lbs) 3.3 kg (7.28 lbs) 4.7 kg (10.36 lbs)
This product family has been tested to applicable product WARNING: EMC Registration is done on this
safety standards by a Nationally Recognized Test Laboratory equipment for business use only. It may cause
(NRTL) and may bear the NRTL’s mark of safety compliance interference when the product would be used
to those applicable standards. in home. This warning statement applies a
product for business use.
CEEMC EN 61326-1
Power Socket,
Fuse Housing,
Power Switch
Block Lifter - Use
to lift blocks
Figure 2. Back Unit
Digital Display
• Heat Light OFF (Unit Stopped) = “Set/Cal” button used 1. Connect power cord to appropriate electrical supply.
for Setpoint adjustment. 2. Make Power switch, located near power cord on rear, to
• Heat Light blinks/flashes as Block is heating or cooling the ‘ON’ position.
toward setpoint. 3. Confirm the Heat-indicator is ‘off’. If 'ON' or 'blinking’,
• Heat Light ON (Unit Started and stable at setpoint) = press “Stop”.
“Set/Cal” used for Calibration. 4. Set-up Temperature:
a. Press the “SET/CAL” button. The instrument is in the
Temperature Setpoint mode with the display window
Examination Before showing the previous temperature set value, and the left
2. Make sure the plug has been fully plugged into the power 5. Setting Hold-Time Setpoint:
socket. a. After the temperature is selected, press “SET/CAL”
3. The power line is reliably grounded. button. The instrument is in the Hold-time Setpoint mode
with the display window showing the previous Hold-time
set value, and the left digit flashing.
CAUTION: If the display of the instrument is
abnormal after startup, please turn off the power b. Press “ ” to increase or “ ” to decrease the flashing
source immediately and contact with the supplier. digit of the Hold-Time. Choose a time from 00:01 to
99:59, hours and minutes.
Start and Stop Instrument 9. After the instrument stable operated for additional
10 minutes, compare the display temperature and
measurement again.
START: Press “START/STOP” to start operation, the indicator If the temperature gap is > 0.2°C, repeat all steps of 6.
light “HEAT” will be flashing. When the instrument achieves
the setting temperature, the “HEAT” light will stop flashing.
This indicates that the unit will enter hold temperature status. CAUTION:
The digital display window will display the real-time • To ensure the accuracy of temperature, the
temperature and the timing function by pressing “SHIFT” instrument shall be calibrated after
button. When the holding time equals to the setting time, the temperature stabilizes for 30 minutes.
instrument will alarm 5 beeps.
• Calibrate the instrument with qualified Grade
STOP: Under the working status, press “START/STOP” to II standard filled thermometer.
stop working. At that time, corresponding working indicator
light “HEAT” will be turned off. The digital display window • Calibration point: Fill the middle hole of the
displays the real-time temperature. heat block with paraffin oil and soak the
thermometer bulb.
Temperature Measurement
and Calibration
1. The instrument has been operational for minimum of
30-minutes at desired setpoint.
2. An independent temperature measurement device is in
the designed 5 mm hole in Block, along with paraffin oil,
or similar heat-transfer oil.
3. Confirm the Heat-indicator is ‘ON’ steady, not ‘blinking’.
4. Compare the Display and independent measurements.
And check if the temperature gap is > 0.2°C or not.
5. If less than 0.2°C, no calibration should be made.
6. If more than 0.2°C, calibrate the Temperature by these
steps:
a. Press and hold the “SET/CAL” button. The instrument is
in the Temperature Calibration mode with the display
window showing the measured temperature value, and
the left digit flashing.
b. Press “ ” to increase, or “ ” to decrease, the flashing
digit to match the same digit of the independent
measurement.
Temperature display does not Sensor is damaged or in poor Contact Thermo Fisher
4
conform to actual temperature contact Scientific.
Block Dimensions
Catalog No. Heating Blocks No. of Holes
W x D x H (mm/in)
124 x 76 x 39 /
88870101 For 6 mm test tubes 46
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870102 For 10 mm test tubes 28
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870103 For 1.5 ml test tubes 28
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870104 For 2.0 ml test tubes 28
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870105 For 12 mm & 13 mm test tubes 24
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870106 For 15 mm & 16 mm test tubes 15
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870107 For 17 mm & 18 mm test tubes 12
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870108 For 20 mm test tubes 8
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870109 For 25 mm test tubes 6
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870110 For 0.5 ml test tubes 40
4.9 x 3.0 x 1.5
124 x 76 x 39 /
88870111 For 96 well Elisa plate 1
4.9 x 3.0 x 1.5
Block Dimensions
Catalog No. Heating Blocks No. of Holes
W x D x H (mm/in)
124 x 76 x 113 /
88870116 For 15 ml flat bottom test tubes 15
4.9 x 3.0 x 4.55
124 x 76 x 113 /
88870117 For 50 ml flat bottom test tubes 4
4.9 x 3.0 x 4.55
124 x 76 x 113 /
88870118 For 15 ml conical bottom test tubes 15
4.9 x 3.0 x 4.55
124 x 76 x 113 /
88870119 For 50 ml conical bottom test tubes 4
4.9 x 3.0 x 4.55
124 x 76 x 39 /
88870120 For 96 well Non-skirted PCR plate 1
4.9 x 3.0 x 1.5
124 x 76 x 55 /
88870121 For 96 well half/full skirted PCR plate 1
4.9 x 3.0 x 2.17
Scenario 2
Suppose the unit is running a set program and there is power
outage. Once the power resumes-
Q-Would unit restart automatically once power is restored
A-The instrument will automatically turn on.
Q-Would unit resume the program from where it was left?
A-It can't.
Q-Would customer need to start the program manually?
A-Yes.
Q-Would the unit keep the saved program in memory?
A-The unit has no memory function.
Q-Would customer need to create new program?
A-Need to reset the program.
WEEE Konformittät. Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE)
Deutschland
Richtlinie 2012/19/EU erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet. Thermo Fisher
Scientific hat Vereinbarungen getroffen mit Verwertungs-/Entsorgungsanlagen in allen EU-
Mitgliederstaaten und dieses Produkt muss durch diese Firmen widerverwetet oder entsorgt werden.
Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo Fisher Scientific, dieVerwerter
und Hinweise die Ihnen nützlich sein können, dsss3Axdie Thermo Fisher Scientific Produkte zu
identizfizieren, die unter diese RoHS. Anweisungfallen, finden Sie unter thermofisher.com/WEEERoHS.
Conformità WEEE. Questo prodotto deve rispondere alla direttiva dell’ Unione Europea 2012/19/EU in
Italia
merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici (WEEE). È marcato col seguente simbolo.Thermo
Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio/smaltimento in ognuno degli
Stati Membri Europei. Questo prodotto verrà smaltito o riciclato tramite queste medesime. Ulteriori
informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive, l’elenco delle ditte di
riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Fisher Scientific che possono essere utili
alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito thermofisher.com/
WEEERoHS.
Conformité WEEE. Ce produit doit être conforme à la directive euro-péenne (2012/19/EU) des Déchets
France
d’Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant. Thermo Fisher
Scientific s’est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de
l’union européenne et ce produit devraitêtre collecté ou recyclé par celles-ci. Davantage d’informations sur
laconformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les
informations sur les produits Thermo Fisher Scientific qui peuvent aider le détection des substances
sujettes à la directive RoHS sont disponibles sur thermofisher.com/WEEERoHS.
North America’s Technical Support Team
Phone: 1-866-984-3766, Option#2
Service Email: Servicesupport.led.asheville@thermofisher.com
Presales Email: LED.Presales@thermofisher.com
Webpage: www.unitylabservices.com
© 2021 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its
subsidiaries unless otherwise specified. 1021