SM06FEN - 0908 VibrationDevice

Download as pdf or txt
Download as pdf or txt
You are on page 1of 32

Service-Info

Liquiphant M / Liquiphant S
Level limit switch FTL50, FTL51, FTL51C, FTL50H, FTL51H FTL70, FTL71

FTL50/ FTL50H

FTL51/ FTL51H/ FTL51C FTL70

FTL71

SM06F/00/en/09.08 Release: 5

SM06F/00/en/09.08

Liquiphant M/S

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Table of contents General information


Safety hints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Product family Liquiphant M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Product family Liquiphant S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Function principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Electronics and wiring


Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 The electronics of Liquiphant M/S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Electronic inserts FEL5x, connection data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Test circuit for PFM devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Exchange of assemblies
Tooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Disassembly of Liquiphant M/S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Assembly of Liquiphant M/S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Exchange of electronic insert for housing T13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Operation
Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Sensor signals at the plug connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Trouble-shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Spare parts
Spare parts Liquiphant M/S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Supplementary documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Endress+Hauser Maulburg

SM06F/00/en/09.08

Liquiphant M/S

Safety hints
For measuring systems used in hazardous areas, the legal requirements have to be observed! Ensure that all personnel are suitably qualified. Observe the specifications in the certificate as well as national and local standards and regulations. ATTENTION: To avoid electrostatic discharge damage to the device, make sure you are well grounded before touching the electronics.

Device certified for use in explosion hazardous area

0
Advice for repair Spare parts

If the device has this symbol embossed on its name plate it can be installed in an explosion hazardous area.

The instrument may only be maintained and repaired by qualified personnel. The device documentation, applicable standards and legal requirements as well as any certificates have to be observed!

Only modular assemblies may be exchanged against identical, original Endress+Hauser spare parts! Cable glands for Ex-devices may only be replaced with the same type (observe marking on packaging and permissible temperature range, T>75 C).

Test

Ex d-devices:

Attention has to be paid not to damage the threads. In other case, the part has to be exchanged. The repair has to be performed such, that the voltage isolation of the Ex ia circuits against ground is maintained. If required, a test can be performed with 500 Veff over a time period of 60 s.

Ex i-devices:

Documentation

Any repair of a certified device has to be documented! This includes stating the device serial number, date of repair, type of repair and name of the repair technician.

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

The product family of Liquiphant M FTL5x


The Liquiphant M is a level limit switch for applications in all kind of liquids. Small process connections and dimensions allow an operation in pipes with diameters as small as DN50. Liquiphant M is also available in compact housing design. This version can not be repaired because it is filled with Silgel and welded. It is not specified in this Service Manual.
Electronic inserts: FEL51 FEL52 FEL54 FEL55 FEL56 FEL58 FEL57 FEL50A

Housing F15

Housing F16

Housing F17

Housing F13

Housing T13

FTL50
Compact design as example with thread

FTL50
Compact design as example with thread

FTL50/51H
Compact and extension pipe with hygienic process connections

FTL51/51C
Extension pipe without/ with coating as example with flange

FTL50/51H
Compact and extension pipe with hygienic process connections

FTL51/51C
Extension pipe without/ with coating as example with flange

General technical data


Type Design Sensor length Process temperature Process pressure Density Viscosity Solid content -1 ... +100 bar 0,5 g/cm3, 0,7 g/cm3 FTL50 / FTL51 Compact / Extension pipe FTL50H / FTL51H Compact / Extension pipe in hygienic design FTL51C Extension pipe with coating 50,5 ... 3000 mm (depending on coating) -50 ... +150 C (230 C optional) -1 ... +64 bar

50,5 ... 6000 mm -50 ... +150 C

max. 10.000 mm2/s max. 5 mm

Endress+Hauser Maulburg

SM06F/00/en/09.08

Liquiphant M/S

The product family of Liquiphant S FTL7x


The Liquiphant S is a level limit switch for high process temperatures and pressures in all kind of liquids.
Electronic inserts: FEL51 FEL52 FEL54 FEL55 FEL56 FEL58 FEL57 FEL50A

Housing F16

Housing F17

Housing F13

Housing T13

FTL70
Compact design as example with thread

FTL70
Compact design as example with thread

FTL71
Extension pipe as example with flange

FTL71
Extension pipe as example with flange

General technical data


Type Design Sensor length Process temperatures Process pressures Density Viscosity Solid content FTL70 Compact 50.5 ... 69 mm (depending on process connection) FTL71 Extension pipe 148 ... 3000 mm (depending on process connection)

-60 ... +280 C -1 ... +100 bar 0,5 g/cm3 max. 10.000 mm2/s max. 5 mm

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Operation principle of the Liquiphant M/S


The sensor, shaped like a tuning fork, is excited in its resonance frequency (oscillation circuitry) by means of a piezo driver. An immersion into a liquid medium causes the resonance frequency to decrease. This frequency change is evaluated in the CPU and transmitted to the output stage as a binary signal via an opto-coupler (galvanical separation). An external load like a relay, PLC or separate switching device can then generate a safety reaction. The limit switch can be set to either minimum or maximum failsafe mode, depending on the application. In case of a power fail, the electronic switch will be "open" or a relay will be de-energized.

Block diagram (for example FEL51)


Power indicator (green)

DC DC

+5 V

Switching status (red/yellow)

L1

Oscillation circuitry

CPU
2 Min/ Max Density Diagnostic Output Oscillation frequency Switch signal w/o delay Identifier GND This part is different for electronics FEL52 FEL54 FEL55 FEL56 FEL57 FEL58

EEPROM (inside)
Tuning fork

Endress+Hauser Maulburg

SM06F/00/en/09.08

Liquiphant M/S

The electronics of Liquiphant M/S


Nine electronics are available for Liquiphant M and Liquiphant S. They are used depending on the application and safety requirements (certificates). FEL51: 2-wire AC connection 19...253 V AC FEL52: 3-wire DC connection 10...55 V DC FEL54: Universal current connection 19...253 V AC or 19...55 V DC with relay output FEL55: 2-wire connection for separate switching unit e.g. PLC FEL56: 2-wire connection for separate switch amplifier according to NAMUR (L-H edge) FEL58: 2-wire connection for separate switch amplifier according to NAMUR (H-L edge) FEL57: 2-wire connection, PFM signal for separate switching unit Nivotester FEL50A: 2-wire connection for power supply and data transfer, connection to PROFIBUS PA FEL50D: 2-wire connection for connecting to the density and concentration computer FML621. ATTENTION: This electronic insert cannot be installed in devices that were originally used as a limit switch.

Electronic inserts FEL5x, connection data


The illustrations show the functional wiring in an installation. Hints regarding the applicable national or international installation guidelines for explosion hazardous areas must be observed. It is mandatory to take into account the grounding concept used by the customer. Connection data (load) FEL51 Relays with a minimum holding power/rated power >2.5 VA at 253 V AC (10 mA) or >0.5 VA at 24 V AC (20 mA) Relays with a lower holding power/rated power can be operated by means of an RC module connected inparallel. For example: Fischer & Tausche Type 9725 A, 0.5 m 220 , 250 V~, HPF565 P1 Relays with a maximum holding power/rated power <89 VA at 253 V AC or <8.4 VA at 24 V AC Voltage drop across FEL51 max. 12 V Residual current with blocked thyristor max. 3,8 mA Max. load current 375 VA at 253 V or max. 36 VA at 24 V (not shortcircuit proof) The external load must not be to small. Due to the residual current (max. 3.8 mA) required for the operation of the electronics, a relay with a low energizing current would be correctly energized but not properly return to the non-energized failsafe position. In such cases, an additional shunt resistor can be used to reduce the relay current. Example: A shunt resistor of 39 k / 2W is recommended for an OMRON relay type MY4/220V.

FEL51

1 2

*
min. 19 V

External load R must be connected

1A

*
R

L1

U~ max. 253 V AC 50/60 Hz

PE (Ground)

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

FEL52

1 2

(+)

Connection data (load) FEL52 Load switched via the transistor and separate PNP connection, max. 55 V DC Current consumption max.15 mA Load current max. 350 mA (pulsed overload and short-circuit protection) Residual current <100 A (with transistor blocked) Capacitive load max. 0.5 F at 55 V, max. 1.0 F at 24 V Residual voltage <3 V (with transistor switched through) Prefereable used with programmable logic controllers (PLC).

0.5 A e.g. Relay PLC

L+ L
... 1055 V DC U

FEL54

1 2

3 4 56 7 8

Connection data (load) FEL54 Loads switched via 2 floating change-over contacts (DPDT) Power consumption 1.3 W I~ max. 6 A (EEx d 4 A) U~ max. 253 V AC P~ max. 1500 VA, cos = 1 P~ max. 750 VA, cos > 0.7 I% max. 6 A (EEx d 4 A) to 30 V I% max. 0.2 A bis 125 V * When jumpered, the relay output works with NPN logic.

*
0.5 A

** Both relay contacts switch simultaneously.

NO C a u L1 N L+ L PE (Ground)

NC r

NO C a u

NC r

**

**

U~ 19253 V AC, 50/60 Hz ... 19 55 V DC U

Endress+Hauser Maulburg

SM06F/00/en/09.08

Liquiphant M/S

FEL55

1 2

Connection data (load) FEL55 For connection to programmable logic controllers (PLC) AI module 4...20 mA to EN 61131-2 Output signal jump from high to low current on limit Load current 16.8 mA Connection voltage 11...36 V DC)

EEx ia

EX

EX

... 1136 V DC U e.g. PLC

FEL56

1 2

Connection data (load) FEL56 For connecting to isolating amplifiers according to NAMUR (IEC 609475-6) or Commutec AI module 4...20 mA according to EN 61131-2 Output signal jump from low to high current on limit Power consumption <6 mW at I < 1 mA; <38 mW at I = 2.2 ...4 mA

EEx ia

Also see technical data of connected isolating amplifier.

EX
H

I
L

EX

Isolating amplifier to IEC 60947-5-6 (NAMUR)

10

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

FEL58

1 2

Connection data (load) FEL58 For connecting to isolating amplifiers according to NAMUR (IEC 60947-5-6) or Commutec AI module 4...20 mA according to EN 61131-2 Output signal jump from high to low current on limit Power consumption <6 mW at I < 1 mA; <38 mW at I = 2.2 ...4 mA See technical data of connected isolating amplifier.

EEx ia

EX
H

I
L

EX

Isolating amplifier to IEC 60947-5-6 (NAMUR)

FEL57

1 2

Connection data (load) FEL57 For connecting to different switching units of Endress+Hauser Output signal jump of PFM signal from high to low frequency when sensor is covered Switching between minimum/maximum safety in the Supply voltage 9.5...12.5 V DC Current consumption 10...13 mA Power consumption <150 mW

EEx ia

EX
PFM 50 / 150 Hz

EX
7 33 37 + 8 34 38 FTL370/372 FTL375P 1CH Input 1 FTL372 FTL375P 2CH Input 2 FTL375P 3CH Input 3 Commutec S SIF101, SIF111 Nivotester FTL320 FTL325P 1CH FTL325P 3CH

d4

d2

z4

z2

z6

d6

Endress+Hauser Maulburg

11

SM06F/00/en/09.08

Liquiphant M/S

FEL50A
Address
Address ON OFF 1 2 3 4 5 6 7 8 SW HW

ON OFF

SW HW

1 2 3 4 5 6 7 8

1 2

FEL50A For the connection to PROFIBUS PA Supply voltage 9.5 to 12.5 V DC The current consumption has been changed with hardware version 02.00. Please observe hardware version printed on the electronics!
Hardware version 01.00 02.00 Software Current consumption version 01.03.00 10.5 mA 0.5 mA 01.03.00 12.5 mA 1.0 mA

EX
Settings for density and Min./Max. are done by the software. Addresses are setted at the DIP switch of the electronic insert.

EX
PA PA+


PROFIBUS PA

... 932 V DC U

Segment coupler

12

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Test circuit dor PFM devices


ATTENTION: This measurement may not be performed in an Ex area! The test circuit only applies to Liquiphant M with electronic insert FEL57. This electronic insert converts the signal from the tuning fork into a pulse frequency modulation (PFM). The output signal jumps from a high to a low PFM frequency when the sensor is covered. The PFM signal can only be processed by the listed E+H Nivotesters. current pulses of approximately 11 mA are superimposed to the supply current, as shown on the oscillogram in order to measure the signal current, a test resistor of approximately 10 Ohms must be inserted into the negative line between the FEL57 and the Nivotester FTL... an oscilloscope must be connected parallel to the resistor for the test, the fork oscillates freely or is manually dampened the Nivotester FTL325P/375P must then change it switching status The PFM signals must then appear as shown, adjustments cannot be made.

FEL57
Supply current approx. 11 mA

E+H-Nivotester

approx. 11 V Supply voltage

1 2

10 W test resistor

FTL320 FTL325P FTL370 FTL372 FTL375P

Pulse width approx. 200 s

t=5 ms/div approx. 0.1Vss / 150 Hz approx. 0.1V 0V

Fork oscillates freely

Pulse width approx. 200 s

t=5 ms/div approx. 0.1Vss / 50 Hz approx. 0.1V 0V

Fork covered

Endress+Hauser Maulburg

13

SM06F/00/en/09.08

Liquiphant M/S

Tooling
Philips screwdriver size 1 Screwdriver for M3/M4 Allen wrench AF3 mm Circlip pliers for external circlips Tweezers

14

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Disassembly of Liquiphant M/S


The Liquiphant M family consists of 5 different instrument types: FTL50/51, FTL51C and FTL50H/51H. The family contains a compact design (FTL50), a version with extension pipe (FTL51), a coated sensor (FTL51C) as well as a polished fork for the use in hygienic applications (FTL50H/51H). The Liquiphant S family covers two different instrument types - the compact design of FTL70 and the version with extension pipe FTL71. Each instrument type of Liquiphant M and Liquiphant S can be combined with various housings, electronics and process connections. The mounting of the electronic insert and the sensor is identical for all instruments. For instruments with housing T13/F13, the sensor and the housing are screwed together.
Housing F15/F16/F17 (as example F15) Housing F13

Disassembly 1 2 Unscrew lid. If available, loosen cover lock. Disconnect cables. Loosen two mounting screws and pull out electronic insert. Loosen screw from mounting adapter (for stainless steel housing F15 only). Loosen external circlip from sensor head using circlip pliers. Then remove mounting parts. Disconnect ground cable from external circlip. Separate sensor from housing (also at extended version) for devices with housing F15/16/17, the sensor must be pulled down out of the housing. for devices with housing F13, the sensor must be screwed out of the housing (12 turns).

3 4 5

2
6

Top view into housing

Removed mounting parts: mounting adapter external circlip

The electronic exchange at Liquiphant M with housing T13 (EEx de) is described separately.

Endress+Hauser Maulburg

15

SM06F/00/en/09.08

Liquiphant M/S

Assembly Liquiphant M/S

Housing F15/F16/F17 (as example F15)

Housing F13

Lightly grease the O-ring and thread (for housing type F15/ F16/ F17). Lightly grease the O-rings and slide them over the sensorhead in the positions shown (for housing type F13).

8
NOTE: Lubrication grease "Syntheso Glep 1; Manufacturer: Klber Lubrication). 2 3 Slide housing over the sensor head respectively screw it onto the sensor head. Place the circlip in the adapter and put both onto the sensor head. Position circlip in the top-most groove of the sensor head using the pliers. All parts (sensor, adapter and housing) should be approx. positioned as shown in the picture below. All parts are now connected and can be turned together (approx. 300 from stop to stop). Secure mounting adapter with screw (for stainless steel housing F15 only). Plug potential ground cable onto the ground connection of the circlip.

3
5 6

5
7

Place electronic insert onto sensor head and tighten the two mounting screws. Screw on cover. If available, fasten cover lock.

16

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Exchange of electronic insert for Liquiphant M/S with housing T13


Instruments with protection class EEx de are only available with T13 housing. The disassembly is limited to the electronic insert. Neither housing and its components, not the sensor are available as spare parts.

Disassembly

1 2 3 4

Loosen cover lock and unscrew cover. Disconnect cable. Loosen the two mounting screws at the electronic insert. Pull out electronic insert using the removal tool.

4
Removal tool

2 3

Assembly The assembly is done in reverse order.

Endress+Hauser Maulburg

17

SM06F/00/en/09.08

Liquiphant M/S

Human interface
With electronic inserts FEL51, FEL51, FEL54, FEL55: 2 switches with safety mode and density change, green LED to indicate operational status, red LED to indicate the switching status, flashes in the event of corrosion damage on sensor or if the electronics are defect
1 2

FEL5x
1 2 Min >0,5 Max >0,7

I max : 350mA

With electronic insert FEL56: 2 switches with safety mode and density change, green LED flashes to indicate operational status, red LED to indicate the switching status, flashes in the event of corrosion damage on sensor or if the electronics is defect

LEDs

Switches

FEL57

With electronic insert FEL57: 2 switches with density change and cyclical checking, green LED to indicate operational status, yellow LED to indicate the covered status, flashes in the event of corrosion damage on sensor or if the electronics is defect

STD

>0,7

EXT

>0,5

I max : 350mA

LEDs

Switches

With electronic insert FEL58: 2 switches with safety mode and density change green LED - flashes quickly to indicate operational status, - flashes slowly to indicate corrosion damage on sensor or if the electronics is defect yellow LED to indicate the switching status, test key - breaks the cable connection

FEL50A
Address ON OFF 1 2 3 4 5 6 7 8 SW HW

PA PA+

With electronic insert FEL50A: 8 switches for configuring the device address, green LED to indicate operational status, pulsing to indicate communication yellow LED to indicate the switching status, flashes in the event of corrosion damage on sensor or if the electronics is defect

LEDs

Switches

18

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Diagnostics at Liquiphant M/S


Each electronics features "diagnostics outputs" providing measuring signals for control purpose. These include the sensor oscillation signal, the switching signal and the identifier of the electronics version. Using small probes (i. e. wire < 1 mm ), the signal frequency and amplitude can be monitores on an oscilloscope. Identification resistor FEL51 = 1k000 FEL52 = 1k540 FEL54 = 2k260 FEL55 = 3k320 FEL56 = 4k870 FEL57 = 6k810 FEL58 = 8k210 FEL50A = 10k000 Switching signal without delay Opposed to the switching output at the terminal block, which is always delayed by 0.5 seconds, this signal is taken directly from the processor. Advantage: eventual erroneous switching caused by external vibration can be detected with an oscilloscope. fork free =0V fork covered = approx. 4.8 V Sensor signal The sensor signal has an offset of approx. 4 V DC. The amplitude is approx. 2 Vss. The signal is in saturation, therefore not a sine wave. The frequency of the fork is: - not covered / in air approx. 900...1200 Hz - covered to switch point approx. 850 Hz (10...15% below resonance frequency) The frequency depends on the insertion depth, the amplitude on the viscosity.

Connection

FEL5x

W
3 4 1 2

The most important operating signals can be checked at the socket via a diagnostics plug, using the appropriate measuring instruments. socket 1 = identification, resistance value over GND socket 2 = switch signal L/H over GND socket 3 = GND socket 4 = fork frequency over GND

approx. 2 Vss

approx. 4 Voffset 0V

Endress+Hauser Maulburg

19

SM06F/00/en/09.08

Liquiphant M/S

Sensor signal at the plug-in connector


Signals at tuning fork
EEPROM in socket > 5 Vss / approx. 1 ms

GND

+5 V

SCL

View on the socket of a sensor

SDA

approx. second

With the sensor removed from the instrument, the "Signal In" and "Signal Out" can be checked against ground with an oscilloscope connected to the socket. Signal Out Signal In The sensor is momentarily excited by carefully tapping the fork. If the piezo and the connecting flexible conductor strip are in order, a dampened sine wave can be measured at the respective test points. This test is only a function test, it does not allow a statement about the quality of the sensor. A check of the data stored in the EEPROM or an exchange of the EEPROM is not possible. ATTENTION: For the coated sensors of the Liquiphant M and the sensors of Liquiphant S, the Signal Out is approx. 4 times higher than the Signal In!

ATTENTION! The socket can be turned, there is no stop. There is a risk of twisting off the flexible conductor strip!

+ +

Piezo of sensor

Signals at the sensor plug of the electronic insert FEL5x At the disassembled electronic insert without fork, the following signals can be diagnosed over an adapter:
GND +5V
+5 V

SDA

SCL 0
+5 V power supply for the EEPROMs against GND

approx. 2.5 ms approx. 900 ms +5 V

approx. 2.5 ms

0
SCL and SDA, Clock- and data line for the EEPROM

approx. 480 Hz...550 Hz

Signal IN

GND

Signal OUT

+6 V...+9 V

Electronic insert, view from down on the plug-contacts 0


Signal OUT, free running transmitting signal

20

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Trouble-shooting
Fault Does not switch Reason No power Faulty signal line Faulty electronic insert - FEL51 connected directly to L1 and N Density of liquid too low Fork encrusted Fork corroded (Indication on FEL: red/yellow flashes, FEL58: green flashes 0.3 Hz) FEL51: Internal resistance of connected relay too large FEL51: Holding current of connected relay too low FEL54: Contacts welded together (after short-circuit) Switches incorrectly Sporadic faulty switching Min-/Max- fail-safe mode set wrongly Thick heavy foam, very turbulent conditions, foaming liquid Extreme RFI Extreme vibration Water in housing FEL52: Output overload Switches incorrectly after power failure FEL57, behaviour during switch-on test (functional test) Remedy Check power Check signal line Exchange - always connect FEL51 via external load Set density to >0.5 at electronic insert Clean fork For Liquiphant M: Exchange fork and process connection For Liquiphant S: Exchange instrument Connect suitable relay

Connected resistor in parallel with relay Exchange FEL54; put pulse in contact circuit Set correct mode at electronic insert Mount Liquiphant in bypass

Use screened cable Decouple, damp, turn fork 90 Screw cover and cable gland tight Reduce load, (cable) capacitance Observe switching behaviour of FEL57; after power failure blockplant control for up to 45 s

Endress+Hauser Maulburg

21

SM06F/00/en/09.08

Liquiphant M/S

Spare parts Liquiphant M/S


The following pages only show an exploded drawing with the respective item number. The corresponding spare part number can be found at the current spare part price list or online at the spare part finding tool (SFT). Spare part numbers also appear on each assembly of a device, which is available as spare part.

Spare parts Liquiphant M FTL5x and Liquiphant S FTL7x with polyester housing F16

20 21

30

12

21 20

10

15

11

50

Sensor complete with O-ring and process seal


Sensor komplett mit O-Ring und Prozessdichtung FTL50X- / FTL50HX- ... FTL51X- / FTL51HX-

57

22

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Spare parts Liquiphant M FTL5x and Liquiphant S FTL7x with aluminum housing F13 (EEx d)

20 21

30

12

20, 21

10

11
Sensor complete with O-ring and process seal Sensor komplett mit O-Ring und Prozessdichtung
FTL50X- / FTL50HX- ... FTL51X- / FTL51HX-

50

57

Endress+Hauser Maulburg

23

SM06F/00/en/09.08

Liquiphant M/S

Spare parts Liquiphant M FTL5x and Liquiphant S FTL7x with aluminum housing F17 (EEx ia)

20

21

30

12

20, 21

10

11

50

Sensor complete with O-ring and process seal


Sensor komplett mit O-Ring und Prozessdichtung FTL50X- / FTL50HX- ... FTL51X- / FTL51HX-

57

24

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Spare parts Liquiphant M FTL5x with housing F15 out of 316L (Ex ia II T6, WHG)

20

20

30

12 20

21

10

15 11

50

Sensor complete with O-ring and process seal Sensor komplett mit O-Ring und Prozessdichtung
FTL50X- / FTL50HX- ... FTL51X- / FTL51HX-

57

Endress+Hauser Maulburg

25

SM06F/00/en/09.08

Liquiphant M/S

Spare parts Liquiphant M FTL5x and Liquiphant S FTL7x with aluminum housing T13 (EEx de)

20

21

30

12

21
Housing T13 is not available as a spare part Gehuse T13 ist nicht als Ersatzteil lieferbar

25

Sensors mounted on housing T13 are not available as a spare part Sensoren zum Anbau an Gehuse T13 sind nicht als Ersatzteil lieferbar

57

26

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Spare parts Liquiphant M FTL5x and Liquiphant S FTL7x


10 11 12 15 20 21 25 30 50 57 Housing Sealing for housing Fixing set Cable entry/Cable gland Cover Sealing for cover Cover for terminal department Electronics Sensor assembly (for FTL5x only) Process gasket (for FTL5x only) Miscellaneous Lubricant "SYNTHESO GLEP1" (tube a 8g) Diagnostics plug FTL5x Liquiphant M

Endress+Hauser Maulburg

27

SM06F/00/en/09.08

Liquiphant M/S

Supplementary documentation
Technical Information Technical Information Technical Information Liquiphant M FTL50(H), FTL51(H) Liquiphant M FTL51C Liquiphant S FTL70, FTL71 TI328F TI347F TI354F

Operating Instructions Operating Instructions Operating Instructions Operating Instructions Operating Instructions

Liquiphant M/S with FEL50A PROFIBUS PA Liquiphant M FTL50, FTL51 Liquiphant M FTL50(H), FTL51(H) Liquiphant M FTL51C Liquiphant S FTL70, FTL71

BA141F KA143F KA144F KA162F KA172F

Above mentioned list is only a short abstract of available documentations, further documents (e.g. Operating Instructions) can be downloaded under www.endress.com.

28

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Endress+Hauser Maulburg

29

SM06F/00/en/09.08

Liquiphant M/S

30

Endress+Hauser Maulburg

Liquiphant M/S

SM06F/00/en/09.08

Endress+Hauser Maulburg

31

Technical changes excepted

Instruments International
Endress+Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel. +41 61 715 81 00 Fax +41 61 715 25 00 www.endress.com info@ii.endress.com

SM06F/en/09.08