ORTEC

With compliments

Helmut Singer Elektronik

DB463

www.helmut-singer.de info@helmut-singer.de
fon +49 241 155 315
fax +49 241 152 066
Feldchen 16-24 D-52070 Aachen Germany

Aligns fast-timing channels that


incorporate coincidence circuits or
TACs
Four independent sections
50- calibrated cable delay for
linear or logic signals
0 to 63.5-ns delay in 0.5-ns steps

The ORTEC Model DB463 furnishes a


50- calibrated cable delay, providing
relative delays from 0 to 63.5 ns with
0.5-ns increments in each of four
identical sections. Longer delays may be
achieved by cascading several Model
DB463 Delay Box sections.
The Model DB463 is convenient for
aligning fast-timing channels that
incorporate coincidence circuits or timeto-amplitude converters.

Delay Box

Specifications
INPUTS (4) 50 ; either polarity; 1500 V
maximum. BNC connectors.
OUTPUTS (4) 50 ; delay between In and
Out is sum of delays. BNC connectors.
DELAY TIMES 0.5, 1, 2, 4, 8, 16, and 32 ns.
DELAY ACCURACY <0.1 ns or 1.0% for
each switch, whichever is greater.
REFLECTIONS <3% at any delay setting for
1-ns rise time step.
CABLE RG-58A/U.
WEIGHT
Shipping 4.0 kg (9 lb).
DIMENSIONS 48.2 cm (19 in.) wide for relay
rack mounting, 8.9 cm (3.5 in.) high, and
20.3 cm (8 in.) deep.

Ordering Information
To order, specify:
Model

Description

DB463

Delay Box

Fig. 1. Typical Schematic for One Section of Model DB463 (Four Sections Included).