PMT Hps 34 TT 03 16 stt750 Spec
PMT Hps 34 TT 03 16 stt750 Spec
PMT Hps 34 TT 03 16 stt750 Spec
Technical Information
Introduction
Reliable measurement
o Supports Namur 89 Wire break All transmitters are available with the above listed
output and communication protocol option.
o Direct entry of Callendar-van Dusen coefficients
R0, α, δ and β for calibrated RTD sensors.
The STT750 SmartLine Temperature Transmitter is Field Device Manager (FDM) Software and FDM Express are
designed and manufactured to deliver high performance also available for managing HART device configurations.
across varying ambient temperature.
Diagnostics
Unique Indication/Display Options SmartLine transmitters all offer digitally accessible
The STT750 modular design accommodates a Standard diagnostics which aid in providing advanced warning of
possible failure events minimizing unplanned shutdowns,
LCD display.
providing lower overall operational costs
Standard LCD Display Features
System Integration
o Modular (may be added or removed in the field).
o 0, 90,180, & 270-degree position adjustments. o All SmartLine products meet the most current published
o Deg C, F, R, Kelvin, Milli volts, and Ohm measurement standards for HART.
Performance Specifications1,3
Reference Accuracy 2 (conformance to +/-3 Sigma)
Input Maximum Range Limits Digital Output D/A Standards
Type Accuracy Accuracy
(+/-) (% of span)
RTD °F
°C °C %
(2,3,4 wire)
Pt255 -200 to 850 -328 to 1562 0.90 0.025 IEC751 (=0.00385)
1. Digital Accuracy is accuracy of the digital value accessed by the Host system and the handheld communicator.
2. Total analog accuracy is the sum of digital accuracy and output D/A Accuracy.
3. Output D/A Accuracy is applicable to the 4 to 20 mA Signal output.
4. For TC inputs, CJ accuracy shall be added to digital accuracy to calculate the total digital accuracy.
5. Custom Callendar-van Dusen not available for Pt25 sensors.
4 STT750 SmartLine Temperature Transmitter
The STT750 device is compliant with IEC compliance EN 61326-1: 2013, EN IEC 61326-1: 2021 (CE) ; IEC 60533: 2015 /
IACS Req. 1991/Rev.8 2021 (Marine) and IEC 61326-3-1: 2017 (SIL)
HART: Worst case deviation < 0.1% of full span (for both Analog and Digital).
Figure 2 – HART Supply voltage and loop resistance chart & calculations
6 STT750 SmartLine Temperature Transmitter
Materials Specifications (see model selection guide for availability/restrictions with various models)
Parameter Description
Mounting Bracket Wall or 2” Pipe, Carbon Steel (Zinc-plated) or 316 Stainless Steel
Pure Polyester Powder Coated Low Copper (<0.4%) – Aluminum.
Electronic Housing Meets Type 4X / IP66 / IP67. All stainless steel housing is optional.
Cover O ring material: Silicone
Sensor/Cable Entry 1/2 NPT electrical connection or M20x1.5
Can be mounted in virtually any position using the standard mounting brackets. Brackets
Mounting
are designed to mount onto a wall or a 2-inch (50 mm) vertical or horizontal pipe.
Wiring Accepts up to 16 AWG (1.5 mm diameter). Preferred 18AWG and above for ease of
wiring.
Dimensions See Figures 4 to 9
Net Weight Lbs (kg) Aluminum housing for transmitter with Display – 2.7 lbs (1.22 kg)
Aluminum housing for transmitter w/o Display – 2.6 lbs (1.18 kg)
Stainless Steel housing for transmitter with Display – 4.9 lbs (2.22 kg)
Stainless Steel housing for transmitter w/o Display – 4.8 lbs (2.18 kg)
HART Protocol
Version:
HART 7
Power Supply
Voltage: 11.8 to 42.4Vdc at terminals
Load: Maximum 1400 ohms See figure 2
Minimum Load: 0 ohms. (For handheld communications, a minimum load of 250 ohms is required)
IEC 61508 Safety Certified SIL 2 and SIL 3
Standard Diagnostics
STT750 top level diagnostics are reported as either critical or non-critical as listed below. All diagnostics are readable via
the DD/DTM tools. All critical diagnostics will appear on the integral display.
Critical Diagnostics
Sensor Module Fault
Communications Module Fault
Sensor Communications Fault
Input Fault
IEC 61508 SIL 2 for non-redundant use and SIL 3 for redundant use according to EXIDA and
SIL 2/3 Certification TÜV Nord Sys Tec GmbH & Co. KG under the following standards: IEC61508-1: 2010; IEC
61508-2: 2010; IEC61508-3: 2010.
SIL EMI/EMC compliance as per Standard: IEC 61326-3-1: 2017
MARINE TYPE
American Bureau of Shipping (ABS); Certificate number: 23-2416104-PD
APPROVAL Lloyd’s Register (LR); Certificate number: LR23389941TA
STT750 SmartLine Temperature Transmitter 7
Wiring Diagram
Approval Certifications:
Ex db IIC T6..T5 Gb ;
Ex tb IIIC T 95oC Db
Intrinsically Safe, Certificate: 2689056:
Class I, II, III, Division 1, Groups A, B, C,
D, E, F, G; T4
CSA-Canada
B Class I Zone 0 AEx ia IIC T4 Ga
and USA
Class I Zone 2 Ex ic IIC T4 Gc -50oC to 70oC
4-20 mA/
Ex ia IIC T4 Ga Note 2 FISCO:
DE/HART/FF
Ex ic IIC T4 Gc -50oC to 45oC
FISCO Field Device (Only for FF Option)
Class I Zone 0 AEx ia IIC T4 Ga
Class I Zone 2Ex ic IIC T4 Gc
Ex ia IIC T4 Ga
Ex ic IIC T4 Gc
Non-Incendive, Certificate: 2689056:
Class I, Division 2, Groups A, B, C, D; T4
4-20 mA/
Note 1 -50oC to 85oC
DE/HART/FF
Class I Zone 2 AEx nA IIC T4 Gc
Ex nA IIC T4 Gc
Standards: CSA C22.2 No. 0-10; CSA 22.2 No. 25-1966 (reaffirmed 2009);
CSA C22.2 No. 30-M1986 (reaffirmed 2012); CSA C22.2 No. 94-M91;
CSA C22.2 No. 61010-1: 2012; CSA-C22.2No.157-92 (reaffirmed 2012);
C22.2 No. 213-2017; C22.2 No. 60529-2016
C22.2 No. CSA 60079-0:2011; C22.2 No. 60079-1: 2011; C22.2 No. 60079-11:2014;
C22.2 No. 60079-15: 2012; C22.2 No. 60079-31:2015;
B
ANSI/ ISA12.12.01-2017; ANSI/ ISA 60079-0 (12.00.01): 2013;
ANSI/UL 60079-1 : 2015; ANSI/ ISA 60079-11(12.02.01) : 2012;
ANSI/ ISA 60079-15(12.12.02) : 2012 ;
ANSI/ ISA 60079-31: 2015;
FM Class 3615: Aug 2006; FM Class 3616: Dec 2011; ANSI/ IEC 60529 : Edition 2.1
ANSI/ UL 913: 2015; ANSI/UL 61010-1: 2016; UL 50: Ed 11
Flameproof, Sira 14ATEX2046X:
T 95oC, T5: Ta= -50oC
II 2 GD 4-20 mA/
Note 1 to 85oC
Ex db IIC T6..T5 Gb DE/HART/FF
T6: Ta= -50oC to 65oC
Ex tb IIIC T 95oC Db
Intrinsically Safe, Sira 14ATEX2046X:
II 1 GD -50oC to 70oC
Ex ia IIC T4 Ga 4-20 mA/
Note 2
Ex ia IIIC T95oC Da DE/HART/FF FISCO:
FISCO Field Device (Only for FF Option) -50oC to 45oC
Ex ia IIC T4 Ga
Enclosure: IP66/ IP67 ALL ALL ALL
C ATEX
Standards: EN 60079-0: 2012/A11:2013; EN 60079-1 : 2014; EN 60079-31 : 2014
EN 60079-11: 2011; EN 60079-26 : 2006; EN 60529 : 2000 + A1
Increase Safety/ Intrinsic Safety, Sira
14ATEX4052X:
II 3 G -50oC to 85oC
4-20 mA/
Ex ec IIC T4 Gc Note 1 FISCO:
DE/HART/FF
Ex ic IIC T4 Gc -50oC to 45oC
FISCO Field Device (Only for FF Option)
Ex ic IIC T4 Gc
Enclosure: IP66/ IP67 ALL ALL ALL
Standards: EN 60079-0: 2012/A11:2013; EN 60079-7:2015; EN 60079-11:2012
Flameproof, SIR 14.0020X T 95oC, T5: Ta= -50oC
4-20 mA/
Ex db IIC T6..T5 Gb Note 1 to 85oC
o DE/HART/FF
Ex tb IIIC T 95 C Db T6: Ta= -50oC to 65oC
Intrinsically Safe, SIR 14.0020X
Ex ia IIC T4 Ga -50oC to 70oC
4-20 mA/
Ex ia IIIC T95oC Da Note 2 FISCO:
DE/HART/ FF
FISCO Field Device (Only for FF Option) -50oC to 45oC
Ex ia IIC T4 Ga
D IECEx Non Sparking, SIR 14.0020X
Ex ecIIC T4 Gc -50oC to 85oC
4-20 mA/
Ex ic IIC T4 Gc Note 1 FISCO:
DE/HART/ FF
FISCO Field Device (Only for FF Option) -50oC to 45oC
Ex ic IIC T4 Gc
Enclosure: IP66/ IP67 ALL ALL ALL
Standards: IEC 60079-0: 2011; IEC 60079-1 : 2014,;
IEC 60079-11 : 2011; IEC 60079-7:2015;
IEC 60079-31 : 2013
10 STT750 SmartLine Temperature Transmitter
J EAC Ex Flameproof:
4-20 mA/
(Russia, 1 Ex d IIC T4 Gb Note 1 -50oC to 85oC
DE/HART/FF
Belarus and Ex tb IIIC T95oC Db
Kazakhstan) Intrinsically Safe:
0 Ex ia IIC T4 Ga -50oC to 70oC
4-20 mA/
Ex ia IIIC T4 Db Note 2 FISCO:
DE/HART/FF
FISCO Field Device (Only for FF Option) -50oC to 45oC
0 Ex ia IIC T4 Ga
Non Sparking: 4-20 mA/
Note 1 -50oC to 85oC
2 Ex nAc IIC T4 DE/HART/FF
Enclosure: IP66/ IP67 ALL ALL ALL
P CCoE Ex ia IIC T4 Ga -50oC to 70oC
4-20 mA/
(India) FISCO Field Device (Only for FF Option) Note 2 FISCO:
DE/HART/FF
Ex ia IIC T4 -50oC to 45oC
Ex d IIC T4 Gb 4-20 mA/
Note 1 -50oC to 85oC
DE/HART/FF
Notes
1. Operating Parameters:
4-20 mA/ HART (Loop Terminal)
Voltage= 11.8 to 42 V Current= 4-20 mA Normal (3.8 – 23 mA Faults)
TRANSMITTER ENCLOSURE CAN BE ROTATED A TOTAL OF 90º FROM THE STANDARD MOUNTING POSITION
Figure 6 – STT750 Pipe Mount with adapter housing - Horizontal & Vertical
*Note 1: Figure 6 and 7. The housing adapter may not be present on all transmitter models. If the housing adapter is not
present, subtract 24,5mm (0,96 inches) from the dimension specified.
14 STT750 SmartLine Temperature Transmitter
millimeters
Reference Dimensions: A
inches E
The Model Selection Guide is subject to change and is inserted into the specification as guidance only.
Section 13
Page: STT7-1
Model STT750 Effective Date: January 12, 2024
Availability
KEY NUMBER INPUT TYPE Selection
Universal Input STT750 *
TABLE III AGENCY APPROVALS (See data sheet for Approval Code Details)
No Approvals Required 0 *
FM Explosion proof, Intrinsically Safe, Non-incendive, & Dustproof A *
CSA Explosion proof, Intrinsically Safe, Non-incendive, & Dustproof B *
ATEX Explosion proof, Intrinsically Safe & Non-incendive C *
IECEx Explosion proof, Intrinsically Safe & Non-incendive D *
Approvals SAEx Explosion proof, Intrinsically Safe & Non-incendive E *
INMETRO Explosion proof, Intrinsically Safe & Non-incendive F *
NEPSI Explosion proof, Intrinsically Safe & Non-incendive G *
KOSHA Explosion proof, Intrinsically Safe & Non-incendive H *
EAC Explosion proof, Intrinsically Safe & Non-incendive J *
CCoE Explosion proof, Intrinsically Safe & Non-incendive P *
TABLE IV TRANSMITTER ELECTRONICS SELECTIONS
Housing and Material Connection Lightning protection
Polyester Powder Coated Aluminum 1/2 NPT Yes C_ _ *
a. Electronic
Housing Material & Polyester Powder Coated Aluminum M20 Yes D_ _ *
Connection Type 316 Stainless Steel (Grade CF8M) 1/2 NPT Yes G_ _ *
316 Stainless Steel (Grade CF8M) M20 Yes H_ _ *
Analog Output Digital Protocol
b. Output/ Protocol
4-20mA dc Ext Zero, Span & HART Protocol _H_ *
Indicator Config Buttons Languages
None None None __0 *
None Yes (Zero/Span Only) None __A *
c. Customer Standard (w/internal
Interface Zero, Span & Conf None English __S *
Selections Buttons)
Standard (w/internal
Zero, Span & Conf Yes English __T *
Buttons)
MODEL RESTRICTIONS
Available Only with Not Available with
Restriction Letter
Table Selection(s) Table Selection(s)
j Vb _ 1,2 _
d VIIa 1,3,5,6
m IVa D,H, _ _
n IVa C,G, _ _
b Select only one option from this group
Singapore
Honeywell Pte Ltd.
Phone: +(65) 6580 3278
Fax: +(65) 6445-3033
South Korea
Honeywell Korea Co Ltd
Phone: +(822) 799 6114
Fax: +(822) 792 9015
Process Solutions
Honeywell