This document provides instructions for calibrating common volumetric glassware used in analytical chemistry including volumetric flasks, pipettes, and burets. It explains that glassware must be calibrated to avoid systematic errors in measurements and outlines the general calibration procedure which involves weighing the amount of water contained or delivered, converting this to a volume using the density of water, and correcting to the standard temperature of 20°C. Tips for proper use of each type of glassware are also provided to reduce random errors.
1 of 10
More Related Content
calibration
1. CHEM 311L
Quantitative Analysis Laboratory
Revision 2.3
Calibration of Volumetric Glassware
In this laboratory exercise, we will calibrate the three types of glassware typically used by an
analytical chemist; a volumetric flask, a volumetric pipet and a buret. Over the course of this
semester, we will use these tools extensively when performing Gravimetric and Titrimetric
Analyses. In order to avoid introducing Systematic Errors into our measurements, each of these
instruments must be properly calibrated. And, to reduce the Random Errors inherent when using
these instruments, their proper use must be thoroughly understood. The quality of the
measurements obtained from these tools depends heavily on the care taken in calibrating and in
using each instrument.
Volumetric Flask (http://en.wikipedia.org/wiki/File:Brand_volumetric_flask_100ml.jpg)
Pipettes (http://www.chem.yorku.ca/courses/chem1000/equipment/pipette.html)
Buret (http://en.wikipedia.org/wiki/File:Burette_vertical.svg)
In precise work it is never safe to assume that the volume delivered by or contained in any
volumetric instrument is exactly the amount indicated by the calibration mark. Instead,
recalibration is usually performed by weighing the amount of water delivered by or contained in
the volumetric apparatus. This mass is then converted to the desired volume using the tabulated
density of Water:
Volume = mass / density (Eq. 1)
2. P a g e | 2
All volumetric apparati should be either purchased with a Calibration Certificate or calibrated by
the analyst in this manner.
Systematic Errors Affecting Volumetric Measurements
The volume occupied by a given mass of liquid varies with temperature, as does the volume of
the device that holds the liquid. 20o
C has been chosen as the normal temperature for calibration
of much volumetric glassware.
Glass is a fortunate choice for volumetric ware as it has a relatively small coefficient of thermal
expansion; a glass vessel which holds 1.00000L at 15o
C holds 1.00025L at 25o
C. If desired, the
volume values (V) obtained at a temperature (t) can be corrected to 20o
C by use of:
V20 = V [1 + 0.000025 (20 - t)] (Eq. 2)
In most work, this correction is small enough it may be ignored.
However, the thermal expansion of the contained liquid is frequently of importance. Dilute
aqueous solutions have a coefficient of thermal expansion of about 0.025%/o
C. A liter of water
at 15o
C will occupy 1.002L at 25o
C. A correction for this expansion must frequently be applied
during calibration procedures.
Parallax is another source of error when using volumetric ware. A correction for this expansion
must frequently be applied during calibration procedures. Frequently, graduation marks encircle
the apparatus to aid in this.
(Quantitative Analysis, 4th
Ed. by Conway Pierce,
Edward L. Haenisch and Donald T. Sawyer; John Wiley
& Sons; 1948.)
Readings which are either too high or too low will result otherwise.
3. P a g e | 3
Tips for Correct Use of Volumetric Glassware
Pipets
The Pipet is used to transfer a volume of solution from one container to another. Most
Volumetric Pipets are calibrated To-Deliver (TD); with a certain amount of the liquid
remaining in the tip and as a film along the inner barrel after delivery of the liquid. The
liquid in the tip should not be blown-out. Pipets of the "blow-out" variety will usually
have a ground glass ring at the top. And, drainage rates from the pipet must be carefully
controlled so as to leave a uniform and reproducible film along the inner glass surface.
Measuring Pipets will be gradated in appropriate units.
Once the pipet is cleaned and ready to use, make sure the outside of the tip is dry. Then
rinse the pipet with the solution to be transferred. Insert the tip into the liquid to be used
and draw enough of the liquid into the pipet to fill a small portion of the bulb. Hold the
liquid in the bulb by placing your fore finger over the end of the stem.
(Quantitative Analysis, 4th
Ed. by Conway Pierce,
Edward L. Haenisch and Donald T. Sawyer; John Wiley
& Sons; 1948.)
Withdraw the pipet from the liquid and gently rotate it at an angle so as to wet all
portions of the bulb. Drain out and discard the rinsing liquid. Repeat this once more.
To fill the pipet, insert it vertically in the liquid, with the tip near the bottom of the
container. Apply suction to draw the liquid above the graduation mark. Quickly place a
fore finger over the end of the stem. Withdraw the pipet from the liquid and use a dry
paper to wipe off the stem. Now place the tip of the pipet against the container from
which the liquid has been withdrawn and drain the excess liquid such that the meniscus is
at the graduation mark.
Move the pipet to the receiving container and allow the liquid to flow out (avoiding
splashing) of the pipet freely. When most of the liquid has drained from the pipet, touch
the tip to the wall of the container until the flow stops and for an additional count of 10.
4. P a g e | 4
Volumetric Flasks
The Volumetric Flask is used to prepare Standard Solutions or in diluting a sample. Most
of these flasks are calibrated To-Contain (TC) a given volume of liquid. When using a
flask, the solution or solid to be diluted is added and solvent is added until the flask is
about two-thirds full. It is important to rinse down any solid or liquid which has adhered
to the neck. Swirl the solution until it is thoroughly mixed. Now add solvent until the
meniscus is at the calibration mark. If any droplets of solvent adhere to the neck, use a
piece of tissue to blot these out. Stopper the flask securely and invert the flask at least 10
times.
Burets
The Buret is used to accurately deliver a variable amount of liquid. Fill the buret to
above the zero mark and open the stopcock to fill the tip. Work air bubbles out of the tip
by rapidly squirting the liquid through the tip or tapping the tip while solution is draining.
The initial buret reading is taken a few seconds, ten to twenty, after the drainage of liquid
has ceased. The meniscus can be highlighted by holding a white piece of paper with a
heavy black mark on it behind the buret.
(Quantitative Analysis, 4th
Ed. by Conway Pierce,
Edward L. Haenisch and Donald T. Sawyer; John Wiley
& Sons; 1948.)
Place the flask into which the liquid is to be drained on a white piece of paper. (This is
done during a titration to help visualize color changes which occur during the titration.)
The flask is swirled with the right-hand while the stopcock is manipulated with the left-
hand.
(Quantitative Analysis, 4th
Ed. by Conway Pierce,
Edward L. Haenisch and Donald T. Sawyer; John Wiley
& Sons; 1948.)
5. P a g e | 5
The buret should be opened and allowed to drain freely until near the point where liquid
will no longer be added to the flask. Smaller additions are made as the end-point of the
addition is neared. Allow a few seconds after closing the stopcock before making any
readings. At the end-point, read the buret in a manner similar to that above.
As with pipets, drainage rates must be controlled so as to provide a reproducible liquid
film along the inner barrel of the buret.
Cleaning Volumetric Glassware
Cleaning of volumetric glassware is necessary to not only remove any contaminants, but to
ensure its accurate use. The film of water which adheres to the inner glass wall of a container as
it is emptied must be uniform.
Two or three rinsings with tap water, a moderate amount of agitation with a dilute detergent
solution, several rinsings with tap water, and two or three rinsings with distilled water are
generally sufficient if the glassware is emptied and cleaned immediately after use.
If needed, use a warm detergent solution (60-70o
C). A buret or test tube brush can be used in the
cleaning of burets and the neck of volumetric flasks. Volumetric flasks can be filled with
cleaning solution directly. Pipets and burets should be filled by inverting them and drawing the
cleaning solution into the device with suction. Avoid getting cleaning solution in the stopcock.
Allow the warm cleaning solution to stand in the device for about 15 minutes; never longer than
20 minutes. Drain the cleaning solution and rinse thoroughly with tap water and finally 2-3
times with distilled water.
Pipets and burets should be rinsed at least once with the solution with which they are to filled
before use.
A General Calibration Procedure
As was noted above, volumetric glassware is calibrated by measuring the mass of Water that is
Contained In or Delivered By the device.
To obtain an accurate mass measurement, buoyancy effects must be corrected for. The amount
of air displaced by the standard weights of the balance is somewhat different than the amount of
air displaced by the weighed water. This difference leads to different buoyancies for these
objects; meaning the balance levels at a point other than when the two objects are of the same
mass. This can be corrected for using:
mtrue = mmeas + da ( (mmeas/d) – (mmeas/ds) ) (Eq. 3)
6. P a g e | 6
where ds is the density of the standard weights (8.47 g/cm3
), da is the density of air ( 0.0012
g/cm3
), and d is the density of the object being measured.
This mass data is then converted to volume data using the tabulated density of Water (See
Appendix) at the temperature of calibration. (In very accurate work, the thermometer must also
be calibrated as an incorrect temperature reading will lead to the use of an incorrect density for
Water. This, in turn, will give an inaccurate volume calibration.)
Finally, this volume data is corrected to the standard temperature of 20o
C. This can be
accomplished using the thermal expansion coefficient of Water; 0.00025/o
C:
V20 = V [1 + 0.00025 (20 - t)] (Eq. 4)
Further details concerning calibration of laboratory glassware can be found in the NIST
publication “The Calibration of Small Volumetric Laboratory Glassware” by Josephine
Lembeck; NBSIR 74-461. This publication can be found at:
http://ts.nist.gov/MeasurementServices/Calibrations/upload/74-461.PDF
Thus, in this exercise we will calibrate a volumetric flask and a pipet and determine a buret
Correction Factor by calibrating each of these devices with Water. In each case, the measured
mass of the calibrating Water will be corrected for buoyancy effects and the resulting volume
will be standardized to 20o
C.
7. P a g e | 7
Procedures
Begin by cleaning a 5 mL or 25 mL Volumetric Pipet, a 50 mL Buret, and a 25 mL
Volumetric Flask according to the procedure outlined above. It is imperative for the
purposes of calibration that these glassware items be cleaned such that Water drains
uniformly and does not leave breaks or droplets on the walls of the glass.
If detergent solutions are not sufficient to clean your glassware, a Cleaning Solution
(Dichromate in Conc. Sulfuric Acid) may be used. Consult you instructor before taking
this step.
Once cleaned, the Buret should be filled with Distilled Water and clamped in an upright
position and stored in this manner until needed. The Volumetric Flask should be clamped
in an inverted position so that it may dry.
Calibration of a Pipet
Use your cleaned pipet. Note if this is a Class A or other device. Weigh a receiving container on
the Analytical Balance; a 100 mL plastic beaker with Aluminum Foil cover. Pipet distilled water
into the plastic beaker and reweigh it.
Record the temperature of the water used.
Repeat the procedure at least 2 more times. Dry the plastic beaker and re-weigh it for each
replication. (Are you pipeting consistently and correctly?)
Calculate the apparent mass and the buoyancy corrected mass of the water delivered for each
time you pipet. From this mass, and the density of water at the given temperature (See
Appendix), calculate the volume of the water delivered. Correct the volume to 20o
C. Calculate
the Average, Standard Deviation and 90% Confidence Interval for your calibration result.
Is your result within the listed tolerance for this pipet? (See Appendix. What is the better
question to ask?)
Calibration of a Buret
Use your cleaned 50mL buret. Note if this is a Class A or other device. Fill the buret with
water. Make sure the tip is free of bubbles. Drain into a waste beaker until it is at, or just below,
the zero mark. Allow 10-20 seconds for drainage. Make an initial reading to a precision of at
least 0.01 mL. Test for tightness of the stopcock by allowing the buret to stand for 5 minutes and
then re-reading the volume. There should be no noticeable change in the reading.
8. P a g e | 8
Once the tightness of the stopcock is assured, refill the buret and again drain into a waste until it
is at, or just below, the zero mark. Allow for drainage. Touch the tip of the buret to the wall of
the waste beaker to remove the pendent drop of water. Make a volume reading.
Weigh a receiving container on the Analytical Balance; a 100 mL plastic beaker with Aluminum
Foil cover. Drain about 5 mL of water from the buret into the beaker. Allow 10-20 seconds for
drainage. Touch the tip of the buret to the wall of the beaker to again remove the pendent drop.
Read the buret and weigh the water.
Calculate the actual volume of water delivered by the buret in the same manner as outlined above
in the procedure on calibrating pipets. Calculate the Correction Factor by subtracting the
apparent volume delivered, as given by the buret readings, from the actual volume delivered.
Repeat the procedure at least once more. The two Correction Factors should agree within 0.04
mL. If they do not, repeat the procedure again. Report the average Correction Factor for 5 mL.
Repeat this process for 15 mL, 25 mL, 35 mL, and 45 mL delivered.
Plot the Average Buret Correction Factor vs. Volume Delivered using Excel or some other
graphing software.
Label and store your buret properly; upright and filled with Distilled Water. This is the
buret you will use for the remainder of the course.
Calibration of a Volumetric Flask
Use your cleaned 25mL volumetric flask. Note if this is a Class A or other device. Weigh the
flask empty. Fill the flask to the mark and re-weigh it. Measure the temperature of the water
used.
Repeat the procedure at least twice more.
Calculate the true volume of the flask using the method outlined above. Report the Average,
Standard Deviation, and 90% Confidence Interval for this result.
Is your result within the listed tolerance for this flask?
9. P a g e | 9
Appendix - Density of Water
Temperature (o
C) Density (g/mL)
10 0.9997026
11 0.9996084
12 0.9995004
13 0.9993801
14 0.9992474
15 0.9991026
16 0.9989460
17 0.9987779
18 0.9985986
19 0.9984082
20 0.9982071
21 0.9979955
22 0.9977735
23 0.9975415
24 0.9972995
25 0.9970479
26 0.9967867
27 0.9965162
28 0.9962365
29 0.9959478
30 0.9956502
10. P a g e | 10
Appendix - Tolerances for Class A Volumetric Glassware at 20o
C
Pipets
Capacity (mL) Tolerances (mL)
0.5 0.006
1 0.006
2 0.006
5 0.01
10 0.02
20 0.03
25 0.03
50 0.05
100 0.08
Volumetric Flasks
Capacity (mL) Tolerances (mL)
5 0.02
10 0.02
25 0.03
50 0.05
100 0.08
250 0.12
500 0.20
1000 0.30
2000 0.50
Burets
Capacity (mL) Tolerances (mL)
5 0.01
10 0.02
25 0.03
50 0.05
100 0.20
With the exception of Graduated Cylinders, the Tolerances for Class B devices is
typically twice that of a Class A device. (ASTM E694)