Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Cro

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

1 Cathode Ray Oscilloscope

CRO
Cathode Ray
Oscilloscope
revised April 21, 2006

(You will do two experiments; this one
and the Charge-to-Mass Ratio of the
Electron experiment. Sections will switch
rooms and experiments half-way through the
lab.)
Learning Objectives:
During this lab, you will
1. learn how to measure time-varying
electronic signals with a cathode ray
oscilloscope (CRO).
2. estimate the uncertainty in measure-
ments made with a CRO and estimate
the uncertainty in a quantities that are
calculated from quantities that are
uncertain.

A. Introduction
The cathode ray oscilloscope, CRO or
simply scope, is used in many fields of basic
and applied research and in electronics
development and repair. It is generally the
tool of choice for examining signals that
change with time on a scale of 1 second to
1 nanosecond. The sketch of the scope in
Figure 1 includes a triangular wave signal, a
voltage that, as a function of time, continu-
ally (and linearly) ramps up and down
between two limiting voltages. The operation
of the CRO is described in detail in
Appendix IX. You should read that appendix
before attempting this lab.
In this lab, you will learn how to use
the CRO to investigate various types of time-
dependant phenomena that you will encoun-
ter in your studies of electricity and magnet-
ism. You should also become more comfort-
able with certain aspects of sine waves, such
as phase differences and the relation of fre-
quency to period, that are critical to under-
standing interference effects and electro-
magnetic radiation.
This experiment requires that you
complete the worksheet worth 30 points that
can be found in Appendix XI. Attach the
worksheet and graphs from Lab #3A to the
paper from Lab #3B and use one cover sheet
for both.

B. Apparatus
You will be using a dual-trace oscillo-
scope, a special scope probe designed to
Figure 1: Cathode ray oscilloscope controls.

Cathode Ray Oscilloscope 2
mate with the scope, a doorbell trans-
former, function generator, microphone and
tuning forks.
Figure 1 is a diagram of the front panel
of the scope, with labels for the important
controls. A photograph of the scope is posted
on the lab web site.
The scope probe resembles a pen with
an alligator clip attached to it by a short wire.
The alligator clip is for the ground connec-
tion and is not needed when the scope probe
is connected to ground-referenced elec-
tronics but is used to establish the ground of
other objects you may be measuring. (DO
NOT connect this alligator clip to a signal
output; it will cause a short circuit!)
The signal connection of the scope
probe is a spring loaded hook located inside
the tip of the probe. This hook is exposed by
retracting its cover (do not unscrew or
remove the cover). Some but not all probes
let you switch between 1X and 10X where
the 10X divides the signal by a factor of 10.
If your probe has this option, be certain to
use the 1X setting.

C. Familiarization and Use
Turn the POWER switch on and leave
it on for the entire lab period. Electronic
devices can produce a lot of waste heat and
their properties can change with temperature;
sensitive instruments are generally left on so
that they give more stable readings.
These are dual trace oscilloscopes (they
have two quasi-independent inputs), so you
must select channel 1 [CH 1 (16)] for the
following measurements. (The numbers in
parentheses refer to the control locations
shown in Figure 1 and are described in
Appendix IX.) Set the SOURCE (21) selec-
tion switch to INT (for internal triggering),
the MODE(25) to AUTO, the MODE (16)
switch to CH1 and the INT TRIG (22) selec-
tion switch to CH1.
The scope is usually used to plot a
changing voltage as a function of time, with
the instantaneous voltage read along the ver-
tical or y-axis while time is measured along
the horizontal or x-axis. The scope face has a
measurement grid that is 8 cm tall by 10 cm
long. The grid lines are referred to as divi-
sions or DIV. Some of the DIV lines have
short markings every 2 mm or 0.2 DIV.
The horizontal sweep control (18/19)
sets the time it takes the scope beam to scan
across the screen horizontally. This control is
labeled as TIME/DIV where TIME may be
measured in seconds (s), milliseconds (ms)
or microseconds (s) depending on the posi-
tion of this control.
The vertical gain controls (10/12 and
11/13) are labeled as VOLTS/DIV. These
control the amplification of the signal or how
large a given signal appears relative to the
vertical or y-axis. The units may be
volts/DIV or millivolts/DIV (mV) depending
on where these knobs are set.
Be sure that the sweep (18/19) and
vertical gain(10/12&11/13) controls are
kept locked in their calibrated positions at
all times. Each of these controls has two
concentric knobs, an inner knob that is
painted red on most of the scopes and an
outer knob. The inner knob lets you vary the
setting continuously but means that you no
longer have a quantitative reading. The outer
knob has calibrated click stops. Dont use the
continuously variable inner knob to make
any of the adjustments described in this
manual; keep them turned fully clockwise
onto their clickstops. You should only use
the outer companion controls which have
discrete, calibrated settings. Otherwise, your
measurements will be meaningless.
Turn the TIME/DIV(18) knob to the
slowest sweep possible (fully counterclock-
wise, CCW). With nothing connected to the
scope inputs, the input voltage is effectively


3 Cathode Ray Oscilloscope
set to zero and the scope will plot zero as a
function of time, a flat line. You should see a
spot moving across the screen. If you dont
see this spot, try adjusting the vertical posi-
tion (14) which should be somewhere near
its midpoint. If this doesnt work, ask a TA
for help. Note what happens as you now
increase the sweep frequency by turning the
TIME/DIV knob clockwise. This increases
the speed as which the scopes electron beam
scans across the face of the scope, letting
you follow signals that themselves are
changing faster, i.e. are of a higher
frequency.
C.1. Square Wave - Time and Voltage
Measurement
The scope has a CAL(26) output tab
which supplies a 0.5 V
peak-to-peak
(V
pp
), 1 kHz
square wave signal. The term peak-to-peak
means that the signal is measured from its
absolute maximum to its absolute minimum,
or top to bottom. This CAL signal can be
used to check the calibration of the scope
settings, but we will actually be using it to
check whether you know how to make
proper measurements with the scope. If the
measurements you are instructed to make
below do not agree with the expected values,
ask for help.
Use the scope probe to connect the
CAL tab to the CH1(6) input of the oscillo-
scope. DONT USE THE ALLIGATOR
CLIP on the probe to connect to this tab!
This shorts it out. Use the hook inside the
retractable tip. Connect the BNC plug on the
other end of the scope probe to the scopes
Channel 1 input. This input is a BNC jack, a
common form of coaxial connector. Align
the slots of the BNC plug of the scope probe
to the pins of the BNC jack on the scope.
Push the plug into place and rotate it 90E to
lock it in position. Set the AC-GND-DC (8)
switch to DC. The TIME/DIV (18) should be
at about 0.5 ms while the channel 1
VOLTS/DIV (10) should be at about 0.2
V/div. These settings will let you view the
signal but you should expect to change them
to optimize your measurements.
Set the trigger LEVEL (24) control to
the full CW (+) position and then decrease
and adjust this knob until you get a stable
image. Setting this triggering is often the
trickiest part of using an oscilloscope. Once
you have a stable image, adjust the Time/Div
switch until you see a sequence of a few
cycles of a square wave on the screen.
Adjust the vertical gain (VOLTS/DIV) and
vertical position so that the signal almost
fills the screen vertically.
You will have to estimate the accuracy
of many of your measurements. You may
assume that any errors in the scope elec-
tronics are negligible and that the only errors
are due to your ability to judge the position
of a signal on the screen. For some arbitrary
signal, how well do you think you can
determine its position, in terms of either mm,
cm, or DIV (your choice of units)? Later,
when you need to convert your estimate of
this error into an error in time or voltage, just
multiply by the setting of the TIME/DIV
knob or VOLTS/DIV knob respectively.
C.1.1. Time Measurement
The period of the square wave, as for
any repetitive wave, is the time it takes to
repeat itself. Measure the period of the cali-
bration square wave by multiplying the
length, in cm or DIV, of one or more periods
times the setting of the TIME/DIV knob.
(Measure as large an image as possible to
obtain the highest possible precision in your
time measurements. If there are 4 full peri-
ods on the screen, measure the time for all 4
and divide by 4 to get the period; in this case
you will also have to divide your error esti-
mate by 4. Alternately, change the
TIME/DIV so that 1-2 periods fill your
screen.) You can shift the signal horizontally

Cathode Ray Oscilloscope 4
using the x-position control (20) to start or
end the sweep at some convenient mark on
the CRT. Use the period you measured to
calculate the frequency (and estimated error)
of the calibration signal. Remember that fre-
quency is just one over the period. To find
the error in the frequency, you should use the
derivative method, (1/T) = (T)/T
2
.
C.1.2. Voltage Measurement
Determine the peak-to-peak voltage of
the square wave by multiplying the measured
height of the square wave by the setting of
the VOLTS/DIV. Note that you can offset
the signal (14) to make this measurement
easier. Remember too that more careful
measurements can be made if you adjust the
calibrated VOLTS/DIV knob so that the sig-
nal almost fills the screen vertically. Com-
pare this result with the expected value of
0.5 V
pp
.
C.2 Measurements of a Sine Wave
You have been looking at a square
wave. Another common signal is a sine wave
[V = V
0
sin(t + )] such as the 110-volt AC
power line. You will use a doorbell trans-
former to reduce the signal to a safer level.
Connect the center and either one of
the two outer terminals of the transformer to
the CH 1 input of the oscilloscope. Adjust
the sweep time, the vertical sensitivity and
other controls until you get a stable picture.
(It probably wont be a very good sine wave,
but thats what you often have at an outlet.)
Sketch the waveform (dont forget to label
the scales on your sketch). Measure and
record the period and calculate the fre-
quency. Measure and record the peak-to-
peak voltage of the signal.
Use a DMM set to measure AC volt-
ages to check the voltage output of the trans-
former. Are your CRO measurements of the
transformer voltage consistent with the
DMM measurements? This probably wont
appear to be the case at first, the meter
should read less than half of your scope
measurement. One reason for this is that
your oscilloscope measurement was a
peak-to-peak voltage. This is twice the
amplitude of the sine wave, the V
0
term the
equation V
0
sin(t + ). Another reason is
that the DMM measures the RMS (root
mean square) voltage, given by

( )
2
0
cos
2
RMS
t dt
V
= < >=
V V
dt
e } (

}
(1)
where the integrals are over one period. V
RMS

is more closely related to the strength of a
signal than is V
0
; although the two are the
same for a DC signal, they vary significantly
for various types of AC signals. The integral
of cos
2
(t)dt divided by the integral of time,
over any number of whole periods, is just .
V
rms
is proportional to the square root of this
factor or the square root of . Taken all
together, V
pp
= 2 2 V
rms
. Knowing this, are
your scope and DMM measurements con-
sistent?
You can view a cleaner sine wave
using a function generator. Connect the
HIGH output of your function generator to
channel 2 of your scope, leaving the trans-
former connected to channel 1. Switch the
scope to CHOP MODE (16) to view both
channels 1 and 2 but set the trigger (22) to
channel 2. Turn on the function generator
and set it to produce sine waves at about 60
Hz with a magnitude similar to that of the
transformer. The controls of your function
generator are poorly marked and calibrated
and should not be trusted to be accurate. Use
your scope to determine that you have the
settings right. To do this, adjust the scope so
that the channel 1 signal appears on the top
half of the screen while channel 2 occupies
the bottom half. The amplitudes and periods
of the two signals should be comparable. If
necessary, adjust the scope and function
generator settings. Try adjusting the trigger


5 Cathode Ray Oscilloscope
level (24) to see how it shifts the point at
which the trace starts. Change the trigger
source to channel 1 to see the effect.
C.3. Lissajous Figures
An alternate method for comparing two
signals is to plot one on the horizontal (X)
and the other on the on the vertical (Y) axis.
This will produce Lissajous figures which
allow for very quick visual comparison of
the relative frequency and phase of two sig-
nals (but which are rarely used except as
special effects in science fiction movies).
Set the oscilloscope to operate in X-Y
mode (TIME/DIV knob (18) fully CW) with
the output of the transformer still connected
to the x-axis (CH1) and the sine-wave output
of the function generator, set to 60 Hz, con-
nected to the y-axis (CH2). Adjust the fre-
quency (slightly) and output amplitude of the
function generator until you repeatedly see a
diagonal straight line at an angle of roughly
45 with the horizontal. Because of changing
phases, the signal will slowly drift from a
line to a circular shape and back again.
To understand what is happening, think
back to your high school trigonometry
course. If someone told you to make an xy
plot as a function of time of a signal given by
x = Acost and y = Acost, where A is some
arbitrary amplitude, is an angular fre-
quency and t is time, hopefully you can see
that you would just trace out a 45 line, since
x = y at every instant of time. What changes
if x = Acost and y = Asint? Now you
would sketch out a circle. Whats the differ-
ence? Its just the phase difference between
the x and y signals, since the sin function is
just the cos function shifted by 90. What
happens if the phase difference between x
and y slowly changes with time? The pattern
slowly drifts from a line to a circle and back.
Sketch the pattern you observe on the scope
at a few representative times as it changes.
Switch the oscilloscope back to 2 ms
TIME/DIV, chop mode and note the gradual
phase change of the oscillator signal relative
to the transformer signal. The signal on
which the scope is triggered should remain
steady while the other sine wave gradually
drifts to the left or right. The speed of this
drift corresponds to the rate at which your
Lissajous pattern changes shape. Try adjust-
ing the frequency to make the drift larger or
smaller, switching quickly back and forth
between XY mode and a 2 ms sweep setting
to observe the corresponding effect on the
Lissajous pattern.
Next, slowly increase the frequency of
the signal from the function generator until it
is approximately doubled, then fine-tune it to
produce a fairly stable Lissajous pattern.
Measure the frequency with the scope, con-
firm that it is roughly 120 Hz and sketch this
Lissajous pattern. (Its a lot harder to
explain the shape of a Lissajous pattern
when the frequencies of the two signals are
different but the drift in the pattern is still
related to a drift in their relative phases.)
Vary the frequency between 60 - 120
Hz and locate the simplest, nearly stable
pattern in this range. Measure this frequency
and sketch the Lissajous pattern. (In princi-
ple, there are an infinite number of patterns
in this range of frequencies but one should
stand out as simpler than any others.)
Try to analyze the experiment and/or
the theory to determine what conditions on
the frequencies are necessary for a relatively
simple, stable pattern to appear, ignoring the
drifts caused by slowly varying phases.
C.4. Sound Waves
Disconnect all of the input leads from
the scope and set the sweep to 1 msec/div.
Attach a microphone to the channel 1 input.
Use AC coupling and a high sensitivity (the
signal will be small) and set the trigger
source to channel 1. Find the frequency of a

Cathode Ray Oscilloscope 6
tuning fork struck with the rubber end of
the mallet supplied to you. (Change the
sweep speed and vertical gain as necessary
to make a good measurement.) Note that the
loudest sound comes from the opening in the
sounding box, not directly from the vibrating
metal tines.
Borrow another fork from a neighbor,
strike both simultaneously and try to hear
beats and see them on the scope (a definition
of beats is given below). Since the beats will
be at a lower frequency than the signal from
a single tuning fork, you will have to slow
your scopes sweep speed by one or two
clicks of the TIME/DIV knob to see them.
(There are a few pairs of tuning forks that
produce particularly clear beat patterns.
These are marked with matching colored
squares or circles.)
Beats are the phenomenon that two
sine waves of similar frequencies add to pro-
duce a signal that looks like a sine wave
whose frequency is the average of the origi-
nal sine waves with an overall modulation at
a frequency given by half the difference in
the original signals (Fig. 2). For example, if
you add a 1000 Hz tone to a 1060 Hz tone,
you will produce a 1030 Hz tone that
increases and decreases in magnitude at 30
Hz. The 30 Hz is sometimes called an enve-
lope that modulates the amplitude of the
1030 Hz signal. (The AM in AM radio refers
to a similar modulation.)
If you have time, you may wish to
investigate the sound of your own voice
using the microphone and scope, although
this is not required for the worksheet. What
is the frequency of your speaking voice?
What are the highest and lowest frequency
tones you can vocalize by singing, humming,
etc.? Can you match the tone of your tuning
fork? Can you and your lab partner(s) sing in
harmony and produce beats? (Only the BEST
lab partners can do this!)

Figure 2: Example of beats.

You might also like