Bs 205 0512e
Bs 205 0512e
Bs 205 0512e
CONTENTS
Introductions . . . . . . . . . .
Technical Specification . .
Dimensions . . . . . . . . . . . . .
Front Panel . . . . . . . . . . . .
Rear Panel . . . . . . . . . . . . . . 9
Parameter . . . . . . . . . . .
16
ZERO Calibration
21
. . . . . . .
. . . 22
Equivalent Calibration . . . 24
Comparator Data
. . . . . . .
26
. . . . . . . . .
34
Key Lock . . . . . . . . . . .
36
Option
37
Hold Function
. . . . . . . . . . .
Error Message
and Trouble Shooting . . .
47
2
Introduction
1. Introduction
Thank you for purchasing BS-205 High precision digital indicator.
Check for any breakdown during transit and discrepancy of the
specification. Always keep this manual at hand.
The BS-205 Series are configured as follows :
Check that there is no discrepancy between the model and its
specifications you have chosen when ordering and the model and its
specifications under your hand.
In order to obtain the highest performance of your BS-205, thoroughly
read this manual before use.
2. Precautions
Do not severely press because the light pressing of keys can incite
the operation.
Keep the scale away from strong EMI noises may cause incorrect
weight readings.
Keep the main body from rain and keep in dry area.
3
The Features of BS-205
1. Features
2. Main Function
Various specification of weight conversion speed.
(Digital Filter Function)
User can set the max. weight which users want to and division
at ones disposal.
4
Technical Specification
1. Analog Input & A/D Conversion
Analog signal input range
0 mV ~ 20 mV
Non-linearity
1/20,000
0.5V/Digit (min.)
Temperature drift
DC 5V 5%, 60
Voltage
Input Noise
Input Impedance
10 (Min.)
A/D converter
50 times/sec
2. Digital Part
Display
7 Segment LED,
5-Digits, 14.1mm(Height)
-minus signal
Display speed
50 times/sec ~ 1 times/sec
Additional symbols
Min. Division
Decimal Point
5
3. Technical
AC adapter
Power consumption
10 VA
Data Memory
10 year
Operating temperature
-10~+50
Humidity
85% Rh Max.
Overall dimensions
Weight
530 g
4. Option
Standard
Option 1
Option 2
Option 3
Option 4
Option 5
6
Dimensions
96
High Precision DIGITAL INDICATOR
BS - 205
OK
HI
HOLD
ZERO
STABLE
48
LO
M
BONGSHIN
136
119
11
92
43
91
44
To mount the BS-205 to the panel, remove its fittings and insert it through
the hole in the front of the panel. From the back of the panel, fix the
product to the panel with the fittings.
7
Front Panel
High Precision DIGITAL INDICATOR
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
4
BONGSHIN
1. Display Lamp (
2. Keyboard
M
M
+
1
+
1
+
4
+
4
9
Rear Panel
F.G
POWER
LO (L1)
OK (L2)
HI (L3)
RELAY OUTPUT
HOLD
ZERO
SHIELD
COM1
INPUT
4
E+
E-
S+
S-
COM2
OPTION
1
Pin No.
13
POWER
14
POWER
15
F.G
10
LOAD CELL :
Pin no.
Signal name
Contents
SIG- (Blue)
SIG+ (Green)
EXC- (White)
EXC+ (Red)
SHIELD
Shield
Maker
1
SIG-
2
SIG+
3
EXC-
4
EXC+
5
SHIELD
BONGSHIN
BLUE
GREEN
WHITE
RED
SHIELD
BLUE
RED
WHITE
WHITE
WHITE
BLACK
WHITE
GREEN
BLACK
WHITE
GREEN
WHITE
GREEN
GREEN
GREEN
WHITE
GREEN
WHITE
WHITE
RED
WHITE
BLACK
BLACK
BLACK
BLACK
BLUE
BLACK
BLACK
BLUE
BLACK
RED
GREEN
RED
RED
RED
RED
RED
RED
RED
GREEN
SHIELD
YELLOW
SHIELD
SHIELD
SHIELD
SHIELD
SHIELD
SHIELD
YELLOW
SHIELD
Cautions 1. When connecting a six-wire type strain gage sensor, shortcircuit (EXC+ and SEN+)(EXC- and SEN-), respectively.
Cautions 2. Applied voltages to the strain gage sensor are 10.
When any sensor rated below the applied voltage is connected,
it may generate heat or be damaged
11
IN-PUT : COM1, ZERO, HOLD
This key is to control a equipment from the outside .
Please connect between COM terminal and each input terminal .
Because the power of input terminal was connected with 12V voltage
From the inside.
* An electric current is about 10mA.
* Please make the minimum time to input a data with over 50msec.
Pin No.
Name
COM1
ZERO
HOLD
Description
Common for all external control terminals.
Control for digital zero function. Enabled when
short-circuited or at the same potential as COM.
Control for hold function. Enabled when shortcircuited or at the same potential as COM.
12
RELAY OUTPUT :
Pin No.
Name
Description
COM2
10
HI (L3)
11
OK (L2)
12
LO (L1)
13
OPTION :
Name
1
2
3
4
V-OUT
COM
COM
A-OUT
Description
Voltage output terminal (0 to 10V)
Common terminal for analog output.
Common terminal for analog output.
Current output terminal (4 to 20mA)
Name
Description
1
2
3
4
RX
TX
SG
NC
RS-232C reception
RS-232C transmission
Common terminal for communications
Do not connect this terminal
Name
1
2
3
4
RXD (-)
RXD (+)
TXD (-)
TXD (+)
Description
RS-422
RS-422
RS-422
RS-422
RX(-)
RX(+)
TX(-)
TX(+)
Name
1
2
3
4
NC
NC
TXD (-)
TXD (+)
Description
Do not connect this terminal
Do not connect this terminal
RS-485
RS-485
14
TERMINAL BLOCK
AC Voltage change
Available to change AC110/220V with multiple.
Before setting up, please confirm the power voltage.
Please change the connect terminal of 110V/220V after opening the
cover. If you need to change. (It was setted with AC220V at the first)
15
AC 220V change
AC 110V change
16
Information on Each Parameter
Indication
Name
Calibration data
Load cell
LC.CAL Rated output
set
Maximum
SCALE
Capacity set
Span weight
At.CAL
set
Function data
biP
Display Mode
Setup options
Default
value
1 mV/V ~3 mV/V
2.0000
0~99999
10000
0~99999
20000
biP
Decimal
0(0)/1(0.0)/2(0.00)/3(0.000)/4(0.0000)
2
Point set
Minimum
ds.1
1/ 2/ 5/ 10/ 20/ 50/ 100/ 200 digit
1
Division set
Number of
average
dU.2
1/ 2/ 5/ 10/ 25/ 50
2
operations
setup
Zero
Zt.oFF
OFF/ON
OFF
Tracking set
Auto zero
AZ.oFF
OFF/ON
OFF
backup set
Hold function OFF(No used)
Hd.oFF
OFF
set
/P(Peak Hold)/ I(Instant Hold)
Manual Tare
tare
0~99999
00000
weight set
Condition data (Serial interface option)
Equipment
Id.0
0 ~15
0
ID setup
Baud rate
1200/2400/4800/9600/1920(19200)/
r.9600
9600
setup
3840(38400)
CommunicaRc.on
ON(Data is required)/OFF(Stream mode) ON
tion mode
dP.2
Analog
output LO
indication
setup
0~99999
00000
17
dA.HI
Analog
output HI
indication
setup
Indication
Name
0~99999
10000
Setup options
Default
value
Comparator data
rLY.1H
rLY.1L
r1H.on
rLY.2H
rLY.2L
r2H.on
rLY.3H
rLY.3L
r3H.on
L1 HI side
judgment
value setup
L1 LO side
judgment
value setup
L1 Relay
motion mode
L2 HI side
judgment
value setup
L2 LO side
judgment
value setup
L2 Relay
motion mode
L3 HI side
judgment
value setup
L3 LO side
judgment
value setup
L3 Relay
motion mode
-9999 ~ 99999
10000
-9999 ~ 99999
10000
r1H.on
-9999 ~ 99999
9999
-9999 ~ 99999
5001
r2H.on
-9999 ~ 99999
5000
-9999 ~ 99999
5000
r3H.on
18
1. Calibration data
1-1 Load cell rated output set (LC.CAL)
Set the rated output value(mV/V value) for the strain gauge sensor.
The setting rage of the rated output value for strain gauge sensor is 0.1 to
3.0 mV/V.
1-2 Maximum Capacity set (SCALE)
Set the indicated value when the rated output value is acquired and
finalized.
Setting range of SCALE set value (indicated value) 100 to 99999.
Acquire both load cell rated output set value and measured value (SCALE
value) and finalize with ENTER key.
1-3 Span weight set (At.CAL)
Actual load is performed by applying actual load to the connected strain
gauge sensor, and by setting the indicated value (SPAN set value) at the
time.
Setting range of SCALE set value (indicated value) 100 to 99999.
Resolution in this meter is 20000 at the time of a 2.0000 mV/V value.
2. Function data
2-1 Display mode (biP)
biP +/- display, uniP + display,
2-2 Decimal point set (dP.2)
Set the decimal point position displayed on the set-value select screen.
Setting items : 0(0), 1(0.0), 2(0.00), 3(0.000), 4(0.0000)
2-3 Minimum Division set (ds.1)
Set the minimum updating width of indicated values.
Set it on the set-value select screen.
Setup items : 1, 2, 5, 10, 20, 50, 100, 200
2-4 Average operations setup
(dU.2)
This function is provided for the moving average of data after A/D
conversion so as to reduce the fluctuation of the indicated values.
As the number of times of moving average is increased, the indicated
value is stabilized. However response becomes slow.
19
20
3-2 Baud rate setup (r.9600)
To enable the communication mode, proper baud rate should be set
within r1200~r3840, in which r1920 means 19200bps and r3840 means
38400bps.
Setup items : 1200, 2400, 4800, 9600, 1920(19200), 3840(38400)
3-3 Communication mode setup (Rc.on)
Rc on : Transmit when data is required
Rc oFF : Stream mode
Serial communication protocol.
21
ZERO Calibration
The digital zero function zeros the indication given at an arbitrary timing.
Thereafter, the function shows the amount of change from the point of
zeroing. However, this function serves as an indication resetting function for
a frequency measuring unit. Thus, the digital zero function can be used to
reset the indication when there is no input signal at all.
Note that, the on/off control of the digital zero function can be achieved by
means of terminal control or front panel keys. In the case of terminal control,
the digital zero function is turned on by short circuiting the ZERO and COM1
terminals or setting both terminals to the same voltage level. The indication
at that moment is zeroed. In the case of control with the front panel keys,
hold down the enter key and press the increment for about 1 second to zero
the indication at that moment.
key
2
Auto zero
backup set
OFF/ON
OFF
22
Actual Load Calibration
Actual load calibration means that calibration is carried out by applying
actually measured pressure to a sensor such as a load cell connected to
the meter.
1. Method of setting
Pressing the
key
2
23
24
Equivalent Calibration
Equivalent calibration means that calibration is carried out according to the
ratings (specifications) of such a sensor as a load cell. It is not necessary
to connect the sensor or to apply pressure to the sensor. Set the rated
value for the strain gauge sensor and the indicated value at the time.
Equivalent calibration can be performed only by setting the rated
output(mV/V) of the strain gauge sensor and the indicated value at the time,
without actual load.
1. Method of setting
Pressing the
CAL 4 : End
key
2
25
26
Comparator Data (Relay setting)
1. Method of setting
Pressing the
key
3
2. Calibration Manu
3. Output Specifications
Conditions for comparison
Indicated value
Judgment result
HI
OK
LO
27
4. Method of Setting Comparator Data (L1 ~ L3)
M
.
3
28
4
29
Indication
Setup options
Relay OFF
Relay ON
r1H.on
Relay 1 Hi
Relay 1 LO
Time
Relay OFF
Relay ON
r1L.on
Relay 1 Hi
Relay 1 LO
Time
The r1r.on mode means the range mode, in which the load
within the limits will activate a relay on and the load beyond
the limits will activate LO LED on as a warning of high NG or
low NG.
Weight
r1r.on
Relay OFF
Relay ON
Relay 1 Hi
Relay 1 LO
Time
30
5. Relay Motion Mode
5-1 HI ON relay motion mode : r1H.on, r2H.on, r3H.on
: The load within the limits will activate a relay ON.
Weight
L3 ON
L3 Hi value
[ HI ]
L3 Lo value
L3 OFF
L2 ON
L2 Hi value
[ OK ]
L2 Lo value
L2 OFF
L1 Hi value
[ LO ]
L1 ON
L1 Lo value
L1 OFF
0
Time
OFF
ON
OFF
OFF
ON
OFF
OFF
ON
OFF
Example 1)
rLY.1H
rLY.1L
HI ON
mode
rLY.2H
rLY.2L
rLY.3H
rLY.3L
L1 HI side
judgment value
L1 LO side
judgment value
L2 HI side
judgment value
L2 LO side
judgment value
L3 HI side
judgment value
L3 LO side
judgment value
31
5-2 LO ON relay motion mode : r1L.on, r2L.on, r3L.on
: The load within the limits will activate a relay OFF.
Weight
L3 OFF
L3 Hi value
[ HI ]
L3 Lo value
L3 ON
L2 OFF
L2 Hi value
[ OK ]
L2 Lo value
L2 ON
L1 Hi value
[ LO ]
L1 OFF
L1 Lo value
L1 ON
0
Time
ON
OFF
ON
ON
OFF
ON
ON
OFF
ON
Example 1)
rLY.1H
rLY.1L
LO ON
mode
rLY.2H
rLY.2L
rLY.3H
rLY.3L
L1 HI side
judgment value
L1 LO side
judgment value
L2 HI side
judgment value
L2 LO side
judgment value
L3 HI side
judgment value
L3 LO side
judgment value
1000
900
2000
1900
3000
2900
32
5-3 Range ON relay motion mode : r1r.on, r2r.on, r3r.on
: Indicates the result of judgment and turns on
if LO judgment value the measured value Hi judgment value.
Weight
L3 OFF
L3 Hi value
[ HI ]
L3 ON
L3 ON
L3 Lo value
L3 OFF
L2 OFF
L2 Hi value
[ OK ]
L2 ON
L2 ON
L2 Lo value
L2 OFF
L1 Hi value
[ LO ]
L1 Lo value
L1 OFF
L1 ON
L1 ON
L1 OFF
0
Time
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
Example 1).
rLY.1H
L1 HI side
judgment value
1000
rLY.1L
L1 LO side
judgment value
900
rLY.2H
L2 HI side
judgment value
2000
rLY.2L
L2 LO side
judgment value
1900
rLY.3H
L3 HI side
judgment value
3000
rLY.3L
L3 LO side
judgment value
2900
Range
mode
33
5-4 Comparator weighing mode : r1L.on, r2r.on, r3H.on
: Indicates the result of judgment and turns on
if L1 judgment value L2 judgment value L3 judgment value.
Weight
[ HI ]
L2 OFF
L3 Hi, Lo value
L2 Hi value
L2 ON
[ OK ]
L2 Lo value
L1 Hi, Lo value
L2 ON
[ LO ]
L2 OFF
Time
ON
OFF
ON
ON
OFF
ON
OFF
ON
OFF
Example 1)
L1
LO ON
mode
L2Range
mode
L3
HI ON
mode
rLY.1H
rLY.1L
L1 HI side
judgment value
L1 LO side
judgment value
1000
1000
rLY.2H
L2 HI side
judgment value
3000
rLY.2L
L2 LO side
judgment value
1000
rLY.3H
L3 HI side
judgment value
3000
rLY.3L
L3 LO side
judgment value
3000
34
Hold Function
1. Hold Function
The Hold function temporarily retains the indication. The hold function is
enabled by short circuiting the HOLD and COM terminals or setting both
terminals to the same voltage level. As a result the display unit retains the
indication given at that moment.
Hold function
set
Hd.oFF
OFF(No used)
/P(Peak Hold)/ I(Instant Hold)
OFF
Indication
Setup options
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
4
BONGSHIN
- Peak HOLD The peak hold function retains one of the maximum
values and provides output for that value. Selection
from these values is made using the hold function
data.
High Precision DIGITAL INDICATOR
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
4
BONGSHIN
35
- Instant HOLD The Hold function temporarily retains the indication.
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
4
BONGSHIN
36
KEY Lock or KEY Lock clear
1. Method of Setting KEY Lock
Press the shift key and enter keys together during measurement.
Protect others.
High Precision DIGITAL INDICATOR
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
4
BONGSHIN
Press the shift key and enter keys together during measurement.
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
4
BONGSHIN
37
Option
1. RS-232C Serial Interface (Option-1)
OPTION
1
Pin No.
Name
Description
1
2
3
4
RXD
TXD
SG
NC
RS-232C reception
RS-232C transmission
Common terminal for communications
Do not connect this terminal
Function
Condition data (Serial interface option)
Equipment
Id.0
0 ~15
0
ID setup
Baud rate
1200/2400/4800/9600/1920(19200)/
r.9600
9600
setup
3840(38400)
CommunicaRc.on
ON(Data is required)/OFF(Stream mode) ON
tion mode
Signal Format
Type
Method :
EIA-RS-232C
Full-Duplex , Asynchronous, Bi-direction
Data Bit
Start/Stop
Parity Bit
Code
:
:
:
:
8 (NO Parity)
1 bit
Even or Odd
ASCII
38
Serial Communication Protocol
The below describe an example on the settings for communication
between an indicator and a PC assuming that current display is 32.567
and the ID of a indicator is 5.
1) Transfer the current display from host PC
Byte 1 (Indicator ID)
35H
Byte 2
R(52H)
Byte 2
35H
Byte 8
6(36H)
Byte 3
+(2BH)
Byte 9
7(37H)
Byte 4
3(33H)
Byte 10
ETX(03H)
Byte 5
2(32H)
Byte 6
.(2EH)
Host (PC)
Byte 1 (Indicator ID)
35H
Byte 1 (Indicator ID)
35H
Byte 1 (Indicator ID)
35H
Indicator
Byte 2
Z(5AH)
Byte 2
H(48H)
Byte 2
L(4CH)
Example) 1(ID)+07.487
ASCII
HEXA
1+07.487
02 31 2B 30 37 2E 34 38 37 03
Example) 1(ID)+07487
ASCII
HEXA
1+07487
02 31 2B 30 37 34 38 37 03
Example) 15(ID)+07486
ASCII
HEXA
15+07486
02 3F 2B 30 37 34 38 36 03
Example) 2(ID)+074.86
ASCII
HEXA
2+074.86
02 32 2B 30 37 34 2E 38 36 03
No response
Transfer again
to enable
Transfer again
to disable
39
Connection of RS-232C
RXD
Transmit Data
TXD
Receive Data
GND
Chassis Ground
Carrier Detect
Request to Send
Clear to Send
RS-232C port of
BS-205
RXD
Transmit Data
TXD
Receive Data
GND
Signal Ground
Request to Send
Clear to Send
Carrier Detect
20
RS-232C port of
BS-205
40
2. RS-422/485 Serial Interface (Option-2)
-
Name
1
2
3
4
RX (-)
RX (+)
TX (-)
TX (+)
Description
RS-422
RS-422
RS-422
RS-422
RX(-)
RX(+)
TX(-)
TX(+)
Name
1
2
3
4
NC
NC
TXD (-)
TXD (+)
Description
Do not connect this terminal
Do not connect this terminal
RS-485
RS-485
41
3. Voltage (0~10V) Analog Output (Option-03)
The BS-205 series of unit meters can output an analog signal for an indicated
value.(when the meter is equipped with an analog output unit)
The analog output of the BS-205 series allows for arbitrary output scaling.
This scaling can be achieved by setting the indication value for an output of the
maximum scale value(10V for 0~10V output range) in the dA.HI parameter of the
scaling data.
The voltage output occurs proportionally the voltage according to
the size of a weight In 0V ~10V.
OPTION
1
Name
1
2
3
4
V-OUT
COM
COM
A-OUT
Description
Voltage output terminal (0 to 10V)
Common terminal for analog output.
Common terminal for analog output.
Current output terminal (4 to 20mA)
SPECIFICATION
output Voltage
0 ~ 10V DC out
Precision
Max 1/1000
Min Impedance
Over 1 k
42
Condition data (Analog output option)
High Precision DIGITAL INDICATOR
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
Analog output LO
indication setup
0~99999
00000
Analog output HI
indication setup
0~99999
10000
4
BONGSHIN
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
4
BONGSHIN
1) Analog Output LO
Set an indicated value when analog output is 0V or 4mA.
Setting range : + 99999
2) Analog Output HI
Set an indicated value when analog output is 10V or 20mA.
Setting range : + 99999
Example of analog output setting
Set the indicated value when analog output is 0V(or 4mA) to 400.
Set the indicated value when analog output is 10V(or 20mA) to 2000.
Note 1) When the indicated value becomes larger than analog output
set value, it is not properly output.
Note 2) When the indicated value is OVER, and analog output is less than
0V, the output is near 0V.
43
4. Electric current (4~20mA) Analog
Output (Option-4)
The BS-205 series of unit meters can output an analog signal for an indicated
value.(when the meter is equipped with an analog output unit)
The analog output of the BS-205 series allows for arbitrary output scaling.
This scaling can be achieved by setting the indication value for an output of the
maximum scale value(20mA for 4~20mA output range) in the dA.HI parameter of
the scaling data.
The current output occurs proportionally the current according to
the size of a weight In 4~20mA.
OPTION
1
Name
1
2
3
4
V-OUT
COM
COM
A-OUT
Description
Voltage output terminal (0 to 10V)
Common terminal for analog output.
Common terminal for analog output.
Current output terminal (4 to 20mA)
SPECIFICATION
output Voltage
Precision
Min Impedance
4 ~ 20 mA DC Current out
Max 1/1000
Under 500
44
Condition data (Analog output option)
High Precision DIGITAL INDICATOR
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
Analog output LO
indication setup
0~99999
00000
Analog output HI
indication setup
0~99999
10000
4
BONGSHIN
BS - 205
LO
OK
HI
HOLD
ZERO
STABLE
4
BONGSHIN
1) Analog Output LO
Set an indicated value when analog output is 0V or 4mA.
Setting range : + 99999
2) Analog Output HI
Set an indicated value when analog output is 10V or 20mA.
Setting range : + 99999
Example of analog output setting
Set the indicated value when analog output is 0V(or 4mA) to 400.
Set the indicated value when analog output is 10V(or 20mA) to 2000.
Note 1) When the indicated value becomes larger than analog output
set value, it is not properly output.
Note 2) When the indicated value is OVER, and analog output is less than
0mA, the output is near 0mA.
45
F.G
OPTION
Pin No.
SIGNAL
Pin No.
SIGNAL
20
4 x 104
1 x 10
21
8 x 104
2 x 100
22
4 x 10
23
8 x 10
24
Decimal Point
101
1 x 10
25
Decimal Point
102
2 x 101
26
Decimal Point
103
4 x 101
27
Decimal Point
104
8 x 10
28
10
1 x 102
29
11
2 x 10
30
12
4 x 10
31
13
8 x 102
32
14
1 x 103
33
15
2 x 10
34
16
4 x 10
35
17
8 x 103
36
18
1 x 10
37
2 x 10
19
POLAR (+/-)
BUSY
GROUND/ (-)COMMON
46
Setting
ID setup
Id.0
r.9600
Communication mode
Rc.oFF
Signal Logic
1)
2)
3)
Specification
1)
2)
47
Error Message and Trouble Shooting
ERROR
CAUSE
A/S
Checking for Input,
Waving a weight
Value.
Insulation
Resistance Value.
resistance
Resistance value of
Load cell.
Checking Insulation
Resistance value of
Load cell damage.
A. Changing a
Load cell.
(Normal Max 100M or
Weight value,
-OL-appear)
B. Not return to
ZERO
Confirm a connect of
Appear "no.LC"
Disconnected to
Load Cell.
Load cell
Checking a single wire
Of load cell cable
Input resistance
: about 1130
Output resistance
: about 1000
Reference.
Appear "Ovr"
"-Ovr"
Insulation
Resistance
: over100M
48
Warranty and After-service
1. Warranty
The warranty period shall be one year from the date of delivery. Any failure
that arises during this period and the cause thereof is judged to be
obviously attributable to BONGSHIN LOADCELL CO., LTD. Shall be remedied
at no cost.
2. After-service
This product is manufactured, tested, inspected, and then shipped under
stringent quality control. Should the product fail, however, contact (or send
the product to) your vendor or BONGSHIN LOADCELL directly.
(It is advisable that you send a memo describing the failure in as much
detail as possible along with the product returned.)