Druck PT Data Sheet

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Data Pack E

Issued September 2000 232-5244

Data Sheet

Pressure transducers and instrumentation

Two ranges of silicon diaphragm gauge pressure transducers, one with a built-in reference tube and one with an integral ventilated cable. The reference tube type allows transducers to be remotely mounted via a connecting pressure tube to the active face of the diaphragm. A second pressure connection must be made from the ambient air pressure in the region of the fluid pressure being measured, to the reference tube (and hence the non-active face of the diaphragm) on the transducer. This allows a true gauge reading to be taken (ie. the pressure above the local ambient of the fluid being measured) by the remote transducer. The integral cable range has the reference, or non-active face of the transducer, vented to the atmosphere via a small pipe inside the cable and should be used when the transducer is mounted directly in contact with the fluid pressure being measured. Both ranges feature silicon diaphragms with a bridge of semiconductor strain gauges integrally diffused which is bonded onto a glass carrier and housed in a titanium body. Titanium is used because of its similar thermal characteristics to silicon and hence, reduced unwanted temperature effects. The transducers can be used with any fluid which will not attack titanium, silicon dioxide or glass. The reference tube, or cable ventilation, can only be connected to a dry, non-corrosive non-conducting gas such as dry air.

Features
q Excellent linearity and hysteresis q High overload capacity q Range of 14 transducers; 7 with integral cable and boot and 7 with PTFE cable and reference tube q 7 operating pressure ranges from 1 to 60bar q Good thermal stability; 0.5% total error band over full temperature range of 0-50C

Dimensions
11 20 28

21 18AF

Electrical connections 6 core shielded/vented cable Supply positive Red Supply negative White Output positive Ye l l o w Blue Output negative N/C to transducer body Screen Any other cores not connected

Integral cable type

18AF

11

44

14

1.6 dia stainless steel reference tube

Reference tube type

Electrical connections 4 core ptfe shielded cable Red Supply positive Blue Supply negative Ye l l o w Output positive Green Output negative Screen N/C to transducer body

232-5244
Technical specification
Max. overpressure _______________4 times rated pressure Max. pressure on reference connection (PDCR 820 types) _____________________2 bar Negative pressure __________________________to vacuum Burst pressure ________________> 10 times rated pressure Excitation voltage __________________________10Vdc 5mA Output voltage ______________________100mV (for 10V in) Common mode voltage________________+ 6.5V (typ) w.r.t. -Ve supply for 10V in Output impedance ___________________________2K nom. Minimum load impedance ___100k for rated performance Resolution _____________________________________Infinite Combined non-linearity, hysteresis and repeatability_________________0.1% B.S.L. (best straight line) Zero offset and span setting _________________3mV max. Operating temp. range __________________-20C to +80C Temperature effects _______0.5% total error 0C to +50C Electrical connections __________________1m integral lead 6 core 810 series 4 core 820 series Pressure connections _________1/4 in BSP 60 internal core

The RS range of pressure transducers


PDCR 810 series - ventilated cable
Pressure range 0 to 1bar 0 to 2bar 0 to 3.5bar 0 to 7bar 0 to 15bar 0 to 35bar 0 to 60bar RS stock no. 646-628 646-634 646-640 646-656 646-662 646-678 646-684

PDCR 820 series - reference tube


Pressure range 0 to 1bar 0 to 2bar 0 to 3.5bar 0 to 7bar 0 to 15bar 0 to 35bar 0 to 60bar RS stock no. 646-690 646-707 646-713 646-729 646-735 646-741 646-757

Panel mounting digital pressure indicator DPI 260 (RS stock no. 646-763)
The digital pressure indicator will measure and indicate pressure to an accuracy of 0.1% full scale. The indicator utilises the PDCR 810 or 820 series pressure transducers. The compact, self contained, unit incorporates an extremely stable polarising supply, signal conditioning and amplifier system which does not degrade the transducer specification. The indicator is housed in a rugged ABS case designed to fit a DIN (92 x 45mm) panel cut-out.

Features
q q q q q q Excellent linearity and hysteresis 7.5mm seven segment display Front panel zero and span adjustment Self contained strain gauge indicator Field ranging/scaling by internal switches Analogue output: 0-2V.

Figure 1 Block diagram

REMOTE TRANSDUCER PRESSURE SOURCE L.E.D. DISPLAY AMPLIFIER SCALING UNIT A.D.C.

19.999

PSI

POWER SUPPLIES

LINEARISATION UNIT

BUFFERED ANALOGUE OUTPUT


-15V

-15V-15V 240V 50HZ

POWER SUPPLIES

TRANSFORMER

232-5244
Specification
Readout
19999.

Long term stability


Zero offset <0.02% F.S. per year. Sensitivity <0.02% F.S. per year.

Display
7.5mm seven segment L.E.D.

Power supplies
110V or 240V, 50-400Hz at 3VA max. Link selectable.

Display overload
Settable anywhere between 1999 and 19999 to give 1 and flashing 0000.

Dimensions
96mm (wide) x 48mm (high) x 175mm (deep).

Resolution
0.005% F.S. maximum.

Weight
500gms. nominal.

Response
333msec to full scale on digital display and BCD output.

Analogue output specification


Output Zero rationalisation Bandwidth In-band noise Zero offset temperature coefficient Zero suppression temperature coefficient reading/C. 0-2V <0.3% F.S. 2kHz <0.02% F.S. pk-pk <0.001% F.S./C. <0.005% of

Position effect
Negligible.

Excitation voltages
5 or 10 Volts into 350 ohm bridge minimum.

Input signal range


10 to 200mV nominal for 19999 maximum display.

Zero suppression/elevation
Capable of 19999 switch settable internally.

Electrical connection
Connector supplied with remote transducer and 1 metre of cable. Pin 1 Supply positive Pin 2 Output positive Pin 3 Output negative Pin 4 Supply negative Pin 5 Screen Pin 6 Sense

Zero and sensitivity control


Front panel trim adjustments.

Operating temperature range


0C to 50C.

Temperature performance
Sensitivity temperature coefficient <0.005% of reading/C. Zero offset temperature coefficient <0.003% F.S./C. (ref. to 50mV input - 5V common mode) Zero suppression temperature coefficient 0.005% of reading/C.

Accuracy
Combined non-linearity, hysteresis and repeatability. 0.1% F.S. for 70mbar to 60bar 0.15% F.S. for 60bar to 700bar

Industrial Pressure Transmitters PTX 1400


Figure 2
33.5 A/F 69 116

18

52

22 A/F

30

BASEEFA Apparatus Certificate Ex97D2058 Approval category: EEx ia IIC T4 zone 0 Two-wire, 4-20 mA pressure sensors designed for continuous monitoring in hazardous areas such as gas/petrochemical and process control environments.

q q q q q q q q

+/-0.15% typical accuracy Zero and span adjustment of +/-5% Absolute and Gauge referenced pressure ranges 4 to 20mA output proportional to applied pressure Compatible with hostile media All stainless steel construction Electron beam welded wetted parts Built-in protection from: power supply fluctuations; reverse polarity; voltage spike, RFI q Intrinsically safe, should be used with a galvanic isolation barrier such as RS stock no. 259-6682. q IP65 rated.

232-5244
Technical specification
Over press.____________1 bar for 100 and 250mbar ranges Supply volt. ____________9 to 30V d.c. (pin 1 +ve, pin 2 -ve) Voltage spike prot.__________withstand 600V to ENV 50142 Zero offset/span setting _________________+/-0.5% / +/-5% Long term stabil. ______________________________+/- 0.2% Combined non-linearity______+/-0.15% typ. +/- 0.25% max. hysteresis and repeat. Operating temp. ________________________-20C to +80C Weight__________________________________________200g Press. connection _____________G 1/4 female (BSP Parallel) Elect. connection ___DIN 43650 plug with fee mating socket.

250mBar - 400Bar LS Pressure Transmitters


Figure 3
27.5 22

49

38

78.5

22. A.F

11

Transducer pressure range


800-1200mbar absolute -1-1.6 bar gauge 250mbar gauge 0-1 bar absolute 0-1 bar gauge 0-4 bar gauge 0-6 bar gauge 0-10 bar gauge 0-16 bar gauge 0-25 bar gauge 0-100 bar gauge 0-250 bar gauge 0-600 bar gauge 313-4140 313-4156 313-4162 313-4178 313-4184 313-4190 313-4207 313-4213 313-4229 313-4235 313-4241 313-4257 313-4263

BASEEFA Apparatus Certificate No.: Ex892251x Approval category: EEx ia IIC T4 Two-wire, 4-20 mA pressure sensors designed for continuous monitoring in hazardous areas such as gas/petrochemical and process control environments. q Gauge style pressure measurement only q Robust stainless steel body with encapsulated electronic and silicon technology sensing element. q Suitable for use in Zone O Gas Group IIC, Temperature Class T4 q G 1/4 BSP pressure connection and DIN 43650 electrical connector (supplied) q Supplied with instruction leaflet with product and copy certificates available on request. q Should be used with a galvanic isolation barrier such as stock number RS 259 6682 q IP65 rated q Available with RS Servicepoint Calibration

Explanation
Absolute pressure is pressure measured with reference to vacuum pressure. An example of absolute pressure measurement is barometric measurement of atmospheric pressure. Gauge pressure is pressure measured with reference to atmospheric pressure. It can be viewed as a form of differential pressure. An example of gauge pressure measurement is measurement of tyre pressure.

Technical specification
Maximum overpressure 250m bar _______________________________________2 bar 1.6 bar to 25 bar __________________________3 x pressure 60 bar to 400 bar __________________________2 x pressure Supply voltage ______________________________9-30V d.c. Output current ________________________________4-20mA Combined non-linearity hysteresis and repeatability _________________10.15% FS Zero offset and span _________________5% site adjustable Long term stability _________________________0.1%, FS p/a Operating temp. ________________________-20C to +80C Process media temp. range _____________-30C to +120C Temperature effects ________________0.5% 9-10 to +50C RFI protection ___Conforms with IT 8839 (10kHZ to 500Mhz)

Range
Uncalibrated RS Servicepoint Calibrated

250m bar 1.6 bar 4 bar 10 bar 25 bar 60 bar 160 bar 400 bar

285-289 285-295 285-302 285-318 285-324 285-330 285-346 285-352

229-6512 229-6528 229-6534 229-6540 229-6562 229-6578 229-6584 229-6590

232-5244 Galvanic isolation barrier for intrinsically safe pressure transmitters


Analogue Input (Smart) RS Stock no. 259-6682 Technical specification
Power supply ____________________Power rail or terminals 7(L+) 8L-) 19(L-) Nominal voltage ___________________________d.c. 20V 35V Ripple __________________________Within supply tolerance Max. power consumption________________________ 1.2W Field circuit (intrinsically safe) Input signal ________________________________4mA 20mA Available voltage at 20mA ____________________d.c. 16.5V Voltage_____________________U = 19.65 - (16.5 x 103 x R R) Details of Certificate of ____________________BASEEFA No Conformity _____________________________EX-94.C.32326 Voltage U0 ______________________________________________d.c. 28V Current I0 _________________________________________________93mA Power P0 ________________________________________________660mW Permissible circuit values _______________________(EEx ia) Ignition protection class, category Explosion group _____________________________IIA/IIB/IIC *Max. external capacitance ______1.016F/0 366F/0 106F *Max. external inductance __________33.6mH/12.6H/4.2mH Fail-safe maximum voltage Urms Power supply________________________________250V rms Output (not intrinsically safe) Output signal _______________________________4mA 20mA Load ________________________500 Ohm (Current output) load of communicator in safe area 200 Ohm Ripple _________________________________________72A Fail-safe maximum voltage u rms _______________250Vrms Transfer characteristics Transfer error at 20C (68F) _______ 10A incl. calibration, linearity, hysteresis, load and nominal voltage Temperature drift per channel_______________ 20p.p.m./K Frequency range (Load = 250 Ohm) Hazardous area to safe area Bandwidth at 1mA - signal ______________0Hz 12kHz (-1dB) Safe area to hazardous area ____________0Hz 30kHz (-6dB) Bandwidth at 1mA - signal _____________1HZ 12kHz (-1dB) 1Hz 30kHz (-6dB) Rise time/fall time ____________________________40s/40s Conformity to standard Isolation co-ordination______________________to EN 50 178 Galvanic isolation__________________________to EN 50 178 Climatical condition__________________________to IEC 721 EMC ____________________________to EN 50 081-2, EN 50 0.82-2, NAMUR NE 21 Weight ________________________________ 150g (5.3oz) Ambient temperature ______________________-20C +60C Max. wire size__________________________2.5mm(14 awg) *Note: Particular attention should be paid to the transmitter and cable capacitances; combined they must not exceed the limits quoted above for explosion groups 11A, 11B and 11C

P.T.B. Cert. No Ex-94.C.2326 Approval category EExia 11a, 11b, 11c The unit is designed to power both 2-wire standard and SMART 420A transmitters in the hazardous area, transferring the output signal to the safe area. KFD2-STC1-Ex1 - Smart Transmitter/Power Supply Single channel Input EEx ia IIC Class 1, Div1, Groups A-G d.c. 24V Nominal power supply SMART compatible EMC in acc. with NAMUR NE 21 As an output signal, 20mA are available, digital information is generated by a processing system, field device or hand-held terminal and bi-directionally transferred by the barrier. This allows the operator to interrogate and reprogram the transmitter from the safe area, using non-certified communication equipment. These operations can be carried out in the hazardous area using suitably certified equipment. Hand-held terminals can be connected. In order to interface a hand-held terminal as easily as possible, these barriers come equipped with KF-STP-BU and KF-STP-GN terminals containing access holes for test probes.

Connections
Figure 4
Input EEx ia IIC Hazardous area Zone O
16.5V

PRESSURE TRANSMITTER

3-

1+

11 +

9+ 10(L-)

7 (L+) 8(L-)

4-20 mA signal to indicator/ 4-20mA panel meter

24Vdc power supply

Front View

1 4

2 5

3 6

7 10

8 11

9 12

Safe Area

232-5244 Druck DPI705, Handheld Portable Pressure Indicators


Easy to use rugged hand held portable pressure indicators, designed to offer accurate and reliable pressure measurement. An absolute pressure measurement version is also available. q 0.1% accuracy q Large 13mm 5 digit LCD display q 16 selectable pressure units q Ambient temperature indication in C or F q Automatic 60 second leak test mode q Averaging filter for unstable pressure measurement q Min/max peak hold q Manual tare facility q Zero pressure correction q Adjustable audio/visual high pressure alarm q Auto power off q Each instrument is supplied complete with 3 off AA alkaline batteries, users guide and soft carry pouch.

DPI 705-IS Indicator, Pressure, Intrinsically Safe

Technical specification
Pressure ranges ____________________200mbar differential 2bar gauge 2bar absolute 20bar gauge Overpressure ______________Display flashes at 110% x FS* Safe to 2 x FS Accuracy: combined non-linearity,______________________+/-0.1% FS hysteresis and repeatability Temperature effects: Span ______________+/- 0.02% rdg/C Zero 1 bar (absolute only)_______________+/-0.5% FS/C >1bar (absolute only) _____________________+/0.2% FS/C Pressure connections ____________G1/8 and 6mm OD hose Pressure units _______Mpa, kPa, Pa, mbar, bar, psi, mmHg, Mhg, inHg, hPa, kgf/cm2, mH2O H2O @ 20C, H2O @ 4C * FS = Full scale

BASEEFA Certificate no: Ex 98E 2332 A compact and rugged range of digital pressure indicators designed for all pressure measurement and approved for use in hazardous areas. q Instrinsically Safe (EEx ia IIC T4) q 16 pressure units q Leak testing q High accuracy (0.1% full scale) q Ambient temperature measurement q Zero/display tare q Peak value hold q Alarm setting q Auto-off q Transient filter q Hinged fold away stand for bench use or hanging q Supplied with 3 x AA Alkaline cells and carry case with belt loop

Technical specification
Pressure ranges ____________________200mbar differential 2 bar gauge 2 bar absolute 20 bar gauge Overpressure __Display flashes at 110% x FS safe to 2 x FS* Accuracy: Combined non-linearity _______________________0.1% FS hysteresis and repeatability Temperature effects: span ________________0.02% rdg/C Zero < 1 bar (absolute only) _______________0.05% FS/C 1 bar (absolute only)______________________0.02% FS/C Pressure connections ____________G1/8 and 6mm OD hose Pressure units _____________MPa, kPa, Pa, mar, psi, mmHg, Mhg, inHg, hPa, kgf/cm2, mH2 O inH20 @ 20, inH2O @ 4C * FS = Full scale

Range
Uncalibrated Differential DPI705-200mbar gauge DPI705-2bar absolute DPI705-2bar gauge DPI705-20bar 312-0494 312-0517 312-0539 312-0523 Calibrated 315-9010 315-9026 316-1215 315-9032

Range
DPI705 I/S 200 mbar DPI705 I/S bar gauge DPI705 I/S 2 bar absolute DPI705 I/S 20 bar Uncalibrated 339-0378 339-0390 339-0384 339-0407 Calibrated 339-4106 339-4112 339-4128

232-5244 Calibrator, Pressure, Intrinsically Safe, DPI 610/IS


q q q q q 24 pressure units Dual display of pressure and electrical readout Loop verification Pressure switch testing Supplied with user handbook, user guide, carrying case, shoulder strap, electrical leads and probes and a blanking plug for the pressure port.

Technical specification
BASEEFA Certificate No: Ex 99E2002X q A rugged, highly versatile, portable pressure calibrator approved for use in hazardous areas-Instrinsically safe (EEx ia IIC T4). q Provides a task driven user interface which facilitates easy setup of calibration, switch testing, leak testing, 4 to 20mA simulation, relief valve testing, data logging and ambient temperature recording modes. q Precision accuracy (0.025% full scale) q Built-in pressure/vacuum pump (-850mbar to +20 bar) q Cyrrent and voltage measurement q Pressure and current displayed simultaneously q Data storage and RS 232 interface Pressure range ________________________________0-2 bar 0-10 bar 0-20 bar Pressure media ______compatible with most common fluids Display ______________________5 digit, 13.6mm LCD digits with additional 16 text characters Resolution _________________________0.005% FS maximum Accuracy _________________________________0.075% FS Electrical ranges ____________________________0-50V d.c. 0-55mA Electrical outputs ___________24V nominal fixed, 25mA max Loop check ___________12mA nominal fixed, source or sink Operating temp. range __________________-10C to +50C Calibrated temp. range ____________________0C to +40C Pressure connection _________________________G18 female Environmental sealing _____________________________IP65 Batteries ___________________4 x size D cells zinc carbide

Technical specification
Pressure
Range 0-2 bar 0-7 bar 0-20 bar Inputs Voltage Current Temp. Overpressure 4 bar 14 bar 35 bar Range +/-50Vd.c. +/- 55mA -10 to 40C 0 to 24 mA Accuracy +/-0.05% rdg +/-0.004% FS +/-0.05% rdg +/-0.004% FS +/-1 C +/-0.1% +/- 5% +/-0.05% rdg +/- 0.1% FS 0.001mA Resolution 100 V 0.001mA 0.1C G 1/8 (female) Pressure port

Range
2 bar 10 bar 20 bar Uncalibrated 215-0826 215-0832 213-0844 Calibrated 215-7158 215-7142 215-7120

Outputs
Current

Calibrator Dirt/Moisture Trap RS stock no. 225-0049


Dirt/moisture trap designed for use with the Druck range of portable pressure calibrators. Regulator use will help ensure only clean dry gas enters the calibrator, thereby preventing contamination of the pneumatic system and reducing calibrator downtime for maintenance. q 1/8 BSPF (parallel) male to 1/8 BSPF (parallel) female pressure connections for in line connection to the calibrator pressure port. q Dismantles to empty collected deposits q Suitable for Druck portable pneumatic pressure field calibrators DPI601, DPI610 AND DPI603 series q Stainless steel/acrylic construction q Max. working pressure 35 bar Note: Must only be used in an upright position and is designed for pneumatic systems only.

Range
Type DPI10 I/S 2bar DPI610 I/S 7bar DPI610 I/S 20bar 339-0413 339-0429 339-0435

DP1 603 Druck Pressure Calibrator


A cost effective rugged portable instrument for calibration and service of pressure instrumentation. The instrument features the capability to generate pressures up to 20 bar and measure to an accuracy of 0.075% FS. If the device under test is a pressure sensor the instrument also measures the current or voltage output and display it together with the pressure reading. An energising voltage of 24V d.c. is provided for powering the device under test if necessary. A fixed current output of 12mA is also provided for loop verification and this can be used in an sink or source mode.

RS Components shall not be liable for any liability or loss of any nature (howsoever caused and whether or not due to RS Components negligence) which may result from the use of any information provided in RS technical literature.

RS Components, PO Box 99, Corby, Northants, NN17 9RS


An Electrocomponents Company

Telephone: 01536 201234 RS Components 1998