FEATURE ARTICLE Neural Interfaces www.afm-journal.de A Materials Roadmap to Functional Neural Interface Design Steven M. Wellman, James R. Eles, Kip A. Ludwig, John P. Seymour, Nicholas J. Michelson, William E. McFadden, Alberto L. Vazquez, and Takashi D. Y. Kozai* Clinical scientists have also applied knowledge gained from basic neuroscience studies to develop interfaces with the nervous system for therapeutic or assistive purposes in patients with injury and disease.[2–4] For example, electrical stimulation using implantable neural interfaces–or neuromodulation devices—have received Food and Drug Administration (FDA) market approval for the treatment of a diverse set of conditions including epilepsy, depression, Parkinson’s Disease, sleep apnea, blindness, deafness, obesity, urinary and fecal incontinence, and hypertension. Patients have also used similar devices to detect and direct brain signals to control robotic limbs, bypassing their injured or degenerated spinal cord.[3,4] While these clinical successes highlight the potential of these technologies, concerns over large variability in therapeutic/ assistive efficacy, long-term reliability, and health risks[5–8] prevent these devices from reaching their full potential. This article reviews the recent advances and continued materials based challenges of this rapidly expanding field. Advancements in neurotechnologies for electrophysiology, neurochemical sensing, neuromodulation, and optogenetics are revolutionizing scientific understanding of the brain while enabling treatments and preventative measures for a variety of neurological disorders. The grand challenge in neural interface engineering is to seamlessly integrate the interface between neurobiology and engineered technology to record from and modulate neurons over chronic timescales. However, the biological inflammatory response to implants, neural degeneration, and long-term material stability diminishes the quality of the interface overtime. Recent advances in functional materials are aimed at engineering solutions for chronic neural interfaces, yet, the development and deployment of neural interfaces designed from novel materials have introduced new challenges that have been largely unaddressed. Many engineering efforts that solely focus on optimizing individual probe design parameters, such as softness or flexibility, downplay critical multidimensional interactions between different physical properties of the device that contribute to overall performance and biocompatibility. Moreover, the use of these new materials present substantial new difficulties that must be addressed before regulatory approval for use in human patients is achievable. In this review, the interdependence of different electrode components is highlighted to demonstrate the current material-based challenges facing the field of neural interface engineering. 1. Introduction 1.1. Neural Signals Implantable neural interfaces are important tools for capturing and modulating the sophisticated computations of the nervous system. This technology has seen an explosion in research, innovation, and potential applications. In order to better understand plastic changes in neural networks, these interface components must maintain stability over the long time periods associated with memory formation and learning.[1] S. M. Wellman, J. R. Eles, N. J. Michelson, W. E. McFadden, Prof. A. L. Vazquez, Prof. T. D. Y. Kozai Department of Bioengineering Center for the Basis of Neural Cognition McGowan Institute of Regenerative Medicine NeuroTech Center University of Pittsburgh Brain Institute Center for Neuroscience at the University of Pittsburgh University of Pittsburgh 208 Center for Biotechnology 300 Technology Drive, Pittsburgh, PA 15219, USA E-mail: tdk18@pitt.edu The success of an implanted neural interface is contingent upon the quality of the signals that the interface can detect from the nervous system, as well as the reliability and precision with which the interface can modulate the nervous system. Extracellular neural recordings can be used to isolate individual neuronal action potentials (also known as “single-units” or spikes), Dr. K. A. Ludwig Department of Neurologic Surgery 200 First St. SW, Rochester, MN 55905, USA Dr. J. P. Seymour Electrical and Computer Engineering 1301 Beal Ave., 2227 EECS, Ann Arbor, MI 48109, USA The ORCID identification number(s) for the author(s) of this article can be found under https://doi.org/10.1002/adfm.201701269. DOI: 10.1002/adfm.201701269 Adv. Funct. Mater. 2017, 1701269 1701269 (1 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de which represent the most basic code in neuron communication. Alternatively, extracellular recordings that do not isolate individual action potentials are called multiunit recordings, which arise when two or more single-units from different nearby neurons are recorded simultaneously. Finally, local field potentials (LFPs) are generated by the summation of neural activity from a large population of neurons within a spatial region. LFPs are often comprised of low frequency oscillations (0.1–120 Hz) that are less sensitive to spatial interference. For neuroprosthetics applications, single-unit and multiunit recordings are often preferred over LFPs, due to the specificity of the information conveyed, but all have demonstrated useful read-out information. Although neuronal signal transduction is electrical in nature, it is propagated at each axon/dendrite terminal, called a synapse, by the release of neurotransmitters. Unlike action potentials and LFPs, neural chemical sensing electrodes are specific to certain electroactive neurotransmitters such as dopamine or adenosine. These biosensors rely on a technique called fastscan cyclic voltammetry (FSCV) to measure the concentration of molecules based on their oxidation and reduction potentials. While these devices do not operate in the same temporal domain as traditional electrodes, their sensitivity to neurotransmitter-specific signaling is critical for assessing neuronal-type specific activity. Neurochemical recordings utilizing FSCV electrodes have been obtained in human patients by multiple groups,[9,10] and have been proposed as a fundamental human research tool as well as a critical sensing component for closedloop neuromodulation systems.[11,12] When combined with traditional electrophysiological recording modalities, FSCV neurochemical measurements can be instrumental in understanding the complex multimodal therapeutic effects and limiting side effects from neuromodulation therapies to treat psychiatric disorders. In addition to sensing electrical current or neurochemicals, electrode-based neural interfaces can also be used to introduce signal into the nervous system by injecting current into local tissue (0.1 to >10 000 Hz). Electrode designs can vary dramatically depending upon the anatomical location and therapeutic application for stimulation. For example, simple bipolar or tripolar electrode configurations are typically used to stimulate the vagus nerve for epilepsy and depression, while indwelling microcortical arrays with hundreds of electrode sites are used to provide sensory feedback for a prosthetic. Recent advances in genetics have expanded the neural interface toolset beyond electrode-based technologies. Optogenetics combines genetics with light-sensitive proteins to instill membrane channels or activity reporters in selective cell populations. Genetic targeting provides a level of control not possible with traditional electrode technology and is attained using several approaches. A common approach is to rely on specific promoters that regulate the expression of genes in cells targeted by the promoter. Major constraints for optogenetics are the requirement for genetic manipulation, the limited availability of promoters, and the potential variability of genetic transcription. Many of the biological challenges for optogenetics are due to poor control over the number of transgenes inserted into the genome, gene insertion loci, promoter enhancers/inhibitors/silencers activity based on insertion Adv. Funct. Mater. 2017, 1701269 Kip A. Ludwig is the associate director of the Mayo Clinic Neural Engineering Laboratories in Rochester, Minnesota. He received his Ph.D. in Biomedical Engineering from the University of Michigan. Prior to Mayo he led translational device programs at the National Institutes of Health under the SPARC, BRAIN, and CREATE Devices Initiatives. He also developed a minimally invasive neuromodulation therapy that is for sale in seven countries and in a U.S. Pivotal Trial. John P. Seymour’s research is in the area of advanced neurotechnology with a focus on circuit mapping tools. His research has addressed reduced tissue reactivity and improved optical, electrical, and mechanical performance. He earned his B.S. with Honors in Engineering Physics from Ohio State University and his M.S. and Ph.D. in Biomedical Engineering from the University of Michigan. His industry experience includes working at NeuroNexus as a principal scientist. He now holds a faculty research position of assistant research scientist in the Department of Electrical Engineering at the University of Michigan developing novel neural interface systems. Takashi D. Y. Kozai is an assistant professor in the Bioengineering Department at the University of Pittsburgh. He received his Ph.D. in Biomedical Engineering from the University of Michigan, and his BAs from the University of Colorado at Boulder in molecular, cellular, and developmental biology and biochemistry. His primary research interest is in seamlessly interfacing neurobiology and neurotechnology through multimodal analysis, including in vivo multiphoton imaging, functional in vivo electrophysiology, and biomaterials engineering. He is also interested in elucidating molecular and cellular pathways of brain injuries and diseases as well as developing intervention strategies. loci, and for virus optogenetics, trade-offs between diffusion radius and transduction rates.[13] In addition, due to biophysics and protein kinetic limitations, high sensitivity opsins will have slower channel kinetics while fast opsins will have lower 1701269 (2 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de sensitivity.[14] For the purpose of this work, the focus will be on the challenges in delivering and extracting light over the desired spatial resolution with implantable devices (Section 2.15). However, it is critical to understand that optogenetic control will be an ongoing challenge for both biological engineering and materials engineering of neural interfaces. In general, neural interfaces intended to stimulate nervous tissue or record neural activity from the nervous system face a similar material based challenge. The fundamental goal is to enable a higher number of input/output channels from the nervous system, while only creating a minimal disruption of tissue. Avoiding tissue damage is important both for the longterm patency of the interface as well as for minimizing safety risks to the patient. As will be explained in subsequent sections, this fundamental trade-off motivates a number of interdependent design choices with interactions that are often not considered. 1.2. Current Challenges Current challenges with chronically implantable neural interfaces can largely be categorized into performance reliability and variability issues. Invasive surgical procedures to implant chronic neural interfaces carry inherent risks, such as brain infection, surgery related hemorrhaging, blood clots, edema, etc. As such these interfaces would ideally be designed to minimize tissue trauma, remain both functional and biocompatible for the life-time of the patient to avoid corrective surgery or reimplantation, and be removable in the event of unforeseen complications. Across all neural implant classes, there is an average decline of the signal-to-noise ratio (SNR) in the communication between the nervous system and neural interfaces over time. The resulting decline in neural interface reliability is believed to be, in part, due to the inflammatory tissue response triggered by the surgical implantation procedure. This initial reaction can progress into a chronic immune response, ultimately turning into a foreign body response and even leading to migration of the tissue or implant.[7,15] While the details of the biological challenges with chronic neural interfaces have been discussed elsewhere,[8] this reactive tissue response degrades the electrical characteristics of the interface by forming a high impedance encapsulation sheath in conjunction with neural degeneration. This response decreases the amplitude of the detected neural signals by increasing the minimum distance from the electrode to the nearest neuron firing action potentials, thereby increasing the applied current necessary for exciting nearby neurons, and limiting the precision of activation via electrical stimulation.[16,17] A number of acute and delayed stab wound studies show that the tissue can recover to some extent when the device is removed,[17–20] suggesting that certain physical properties of chronically implanted neural interfaces are responsible for this tissue reaction.[20] The focus on minimizing the foreign body response has increased the emphasis on studying advanced materials to overcome specific design parameters hypothesized to attenuate the inflammatory tissue response (see reviews[8,21–28]). However, large variability in histology and recording performance[8,29–34] Adv. Funct. Mater. 2017, 1701269 have made it difficult to disentangle contributions from the many physical properties of these devices to the myriad of observed multimodal failure modes.[5,8,35–37] These complex failure modes produce histological outcomes that, in isolation, are poor predictors of actual performance outcomes as measured by brain-machine neural communication with high degrees of freedom and SNR.[34,35] Similarly small amounts of platinum particulates have been shown to break off from the electrode during chronic stimulation resulting in a foreign body response, but multiple studies have demonstrated the presences of these particulates post-mortem without causing clinically significant safety concerns or deterioration of performance.[38–41] Current literature often highlights the advances of individual material properties and therefore has a limited discussion of the broader obstacles limiting the advancement of the field. While excellent reviews can be found on advancing individual design properties[21–28,42] or expanding upon the biological tissue reaction,[8] this review aims to highlight the interdependence of multiple design and material properties as well as to guide the successful of balancing and titrating of these properties. Through this multidimensional cost-benefit analysis, engineers can design functional neural interfaces capable of obtaining selective, high-fidelity, long-lasting readouts of brain activity. Material selection, the design and assembly of the materials, and strategies for surgical delivery of the implantable components will likely provide critical breakthroughs in achieving improved device performance. 2. Material Considerations for Engineering The simplest electrode design consists of a conductive component to transfer electrical charge, insulation to spatially constrain neuronal signals, and a connector that provides external interfacing with signal processing devices. However, there are numerous intrinsic material properties (conductivity, flexibility, material strength, and chemical stability) as well as extrinsic design parameters (geometry, density, insertion, and packaging) that must be considered to achieve long-term biocompatibility and functionality. These individual intrinsic and extrinsic properties are interdependent on one another; optimizing one can negatively impact performance or lead to loss of functionality. The goal of this review is to introduce a number of the key material and design properties as well as their complex interdependence on one another. Understanding these engineering challenges are critical for the development of devices with longterm functionality. 2.1. Electrical Property/Electrode Traces One of the most important components of any electrode technology is the conductor, which forms the electrical connection, also known as electrical trace, that carries current (Figure 1a). Every electrode must include a conductive element which transfers detected signal from the tissue to an integrated wireless transducer or external electrical device, such as a headcap for cortical recording electrodes, or transmits current to the electrode site from a percutaneous connector or implanted pulse 1701269 (3 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Figure 1. Basic Electrode Anatomy. a) Randles circuit representation of the electrode, which includes the electrode trace (red dash), insulation (green dash), and electrode site (blue dash: Relectrode and Celectrode) where neural potentials (Vneuron) are recorded from. Note: Postelectrode equipment is not shown. b) Cross-section of a microthread electrode[43] showing the carbon fiber core and Parylene insulation. c) Layer-by-layer (LbL) assembled composite electrodes using d) Au nanoparticles or e) carbon nanotubes. Reproduced with permission.[44] Copyright 2012, American Chemical Society. generator. The conductance (G) is the material property that describes the ease with which charges can move within it G= 1 A =σ R l R=ρ (1) l A (2) where R is the resistance of the conductor, A is the cross-sectional area of the conductor, l is the length, σ is the electrical conductivity in siemens per meter (S·m−1), and ρ is the electrical resistivity or specific electrical resistance (Ω·m). As shown in Equation (1) and Equation (2), the dimensions of the material can play a crucial role in determining electrical properties. With the emergence of nanoscale and angstrom diameter wires, it is important to consider how the size of the electrical trace can constrain the maximum signal or stimulation current that can safely be passed through the device. The minimum cross-sectional area (A) is limited by the heating power (P, Joule heating, also known as Ohmic or resistive heating), especially for electrical stimulation and chemical sensing electrodes P = I 2R = I 2 ρ l A (3) where I is electrical current. At some finite A the electric current will produce enough heat that it might burn/oxidize or Adv. Funct. Mater. 2017, 1701269 melt the conductor, reducing its conductivity. Given all of these considerations, specific materials and the method of deposition have unique advantages and disadvantages that should be considered when choosing an electrical trace conductor for particular applications. Metals, particularly noble metals, are perhaps the oldest and most common electrical conductors used for neural electrodes.[45,46] Their high electrical conductivity originates from the overlapping valence and conductance bands of metal atoms, allowing for relatively free movement of electrons through the solid “electron sea” lattice.[47] Selection of metal conductors requires careful consideration of conductivity, mechanical stability for handling, chemical stability (namely, corrosion resistance), and biocompatibility. In general, heavy transition metals such as gold (Z = 79), platinum (Z = 78), and iridium (Z = 77), as well as tungsten (Z = 74) or tantalum (Z = 73; for short-term applications) display a number of these desirable properties. The noble metals and alloys can generally remain relatively chemically inert over the chronic timescales that would affect other conductor classes. Advances in microfabrication and photolithography techniques enabling well controlled and reproducible batch fabrication has led to the use of semiconductors and polycrystalline materials which are more compatible with these microfabrication processes. Semiconductors are materials with small bandgaps (between the valence band and conductance band) that can be made conductive by adding electron acceptors (p-dopants) or 1701269 (4 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de electron donors (n-dopants) impurities. Polycrystalline materials are composed of small crystalline regions separated by grain boundaries.[48] These grain sizes dramatically impact the conductivity, making them more conductive than their amorphous material counterparts.[49–51] Polycrystalline silicon (polysilicon) is a highly pure, polycrystalline form of silicon that is easily fabricated into microelectrodes and is compatible with complementary metal-oxide-semiconductor (CMOS)[52] and micro-electromechanical systems (MEMS)[53] technology. Additionally, the conductive properties of polysilicon can be modulated by varying processing conditions or dopant densities.[51] For these reasons, polysilicon is a popular material used in microfabricated silicon arrays. While silicon technology can be more brittle than metal, it offers a greater design space than traditional microwires. Unlike inorganic semiconductors, conductive polymers (CPs) are organic polymers that conduct current.[54–59] CPs are composed of monomeric compounds linked in chains of alternating single and double bonds and doped with a stabilizing counter-ion.[60] Conjugated polymers have narrower band gaps, allowing electrons to move easily between the conducting band and valence band. Dopants increase the conductivity of polymers by either removing electrons from the valence band (p-doping) or adding electrons to the conduction band (n-doping). The most common biocompatible conductive polymers used for neural interfaces are polypyrrole (PPy),[61] poly(ethylene dioxythiophene) (PEDOT),[61] and polyaniline,[62] with PEDOT being the most electrochemically stable.[28] Most conductive polymers, however, do not contain intrinsic charge carriers and must be doped with a counter ion to reduce electrical resistance and heating, and improve conductance. Therefore, they require counter ions that are partially oxidizing (p-doping) electron acceptors (such as LiClO4, I2, or AsF3) or alternatively partially reducing (n-doping) electron donors such as Na or K. While CPs may be either chemically or electrochemically synthesized,[63,64] careful consideration of the counter-ion’s charge or polarity (zwitterions), and concentration is necessary for electrochemical deposition.[28] Conductive polymers generally have lower conductivity than most metals, but other advantageous properties make them popular for coating electrode sites (see Section 2.4).[65] However, recent efforts are aimed at developing conductive polymer composite wires with improved mechanical properties.[66–68] Currently, considerable challenges remain due the brittleness of deposited CPs which complicate the fabrication of electrodes comprised entirely of CPs. Carbon-based materials (graphene, carbon fiber, carbon nanotubes) are also being developed as the conductive components of electrodes. Some carbon allotropes are highly conductive due to the carbon–carbon pi bonds. Carbon-based materials also have useful mechanical properties such as high tensile strength and modulus. Single carbon fibers (CF) are implemented similarly to metal microwires[69–73] (Figure 1b). Graphene, a hexagonal lattice monolayer of carbon,[74] has exceptional flexibility, surface area, and conductivity.[75] Graphene and graphene oxide (GO) have been implemented as bulk conductors,[75] coatings,[76] and dopants.[77,78] The flat structure of graphene makes it ideal for patterning planar implants.[79–81] Recently developed high aspect ratio all-carbon Adv. Funct. Mater. 2017, 1701269 crumpled graphene transistors demonstrate high sensitivity and improved spatial resolution, with no observed increase in source–drain resistance at up to 56% compressive strain.[82] In contrast, carbon nanotubes (CNTs) are cylindrically wrapped graphene sheets. Single-walled carbon nanotubes are comprised of a single layer and multiwalled carbon nanotubes have multiple layers. These CNTs are emerging as novel substrates for a variety of neural engineering applications due to their high mechanical strength and electrical properties.[83–85] CNTs also can have a range of electrical properties based on the orientation in which the graphene is wrapped in a tube. This property is represented by a pair of indices (n,m), and it modulates the CNTs’ conductive properties: from metallic, to quasi-metallic, or semiconductive[86,87] (see review[28]). Due to challenges of fabricating long CNTs, they are generally employed for the electrode site materials (see Section 2.4),[85] however, there has been some effort in developing and patterning CNT containing composite conductors, such as through layer-by-layer deposition methods for neural interfaces (Figure 1c–e).[28,88–91] For applications where stretchability rather than flexibility is required, CNT-doped copolymer matrices can achieve up to 130% strain.[92] These composite conductors are engineered by bringing conductive particles close enough for electrons to jump from one particle to another, known as percolation networks. The small spaces between each particle can absorb more strain before breakage, improving the mechanical properties of the device. However, the tradeoff is that these small spaces lead to increased resistivity compared to solid metals. To summarize, ideal conductors should achieve exceptional conductivity, long-term functional durability, electrochemical stability, low cross-sectional area, and the ability to withstand stress and strain forces. As devices become smaller and vary in their geometries, materials that can be handled easily during microfabrication are preferred since their conductive properties can be altered with processing. Metals wires are traditional materials with good conductivity with limited stretchability and the ability to assemble into large arrays. Some other conductors trade compatibility with microfabrication techniques and increased design space for assembly into arrays with increased brittleness. Composite materials generally trade decreased conductivity with increased stretchability and/or durability. Often, the decreased conductivity needs to be compensated by increasing the cross-sectional surface area. As such, for developing functional implants, it is critically important to balance all of these material properties as well as those in the following sections. 2.2. Insulation Materials The other critical component of chronically implantable microelectrode technology is the dielectric layer, or insulation layer. Some commonly employed insulating materials include varnish, epoxy, parylene, glass, Teflon, polyimide, fused silica, silicon oxide, silicon nitride, or silicone elastomers.[28,93] Dielectric materials generally have a large bandgap between the conductive and valence band, which blocks electrical conduction at low voltages. While the functional component of any electrode is the conductive element, insulating dielectric materials 1701269 (5 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de are equally critical to preventing parasitic capacitance from allowing the signal to leak into the tissue or into adjacent conductors. The insulating material property is defined primarily by its resistivity (ρ) and the dielectric constant (εr) or the relative static permittivity. For simplicity one may consider the effect of a dielectric material in a parallel plate capacitor (C) C = ε rε 0 A d (4) where ε0 is the electrical constant (ε0 ≈ 8.854 × 10−12 F⋅m−1), d is the thickness of the insulation layer, and A is the crosssectional area of the insulation layer along the electrical trace. Insulating materials are characterized as having a relatively low εr, that is, the ratio of the capacitance of a dielectric material to that of a vacuum which tends to 1. The shunt capacitance (Cshunt, Figure 1a) is the capacitance across the insulation layer from the center of the electrode conductor to the conductive electrolyte bath conductor, which results in an inversely frequency dependent impedance across the electrode, behaving as a low pass RC filter with frequency fc[94–96] Vout-shunt = Vin fc = 1 1 + ω 2R 2C 2 1 ωRC (5) (6) where ω is the angular frequency of the waveform (rad s−1). With respect to Equations (3) and (4), materials with high dielectric constants or low dielectric thickness lead to high capacitance, coupling of electrical signal to adjacent traces, and attenuating the recorded signal (Vout-shunt). At very thin dielectric thicknesses, the insulation becomes less effective due to electron tunneling. For high density arrays, current density accumulation in a wire or electrode can cause a change in current density in adjacent wires through capacitive charging with the two wires acting as the capacitor (Ccross-talk: aka, cross-talk capacitance, capacitive coupling or mutual capacitance, Figure 1a). In multichannel electrode arrays, cross-talk is caused by capacitive coupling of one circuit (the “aggressor”) to another (the “victim”) ∆Vvictim = Cadj 1 ⋅ × ∆Vaggressor C grd + Cadj  Ragressor (C grd−a + Cadj )  1 +  R victim (C grd−v + Cadj )   (7) where ∆Vvictim is the potential change induced by the “aggressor” circuit onto the “victim” circuit, ∆Vaggressor is the potential change in the “aggressor circuit,” Cgrd is the total capacitance relative to ground, Cadj describes the capacitor formed between the circuits, Ragressor and Rvictim are the resistive loads of the coupled circuit on the conductors respectively, and Cgrd−a and Cgrd−v are the capacitances between the aggressor and ground and the victim to ground respectively. Equivalent circuit models developed for electrodes implanted into tissue show that crosstalk increases the effective amplifier impedance (comprised of the head-stage impedance and shunt capacitance) above the Adv. Funct. Mater. 2017, 1701269 effective electrode impedance (comprised of the resistance of the electrolyte solution as well as resistance and capacitance of the electrode) with increasing signal frequency.[95,96] As a result, crosstalk can attenuate recorded signals, cause false signal in adjacent electrodes during recording, or result in unwanted accumulation of charge density between adjacent electrodes during stimulation.[97,98] This effect is especially prominent in small multielectrode arrays in which resolution is important, as signal in one wire can affect signals in nearby wires. Inert materials or materials with low dielectric constants, such as ceramics or polyimide, reduce crosstalk compared to other poor insulating materials such as silicon, a more conductive substrate.[99] While the source of cross-talk is traditionally in the interconnects and packaging, it becomes an increasing important issue as engineers attempt to design higher density arrays.[100] A number of polymer insulating materials have become popular due to the ability to uniformly and reproducibly coat complex microscale shapes, lower elastic modulus, attain chronic stability in physiological environments, and achieve compatibility with other microfabrication techniques for selectively controlled de-insulation of active sites.[101] These include parylene and polyimide variants where the side chains (or R groups) alter the material’s dielectric, mechanical, and sometimes even bioactive properties.[101–103] Careful selection or combination of functional R groups can also be used to functionalize bioactive surfaces (see Section 2.14). Polydimethylsiloxane (PDMS) is another insulator for neural tissue interfaces due to its biocompatibility.[104] However, effective PDMS-based dielectrics for implants require substantial thickness due to its high water absorption rate, which increases its dielectric constant and shunt capacitance, and decreases the cutoff frequency. Other polymers with very high water absorption cannot be used as insulation materials for chronic implants. For example, many epoxies are unsuitable for use as insulators for chronic electrodes because they absorb water, negatively impacting dielectric properties as well as allowing for corrosion conditions in devices utilizing dissimilar metal connections within the insulation, especially under stimulation.[105,106] Another property that requires consideration, especially for electrical stimulation, is the material’s dielectric strength, which is a measure of the field strength a material can withstand before electric breakdown occurs. At high voltages (on the order of 106 V cm−1, the limit increases with decreasing relative permittivity and thickness) dielectric materials “break-down” and become conductive.[107] Dielectric strength, in contrast to the dielectric constant, decreases with surface defects (pinhole defects; Figure 2a), increases with temperature, increases with signal frequency, and can decrease with water absorption. Careful consideration of the dielectric property, water absorption, and insulation thickness are important for chronic neural interfaces. Exceeding these voltages will usually also lead to other failure modes such as gross degradation or melting.[108] However, for chronic electrical stimulation, lower DC voltages that are either deliberately induced to power active components on the electrode, or naturally arise during stimulation can exacerbate failure of the insulation, especially at points of defects. For example, the increase in time to failure for an Al2O3 and parylene bilayer is, in some situations, 13 times faster when 1701269 (6 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Figure 2. Insulation Failures. a) Strong stimulation currents lead to insulation failure that start as pin hole defects (red arrow) in the insulation (cyan). Complete Insulation failure (green arrow) can be identified over regions of PEDOT/PSS electrochemical deposition through the insulation (cyan). b) Many dielectric polymers such as parylene are susceptible to cracking (magenta arrow) and delamination (cyan arrow). Reproduced with permission.[111] Copyright 2012, IOP Publishing. c) Planar arrays made from crystalline polymers form triple junctions (red dashed circle) when implanted. The two crystalline layers cannot be perfectly aligned due to entropy (blue/white angles). d) Ions and water molecules in the tissue thermodynamically degrade the interface of the two crystalline layers a crack forms. e) Cracks thermodynamically propagate until insulation fails. Note: this is not an issue if the triple junction crack propagation failure rate is slower than the lifetime of the patient. f) Self-healing smart polymers with self-annealing functional groups and reseal delamination. g) Insulation adhesion layers (e.g., Au) are sometimes used to reduce insulation delamination failures. h) In the tissue, the wire (W), exposed adhesion layers (Au), and the tissue environment creates a galvanic cell, and accelerates corrosion. Reproduced with permission.[112] Copyright 2011, Elsevier. a 5 V bias is applied.[109] Also, thin dielectric films risk the possibility of forming a simple low-pass RC circuit. An RC low-pass filter allows frequencies from 0 Hz (DC) to its cutoff frequency to pass while attenuating higher frequencies, which can contain critical single-unit information. In this case, a negative capacitance amplifier can help correct for the various capacitive impedances introduced by the conductor/insulator capacitors.[94,110] Dielectrics are essential in ensuring electrical current travels unimpeded through the conductor preventing any loss in electrical quality through shunting or attenuation. As a result, an ideal insulator would be thin and have a low dielectric constant, high dielectric strength, achieve virtually no capacitance, have minimal potential for water absorption, and simultaneously thick enough to occlude fluid ingress. These optimal properties would allow small electrical traces to be packed close together while limiting parasitic capacitance and cross-talk, leading to limited signal loss or attenuation. In addition, the material must be flexible (low Young’s modulus) and remain durable under chronic cycling. Since a wide range of materials fit into the insulator class, other mechanical and adhesion properties must also be considered for applicability to specific electrode designs and functionality requirements. 2.3. Adhesion of the Insulator During microelectrode production, the insulator is typically either deposited uniformly over the entire conductive material, Adv. Funct. Mater. 2017, 1701269 as in the case of fabricated microelectrodes, or hand-made using backfill in the case of neuromodulatory devices. Active electrode sites are formed by removing insulation in a controlled fashion or by controlling the wire length that protrudes from the insulation sleeve. Exposure methods for microwires, such as mechanical impact, cutting with a razor, pyrolysis (flame), electric arc, plasma arc, and laser can lead to large variability in site size, impedance, and recording quality.[113–117] These methods can also create collateral damage to the surrounding insulation layer, such as cracking and delamination (Figure 2b). Cracking and delamination lead to electrode failure as the geometric area of the exposed conductor becomes too large.[118] In turn, this averages the recorded potentials across a larger electrode surface area and attenuates the neural signal (see Section 2.4). This underscores the importance of carefully consider the biocompatibility of the underlying materials. For example, many semiconductors (i.e., gallium or arsenide) are toxic in vivo.[119] While adhesion layer coatings such as Au and Ti have been employed to limit delamination, this can lead to rapid corrosion of the electrode site material (Figure 2g,h; see Section 2.5).[120] A much more controlled approach for exposing the electrode site or depositing the same electrode site is through lithographic techniques, such as through the use of photoresist masks.[121] However, unlike microwires that have conformal coatings of insulation, planar microelectrodes are generally created by sandwiching conductors between two insulation materials. With silicon technology, low pressure chemical vapor 1701269 (7 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de deposition of silicon oxide is used to conformally insulate the polycrystalline silicon traces,[122] and provides one of the highest quality SiO2 films available. However, thin films of silicon oxide have a tendency to bow from compressive internal stresses.[123] For this reason, an additional silicon nitride dielectric layer or silicon nitride followed by another silicon oxide dielectric layer are used to balance the bowing and provide additional biostability compared to a single SiO2 layer.[124,125] When two crystalline materials are used to sandwich electrical traces and then are implanted into tissue, a triple junction forms (Figure 2c) often leading to delamination (Figure 2d,e). The thermodynamics of crack propagation at triple-junction sites between crystalline materials has been well studied, especially in electrolytic solutions such as the cerebral spinal fluid.[35,126–128] One approach to reducing delamination is through thermal annealing, a method that utilizes heat and pressure to improve the seal between two polymer layers.[129] However, this has been shown to impact device performance and sensitivity by influencing bulk polymer properties.[129,130] Polymers that can withstand temperatures near their melting points for longer periods of time such as amorphous (noncrystalline) polymers may be another alternative to forming better thermally annealed insulators for avoiding triple-junctions. Another approach to prevent crack propagation is through the use of “smart” self-healing materials that can respond to incidences of damage.[131] For example, polymers can be functionalized with reactive chemical species,[132–144] two polymer layers with appropriate functional groups (CONH2 and COOH) have been shown to self-anneal into imine bonds in vitro that chemically seal two layers together over time without the need for thermal annealing (Figure 2f).[145] In the event of material degradation or delamination, the exposed functional groups may reform covalently through thermodynamic reactions. However, further investigations are required to determine the self-healing potential of these functionalized polymers in vivo. Unlike planer probes that involve polymer–polymer delamination failure, designs such as microwires and bed of needle arrays benefit from being coated with a single conformal layer of insulation. However, these devices still suffer from delamination and cracking at the conductor–insulator interface near their recording sites, especially if the conductor microactuates during stimulation.[36,118,146,147] Careful material selection and deposition methods are critical for long-term functional performance. 2.4. Electrode Size and Materials Electrode site materials and site sizes are important aspects that impact the sensitivity and performance capabilities of neural microelectrodes. Electrode site designs need to carefully consider the electric field strength to selectively attenuate signals resulting from more distant neurons Vmonopole = Vdipole = 1 q ⋅ 4 πε 0 r 1 qdcos (θ ) ⋅ 4 πε 0 r2 Adv. Funct. Mater. 2017, 1701269 (8) (9) where q is the magnitude of the action potential, r is the distance from the source, d is the thickness of the cell membrane, ε0 is the permittivity constant of the tissue, and θ, the angle between the current source and the electrode, is equal to 0. In reality, extracellular action potential characterization depends on the distance between the electrode and the cell, and the radius and dendritic density of the cell.[148] In general, empirical measurements indicate that the voltage drop-off is between that of Vmonopole and Vdipole.[149–152] This voltage can be expressed simply by n 1 I ⋅ xi 4 r πσ i e i =1 V =∑ ( where 1 < x < 2) (10) where σe is the conductivity of the extracellular space, n is the number of current point sources which the electrode is exposed to, and x is 1 < x < 2 depending on the size, geometry and type of neuron. Small contact sites (<<1000 µm2) have the advantage of greater selectivity, and can potentially distinguish between multiple single-units with higher signal amplitudes. This is because only a small number of neurons will be within close proximity to small recording sites and detected voltage of distant neurons decay rapidly with r (Figure 3a). However, small electrode sites are impaired by increased impedances and thermal noise.[150,153] On the other hand, lowering the impedance by increasing the recording site surface area leads to attenuation of the signal amplitude (Figure 3a,b).[150,154] Large sites can be modeled as small sites in parallel with increasing r with distant parallel elements detecting very low V. Voltage elements in parallel are calculated as the average of all parallel elements. As a result, the fall-off of the electric field generated by a nearby neuron is steep; larger electrodes tend to average strong signal from the nearby neuron with weak signals that are more distant, effectively decreasing the signal recorded. Therefore, it is important to select electrode site materials that lower impedance without increasing the geometric site size. For single-unit recordings, the ideal electrode has the lowest possible surface area and impedance. Stimulation electrodes require additional material considerations. The electrode site material and stimulation amplitude on stimulating electrodes determine whether the electrode is capacitive (non-Faradaic), involving electrostatic or electrolytic interactions between the electrode–electrolyte interface, or Faradaic, which involves redox reactions between the electrode and surrounding medium.[94,158] Non-Faradic electrodes stimulate neurons through the diffusion of ions by charging and discharging the capacitive layer that forms at the electrode surface, which is considerably safer than Faradic charge injection methods since the electrode material is chemically stable. Non-noble metals, such as titanium, tantalum, and stainless steel metals inject charge mainly through this double layer capacitance.[159] However, there are charge transfer limitations compared to Faradaic electrode materials.[160,161] Irreversible Faradic interactions occur when oxidized ions diffuse away from the electrode before they can be reduced and redeposited onto the electrode site, which is called electrode dissolution and can have harmful effects on the surrounding tissue.[162] This is particularly true with metals that generate 1701269 (8 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Figure 3. Electrode site size and material properties influence recording and stimulating performance. a) A small contact site surface area b) compared to a larger contact site surface area records larger voltage amplitudes (VT) from a single point source (a neuron) provided by the inverse square law of electric field strength. Note: I2 and I3 are negative. c–f) Charge transfer during electrical stimulation with varying geometric surface areas at the electrode interface. Increasing surface area from (c) to (d) increases the capacitance (Ce2 > Ce1) and halves the resistance (Re1 > Re2). e) Creating a pore introduces a resistance (Rp’) with a cross-sectional area of A and length l. Increasing pore length from (e) to (f) begins to impact ion movement within the electrodetissue interface and attenuates electrode performance. g–j) SEM images of electrode site materials: g) PEDOT/CNT, scale bar 1 µm, h) PEDOT/GO, scale bar 1 µm, i) PEDOT/PSS, scale bar 1 µm, and j) IrOx, scale bar 500 nm. g,i Reproduced with permission.[155] Copyright, 2016, IEEE. h Reproduced with permission.[77] Copyright 2013, the Royal Society of Chemistry. j Reproduced with permission.[156] Copyright 2010, Elsevier. Distribution of charge density representing the edge effect for k) a planar high-surface area electrode and l) circular electrode. Adapted with permission from.[157] Copyright 2009, Frontiers Media SA. m) Recording site electrochemically deposited with polypyrrole (PPy). The octagonal recording site geometry allowed the polymer to grow into eight teardrop segments with an absence of coating from the center. c–f,m) Reproduced with permission.[28] Copyright 2014, Springer. Adv. Funct. Mater. 2017, 1701269 1701269 (9 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de low molecular weight ions during stimulation. Noble metals such as platinum (Pt), iridium (Ir), palladium (Pd), and rhodium (Rh) are used during neural stimulation due to their corrosion resistance.[163,164] While platinum and iridium have limitations on charge injection capacity,[165,166] platinum–iridium alloys improve mechanical and electrochemical properties, making them suitable as implantable interfaces. Another concern during tissue stimulation is the formation of hydrogen and oxygen gas through the electrolysis of water. The potential range that occurs before the reduction of water (H2 formation) or oxidation of water (O2 formation) is defined as the water window. Platinum electrodes have reported charge injection limits between 50 and 150 µC cm−2 using biphasic 0.2 ms pulses, safely within limits of the water window.[167] However, platinum dissolution has been shown to occur at as little as 20 µC cm−2 so alternative parameters during stimulation must be carefully considered.[168] On the other hand, pseudocapacitive electrodes are made from electrode materials that transfer charge through Faradaic modes, but some of these metal ions can be reversed back onto the electrode before the ion diffuses away by rapidly applying the opposite charge. These electrode materials are sometimes considered ideal stimulating electrode materials because their method of charge injection involves reversible redox reactions that avoid the harmful dissolution of metal ions but have high charge transfer capabilities. For example, iridium oxide (IrO2) electrodes are used for their mixed conductivity and reversible Ir3+/Ir4+ redox reactions that permit charge injection capacities of about 0.5–8 mC cm−2.[169,170] Sputtered iridium oxide films, electrodeposited iridium oxide films, and activated iridium oxide films generate the same iridium oxide as an electrode material, but with widely different surface morphologies.[171–173] Other common materials for pseudocapacitors include transition metal oxides, nitrides, and sulfides such as manganese (MnO2), iron oxide (Fe3O4), titanium nitride (TiN), and titanium sulfide (TiS2).[174–180] While many safe stimulation electrodes exist, they are limited in their ability to spatially and selectively activate small populations of neurons.[181] Ideally, a stimulation electrode will have a very small surface area such that it minimizes activation of nearby axons and neurons to a locally confined area.[28,182–184] However, attempting to increase specificity by lowering the geometric area results in higher charge densities (Q/A) that can damage tissue.[185–187] To address tissue safety issues, most stimulating electrodes are designed to have large surface areas. One way to safely alter the electrical properties of both recording and stimulating electrodes is to increase the electrochemical surface area (ESA) relative to the geometric surface area (GSA) (Figure 3b–d). The ESA of a probe involves introducing micro and nanoscaled features that alter the roughness and porosity of the surface of the electrode, improving charge transfer with the surrounding electrolyte solution. Likewise, increasing ESA lowers impedance,[188] so charges of equivalent amplitude can be injected with lower electrode potentials, decreasing the risk of tissue damage. However, there is an upper limit to increasing the ESA, since at some point ions diffusion through the pore become negatively impacted (Figure 3e). For a sinusoidal AC waveform, the Adv. Funct. Mater. 2017, 1701269 penetration depth λ into the pore of a rough surface can be modeled as λ= 1 κr 2 c dω (11) where r is the pore radius (cm), κ is the electrolyte conductivity (Ω−1 cm−1), cd is the surface capacitance (F cm−2), and ω is the angular frequency of the waveform (rad s−1). For a given frequency ω, there will be a maximum penetration depth λ at which increasing the surface area beyond this distance will not influence charge transfer, due to resistance to ionic current as a result of increasing pore length (Figure 3e).[189] Likewise in chronic implants, resistance to ion diffusion across the coating can be exacerbated in vivo as reactive tissue and cells block pores across the surface of the electrode, likely overriding any beneficial enhancements over time from increasing the electrochemical surface area.[190] Lastly, it may be worth considering how the geometry of the electrode site leads to an “edge effect,” where charges are more concentrated on the edges of an electrode during stimulation due to repelling like charges,[157,191] leading to preferential corrosion at the edges of the electrode site (Figure 3j,k).[192] Therefore, even when relatively safe stimulation parameters are followed, irreversible redox reaction may still occur at the edges which may lead to corrosion of the electrode, damage of the tissue, as well as facilitate the delamination of the electrode sites from around its edges. One method of increasing the ESA:GSA ratio to reduce impedance is with CNT, graphene, CP coatings, or other nanostructured surfaces (Figure 3f–j).[28,83,85,121,193] While these materials generally have lower impedances per electrochemical surface area compared to many metals, they are easier to deposit with high ESA. For stimulation, CPs require more careful consideration of the counter ion.[194,195] Among the CPs, PEDOT is considered to be the most stable electrically conducting polymer coating that has demonstrated decreased impedance, similar charging injecting capabilities to iridium oxide electrodes, biocompatibility with tissue, and long-term mechanical stability in vivo.[28,196,197] Doped CPs have the potential for high storage capacities. PEDOT doped with paratoluene sulfonate (pTS) can produce charge storage capacities of about 130 mC cm−2.[198] The specific dopant and deposition parameters can dramatically alter the surface morphology of the electrode thus influencing charge capacities. For example, poly(styrene sulfonate) (PSS)doped PEDOT electrodes display a reduced storage capacity of about 100 mC cm−2.[198] These polymers can also be chemically modified to form a plethora of derivatives, each with unique electrochemical and biological properties(see Section 2.14).[199] In contrast to stimulating electrodes or electrodes to record electrophysiological signals, electrodes used to measure electroactive neurochemical concentrations using FSCV are critically dependent on the interaction between the electrode surface and adsorbed electroactive neurochemicals of interest.[200] FSCV measurements typically consists of two steps: 1) a small DC holding potential is applied to the electrode to promote adsorption of neurochemicals of interest to the electrode surface while repelling confounding interferents and 2) the application of an approximately 10 ms triangular voltage pulse that sweeps from −0.4 to 1.5 V back to −0.4 V (the cathodic/anodic 1701269 (10 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de limits and sweep rate may vary based off of the neurochemical of interest and electrode material). The triangle voltage pulse causes electroactive neurochemicals adsorbed to the surface to oxidize and reduce at specific potentials, creating brief peaks in Faradaic current that can be used to identify specific neurochemicals and changes in their concentration. Consequently, materials used for FSCV materials must (1) enable small electrode designs that minimize biofouling/scarring that would prevent the adsorption of the neurochemical to the surface, (2) minimize the number of other electrochemical reactions that occur at the surface that could confound measurement of the neurochemical of interest, (3) have minimal drift in open circuit potential that would generate variability in the true voltage applied during the FSCV pulse, and (4) be able to both adsorb the neurochemical of interest while allowing the neurochemical to desorb once oxidized/reduced so that changes in local concentrations of the neurochemical can be monitored over time.[201] The ultimate goal is for the FSCV electrode to have both high sensitivity (detect low concentrations of a neurochemical) and high selectivity (able to differentiate the neurochemical of interest from interferents). The traditional electrode material for FSCV recordings are small carbon fiber electrodes, which can also be cycled at high levels to deliberately cause dissolution of carbon at the surface. This removes biofouling and uncovers new unoccupied sites for adsorption to enhance signal in a chronic environment.[202] Numerous alternatives to traditional carbon have been studied including noble metals like gold, platinum/iridium, iridium oxide, conductive polymers, boron-doped diamond, graphene, and pyrolyzed photoresist.[203–211] Boron-doped diamond is a promising material for chemical sensing due to its good conductivity, chemical stability, and biocompatibility, but it has a low double-layer capacitance that is not ideal for electrical stimulation.[212,213] Some of these newer materials like borondoped diamond have yet to demonstrate sensitivity comparable to carbon fibers. For other new materials the inherent drift in electrochemical reactions at the surface of the material over time ultimately limit sensitivity, or the inability of the neurochemicals of interest to desorb after oxidation/reduction once adsorbed dramatically decrease sensitivity over time. FSCV waveforms and subsequent analysis to maximize sensitivity and selectivity utilizing a small carbon fiber electrode was optimized over the course of decades, while the same degree of optimization for other material candidates has yet to be performed. Of the recently explored new materials, pyrolyzed photoresist, which behaves the most like traditional carbon but is more amenable to modern multielectrode fabrication processes, is arguably the most promising. 2.5. Chemical Stability of Materials Developing chronically implantable devices require careful selection or development of materials and assemblies that outlast the lifetime of the subject. When the adhesion layer (e.g., Ti underneath Au or Ir) or the interface of the conductor and electrode site material becomes exposed to the tissue (due to dielectric cracking, delamination, electrode site dissolution,[214,215] or general preparation), the tissue fluid and the two separate Adv. Funct. Mater. 2017, 1701269 metals form a short circuit galvanic cell that continuously corrodes the electrode material and degrades the electrical characteristics of the microelectrode. Gold-plated tungsten wires, for instance, have been shown to corrode readily in phosphatebuffered saline (Figure 2g,h).[112,111] Not only does this negatively impact the electrical properties of the electrode, but it can also lead to toxic metal ion generation.[216] Silver electrodes, for example, have been shown to rapidly oxidize in the body, leading to large implant mass loss and rampant tissue inflammation.[217–219] Corrosion in tissue commonly occurs between two metals, metals of varying purity, or a heterogeneous alloy surface, where one metal dissolves and deposits onto the surface of the other metal through ion migration facilitated by the electrolyte solution.[220] This can be avoided through homogenously alloying multiple metal elements together. Corrosion resistant, higher atomic weight transition metals with good biocompatibility, such as platinum or iridium, are chosen as preferred electrode materials.[21] In addition to improving corrosion resistance, alloys combine the mechanical and electrochemical properties of two metals, allowing for tunable mechanical and electrical properties.[221–224] For example, pure platinum is too soft for use as a cortical penetrating microwire, so it is commonly implemented in an alloy with the stiffer metal iridium, thus resulting in a stiffer, lower impedance electrode.[223,224] Even without metal adhesion layers, some metals can be prone to delamination and corrosion in physiological environments. Platinum black,[225] iridium oxide,[226] and nanoporous materials[227] can all be used to increase the ESA of the probe. However, careful selection of material selection, site size, and deposition parameters are necessary as these coatings are often prone to delamination or electrode dissolution.[214,215,228] For example, platinum black suffers from low mechanical stability and poor adhesion to the electrode substrate,[229] while iridium oxide films become more prone to degradation and instability during electrical stimulation as electrode site sizes becomes smaller and charge densities increase.[214] Electrodeposition of CPs can also control corrosion. PEDOT is especially promising due to its stability in oxygen-rich, hydrated environments,[61] since oxidation further polymerizes the CP rather than corroding the metal.[28] Likewise, undoped PPy coatings have been shown to impart corrosion resistance on steel.[230] However, while CP coatings offer electrochemical benefits and a wide array of options for functionalization, they have difficulty maintaining lamination[231] and are brittle.[232] Therefore, when using CP coatings, it is also important to consider the underlying conductor material. Care must be taken to ensure that the CP adheres well to the electrode site material to prevent delamination. Adhesion can also be improved by appropriate dopant selection or by surface premodification such as roughening.[233] For example, addition of self-assembled alkylsilane monolayers improved PPy adhesion.[234] The size of the dopant molecule is an important factor to consider. For example, PEDOT doped with large PSS (M.W. 100,000) had an increased tendency to delaminate due to smoother electrode morphology when compared to doping with smaller pTS (M.W. 171.19) and perchlorate (ClO4; M.W. 99.45).[235] However, small dopants can leach out of the CP matrix eventually leading to a decrease in conductivity and long-term instability in vivo.[236] On the other hand, larger dopants such as PSS (M.W. 70,000)[212] 1701269 (11 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de typically remain trapped within the CP matrix, and therefore do not contribute to a loss in conductivity.[237] Unsurprisingly, appropriate selection of the conductive polymers (monomers and/or oligomers) used for electrochemical deposition is also critically important for the success of the electrode. Although, PPy is commonly used, due to its ease of growth and low toxicity, PEDOT exhibits superior chemical and electrochemical stability. Coating thickness is also an important consideration, with thicker PEDOT coatings displaying lower impedance measurements on average compared to thinner coatings, but are more prone to delamination.[196] It has been suggested that rough-patterned surfaces offer better adhesion of polymers to the substrate than smooth surfaces.[233,235,238] Furthermore, the applied potentials required for electrochemical deposition can cause material dependent corrosion which can disturb the integrity of CP coating, where oxidation of the metal can block electron transfer resulting in a nonuniform, patchy coverage.[239] The risk of corrosion can also vary depending on whether the electrode is used for stimulation or recording: excessive electrode polarization during stimulation has been shown to produce irreversible electrochemical reactions that can degrade metals (such as stainless steel[240]) which are biologically stable otherwise. This has led to the exploration of conductive polymers and CNT composite materials for recording and electrical stimulation electrode materials.[84,155,241–245] CNT doping improves the charge injection capacity through several mechanisms, including increased coating surface area, CNT electrical conductivity, and the additional charge transfer mechanism of small cationic molecules entering the coating to serve as counter-ions against the trapped anionic CNTs when the CP is reduced. These attributes provide the coating with substantial non-Faradaic (capacitive) and Faradaic character in both oxidized and reduced states.[246] Mechanical stability tests following both acute and chronic stimulation, as well as chronic soaking, revealed PEDOT doped with multiwalled nanotubes (PEDOT/MWNT) and layer-by-layer CNT composites exhibited none of the delamination or cracking that are typical of PEDOT coatings or IrOx sites.[84,241,247] Interestingly, PEDOT doped with carboxylic acid functionalized CNT maintained recording quality despite increases in impedance over 3–8 weeks post implant.[248] However, careful consideration of size and concentration are necessary since ejected nanoparticles can cause frustrated phagocytosis, cell cycle arrest, apoptosis, and membrane lysis.[249,250] Lastly, the biological response to the implant contributes to electrode failure in a variety of ways. Initially, biofouling, or protein adsorption on the electrode surface, leads to encapsulation, which increases the resistance and capacitance of the electrode, while reducing the electrochemical surface area and attenuating the high frequency content of the electrical stimulus.[237] This is particularly true for high ESA electrode materials where the protein and glial cells restrict ionic diffusion.[251] A study for cardiac pacing electrodes showed that the advantages for having a high ESA electrode were lost 3–8 weeks after implant unless antiinflammatory steroids were released from the electrode site.[252] Biofouling has also been shown to be exacerbated with stimulating electrodes. For example, IrOx electrode sites used for chronic in vivo stimulation or CV testing demonstrated Adv. Funct. Mater. 2017, 1701269 more biological adhesion compared to nonstimulating electrode sites or sites used only for electrochemical impedance spectroscopy.[226] Besides decreasing electrode performance, the tissue response also continuously facilitates the degradation of the insulation and electrode site materials. Cracked or delaminated insulation is further exacerbated by the body heat and infiltrating immune cells that produce reactive oxygen species.[5,253,254] Activated macrophages have been speculated to break down cracked parylene coatings.[101] Efforts toward reducing the biological factors’ contribution to material degradation and recording failure are aimed at reducing biofouling, and improving the stability of the underlying materials. For intracortical electrodes, biofouling can be minimized by modifying the electrode surface properties and minimizing damage to the blood–brain barrier (BBB) upon electrode insertion. Various hydrogel and polymer coatings can be applied to electrode surfaces to increase hydrophilicity or incorporate anti-biofouling molecules or drugs. Common anti-biofouling polymers used in surface functionalization include polyethylene glycol (PEG) and PEG methacrylate (PEGMA). Boron-doped-diamond electrodes have also been shown to exhibit reduced biofouling and more stable electrochemical properties than conventional TiN electrodes.[43,255] Combining anti-biofouling surface chemistries with image guidance technology to avoid tearing major blood vessels during insertion may minimize overall inflammation and glial encapsulation.[33] A combination of materials and design selection will help improve the long-term performance of implantable interfaces. 2.6. Size and Geometry: Surface Area and Diffusion Initially, the size and geometry of implants were believed to play a small role in recording performance and tissue integration when comparing three microelectrode designs with crosssectional areas ranging from 1450–16,900 µm2.[256] However, in vitro and in vivo studies with polymer fibers show glial attachment significantly decreased when the fiber diameters were less than 12–5.9 µm, respectively.[257,258] In the brain, it was shown that 20, 50, and 150 µm2 cross-sectional areas significantly improved neural health and reduced glial encapsulation compared to 3264 µm2 (Figure 4a–c).[259] Lastly, microelectrodes with 55.4 µm2 cross-sectional area demonstrated significant improvement in chronic single-unit recording performance when compared to traditional devices with a 1875 µm2 crosssectional footprint.[132] This has led to a number of hypotheses on how size and geometry may be designed to enhance chronic recording performance.[8,248] For example, devices with holes or lattice structures might improve recording performance by allowing trophic factors to diffuse from one side of the probe to another, which can potentially improve the health of the tissue around the probe.[8,248] In addition, tissue can regenerate through the probe holes, which both improve electrical and chemical signaling between the tissues on either side of the probe as well as prevent probe migration by providing a soft anchor once the tissue heals after the initial insertion injury.[260] These, in turn, have been demonstrated to impact chronic neural recordings and the efficacy of stimulation therapies.[261] Another hypothesis is that these lattice structures reduces the 1701269 (12 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Figure 4. Implant size impacts acute and chronic tissue integration. a) Tissue around the probe’s (b) thin polymer lattice structure showed significant reduction in encapsulating cells and improved neural density (c). Reproduced with permission.[259] Copyright 2007, Elsevier. Copyright 2017, Nature Publishing Group.[267] d,e) Two photon imaging of tissue strain in vivo from a Michigan Electrode Array (d) and carbon fiber microthread electrode (e). Neurons are green, while recording sites and astrocytes are red. Cyan outline highlights microthread electrode. Neurons in panel (d) are much more compressed and oval/elliptical than neurons in panel (e), indicating increased mechanical strain from the embedded electrode volume. Reproduced with permission.[34] Copyright 2014, Elsevier. f,g) Chronic histology shows carbon fiber microthreads with an 8 µm diameter reduced tissue reactivity and improved neuron density of microthread (g) compared to silicon electrodes (f). Reproduced with permission.[132] Copyright 2012, Nature Publishing Group. surface area for biofouling,[132] as well as the diffusion barrier for proinflammatory cytokines, which cause an accumulation of these cytokines around the probe.[262] Similarly, electrocorticography (ECoG) arrays also benefit from a more open architecture, due to tissue growth through the holes, which minimizes the amount of scarring formed between the electrode sites and the brain.[263,264] Reducing the overall cross-sectional area of the implant reduces the overall volume of the implant. For implantable probes, lower implant volumes reduces the tissue displacement volume that occurs during device implantation, which in turn decrease the mechanical strain experienced by neighboring neurons (Figure 4d,e).[8,34] This strain has been shown to inhibit activity of nearby neurons and cause additional inflammation that negatively impacts the nearby tissue health.[265,266] Finally, the location of the electrode sites on the implantable probe is thought to influence recording characteristics. For instance, electrode sites located at the tip of the probe are hypothesized to have a larger viewing radius (see Section 2.7), or volume of tissue in which extracellular action potentials can be detected, compared to sites located along the shank.[121] In contrast to tip electrodes, sites along the shank have less visibility to tissue located directly behind or to the side of the electrode that is partially obstructed by the insulating shank. The benefits of minimizing the volume and cross-sectional area of implants has motivated advances in microscale and Adv. Funct. Mater. 2017, 1701269 nanoscale electrode technology. Wire shank devices with 1–10 µm diameters have been fabricated with silicon, gold, carbon fiber, and conductive polymer composite substrates with excellent electrical properties and minimal host-tissue response.[268–272] On the nanoscale, etching and chemical vapor deposition of silicon has enabled the creation of devices <150 nm diameter for in vitro and in vivo applications while using tobacco mosaic virus as a biotemplate has enabled metal wires to be as thin as <20 nm in diameter.[273–279] However, these introduce new challenges,[28] including handling, packaging, resistive heating, and in vivo chronic implantation requirements.[176,177] In addition, decreasing the size of the implant corresponds with a smaller electrode site, which increases impedance. This can be addressed by increasing the ESA:GSA ratio(see Section 2.4).[43,72,73,193] Given that FSCV measurements require that neurochemicals of interest either directly adsorb to the electrode surface or diffuse into the vicinity directly adjacent to the electrode, there has been a fundamental push toward the use of FSCV electrodes with diameters of 10 µm or less.[280] The diameter of the carbon fiber electrode is important to decrease local damage to the synaptic end-terminals which are the neurochemical sources, as well as decrease biofouling and/or reactive gliosis which may create a barrier preventing the adsorption/diffusion of neurochemicals to the surface of the electrode.[280] Prior studies have demonstrated in vitro that biofouling related to 1701269 (13 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de protein adsorption to the surface of the electrode can dramatically reduce sensitivity to dopamine and other neurochemicals of interest.[281] As carbon fibers are mechanically brittle, a stringent requirement of the use of carbon fiber FSCV electrodes with diameters 10 µm or less may be problematic in terms of handling and long-term implantation.[282] There is no published work directly comparing different diameter sizes for FSCV electrodes and neurochemical recording quality in a chronic in vivo experimental preparation. Nevertheless, difficulties have been observed in obtaining acute dopamine measurements when implanting a 50 µm diameter boron-doped diamond FSCV electrode, likely due to trauma to the adjacent tissue preventing dopamine release locally.[283] However, using a dopamine reuptake inhibitor to promote dopamine diffusion from intact tissue into traumatized tissue did enable dopamine detection with these larger devices.[283] These data support a substantial body of work demonstrating that (1) implantation of a 220 µm diameter dialysis probe can dramatically impact the release and reuptake of neurochemicals at least 220–250 µm from the microdialysis probe,[284,285] (2) tissue near probe tracks created by 280 µm diameter microdialysis probes exhibit ischemia and endothelial cell debris which is most prominent at 4 h post implant,[286] (3) at 24 h the microdialysis probe tracks are surrounded by hyperplasic and hypertrophic glial and glia processes,[286] and (4) 7 µm diameter carbon fiber probes produce a diffuse disruption of nanobead labeling, but no focal disruption of blood vessels, evidence of endothelial cell debris, or glial activation.[286] Across all interface modalities, there is increasing evidence that subcellular geometries improve long-term tissue integration. is possible to avoid penetrating surface pial vessels and large penetrating blood vessels,[289] but with multiple, densely packed shanks, there is a higher probability for vascular injury, microhemorrhage, and exacerbated scarring and cell death, which further increases with greater shank density.[290–293] Despite this, there are few studies that define a “critical” shank spacing to avoid compounding damage. Both electrophysiological performance and gliosis are progressively worse for Michigan style devices and microwire devices at tip-to-tip spacings <600 µm, while wire shanks that are 1 mm apart are indistinguishable from solitary electrodes.[294–296] For tapered devices, the base-tobase spacing of shanks should also be considered. The blood supply for the underlying cortex enters through the surface of the brain, and it is possible that the high-strain of narrow baseto-base separation could disrupt that blood flow.[297] Evaluating the effect of shank density on the hemodynamics of the underlying cortex should be examined in future studies. Common types of multishank arrays (silicon planar Michigan arrays, silicon bed-of-needles Utah arrays, and metal microwire arrays) have been compared in 1–4 monthlong rodent studies.[298–300] These studies show that metal wire shanks tend to outperform planar or bed-of-needles style silicon devices in chronic electrophysiology with higher SNRs and lower impedances, as well as with less long-term BBB dysfunction. In addition to different materials, these devices vary in shank density, shank shape, and platform area, complicating the interpretation of these results. Future studies should systematically vary these parameters to determine their significance for long-term probe viability. 2.8. Flexibility: Softness/Geometry 2.7. Volumetric Density and Channel Count For recording applications, the demand for a greater number of channels has increased. More recording channels allows for a greater volume of tissue to be sampled, and further, multiple channels located closely together allows for the sampling of overlapping volumes of tissue.[152,287,288] This makes triangulation of the signal across multiple electrodes possible. An important consideration for multishank devices is the impact of the overall footprint of the device on the tissue health. More specifically, the volumetric density that the device occupies in the brain over a given amount of space can impact performance. Even if the same volume is implanted into the brain, the maximum strain on the tissue can be reduced by distributing the implant volume over a larger tissue volume as demonstrated by finite element modeling (Figure 5). For example, a single shank tetrode made from four 50 µm diameter electrodes may generate a tissue response similar to a single 100 µm diameter wire, but would generate a very different tissue reaction when compared to four 50 µm diameter wires that are implanted with 1 mm spacing over a large region of the cortex. However, additional considerations may be required based on how the shank spacing impacts injury (tearing and compressing) to the underlying vasculature. When considering configurations of multiple penetrating shanks, the most important factors are the density of the shanks and the configuration and footprint of the platform piece where the individual shanks come together. With a single shank, it Adv. Funct. Mater. 2017, 1701269 Once in the brain, a neural probe spends the rest of its life battling a wet, chemically dynamic, and mechanically tumultuous environment. This mechanical environment is dominated by “micromotions,” which are the movements of the brain in the cavity of the skull in response to respiration, pulsation, and everyday movement.[301] This effect is likely to be greater in larger primates than rodents due to larger and more intricate brains and interstitial space.[302] While the material composition of the probe will largely determine how it fares in the mechanical and chemical milieu of the brain, many groups have also shown that the flexibility of the probe can have an impact on how the host-tissue response and micromotion-related strain is generated by the probe. Flexible implantable devices are believed to reduce the mechanical strain experienced by tissue due to motion especially in the brain when devices are anchored to the skull. The brain is normally floating inside a cerebrospinal fluid filled dural sac which in turn floats inside of the skull. In humans, the brain and skull are usually separated by about 2–4 mm while this number is closer to 100 µm in mice.[35] As such, large forceful movements or impacts to the skull can lead to substantial brain tissue displacement and mechanical strain to the tissue surrounding rigid implants. In addition, there are smaller micromotions and physiological motions in the brain due to respiration and heart-related blood flow pulsation.[303,304] These small movements can increase when craniotomies and 1701269 (14 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Figure 5. Probe insertion results in high strain between shanks of multishank devices, with more strain generated between densely packed shanks. A finite element simulation of Michigan-style planar silicon probe insertion into a viscoelastic brain phantom is shown. The maximum strain value of each element following insertion is displayed. a) When the probe shanks are spread far apart (500 µm tip-to-tip spacing), there is little overlapping strain between shanks. b,c) As tip-to-tip spacing gets tighter, there is increased tissue strain between shanks (300 and 150 µm). Scale bar = 100 µms. duratomies are left open and unsealed.[305] Finite element modeling studies have shown that stiff or rigid devices that are anchored to the skull resist complying with these brain movements.[306,307] In turn, this can lead to a perpetual generation of mechanical strain and proinflammatory cytokines or even reinjure the tissue resulting in device failure.[5,8,34,308] As such, there has been increased focus on developing implantable technologies that are much more compliant, which can be achieved in several ways kc _ Normal = l E⋅A (12) kc _ planar = 4l 3 Ewt 3 (13) kc _ lateral = 4l 3 Etw 3 (14) kc _ radial = 64l 3 3π E (do4 − di4 ) (15) k= 1 kc (16) where kc is compliance, l is the device length, A is the crosssectional area, E is the elastic modulus (or Young’s modulus), w is the width of a planar devices, t is the thickness of a planar devices, do is the outer diameter of a cylindrical devices, di is the inner diameter of a cylindrical devices, and k is stiffness. The compliance of a device can be improved by employing materials with a low elastic modulus, reducing the cross-sectional area of the device, or increasing the length of the device. In order to reduce the rigidity of neural implants from silicon backbone substrates and large metal backbones (100’s GPa), engineers have turned to polymers and thin-film technology.[309,310] Polymers can have softer material properties (Young’s modulus), low thermal expansion, exceptional solvent tolerance, and good dielectric constants. However, polymers can also have high water absorption, poor solvent tolerance, and changing Adv. Funct. Mater. 2017, 1701269 dielectric properties that increase when it has absorbed water or ions, which negatively impact the chronic performance of implanted devices. Therefore, polymers selected for device design must satisfy these broad range of criteria. Hydrogels are among the few materials that can be engineered to have similar mechanical properties to brain tissue due to their high water content.[311–313] Microglia have been shown to be responsive to the stiffness of a material, migrating from softer to stiffer surfaces.[314] However, hydrogels lack many of the other material properties necessary for functional performance, such as electrical, strength, and durability properties. Furthermore, the swelling of hydrogel materials can lead to additional neural cell death if the hydrophilicity, crosslink density, and weight of the polymer is not properly optimized.[315,316] This greatly limits the ability to construct functional devices from low elastic modulus hydrogels. Reducing the cross-sectional area of the probe track has been shown to yield significant improvement in chronic neural recording performance, but as the dimensions decrease, other challenges emerge such as the mechanical handing requirement related to insertion and durability.[73] Similarly, overall increases in device length lead to new challenges. For specific anatomical targets, length can be increased by entering the tissue at a very distant region, though this substantially increases tissue injury. Alternatively, a meandering design, such as a sinusoid, could be employed to increase the length of the device without dramatically increasing the implantation related injury.[317–320] The placement of trifilar coils along the lead combined with strategic anchor points to prevent migration, such as with suture pads, may improve device performance, similar to pacemakers. The last decade has seen intense excitement in the development of wearable and soft electronics[321–325] but rarely are these new materials used in vivo. Little is known about the performance of wearable systems for long-term in vivo use and the number of papers showing in vitro monitoring or actuation is quite limited. Another major challenge is that all of these devices currently require much larger substrate thickness and feature sizes relative to thin flexible arrays patterned with photolithographic approaches such as polyimide, parylene, or SU-8 substrates.[129,130,326–340] Most applications currently favor thin substrates and small feature sizes because the tradeoffs for biomimetic elasticity at the expense of size are still unproven. 1701269 (15 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de In addition, as implantable probes become more flexible, motion related electromagnetic artifacts are likely to manifest within the probe. Motion artifacts are caused by the movement or bending of a current carrying conductor, resulting in a change in the induced electromagnetic field. Although these artifacts are typically observed in the external connector cable (see Section 2.12), increasing implant flexibility may increase the likelihood of motion artifact occurrences due to perturbances from physiological micromotion. These motion artifacts are further exacerbated by changes in the internal resistance of soft conductors or in the capacitance of the insulation due to mechanical strain, which alters the alignment of the material’s conductive molecules and the cross-sectional area of the material. This, in turn, can dramatically impact the amplitude and waveform characteristics of the electrical signal. While improving the softness and flexibility of the device is hypothesized to improve chronic tissue interfaces, it can negatively impact conductivity, dielectric properties, impedances, handling requirements for insertion, material strength, packaging, and scaling.[66,341] Balancing these properties with advanced composite materials or a blend of multiple materials has proven to be the critical challenge for developing functional implantable devices.[66,68] Ideally these devices will be anchored to the region which they are intended to sample and deliver their signal reliably amid all the stresses and strains produced by normal brain motion. 2.9. Implantation Technology Well controlled probe insertion techniques play a critical role in optimizing recording performance and minimizing injury during surgery.[342] This is particularly true for thin, flexible polymer probes that experience buckling when penetrating tough pia mater and, once successful, deflect from original site of implantation.[343,344] Insertion must generate enough strain to puncture the leptomeninges and brain tissue (as little as 0.05 shear strain or 1 mN penetration force), but without causing damage to the surrounding neural tissue, which occurs with as little as 0.1 strain at a 10 s−1 strain-rate or 3% cyclic strain at 0.2 Hz.[344–349] The pia mater is relatively heterogeneous with differing regions of thickness and surface vasculature, so the location of insertion has implications on insertion success. Implanting through thicker pial blood vessels increases compression on the cortex before penetration occurs, while insertions through smaller vessels demonstrate reduced compression on the brain.[350] In pursuing designs that minimize insertion injury, in vivo and in vitro probe insertion experiments consistently show that rigid devices with sharper-tipped and lower cross-sectional area devices have lower penetration forces, less tissue dimpling, and less tissue strain during insertion.[346,351–355] Considering a single penetrating shank, the major implicated parameters are tip shape, presence of edges, and cross-sectional area and profile. Devices with smaller profiles display reduced cortical insertion injury in postmortem histological analysis <1 week postimplant.[356] However, sharper tips have other trade-offs such as leading to more vascular damage, especially at lower speeds in in vitro models.[346] In addition to device geometry, implant Adv. Funct. Mater. 2017, 1701269 surface chemistry has been shown to play an important role in insertion. Notably, both hydrophobic (silane and Parylene C) or hydrophilic (acid functionalization) surface treatments were able to reduce penetration force (silane and acid treatment) or friction force during insertion (Parylene C).[352,354] There is less consensus about the effect of insertion speed on implantation damage. The brain is thought to be a hyperviscoelastic material, which means the faster the tissue deforms, the higher the Young’s modulus (stiffer).[352,355,357–361] Thus, at higher insertion speeds it may require more stress to generate enough strain to puncture. In vivo puncture studies of single shank devices have corroborated this, showing more dimpling, higher penetration force, greater penetration damage, and ultimately more long-term gliosis at higher insertion speeds.[352,355,362,363] Interestingly, lower insertion speeds could lead to more vascular damage, which may be due to differing mechanical properties of brain and blood vessels. While strategies to map and avoid larger blood vessels with two-photon microscopy can minimize most of this vascular damage, this may be impossible for devices with multiple shanks.[289] For these multishank devices, high-speed insertion velocities >8.3 m s−1 have demonstrated improved insertion success, reduced hemorrhaging, and improved gliosis.[290,291,360] So far insertion has been reviewed with respect to rigid, brittle substrates that can be machined into pointed shanks for facile penetration. However, stiff devices result in mechanical mismatch with the surrounding tissue (discussed previously in Section 2.8).[306,364] In contrast, softer, more flexible substrates are more prone to buckling upon insertion.[344,365] To address this, groups have proposed using delivery shuttles to deliver devices into the brain. Some simple designs use electrostatic interactions to adhere flexible probes to a silicon substrate.[366] Following implantation, moisture in the brain can disrupt the interactions between the silicon substrate and the flexible probe to allow for shuttle retrieval. This design can be scaled by using a “sewing machine” method, which uses an insertion needle that can mechanically attach and detach from a flexible probe for more rapid implantation.[367] However, these require additional design considerations for the coupling mechanism. Alternative approaches involve fully or partially coating a flexible device with a sacrificial, resorbable polymer.[368–374] In vivo studies have shown that insertion wounds around resorbable insertion shuttles can fully close with limited evidence of an inflammatory lesion at 4 weeks postimplant.[375] While this is a promising strategy, current fabrication methods are costly and labor intensive, and finished products are often very sensitive to moisture and handling.[318,371] These materials and design issues of scalability and ease-of-use must be addressed before these next generation techniques can be clinically realized. 2.10. Mechanical Strength Materials must maintain robust mechanical strength despite implementation in smaller and more flexible devices, demanding careful material selection and geometry design. As devices become smaller and more flexible with material selection and design geometry, the strength of the material becomes an increasing problem. Mechanical strength varies with 1701269 (16 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de material, size, orientation, and geometry. Therefore, mechanical properties are often quantified in different ways: yield strength, compressive strength, shear strength, tensile strength, fatigue strength, and impact strength. Decreasing cross-sectional area (A) leads to increased stress on the implant σ stress = F A (17) where the stress on the device (σstress) is a function of the force (F) applied on the device area (A). The stress on the implant causes strain (ε) or deformation of the material, while the stress–strain relationship is linear σ stress = Eε (18) As the relationship becomes nonlinear, elastic deformation of the implant becomes plastic, resulting in permanent deformation. Brittle materials such as varnish, glass, or even diamond undergo limited plastic deformation before failing.[376,377] In contrast, tough metals can undergo extensive plastic deformation before reaching failure, making them more resistant to mechanical stresses.[378] However, over time, small plastic deformations can cause failure through fatigue. Elastomers are a class of polymers that are capable of enduring large amounts of elastic deformations, and are the preferred choices for nextgeneration devices.[22,27] While strong polymers such as aramid or polybenzoxazole exist, their high resistivities, high dielectric properties, and sometimes poor chemical stabilities as thin films in physiological solutions make them difficult to implement as building blocks for neural interface technologies.[379,380] As such, a major challenge becomes balancing size with multiple mechanical, electrical, and chemical properties of both conductors and insulators. One approach to balance mechanical strength with conductivity is to incorporate carbon-based materials, such as graphene or carbon fibers into composite materials.[28,132,381] These composite conductors represent flexible alternatives to brittle thin-film metal traces.[28,84,89,90,382–385] For example, layerby-layer assembled polyelectrolytes in particular have demonstrated some limited self-healing capacity.[386] The noncovalently bonded individual particles display some limited mobility within the composite material resulting in the material’s flexibility and capacity to heal defects. Alternatively, controlled addition of defects (adding tears or holes in strategic locations) in thin-film metal traces allow greater distribution of stress, allowing the device to undergo greater amounts of strain before failure.[387] One material with exceptional mechanical strength and stiffness is silicon carbide (SiC). SiC is a wide band gap semiconductor with highly saturated electron drift velocity (leading to high electron mobility and good conductivity at high current frequencies) and a high thermal conductivity.[388] While SiC has more limitations in lithography and microfabrication compared to other forms of silicon, it can be grown readily and doped with p-type or n-type dopants.[388] It is also extremely resistant to physical wear and corrosion[388,389] and has improved biocompatibility compared to silicon[48] and surgical steel.[390] This has renewed interest in applying silicon carbide for chronic neural Adv. Funct. Mater. 2017, 1701269 interfaces. The amorphous phase of SiC is used as an insulating coating, while the polycrystalline phase is conductive.[48] SiC has also been incorporated into nanocomposites with conductive polymers, augmenting the conductivity of SiC and the mechanical properties of the employed conductive polymer.[391] Additionally, the electrical conductivity of this composite can be “tuned” by varying the particle loading.[391] 2.11. Mechanical Mismatch of Subcomponents and Geometry While the mechanical mismatch between the implant and tissue has been well characterized,[306,307] mechanical mismatch of implant subcomponents has recently been emphasized as critical points of failure.[35] For example, iridium and bulk silicon have elastic modulus’ of 528 and 200 GPa, respectively.[35] As a result of this mismatch, mechanical strain occurs, especially on the lower modulus material during implant movement, such as from physiological motion. Finite element modeling revealed that these strains are further focused on protruding geometries, leading to material failure at these mechanical mismatched interfaces as demonstrated by in vivo experiments (Figure 6).[35] These finite element models also demonstrated that geometry and placement of these subcomponents impact the durability and mechanical strain experienced. Specifically, the center of the substrates experience the least amount of strain, while outer surfaces, especially on protrusions and edges, experience the greatest amount of strain. In turn, these impact the integrity of the electrical trace and insulation material. Furthermore, these models emphasized that with simple and uniform geometries, these strain profiles are more uniformly distributed and shared more evenly across the entire device rather than along sharp edges.[35] In some situations, simple and uniform geometries may have improved strength and durability compared to complex lithographic designs (Figure 6f).[35] 2.12. Tethering and Packaging The penetrating shaft of an intracortical electrode brings electrode sites close to neurons of interest, but in many applications, the neural signal must be transferred to an internal pulse generator or an external system. To accomplish this, the shanks and electrode traces of the device are joined at a platform at the surface of the brain (Figure 7). The platform interfaces with the skull-anchored connector, which interfaces with external devices. The design of the platform and connector (jointly referred to as the device’s packaging) has significant ramifications on the long-term viability of the device. Extracorporeal connectors are at risk for traumatic damage and material failure due to environmental exposure.[253,392] Metals, ceramics, and polymers have all been used for the packaging of subcutaneous implants, but they vary in their benefits: ceramics are known for biocompatibility and corrosion resistance yet lack proper mechanical strength and reliability, metals have high ductility and durability but are usually bulky and difficult to scale down in size, and polymers allow for tunable mechanical properties but suffer from absorption of fluids and swelling.[393] The current gold standard is to use a titanium casing with insulated 1701269 (17 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Figure 6. Mechanical failure from nonhomogenous components and geometries. a) A finite element model (FEM) of mechanical strain on a planar silicon array with iridium electrode sites under a 1 µm micromotion in tissue. b) A strain profile along the protruding electrical trace (red dashed line), substrate surface (black solid line), and at the center of the substrate (blue dash dot line) as indicated in (a). c) Strain profile of (a) from the top view. Lowest strain (black) and highest strain (purple) regions are indicated with arrows. d) SEM of a planar silicon array explanted 189 d post implant in mice V1 cortex. e) 2× zoom in of (d). f) FEM showing an order of magnitude less, evenly distributed strain profiles along simple elegant geometries. Reproduced with permission.[35] Copyright 2015, Elsevier. wiring to hermetically seal electrical components from exposure to biological fluids,[394] and help protect device circuitry from electromagnetic interference.[395] Packaging choices for Adv. Funct. Mater. 2017, 1701269 size and form-factor of the platform and connector can also enable on-board electronics such as preamplifiers and filters or incorporate wireless chips as modules.[267,396] This is especially 1701269 (18 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Figure 7. Schematic of different neural probe packaging strategies. Most commercially available technologies are anchored to the skull by cement or bone screws in order to achieve mechanical stability (top). The platforms and shanks of a device can either be directly attached to the connector (rigidly anchored) or attached to a wire bundle intermediary (tethered). Next generation devices could be unanchored (Bottom). These devices would use telemetry to send neural data to external neural processors. Inset: meningeal fibrosis of the wire bundle of a Utah array that was implanted in a nonhuman primate for ≈4 years (inset, top). Second-harmonic generation imaging of fibrous tissue around wire bundle shows thick collagenous tissue (≈500 µm thick) surrounding wire bundle (inset, bottom). Nonhuman primate data is courtesy of the Batista Lab, University of Pittsburgh. true of fully subcutaneous deep brain stimulating (DBS) systems, which house their power source and electronics in an Implanted Pulse Generator (IPG). The IPG is typically secured in a distant tissue compartment, necessitating durable material choices to withstand movements. One of the more common issues for patients using DBS[397] and pacemaker[398] devices is lead breakage and infection. The mechanical mismatch between the hermetically sealed casing and the insulated wires make this joint a preferential failure point (see Section 2.11). Efforts to minimize potential failure modes involve wireless devices that power IPGs through radiofrequency (RF) telemetry. Still, this strategy requires significantly more power to operate, bulky device components, and is complicated by tissue that impedes wireless coupling between subcutaneous implants and external devices.[399] Transparent ceramics can avoid this issue as well as enable optical surveying of neural activity and optogenetics (Section 2.15).[400] Packaging choices also affect the biological response to implants. Devices with larger platforms such as 10 × 10 or 4 × 4 shank Utah arrays are prone to fibrous encapsulation by the surrounding meninges. While the burden of penetrating shafts likely plays a role in this outcome, a similar encapsulation Adv. Funct. Mater. 2017, 1701269 process occurs for platforms without penetrating shafts, such as silicone ECoG implants, which sit sub or epidurally on the surface of the brain.[264,401,402] The platform can become enveloped in a fibrous capsule as early as 7 d postimplant, with encapsulation persisting for the lifetime of the device.[263,264,401] For these implants, fibrosis can be mitigated by leaving fenestrations throughout the substrate or using a resorbable substrate that leaves only electrode sites and traces on the surface of the brain.[263] Additionally, one study[403] examining nonfenestrated ECoG grids, found that hematomas formed in 14% of cases after subdural implantation of electrocorticography grids in human children. Subdural hematomas located beneath an ECoG array can interfere with recording. These grids can also form depressions on the surface of the brain, indicating that the weight of an ECoG array is an important consideration. However, mesh-based micro-ECoG devices with a more open architecture suggest that they allow tissue vasculature to wrap up and around electrode sites, reducing the formation of collagen scar tissue between electrodes and the brain in at least one study.[263] In devices with low-profile or 2D platforms that can be fully embedded in dental acrylic, there are no reports of fibrous encapsulation, but possible fibroblast and circulating white blood cell influx into the parenchyma from the meningeal space.[263,403–405] Cell influx can be mitigated by implanting devices fully in the cortex, eliminating any device-meninges interface.[404,406] While the significance of this cell population is unknown, it presents as an important consideration for designing the meninges-device interface in future probes. Platforms are also important fixation points. Platforms that are not fixed can migrate during the brain’s recovery from implantation surgery. Anchoring can either be accomplished by directly attaching the platform to the connector (rigidly anchored) or held in place by a wire bundle that attaches to the connector (tethered) (Figure 7). In the latter scenario, the connector is cemented and/or screwed into nearby skull bone. The method of tethering and packaging of the wire bundle cables, however, can affect the long-term viability of an implant. For one, due to their proximity and thin insulation layer, wires in the bundle are the primary source of parasitic capacitance (Section 2.2, Equation (6)). Additionally, several studies in rodents have shown that tethering the device leads to increased reactive microglial and astrocytic gliosis and neuronal cell death compared to untethered devices in both planar, Michigan style arrays[404,406] and microwire style arrays.[300] Untethered microwire arrays also outperform tethered microwire arrays in long-term electrophysiology SNRs.[300] This is likely due to slack in the wire bundle, which can allow for more gross movement of the implanted device during micromotion and impacts. Analysis of the implant tracks for tethered devices corroborate this, which appear to be more elongated than tracks around untethered devices.[407] Movement of the wire bundle can also be a source of motion artifact and lead fracture.[408–412] This could be somewhat reduced by placing the bundle in a trifilar coil, which can allow expansion and contraction of the bundle. It is also possible that the wire bundle generates its own tissue response, which can further affect the relative motion of the brain and the implant. This is seen in the wire bundles of Utah arrays in nonhuman primates, which can be become fully encapsulated 1701269 (19 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de in fibrous tissue after 4 years of implantation (Figure 7). The potential for additional fibrous tissue buildup at the interface of the wire bundle and platform may explain why Utah arrays in rats are frequently slanted and partially explanted.[411] The packaging material and design are critically important components that impact overall functional performance that are often underappreciated. 2.13. Wireless Transducers While untethered devices have better outcomes than tethered devices, they still must be anchored to the skull, which causes incongruous motion of the brain and implanted device (Figure 7). It is no surprise that devices that are not anchored to the skull or to a gel sealant have reduced astrocytic and microglial gliosis than their anchored counterparts.[404,406] The percutaneous nature of the connector also presents an infection risk that is much lower for fully subcutaneous deep brain stimulators.[406,413,414] These issues press the need for fully implantable, wireless recording systems that are not anchored to the skull. While completely wireless, closed-loop systems for motor cortex recording and spinal stimulation have been achieved and able to treat gait dysfunction, the brain–computer interface is still bulky and transcranial.[415,416] Smaller, fully implantable wireless modules have been fabricated that can directly interface with commercially available neural probes.[417–420] These devices still require in vivo validation, but are a promising strategy to adapt mature technology to the next generation. Less invasive options include fully injectable millimeter scale electrodes which can interface with neural signal processors via RF or ultrasound.[25,421] Similar devices could also be incorporated on an intravascular stent-electrode device to minimize invasiveness even further.[422] It is unclear, however, if these devices can record single neuronal unit activity. There are also challenges involved with wireless stimulation of neurons. Clinically available deep brain stimulators require subcutaneous stimulation generators and batteries, but these are bulky and batteries must be periodically replaced. One possible solution for this is to use an external energy source to stimulate neurons. Toward this end, RF or ultrasound stimulation can be targeted to electrode devices to stimulate neurons, though in vivo validation is required.[423–425] Infrared light can be used to thermally stimulate neurons, but it must be delivered through an optically transparent “optrode” device to achieve localized delivery and avoid light scattering.[424,426,427] Visible light can also be used to stimulate neurons by using a photovoltaic substrate, which can convert light into electric current, though they often require toxic materials and have poor stimulation efficiency.[408,428–431] A somewhat less invasive strategy is to deliver nanomaterials such as gold or magnetic iron oxide nanoparticles to a neuronal population and stimulate through IR and near-IR light, ultrasound, or magnetic fields.[432–435] This also offers advantages over conventional stimulation through infrared or transcranial magnetic stimulation in that nanoparticles can be targeted to specific populations of neurons through localized injection, magnetic targeting, or modified surface chemistry.[427,434,436,437] Many nanomaterials that can be incorporated into wireless stimulation can also be Adv. Funct. Mater. 2017, 1701269 used as stimulated drug-release carriers to further increase the therapeutic flexibility of wireless stimulation.[434,437–441] However, nanoparticles can undergo photothermal reshaping at relatively low energies (0.6–1.5 eV) driven by curvature-induced surface diffusion at temperatures well below the melting point threshold.[442] This is particularly important for nanoparticles that have specific shape and aspect ratio that allow tuning to specific ranges of wavelengths or frequencies.[443] For chronic applications, the long-term stability of nanorods and nanoparticles on stimulation performance and biocompatibility need to be considered at temperatures well below the melting point of these particles. More recently, using carbon fibers as wireless antennas, carbon fiber photoconverters have been stimulated using photovoltaic and photothermal principles and with better spatially localized neural activation pattern than achievable by optogenetics or electrical stimulation.[182] Emerging materials and new wireless transduction technology together promise novel modalities for interfacing with the nervous system and mitigation of the risks of percutaneous devices. However, these new systems must be thoroughly characterized to identify unique challenges and limitations. 2.14. Molecular and Biochemical (Coatings) Engineering The chemical and mechanical properties of the probe’s surface play a large role in determining how the brain and device will interact with one another. Researchers exploit surface chemistry to mitigate molecular and cellular fouling of the probe, modulate the neurodegenerative and corrosive behavior of encapsulating cells, and deliver therapeutics. As protein adsorption and cellular attachment to the neural probe’s surface are the first steps in forming scar tissue, many groups are pursuing surface chemistry approaches to prevent molecular and cellular adhesion to the probe.[182,356,444–446] The most basic approach to this is to modulate the hydrophobicity of the probe’s surface.[447–449] While hydrophobic surfaces prevent hydrophilic domains of proteins from binding to the surface of the probe, hydrophobic domains can still attach.[450] Hydrophilic and zwitter-ionic coatings can prevent both hydrophobic and hydrophilic proteins from attaching by using hydrogen bonds and electrostatic interactions, respectively, to hold a water layer at the interface of the probe and brain. This displaces proteins and cell binding domains, preventing attachment.[446,451] Various PEG, poly(OEGMA), and hyaluronic acid hydrophilic coatings have been incorporated onto neural probes to prevent fouling and gliosis in vitro and in vivo.[271,452,453] As an added benefit, hydrophilic or zwitterionic polymer chains can also sterically prevent adsorption and adhesion.[271,454,455] Unsurprisingly, coatings are more effective in preventing gliosis in vivo if they are applied to the whole penetrating shank as opposed to just the electrode sites, however electrode site coatings can suppress increases in impedance after initial implantation.[453,455] Despite promising results, many current antifouling coatings fail over time as coating materials oxidize and molecules and cells find vulnerabilities within the coating.[446,450,455] Alternative coating strategies seek to modulate the phenotype of contacting cells instead of preventing attachment. These strategies seek to improve neuronal attachment and regeneration while 1701269 (20 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de preventing microglial and astrocyte adhesion and activation. By reducing inflammatory cell activation, reactive oxidative species (ROS) output can also be reduced, which has been shown to cause material degradation of most commercially available probe types.[254,456–458] Cells can be modulated by modifying the surface topography of the probe, immobilizing bioactive molecules to the surface, and/or applying drug-releasing coatings to the probe. Coating stiff devices with a low density or soft material with a Young’s modulus closer to that of brain tissue can reduce gliosis and improve neuronal survivability.[314,459,460] Because hydrogel materials lack the electrical, strength, and durability properties necessary for functional devices, they are commonly used as surface coatings with and without bioactive molecules which have been shown to improve the tissue response.[314,316,461–463] One disadvantage of hydrogel coatings is that they can push neurons away increasing the distance to the nearest neuron, especially when it swells, which decreases recorded amplitude. Furthermore, depending on the hydrogel crosslink density, it can provide a scaffold for infiltrating macrophages and neutrophils. Hydrophobic, silicone based gels can avoid swelling issue and still yield lower glial attachment.[464] Macroscopic gel coatings that cover electrode sites may additionally increase device impedance, though the effect is minimal for swollen hydrogels. Furthermore, conductive polymers can be electropolymerized within the hydrogel to improve the electrical properties.[465] Like soft coatings, patterned surfaces can also have an impact on adherent cell behavior. Etched silicon substrates with 3D grid patterns or pore features reduced microglial attachment while improving neuronal attachment in vitro and in vivo, respectively.[466–468] Improved and directed neuronal cell attachment have also been demonstrated in vitro for synthetic polymer, gold, and extracellular-matrix based micro and nanopatterned surfaces.[469,470] Conductive-polymer and carbon nanotube or graphene oxide composite coatings have also been shown to improve cell attachment and neural stem cell differentiation, though it is unclear if this is due to the topographical features of the coating, or if there are other chemical or mechanical pathways in play.[470,471] Attaching bioactive molecules to the surfaces of neural probes can also modulate cell behavior. Molecules can either be attached directly to the probe’s surface, or can be chemically tethered[472] or mechanically interlocked[236] to polymer coatings. Since immobilized molecules cannot directly act on intracellular targets, most coating strategies use molecules that target extracellular targets. Attaching extracellular matrix molecules such as laminin, peptide sequences from laminin, or the proteoglycan hyaluronic acid to the surface can “camouflage” probes such that host cells do not recognize the devices as foreign. These coatings have been shown to reduce gliosis and prevent neuronal cell death.[177,236,472,473] Immobilized cell adhesion proteins like L1, which brokers neuron–neuron binding, have been shown to prevent gliosis and neuron death as well as to encourage neurite regeneration in vivo (Figure 8).[452,474–477] Other immobilization strategies seek to actively suppress inflammation. Attaching antagonists of proinflammatory surface receptors or ROS scavengers to remove extracellular inflammatory species can reduce gliosis and spare neurons in vivo.[478,479] Small bioactive molecules, such as superoxide dismutase mimic ROS scavengers, have the added benefit of lower immunogenicity and better in vivo stability compared with fragile peptides and proteins.[479] On the other hand, ligands that mimic the physiological signals for transcytotic transport can be attached to nanoparticles to enable transport across the blood–brain barrier.[480] Drug eluting coatings have the advantage of modulating cell behavior at greater distances and acting on intracellular targets, whereas surface-bound bioactive molecules can only directly act upon membrane targets of adherent cells. Perhaps the most explored drug release target is the antiinflammatory Figure 8. Microglial encapsulation of neural probes is attenuated by L1 cell adhesion molecule coating. Both ultrasmall carbon fiber microthread electrodes (left, blue outline) and silicon Michigan microelectrodes (middle, blue outline) are encapsulated by microglial processes (green) after 6 h of implantation in mice. Electrode devices can be “camouflaged” by covalently attaching bioactive molecules to their surfaces, as shown by an L1 cell adhesion molecule coated Michigan electrode (left, blue outline), which is relatively unburdened by microglial encapsulation. Reproduced with permission.[8] Copyright 2015, American Chemical Society. Reproduced with permission.[475] Copyright 2017, Elsevier. Adv. Funct. Mater. 2017, 1701269 1701269 (21 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de drug dexamethasone. Dexamethasone, which has been effective in reducing gliosis, sparing gliosis, and reducing electrode impedance, can be administered passively from polymer coatings or actively administered from conductive polymer coatings or porous metal electrode sites by electrical stimulation, systemic injection, and microdialysis infusion.[480–489] It is unclear, however, if the effect extends beyond the release period, or if the standard host–tissue response is simply delayed by drug release. Similar dexamethasone-eluting pacemaker leads can have long-periods of drug release (>10 years) with reduction in fibrosis and lower pacing thresholds.[490,491] Other small molecules are being explored as well. Antiinflammatory cucurmin and resveratrol delivered through systemic injection or resorbable polymeric coatings on devices are effective in preserving neurons and repelling glial scarring, but only for as long as the drug is being released.[491–493] Protein release from resorbable polymers, conductive polymers, and hydrogel coatings has enabled the release of alpha-melanocyte stimulating hormone to reduce inflammation as well as neural growth factor, neurotropin-3, and brain-derived growth factor to stimulate neurite extension toward devices.[494–498] This was demonstrated in glass cone electrodes loaded with growth factor have shown stable neural recordings for at least 5 years, though the glass is brittle and susceptible to shattering; it remains unclear if the stability of the recordings can be attributed to the growth factors or other design features.[499,500] Therefore, glass devices should be pursued with extreme caution due to their fragility. In addition to drug delivery, stem cell delivery has also been achieved by seeding stem cells onto laminin or alginate coated probes, though once in the brain most cells differentiate along a non-neuronal fate.[499,501] Lastly, conductive polymers have long been used as a method to attach bioactive molecules to the electrode site and for electrically stimulated controlled drug release. The electrical advantages of CPs[501,502] can be further enhanced by co-depositing conductive macromolecules into the coating such as graphene oxide or carbon nanotubes.[502,503] In vivo, these loaded conductive polymer coatings have proven to be safe, chemically and mechanically durable, and beneficial for long-term neural recordings, with larger spike amplitudes, lower noise floors,[155,271,504–506] and are stable over long periods of stimulation.[155,271] They have also been implicated in encouraging neuronal attachment and neural stem cell differentiation.[471,507,508] Combined, these features make conductive polymers a versatile tool that can be incorporated into any electrode device, including those with soft substrates.[509] There are molecular and biochemical coating strategies available for devices of all materials, conformations, and dimensions. Each device should consider how to leverage surface chemistry to improve the host–tissue response while maintaining electrical properties. 2.15. Optogenetics and Optical Engineering The development of protein based light-sensitive ion channels (opsins) and fluorescent sensors (voltage sensors, calcium sensors, sensing fluorescent reporters) have become a critical tool for achieving cell-specific interrogation. Light delivery and virally targeted opsins have gained increasing scientific and Adv. Funct. Mater. 2017, 1701269 commercial interest as a potential technology for neuromodulation. Optogenetic tools have encountered material challenges related to achieving high-density and efficient light control (see ref. [267] for review), but these technologies are only reliable for months and intended for animal use. This section discusses the material challenges of long-term optical stimulation that includes the packaging of light sources. It is important to note that optogenetics also faces important molecular biology challenges[14,510,511] including transduction reliability and regulatory concerns.[512] The central component of an optical stimulation device is typically a light emitting diode (LED) or laser diode, and depending on the application, the use of materials such as quantum dots or photoluminescence elements might be desirable. One key advantage of LEDs (including µLEDs and organic LEDs [OLEDs]) is this element can directly illuminate the target tissue and thereby simplify the optical system. LEDs provide a wide illumination profile (typically a cosine function, i.e., Lambertian) and thus are an efficient and effective way to illuminate reasonable volumes of tissue close to the component surface. However, if the presence of the LED(s) interferes with nearby sensors, generates too much heat, or if the required LED lifetime is long (e.g., ≥1 year), then the use of waveguides to deliver the light from an external light source such as a laser diode might be a better long-term strategy. Coupling light into a waveguide can be very efficient when using a laser diode since those have pinpoint directed emission, while the coupling efficiency of an LED is exceptionally poor by comparison unless the waveguide is larger than the surface area of the LED.[513] A common challenge for all light sources is their energy requirement to generate sufficient irradiance, while minimizing heat generation due to inefficiencies in the current driver, quantum efficiency, and light outcoupling. It has been estimated that optogenetic stimulation requires as much as 1000fold more power than electrical stimulation to excite matching tissue volumes.[512] Therefore, the electrical insulation must have high dielectric strength to compensate for the additional electrical power necessary to drive LEDs and be robust against insulation failure. This is especially true in locations outside of the brain that are prone to mechanical forces or movement (e.g., muscle). However, power requirements also depend on the distance to the target, opsin sensitivity, and the irradiance required to cause a therapeutic effect. As the efficiency of LEDs and opsins improve, currently available low power µLEDs and OLEDs become practical for in vivo implantation (Figure 9). One inorganic LED system having a pixel size of 10 × 15 µm has been demonstrated to control neural activity with as little as 60 nW (optical power), so even though the power efficiency is only ≈1%, the heat generation does not exceed safety limits.[514] Organic LEDs having a pixel size of 6 × 9 µm have also been demonstrated to drive cellular activity.[515,516] Many challenges still remain. Solid state light emission has yet to achieve longevity for at least a year when being used in vivo. Devices also need to be tested for biocompatibility given the potentially reactive materials such as Ag, Al, ITO, GaN, AlGaNP and many other possible light emitting diode layer materials. Incoherent µLEDs and coherent vertical cavity surface emitting lasers (VCSELs) are rigid inorganic devices but still have the potential to achieve a flexible form factor suitable 1701269 (22 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Figure 9. Sample light delivery systems for optogenetic stimulation. a) Optroelectrode design using gallium nitrate (GaN) microlight emitting diodes (µLEDs) and electrode recording sites stacked together using heterogenous packaging. Reproduced with permission[322] Copyright 2013, American Association for the Advancement of Science. b) Cross-section along a monolithically fabricated µLED and microelectrode (dash-line on top inset, bottom inset shows the layers along the cross-section). The electrode typically consists of low impedance materials like iridium deposed on layers of SiO2 (green). Gallium nitride (GaN) multiquantum wells (MQW) are connected to an anode and cathode to form the LED and silicone shank is the base layer. Reproduced with permission.[514] Copyright 2015, Elsevier. c) Organic LED (OLED) layers have also been recently demonstrated in vitro using these dimensions. Potential failure of LED-based illuminating systems is the possibility of failure of the deposited materials and water ingress to leak the voltage to power the LED into the surrounding tissue. This will electrically stimulate the tissue instead of optically. Reproduced with permission.[516] Copyright 2016, the American Association for the Advancement of Science. d,e) Another prototype of optrode where a waveguide is used to deliver light. e) Waveguide cross-section showing the transmission, cladding, and silicon substrate layers. Reproduced with permission.[517] Copyright 2015, Nature Publishing Group. for the surface of the brain or conformal wrapping around a nerve.[518,519] To do so, one must thin the light source and heterogeneously package each component onto a substrate to provide flexibility and insulation.[322,520] The challenge here is to electrically bond the light source to a substrate, eliminate any voids at the interface, and apply a pin-hole-free conformal coating.[521] In practice, this approach usually fails because of defects at sharp edges and voids at interfaces.[522] The usual failure modes of all implantable devices are applicable here including water and ion ingress, hydrolysis, and delamination. Since these are power consuming elements, the failure is accelerated from the electric fields generated with every pulse of current used to generate the light (Figure 9).[109] To be practical in human use, one must demonstrate stability after millions or billions of pulses and many years of soaking, which to date has not be done in any optogenetic stimulation system. Adv. Funct. Mater. 2017, 1701269 OLEDs have the potential to be highly flexible, conformal, and multicolor. However, they are also the most challenging since they are additionally prone to fail in the presence water vapor or oxygen, which have high transmission rates in polymers.[523] Still, consumer demand for flexible displays and wearable electronics has spurred large R&D investments in this technology. The most common approach to reduce the permeability of water and oxygen without compromising flexibility is the deposition of a hybrid stack of alternating organic/inorganic layers (Figure 9c).[523] Atomic layer deposition of Al2O3 and ZnO with low water absorption polymers (e.g., Barix) is one example of the hybrid approach.[516,524] If the light source and electronics must be implanted and designed for long-term use, then hermetic packaging is currently the most feasible solution to prevent in vivo failure.[522,525] A hermetic feedthrough for light is also required, which is another material challenge and unlike electrical feedthroughs is 1701269 (23 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de not a mature medical technology. Usually hermetic enclosures are large and must be positioned many millimeters or centimeters from the therapeutic target, and a waveguide must be engineered with high transmission and biostability. The core and cladding of a waveguide are equally important material choices and both have an optical requirement and lifetime requirements. The index of refraction is the most critical optical parameter and should ideally maximize the transmission efficiency (and reduce Fresnel losses if covering either end of the waveguide). Some common waveguide core and cladding combinations that may be lithographically patterned include SU8/PMMA,[526] SU8/tissue[527] polysiloxane/polysiloxane,[528] and SiON/SiO2 (Figure 9d,e).[529,530] A simpler manufacturing technique is fiber pulling using polymer/polymer and many forms of fused silica. Unlike the microfabricated waveguides, these can be combined as bundles and that may increase flexibility. Conversely, the microfabricated systems are particularly useful for multimodal device design like optoelectrodes and drug delivery.[517] One group has also developed a technique for co-polymer-metal extrusion so that the tip becomes a moderate-density optoelectrode tool.[269] Many questions remain about all of these optical systems. A few studies have tested SU8 and EpoCore and found optical losses become stable out to 3–6 months of soaking[326,527] but longer testing is still required and for each optical system. Polymer fibers (extruded cylinders) have the most potential as a flexible waveguide core, which would enable a similar form factor as used in conventional neuromodulation leads. When combined with a hermetically packaged laser diode and accompanying electronics, it is similar to an IPG and flexible lead, e.g., spinal cord stimulator and DBS stimulators. A stylet could provide the stiffness to reach the spinal cord or deep brain structure, then is removed to allow the waveguide to move more naturally with the tissue. An optical waveguide, especially one designed for peripheral nervous system, should be flexible in all directions so as to minimize tissue reactivity. Polymer fibers are typically made of polycarbonate, PMMA, polystyrene,[531] and Cytop (fluorinated polymer) but only Cytop has good transmission between 400 and 550 nm.[532] Regardless, termination to a hermetically packaged laser diode remains a major challenge. Another constraint of optical stimulation is heat generation. Light requirements usually depend on the tissue volume and distance from the light source. A Monte Carlo simulation tool available online provides an excellent resource for predicting heat generation in tissue due to light absorption and irradiance profiles as a function of the waveguide diameter, optical power, and wavelength.[533] Fortunately, the duty cycle and pulse width for ChR2 stimulation is typically very low (e.g., 5%–15% duty cycle, 2–5 ms duration) and can become even lower at shorter pulse durations. Recent work demonstrated that even when using pulse-width modulation within this small time window the electrophysiological effect is equivalent and therefore power consumption can be significantly reduced.[534] When optical pulses are in the range of 1 s however, then the rate of heat dissipation is no longer high enough to prevent local tissue heating. Additionally, heat from the light source and any nearby electronics inside the implanted electronics enclosure will have to be added to the radiation heat. Although the heat deposition at these levels might not adversely impact tissue, it is sufficient Adv. Funct. Mater. 2017, 1701269 to elicit changes in tissue perfusion to scavenge increases in heat.[535,536] One recent report stated a maximum energy of 0.5 mJ to avoid these physiological responses.[535] Lastly, there is increased desire for electrode materials that are transparent in tissue for optical stimulation through an electrode, read-out of fluorescent neural activity sensors, and compatibility with optical imaging modalities. In order to image through a probe, the material needs to be both transparent, and have the same refractive index as tissue. This is particularly true with multiphoton microscopy and holography, where nonuniform refraction of light can alter the convergence of the twophoton light path. This can be addressed with uniform “sheet” or “slab” arrays such as an ECoG array, though these devices are prone to dense fibrous encapsulation quickly after implantation (Section 2.12). Transparent electrode site and electrical trace materials have also become desirable. Thin-film indium tin oxide (ITO) is a mostly (≈80%) transparent conductor that has been gaining attention for optogenetic applications.[537,538] More recently, graphene conductors with ≈90% transparency have also become a popular electrode material for optogenetic applications due to improved strength and lower resistivity.[80,81] However, even with one to four layer thick graphene, these transparent materials generate photoelectric artifacts[80] (see[182] review). Even when light source placement is carefully considered, tissue is a very turbid medium that readily scatters photons. The resulting photoelectrical activation can impact the detected signal[182] and can even cause photoelectrochemical Faradaic redox reactions if proper material selections are not made.[539] A long-term challenge for ITO with in vivo optogenetic application is that ITO oxidizes overtime, which results in darkening of the film and a dramatic rise in impedance.[540] Development of new materials for optogenic stimulation and detection need to consider these additional challenges beyond the electrode consideration discussed in Sections 2.1–2.14. Although numerous issues involving optogenetics are highlighted, its future is bright especially for basic-science researchers mapping neural circuits. Solutions for long-term use of optogenetic neuromodulation will require new advances in materials to provide robust electrical insulation, chemically stable, flexible, and safe light delivery via LEDs or waveguides, and for some applications, near-transparent materials. The opsin transduction challenges will also need to be resolved along with the regulatory challenges, and the therapy will have to be highly efficacious relative to electrical stimulation or pharmaceutical interventions for its long-term translational viability. 3. Challenges and Opportunities for NeuralInterfaces Located Outside the Brain Although implants in the brain often capture the academic and public imagination, neural interfaces for use outside the brain are by far the largest current therapeutic market. The cochlear prosthetic for deafness is one of the earliest neural interfaces FDA approved in the United States. Vagal nerve stimulators for epilepsy, spinal cord stimulators for pain, and Finetech– Brindley stimulator bladder control stimulators were all FDA market approved well before DBS for Parkinson’s disease. Currently, the single largest implantable neural interface market is 1701269 (24 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de the epidural spinal cord stimulators for pain. However, recent FDA approvals include: the Argus II Retinal Prosthetic from Second Sight,[541] the CVRx Rheos Barostim system for hypertension,[542] the Inspire hypoglossal nerve stim system for sleep apnea,[543] the Enteromedics abdominal vagus block for obesity,[544] the Medtronic InterStim sacral nerve stimulator for urinary/fecal incontinence,[545] and the StimWave percutaneously deliverable and wirelessly powered spinal cord stimulation system.[546] In general, many of the same design constraints and material trade-offs for neural interfaces in the brain apply to neural interfaces placed outside the brain (Figure 10), but the risks/ benefits of each trade-off changes for the specific therapeutic application and implant location. More severe side effects such as intracranial hemorrhage leading to paralysis, stroke, and seizures that are risks in brain implants are dramatically less prevalent in the periphery, and replaced with other therapy limiting side-effects such as nerve damage leading to motor dysfunction, unintended muscle activation, bradycardia/tachycardia, dyspnea, pain and paresthesia. Whereas devices such as the CVRx baroreflex activation system or the Finetech–Brindley bladder stimulator work very well as a single minimally invasive stimulating electrode or bipolar/tripolar epineural cuff, systems intended to generate multifaceted motor signal decoding and tactile/proprioceptive feedback for control of an artificial arm may require dozens if not hundreds of precisely located stimulating and recording electrodes. Similarly, an indwelling longitudinal intrafascicular electrode that violates the epineurium, or the flat interface electrode (FINE) which redefines a nerve’s geometry by flattening the nerve to allow for closer contact with central nerves may be completely suitable for use in the axon fibers.[547,548] However, penetrating electrodes increase the risk of disrupting the blood–nerve barrier (BNB), which is essential in circulating oxygen and nutrients to individual nerve fascicles.[549–551] Likewise, FINE or book-cuff electrodes have been shown to increase BNB permeability,[552] which allows for the leakage of inflammatory plasma proteins into the extracellular space.[545] Additional design considerations are necessary to minimize device related trauma. More recently, regenerative peripheral nerve interfaces (RPNIs) are currently being developed to provide seamless integration of transected peripheral nerve with electronic interfaces by minimizing fibrotic encapsulation, promoting revascularization, and improving signal transduction for neuroprosthetic control.[553,554] Similar to Figure 10. Device considerations and challenges encountered during electrode design in the central nervous system and peripheral nervous system. Axes rank design components from lowest to highest priority for both PNS and CNS devices. Note that device challenges between the two systems are not mutually exclusive. For example, foreign body response is not as disruptive in the PNS (moderate priority) than the CNS (high priority). Adv. Funct. Mater. 2017, 1701269 1701269 (25 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de targeted reinnervation, when the nerve terminates on a muscle graft then the action potential is transduced into a large amplitude coordinated muscle response that is read out as an electromyogram. However, unlike targeted reinnervation, RPNIs treat even small diameter nerves in situ using small muscle grafts and so can potentially achieve a large number of afferent and efferent interfaces. Stresses on the neural interface and leads connecting to the implantable pulse generator (IPG), as well as relative movement of the targeted tissue with respect to the implanted electrode, also vary greatly based on location of the implant. Epidurally placed spinal cord stimulation electrodes for pain are known to move relative to the underlying spinal cord during changes in postures, motivating strategies to anchor and stimulate on the dorsal root ganglia to provide more consistent therapy. Spinal cord stimulation systems, cochlear implants, and occipital nerve stimulators, suffer most frequently from electrode displacement, extrusion, or breakage.[555–559] Relative movement of nearby tissue inducing migration is less of an issue for vagal nerve stimulators or carotid sinus nerve stimulators, but insulation to direct current preferentially to the underlying nerve of epineural cuffs is more important. Spiral cuff electrodes, such as those used to stimulate the vagus nerve in epilepsy patients, allow for multiple contact points with a flexible geometry that wraps around the nerve and tethers it in place.[560–563] Efforts to reduce migration of PNIs involve device redesign such as suture pads along the lead or mesh electrodes which provide additional points of suturing to the fascia, connective tissue that encloses nerve and muscle tissue.[564] Modifications to surgical techniques have also seen success, such as the use of a silicone elastomer adhesive to cement the electrode to the nerve cable.[565] Electrodes placed on or in peripheral nerves in the arm to record/stimulate for prosthetic control may have leads that are tunneled back over the shoulder joint back to an IPG or wireless base station for control, and relative movement across that joint is particularly problematic for wear and tear on tissue and lead insulation. Moreover, electrodes placed outside the brain and tunneled back to an IPG tend to be much more consciously perceptible to patients, leading to a host of “irritation” related patient manipulation of the electrode, tunneled lead or pulse generator. For example, there is a common condition known as “IPG Twiddler’s Syndrome” where the patient attempts to spin the IPG on a 2-suture axis within the subcutaneous pocket, which has later led to 3- and 4-suture fixation of the IPG.[566–568] Similarly, one of most common failure points in the cardiac pacemaker or DBS system is at the ISO-1 connector (International Organization for Standardization) which connects the electrode lead to the IPG. The mechanical mismatch between the hermetically sealed titanium housing of the IPG placed subcutaneously or submuscular in the chest, the bore in the RF-transparent header section, and the lead insulation wears the insulation over time, slowly exposing the underlying trifilar coil leading to extraneous stimulation of the muscles/ nerves in the chest. Although more clinically available, PNS devices face challenges similar to those faced by CNS systems. Flexible materials are required to bend with macromotion and retain fit to nearby anatomy. Greater stresses on implants also demand more mechanically robust materials, conductors that will not Adv. Funct. Mater. 2017, 1701269 fail due to fracture, insulation that will maintain closure, and traces and connectors that can withstand stresses induced by mechanical mismatch. More creativity is required in defining device geometry, in order to complement the specific strengths of the materials involved. All of the considerations regarding chemical stability of exposed materials also apply as modified for a peripheral environment. Ultimately, materials for PNS devices must work synergistically to maintain functionality within some of the most active regions of the body. 4. Regulatory, Reimbursement, and Other Practical Considerations Despite numerous advances in electrode fabrication processes, electrode and insulation materials, and bioactive coatings within the academic realm, existing clinical deep brain stimulators, spinal cord stimulators, and vagal nerve stimulators still draw heavily from the designs, materials, and underlying concepts from the cardiac pacemaker/defibrillator. There are two primary drivers for the slow speed of adoption of new fabrication processes, passive materials, and biologically active materials. The first is the large jump from low volume animal studies across a few academic labs, to widespread clinical use in hundreds of thousands of patients at facilities across the world. Even the cardiac pacemaker/defibrillator, which is implanted in 900,000 patients a year worldwide and relies on a simple surgical procedure and well understood materials that have been in practice for 50 years, has a significant rate of device failure. The comprehensive FOLLOWPACE study, which followed over 1500 pacemaker patients over an average of 5.8 years, found that 12.4% of patients developed pacemaker complications within the first two months after surgery, while 9.2% of patients experienced complications after two months.[569] Many of the novel materials discussed in this review such as conductive polymers and carbon fibers are manufactured using processes where the intermediate stages of the material may be toxic, or where potentially toxic substances are used to treat the material to achieve the final result. In an industry supply chain designed to manufacture a product at scale, many of the manufacturing steps or constituent materials are provided by an outside supplier. When manufacturing at the scale of 10,000 implants per year, it can be difficult for the end-system manufacturer to guarantee that the supplier has not changed their process in some small way that might accidentally introduce a leachable toxic substance into material in contact with tissue. Moreover, simple unknowns like shipping the electrode coated with IrOx overseas, or understanding the potential consequences if a surgeon drops the sterile package with the electrode and leads before implantation have to be carefully understood for any change in design. Finally, the introduction of biologically active components can greatly complicate the regulatory pathway. The “combined” devices often have to go through the FDA Office of Combination Products and review from both the FDA Center for Devices and Radiological Health and the FDA Center for Biologics Evaluation and Research, dramatically increasing the regulatory burden and bar for safety. Given the biological component, there is understandably a concern about variability in manufacturer process/supply chain/ handling leading to unforeseen consequences in the patient. 1701269 (26 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Table 1. Constraints on electrode device design and ideal solutions for recording and stimulating applications. Constraint columns represent opposing extremes for different engineering parameters. Constraint Ideal Constraint Conductor Low conductivity High electrical resistance Ohmic Heating Infinite conductivity Zero cross-sectional area Electrochemically stable materials Flexibility Mechanical strength High cross-sectional area Increased stiffness Low stretchability Increased device footprint Dielectric thickness Ultrathin Low dielectric strength Dielectric constant of vacuum (1) High shunt capacitance Greater impact from water absorption Electron tunneling Parasitic capacitance No capacitance No water absorption Low dielectric constant Self-healing Thick Insulation Large device size & footprint Increased susceptibility for delamination Increased stiffness Site size Very small High impedance Charge density safety limit (stimulation) Zero surface area electrode site Zero impedance Very large Signal is distributed and attenuated Large activation area (stimulation) Site coating Resistive materials Corrosion Material dissolution High ESA:GSA ratio High charge injection capability Non-Faradaic charge transfer Good biocompatibility Capacitive materials Low charge injection capabilities Size and Geometry Small cross-sectional area Very flexible Poor Strength Low device volume High device strength Large cross-sectional area Increased stiffness Increased tissue strain Increased glial encapsulation Volumetric density Low shank density Discontinuous sampling volume Spatially continuous recording/stimulating radius Large spacing between electrode shanks Low tissue strain High shank density Increased tissue strain BBB disruption Meningeal gliosis Flexibility Very stiff Mechanical mismatch with soft tissue Stiff enough to insert/handle Flexible enough to not activate inflammatory mechanosensitive receptors Rigid enough to minimize motionrelated electromagnetic artifact No migration Very flexible Difficult to penetrate pia mater Greater motion artifact Device migration Insertion Slow insertion BBB Compression and Disruption Tissue Strain and Dimpling No vascular disruption No tissue strain Controlled insertion speed and angle Fast insertion Ballistic trauma Increased gliosis BBB Severing Very rigid Mechanical mismatch with soft tissue Activates inflammatory mechanosensitive receptors Elastic Modulus of tissue Good conductivity at low crosssectional area Good dielectric at thin insulation layer High yield strength Very soft Low Conductivity Low Dielectric Constant Difficult to insert Packaging Rigidly anchored device Increased tissue strain Impact related trauma Infection route Unanchored to the skull Wireless device Flexibly Tethered device Increased fibrotic encapsulation Implant migration Infection route Motion artifacts Molecular and Cellular Pharmaceutical Drug release side effects Toxic polymer degradation Hydrogels swell No biofouling No inflammation or scarring No neurodegeneration Neurotrophic, Neuroregeneration, Neural Growth Factors Neurocamouflaging/Neural Adhering Surface Biological Hydrogels swell Stem cell rejection by immune system Proteins degrade quickly Elastic modulus (softness) Adv. Funct. Mater. 2017, 1701269 1701269 (27 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de The second driver of the slow speed of adoption of new materials/concepts/process into clinical use is the high cost in both time and money to obtain market approval from the FDA, and the realities of insurance reimbursement. A pivotal trial in the United States for an implantable neuromodulation/neuroprosthetic device to obtain FDA premarket approval often takes more than 5 years to complete and can cost well in excess of 100 million dollars. Moreover, obtaining FDA market approval does not guarantee that insurance payers will reimburse for the therapy, or reimburse at a price necessary to sustain a commercial manufacturer. As new materials/processes introduce significant unknowns into the regulatory and reimbursement pathways, the prospective reward for adopting these new concepts must be both highly likely and offset the large costs and risks inherent in regulatory approval given the price for reimbursement set by insurance payers. Despite these drivers slowing the adoption of new materials/concepts/processes, in 2015 RetroSense Therapeutics LLC received an investigational new drug exemption from the FDA to transduce a very limited number of human subjects in order to restore light sensitive channels in the retina of patients blinded by retinitis pigmentosis.[570] Although this is a Phase 0–1 study to determine safety, and only ambient light is used for illumination, this is a critical step in moving toward optogenetic therapies.[571] The jump from animal models is quite large, however, and demonstrating transduction, specificity of transduction, and maintenance of transduction utilizing an adeno-associated virus gene vector to enable optogenetic therapies in humans has yet to be demonstrated. To date optogenetics has primarily been used in genetically homogenous murine models, with no studies as of yet demonstrating consistent specificity and maintenance of transduction over a long period of time across a large cohort in a large animal model. A study using adeno-associated virus ocular subretinal injection to address c-MER proto-oncogene tyrosine kinase (MERTK) mutations in six human patients blinded by retinitis pigmentosis, for example, failed to demonstrate any transduction in three of the patients, and only one patient maintained expression out to 2 years.[571] Figure 11. Complexity of Functional Neural Interface Engineering. There is a complex network of interdependence between design properties and material constraints, which must be carefully considered when engineering functional neural interface devices. Optimizing a single material property can often negatively impact another material characteristics directly or indirectly through downstream relationships. This can push the overall design outside functional performance constraint windows, ultimately resulting in a nonfunctional device. Therefore, neural interface engineering requires precise balancing of between intrinsic and extrinsic material properties, and their impact on overall performance. Bolded words represent tunable design parameters, and italics denote influencing factors or constraints. Functional performance characteristics are listed in the center, where the biological, electrical and mechanical domains overlap. Blue and red arrows represent positive or negative correlations between terms. Abbreviations include blood–brain barrier (BBB), PC (pseudocapacitive), µ/n (micro/nano surface), electrochemical to geometric surface area ratio (ESA:GSA), volumetric density of tissue recording (VD). Note: Impedance increase and decrease can both impact performance positively or negatively, depending on the cause. See Table 1 for simplified form. Adv. Funct. Mater. 2017, 1701269 1701269 (28 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de Although the retina is attractive for initial demonstrations of optogenetics (as photoreceptors can be illuminated without an implantable device), more diverse therapeutic applications will require implantation of yet to be developed human-ready fiber optics or µLEDs for illumination. This will require a host of developments to enable a safe, reliable, and cost-efficient system at scale, as outlined in the excellent review “From optogenetic technologies to neuromodulation therapies.”[512] Even given a safe and reliable implantable optogenetic transduction and illumination system, clinical adoption will be key for a successful optogenetics business case. At present only 5%–20% of the patients indicated for standard DBS for Parkinson’s elect to receive an implant; optogenetics will have to demonstrate a large improvement over electrode technologies to overcome hesitance to undergo both manipulation of the genome in addition to surgical implantation of a fiber optic or µLED system. 5. Conclusion The complexity of neural interface design, coupled with regulatory and reimbursement hurdles, forecasts a challenging road forward for medical device innovators. Understanding how individual material properties contribute to the success or failure of a neural interface has allowed researchers to experiment with the optimization of specific design variables. Although this has resulted in advancements in materials science and the current understanding of ideal design (Table 1), major obstacles remain for translation to functional implantable devices. This is because single parameter optimization alone is insufficient in producing long-term functional neural interfaces, due to unavoidable tradeoffs that result from the interdependence of design variables. For example, hydrogels can be engineered to have a similar elastic modulus as the brain tissue, however they lack many other important characteristics, including electrical and insulating properties, chemical stability, or material strength and durability. Conductive polymers provide benefits of corrosion resistance, high ESA, and favorable electrical properties, but they are lacking in mechanical stability, biostability, and softness. And subcellular devices have favorable softness and flexibility, and induce minimal strain on the surrounding tissue, yet they suffer from high resistance and low mechanical strength. It is clear that there is a complex network of interplay between intrinsic and extrinsic material and design properties, and the resulting biological, mechanical and electrical responses (Figure 11). Therefore, the ultimate challenge in developing functional devices is understanding interdependence of these variables, and then carefully balancing and titrating these variables with existing or emerging biomaterials. Alternatively, new modes of technology (e.g., wireless transducers) could eliminate the need for some of these variables. These innovations, however, require careful perspective and consideration to critically analyze what new limiting variables are introduced with these new signal transduction modes. Only through honest evaluation of limitations and failure modes, as well as outliers, are we able to engineer transformative next generation technologies. Adv. Funct. Mater. 2017, 1701269 Acknowledgements T.D.Y.K., S.M.W., A.L.V., and N.J.M. were supported by NIH NINDS 1R01NS094396. K.A.L. was supported by The Grainger Foundation. J.P.S. was supported by NIH (U01-NS090526-01) and the NSF (1545858). A.L.V. was supported by NIH (R01-NS094404). The authors would like to thank Dr. Aaron Batista Lab for providing the histology samples used in Figure 7. Conflict of Interest The authors declare no conflict of interest. Keywords bioelectronics, electrodes, microelectromechanical systems, photonics, sensors Received: March 8, 2017 Revised: April 10, 2017 Published online: [1] D. Huber, D. A. Gutnisky, S. Peron, D. H. O’Connor, J. S. Wiegert, L. Tian, T. G. Oertner, L. L. Looger, K. Svoboda, Nature 2012, 484, 473. [2] J. L. Collinger, M. A. Kryger, R. Barbara, T. Betler, K. Bowsher, E. H. Brown, S. T. Clanton, A. D. Degenhart, S. T. Foldes, R. A. Gaunt, F. E. Gyulai, E. A. Harchick, D. Harrington, J. B. Helder, T. Hemmes, M. S. Johannes, K. D. Katyal, G. S. Ling, A. J. McMorland, K. Palko, M. P. Para, J. Scheuermann, A. B. Schwartz, E. R. Skidmore, F. Solzbacher, A. V. Srikameswaran, D. P. Swanson, S. Swetz, E. C. Tyler-Kabara, M. Velliste, W. Wang, D. J. Weber, B. Wodlinger, M. L. Boninger, Clin. Transl. Sci. 2014, 7, 52. [3] L. R. Hochberg, M. D. Serruya, G. M. Friehs, J. A. Mukand, M. Saleh, A. H. Caplan, A. Branner, D. Chen, R. D. Penn, J. P. Donoghue, Nature 2006, 442, 164. [4] J. L. Collinger, B. Wodlinger, J. E. Downey, W. Wang, E. C. Tyler-Kabara, D. J. Weber, A. J. McMorland, M. Velliste, M. L. Boninger, A. B. Schwartz, Lancet 2013, 381, 557. [5] J. C. Barrese, N. Rao, K. Paroo, C. Triebwasser, C. Vargas-Irwin, L. Franquemont, J. P. Donoghue, J. Neural Eng. 2013, 10, 066014. [6] C. A. Chestek, V. Gilja, P. Nuyujukian, J. D. Foster, J. M. Fan, M. T. Kaufman, M. M. Churchland, Z. Rivera-Alvidrez, J. P. Cunningham, S. I. Ryu, K. V. Shenoy, J. Neural Eng. 2011, 8, 045005. [7] T. D. Y. Kozai, Z. Du, Z. V. Gugel, M. A. Smith, S. M. Chase, L. M. Bodily, E. M. Caparosa, R. M. Friedlander, X. T. Cui, J. Neurosci. Methods 2015, 242, 15. [8] T. D. Y. Kozai, A. Jaquins-Gerstl, A. L. Vazquez, A. C. Michael, X. T. Cui, ACS Chem. Neurosci. 2015, 6, 48. [9] S. Y. Chang, I. Kim, M. P. Marsh, D. P. Jang, S. C. Hwang, J. J. Van Gompel, S. J. Goerss, C. J. Kimble, K. E. Bennet, P. A. Garris, C. D. Blaha, K. H. Lee, Mayo Clin. Proc. 2012, 87, 760. [10] K. T. Kishida, S. G. Sandberg, T. Lohrenz, Y. G. Comair, I. Saez, P. E. Phillips, P. R. Montague, PLoS One 2011, 6, e23291. [11] S.-Y. Chang, C. J. Kimble, I. Kim, S. B. Paek, K. R. Kressin, J. B. Boesche, S. V. Whitlock, D. R. Eaker, A. Kasasbeh, A. E. Horne, J. Neurosurg. 2013, 119, 1556. [12] J. K. Trevathan, A. Yousefi, H. O. Park, J. J. Bartoletta, K. A. Ludwig, K. H. Lee, J. L. Lujan, ACS Chem. Neurosci. 2017, 8, 394. 1701269 (29 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de [13] X. W. Yang, S. Gong, Current Protocols in Neuroscience 2005, Ch. 5, p. 20. [14] J. Mattis, K. M. Tye, E. A. Ferenczi, C. Ramakrishnan, D. J. O’Shea, R. Prakash, L. A. Gunaydin, M. Hyun, L. E. Fenno, V. Gradinaru, Nat. Methods 2012, 9, 159. [15] P. J. Rousche, R. A. Normann, J. Neurosci. Methods 1998, 82, 1. [16] T. Roitbak, E. Sykova, Glia 1999, 28, 40. [17] G. C. McConnell, H. D. Rees, A. I. Levey, C. A. Gutekunst, R. E. Gross, R. V. Bellamkonda, J. Neural Eng. 2009, 6, 56003. [18] R. Biran, D. C. Martin, P. A. Tresco, Exp. Neurol. Exp. Neurol. 2005, 195, 115. [19] K. A. Potter, A. C. Buck, W. K. Self, J. R. Capadona, J. Neural Eng. 2012, 9, 046020. [20] T. D. Kozai, Z. Gugel, X. Li, P. J. Gilgunn, R. Khilwani, O. B. Ozdoganlar, G. K. Fedder, D. J. Weber, X. T. Cui, Biomaterials 2014, 35, 9255. [21] M. Jorfi, J. L. Skousen, C. Weder, J. R. Capadona, J. Neural Eng. 2015, 12, 011001. [22] S. P. Lacour, G. Courtine, J. Guck, Nat. Rev. Mater. 2016, 1, 16063. [23] C. H. Thompson, M. J. Zoratti, N. B. Langhals, E. K. Purcell, Tissue Eng, Part B 2016, 22, 125. [24] A. C. Patil, N. V. Thakor, Med. Biol. Eng. Comput. 2016, 54, 23. [25] C. Im, J.-M. Seo, Biomed. Eng. Lett. 2016, 6, 104. [26] J.-W. Jeong, G. Shin, S. I. Park, K. J. Yu, L. Xu, J. A. Rogers, Neuron86, 175. [27] J. H. Lee, H. Kim, J. H. Kim, S.-H. Lee, Lab Chip 2016, 16, 959. [28] T. Kozai, N. Alba, H. Zhang, N. Kotov, R. Gaunt, X. Cui, in Nanotechnology and Neuroscience: Nano-Electronic, Photonic and Mechanical Neuronal Interfacing (Eds: M. D. Vittorio, L. Martiradonna, J. Assad), Springer, New York, NY 2014, Ch. 4, p. 71. [29] P. J. Rousche, D. S. Pellinen, D. P. Pivin Jr., J. C. Williams, R. J. Vetter, D. R. Kipke, IEEE Trans. Biomed. Eng. 2001, 48, 361. [30] J. C. Williams, J. A. Hippensteel, J. Dilgen, W. Shain, D. R. Kipke, J. Neural Eng. 2007, 4, 410. [31] J. C. Williams, R. L. Rennaker, D. R. Kipke, Brain Res. Protoc. 1999, 4, 303. [32] T. D. Kozai, A. S. Jaquins-Gerstl, A. L. Vazquez, A. C. Michael, X. T. Cui, Biomaterials 2016, 87, 157. [33] T. D. Kozai, T. C. Marzullo, F. Hooi, N. B. Langhals, A. K. Majewska, E. B. Brown, D. R. Kipke, J. Neural Eng. 2010, 7, 046011. [34] T. D. Y. Kozai, X. Li, L. M. Bodily, E. M. Caparosa, G. A. Zenonos, D. L. Carlisle, R. M. Friedlander, X. T. Cui, Biomaterials 2014, 35, 9620. [35] T. D. Y. Kozai, K. Catt, X. Li, Z. V. Gugel, V. T. Olafsson, A. L. Vazquez, X. T. Cui, Biomaterials 2015, 37, 25. [36] P. J. Gilgunn, O. X. Chuan, S. N. Flesher, A. B. Schwartz, R. A. Gaunt, presented at 6th International IEEE/EMBS Conference on Neural Engineering (NER), San Diego, USA November 2013. [37] D. McCreery, S. Cogan, S. Kane, V. Pikov, J. Neural Eng. 2016, 13, 036012. [38] J. B. Nadol Jr., J. T. O’Malley, B. J. Burgess, D. Galler, Hear Res. 2014, 318, 11. [39] K. Spiers, T. Cardamone, J. B. Furness, J. C. Clark, J. F. Patrick, G. M. Clark, Cochlear Implants Int. 2016, 17, 129. [40] G. M. Clark, J. Clark, T. Cardamone, M. Clarke, P. Nielsen, R. Jones, B. Arhatari, N. Birbilis, R. Curtain, J. Xu, Cochlear Implants Int. 2014, 15, S1. [41] G. M. Clark, J. C. Clark, J. B. Furness, J. Am. Med. Assoc. 2013, 310, 1225. [42] B. Gunasekera, T. Saxena, R. Bellamkonda, L. Karumbaiah, ACS Chem. Neurosci. 2015, 6, 68. [43] T. D. Kozai, N. B. Langhals, P. R. Patel, X. Deng, H. Zhang, K. L. Smith, J. Lahann, N. A. Kotov, D. R. Kipke, Nat. Mater. 2012, 11, 1065. Adv. Funct. Mater. 2017, 1701269 [44] H. Zhang, J. Shih, J. Zhu, N. A. Kotov, Nano Lett. 2012, 12, 3391. [45] S. Woldring, M. N. Dirken, Acta Physiol. Pharmacol. Neerl. 1950, 1, 369. [46] B. Renshaw, A. Forbes, B. R. Morison, J. Neurophysiol. 1940, 3, 74. [47] N. F. Mott, H. Jones, The Theory of the Properties of Metals and Alloys, Courier Corporation, Toronto, Canada 1958. [48] A. Oliveros, A. Guiseppi-Elie, S. E. Saddow, Biomed. Microdevices 2013, 15, 353. [49] A. Tschöpe, Solid State Ionics 2001, 139, 267. [50] S. P. S. Badwal, J. Mater. Sci. 1984, 19, 1767. [51] J. Y. W. Seto, J. Appl. Phys. 1975, 46, 5247. [52] R. Saha, J. Muthuswamy, Biomed. Microdevices 2007, 9, 345. [53] J. Muthuswamy, M. Okandan, T. Jain, A. Gilletti, IEEE Trans. Biomed. Eng. 2005, 52, 1748. [54] M. Berggren, A. Richter-Dahlfors, Adv. Mater. 2007, 19, 3201. [55] K. Svennersten, K. C. Larsson, M. Berggren, A. Richter-Dahlfors, Biochim. Biophys. Acta, Gen. Subj. 2011, 1810, 276. [56] D. T. Simon, E. O. Gabrielsson, K. Tybrandt, M. Berggren, Chem. Rev. 2016, 116, 13009. [57] R. Green, M. R. Abidian, Adv. Mater. 2015, 27, 7620. [58] D. C. Martin, MRS Commun. 2015, 5, 131. [59] J. Rivnay, R. M. Owens, G. G. Malliaras, Chem. Mater. 2013, 26, 679. [60] U. A. Aregueta-Robles, A. J. Woolley, L. A. Poole-Warren, N. H. Lovell, R. A. Green, Front. Neuroeng. 2015, 7, 57. [61] X. Cui, J. Wiler, M. Dzaman, R. A. Altschuler, D. C. Martin, Biomaterials 2003, 24, 777. [62] P. R. Bidez, S. Li, A. G. MacDiarmid, E. C. Venancio, Y. Wei, P. I. Lelkes, J. Biomater. Sci., Polym. Ed. 2006, 17, 199. [63] R. A. Green, N. H. Lovell, G. G. Wallace, L. A. Poole-Warren, Biomaterials 2008, 29, 3393. [64] D. Svirskis, J. Travas-Sejdic, A. Rodgers, S. Garg, J. Controlled Release 2010, 146, 6. [65] J. Kim, J. Jung, D. Lee, J. Joo, Synth. Met. 2002, 126, 311. [66] C. L. Kolarcik, S. D. Luebben, S. A. Sapp, J. Hanner, N. Snyder, T. D. Y. Kozai, E. Chang, J. A. Nabity, S. T. Nabity, C. F. Lagenaur, X. T. Cui, Soft Matter 2015, 11, 4847. [67] D. Esrafilzadeh, J. M. Razal, S. E. Moulton, E. M. Stewart, G. G. Wallace, J. Controlled Release 2013, 169, 313. [68] Z. J. Du, C. L. Kolarcik, T. D. Y. Kozai, S. D. Luebben, S. A. Sapp, X. S. Zheng, J. A. Nabity, X. T. Cui, Acta Biomater. 2017, 53, 46. [69] M. Armstrong-James, J. Millar, J. Neurosci. Methods 1979, 1, 279. [70] T. D. Y. Kozai, N. B. Langhals, P. R. Patel, X. Deng, H. Zhang, K. L. Smith, J. Lahann, N. A. Kotov, D. R. Kipke, Nat. Mater. 2012, 11, 1065. [71] G. Guitchounts, J. E. Markowitz, W. A. Liberti, T. J. Gardner, J. Neural Eng. 2013, 10, 046016. [72] P. R. Patel, H. Zhang, M. T. Robbins, J. B. Nofar, S. P. Marshall, M. J. Kobylarek, T. D. Y. Kozai, N. A. Kotov, C. A. Chestek, J. Neural Eng. 2016, 13, 066002. [73] P. R. Patel, K. Na, H. Zhang, T. D. Y. Kozai, N. A. Kotov, E. Yoon, C. A. Chestek, J. Neural Eng. 2015, 12, 046009. [74] K. S. Novoselov, A. K. Geim, S. V. Morozov, D. Jiang, Y. Zhang, S. V. Dubonos, I. V. Grigorieva, A. A. Firsov, Science 2004, 306, 666. [75] C.-H. Chen, C.-T. Lin, W.-L. Hsu, Y.-C. Chang, S.-R. Yeh, L.-J. Li, D.-J. Yao, Nanomed.: Nanotechnol. Biol. Med. 2013, 9, 600. [76] B. Li, X. Cao, H. G. Ong, J. W. Cheah, X. Zhou, Z. Yin, H. Li, J. Wang, F. Boey, W. Huang, H. Zhang, Adv. Mater. 2010, 22, 3058. [77] X. Luo, C. L. Weaver, S. Tan, X. T. Cui, J. Mater. Chem. B 2013, 1, 1340. [78] H.-C. Tian, J.-Q. Liu, D.-X. Wei, X.-Y. Kang, C. Zhang, J.-C. Du, B. Yang, X. Chen, H.-Y. Zhu, Y.-N. NuLi, C.-S. Yang, Biomaterials 2014, 35, 2120. [79] N. Rakhilin, B. Barth, J. Choi, N. L. Munoz, S. Kulkarni, J. S. Jones, D. M. Small, Y. T. Cheng, Y. Cao, C. LaVinka, E. Kan, X. Dong, 1701269 (30 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com [80] [81] [82] [83] [84] [85] [86] [87] [88] [89] [90] [91] [92] [93] [94] [95] [96] [97] [98] [99] [100] [101] [102] [103] [104] [105] [106] [107] [108] [109] [110] [111] [112] www.afm-journal.de M. Spencer, P. Pasricha, N. Nishimura, X. Shen, Nat. Commun. 2016, 7, 11800. D. W. Park, A. A. Schendel, S. Mikael, S. K. Brodnick, T. J. Richner, J. P. Ness, M. R. Hayat, F. Atry, S. T. Frye, R. Pashaie, S. Thongpang, Z. Ma, J. C. Williams, Nat. Commun. 2014, 5, 5258. D. Kuzum, H. Takano, E. Shim, J. C. Reed, H. Juul, A. G. Richardson, J. de Vries, H. Bink, M. A. Dichter, T. H. Lucas, D. A. Coulter, E. Cubukcu, B. Litt, Nat. Commun. 2014, 5, 5259. L. Yang, Y. Zhao, W. Xu, E. Shi, W. Wei, X. Li, A. Cao, Y. Cao, Y. Fang, Nano Lett. 2017, 17, 71. A. Fabbro, M. Prato, L. Ballerini, Adv. Drug Delivery Rev. 2013, 65, 2034. E. Jan, J. L. Hendricks, V. Husaini, S. M. Richardson-Burns, A. Sereno, D. C. Martin, N. A. Kotov, Nano Lett. 2009, 9, 4012. E. W. Keefer, B. R. Botterman, M. I. Romero, A. F. Rossi, G. W. Gross, Nat. Nano 2008, 3, 434. T. W. Odom, J.-L. Huang, P. Kim, C. M. Lieber, Nature 1998, 391, 62. R. Saito, M. Fujita, G. Dresselhaus, M. S. Dresselhaus, Appl. Phys. Lett. 1992, 60, 2204. E. Jan, N. A. Kotov, Nano Lett. 2007, 7, 1123. Y. X. Bai, S. S. Ho, N. A. Kotov, Nanoscale 2012, 4, 4393. H. Zhang, P. R. Patel, Z. Xie, S. D. Swanson, X. Wang, N. A. Kotov, ACS Nano 2013, 7, 7619. F. Vitale, S. R. Summerson, B. Aazhang, C. Kemere, M. Pasquali, ACS Nano 2015, 9, 4465. T. Sekitani, Y. Noguchi, K. Hata, T. Fukushima, T. Aida, T. Someya, Science 2008, 321, 1468. J. D. Rolston, S. A. Desai, N. G. Laxpati, R. E. Gross, Neurosurg. Clin. N. Am. 2011, 22, 425. D. A. Robinson, Proc. IEEE 1968, 56, 1065. M. E. J. Obien, K. Deligkaris, T. Bullmann, D. J. Bakkum, U. Frey, Front. Neurosci. 2014, 8, 423. M. J. Nelson, P. Pouget, E. A. Nilsen, C. D. Patten, J. D. Schall, J. Neurosci. Methods 2008, 169, 141. R. Wilke, G. K. Moghadam, N. Lovell, G. Suaning, S. Dokos, J. Neural Eng. 2011, 8, 046016. K. Najafi, J. Ji, K. Wise, IEEE Trans. Biomed. Eng. 1990, 37, 1. J. J. Burmeister, K. Moxon, G. A. Gerhardt, Anal. Chem. 2000, 72, 187. R. Wilke, G. K. Moghadam, N. Lovell, G. Suaning, S. Dokos, J. Neural Eng. 2011, 8, 6016. G. E. Loeb, M. J. Bak, M. Salcman, E. M. Schmidt, IEEE Trans. Biomed. Eng. 1977, BME-24, 121. P. K. Campbell, K. E. Jones, R. J. Huber, K. W. Horch, R. A. Normann, IEEE Trans. Biomed. Eng. 1991, 38, 758. K.-K. Lee, J. He, A. Singh, S. Massia, G. Ehteshami, B. Kim, G. Raupp, J. Micromech. Microeng. 2003, 14, 32. D. S. Lee, S. J. Kim, J. H. Sohn, I. G. Kim, S. W. Kim, D. W. Sohn, J. H. Kim, B. Choi, Chin. J. Polym. Sci. 2012, 30, 242. K.-I. Naka, R. Kido, Brain Res. 1967, 5, 422. V. Sankar, E. Patrick, R. Dieme, J. C. Sanchez, A. Prasad, T. Nishida, Front. Neuroeng. 2014, 7, 13. J. McPherson, J. Y. Kim, A. Shanware, H. Mogul, Appl. Phys. Lett. 2003, 82, 2121. L. A. Dissado, J. C. Fothergill, Electrical Degradation and Breakdown in Polymers, IET, London, UK 1992. X. Xie, L. Rieth, R. Caldwell, S. Negi, R. Bhandari, R. Sharma, P. Tathireddy, F. Solzbacher, Biomed. Microdevices 2015, 17, 1. T. D. Kozai, A. L. Vazquez, J. Mater. Chem. B Mater. Biol. Med. 2015, 3, 4965. A. Prasad, Q.-S. Xue, V. Sankar, T. Nishida, G. Shaw, W. J. Streit, J. C. Sanchez, J. Neural Eng. 2012, 9, 056015. E. Patrick, M. E. Orazem, J. C. Sanchez, T. Nishida, J. Neurosci. Methods 2011, 198, 158. Adv. Funct. Mater. 2017, 1701269 [113] E. Marg, Nature 1964, 202, 601. [114] E. Schmidt, J. McIntosh, M. Bak, Med. Biol. Eng. Comput. 1988, 26, 96. [115] A. J. Bard, M. V. Mirkin, Scanning Electrochemical Microscopy, CRC Press, Boca Raton, USA 2012. [116] K. C. Morton, C. A. Morris, M. A. Derylo, R. Thakar, L. A. Baker, Anal. Chem. 2011, 83, 5447. [117] I. W. Boyd, Laser Processing of Thin Films and Microstructures: Oxidation, Deposition and Etching of Insulators, Springer Science & Business Media, Berlin, Heidelberg 2013. [118] A. Prasad, Q. S. Xue, V. Sankar, T. Nishida, G. Shaw, W. J. Streit, J. C. Sanchez, J. Neural Eng. 2012, 9, 056015. [119] M. Omura, M. Hirata, A. Tanaka, M. Zhao, Y. Makita, N. Inoue, K. Gotoh, N. Ishinishi, Toxicol. Lett. 1996, 89, 123. [120] S. Cherevko, A. A. Topalov, A. R. Zeradjanin, I. Katsounaros, K. J. J. Mayrhofer, RSC Adv. 2013, 3, 16516. [121] J. P. Seymour, N. B. Langhals, D. J. Anderson, D. R. Kipke, Biomed. Microdevices 2011, 13, 441. [122] J. Batey, E. Tierney, J. Appl. Phys. 1986, 60, 3136. [123] T. C. Yang, K. C. Saraswat, IEEE Trans. Electron Devices 2000, 47, 746. [124] Y. Liu, I. K. Lin, X. Zhang, Mater. Sci. Eng., A 2008, 489, 294. [125] J. F. Hetke, J. L. Lund, K. Najafi, K. D. Wise, D. J. Anderson, IEEE Trans. Biomed. Eng. 1994, 41, 314. [126] T. O. Ogurtani, J. Chem. Phys. 2006, 124, 144706. [127] O. Schenk, J. L. Urai, S. Piazolo, Geofluids 2006, 6, 93. [128] O. Schenk, J. Urai, Contrib. Mineral. Petrol. 2004, 146, 671. [129] B. J. Kim, E. P. Washabaugh, E. Meng, presented at 2014 IEEE 27th International Conference on Micro Electro Mechanical Systems (MEMS), San Francisco, USA January 2014. [130] B. J. Kim, C. A. Gutierrez, E. Meng, J. Microelectromech. Syst. 2015, 24, 1534. [131] M. Yoshida, J. Lahann, ACS Nano 2008, 2, 1101. [132] T. D. Y. Kozai, N. B. Langhals, P. R. Patel, X. Deng, H. Zhang, K. L. Smith, J. Lahann, N. A. Kotov, D. R. Kipke, Nat. Mater. 2012, 11, 1065. [133] H. Nandivada, H. Y. Chen, L. Bondarenko, J. Lahann, Angew. Chem. Int. Ed. Engl. 2006, 45, 3360. [134] J. Lahann, Polym. Int. 2006, 55, 1361. [135] Y. Elkasabi, H. Y. Chen, J. Lahann, Adv. Mater. 2006, 18, 1521. [136] J. Lahann, H. Y. Chen, Abstr. Pap. Am. Chem. Soc. 2005, 230, U4292. [137] J. Lahann, Abstr. Pap. Am. Chem. Soc. 2005, 229, U1106. [138] J. Lahann, M. Balcells, H. Lu, T. Rodon, K. F. Jensen, R. Langer, Anal. Chem. 2003, 75, 2117. [139] J. Lahann, S. Mitragotri, T. N. Tran, H. Kaido, J. Sundaram, I. S. Choi, S. Hoffer, G. A. Somorjai, R. Langer, Science 2003, 299, 371. [140] J. Lahann, H. Hocker, R. Langer, Angew. Chem., Int. Ed. Engl. 2001, 40, 2947. [141] J. Lahann, D. Klee, W. Pluester, H. Hoecker, Biomaterials 2001, 22, 817. [142] J. Lahann, I. S. Choi, J. Lee, K. F. Jenson, R. Langer, Angew. Chem. Int. Ed. 2001, 40, 3166. [143] F. Bally, K. Cheng, H. Nandivada, X. Deng, A. M. Ross, A. Panades, J. Lahann, ACS Appl. Mater. Interfaces 2013, 5, 9262. [144] X. Deng, T. W. Eyster, Y. Elkasabi, J. Lahann, Macromol. Rapid Commun. 2012, 33, 640. [145] J. P. Seymour, Y. M. Elkasabi, H. Y. Chen, J. Lahann, D. R. Kipke, Biomaterials 2009, 30, 6158. [146] A. Prasad, Q. S. Xue, R. Dieme, V. Sankar, R. C. Mayrand, T. Nishida, W. J. Streit, J. C. Sanchez, Front. Neuroeng. 2014, 7, 2. [147] J. S. Ordonez, L. Rudmann, P. Cvancara, C. Bentler, T. Stieglitz, presented at Engineering in Medicine and Biology Society (EMBC), 1701269 (31 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com [148] [149] [150] [151] [152] [153] [154] [155] [156] [157] [158] [159] [160] [161] [162] [163] [164] [165] [166] [167] [168] [169] [170] [171] [172] [173] [174] [175] [176] [177] [178] [179] [180] [181] www.afm-journal.de 2015 37th Annual International Conference of the IEEE, Milan, Italy, August 2015. F. Mechler, J. D. Victor, J. Comput. Neurosci. 2012, 32, 73. J. Du, M. L. Roukes, S. C. Masmanidis, J. Micromech. Microeng. 2009, 19, 075008. M. A. Moffitt, C. C. McIntyre, Clin. Neurophysiol. 2005, 116, 2240. N. V. Swindale, M. A. Spacek, Front.Syst. Neurosci. 2014, 8, 6. T. J. Blanche, M. A. Spacek, J. F. Hetke, N. V. Swindale, J. Neurophysiol. 2005, 93, 2987. S. F. Lempka, M. D. Johnson, M. A. Moffitt, K. J. Otto, D. R. Kipke, C. C. McIntyre, J. Neural Eng. 2011, 8, 045006. I. D. Dryg, M. P. Ward, K. Y. Qing, H. Mei, J. E. Schaffer, P. P. Irazoqui, IEEE Trans. Neural Syst. Rehabil. Eng. 2015, 23, 562. T. D. Y. Kozai, K. Catt, Z. Du, K. Na, O. Srivannavit, R.-U. M. Haque, J. Seymour, K. D. Wise, E. Yoon, X. T. Cui, IEEE Transactions on Bio-medical Engineering 2016, 63, 111. N. S. Dias, J. P. Carmo, A. F. da Silva, P. M. Mendes, J. H. Correia, Sens. Actuators, A 2010, 164, 28. X. F. Wei, W. M. Grill, Front. Neuroeng. 2009, 2, 15. A. M. Dymond, IEEE Trans. Biomed. Eng. 1976, BME-23, 274. S. F. Cogan, D. J. Garrett, R. A. Green, Neurobionics: The Biomedical Engineering of Neural Prostheses, John Wiley & Sons, Hoboken, USA 2016, p. 55. J. V. Macpherson, Phys. Chem. Chem. Phys. 2015, 17, 2935. H. B. Martin, A. Argoitia, U. Landau, A. B. Anderson, J. C. Angus, J. Electrochem. Soc. 1996, 143, L133. J. T. Mortimer, D. Kaufman, U. Roessman, Ann. Biomed. Eng. 1980, 8, 235. R. L. White, T. J. Gross, IEEE Trans. Biomed. Eng. 1974, BME-21, 487. P. F. Johnson, L. L. Hench, Brain Behav. Evol. 1977, 14, 23. S. F. Cogan, P. R. Troyk, J. Ehrlich, T. D. Plante, IEEE Trans. Biomed. Eng. 2005, 52, 1612. T. L. Rose, E. M. Kelliher, L. S. Robblee, J. Neurosci. Methods 1985, 12, 181. T. Rose, L. Robblee, IEEE Trans. Biomed. Eng. 1990, 37, 1118. L. S. Robblee, J. McHardy, W. F. Agnew, L. A. Bullara, J. Neurosci. Methods 1983, 9, 301. X. Beebe, T. L. Rose, IEEE Trans. Biomed. Eng. 1988, 35, 494. S. F. Cogan, P. R. Troyk, J. Ehrlich, T. D. Plante, D. E. Detlefsen, IEEE Trans. Biomed. Eng. 2006, 53, 327. S. F. Cogan, T. D. Plante, J. Ehrlich, presented at Engineering in Medicine and Biology Society, 2004. IEMBS ’04. 26th Annual International Conference of the IEEE, San Francisco, USA September 2004. R. D. Meyer, S. F. Cogan, T. H. Nguyen, R. D. Rauh, IEEE Trans. Neural Syst. Rehabil. Eng. 2001, 9, 2. S. Negi, R. Bhandari, F. Solzbacher, J. Sens. Technol. 2012, 2, 138. M. Jayalakshmi, K. Balasubramanian, Int. J. Electrochem. Sci. 2008, 3, 1196. J. D. Weiland, D. J. Anderson, M. S. Humayun, IEEE Trans. Biomed. Eng. 2002, 49, 1574. U. Egert, D. Heck, A. Aertsen, Exp. Brain Res. 2002, 142, 268. W. Franks, I. Schenker, P. Schmutz, A. Hierlemann, IEEE Trans. Biomed. Eng. 2005, 52, 1295. L. Q. Jiang, L. Gao, J. Am. Ceram. Soc. 2006, 89, 156. J. D. Weiland, D. J. Anderson, M. S. Humayun, IEEE Trans. Biomed. Eng. 2002, 49, 1574. M. Janders, U. Egert, M. Stelzle, W. Nisch, presented at Engineering in Medicine and Biology Society, Proceedings of the 18th Annual International Conference of the IEEE, Bridging Disciplines for Biomedicine, Amsterdam, The Netherlands October–November 1996. J. M. Gaylor, G. Raman, M. Chung, J. Lee, M. Rao, J. Lau, D. S. Poe, JAMA Otolaryngol. 2013, 139, 265. Adv. Funct. Mater. 2017, 1701269 [182] [183] [184] [185] [186] [187] [188] [189] [190] [191] [192] [193] [194] [195] [196] [197] [198] [199] [200] [201] [202] [203] [204] [205] [206] [207] [208] [209] [210] [211] [212] [213] [214] [215] [216] [217] [218] T. D. Y. Kozai, A. L. Vazquez, J. Mater. Chem. B 2015, 3, 4965. M. H. Histed, V. Bonin, R. C. Reid, Neuron 2009, 63, 508. F. Rattay, C. Wenger, Neuroscience 2010, 170, 399. S. F. Cogan, K. A. Ludwig, C. G. Welle, P. Takmakov, J. Neural Eng. 2016, 13, 021001. D. R. Merrill, M. Bikson, J. G. R. Jefferys, J. Neurosci. Methods 2005, 141, 171. R. V. Shannon, IEEE Trans. Biomed. Eng. 1992, 39, 424. S. Negi, R. Bhandari, F. Solzbacher, Conf. Proc. IEEE Eng. Med. Biol. Soc. 2012, 2012, 5142. A. Norlin, J. Pan, C. LeygrafJ. Electrochem. Soc. 2005, 152, J85. H. K. Song, Y. H. Jung, K. H. Lee, L. H. Dao, Electrochim. Acta 1999, 44, 3513. J. Han, J. Han, B. S. Lee, J. Lim, S. M. Kim, H. Kim, S. Kang, J. Micromech. Microeng. 2012, 22, 065004. B. Wang, A. Petrossians, J. D. Weiland, IEEE Trans. Biomed. Eng. 2014, 61, 2254. K. A. Ludwig, N. B. Langhals, M. D. Joseph, S. M. RichardsonBurns, J. L. Hendricks, D. R. Kipke, J. Neural Eng. 2011, 8, 014001. X. Y. Cui, J. Wiler, M. Dzaman, R. A. Altschuler, D. C. Martin, Biomaterials 2003, 24, 777. R. A. Green, R. T. Hassarati, L. Bouchinet, C. S. Lee, G. L. M. Cheong, J. F. Yu, C. W. Dodds, G. J. Suaning, L. A. Poole-Warren, N. H. Lovell, Biomaterials 2012, 33, 5875. X. T. Cui, D. D. Zhou, IEEE Trans. Neural Syst. Rehabil. Eng. 2007, 15, 502. S. F. Cogan, Annu. Rev. Biomed. Eng. 2008, 10, 275. S. Baek, R. A. Green, L. A. Poole-Warren, Acta Biomater. 2014, 10, 3048. D. C. Martin, K. E. Feldman, US Patent 8586702, 2013. K. T. Kawagoe, J. B. Zimmerman, R. M. Wightman, J. Neurosci. Methods 1993, 48, 225. B. K. Swamy, B. J. Venton, Anal. Chem. 2007, 79, 744. P. Takmakov, M. K. Zachek, R. B. Keithley, P. L. Walsh, C. Donley, G. S. McCarty, R. M. Wightman, Anal. Chem. 2010, 82, 2020. M. K. Zachek, A. Hermans, R. M. Wightman, G. S. McCarty, J. Electroanal. Chem. 2008, 614, 113. K. Jackowska, P. Krysinski, Anal. Bioanal. Chem. 2013, 405, 3753. J. G. Roberts, K. L. Hamilton, L. A. Sombers, Analyst 2011, 136, 3550. M. D. Johnson, O. E. Kao, D. R. Kipke, J. Neurosci. Methods 2007, 160, 276. K. Pihel, Q. D. Walker, R. M. Wightman, Anal. Chem. 1996, 68, 2084. R. F. Vreeland, C. W. Atcherley, W. S. Russell, J. Y. Xie, D. Lu, N. D. Laude, F. Porreca, M. L. Heien, Anal. Chem. 2015, 87, 2600. B. Sarada, T. N. Rao, D. Tryk, A. Fujishima, Anal. Chem. 2000, 72, 1632. M. Zhu, C. Zeng, J. Ye, Electroanalysis 2011, 23, 907. M. K. Zachek, P. Takmakov, B. Moody, R. M. Wightman, G. S. McCarty, Anal. Chem. 2009, 81, 6258. J. M. Halpern, M. J. Cullins, H. J. Chiel, H. B. Martin, Diamond Relat. Mater. 2010, 19, 178. X. Xiao, J. Wang, C. Liu, J. A. Carlisle, B. Mech, R. Greenberg, D. Guven, R. Freda, M. S. Humayun, J. Weiland, O. Auciello, J. Biomed. Mater. Res., Part B 2006, 77B, 273. S. F. Cogan, A. A. Guzelian, W. F. Agnew, T. G. H. Yuen, D. B. McCreery, J. Neurosci. Methods 2004, 137, 141. L. S. Robblee, J. McHardy, W. F. Agnew, L. A. Bullara, J. Neurosci. Methods 1983, 9, 301. E. Patrick, M. E. Orazem, J. C. Sanchez, T. Nishida, J. Neurosci. Methods 2011, 198, 158. R. H. Pudenz, Surgery 1942, 12, 791. A. M. Dymond, L. E. Kaechele, J. M. Jurist, P. H. Crandall, J. Neurosurg. 1970, 33, 574. 1701269 (32 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com [219] [220] [221] [222] [223] [224] [225] [226] [227] [228] [229] [230] [231] [232] [233] [234] [235] [236] [237] [238] [239] [240] [241] [242] [243] [244] [245] [246] [247] [248] [249] [250] www.afm-journal.de F. R. Robinson, M. T. Johnson, DTIC Document, 1961. J. T. McFadden, J. Neurosurg. 1969, 31, 373. B. J. Piersma, W. Greatbarch, J. Electrochem. Soc. 1987, 134, 2458. L. S. Robblee, T. L. Rose, in Neural Prostheses: Fundamental Studies, (Eds: W. F. Agnew, D. B. McCreery), Prentice-Hall, Englewood Cliffs, USA 1990, p. 25. S. B. Brummer, L. S. Robblee, F. T. Hambrecht, Ann. N. Y. Acad. Sci. 1983, 405, 159. A. Petrossians, J. J. Whalen, J. D. Weiland, F. Mansfeld, J. Electrochem. Soc. 2011, 158, D269. C. A. Marrese, Anal. Chem. 1987, 59, 217. J. D. Weiland, D. J. Anderson, IEEE Trans. Biomed. Eng. 2000, 47, 911. K. A. Moxon, N. M. Kalkhoran, M. Markert, M. A. Sambito, J. L. McKenzie, J. T. Webster, IEEE Trans. Biomed. Eng. 2004, 51, 881. S. A. Desai, J. D. Rolston, L. Guo, S. M. Potter, Front. Neuroeng. 2010, 3, 5. M. Heim, B. Yvert, A. Kuhn, J. Physiol. 2012, 106, 137. P. Ocón, A. B. Cristobal, P. Herrasti, E. Fatas, Corros. Sci. 2005, 47, 649. T. Boretius, M. Schuettler, T. Stieglitz, Artif. Organs 2011, 35, 245. R. A. Green, N. H. Lovell, L. A. Poole-Warren, Acta Biomater. 2010, 6, 63. X. Cui, D. C. Martin, Sens. Actuators, B 2003, 89, 92. M. Nishizawa, H. Nozaki, H. Kaji, T. Kitazume, N. Kobayashi, T. Ishibashi, T. Abe, Biomaterials 2007, 28, 1480. R. A. Green, R. T. Hassarati, L. Bouchinet, C. S. Lee, G. L. Cheong, J. F. Yu, C. W. Dodds, G. J. Suaning, L. A. Poole-Warren, N. H. Lovell, Biomaterials 2012, 33, 5875. R. A. Green, R. T. Hassarati, L. Bouchinet, C. S. Lee, G. L. Cheong, F. Y. Jin, C. W. Dodds, G. J. Suaning, L. A. Poole-Warren, N. H. Lovell, Biomaterials 2012, 33, 5875. T. D. Y. Kozai, N. A. Alba, H. Zhang, N. A. Kotov, R. A. Gaunt, X. T. Cui, in Nanotechnology and Neuroscience: Nano-Electronic, Photonic and Mechanical Neuronal Interfacing (Eds: M. De Vittorio, L. Martiradonna, J. Assad), Springer, New York 2014, p. 71. M. Burke, B. Clarke, Y. Rochev, A. Gorelov, W. Carroll, J. Mater. Sci. Mater. Med. 2008, 19, 1971. D. E. Tallman, C. Vang, G. G. Wallace, G. P. Bierwagen, J. Electrochem. Soc. 2002, 149, C173. J. T. Mortimer, D. Kaufman, U. Roessmann, Ann. Biomed. Eng. 1980, 8, 235. X. Luo, C. L. Weaver, D. D. Zhou, R. Greenberg, X. T. Cui, Biomaterials 2011, 32, 5551. X. T. Cui, D. D. Zhou, IEEE Trans. Neural Syst. Rehabil. Eng. 2007, 15, 502. H. Zhou, X. Cheng, L. Rao, T. Li, Y. Y. Duan, Acta Biomater. 2013, 9, 6439. S. Wilks, S. M. Richardson-Burns, J. L. Hendricks, D. C. Martin, K. J. Otto, Front. Neuroeng. 2009, 2, 7. C. L. Kolarcik, K. Catt, E. Rost, I. N. Albrecht, D. Bourbeau, Z. Du, T. D. Kozai, X. Luo, D. J. Weber, X. T. Cui, J. Neural Eng. 2015, 12, 016008. G. Z. Chen, M. S. P. Shaffer, D. Coleby, G. Dixon, W. Zhou, D. J. Fray, A. H. Windle, Adv. Mater. 2000, 12, 522. K. Wang, H. A. Fishman, H. Dai, J. S. Harris, Nano Lett 2006, 6, 2043. T. D. Y. Kozai, K. Catt, Z. Du, N. Kyounghwan, O. Srivannavit, R. M. Haque, J. P. Seymour, E. Yoon, X. T. Cui, IEEE Trans. Biomed. Eng. 2015, 63, 111. T. J. Webster, M. C. Waid, J. L. McKenzie, R. L. Price, J. U. Ejiofor, Nanotechnology 2003, 15, 48. S. Sharifi, S. Behzadi, S. Laurent, M. L. Forrest, P. Stroeve, M. Mahmoudi, Chem. Soc. Rev. 2012, 41, 2323. Adv. Funct. Mater. 2017, 1701269 [251] H. Song, Y. Jung, K. Lee, L. Dao, Electrochim. Acta 1999, 44, 3513. [252] M. Santini, F. De Seta, T. I. M. S. G. o. L. O. Stimulation, Pacing Clin. Electrophysiol. 1993, 16, 722. [253] C. B. James, A. Juan, P. D. John, J. Neural Eng. 2016, 13, 026003. [254] P. Takmakov, K. Ruda, K. S. Phillips, I. S. Isayeva, V. Krauthamer, C. G. Welle, J. Neural Eng. 2015, 12, 026003. [255] S. Meijs, M. Alcaide, C. Sorensen, M. McDonald, S. Sorensen, K. Rechendorff, A. Gerhardt, M. Nesladek, N. J. Rijkhoff, C. P. Pennisi, J. Neural Eng. 2016, 13, 056011. [256] D. H. Szarowski, M. D. Andersen, S. Retterer, A. J. Spence, M. Isaacson, H. G. Craighead, J. N. Turner, W. Shain, Brain Res. 2003, 983, 23. [257] S. F. Bernatchez, P. J. Parks, D. F. Gibbons, Biomaterials 1996, 17, 2077. [258] J. E. Sanders, C. E. Stiles, C. L. Hayes, J. Biomed. Mater. Res. 2000, 52, 231. [259] J. P. Seymour, D. R. Kipke, Biomaterials 2007, 28, 3594. [260] V. S. Polikov, P. A. Tresco, W. M. Reichert, J. Neurosci. Methods 2005, 148, 1. [261] A. Ersen, S. Elkabes, D. S. Freedman, M. Sahin, J. Neural Eng. 2015, 12, 016019. [262] J. L. Skousen, M. J. Bridge, P. A. Tresco, Biomaterials 2015, 36, 33. [263] A. a. Schendel, M. W. Nonte, C. Vokoun, T. J. Richner, S. K. Brodnick, F. Atry, S. Frye, P. Bostrom, R. Pashaie, S. Thongpang, K. W. Eliceiri, J. C. Williams, J. Neural Eng. 2014, 11, 046011. [264] A. A. Schendel, S. Thongpang, S. K. Brodnick, T. J. Richner, B. D. B. Lindevig, L. Krugner-Higby, J. C. Williams, J. Neurosci. Methods 2013, 218, 121. [265] W. Wang, W. Tan, D. Luo, J. Lin, Y. Yu, Q. Wang, W. Zhao, B. Wu, J. Chen, J. He, BMC Neurosci. 2012, 13, 147. [266] T. L. Roth, D. Nayak, T. Atanasijevic, A. P. Koretsky, L. L. Latour, D. B. McGavern, Nature 2014, 505, 223. [267] J. P. Seymour, F. Wu, K. D. Wise, E. Yoon, Microsyst. Nanoeng. 2017, 3, 16066. [268] A. Fujishiro, H. Kaneko, T. Kawashima, M. Ishida, T. Kawano, Sci. Rep. 2014, 4, 4868. [269] A. Canales, X. Jia, U. P. Froriep, R. A. Koppes, C. M. Tringides, J. Selvidge, C. Lu, C. Hou, L. Wei, Y. Fink, P. Anikeeva, Nat. Biotechnol. 2015, 33, 277. [270] D. E. Gunning, J. M. Beggs, W. Dabrowski, P. Hottowy, C. J. Kenney, A. Sher, A. M. Litke, K. Mathieson, J. Neural Eng. 2013, 10, 016007. [271] T. D. Y. Kozai, N. B. Langhals, P. R. Patel, X. Deng, H. Zhang, K. L. Smith, J. Lahann, N. A. Kotov, D. R. Kipke, Nat. Mater. 2012, 11, 1065. [272] S. M. Ojovan, N. Rabieh, N. Shmoel, H. Erez, E. Maydan, A. Cohen, M. E. Spira, Sci. Rep. 2015, 5, 14100. [273] D. B. Suyatin, L. Wallman, J. Thelin, C. N. Prinz, H. Jörntell, L. Samuelson, L. Montelius, J. Schouenborg, PLoS ONE 2013, 8, e56673. [274] C. Xie, Z. Lin, L. Hanson, Y. Cui, B. Cui, Nat. Nano 2012, 7, 185. [275] M. Wne˛k, M. Ł. Górzny, M. B. Ward, C. Wälti, A. G. Davies, R. Brydson, S. D. Evans, P. G. Stockley, Nanotechnology 2013, 24, 025605. [276] W. Shenton, T. Douglas, M. Young, G. Stubbs, S. Mann, Adv. Mater. 1999, 11, 253. [277] E. Dujardin, C. Peet, G. Stubbs, J. N. Culver, S. Mann, Nano Lett. 2003, 3, 413. [278] J. S. Lim, S. M. Kim, S. Y. Lee, E. A. Stach, J. N. Culver, M. T. Harris, J. Colloid Interface Sci. 2010, 342, 455. [279] M. Knez, A. M. Bittner, F. Boes, C. Wege, H. Jeske, E. Maiβ, K. Kern, Nano Lett. 2003, 3, 1079. 1701269 (33 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de [280] J. J. Clark, S. G. Sandberg, M. J. Wanat, J. O. Gan, E. A. Horne, A. S. Hart, C. A. Akers, J. G. Parker, I. Willuhn, V. Martinez, Nat. Methods 2010, 7, 126. [281] Y. S. Singh, L. E. Sawarynski, P. D. Dabiri, W. R. Choi, A. M. Andrews, Anal. Chem. 2011, 83, 6658. [282] K. E. Bennet, J. R. Tomshine, H.-K. Min, F. S. Manciu, M. P. Marsh, S. B. Paek, M. L. Settell, E. N. Nicolai, C. D. Blaha, A. Z. Kouzani, Front. Hum. Neurosci. 2016, 10, 102. [283] P. U. Arumugam, H. Zeng, S. Siddiqui, D. P. Covey, J. A. Carlisle, P. A. Garris, Appl. Phys. Lett. 2013, 102, 253107. [284] L. M. Borland, G. Shi, H. Yang, A. C. Michael, J. Neurosci. Methods 2005, 146, 149. [285] A. S. Khan, A. C. Michael, TrendsAnal. Chem. 2003, 22, 503. [286] A. Jaquins-Gerstl, A. C. Michael, J. Neurosci. Methods 2009, 183, 127. [287] M. Frédéric, A. Arno, H. Tobias, R. Patrick, B. Gustaaf, M. Bruce, K. Fabian, J. Neural Eng. 2016, 13, 046018. [288] C. M. Gray, P. E. Maldonado, M. Wilson, B. McNaughton, J. Neurosci. Methods 1995, 63, 43. [289] T. D. Y. Kozai, T. C. Marzullo, F. Hooi, N. B. Langhals, A. K. Majewska, E. B. Brown, D. R. Kipke, J. Neural Eng. 2010, 7, 046011. [290] E. Fernández, B. Greger, P. A. House, I. Aranda, C. Botella, J. Albisua, C. Soto-Sánchez, A. Alfaro, R. A. Normann, Front. Neuroeng. 2014, 7, 24. [291] P. A. House, J. D. MacDonald, P. A. Tresco, R. A. Normann, Neurosurg. Focus. 2006, 20, E4. [292] N. F. Nolta, M. B. Christensen, P. D. Crane, J. L. Skousen, P. A. Tresco, Biomaterials 2015, 53, 753. [293] A. Y. Shih, P. Blinder, P. S. Tsai, B. Friedman, G. Stanley, P. D. Lyden, D. Kleinfeld, Nat. Neurosci. 2012, 16, 55. [294] G. Lind, L. Gällentoft, N. Danielsen, J. Schouenborg, L. M. E. Pettersson, PLoS ONE 2012, 7, e47509. [295] G. C. McConnell, H. D. Rees, A. I. Levey, C.-A. Gutekunst, R. E. Gross, R. V. Bellamkonda, J. Neural Eng. 2009, 6, 056003. [296] X. Liu, D. B. McCreery, L. A. Bullara, W. F. Agnew, IEEE Trans. Neural Syst. Rehabil. Eng. 2006, 14, 91. [297] G. L. Baumbach, J. E. Siems, F. M. Faraci, D. D. Heistad, Am. J. Physiol. 1989, 256, H493. [298] M. P. Ward, P. Rajdev, C. Ellison, P. P. Irazoqui, Brain Res. 2009, 1282, 183. [299] T. Saxena, L. Karumbaiah, E. A. Gaupp, R. Patkar, K. Patil, M. Betancur, G. B. Stanley, R. V. Bellamkonda, Biomaterials 2013, 34, 4703. [300] L. Karumbaiah, T. Saxena, D. Carlson, K. Patil, R. Patkar, E. A. Gaupp, M. Betancur, G. B. Stanley, L. Carin, R. V. Bellamkonda, Biomaterials 2013, 34, 8061. [301] A. Gilletti, J. Muthuswamy, J. Neural Eng. 2006, 3, 189. [302] T. D. Y. Kozai, K. Catt, X. Li, Z. V. Gugel, V. T. Olafsson, A. L. Vazquez, X. T. Cui, Biomaterials 2015, 37, 25. [303] M. Paukert, D. E. Bergles, J. Physiol. 2012, 590, 2955. [304] D. A. Feinberg, A. S. Mark, Radiology 1987, 163, 793. [305] A. Gilletti, J. Muthuswamy, J. Neural Eng. 2006, 3, 189. [306] J. Subbaroyan, D. C. Martin, D. R. Kipke, J. Neural Eng. 2005, 2, 103. [307] H. Lee, R. V. Bellamkonda, W. Sun, M. E. Levenston, J. Neural Eng. 2005, 2, 81. [308] P. Moshayedi, G. Ng, J. C. Kwok, G. S. Yeo, C. E. Bryant, J. W. Fawcett, K. Franze, J. Guck, Biomaterials 2014, 35, 3919. [309] T. Someya, Z. Bao, G. G. Malliaras, Nature 2016, 540, 379. [310] C. Liao, M. Zhang, M. Y. Yao, T. Hua, L. Li, F. Yan, Adv. Mater. 2015, 27, 7493. [311] U. A. Aregueta-Robles, A. J. Woolley, L. A. Poole-Warren, N. H. Lovell, R. A. Green, Front. Neuroeng. 2014, 7, 15. Adv. Funct. Mater. 2017, 1701269 [312] R. A. Green, S. Baek, L. A. Poole-Warren, P. J. Martens, Sci. Technol. Adv. Mater. 2010, 11, 014107. [313] Y. Zhong, R. V. Bellamkonda, J. R. Soc., Interface 2008, 5, 957. [314] L. Bollmann, D. E. Koser, R. Shahapure, H. O. B. Gautier, G. A. Holzapfel, G. Scarcelli, M. C. Gather, E. Ulbricht, K. Franze, Front. Cell. Neurosci. 2015, 9, 363. [315] S. M. Gutowski, K. L. Templeman, A. B. South, J. C. Gaulding, J. T. Shoemaker, M. C. LaPlaca, R. V. Bellamkonda, L. A. Lyon, A. J. García, J. Biomed. Mater. Res., Part A 2014, 102, 1486. [316] L. Rao, H. Zhou, T. Li, C. Li, Y. Y. Duan, Acta Biomater. 2012, 8, 2233. [317] P. J. Gilgunn, R. Khilwani, T. D. Y. Kozai, D. J. Weber, X. T. Cui, G. Erdos, O. B. Ozdoganlar, G. K. Fedder, presented at 2012 IEEE 25th International Conference on Micro Electro Mechanical Systems (MEMS), Paris, France January–February 2012. [318] R. Khilwani, P. J. Gilgunn, T. D. Y. Kozai, X. C. Ong, E. Korkmaz, P. K. Gunalan, X. T. Cui, G. K. Fedder, O. B. Ozdoganlar, Biomed. Microdevices 2016, 18, 97. [319] H. S. Sohal, A. Jackson, R. Jackson, G. J. Clowry, K. Vassilevski, A. O’Neill, S. N. Baker, Front. Neuroeng. 2015. [320] J. Subbaroyan, Investigations of Tethering Induced Injury Response in Brain Tissue by Intracortical Implants through Modeling and In Vivo Experiments, University of Michigan, Ann Arbor, USA 2007. [321] U.-H. Shin, D.-W. Jeong, S.-M. Park, S.-H. Kim, H. W. Lee, J.-M. Kim, Carbon 2014, 80, 396. [322] T.-i. Kim, J. G. McCall, Y. H. Jung, X. Huang, E. R. Siuda, Y. Li, J. Song, Y. M. Song, H. A. Pao, R.-H. Kim, Science 2013, 340, 211. [323] J. J. Norton, D. S. Lee, J. W. Lee, W. Lee, O. Kwon, P. Won, S.-Y. Jung, H. Cheng, J.-W. Jeong, A. Akce, Proc. Natl. Acad. Sci. USA 2015, 112, 3920. [324] K.-Y. Chun, Y. Oh, J. Rho, J.-H. Ahn, Y.-J. Kim, H. R. Choi, S. Baik, Nat. Nanotechnol. 2010, 5, 853. [325] D.-H. Kim, N. Lu, R. Ma, Y.-S. Kim, R.-H. Kim, S. Wang, J. Wu, S. M. Won, H. Tao, A. Islam, Science 2011, 333, 838. [326] E. Fiedler, N. Haas, T. Stieglitz, presented at Engineering in Medicine and Biology Society (EMBC), 36th Annual International Conference of the IEEE, Chicago, Il, USA August 2014. [327] S. Loffler, Y. Xie, P. Klimach, A. Richter, P. Detemple, T. Stieglitz, A. Moser, U. G. Hofmann, Biomed. Technol. 2012, 57, 104. [328] C. Hassler, T. Boretius, T. Stieglitz, J. Polym. Sci., Part B: Polym. Phys. 2011, 49, 18. [329] C. Hassler, R. P. von Metzen, P. Ruther, T. Stieglitz, J. Biomed. Mater. Res., Part B 2010, 93B, 266. [330] C. Hassler, R. Metzen, T. Stieglitz, in 4th European Conf. of the Int. Federation for Medical and Biological Engineering, Vol. 22 (Eds: J. Sloten, P. Verdonck, M. Nyssen, J. Haueisen), Springer, Berlin 2009, Ch. 585, p. 2439. [331] S. Kisban, S. Herwik, K. Seidl, B. Rubehn, A. Jezzini, M. A. Umilta, L. Fogassi, T. Stieglitz, O. Paul, P. Ruther, Conf. Proc. IEEE Eng. Med. Biol. Soc. 2007, 2007, 175. [332] F. J. Rodriguez, D. Ceballos, M. Schuttler, A. Valero, E. Valderrama, T. Stieglitz, X. Navarro, J. Neurosci. Methods 2000, 98, 105. [333] T. Stieglitz, H. r. Beutel, M. Schuettler, J.-U. Meyer, Biomed. Microdevices 2000, 2, 283. [334] T. Stieglitz, H. Beutel, J. U. Meyer, Sens. Actuators, A 1997, 60, 240. [335] J. T. Kuo, B. J. Kim, S. A. Hara, C. D. Lee, C. A. Gutierrez, T. Q. Hoang, E. Meng, Lab Chip 2013, 13, 554. [336] B. J. Kim, J. T. Kuo, S. A. Hara, C. D. Lee, L. Yu, C. A. Gutierrez, T. Q. Hoang, V. Pikov, E. Meng, J. Neural Eng. 2013, 10, 045002. [337] T. M. Fu, G. Hong, T. Zhou, T. G. Schuhmann, R. D. Viveros, C. M. Lieber, Nat. Methods 2016, 13, 875. [338] X. Dai, W. Zhou, T. Gao, J. Liu, C. M. Lieber, Nat. Nanotechnol. 2016, 11, 776. 1701269 (34 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de [339] C. Xie, J. Liu, T. M. Fu, X. Dai, W. Zhou, C. M. Lieber, Nat. Mater. 2015, 14, 1286. [340] L. Luan, X. Wei, Z. Zhao, J. J. Siegel, O. Potnis, C. A. Tuppen, S. Lin, S. Kazmi, R. A. Fowler, S. Holloway, Sci. Adv. 2017, 3, e1601966. [341] T. D. Y. Kozai, D. R. Kipke, J. Neurosci. Methods 2009, 184, 199. [342] R. L. Rennaker, S. Street, A. M. Ruyle, A. M. Sloan, J. Neurosci. Methods 2005, 142, 169. [343] J. P. Harris, A. E. Hess, S. J. Rowan, C. Weder, C. A. Zorman, D. J. Tyler, J. R. Capadona, J. Neural Eng. 2011, 8, 046010. [344] S. Singh, M.-C. Lo, V. Damodaran, H. Kaplan, J. Kohn, J. Zahn, D. Shreiber, Sensors 2016, 16, 330. [345] L. Karumbaiah, S. E. Norman, N. B. Rajan, S. Anand, T. Saxena, M. Betancur, R. Patkar, R. V. Bellamkonda, Biomaterials 2012, 33, 5983. [346] C. S. Bjornsson, S. J. Oh, Y. A. Al-Kofahi, Y. J. Lim, K. L. Smith, J. N. Turner, S. De, B. Roysam, W. Shain, S. J. Kim, J. Neural Eng. 2006, 3, 196. [347] B. Morrison, H. L. Cater, C. C.-B. Wang, F. C. Thomas, C. T. Hung, G. A. Ateshian, L. E. Sundstrom, Stapp Car Crash J. 2003, 47, 93. [348] B. Morrison, H. L. Cater, C. D. Benham, L. E. Sundstrom, J. Neurosci. Methods 2006, 150, 192. [349] E. Bar-Kochba, M. T. Scimone, J. B. Estrada, C. Franck, Sci. Rep. 2016, 6, 30550. [350] C. S. Bjornsson, S. J. Oh, Y. A. Al-Kofahi, Y. J. Lim, K. L. Smith, J. N. Turner, S. De, B. Roysam, W. Shain, S. J. Kim, J. Neural Eng. 2006, 3, 196. [351] D. J. Edell, V. V. Toi, V. M. McNeil, L. D. Clark, IEEE Trans. Biomed. Eng. 1992, 39, 635. [352] A. Andrei, M. Welkenhuysen, B. Nuttin, W. Eberle, J. Neural Eng. 2012, 9, 016005. [353] N. Haj Hosseini, R. Hoffmann, S. Kisban, T. Stieglitz, O. Paul, P. Ruther, in 2007 29th Ann. Int. Conf. of the IEEE Engineering in Medicine and Biology Society, Lyon, France 2007, 4711. [354] W. Jensen, K. Yoshida, U. G. Hofmann, IEEE Trans. Biomed. Eng. 2006, 53, 934. [355] A. A. Sharp, A. M. Ortega, D. Restrepo, D. Curran-Everett, K. Gall, IEEE Trans. Biomed. Eng. 2009, 56, 45. [356] D. H. Szarowski, M. D. Andersen, S. Retterer, A. J. Spence, M. Isaacson, H. G. Craighead, J. N. Turner, W. Shain, Brain Res. 2003, 983, 23. [357] B. S. Elkin, B. Morrison, J. Biomech. Eng. 2013, 135, 114507. [358] H. Lee, R. V. Bellamkonda, W. Sun, M. E. Levenston, J. Neural Eng. 2005, 2, 81. [359] D. B. MacManus, B. Pierrat, J. G. Murphy, M. D. Gilchrist, Sci. Rep. 2016, 6, 21569. [360] P. J. Rousche, R. A. Normann, Ann. Biomed. Eng. 1992, 20, 413. [361] J. Subbaroyan, D. C. Martin, D. R. Kipke, J. Neural Eng. 2005, 2, 103. [362] F. Casanova, P. R. Carney, M. Sarntinoranont, J. Neurosci. Methods 2014, 237, 79. [363] A. A. Sharp, H. V. Panchawagh, A. Ortega, R. Artale, S. RichardsonBurns, D. S. Finch, K. Gall, R. L. Mahajan, D. Restrepo, J. Neural Eng. 2006, 3, L23. [364] J. Thelin, H. Jorntell, E. Psouni, M. Garwicz, J. Schouenborg, N. Danielsen, C. E. Linsmeier, PLoS One 2011, 6, e16267. [365] J. P. Harris, A. E. Hess, S. J. Rowan, C. Weder, C. A. Zorman, D. J. Tyler, J. R. Capadona, J. Neural Eng. 2011, 8, 046010. [366] T. D. Y. Kozai, D. R. Kipke, J. Neurosci. Methods 2009, 184, 199. [367] T. L. Hanson, M. M. Maharbiz, P. N. SABES, Patent Application WO2016126340 A3, 2016. [368] C. H. Chen, S. C. Chuang, H. C. Su, W. L. Hsu, T. R. Yew, Y. C. Chang, S. R. Yeh, D. J. Yao, Lab Chip 2011, 11, 1647. Adv. Funct. Mater. 2017, 1701269 [369] D. Lewitus, K. L. Smith, W. Shain, J. Kohn, Acta Biomater. 2011, 7, 2483. [370] K. B. Neeves, C. T. Lo, C. P. Foley, W. M. Saltzman, W. L. Olbricht, J. Controlled Release 2006, 111, 252. [371] T. D. Y. Kozai, Z. Gugel, X. Li, P. J. Gilgunn, R. Khilwani, O. B. Ozdoganlar, G. K. Fedder, D. J. Weber, X. T. Cui, Biomaterials 2014, 35, 9255. [372] M. A. McClain, I. P. Clements, R. H. Shafer, R. V. Bellamkonda, M. C. LaPlaca, M. G. Allen, Biomed. Microdevices 2011, 13, 361. [373] P. R. Patel, K. Na, H. Zhang, T. D. Y. Kozai, N. A. Kotov, E. Yoon, C. A. Chestek, J. Neural Eng. 2015, 12, 046009. [374] M.-c. Lo, S. Wang, S. Singh, V. B. Damodaran, H. M. Kaplan, J. Kohn, D. I. Shreiber, J. D. Zahn, Biomed. Microdevices 2015, 17, 34. [375] T. D. Y. Kozai, Z. Gugel, X. Li, P. J. Gilgunn, R. Khilwani, O. B. Ozdoganlar, G. K. Fedder, D. J. Weber, X. T. Cui, Biomaterials 2014, 35, 9255. [376] M. L. Wolbarsht, E. F. MacNichol, H. G. Wagner, Science 1960, 132, 1309. [377] R. M. Dowben, J. E. Rose, Science 1953, 118, 22. [378] P. T. McCarthy, K. J. Otto, M. P. Rao, Biomed. Microdevices 2011, 13, 503. [379] Y. Maruyama, Y. Oishi, M. Kakimoto, Y. Imai, Macromolecules 1988, 21, 2305. [380] G. Srinivasan, D. H. Reneker, Polym. Int. 1995, 36, 195. [381] G. Guitchounts, J. E. Markowitz, W. A. Liberti, T. J. Gardner, J. Neural Eng. 2013, 10, 046016. [382] T. D. Y. Kozai, N. B. Langhals, P. R. Patel, X. Deng, H. Zhang, K. L. Smith, J. Lahann, N. A. Kotov, D. R. Kipke, Nat. Mater. 2012, 11, 1065. [383] B. S. Shim, J. Zhu, E. Jan, K. Critchley, N. A. Kotov, ACS Nano 2010, 4, 3725. [384] B. S. Shim, J. Zhu, E. Jan, K. Critchley, S. Ho, P. Podsiadlo, K. Sun, N. A. Kotov, ACS Nano 2009, 3, 1711. [385] B. S. Shim, W. Chen, C. Doty, C. Xu, N. A. Kotov, Nano Lett. 2008, 8, 4151. [386] E. V. Skorb, D. V. Andreeva, Polym. Chem. 2013, 4, 4834. [387] S. P. Lacour, D. Chan, S. Wagner, T. Li, Z. Suo, Appl. Phys. Lett. 2006, 88, 204103. [388] S. E. Saddow, A. Agarwal, Advances in Silicon Carbide Processing and Applications, Artech House, Norwood, USA 2004. [389] J. D. Reddy, A. Volinsky, C. Frewin, C. Locke, S. Saddow, presented at Mater. Res. Soc. Symp. Proc., Strasburg, USA 2008. [390] A. Rzany, C. Harder, M. Schaldach, Progress in Biomedical Research 2000, 5, 168. [391] P. Mavinakuli, S. Wei, Q. Wang, A. B. Karki, S. Dhage, Z. Wang, D. P. Young, Z. Guo, J. Phys. Chem. C 2010, 114, 3874. [392] J. C. Barrese, N. Rao, K. Paroo, C. Triebwasser, C. VargasIrwin, L. Franquemont, J. P. Donoghue, J. Neural Eng. 2013, 10, 066014. [393] A. J. Teo, A. Mishra, I. Park, Y.-J. Kim, W.-T. Park, Y.-J. Yoon, ACS Biomater. Sci. Eng. 2016, 2, 454. [394] N. Amanat, N. L. James, D. R. McKenzie, Med. Eng. Phys. 2010, 32, 690. [395] O. Erdogan, Indian Pacing Electrophysiol. J. 2002, 2, 74. [396] A. Nurmikko, D. Borton, M. Yin, in Neurobionics: The Biomedical Engineering of Neural Prostheses (Ed: R. K. Shepherd), John Wiley & Sons, Inc. Hoboken, USA 2016, Ch. 5, p. 123. [397] A. L. Benabid, S. Chabardes, J. Mitrofanis, P. Pollak, Lancet Neurol. 2009, 8, 67. [398] R. G. Hauser, W. T. Katsiyiannis, C. C. Gornick, A. K. Almquist, L. M. Kallinen, Europace 2010, 12, 395. [399] J. E. Ferguson, A. D. Redish, Expert Rev. Med. Devices 2011, 8, 427. 1701269 (35 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de [400] Y. Damestani, C. L. Reynolds, J. Szu, M. S. Hsu, Y. Kodera, D. K. Binder, B. H. Park, J. E. Garay, M. P. Rao, G. Aguilar, Nanomed.: Nanotechnol. Biol. Med. 2013, 9, 1135. [401] A. D. Degenhart, J. Eles, R. Dum, J. L. Mischel, I. Smalianchuk, B. Endler, R. C. Ashmore, E. C. Tyler-Kabara, N. G. Hatsopoulos, W. Wang, A. P. Batista, X. T. Cui, J. Neural Eng. 2016, 13, 046019. [402] E. Ryapolova-Webb, P. Afshar, S. Stanslaski, T. Denison, C. de Hemptinne, K. Bankiewicz, P. A. Starr, J. Neural Eng. 2014, 11, 016009. [403] C. Onal, H. Otsubo, T. Araki, S. Chitoku, A. Ochi, S. Weiss, I. Elliott, O. C. Snead 3rd, J. T. Rutka, W. Logan, J. Neurosurg. 2003, 98, 1017. [404] N. T. Markwardt, J. Stokol, R. L. RennakerII, J. Neurosci. Methods 2013, 214, 119. [405] M. Ravikumar, S. Sunil, J. Black, D. S. Barkauskas, A. Y. Haung, R. H. Miller, S. M. Selkirk, J. R. Capadona, Biomaterials 2014, 35, 8049. [406] Y.-T. Kim, R. W. Hitchcock, M. J. Bridge, P. A. Tresco, Biomaterials 2004, 25, 2229. [407] R. Biran, D. C. Martin, P. A. Tresco, J. Biomed. Mater. Res., Part A 2007, 82A, 169. [408] A. Abdo, V. Jayasinha, P. S. Spuhler, M. Unlu, M. Sahin, Conf. Proc. IEEE Eng. Med. Biol. Soc. 2009, 2009, 626. [409] P. G. Musial, S. N. Baker, G. L. Gerstein, E. A. King, J. G. Keating, J. Neurosci. Methods 2002, 115, 29. [410] K. J. Paralikar, C. R. Rao, R. S. Clement, J. Neurosci. Methods 2009, 181, 27. [411] A. Garg, A. L. Mohan, P. C. Garell, Parkinson’s Dis. 2010, 2010, 189371. [412] F. S. Fernández, M. A. A. Vega, A. A. Ramos, F. F. González, B. L. Aragoneses, Parkinson’s Dis. 2010, 2010, 409356. [413] S. Bjerknes, I. M. Skogseid, T. Sæhle, E. Dietrichs, M. Toft, PLoS One 2014, 9, e105288. [414] S. Suner, M. R. Fellows, C. Vargas-Irwin, G. K. Nakata, J. P. Donoghue, IEEE Trans. Neural Syst. Rehabil. Eng. 2005, 13, 524. [415] D. A. Borton, M. Yin, J. Aceros, A. Nurmikko, J. Neural Eng. 2013, 10, 026010. [416] M. Capogrosso, T. Milekovic, D. Borton, F. Wagner, E. M. Moraud, J.-B. Mignardot, N. Buse, J. Gandar, Q. Barraud, D. Xing, E. Rey, S. Duis, Y. Jianzhong, W. K. D. Ko, Q. Li, P. Detemple, T. Denison, S. Micera, E. Bezard, J. Bloch, G. Courtine, Nature 2016, 539, 284. [417] R. R. Harrison, R. J. Kier, S. Kim, L. Rieth, D. J. Warren, N. M. Ledbetter, G. A. Clark, F. Solzbacher, C. A. Chestek, V. Gilja, P. Nuyujukian, S. I. Ryu, K. V. Shenoy, in 2008 IEEE Biomedical Circuits and Systems Conf., Baltimore, USA 2008, 125. [418] S. Kim, R. Bhandari, M. Klein, S. Negi, L. Rieth, P. Tathireddy, M. Toepper, H. Oppermann, F. Solzbacher, Biomed. Microdevices 2009, 11, 453. [419] L. Rafflenbeul, R. Werthschützky, A. Gail, Procedia Eng. 2012, 47, 1097. [420] A. Kiourti, C. W. L. Lee, J. Chae, J. L. Volakis, IEEE Trans. Biomed. Eng. 2016, 63, 131. [421] H. N. Schwerdt, W. Xu, S. Shekhar, A. Abbaspour-Tamijani, B. C. Towe, F. A. Miranda, J. Chae, J. Microelectromech. Syst. 2011, 20, 1119. [422] T. J. Oxley, N. L. Opie, S. E. John, G. S. Rind, S. M. Ronayne, T. L. Wheeler, J. W. Judy, A. J. McDonald, A. Dornom, T. J. H. Lovell, C. Steward, D. J. Garrett, B. A. Moffat, E. H. Lui, N. Yassi, B. C. V. Campbell, Y. T. Wong, K. E. Fox, E. S. Nurse, I. E. Bennett, S. H. Bauquier, K. A. Liyanage, N. R. van der Nagel, P. Perucca, A. Ahnood, K. P. Gill, B. Yan, L. Churilov, C. R. French, P. M. Desmond, M. K. Horne, L. Kiers, S. Prawer, S. M. Davis, A. N. Burkitt, P. J. Mitchell, D. B. Grayden, C. N. May, T. J. O’Brien, Nat. Biotechnol. 2016, 34, 320. Adv. Funct. Mater. 2017, 1701269 [423] G. Ciofani, S. Danti, D. D’Alessandro, L. Ricotti, S. Moscato, G. Bertoni, A. Falqui, S. Berrettini, M. Petrini, V. Mattoli, A. Menciassi, ACS Nano 2010, 4, 6267. [424] P. J. Larson, B. C. Towe, presented at 5th International IEEE/EMBS Conference on Neural Engineering (NER), Cancun, Mexico April– May 2011. [425] R. R. Harrison, P. T. Watkins, R. J. Kier, R. O. Lovejoy, D. J. Black, B. Greger, F. Solzbacher, IEEE J. Solid-State Circuits 2007, 42, 123. [426] T. V. F. Abaya, M. Diwekar, S. Blair, P. Tathireddy, L. Rieth, G. A. Clark, F. Solzbacher, Biomed. Opt. Express 2012, 3, 2200. [427] M. G. Shapiro, K. Homma, S. Villarreal, C.-P. Richter, F. Bezanilla, Nat. Commun. 2012, 3, 736. [428] A. Abdo, M. Sahin, IEEE Trans. Biomed. Circuits Syst. 2011, 2011, 1. [429] A. Abdo, M. Sahin, D. S. Freedman, E. Cevik, P. S. Spuhler, M. S. Unlu, J. Neural Eng. 2011, 8, 056012. [430] M. Sahin, S. S. Ur-Rahman, IEEE Trans. Neural Syst. Rehabil. Eng. 2007, 15, 227. [431] E. Ç. Seymour, D. S. Freedman, M. Gökkavas, E. Özbay, M. Sahin, M. S. Ünlü, Front. Neuroeng. 2014, 7, 00005. [432] G. Romero, M. G. Christiansen, L. S. Barbosa, F. Garcia, P. Anikeeva, Adv. Funct. Mater. 2016, 26, 6471. [433] R. Chen, G. Romero, M. G. Christiansen, A. Mohr, P. Anikeeva, Science 2015, 347, 1477. [434] Y. Wang, L. Guo, Front. Neurosci. 2016, 10, 69. [435] F. Lavoie-Cardinal, C. Salesse, É. Bergeron, M. Meunier, P. De Koninck, Sci. Rep. 2016, 6, 20619. [436] A. V. Chervyakov, A. Y. Chernyavsky, D. O. Sinitsyn, M. A. Piradov, Front. Hum. Neurosci. 2015, 9, 303. [437] M. Hallett, Nature 2000, 406, 147. [438] X. Luo, X. T. Cui, Electrochem. Commun. 2009, 11, 1956. [439] S. Jiang, Y. Sun, X. Cui, X. Huang, Y. He, S. Ji, W. Shi, D. Ge, Synth. Met. 2013, 163, 19. [440] M. S. Freedman, X. T. Cui, Phys. Chem. Commun. 2014, 1, 15. [441] Y. Cho, R. Shi, A. Ivanisevic, R. Ben Borgens, Nanotechnology 2009, 20, 275102. [442] A. B. Taylor, A. M. Siddiquee, J. W. Chon, ACS Nano 2014, 8, 12071. [443] J. Yong, K. Needham, W. G. Brown, B. A. Nayagam, S. L. McArthur, A. Yu, P. R. Stoddart, Adv. Healthcare Mater. 2014, 3, 1862. [444] T. D. Y. Kozai, A. L. Vazquez, C. L. Weaver, J. Neural Eng. 2012, 9, 066001. [445] E. K. Purcell, J. P. Seymour, S. Yandamuri, D. R. Kipke, J. Neural Eng. 2009, 6, 026005. [446] A. W. Bridges, A. J. García, J. Diabetes Sci. Technol. 2008, 2, 984. [447] R. Biran, M. D. Noble, P. A. Tresco, J. Biomed. Mater. Res. 1999, 46, 150. [448] B. K. Leung, R. Biran, C. J. Underwood, P. A. Tresco, Biomaterials 2008, 29, 3289. [449] M. E. Manwaring, R. Biran, P. A. Tresco, Biomaterials 2001, 22, 3155. [450] H. Zhang, M. Chiao, J. Med. Biol. Eng. 2015, 35, 143. [451] J. L. Harding, M. M. Reynolds, Trends Biotechnol. 2014, 32, 140. [452] E. Azemi, W. R. Stauffer, M. S. Gostock, C. F. Lagenaur, X. T. Cui, Acta Biomater. 2008, 4, 1208. [453] J. Y. Lee, Z. Z. Khaing, J. J. Siegel, C. E. Schmidt, RSC Adv. 2015, 5, 39228. [454] H. Zhao, B. Zhu, S.-C. Luo, H.-A. Lin, A. Nakao, Y. Yamashita, H.-h. Yu, ACS Appl. Mater. Interfaces 2013, 5, 4536. [455] Y. Hu, G. Yang, B. Liang, L. Fang, G. Ma, Q. Zhu, S. Chen, X. Ye, Acta Biomater. 2015, 13, 142. [456] E. Patrick, M. E. Orazem, J. C. Sanchez, T. Nishida, J. Neurosci. Methods 2011, 198, 158. [457] A. Prasad, Q.-S. Xue, R. Dieme, V. Sankar, R. C. Mayrand, T. Nishida, W. J. Streit, J. C. Sanchez, Front. Neuroeng. 2014, 7, 2. [458] J. C. Sanchez, N. Alba, T. Nishida, C. Batich, P. R. Carney, IEEE Trans. Neural Syst. Rehabil. Eng. 2006, 14, 217. 1701269 (36 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de [459] A. Sridharan, J. K. Nguyen, J. R. Capadona, J. Muthuswamy, J. Neural Eng. 2015, 12, 036002. [460] G. Lind, C. E. Linsmeier, J. Schouenborg, Sci. Rep. 2013, 3, 11830. [461] A. W. Bridges, R. E. Whitmire, N. Singh, K. L. Templeman, J. E. Babensee, L. A. Lyon, A. J. García, J. Biomed. Mater. Res., Part A 2010, 94A, 252. [462] Y. Lu, D. Wang, T. Li, X. Zhao, Y. Cao, H. Yang, Y. Y. Duan, Biomaterials 2009, 30, 4143. [463] M. Li, H.-H. Zhou, T. Li, C.-Y. Li, Z.-Y. Xia, Y. Y. Duan, Neural Regener. Res. 2015, 10, 2048. [464] A. K. H. Achyuta, V. S. Polikov, A. J. White, H. G. P. Lewis, S. K. Murthy, Macromol. Biosci. 2010, 10, 872. [465] D.-H. Kim, M. Abidian, D. C. Martin, J. Biomed. Mater. Res. 2004, 71A, 577. [466] Y. W. Fan, F. Z. Cui, S. P. Hou, Q. Y. Xu, L. N. Chen, I.-S. Lee, J. Neurosci. Methods 2002, 120, 17. [467] K. A. Moxon, N. M. Kalkhoran, M. Markert, M. A. Sambito, J. L. McKenzie, J. T. Webster, IEEE Trans. Biomed. Eng. 2004, 51, 881. [468] K. A. Moxon, S. Hallman, A. Aslani, N. M. Kalkhoran, P. I. Lelkes, J. Biomater. Sci., Polym. Ed. 2007, 18, 1263. [469] S. Chen, M. G. Allen, MRS Bull. 2012, 37, 606. [470] C. A. R. Chapman, L. Wang, H. Chen, J. Garrison, P. J. Lein, E. Seker, Adv. Funct. Mater. 2017, 27, 1604631. [471] X. Luo, C. L. Weaver, D. D. Zhou, R. Greenberg, X. T. Cui, Biomaterials 2011, 32, 5551. [472] D. E. Labaye, C. Jérôme, V. Geskin, P. Louette, R. Lazzaroni, L. Martinot, R. Jérôme, Langmuir 2002, 18, 5222. [473] W. R. Stauffer, X. T. Cui, Biomaterials 2006, 27, 2405. [474] C. L. Kolarcik, D. Bourbeau, E. Azemi, E. Rost, L. Zhang, C. F. Lagenaur, D. J. Weber, X. T. Cui, Acta Biomater. 2012, 8, 3561. [475] J. R. Eles, A. L. Vazquez, N. R. Snyder, C. F. Lagenaur, M. C. Murphy, T. D. Y. Kozai, X. T. Cui, Biomaterials 2017, 113, 279. [476] K. Webb, E. Budko, T. J. Neuberger, S. Chen, M. Schachner, P. A. Tresco, Biomaterials 2001, 22, 1017. [477] E. Azemi, C. F. Lagenaur, X. T. Cui, Biomaterials 2011, 32, 681. [478] A. H. Taub, R. Hogri, A. Magal, M. Mintz, Y. Shacham-Diamand, J. Biomed. Mater. Res., Part A 2012, 100A, 1854. [479] K. A. Potter-Baker, J. K. Nguyen, K. M. Kovach, M. M. Gitomer, T. W. Srail, W. G. Stewart, J. L. Skousen, J. R. Capadona, J. Mater. Chem. B 2014, 2, 2248. [480] J. Chang, Y. Jallouli, M. Kroubi, X. B. Yuan, W. Feng, C. S. Kang, P. Y. Pu, D. Betbeder, Int. J. Pharm. 2009, 379, 285. [481] T. D. Y. Kozai, A. S. Jaquins-gerstl, A. L. Vazquez, A. C. Michael, X. T. Cui, Biomaterials 2016, 87, 157. [482] A. Mercanzini, S. T. Reddy, D. Velluto, P. Colin, A. Maillard, J.-C. Bensadoun, J. A. Hubbell, P. Renaud, J. Controlled Release 2010, 145, 196. [483] W. Shain, L. Spataro, J. Dilgen, K. Haverstick, S. Retterer, M. Isaacson, M. Saltzman, J. N. Turner, IEEE Trans. Neural Syst. Rehabil. Eng. 2003, 11, 186. [484] L. Spataro, J. Dilgen, S. Retterer, A. J. Spence, M. Isaacson, J. N. Turner, W. Shain, Exp. Neurol. 2005, 194, 289. [485] Y. Zhong, R. V. Bellamkonda, Brain Res. 2007, 1148, 15. [486] R. Wadhwa, C. F. Lagenaur, X. T. Cui, J. Controlled Release 2006, 110, 531. [487] D.-H. Kim, D. C. Martin, Biomaterials 2006, 27, 3031. [488] C. L. Weaver, J. M. LaRosa, X. Luo, X. T. Cui, ACS Nano 2014, 8, 1834. [489] H. Mond, K. Stokes, J. Helland, L. Grigg, P. Kertes, B. Pate, D. Hunt, Pacing Clin. Electrophysiol. 1988, 11, 214. [490] G. H. Crossley, J. A. Brinker, D. Reynolds, W. Spencer, W. B. Johnson, H. Hurd, L. Tonder, M. Zmijewski, Circulation 1995, 92, 2935. Adv. Funct. Mater. 2017, 1701269 [491] V. Kutyifa, E. Zima, L. Molnar, C. Kuehne, S. Theiss, G. Herrmann, L. Geller, B. Merkely, Cardiol. J. 2013, 20, 431. [492] J. K. Nguyen, M. Jorfi, K. L. Buchanan, D. J. Park, E. J. Foster, D. J. Tyler, S. J. Rowan, C. Weder, J. R. Capadona, Acta Biomater. 2016, 29, 81. [493] K. A. Potter-Baker, W. G. Stewart, W. H. Tomaszewski, C. T. Wong, W. D. Meador, N. P. Ziats, J. R. Capadona, J. Neural Eng. 2015, 12, 046002. [494] Y. Zhong, R. V. Bellamkonda, J. Controlled Release 2005, 106, 309. [495] N. Han, S. S. Rao, J. Johnson, K. S. Parikh, P. A. Bradley, J. J. Lannutti, J. O. Winter, Front. Neuroeng. 2011, 4, 2. [496] S. J. Jhaveri, M. R. Hynd, N. Dowell-Mesfin, J. N. Turner, W. Shain, C. K. Ober, Biomacromolecules 2009, 10, 174. [497] A. J. Evans, B. C. Thompson, G. G. Wallace, R. Millard, S. J. O’Leary, G. M. Clark, R. K. Shepherd, R. T. Richardson, J. Biomed. Mater. Res., Part A 2009, 91A, 241. [498] P. R. Kennedy, S. S. Mirra, R. A. Bakay, Neurosci. Lett. 1992, 142, 89. [499] P. Kennedy, D. Andreasen, J. Bartels, P. Ehirim, H. Mao, M. Velliste, T. Wichmann, J. Wright, Prog. Brain Res. 2011, 194, 1. [500] J. Bartels, D. Andreasen, P. Ehirim, H. Mao, S. Seibert, E. J. Wright, P. Kennedy, J. Neurosci. Methods 2008, 174, 168. [501] E. Azemi, G. T. Gobbel, X. T. Cui, J. Neurosurg. 2010, 113, 673. [502] X. Cui, J. F. Hetke, J. A. Wiler, D. J. Anderson, D. C. Martin, Sens. Actuators, A 2001, 93, 8. [503] C. L. Weaver, H. Li, X. Luo, X. T. Cui, J. Mater. Chem. B 2014, 2, 5209. [504] N. A. Alba, Z. J. Du, K. A. Catt, T. D. Y. Kozai, X. T. Cui, Biosensors 2015, 5, 618. [505] P. Fattahi, G. Yang, G. Kim, M. R. Abidian, Adv. Mater. 2014, 26, 1846. [506] C. L. Kolarcik, K. Catt, E. Rost, I. N. Albrecht, D. Bourbeau, Z. Du, T. D. Y. Kozai, X. Luo, D. J. Weber, X. T. Cui, J. Neural Eng. 2015, 12, 016008. [507] C. L. Weaver, X. T. Cui, Adv. Healthcare Mater. 2015, 4, 1408. [508] K. Matsumoto, C. Sato, Y. Naka, R. Whitby, N. Shimizu, Nanotechnology 2010, 21, 115101. [509] T. Moser, C. Celma, A. Lebert, E. Charrault, R. Brooke, P. J. Murphy, G. Browne, R. Young, T. Higgs, D. Evans, ACS Appl. Mater. Interfaces 2016, 8, 974. [510] V. Gradinaru, F. Zhang, C. Ramakrishnan, J. Mattis, R. Prakash, I. Diester, I. Goshen, K. R. Thompson, K. Deisseroth, Cell 2010, 141, 154. [511] B. Y. Chow, E. S. Boyden, Sci. Transl. Med. 2013, 5, 177ps5. [512] J. C. Williams, T. Denison, Sci. Transl. Med. 2013, 5, 177ps6. [513] M. C. Hudson, F. L. Thiel, Appl. Opt. 1974, 13, 2540. [514] F. Wu, E. Stark, P.-C. Ku, K. D. Wise, G. Buzsáki, E. Yoon, Neuron 2015, 88, 1136. [515] A. Steude, M. Jahnel, M. Thomschke, M. Schober, M. C. Gather, Adv. Mater. 2015, 27, 7657. [516] A. Steude, E. C. Witts, G. B. Miles, M. C. Gather, Sci. Adv. 2016, 2, e1600061. [517] Y. Son, H. J. Lee, J. Kim, H. Shin, N. Choi, C. J. Lee, E. S. Yoon, E. Yoon, K. D. Wise, T. G. Kim, I. J. Cho, Sci. Rep. 2015, 5, 15466. [518] D. H. Kim, J. Viventi, J. J. Amsden, J. Xiao, L. Vigeland, Y. S. Kim, J. A. Blanco, B. Panilaitis, E. S. Frechette, D. Contreras, D. L. Kaplan, F. G. Omenetto, Y. Huang, K. C. Hwang, M. R. Zakin, B. Litt, J. A. Rogers, Nat. Mater. 2010, 9, 511. [519] C. Choi, L. Lin, Y. Liu, J. Choi, L. Wang, D. Haas, J. Magera, R. T. Chen, J. Lightwave Technol. 2004, 22, 2168. [520] M. Schwaerzle, F. Pothof, O. Paul, P. Ruther, presented at 18th International Conference on Solid-State Sensors, Actuators, and Microsystems (Transducers), Anchorage, USA June 2015. 1701269 (37 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.advancedsciencenews.com www.afm-journal.de [521] J. Wu, R. T. Pike, C. Wong, N. P. Kim, M. H. Tanielian, IEEE Trans. Adv. Pack. 2000, 23, 721. [522] G. Jiang, D. D. Zhou, in Implantable Neural Prostheses 2, Springer, New York, USA 2009, 27. [523] J. S. Lewis, M. S. Weaver, IEEE J. Sel. Top. Quantum Electron. 2004, 10, 45. [524] E. G. Jeong, Y. C. Han, H.-G. Im, B.-S. Bae, K. C. Choi, Org. Electron. 2016, 33, 150. [525] Y.-H. Joung, Int. Neurourol. J. 2013, 17, 98. [526] A. Borreman, S. Musa, A. Kok, M. Diemeer, A. Driessen, IEEE/ LEOS Benelux Chapter 2002 Annual Symposium, Amsterdam, The Netherlands December 2002. [527] Y. Son, H. J. Lee, J. Kim, C. J. Lee, E.-S. Yoon, T. G. Kim, I.-J. Cho, presented at 28th IEEE International Conference on Micro Electro Mechanical Systems (MEMS), Estoril, Portugal January 2015. [528] T. Han, S. Madden, M. Zhang, R. Charters, B. Luther-Davies, presented at Conference on Optoelectronic and Microelectronic Materials and Devices (COMMAD), Sydney, Australia July–August 2008. [529] M. Im, I.-J. Cho, F. Wu, K. D. Wise, E. Yoon, presented at IEEE 24th International Conference on Micro Electro Mechanical Systems (MEMS), Cancun, Mexico January 2011. [530] K. Kampasi, E. Stark, J. Seymour, K. Na, H. G. Winful, G. Buzsáki, K. D. Wise, E. Yoon, Sci. Rep. 2016, 6, 30961. [531] N. Ioannides, E. Chunga, A. Bachmatiuk, I. Gonzalez-Martinez, B. Trzebicka, D. Adebimpe, D. Kalymnios, M. Rümmeli, Mater. Res. Express 2014, 1, 032002. [532] Y. Yang, G. Turnbull, I. Samuel, Appl. Phys. Lett. 2008, 92, 150. [533] J. M. Stujenske, T. Spellman, J. A. Gordon, Cell Rep. 2015, 12, 525. [534] F.-Y. B. Chen, D. M. Budgett, Y. Sun, S. Malpas, D. McCormick, P. S. Freestone, IEEE Trans. Biomed. Circuits Syst. 2016, 11, 28. [535] R. L. Rungta, B.-F. Osmanski, D. Boido, M. Tanter, S. Charpak, Nat. Commun. 2017, 8, 14191. [536] I. N. Christie, J. A. Wells, P. Southern, N. Marina, S. Kasparov, A. V. Gourine, M. F. Lythgoe, Neuroimage 2013, 66, 634. [537] M. Zochowski, M. Wachowiak, C. X. Falk, L. B. Cohen, Y. W. Lam, S. Antic, D. Zecevic, Biol. Bull. 2000, 198, 1. [538] G. W. Gross, B. K. Rhoades, D. L. Reust, F. U. Schwalm, J. Neurosci. Methods 1993, 50, 131. [539] A. Khurram, J. P. Seymour, presented at Engineering in Medicine and Biology Society (EMBC), 35th Annual International Conference of the IEEE, Osaka, Japan July 2013. [540] D. B. Fraser, H. D. Cook, J. Electrochem. Soc. 1972, 119, 1368. [541] US FDA Second Sight Argus II Retinal Prosthesis HDE H110002 System Summary of Safety and Probable Benefit, http://www. accessdata.fda.gov/cdrh_docs/pdf11/H110002b.pdf (accessed: February 2017). [542] US FDA CVRx Barostim neo Legacy System HDE H130007, www. accessdata.fda.gov/cdrh_docs/pdf13/H130007a.pdf (accessed: February 2017). [543] US FDA Inspire Upper Airway Stimulation PMA P130008: FDA Summary of Safety and Effectiveness Data, http://www.accessdata.fda.gov/cdrh_docs/pdf13/P130008b.pdf (accessed: February 2017). Adv. Funct. Mater. 2017, 1701269 [544] US FDA Enteromedics Maestro Rechargeable System PMA P130019, www.accessdata.fda.gov/cdrh_docs/pdf13/P130019a.pdf (accessed: February 2017). [545] L. Y. Liu, H. Zheng, H. L. Xiao, Z. J. She, S. M. Zhao, Z. L. Chen, G. M. Zhou, Neurosci. Lett. 2008, 434, 155. [546] US FDA Stimwave Freedom Spinal Cord Stimulator (SCS) System K162161, http://www.accessdata.fda.gov/cdrh_docs/ pdf16/K162161.pdf (accessed: February 2017). [547] K. Yoshida, K. Horch, IEEE Trans. Biomed. Eng. 1993, 40, 492. [548] D. K. Leventhal, D. M. Durand, Ann. Biomed. Eng. 2003, 31, 643. [549] M. B. Christensen, S. M. Pearce, N. M. Ledbetter, D. J. Warren, G. A. Clark, P. A. Tresco, Acta Biomater. 2014, 10, 4650. [550] A. Branner, R. A. Normann, Brain Res. Bull. 2000, 51, 293. [551] T. W. Bouldin, T. S. Earnhardt, N. D. Goines, J. Neuropathol. Exp. Neurol. 1991, 50, 719. [552] D. J. Tyler, D. M. Durand, Ann. Biomed. Eng. 2003, 31, 633. [553] T. A. Kung, N. B. Langhals, D. C. Martin, P. J. Johnson, P. S. Cederna, M. G. Urbanchek, Plast. Reconstr. Surg. 2014, 133, 1380. [554] M. G. Urbanchek, T. A. Kung, C. M. Frost, D. C. Martin, L. M. Larkin, A. Wollstein, P. S. Cederna, BioMed. Res. Int. 2016, 2016, 5736730. [555] T. Cameron, J. Neurosurg. 2004, 100, 254. [556] A. Dietz, M. Wennstrom, A. Lehtimaki, H. Lopponen, H. Valtonen, Eur. Arch. Oto-Rhino-Laryngol. 2016, 273, 1411. [557] T. J. Schwedt, D. W. Dodick, J. Hentz, T. L. Trentman, R. S. Zimmerman, Cephalalgia 2007, 27, 153. [558] D. W. Dodick, S. D. Silberstein, K. L. Reed, T. R. Deer, K. V. Slavin, B. Huh, A. D. Sharan, S. Narouze, A. Y. Mogilner, T. L. Trentman, J. Ordia, J. Vaisman, J. Goldstein, N. Mekhail, Cephalalgia 2015, 35, 344. [559] S. Falowski, D. Wang, A. Sabesan, A. Sharan, Neuromodulation 2010, 13, 121. [560] E. Ben-Menachem, Lancet Neurol. 2002, 1, 477. [561] W. M. Grill, J. T. Mortimer, IEEE Trans. Rehabil. Eng. 1998, 6, 364. [562] G. G. Naples, J. T. Mortimer, A. Scheiner, J. D. Sweeney, IEEE Trans. Biomed. Eng. 1988, 35, 905. [563] J. Espinosa, M. T. Aiello, D. K. Naritoku, Surg. Neurol. 1999, 51, 659. [564] R. J. Mobbs, P. Blum, R. Rossato, J. Clin. Neurosci. 2003, 10, 476. [565] V. M. Renard, R. B. North, J. Neurosurg. Spine 2006, 4, 300. [566] G. Geissinger, J. H. Neal, Surg. Neurol. 2007, 68, 454. [567] A. P. Burdick, M. S. Okun, I. U. Haq, H. E. Ward, F. Bova, C. E. Jacobson, D. Bowers, P. Zeilman, K. D. Foote, Stereotact Funct. Neurosurg. 2010, 88, 353. [568] G. L. Ellis, Am. J. Emerg. Med. 1990, 8, 48. [569] E. O. Udo, N. P. Zuithoff, N. M. van Hemel, C. C. de Cock, T. Hendriks, P. A. Doevendans, K. G. Moons, Heart Rhythm 2012, 9, 728. [570] US FDA RST-001 Phase I/II Trial for Retinitis Pigmentosa, https:// clinicaltrials.gov/ct2/show/NCT02556736 (accessed: March 2017). [571] N. G. Ghazi, E. B. Abboud, S. R. Nowilaty, H. Alkuraya, A. Alhommadi, H. Cai, R. Hou, W.-T. Deng, S. L. Boye, A. Almaghamsi, Hum. Genet. 2016, 135, 327. 1701269 (38 of 38) © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim