Infusion Pump Te-172 Inspection Record: N o Check Item Result Yes No

Download as doc, pdf, or txt
Download as doc, pdf, or txt
You are on page 1of 2

INFUSION PUMP TE-172

INSPECTION RECORD

N
o
1
2
3

Check item

Is there any damage, etc in the main unit or pole clamp


Can it be operated with battery power
Is normal self-check is provided when the power switch
is ON
4
If the AC power cable is connected, does the BATTERY
indicator go ON
5
Was the [AC/DC] indicator lit when AC power was turned
on ?
6
With the door open, did the MIDPRESS finger section
move momentarily
7
If the door is closed and then opened. Is the tube clamp
closed
8
Does the START/STOP/SILENCE button function
normally?
9
Is the tube firmly clamped
10 Does the air-in-line sensor work normally
11 Was the Occlusion alarm activated in the following range
?
Upstream : - 30 ~ - 80 kPa ( - 0.3 ~ - 0.8 kgf/cm2 )
Downstream : 30 ~ 90 kPa ( 0.3 ~ 0.9 kgf/cm2 )
Setting on Pump:
Flow rate : 120 mL/h
Delivery Limit : 100 mL
Distance from pump to pressure gauge :
o 50 cm for Upstream Occlusion Sensor
o 100 cm for Downstream Occlusion Sensor
12 Was the [ FLOW RATE ] accuracy within the following
range ?
9-11 mL

Result
Yes
No

Value:
Upstream:
.........................
Downstream:
........................

Value:

Setting on Pump :
........................
Flow Rate : 300 mL/h
Delivery Limit : 5 mL
Test Time : 1 minutes
Remarks:
.......................................................................................

Serial
Number:

Date:

Checked by:

Signature: