JIF 2002 User Manual v01
JIF 2002 User Manual v01
INTRODUCTION
1-2 Features
JIF-2002 Weighing Indicator & Controller Features:
SETTING FUNCTIONS:
SETTING GENERAL FUNCTIONS (F0XX, FCXX)
SETTING CONTROLS (F1XX)
SETTING SERIAL INTERFACE (F2XX)
SETTING PRINTER (F3XX)
PARAELLE OUTPUT (BCD) (F4XX)
ANALOGE OUTPUT (F5XX) (420 mA , 010 V)
CALIBRATION:
CAL
Setting Minimum Division
Setting Decimal
Setting Maximum Capacity
ZERO Adjust
SPAN Adjust
KEY LOCK:
Disable unimportant or useless keys.
SYSTEM INITIALIZE:
Re-install resets the JIF-2002 to the initial factory setting and starts the
operation.
Screw Signal
1 Positive Excitation Voltage, (EXC+)
2 Positive Sense Voltage, (SEN+)
3 Negative Sense Voltage, (SEN-)
4 Negative Excitation Voltage, (EXC-)
5 Positive Signal Voltage, (SIG+)
6 Negative Signal Voltage, (SIG-)
7 Shield, (SHD)
LTo connect your load cell to the weighing Indicator use a six-wire cable
with shield-connect the wires as indicated above. If the JIF-2002 is located
near the Load Cells (Within five meters or a few yards) you may use a
4-wire cable with shield, but first connect screws 1&2 and 3&4 with
independent jumper leads.
LThe analogue output from the Load Cell and input/output signals are
sensitive to electrical noise. Do not bind these cables together as it could
result in cross-talk interface. Please also keep them away from AC power
cables.
It could be dangerous by using improper battery or wrong connection of
battery.
Front Panel
Rear Panel
Side View
For Mounting
3-2 General
General
Type JIF-2002
Power Requirements AC 110V or AC 220V 10%
5060Hz, Approx. 17VA
Net weight Approx. 3.2 kg7.054 lb
Operation Temperature 10 45
Maximum Humidity 85%non-condensing
Physical Dimensions 240 (D) 190 (W) 104(H) mm
DIGITAL SECTION
Main Display 7-segment7-digit display, VFD screen with a
[Green Tube] 13mm character size, displays the weight.
Minimum Division 125102050
Maximum Display +800450
Under ZERO Indicator minus sign
ZEROAnnunciator Center of Zero
MDAnnunciator Motion Detected = Unstable
GROSSAnnunciator Gross Weight displayed
NETAnnunciator Net Weight displayed
TARE ENTERED Tare has been entered
Annunciator
Annunciator Status Annunciator
kg Annunciator Kilograms Displayed
lb AnnunciatorPonds Displayed
ZERO Key Stable ZEROLeft shift key
TAREKey TAREs when stable-in Net, display ZERO
Increase the number
GROSSNETKey Changes from Gross to Net and vise versa
Decrease the number
TARE CLEARKey Tare is clearedRight shift Key
PRINTACCKey Print data will send one timePrint
Accumulation
ACCCLEARKey Display AccumulatorClear Accumulation
SET POINTCODEKey Setting set point values/ Change set point code
FnKey Unused
MODEKey Change ModeEnter Key
STANDBYESC Key Standby StatusLeave the current status
PRINTER OP-03
Setting Date, Setting Year. Month, day, hour, minute, second
F 300
Time
F 301 Data Format
Not print 2Only print above the latest data
Date
3Print on all
1 Not print Only print above the latest data
Time
3Print on all
BCD OP-04
Same as display 2Gross data 3NET data
F 400 Data type
4TARE data
Stream 2 Stable and print 3Manual print
F 401 Output Mode
mode4Accumulate and print
F 402 Output Logic Positive Logic 2Negative Logic
Analog Output
Analog Output
F 500 Output 4~20 mA Output 0~+10 V
Data
F 501 Output Mode Same as display 2Gross data 3NET data
Loss-in-weight Not read Absolute Value
F 502
Absolute Value 1 BDI-2002 reads Absolute Value
Output current
F 503 when display 0.0mA through 9.99mA (Initial 0.40 mA)
ZERO
Output current
F 504 0.0mA through 9.99mA (Initial 20.0 mA)
at Full Capacity
Output Volt
F 505 when display -2.5V through +59.9V (Initial 00.0 V)
ZERO
Output Volt at -2.5V through +59.9V (Initial 10.0 V)
F 506
Full Capacity
STEP 2: Turn the power supply ON, the display will show blinking
SELECT.
STEP 3: Press the ZERO key and screen will show blinking CHEC,
and press key to start system check.
STEP 4: The system will check Green Tube and LED in sequence.
4-2 Functions
STEP 1: Turn the Power Switch OFF on the Rear Panel, Slide the SET
switch to the set side.
STEP 2: Turn the power supply ON, the display will show blinking
SELECT.
STEP 3: Press TARE key and screen will show blinking FUNC and
blinking F000 afterward.
General Functions
F105 COM 8
0 Unstable
1 Error
SERIAL (RS-232)
F202 Parity
1 Same as display
2 GROSS Weight
3 NET Weight
4 TARE Weight
5 GROSS Weight, NET Weight, TARE Weight
Printer
Only Print
NET
3 above the
Weight
Latest data t
Print on all
TARE
4 according to
Weight
F001
GROSS
Weight,
NET
5 Print on all g
Weight,
TARE
Weight
GROSS
Weight,
NET
6 Print on all t
Weight,
TARE
Weight
Initial 1 2 1 0 1 1
BCD
F400 Data Type
1 Same as display
2 GROSS Weight
3 NET Weight
4 TARE Weight
Analog Output
F 500 Analog Output Data
1 Output 4~20 mA
2 Output 0~+10 V
4-3 CALIBRATION
1. Select FULL CALIBRATION
STEP 1: Turn the Power Switch OFF on the rear panel. Slide the SET
switch to the set side.
STEP 2: Turn the power switch ON. The screen will show blinking
SELECT.
STEP 3: Please press GROSS/NET key and a blinking CAL will show
on the screen. Then press the key.
STEP 4: The screen will show F-CAL. Please press the key.
4ZERO Adjust
The Screen will display ZERO . Please move the calibration
mass and objects away on the Weighing device then press
key. A display of ....... means finishing the Adjustment.
5SPAN Calibration
The screen will show SPAN . Press key and place your
calibration mass on the weighing device and input weight value.
Use the or key to set the available value, and the or key
to move through digits. Please press the key to finish the
calibration. The screen will show ........
STEP 1: Turn the Power Switch OFF on the rear panel. Slide the SET
switch to the set side.
STEP 2: Turn the power switch ON. The screen will show blinking
SELECT.
STEP 3: Please press GROSS/NET key and a blinking CAL will show
on the screen. Then press the key. The screen will show
(d-CAL).
STEP 4: The screen will show F-CAL. Please use the or key to
choose digital Calibration (d-CAL). Please press the key.
5ZERO Adjust
The Screen will display ZERO . Please move the calibration
mass and objects away on the Weighing device then press
key. A display of ....... means finishing the Adjustment.
6d-SPAN Calibration
The screen will show d-SPAN . Press key and place your
calibration mass on the weighing device and input weight value.
Use the or key and the or key to enter Load Cell O/P
Volt. Please press the key to finish d-SPAN. The screen will
show ........
STEP 1: Turn the Power Switch OFF on the rear panel. Slide the SET
switch to the set side.
STEP 2: Turn the power switch ON. The screen will show blinking
SELECT.
STEP 4: Use the or key to choose lock or unlock the key and press
key for confirmation.
STEP 5 : Slide the SET switch to the original side for finishing the step.
System Parameter:
includes functions FXXX, Calibration parameters, disable keys.
STEP 1: Turn the Power Switch OFF on the rear panel. Slide the SET
switch to the set side.
STEP 2: Turn the power switch ON. The screen will show blinking
SELECT.
STEP 3: Please press ACC/CLEAR key and a blinking COPY will show
on the screen. Please enter key.
STEP 5 : Slide the SET switch to the original side for finishing the step.
happens.
System Parameter:
includes functions FXXX, Calibration parameters, disable keys.
Set-point Parameter: includes Final, SP1, SP2, Free Fall, Hi, Lo.
STEP 1: Turn the Power Switch OFF on the rear panel. Slide the SET
switch to the set side.
STEP 2: Turn the power switch ON. The screen will show blinking
SELECT.
STEP 5 : Slide the SET switch to the original side for finishing the step.
STEP 1: Turn the Power Switch OFF on the rear panel. Slide the SET
switch to the OFF side (The SET switch remain in OFF status).
STEP 2: Turn the power switch ON. The screen will show Normal
operation condition.
STEP 3: Please press STANDBY key and hold SET POINT/ CODE key
at the same time until the screen shows C Lr cd .Please
release the STANDBY key ( SET POINT/CODE key still
holding). Please release the SET POINT/CODE key in
sequence. Please press the key and the JIF-2002 will
subsequently ask the operator to clear set point data.
STEP 4: Use the or key to choose NO or YES and press the key to
confirm.
STEP 1: Turn the Power Switch OFF on the Rear Panel, and slide SET
switch to he set side.
STEP 2: Turn the power switch ON. The screen will show blinking
SELECT.
STEP 5 : Slide the SET switch to the original side for finishing the step.
Calibration Errors
C.Err 7: The load cell signal pins are reversed, or the load cell output
voltage is too low.
Check the load cell connections if reversed or load cell failure.
C.Err 8: The load cell output voltage at maximum capacity is too high.
Check the load cell specification or load cell failure.
CURRENT LOOP
1 20 mA
0 0 mA
Pin Assignment
How to change Set point Code: After press SET POINT/ CODE key,
STEP 2: The screen will show Final. Please press the key, and the
screen will show 6-digit value with blinking 0 000.000. Use the
or key to change position and or key to change
value. Please press the key to confirm FINAL value.
STEP 4: The screen will show SP2, please press the key. The screen
will show 6-digit value with blinking 0 000.000. Please use the
or key to change position and the or key to
change value. Please press the key to confirm SP2 value.
STEP 5 : The screen will show Free, please press the key. The screen
will show 6-digit value with blinking 0 000.000. Please use the
or key to change position and the or key to
change value. Please press the key to confirm Free Fall
value.
STEP 6: The screen will show Hi, please press the key. The screen
will show 6-digit value with blinking 0 000.000. Please use the
or key to change position and the or key to
change value. Please press the key to confirm Hi (Over
Limit) value.
STEP 7 : The screen will show Lo, please press the key. The screen
will show 6-digit value with blinking 0 000.000. Please use the
or key to change position and the or key to
change value. Please press the key to confirm Lo (Under
Limit) value.
STEP 8: The screen will show CODE 00, please press the STANDBY
key to leave set point codes. If you wish to continue reset other
SET POINT CODES. Please enter values according to step1 to
step 8 again.
1. The Weighing Hopper is empty, the display shows "0", and all Gates are
closed. If the display is not at ZERO, input a TARE signal (Pin 24) to
re-ZERO the display.
3. When the display reaches "Final - SP 1", the SP 1 Output (Pin 12) signal will
come ON. Closed the Full-Flow Gate by using the SP 1 Output ON signal.
4. When the display reaches "Final - SP 2", the SP 2 Output (Pin 11) signal will
come ON.
6. After Free Fall has stopped - check if the HI and LO (Pin 9, 8) signals are
OFF. If both outputs are OFF then the batch is completed correctly.
8. Use the FREE (Pin 10) signal to delay a time period as the control signal is
processing empty the Weighing Hopper.
9. When the GROSS weight is below the ZERO band, the ZERO Band Output
will come ON -signifying the Weighing Hopper is empty. Closed the
Weighing Hopper Discharge Gate by using the ZERO Band (Pin 13) Output
ON signal.
10. You are now ready for your next batching event.
1. The Weighing Hopper is empty as is the Receiving Bin. The display shows
"0", and all Gates are closed.
3. When the GROSS Weight reaches "SP 1", the SP 1 Output (Pin 12) signal
will come ON. Closed the Supplying Bin Gate by using the SP 1 Output ON
signal.
4. The displayed weight will exceed the SP 1 value by the Free Fall value.
This weight is not necessarily accurate - but accuracy is not needed at this
moment since the purpose of this event is to fill up the Weighing Hopper.
The SP 1 value is always compared to GROSS weight.
6. Open the Full-Flow Gate and the Dribble-Flow Gate for Full-Flow filling into
the Receiving Bin.
7. When the display reaches "Final - SP 2", the SP 2 Output (Pin 11) signal will
come ON.
9. After Free Fall has stopped - check to see if the HI and LO (Pin 9, Pin 8)
signals are OFF. If both outputs are OFF then the batch is completed
correctly.
10. An Automatic Free Fall Compensation Command (Min. 200ms pulse to Pin
21) may be given at this time.
11. If the GROSS weight of the Weighing Hopper is below the ZERO Band
(Pin 13), the ZERO Band Output will be ON. The ZERO Band Output will
refill Weighing Hopper if needed.
1. The Weighing Hopper is empty, the display shows "0", and all Gates are
closed. If the display is not at ZERO, input a TARE signal (Pin 24) to
re-ZERO the display.
2. Check if the Weighing Hopper is empty using the ZERO Band Output
(Pin 13).
3. Input the Start signal via the Control I/O Interface connector (Pin 22).
When the Start signal is received, then SP 1, SP 2, and Free Output
signals will "come ON".
Note: When the Final Weight is 0, the Pin 12, 11 and 10 are kept OFF.
5. When the display reaches "Final - SP 1", the SP 1 Output (Pin 12) signal
will come OFF. Closed the Full-Flow Gate by using the SP 1 Output OFF
signal.
6. When the display reaches "Final - SP 2", the SP 2 Output (Pin 11) signal
will come OFF.
8. Batch Finish signal is sent after the set time period (F103) or when the
display is stable.
9. After Free Fall has stopped - check to see if the HI and LO (Pin 9, 8)
signals are OFF. If both outputs are OFF then the batch is completed
correctly.
10. Automatic Free Fall is now recalculated for the next event.
11. The Weighing Hopper Discharge Gate will be opened using the Finish
Output (Pin7) ON signal.
12. Data Output is sent (Auto print Mode: BCD, RS-232C, RS-422/485,
Printer or Current Loop). The NET Weight data will be accumulated.
14. If an Abort signal is sent (Pin 21) anytime after the Start signal is
received, then:
(1) SP 1, SP 2 and Free signals will go OFF, and Gates will be closed.
(2) Batch Finish and Data Output signals will be sent.
(3) NET Weight data will be accumulated.
1. The Weighing Hopper is empty as is the Receiving Bin. The display shows
"0", and all Gates are closed.
3. When the GROSS Weight reaches "SP 1", the SP 1 Output (Pin 12) signal
will come ON. Closed the Supplying Bin Gate by using the SP 1 Output ON
signal.
4. The displayed weight will exceed the SP 1 value by the Free Fall value.
This weight is not necessarily accurate - but accuracy is not needed at this
moment since the purpose of this event is to fill up the Weighing Hopper.
The SP 1 value is always compared to GROSS weight.
6. Input the Start signal via the Control I/O interface connector (Pin 22). When
the Start signal is received, the SP 2 and Free Outputs "come ON".
Note : When the Final Weight is 0, the Pin 11 and 10 are kept OFF .
7. Open the Full-Flow Gate and the Dribble-Flow Gate for Full-Flow filling into
the Receiving Bin.
9. When the display reaches "Final - FREE", the FREE Output (Pin 10) signal
will come OFF. Closed the Dribble-Flow Gate by using the FREE Output
OFF signal.
10. Batch Finish signal is sent after the set time period (F103) or when the
display is stable.
11. After Free Fall has stopped - check if the HI and LO (Pin 9, 8) signals are
OFF. If both outputs are OFF then the batch is completed correctly.
12. Automatic Free Fall is now recalculated for the next event.
13.The Weighing Hopper Discharge Gate will be opened using the Finish
Output (Pin 7) ON signal.
14. Data Output is sent (Auto print Mode: BCD, RS-232C, RS-422/485, Printer
or Current Loop). The NET Weight data will be accumulated.
15. Signal (Pin 13) will refill using ZERO Band Output if needed.
17. If an Abort signal is sent (Pin 21) anytime after the Start signal is received,
then:(1) SP 1, SP 2 and Free signals will go OFF, and Gates will be closed.
(2) Batch Finish and Data Output signals will be sent.
(3) NET Weight data will be accumulated.
1. The Weighing Hopper is empty, the display shows "0", and all Gates are
closed. If the display is not at ZERO, input a TARE signal (Pin 24) to
re-ZERO the display.
4. When the display reaches "Final - SP 1", the SP 1 Output (Pin 12) signal
will come ON. Closed the Full-Flow Gate by using the SP 1 Output ON
signal.
5. When the display reaches "Final - SP 2", the SP 2 Output (Pin 11) signal
will come ON. Closed the Medium-Flow Gate by using the SP 2 Output
ON signal.
8. After Free Fall has stopped - check to see if the HI and LO (Pin 9, 8)
signals are OFF. If both outputs are OFF then the batch is completed
correctly.
9. Please input a TARE signal (Pin 24) and set point code, then preparing
another substance batching.
10. Load the next substance into the Supplying Bin. Prepare the proper
Program, Operator Settings (if needed). Repeat Steps 3 ~ 8.
11. Use the FREE (Pin 10) signal to delay a time period as the control signal
is processing to empty the Weighing Hopper.
12. When the GROSS weight is below the ZERO band, the ZERO Band
Output will come ON which signifying the Weighing Hopper is empty.
Closed the Weighing Hopper Discharge Gate by using the ZERO Band
(Pin 13) Output ON signal.
13. You are now ready for your next batching event.
FREE FALL
Close the Weighing Hopper Dribble Flow Gate with six-digit value.
HIOver Limit
Please enter six-digit Over Limit Value NET weight > Final + Over Limit
Value
LOUnder Limit
Please enter six-digit Under Limit Value NET weight < Final - Under Limit
Value
CODE INPUT
Pin Pin Name Pin Pin Name
1 11 6 210
2 21 7 410
3 41 8 810
4 81 9 Common
5 110
Output
JIF-2002
Pin3.
Output 12V
Pin4
Pin 1.
Output Common
Pin2
Specifications
Type EIA-RS-232C 12V
Transmission Half Duplex, Asynchronus Transmission
Baud Rate 1200BPS2400BPS4800BPS9600BPS
Bit
8 bit non- parity 7 bit even parity, odd parity
Parity
Stop Bit 1 bit
Output Code ASC II
I/O Specifications
25 Pin D- Shape
HEADER 1
t Over Max. Capacity or under
O L
Min. Capacity
S T t STABLE
U S t UNSTABLE
HEADER 2
N T t NET
G S t GROSS
T R t TARE
UNIT
k G t Kilogram
l B t Pound
1 2 3 4 1 2 3 4 1 2 3 4 Cr Lf
y Free Fall y HI yLO
Command Format SA Cr Lf
0 0 3 4 5 6 Cr Lf
Zero Band Range
When Output at Full Scale is set at 36mA, and Output Current at Display
ZERO is set at 4mA, then at 1/2 Capacity (5000kg) the Output Current will be