Pacemaker
Pacemaker
Pacemaker
Management
Cardiac Pacemaker
Introduction
A pacemaker system is a device capable of generating artificial pacing impulses and delivering them to the heart. It consists of a pulse generator and appropriate electrodes. In the past few years electronic pacemaker systems have become extremely important in saving and sustaining the lives of cardiac patients whose normal pacing function of the heart have been impaired.
2
A pacemaker stimulates the heart muscle with precisely timed discharges of electricity that cause the heart to beat in a manner very similar to a naturally occurring heart rhythm.
3
Indications
Conduction disorders
2nd and 3rd degree heart block
Rate disorders
Asystole Symptomatic sinus bradycardia Sick sinus syndrome
Prophylaxes
Post cardiac surgery Back up
4
Pacing system
Unipolar Bipolar
10
Menu
11
12
Insertion
External: in emergency at the bedside
14
Setting Parameters
Rate Sensitivity threshold Stimulation threshold (output)
15
Setting Parameters
Rate
Fixed: stimulus provided at a preset rate (grater than patients rate) Demand: stimulus provided when the patients heart falls below a predetermined rate (proper sensing is required)
16
Setting Parameters
Sensitivity threshold:
regulates the ability of the pacemaker to detect the hearts intrinsic electrical activity (senses R wave) Measured in mV Sense indicator flashes when sensing inherent patients signal Sensitivity set 2-3 times (or half) lower than Sense indicator Threshold
17
Setting Parameters
Stimulation threshold (output)
The electrical current that is delivered to the heart to initiate depolarization It is measured in milliampers (mA) Threshold: the point at which depolarization occurs and identified by a myocardial response to pacing (capture) Set to double the minimum amount of mA needed to achieve 100% capture
18
19
0=None A=Atrium
V=Ventricle
D=Dual (A+V)
V=Ventricle
D=Dual (A+V)
I=Inhibited
D=Dual (T+I)
20
Universal (Dual) DDD Atrial & Ventricular pacing & sensing, inhibition for both intrinsic P & QRS, triggered response to sensed P wave to allow for rate responsive ventricular pacing
22
23
Pacing srtips
A: atrial pacing
B: Ventricular pacing
25
Failure to Fire
Spikes are not noted during period of asystole or bradycardia Causes
Loose connection son the system Failure of battery or pulse generator Broken lead wires Lead wire dislodgment Assure pacing connections Replace battery or generator as appropriate Reposition leading wire Attempt pacing with another pacing system
26
Interventions
Failure to Fire
27
Failure to Capture
Spike is not followed by a P or QRS complex as appropriate Causes
Loose connections on the system Failure of battery or pulse generator Broken lead wires Lead wire dislodgment or fibrous at site of electrodes low pacing threshold (output)
Interventions
Assure pacing connections Check threshold and increase output mAm Repositioning the patient may also resolve the problem
28
Failure to Capture
29
Failure to capture
30
31
Undersensing
Inability of the pacemaker to sense spontaneous myocardial depolarization Pacemaker looses its ability for self-inhibition Competition between the paced complexes and intrinsic hearts rhythm occur Demonstrated on the ECG by a pacing spike occurring after or unrelated to intrinsic QRSs It is a serious malfunction that could lead to dangerous ventricular dysrhythmia as VT & VF
32
Undersensing
Causes
Inadequate QRS signal (QRS signal not detected by pacemaker, low sensitivity) Myocardial ischemia, fibrosis, electrolytes disturbances Inappropriate mode selection (asynchronous)
Intervention
Increase sensitivity (moving the sensitivity dial toward its lowest setting)
33
Undersensing
Undersensing. This transcutaneous temporary pacemaker set in the ventricular demand mode fires and paces appropriately in the beginning of the strip. The 9th complex is the patients inherent QRS complex which should have been sensed by the pacemaker. Instead the pacemaker fired.
34
Undersensing
35
Oversensing
Inappropriate sensing of patients QRS The pacemaker thinks it detects a QRS complex so it inhibit itself and does not fire Result in unexplained pauses in the ECG traces Causes
Tall P or T waves Electrical signals produced by skeletal muscle contractions (during shivering or seizures) EMI
Intervention
Decrease sensitivity (moving the sensitivity dial toward its highest setting, 20 mV) Place a magnet over the generator to restore pacing (in permanent pacemakers only)
36
Oversensing
37
Menu
38
Patients Management
ECG monitoring Hemodynamic monitoring Frequent assessment of pacemaker Electrical safety Pacing insertion site care
Cleaning, dressing, signs of infection