HI 96753C Chloride ISM: Instruction Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 13

Instruction Manual

HI 96753C Chloride ISM

www.hannainst.com

Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.

PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please notify your Dealer. Each HI 96753 Ion Selective Meter is supplied complete with: Two Sample Cuvettes and Caps CAL CHECK standard cuvettes 9V Battery Cloth for wiping cuvettes Instrument quality certificate Instruction Manual Rigid carrying case

TABLE OF CONTENTS
PRELIMINARY EXAMINATION .............................................................................................. 3 GENERAL DESCRIPTION .....................................................................................................4 ABBREVIATIONS ................................................................................................................ 4 SPECIFICATIONS .................................................................................................................5 PRECISION AND ACCURACY ................................................................................................ 5 PRINCIPLE OF OPERATION .................................................................................................6 FUNCTIONAL DESCRIPTION ................................................................................................ 7 ERRORS AND WARNINGS .................................................................................................. 9 GENERAL TIPS FOR AN ACCURATE MEASUREMENT ...........................................................11 STARTUP .........................................................................................................................12 MEASUREMENT PROCEDURE ...........................................................................................12 VALIDATION PROCEDURE .................................................................................................15 CALIBRATION PROCEDURE ...............................................................................................17 GLP .................................................................................................................................20 BATTERY MANAGEMENT ..................................................................................................21 BATTERY REPLACEMENT ..................................................................................................22 ACCESSORIES ...................................................................................................................22 WARRANTY ......................................................................................................................23

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.

Note: Save all packing material until you are sure that the instrument works correctly. Any defective item must be returned in its original packing. 3

GENERAL DESCRIPTION
The HI 96753 is an auto diagnostic portable microprocessor meter that benefits from Hannas years of experience as a manufacturer of analytical instruments. It has the advanced optical system based on a special Light Emitting Diode (LED) and a narrow band interference filter that allows most accurate and repeatable readings. All instruments are factory calibrated and the electronic and optical design minimizes the need of frequent calibration. With the powerful CAL CHECKTM validation function, you are able to validate good performance of your instrument at any time. The validation procedure is extremely user friendly. Just use the exclusive HANNA ready-made, NIST traceable standards to verify the performance of the instrument and recalibrate if necessary. All instruments are splash waterproof and the LED and filter units are protected from dust or dirt by a transparent cup. This makes the instruments fulfill field applications. Display messages aid the user in routine operation. The meter has an auto-shut off feature that will turn off the instrument after 10 minutes of non use in measurement mode or after 1 hour if left in calibration mode. The meter uses an exclusive positive-locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell. It is designed to fit a cuvette with a larger neck making it easier to add both sample and reagents. The cuvette is made from special optical glass to obtain best results. The HI 96753 meter measures the chloride (Cl) content in water and wastewater samples in the 0.0 to 20.0 mg/L (ppm) range. The method is an adaptation of the mercury(II) thiocyanate method. The reagents are in liquid form and are supplied in bottles. The amount of reagent is precisely dosed to ensure the maximum repeatability. Range Resolution Accuracy Typical EMC Deviation Light Source Light Detector Method

SPECIFICATIONS
0.0 to 20.0 mg/L 0.1 mg/L 0.5 mg/L 6% of reading @ 25C 0.1 mg/L Light Emitting Diode Silicon Photocell with narrow band interference filter @ 466nm Adaptation of the mercury(II) thiocyanate method. The chloride ion displaces thiocyanate ion from mercury(II). The iron(III) present forms with thiocyanate an orange colored complex. The intensity of color is proportional to the chloride ion concentration. 0 to 50C (32 to 122F); max 95% RH non-condensing 1 x 9 volt After 10' of non-use in measurement mode; after 1 hour of non-use in calibration mode; with last reading reminder. 192 x 104 x 69 mm (7.6 x 4.1 x 2.7) 260 g (12.7 oz.). Description Displacing Reagent Complexing Reagent Quantity/test 1.0 mL 1.0 mL

Environment Battery Type Auto-Shut off

Dimensions Weight REQUIRED REAGENTS Code HI 93753A-0 HI 93753B-0

PRECISION AND ACCURACY


Precision is how closely repeated measurements agree with each other. Precision is usually expressed as standard deviation (SD) . Accuracy is defined as the nearness of a test result to the true value. Although good precision suggests good accuracy, precise results can be inaccurate. The figure explains these definitions. In a laboratory using a standard solution of 10.0 mg/L chloride and a representative lot of reagent, an operator obtained with a single instrument a standard deviation of 0.3 mg/L. 5

ABBREVIATIONS
C: F: mg/L: mL: mV: degree Celsius degree Fahrenheit milligrams per liter. mg/L is equivalent to ppm (parts per million) milliliter millivolts

PRINCIPLE OF OPERATION
Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or crystal lattices. If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the matter and on the physical-chemical characteristics of the substance according to the Lambert-Beer Law: -log I/Io = c d or A = c d Where: -log I/Io Io I c d = = = = = = Absorbance (A) intensity of incident light beam intensity of light beam after absorption molar extinction coefficient at wavelength molar concentration of the substance optical path through the substance

A microprocessor controlled special LED emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette. The optical path is fixed by the diameter of the cuvette. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity Io or I. The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an electric current, producing a potential in the mV range. The microprocessor uses this potential to convert the incoming value into the desired measuring unit and to display it on the LCD. The measurement process is carried out in two phases: first the meter is zeroed and then the actual measurement is performed. The cuvette has a very important role because it is an optical element and thus requires particular attention. It is important that both, the measurement and the calibration (zeroing) cuvettes, are optically identical to provide the same measurement conditions. Whenever possible use the same cuvette for both. It is necessary that the surface of the cuvette is clean and not scratched. This to avoid measurement interference due to unwanted reflection and absorption of light. It is recommended not to touch the cuvette walls with hands. Furthermore, in order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the cuvette to prevent any contamination.

Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other factors are known. Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents. Given that the absorption of a compound strictly depends on the wavelength of the incident light beam, a narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize measurements. The optical system of Hanna's HI 96753 colorimeter is based on special LED and narrow-band interference filters to guarantee both high performance and reliable results.

FUNCTIONAL DESCRIPTION
INSTRUMENT DESCRIPTION

HI 96753 block diagram (optical layout)

1) 2) 3) 4) 5) 6) 7) 8)

GLP/ key CAL CHECK key ZERO/CFM key READ/ /TIMER key ON/OFF key Liquid Crystal Display (LCD) Cuvette alignment indicator Cuvette holder

KEYPAD DESCRIPTION ON/OFF: to turn the meter on and off. ZERO/CFM: to zero the meter prior to measurement, to confirm edited values or to confirm factory calibration restore. READ/ /TIMER: this is a multi-functional key. In measurement mode, press to make a measurement, or press and hold for three seconds to start a pre-programmed countdown prior to measurement. In GLP mode press to view the next screen. CAL CHECK: this is a bi-functional key. Just press to perform the validation of the meter, or press and hold for three seconds to enter calibration mode. GLP/ : this is a bi-functional key. Just press to enter GLP mode. In calibration mode press to edit the date and time. OPERATING MODES Measurement mode: default operation mode, enables both validation and measurement. Calibration mode: may be entered by keeping CAL CHECK pressed for three seconds (the CAL tag appears), it enables calibration of the instrument. GLP mode may be entered by pressing GLP/ (GLP appears), it enables consulting of user calibration date or restore factory calibration. DISPLAY ELEMENTS DESCRIPTION

ERRORS AND WARNINGS


The instrument shows clear messages when erroneous condition appears. Messages are also displayed when the obtained values are outside expected range. The beeper is playing a beep on errors. a) on zero reading Light High: There is too much light to perform a measurement. Please check the preparation of the zero cuvette.

Light Low: There is not enough light to perform a measurement. Please check the preparation of the zero cuvette.

No Light:The instrument cannot adjust the light level. Please check that the sample does not contain any debris.

b) on sample reading Inverted cuvettes: The sample and the zero cuvette are inverted. 1) The measuring scheme (lamp, cuvette, detector), appears during different phases of zero or reading measurement 2) Error messages and warnings 3) The battery icon shows the charging level of the battery 4) The hourglass appears when an internal checking is in progress 5) Status messages 6) The chronometer appears when the reaction timer is running 7) The month, day and date icons appear when a date is displayed 8) Four digit main display 9) Measuring units 10) Four digit secondary display 8

Zero: A zero reading was not taken. Follow the instructions of the measurement procedure for zeroing the meter.

Under Range: A blinking 0.0 indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvette for reference (zero) and measurement.

Battery low: The battery must be replaced soon.

Over Range: A flashing value of the maximum concentration indicates an over range condition. The concentration of the sample is beyond the programmed range: dilute the sample and re-run the test.

Dead battery: This indicates that the battery is dead and must be replaced. Once this indication is displayed, the meter will lock up. Change the battery and restart the meter.

c) during calibration procedure Standard Low: The standard reading is less than expected.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT


The instructions listed below should be carefully followed during testing to ensure best accuracy. Color or suspended matter in large amounts may cause interference, therefore these should be removed by treatment with active carbon and by prior filtration. For a correct filling of the cuvette: the liquid in the cuvette forms a concavity on the top; the bottom of this concavity must be at the same level of the 10 mL mark. In order to measure exactly 0.5 mL of reagent with the 1 mL syringe: (a) push the plunger completely into the syringe and insert the tip into the solution. (b) pull the plunger up until the lower edge of the seal is exactly on the 0.0 mL mark. (c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above the cuvette, push the plunger down into the syringe until the lower edge of the seal is exactly on the 0.5 mL mark. Now the exact amount of 0.5 mL has been added to the cuvette, even if the tip still contains some solution.

Standard High: The standard reading is higher than expected.

d) other errors and warnings Cap error: Appears when external light enters in the analysis cell. Assure that the cuvette cap is present.

Cooling lamp: The instrument waits for the lamp to cool down.

(a)

(b)

(c)

10

11

It is important that the sample does not contain any debris. This would corrupt the reading. Each time the cuvette is used, the cap must be tightened to the same degree. Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free cloth prior to insertion. Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette. Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost. It is possible to take multiple readings in a row, but it is recommended to take a new zero reading for each sample and to use the same cuvette for zeroing and measurement. After the reading it is important to discard immediately the sample, otherwise the glass might become permanently stained. All the reaction times reported in this manual are referred to 20C (68F). As a general rule of thumb, they should be doubled at 10C (50F) and halved at 30C (86F). In order to maximize accuracy, prior to a measurement follow the validation procedure to be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

When the beeper sounds briefly and the LCD displays dashes, the meter is ready. The blinking ZERO indicates that the instrument needs to be zeroed first.

Fill one cuvette (#1) with 10 mL of distilled water (up to the mark).

Fill another cuvette (#2) with 10 mL of sample (up to the mark) Note: for low chloride ions concentration, rinse the cuvette a couple of times with sample before filling it with 10 mL of sample (up to the mark). Note: for most accurate results, use two graduated pipettes to deliver exactly 10 mL of distilled water and 10 mL of sample to the cuvettes. Using the 1 mL syringe, add 0.5 mL of HI 93753A-0 Displacing Reagent to each cuvette. Replace the caps and mix each cuvette by turning the cuvette upside down for approximately 30 seconds.

STARTUP
Prepare the instrument for measurement as follows: Unpack the instrument by removing the dust protection sleeve from the instrument cuvette holder. Place the battery in the instrument as described in the BATTERY REPLACEMENT chapter. Place the instrument on a flat table. Do not place the instrument under direct sun light.

MEASUREMENT PROCEDURE
To compensate the meter for the sample turbidity or color, the measurement takes place in two phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the reacted sample is measured. Turn the meter on by pressing ON/OFF. The display briefly shows all tags on.

12

13

Remove the caps and use the second 1 mL syringe to add 0.5 mL of HI 93753B-0 Complexing Reagent to each cuvette.

Press READ/ /TIMER. The lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. At the end of measurement, the instrument directly displays concentration in mg/L of chloride on the LCD. INTERFERENCES The pH of sample after addition of reagents should be about 2. For alkaline samples, neutralize before adding reagents. Intensely colored samples will cause interference, therefore they should be adequately treated before performing the test. Suspended matter in large amounts should be removed by prior filtration.

Replace the caps and mix each cuvette by turning the cuvette upside down for approximately 30 seconds.

Place the cuvette with the reacted distilled water (#1) into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.

#1

Press READ/ /TIMER. The display will show the countdown prior to zeroing. The beeper is playing a beep at the end of countdown period. Alternatively, wait for 2 minutes then just press ZERO/CFM. In both cases, the lamp, cuvette and detector icons will appear on the display, depending on the zeroing phase.

VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated. Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL CHECKTM Standards, otherwise erroneous results will be obtained. Turn the meter on by pressing ON/OFF.

When the beeper sounds briefly and the LCD displays dashes, the meter is ready. After a few seconds, the display will show -0.0-. The meter is now zeroed and ready for measurement.
#2

Remove the cuvette.

Place the CAL CHECK TM Standard HI 96753-11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.

Place the other cuvette with the reacted sample (#2) into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. 14 15

Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. After a few seconds, the display will show -0.0-. The meter is now zeroed and ready for validation. Remove the cuvette. Place the CAL CHECK TM Standard HI 96753-11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. Press CAL CHECK and the lamp, cuvette and detector icons together with CAL CHECK will appear on the display, depending on the measurement phase. At the end of the measurement the display will show the validation standard value.

CALIBRATION PROCEDURE
Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON/OFF keys. Warning: Do not calibrate the instrument with standard solutions other than the HANNA CAL CHECKTM Standards, otherwise erroneous results will be obtained. Turn the meter on by pressing ON/OFF.

When the beeper sounds briefly and the LCD displays dashes, the meter is ready.

Press and hold CAL CHECK for three seconds to enter calibration mode. The display will show CAL during calibration procedure. The blinking ZERO asks for instrument zeroing.

Place the CAL CHECK TM Standard HI 96753-11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. The reading should be within specifications as reported in the CAL CHECKTM Standard Certificate. If the value is found out of the specifications, please check that the cuvettes are free of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications, then recalibrate the instrument.

Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase.

16

17

After a few seconds the display will show -0.0-. The meter is now zeroed and ready for calibration. The blinking READ asks for reading calibration standard. Remove the cuvette. Place the CAL CHECKTM Standard HI 96753-11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.

DATE INPUT Press GLP/ to edit the desired year (2000-2099). If the key is kept pressed, the year number is automatically increased. When the correct year has been set, press ZERO/CFM or READ/ /TIMER to confirm. Now the display will show the month blinking. Press GLP/ to edit the desired month (01-12). If the key is kept pressed, the month number is automatically increased.

or

Press READ/ /TIMER and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase.

When the correct month has been set, press ZERO/CFM or READ/ /TIMER to confirm. Now the display will show the day blinking. Press GLP/ to edit the desired day (01-31). If the key is kept pressed, the day number is automatically increased. Note: It is possible to change the editing from day to year and to month by pressing READ/ /TIMER.

or

After measurement the instrument will show for three seconds the CAL CHECKTM Standard value. Note: If the display shows STD HIGH, the standard value was too high. If the display shows STD LOW, the standard value was too low. Verify that both CAL CHECKTM Standards HI 96753-11 Cuvettes, A and B are free from fingerprints or dirt and that they are inserted correctly. Then the date of the last calibration (e.g.: 01.08.2008) appears on the display, or 01.01.2008 if the factory calibration was selected before. In both cases the year number is blinking, ready for date input. 18

Press ZERO/CFM to save the calibration date.

The instrument displays Stor for one second and the calibration is saved. 19

The instrument will return automatically to measurement mode by displaying dashes on the LCD.

Press READ/ /TIMER to enter in the factory calibration restore screen. The instrument asks for confirmation of user calibration delete.

GLP
In the GLP mode, the last user calibration date can be consulted and the factory calibration can be restored. LAST CALIBRATION DATE To display the calibration date: Press GLP/ to enter GLP mode. The calibration month and day will appear on the main display and the year on the secondary display.

Press ZERO/CFM to restore the factory calibration or press GLP/ again to abort factory calibration restore. The instrument briefly notifies done when restores factory calibration and returns to measurement mode.

BATTERY MANAGEMENT
To save battery, the instrument shuts down after 10 minutes of non-use in measurement mode and after 1 hour of non-use in calibration mode. If a valid measurement was displayed before auto-shut off, the value is displayed when the instrument is switched on. The blinking ZERO means that a new zero has to be performed. One fresh battery lasts for around 750 measurements, depending on the light level. The remaining battery capacity is evaluated at the instrument startup and after each measurement. The instrument displays a battery indicator with three levels as follows: 3 lines for 100 % capacity 2 lines for 66 % capacity 1 line for 33 % capacity Battery icon blinking if the capacity is under 10 %. If the battery is empty and accurate measurements cant be taken anymore, the instrument shows dead batt and turns off. To restart the instrument, the battery must be replaced with a fresh one. 20 21

If no calibration was performed, the factory calibration message, F.CAL will appear on the main display and the instrument returns to measurement mode after three seconds.

FACTORY CALIBRATION RESTORE It is possible to delete the calibration and restore factory calibration. Press GLP/ to enter GLP mode.

BATTERY REPLACEMENT
To replace the instruments battery, follow the steps: Turn the instrument off by pressing ON/OFF. Turn the instrument upside down and remove the battery cover by turning it counterclockwise.

WARRANTY
HI 96753 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact your dealer. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.

Extract the battery from its location and replace it with a fresh one. Insert back the battery cover and turn it clockwise to close.

ACCESORIES
REAGENT SET HI 93753-01 HI 93753-03

Reagents for 100 Reagents for 300

OTHER ACCESORIES HI 96753-11 CAL CHECKTM Standard Cuvettes (1 set) HI 721310 9V battery (10 pcs.) HI 731318 Cloth for wiping cuvettes (4 pcs.) HI 731331 Glass cuvettes (4 pcs.) HI 731335 Caps for cuvettes HI 93703-50 Cuvette cleaning solution (230 mL) HI 741218 Carrying case

Recommendations for Users Before using these products, make sure that they are entirely suitable for your specific application and for the environment in which they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipments, this requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument safety do not use or store the instrument in hazardous environments.

Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice. 22 23

Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.com

Local Sales and Customer Service Office

Printed in EUROPE (ROMANIA)

MAN96753 01/11

24