Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

ADT3000 服务手册 (2)

Download as pdf or txt
Download as pdf or txt
You are on page 1of 18

1

Contents
1. Product Introduce ............................................................................................. 1
1.1 Product Name : ..................................................................................................2
1.2 REF/Model: ....................................................................................................2
1.3 Intended Use: .................................................................................................... 2
1.4 Principle of measurement ................................................................................. 2
2. Specifications ..................................................................................................... 3
2.1 Technical specification ......................................................................................3
2.2 Performance characteristics .............................................................................. 4
3. Package contents ............................................................................................... 4
4. Instruction of the analyzer ............................................................................... 6
5. Installation of the analyzer ...............................................................................7
6. Quick Use Guide ............................................................................................... 8
6.1 Sample Test ....................................................................................................... 8
6.2 QC Test ............................................................................................................11
7. Warning and Handling ................................................................................... 13
8. Daily Maintenance .......................................................................................... 14
8.1 Maintenance .................................................................................................... 14
8.2 Cleaning and storage .......................................................................................14
8.3 Software Update ..............................................................................................15
9. Trouble shooting ..............................................................................................16

1
1. Product Introduce

1.1 Product Name :


Handheld DOA Reader Service Manual

1.2 REF/Model:
ADT3000

1.3 Intended Use:


The Handheld DOA Reader is designed to qualitatively measure drugs of abuse with urine or
saliva sample. The system is intended for IVD use as an aid to monitor the effectiveness of
drug abuse. It must be used with its corresponding drug test strip. For professional in vitro
diagnostic use.

1.4 Principle of measurement


The Handheld DOA Reader adopts image sensor and opitics combination method for the test
strip after loading and then digital image processing of the photo. The intensity of color in the
test region (T) may vary depending on the concentration of analytes present in the specimen.
Therefore, any shade of color in the test region (T) should be considered negative. Please note
that this is a qualitative test only, and cannot determine the concentration of analytes in the
specimen.Test samples include urine and saliva.

2
2. Specifications

2.1 Technical specification


Description Specification
CPU Qualcomm SDM450
Memory 16GB eMMC + 2GB LPDDR3
Operation system Android 7.1
Touch screen 4.3’,480*800,show the operating status and results
Thermal printer Print test results
Testing Slot Location to insert or remove the device
WIFI Connect the WLAN network, support 2.4GHz and 5GHz
For manufacturer's use only
USB port
No charging and data transfer function
Power switch key Power on/off
Power outlet Power supply
CMOS Image Sensor Take a strip photo, 8M pixel
Position the longitude & latitude, support
GPS
GPS/GLONASS/BeiDou
Control card Quality control
Used Indoors
RTC Real time clock
Charging indicator Charging animation indication when power off
Rear camera 5M pixel, take a testee’s photo, scanning QR code, etc
Light source LED
External memory TF Card
Printer Built-in thermal printer
Printing paper 57.5±0.5mm
Data storage 10000 records
Power voltage Li Ion Battery 7.2V 2500mAh
Rated power <36VA
Cassette Size (60~150)mm×(18~20)mm×(5~6.5)mm
On-line operation WIFI
Body 247x98x100 mm

3
Panel holder (92x89x116)mm
Sample holder Urine cup holder(147x77x104)mm
Saliva cup holder(113x85x116)mm
Model: GST36U12
Adaptor Input: 100-240VAC,50/60Hz,0.8A
Output: 12V-3A, 36VA (max)
COM interface Male and female link thermal printer
Ingress Protection IPX0
Overvoltage category CAT II
Pollution Degree II

2.2 Performance characteristics


 Linearity:r ≥0.990
 Repeatability: Relative standard deviation (CV) ≤5%
 Stability: Relative deviation (a)≤±10%
 Accuracy: Relative deviation (B)≤ ±15%
 A complete measurement time < 5s

3. Package contents

NO. Description Quantity photo Unit

1 Host machine 1 Set

2 Panel holder 1 Set

3 Urine cup holder 1 Set

4 Saliva cup holder 1 Set

4
5 Power adapter 1 Set

6 TF Card 1 PC

7 User manual 1 PC

8 Thermal printer paper 1 Roll

9 TF Card reader 1 Set

Positive and Negative QC


Card of Panel(Two-fold)
10 1 PC

Positive and Negative QC


11 Card of Saliva Cup 1 PC
(Two-fold)
Positive QC Card of Urine
12 Cup 1 PC

Negative QC Card of Urine


13 Cup 1 PC

14 Lanyard 1 PC

15 Power plug 1 PC

5
4. Instruction of the analyzer

The Handheld DOA Reader is composed of optical unit, mechanical unit, control unit, display
unit, thermal printer unit, and power key.

Host machine of DOA Reader Panel holder Saliva cup holder Urine cup holder

RS232
6
5. Installation of the analyzer

Plug the instrument in,Keep pushing the power key down until the touch screen show boot
image.
5.1 Power on
Keep pushing the power key down until the touch screen show boot image.
Warning:
If you push the power key more than 5 seconds, the touch screen does not show the boot image, please use the AC adapter
charging the inner Li battery.

5.2 Install printing paper


The Handheld Reader has designed a removable thermal printer. User can remove or install printer manually. The printer
locker in the back of reader, as shown follow;
Put the printing paper into the printer, facing down to the hot sensor of the printer. Leave the rest part of the paper outside
of the door and close the printer door.

5.3 Install test holder


Press and hold the Lock on the top of the instrument and insert the test head into the instrument to complete the assembly.

7
6. Quick Use Guide

6.1 Sample Test

①Power on ②Enter the test APP

③Click System Setting , select Configuration File, and Scanning configuration


files

8
④Enter Standard Test, Fill sample information ⑤Select reagent type

⑥Test a sample
A.Panel Test

B.Saliva cup Test

9
C.Urine cup Test

⑦Change another sided test(Panel and Saliva cup need do this)

⑧Read the results

10
6.2 QC Test
①Scan QC file(QR code in the foil bag of QC Card)

②Select QC model

11
③Insert the QC Card
A. QC Card of Panel

B.QC Card of Saliva cup

C.QC Card of Urine cup

NOTE: Panel and Saliva cup QC card need to be flipped to complete negative and positive
QC test,Urine cup have two cup,negative and positive QC Cup

12
④Click “Start test”and View QC result

PASS

PASS

7. Warning and Handling

 Before using the reader, read this manual and be sure you thoroughly understand its
contents.
 Please follow the instructions in this manual to assure proper operation of the reader.
 When not in use, unplug the reader from the electrical outlet.
 Perform regular maintenance checks on the mainframe of the reader. Inspect all
components.
 After long term storage, take extra care to be sure the reader operates correctly.
 Avoid handling or storing liquid on or near the reader.
 Do not open the reader at any time while it is operating or when it is plugged into the wall
outlet.
 Keep the reader away from fire or smoking.
 When moving the reader, unplug it and move it carefully and gently.
 The reader should be placed on a flat and stable surface.
 Refer to this manual for handling and/or maintenance.
 Keep the reader (including the cable and the plug) out of reach of children.
 Do not attempt to disassemble/modify the reader.
 Do not handle the reader or the power cord with wet hands.
 Do not store the reader in direct sunlight, high/low temperatures or high humidity.
 Do not throw, drop, hit or push the reader.

13
 Do not use the reader or the plug during an active thunder and/or lightning storm.
 Do not touch the LCD panel with sharp objects.
 Do not use the reader, if it has any damaged parts (including the power cord or the plug).
 Do not use in a manner not specified by the manufacturer, otherwise it may cause unknown
risks.

8. Daily Maintenance

8.1 Maintenance
To assure reliable and accurate test results, please check the following:
 Observe the external appearance and screen of Handheld DOA Reader every time you use
the device.
 Please operate the reader following the requirements of the user manual.
 Calibrate the reader periodically by control cards to make sure the results accurate.
 Reader should be put in stable temperature, dry place.
 Do not try to disassemble the reader. Operation by laypeople may damage the reader.
 If there are any problems found in the results after the checkup, stop using the reader
immediately and contact the distributor.
The maintenance should be carried out by members of the professional equipment
maintenance team. After equipment maintenance, random inspections and tests will be carried
out. Only those that meet the technical requirements will be released.

8.2 Cleaning and storage


 Do not clean the reader with strong or caustic cleansing agents, chemicals or cleansing
tissues impregnated with chemicals.
 The reader only requires external cleaning. Please do not clean any internal parts or inner
surfaces.
 The power switch should be turned off before the reader is cleaned, to avoid short circuit
and electric shock hazard.
 If you want to store for a lengthy period, keep the reader covered in a dry place. Do not
expose the reader to direct sunlight.
 Recommended storage conditions are as follows: Temperature (10~40°C), Humidity (≤
80% RH,no condensation), Air pressure (86.0 ~ 106.0 kPa).
 Do not place the reader near chemicals and/or corrosive substances.
 Keep the reader, cable and other accessories well-organized.
 Keep the reader clean for the next use.

14
 To move the reader, carefully pick it up and put it down gently to avoid impact.
 Do not place heavy materials on top of the reader.

8.3 Software Update


When the Handheld DOA Reader has been connected to the WIFI, in below figure, touch
“Check update” to check if there is a new version update.You can click“update right now” to
update the software.

When download finished, LCD show menu,and touch “Install” to install new version
software.

15
9. Trouble shooting

NO. Error description Troubleshooting

a) Long pressed the power switch to power off;


the button in the interface does not respond, b) Release the power key, then wait for a while;
1
or a phenomenon of crash occurs c) Long pressed the power switch to power on.

a) Pull up the top of the printer, check whether paper

2 the thermal printer is jammed is fold;


b) Restart the reader.
please check if the printer is out of paper, or if the
paper is installed correctly, if there is paper but the
3 the data can not be printed above problems occur, please contact manufacturer
for repair.
his may cause the reader fails to execute the
"detection" function. The data backup and
transmission functions are not affected. In this case,
4 Error occurs on the user interface. you can upload the backed up records to the host
computer software and return the reader to the factory
for repair.
The occurrence rate of this error is very low.
When it happens, the latest test data is lost, but

5 Error of logic of the software the historical data backed up in the reader will not
be lost. At this time, the reader should be powered
off and restarted and can be used continuously.

When the system or network resources can


not be used, the reader's self-diagnosis When the system or network resources become
6 function and data backup function may stable, the function will be automatically restored.
continue to be used, which may affect the
record upload to the host computer software