JMDH 14 125
JMDH 14 125
JMDH 14 125
1
Mulugeta Teshome Background: Currently, the use of clinical laboratory tests is growing at a promising rate
Abebaw Worede 2 and about 80% of the clinical decisions made are based on the laboratory test results.
Daniel Asmelash 2 Therefore, it is a major task to achieve quality service. This study was conducted to assess
1
the magnitude of errors in the total testing process of Clinical Chemistry Laboratory and to
Department of Medical Laboratory,
Dessie Comprehensive Specialized evaluate analytical quality control using sigma metrics.
Hospital, Dessie, Ethiopia; 2Department Methods: A cross-sectional study was conducted at Dessie Comprehensive Specialized
of Clinical Chemistry, School of Hospital Clinical Chemistry Laboratory, Northeast Ethiopia, from 10 February 2020 to
Biomedical and Laboratory Sciences,
College of Medicine and Health Sciences, 10 June 2020. All Clinical Chemistry Laboratory test requests with their respective samples,
University of Gondar, Gondar, Ethiopia external quality control and all daily internal quality control data during the study period
were included in the study. Data were collected using a prepared checklist and analyzed
using SPSS version 21.
Results: A total of 4719 blood samples with their test requests were included in the study. Out of
145,383 quality indicators, an error rate of 22,301 (15.3%) was identified in the total testing
process. Of the total errors, 76.3% were pre-analytical, 2.1% were analytical and 21.6% were
post-analytical errors (p<0.0001). Of the total 14 analytes in the sigma metric evaluation, except
ALP, all routine clinical chemistry tests were below the standard (<3). In multivariate logistic
regression, the location of patients in the inpatient department was significantly associated with
the specimen rejection ((AOR=1.837, 95% CI (1.288–2.618), p=0.001).
Conclusion: The study found a higher frequency of errors in the total testing process in the
Clinical Chemistry Laboratory and almost all test parameters had an unsatisfactory sigma
metric value.
Keywords: analytical errors, Clinical Chemistry Laboratory, post-analytical errors, pre-
analytical errors, sigma metrics, Ethiopia
Introduction
The use of clinical laboratory test results in clinical decisions has become an
integral part of clinical medicine and studies showed that laboratory medicine
determines 70% of clinical decisions; however, small variations around this figure
(60–80%) were reported.1,2 The quality laboratory service ensures accurate, precise
Correspondence: Daniel Asmelash and timely results. The total testing process (TTP) of a laboratory is a complex
Department of Clinical Chemistry, School procedure that includes three phases: the pre-analytical, analytical and the post-
of Biomedical and Laboratory Sciences,
College of Medicine and Health Sciences, analytical. In all phases of TTP, quality indicators (QIs) are used by health
University of Gondar, P.O Box 196, laboratories based on the requirements of the International Organization for
Gondar, Ethiopia
Email daniel.asmelash111@gmail.com Standardization (ISO) 15,189:2012.3,4
submit your manuscript | www.dovepress.com Journal of Multidisciplinary Healthcare 2021:14 125–136 125
DovePress © 2021 Teshome et al. This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.
http://doi.org/10.2147/JMDH.S286679
php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the
work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For
permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
Teshome et al Dovepress
Laboratory errors can occur at any stage of the pre- a major issue. For example, studies in Ghana revealed
analytical phase to the post-analytical phase of the TTP.5 unsatisfactory sigma level (<3) for all analytes.15
Pre-analytical errors include all errors that occur prior to Laboratory personnel should control the whole TTP
analysis. Some of the pre-analytical errors include hemo using QIs and focus on improving extra-laboratory proce
lyzed sample, insufficient sample, incorrect label, incorrect dures, in particular test selection and interpretation in
requisition, clotted sample and tube broken in centrifuges. coordination with other medical staff.16
The impact of pre-analytical error occurs in the analytical The achievement of sustainable laboratory perfor
and the post-analytical stage. At the analytical phase mance in accreditation has also been a major challenge
where analysis takes place, non-conformity with quality in African countries and there are still gaps in strength
control, calibration failure, random and systematic errors ening laboratory services.17 According to previous stu
can be occurred.6 Common post-analytical errors include dies, the majority of Ethiopian clinical laboratories
failure to report test results, delay in reporting, incorrect provide sub-optimal service and the quality of the ser
calculation, critical results not reported or delayed, and vice is severely compromised, which may lead to the
results sent to the wrong patient.7 occurrence of several laboratory errors.18 Despite very
Sigma metric quantifies the analytical performance of few studies on the overall distribution of laboratory
a laboratory process as a rate of Defects-Per-Million errors in Clinical Chemistry, no studies have yet been
Opportunities (DPMO). The evaluation of laboratory
conducted to evaluate analytical performance using
errors in terms of sigma metric is more meaningful than
sigma metric in Ethiopia.19,20 Therefore, the current
the number of defects alone. For the analytical process of
study will assess the distribution of errors in each
the laboratory system, the sigma metric analysis identifies
phase of the clinical chemistry testing process and the
errors in quality indicators of the process and provides
analytical performance of routine chemistry tests using
error corrections based on results. Using the Six Sigma
sigma metric.
principles, it is possible to assess the quality of laboratory
testing processes and the number of quality controls
needed to ensure the desired quality.8 Materials and Methods
Attainment of Six Sigma performance represents 3.4 Study Design and Setting
DPMO and the achievement of 3 sigma values is the mini A cross-sectional study was conducted at the Dessie
mum acceptable quality for a process to be applied.9 Comprehensive Specialized Hospital Clinical Chemistry
A higher sigma metric value means fewer analytical errors Laboratory from 10 February 2020 to 10 June 2020.
and fewer acceptable test results are falsely rejected, a lower Dessie Comprehensive Specialized Hospital laboratory
sigma metric value of the parameters indicates higher ana was involved in the Stepwise Laboratory Quality
lytical errors and many acceptable test results are falsely Improvement Process towards Accreditation, although its
rejected, which is more difficult to use in the analysis of performance was not satisfactory.
patient samples. However, low sigma metrics could also
mean that there are high true rejections that may occur as
a result of inadequate QC data and the short study period.10 Study Variables
Laboratory errors that occur at any phase of the TTP The study variables included all phlebotomists, laboratory
can directly contribute to increased healthcare costs, professionals in the Clinical Chemistry laboratory, Clinical
decreased patient satisfaction, delayed diagnosis, misdiag Chemistry test requests with their respective blood samples
nosis and a serious risk to the patient’s health.11 Studies for routine Clinical Chemistry tests, quarterly external qual
showed that 6.4% to 12% risk of inappropriate care and ity control (EQC) and all daily internal quality control
death occurs due to laboratory errors.12 Studies in the (IQC) data from the Clinical Chemistry laboratory during
United States indicated that diagnostic errors occur in the specified study period. Therefore, the study covered the
5% of the outpatients and about half of the errors may pre-analytical, analytical and post-analytical phases of the
cause severe harm to patients. This is supported by a study TTP in the Clinical Chemistry Laboratory. However, body
conducted in Malaysia with 3.6% errors.13,14 fluids other than venous blood samples and not routine
Although analytical errors contain fewer errors than clinical chemistry tests were excluded from the sigma
pre-and post-analytical phases, analytical quality is still metric evaluation.
Data Collection Procedure of errors in the Clinical Chemistry Laboratory was sum
A standardized pre-tested assessment checklist was used to marized by descriptive statistics using frequency tables.
evaluate the pre-analytical, analytical and post-analytical The sigma metrics value of each analyte was summarized
phases of the Clinical Chemistry Laboratory. The checklist in tables after extracting Total Allowable Error (%TEa)
was prepared using variables from different studies and value of each analyte from Clinical Laboratory
International Federation of Clinical Chemistry.21 The Improvement Amendment (CLIA).23 A chi square test
selected data collectors were trained on how to collect all was used to check the presence of statistical difference
the necessary data in the assessment of the completeness between the three phases of Clinical Chemistry testing
of the standard laboratory request forms (such as, patient process.
age, gender, signature of physicians, location, date and In addition, the chi square test was also used to assess
authorized request formats), in the assessment of specimen the association between selected independent variables
quality, analytical and post-analytical QIs based on the with pre-analytical and post-analytical errors. Bivariate
checklist. and multivariate logistic regression were used to assess
All required data were collected using a checklist in all the association between work shift and location of patients
phases of the TTP. In addition, face-to-face interviews with prolonged turnaround time (TAT). The strength of the
were also used to collect socio-demographic data. Data association was measured using the adjusted odds ratio
on pre-analytical variables, specifically laboratory speci (AOR) and 95% confidence interval (CI). The p-value
mens, were collected through observation. Color charts <0.05 was considered statistically significant. The chi
were used to assess the quality of the specimens, which square test and logistic regression were conducted for
were used to standardize the reports of each data collector. selected independent variables (work shift and patient
Furthermore, a routine automatic biochemistry analy location) because the assumptions were not fulfilled by
zer (DIRU CS-T240, Dirui Industrial Co., Ltd. China) was other independent variables.
evaluated by sigma metric and important data generated by Analytical quality control performance for routinely
tested 14 analytes [aspartate transaminase (AST), alanine
the analyzer were extracted by using observation. Daily
transaminase (ALT), alkaline phosphatase (ALP), urea
IQC data collected during the analytical phase were
(UR), creatinine (CRE), glucose (GLU), total cholesterol
entered in the checklist and record formats. The laboratory
(TC), triglyceride (TG), high-density lipoprotein (HDL),
was involved in the EQC program by analyzing five dif
low-density lipoprotein (LDL), total protein (TP), albumin
ferent concentration proficiency test samples provided by
(ALB), total bilirubin (TBIL) and uric acid (UA)] was
one world Accuracy and the most recent quarterly EQC
evaluated using a sigma metric quality monitoring tool.
laboratory performance report was taken and transferred to
The TEa source for each of the analytes was CLIA. The
the checklist. Post-analytical QI outcomes were collected
bias of all analytes except LDL was calculated using the
before the results were delivered to clients, and each vari
performance of the Laboratory in the EQC and the peer
able was entered as a record in the checklist.
groups mean. Since LDL was not properly documented in
the EQC documentation chart, its bias was calculated
Data Quality and Statistical Analysis using the observed mean from the daily IQC and the target
The standardized data collection checklist was pre-tested value from the manufacturer. The CV for all analytes was
on 30 venous blood specimens with their corresponding calculated from the observed mean and SD of the daily
test requests for Clinical Chemistry at Boru-Meda Hospital IQC in the study period. For all analytes, two levels
to check its feasibility and ensure the validity of the study (normal and pathological) of IQC materials were used,
tool.22 The checklist was modified based on the pre-test and sigma metrics value was calculated separately. The
results. The investigator undertook a continuous monitor estimates of sigma metrics were optimistic because the
ing and support to ensure complete, consistent, clear and quality control materials were not from third parties. The
accurate data. cause of low sigma value (<6) in all parameters was
The data were checked for completeness and reliability. identified using the calculated Quality goal index (QGI).
After ensuring this, data were entered to EPI info version 7 The sigma metrics was calculated with the Sigma =
and transferred to SPSS version 21 (IBM Corporation, (TEa – Bias%)/CV. In addition, the QGI ratio was calcu
Armonk, NY, USA) for analysis. The overall distribution lated using a Bias%/1.5 × CV%.24 A sigma value between
3 and 4 quality requires a multi-rule procedure 13S/22S/R4S Table 1 Frequency of Missed Information on Laboratory
/41S/10x, with four levels of control measurement in two Request Forms
runs (N=4, R=2) or two levels of control measurement in S. No. Variables Missed
four runs (N=2, R=4). However, <3-sigma cannot be con Information
but the remaining 3142 (66.6%) test requests were with Abbreviation: MRN, medical record number.
Total pre-analytical errors from all pre analytical QIs (Including laboratory test 18(17,017/94,380)
requests form and specimens)
Abbreviation: QIs, quality indicators.
%(Error/Total)
the second observed error was 4827 (21.6%) in the post- missed information on Clinical Chemistry Laboratory
analytical phase, but the lowest frequency 457 (2.1%) was request formats (p<0.0001), hemolysis (p=0.04), icteric
identified in the analytical phase (Table 5). (p<0.0001), specimen rejection (p=0.001) and failure to
record results (p=0.01) (Table 6).
Location of Sample Collection with Pre-
and Post-Analytical Errors Factors Associated with Prolonged TAT
Using the chi square analysis, the location of sample The multivariate logistic regression showed that pro
collection was found to be statistically associated with longed TAT was independently associated with
%(Error/Total)
Yes No
the second shift (AOR=8.354, 95% CI (4.453–15.670), Sigma Value and Quality Goal Index
p< 0.0001) in which the occurrence of prolonged TAT (QGI) of the Clinical Chemistry Tests
was 8 times higher in the second shift workers com Using CLIA as a source of TEa, the sigma metric values of
pared to the first shift workers (Table 7). 14 routine analytes were measured. Based on the measured
Patient age missed 9 144 116 763 21 524 1125 2017 392*(p<0.0001)
Gender of patient missed 5 148 72 807 17 528 1044 2098 432*(p<0.0001)
Signature of the physicians missed 69 84 599 280 318 227 2672 470 353.5*(p<0.0001)
Location of the patient missed 5 148 6 873 8 537 3103 39 4463.5*(p<0.0001)
Date of ordered missed 62 91 335 544 161 384 2157 985 487.5*(p<0.0001)
Unauthorized requests 0 153 9 870 1 544 70 3072 17.8*(p<0.0001)
Hemolyzed specimen 6 147 17 862 21 524 67 3075 8.3*(p=0.04)
Icteric specimen 4 149 9 870 23 522 61 3081 17.77*(p<0.0001)
Mislabeled specimen 1 152 4 875 3 542 37 3105 5.06(p=0.167)
Specimen rejection 9 144 31 848 43 502 140 3002 15.9*(p=0.001)
Unrecorded results 1 152 21 858 9 536 31 3111 11.4*(p=0.010)
Results released out of TAT 3 150 14 865 10 535 26 3116 7.67(p=0.053)
Note: *Statistically significant association (p<0.05).
Abbreviations: IPD, impatient department; OPD, outpatient department; TAT, turnaround time.
Yes No
values, only one analyte (ALP) had above 3 sigma values than studies in Nigeria,28 Kenya29 and Uganda30 with an
and the remaining 13 analytes were found to be below 3 incompleteness rate of 14.3%, 22.7% and 17.9%, respec
sigma values. The lowest sigma value (0.01) was observed tively. This disparity may be due to variation in quality
in ALB (level-I). In addition, the calculated QGI showed indicators, study designs, study periods, awareness and
that the presence of imprecision, inaccuracy and both experience of clinicians on the value of patient
imprecision and inaccuracy. It was found that the majority information.
of the poor performance of analytes in the sigma value was Of the total number of QIs, 39.3% showed missed
due to the imprecision of the QC (Table 8). information on Clinical Chemistry test requests. This find
ing was higher than the study conducted in Ethiopia31 and
Nigeria28 with an overall incompleteness of 8.7% and
Quality Control Strategy Based on Sigma
10.5%, respectively. This large variation may be due to
Values differences in sample size, operational definition, test
Of all the analytes, only ALP had a sigma value of >3. The
requests ordered by inexperienced staffs and lack of com
Westgard rule was established for ALP only and the
mitment to complete the required information.
remaining analytes showed poor performance in which
Based on our study, the clinical history of the patient
the methods should be rechecked prior to the analysis of
was not stated in all 4719 (100%) requests which were
patient samples. The Westgard multi rule selected for ALP
consistent with the study conducted in Gondar, Ethiopia
was 13S/22S/R4S/41S/10x. Although a multi rule was
with 99% incompleteness,20 but higher than studies con
applied for ALP, the QGI indicated an imprecision that
ducted in Addis Ababa, Ethiopia (72.6%).19 In addition,
requires a close monitoring and troubleshooting in the
signature of physicians (77.5%) and location of patients
daily IQC (Table 9).
(66.2%) information were also incomplete in the test
request. These findings showed a wide difference from
Discussion previous studies conducted in Gondar (38.7% and
Although most modern medical laboratories have joined 1.8%),20 Addis Ababa (30.4% and 1.1%)19 and Nigeria
automation, it remains a challenge to ensure accurate and (19.8% and 20.1%)32 respectively. The higher incomplete
precise laboratory results.26 Therefore, the current study information in the current study may be due to lack of
used relatively comprehensive QIs to assess the total errors commitment, failure to create awareness and workload.
in the Clinical Chemistry Laboratory. In this study, hemolysis was the major cause of sample
In the current study, the majority of the pre-analytical rejection (35.4%) which was inconsistent with studies
errors found were incompleteness of the required informa conducted in Ghana (17.6%).33 The increased percentage
tion on the test request form (39.3%). This finding was of hemolysis may be due to untrained phlebotomists, high
consistent with the study conducted in Hawassa, workload and lack of regular monitoring and support.
Ethiopia27 (39.4%), but lower than a study conducted in According to this study, the prevalence of sample
Gondar, Ethiopia20 with an overall incompleteness of rejection was 4.7%, which was consistent with studies in
49.9%. In contrast, the current finding was much higher Egypt34 with a rejection rate of 4.6%. However, it was
Table 8 Total Allowable Error (TEa), CV, Bias, Sigma Value and QGI of the Routine Clinical Chemistry Tests
Analyte IQC Level TEa% Average Bias (%) CV (%) Sigma QGI Problem
Table 9 Quality Control Strategy Based on the Sigma Value of the Test Parameters
Parameters Sigma Levels of Run Westgard Rules Status of the Decision
Metrics Control Method
AST, ALT, UR, CRE, GLU, TC, TG, HDL. LDL, <3 – – – Poor Needs
TP, ALB, TBIL, UA rechecking
higher than studies conducted in India (3.45%),35 South rejection may be due to untrained phlebotomists, lack of
Africa (1.46%),36 Turkey (0.65%),37 Saudi Arabia cooperation and communication, incompetence, improper
(2.07%)38 and Ethiopia (1.4%).39 The higher rate of processing of specimens.
In the current study, the total pre-analytical error (18%) each test parameter showed that analytical quality is still
was lower than studies conducted in Iraq (39%)40 and a major issue.
Egypt (43.7%),41 but it was much higher than studies in In the current study, the sigma metric value of most
Ghana (3.7%),33 Saudi Arabia (3.15%),42 Tunisia analytes was unsatisfactory. ALP (Level-I (3.9) and II
(7.7%),43 Greece (1.94%)44 and India (0.15%).45 This (3.8) IQC) was the only analyte with a sigma value
may be due to differences in the QIs, sample size, labora above 3. The remaining 13 analytes (AST, ALT, UR,
tory facilities and the experience of health professionals. CRE, GLU, TG, TC, HDL, LDL, TP, ALB, TBIL and
Our study found a lower frequency of 6.2% IQC failure UA) had low sigma values <3. This finding was incon
compared to the study conducted in Pakistan (32%),46 but sistent with a study conducted in China51 that reported >3
higher than the study conducted in Ethiopia (2.95%)19 and sigma values for all test parameters (AST, ALT, ALP, UR,
India (0.6%).47 The difference between findings may be due to CRE, GLU, TC, TG, HDL, LDL, TBIL, UA, TP and
variation in equipment and laboratory personnel performance, ALB). This may be due to differences in the performance
storage condition of QC materials, the reconstitution of lyo of analyzer, poorly stored reagents and QC materials,
philized QC materials and availability of in-house mean/SD. study period, failure to perform regular maintenance of
The current study found a 23.1% failure of the proficiency analyzers and competency of laboratory staffs. In addition,
testing, which was inconsistent with a study conducted in the calculated QGI showed that the presence of impreci
Pakistan (5.4%).48 This higher level of non-conformity with sion, inaccuracy and both imprecision and inaccuracy.
EQC proficiency testing may be due to improper transport/ This may be due to power fluctuation, incorrect calibra
tion, improper reconstitution of QC/calibration materials
storage of EQC samples, improper reconstitution of the sam
and interferences.
ples, analyzing by faulty instruments, untrained and incompe
The current study found <2 sigma value for 11 test
tent laboratory staffs, reporting with unacceptable unit.
parameters at two IQC levels for (UR, CRE, GLU, TG,
In the current study, 99.5% of the test results were not
TC, HDL, LDL, TP, ALB, TBIL and UA). This result was
verified and signed by independent reviewers. This finding
consistent with a study conducted in Ghana15 for (GLU,
was far higher than the study conducted in Kenya
TG, TC, HDL, UR, CRE, TP and AST) and in Sudan52 for
(13.1%).49 This may be related to non-committed labora
UR and CRE. It was also comparable to the study con
tory staffs, the workload of the laboratory and failure to
ducted in India53 for ALT (Level-I), TBIL (Level-I), CRE
adhere to the laboratory quality policy. In addition, 1.1%
(Level-I) and UR (Level-I and II). In contrast, higher
of laboratory results were reported out of the established
sigma values >3 were reported in China51 and Turkey54
TAT, which was higher than the frequency of delayed TAT
for ALB, GLU, UA, TC, TG and CRE. This disparity in
in Pakistan (0.003%).48 The prolonged TAT in this study
the sigma metrics value may be due to differences in
may be due to the unavailability of Laboratory Information analytical methods, different IQC materials, different pro
System (LIS), electric interruption, equipment malfunction ficiency testing bodies and manufacturers.
and stock out of distilled waters.
The current study found that the prevalence of errors in Conclusion
the TTP was 15.3%. This finding was lower than the study The study found a high frequency of errors in the Clinical
conducted in Addis Ababa, Ethiopia (33.1%)19 but higher Chemistry Laboratory and most of the errors were observed
than the study conducted in Saudi Arabia (4.35%)42 and in the pre-analytical phase. The majority of pre-analytical
Ghana (4.7%).33 This may be due to difference in the errors were significantly associated with the location of the
infrastructure of the laboratory, regular monitoring and sample collection. In addition, with the exception of the ALP,
evaluation, availability of functional LIS, competency of the sigma metric values for all routine clinical chemistry tests
laboratory staffs, in-house mean/SD, variation in the study were below the standard (<3). Although the sigma metric
designs, operationalization of variables and QIs. value of ALP was within an acceptable range, a multi-
Although IQC ensures a continuous monitoring of the Westgard rule (13S/22S/R4S/41S/10x) should be implemented
analytical method, the exact number of errors that under close monitoring rather than a single rule.
occurred during the analytical phase cannot be Specimen collectors, particularly in the IPD should be
evaluated.50 As a result, an effective quantifying sigma trained in the laboratory quality management system.
metrics tool was selected and the sigma metrics value for Besides the existing manual recording system, the
14. M.V R. Are diagnostic and laboratory errors causing fatal deaths? 31. Addis Z, Wondimagegn T, Tachebele B. Types and frequency of
Int J Curr Adv Res. 2017;7(1(A)). pre-analytical errors at University of Gondar hospital laboratory.
15. Afrifa J, Gyekye SA, Owiredu WK, et al. Application of sigma Elective Med . 2015;2(4):363–365. doi:10.18035/emj.v2i4.160
metrics for the assessment of quality control in clinical chemistry 32. Oladeinde BH, Omoregie R, Osakue EO, Onifade AA. Evaluation of
laboratory in Ghana: a pilot study. Niger Med J. 2015;56(1):54. laboratory request forms for incomplete data at a rural tertiary hospi
doi:10.4103/0300-1652.149172 tal in Nigeria. N Z J Med Lab Sci. 2012;66(2):39.
16. Mrazek C, Lippi G, Keppel MH, et al. Errors within the total 33. Sakyi A, Laing E, Ephraim R, Asibey O, Sadique O. Evaluation of
laboratory testing process, from test selection to medical analytical errors in a clinical chemistry laboratory: a 3 year
decision-making - A review of causes, consequences, surveillance experience. Ann Med Health Sci Res. 2015;5(1):8–12. doi:10.4103/
and solutions. Biochem Med. 2020;30(2):020502. doi:10.11613/ 2141-9248.149763
BM.2020.020502 34. Rizk MM, Zaki A, Hossam N, Aboul-Ela Y. Evaluating laboratory
17. Mate KS, Rooney AL, Supachutikul A, Gyani G. Accreditation as key performance using quality indicators in Alexandria University
a path to achieving universal quality health coverage. Global Health. Hospital Clinical Chemistry Laboratories. J Egypt Public Health
2014;10(1):68. doi:10.1186/s12992-014-0068-6 Assoc. 2014;89(3):105–113. doi:10.1097/01.EPX.0000453262.8
18. Girma M, Desale A, Hassen F, Sisay A, Tsegaye A. Survey-defined 5383.70
and interview-elicited challenges that faced Ethiopian Government 35. Sushma B, Shrikant C. Study on “pre-analytical errors in a clinical
Hospital Laboratories as they applied ISO 15189 accreditation stan biochemistry laboratory”: the hidden flaws in total testing. Biochem
dards in resource-constrained settings in 2017. Am J Clin Pathol. Anal Biochem. 2019;8(374):2161–1009.19.
2018;150(4):303–309. doi:10.1093/ajcp/aqy049 36. Jacobsz LA, Zemlin AE, Roos MJ, Erasmus RT. Chemistry and
19. Tadesse H, Desta K, Kinde S, Hassen F, Gize A. Clinical chemistry haematology sample rejection and clinical impact in a tertiary labora
laboratory errors at St. Paul’s Hospital Millennium Medical College tory in Cape Town. Clin Chem Lab Med. 2011;49(12):2047–2050.
(SPHMMC), Addis Ababa, Ethiopia. BMC Res Notes. 2018;11 doi:10.1515/CCLM.2011.743
(1):789. doi:10.1186/s13104-018-3893-5 37. Atay A, Demir L, Cuhadar S, et al. Clinical biochemistry laboratory
20. Ambachew S, Adane K, Worede A, et al. Errors in the total testing rejection rates due to various types of preanalytical errors. Biochem
process in the clinical chemistry laboratory at the University of Med. 2014;24(3):376–382. doi:10.11613/BM.2014.040
Gondar Hospital, Northwest Ethiopia. Ethiop J Health Sci. 2018;28 38. Zaini R, Dahlawi H, Siddiqi A. Identification of the types and
(2):235–244. doi:10.4314/ejhs.v28i2.15 frequencies of pre-analytical errors in the clinical biochemistry
21. Plebani M, Sciacovelli L, Aita A. Quality indicators for the total testing laboratory: 1-year study at Hera’a General Hospital. Archiv Med.
process. Clin Lab Med. 2017;37(1):187–205. doi:10.1016/j.cll.2016.09.015 2016;8(4):1. doi:10.21767/1989-5216.1000152
22. Bolarinwa OA. Principles and methods of validity and reliability 39. Tesfaw HM, Segaye A, Hassen F. Frequency of Specimen Rejection
testing of questionnaires used in social and health science and A ssociated F actors at St. Paul’s Hospital Millennium Medical
researches. Niger Postgrad Med J. 2015;22(4):195. doi:10.4103/ College, Addis Ababa Ethiopia. J Multidiscip Healthc. 2015;2
1117-1936.173959 (1):1–16. doi:10.15415/jmrh.2015.21001
23. Westgard JO, Westgard SA. The quality of laboratory testing today: 40. Najat D, Szecsi PB. Prevalence of pre-analytical errors in clinical
an assessment of sigma metrics for analytic quality using perfor chemistry diagnostic labs in Sulaimani city of Iraqi Kurdistan. PLoS
mance data from proficiency testing surveys and the CLIA criteria One. 2017;12(1):1. doi:10.1371/journal.pone.0170211
for acceptable performance. Am J Clin Pathol. 2006;125(3):343–354. 41. Sharaki O, Abouzeid A, Hossam N, Elsherif Y. Self assessment of
doi:10.1309/V50H4FRVVWX12C79 pre, intra and post analytical errors of urine analysis in clinical
24. Kumar BV, Mohan T. Sigma metrics as a tool for evaluating the chemistry laboratory of Alexandria Main University Hospital. Saudi
performance of internal quality control in a clinical chemistry J Health Sci. 2014;3(2):96. doi:10.4103/2278-0521.134863
laboratory. J Lab Physicians. 2018;10(2):194. doi:10.4103/JLP. 42. Pothula Y, Al-Marzooq YM, Salem RA, Al-Jasem W, Al-Hajji A.
JLP_102_17 A retrospective study of quality improvement in clinical biochemistry
25. Mahmood B, Rasheed MK, Khazal A, Rasheed MK. Assessment of laboratory. Parameters. 2016;2017:2018.
sigma metric results of serum parameters of liver and kidney function 43. Bouzid K, Zrelli S, Rebeh A, et al. Preanalytical phase management:
tested by automated chemistry analyzer in Medical City Hospital. identification and control of nonconformities in the sampling room of
Iraqi Acad Sci J. 2018;17(3):307–314. a clinical laboratory in Tunisia. J Lab Precis Med. 2018;3:38.
26. Abdollahi A, Saffar H, Saffar H. Types and frequency of errors doi:10.21037/jlpm.2018.03.14
during different phases of testing at a clinical medical laboratory of 44. Englezopoulou A, Kechagia M, Chatzikiriakou R, Kanellopoulou M,
a teaching hospital in Tehran, Iran. N Am J Med Sci. 2014;6(5):224. Valenti M, Masedu F. Pre analytical errors as quality indicators in
doi:10.4103/1947-2714.132941 clinical laboratory. Austin J Public Health Epidemiol. 2016;3(5):8.
27. Ali M Assessment of medical laboratory request form completeness 45. Tiwari E, Pallipady A, Mishra S. Preanalytical, analytical and post
and non-communicated result to clinicians in the two public hospitals analytical errors in chemical laboratory. Int j Sci Res. 2015;4
found in Hawassa city, Southern Ethiopia: Addis Ababa University; (3):2279–2281.
2015. 46. Sadiq F, Yasmeen F, Mumtaz A, et al. Frequency of errors in clinical
28. Jegede F, Mbah HA, Dakata A, Gwarzo DH, Abdulrahman SA, Kuliya- laboratory practice. Iran J Pathol. 2014;9(1):45–49.
Gwarzo A. Evaluating laboratory request forms submitted to haematology 47. Trivedi P, Shah N, Ramani K Managing clinical laboratories: monitor
and blood transfusion departments at a hospital in Northwest Nigeria. Afr and control lab errors to improve lab performance: Indian Institute of
J Lab Med. 2016;5(1):1–6. doi:10.4102/ajlm.v5i1.381 Management Ahmedabad; 2011.
29. Kipkulei JC, Lotodo TC. Evaluation of the completeness in the filling 48. Jafri L, Khan AH, Ghani F, Shakeel S, Raheem A, Siddiqui I. Error
of laboratory request forms submitted to the haematology laboratory identification in a high-volume clinical chemistry laboratory:
at a Tertiary Hospital in Kenya. Health. 2019;11(7):862–868. five-year experience. Scand J Clin Lab Invest. 2015;75(4):296–300.
doi:10.4236/health.2019.117069 doi:10.3109/00365513.2015.1010175
30. Namwase B Pre-analytical errors affecting quality of final results at 49. Kimengech KK, Waithaka SK, Onyuka J, Kigondu CS.
the pathology laboratory college of health sciences, Makerere Determination of errors that compromise the quality of laboratory
University Kampala Uganda: international health sciences service in a tertiary hospital. Asian J Med Sci. 2017;8(1):64–70.
university; 2018. doi:10.3126/ajms.v8i1.14740
50. Sayeed S, Ganji SB, Mopuri R. A Short-term assessment of routine 53. Adiga US, Preethika A, Swathi K. Sigma metrics in clinical chem
chemistry parameters by sigma metrics and quality goal index ratio in istry laboratory-A guide to quality control. Al Ameen J Med Sci.
a tertiary care hospital laboratory. J Evol Med Dent Sci. 2019;8 2015;8(4):281–287.
(29):2303–2306. doi:10.14260/jemds/2019/505 54. Gülbahar Ö, Kocabıyık M, Çıracı MZ, et al. The use of six sigma
51. Mao X, Shao J, Zhang B, Wang Y. Evaluating analytical quality in methodology to evaluate the analytical performances of clinical
clinical biochemistry laboratory using six sigma. Biochem Med. chemistry analyzers. Turk J Biochem. 2016;43(1):1–8. doi:10.1515/
2018;28(2):253–256. doi:10.11613/BM.2018.020904 tjb-2016-0223
52. Affram K, Udofot O, Cat A, Agyare E. Application of sigma metrics
for the assessment of analytical quality in clinical biochemistry
laboratories in Sudan: a pilot study. Int J Adv Res. 2015;3(Issue
10):859–874.