ARTICLE IN PRESS
Nuclear Instruments and Methods in Physics Research A 518 (2004) 448–451
Imaging with the leak microstructures: preliminary results
C. Baiocchia, G Balbinotb, A. Battistellaa, G. Galeazzic, F.S. Lombardid,
M. Lombardia,*, G. Pretea, A. Simona
a
INFN Laboratori Nazionali di Legnaro, viale dell’Universita" 2, Legnaro, Padova, Italy
b
Italstructures, via Monte Misone 11/d, Riva del Garda, Trento, Italy
c
Universita" degli studi di Padova, Dipartimento di Fisica Galileo Galilei, Padova, Italy
d
Osservatorio Astronomico S. Lucia Stroncone, Terni, Italy
Abstract
We report on some images obtained with the Leak Microstructures (LM), which are elements for the detection of gas
ionizing particles based on needle-point anodes. X-ray images of a mask with seven holes, 300 mm in diameter and
100 mm separated, drilled on a tantalum sheet were obtained with a very good spatial resolution. Varying the potential
to the drifting electrode we put in evidence the possibility to perform a ‘‘zoom’’ of the image. X-ray images of a
tantalum sheet of about 20 20 mm2 were obtained using a matrix of 9 9 LMs read by only 6 electronic chains (6
ADCs).
r 2003 Elsevier B.V. All rights reserved.
PACS: 29.40.Cs; 85.60.Gz; 87.59.e
Keywords: Gaseous detector
1. Introduction
We would like to remember that the Leak
Microstructures (LM), widely reported elsewhere
[1–7], are elements for the detection of gas ionizing
particles based on the point of needles as anodes.
For every single ionizing radiation detected they
give a pair of ‘‘induced’’ charges of the same
amount (pulses of the same amplitude), with
opposite sign, the same collection time and,
essentially, in time coincidence that are proportional to the primary ionization collected. Gas
*Corresponding author. Osservatorio Astronomico S. Lucia
Stroncone, Terni, Italy. Fax: +39-49641925.
E-mail address: mariano.lombardi@lnl.infn.it
(M. Lombardi).
multiplication more than 105 with more than 108
electrons in the primary avalanches, working in
proportional region, are possible [3]. The complete
lack of insulating materials in the active volume of
these microstructures avoids problems of charging-up and makes their behaviour stable and
repeatable.
In our attempts to get images with the LMs we
had well in mind the works reported in Ref. [8,9].
2. Imaging with alpha particles
To begin with imaging using LM and to set in
order our electronic set-up and data acquisition
system we used an alpha particles source 241Am.
0168-9002/$ - see front matter r 2003 Elsevier B.V. All rights reserved.
doi:10.1016/j.nima.2003.11.043
ARTICLE IN PRESS
C. Baiocchi et al. / Nuclear Instruments and Methods in Physics Research A 518 (2004) 448–451
449
Fig. 1. Electronic set-up.
Fig. 3. Image of two wires.
Fig. 2. Experimental set-up (not to scale).
The electronic and experimental set-up used are
reported in Figs. 1 and 2. The needle (anode) and
the four pads (10 10 mm2 each) of the detector
were each equipped with an inverting current
preamplifier
(input
impedance
Z ¼ 100 O,
Vout =Vin ¼ 3 or 0.3 mV/mA) [3], followed by main
amplifier (Silena 7612) and by ADC (Silena 4419/
V). A trigger pulse has been generated from anodic
signal in order to start with the acquisition. The
height of the point, to respect the plane of the
coplanar pads, was 100 mm.
For every detected particle there are five digital
numbers that depend on the amplitude of the
signals; four of these numbers must be elaborated
with the charge-ratio method in order to calculate
the avalanche position. The algorithm we used is
the following:
PX ¼
X1 X2
;
X1 þ X2
PY ¼
Y1 Y2
;
Y1 þ Y2
where X1 ; X2 ; Y1 ; Y2 are the four ‘‘induced’’
charges (pulses) on the pads.
After having equalized the five electronic chains
we got our first data with the experimental set up
reported in Fig. 2. Two copper wires, set at Vdrift ;
1 mm in diameter and 1 mm spaced, were placed
between the detector and the source at 2 mm from
pads.
Fig. 3 reports the distribution of data (4110
detected alpha particles) on XY plane when VLM
was 1050 V and Vdrift 300 V in 760 Torr of
isobutane. Alpha particles from the extended
241
Am source that hit wires are certainly adsorbed
while others, which ionize the gas under them,
make a map of electric field and so of the wires.
These results permit to conclude that it is
possible to make images with a needle and four
coplanar pads, that is a LM, with a spatial
resolution which is certainly better than 1 mm.
Considering that the distance between the two axes
of the wires is 2 mm, the field of vision results to be
of several millimetres in diameter.
3. Imaging with a
55
Fe source
The experimental set-up consisted in a window
drift electrode set at 3 mm from the pads and
above it a tantalum sheet, thickness 0.2 mm, on
which seven holes 300 mm in diameter and
separated by 100 mm were drilled, was leaned as
mask. Above this mask was set the 55Fe source.
The LM used and the electronic set-up were still
the same of Fig. 1. Because the source we used is
punctiform, to reduce parallax error, we put it at
1 mm above the mask.
ARTICLE IN PRESS
450
C. Baiocchi et al. / Nuclear Instruments and Methods in Physics Research A 518 (2004) 448–451
Fig. 4. Zoom effect due to different drifting electric field;
starting from left, the drifting electrode was at 50, 100, 200
and 500 V, respectively.
The working conditions were VLM 1150 V
in 760 Torr of isobutane. Fig. 4 reports results
when Vdrift was 50, 100, 200 and 500 V
respectively and the other working conditions were
the same for all the images.
The evaluated mean spatial resolution across the
edges of images is 118 mm FWHM.
It should be noted that increasing the drifting
electric field a zoom effect is obtainable.
Fig. 5. Electronic set-up to get the image of Fig. 6.
4. Imaging with a X-ray generator
With a printed board for 256 LMs, reported in
Ref. [7], we started to make images with an
extended X-ray generator. Each LM of this board
has 4 pads, two for the X position (left and right)
and two for the Y position (Up and Down). We
found that, connecting (electrically) together all
the 256 Left pads, as well as the 256 Right, Up and
Down pads to be all read with only four channels
(4 Peamp., 4 Main Amp. and 4 ADCs) and making
again the image of Fig. 4, the mean spatial
resolution was still of the order of 100 mm FWHM.
To perform a two-dimensional image, with a
multi-LM matrix, it remained to read the anodepoints one by one and to give them an address.
This was accomplished reading each point through
two diodes, Fig. 5, one for the X position and one
for the Y ; supplying two delay lines. In this
manner it was enough to use only six electronic
chains, four for the pads as said before and
another two to read the two time-to-amplitude
converters for the addresses, to get the image of
Fig. 6. This was achieved in 760 Torr of isobutane
(C4H10), at VLM ¼ 1900 V (this voltage is common
to all the points, diodes and delay lines) when to
Fig. 6. Shadowgram of a tantalum sheet of about 20 20 mm2.
the window drift electrode, 3 mm above the pads,
was applied 200 V. In this image, in which there
is the shadowgram of a sheet of tantalum of about
20 20 mm2 and 300 mm thick, only a matrix of 81
LMs were instrumented. The image was taken
using the primary beam of a tungsten-anode X-ray
generator setting the HV at 20 kV. In this figure 81
domains (9 9 LMs) are well distinguishable. The
pitch of the points was 5.08 mm. Unfortunately we
were not able to light the points in such a manner
that they would have a field of vision of
5.08 5.08 mm2 but only about 3 3 mm2. To
ARTICLE IN PRESS
C. Baiocchi et al. / Nuclear Instruments and Methods in Physics Research A 518 (2004) 448–451
overcome this problem now we are evaluating the
proper pitch to be used to obtain a continuous
image.
References
[1] M. Lombardi, et al., Proceedings of the International
Conference-Brolo (Messina) 15–19 Oct., World Scientific,
Singapore, 1996, pp. 459–465.
[2] M. Lombardi, F.S. Lombardi, Nucl. Instr. and Meth. A 392
(1997) 23.
451
[3] M. Lombardi, et al., Nucl. Instr. and Meth. A 388 (1997)
186.
[4] M. Lombardi, et al., Nucl. Instr. and Meth. A 409 (1998)
65.
[5] M. Lombardi, et al., Nucl. Instr. and Meth. A 461 (2001)
91.
[6] M. Lombardi, et al., PRAMANA (Indian Acad. Sci.) 57 (1)
(2001) 115.
[7] M. Lombardi, et al., Nucl. Instr. and Meth. A 477 (2002)
64.
[8] J.E. Bateman, Nucl. Instr. and Meth. A 240 (1985) 177.
[9] M. Lampton, R.F. Malina, Rev. Sci. Instrum. 47 (11) (1976)
1360.