FIELD DEVICES - FLOW

Product Specifications L
o
g
o
PSS 1-6F9 A
9100A Series Magnetic Flowtubes with Ebonite Liners
1 to 78 in (25 to 2000 mm) Nominal Flowtube Sizes
The Foxboro

brand 9100A Series Magnetic Flowtube together with an IMT25 or IMT25L Magnetic Flow
Transmitter, combine to form a Magnetic Flowmeter with pulsed dc excitation. It is designed for use with
conductive fluids in general purpose, and waste and water applications. The transmitter converts the low
level, high impedance signal from the lined flowtube to a scaled transmission signal, either 4 to 20 mA,
digital, or pulse output, that is proportional to volumetric flow rate.
FEATURES
Reliability and Dynamic Range meet the Heavy
Demands of the Water Treatment Industry
Field-Proven Electrode Seal Design
Ebonite Liner
Enclosure meets IEC IP67 and NEMA Type 4X;
Upgradeable to IEC IP68
Withstands Temporary Flooding before, during,
and after Installation; Can be upgraded on Site to
withstand Continuous Flooding
Each Flowtube has Built-in Grounding (Earthing)
Electrodes
ANSI/NSF 61 (US) Certified
Large Selection of Flowtube Sizes and Flanges
Unidirectional or Bidirectional Flow
Optional Liner Protector offered for Processes
with High Velocity and Abrasive Fluids
Conforms to Applicable European Union
Directives (Product Marked with CE Logo)
PSS 1-6F9 A
Page 2



SUPERIOR REPUTATION FOR
DEPENDABILITY AND QUALITY
Invensys introduced magnetic flow measurement
systems to the process industries in 1954 and has
demonstrated the broadest and most time-proven
application expertise with tens of thousands of
successful small and large size flowtube installations.
INTELLIGENT PATH TO MAG FLOW SYSTEMS
The merging of the latest technology in flowtube
design, in conjunction with the Intelligent IMT25
Magnetic Flow Transmitter, provides the waste and
water industries with a significant advancement in
conductive liquid flow measurement. In addition to
4 to 20 mA and pulse output signals, remote digital
communication is provided using either FoxCom,
HART, or FOUNDATION fieldbus protocols.
PULSED DC FLOWTUBES USED WITH
REMOTE MOUNTED TRANSMITTERS
The 9100A Series Magnetic Flowtubes are calibrated
for use with pulsed dc coil excitation. As stated
above, Foxboro offers the Intelligent I/A Series IMT25
Magnetic Flow Transmitter for use with these
flowtubes. This transmitter may be remote mounted
on a pipe or to a flat surface for distances up to
1000 ft (300 m) away from the flowtube.
LARGE SELECTION OF FLOWTUBE SIZES
Flowtubes are offered with a selection of sizes from
1 to 78 in or from DN 25 to DN 2000. Consistent
with the size selected, ANSI Class 150, PN 6, PN 10,
PN 16, PN 40, and AWWA
(1)
C-207 Class D flanges
are provided as end connections.
WEATHERPROOF AND CORROSION
RESISTANT CONSTRUCTION - SUBMERSIBLE
FLOWTUBES
These flowtubes are designed to operate in harsh
outdoor environments. The enclosure is weatherproof
as defined by IEC IP67, and provides the watertight
and corrosion resistant protection of NEMA Type 4X.
The standard IEC IP67 flowtube can also withstand
submersion at a 1.5 m (5 ft) depth for a period up to
72 hours. This depth can be increased to 10 m (33 ft)
and continuous flooding by using a field submersion
kit. Refer to the Optional Selections section.
EXCELLENT ZERO STABILITY
Excellent zero stability is inherent in the design. The
mechanical design and electronic package feature
accurately located and securely mounted coil and
electrode assemblies, spring-loaded electrodes,
optimized field characterization, and power-driven
screens (shields) on the signal leads. All of these
features ensure the ultimate in long-term stability,
signal integrity, and accurate measurement.
STRAIGHT-THROUGH TUBE DESIGN
The straight-through, unobstructed tube design
makes these flowtubes ideal for waste water and
sludge fluids where a minimal pressure drop is
required. This means lower fluid movement costs.
GROUNDING (EARTHING) ELECTRODES
The flowtube design incorporates built-in grounding
electrodes as standard. This means that these
flowtubes can be installed in metallic, nonmetallic, or
lined pipelines without the need for grounding rings.
(1) AWWA = American Water Works Association.
PSS 1-6F9 A
Page 3



LINER PROTECTION
Optional liner protection is offered using grounding
rings on the upstream side of the flowtube to
mechanically protect the liner from high velocity and
very abrasive fluids.
COMPACT FLOWTUBE
The 9100A flowtube has a compact design that
provides face-to-face overall lengths in each size that
conform to recommended flowtube dimensions
contained in ISO/DIS 13359.
LOW POWER CONSUMPTION
All Flowmeter configurations are designed to
consume less than 24 W of power at reference
voltage and frequency.
FLOWTUBE CALIBRATION
All flowtubes are wet calibrated to verify their
specified accuracy. The calibration facilities have
been accredited to the EN 45001 standards.
CE COMPLIANCE
These flowtubes conform to the applicable European
Community Standards when used in conjunction with
the IMT25 Magnetic Flow Transmitter.
OPERATI NG CONDI TI ONS
PERFORMANCE SPECI FI CATI ONS
The Model 9100A Flowtube is used with I/A Series Magnetic Flow Transmitters to form a magnetic flow
system. System performance specifications are listed in the following Magnetic Flow Transmitter PSSs.
Influence
Reference
Operating Conditions
Normal Operating
Condition Limits
Ambient Temperature 20 2C (68 3F) 40 and +100C (40 and +212F)
Process Temperature 20 2C (68 3F) with Ebonite Liner:
0 and 70C (32 and 158F)
Process Pressure 3 bar gauge (43 psig) with Ebonite Liner:
Full Vacuum and Flange Rating
Vibration Negligible 0 and 32 m/s
2
(0 and 3.2 g)
from 18 to 1000 Hz in all directions
PSS 1-6F5 A IMT25 Magnetic Flow Transmitter with FoxCom or HART Communications Protocol
PSS 1-6F5 B IMT25 Magnetic Flow Transmitter with FOUNDATION Fieldbus Communication Protocol
PSS 1-6F5 C IMT25L Magnetic Flow Transmitter with FoxCom Communication Protocol
PSS 1-6F9 A
Page 4



FUNCTI ONAL SPECI FI CATI ONS
Nominal Flowtube Sizes
1 to 78 in (DN 25 to DN 2000). Refer to Model Code
section for a complete listing of flowtube sizes.
End Connections
ANSI, AWWA, and BS 4504 (EN 1092-1) flanges.
Refer to Model Code section for a listing of flange
types and ratings available with each flowtube size.
Process Fluid Conductivity
The minimum process fluid conductivity required is
5 S/cm. For empty pipe detection, the minimum
process fluid conductivity is 20 S/cm. Refer to
TI 027-072 for conductivities of various process
liquids. Also see Table 1.
Signal Cable Length
The maximum allowable cable length is a function of
the cable type, process fluid conductivity, and
whether the cables are in the same or separate
conduits. Standard magnetic flow system accuracy is
maintained when the installations are in accordance
with Table 1.
Ebonite Liner
An Ebonite liner is the best choice when used with
clean, mildly corrosive, or mildly abrasive fluids. It is
not recommended when used with severe corrosive
or severe abrasive fluids. Generally, Ebonite is
selected for general purpose, sewage, drinking
water, and district heating applications.
Flowtube Process Pressure-Temperature Limits
The maximum pressure limit of the flowtube is equal
to the pressure limit of the flanges selected. The
flowtube temperature limit depends on the liner
material. See Figure 1 for Pressure-Temperature
limits of the 9100A Series Flowtubes.
The shaded area beneath each flange curve
represents the flowtube pressure-temperature limits
for each flange type with the selected liner. The
vertical dashed lines represent the temperature limits
of the Ebonite liners. Also, as shown in Figure 1, for
AWWA C207 Class D flanges, the pressure rating is
150 psig within the temperature limits of the liner.

(a) Values in table are fluid conductivity minimums, and maximum distance between transmitter and flowtube. Refer to TI 027-072 for
conductivities of various process liquids.
Table 1. Conductivities of Various Process Liquids (a)
Maximum
Cable Length
Minimum
Fluid Conductivity Signal and Coil Drive Cables
300 m (1000 ft) 5 S/cm Signal and coil drive cables in separate conduit. Signal Cable to be Foxboro
Part No. R0101ZS (feet) or B4017TE (meters).
225 m (700 ft) 5 S/cm Signal and coil drive cables in same conduit. Signal Cable to be Foxboro Part
No. R0101ZS (feet) or B4017TE (meters).
150 m (500 ft) 20 S/cm Signal cable may be in same conduit as coil drive cable. Signal cable to be
good quality twisted shielded pair, preferable no smaller than 1.0 mm
2
(or
18 AWG) for mechanical considerations (Belden 8760 or 9318,
Alpha 5610/1801 or 5611/1801, or equivalent).
PSS 1-6F9 A
Page 5



FUNCTI ONAL SPECI FI CATI ONS ( CONT. )
Figure 1. Flowtube Process Pressure-Temperature Limits with Various Liners, and ANSI/EN 1092-1/AWWA Flanges
Flow Velocity (Figures 2 and 3)
MINIMUM VELOCITY FLOW RANGE
0 to 1.65 ft/s (0 to 0.50 m/s)
MAXIMUM VELOCITY FLOW RANGE
0 to 33 ft/s (0 to 10 m/s)
RECOMMENDED OPERATING VELOCITY
General Liquids
3 to 15 ft/s (0.9 to 4.6 m/s)
Erosive Slurries
3 to 6 ft/s (0.9 to 1.8 m/s)
Liquids that Coat Flowtube Surface
6 to 15 ft/s (1.8 to 4.6 m/s)
Flow Rate (Figures 2 and 3)
Figure 2 and 3 show the relationship between volume
flow rate, flow velocity, and flowtube size in metric
and U.S. Customary units. Also see Table 2 for unit
conversion factors, and Table 3 for Upper Range
Values.
45
40
35
30
25
20
15
10
5
0
360
320
280
240
200
160
120
80
40
0
F C PN 40 LIMIT
F
C
ANSI CL, 150 LIMIT (285 psig)
AWWA LIMIT (150 psig)
EBONITE
LINER LIMITS
P
R
O
C
E
S
S

P
R
E
S
S
U
R
E
,

p
s
i
g
P
R
O
C
E
S
S

P
R
E
S
S
U
R
E
,

b
a
r

(
g
a
u
g
e
)
EBONITE
LINER LIMITS
PN 16 LIMIT
PN 10 LIMIT
PN 6 LIMIT
D E D E
-20
95
80 60 40
-10
20 0 100
32
160 140 120 0 150 120 90 60 30 180 270 240 210
14 203
PROCESS TEMPERATURE, C PROCESS TEMPERATURE, F
45
40
35
30
25
20
15
10
5
0
360
320
280
240
200
160
120
80
40
0
F C PN 40 LIMIT
F
C
ANSI CL, 150 LIMIT (285 psig)
AWWA LIMIT (150 psig)
EBONITE
LINER LIMITS
P
R
O
C
E
S
S

P
R
E
S
S
U
R
E
,

p
s
i
g
P
R
O
C
E
S
S

P
R
E
S
S
U
R
E
,

b
a
r

(
g
a
u
g
e
)
EBONITE
LINER LIMITS
PN 16 LIMIT
PN 10 LIMIT
PN 6 LIMIT
D E D E
-20
95
80 60 40
-10
20 0 100
32
160 140 120 0 150 120 90 60 30 180 270 240 210
14 203
PROCESS TEMPERATURE, C PROCESS TEMPERATURE, F
Table 2. Unit Conversion Factors
To Convert: Multiply by: To Obtain:
m
3
/h 16.67 L/min
m
3
/h 0.2778 L/s
m
3
h 4.403 U.S. gpm
L/min 0.06 m
3
/h
L/min 0.0167 L/s
L/s 3.60 m
3
/h
L/s 60.0 L/min
U.S. gpm 0.8327 IMP gpm
U.S. gpm 8.021 ft
3
/h
U.S. gpm 0.2271 m
3
h
U.S. gpm 60.0 U.S. gph
U.S. gph 0.0167 U.S. gpm
IMP gpm 1.201 U.S. gpm
ft
3
/h 0.1247 U.S. gpm
PSS 1-6F9 A
Page 6



FUNCTI ONAL SPECI FI CATI ONS ( CONT. )
Figure 2. Flowtube Sizing Curves in Metric Units - Also See Table 2
Sizing table (DN15 DN 2000)
The table shows the relationship between fow velocity v, fow
quantity Q and sensor dimension DN.
Guidelines for selection of sensor
Min. measuring range: 0 to 0.25 m/s
Max. measuring range: 0 to 10 m/s
Normally the sensor size is selected so that the nominal fowve-
locity v lies within the measuring range 1 to 3 m/s.
Example:
Flowquantity of 50 m
3
/handa sensor dimensionof DN 80 gives
a fow velocity of 2.7 m/s, which is within the recommended
measuring range of 1 to 3 m/s.
100.000
50.000
m /h
3
10.000
20.000
5.000
2.000
1.000
500
200
100
50
20
10
5
2
1
0.5
0.2
0.1
0.05
0.02
0.01
0.005
0.002
0.001
0.0005
0.0002
I/min.
D
N
1
8
0
0
D
N
2
0
0
0
D
N
1
6
0
0
D
N
1
4
0
0
D
N
1
2
0
0
D
N
1
1
0
0 D
N
1
0
0
0
D
N
8
0
0
D
N
9
0
0
D
N
7
0
0 D
N
6
0
0
D
N
5
0
0
D
N
4
0
0
D
N
3
0
0
D
N
2
0
0
D
N
1
0
0
D
N
4
5
0
D
N
3
5
0
D
N
2
5
0
D
N
1
5
0
D
N
1
2
5
D
N
8
0
D
N
6
5
D
N
5
0
D
N
4
0
D
N
2
5
D
N
1
5
0.05 0.1 m/s 0.15 0.2 0.3 1 m/s 0.5 1.5 2 5
0.00005
0.0001
0.0002
0.0005
0.001
0.005
0.01
0.02
0.05
0.002
0.005
0.01
0.02
0.05
0.1
0.1
0.2
0.5
0.2
0.5
1
1
2
5
10
20
50
10
5
2
20
50
100
100
500
200
200
500
1000
2000
5000
2000
5000
10
2 10
10
I/s
5
10
4
4
4
2 10
6
5 10
5
2 10
5
5 10
4
2 10
4
1000
10 m/s
10
6
Flow velocity calculation formula Units
v = 1273.24 Q / DN
2
or v : [m/s], Q : [l/s], DN : [mm]
v = 353.68 Q / DN
2
v : [m/s], Q : [m
3
/h], DN : [mm]
PSS 1-6F9 A
Page 7



FUNCTI ONAL SPECI FI CATI ONS ( CONT. )
Figure 3. Flowtube Sizing Curves in U.S. Customary Units - Also See Table 2
Sizing table (
1
/
2
78)
The table shows the relationship between fow velocity v, fow
quantity Q and sensor dimension size.
Guidelines for selection of sensor
Min. measuring range: 0 to 0.8 ft/s
Max. measuring range: 0 to 33 ft/s
Normally the sensor size is selected so that the nominal fowve-
locity v lies within the measuring range 3 to 10 ft/s.
Example:
Flow quantity of 500 GPM and a sensor dimension of 6" gives a
fow velocity of 5.6 ft/s, which is within the recommended meas-
uring range of 3 to 10 ft/s.
Flow velocity
MGD
0.000001
0.1 0.15 0.2 0.3 1 f/s 0.5 1.5 15 20 2 5
0.0005
0.001
0.005
0.01
0.02
0.001
0.002
0.05
0.002
0.05
0.1
0.2
0.5
0.1
0.2
0.5
1
1
2
5
10
20
50
10
5
2
20
50
100
100
500
200
500 000
I/min.
0.8 3 10 f/s 3033 f/s
1
/2
"
1

"
2

"
1
"
0.000002
0.000005
0.00001
0.00002
0.00005
0.0001
0.0002
0.0005
0.001
0.002
0.005
0.01
0.02
0.05
0.1
0.2
0.5
1
2
5
10
20
50
100
200
500
1,000
GPM
200 000
100 000
50 000
20 000
10 000
5 000
2 000
1 000
500
200
500 000
200 000
100 000
50 000
20 000
10 000
5 000
2 000
1 000
1 000 000
2 000 000
7
8
"
7
2
"
6
6
"
6
0
" 5
4
"
4
8
" 4
4
"
2
"
3
"
4
"
5
"
6
"
8
"
1
0
"
1
2
"
4
0
"/4
2
"
3
6
"
3
2
"
3
0
"
2
8
" 2
4
"
2
0
" 1
8
"
1
6
" 1
4
"
Flow velocity calculation formula Units
v = 0.408 Q / (Pipe I.D.)
2
or v : [ft/s], Q: [GPM], Pipe I.D. : [inch]
v = 283.67 Q / (Pipe I.D.)
2
v : [ft/s], Q: [MGD], Pipe I.D. : [inch]
PSS 1-6F9 A
Page 8



FUNCTI ONAL SPECI FI CATI ONS ( CONT. )
Table 3. Upper Range Values
Flowtube Size Metric Units U.S. Customary Units
DN
Flange Size
Inch
Flange Size
Minimum
URV(
(a)
(a) URV = Upper range value. See Table 1 for unit conversions.
Maximum
URV
(a)
Minimum
URV
(a)
Maximum
URV
(a)
DN 25 1 in 0.80 m
3
/h 16 m
3
/h 3.5 U.S. gpm 70 U.S. gpm
DN 40 1 1/2 in 2.1 m
3
/h 42 m
3
/h 9.0 U.S. gpm 180 U.S. gpm
DN 50 2 in 2.9 m
3
/h 58 m
3
/h 12.5 U.S. gpm 250 U.S. gpm
DN 65 2 1/2 in 5.0 m
3
/h 100 m
3
/h 22 U.S. gpm 440 U.S. gpm
DN 80 3 in 7.1 m
3
/h 142 m
3
/h 31 U.S. gpm 625 U.S. gpm
DN 100 4 in 12.0 m
3
/h 250 m
3
/h 55 U.S. gpm 1100 U.S. gpm
DN 125 5 in 19.3 m
3
/h 385 m
3
/h 85 U.S. gpm 1700 U.S. gpm
DN 150 6 in 27.7 m
3
/h 554 m
3
/h 122 U.S. gpm 2440 U.S. gpm
DN 200 8 in 58.0 m
3
/h 1160 m
3
/h 255 U.S. gpm 5100 U.S. gpm
DN 250 10 in 93.0 m
3
/h 1860 m
3
/h 410 U.S. gpm 8200 U.S. gpm
DN 300 12 in 133 m
3
/h 2660 m
3
/h 585 U.S. gpm 11 700 U.S. gpm
DN 350 14 in 165 m
3
/h 3300 m
3
/h 730 U.S. gpm 14 600 U.S. gpm
DN 400 16 in 220 m
3
/h 4400 m
3
/h 960 U.S. gpm 19 200 U.S. gpm
DN 450 18 in 280 m
3
/h 5600 m
3
/h 1200 U.S. gpm 24 000 U.S. gpm
DN 500 20 in 340 m
3
/h 6800 m
3
/h 1500 U.S. gpm 30 000 U.S. gpm
DN 600 24 in 490 m
3
/h 9800 m
3
/h 2150 U.S. gpm 43 000 U.S. gpm
DN 700 28 in 690 m
3
/h 13 800 m
3
/h 3000 U.S. gpm 60 000 U.S. gpm
30 in 3400 U.S. gpm 68 000 U.S. gpm
DN 800 32 in 900 m
3
/h 18 000 m
3
/h 3900 U.S. gpm 78 000 U.S. gpm
DN 900 36 in 1150 m
3
/h 23 000 m
3
/h 5000 U.S. gpm 100 000 U.S. gpm
DN 1000 40 in 1400 m
3
/h 28 000 m
3
/h 6200 U.S. gpm 124 000 U.S. gpm
42 in 6800 U.S. gpm 136 000 U.S. gpm
44 in 7500 U.S. gpm 150 000 U.S. gpm
DN 1200 48 in 2050 m
3
/h 41 000 m
3
/h 9000 U.S. gpm 180 000 U.S. gpm
DN 1400 54 in 2800 m
3
/h 56 000 m
3
/h 12 000 U.S. gpm 240 000 U.S. gpm
60 in 14 000 U.S. gpm 280 000 U.S. gpm
DN 1600 66 in 4000 m
3
/h 80 000 m
3
/h 17 500 U.S. gpm 350 000 U.S. gpm
DN 1800 72 in 4800 m
3
/h 96 000 m
3
/h 21 000 U.S. gpm 420 000 U.S. gpm
DN 2000 78 in 5600 m
3
/h 112 000 m
3
/h 25 000 U.S. gpm 500 000 U.S. gpm
PSS 1-6F9 A
Page 9



PHYSI CAL SPECI FI CATI ONS
Flowtube Enclosure (Including Terminal Box)
The overall flowtube enclosure construction meets
IEC IP67 and provides the environmental protection
of NEMA Type 4X. When used with an optional
submersion kit, the enclosure meets IP68 and NEMA
Type 6. See OPTIONAL SELECTIONS AND
ACCESSORIES section.
Housing Material
Carbon steel with corrosion-resistant, two
component coating.
Flowtube Material
AISI Type 304 ss
Flange Material
Carbon Steel with corrosion resistant, two
component coating.
Terminal Box Enclosure Material
Fiberglass reinforced polyamide, or aluminum, as
specified.
Process Wetted Materials
FLOWTUBE LINER
Ebonite.
ELECTRODES
Hastelloy C-276.
FLANGE GASKETS
Provided by user.
Installation Requirements
During measurement, the flowtube must remain full
with the process fluid to achieve the stated
performance. Installation in a vertical pipe with flow
going upward is ideal. For horizontal installations,
position the flowtube so that the electrodes are in a
horizontal plane. The minimum upstream straight
pipe length recommended is five pipe diameters, and
the minimum downstream straight pipe length
recommended is three pipe diameters. The pipe
length is measured outward from the center of the
flowtube. Refer to the applicable installation
instructions for more details.
Electrical Connections
Four 1/2 NPT holes with cable glands for cable entry
are provided on the plastic terminal box, and two
1/2 NPT holes for conduit are provided on the
aluminum terminal box. Separate screw type
terminals within the terminal box are provided for
electrode and coil cable terminations. Refer to
DIMENSIONS-NOMINAL section.
Approximate Mass
Refer to Table 4.
Dimensions-Nominal
Refer to DIMENSIONS-NOMINAL section.
PSS 1-6F9 A
Page 10



PHYSI CAL SPECI FI CATI ONS ( CONT. )
Table 4. Approximate Mass
Nominal Flowtube Size Approximate Mass
(a)
DN Flange Size Inch Flange Size kg lb
DN 25 1 in 5 11
DN 40 1 1/2 in 8 18
DN 50 2 in 9 20
DN 65 2 1/2 in 11 24
DN 80 3 in 12 26
DN 100 4 in 16 35
DN 125 5 in 19 42
DN 150 6 in 27 60
DN 200 8 in 40 88
DN 250 10 in 60 132
DN 300 12 in 80 176
DN 350 14 in 110 243
DN 400 16 in 125 276
DN 450 18 in 175 386
DN 500 20 in 200 441
DN 600 24 in 287 633
DN 700 28 in 330 728
DN 750 30 in 360 794
DN 800 32 in 450 992
DN 900 36 in 530 1168
DN 1000 40 in 660 1455
DN 1100 44 in 1140 2513
DN 1200 48 in 1180 2601
DN 1400 56 in 1600 3527
DN 1500 50 in 2460 5423
DN 1600 64 in 2140 4718
DN 1800 72 in 2930 6460
DN 2000 78 in 3665 8080
(a) Approximate Mass includes the plastic terminal box.
PSS 1-6F9 A
Page 11



ELECTRI CAL SAFETY SPECI FI CATI ONS
NOTE
Flowtubes have been designed to meet electrical safety descriptions listed above. For detailed
information, or status of testing laboratory approvals or certifications, contact Invensys.
OPTI ONAL SELECTI ONS AND ACCESSORI ES
Submersion Kit to IEC IP68
The standard flowtube enclosure meets IEC IP67 and
may be immersed in water. A submersion kit option
is offered to field upgrade the protection from IEC
IP67 to IEC IP68. The table below compares the
standard submersion protection, and improved
protection using the optional submersion kit. Specify
Part Number 085U0220 for the optional IP68
Submersion Kit.
Liner Protection
Grounding rings are offered for the protection of the
flowtube liner to protect the liner against high velocity
and/or very abrasive fluids. A Type C 304 ss ring can
be used with Ebonite lined flowtubes. The 9100A
flowtube does not require grounding rings to provide
accurate flow measurements when equipped with an
Ebonite liner. See Table 5 for ring specifications and
Table 6 for ring part numbers, as required for liner
protection.
Testing Laboratory, Types of Protection,
and Area Classification Application Conditions
Electrical Safety
Design Code
FM/CSA nonincendive Class I, Division 2, Groups A, B,
C, and D hazardous locations.
Temperature Class T4; Ta = 100C. N
Designation
Submersion
Depth
Submersion
Duration
IP67; NEMA 4X
(Standard)
1.5 m Water
(5 ft Water)
72 hours
IP68; NEMA 6
(Optional)
10 m Water
(33 ft Water)
Continuous
PSS 1-6F9 A
Page 12



OPTI ONAL SELECTI ONS AND ACCESSORI ES ( CONT. )
(a) Type C 304 ss ring is used with Ebonite lined flowtubes for liner protection. Rings are not needed for fluid reference ground since the
flowtube includes grounding electrodes.
Table 5. Type C Liner Protection Ring Information (a)
Flowtube Size Type C Liner Protection Ring - 304 ss
DN
Flowmeter
Size
Inch
Flowmeter
Size
Dim. T1 Dim. T2
Approx.
Mass
Liner Protection Ring Configuration mm in mm in kg lb
DN 25 1 in 1.2 0.047 15 0.59 0.03 0.07
DN 40 1 1/2 in 1.2 0.047 15 0.59 0.08 0.18
DN 50 2 in 1.2 0.047 15 0.59 0.12 0.26
DN 65 2 1/2 in 1.2 0.047 15 0.59 0.16 0.35
DN 80 3 in 1.2 0.047 15 0.59 0.20 0.44
DN 100 4 in 1.2 0.047 15 0.59 0.25 0.55
DN 125 5 in 1.2 0.047 15 0.59 0.29 0.64
DN 150 6 in 1.2 0.047 15 0.59 0.33 0.73
DN 200 8 in 1.2 0.047 15 0.59 0.37 0.81
DN 250 10 in 1.2 0.047 15 0.59 0.4 0.88
DN 300 12 in 1.6 0.063 20 0.79 0.6 1.3
DN 350 14 in 1.6 0.063 20 0.79 1.0 2.2
DN 400 16 in 1.6 0.063 20 0.79 1.4 3.1
DN 450 18 in 1.6 0.063 20 0.79 1.8 4.0
DN 500 20 in 1.6 0.063 20 0.79 2.2 4.8
DN 600 24 in 1.6 0.063 20 0.79 2.6 5.7
DN 700 28 in 2.0 0.079 25 0.98 3.0 6.6
30 in 2.0 0.079 25 0.98 3.3 7.3
DN 800 32 in 2.0 0.079 25 0.98 3.7 8.1
DN 900 36 in 2.0 0.079 25 0.98 4.0 8.8
DN 1000 40 in 2.0 0.079 25 0.98 4.4 9.7
42 in 2.0 0.079 25 0.98 4.5 10.0
44 in 2.0 0.079 25 0.98 4.7 10.3
DN 1200 48 in 2.0 0.079 25 0.98 5.0 11.0
DN 1400 54 in 3.0 0.12 40 1.6 9.0 20
60 in 3.0 0.12 40 1.6 10.0 24
DN 1600 66 in 3.0 0.12 40 1.6 12.5 28
DN 1800 72 in 3.0 0.12 40 1.6 14.3 32
DN 2000 78 in 3.0 0.12 40 1.6 16 35
RING
GASKET
LINER
FLANGE
BOLTS
"T1"
"T2"
PSS 1-6F9 A
Page 13



OPTI ONAL SELECTI ONS AND ACCESSORI ES ( CONT. )
(a) Contact Invensys.
Table 6. Ring Part Numbers
Nominal Flowtube Size
Part Number for Liner Protection Ring for Each Flange Type - 304 ss
EN1092
PN 6
EN1092
PN 10
EN1092
PN 16
EN1092
PN 40
ANSI
Cl. 150
AWWA
C-207
DN 25 1 in 083N8361 083N8361
DN 40 1 1/2 in 083N8362 083N8362
DN 50 2 in 083N8344 083N8344
DN 65 2 1/2 in 083N8345 083N8345
DN 80 3 in 083N8347 083N8347
DN 100 4 in 083N8025 083N8025
DN 125 5 in 083N8071 083N8071
DN 150 6 in 083N8008 083N8008
DN 200 8 in 083N8011 083N8011 083N8011
DN 250 10 in 083N8013 083N8013 083N8013
DN 300 12 in 083N8012 083N8012 083N8012
DN 350 14 in 083N8039 083N8039 083N8039
DN 400 16 in 083N8100 083N8100 083N8100
DN 450 18 in 083N8103 083N8104 083N8104
DN 500 20 in 083N8107 083N8108 083N8107
DN 600 24 in 083N8111 083N8112 083N8113
DN 700 28 in 083N8294 083N8294
(a)
30 in
(a)
DN 800 32 in 083N8304 083N8304
(a)
DN 900 36 in 083N8307 083N8307
(a)
DN 1000 40 in 083N8310 083N8310
(a)
42 in
(a)
44 in
(a)
DN 1200 48 in 083N8313 083N8313
(a)
DN 1400 54 in 083N8349 083N8353 083N8357
(a)
60 in
(a)
DN 1600 66 in 083N8350 083N8354 083N8358
(a)
DN 1800 72 in 083N8351 083N8355 083N8359
(a)
DN 2000 78 in 083N8352 083N8356 083N8360
(a)
PSS 1-6F9 A
Page 14



MODEL CODE
9100A Series Magnetic Flowtubes
Nominal Flowtube Size (a) Model
DN Flowmeter Size Inch Flowmeter Size
DN 25 1 in 9101A
DN 40 1 1/2 in 911HA
DN 50 2 in 9102A
DN 65 2 1/2 in 912HA
DN 80 3 in 9103A
DN 100 4 in 9104A
DN 125 5 in 9105A
DN 150 6 in 9106A
DN 200 8 in 9108A
DN 250 10 in 9110A
DN 300 12 in 9112A
DN 350 14 in 9114A
DN 400 16 in 9116A
DN 450 18 in 9118A
DN 500 20 in 9120A
DN 600 24 in 9124A
DN 700 28 in 9128A
30 in 9130A
DN 800 32 in 9132A
DN 900 36 in 9136A
DN 1000 40 in 9140A
42 in (42 inch AWWA Flange on a 40 inch Tube) 9142A
44 in 9144A
DN 1200 48 in 9148A
DN 1400 54 in (54 inch AWWA Flange on a 56 inch Tube) 9154A
60 in 9160A
DN 1600 66 in (66 inch AWWA Flange on a 64 inch Tube) 9166A
DN 1800 72 in 9172A
DN 2000 78 in (78 inch AWWA Flange on a 80 inch Tube) 9178A
Tube Construction
AISI Type 304 Stainless Steel Tube (304 ss) SI
End Connections
ANSI Class 150, Carbon Steel Flange - 1 to 24 in Line Sizes CA
AWWA C-207, Class D, Carbon Steel Flange - 28 to 78 in Line Sizes WC
EN 1092-1, PN 6 (DN1400-2000) Carbon Steel CZ
EN 1092-1, PN 10 (DN200-2000) Carbon Steel CX
EN 1092-1, PN 16 (DN65-1200) Carbon Steel CE
EN 1092-1, PN 40 (DN25-50) Carbon Steel CG
Liner Material
Ebonite -B
Model Code continued on next page
PSS 1-6F9 A
Page 15



(a) See End Connections selections further in Code to determine ANSI, AWWA, and BS (EN 1092-1) flanges applicable to each
flowtube size.
MODEL CODE ( CONT. )
9100A Series Magnetic Flowtubes (Cont.)
Electrodes Model
Hastelloy C-276 - Fluid Reference Electrode Included H
Coil Drive
Pulsed dc J
Housing Construction/Transmitter Mounting
Coated Carbon Steel Housing with Aluminum Terminal Box
Terminal Box has two 1/2 NPT Conduit Threads
Remote Mounted Transmitter
G
Coated Carbon Steel Housing with Polyamide Terminal Box
Terminal Box has four 1/2 NPT Threads with Cable Glands
Remote Mounted Transmitter
F
Coated Carbon Steel Housing with Polyamide Terminal Box, M20 Conduit V
Electrical Safety (Also see Electrical Safety Specifications section)
FM/CSA, Nonincendive, Class I, Division 2 N
Example: 9116A-SICA-BHJ-GN
PSS 1-6F9 A
Page 16



DI MENSI ONS - NOMI NAL
"OD"
"D"
PLASTIC OR ALUMINUM
TERMINAL BOX CAN BE
ROTATED 90 IN EITHER
DIRECTION
PLUG UNUSED
CABLE ENTRY
HOLES
CAPTIVE SCREWS
ALLOW REMOVAL OF
TERMINAL BOX COVER
1/2 NPT GLAND WITH
PLASTIC TERMINAL BOX.
1/2 NPT CONDUIT THREAD WITH
ALUMINUM TERMINAL BOX.
MEASURING
ELECTRODES
(2 PLACES) IN
HORIZONTAL PLANE
NOTES
1. SEE TABLE THAT FOLLOWS FOR FLOWTUBE DIMENSIONS. FLANGE "OD" AND NUMBER OF
FLANGE BOLT HOLES ARE PER THE APPLICABLE FLANGE STANDARDS.
2. IN THE FIGURES ABOVE, THE PLASTIC TERMINAL BOX IS SHOWN. THE ALUMINUM TERMINAL
BOX HAS A SLIGHTLY DIFFERENT CONFIGURATION AND SIZE.
3. THE PLASTIC TERMINAL BOX IS PROVIDED WITH FOUR 1/2 NPT THREADED HOLES
WITH CABLE GLANDS. THE ALUMINUM TERMINAL BOX IS PROVIDED WITH TWO 1/2 NPT
THREADED HOLES FOR CONDUIT.
4. THE BUILT-IN GROUNDING (EARTHING) ELECTRODES ARE NOT SHOWN. THEY ARE LOCATED
BELOW THE MEASURING ELECTRODES SHOWN IN THE FIGURE ABOVE.
"L"
"B"
"A"
mm
in
PSS 1-6F9 A
Page 17



DI MENSI ONS - NOMI NAL ( CONT. )
Flowtube Size Nominal Dimensions
DN
Flowmeter
Size
Inch
Flowmeter
Size Dim. A
(a)
(a) Dimension A is for a flowtube with the plastic terminal box. This dimension is reduced by approximately 30 mm (1.2 in) when the aluminum
terminal box is used.
Dim. B Dim. D
Dimension L for Flange Type
(b)
(b) Underlined length dimensions (Dimension L) comply with ISO/DIS Standard 13359.
PN 6
(c)
(c) Model code option 91xxx-xxCZ-xxx-xx
PN 10
(d)
(d) Model code option 91xxx-xxCX-xxx-xx
PN 16
(e)
(e) Model code option 91xxx-xxCE-xxx-xx
PN 40
(f)
(f) Model code option 91xxx-xxCG-xxx-xx
ANSI
Class
150
(g)
(g) Model code option 91xxx-xxCA-xxx-xx
AWWA
(h)
(h) Model code option 91xxx-xxWC-xxx-xx
mm in mm in mm in mm in mm mm mm mm in in
DN 25 1 187 7.4 59 2.3 104 4.1 200 7.9
DN 40 1 1/2 197 7.8 82 3.2 124 4.9 200 7.9
DN 50 2 205 8.1 72 2.8 139 5.5 200 7.9
DN 65 2 1/2 212 8.3 72 2.8 154 6.1 200 200 200 200 7.9
DN 80 3 222 8.7 72 2.8 174 6.9 200 200 200 272 10.71
DN 100 4 242 9.5 85 3.3 214 8.4 250 250 250 250 9.84
DN 125 5 255 10.0 85 3.3 239 9.4 250 250 250 250 9.84
DN 150 6 276 10.9 85 3.3 282 11.1 300 300 300 300 11.81
DN 200 8 304 12.0 137 5.4 338 13.3 350 350 350 350 13.8
DN 250 10 332 13.1 157 6.2 393 15.5 450 450 450 450 17.7
DN 300 12 357 14.1 157 6.2 444 17.5 500 500 500 500 19.7
DN 350 14 362 14.3 270 10.6 451 17.8 550 550 550 550 21.7
DN 400 16 387 15.2 270 10.6 502 19.8 600 600 600 600 23.6
DN 450 18 418 16.5 310 12.2 563 22.2 600 600 600 23.6
DN 500 20 443 17.4 350 13.8 614 24.2 600 600 600 680 23.6
DN 600 24 494 19.4 430 16.9 715 28.1 600 600 600 800 23.6
DN 700 28 544 21.4 500 19.7 816 32.1 700 700 875 27.6
30 571 22.5 556 21.9 869 34.2 29.5
DN 800 32 606 23.9 560 22.0 927 36.5 800 800 1000 31.5
DN 900 36 653 25.7 630 24.8 1032 40.6 900 900 1125 35.4
DN 1000 40 704 27.7 670 26.4 1136 44.7 1000 1000 1250 39.4
DN 1100 44 755 29.7 770 30.3 1238 48.7 43.3
DN 1200 48 810 31.9 792 31.2 1348 53.1 1200 1200 1500 47.2
DN 1400 54 925 36.4 1000 39.4 1675 65.9 1400 1400 55.1
DN 1500 60 972 38.3 1020 40.2 1672 65.8 1500 1500 59.1
DN 1600 66 1025 40.4 1130 44.5 1915 75.4 1600 1600 63.0
DN 1800 72 1123 44.2 1250 49.2 1974 77.7 1800 1800 70.9
DN 2000 78 1223 48.1 1375 54.1 2174 85.6 2000 2000 78.7
PSS 1-6F9 A
Page 18



PSS 1-6F9 A
Page 19



PSS 1-6F9 A
Page 20


Invensys Operations Management
5601 Granite Parkway Suite 1000
Plano, TX 75024
United States of America
http://iom.invensys.com
Global Customer Support
Inside U.S.: 1-866-746-6477
Outside U.S.: 1-508-549-2424 or contact
your local Invensys representative.
Email: iom.support@invensys.com
Website: http://support.ips.invensys.com
Invensys, Foxboro, FoxCom, and I/A Series are
trademarks of Invensys plc, its subsidiaries, and
affiliates.
All other brand names may be trademarks of their
respective owners.
Copyright 1997-2010 Invensys Systems, Inc.
All rights reserved
MB 010 1110
ORDERI NG I NSTRUCTI ONS
OTHER M&I PRODUCTS
1. Model Number
2. Operating Flow Range (Also see Figures 2 and 3 and Table 2)
3. Process Temperature and Pressure Range. Specify minimum, normal, and
maximum values.
4. Process Fluid Composition and Conductivity (Also see Table 3)
5. Options or Accessories - See Options and Accessories section for
Description and Ordering Instructions.