PIXIS System Manual
PIXIS System Manual
PIXIS System Manual
Issue 4
April 20, 2016
Copyright 2004-2016 Princeton Instruments, a division of Roper Scientific, Inc.
3660 Quakerbridge Rd
Trenton, NJ 08619
TEL: 800-874-9789 / 609-587-9797
FAX: 609-587-1970
All rights reserved. No part of this publication may be reproduced by any means without the written
permission of Princeton Instruments, a division of Roper Scientific, Inc. ("Princeton Instruments").
Printed in the United States of America.
IntelliCal and PICam are trademarks and eXcelon, LightField, and PVCAM are registered trademarks of
Roper Scientific, Inc.
LabVIEW is a registered trademark of National Instruments, Inc.
LEMO is a registered trademark of INTERLEMO HOLDING SA
Nikon is a registered trademark of Nikon, Inc.
Scientific Imaging ToolKit and SITK are trademarks of R Cubed Software Consultants, LLC.
SpectraPro is a trademark of Acton Research Corporation.
Windows and Windows Vista are registered trademarks of Microsoft Corporation in the United States
and/or other countries.
The information in this publication is believed to be accurate as of the publication release date. However,
Princeton Instruments does not assume any responsibility for any consequences including any damages
resulting from the use thereof. The information contained herein is subject to change without notice.
Revision of this publication may be issued to incorporate such change.
Table of Contents
Chapter 1 Introduction .........................................................................................7
PIXIS .................................................................................................................................. 7
eXcelon® ............................................................................................................................. 7
Advanced Design ................................................................................................................ 7
Grounding and Safety ......................................................................................................... 8
Precautions .......................................................................................................................... 8
UV Coating ......................................................................................................................... 8
Cleaning .............................................................................................................................. 9
Repairs ................................................................................................................................ 9
About this Manual .............................................................................................................. 9
Chapter 2 System Component Descriptions .................................................. 11
System Components ......................................................................................................... 11
PIXIS Camera ................................................................................................................... 12
Power Supply (Air-cooled and CoolCUBEII Liquid-cooled systems) ................................... 13
Coolant Hoses (Liquid-cooled systems) ........................................................................... 13
Cables ............................................................................................................................... 14
Certificate of Performance ................................................................................................ 14
User Manuals .................................................................................................................... 14
Optional Components ....................................................................................................... 15
Chapter 3 Initial System Verification ............................................................... 17
Chapter 4 System Setup ................................................................................... 21
Introduction....................................................................................................................... 21
Unpacking the System ...................................................................................................... 21
Checking the Equipment and Parts Inventory................................................................... 22
System Requirements ....................................................................................................... 22
Software Installation ......................................................................................................... 24
Making the Camera-Circulator Connections for a CoolCUBEII ....................................... 25
Entering the Default Camera System Parameters ............................................................. 27
Attaching Lenses to C- and F-Mount Adapters ................................................................ 28
Mounting the Adjustable C- to Spectroscopy-Mount Kinetics Adapter ........................... 29
Mounting the Camera to a Spectrograph .......................................................................... 30
Connecting an External Shutter ........................................................................................ 31
Chapter 5 Operation .......................................................................................... 33
Introduction....................................................................................................................... 33
System On/Off Sequences ................................................................................................ 34
WinX First Light Instructions ........................................................................................... 34
LightField First Light Instructions .................................................................................... 43
Exposure and Signal ......................................................................................................... 54
Readout ............................................................................................................................. 57
Digitization (Rate) ............................................................................................................ 63
Chapter 6 Advanced Topics ............................................................................. 65
Introduction....................................................................................................................... 65
Timing Modes ................................................................................................................... 66
iii
iv PIXIS System Manual Issue 4
Acton Series Spectrograph (PIXIS with 3.60/3.88 Bolt Circles) .................................... 113
Acton Series Spectrograph (PIXIS with C-Mount) ........................................................ 114
Acton SP-2350/SP-2550 Adjustable C- to Spectroscopy-Mount Adapter (Adapter
Kit 7050-0104) ......................................................................................................... 115
Acton SP-2150/SP-2750 Adjustable C- to Spectroscopy-Mount Adapter (Adapter
Kit 7050-0107) ......................................................................................................... 116
IsoPlane SCT-320 (PIXIS with Flange) ......................................................................... 117
Appendix E Cross-Referencing of WinX and LightField Terms .................... 119
WinX-to-LightField ........................................................................................................ 119
LightField-to-WinX ........................................................................................................ 121
Warranty & Service ......................................................................................... 123
Limited Warranty ............................................................................................................ 123
Contact Information ........................................................................................................ 126
Index ................................................................................................................. 127
Figures
Figure 1. Typical System Components ............................................................................ 11
Figure 2. Typical Imaging Experiment Layout (Air-cooled Camera) ............................. 18
Figure 3. Typical Spectroscopy Experiment Layout (Air-cooled Camera) ..................... 18
Figure 4. Typical Imaging Experiment Layout (Liquid-cooled Camera with
CoolCUBEII) ................................................................................................. 19
Figure 5. Typical Spectroscopy Experiment Layout (Liquid-cooled Camera with
CoolCUBEII) ................................................................................................. 19
Figure 6. WinView Installation: Select Installation Type dialog..................................... 24
Figure 7. LightField Installation Wizard dialog .............................................................. 25
Figure 8. Camera Detection Wizard - Welcome dialog ................................................... 27
Figure 9. LightField Experiment Workspace................................................................... 28
Figure 10. Adjustable C- to Spectroscopy-Mount Kinetics Adapter ............................... 30
Figure 11. Acton Series Spectrograph Entrance Slit Mount ............................................ 31
Figure 12. Block Diagram of PIXIS System .................................................................. 33
Figure 13. Available Devices Area .................................................................................. 43
Figure 14. Experiment Devices Area ............................................................................... 44
Figure 15. View Area....................................................................................................... 45
Figure 16. View Area Displaying an Image .................................................................... 45
Figure 17. Available Devices Area .................................................................................. 47
Figure 18. Experiment Devices Area ............................................................................... 48
Figure 19. View Area....................................................................................................... 50
Figure 20. Spectrometer Alignment: Before Rotational Alignment ................................ 51
Figure 21. Spectrometer Alignment: After Rotational Alignment .................................. 52
Figure 22. Exposure of the CCD with Shutter Compensation ......................................... 54
Figure 23. WinX Detector Temperature dialog .............................................................. 55
Figure 24. Array Terms for a CCD with a Dual Output Amplifier ................................ 57
Figure 25. Full Frame at Full Resolution ......................................................................... 58
Figure 26. 2 × 2 Binning .................................................................................................. 59
Figure 27. Binning and Array Orientation ....................................................................... 61
Figure 28. Free Run {No Response} Timing Chart, part of the chart in Figure 36 ......... 67
Figure 29. Free Run {No Response} Timing Diagram.................................................... 67
Figure 30. Chart Showing Two External Sync Timing Options ...................................... 68
v
vi PIXIS System Manual Issue 4
Figure 31. Timing Diagram for External Sync Mode (+ edge trigger) ............................ 69
Figure 32. Continuous Cleans {Clean Until Trigger} Flowchart .................................... 70
Figure 33. WinX Continuous Cleans Timing Diagram ................................................... 71
Figure 34. LightField Clean Until Trigger (CUT) Timing Diagram ............................... 71
Figure 35. Rear of PIXIS Camera .................................................................................... 71
Figure 36. Chart of Safe and Fast Mode Operation ......................................................... 73
Figure 37. Comparison of NOT SCAN {Not Reading Out}, SHUTTER {Shutter
Open}, and NOT READY {Busy} Logic Output Levels.............................. 74
Figure 38. Kinetics Readout ............................................................................................ 75
Figure 39. Hardware Setup dialog ................................................................................... 76
Figure 40. Experiment Setup dialog ................................................................................ 76
Figure 41. Sensor Readout expander: Kinetics Readout Mode ...................................... 76
Figure 42. Shutter and Trigger expanders: No Response ............................................... 76
Figure 43. Free Run Timing Diagram.............................................................................. 77
Figure 44. Single Trigger Timing Diagram ..................................................................... 77
Figure 45. Multiple Trigger Timing Diagram.................................................................. 78
Figure 46. WinX: Custom Chip tab ................................................................................. 79
Figure 47. LightField: Custom Sensor pane .................................................................... 79
Figure 48. LightField: Custom Timing ............................................................................ 80
Figure 49. WinX: Vertical Shift ...................................................................................... 80
Figure 50. Acquisition Display ........................................................................................ 82
Figure 51. Camera1 in Camera Name Field .................................................................... 83
Figure 52. Data Overrun Due to Hardware Conflict dialog ............................................. 86
Figure 53. Devices Missing dialog .................................................................................. 87
Figure 54. Occupied Device icon .................................................................................... 88
Figure 55. Error Creating Controller dialog .................................................................... 88
Figure 56. Program Error dialog ...................................................................................... 89
Figure 57. Serial Violations Have Occurred dialog ......................................................... 90
Figure 58. Adjustable C-Mount: Internal Shutter (Air-cooled) ....................................... 95
Figure 59. Fixed C-Mount: Internal Shutter (Air-cooled) ............................................... 96
Figure 60. Adjustable C-Mount: Internal Shutter (Liquid-cooled) .................................. 97
Figure 61. Fixed C-Mount: Internal Shutter (Liquid-cooled) .......................................... 98
Figure 62. F-Mount: Internal Shutter (Air-cooled) .......................................................... 99
Figure 63. F-Mount: Internal Shutter (Liquid-cooled) .................................................. 100
Figure 64. F-Mount: 2048x2048, Internal Shutter (Air-cooled) .................................... 101
Figure 65. F-Mount: 2048x2048, Internal Shutter (Liquid-cooled)............................... 102
Figure 66. Spectroscopy-Mount: No Internal Shutter, 3.60 bolt circle (Air-cooled)..... 103
Figure 67. Spectroscopy-Mount: No Internal Shutter, 3.60 bolt circle (Liquid-cooled) 104
Figure 68. Spectroscopy-Mount: Internal Shutter (3.60" and 3.88" bolt circles)
(Air-cooled) ................................................................................................ 105
Figure 69. CoolCUBEII Circulator................................................................................. 106
Figure 70. F-mount Adjustment..................................................................................... 108
Tables
Table 1. USB Driver Files and Locations ........................................................................ 24
Table 2. Example of Controller Gain {Analog Gain} vs. Readout Port .......................... 62
Table 3. Camera Timing Modes ...................................................................................... 66
Table 4. Focal Plane Distances ........................................................................................ 91
Table 5. Typical Deepest Operating Temperature ........................................................... 93
Chapter 1
Introduction
PIXIS
Thank you for purchasing a PIXIS camera system
from Princeton Instruments. For over two decades
Princeton Instruments has been the legendary name
behind the most revolutionary spectroscopy and
imaging products for cutting edge research.
PIXIS represents the most advanced camera design and utilizes years
of experience and expertise in low-light detection. Whether your
application involves Raman spectroscopy in the near infrared or
semiconductor imaging in the ultraviolet, PIXIS has everything you need to tackle the
most demanding applications.
Among the many state of the art features are its maintenance-free permanent vacuum,
integrated controller, deep thermoelectric air-cooling, and compact design. Currently the
platform supports several imaging and spectroscopy CCDs, including eXcelon® enabled
back-illuminated spectroscopy CCDs. Please visit www.princetoninstruments.com for the
current list of supported CCDs.
eXcelon®
eXcelon is a new CCD/EMCCD sensor technology jointly developed by Princeton
Instruments, e2v, and Photometrics. Spectroscopy CCDs using this technology provide
three significant benefits:
Improved sensitivity – improved QE over broader wavelength region compared
to back-illuminated sensors,
Reduced etaloning – up to 10 times lower etaloning or unwanted fringes in near
infrared (NIR) region compared to standard back-illuminated CCDs,
Lower dark current – similar to back-illuminated CCDs or 100 times lower
than the deep depletion CCDs.
Advanced Design
PIXIS is a fully integrated camera system. The camera contains all of the electronics
necessary to read out and control the CCD device. For instance, it houses precision analog-
to-digital converters (ADCs) positioned close to the CCD for lowest noise and has USB 2.0
electronics to interface with the host computer.
The easy-to-use PIXIS camera system offers all basic CCD camera functions such as
region-of-interest (ROI) selection and binning --- all under software control. It also
provides advanced triggered operation as well as programmable TTL output.
To utilize the full potential of the PIXIS camera system, please read the manual completely.
7
8 PIXIS System Manual Issue 4
WARNING! If the equipment is damaged, the protective grounding could be disconnected. Do not use
damaged equipment until its safety has been verified by authorized personnel.
Disconnecting the protective earth terminal, inside or outside the apparatus, or any
tampering with its operation is also prohibited.
Inspect the supplied power cord. If it is not compatible with the power socket, replace the
cord with one that has suitable connectors on both ends.
WARNING! Replacement power cords or power plugs must have the same polarity and power rating
as that of the original ones to avoid hazard due to electrical shock.
Precautions
To prevent permanently damaging the system, please observe the following precautions:
The CCD array is very sensitive to static electricity. Touching the CCD can
destroy it. Operations requiring contact with the device can only be performed at
the factory.
If you are using high-voltage equipment (such as an arc lamp) with your camera
system, be sure to turn the camera power ON LAST and turn the camera power
OFF FIRST.
Use caution when triggering high-current switching devices (such as an arc lamp)
near your system. The CCD can be permanently damaged by transient voltage
spikes. If electrically noisy devices are present, an isolated, conditioned power
line or dedicated isolation transformer is highly recommended.
Do not block air vents on the camera. Preventing the free flow of air overheats
the camera and may damage it.
UV Coating
Caution If you have a camera with a UV (Lumogen or Unichrome) coated CCD, protect it from
unnecessary exposure to UV radiation. This radiation slowly bleaches the coating,
reducing sensitivity.
Chapter 1 Introduction 9
Cleaning
WARNING! Turn off all power to the equipment and secure all covers before cleaning the units.
Otherwise, damage to the equipment or injury to you could occur.
Camera
Although there is no periodic maintenance that needs to be performed on a PIXIS
camera, users are advised to wipe it down with a clean damp cloth from time to time.
This operation should only be done on the external surfaces and with all covers secured.
In dampening the cloth, use clean water only. No soap, solvents or abrasives should be
used. Not only are they not required, but they could damage the finish of the surfaces on
which they are used.
Optical Surfaces
As a good practice, the camera must be closed/capped off with the supplied dust cover or
lens cap when not in use. Should a need to clean the optical window arise due to the
accumulation of atmospheric dust, we advise that the drag-wipe technique be used. This
involves dipping a clean cellulose lens tissue into clean anhydrous methanol, and then
dragging the dampened tissue over the optical surface to be cleaned. Do not allow any
other material to touch the optical surfaces.
Repairs
Because the PIXIS camera system contains no user-serviceable parts, repairs must be
performed by Princeton Instruments. Should your system need repair, contact Princeton
Instruments customer support for instructions. For contact information, refer to page 126
of this manual.
Save the original packing materials and use them whenever shipping the system or
system components.
9
10 PIXIS System Manual Issue 4
11
12 PIXIS System Manual Issue 4
PIXIS Camera
CCD Array: The PIXIS camera system offers both front- and back-
illuminated CCDs in a variety of array sizes that allow you to precisely
match the sensor to your application. Only scientific-grade devices are
used in order to ensure the highest image fidelity, resolution, and
acquisition flexibility required for scientific imaging. Princeton Instruments has
developed exclusive CCDs with unmatched quantum efficiency and low noise to offer the
utmost in sensitivity. Large full wells, square pixels, and 100% fill factors provide high
dynamic range and excellent spatial resolution. Unichrome (exclusive Princeton
Instruments technology) and other UV-enhancement coatings can be used to further
improve the quantum efficiency of these CCDs in the ultraviolet. Your choice of CCD is
already installed in the camera that you received and has been individually tested.
Cooling: Dark current is reduced in PIXIS camera systems through thermoelectric
cooling of the CCD arrays. Cooling by this method uses a four-stage Peltier cooler in
combination with circulating air or coolant. To prevent condensation and contamination
from occurring, cameras cooled this way are evacuated. Due to CCD size/packaging
differences, the lowest achievable temperature can vary from one PIXIS model to the
next. Please refer to the specific system’s data sheet for cooling performance.
Connectors:
USB 2.0: Control signals and data are transmitted between
the camera and the host computer via the USB port located on
the rear of the camera. As of this printing, you can hot plug
the PIXIS camera whenever the WinX application is not
running (i.e., connect or disconnect from the camera or the
host computer while the camera is powered ON). In the
case of cameras built before November 1, 2005, you must
exit the WinX application and turn the camera power OFF
before connecting the USB cable to or disconnecting it from the camera or host
computer.
Shutter: LEMO® connector provides the shutter drive pulses for driving a Princeton
Instruments-supplied external shutter (for example, a shutter at the entrance slit of a
spectrograph). Camera power must be OFF before connecting to or disconnecting
from this connector.
Note: When there is an installed internal shutter, this connector cannot drive an
external shutter.
LOGIC OUT: 0 to +3.3V programmable logic level output (TTL-compatible). The
output of this connector can be programmed and can also be inverted via the
application software. For detailed information about each output signal, please see
“LOGIC OUT Control” (page 74).
EXT SYNC: 0-+3.3V logic level input (TTL-compatible) that has a 10 k pullup
resistor. Allows data acquisition and readout to be synchronized with external events.
Through software, positive or negative (default) edge triggering can be selected.
Power: 12 VDC (6.6A max) input from power supply.
Chapter 2 System Component Descriptions 13
WARNING! Use only the hoses and circulator shipped with your system. Attaching any other hoses or
circulator voids the warranty.
13
14 PIXIS System Manual Issue 4
Cables
USB 2.0 Cable: The standard 16.4' (5 m) cable (6050-0494) has USB connectors that
interconnect the "USB 2.0" connector on the rear of the PIXIS with a USB card installed
in the host computer.
MCX to BNC Adapter Cables: Two MCX to BNC adapter cables are provided with
the PIXIS system. These mount to the EXT SYNC and the LOGIC OUT connectors on
the rear of the PIXIS.
Certificate of Performance
Each PIXIS camera has a Certificate of Performance. This certificate states that the
camera system was assembled and tested according to approved Princeton Instruments
procedures. It documents the camera performance data as measured during the testing of
your PIXIS and lists the Sales Order, Purchase Order, and Camera Serial numbers (useful
if you ever need to contact Princeton Instruments Customer Support).
User Manuals
PIXIS System User Manual: This manual describes how to install and use the PIXIS
system components.
WinView/32 or WinSpec/32 User Manual: This manual describes how to install and
use the application program. A PDF version of this manual is provided on the installation
CD. Additional information is available in the program's on-line help.
LightField® User Manual: This manual describes how to install and use the LightField
application program (for 64-bit Windows Vista® and Windows® 7 operating systems).
The manual is provided in PDF version on the installation CD and will be installed in the
Princeton Instruments/LightField/Documents subdirectory. Acrobat 7.0 or higher is
required. Additional information is available in the program's on-line help.
Note: You can download current versions of Princeton Instruments manuals at
ftp://ftp.princetoninstruments.com/Public/Manuals/Princeton Instruments/. The most
current versions of Acton manuals are located at
ftp://ftp.princetoninstruments.com/Public/Manuals/Acton/.
Chapter 2 System Component Descriptions 15
Optional Components
Application Software
WinX: The PIXIS camera can be operated by using either WinView/32 or
WinSpec/32, Princeton Instrument's 32-bit Windows® software packages designed
specifically for high-end imaging and spectroscopy, respectively. The Princeton
Instruments' software provides comprehensive image/spectral capture and display
functions. The package also facilitates snap-ins to permit advanced operation. Using
the optional built-in macro record function, you can also create and edit your own
macros to automate a variety of operations. WinView and WinSpec take full advantage
of the versatility of the PIXIS camera and even enhance it by making integration of the
detection system into larger experiments or instruments an easy, straightforward
endeavor.
PVCAM®: The standard software interface for cooled CCD cameras from Princeton
Instruments. It is a library of functions that can be used to control and acquire data
from the camera when a custom application is being written. For example, in the case
of Windows, PVCAM is a dynamic link library (DLL). Also, it should be understood
that PVCAM is solely for camera control and image acquisition, not for image
processing. PVCAM places acquired images into a buffer, where they can then be
manipulated using either custom written code or by extensions to other commercially
available image processing packages.
Scientific Imaging ToolKit™: SITK™ is a collection of LabVIEW® VIs for scientific
cameras and spectrographs. This third party software can be purchased from Princeton
Instruments.
LightField®: The PIXIS can be operated using LightField, Princeton Instrument’s
64-bit Windows Vista® and Windows® 7 compatible software package. LightField
combines complete control over Princeton Instruments’ cameras and spectrographs
with easy-to-use tools for experimental setup, data acquisition and post-processing.
LightField makes data integrity priority #1 via automatic saving to disk, time stamping
and retention of both raw and corrected data with full experimental details saved in
each file. LightField works seamlessly in multi-user facilities, remembering each user’s
hardware and software configurations and tailoring options and features accordingly.
The optional, patent-pending IntelliCal™ package is the highest-performance
wavelength calibration software available, providing up to 10X greater accuracy across
the entire focal plane than competing routines.
PICam™: The standard 64-bit software interface for cooled CCD cameras from
Princeton Instruments. PICam is an ANSI C library of camera control and data
acquisition functions. Currently, the interface supports Windows Vista and Windows 7.
Note: PIXIS may also be operated by several other third-party software packages.
Please check with the providers of the packages for compatibility and support
information.
15
16 PIXIS System Manual Issue 4
Internal Shutter
Optional 25 or 45 mm internal shutter (dependent on CCD array size). Shutters are
mechanical devices with a finite lifetime, typically of the order of a million cycles,
although some individual shutters may last a good deal longer. How long a shutter lasts in
terms of experimental time will, of course, be strongly dependent on the operating
parameters. High repetition rates and short exposure times will rapidly increase the
number of shutter cycles and hasten the time when the shutter will have to be replaced.
F-Mount Adapter
An F-mount adapter (7050-0009) is available for modifying a spectroscopy-mount PIXIS
with internal shutter (see Figure 68, page 105).
Action Reference
1. If the system components have not already been unpacked, unpack Chapter 4 System Setup,
them and inspect their carton(s) and the system components for page 21
in-transit damage.
2. Verify that all system components have been received. Chapter 4 System Setup,
page 22
3. If the components show no signs of damage, verify that the Chapter 4 System Setup,
appropriate power cord has been supplied with the power supply. page 22
4. If the application software is not already installed in the host Chapter 4 System Setup,
computer, install it. page 24 & Software manual
5. If a USB 2 interface card is not already installed in the host
computer, install it. Follow the manufacturer’s instructions.
6. Depending on application, attach lens to the camera or mount the Chapter 4 System Setup,
camera to a spectrograph. page 27
7. With the power supply disconnected from the camera, connect the USB
cable to the USB port at the rear of the camera and to the USB port at
the computer.
8. Air-Cooled System: Plug the power supply into the rear of the
camera and plug the power supply into the power source.
Liquid-Cooled System: Plug the power supply into the rear of the
Chapter 4 System Setup,
camera and plug the power supply into the power source. Make the
page 25
hose connections to the camera. Plug the circulator into the power
source. Add coolant if necessary. Turn on the circulator.
9. Turn the camera ON.
10. Turn on the computer and begin running the application software. Software manual
11. Enter the hardware setup information or load the defaults from the Chapter 5 Operation,
camera. page 27
12. Set the target array temperature. Chapter 5 Operation,
page 55
13. When the system reaches temperature lock, wait an additional 20 Chapter 5 Operation,
minutes and then begin acquiring data in focus mode. page 36 or page 39
17
18 PIXIS System Manual Issue 4
Action Reference
14. Adjust the focus for the best image or spectral lines. If you are using Chapter 5 Operation,
WinSpec/32, you may want to use the Focus Helper function for page 43 or page 46
spectroscopy applications. If you are using LightField, you may
want to use the Align Spectrometer function.
19
20 PIXIS System Manual Issue 4
To minimize risk to users or to system equipment, turn the system OFF before any cables
are connected or disconnected.
Introduction
A PIXIS camera system consists of three hardware components:
Camera head
Power supply
Cables
All of the components and cables required for your configuration are included with your
shipment. Your PIXIS system has been specially configured and calibrated to match the
camera options specified at the time of purchase. The CCD and coating you ordered have
been installed in the camera head.
Keep all of the original packing materials so you can safely ship the PIXIS system to
another location or return it for service if necessary. If you have any difficulty with any
step of the instructions, call Princeton Instruments Customer Support. For contact
information, refer to page 126.
Hardware installation may consist of:
Installing an interface card, if the appropriate card is not already resident.
Attaching a lens to a C-mount on the camera or to an F-mount adapter.
Connecting the camera to an external shutter, if one is required.
Mounting the camera to a spectrograph.
Software installation depends on the application software you will be using to run the
system. Refer to the manual supplied with the software for information about installing
and setting it up.
21
22 PIXIS System Manual Issue 4
Ventilation: Allow at least one inch clearance for the side air vents. Where the camera
is inside an enclosure, > 30 cfm air circulation and heat dissipation of 100W is required.
Power: The PIXIS camera receives its power from the supplied power supply which in
turn plugs into an AC power source.
Host Computer
Note: Computers and operating systems all undergo frequent revision. The following
information is only intended to give minimum computer requirements. Please contact the
factory to determine your specific needs.
WinX Requirements
Windows® XP (32-bit with SP3 or later), Vista® (32-bit), or Windows® 7 (32-bit)
2 GHz Pentium® 4 (or greater).
Native USB 2.0 support on the mother board or USB 2.0 Interface Card (Orange
Micro 70USB90011 USB2.0 PCI is recommended for desktop; SIIG, Inc. USB
2.0 PC Card, Model US2246 for laptop)
Minimum of 1 GB RAM (or greater).
CD-ROM drive.
Hard disk with a minimum of 1 Gbyte available. A complete installation of the
program files takes about 17-50 Mbytes and the remainder is required for data
storage, depending on the number and size of images/spectra collected. Disk level
compression programs are not recommended. Drive speed of 10,000 RPM
recommended.
Super VGA monitor and graphics card supporting at least 65,535 colors with at
least 128 Mbyte of memory. Memory requirement is dependent on desired
display resolution.
Mouse or other pointing device.
LightField Requirements
Windows Vista® (64-bit) or Windows® 7 (64-bit)
2 GHz dual core processor
4 GB RAM (or greater)
CD-ROM drive
Super VGA monitor and graphics card supporting at least 65535 colors with at
least 128 Mbyte of memory. Memory requirement is dependent on desired
display resolution.
Hard disk with a minimum of 1 Gbyte available for installation. Additional space
is required for data storage: the amount of space required depends on the number
and size of images/spectra collected. Disk level compression programs are not
recommended. Drive speed of 10,000 RPM recommended.
Mouse or other pointing device.
Note: The above requirements are the minimum for operating a PIXIS camera. A
faster computer with 5GB or larger memory (RAM) will greatly enhance the
software performance during live mode operations.
23
24 PIXIS System Manual Issue 4
Software Installation
WinX
Notes:
1. Before proceeding, please check to see if your computer supports USB 2.0. If it does
not, install a USB 2.0 interface card. Follow the manufacturer’s instructions.
2. Leave the USB cable disconnected from the camera until you have installed
WinView/32 or WinSpec/32.
The following installation is performed via the WinView/32 or WinSpec/32 software
installation CD.
1. On the Select Installation Type
dialog (see Figure 6), click on
Typical radio button to install the
required drivers and the most
commonly installed program files.
Select the Custom radio button if
you would like to choose among the
available program files or do not want
to install the drivers. Complete
installs all of the application features.
The required INF, DLL, and USB
driver files will be placed in the
appropriate "Windows" directories Figure 6. WinView Installation:
(see Table 1 for locations). Select Installation Type dialog
2. Make sure the camera is connected to the host computer and that the camera power
supply is turned on.
3. Reboot the computer.
4. At bootup, Windows will detect the Princeton Instruments USB2 Interface hardware
in the PIXIS. You may be prompted to enter the directory path(s) for the
apausbprop.dll and/or the apausb.sys file(s), either by keyboard entry or by using the
browse function.
Windows Version USB INF Filename USB Properties DLL USB Device Driver Name
Located in Located in Located in
"Windows"/INF "Windows"/System3 "Windows"/System32/Drivers
directory* 2 directory directory
®
Windows XP, rsusb2k.inf (in apausbprop.dll (in apausb.sys (in WINDOWS/
Vista (32-bit), WINDOWS/INF, WINDOWS/System3 System32/Drivers, for
Windows 7 (32-bit) for example) 2, for example) example)
LightField
The following installation is performed via the LightField software installation CD.
1. Before starting the installation:
Verify that the computer operating system is Windows Vista (64-bit) or
Windows 7 (64-bit).
Confirm that your computer supports USB 2.0. If it does not, please refer to the
manufacturer’s instructions for installing a USB 2.0 interface card.
Verify that your computer is connected to the Internet. Internet connection is
required for product activation.
2. Insert the CD and follow the installation wizard prompts.
25
26 PIXIS System Manual Issue 4
Notes:
1. Make sure that there are no kinks in the hoses that impede the coolant flow. Lack
of sufficient flow can seriously harm the detector and any resulting damage is not
covered under warranty.
2. Damage caused by water leaking into the PIXIS voids the warranty.
5. Unscrew the reservoir cap (on top of the CoolCUBEII) and make
sure that the coolant reservoir contains coolant. If additional
coolant is required, fill with a 50:50 mixture of water and ethylene
glycol.
6. Screw the reservoir cap back in.
7. Plug the circulator into a 100-240 VAC, 47-63 Hz power source.
8. Turn the circulator on. Make sure there are no leaks or air bubbles in the hoses.
Note: Small air bubbles (about the size of bubbles in soda) are common in the
CoolCUBEII especially at start up and do not prevent proper operation.
WinX
1. Make sure the PIXIS is connected to the host computer and that it is turned on.
2. Run the WinX application. The Camera Detection wizard will automatically run if
this is the first time you have installed a Princeton Instruments WinX application
(WinView/32, WinSpec/32, or WinXTest/32) and a supported camera. Otherwise, if
you installing a new camera type, click on the Launch Camera Detection
Wizard… button on the Controller/CCD tab to start the wizard.
3. On the Welcome dialog (Figure 8), leave the checkbox unselected and click on
Next.
LightField
1. Make sure the PIXIS (and spectrograph, if this is a spectroscopy system) is
connected to the host computer and that the camera (and spectrograph) power
supply is turned on.
2. Start LightField.
3. While LightField is starting up, it will detect the available device(s) and load the
appropriate icons into the Available Devices area in the Experiment workspace.
4. When you drag an icon into the Experiment Devices area, the appropriate
expanders will be loaded into the Experiment Settings stack on the lefthand side
of the window.
27
28 PIXIS System Manual Issue 4
29
30 PIXIS System Manual Issue 4
6. When you have finished with focusing and rotational alignment, secure the
sliding tube in place with the spectrograph setscrews.
7. If they have already been installed, loosen the adapter’s two recessed locking
setscrews (use a 0.050 hex key). If they have not been installed, insert the
setscrews into the holes at the sides of the adapter and screw them in a couple of
turns.
8. When the camera opening is centered within the sliding tube, you can adjust the
camera up or down .4” (4.06 mm) by tightening or loosening the vertical
adjustment setscrew (use a 3/32” hex key).
9. After you have completed the adjustment, tighten the recessed locking setscrews
in the adapter flange to lock the position.
Spectrograph Adapter
Refer to Appendix C, page 111 for instructions for a variety of adapters. Other adapters
may be available. Consult the factory for specific information relating to your needs.
Chapter 4 System Setup 31
Introduction
Typically, PIXIS cameras for imaging applications are shipped with an internal shutter.
The 25 or 45 mm internal shutter (depending on the CCD array size) is housed in the
main body. Typically, cameras for spectroscopy applications do not incorporate internal
shutter. However, for cameras without an internal shutter, there is provision for
connecting an external 25 or 45 mm shutter (supplied by Princeton Instruments) at the
Shutter connector on the rear of the PIXIS.
Cautions 1. DO NOT connect a Princeton Instruments-supplied external shutter when there is an
installed internal shutter. Permanent damage to the shutter driver may occur.
2. Electromechanical shutters typically have a lifetime of about a million cycles. Avoid
running the shutter unnecessarily.
3. A shutter can become overheated when short, rapidly repeated exposures are used or if
the shutter is held open for an extended period of time. Newer versions of the
WinView/WinSpec software and camera hardware monitor the temperature of 45 mm
internal shutters to prevent overheating conditions. The temperature of external
shutters is NOT monitored!
External Shutter
A Princeton Instruments-supplied external shutter may be used with a PIXIS camera that does
not have an internal shutter. In most cases, the external shutter will be mounted on the
entrance slit of a spectrograph. The shutter mount used with all Acton Series spectrographs
requires no disassembly. Mount it to the spectrograph as shown in Figure 11. In the case of
the IsoPlane SCT-320 spectrograph, the shutter assembly is mounted internally: refer to the
instructions provided in the IsoPlane manual.
Shutter Cable Connection
1. Verify that the PIXIS camera is turned OFF (i.e., the
power supply is switched OFF).
2. Verify that there is no internal shutter.
Caution If the camera is equipped with an internal shutter, DO
NOT USE the Shutter connector to drive an external
(second) shutter. Such a configuration will result in
under-powering both shutters and may cause damage to
the system.
3. Connect the shutter cable to the LEMO® connector at
the rear of the camera.
4. Power the PIXIS camera ON.
Figure 11. Acton Series Spectrograph
Entrance Slit Mount
31
32 PIXIS System Manual Issue 4
Overexposure Protection
Cameras that are exposed to room light or other continuous light sources will quickly
become saturated. This most often occurs when operating without a shutter. If the camera
is mounted to a spectrograph, close the entrance slit of the spectrograph to reduce the
incident light.
Note: If the CCD is cooled to low temperatures (below -50°C), exposure to ambient
light will over-saturate it. This may increase dark charge significantly. If the camera
remains saturated after all light sources are removed, you may have to bring the camera
back to room temperature to restore dark charge to its original level.
Chapter 5
Operation
Introduction
Once the PIXIS camera has been installed as explained in
the preceding chapters, operation of the camera is
straightforward. In most applications you simply establish
optimum performance using the Focus mode (in WinX,
for example), set the target camera temperature, wait until
the temperature has stabilized, and then do actual data
acquisition in the Acquire mode. Additional
considerations regarding experiment setup and equipment
configuration are addressed in the software manual.
During data acquisition, the CCD array is exposed to a
source and charge accumulates in the pixels. After the
defined exposure time, the accumulated signal is readout
of the array, digitized, and then transferred to the host
computer. Upon data transfer, the data is displayed and/or
stored via the application software. This sequence is
illustrated by the block diagram shown in Figure 12.
Whether or not the data is displayed and/or stored depends
on the data collection operation that has been selected in
the application software.
In WinX and LightField, the data collection operations Figure 12. Block Diagram of
use the Experiment Setup parameters to establish the PIXIS System
exposure time (the period when signal of interest is
allowed to accumulate on the CCD). Focus {Preview} is more likely to be used in
setting up the system (see the "First Light" discussions) and Acquire is then used for the
collection and storage of data. Briefly:
In Focus {Preview} mode, the number of frames is ignored. A single frame is
acquired and displayed, another frame is acquired and overwrites the currently
displayed data, and so on until Stop is selected. In WinX, the last frame acquired
before Stop is selected can be stored; in LightField, this frame cannot be stored.
Focus {Preview} mode is particularly convenient for familiarization and setting up.
For ease in focusing, the screen refresh rate should be as rapid as possible, achieved
by operating with axes and cross-sections off, and with Zoom 1:1 selected.
In Acquire mode, every frame of data collected can be automatically stored (the
completed dataset may include multiple frames with one or more accumulations).
This mode would ordinarily be selected during actual data collection. One limitation
of Acquire mode operation is that if data acquisition continues at too fast a rate for it
to be stored, data overflow may occur. In WinX, this could only happen in Fast Mode
operation.
33
34 PIXIS System Manual Issue 4
The remainder of this chapter provides "First Light" procedures (these provide step-by-
step instruction on how to initially verify system operation) and discusses factors that
affect exposure, readout, and digitization of the incoming signal. By understanding the
exposure, readout, and digitization factors and making adjustments to software settings
you can maximize signal-to-noise ratio. For information about synchronizing data
acquisition with external devices, please refer to Chapter 6, Advanced Topics.
Getting Started
1. Mount a test target in front of the camera.
2. Power ON the camera (i.e., switch the power supply ON).
Note: The camera must be turned on before WinX is opened, and WinX must be
closed before the camera is turned off.
3. Turn on the computer power.
4. Start the application software.
Setting the Parameters
Note: The following procedure is based on WinView/32: you will need to modify it
if you are using a different application. Basic familiarity with the WinView/32
software is assumed. If this is not the case, you may want to review the software
manual or have it available while performing this procedure.
Set the software parameters as follows:
Environment dialog (Setup|Environment): Check the DMA Buffer size.
Large arrays (2048x2048, for example) require a buffer size on the order of
32 Mbytes. If you change the buffer size, you will have to reboot the
computer for this memory allocation to be activated, and then restart
WinView.
Controller|Camera tab (Setup|Hardware): These parameters should be set
automatically to the proper values for your system.
Controller type: This information is read from the camera.
Camera type: This information is read from the camera.
Shutter type: None or Small (System dependent).
Readout mode: Available modes are read from the camera. Select Full
frame.
Detector Temperature (Setup|Detector Temperature…): The default
temperature setting is read from the camera. When the array temperature
reaches the set temperature, the Detector Temperature dialog will report
that the temperature is LOCKED. Note that some overshoot may occur. This
could cause temperature lock to be briefly lost and then quickly re-
established. If you are reading the actual temperature reported by the
application software, there may be a small difference between the set and
reported temperature when lock is established. This is normal and does not
indicate a system malfunction. Once lock is established, the temperature will
be stable to within ±0.05°C.
Note: The Detector Temperature dialog will not display temperature
information while you are acquiring data.
Cleans and Skips tab (Setup|Hardware): Click on Load Default Values
and click on Yes.
35
36 PIXIS System Manual Issue 4
Spectroscopy
The following paragraphs provide step-by-step instructions for operating PIXIS in a
spectroscopy setup for the first time. The intent of this simple procedure is to help you gain
basic familiarity with the operation of your system and to show that it is functioning
properly. Once basic familiarity has been established, then operation with other operating
configurations, ones with more complex timing modes, can be performed. An underlying
assumption for the procedure is that the camera is to be operated with a spectrograph (such
as an Acton Series 2300 spectrograph) on which it has been properly installed (refer to
Appendix D for instructions for mounting a spectrograph adapter to the PIXIS). A
suitable light source, such as a mercury pen-ray lamp, should be mounted in front of the
entrance slit of the spectrograph. Any light source with line output can be used. Standard
fluorescent overhead lamps have good calibration lines as well. If there are no "line"
sources available, it is possible to use a broadband source such as tungsten for the
alignment. If this is the case, use a wavelength setting of 0.0 nm for alignment purposes.
Assumptions
The following procedure assumes that
1. You have already set up your system in accordance with the instructions in
Chapter 4.
2. You have read the previous sections of this chapter.
3. You are familiar with the application software.
4. The system is being operated in spectroscopy mode.
5. The PIXIS does not have an internal shutter.
6. The spectrograph has an entrance slit shutter that is being controlled by the
PIXIS via the Shutter connector.
Caution If the PIXIS has an internal shutter, DO NOT USE the Shutter connector to
drive a spectrograph entrance slit shutter. If both the camera and the spectrograph
have shutters, the entrance slit shutter will have to be removed or a Princeton
Instruments SHC-EXT Shutter Control Box may be used to control the opening
and closing of the entrance slit shutter. See the spectrograph manual.
Getting Started
1. Set the spectrograph entrance slit width to minimum (10 µm if possible).
2. Power ON the spectrograph (i.e., switch the power supply ON).
3. Mount a light source at the spectrograph entrance slit.
4. Mount the camera to the spectrograph exit port.
5. Connect the shutter cable between the entrance slit shutter and the PIXIS Shutter
connector.
External Slit Shutter: A shutter assembly mounted externally to the
spectrograph has shutter cable that plugs into the Shutter connector.
37
38 PIXIS System Manual Issue 4
39
40 PIXIS System Manual Issue 4
2. With the spectrograph properly connected to the camera, turn the power on, wait for
the spectrograph to initialize.
3. Select Define from the Spectrograph menu, click on Install/Remove
Spectrograph, highlight the appropriate spectrograph name (for example, Acton
SP300i for an Acton SP-2300i spectrograph) in the Supported Spectrographs list,
and click on Install Selected Spectrograph.
4. Select Move from the Spectrograph menu, choose the grating to be move, and then
set it to 500 nm if using a mercury lamp or to 0.0 nm if using a broadband source.
Hint: Overhead fluorescent lights produce a mercury spectrum. Use a white card
tilted at 45 degrees in front of the entrance slit to reflect overhead light into the
spectrograph. Select 500 nm as the spectral line.
5. Set the slit to 10 µm at a minimum. If necessary, adjust the Exposure Time to
maintain optimum (near full-scale) signal intensity.
6. Wait until the detector temperature locks at its default temperature.
7. In WinSpec, select Focus (on the Acquisition menu or on the Experiment Setup
dialog) to begin data accumulation. Data will be continuously acquired and displayed but
will not be stored until you stop acquisition and use the Save function on the File menu.
8. Slowly move the camera in and out of focus. You should see the spectral line go from
broad to narrow and back to broad. Leave the camera set for the narrowest achievable
line. You may want to use the Focus Helper function (Process|Focus Helper…)
to determine the narrowest line width: it can automatically locate peaks and generate
a report on peak characteristics during live data acquisition (see the WinSpec/32
on-line help for more information).
Note that the way focusing is accomplished depends on the spectrograph, as follows:
Long focal-length spectrographs such as the Acton SP-2300: The
mounting adapter includes a tube that slides inside another tube to move the
camera in or out as required to achieve optimum focus.
Short focal-length spectrographs: There is generally a focusing
mechanism on the spectrograph itself which, when adjusted, will move the optics
as required to achieve proper focus.
No focusing adjustment: If there is no focusing adjustment, either provided
by the spectrograph or by the mounting hardware, then the only recourse will be
to adjust the spectrograph’s focusing mirror.
9. Next adjust the rotation. You can do this by rotating the camera while watching a live
display of the line (you may need to loosen two setscrews securing the spectrograph
adapter). Choose a peak to monitor during the rotational alignment. This peak will go
from broad to narrow and back to broad. Leave the camera rotation set for the
narrowest achievable peak.
Alternatively, take an image, display the horizontal and vertical cursor bars, and
compare the vertical bar to the line shape on the screen. Rotate the camera until the
line shape on the screen is parallel with the vertical bar.
41
42 PIXIS System Manual Issue 4
Note: When aligning other accessories, such as fibers, lenses, optical fiber adapters, first
align the spectrograph to the slit. Then align the accessory without disturbing the camera
position. The procedure is identical to that used to focus the spectrograph (i.e., do the
focus and alignment operations while watching a live image).
10. Tighten the spectrograph set screws to secure the spectrograph adapter. Then stop
data acquisition.
IsoPlane SCT-320 Spectrograph
Because the PIXIS is mounted directly to the mounting plate on the IsoPlane, focusing
and alignment is different from the way that focusing and alignment are performed for an
Acton Series spectrograph. The following information assumes that you are familiar with
the locations of the mounting plate, Micrometer Compartment, and the locking set screw.
If not, refer to the IsoPlane manual supplied with the spectrograph.
1. Mount an Acton light source such as the dual HG/NeAr source in front of the
entrance slit of the spectrograph.
2. With the spectrograph properly connected to the computer, turn the power on, wait
for the spectrograph to initialize.
3. With the PIXIS mounted to the spectrograph and connected to the computer, turn on
the power and wait for the detector to initialize.
4. Boot the WinSpec software. Select Define from the Spectrograph menu, click on
Install/Remove Spectrograph, highlight “Acton SCT320” in the Supported
Spectrographs list, and click on Install Selected Spectrograph.
5. Select Move from the Spectrograph menu, choose the grating to be move, and then
set it to 500 nm if using a mercury lamp or to 0.0 nm if using a broadband source.
6. Set the slit to 10 µm at a minimum. If necessary, adjust the Exposure Time to
maintain optimum (near full-scale) signal intensity.
7. Wait until the detector temperature locks at its default temperature.
8. Remove the cover from the Micrometer Compartment.
9. Using a 3/32” hex wrench, loosen the locking set screw.
10. Turn on Focus mode.
11. While continuously acquiring data, adjust the micrometer until you maximize the
intensity level of a selected peak or peaks.
12. Tighten down the locking set screw.
13. Place the Micrometer Cover on the spectrograph. Replace and tighten all of the cover
screws.
14. Next adjust the rotation. First, use a 9/64” hex wrench to loosen the four screws at the
corners of the detector mounting plate. While watching a live display of the
spectrum, select a peak to monitor and then rotate the detector (up to 4 degrees of
rotation are possible). The peak will go from broad to narrow and back to broad.
Leave the detector rotation set for the narrowest achievable peak.
Alternatively, take an image, display the horizontal and vertical cursor bars, and
compare the vertical bar to the line shape on the screen. Rotate the detector until the
line shape on the screen is parallel with the vertical bar.
15. After completing the rotational alignment, re-tighten the four mounting plate screws.
Then stop acquisition.
Chapter 5 Operation 43
43
44 PIXIS System Manual Issue 4
1. After LightField opens, you should see an icon representing your camera in the
Available Devices area. In the figure above, the camera is a PIXIS:512B
eXcelon®.
2. Drag the icon into the Experiment Devices area.
3. Click on the Run button to start Preview mode. In this mode, images will
be continuously acquired and displayed.
45
46 PIXIS System Manual Issue 4
Check the brightest regions of the image to determine if the A/D converter is at
full-scale. A 16-bit A/D is at full scale when the brightest parts of the image
reach an intensity of 65535. Adjust the aperture to where it is just slightly
smaller (higher f-stop) than the setting where maximum brightness on any part
of the image occurs.
Set the focus adjustment of the lens for maximum sharpness in the viewed
image.
5. After you have focused the camera, you can stop Preview mode, continue
Preview mode, or begin Acquire mode.
Spectroscopy
The following paragraphs provide step-by-step instructions for operating a PIXIS in a
spectroscopy setup in LightField for the first time. The intent of this simple procedure is
to help you gain basic familiarity with the operation of your system and to show that it is
functioning properly. Once basic familiarity has been established, then operation with
other operating configurations, ones with more complex timing modes, can be performed.
An underlying assumption for the procedure is that the camera is to be operated with a
spectrograph (such as an Acton Series 2300 spectrograph) on which it has been properly
installed (refer to Appendix D for instructions for mounting a spectrograph adapter to the
PIXIS). A suitable light source, such as a mercury pen-ray lamp, should be mounted in
front of the entrance slit of the spectrograph. Any light source with line output can be
used. Standard fluorescent overhead lamps have good calibration lines as well. If there
are no "line" sources available, it is possible to use a broadband source such as tungsten
for the alignment. If this is the case, use a wavelength setting of 0.0 nm for alignment
purposes.
Caution Overexposure Protection: Cameras that are exposed to room light or other
continuous light sources will quickly become saturated. If the camera is mounted to a
spectrograph, close the entrance slit of the spectrograph to reduce the incident light.
Assumptions
The following procedure assumes that
1. You have already set up your system in accordance with the instructions in
Chapter 4.
2. You have read the previous sections of this chapter.
3. You are familiar with the application software.
4. The system is being operated in spectroscopy mode.
5. The PIXIS does not have an internal shutter.
Chapter 5 Operation 47
6. The spectrograph has an entrance slit shutter that is being controlled by the
PIXIS via the Shutter connector.
Caution If the PIXIS has an internal shutter, DO NOT USE the Shutter connector to drive a
spectrograph entrance slit shutter. If both the camera and the spectrograph have
shutters, the entrance slit shutter will have to be removed or a Princeton Instruments
SHC-EXT Shutter Control Box may be used to control the opening and closing of the
entrance slit shutter. See the spectrograph manual.
Getting Started
1. Set the spectrograph entrance slit width to minimum (10 µm if possible).
2. Power ON the spectrograph (i.e., switch the power supply ON).
3. Mount a light source at the spectrograph entrance slit.
4. Mount the camera to the spectrograph exit port.
5. Connect the shutter cable between the entrance slit shutter and the PIXIS Shutter
connector.
External Slit Shutter: A shutter assembly mounted externally to the
spectrograph has shutter cable that plugs into the Shutter connector.
Internal Slit Shutter: A shutter mounted internally has an external shutter
connector in the sidewall of the spectrograph. Connect a shutter cable from
the PIXIS Shutter connector to that connector.
6. Power ON the camera (i.e., switch the power supply ON).
7. Turn on the computer power.
8. Start the application software.
Setting the Parameters
47
48 PIXIS System Manual Issue 4
1. After LightField opens, you should see icons representing your camera and the
spectrograph in the Available Devices area. In the figure above, the camera is a
PIXIS:400BR eXcelon® and the spectrograph is an SP-2356.
2. Drag the icons into the Experiment Devices area.
49
50 PIXIS System Manual Issue 4
Focusing
The detector mounting hardware provides two degrees of freedom: focus and rotation.
In this context, focus means to physically move the detector back and forth through the
focal plane of the spectrograph. The approach taken is to slowly move the detector in and
out of focus and adjusting for optimum while watching a live display on the monitor,
followed by rotating the detector and again adjusting for optimum. The following
procedure, which describes the focusing operation with an Acton SP-2300i spectrograph,
can be easily adapted to other spectrographs. For IsoPlane SCT-320 related focusing
information, see “IsoPlane SCT-320 Spectrograph” on page 53.
Acton Series Spectrograph
1. Click on the View tab, just above Experiment Devices, to change to the display
area.
6. Make sure that the spectroscopy-mount adapter moves freely at the spectrograph.
7. Select Align Spectrometer… from the Experiment Options menu. Review
the displayed information and then click on the Begin button. Typically, this
feature creates three 1-row high ROIs (one near the top of the array, one in the
middle, and one near the bottom) and begins data acquisition. Data will be
continuously acquired and displayed but will not be stored.
51
52 PIXIS System Manual Issue 4
9. Next adjust the rotation. You can do this by rotating the camera while watching a
live display of the line (you may need to loosen two setscrews securing the
spectrograph adapter). Click on the peak you want to monitor during the
rotational alignment. This positions the large cursor to provide a vertical
reference line across all of the ROIs.
10. Rotate the camera while watching the live display of the lines until the selected
peak is aligned horizontally in all of the ROIs. Tighten the setscrews securing the
spectrograph adapter at the spectrograph.
53
54 PIXIS System Manual Issue 4
Exposure Time
Exposure time, which is set on the Experiment Setup|Main tab {Common
Acquisition Settings expander}, is the time between start acquisition and stop
acquisition commands sent by the application software to the camera. In combination
with triggers, these commands control when continuous cleaning of the CCD stops and
when the accumulated signal will be readout. The continuous cleaning prevents buildup
of dark current and unwanted signal before the start of the exposure time. At the end of
the exposure time, the CCD is readout and cleaning starts again.
Because some PIXIS cameras do not incorporate an internal shutter, some signal may
accumulate on the array while it is being readout. This continuous exposure of the array
during readout may result in some smearing. However, exposures that are significantly
longer than the readout time can be performed without a shutter, as the amount of
smearing will be low.
If smearing or other factors require a shutter, the NOT SCAN {Not Reading Out} or the
SHUTTER {Shutter Open}signal at the LOGIC OUT connector (on the rear of the PIXIS)
can be used to control a customer-supplied external shutter. By using one of the signals to
synchronize the shutter operation with exposure, the CCD can be read out in darkness.
CCD Temperature
As stated before, lowering the temperature of the CCD
will generally enhance the quality of the acquired
signal. When WinX is the controlling software,
temperature control is done via the Detector
Temperature dialog (see Figure 23) accessed from the
Setup menu. When LightField is being used,
temperature control is done on the Sensor
Figure 23. WinX
expander.
Detector Temperature dialog
55
56 PIXIS System Manual Issue 4
Once the target array temperature {Temperature Setpoint} has been set, the software
controls the camera's cooling circuits to reach set array temperature. On reaching that
temperature, the control loop locks to that temperature for stable and reproducible
performance. When temperature lock has been reached (temperature within 0.05°C of set
value), the current temperature is Locked. The on-screen indication allows easy
verification of temperature lock.
The time required to achieve lock can vary over a considerable range, depending on such
factors as the camera type, CCD array type, ambient temperature, etc. Once lock occurs,
it is okay to begin focusing. However, you should wait an additional twenty minutes
before taking quantitative data so that the system has time to achieve optimum thermal
stability.
The deepest operating temperature for a system depends on the CCD array size and
packaging. Refer to Table 5, on page 93, for typical deepest cooling temperatures.
Note: In WinX, the Detector Temperature dialog will not display temperature
information while you are acquiring data.
Dark Charge
Dark charge (or dark current) is the thermally induced buildup of charge in the CCD over
time. The statistical noise associated with this charge is known as dark noise. Dark charge
values vary widely from one CCD array to another and are exponentially temperature
dependent. In the case of cameras with MPP type arrays, the average dark charge is
extremely small. However, the dark-charge distribution is such that a significant number
of pixels may exhibit a much higher dark charge, limiting the maximum practical
exposure. Dark charge effect is more pronounced in the case of cameras having a non-
MPP array (such as deep-depletion devices).
With the light into the camera completely blocked, the CCD will collect a dark charge
pattern, dependent on the exposure time and camera temperature. The longer the
exposure time and the warmer the camera, the larger and less uniform this background
will appear. Thus, to minimize dark-charge effects, you should operate with the lowest
CCD temperature possible.
Note: Do not be concerned about either the DC level of this background. What you see
is not noise. It is a fully subtractable bias pattern. Simply acquire and save a dark charge
"background image" under conditions identical to those used to acquire the "actual"
image. Subtracting the background image from the actual image will significantly reduce
dark-charge effects.
WARNING! If you observe a sudden change in the baseline signal, there may be excessive humidity in
the camera vacuum enclosure. Turn off the camera and contact Princeton Instruments
Customer Support. See page 126 for contact information.
Saturation
When signal levels in some part of the image are very high, charge generated in one pixel
may exceed the "well capacity" of the pixel, spilling over into adjacent pixels in a process
called "blooming." In this case a shorter exposure is advisable, with signal averaging to
enhance S/N (Signal-to-Noise ratio) accomplished through the software.
Chapter 5 Operation 57
For signal levels low enough to be readout-noise limited, longer exposure times, and
therefore longer signal accumulation in the CCD, will improve the S/N ratio
approximately linearly with the length of exposure time. There is, however, a maximum
time limit for on-chip accumulation, determined by either the saturation of the CCD by
the signal or the loss of dynamic range due to the buildup of dark charge in the pixels.
Readout
Introduction
After the exposure time has
elapsed, the charge
accumulated in the array pixels
needs to be read out of the
array, converted from electrons
to digital format, and
transmitted to the application
software where it can be
displayed and/or stored.
Readout begins by moving
charge from the CCD image
area to the shift register. The
charge in the shift register Figure 24. Array Terms for a CCD with a
pixels, which typically have Dual Output Amplifier
twice the capacity of the image
pixels, is then shifted into the output node and then to the output amplifier where the
electrons are grouped as electrons/count. This result leaves the CCD and goes to the
preamplifier where gain is applied.
WinX and LightField allow you to specify the type of readout (full frame or binned), the
output amplifier, and the gain (the number of electrons required to generate an ADU).
57
58 PIXIS System Manual Issue 4
t R texp tc (1)
where
tR is the CCD readout time,
texp is the exposure time, and
tc is the shutter compensation time.
Chapter 5 Operation 59
t N x N y t sr t v N x t i
R (2)
where
Nx is the smaller dimension of the CCD
Ny is the larger dimension of the CCD
tsr is the time needed to shift one pixel out of the shift register
tv is the time needed to digitize a pixel
ti is the time needed to shift one line into the shift register
A subsection of the CCD can be read out at full resolution, sometimes dramatically
increasing the readout rate while retaining the highest resolution in the region of interest
(ROI). To approximate the readout rate of an ROI, in Equation 2 substitute the x and y
dimensions of the ROI in place of the dimensions of the full CCD. Some overhead time,
however, is required to read out and discard the unwanted pixels.
Binning
Binning is the process of adding the data from adjacent pixels together to form a single
pixel (sometimes called a super pixel), and it can be accomplished in either hardware or
software. Rectangular groups of pixels of any size may be binned together, subject to
some hardware and software limitations.
Hardware Binning
Hardware binning is
performed on the CCD
array before the signal is
read out of the output
amplifier. For signal levels
that are readout noise
limited this method
improves S/N ratio linearly
with the number of pixels
grouped together. For
signals large enough to
render the camera photon
shot noise limited, the S/N
ratio improvement is
roughly proportional to the
square-root of the number
of pixels binned.
Binning also reduces
readout time and the
burden on computer
memory, but at the
expense of resolution.
Since shift register pixels
typically hold only twice
as much charge as image Figure 26. 2 × 2 Binning
59
60 PIXIS System Manual Issue 4
pixels, the binning of large sections may result in saturation and “blooming”, or spilling
of charge back into the image area.
Figure 26 shows an example of 2 2 binning. Each pixel of the image displayed by the
software represents 4 pixels of the CCD array. Rectangular bins of any size are possible.
Binning also reduces readout time and the burden on computer memory, but at the
expense of resolution. Since shift register pixels typically hold only twice as much charge
as image pixels, the binning of large sections may result in saturation and "blooming", or
spilling of charge back into the image area.
The readout rate for n n binning is approximated using a more general version of the
full resolution equation. The modified equation is:
t
sr t
tR
Nx N y v2 N x ti (3)
n n
Software Binning
One limitation of hardware binning is that the shift register pixels and the output node are
typically only 2-3 times the size of imaging pixels. Consequently, if the total charge binned
together exceeds the capacity of the shift register or output node, the data will be corrupted.
This restriction strongly limits the number of pixels that may be binned in cases where there
is a small signal superimposed on a large background, such as signals with a large
fluorescence. Ideally, one would like to bin many pixels to increase the S/N ratio of the weak
peaks but this cannot be done because the fluorescence would quickly saturate the CCD.
The solution is to perform the binning in software. Limited hardware binning may be used
when reading out the CCD. Additional binning is accomplished in software, producing a
result that represents many more photons than was possible using hardware binning.
Software averaging can improve the S/N ratio by as much as the square-root of the
number of scans. Unfortunately, with a high number of scans, i.e., above 100, camera 1/f
noise may reduce the actual S/N ratio to slightly below this theoretical value. Also, if the
light source used is photon-flicker limited rather than photon shot-noise limited, this
theoretical signal improvement cannot be fully realized. Again, background subtraction
from the raw data is necessary.
This technique is also useful in high light level experiments, where the camera is again
photon shot-noise limited. Summing multiple pixels in software corresponds to collecting
more photons, and results in a better S/N ratio in the measurement.
Array Orientation
For square format CCDs (for example, 512 × 512B or 1024 × 1024F/B) you may orient
the CCD to achieve binning along either direction of the CCD.
Binning along columns provides maximum scan rate.
Binning along the rows minimizes crosstalk and is therefore better for multi-
spectral applications.
Chapter 5 Operation 61
61
62 PIXIS System Manual Issue 4
Controller Gain
Controller gain (a function of the preamplifier) is software-selectable and is used to
change the relationship between the number of electrons acquired on the CCD and the
Analog-to-Digital Units (ADUs or counts) generated. Selecting the amount of gain is
done on the Acquisition|Experiment Setup…|ADC tab {Analog to Digital
Conversion expander}. The choices are 1 {Low}, 2 {Medium}, and 3 {High}. Users
who measure high-level signals may wish to select Low to allow digitization of larger
signals. Medium is suitable for experiments within the mid-level intensity range. Users
who consistently measure low-level signals may wish to select High, which requires
fewer electrons to generate an ADU and reduces some sources of noise.
The "Certificate of Performance" supplied with the camera lists the measured gain values
at all settings.
Example: The following descriptions assume the Low Noise Readout Port has been
selected and that the actual incoming light level is identical in all three instances. The
numbers used illustrate the effect of changing a controller gain setting and may not
reflect actual performance: gain at the 1, 2, and 3 settings depends on the CCD
installed.
1 {Low} requires four electrons to generate one ADU. Strong signals can be acquired
without flooding the CCD array. If the gain is set to Low and the images or spectra
appear weak, you may want to change the gain setting to Medium or High.
2 {Medium} requires two electrons to generate one ADU. If the gain is set to
Medium and the images or spectra do not appear to take up the full dynamic range of
the CCD array, you may want to change the gain setting to High. If the CCD array
appears to be flooded with light, you may want to change the setting to Low.
3 {High} requires one electron to generate one ADU and some noise sources are
reduced. Because fewer electrons are needed to generate an ADU, weaker signals can
be more readily detected. Lower noise further enhances the ability to acquire weak
signals. If the CCD array appears to be flooded with light, you may want to change
the setting to Medium or Low.
Chapter 5 Operation 63
Digitization (Rate)
Introduction
After gain has been applied to the signal, the Analog-to-Digital Converter (ADC)
converts that analog information (continuous amplitudes) into a digital data (quantified,
discrete steps) that can be read, displayed, and stored by the application software. The
number of bits per pixel is based on both the hardware and the settings programmed into
the camera through the software (see "Readout", page 57).
Factors associated with digitization include the digitization rate and baseline offset. The
speed at which digitization occurs is software-selectable but baseline offset is factory-set.
These factors are discussed in the following paragraphs.
63
64 PIXIS System Manual Issue 4
65
66 PIXIS System Manual Issue 4
Timing Modes
Overview
The basic PIXIS timing modes are Free Run {No Response}, External Sync {Readout
Per Trigger}, and External Sync {Readout Per Trigger} with Continuous Cleans {Clean
Until Trigger}. These timing modes are combined with the Shutter options to provide the
widest variety of timing modes for precision experiment synchronization.
Mode Shutter Condition
Free Run {No Response} Normal
External Sync {Readout Per Trigger} Normal
External Sync{Readout Per Trigger} PreOpen {Open Before Trigger}
External Sync {Readout Per Trigger} with Continuous
Normal
Cleans {Clean Until Trigger}
External Sync {Readout Per Trigger} with Continuous
PreOpen {Open Before Trigger}
Cleans {Clean Until Trigger}
67
68 PIXIS System Manual Issue 4
The PreOpen {Open Before Trigger} mode is useful in cases where an External Sync
pulse cannot be provided 8 ms or ~ 20 ms (the length of time the 25 mm or 45 mm
mechanical shutter takes to open) before the actual signal occurs. Its main drawback is
that the CCD is exposed to any ambient light while the shutter is open between frames. If
this ambient light is constant, and the triggers occur at regular intervals, this background
can also be subtracted, providing that it does not saturate the CCD. As with the Normal
Shutter {Normal} mode, accurate background subtraction may not be possible for the
first frame.
Also note that, in addition to signal from ambient light, dark charge accumulates during
the "wait" time (tw). Any variation in the external sync frequency also affects the amount
of dark charge, even if light is not falling on the CCD during this time.
Figure 31. Timing Diagram for External Sync Mode (+ edge trigger)
69
70 PIXIS System Manual Issue 4
71
72 PIXIS System Manual Issue 4
73
74 PIXIS System Manual Issue 4
Figure 37. Comparison of NOT SCAN {Not Reading Out}, SHUTTER {Shutter Open}, and NOT
READY {Busy} Logic Output Levels
Chapter 6 Advanced Topics 75
Kinetics Mode
In the past, Kinetics Mode was a purchased option for WinView/32. Kinetics Mode
is now included in the standard LightField and WinX applications.
Introduction
Kinetics mode uses the CCD to expose and store a limited number of images in rapid
succession. The time it takes to shift each line (or row) on the CCD is as short as a few
hundred nanoseconds to few microseconds, depending on the CCD. Therefore the time
between images can be as short as a few microseconds. Kinetics mode allows full
frame CCDs to take time-resolved images/spectra. Optical or mechanical masking of
the array is required.
Note: In WinX, if the Kinetics option has been installed in the PIXIS, this readout mode
will be made available when you select the appropriate camera type on the Hardware
Setup dialog.
Below is a simplified illustration of kinetics mode. Returning to our 4 × 6 CCD example,
in this case 2/3 of the array is masked, either mechanically or optically. The shutter opens
to expose a 4 × 2 region. While the shutter remains open, charge is quickly shifted just
under the mask, and the exposure is repeated. After a third image is collected the shutter
is closed and the CCD is read out. Since the CCD can be read out slowly, very high
dynamic range is achieved. Shifting and readout are shown in Figure 38.
75
76 PIXIS System Manual Issue 4
Figure 39. Hardware Setup dialog Figure 40. Experiment Setup dialog
Figure 41. Sensor Readout expander: Figure 42. Shutter and Trigger expanders:
Kinetics Readout Mode No Response
Free Run (Non-Triggered Mode)
In the Free Run {No Response} kinetics mode, the PIXIS takes a series of images, each with
the Exposure time set through the software (in WinX, the exposure time is set on the
Experiment Setup|Main tab; in LightField on the Common Acquisition Settings
expander). The time between image frames, which may be as short as a few microseconds, is
limited by the time required to shift an image under the mask: this interimage time equals the
Vertical Shift rate (specified in ns/row) multiplied by the Window Size (the number of rows
allocated for an image frame). The exact number of frames depends on the selected Window
Size and is equal to the number of pixels perpendicular to the shift register divided by the
Chapter 6 Advanced Topics 77
Window Size. Integrate signals (SHUTTER {Shutter Open}) or Readout signals (NOT
SCAN {Not Reading Out}) are provided at the LOGIC OUT connector for timing
measurements.
Example: Referring to the readout shown in Figure 38, there are 6 pixels perpendicular to
the shift register and the Window Size is 2 pixels high. The number of frames is 3. If the
Vertical Shift rate for the CCD is 1600 ns/row, the Shift time will be 3200 ns per frame.
77
78 PIXIS System Manual Issue 4
In multiple trigger mode, the shutter is opened when Acquire or Focus {Run} is clicked
and each “exposure-shift” cycle in the acquisition is triggered independently by a pulse
applied at the EXT SYNC connector. This mode is useful when each subframe needs to
be synchronized with a pulsed external light source such as a laser. Once the series is
complete, the shutter closes and readout begins. Since the shutter is open during the entire
series of images, irregularly spaced external pulses will result in exposures of different
lengths. Once the series has been read out, the camera is ready for the next series. This
timing is shown in Figure 45, where a series of 6 frames is collected with 6 External Sync
pulses.
Custom Modes
Introduction
Custom Chip and Custom Timing modes are available as an option for the WinX
applications; {Custom Sensor} and Custom Timing are standard (sensor and readout
mode-dependent) in LightField. These modes are intended to allow data acquisition at the
fastest possible rates for your camera. Custom Chip {Custom Sensor} allows you to
reduce the apparent size of the CCD array and Custom Timing allows you to select a
faster vertical shift time.
Custom Chip
Introduction
In addition to Binning and ROI (previously discussed in the manual), there is a third way
to reduce Readout Time – Custom Chip {Custom Sensor}. This feature allows you to
redefine the size of the CCD’s active area via software. Unlike setting a smaller region of
interest (ROI), which also involves reading out fewer pixels, this mode does not incur
overhead from discarding or skipping the rest of the rows. And, unlike both Binning and
ROI, Custom Chip {Custom Sensor} also relies on some form of array masking to ensure
that no light falls outside the currently set active area.
Note: Custom Chip is a WinX option. Custom Sensor is standard with LightField:
availability is sensor-dependent.
Chapter 6 Advanced Topics 79
Software Settings
Princeton Instruments does not encourage users to change these parameter settings. For
most applications, the default settings will give the best results. We strongly advise
contacting the factory for guidance before customizing the chip definition.
79
80 PIXIS System Manual Issue 4
Custom Timing
Notes:
1. This mode is standard with LightField for full frame CCD cameras.
2. This option is fully supported by WinX Version 2.5.18.1 (and higher).
3. This option is not supported by WinX for 5 MHz systems..
In LightField, Custom Timing is accessed via the Custom Sensor button on the
Sensor expander. In the Custom Timing panel, you can select from among the listed
vertical shift rate choices (see Figure 48). Vertical shift rate is the time required to shift
one row into the serial register. The smaller the value, the faster charge will be shifted up
one row at a time toward the serial register. In WinX, if the Custom Timing option is
present and selected, the equivalent function is located on the Custom Timing tab on
the Hardware Setup dialog (see Figure 49).
Figure 48. LightField: Custom Timing Figure 49. WinX: Vertical Shift
Chapter 7
Troubleshooting
WARNING! Do not attach or remove any cables while the camera system is powered on.
Introduction
The following issues have corresponding troubleshooting sections in this chapter.
81
82 PIXIS System Manual Issue 4
6. If you later re-run the Camera Detection Wizard, the name will be changed back to
the default name (i.e., Camera1).
83
84 PIXIS System Manual Issue 4
1. Unscrew the reservoir cap (on top of the CoolCUBEII) and make sure that the coolant
reservoir contains coolant. If additional coolant is required, fill with a 50:50 mixture
of water and ethylene glycol.
2. Screw the reservoir cap back in.
3. Make sure the power switch is turned off before plugging the circulator in.
4. Plug the circulator into a 100-240 VAC, 47-63 Hz power source.
5. Turn the circulator on. Make sure there are no leaks or air bubbles in the hoses.
Note: Small air bubbles (about the size of bubbles in soda) are common in the
CoolCUBEII especially at start up and do not prevent proper operation.
Cooling Troubleshooting
Temperature Lock Cannot be Achieved or Maintained.
Possible causes for not being able to achieve or maintain lock could include:
Ambient temperature greater than +23°C. This condition affects TE-cooled
cameras. If ambient is greater than +23°C, you will need to cool the camera
environment or raise the set temperature.
Airflow through the camera and/or circulator is obstructed. The camera needs to
have approximately two (2) inches (50 mm) clearance around the vented covers.
If there is an enclosure involved, the enclosure needs to have unrestricted flow to
an open environment. The camera vents its heat out the vents near the nose. The
air intake is near the rear of the camera.
A hose is kinked. Unkink the hose.
Coolant level is low. Add coolant. See “CoolCUBEII: Low Coolant (Air in Hoses)”
on page 84.
There may be air in the hoses. Add coolant. See “CoolCUBEII: Low Coolant
(Air in Hoses)” on page 84.
Circulator pump is not working. If you do not hear the pump running when the
CoolCUBEII is powered on, turn off the circulator and contact Customer Support.
The circulator is higher than the camera. Reposition the circulator so that it is 6
inches (150 mm) or more below the camera. The vertical distance should not exceed
10 feet (3 m). Typically, the camera is at table height and the circulator is on the
floor.
The camera vacuum has deteriorated and needs to be refreshed. Contact
Customer Support.
The target array temperature is not appropriate for your particular camera and CCD
array.
For a TE-cooled camera, the camera's internal temperature may be too high, such as
might occur if the operating environment is particularly warm or if you are
attempting to operate at a temperature colder than the specified limit. TE-cooled
cameras are equipped with a thermal-protection switch that shuts the cooler circuits
down if the internal temperature exceeds a preset limit. Typically, camera operation
is restored automatically in about ten minutes. Although the thermo-protection
switch will protect the camera, you are nevertheless advised to power down and
correct the operating conditions that caused the thermal-overload to occur.
85
86 PIXIS System Manual Issue 4
87
88 PIXIS System Manual Issue 4
Cancel the load. Cancelling a load means that the last used experiment will not be
loaded automatically when LightField opens. However, you can load the experiment
after all the devices are available, you can start a new experiment design, or you can
load a different experiment that matches the devices you are using.
Device is Occupied
Multiple instances of LightField can be running at the same time.
However, a device currently being used by one instance of LightField
will be shown in the Available Devices area as “Occupied” for all other
instances of LightField. To make a device available to the current
instance of LightField, either remove it from the Experiment Devices
area in the other instance or close the instance that is using the device. Figure 54. Occupied
Device icon
Error Creating Controller message
89
90 PIXIS System Manual Issue 4
Shutter Failure
See "Overexposed or Smeared Images", page 88.
Appendix A
Basic Specifications
Note: This appendix provides some of the basic specifications of a PIXIS system. If the
information you are looking for is not here, it may be available in Appendix B, "Outline
Drawings" or on the appropriate data sheet. Data sheets can be downloaded from the
Princeton Instruments website (www.princetoninstruments.com).
Window
SI-UV fused-silica quartz (.125"/3.17 mm thick)
CCD Arrays
Princeton Instruments offers a wide-variety of CCDs for scientific imaging and
spectroscopy. Contact Princeton Instruments or visit www.princetoninstruments.com for an
up-to-date list of arrays supported by the PIXIS.
Mounts
C-mount: Standard threaded video mount. Optional adjustable C- to Spectroscopy-mount
kinetics adapter
F-mount: Standard Nikon® bayonet mount.
Spectroscopy mount: 3.60" (91.44 mm) or 3.88" (98.55 mm) bolt circle. Optional
Spectroscopy-to-F mount adapter
91
92 PIXIS System Manual Issue 4
Shutter
Typically, PIXIS cameras for imaging applications are shipped with an internal shutter.
PIXIS:512/1024: Internal, windowless, 1 in (25 mm) aperture, 8 ms open time, 8 ms
close time
PIXIS:2048: Internal, windowless, 1.8 in (45 mm) aperture, ~ 20 ms open time, ~48 ms
close time
Typically, PIXIS cameras for spectroscopy applications do not have an installed internal
shutter. An optional Princeton Instruments-supplied 25 or 45 mm external shutter can be
ordered if the camera has no internal shutter.
Camera
Cooling: Thermoelectric; air or circulating coolant
Coolant Ports: The inlet/outlet ports on the side of the camera allow you to connect the
camera to a CoolCUBEII with PIXIS-compatible hoses (PN 7567-0002). Use only the
hoses and circulator shipped with your system. Attaching any other hoses or
circulator voids the warranty
Gain: Software-selectable (high, medium, low)
Dimensions: See Appendix B.
Connectors:
EXT SYNC (MCX): 0 to +3.3 V logic level input to allow data acquisition to be
synchronized with external events. Trigger edge can be positive- or negative-
going as set in software. Synchronization and Timing Modes are discussed in
Chapter 6. MCX-to-BNC adapter cable supplied with system.
LOGIC OUT (MCX): 0 to +3.3 V logic level output for monitoring camera status.
Logic output is software-selectable as NOT SCAN, SHUTTER, NOT READY,
LOGIC 0, or LOGIC 1. Logic Out Control is discussed in Chapter 6. MCX-to-
BNC adapter cable supplied with system.
NOT SCAN {Not Reading Out}: Logic low when CCD is being read;
otherwise high.
SHUTTER {Shutter Open}: The output precisely brackets shutter-open
time (exclusive of shutter compensation) and can be used to control an
external shutter or to inhibit a pulser or timing generator.
NOT READY {Busy}: The output changes state on completion of the array
cleaning cycles that precede the first exposure. Initially high, it goes low to
mark the beginning of the first exposure. In free run operation, it remains low
until the system in halted. If a specific number of frames have been
programmed, it remains low until all have been taken, then returns high.
LOGIC 0 {Always Low) and LOGIC 1 {Always High}: Logic low and
logic high, respectively.
Shutter (LEMO): Optional. Used for connecting to a Princeton Instruments-supplied
external shutter. Cable not supplied.
Appendix A Basic Specifications 93
USB 2.0 (USB B Female): Data link to computer via USB cable inserted at this
connector. Cable length of 5 meters is standard. Other lengths may be available.
Contact Customer Support for more information. USB A/B cable supplied with
system.
Power (DIN): 12 VDC (6.6A max) input from power supply.
Pin 3, 5: 12 VDC
Pin 1, 2, 4: Return
Shell: Ground
93
94 PIXIS System Manual Issue 4
Options
A partial listing of options includes: internal 25 mm shutter, internal 45 mm shutter, C-to-
F-mount adapter, adjustable C- to Spectroscopy-mount kinetics adapter, and a fiber optic
extender kit. Contact the factory for more information regarding options available for
your system.
Appendix B
Outline Drawings
NOTE: Dimensions are in inches [mm].
95
96 PIXIS System Manual Issue 4
97
98 PIXIS System Manual Issue 4
99
100 PIXIS System Manual Issue 4
101
102 PIXIS System Manual Issue 4
103
104 PIXIS System Manual Issue 4
Figure 68. Spectroscopy-Mount: Internal Shutter (3.60" and 3.88" bolt circles) (Air-cooled)
105
106 PIXIS System Manual Issue 4
CoolCUBEII Circulator
Tools Required
Spanner (supplied)
Small flat blade screwdriver
Procedure
1. Using a small flat blade screwdriver, loosen the
lock screws.
2. Use the spanner to rotate the C-mount adapter the
desired distance.
3. Tighten the lock screws to lock the C-mount
adapter in place.
Note: To tighten down the screws, the face of the
adapter should be no further than .06" (1.5 mm)
out from the front surface of the camera nose.
Caution
The C-mount lens thread-depth should be .21" (5.3 mm) or less. Otherwise, depending on
the adapter in-out location, the lens could bottom out and damage the shutter, if there is
one installed. If you are not certain of the thread depth, remove the adapter from the
camera, thread the lens into the adapter until the lens threads are flush with the back
surface of the adapter. Note the depth at the front surface, remove the lens, and then re-
insert the adapter into the camera nose.
107
108 PIXIS System Manual Issue 4
* The screws are #4-40 setscrews. A 0.050 hex key is required to loosen or tighten them.
Appendix C Adapter Adjustment and Focusing Procedures 109
109
110 PIXIS System Manual Issue 4
111
112 PIXIS System Manual Issue 4
113
114 PIXIS System Manual Issue 4
115
116 PIXIS System Manual Issue 4
117
118 PIXIS System Manual Issue 4
119
120 PIXIS System Manual Issue 4
WinX LightField
Minimum Block Size Final Section Height
Multiple Trigger (Kinetics) Shift Per Trigger (Kinetics)
Normal Shutter Normal (Shutter)
Number of Blocks Final Section Count
Number of Cleans Number of Clean Cycles
Number of Strips per Clean Clean Cycle Height
Post-Dummy Rows Parallel to Shift Register Active Area: Bottom Margin
Post-Dummy Shift Register Columns Active Area: Right Margin
Pre-Dummy Rows Parallel to Shift Register Active Area: Top Margin
Pre-Dummy Shift Register Columns Active Area: Left Margin
PreOpen (Shutter) Open Before Trigger (Shutter)
Readout Port Quality
Shutter Close Compensation Time Closing Delay
Shutter Control Shutter Mode
Single Trigger (Kinetics) Readout Per Trigger
Skip Serial Register Clean (deselected) Clean Serial Register
Target Temperature Temperature Setpoint
Timing Mode Trigger Response
Window Size Kinetics Window Height
Appendix E Cross-Referencing of WinX and LightField Terms 121
LightField-to-WinX
LightField WinX
Active Area: Bottom Margin Post-Dummy Rows Parallel to Shift Register
Active Area: Left Margin Pre-Dummy Shift Register Columns
Active Area: Right Margin Post-Dummy Shift Register Columns
Active Area: Top Margin F.T. Dummies or Frame Transfer Dummies
Active Area: Top Margin Pre-Dummy Rows Parallel to Shift Register
Active Height Active Rows Parallel to Shift Register
Active Width Active Shift Register Columns
Always Closed (Shutter) Disabled Closed (Shutter)
Always Open (Shutter) Disabled Open (Shutter)
Analog Gain Controller Gain
Bit Depth ADC Resolution
Clean Cycle Height Number of Strips per Clean
Clean Serial Register Skip Serial Register Clean (deselected)
Clean Until Trigger Continuous Cleans
Closing Delay Shutter Close Compensation Time
Custom Sensor Custom Chip
Custom Timing Custom Timing
Final Section Count Number of Blocks
Final Section Height Minimum Block Size
Kinetics Window Height Window Size
N/A Disable Pixel Bias Correction (PBC)
No Response Free Run
No Response (Kinetics) Free Run (Kinetics)
Normal (Shutter) Normal Shutter
Number of Clean Cycles Number of Cleans
Open Before Trigger (Shutter) PreOpen (Shutter)
Output Signal Logic Out
Output Signal: Always High Logic Out: Logic 1
Output Signal: Always Low Logic Out: Logic 0
Output Signal: Busy Logic Out: Not Ready
121
122 PIXIS System Manual Issue 4
LightField WinX
Output Signal: Not Reading Out Logic Out: Not Scan
Output Signal: Shutter Open Logic Out: Shutter
Preview Focus
Quality Readout Port
Readout Per Trigger External Sync
Readout Per Trigger (Kinetics) Single Trigger (Kinetics)
Sensor Readout Region expander functions Easy Bin
Shift Per Trigger (Kinetics) Multiple Trigger (Kinetics)
Shutter Mode Shutter Control
Speed ADC Rate
Temperature Setpoint Target Temperature
Trigger Determined By Edge Trigger
Trigger Response Timing Mode
Warranty & Service
Limited Warranty
Princeton Instruments, a division of Roper Scientific, Inc. ("Princeton Instruments", "us",
"we", "our") makes the following limited warranties. These limited warranties extend to
the original purchaser ("You", "you") only and no other purchaser or transferee. We have
complete control over all warranties and may alter or terminate any or all warranties at
any time we deem necessary.
Basic Limited One (1) Year Warranty
Princeton Instruments warrants this product against substantial defects in materials and /
or workmanship for a period of up to one (1) year after shipment. During this period,
Princeton Instruments will repair the product or, at its sole option, repair or replace any
defective part without charge to you. You must deliver the entire product to the Princeton
Instruments factory or, at our option, to a factory-authorized service center. You are
responsible for the shipping costs to return the product. International customers should
contact their local Princeton Instruments authorized representative/distributor for repair
information and assistance, or visit our technical support page at
www.princetoninstruments.com.
123
124 PIXIS System Manual Issue 4
Your Responsibility
The above Limited Warranties are subject to the following terms and conditions:
1. You must retain your bill of sale (invoice) and present it upon request for service
and repairs or provide other proof of purchase satisfactory to Princeton
Instruments.
2. You must notify the Princeton Instruments factory service center within (30)
days after you have taken delivery of a product or part that you believe to be
defective. With the exception of customers who claim a "technical issue" with the
operation of the product or part, all invoices must be paid in full in accordance
with the terms of sale. Failure to pay invoices when due may result in the
interruption and/or cancellation of your one (1) year limited warranty and/or any
other warranty, expressed or implied.
3. All warranty service must be made by the Princeton Instruments factory or, at our
option, an authorized service center.
4. Before products or parts can be returned for service you must contact the
Princeton Instruments factory and receive a return authorization number (RMA).
Products or parts returned for service without a return authorization evidenced by
an RMA will be sent back freight collect.
5. These warranties are effective only if purchased from the Princeton Instruments
factory or one of our authorized manufacturer's representatives or distributors.
6. Unless specified in the original purchase agreement, Princeton Instruments is not
responsible for installation, setup, or disassembly at the customer’s location.
7. Warranties extend only to defects in materials or workmanship as limited above
and do not extend to any product or part which:
has been lost or discarded by you;
has been damaged as a result of misuse, improper installation, faulty or
inadequate maintenance or failure to follow instructions furnished by us;
has had serial numbers removed, altered, defaced, or rendered illegible;
has been subjected to improper or unauthorized repair;
has been damaged due to fire, flood, radiation, or other "acts of God" or other
contingencies beyond the control of Princeton Instruments; or
is a shutter which is a normal wear item and as such carries a onetime only
replacement due to a failure within the original 1 year Manufacturer
warranty.
8. After the warranty period has expired, you may contact the Princeton Instruments
factory or a Princeton Instruments-authorized representative for repair
information and/or extended warranty plans.
9. Physically damaged units or units that have been modified are not acceptable for
repair in or out of warranty and will be returned as received.
125
126 PIXIS System Manual Issue 4
10. All warranties implied by state law or non-U.S. laws, including the implied
warranties of merchantability and fitness for a particular purpose, are expressly
limited to the duration of the limited warranties set forth above. With the exception
of any warranties implied by state law or non-U.S. laws, as hereby limited, the
forgoing warranty is exclusive and in lieu of all other warranties, guarantees,
agreements, and similar obligations of manufacturer or seller with respect to the
repair or replacement of any parts. In no event shall Princeton Instruments' liability
exceed the cost of the repair or replacement of the defective product or part.
11. This limited warranty gives you specific legal rights and you may also have other
rights that may vary from state to state and from country to country. Some states
and countries do not allow limitations on how long an implied warranty lasts,
when an action may be brought, or the exclusion or limitation of incidental or
consequential damages, so the above provisions may not apply to you.
12. When contacting us for technical support or service assistance, please refer to the
Princeton Instruments factory of purchase, contact your authorized Princeton
Instruments representative or reseller, or visit our Support page at
www.princetoninstruments.com.
Contact Information
Roper Scientific's manufacturing facility for this product is located at the following
address:
Princeton Instruments
3660 Quakerbridge Road
Trenton, NJ 08619 (USA)
Tel: 1-800-874-9789 / 1-609-587-9797
Fax: 1-609-587-1970
Customer Support E-mail: techsupport@princetoninstruments.com
For immediate support in your area, please call the following locations directly:
North America 1-800- 899-1144 (toll free) or 1-609-587-9797
France +33 (1) 60 86 03 65
Germany +49 (0) 89 660 7793
Japan +81 (3) 5639 2741
UK & Ireland +44 (0) 1628 472 346
Singapore +65 6408 6240
China +86 10 659 16460
Otherwise, see our Support web page at www.princetoninstruments.com.
An up-to-date list of addresses and telephone numbers is posted on the
www.princetoninstruments.com/Support page. In addition, links on this
page to support topics allow you to send e-mail based requests to the
Customer Support group.
Index
A Binning (continued)
A/D converters 63, 93 on-chip 59
AC power requirements 23 readout time 59, 60
Accessories, alignment of 42, 52 resolution loss 59, 60
Acquisition started, viewer contents do not update 82 software 60
Adapter instructions 112, 113, 114 effect on S/N ratio 60
Acton Series spectrographs 112, 113, 114, 115, 116 high light level measurements 60
adjustable C- to spectroscopy-mount shot-noise limited measurements 60
7050-0104 115 well capacity 60
7050-0107 116 Blooming 56
IsoPlane SCT-320 spectrograph 117 Busy output level 74
Adapters
adjustable C- to Spectroscopy-mount kinetics adapter C
7050-0104 115 C- to Spectroscopy-mount kinetics adapter 29
7050-0107 116 7050-0104 115
ADC offset 63 7050-0107 116
Adjustable C- to Spectroscopy-mount kinetics Cables
adapter 16 MCX to BNC adapter 14
Adjustable C- to Spectroscopy-mount kinetics USB 2.0 14
adapter 29 Calibration
Adjustable C- to Spectroscopy-mount kinetics spectroscopy
adapter suitable light sources 42, 53
7050-0104 115 Calibration, spectrometer
Adjustable C- to Spectroscopy-mount kinetics suitable light sources 40, 50
adapter Camera Detection wizard 27
7050-0107 116 Cautions
Adjustable C-mount adapter 28, 107 excessive humidity in CCD chamber 56, 63
Air-circulation requirement 13 scintillator & UV 8
Always High output level 74 CCD array
Always Low output level 74 blooming 56
Analog gain See Controller gain dark charge effects 56
ANSI C library 15 functions performed 54
maximum on-chip integration 57
B readout theory 58
Background DC level 56 shift register 58
Background subtraction 68 signal-to-noise ratio vs on chip integration time 57
Baseline square format 60
offset 63 theory of operation 54
signal 56, 63 well capacity 56
troubleshooting 82 Certificate of Performance 14
Baseline signal 56 Clean Until Trigger 69
Binning Cleaning
along columns 60 detector 9
along rows 60 optical surfaces 9
array orientation 60 C-mount
computer memory burden 59, 60 adapters 28, 107
hardware 59 adjustable 107
lens installation/removal 28
127
128 PIXIS System Manual Issue 3
E I
Environmental conditions 22 Installation
Environmental requirements 22 PCI drivers 24, 25
Error Creating Controller message 88 software 24, 25
eXcelon 7 IntelliCal 15
Exposure 54 Interface card
shutter 54 driver installation 24, 25
time 66 Internal shutter 16
Exposure and Readout 33 IsoPlane SCT-320 spectrograph adapter instructions
Exposure time 55 117
Index 129
K P
Kinetics Performance certificate 14
multiple trigger 78 PICam 15
single trigger 77 Power cord 13
Kinetics adapter 29 Power requirements 23
7050-0104 115 Preopen Shutter mode 68
7050-0107 116 Procedures
C-mount adjustment 107
L F-mount focusing 108
LEMO connector 12 lens focusing 109
Lenses, mounting 29 spectrograph adapter installation 111
LightField 15
LOGIC 0 output level 74, 92 R
LOGIC 1 output level 74, 92 Readout
Logic Out levels binning 59
Always High 74 hardware 59
Always Low 74 software 60
Busy 74 subsection of array 59
LOGIC 0 74, 92 time 66
LOGIC 1 74, 92 Readout Per Trigger 67, See also External Sync
Not Reading Out 74 Readout rate
NOT READY 74, 92 control of 63
NOT SCAN 74, 92 precision vs speed tradeoff 63
SHUTTER 74, 92 Requirements
Shutter Open 74 environmental 22
Low noise amplifier 61 ventilation 13
Resolution, loss of with binning 59, 60
M
Maintenance 9 S
MCX to BNC adapter cables 14 S/N ratio 57, 60
Mercury spectrum, fluorescent lights 41, 48, 50 Safe mode
Mounting as used for setting up 72
C- to Spectroscopy-mount kinetics adapter 29 fast image update 72
spectrograph 30 flowchart 73
Multiple Trigger mode 78 missed events 72
Safety related symbols used in manual 10
N Saturation 56
Not Reading Out 67 Serial violation 86
output level 74 Serial Violations message 90
NOT READY output level 74, 92 Shift register 58
NOT SCAN Shutter
output level 74, 92 compensation time 54
timing 54 effect on exposure 54
exposure 54
O external shutter connector 12
Operating temperature 93 lifetime 16, 31
Option modes
Custom Chip 78 Disable 66
Custom Timing 80 Normal 66
Outline drawings 95 Preopen 66, 68
Output amplifier signs of failure 88
dual 57, 61 Shutter mode
High Capacity 61 Open Before Trigger 69
Low Noise 61 PreOpen 69
Overexposure protection 28, 32, 46 Shutter Open output level 74
129
130 PIXIS System Manual Issue 3
T V
Technical support 126 Vacuum deterioration 86
Temperature Ventilation requirements 13
control problems 85 Vertical shift rate 80
operating 93
specifications 93 W
thermal cutout switch 85 Warnings
Temperature control 55 cleaning 9
effect of vacuum deterioration 86 protective grounding 8
Temperature lock 56 replacement power cord 8
Thermal cutout switch 85 touching the CCD array 8
Timing control 72 Warranties
Timing modes 66 image intensifier detector 124
Trigger modes one year 123
multiple trigger (Kinetics) 78 one year on refurbished/discontinued products 123
single trigger (Kinetics) 77 owner's manual and troubleshooting 125
Troubleshooting sealed chamber 124
acquisition start but viewer contents do not update 82 software 124
baseline change 82 vacuum integrity 124
camera stops working 82 XP vacuum chamber 123
controller not responding 84 x-ray detector 124
CoolCUBEII your responsibility 125
air in hoses 84 Website 126
low coolant 84 Well capacity 56
data loss 86 blooming 56
data overrun (hardware conflict) 86 restrictions on hardware binning 60
data overrun message 87 saturation 56
Device is not found message 87 Wizard, Camera Detection 27