Manuale HPP108
Manuale HPP108
Manuale HPP108
IPP 200-500
Cooled incubator with Peltier technology
HPP 108
Constant Climate Chamber with humidification
Contents
1.
1.1
1.2
1.3
1.4
4
4
4
4
14.
15.
2.
Safety regulations
3.
15.1
15.2
15.3
15.4
4.
4.1
4.2
Overview
Construction
Control unit
8
8
10
5.
5.1
5.2
Removing covers
Removing top panel
Removing the rear panel
11
11
12
6.
6.1
Mainboard
Check switching function of
mainboard
Replace fuses
Replace mainboard
13
6.2
6.3
13
14
14
7.
8.
9.
Replace controller
17
10.
18
11.
19
12.
21
13.
Replacing a Peltier
cooling/heating module
23
15
25
25
25
26
26
26
27
28
30
30
30
18.
Humidification and/or
dehumidification does
not work properly
33
18.1 Humidification is not
working properly
33
18.1.1 Replac steam generator
35
18.1.2 Replacing the pump
36
18.2 Dehumidification is not
working properly
36
18.2.1 Replace fan and dehumidification Peltier module
37
18.2.2 Replace the dehumidification power supply unit 38
18.3 Replace control module
39
18.4 Replacing the humidity sensor 40
19.
19.1
19.2
19.3
19.4
Circuit diagrams
Legend
IPP 200/300/400
IPP 500
HPP 108
42
42
43
44
45
2. Safety regulations
Danger!
After removing covers, voltage-carrying parts
may be exposed if the mains plug has not
been disconnected. Any work that involves
opening the appliance may only be performed
by trained electricians who are familiar with
the safety regulations for electrical engineering.
Caution!
When dismantling the appliance you could be
injured on sharp (unprocessed) metal edges.
You should wear gloves.
Please also observe the safety regulations in
the operating instructions for the respective
appliance.
After all maintenance or repair work, the appliance must be subject to a function test. The appliance must not be left unattended
until it has reached the steady state.
If the fault or malfunction cannot be rectified by the repair work:
Contact the manufacturer.
Cause of error
Rectifying errors
Page
Fuse 100 mA
(200 mA/115 V) faulty
14
14
13
General faults
Controller display does
not light up
Mainboard faulty
Replace mainboard
Temperature cannot be
adjusted and
appliance can in general
not be operated with the
push/turn control
16
21
15
19
15
15
15
Replace mainboard
13
15
Increase temperature
difference between
monitoring temperature and
working temperature.
Replace Pt100 temperature
sensor of monitoring
controller if necessary
18
Error description
Cause of error
Rectifying errors
Page
18
Replace controller
17
17
13
18
Replace mainboard
13
Mainboard faulty
Replace mainboard
13
Malfunctions of the humidifier and/or dehumidifier (only for the constant climate chambers HPP 108)
Neither humidification nor dehumidification are working properly
33
33
36
4.
Overview
4.1
Construction
1
9
8
10
11
5
4
6
Fig. 1: Construction (shown in figure: Maximum fittings for HPP 108; the actual fittings may vary, depending
on the appliance type)
Item Designation
Top panel
11
Rear panel
12
Right-side panel
11
Door
26
25
19
19
Left-side panel
10
11
11
21
37
1
21
only for types IPP 500 and HPP 108; types IPP 200-400 have just one interior fan and one Peltier heating/
cooling module on the rear side
4.2
Control unit
5 6
10
10
5.
Removing covers
For some service work it is necessary to remove case panels (top,
side, rear panels). How these panels are removed is described
below. What panel(s) need(s) to be removed for the respective
service work is described in the relevant chapter.
5.1
Undo the 4 screws to the left and right on the rear side of
the appliance holding down the top panel (1).
Slightly raise the panel and pull it out backwards (2).
Removing the
side panels
You can now also remove the left and/or right side panel. They
are not screwed down, and can simply be pulled out by pulling
upwards (3). Make sure when doing this that they do not jam,
and disconnect the PE conductor connection.
Insert side panels carefully from above into the guide rails and
push downwards until they slot into place. Make sure that the
side panels do not jam when doing this. Reconnect the PE conductor.
Put on the top panel and push forwards into the support. Insert
the screws with retaining washers into the openings in the rear
panel and screw down the top panel (1).
The retaining washers are necessary to create a safe PE conductor
connection with the top panel.
11
5.2
4.
1
2
1.
2.
12
Insert the rear panel carefully from above into the guide rails
and push downwards until it slots into place. Make sure that
the rear panel does not bend when doing this.
Screw on the grid covers of the Peltier modules on the right
and left (1).
In addition for the constant climate chambers HPP 108:
Screw on the cover grid of the dehumidification Peltier module and of the steam generator in the centre (2).
6.
Mainboard
8
7
6
1 2
5
3
4
Fig. 6: Mainboard (see also control unit overview on page 10)
1
2
3
4
5
6
7
8
6.1
13
6.2
Replace fuses
The fuses for controller/display and appliance protection are located on the mainboard (see page13).
1. Disconnect the mains plug.
2. Remove the top panel (see page 11).
3. Measure the fuse resistance with a continuity tester. If the
fuse is in order the resistance should only be a few Ohms.
4. To replace the fuse, push an insulated flat screwdriver beneath the side of the fuse and lever it out of the mounting
bracket. Press the new fuse into the mounting bracket from
above.
Appliance fuse:
15 A quick-blow 32 x 6.3 mm
(American)
Fuse for controller/display:
100 mA slow-blowing
20 x 5 mm
200 mA slow-blowing
20 x 5 mm (115V)
Spare part
Variants
Appliance fuse 15 A
Fuse for controller/display
6.3
E02387
100 mA/230 V E02497
200 mA/115 V E01007
Replace mainboard
1.
2.
3.
4.
5.
6.
7.
8.
9.
Spare part
Mainboard
14
Variants
230 V
B04098
115 V
B04113
7.
2
2
Connections
Screws
Spare part
E04016
15
8.
4
3
Fig. 8: Main switch module (see also control unit overview on page 10)
1
2
3
4
16
Push/turn control
Mounting bolts (left and right)
Main switch
Plug (controller slot J2)
Spare part
Variants
230 V
B04425
9.
Replace controller
1.
2.
3.
4.
J13 2
J4 J3 J11 3
J2
J9
2 J8
Fig. 9: Replace controller (see also control unit overview on page 10)
Spare part
Controller RS 232
Controller RS 485
Variants
IPP 200/300/400
B04279
B04281
IPP 200/300/400
B04280
B04282
17
3
1,2
1
2
Fig. 10: Replace temperature sensor (see also control unit overview on page 10)
1 Temperature sensor for operating controller
2 Temperature sensor for monitoring sensor
3 Clamp
4 Feed-through through the control unit floor/chamber ceiling
5 Cable connection to the controller in the control unit
Spare part
E04822
18
4
2
19
Spare part
Variants
all
B04004
IPP 200/300
E03995
IPP 400
E03994
IPP 500
E04656
Fan cover
HPP 108
20
left
E05401
right
E05402
21
6
5
Fig. 12: Remove the fan from the rear panel (actual view may differ from this representation, depending on the
appliance type)
1
2
3
4
5
6
Spare part
Cooling fan 120 mm
22
Variants
IPP 200/300
B04002
IPP 400/500/HPP108
B04423
Peltier module
Variants
IPP 200/300
E04002
IPP 400
E03986
E04641
IPP 200/300
E03973
E03972
E05231
HPP 108
E03972
23
1
6
24
Connection cable
Plastic wedge
Foam ring (cable seal)
Loosen screws (1); do not remove screws completely, as otherwise the mounting block (inside) will fall out.
2. Check the screw (2) in the hinge and tighten if necessary.
3. Adjust the inner door (4) and tighten screws (1) again.
If a mounting block has fallen out, you can reach it by pulling out
the side section and then mount it again.
1
2
3
4
Screws
Screw
Inner door hinge
Inner door
Outer door (open)
Spare part
Variants
IPP 200
B03395
IPP 300
B03396
IPP 400
B03397
B03398
all
B02902
all
B02903
Inner door
25
Open door 180SDgr and undo screws (4) at the two eccentric
sockets (1) at the top and bottom.
2.
Grease the moving parts of the door hinges with thin silicone
grease (see also operating instructions "Maintenance").
Inset a hard paper disk on the lower door hinge between the
eccentric tappet (3) and door (5).
Insert the door. When doing this, make sure that the adjuster
cam of the eccentric tappet (3) is pointing forwards (1a).
Put screws back in (4) and tighten up.
1b
1a
1
2
3
5
Fig. 16: Installing/removing door
1
2
3
4
5
26
Eccentric socket
1a Eccentric tappet moved fowards: factory settings for door
1b Eccentric tappet moved to the right: Doors are therefore also adjusted to the right
Allen set screw
Eccentric tappet
Screws
Door
1
2
3
4
5
6
7
Screw
Locking plate
27
2
3
28
Door
Door panel
Door seal
Locking mechanism
Lock cover
Screws
9 10 11
12
Door-handle cover
Circlip (large)
Door-handle
Guide screw
Sleeve
Locking bracket
Locking pin
Locking bolt
Locking arm (brass)
Circlip (small)
Axles
Tension spring
Spare part
Model
IPP 200
B00408
IPP 300
B00430
IPP 400
B00432
B00444
all
B03749
all
B00995
all
B02600
IPP 200
B02419
IPP 300
B02603
IPP 400
B02420
B02421
all
B02720
Door
Locking plate
29
Screw
Lock cover
Door seal
Door panel
1
2
30
Chamber seal
Cage seal
Seal cross-section
Spare part
Door seal
Cage seal
Chamber seal
Model
IPP 200
Z00691
IPP 300
Z00692
IPP 400
Z00693
Z00694
IPP 200
Z00713
IPP 300
Z00714
IPP 400
Z00715
Z00716
IPP 200
Z00699
IPP 300
Z00700
IPP 400
Z00701
Z00702
31
4.
5.
6.
7.
32
Plug
Screw connection
Spare part
B04118
Ethernet module
B04121
1
3
Fig. 24: Checking the tube connections
1
2
3
Water tank/canister
Tube connections
Steam generator
33
2 3
Pump
Control module
Fuse for control module
Power supply unit
2.
3.
4.
5.
34
2
3
1
1
Steam generator
Tube connections
Screw connections (four screws)
Spare part
Steam generator
B04436
35
2
Fig. 27: Replace the pump
1
2
3
Remove hoses
Undo plug connector
Undo screws
Spare part
Pump
E04504
36
If the fan is running, but the humidity inside the appliance does
not fall, you should proceed as follows:
1. Remove the top panel (see page 11).
2. Check that all plugs in the control unit are connected.
3. Check that the fuse for the control module is OK (see page
34). If not: replace the fuse.
4. Check that the green LED on the power supply unit lights up
for dehumidification (see page 34). If not: Replace the power
supply unit (see page 38).
5. If no errors can be found, either in the fan or the control
module or power supply unit, the dehumidification Peltier
module with the fan must be replaced (see next chapter).
18.2.1 Replace fan and dehumidification Peltier module
1.
2.
3.
4.
1
2
Plug connection
Screw connections at the top and bottom (2 screws)
Spare part
37
2
Fig. 29: Replace the power supply unit for dehumidification
1
2
3
38
Spare part
E05400
Control module
Screw connection (four screws)
Spare part
Control module
B04376
39
40
3
1
Fig. 31: Replacing the humidity sensor
1
2
3
4
5
Humidity sensor in the insulation material on the rear panel of the appliance
Connection cable in the control unit
Allen set screw
Undo the Allen set screw with the 1.5 mm Allen key through the opening in
the insulation material on the side wall
Pull the humidity sensor out and upwards through the opening in the floor
of the control unit
Spare part
Humidity sensor
E05249
41
42
Component
drawing
Component
B1
B2
J15
N1
N2
Mainboard 55162
N5
Chipcard reader
N7+N9
N8+N10
S1
Main switch
X1
X2
43
44
45
Notes:
Memmert GmbH + Co. KG | Postfach 1720 | D-91107 Schwabach | Tel. +49 (0) 9122 / 925 - 0 | Fax +49 (0) 9122 / 145 85 | E-Mail: service@memmert.com | www.memmert.com