Abstract
Flexible pressure sensors can convert mechanical stimuli to electrical signals to interact with the surroundings, mimicking the functionality of the human skins. Piezocapacitive pressure sensors, a class of most widely used devices for artificial skins, however, often suffer from slow response-relaxation speed (tens of milliseconds) and thus fail to detect dynamic stimuli or high-frequency vibrations. Here, we show that the contact-separation behavior of the electrode-dielectric interface is an energy dissipation process that substantially determines the response-relaxation time of the sensors. We thus reduce the response and relaxation time toâ~0.04âms using a bonded microstructured interface that effectively diminishes interfacial friction and energy dissipation. The high response-relaxation speed allows the sensor to detect vibrations over 10âkHz, which enables not only dynamic force detection, but also acoustic applications. This sensor also shows negligible hysteresis to precisely track dynamic stimuli. Our work opens a path that can substantially promote the response-relaxation speed of piezocapacitive pressure sensors into submillisecond range and extend their applications in acoustic range.
Similar content being viewed by others
Introduction
The perception of touch of the human skin is enabled by mechanoreceptors that respond to not only static forces (by slow adaptors) but also vibrational stimuli (by fast adaptors)1. Electronic skins or flexible pressure sensors are emerging devices that mimic the functionalities of the mechanoreceptors2,3,4,5, which have been widely studied because of their potential applications in the fields of robot haptics6,7,8, human-machine interfaces9,10, intelligent wearables11,12,13, and metaverse14,15,16. Many applications, such as texture recognition, sound recognition, and pressure/vibration detection, require sensors to respond to both static pressure and high-frequency vibrations up to thousands of hertz.
Piezocapacitive flexible pressure sensors are a class of the most widely studied sensing devices that can detect static pressure, but these devices perform insufficiently in responding to dynamic stimuli. While elastic elastomers can respond to mechanical stimuli in nanoseconds17,18,19, conventional piezocapacitive flexible pressure sensors often exhibit a response-relaxation time on the level of tens of milliseconds, corresponding to a narrow frequency range up to tens of hertz. This low response-relaxation speed is primarily attributed to energy dissipation associated with viscoelastic materials and interfacial frictions. Soft dielectrics are typically viscoelastic materials that dissipate energy during loading-unloading cycles. Such an energy loss gets more pronounced when softer materials are used for detecting subtle pressures20,21. Also, during the contact-separation process, the interfacial friction and adhesion between the electrode and dielectric further contribute to energy loss22,23. To improve the response-relaxation speed, a common strategy is to engineer the dielectric layer with microstructured surfaces20. This strategy works through two principles. First, the microstructures reduce the bulk viscoelasticity of the dielectric by storing more elastic energy in smaller deformations. Second, they reduce the contact area between the dielectric and electrode, thereby lowering energy dissipation due to interfacial friction and adhesion. However, despite the reduced energy dissipation by introducing microstructures, the response-relaxation time remains largely above 1âms to date. This limitation seems to be unreconcilable as long as viscoelastic materials are used and interfacial gaps persist. Although there are a few very recent advances reporting sensors with a shorter response-relaxation time down to a few milliseconds6,24,25,26, such sensors can still not be used to detect high-frequency vibrations of hundreds or thousands of hertz, and thus the application of the sensors to high-frequency or acoustic purpose is still unavailable.
In this work, we present a strategy for downscaling the response-relaxation time of flexible piezocapacitive pressure sensors to ~0.04âms by seamlessly bonding a low-viscosity microstructured dielectric with the electrode. The dielectric is made by dispersing 2âwt.% carbon nanotubes (CNTs) within a polydimethylsiloxane (PDMS) matrix, which reduces the material viscosity and surface adhesion. Without interfacial gaps, the bonded microstructured interfaces substantially diminish the friction-induced energy dissipation. We show that our sensor can quickly respond to stimuli from steady pressures to high-frequency vibrations over 10âkHz. In addition, the sensor also exhibits a high frequency-resolution of 0.2âHz at 1000âHz, and negligible capacitance-pressure hysteresis. Such behaviors enable its applications for dynamic pressure detection including acoustic scenarios. We further designed an artificial ear system based on the sensor and used the system for sound detection. We expect that our sensor to be used in more applications that require the detection of both static pressure and high-vibrational stimuli, and the method of using a bonded interface to improve response-relaxation speed might be extended to other devices.
Results
Microstructured pressure sensors with a bonded interface
Conventional microstructured pressure sensors have a gap between the electrode and the dielectric (Fig. 1a), while our sensors have a bonded electrode-dielectric interface (Fig. 1b). To compare their different response-relaxation behavior, we fabricated two sensors. The top and bottom electrodes are made by dispersing 7âwt.% carbon nanotubes (CNTs) in polydimethylsiloxane (PDMS), while the middle dielectric layer is made of PDMS with 2âwt.% CNT fillers. Microstructured dielectric layers were fabricated by molding a 3D-printed template of microcones (Supplementary Fig. 1). The bonded interface is further realized by immersing electrodes and dielectric in trichloromethane solvent with uncured PDMS networks, followed by a curing process to form topological interlinks (Supplementary Fig. 2)27.
To elucidate the reduced energy dissipation of bonded interfaces, we performed finite element analysis (FEA) in which two sensors are compressed to 100âkPa and then recovered (Fig. 1c and Supplementary Fig. 3). The bottom area of the microcone and the increased electrode-dielectric contact area is denoted as D and ÎA, respectively. For the non-bonded sensor with a gap, ÎA increases quickly upon loading, as shown in Fig. 1d, and the stress concentrates at contacted regions (Fig. 1c). The corresponding unloading process is shown in Supplementary Fig. 3. As a result, the friction contributes to substantial energy loss (denoted as Wloss) that is higher than the maximum elastic energy (denoted as Welastic) during loadingâunloading process (Fig. 1e). On the contrary, the bonded pressure sensor only shortens under compression with negligibly smaller ÎA than that of the non-bonded counterpart (Fig. 1d). Therefore, the energy loss is significantly reduced (Fig. 1e), leading to an elevated response-relaxation time of the bonded sensor.
The bonded pressure sensor exhibits a variation of capacitance ÎC to a broad pressure range of 0â350âkPa (Fig. 1f) with exceptionally rapid response-relaxation speed. Figure 1g shows that both the response time and the relaxation time are 0.04âms, which is tested by a customized circuit board (Supplementary Figs. 4 and 5, Supplementary Text 1). The maximum frequency that a sensor can detect is determined as: \({{{{{{\rm{f}}}}}}}_{\max }=1/\left({{{{{{\rm{t}}}}}}}_{{{{{{\rm{res}}}}}}}+{{{{{{\rm{t}}}}}}}_{{{{{{\rm{rel}}}}}}}\right)\), where tres and trel represent response time and relaxation time, respectively. Existing piezo-capacitive sensors mostly exhibit a response-relaxation time of tens of milliseconds, and the corresponding fmax is on the order of tens of hertz. Here, the total response-relaxation time of 0.08âms allows the sensor to respond to high frequencies up to 12,500âHz. By contrast, the sensor with a non-bonded interface exhibits a much longer response time of 4.76âms and a relaxation time of 6.55âms, which allows the sensor to respond to a frequency limit of only 88âHz. Moreover, our sensor outperforms existing capacitive pressure sensors in both response and relaxation speed (Fig. 1h) by one to two orders of magnitude compared with existing capacitive pressure sensors with a non-bonded interface3,7,24,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42. Note that the response time and relaxation time of piezocapacitive sensors are correlated and exhibit close values (the dashed line in Fig. 1h). In general, the ratio of relaxation time to response time is usually greater than or close to 1. In addition to the wide frequency bandwidth up to 12,500âHz, our sensor exhibits a low detection limit of 0.007âPa, which allows the sensor to detect tiny stimuli such as sound. By contrast, existing sensors can hardly balance both a low detection limit and a wide frequency bandwidth (shown as in Fig. 1i)28,35,43,44,45,46,47,48,49.
Effect of microcones on response-relaxation time
The bonded microcones play a key role in reducing the energy dissipation during the contact-separation process, thus increasing the response-relaxation speed. The effect of microcone structure on the response-relaxation time is further investigated by FEA. The structure of a single microcone in 2D configuration can be described by three parameters: height H, top surface A0 (i.e., initial contact area), and bottom surface D (Fig. 2a). Twenty different microcones are investigated by varying A0/Dâ=â0.2, 0.4, 0.6, 0.8, 1 and H/Dâ=â0.5, 1, 1.5, 2. With a fixed A0/Dâ=â0.4, we first study the effect of height H on the energy dissipation by varying H/D. Results in Fig. 2b show that sharper microcones (i.e., larger H/D) show less energy loss. However, sharp microcones may also undergo buckling instability that undermines the mechanical stability of the sensor. For a tapered column with clamped top/bottom surfaces, the critical buckling force can be calculated by \(\frac{4{\pi }^{2}E{I}_{D}}{{H}^{2}}\frac{{A}_{0}^{2}}{{D}^{2}}\), where E is the Youngâs modulus and \({I}_{D}=\frac{\pi {D}^{4}}{64}\)50. The critical buckling pressure (denoted as \({P}_{c}\)) of 20 microcones are summarized in Fig. 2c. Considering that the maximum applied pressure should be greater than 300âkPa, microcones with a small A0/D and a large H/D should not be adopted (marked with blue in Fig. 2c).
Second, we analyze how initial contact area A0 affects the energy dissipation by setting H/Dâ=â1 and varying A0/D. The normalized energy loss in Fig. 2d suggests that larger A0/D yields a lower energy loss. This conclusion is expected because stout microcones (i.e., larger A0/D) create a smaller contact area under the same pressure while preserving a larger elastic energy Welastic. However, we may not rush to conclude that microcone with A0/Dâ=â1 (i.e., a cylindrical micropillar) is the best by far, because the flat dielectric is notorious for its low sensitivity20, i.e., the slope of ÎC/C0 versus pressure curve. As also revealed by our models in Fig. 2e, we show that the sensitivity gradually drops when A0/D increases (Supplementary Text 2), leading to an undermined detection range due to the saturated signals, as validated by experiments (dotted data in Fig. 2e). Therefore, the tradeoff between response-relaxation speed and sensitivity should also be considered. By summarizing the normalized energy loss of 20 microcones in Fig. 2f with background colors indicating other metrics, we show that microcones with moderate values of (A0/D, H/D)â=â(0.4, 1), (0.6, 1) and (0.6, 1.5) can simultaneously achieve low energy dissipation, high sensitivity, and high mechanical stability. Note that it is difficult to fabricate bonded microcones with perfect geometries, and microcones used in this work have close values of (A0/D, H/D)âââ(0.4, 1) (see SEM in Fig. 1b). In addition to the structure of microcones, the bonding ratio of microcone arrays also affect the response-relaxation time of the sensor. In this regard, we compare the performance of six sensors with bonding ratios of 0, 20, 40, 60, 80, and 100% (Fig. 2g). Results in Fig. 2h manifest that the 100% bonded ratio provides the lowest response-relaxation time of ~0.04âms, which is higher than previous sensors without fully bonded interfaces (Supplementary Fig. 6)27.
Effect of materials on response-relaxation time
Viscoelastic materials are known for their energy dissipation during loading-unloading cycles. To reduce the viscosity of the dielectric, we employ a highly crosslinked PDMS matrix with a low base-to-curing agent ratio of 5:1, into which CNT fillers are further introduced. The choice of CNTs is due to their stable mechanical and superior electrical properties. Notably, the PDMS-CNTs compositeâs relative permittivity is significantly higher than that of pure PDMS, thereby enhancing the capacitance signal51. As shown in Fig. 3a, the 5:1 PDMS has a lower ratio of loss modulus to storage modulus (Eâ/Eâ) compared to the 15:1 PDMS. By further doping CNTs, the PDMS-CNTs composite exhibits even lower Eâ/Eâ as the doping wt.% increases. The high viscosity usually leads to a large adhesion force, thereby prolonging the separation (Supplementary Fig. 7). This phenomenon is further supported by the measured adhesion force of various materials using an atomic force microprobe (Fig. 3b). Additionally, the bonded microstructured interface contributes to the reduced interfacial adhesion, facilitating faster separation. We measured the adhesion strength between the 2âwt.% PDMS-CNTs dielectric and the electrode (7âwt.% PDMS-CNTs) under a 100âkPa load over 10âmin. As shown in Fig. 3c, the adhesion strength of the bonded interface is negligibly small (~0.01âkPa), whereas the flat interface and the non-bonded interface exhibit much higher adhesion strength values of 28âkPa and 0.23âkPa, respectively.
The combination of low-viscosity 2âwt.% PDMS-CNTs dielectric with a bonded microstructured interface leads to ultrafast response and relaxation time of the sensor, as shown in Fig. 3d, e. Our sensor stands out substantially compared with sensors with higher viscosity and non-bonded interfaces. It achieves both response and relaxation times as low as 0.04âmsâthis is an order of magnitude faster than its counterparts. Low viscosity and bonded interface also contribute to low hysteresis and high mechanical stability of the sensor. Figure 3f, g shows substantially low hysteresis during a loading-unloading cycle to 100âkPa, and there is barely a capacitance-pressure hysteresis loop. The low hysteresis is related to the negligible viscous dissipation of the interface. Furthermore, the bonded microstructured interface exhibits a high interfacial toughness of 530âJmâ2 measured by 180° peeling  test (Supplementary Fig. 8). Such high interfacial toughness allows the sensor to work stably under certain harsh mechanical conditions. Figure 3h shows that the sensor exhibits a highly stable signal when subjected to repeated rubbing with a shear stress of ~45âkPa and a pressure of 200âkPa over a displacement of 2âmm for 10,000 cycles.
Frequency bandwidth and frequency resolution of the pressure sensor
Sensors with rapid response and relaxation times can detect high-frequency vibrations. We use a testing system (Fig. 4a) to verify the frequency detection range. The experimental setup includes a computer-controlled function generator, a power amplifier, and a vibration generator. When vibrations of 12,500âHz are generated, the sensor can detect the signal with a period of 0.08âms (Fig. 4b, Supplementary Fig. 9), and the corresponding Fourier transform result shows a peak at 12,500âHz (Fig. 4c). Furthermore, the sensor can detect signals of vibrations at different amplitudes at 12,500âHz (Supplementary Fig. 10).
Sensors with a non-bonded interface have a much narrower frequency range compared to those with bonded interfaces. Figure 4d shows the capacitance response to vibrations with frequencies of 200, 400, 1000, 3000, 6000, 9000, and 12,000âHz of the two sensors, which use the same materials and similar device configurations except that the former has bonded interfaces. Whereas the sensor with bonded interfaces can detect all those vibrations, the sensor with a non-bonded interface fails to detect vibrations of 400âHz with high fidelity, as indicated in both time domain signals and corresponding frequency domain analysis by short-time Fourier transform (Fig. 4e) or by fast Fourier transformation (Supplementary Fig. 11). Note that although the frequencies of 200âHz and 400âHz can be detected using the sensor with non-bonded interface, the output signals are distorted since fmax of the sensor is only 73âHz.
Our sensor also identifies high-frequency vibrations under pre-applied static pressure or low-frequency stimuli. At a static pressure of 100âkPa, we apply an extra vibrational signal of 500, 4000, 8000, and 12,500âHz, and the sensor effectively detects the superimposed static and vibrational signals for all cases (Fig. 4f). Corresponding results in the frequency domain also well reflect vibrational signals of these frequencies (Supplementary Fig. 12). In addition, our sensor can detect vibrational stimuli with two substantially different frequencies (30 and 1000âHz) and amplitudes (Supplementary Fig. 13), which may enable the recognition of complex vibrational signals, such as sound recognition.
Frequency resolution is an index that determines the performance of the sensor to discriminate close frequencies. Figure 4g shows vibrational signals with frequencies of 1000.0, 1000.2, 1000.4, and 1000.6âHz from the first to the 1000th period in the time domain detected using the sensor. Although the signals have close frequencies, they can be resolved in both the time domain and frequency domain (Fig. 4g), which clearly shows the peaks of 1000.0, 1000.2, 1000.4, and 1000.6âHz. The result indicates a superior frequency resolution of 0.2âHz at 1000âHz, or 2âÃâ10â4 (Fig. 4h).
Sound detection using the sensor
We use the sensor in an artificial ear for sound detection. The human ear receives, transmits, and converts sound to physiological signals by the eardrum, the auditory ossicle, and the cochlea, respectively. Here, our artificial system uses a polyethylene naphthalate (PEN) membrane as the eardrum, the sensor as the cochlea, and a 3D printed aluminum auditory ossicle being connected to the artificial eardrum and the sensor, as shown in Fig. 5a. Sound waves cause the artificial tympanic membrane to vibrate, and the sensor converts the vibration to capacitance signals. The artificial tympanic membrane (area: 314âmm2) and the aluminum auditory ossicle (with an end dimension of 4âmm) can also amplify the sound to enhance the pressure detection capability of the sensor. Figure 5b and Supplementary Fig. 14 show that the system can respond to a low sound pressure of 0.007âPa, and the signal amplitude (âC, change in capacitance) increases linearly with increasing sound pressure in the range of 0.007â0.16âPa. Similarly, at a given frequency (we select 200âHz because it is the resonant frequency of the system, which will be discussed hereinafter), âC also increases monotonically with increasing sound intensity ranging from 50 to 110âdB because a higher sound intensity generates a larger sound pressure. Note that no hysteresis is observed during the process of increasing and decreasing sound intensity (Fig. 5c). Furthermore, at a fixed sound intensity of 98âdB, the system exhibits a nonlinear response with a frequency up to 5000âHz, and a resonant frequency of 200âHz is observed (Fig. 5d).
We have used our artificial ear for audio recording and compared the results with that of a commercially available microphone and that of the system using a non-bonded sensor. Figure 5e presents the recorded signals of the song âBury Me Not On The Lone Prairieâ using the microphone, the sensor with a bonded interface, and a sensor with a non-bonded interface. The waveforms of the first two are found to be generally consistent, while the waveform of the signal recorded using the sensor with a non-bonded interface is markedly distorted. Our wavelet transform analysis further verifies that the signals recorded using the microphone and the sensor with bonded interfaces are highly consistent in the frequency domain, whereas the signal recorded using the sensor with a non-bonded interface fails to record high-frequency information of a few hundred hertz as a result of the inability of the sensor to capture high-frequency vibrations (Fig. 5f). The results verify the potential of our sensor in acoustic applications. Although our artificial system can effectively showcase the practical feasibility of using this sensor in artificial ear applications, there is a large frequency-response variation due to the single structure. The frequency-response behavior can be optimized by integrating multiple tympanic membranes of different thicknesses or by incorporating frequency-dependent gain adjustments to compensate for the frequency-response variations.
Discussion
In conventional capacitive soft pressure sensors, a common issue is the presence of gaps between the electrode and the viscoelastic dielectric layer. This often leads to high energy dissipation, resulting in response-relaxation times exceeding 10âms. Such durations limit their effectiveness in detecting high-frequency signals. In our study, we address this limitation by reducing the viscosity of the PDMS dielectric through the incorporation of 2âwt.% CNTs. Additionally, we have engineered a bonding between microstructured microcones and the electrode, effectively reducing the response-relaxation time to ~0.04âms.
We observed that the friction-induced energy dissipation is directly proportional to the contact area created during deformation. Our findings reveal that sharp microcones with larger bonded heads are more efficient in minimizing energy dissipation upon compression. However, a drawback is that these sharp microcones are prone to buckling instability, which can compromise mechanical stability. Furthermore, a larger bonded head can reduce the sensorâs sensitivity. Through our analysis, we conclude that microcones with a moderate height and bonded head size offer an optimal balance, achieving rapid response-relaxation times, high mechanical stability, and enhanced sensitivity.
Our sensor design represents a significant advancement over existing PVDF-based flexible sensors, which may possess a broader frequency bandwidth52. Notably, our sensors can detect both static and dynamic pressures simultaneously. They are also softer and more flexible, broadening their practical applications in fields such as robotics, the metaverse, and biomedical engineering.
It is also important to highlight that standard commercial LCR meters may not be able to detect the limits of response time. For instance, the 6âms response time of our previous sensors also represented the detection threshold of the LCR meter used27. Interestingly, the remarkably short 0.04âms response time of our current sensor is partly attributed to the 25,000âHz bandwidth of our customized circuit board. This underscores the necessity of upgrading the testing systems in tandem with sensor improvements in the future.
Note that there are sensors that have similar or wider bandwidth ranges, including traditional silicon-based capacitive sensors53, and flexible piezoelectric and triboelectric sensors54,55. However, the traditional silicon-based capacitive sensors are stiff, while the piezoelectric and triboelectric sensors are limited to the detection of dynamic signals. By contrast, our capacitive sensor is soft and has little signal drift56; it can not only detect static pressure but also record high-frequency vibrations over 10âkHz, making it potential for a wider range of applications.
Methods
Fabrication of the microstructured mold
A resin template with a microcone array with cone diameter of 50âμm and height of 40âμm was fabricated using high-precision 3D printing (NanoArch S130, BMF Precision Tech, Inc.). The density of the microcones is 4âÃâ104âcmâ2. The bonding joint is about 2âμm in diameter. A mixture of PDMS base and curing agent (Sylgard 184, Dow Corning Co., Ltd.) with a mass ratio of 5:1 was cast onto the microcone array mold. After curing at 80â°C for 30âmin, the templated PDMS layer was peeled off, serving as a reverse template. Next, the reverse template was subject to air plasma treatment (TS-PL05, Dongxingaoke Co., Ltd) at 50âW for 3âmin to realize an easy separation between the template and the PDMS-CNTs composite.
Preparation of the PDMS-CNTs composites
The PDMS (Sylgard 184, Dow Corning Co., Ltd.) and CNTs (purity: 95%, diameter: 10â20ânm, length: 10â30âμm, Nanjing Xianfeng Nano-technology Co., Ltd.) were dispersed in trichloromethane (Aladdin, 99%) for 2âh by ultrasonication. PDMS base and curing agent with a base-to-curing agent ratio of 5:1 and CNTs (2âwt.% or 7âwt.%) were directly mixed to form a homogeneous PDMS-CNTs solution. Next, the solution was cast on a specific mold. After the trichloromethane was evaporated, the degassed PDMS-CNTs composites were cured at 80â°C for 2âh. For the electrode, the weight ratio of CNTs to PDMS was controlled to be 7âwt.% with 50âμm thickness. For the dielectric, the uncured PDMS-CNTs solution was cast on the reverse template. The weight ratio of CNTs to PDMS was controlled to 2âwt.% with a 100âμm thickness.
Preparation of the pressure sensor
PDMS base (10âg) and curing agent (2âg) were added to (100âmL) trichloromethane, and the mixture was sonicated for 30âmin to form a solution. The pre-prepared electrodes and dielectric layer were then immersed in the solution for 6âh at room temperature for swelling. Next, the swollen top electrode, microstructured dielectric, and bottom electrode were stacked in sequence and pressed under 20âkPa to allow for curing to form bonded interfaces. The trilayer was cut to be 10âmmâÃâ10âmm in area and used as a sensor.
Characterization and measurements
The morphologies of the sensor were characterized using a field-emission scanning electron microscope (TESCAN MIRA3). The external pressure was applied and measured using a force gauge controlled with a computer-programed stage (XLD-500E, Jingkong Mechanical Testing Co., Ltd). The pull-off force test was tested by atomic force microscopy (Nanoscope IIIa, Multimode, Picoforce force control module, DI-Veeco Instruments Inc.) in the tapping mode. A rheometer (TA Q850) was used to carry out dynamic mechanical analysis in a tension mode. The storage modulus Eâ and the loss modulus Eâ were measured as a function of temperature. The capacitance response was measured using an LCR meter (E4980AL, KEYSIGHT) at a testing frequency of 1âMHz if not specified. The loading and unloading cycles were conducted at a loading rate of 100âkPasâ1. Unless a preload is specified, a vibration generator (Model BL-ZDQ-2185, Hangzhou Peilin Instrument Co. Ltd.) was adopted to apply vibration of given frequencies to the sensor. The capacitance responses for response time, relaxation time, and mechanical vibration were measured using a home-made circuit with a sampling frequency of 25âkHz. Adhered samples of 10âmm in width and 30âmm in length were used to measure the interfacial toughness of the sample, tested through a standard 180° peel test with a force gauge and computer-controlled stage (XLD-500E, Jingkong Mechanical Testing Co., Ltd). The peeling speed was maintained at 50âmmminâ1. The sensors were affixed to a polyimide sheet using a cyanoacrylate glue (Krazy Glue) serving as a rigid backing.
Detection of sound
The artificial ear was designed following the structure of the human ear, consisting of a tympanic membrane, an auditory ossicle, and our sensor. The auditory ossicle was fabricated by 3D printing of aluminum alloy. The tympanic membrane was made of a fluorinated ethylene propylene film with low damping (diameter: 50âmm). An aluminum-alloy rod of dumbbell shape was designed to mimic the auditory ossicle. The acoustic measurement was conducted in an acoustic chamber to ensure immunity from environmental noise interference.
For the acoustic sound detection, we used a computer-controlled speaker as an acoustic source. The sound pressure level produced by the speaker was measured using a sound meter. The sound pressure P is correlated to SPL and expressed as SPLâ=â20lg (P/P0), where P0ââ=ââ20âµPa is the reference sound pressure in the air. The response as a function of sound pressure and the capacitance changes with increasing sound intensity ranging from 50 to 110âdB were measured at 200âHz.
Reporting summary
Further information on research design is available in the Nature Portfolio Reporting Summary linked to this article.
Data availability
The data generated in this study are provided in the Main Text and the Supplementary Information. Additional data are available from the corresponding author upon request.
Code availability
The code that supports the fundings of this study is available in GitHub with the identifier https://github.com/tao-sun2/SUSTEch-frequency-domain-diagrams.
References
Dargahi, J. & Najarian, S. Human tactile perception as a standard for artificial tactile sensingâa review. Int. J. Med. Robot. 1, 23â35 (2004).
Dagdeviren, C. et al. Conformable amplified lead zirconate titanate sensors with enhanced piezoelectric response for cutaneous pressure monitoring. Nat. Commun. 5, 4496 (2014).
Schwartz, G. et al. Flexible polymer transistors with high pressure sensitivity for application in electronic skin and health monitoring. Nat. Commun. 4, 1859 (2013).
Huang, Y.-C. et al. Sensitive pressure sensors based on conductive microstructured air-gap gates and two-dimensional semiconductor transistors. Nat. Electron. 3, 59â69 (2020).
Yao, H. et al. Near-hysteresis-free soft tactile electronic skins for wearables and reliable machine learning. Proc. Natl. Acad. Sci. USA 117, 25352â25359 (2020).
Shi, J. et al. Embedment of sensing elements for robust, highly sensitive, and cross-talk-free iontronic skins for robotics applications. Sci. Adv. 9, eadf8831 (2023).
Park, K. et al. A biomimetic elastomeric robot skin using electrical impedance and acoustic tomography for tactile sensing. Sci. Robot. 7, eabm7187 (2022).
Liu, F. et al. Neuro-inspired electronic skin for robots. Sci. Robot. 7, eabl7344 (2022).
Yu, Y. et al. All-printed soft human-machine interface for robotic physicochemical sensing. Sci. Robot. 7, eabn0495 (2022).
Liu, Y. et al. Electronic skin as wireless human-machine interfaces for robotic VR. Sci. Adv. 8, eabl6700 (2022).
Shih, B. et al. Electronic skins and machine learning for intelligent soft robots. Sci. Robot. 5, eaaz9239 (2020).
Xu, S., Kim, J., Walter, J. R., Ghaffari, R. & Rogers, J. A. Translational gaps and opportunities for medical wearables in digital health. Sci. Transl. Med. 14, eabn6036 (2022).
Kang, S. et al. Transparent and conductive nanomembranes with orthogonal silver nanowire arrays for skin-attachable loudspeakers and microphones. Sci. Adv. 4, eaas8772 (2018).
Kim, K. K. et al. A substrate-less nanomesh receptor with meta-learning for rapid hand task recognition. Nat. Electron. 6, 64â75 (2023).
Lin, W. et al. Super-resolution wearable electrotactile rendering system. Sci. Adv. 8, eabp8738 (2022).
Niu, H. et al. Cocklebur-inspired âbranch-seed-spininessâ 3D hierarchical structure bionic electronic skin for intelligent perception. Nano Energy 107, 108144 (2023).
Schuler, K. W. & Nunziato, J. W. The dynamic mechanical behavior of polymethyl methacrylate. Rheol. Acta 13, 265â273 (1974).
Armstrong, M. R. et al. Ultrafast shock compression of PDMS-based polymers. J. Polym. Sci. Part B Polym. Phys. 56, 827â832 (2018).
Å ilhavÃk, M. et al. Anomalous elasticity and damping in covalently cross-linked graphene aerogels. Commun. Phys. 5, 27 (2022).
Mannsfeld, S. C. B. et al. Highly sensitive flexible pressure sensors with microstructured rubber dielectric layers. Nat. Mater. 9, 859â864 (2010).
Kim, H. et al. Transparent, flexible, conformal capacitive pressure sensors with nanoparticles. Small 14, 1703432 (2018).
Lipomi, D. J. et al. Skin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes. Nat. Nanotechnol. 6, 788â792 (2011).
Jin, L. et al. Microstructural origin of resistance-strain hysteresis in carbon nanotube thin film conductors. Proc. Natl. Acad. Sci. USA 115, 1986â1991 (2018).
Li, L. et al. A flexible and ultrasensitive interfacial iontronic multisensory sensor with an array of unique âcup-shapedâ microcolumns for detecting pressure and temperature. Nano Energy 105, 108012 (2023).
Bai, N. et al. Graded interlocks for iontronic pressure sensors with high sensitivity and high linearity over a broad range. ACS Nano 16, 4338â4347 (2022).
Nguyen, T.-H., Ngo, B.-V., Nguyen, T.-N. & Vu, C. C. Flexible pressure sensors and machine learning algorithms for human walking phase monitoring. Micromachines 14, 1411 (2023).
Zhang, Y. et al. Highly stable flexible pressure sensors with a quasi-homogeneous composition and interlinked interfaces. Nat. Commun. 13, 1317 (2022).
Li, Z. et al. Gelatin methacryloyl-based tactile sensors for medical wearables. Adv. Funct. Mater. 30, 2003601 (2020).
Lee, S. et al. Nanomesh pressure sensor for monitoring finger manipulation without sensory interference. Science 370, 966â970 (2020).
Yang, J. C. et al. Microstructured porous pyramid-based ultrahigh sensitive pressure sensor insensitive to strain and temperature. ACS Appl. Mater. Interfaces 11, 19472â19480 (2019).
Zhang, Y. et al. Dual-mode electronic skin with integrated tactile sensing and visualized injury warning. ACS Appl. Mater. Interfaces 9, 37493â37500 (2017).
Guo, X. et al. Biologically emulated flexible sensors with high sensitivity and low hysteresis: toward electronic skin to a sense of touch. Small 18, 2203044 (2022).
Shen, Z., Zhu, X., Majidi, C. & Gu, G. Cutaneous ionogel mechanoreceptors for soft machines, physiological sensing, and amputee prostheses. Adv. Mater. 33, 2102069 (2021).
Bai, N. et al. Graded intrafillable architecture-based iontronic pressure sensor with ultra-broad-range high sensitivity. Nat. Commun. 11, 209 (2020).
Niu, H. et al. Perception-to-cognition tactile sensing based on artificial-intelligence-motivated human full-skin bionic electronic skin. Adv. Mater. 34, 2202622 (2022).
Bae, G. Y. et al. Pressure/temperature sensing bimodal electronic skin with stimulus discriminability and linear sensitivity. Adv. Mater. 30, 1803388 (2018).
Fu, M. et al. A highly sensitive, reliable, and high-temperature-resistant flexible pressure sensor based on ceramic nanofibers. Adv. Sci. 7, 2000258 (2020).
Li, R. et al. Supercapacitive iontronic nanofabric sensing. Adv. Mater. 29, 1700253 (2017).
Yang, C.-R. et al. Highly sensitive and wearable capacitive pressure sensors based on PVDF/BaTiO3 composite fibers on PDMS microcylindrical structures. Measurement 202, 111817 (2022).
Wang, X., Deng, Y., Jiang, P., Chen, X. & Yu, H. Low-hysteresis, pressure-insensitive, and transparent capacitive strain sensor for human activity monitoring. Microsyst. Nanoeng. 8, 113 (2022).
Lee, S. et al. A transparent bending-insensitive pressure sensor. Nat. Nanotechnol. 11, 472â478 (2016).
Luo, Y. et al. Flexible capacitive pressure sensor enhanced by tilted micropillar arrays. ACS Appl. Mater. Interfaces 11, 17796â17803 (2019).
Luo, Y. et al. Gecko-inspired slant hierarchical microstructure-based ultrasensitive iontronic pressure sensor for intelligent interaction. Research 2022, 9852138 (2022).
Ha, K.-H. et al. Highly sensitive capacitive pressure sensors over a wide pressure range enabled by the hybrid responses of a highly porous nanocomposite. Adv. Mater. 33, 2170382 (2021).
Yu, P. et al. All-fabric ultrathin capacitive sensor with high pressure sensitivity and broad detection range for electronic skin. ACS Appl. Mater. Interfaces 13, 24062â24069 (2021).
Zhang, Z. et al. Highly sensitive capacitive pressure sensor based on a micropyramid array for health and motion monitoring. Adv. Electron. Mater. 7, 2100174 (2021).
He, Z. et al. Capacitive pressure sensor with high sensitivity and fast response to dynamic interaction based on graphene and porous nylon networks. ACS Appl. Mater. Interfaces 10, 12816â12823 (2018).
Shi, X. et al. Pushing detectability and sensitivity for subtle force to new limits with shrinkable nanochannel structured aerogel. Nat. Commun. 13, 1119 (2022).
Chhetry, A., Kim, J., Yoon, H. & Park, J. Y. Ultrasensitive interfacial capacitive pressure sensor based on a randomly distributed microstructured iontronic film for wearable applications. ACS Appl. Mater. Interfaces 11, 3438â3449 (2019).
Wan, J., Liu, S., Li, X., Zhang, L. & Zhao, H. Buckling analysis of tapered piles using non-prismatic beam-column element model. Comput. Geotech. 139, 104370 (2021).
Sun, K. et al. Flexible polydimethylsiloxane/multi-walled carbon nanotubes membranous metacomposites with negative permittivity. Polymer 125, 50â57 (2017).
Gerardo, C. D., Cretu, E. & Rohling, R. Fabrication and testing of polymer-based capacitive micromachined ultrasound transducers for medical imaging. Microsyst. Nanoeng. 4, 19 (2018).
Xu, J., Wood, G. S., Mastropaolo, E., Newton, M. J. & Cheung, R. Realization of a graphene/PMMA acoustic capacitive sensor released by silicon dioxide sacrificial layer. ACS Appl. Mater. Interfaces 13, 38792â38798 (2021).
Zhang, Q. et al. Multifunctional and wearable patches based on flexible piezoelectric acoustics for integrated sensing, localization, and underwater communication. Adv. Funct. Mater. 33, 2209667 (2023).
Lee, S. & Park, J.-W. Fingerprint-inspired triboelectric nanogenerator with a geometrically asymmetric electrode design for a self-powered dynamic pressure sensor. Nano Energy 101, 107546 (2022).
Sun, J.-Y., Keplinger, C., Whitesides, G. M. & Suo, Z. Ionic skin. Adv. Mater. 26, 7608â7614 (2014).
Acknowledgements
The work was funded by the National Natural Science Foundation of China (No. T2225017 and 52073138), the Science Technology and Innovation Committee of Shenzhen Municipality (No. JCYJ20210324120202007), the Shenzhen Sci-Tech Fund (No. YTDPT20181011104007), and the Guangdong Provincial Key Laboratory Program (No. 2021B1212040001).
Author information
Authors and Affiliations
Contributions
C.F.G. and Y.Z. conceived the idea, C.F.G., Y.C., and L.W. directed the study. Y.Z. conducted most of the experiments. X.Z. designed and manufactured the circuit board. L.W. and N.Z. performed the finite element analysis. J.Z. and N.B. wrote the code of MATLAB for acquiring frequency-domain diagrams. T.S. performed the recognition of acoustic sound detection. X.H., G.L., and L.Z. also participated in the experiments. C.F.G., L.W., and Y.Z. wrote the manuscript. All authors contributed to the writing of the final version of the manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Peer review
Peer review information
Nature Communications thanks Yutian Zhu, Sheng Xu, and the other, anonymous, reviewer(s) for their contribution to the peer review of this work. A peer review file is available.
Additional information
Publisherâs note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the articleâs Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the articleâs Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhang, Y., Zhou, X., Zhang, N. et al. Ultrafast piezocapacitive soft pressure sensors with over 10âkHz bandwidth via bonded microstructured interfaces. Nat Commun 15, 3048 (2024). https://doi.org/10.1038/s41467-024-47408-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41467-024-47408-z