Energy_gap_in_semiconductors (1)
Energy_gap_in_semiconductors (1)
Energy_gap_in_semiconductors (1)
2 Apparatus
The experiment requires a constant current source, a voltmeter with high input
resistance, a thermometer, and a device for changing the sample’s temperature
at a small and constant rate.
1
3 Observation Table
Current is kept constant at 8mA. The separation between the probes is 2mm.
2
4 Plot and Calculations
5 Error analysis
Relative error in measurement of resistivity is given by
∆ρ ∆V ∆I ∆S
= + + (1)
ρ V I S
0.001 0.01 0.01
= + + (2)
0.338 8.00 2.00
= 0.0092 (3)
∆Eg ∆ρ ∆T
= + (4)
Eg ρlog(ρ) T
0.0092 1
= + (5)
1.669 315
= 0.0087 (6)
⇒ ∆Eg = 0.0087 × 0.789 = 0.0069eV
3
6 Sources of Error and Discussions
The band gap of Germanium was found to be (0.789±0.0069) eV, while the
actual bandgap is found to be 0.7 eV.
Temperature may not be high enough for intrinsic behaviour approximation to
hold true during the start of the experiment when the temperature is around
30◦ C.
Voltmeter’s internal resistance also affects the readings