5320A Multifunction Electrical Tester Calibrator: Verify and Calibrate Electrical Test Tools With A Single Instrument
5320A Multifunction Electrical Tester Calibrator: Verify and Calibrate Electrical Test Tools With A Single Instrument
5320A Multifunction Electrical Tester Calibrator: Verify and Calibrate Electrical Test Tools With A Single Instrument
Broad workload coverage The 5320A calibrates a broad range of equipment, including: Hipot testers Insulation resistance testers (megohm meters) Loop/line impedance testers Continuity testers Earth resistance testers Ground bond testers Leakage current testers Circuit breaker testers (RCD/GFCI) Multifunction installation testers Portable appliance testers (PATs) Medical electrical safety testers
Multifunction Hipots
Insulation Testers
provide additional assistance if you need it. For even greater efficiency, the 5320A can be automated with MET/CAL Plus Calibration Management Software. MET/CAL Plus software has become the industry standard for automating the calibration process and managing the inventory of your cal lab. MET/CAL Plus is a complete, scalable and affordable solution. LAN, IEEE-488 and RS-232 interfaces on the 5320A provide convenient, industry standard connectivity.
Hipot testers
Electrical safety testing with hipots is an integral part of development and manufacturing of electronic and electrical products, ranging from refrigerators to power supplies. Such testing is often required by government regulations to ensure product safety. The Fluke 5320A provides bestin-class hipot calibration of ac and dc voltage. The built-in 1000 V meter measures voltage and current. For voltages over 1 kV, Fluke provides a standard 10 kV adapter or an optional 40 kV probe. With the 10 kV adapter, ac voltages to 10 kV can be calibrated with 0.5 % uncertainty. Either item can be used with the built-in meter to measure voltage over 1000 V. For calibration of hipot current up to 100 mA, Fluke offers a load adapter accessory. Use the load adapter in conjunction with the 5320A built-in current meter for full calibration of hipots. (Load adapter available early 2007).
Voltmeters
Many testers have built-in voltmeters. But you dont need an additional calibrator to test this part of the workload. The 5320A/VLC calibrator adds a precision voltage source to 600 V with 0.1 % uncertainty, to calibrate ac/dc voltmeters.
Active terminal display B. Always know which calibrator terminals are active. When a function has been selected, the graphical display shows the active terminals.
1 3
Select a function at the press of a button. Functions are clearly organized into Outputs (HI Ohms, LO Ohms, Leakage Current and Voltage Source), discrete high current resistors (for loop/line impedance, ground bond and RCD/GFCI) and the meter (ac/dc voltage and current).
Soft menu keys C. Soft menu keys adapt to the active function, so the menu structure is intuitive and easy to learn.
B E C
Output knob, numeric keypad D. To select an output value or measurement range, use the numeric keypad or rotary knob.
Graphical help guide E. Find out what connections to make in an easy-to-understand graphical format. The help guide is available through the Mode softkey.
2
LAN, GPIB, RS-232 connectors Makes it convenient to connect the 5320A to your PC for automation and data exchange
Enter values. A calculator-style keypad makes it easy to enter values. You can also use the edit knob to vary the values entered.
Spec readout F. The spec readout lets you view the uncertainty of the sourced or measured primary value.
Connect the unit under test. Not sure of how to connect the UUT properly? The active terminal window and the graphical help guide shows you exactly how to do it. Just press the Operate key and youre ready to go!
Sales Utilities CD
de
in
S.A U.
Summary Specifications
Measurement functions
Voltmeter
Range of input voltage: 0 to 1000 V RMS (ac or dc) Resolution: 4 digits Frequency range: DC, 20 Hz to 2 kHz Reading/second: 2 Range 10 V 100 V 1100 V Uncertainty
(% of reading + mV)
Maximum Voltage
(ac+dc) Peak
Uncertainty
(tcal 5 C)
[1]
Maximum test current ac (RMS) or dc 30 A 28 A 25 A 18 A 10 A 8A 6A 3.2 A 2.0 A 1.5 A 0.8 A 0.5 A 0.25 A 0.1 A 0.05 A 0.03 A
Maximum short term test current ac (RMS) or dc* 40 A 40 A 40 A 40 A 40 A 40 A 30 A 16 A 10 A 7.5 A 4.0 A 2.5 A 1.25 A 0.5 A 0.25 A 0.15 A
Test current uncertainty 1.5 % + 0.7 A 1.5 % + 0.5 A 1.5 % + 0.35 A 1.5 % + 0.3 A 1.5 % + 0.2 A 1.5 % + 0.15 A 5 % + 0.1 A 1.5 % + 70 mA 1.5 % + 50 mA 1.5 % + 30 mA 1.5 % + 20 mA 1.5 % + 10 mA 1.5 % + 5 mA 1.5 % + 3 mA 1.5 % + 2 mA 1.5 % + 2 mA
Resolution 1 mV 10 mV 100 mV
Ammeter
Range of input current: 0 to 30 A RMS Resolution: 4 digits Frequency range: DC, 20 to 400 Hz Reading/second: 2 Range 300 mA 3A 30 A Uncertainty
(% of reading + mA)
100 k
[1] Uncertainty is valid at maximum test voltage to 500 V. For test voltage over 500 V add 0.1 % per each 200 V. [2] 100 GW is a single value resistor. [3] With resistance multiplier adapter. [4] Uncertainty of resistor to be multiplied by 1000 + 1 % of displayed value to 99.99 GW, or + 2 % of displayed value to 999.9 GW, or + 3 % of displayed value to 10 TW.
Resolution 0.1 mA 1 mA 10 mA
1 kW 1.8 kW
VA measurement
*Maximum short term test current is dened as RMS value of halfwave or fullwave test current from the UUT. Maximum time of test current is 200 ms (represents 10 full waves of power line voltage at 50 Hz and 12 full waves at 60 Hz).
Range: 0 to 10 kV ac peak/dc Uncertainty: 0.3 % of reading + 5 Vdc 0.5 % of reading + 5 Vac (at 50 or 60 Hz) Range: 0 to 40 kV ac peak/dc Uncertainty: 0.5 % of reading + 10 Vdc 0.5 % of reading + 10 Vac (at 50 or 60 Hz)
Output functions
Voltage calibrator (5320A/VLC only)
Range of output voltage: 3 V to 600 V ac or dc Voltage Resolution: 4 digits Frequency range: 40 Hz to 400 Hz Frequency resolution: 3 digit Distortion of ac output signal: 0.2 % + 10 mV Range 3 V to 29.99 V 30 V to 99.99 V 100 V to 299.99 V 300 V to 600 V
10
Trip current Range of trip current in 0.5xI and 1xI mode: 3 to 3000 mA, in 1 mA steps Range of trip current in 1.4xI and 2xI mode: 3 to 1500 mA, in 1 mA steps Range of trip current in 5xI mode: 3 to 600 mA, in 1 mA steps Uncertainty of measured trip current: 1 % RMS (0.5xI and 1xI mode), 2 % RMS (1.4xI and 2xI mode), 5 % RMS (5xI mode) Trip time Range of trip time: 10 to 5000 ms Uncertainty of trip time: 0.25 ms Output voltage on RCD terminals: Power line voltage, 115 V or 230 V
[1] Uncertainty is valid to 200 m. For higher power rating, add 0.1 % per each 300 m above 200 m.
Uncertainty in ac mode
(% of reading + mV)
Uncertainty in dc mode
(% of reading + mV)
General specications
Operating temperature: 18 C to 28 C Calibration (Tcal): 23 C Storage temperature: -20 C to 70 C Temperature Coefcient: Temperature coefcient for temperatures outside of Tcal 5 C between +5 C and +40 C is 0.1/ C Relative humidity (operating): < 70 % to 28 C Altitude (operating): 3,050 meters (10,000 ft) maximum Altitude (non-operating): 12,200 meters (40,000 ft) maximum Specications condence interval: 99 %
Environmental
EMI/RFI: Designed to comply with Class B per EN61326 Safety: Designed to comply with EN61010 Electro static discharge: This instrument meets class 1 for ESD requirements per EN61326
Standard interfaces: LAN, IEEE 488 (GPIB), RS-232 Dimensions: (D x W x H) 450 mm x 480 mm x 170 mm (17.7 in x 18.9 in x 6.7 in). Mounts within industry-standard 483 mm (19 in) rack-mount frames when tted with the rack mounting kit. Shipping weight: 25 kg, (55 lbs) Net weight: 18 kg, (39.7 lbs)
Power
Line voltage: 115/230 V 10 %, 50/60 Hz nominal Line frequency: 47 to 63 Hz Power consumption: 150 VA maximum
Ordering information
Models 5320A 5320A/40 5320A/VLC 5320A/VLC/40 Description Multifunction Electrical Tester Calibrator Calibrator with 40 kV Probe Calibrator with 600 V Source and Active Loop Compensator 5320A/VLC Calibrator with 40 kV Probe
Rugged Transit Case Rack Mount Kit (Slides) US Mains Plug-to-Banana Leads (available early 2007) Hipot Current Calibration Load Resistors (available early 2007)