Vtools User Manual
Vtools User Manual
User Manual
Vision Tools
User Manual
Doc Ver : 1.15
Note :
This manual is intended for use with all Vision Tools Family of products , describing all features and
modules available in the application .
According to the specific Vision Tools product you have , features and modules will be disabled or
enabled according to your licensed application and the product family member (e.g. multiview ,
radview, sonopro, mammoview ,.....etc.) most of the features will be available in the MultiView
version of the product as it is a general viewer for all modalities.
© Copyright by MILLENSYS
MILLENSYS Contacts
Web site
- www.millensys.com
Email
- Sales : sales@millensys.com
- Support : support@millensys.com
©MILLENSYS 2 / 136
All rights reserved
Vision Tools Workstations
User Manual
Table of Contents
TABLE OF CONTENTS.............................................................................................. 3
1 INTRODUCTION ........................................................................................ 6
1.1 VISION TOOLS FAMILY OF PRODUCTS............................................... 7
• VISION TOOLS MULTIVIEW | PACS WORKSTATION ........................... 7
• VISION TOOLS RADVIEW | FOR RADIOGRAPHY CR/ DR MACHINES ... 7
• VISION TOOLS MAMMOVIEW | FOR MAMMOGRAPHY MG / CR/ DR
MACHINES ................................................................................................ 7
• VISION TOOLS OPERA PRO | FOR X-RAY AND ANGIOGRAPHY
MACHINES ................................................................................................ 8
• VISION TOOLS CTVIEW | FOR COMPUTED TOMOGRAPHY "CT"
MODALITIES.............................................................................................. 8
• VISION TOOLS MRVIEW | FOR MRI MODALITIES .............................. 8
• VISION TOOLS CATHVIEW | FOR VASCULAR CATHLAB MODALITIES .. 8
• VISION TOOLS 3DVIEW | ADVANCED VISUALIZATION 3D
APPLICATION. ........................................................................................... 9
• VISION TOOLS SONOPRO | FOR COLOR DOPPLER AND ECHO
ULTRASOUND MACHINES.......................................................................... 9
• VISION TOOLS ENDO | FOR ENDOSCOPY, LAPAROSCOPY,
ARTHROSCOPY, …, MACHINES................................................................... 9
• VISION TOOLS GVIEW | FOR NM MACHINES .................................... 9
2 LOGIN SCREEN ........................................................................................ 10
2.1 MAIN SCREEN AND NAVIGATION ................................................... 10
3 EXAM LIST (WORKLIST\DIRECTORY) ....................................................... 11
3.1 MAIN SCREEN AND NAVIGATION ................................................... 11
3.2 QUICK SEARCH TOOL ..................................................................... 12
3.3 QUICK TOOLS................................................................................. 12
3.4 ADVANCED SEARCH ....................................................................... 13
3.5 CREATE USER FILTERS .................................................................... 13
3.6 IMPORTING STUDIES...................................................................... 14
3.7 VIEW MENU................................................................................... 15
©MILLENSYS 3 / 136
All rights reserved
Vision Tools Workstations
User Manual
©MILLENSYS 5 / 136
All rights reserved
Vision Tools Workstations
User Manual
1 Introduction
Vision Tools is a powerful stand-alone PACS workstation for viewing, processing and managing medical images
from different modalities (FDA 510K cleared).
With its archiving capabilities, static and dynamic image viewing, variety of processing tools, printing features and
reporting, the Vision Tools MultiView application can be considered as an ideal solution for most doctors and users
who need to increase their daily diagnosis productivity aiming to an improve of the provided patient healthcare
service.
Vision Tools suites almost all of the daily activities of any physician in a low cost of ownership and high content
solution that is scalable to any organization.
Stay up-to-date with MILLENSYS automatic updates feature that keeps the Vision Tools application updated to the
latest version with new features and bug-fixes.
Combining Vision Tools with MILLENSYS Cloud PACS "Vision Tools Workspace" decreases the diagnosing time
through the powerful included tools leading to an increase the productivity of all the users dealing with PACS.
©MILLENSYS 6 / 136
All rights reserved
Vision Tools Workstations
User Manual
Vision Tools has a wide variety of workstation software that suites almost all modalities
described as the following
©MILLENSYS 7 / 136
All rights reserved
Vision Tools Workstations
User Manual
©MILLENSYS 8 / 136
All rights reserved
Vision Tools Workstations
User Manual
©MILLENSYS 9 / 136
All rights reserved
Vision Tools Workstations
User Manual
2 Login Screen
The multilingual core of the Vision Tools is very important to have a localized version of
Workspace to suite every country.
©MILLENSYS 10 / 136
All rights reserved
Vision Tools Workstations
User Manual
Default patient list grid is showing default\most commonly used columns including
demographics and exam details
Users can customize their own grids, searching filters, customize icon features, opening
patient history, burning CD's, importing studies and sending studies to other DICOM
server from their user profile, this will be described later in the manual.
©MILLENSYS 11 / 136
All rights reserved
Vision Tools Workstations
User Manual
Quick search above the grid is a tool helps query studies using most common fields
such as modality type, facility\branch, study\exam date...etc.
©MILLENSYS 12 / 136
All rights reserved
Vision Tools Workstations
User Manual
This tool is used to define more search criteria to get specific studies for diagnosing and
reporting, after applying your search criteria
From advanced search (Figure 1.6) - Users will define any search criteria then select
[Add] to type a name for your own fast search filters (as shown in Figure 1.7).
©MILLENSYS 13 / 136
All rights reserved
Vision Tools Workstations
User Manual
All saved search filters will be saved automatically to each user's profiles and will be
displayed as buttons above the grid as shown on Figure 1.8
In order to activate them just press on the button, this will trigger the search criteria
defined
©MILLENSYS 14 / 136
All rights reserved
Vision Tools Workstations
User Manual
1. Open Patient Data: This option will open the selected cine loop from
the data list box in the cine tool.
2. Open Image Display Tool: This option will open the images of the selected
patient in the display tool.
3. View Selected patient Images: This option will open the selected patient
images in a list beside the patient list.
©MILLENSYS 15 / 136
All rights reserved
Vision Tools Workstations
User Manual
Charts Settings
1. Legend: Show or hide the chart details which appear in a box to the left
of the chart.
2. Wall: Show or hide the background.
3. 3D: View the chart in 2 dimensions or volumetric 3 dimensions.
4. Colored: View the chart with colors or with grey color only.
5. Back Color: Change the background color.
6. Axis: Show or hide the axis of the chart.
7. Include empty items: Show or hide the items with no data.
8. Chart Type: View the chart as column or pie.
9. Rotate: Rotate the pie chart.
The charts can be printed from the Print button and can be copied
to the clipboard so that the user can use it in any application, this
can be done from the Copy button in the left bottom of chart form.
©MILLENSYS 16 / 136
All rights reserved
Vision Tools Workstations
User Manual
©MILLENSYS 17 / 136
All rights reserved
Vision Tools Workstations
User Manual
Note:
To create an image DICOM CD select a patient then right click the mouse and
choose the Archive to CD-R option. A message will appear to inform total
patient data size in megabytes.
©MILLENSYS 18 / 136
All rights reserved
Vision Tools Workstations
User Manual
©MILLENSYS 19 / 136
All rights reserved
Vision Tools Workstations
User Manual
1. Send Selected Images to Dicom Server: This option will open the DICOM
Connections form which enables the user to send the selected images to the required
DICOM server from the list, the user can add or delete any DICOM server from this
form. After choosing the required server press the Send button or cancel the whole
process from the Abort button or by closing the form.
©MILLENSYS 20 / 136
All rights reserved
Vision Tools
User Manual
4 Image Viewer
The Image Tool will open on the assigned monitors (According to default settings
or default hanging Protocol)
The user can choose to view more than one image
in the main screen by choosing the Windows menu
then specifying the number of images to appear in
the main screen.
• Study Navigation:
-In case of more than one study found in TimeLine: Viewer will display
first the study that user already selected from patients list\worklist.
Other studies could be opened as mentioned in section 7.2
• Series Navigation:
- To view any series just “Drag & Drop” by left click.
-“Series Restricted Display”: If enabled, series will be displayed only on the
Window to which user dragged and dropped the series not to all windows.
Also user will not be able to browse the next series in the same window.
• Series Description:
©MILLENSYS 22 / 136
All rights reserved
Vision Tools
User Manual
Under each series there is a description written in the form “(XX)|YY”
where “XX” represents the series number and “YY” represents number
of images of the series.
Example: (4750)|73 means that series number is 4750 and this series
has 73 images.
• Image Navigation:
To navigate images: By mouse scroll, Image Navigation Buttons or
keyboard shortcuts. Keyboard shortcuts could be configured in profile
settings.
- Image Navigation Buttons:
©MILLENSYS 23 / 136
All rights reserved
Vision Tools
User Manual
- Save as: Save selected image to directory without annotations and
modifications like zooming and without image information.
- Save in Current Series: Save modified image (with annotations, zooming &
WW/WL) to current selected series.
- Save Display as: Save image to directory with annotations and modifications
like zooming and with image information.
©MILLENSYS 24 / 136
All rights reserved
Vision Tools
User Manual
-
- Maxi
mize/Minimize:
Double click to maximize and double click to minimize. Also from
- WW/WL:
• By mouse: Change WW\WL or (B\C) of selected image(s)by clicking
and holding the right click button.
• WW/WL Presets:
• In
addition to using mouse in changing WW/WL, user can select window
from “Window Presets” menu. User also can edit, delete and create
new ones.
©MILLENSYS 25 / 136
All rights reserved
Vision Tools
User Manual
• In case of image have 2 or 3 ww\wl sent in the DICOM images and
created by the modality, user will be able to switch between these
windows through “W1”, “W2” and “W3” buttons.
- Image Restore:
To restore original image, select “Restore Original” from right click menu.
©MILLENSYS 26 / 136
All rights reserved
Vision Tools
User Manual
- User can select\deselect one or more windows\images from the top left
selection mark.
©MILLENSYS 27 / 136
All rights reserved
Vision Tools
User Manual
Lock WW\WL feature is used to fix WW\WL adjustment while user browsing of
images. To enable\disable it go to Menu->Edit menu->Select Lock Window/Color
Adjustments or Press Button Lock WW\WL on the Toolbar
LOCK ZOOM
Lock Zoom feature is used to fix zooming adjustment while user browses images.
To enable\disable it go to Menu->Edit menu->Select Lock Zoom/Pan Adjustments
or Press Button Lock Zoom/Pan on the Toolbar
ANNOTATION SETTINGS
In Order to change font and drawing settings From to Menu->Edit menu->Drawing
Settings and Menu->Edit menu->Font Settings.
Drawing Settings for the drawings like arrows and rectangles.
Font settings used to change settings of the text font of annotation.
©MILLENSYS 28 / 136
All rights reserved
Vision Tools
User Manual
Drawing Settings Font Settings
ANNOTATION DRAWING
Select the annotation from “Annotations” menu or annotations shortcut in the
toolbar then start draw it directly on image using mouse left button click.
Note: Shortcuts could be configured from profile settings.
©MILLENSYS 29 / 136
All rights reserved
Vision Tools
User Manual
ANNOTATION COPY\PASTE\SAVING
- Select “Selection Tool” from Annotations toolbar icons to select the drawn
annotation.
- Right click on any annotation and select “Copy”.
- Go to target image if the copied annotation will be pasted to another image.
- Select “Selection Tool” again.
- Left click and select “Paste”.
©MILLENSYS 30 / 136
All rights reserved
Vision Tools
User Manual
4.8 Series Synchronization (for CT and MRI)
In right click menu select “Link/unlink selected windows” to apply/disable
synchronized viewing.
After applying:
After applying linking you will notice change in highlighted image border, for
synchronized windows
©MILLENSYS 31 / 136
All rights reserved
Vision Tools
User Manual
Note: In order to link series, the windows should me Series Restricted Mode,
means each window would be configured to load certain series
©MILLENSYS 32 / 136
All rights reserved
Vision Tools
User Manual
Step 2: Click “New” to add new hanging protocol
Step 3: Add screen in order to display this protocol on multiple screen , default it
will be displayed in only one screen
©MILLENSYS 33 / 136
All rights reserved
Vision Tools
User Manual
Step 4: modify screen, this function is used to define no of windows in each screen,
select screen from the tree view on left , then modify no of rows and columns to
specify horizontal and vertical windows.
©MILLENSYS 34 / 136
All rights reserved
Vision Tools
User Manual
Step5: select monitors, select each screen will be displayed on which physical
computer monitor, select screen from left tree, the press Screen Display, select
from Run This Screen On to select the monitor
©MILLENSYS 35 / 136
All rights reserved
Vision Tools
User Manual
Step 6: in order to specify special selection of image to be displayed at which
window,Assign screen to current\prior study (User could compare current study
with up to 4 prior studies).
Step7:
Click “Next” 2 times then write hanging protocol name and save.
This new hanging protocol will appear in “H.Protocols” menu where user can select
and apply
©MILLENSYS 36 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 37 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 38 / 136
All rights reserved
Vision Tools
User Manual
To select other design not found in the menu, select “Other Layouts” from the
menu then “Insert_Rows_Col”.
©MILLENSYS 39 / 136
All rights reserved
Vision Tools
User Manual
Also from “Windows” menu then “Divide selected Windows To” or from the
shortcut button.
VIEW MPR\3D
©MILLENSYS 40 / 136
All rights reserved
Vision Tools
User Manual
MPR\3D modules could be viewed from “Windows” menu as well as shortcuts
toolbar.
Note: this will require to have license to open MPR\3D modules, please system
administrator is giving the logged in user to open these modules from system
administration tool
©MILLENSYS 41 / 136
All rights reserved
Vision Tools
User Manual
From “Process” menu select and apply filters on the selected images.
And once user selects any option in the menu,
“Filters” window will open, in the tools panel
Either to apply predefined filters, from dropdown menu, and also used can select
filter save it as preset filter to be used in the future.
“Filters” window
©MILLENSYS 42 / 136
All rights reserved
Vision Tools
User Manual
4.11.2 Image Invert\Rotate\Resize\Cropping
Image Invert, Rotate, Resize and cropping tools are found in “Process” menu and
could be added to shortcuts toolbar.
Rotate tools:
©MILLENSYS 43 / 136
All rights reserved
Vision Tools
User Manual
• Distance
- Select Distance from the list.
- Move the mouse to the start point.
- Left click with holding.
- Move the mouse to the end point.
- Stop left click holding and measured distance will be written by the end
of the line.
• Angle
- Select Angle from the list.
- 3 left clicks: first one at the angle head point, second one at the end of
the first line and the third one at the end of second line.
• Calibration tool
If image is not calibrated, measurements couldn’t be taken until user
calibrates the image.
- Select calibration tool from the list (Or user will be prompted to open it
when trying do measurements on non-calibrated image).
- Draw distance line on the image and write the distance
- After that you can use the measurement tools
©MILLENSYS 44 / 136
All rights reserved
Vision Tools
User Manual
4.13 Display Image Inset\Scan Lines (for CT and MRI)
“Image Inset & Scan Lines” are found in “Process” menu and also are added by
default to shortcuts toolbar.
©MILLENSYS 45 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 46 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 47 / 136
All rights reserved
Vision Tools
User Manual
4.14 Filming
To create new: Click “Filming Menu” and select “Film Designer” or click “Film
Designer” directly.
©MILLENSYS 48 / 136
All rights reserved
Vision Tools
User Manual
4.14.2 Add image to film
To add a specific image to specific cell: Just 2 mouse clicks in order, the first one on
the film cell and the second one on the image number or image thumbnail (Images
could be listed as numbers or thumbnails according to user profile settings).
©MILLENSYS 49 / 136
All rights reserved
Vision Tools
User Manual
• To print the current film only click the “Print Film” button
• User can create more than one film and print all of them by
one click on the “Print All Films” button.
User can add films to print queue to print them later by the
“Queue Current Film” button or “Queue All Films” button.
To check status of the print queue click the “Open Print Queue”
button and the next screen will open
User can select the printer, change film format and change film size
©MILLENSYS 50 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 51 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 52 / 136
All rights reserved
Vision Tools
User Manual
4.14.7 Customized template printing
User is able to create new templates through “Film Template Design” tool.
To open “Film Template Design” tool:
©MILLENSYS 53 / 136
All rights reserved
Vision Tools
User Manual
Once the user click on the button two windows will open, Design window & saved
templates list.
Design Controls:
Her how to create a template with 2 windows. The first window for the scout
image and the second window for the axial series (2 * 2 images).
©MILLENSYS 54 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 55 / 136
All rights reserved
Vision Tools
User Manual
Select the second window and divide it to 4 images
©MILLENSYS 56 / 136
All rights reserved
Vision Tools
User Manual
Click “Save Template” button then write the template name and
click “OK”
This new template will be added to the saved templates list so that user
could select it later directly.
©MILLENSYS 57 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 58 / 136
All rights reserved
Vision Tools
User Manual
5 Optional Advanced Post Processing Tools
5.1 MPR (Multi-Planar Reconstruction) (for CT only)
Multi Planar Reconstruction (MPR) allows images to be created from the original
axial plane in either the coronal, sagittal, or oblique plane as well as special curved
and rotating plans.
MPR offer additional advantages for radiologist towards easier approach for
diagnosis as:
View Lesions from different angles
Better localization of lesions within the volume
Measurement of lesions on different axis
To open the MPR Module you need to select the required series from the left
series panel then Press the MPR button from the top toolbar
©MILLENSYS 59 / 136
All rights reserved
Vision Tools
User Manual
1- Orientation cube: showing the image ordination in 3D to give a precise
information to the user.
2- Cross cut lines (color coded) each line will be represented on the view
port of the same color
3- Slice thickness: that will be used in during the image reconstruction and
rendering, this can be changed editing this number or manipulating the
vertical slider on the right of the image box.
4- Blend Mode: the is the blending method used while blending pixels of the
selected slice thickness
a. Max: The highest intensity pixels will be shown (contrast
enhanced objects, calcium plaques…)
b. Min: The lowest intensity pixels will be shown (Air, Fat…)
c. Avg: The average of all pixel intensities will be shown which will
give an overlapped image similar to the ordinary X-ray images.
And this image showing average blending mode with X-ray like effect.
©MILLENSYS 60 / 136
All rights reserved
Vision Tools
User Manual
The screen shot bellow shows to apply the double oblique concept to visualize the
vertebral column as whole in one slice.
And this shows the ability of curved MPR to straighten and visualize the Aorta in
one slice
©MILLENSYS 61 / 136
All rights reserved
Vision Tools
User Manual
For Orthogonal series creation we can set the start and the end of the series and
fine tune from other viewports to have the exact location needed, then adjust the
slice thickness, spacing and the series label then click generate the application will
automatically calculate the reset and generates the series images, from the forth
viewport, user can review the series if it is ok, the pressing save will physically save
this newly generated series to this patient study.
©MILLENSYS 62 / 136
All rights reserved
Vision Tools
User Manual
And in this screen shows how to create a new series from a curved plan for better
capturing of curved objects centered throughout the series.
And here we used the Spiral to rotate around the Aorta and get all possible capture
from 360 degree view
©MILLENSYS 63 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 64 / 136
All rights reserved
Vision Tools
User Manual
For better clinical diagnosis user can use the stitching option in Image Tool to
create composite image starting from its component images to get a complete
panoramic image which cannot be captured in a single scan .
For two images like below , stitching is used to create one image combining both
images in a smooth way
The user should first choose Image Stitch option from Windows menu
2- The user then should select the first (upper ) from the tool box (Thumbnails)
3- Then selecting the lower image(s) maximum 2 images ( Total 3 individual
images)
4- Then using the stitch tools to align images and stitching them into one Image
5- After stitching the selected images user can save the output image (Stitched
image) using the save button.
©MILLENSYS 65 / 136
All rights reserved
Vision Tools
User Manual
Please refer to the below figure for discovering Stitch tools
Move Image
To move image Right, Left , Up and Down (Also are done by mouse left click )
Zoom Image
To Zoom Image In and Out (Also are done by mouse middle click )
Rotate Image
To rotate image Clockwise and Anticlockwise (Also are done by mouse right click )
Auto Crop
Eliminates the unwanted black area around the image.
Save image
To save constructed (Stitched )image in a new series in the original study.
Reset
To clear the stitch window
Exit Stitch
To Exit Stitch window and go back to Windows mode
©MILLENSYS 66 / 136
All rights reserved
Vision Tools
User Manual
5.4 QCA (Quantitative Coronary Analysis) (Option for
Cathlab only)
For using QCA feature you should open an angiography case and do the following,
a. Select any image from series or a frame from a cine loop (by pausing cine
loop using double click mouse action) that you would like to apply QCA
upon it,
©MILLENSYS 67 / 136
All rights reserved
Vision Tools
User Manual
c. In figure below, you should scroll to reach desired frame then press on
QCA button.
d. After that you will calibrate the image by using left click mouse on
catheter area and type the number of catheter used during the operation.
e. Finally you will draw the path that required to be analyzed from point of
perspective.
©MILLENSYS 68 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 69 / 136
All rights reserved
Vision Tools
User Manual
5.5 QLVA (Quantitative Left Ventricle Analysis) (Option
for Cathlab only)
For using QLVA feature you should open a cardiac case and do the following,
a. First open the study and select required cine loop.
©MILLENSYS 70 / 136
All rights reserved
Vision Tools
User Manual
b. To open QLVA go to Process -> Cathlab then select QLVA.
c. In figure below, you will follow few steps to get QLVA analysis,
d. First you should select ESV (End Systolic Volume) frame (by scrolling in
cine loop) then press on ESV button,
©MILLENSYS 71 / 136
All rights reserved
Vision Tools
User Manual
e. Then you should scroll to get EDV (End Diastolic Volume) frame (by
scrolling in cine loop) then press on EDV button
©MILLENSYS 72 / 136
All rights reserved
Vision Tools
User Manual
f. Then press this button Calculate LVA to continue,
h. After that you will need to draw ESV & EDV outline on selected frame as
below, press on 1 button then draw outline for ESV.
©MILLENSYS 73 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 74 / 136
All rights reserved
Vision Tools
User Manual
j. Then Press done button to get your required analysis,
©MILLENSYS 75 / 136
All rights reserved
Vision Tools
User Manual
5.6 Volume Rendering Technique (VRT) (Option for CT
only)
VRT module is one of MillenSys product, in scientific visualization and computer
graphics, volume rendering is a set of techniques used to display a 2D projection of
a 3D discretely sampled data set.
A typical 3D data set is a group of 2D slice images acquired by a CT, MRI, or
MicroCT scanner. Usually these are acquired in a regular pattern (e.g., one slice
every millimeter) and usually have a regular number of image pixels in a regular
pattern. This is an example of a regular volumetric grid, with each volume element,
or voxel represented by a single value that is obtained by sampling the immediate
area surrounding the voxel.
©MILLENSYS 76 / 136
All rights reserved
Vision Tools
User Manual
5.6.1 Gallery
3Dview application module comes with a wide variety of presets that store the
settings used to create the 3D images. 3Dview has Real-time Preview of presets;
when you browse a preset, the preview window will show the effect of applying
that preset to your currently loaded volume dataset. You can create customized
presets. Presets can store more than just window width and level; all imaging
parameters can be saved in a preset allowing for highly customized capabilities.
©MILLENSYS 77 / 136
All rights reserved
Vision Tools
User Manual
5.6.2 Understanding GUI
5.6.3 Segmentation
©MILLENSYS 78 / 136
All rights reserved
Vision Tools
User Manual
reconstructions with the help of interpolation algorithms.
So we will start using our segmentation threshold tool, which know as one of the
simplest method of image segmentation. This method is based on a clip-level (or a
threshold value) to turn a gray-scale image into a binary image. The key of this
method is to select the threshold value (or values when multiple-levels are
selected).
©MILLENSYS 79 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 80 / 136
All rights reserved
Vision Tools
User Manual
3- Next step is to modify CT number using threshold tool (for not loosing time
details of the segmented object or vessel):-
©MILLENSYS 81 / 136
All rights reserved
Vision Tools
User Manual
4- In some cases user likes to see the removed object (with different opacity
level):-
©MILLENSYS 82 / 136
All rights reserved
Vision Tools
User Manual
5- Finally user can construct different views from the constructed 3D volume:-
©MILLENSYS 83 / 136
All rights reserved
Vision Tools
User Manual
5.6.4 Annotations and Measurements
Annotation & Measurement tools allow users to add annotation, measure distance
and adding bookmarks, which help users saving their work on safe environment.
©MILLENSYS 84 / 136
All rights reserved
Vision Tools
User Manual
5.6.5 Capturing Reconstruction
©MILLENSYS 85 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 86 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 87 / 136
All rights reserved
Vision Tools
User Manual
5.6.6 Virtual Colonoscopy
©MILLENSYS 88 / 136
All rights reserved
Vision Tools
User Manual
5.6.6.1 Gallery
3Dview application module comes with a wide variety of presets that store the
settings used to create the 3D images. 3Dview has Real-time Preview of presets;
when you browse a preset, the preview window will show the effect of applying
that preset to your currently loaded volume dataset. You can create customized
presets. Presets can store more than just window width and level; all imaging
parameters can be saved in a preset allowing for highly customized capabilities.
©MILLENSYS 89 / 136
All rights reserved
Vision Tools
User Manual
5.6.6.2 Understanding GUI
©MILLENSYS 90 / 136
All rights reserved
Vision Tools
User Manual
©MILLENSYS 91 / 136
All rights reserved
Vision Tools
User Manual
User should select the colon by clicking on any part of the colon with the left
mouse click while holding the control button
The colon then will be selected automatically as shown in the above image.
The user can repeat the same procedure to manually add unselected parts of the
colon
When colon is completely selected the user should click Accept Path.
5.6.6.4 Fly-Through
This option allows the user to move through the bowel looking for abnormalities as
if performing a normal colonoscopy.
©MILLENSYS 92 / 136
All rights reserved
Vision Tools
User Manual
1- Defining the volume (Prone, Supine or Both)
2- Defining the Direction of the Fly-Through (Anus-Caecum or Caecum-Anus)
3- Defining the speed
Then click
User can add annotations to the generated Virtual Colonoscopy preview as well as
using the ruler tool to measure needed measurements
(Please refer to the below image).
©MILLENSYS 93 / 136
All rights reserved
Vision Tools
User Manual
5.6.7 Capturing Reconstruction
Using Fly-Through capture the user can capture the resulted Virtual Colonoscopy in
the form of cine or a series of images
To start capturing the user should click the Capture tab in the tool box, then the
user should select whether to save output as Images or Cine.
User can also define the frame rate of the output cine in frames per second.
©MILLENSYS 94 / 136
All rights reserved
Vision Tools
User Manual
The user can also capture all the views presented on the monitor by checking in
the Capture All Views
Check box, the result will be similar to the below image.
©MILLENSYS 95 / 136
All rights reserved
Vision Tools
User Manual
5.6.8 Vessel Analysis
Vessel Analysis Module is One of MillenSys Products, Vessel Analysis for CT comes
with new analysis & follow-up tools (thrombus segmentation, pre/post abdominal
stent analysis), and facilitates workflow with new tools like aorta & iliac auto
tracking or one-click vessel analysis. Decrease your runoff processing time by 75 %.
©MILLENSYS 96 / 136
All rights reserved
Vision Tools
User Manual
5.6.8.1 Gallery
3Dview application module comes with a wide variety of presets that store the
settings used to create the 3D images. 3Dview has Real-time Preview of presets;
when you browse a preset, the preview window will show the effect of applying
that preset to your currently loaded volume dataset. You can create customized
presets. Presets can store more than just window width and level; all imaging
parameters can be saved in a preset allowing for highly customized capabilities.
©MILLENSYS 97 / 136
All rights reserved
Vision Tools
User Manual
• N.B.: In order to obtain best result for vessel analysis, follow these
recommendations :-
©MILLENSYS 98 / 136
All rights reserved
Vision Tools
User Manual
5.6.8.3 Segmentation
©MILLENSYS 99 / 136
All rights reserved
Vision Tools
User Manual
3. Choose any view from the left side and enlarge it (By double click on the view),
as from this enlarged view we will choose our start point and End point of the
vessel.
In our Example we will work on Coronal view. In some studies we need to do minor
changes in Threshold in order to isolate vessel from bone.
4. Click on Add Point to start selecting Vessel (start and End points), then do Mouse
Left Click at the Start Point then scroll up or down and Mouse left click at the End
point, Notice that when you are selecting tracking points on the vessel, there is a
vertical view for the vessel is being created.
After applying measurements on vessels, users can capture resultant image from
Capture TAB (Check image below).
Save Report
Click this button to save the report partially (Report work flow)
Verified Button
Click this button to verify the report (Report work flow ).
Image Tab
Click this tab to view a thumbnail of the images or key images to add it to the report.
Templates Tab
Click on this tab view, add, edit or delete the templates of the report.
Reports tab
Click this tab to view the status of the report for the study.
Search for the template desired either by scrolling between them or by writing its
name or part of its name and system will filter the list of templates
To use the template double click on it, to load in the word editor
In Order Merge template to the existing word editor content select the template
you need to merge its content, then press right click then press Merge Template
to Current
Note: Voice recording module could be integrated with Phillips Mike and foot
pedal.
For recording and transcript functions
In order to start recording either press on Record button, or keyboard shortcut can
be assigned from the profile settings in order to start recording.
Once start recording, recoding minutes counter will be displaying total time during
your recording session
Once finished press Stop Button, system will then upload the voice note file to the
server and refresh the list with new Voice Note line, with its time
Recorded voice could be edited after saving where user can play the recorded voice,
stop at the point from which record needs to be edited, start recording over it.
Each saved version is saved by the name of the User who saved this version, time
and the status of the report while saving this version
To view these version check the reports grid, as the left there is Versions button
Use Microsoft Word, checked to use Microsoft Word Editor as our word processor
if not checked system will use built in word processor
Download Templates (Automatic at startup), this option will cache list of templates
assigned to the login user at the startup of the application, if not enabled the
template section will be empty list
Download Study Status (For Report Status like study status), this will download the
translated names of the Partial\Complete Not Verified\Verified
Download Signature, when this option is enabled, at the startup of the application
system will download user’s signature, the signature is used in the Verify Process
Prompt to save while template navigation, this option will save the current written
report incase user start loading another template to avoid losses of written report.
Template Override Current Report, using this option if template is loaded it will
replace curent written report
Remove report temp file if the report uploaded successfully, option to save local
computer storage area, once report was uploaded successfully system will delete
the local cached report Default Location for local cache is
C:\Millensys\MiWebView\Reports
Make Mireport Application always on top, this will make the application to be on
the top of all other application on the user desktop.
Send Report as HL7(in case of HIS Integration), option in case user need to send
the verified report to another third party application such as HIS or PACS system
Prompt on Re-Printing Printed Reports, this option is enabled it will notify user
when printing already printed report, that the report was already printed
Default Report file type, This defines the Microsoft Word File type used incase we
have MS Word 2003 and Less, Only valid format is .doc, starting from 2007 version
we can use .docx version, please take in consideration that if you have only one PC
Patient Panel Color, This is the background color under the patient name in the
main report interface
Speed buttons Panel, This number of rows for the toolbar tools on the main report
interface, One Row is the default option, it can be increased
Active Monitor, This is the default monitor user needs the report to start on,
specially designed incase local computer having multiple monitors attached to it
Enable History Mode, This option will enable the prior reports tab, once enabled a
new tab will be displayed after next startup of the application, showing the patient
old written reports
Prompt if
the printing
report is not verify, this option will notify user if he try to print a report which is
not saved yet
After print close file and return to patients list, this option is enabled once print
report tool will be minimized and give user ability to return again to launching
system with website or another application
Save before print, this option will trigger the save option before printing to avoid
printing unsaved reports
Use default Image Format, if enabled inserted image from clipboard will be
inserted in its default width\height this option is not recommended since
sometimes the pasted images are big
It this option is No, so the image will be resized before adding to the report
Layout, define the image style in the word
Width, define max width for the image, system will calculates the height
Resolution, this is the inserted image resolution in dot per inch (DPI)
In this setup page user can define the Word Page, Custom Setup
Define default header and footer distances from edges
Define either page is Landscape or Portrait (Default portrait)
Report Tab,
Report Header
In this setup screen user can define the Patient Header, header displayed at the
top of the report
The settings are divided into three sections
- Text before Patient Data
Insert Line before header, if this option is enabled a line will be drawn using the
Height settings, and color settings
Insert File, this settings is used to put logo or preceding text at the top of the
header, what is written in the textbox is the location where the file of the header
should be located at the local computer, make sure that the file is always located
in this location on all PCs is used by this user
Text, in this text box user can but instead of file some text to be added at the top
of the header in order to have enter some text and you need new line please put
#13 it will be replaced by New Line in the header
Ex:
If we need to make report text as follows
Radiology Department
Patient Report
So this will put text and add line after the text before the header
Insert Patient header automatically, if this option is enabled the system will insert
header at the top of the report, if not enabled nothing will be inserted at the top
Headers and footers different first page, if enabled first report page header will
have the header and other pages will not contains the report header (this is not
recommended)
Patient Info in table\Patient Info In Lines, select which is desired either to display
patient data in the header in a table view or just write data in lines and no outer
table desired
In case Patient Info in table is enabled some settings can be then defined
Show Inner Lines, Either displays Inner border for the table
Show Outer Borders, either display table outer border of not
Table auto Fit behavior, either fit content to the table or fit table to the window
% Width of the first column, this is the percentage of the first column, this is used
if the information displayed in the first column is larger in width to the data
displayed in the second one
Background color, is the color of the table background
Table Style, these settings is to apply table styles in Microsoft word, when style 1 is
the first table template style in Microsoft word
In case of Patient Info in Lines is enabled some settings can be then defined
Shift Header Content by, this is no of pixels to shift header with while displayed in
lines this is used to center the header at the middle of the page
Line Spacing, this define the spacing between patient header lines
Insert Line after header, if this option is enabled a line will be drawn using the
Height settings, and color settings
Text, in this text box user can add after the header and you need new line please
put #13 it will be replaced by New Line in the header
Ex:
If we need to make report text as follows
Radiology Department
Patient Report
Study Date, This option define the way to display study date, either display day and
time of just the Day Only
Hide ref. phy. Field if no data in it, this option when enabled, if the report data
does not include referring physician name, so the filled will be displayed in the grid
as empty filed, or it can show the Ref.Phy. Prefix with no data after that
Show warning Message for Empty Header Content, this option will give warning
message to the user when creates new report for the patient, incase settings of
the header is enabled but no field was defined to be inserted automatically
Defined report font settings such as , Paragraph Alignment, Text Direction, Text
Font
Lines before text, no of lines added at the top of the body before the text
Write “Not Approved” word while printing “Unverified Report” , this will write the
word not approved at the end of the report, just before printing and it will be
removed after wards
Same as report header, you can define fields that can be displayed in the Footer
Section, define what should be added in the Footer Section
Insert Page Number, this will automatically insert page number in footer section
Insert Signature, if enabled system is asking when to add signature, you can define
ether by Verify Action or with Complete Not Verified and with Partial Save
In these settings, will pressing print, does system automatically print all report
pages or just select the current page or custom selection by the mouse
Save temporary voice notes in voice dictation tool, if enabled system will just save
voice notes and it will not be saved on the server, I will be just saved in the location
defined
F/R Step, this is the playback speed factor, 1 is default without slowness or
speediness
Manage Toolbars
To manage toolbar click Options and select Customize Toolbar
Just drag and drop required function from “Actions” menu to the toolbar
General Tab,
Cache directory; This Path for cache Studies on local PC used when open the
case in advanced viewer
Empty cache directory now; press this button to trigger the Clear the Cache
directory immediately
Use Fast Local Mode; this option to retrieve images using shared network
access it is faster mode of accessing the file, it is enabled by default, it is only
active if you are using Advanced Viewer(Local) Mode in the user login
Smart switch to Remote Mode When local Network Can't Reached, this option
is used to dedicate if you left your location inside the local network, so system
will start getting files using normal http communication protocol.
General
Load Thumbnails with Image Tool; this will load thumbnails for the series
when open new study for viewing.
Show Slider (Track bar) on each Image windows; this option will enable the
trackbar slider with every windows inside the image tool, this bar is used in
browsing of images and also used in case of cine loops to control the preview
and scroll in the cine loop.
Cache Image in memory; if this mode is enabled, images after loading them in
the viewer they will be cached in the PC memory, in order to be accessed
quickly if I will browse them continuously in this session, It enabled whenever
the image will be request again it will be recalled from local cache directory
Lock WW/WL; if enabled it will make default settings for WW\WL is locked,
user can change this settings from the viewer after opening the studies.
Center WW_WL if LUT values exists; This will auto adjust the ww\wl once open
an image if it was not provided in DICOM file.
Lock Zoom; if enabled it will make default settings for Zoom\Pan Option is
locked, user can change this settings from the viewer after opening the
studies.
Reverse Series Sort; normally series are displayed sorted by series numbers
from small number to large number, if enabled this will reverse the order
Automatically Split Multiphase series (CT only), this will automatically split CT
cases for Multiphase cases.
Auto Detect MPR Module; if enabled system will check PC graphics card , if it
have good capabilities to launch our MPR module it will be launched
Use enhancement MPR Module; this option will force usage of enhanced MPR
module instead of using basic MPR module
Person Name Encoding; using this module to change the way of displaying the
patient name either display it First Middle Last, or Leave it at it is in Dicom
Files
Auto Fit on Load; if enabled image will be displayed fitted not zoomed or
panned
Smart Mammo Image Fit on Load; if enabled mammo cases (Modality Type
MG), related to special modalities to remove and trim spaces on the outer side
spaces
Auto Load Image When Change Windows; load image when Increase or
Decrease Number of Windows
Display Image Overlay; show Dicom displayed diagram (Dicom LUT) if exists in
the Dicom Image
Load to selected Images only; this will load image on the selected images only
not on all the displayed images
Hide Image Borders; this will hide gray frame outside the images which
contains the slidebar and some buttons
Select All windows after open; select all window of image tool, to be used
directly with mouse actions once the image tool was loaded
Keep WW/WL; Zoom and flip when series change ,if make change on images
(WW/WL,Zoom and flip) and scroll in new series last configure ww\wl and
zoom\pan configuration still not change on displayed images
Use PT Specific WW/WL for PT Image; when open PT Image load with Specific
Width and window level
Info Back Color; this will select the default background color of image information,
incase it is requested to be not transparent
Font Types; normal the image information, is displayed in three levels of fonts,
from this settings user can define the setup of each of them
Image Tab->Measurements
Measurement Text; this option to change text Font, background and Color
Link style; Draw Link Line in the center of textbox or nearest corner of the textbox
Automatic Font size; Adjust font size automatically relational to the image size
Image Scale option; Select either we need to display the scale ruler on which side
of image , also we define line width; Line shift percentage from the border lines of
the image
Home Directory; This Path For folder include Body Mark Icons
Image Output Format; specify default output image format if we scan an image.
Image Output Bits Per Pixel; Define the resolution of the scanned image bits per
pixel, normally it is not more than 12 bits per pixel
Image->Image Render
Settings used for exported images during save display as\ Image Printing
Enable frame around Annotation; incase annotation while printing is not visible
this will make a border around it to show the content “this will be slow operation”
Select background color; incase background frame enabled , select its color from
this settings
Enable Inset On Top; this will always make inset to be displayed over the image
incase we exported image including inset displayed to it.
User from this setup can delete film printing formats that is no longer needed by
the user
Select the format , press delete
Inorder to add format, select the format from combobox, and press add button to
add it to the current selection