Infrared Physics & Technology 45 (2004) 315–321
www.elsevier.com/locate/infrared
Normalisation procedure in thermal wave approach
of thermal diffusivity measurement of solids using
pyroelectric sensor
B.Z. Azmi *, H.S. Liaw, W.M.M. Yunus, M. Hashim,
M.M. Moksin, W.M.D.W. Yusoff
Photoacoustic Laboratory, Department of Physics, Universiti Putra Malaysia, 43400 Serdang, Selangor Darul Ehsan, Malaysia
Received 7 May 2003
Abstract
The theory of thermal wave interference in generation of photoacoustic signal suggested by Bennett and Patty [Appl.
Opt. 21 (1982) 49] was applied to measure thermal diffusivity of optically opaque sample using photopyroelectric
technique. To determine thermal diffusivity with precision, one needs to work with frequencies high enough in order to
be in a thermally thick regime. However, this condition cannot always easily be fulfilled due to the great attenuation of
the amplitude signal at high frequency resulting in decrease in signal-to-noise ratio. In this work, we proposed thermal
diffusivity measurements to be carried out with the thermal wave probing starting from the sample–pyroelectric
interface towards the thermally thick region of the sample. Reflection of thermal wave in the pyroelectric coatings,
which was ignored in previous models, was considered in the generation of photopyroelectric signal. Normalisation
procedure was used to eliminate the number of media parameters of photopyroelectric cell that otherwise need to be
known before one can determine the thermal diffusivity of the sample. With the appropriate sample–pyroelectric
detector dimension, the thermal diffusivity of any solid sample is readily being determined. The method was experimentally tested for aluminum, copper, and nickel, and the values obtained were close to the literature values.
2004 Elsevier B.V. All rights reserved.
PACS: 77.70.+a; 66.10.Cb; 66.70.+f; 78.20.Nv
Keywords: Photothermal; Photopyroelectric; Thermal wave; Thermal diffusivity; PVDF
1. Introduction
*
Corresponding author. Tel.: +60-3894-66650; fax: +603894-54454.
E-mail address: azmizak@fsas.upm.edu.my (B.Z. Azmi).
Recently, there has been considerable interest in
polyvinylidene diflouride (PVDF) polymer since
the discovery of its strong pyroelectric (PE) effect
coefficient by Bergman et al. in 1971 [1]. The low
cost, light weight, flexible, insensitive to acoustics
and mechanical noise, and good electrical prop-
1350-4495/$ - see front matter 2004 Elsevier B.V. All rights reserved.
doi:10.1016/j.infrared.2004.01.002
316
B.Z. Azmi et al. / Infrared Physics & Technology 45 (2004) 315–321
erties of PVDF film make the film an excellent
choice for signal detection in photopyroelectric
(PPE) technique.
In PPE technique, the test sample is usually
deposited on or placed in good thermal contact
with the PE sensor and is exposed to intensity
modulated light. The theory of this configuration
in PPE signal generation has been presented by
Mandelis et al. [2] and thermal diffusivity measurement of optically opaque sample has been
carried out by Marinelli et al. [3]. It is inevitable to
work at high frequencies or to operate in a thermally thick regime as being predicted by the theory
[2]. However, this condition cannot always easily
be fulfilled due to the great attenuation of the
amplitude signal at high frequency resulting in
decrease in the signal-to-noise ratio [4]. On the
other hand, in the case of thermally very thin
sample, the PPE signal is independent of sample
thermal property [2]. Therefore, we examine the
suitability of a new approach in thermal diffusivity
measurement that has capability to probe from the
near sample–PE interface and continuously
through the thermally thick region of the sample.
In this paper, the theory of thermal wave
interference in photoacoustic (PA) signal generation suggested by Bennett and Patty [5] has been
adopted and applied to thermal diffusivity measurement of optically opaque solids using a standard PPE configuration. In the PPE signal
generation, we have considered the transmitted
thermal waves from the PVDF film coatings into
the film. The thermal diffusivity value of Al, Cu,
and Ni samples are then obtained by fitting the
theory to the experimental data.
mitted to the PE detector, instead of the transmitted terms that transmitted through the gas to
the detector in PA technique. If the test sample is
highly opaque, in which the optical absorption
length is much smaller than sample thickness, this
approach is further simplified by considering only
the heat source generated on the sample surface
[3]. Then, the resulting thermal wave is independent of the optical properties of the sample.
A one-dimensional configuration of PPE cell is
shown in Fig. 1. To understand the route and effect of the thermal wave, consider now the thermal
wave generated on sample surface and propagates
to the left. It will be partially transmitted and reflected upon striking the sample–coating interface,
A. Generally for a thermal wave propagating in
medium 1 towards medium 2, the thermal wave
reflection coefficient R12 and the thermal wave
transmission coefficient T12 at the interface are
respectively given by [5,6]
R12 ¼
1 b12
;
1 þ b12
T12 ¼
2
;
1 þ b12
ð1Þ
where b12 is the ratio of the thermal diffusivity of
the medium 2 to medium 1. The wave partially
reflected back into sample Rsc A ers Ls , where A ¼
Qo =2ks rs , Qo is the source intensity, ks is thermal
conductivity of the sample, rs ¼ ð1 þ iÞ=ls , ls ¼
1=2
ðas =pf Þ is the thermal diffusion length of sample
at light modulation frequency f , Ls is the sample
thickness, and as is the thermal diffusivity of the
sample. Thermal diffusivity is defined by as ¼ ks =
qs cs , where qs is the density and cs is the specific
heat. This reflected thermal wave will eventually
2. Theory
The thermal wave generated within the sample
will be partially reflected and transmitted upon
striking at the interface of two media. The effect of
infinite multiple reflections of the thermal wave
within the sample was extensively discussed by
Bennett and Patty in the PA signal generation [5].
In this work, a similar approach was used to obtain the complex PPE signal by adding all the
transmitted terms of thermal waves that trans-
Fig. 1. Thermal wave presence in each region of PPE cell. The
thermal waves are partially reflected and transmitted upon
striking the boundaries. Subscript g, s, c, p, and b stand for gas,
sample, coating, PE detector, and backing, respectively.
B.Z. Azmi et al. / Infrared Physics & Technology 45 (2004) 315–321
died off in short thermal diffusion length operations. However, the transmitted wave at A,
hs ¼ Tsc A ers Ls ;
ð2Þ
will travel across the coating [5,6]. Since the coating thickness is thermally much more thinner
compare to that of PVDF film, there will be a
series of partially reflected waves hs Rcp erc Lc ,
hs R2cp Rcs e3rc Lc , hs R3cp R2cs e5rc Lc , etc. and of partially
transmitted waves hs Tcp erc Lc , hs Rcp Rcs Tcp e3rc Lc ,
hs R2cp R2cs Tcp e5rc Lc , etc. at the coating–PVDF interface, x ¼ 0. The resultant of the transmitted wave
in PVDF emerging from the right coating is given
by,
hpr ¼ hs Tcp
erc Lc
;
1 Rcp Rcs e2rc Lc
ð3Þ
and it contributes to the temperature distribution
in PVDF film.
Then, if we stepped backward and trace the first
transmitted wave hs Tcp erc Lc from point B after
traveling through a distance Lp in the PVDF film,
see Fig. 1, it will be partially transmitted into left
coating hs Tcp erc Lc Tpc erp Lp and partially reflected
back into PVDF hs Tcp erc Lc Rpc erp Lp at the PVDF–
coating interface, C. Similarly at the interface
points D and C0 and so on, there will be many
partially transmitted and reflected waves. As a result of these, the resultant wave emerges from the
left coating at x ¼ Lp into PVDF film is given by
Tpc Tcp Rcb e2rc Lc
1
rc Lc rp Lp
hpl ¼ hs Tcp e
:
e
Rpc þ
1 Rcb Rcp e2rc Lc
ð4Þ
Similarly as the first transmitted wave, the transmitted waves from the right that are of factors
Rcp Rcs e2rc Lc , R2cp R2cs e4rc Lc and so on, of the first
wave hs Tcp erc Lc will produce multiple transmissions from the left coating. The resultant wave hpl
in PVDF is therefore given by
hpl ¼ hs Tcp e
rc Lc
e
Tpc Tcp Rcb e2rc Lc
Rpc þ
1 Rcb Rcp e2rc Lc
:
2rc Lc
rp Lp
1
1 Rcp Rcs e
ð5Þ
317
This amplitude of thermal wave is equivalent to
the amplitude of thermal wave transmitted to the
gas treated by Bennett and Patty [5].
In the analysis of thermal wave propagation in a
sample, Opsal and Rosencwaig [7] suggested a
multi-layer model that provides expressions for
temperature at the surface of each layer. The model
is applicable both to the measurement of surface
temperature variations, as in the cases of gasmicrophone photoacoustics and photothermal
detection, and to the measurement of subsurface
thermoelastic response, as in the cases of piezoelectric and thermal-wave experiments. Minamide
et al. [8] then applied the model in their 3-layer PPE
cell, which was the layer of sample, air gap and
PVDF film, to obtain the surface temperature of
the film. The averaged temperature hp in the film
with no reflection in PVDF–coating interface is
given by
Z Lp
1
hc erp x dx;
ð6Þ
hp ðxÞ ¼
Lp 0
where hc is the surface temperature of the PVDF
film. In our case, by considering multiple thermal
wave transmission from the left coating of
expression stated in Eq. (5), as well as that from
the right, Eq. (3), the averaged temperature hp in
the PVDF film is given by
Z Lp
1
hp ðxÞ ¼
½hpr erp x þ hpl erp Lp þrp x dx:
ð7Þ
Lp 0
The average PE voltage produced by the PVDF
film is given by [2]
V ðxÞ ¼
pLp hp
;
ee0
ð8Þ
where p is the PE coefficient, e is the dielectric
constant of PE detector, and e0 is the permittivity
constant of vacuum. Substituting the result of (7)
into Eq. (8) leads to
V ðxÞ ¼
Qo p
Tcp Tsc eðrs Ls þrc Lc Þ
ð1 erp Lp Þ
2eeo ks rs rp 1 Rcp Rcs e2rc Lc
Tpc Tcp Rcb e2rc Lc
:
þ erp Lp ðerp Lp 1Þ Rpc þ
1 Rcp Rcb e2rc Lc
ð9Þ
The a values of the samples can be determined
from either the amplitude or the phase of Eq. (9).
318
B.Z. Azmi et al. / Infrared Physics & Technology 45 (2004) 315–321
However, a number of pertinent values for gas, PE
coating, PE, and backing need not to be used by
normalising the signal of the test sample to that of
the reference sample. This simplifies Eq. (9) to
V1
T1c er1 L1 1 Rcp Rc2 e2rc Lc
ðxÞ ¼ b12
;
ð10Þ
V2
T2c er2 L2 1 Rcp Rc1 e2rc Lc
where subscripts 1 and 2 represent the test and
reference samples, respectively.
A similar normalisation procedure was reported
by Delenclos et al. [4], in which the measured PPE
signal was normalised to the one obtained with the
sensor alone, or to the signal obtained with a reference sample. However, laborious work, which
involved solving 10 · 10 matrix by CramerÕs rule,
was needed in obtaining the average PE voltage,
and it still constraints to work at high frequency in
thermal diffusivity measurement. In this paper, an
easier approach of thermal wave interferometry
presented here has greatly reduced the work in
getting the average PE voltage.
To further simplify Eq. (10), the term in the
square bracket can be neglected due to its small
attenuation of phase signal. Therefore the phase
signal of Eq. (10) can be written as
/1 /2 ¼
pf
a2
1=2
L2
pf
a1
1=2
L1 ;
ð11Þ
where subscripts 1 and 2 refer to the test sample
and the reference sample respectively. From Eq.
(11), we should expect a linear dependence of the
normalised phase signal with f 1=2 . Hence, if m1 and
m2 are the gradients for the test and the reference
samples respectively, it can be written as
1=2
1=2
p
p
L2
L1 :
ð12Þ
m1 m2 ¼
a2
a1
This means that by obtaining gradients of plot
phase signal versus f 1=2 , that is m1 of test and m2 of
reference sample, a1 can be calculated by substituting the known value a2 of the reference sample.
In this model, multiple reflections in the sample
are neglected because the chopping frequencies are
made for thermally thick operation for the sample.
Meaning the thermal wave is very short and dies
off after one reflection at sample–coating interface.
Similarly single reflection is only considered in
PVDF film due to its extremely small a ( 104
cm2 s1 ) compare to metal sample ( 100 cm2 s1 ).
The thermal diffusion length in Al sample with a of
0.979 cm2 s1 and in PVDF film with a of
5.4 · 104 cm2 s1 at 30 Hz are 1020 and 24 lm,
respectively.
3. Experimental procedures
The schematic diagram of PPE experimental
setup system is shown in Fig. 2. The beam of 30mW He–Ne laser (05-LHR-991) modulated by
variable frequency optical chopper (SR540) is
impinged onto the sample surface which is in
thermal contact with 36-lm-thick PVDF film
supported by 1.2-cm-thick Perspex backing. A
very thin layer of thermal conductive grease was
used as a coupling fluid to optimize the sample–
detector thermal contact. The grease layer does
affect the measured signal; moreover, it can be
eliminated via the process of normalisation. The
output from the PE detector is fed into a low-noise
preamplifier for signal amplification and then into
a lock-in amplifier (SR530) for signal analysis. The
effects of the wrap of PVDF film and piezoelectricity due to expansion and contraction of sample
were effectively eliminated by clamping the sample
and PVDF film between two Perspex plates [9].
The front plate has a circular opening to allow
uninterrupted illumination of the laser beam to the
sample. The measurement was carried out at room
temperature on aluminum, copper, and nickel
samples and the PPE signal was recorded over a
frequency range that suits the model proposed in
the theory.
Fig. 2. Schematic diagram of PPE experimental setup.
B.Z. Azmi et al. / Infrared Physics & Technology 45 (2004) 315–321
4. Results and discussion
319
In the experiment care has been taken, firstly, to
produce a reasonably short thermal diffusion
length so that the reflected thermal wave in the
sample will die off quickly in the sample after first
or before second reflection at sample–coating
interface. As evidence at starting frequency for Al
sample, for thermal diffusion length at even
slightly larger than sample thickness Ls the plot
still produces a good straight line. This is because
the reflected thermal wave will diminish in sample
as long as the thermal diffusion length is shorter
than 1:5Ls . The scanning in the thermally thick
regime or towards the shorter thermal diffusion
lengths will ensure the reflection of the thermal
wave in these samples is only due to the weak
exponential tail of the wave. Secondly, the gradient difference between the test and the reference
samples has to be reasonably large. If it is small the
normalised gradient will close to zero and thus will
introduce greater errors in calculating the thermal
diffusivity of the test sample.
Fig. 3 shows the plot for the phase of PE voltage versus the same range of square-root frequency generated by the infinite terms of thermal
wave in the right coating, which is in contact with
Cu sample, that is the term 1 Rcp Rc2 e2rc Lc of Eq.
(10). The gradient m of the plot is in the order of
104 , which can be neglected in our following
calculation of a.
Figs. 4–5 show the plot of linear dependant of
unnormalised PE phase signal as a function of f 1=2
for Al (Ls ¼ 670 lm), Cu (Ls ¼ 940 lm), and Ni
(Ls ¼ 500 lm) recorded over the same range of
frequency, where m is the slope of linearly fit line.
The phase signal was chosen in present study because it does not fall off as quickly as the magnitude signal over the frequency range. From these
figures, the Ni sample gives the highest magnitude
of slope among the three samples. Thus, one can
expect that Ni gives the lowest value of a.
By using Eq. (12), the calculated values of a
with respect to different reference sample were
shown in Table 1, and the results obtained are
reasonably close to the literature values [10], which
is less than 4%. The experimental error is only due
to errors in the sample thickness measurement and
the plot gradient.
The effect of gradient difference on a values can
be seen in Table 1. The gradient for Cu is in between gradients for Ni and Al. By using Ni as
reference, a of Al is closer to the literature values
compared to the case of using Cu as reference.
Similarly by using Al as reference, a of Ni is closer
to the literature value compared to when Cu is
Fig. 3. Theoretical PPE phase signal caused by the infinite
terms of thermal wave in the left PVDF coating in contact with
Cu sample of 940 lm thick. The solid line represents the best-fit
line with slope m.
Fig. 4. PPE phase signal of aluminum as a function of square
root chopping frequency. The solid line represents the best-fit
line with slope m.
320
B.Z. Azmi et al. / Infrared Physics & Technology 45 (2004) 315–321
Fig. 5. PPE phase signal of (a) copper and (b) nickel as a function of square root chopping frequency. The solid line represents the
best-fit line with slope m.
Table 1
The a values for materials obtained using phase signal
Sample
Al
Cu
Ni
Calculated thermal diffusivity a (cm2 s1 )
Al
(as reference)
Cu
(as reference)
Ni
(as reference)
–
1.138 ± 0.046
0.233 ± 0.006
1.006 ± 0.040
–
0.237 ± 0.008
0.956 ± 0.027
1.117 ± 0.042
–
used as reference. The effect of thin thermal grease
layer even though is not included in the present
analysis this technique still produces reasonably
good result. This is because it can be excluded after
normalisation similarly as the considered PVDF
coating layer.
Application of this normalisation procedure to
amplitude signal does not give the thermal diffusivity value as can be seen in Table 2. This is be-
Table 2
The a values for materials obtained using amplitude signal
Sample
Al
Cu
Ni
Calculated thermal diffusivity a (cm2 s1 )
Al
(as reference)
Cu
(as reference)
Ni
(as reference)
–
3.0170
1.4412
0.4892
–
0.9628
0.3508
0.8540
–
Average
(cm2 s1 )
Literature
value
(cm2 s1 )
Deviation
(%)
Reference
0.981
1.128
0.235
0.979
1.163
0.229
0.2
3.0
2.6
[10]
[10]
[10]
cause larger noise to signal ratio in the system as
such after normalisation, the data were too scattered hence the thermal diffusivity value are completely off from the actual value.
5. Conclusion
In short thermal diffusion length operations of
samples, a series of transmitted waves from right
and left coatings, and hence resultants of transmitted waves from right and left coatings into
PVDF film were taken into account in the generation of PPE signal. A normalisation procedure
was used as a convenient way to eliminate the effect of a number of unknown parameters in other
media of PPE cell. The obtained value of thermal
diffusivity for aluminum, copper, and nickel samples were reasonably close to the literature values.
B.Z. Azmi et al. / Infrared Physics & Technology 45 (2004) 315–321
Acknowledgements
The authors are grateful to the Ministry of
Science, Technology and the Environment of
Malaysia for supporting this work under IRPA
grant no. 02-02-04-0132-EA001.
References
[1] J.B. Bergman, J.H. McFee, G.R. Crane, Appl. Phys. Lett.
18 (5) (1971) 203.
[2] A. Mandelis, M.M. Zver, J. Appl. Phys. 57 (9) (1985) 4421.
321
[3] M. Marinelli, U. Zammit, F. Mercuri, R. Pizzoferrato, J.
Appl. Phys. 72 (3) (1992) 1096.
[4] S. Delenclos, M. Chirtoc, A.H. Sahraoui, C. Kolinsky,
J.M. Buisine, Anal. Sci. 17 (2001) 161.
[5] C.A. Bennett, R.R. Patty, Appl. Opt. 21 (1982) 49.
[6] D.P. Almond, P.M. Patel, Photothermal Science and
Techniques, Chapman & Hall, London, 1996.
[7] J. Opsal, A. Rosencwaig, J. Appl. Phys. 53 (6) (1982)
4240.
[8] A. Minamide, M. Shimaguchi, Y. Tokunaga, Jpn. J. Appl.
Phys. 37 (1998) 3144.
[9] M. Aravind, P.C.W. Fung, Meas. Sci. Technol. 10 (1999)
979.
[10] R.D. Lide, CRC Handbook of Chemistry and Physics,
78th ed., CRC Press Inc., Florida, 1997.